ORIGINAL RESEARCH article

Front. Cardiovasc. Med., 25 March 2024
Sec. Aortic Surgery and Endovascular Repair
Volume 11 - 2024 | https://doi.org/10.3389/fcvm.2024.1351342

Treatment for Stanford type B aortic dissection with insufficient anchoring region using castor integrated branched aortic stent graft

Weiqing Chen1,† Dabing Liu2,† Tao Chen1 Jian Liu1 Yi Guo1 Bo Ye1*
  • 1Department of Vascular Surgery, Ganzhou People’s Hospital, Ganzhou, Jiangxi, China
  • 2Department of General Surgery, The People’s Hospital of Ganxian District, Ganzhou, Jiangxi, China

Background: To investigate the clinical efficacy of Castor integrated branched aortic stent graft for the treatment of Stanford type B aortic dissection with insufficient anchoring area.

Methods: Retrospective analysis of clinical data of 26 patients with Stanford type B aortic dissection with insufficient anchoring region (<15 mm) treated by Castor branched aortic stent graft from September 2018 to June 2022 at Ganzhou People's Hospital, including 23 acute cases and 3 chronic cases.

Results: Surgical procedures were successfully performed in all 26 patients, and during the perioperative period no complications occurred, such as cerebrovascular accident, stenosis or occlusion of left subclavian artery, progression of reverse avulsion of aortic dissection, and paraplegia. During the operation 2 patients had a small amount of type I endoleak, which disappeared during the postoperative follow-up. The other patients had good postoperative follow-up results. Review of the aortic CTA indicated good stent morphology with patency of the left subclavian artery.

Conclusions: The Castor integrated branched aortic stent graft expanded the indications for endoluminal treatment for Stanford type B aortic dissection, which can avoid open surgery and has good clinical outcomes.

1 Introduction

Aortic dissection, a highly dangerous medical condition, is a common emergency in vascular surgery. With rapid development of endovascular technology and related equipment, thoracic endovascular aortic repair (TEVAR) has become the first choice for the treatment of Stanford B aortic dissection (Type B aortic dissection, TBAD) because of its minimal trauma, quick recovery, good therapeutic effect and fewer complications (1). In normal circumstances, the proximal healthy anchoring zone is required to be longer than 15 mm, so that TEVAR can be performed with conventional aortic covered stent (2, 3). Insufficient proximal healthy anchoring area (<15 mm) may lead to the reverse avulsion of aortic dissection, endoleak and occlusion of aortic branches. Therefore, when the healthy anchoring area is shorter than 15 mm, the proximal anchoring area should be expanded in endovascular surgery, mostly around the left subclavian artery (LSA). Techniques such as carotid-subclavian or axillary-axillary artery artificial vascular bypass, parallel stent technique, fenestration technique, LSA endovascular reconstruction can be adopted, with the main purpose of providing sufficient healthy anchoring area for the covered stent and keeping the LSA unobstructed. In recent years, our department has adopted the Castor integrated branched aortic covered stent to treat Stanford B aortic dissection with insufficient anchoring area (4, 5), which is reported as follows.

2 Materials and methods

2.1 The research object

We retrospectively analyzed the clinical data of Stanford type B aortic dissection patients with insufficient anchoring zone treated by Castor integrated branched aortic covered stent in Ganzhou People's Hospital from September 2018 to September 2022. All patients in this group were examined by CTA before operation, and the diagnosis of Stanford B aortic dissection was confirmed, with healthy anchoring area shorter than 15 mm (the distance from the rupture of aortic intima or the avulsion dissection to the opening of LSA).

2.2 Pre-operative preparation

Acute cases in this group were demanded strict confinement in bed and given active pain relieving after admission; Strict control of blood pressure and heart rate: blood pressure was quickly brought down below 120/80 mmHg (1 mmHg = 0.133 kPa) and heart rate maintained at 60∼70 bpm. Close observation of patient's state of mind, vital signs, skin temperature of limbs, pain changes, urine volume, existence of dyspnea or displegia, smooth defecation, etc. CTA for the aorta, ECG, color Doppler echocardiography and biochemical tests were performed for all patients in this group before operation. The aortic diameter at the opening of LSA, LSA diameter, descending aorta diameter, the length of healthy anchoring zone and the length from the left common carotid artery to LSA were measured and preoperative evaluation was actively completed (Figure 1A). Covered stents with 10%–20% oversize in diameter, and branch stents with 1–2 mm enlargement were often chosen for the procedure.

Figure 1
www.frontiersin.org

Figure 1. Results of aortic imaging in patients at different periods. (A) Is the preoperative CTA image of aorta. (B) Is the result of DSA angiography of aorta during operation, and the intimal rupture is visible near the root of LSA. (C) Is the result of DSA after the release of Castor integrated branched aortic covered stent. The intimal rupture is closed, with no visible endoleak, and LSA is unobstructed. (D) Is CTA imaging review of aorta more than one month after operation, which shows that the stent is well extended, the original false lumen is thrombed and LSA is unobstructed.

2.3 Operation methods

Patients in this group were given general anesthesia and endotracheal intubation. Inguinal region and left elbow incision were taken to expose the femoral artery and the left brachial artery, and heparin sodium was injected intravenously for systemic heparinization (0.5 mg/kg). Vascular sheath was inserted via femoral artery access, and gold-labeled pig tail catheter were deployed into ascending aorta through the descending aorta true lumen, and the location of LSA opening were confirmed by angiography (Figure 1B). Seldinger technique was used to puncture the left brachial artery under direct vision, and a 7F vascular sheath was inserted. Through this vascular sheath, a catheter guidewire was inserted and led out of the body from the femoral artery vascular sheath through the descending aorta, and the guidewire was removed while the catheter remained in the blood vessel. Lunderquist guide wire was placed in ascending aorta through femoral artery gold-labeled pig tail catheter, and the gold-labeled pig tail catheter was withdrawn. After air expulsion, Castor integrated branched aortic covered stent was deployed through Lunderquist guide wire of femoral artery. At the mean time, the branched guide wire was placed in femoral artery catheter and led out of the body from the left brachial artery vascular sheath, and the stent body was placed along Lunderquist guide wire. Meanwhile, the branched guide wire was pulled, and the conical head of the delivery system was led into the upper segment of descending aorta. Adjust the 8-shaped marker at the head of the soft sheath to the lesser curvature side of the aortic arch under fluoroscopy, and transform the marker into an I-shape. Continue to push the distal end of the conical head of the delivery system to the same height of LSA opening, and pay attention to avoid entanglement between the branch guide wire and the delivery system. If entanglement occurs, the delivery system needs to be withdrawn to the straight section of the descending aorta and then the handle of the outer tube needs to be rotated to unwind the entanglement. Retract the soft sheath which wrapped around the stent to the limit position, to expose the main stent and the branch stent, then continue to push the stent body and pull the branch guide wire at the same time, so that the branch stent can be placed in LSA. Confirm that the branch stent is located on the greater aortic curvature side. Fine-tune the position of the main body up and down to make the LSA branch stent completely attach to the LSA blood vessel wall. Pull the branch guide wire and stabilize the delivery system, and release the stent body. After the main stent is released, pull the branch guide wire and catheter to release the branch stent. Perform another angiography to clarify the stent position, whether there is endoleak and the patency of LSA branch stent (Figure 1C). Withdraw the arterial sheath and catheter, suture the puncture of femoral artery and left brachial artery, and suture each surgical incision layer by layer.

2.4 Follow-up and observation indicators

Patients were routinely followed up 1 month, 3 months, 6 months and every year after discharge. At each follow-up, CTA imaging of aorta were performed, blood pressure and heart rate were monitored and discomforts with left upper limb, head, chest or back were recorded. CTA examination of aorta confirmed the stent's position and shape, existence of endoleak, patency of branch stents, false lumen thrombosis, etc.

2.5 Statistical analysis

Spss 25.0 statistical software was used, and the measurement data were represented as X ± S. Paired T test was used to compare the changes of true and false aortic lumen diameters at different levels before and one year after TEVAR, and P < 0.05 indicated that the difference was statistically significant.

3 Results

26 patients were included in this group, including 20 males and 6 females. There were 13 cases of coronary heart disease, 5 cases of diabetes, 3 cases of chronic renal insufficiency and 8 cases of chronic obstructive pulmonary disease. 17 cases had a history of smoking. All the 26 patients were successfully operated. During the operation, 2 patients had a small amount of type I endoleak, and there were no perioperative complications such as cerebrovascular accident, left subclavian artery stenosis or occlusion, aortic dissection and paraplegia. In this group, CTA of aorta was reviewed 1 month, 3 months, 6 months and every year after operation. The total follow-up duration was 12–48 months, with an average of 24 ± 6.4 months. Two patients had a small amount of type I endoleak, which disappeared in subsequent follow-ups. The follow-ups of the other 24 patients all showed satisfactory results. The CTA of aorta showed that the stent was in good shape, the left subclavian artery was unobstructed, and the false lumen covered by the original thoracic aorta stent was absorbed after thrombosis. The changes of true and false lumen diameters of each segment of aorta before and 1 year after operation are shown in Table 1.

Table 1
www.frontiersin.org

Table 1. Changes of true and false lumen diameters in 26 patients before and 1 year after operation.

4 Discussion

At present, TEVAR is the first choice for the treatment of TBAD. When the proximal healthy anchoring zone is shorter than 15 mm, the use of conventional aortic covered stent will cause partial or complete occlusion of LSA, which will lead to LSA steal syndrome, higher risk of vascular injury, stent displacement and endoleak (3, 68). LSA provides extensive blood supply to the brain, left upper limb and spinal cord. If LSA is completely blocked during TEVAR operation, it will increase the risk of postoperative complications such as insufficient blood supply to left upper limb artery and vertebrobasilar artery. In severe cases, stroke, paraplegia, upper limb ischemia and necrosis are also likely to occur (911). Therefore, in 2009, the American Society of Vascular Surgery issued guidelines recommending LSA revascularization be performed whenever possible (7). It is now considered that keeping LSA unobstructed can reduce postoperative complications and improve long-term prognosis of TEVAR (12).

Currently, there are several approaches to reconstruct LSA during TEVAR, such as hybrid procedure, parallel stent technique, fenestration technique, single-branch stent technique, etc. Hybrid techniques, such as carotid-subclavian or axillary-axillary artery artificial vascular bypass, can rebuild LSA blood supply by external anatomical means, which requires fine surgical skills of the operator to avoid surgery-related complications, such as bleeding, neurolymphatic injury, infection, etc. (13). Parallel stent technique: it includes the anterograde chimney technique and the retrograde periscope technique, and the chimney technique is more commonly used at present. The operation of this technique is relatively simple, but branched vascular stents and aortic covered stents are placed in parallel in the aorta, and there is a greater possibility of endoleak due to the gap between stents (14). In addition, the blood flow pattern at the parallel stent is abnormal, which tends to form blood vortex and cause thrombosis of the branch stent (15). Fenestration techniques: mainly includes in-vitro pre-fenestration and in-situ fenestration. In in-vitro pre-fenestration procedures, the operater is required to make accurate measurement according to CTA results before operation, open the covered stent in vitro, and make a hole with the same size as the LSA in the corresponding position of the covered stent in advance. When the stent is released, align the hole with the LSA opening to rebuild the LSA. In-situ fenestration is to cover the LSA opening when the aortic covered stent is placed, puncture the left brachial artery, and with the assistance of vascular sheath, use puncture needle or laser to fenestrate on the aortic covered stent, place the stent, and rebuild LSA. Because LSA is reconstructed through anatomical approach, the original hemodynamic characteristics can be preserved. This technique is widely used in clinical practice due to its minimal trauma and low cost, but it also has disadvantages because of its complicated operation, high requirments for skills, and increased risk of endoleak due to the damage of integrity and stability of the stent (16). Single-branch stent technique: Castor integrated branched aortic covered stent is the first branched covered stent approved for clinical use in China. The integrated design makes the main body and branches firmly anchored in the vascular lumen, which is more in line with the anatomical and physiological characteristics of human blood vessels, and can reduce the risk of stent displacement and endoleak and improve the patency rate of stent blood vessels (17). Castor integrated branched aortic covered stent is reasonable in design and simple in operation. There are four specifications for the covered length from the front end of the main stent to the beginning of the branch: 5 mm, 10 mm, 15 mm and 30 mm, which is suitable for the interval between most LSAs and the left common carotid artery. In addition, the Castor integrated branched aortic covered stent has a vertebral body design, which can better adapt to various aortic shapes and reduce the risk of endoleak. In this study, there were 2 cases with a small amount of type I endoleak, which disappeared in subsequent follow-ups. That might be due to further adaptation of the stent to the blood vessel wall in postoperative remodeling. In this group of patients, the first intima of aorta was well isolated during the operation. During the follow-ups, results show that the true lumen of aorta increased, the false lumen decreased and thrombed after TEVAR, and the changes in true and false lumen diameter were statistically significant. During the release process of Castor integrated branched aortic covered stent, attention should be paid to the following issues: (1) Both the main body of covered stent and the guide wire of branched stent should be in the aortic lumen. (2) If the branched guide wire is entangled with the delivery system during the operation, the delivery system should be withdrawn to the straight segment of descending aorta for unwinding. (3) If stenosis exists after the branch stent is released, balloon expansion can be applied, and stents can be placed, if necessary, to keep the LSA unobstructed.

Castor integrated branched aortic covered stent entered the market in 2017, and has not been in use for quite long. However, based on our experience in treating this group of cases, we found that this stent is easy to operate, stable after release, with less risk of endoleak and higher rate of LSA patency. The short-term follow-ups showed good results, but long-term prognosis needs further follow-ups (5, 18).

To sum up, in the treatment of TBAD with insufficient proximal healthy anchoring zone, Castor integrated branched aortic covered stent effectively prolongs the healthy anchoring zone and expands the surgical indications of TEVAR. This technique reconstructs LSA and maintains the stability of the stent when treating aortic dissection, thus open surgery is avoided. The outcome is satisfactory.

5 Conclusion

In the treatment of TBAD with insufficient proximal healthy anchoring zone, Castor integrated branched aortic covered stent effectively prolongs the healthy anchoring zone and expands the surgical indications of TEVAR.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

The studies involving humans were approved by Ethics Committee of Ganzhou People's Hospital. The studies were conducted in accordance with the local legislation and institutional requirements. Written informed consent for participation was not required from the participants or the participants’ legal guardians/next of kin in accordance with the national legislation and institutional requirements.

Author contributions

WC: Conceptualization, Data curation, Writing – original draft, Writing – review & editing. DL: Resources, Validation, Writing – original draft. TC: Methodology, Software, Writing – original draft, Writing – review & editing. JL: Methodology, Visualization, Writing – original draft. YG: Investigation, Writing – original draft. BY: Project administration, Supervision, Writing – review & editing.

Funding

The author(s) declare that no financial support was received for the research, authorship, and/or publication of this article.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Parsa CJ, Williams JB, Bhattacharya SD, Wolfe WG, Daneshmand MA, McCann RL, et al. Midterm results with thoracic endovascular aortic repair for chronic type b aortic dissection with associated aneurysm. J Thorac Cardiovasc Surg. (2011) 141(2):322–7. doi: 10.1016/j.jtcvs.2010.10.043

PubMed Abstract | Crossref Full Text | Google Scholar

2. Chen S, Larion S, Ahanchi SS, Ammar CP, Brandt CT, Panneton JM. Erratum to: A novel anatomic severity grading score for acute type B aortic dissections and correlation to aortic reinterventions after thoracic endovascular aortic repair. Cardiothorac Surg. (2017) 12(1):53–6. doi: 10.1186/s13019-017-0616-2

Crossref Full Text | Google Scholar

3. Bradshaw RJ, Ahanchi SS, Powell O, Larion S, Brandt C, Soult MC, et al. Left subclavian artery revascularization in zone 2 thoracic endovascular aortic repair is associated with lower stroke risk across all aortic diseases. J Vasc Surg. (2017) 65(5):1270–9. doi: 10.1016/j.jvs.2016.10.111

PubMed Abstract | Crossref Full Text | Google Scholar

4. Lu Q, Feng J, Zhou J, Zhao Z, Li H, Teng Z, et al. Endovascular repair by customized branched stent-graft: a promising treatment for chronic aortic dissection involving the arch branches. J Thorac Cardiovasc Surg. (2015) 150:1631–8. doi: 10.1016/j.jtcvs.2015.08.032

PubMed Abstract | Crossref Full Text | Google Scholar

5. Jing Z, Lu Q, Feng J, Zhou J, Feng R, Zhao Z, et al. Endovascular repair of aortic dissection involving the left subclavian artery by castor stent graft: a multicentre prospective trial. Eur J Vasc Endovasc Surg. (2020) 60(6):854–61. doi: 10.1016/j.ejvs.2020.08.022

PubMed Abstract | Crossref Full Text | Google Scholar

6. Waterford SD, Chou D, Bombien R, Uzun I, Shah A, Khoynezhad A. Left subclavian arterial coverage and atroke during thoracic aortic endografting: a systematic review. Ann Thorac Surg. (2016) 101(1):381–9. doi: 10.1016/j.athoracsur.2015.05.138

PubMed Abstract | Crossref Full Text | Google Scholar

7. Matsumura JS, Lee WA, Mitchell RS, Farber MA, Murad MH, Lumsden AB, et al. The society for vascular surgery practice guidelines: management of the left subclavian artery with thoracic endovascular aortic repair. J Vasc Surg. (2009) 50(5):1155–8. doi: 10.1016/j.jvs.2009.08.090

PubMed Abstract | Crossref Full Text | Google Scholar

8. Mesar T, Alie-Cusson FS, Rathore A, Dexter DJ, Stokes GK, Panneton JM. A more proximal landing zone is preferred for thoracic endovascular repair of acute type B aortic dissections. J Vasc Surg. (2022) 75(1):38–46. doi: 10.1016/j.jvs.2021.06.036

PubMed Abstract | Crossref Full Text | Google Scholar

9. Fang C, Wang C, Liu K, Pang X. Early outcomes of left subclavian artery revascularization using castor single-branched stent-graft in the treatment of type B aortic dissection or intramural hematoma. Ann Thorac Cardiovasc Surg. (2021) 27(4):251–9. doi: 10.5761/atcs.oa.20-00166

PubMed Abstract | Crossref Full Text | Google Scholar

10. Zamor KC, Eskandari MK, Rodriguez HE, Ho KJ, Morasch MD, Hoel AW. Outcomes of thoracic endovascular aortic repair and subclavian revascularization techniques. J Am Coll Surg. (2015) 221(1):93–100. doi: 10.1016/j.jamcollsurg.2015.02.028

PubMed Abstract | Crossref Full Text | Google Scholar

11. Riambau V, Böckler D, Brunkwall J, Cao P, Chiesa R, Coppi G, et al. Editor’s choice-management of descending thoracic aorta diseases: clinical practice guidelines of the European society for vascular surgery (ESVS). Eur J Vasc Endovasc Surg. (2017) 53(1):4–52. doi: 10.1016/j.ejvs.2016.06.005

PubMed Abstract | Crossref Full Text | Google Scholar

12. Awad H, Ramadan ME, El Sayed HF, Tolpin DA, Tili E, Collard CD. Spinal cord injury after thoracic endovascular aortic aneurysm repair. Can J Anaesth. (2017) 64(12):1218–35. doi: 10.1007/s12630-017-0974-1

PubMed Abstract | Crossref Full Text | Google Scholar

13. Protack CD, Smith A, Moennich LA, Hardy D, Lyden SP, Farivar BS. Midterm outcomes of subclavian artery revascularization in the setting of thoracic endovascular aortic repair. J Vasc Surg. (2020) 72(4):1222–8. doi: 10.1016/j.jvs.2019.11.049

PubMed Abstract | Crossref Full Text | Google Scholar

14. Xue Y, Sun L, Zheng J, Huang X, Guo X, Li T, et al. The chimney technique for preserving the left subclavian artery in thoracic endovascular aortic repair. Eur J Cardiothorac Surg. (2015) 47(4):623–9. doi: 10.1093/ejcts/ezu266

PubMed Abstract | Crossref Full Text | Google Scholar

15. Zhu Y, Guo W, Liu X, Jia X, Xiong J, Wang L. The single-centre experience of the supra-arch chimney technique in endovascular repair of type B aortic dissections. Eur J Vasc Endovasc Surg. (2013) 45(6):633–8. doi: 10.1016/j.ejvs.2013.02.016

PubMed Abstract | Crossref Full Text | Google Scholar

16. Percy ED, Sabe AA. Commentar: A new chapter in chronic type B aortic dissection: balloon fracture fenestration and remodeling. J Thorac Cardiovasc Surg. (2020) 17(20):22–3. doi: 10.1016/j.jtcvs.2020.10.036

Crossref Full Text | Google Scholar

17. Shen Y, Zhang S, Zhu G, Chen Y, Chen Z, Jing Z, et al. Risk factors ofdistal segment aortic enlargement after complicated type B aortic dissection. J Interv Med. (2019) 2(4):154–9. doi: 10.1016/j.jimed.2019.10.003

PubMed Abstract | Crossref Full Text | Google Scholar

18. Chang H, Wang Y, Liu B, Wang W, Li Y. Endovascular repair for acute type B aortic dissection with unfavorable proximal landing zone. Ann Thorac Surg. (2022) 113(2):545–53. doi: 10.1016/j.athoracsur.2021.02.092

PubMed Abstract | Crossref Full Text | Google Scholar

Keywords: integrated branched stent, aortic dissection, thoracic endovascular aortic repair (TEVAR), anchoring region, castor

Citation: Chen W, Liu D, Chen T, Liu J, Guo Y and Ye B (2024) Treatment for Stanford type B aortic dissection with insufficient anchoring region using castor integrated branched aortic stent graft. Front. Cardiovasc. Med. 11:1351342. doi: 10.3389/fcvm.2024.1351342

Received: 6 December 2023; Accepted: 20 February 2024;
Published: 25 March 2024.

Edited by:

Jiaxuan Feng, Second Military Medical University, China

Reviewed by:

Huibo Ma, The Affiliated Hospital of Qingdao University, China
Xianhao Bao, Shanghai General Hospital, China

© 2024 Chen, Liu, Chen, Liu, Guo and Ye. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Bo Ye yebobo2008@163.com

These authors have contributed equally to this work and share first authorship

Download