Abstract

Diabetes mellitus is characterized by elevated blood glucose levels, however patients with diabetes may also develop hypoglycemia due to treatment. There is an increasing demand for non-invasive blood glucose monitoring and trends detection amongst people with diabetes and healthy individuals, especially athletes. Wearable devices and non-invasive sensors for blood glucose monitoring have witnessed considerable advances. This review is an update on recent contributions utilizing novel sensing technologies over the past five years which include electrocardiogram, electromagnetic, bioimpedance, photoplethysmography, and acceleration measures as well as bodily fluid glucose sensors to monitor glucose and trend detection. We also review methods that use machine learning algorithms to predict blood glucose trends, especially for high risk events such as hypoglycemia. Convolutional and recurrent neural networks, support vector machines, and decision trees are examples of such machine learning algorithms. Finally, we address the key limitations and challenges of these studies and provide recommendations for future work.
1 Introduction
Blood glucose levels are closely regulated within a desirable range by several hormones released primarily by the pancreas. Diabetes mellitus is a group of metabolic diseases characterized by elevated glucose levels (). The 2021 International Diabetes Federation (IDF) atlas has estimated that there are 537 million adults with diabetes and that number is expected to increase to 784 million by 2045. Type 1 diabetes affects between 5 and 10% of patients and is characterized by a lack of insulin production and a higher variability in blood sugars, which requires exogenous insulin and more regular monitoring of blood glucose (). Type 2 diabetes affects up to 90% of all patients diagnosed with diabetes and can be further subdivided into a distinct number of subtypes with varying degrees of insulin deficiency, insulin resistance, and propensity for developing complications (; ).
Normal blood glucose concentrations should lie within a range of 4.0–5.5 mmol/L (72–99 mg/dl) after an 8 h fast and should be 7.8 mmol/L (140 mg/dl) 2 h after eating (Güemes et al., 2015). Chronic hyperglycemia is associated with long-term microvascular and macrovascular complications (). Hypoglycemia occurs due to excess insulin which can be exogenous or endogenous leading to glucose dropping below the normal range. Hypoglycemia occurs in both type 1 diabetes and type 2 diabetes (; Zammitt and Frier, 2005; Miller et al., 2001). There are clear American Diabetes Association (ADA) clinical practice recommendations to achieve optimal blood glucose targets using different therapies (; ), but are inherently associated with an increased risk of hypoglycemia (Ohkuma et al., 2021), especially in patients who have reduced awareness of hypoglycemia (Yeoh et al., 2015).
The ADA categorizes hypoglycemia into: Level 1 with a blood-glucose value between 54 mg/dl and 70 mg/dl; level 2 with a blood-glucose value less than 54 mg/dl; and level 3 with severe hypoglycemia characterized by altered mental and/or physical status requiring external assistance (). The symptoms of hypoglycemia are categorized as autonomic, with adrenergic (tremor, palpitations, tachycardia, and anxiety) and cholinergic (sweating, hunger, and paresthesia) manifestations. Neuroglycopenic symptoms may include dizziness, weakness, drowsiness, delirium, confusion, seizure, and coma. The blood-glucose threshold at which an individual patient will experience hypoglycemic symptoms depends on overall glycemic control as the threshold for hypoglycemia will be higher in patients with consistently elevated blood-glucose. Thus, self-monitoring of blood glucose is a key part of diabetes management ().
The most widely used glucose monitoring systems rely on disposable test strips that read the blood glucose from a sample obtained using the finger prick approach. This method is associated with discomfort, especially when it is undertaken frequently in patients with type 1 diabetes (Heinemann, 2008). Continuous glucose monitoring (CGM) devices measure interstitial glucose and have been shown to marginally improve overall glycemic control and reduce the incidence of hypoglycemia (Nathanson et al., 2021; Lee I. et al., 2021), but suffer from several inherent limitations. There is a time lag between the interstitial glucose measured by CGM and actual blood glucose and there are inconsistencies in the readings of different CGM devices (Kamath et al., 2009; ). Furthermore, most CGMs are relatively expensive, which limits the wider adoption of such technology among diabetic patients.
Wearables are lightweight devices capable of measuring different vital signs and modalities such as heart rate, temperature, respiration rate, activity level, and skin conductance (). Wearable technologies have been proposed to improve the quality of life by collecting and sharing data between people and their carers (Seneviratne et al., 2017). Wearable devices have different functions and may include patches, belts, lenses, earphones, socks, glasses, watches, wrist bands, and bracelets (). Wearables have been recommended in the elderly to increase their independence and have been proposed for autism screening and therapy (Kekade et al., 2018; ; ). Recently, the readings from wearable technologies have been associated with blood glucose levels (; ; Maritsch et al., 2020). Using such devices to indirectly monitor glucose represents a non-invasive, convenient, and inexpensive method over CGM. A typical glucose monitoring system consists of wearable devices and machine learning models assessed against FDA established evaluation metrics (Figure 1).
FIGURE 1
Siddiqui et al. surveyed different types of non-invasive blood glucose monitoring equipment and highlighted the potential role of artificial intelligence in solving some of the existing challenges (Siddiqui et al., 2018). Non-invasive monitoring of glucose levels from breath samples and epidermal electrochemical glucose sensors hold promise (Lekha and Suchetha, 2020; Kim et al., 2018). Other non-invasive sensors that measure physiological parameters such as pulse and pulse pressure may help to improve glucose management among diabetic patients, especially during physical activity (). A wide range of non-invasive techniques for blood glucose monitoring have been developed utilizing glucose monitoring in bodily fluids (Nawaz et al., 2016; ).
Despite the existence of many studies related to non-invasive glucose monitoring, several questions remain unanswered:
1. What technologies can be used in blood glucose prediction and trends detection?
2. Which machine learning algorithms can be used?
3. To what extent are these technologies and techniques successful?
4. What are the limitations of the existing approaches?
To answer these questions, we have reviewed the advances over the past five years in non-invasive sensors, wearable technologies, and machine learning approaches to estimate blood glucose, especially to identify hypoglycemia. We have identified the major limitations and challenges faced by these studies and provided recommendations to overcome these constraints. We have reviewed studies that have developed non-invasive sensors and wearable devices as well as contributions that have employed existing commercially available devices. Studies without working prototypes were excluded. The considered search keywords included non-invasive or noninvasive, sensor, wearable, glucose, levels, trends, hypoglycemia, monitoring, machine learning, and prediction.
2 Non-invasive Sensors and Wearables
This section presents the advances in wearables and sensors over the past five years. The technologies used have varied from direct measurement of glucose from bodily fluids such as saliva, sweat, and tears to the detection of physiological variables which change with blood glucose (Figure 2). Examples of such wearable devices and sensors are illustrated in Figure 3. A summary of the surveyed studies is provided in Table 1.
FIGURE 2
FIGURE 3
TABLE 1
| Study | Device | Experiments | Key Findings | Limitations |
|---|---|---|---|---|
| Susana et al. (2022) | PPG device | Collected data from 80 participants | The classification of blood glucose levels into normal and diabetic | Limited experimental settings and no reported instances of hypoglycemia |
| PPG device | PPG and physiological data from 2,538 participants | Promising blood glucose prediction performance for the group without medication | Limited PPG data collected from each participant with limited results for the group with medication | |
| Maritsch et al. (2020) | PPG wrist wearable | The data was obtained from one participant with T1DM | The detection of hypoglycemia using HRV time features | Limited to one participant |
| Porumb et al. (2020b) | ECG chest wearable device | The study acquired the ECG data from healthy participants monitored for up to 14 days | The ability to detect nocturnal hypoglycemia relying on raw ECG signal | No participants with diabetes and limited instances of hypoglycemia |
| ECG patch wearable device | The data was acquired from patients with T1DM | The identification of HRV patterns in relation to early detection of hypoglycemia | Limited instances of hypoglycemia | |
| 12-lead ECG | ECG, CGM, and diary from 17 patients with type 1 diabetes | Identifying a negative relationship between QTC and hypoglycemia | No female participants and no investigation on the effects of medications | |
| Li et al. (2021) | ECG device | ECG and glucose data from 21 participants | An overall accuracy of 81.69% in classifying blood glucose to three groups | Limited to young participants and small sample size |
| Saha et al. (2017) | EM patch antennas | Tested with water-based samples and with blood samples from humans | The detection of glucose spikes in humans | Limited to experimental settings and susceptible to noise |
| Hanna et al. (2020) | EM wearable glove | Glucose solutions and healthy participants | High correlation with glucose changes and blood glucose trends | No diabetic patients and limited testing conditions |
| Raj et al. (2021) | EM antenna | Tested with plasma glucose | Deviations in the reflection coefficient | Limited experiments with participants |
| Zidane et al. (2021) | EM ring | Tested with glucose solutions | The ability to detect glucose with a 1.665 mmol/L resolution | No experiments with participants were reported |
| Zapasnoy et al. (2021) | Microwave sensor | Tested with glucose solutions | Detecting glucose concentrations in sodium chloride solutions with a 1 mmol/L resolution | No testing with participants were reported |
| Liu et al. (2016) | Bioimpedance antenna | In vitro test in a pork slab | Glucose levels changes can be picked up by bioimpedance parameters | Limited testing conditions with a prototype |
| Nanayakkara et al. (2018) | Bioimpedance and near-infrared | Datasets from one subject | High correlation was achieved | Testing was limited to one participant and no blood trends |
| Tronstad et al. (2019) | Bioimpedance | Data were collected from 20 patients with T1DM | Bioimpedance aids in the prediction | Limited experimental settings and no blood trends |
| Snekhalatha et al. (2018) | Bioimpedance to measure galvanic skin response | Data were collected from 50 participants with diabetes and 50 healthy participants | Negative correlation was observed with blood glucose levels | Limited to experimental settings and no blood trends |
| Sanai et al. (2021) | Bioimpedance wearable ring | Tested with 14 patients with type 2 diabetes | Reported accurate prediction of blood glucose levels | No testing with patients with type 1 diabetes |
| Xiao et al. (2019) | Sweat glucose wearable sensor | The device was tested with 4 participants | Dynamic detection range | Testing only with healthy participants and no blood trends |
| Xuan et al. (2018) | Sweat glucose biosensor | Tested with sweat samples | High linearity and short response time | Limited experimental testing with a prototype |
| Zhu et al. (2018b) | Wristband sweat glucose wearable | Tested using sweat samples and with volunteers | Continuous sweat glucose monitoring with a mobile app | Limited testing conditions, no patients with diabetes, and no blood trends |
| Zhao et al. (2019) | Sweat glucose smartwatch | Tested using sweat samples and with 6 healthy participants | Real-time glucose monitoring using a smartwatch | No participants with diabetes and no blood trends |
| Katseli et al. (2021) | Sweat wearable ring | Tested with one healthy participant | Detecting sweat glucose in the range of 12.5–400 μmol/L | Limited testing conditions |
| Sempionatto et al. (2021) | Touch-based sweat sensor | Tested with three healthy participants | High correlation with blood glucose | Not tested with patients with diabetes |
| Kim et al. (2017) | Contact lens tears glucose sensor | Tested in rabbit and bovine eyes | Wireless detection of tears glucose | No reported tests with participants |
| Park et al. (2018) | Soft and smart tears glucose lens | Tested on a live rabbit | Real-time detection of tears glucose wirelessly | No reported tests with participants |
| Elsherif et al. (2018) | Optical tears glucose sensor | Tested with an artificial eye and different glucose concentrations | Detection of glucose concentration using smartphone camera | No in vivo tests or with participants |
| Kownacka et al. (2018) | Flexible tears glucose biosensor | Tested with 6 participants | Achieved decent performance on Clarke’s error grid | No blood trends and no diabetic participants |
| Lee et al. (2021b) | Tears glucose contact lens | Tested using glucose solutions and human tears from three participants | Detecting tear glucose with a detection limit of 211 nM | No tests with diabetic patients |
| Geelhoed-Duijvestijn et al. (2021) | Tears glucose biosensor | Evaluated with 24 patients with type 1 diabetes | Comparable performance to that of CGM | Requires further evaluations with blood glucose trends |
| Lin et al. (2017) | Saliva glucose sensor | Tested with saliva samples from 9 participants | Saliva glucose detection range that corresponded to blood glucose changes | Limited testing conditions and no tests with diabetic patients |
| Saliva glucose smart toothbrush | Saliva samples collected from 5 subjects | Linear detection range and reflected with blood glucose changes | Limited to healthy participants | |
| Saliva glucose mouthguard | Tested with artificial saliva | Detected glucose concentration | Limited testing conditions and no in vivo tests | |
| Zhang et al. (2021) | Salivary glucose biosensor | Evaluated with saliva samples | Good linearity for glucose concentration between 0 and 50 mg/L | Further evaluations are required with patients with diabetes |
| Fingertip wearable with an accelerometer | Simulated hand tremors | Detection of tremors that are similar to that exhibited during hypoglycemia | Limited tests and no patients with diabetes | |
| Wrist wearable with an accelerometer | Data collected from seven subjects with T1DM and T2DM | Identifying tremors under fatigue | No hypoglycemic events | |
| Jaggers et al. (2019) | Smartwatch with an accelerometer | Data was acquired from 10 patients with T1DM | Activity recognition helps in the prediction of nocturnal hypoglycemia | Limited to adolescent subjects with T1DM and nocturnal hypoglycemia |
A summary of the advances in non-invasive sensors and wearable technologies for blood glucose monitoring.
2.1 Photoplethysmography
Heart rate variability (HRV) reflects physiological changes in the human body and provides insights on cardiac health and the autonomic nervous system, which can be used as an indicator of current or impending diseases (Yuan et al., 2020). For example, it was considered as a predictor in rapid renal function deterioration (
Estimating HRV parameters from wearables to predict hypoglycemia has been investigated in previous studies with changes in HRV occurring up to 90 min prior to hypoglycemia (
A recent study demonstrated the application of a PPG device in the classification of blood glucose levels into normal and diabetic (Susana et al., 2022). With PPG signal and blood glucose data collected from 80 participants, the study reported an accuracy of 98% with a decision tree based classifier. Another recent study investigated blood glucose prediction based on PPG and physiological data acquired from 2,538 participants split into two groups with or without medication (
2.2 Electrocardiogram
Electrical signals arise in the SA node in the right atrium, travel to the atrioventricular node in the interatrial septum and diverge through the left and right bundle of His to the Purkinje fibers terminating in the endocardium and ventricular epicardium causing ventricular contraction. Cardiac electrical activity can be monitored using an Electrocardiogram (ECG). A typical heartbeat is partitioned into three segments, namely PR segment, QRS complex, and ST segment and each segment can be used to identify underlying cardiac conduction defects such as short QT-interval or PR-interval (Nikolaidou et al., 2017; Hasegawa et al., 2016; Kališnik et al., 2019).
The advancement in sensor miniaturization has enabled current wearables to measure changes in cardiac conduction and contraction with good accuracy. The Fitbit Sense smartwatch and Apple watch series 4 are examples of wrist wearables that can measure the heart rate and ECG signal to identify an irregular heart rhythm, e.g., atrial fibrillation. Chest sensors are another trend in wearables that can measure electrical signals from the heart. The Bioharness (3.0, Zephyr Technology, US) is a lightweight, portable chest wearable device that can provide live access to a variety of physiological parameters, including ECG and it has been tested to predict hypoglycemia. The HealthPatch (VitalConnect, San Jose, CA) is another example of an ECG wearable used to derive HRV to investigate hypoglycemia (
Changes in blood glucose can alter cardiac repolarization and induce prolongation of the QT interval with an increased risk of cardiac arrhythmia (Kališnik et al., 2019). These changes in cardiac conduction prior to the development of arrythmia could be used to predict the occurrence of hypoglycemia (Ling et al., 2016; Porumb et al., 2020b). In a recent work, the relationship between ECG data and blood glucose in adults has been investigated (
A recent study utilized an ECG device (i.e., Bioharness) and a glucose monitoring system (i.e., FreeStyle Libre Flash) to develop personalized models to predict the occurrence of nocturnal hypoglycemic events (Porumb et al., 2020b). The study acquired the ECG data from healthy participants monitored for up to 14 days in relation to blood glucose levels below 4 mmol/L to define a hypoglycemia threshold. Using machine learning techniques, the study showed the feasibility of predicting nocturnal hypoglycemia from raw ECG signals. However, the participants in the study were all healthy, hence, low blood glucose instances were limited. Furthermore, the collected data were limited to nocturnal events, thus, ECG circadian changes during the day in relation to glucose concentration were not investigated. Another study acquired ECG and glucose data from 16 healthy adults and 5 with prediabetes to establish a machine learning model based on CNN to classify blood glucose values into three groups, namely, low (i.e., below 6.0 mmol/L), moderate (i.e., above 6.0 mmol/L and below 7.7 mmol/L), and high (i.e., above 7.7 mmol/L) (Li et al., 2021). The results of the best trained model showed an accuracy of 81.69% in classifying the blood glucose values.
2.3 Electromagnetic
Non-ionizing parts of the electromagnetic (EM) spectrum (e.g., visible light, radio, and ultraviolet) ranging from 0 to 3.0 PHz have been used for diagnostic and therapeutic medical applications Mattsson and Simkó (2019). Non-invasive estimation of blood glucose has been investigated using electromagnetic waves and near infrared (NIR) waves based on the unique absorption spectrum of glucose (Zhang et al., 2019; Jiang et al., 2017; Hotmartua et al., 2015; Yadav et al., 2014; Tronstad et al., 2019). Transmittance, which measures the scattered light after penetrating the tissue, and reflectance, which measures the reflected light from the skin surface, are two methods that rely on light to acquire information about a substance (Hotmartua et al., 2015). By investigating the properties of both the reflected and transmitted waves, the level of glucose can be estimated (Yadav et al., 2014).
Several devices have been developed that show that changes in EM correlate with glucose concentrations (
Recently, an innovative EM-based glove wearable device has been developed to monitor blood glucose levels (Hanna et al., 2020). The system consists of two flexible sensors (i.e., slot antenna and a reject filter) and operates in the frequency range of 500 MHz and 3 GHz, which reaches subcutaneous veins and arteries (
2.4 Bioimpedance
Bioimpedance measures the response of a biological medium (e.g., human body) to an electric current. The composition of the biological mediums affects the bioimpedance based on whether they act as insulators, dielectrics, or conductors (Naranjo-Hernández et al., 2019). Hence, bioimpedance measurements can be used to acquire information about body composition, such as fat, muscle, and water and thereby assess obesity (Zhang et al., 2018); sarcopenia in patients with renal disease (Lai et al., 2019), and gastrointestinal disease (Ruiz-Vargas et al., 2016). Bioimpedance analysis has been suggested as a non-invasive method to screen for diabetes mellitus (Jun et al., 2018).
Some studies have investigated the correlation between changes in glucose levels and bioimpedance (
In relation to blood trends, a recent study assessed non-invasive sensors that included bioimpedance measurements in detecting hypoglycemia among 20 patients with type 1 diabetes that underwent clamp procedures (Tronstad et al., 2019). The study revealed that bioimpedance plays a correcting role in the prediction when paired with other sensors. Bioimpedance of the skin has also been used to detect nocturnal hypoglycemic events (Lesko et al., 2018). Another study investigated the galvanic skin response (GSR) and correlated it with blood glucose levels (Snekhalatha et al., 2018), where a device was developed to acquire the GSR resistance and voltage data from 100 participants (i.e., 50 with diabetes). A negative correlation with GSR voltage and resistance was observed among diabetic patients. A recent study in 14 patients with type 2 diabetes collected bioimpedance data using a wearable ring over 2 weeks and used a gradient boosted model to estimate blood glucose levels and trends and showed excellent prediction performance with 99% of the values in zones A and B of the Parkes error grid (Sanai et al., 2021).
2.5 Sweat
Sweating is a normal physiological mechanism to regulate body temperature through evaporation, but it is also a key autonomic feature of hypoglycemia (Escolar et al., 2016). Sweat consists mostly of water but also contains sodium, chloride, potassium, lactate, and urea (
Wearable sensors can take advantage of the non-invasive nature of using sweat as a predictor of the human health status (Upasham and Prasad, 2020). There is a growing interest in developing sweat-based sensors and systems that are aimed to monitor health to help in the management of patients with diabetes (Hojaiji et al., 2020; Lee et al., 2017; Karpova et al., 2019). One study developed a microfluidic device using a cotton thread and filter paper paired with a smartphone to sense sweat glucose (Xiao et al., 2019). The device showed a linear trend in the 50–250 μM range with a detection limit of 35 μM and the nature of the sensor construction enabled it to be flexible, easy to integrate, and to be produced at relatively low-cost. Another study developed a biosensor based on a graphene oxide nanostructured composite deposited with gold and platinum nano-particles to detect glucose in human sweat (Xuan et al., 2018). When tested with sweat samples, the device showed a short response time and high linearity.
Katseli et al. (2021) demonstrated that a 3D printed electrochemical sweat sensor shaped like a ring was capable of monitoring sweat glucose in the range of 12.5–400 μmol/L, which could be read by a smartphone. However, prototype testing was limited to one healthy volunteer. Similarly, Sempionatto et al. (2021) developed a touch-based sweat glucose sensor to estimate blood glucose. The electrochemical sensor consists of a sweat collecting layer, glucose biosensor, and a substrate that requires no sweat stimulation. The sensor achieved a high correlation (i.e., 0.95) with blood glucose and with all the points in the A and B regions on the Clarke error grid. Another study developed a tandem catalytic system for sweat glucose detection based on chemiluminescence with a high sensitivity and detection limit of 0.1 μM when compared to solutions containing different glucose concentrations (Gao et al., 2021).
Smart wrist wearables are becoming an essential part of fitness and health monitoring. The convenience of wearing watches on the wrist only widened the adoption of such wearable devices (Guler et al., 2016). Several wrist wearable sensors have been developed to detect sweat glucose (Hong et al., 2018; Xuan et al., 2018; Lu et al., 2019). One study demonstrated a nonenzymatic wearable sensor that allowed the analysis of sweat glucose (Zhu X. et al., 2018). The sensor was made from a treated silver electrode coated with fluorocarbon-based materials. An integrated wristband containing the sensor provided continuous monitoring of sweat glucose and showed the results on a smartphone App. Their solution demonstrated the possibility of detecting glucose in the range of 30–1,100 μM. However, the developed wearable was tested with samples acquired from participants only and no correlations with blood glucose were made. Another research group developed a fully integrated device to provide continuous monitoring of sweat glucose (Zhao et al., 2019). The device consists of flexible rechargeable batteries and photovoltaic cells that are used to power up the device (i.e., signal processing and display) using solar energy. Monitoring of sweat glucose is based on an electrochemical sensor connected to a controlling module. A small display is used to provide real time monitoring. The wearable displayed potential in detecting sweat glucose changes in the range of 50–200 μM during different activities (e.g., running vs. biking).
2.6 Tears
The human eyes produce tears as a response to irritants, due to intense emotions, and to keep the ocular surface lubricated and protected. Tears are made of water, protein, lipids, and electrolytes (
A ratiometric fluorescent membrane capable of sensing tear glucose in the range of 0.1–10 mM has been developed (
The development of wearable contact lenses capable of monitoring different physiological signs has gained considerable interest in recent years (Elsherif et al., 2018; Park et al., 2018; Kim et al., 2017). A contact lens comprised of three layers (silk fibroin, silver nanowires, and protected film) capable of sensing tear glucose in the range of 500 nM to 1 mM with a detection limit of 211 nM was developed by Lee W.-C. et al. (2021). Another contact lens was developed by Kim et al. (2017) which monitors tear glucose based on glucose oxidase linked to a graphene channel that can be read wirelessly by a coil. The in vivo experiments showed that the wearable sensor can detect tear glucose concentration when placed in a rabbit’s eye. Another study developed a fully integrated soft contact lens that contains glucose sensors, wireless circuits, and a display (Park et al., 2018). The developed smart contact lens is supposed to overcome some of the limitations of existing contact lens such as being brittle, blocking the vision, and requiring extra equipment to read the lens. The wearable was able to detect tear glucose when tested in vivo on a live rabbit.
An optical sensor embedded in a wearable contact lens was developed to provide continuous glucose monitoring (Elsherif et al., 2018). The reading of the sensor was based on smartphone camera readouts that correlated the reflected power of the diffraction with glucose concentration. The developed sensor could detect glucose concentration less than 50 mM with a 12 nm mM−1 sensitivity. However, no in vivo experiments were reported. They subsequently developed a bifocal contact lens containing a hydrogel glucose sensor that could detect tear glucose within the 0–3.3 mM range in artificial tears (Elsherif et al., 2021). Another study developed and clinically tested a flexible tear glucose biosensor (Kownacka et al., 2018). The coil-shaped sensor is 1.3 mm in diameter and 15 mm long and consists of a flexible coil made of electrodes arranged in parallel and an antenna wire that can transfer the readings to an external device through telemetry. The sensor was designed to be placed under the lower eyelid and was reported to not cause any irritation or abnormalities in a sheep’s eye. Clinical testing was conducted with six subjects who wore the sensors and CGMs (i.e., Abbott FreeStyle Libre) for 5 hours in a fasted state. On Clarke’s error grid, 95% of the data points for the sensor were in the A and B regions with 70% in region A, which was comparable to the performance of the CGM. Geelhoed-Duijvestijn et al. (2021) evaluated the same biosensor in 24 patients with type 1 diabetes by simultaneously measuring blood and interstitial glucose levels every 15 min compared to continuous tear glucose and a neural network based regression model was used to convert tear glucose to blood glucose. The performance of the device was comparable to the CGM with a mean absolute relative difference in glucose of 16.7 mg/dl.
2.7 Saliva
Saliva is a clear and slightly acidic secretion originating from the sublingual, submaxillary, parotid, and minor mucous glands and serves to lubricate and clean the oral tissues and assist in taste and digestion (Humphrey and Williamson, 2001;
Interest in the use of saliva as a diagnostic fluid has grown considerably and several sensors have been developed (Nunes et al., 2015;
There have been attempts to incorporate saliva glucose sensors into devices used daily to measure glucose levels. One study embedded a saliva glucose sensor into a smart toothbrush which integrated a bronze based sensor to provide non-enzymatic electrochemical measurement of salivary glucose (
2.8 Acceleration
Involuntary shaking part of the human body, such as the hand, is one manifestation of the symptoms that are associated with hypoglycemia (Wild et al., 2007). The tremor that occurs during hypoglycemia is categorized under enhanced physiologic tremor and results from different mechanical and neuromuscular interactions (Rana and Chou, 2015). The enhanced physiologic tremor is usually more visible compared to normal tremors with a frequency that was estimated to be in the range of 5–14 Hz (Puschmann and Wszolek, 2011; Rana and Chou, 2015). Smart wearables can detect tremors and identify different patterns with the help of machine learning techniques, but have not been widely assessed to predict hypoglycemia (San-Segundo et al., 2020; Lonini et al., 2018; Zahed et al., 2018). A low-cost wearable device based on an accelerometer mounted on the index finger showed that it can detect tremor in the range of 10–14 Hz, but was not validated in patients with diabetes in relation to hypoglycemia (
Acceleration can be used to provide information about human activities and wearables with an accelerometer have found increasing use in health care applications such as detecting falls among the elderly (Janidarmian et al., 2017;
3 Machine Learning Techniques
A typical machine learning algorithm uses data to build a predictive model that can map a set of inputs to a desired output. The core element in machine learning is collecting enough data to be used in the training and evaluation of the predictive model. Once enough data is collected, a portion of this data is typically selected to be used to train and optimize a model’s parameters, and the remaining part is then used to evaluate the performance of the learned model. In glucose monitoring applications, wearable devices can be used to acquire physiological data as inputs while a CGM device is used to acquire the output or target values. Machine learning techniques have been considered for several glucose monitoring applications such as predicting current glucose levels, forecasting future values, and classifying ongoing trends.
This section presents the different machine learning techniques that have been used for blood glucose monitoring and trends detection (Figure 4). A summary of the surveyed studies is provided in Table 2.
FIGURE 4

Machine learning algorithms utilized in glucose monitoring applications over the past five years.
TABLE 2
| Study | Inputs | Data Used | Algorithm | Key Results |
|---|---|---|---|---|
| G et al. (2018) | ECG signals used in the analysis of HRV | Data acquired from 20 healthy participants and 20 with diabetes in supine position | CNN and a hybrid network of CNN-LSTM | CNN-LSTM achieved 95.1% in distinguishing diabetes |
| Blood glucose levels using a CGM | Dataset from 10 T1DM patients | CNN and LSTM | CNN outperformed LSTM in blood glucose prediction tests | |
| Zhu et al. (2018a) | Dataset containing glucose levels, insulin dosages, and carb intake | Data of 6 participants with T1DM collected over 8 weeks | CNN | The best CNN model achieved an RMSE of 21.72 mg/dl in predicting glucose levels |
| Porumb et al. (2020a) | ECG signals | Data acquired from 8 healthy elderly participants | CNN with autoenecoders | Achieved 90% in nocturnal hypoglycemia detection |
| Shahid et al. (2021) | CGM, carbohydrates, and insulin | Simulated data representing 10 patients with T1DM | CNN with gated recurrent unit neural networks | RMSE of 6.04 mg/dl for the 30 min prediction horizon |
| Samir et al. (2021) | CGM, food, and human activity | Data acquired from patients with diabetes | CNN | Classifying hyperglycemia with 93.2% accuracy |
| Zhu et al. (2020b) | Meal intake, glucose levels, and insulin dosage | Used simulated and real datasets of T1DM subjects | Dilated RNN and transfer learning | Forecasting future glucose levels with an RMSE of 18.9 mg/dl |
| Faruqui et al. (2019) | Data contained glucose levels, diet, and physical activity | Data acquired from 10 subjects with T2DM monitored for 6 months | LSTM based structure | 84.12% of next day predictions in zone A of Clark’s error grid |
| Gu et al. (2017) | CGM, Insulin, food, and physical activity | Dataset collected from 112 participants (35 healthy, 38 with T1DM, and 39 with T2DM) | RNN | Achieved a blood glucose inference accuracy of 82.14% |
| Nasser et al. (2021) | CGM | Data acquired from 10 patients with T1DM | RNN with restricted boltzmann machines | RMSE value of 15.59 mg/dl for 30 min prediction horizon |
| Eom and Lee (2022) | CGM, insulin, and carbohydrate | Dataset of 30 patients | Weibull Time To Event RNN | Predicting future episodes of hypoglycemia with an RMSE of 12.56 mg/dl |
| Zhang et al. (2017) | Physiological parameters and metabolic rate | Data collected from healthy subjects, senior citizens, and patients with diabetes | DT and ANN | Blood glucose prediction with 88.53% accuracy |
| Tsai et al. (2020) | PPG signals | 9 subjects with T2DM | DT based structure (Random forest) | 90% accuracy in predicting glucose |
| Maritsch et al. (2020) | HRV using PPG smartwatch | One participant with T1DM | DT based structure (Gradient boosting) | 82.7% accuracy in detecting hypoglycemia |
| Reddy et al. (2019) | HR, physical activity, and glucose levels | 43 subjects with T1DM during aerobic exercise | DT | 86.7% accuracy in predicting hypoglycemia during exercises |
| CGM | Simulated CGM data from 40 participants with T1DM | DT and Adaboost | RMSE value of 2.204 mg/dl in predicting blood glucose levels | |
| Sankhala et al. (2022) | Sweat glucose | Three participants | DT | RMSE value of 0.1 mg/dl in estimating sweat glucose |
| Habibullah et al. (2019) | NIR signals | Used artificial blood samples with different glucose concentration | SVM | 77.5% accuracy with PCA |
| Malik et al. (2016) | Salivary electrochemical properties | 175 participants including participants with T2DM | SVM, ANN, and logistic regression | SVM achieved 85% accuracy in detecting fasting blood glucose |
| Marling et al. (2016) | Heart rate, galvanic skin response, CGM, and temperature | One participant with T1DM | SVM | SVM with a linear kernel showed a promising performance |
| CGM, HR, steps, sleep, and calories | 10 subjects with T1DM | SVM and MLP | SVM achieved the best results in predicting nocturnal hypoglycemia | |
| Shokrekhodaei et al. (2021) | Intensity data of an optical sensor | In aqueous solutions containing glucose | SVM | RMSE value of 12.5 mg/dl and 99.55% accuracy |
| Yang et al. (2019) | CGM | 100 participants with T1DM and T2DM | ARIMA | 9.4% false rate in predicting future blood glucose trends |
| Prendin et al. (2021) | CGM | Partial data from 141 patients with T1DM | ARIMA | RMSE of 22.15 mg/dl in 30 min prediction horizon and also in hypoglycemia detection with 64% precision and 82% recall |
A summary of machine learning based blood glucose monitoring contributions.
3.1 Artificial Neural Network
An ANN consists of interconnected layers of perceptrons that can learn patterns of data by adjusting numerical weights attached to each connection. Several popular ANN architectures have been considered for the purpose of blood glucose monitoring (
3.1.1 Convolutional Neural Network
CNN is a type of ANN mainly used for processing and recognition of grid like data (e.g., images) and typically consists of three types of building blocks (LeCun et al., 2015). The first two building blocks (i.e., convolution and pooling layers) perform feature extraction while the third (i.e., fully connected layer) matches the extracted features with outputs or classes. The stacking of these blocks constitutes the CNN architecture (Figure 5A). To make a prediction, the first layer in CNN receives a vector of input values that can either be spatially related such as images, or short sequences of time series data such as multi-dimensional biometric data. These values are then passed to several layers that perform two operations; convolutions, and downsampling. In the convolution layers, specific features are extracted from nearby inputs by matching learned meaningful patterns with the sequence of data that are fed into the layer. The results of this operation are then forwarded to a next layer which chooses the patterns that were most apparent (i.e., max-pooling). This operation is repeatedly done depending on the number of layers used. Finally, the resulting patterns are fed into a fully connected ANN that produces a prediction based on the information presented on the last preceding layer. The number of patterns to match in each layer (i.e., filters) and number of stacked layers are hyper-parameters that are tuned based on the desired prediction performance and computational cost.
FIGURE 5

Illustrations of commonly considered artificial neural networks in glucose monitoring applications and an example of a study that considered a combined architecture. (A) Illustration of a CNN model. The first layer is the input layer which holds values of the input data that is followed by a convolution layer, which serves to extract meaningful patterns in partial regions of the input. Next, comes the pooling layer that will perform a downsampling operation that reduces the number of parameters. The extracted features are then passed to a fully connected layer to make the final prediction. (B) Illustration of RNN model. The RNN model consists of an input layer (X), hidden layers to model sequential information (from h0 to hn where n is the number of hidden layers), and an output layer (O). The structure is connected with weights that link the input layer to the first hidden layer (Wxh), the hidden layers together (Whh), and the last hidden layer to the output layer (Why). (C) An example of a study that considered the application of CNN and RNN to predict the occurrence of hypoglycemia based on ECG data (Adapted with permission from Porumb et al. (2020b)). The isolated heartbeats were combined into segments of 5 min each and each segment has been assigned a label (i.e., low or normal glucose level) based on the recorded CGM value.
CNN has been utilized in many applications, e.g., heart-beat classification (
A more recent study considered a combination of CNN with autoencoders to detect nocturnal hypoglycemia (Porumb et al., 2020a). The study used a non-invasive wearable (i.e., ECG) and CGM to acquire data from 25 elderly subjects under controlled conditions for up to 36 h. The collected data were used to train and evaluate personalized deep learning models. The best trained model was able to distinguish low blood glucose trends with an accuracy of 90% in the test dataset. Samir et al. (2021) proposed another CNN based model to predict blood glucose trends by classifying food and human activity and classified hyperglycemia with an accuracy of 93.2% based on the D1NAMO dataset of diabetic patients using CGMs alongside food images and acceleration data (
3.1.2 Recurrent Neural Network
RNN is a class of ANNs with feedback signals developed to learn sequential ordered data (i.e., time series) or time-varying patterns such as that found in speech (Medsker and Jain, 1999). The prediction in RNN relies on previous information maintained internally. The hidden layers act like a memory that captures the information about a sequence. RNN models consist of interconnected layers of neurons just as in normal ANNs (Figure 5B). The difference in RNN is their ability to take into account information from previous predictions. Specifically, each hidden layer also considers the outputs of hidden layers from preceding predictions. This allows the network to capture information from ordered sequences of data. The learning in a standard RNN structure might be limited and hindered due to the vanishing gradients problem, hence, a structure based on RNN (e.g., long-short term memory (LSTM)) is usually used to combat that (Hochreiter and Schmidhuber, 1997).
Several applications used RNN based models to recognize different sequences in human activities (Pienaar and Malekian, 2019), emotion recognition based on videos (Fan et al., 2016), and the characterization of unwanted interactions with small robots (
In another work, an LSTM-based algorithm was applied to 6-months data on diet, glucose levels, and physical activity in 10 patients with type 2 diabetes to forecast daily glucose concentrations (Faruqui et al., 2019). They were able to predict the next day glucose concentrations with 84.12% of the predicted values in zone A based on Clark error grid when compared to the true values of blood glucose. However, the study was limited to a small sample size and was affected by individual variations and data collection challenges. An inference system based on a smartphone to monitor blood glucose non-invasively was developed by Gu et al. (2017). They collected data about insulin, drug dosage, food intake, sleep quality, and physical activities along with CGM. The authors evaluated their system on 112 subjects and the implementation of RNN achieved an accuracy of 82.14% in tracking blood glucose levels into four classes. A recent study also assessed the ability to estimate future (i.e., 30 min prediction horizon) blood glucose levels using an IoT device with CGM sensor, an application layer protocol, and prediction model on the cloud (Nasser et al., 2021). Based on a model consisting of RNNs and restricted boltzmann machines, the proposed system achieved an RMSE value of 15.59 mg/dl in data acquired from ten patients with type 1 diabetes. Another recent study proposed using Weibull Time To Event RNN (i.e., WTTE-RNN (Martinsson, 2017)) to predict future episodes of hypoglycemia and reported an RMSE of 12.56 mg/dl (Eom and Lee, 2022).
3.2 Decision Tree
A decision tree (DT) is a divide-and-conquer method that partitions data such that it becomes easy to classify. A typical DT model consists of nodes that split the data based on attribute-value combinations. Data are split repeatedly until a given criteria is satisfied (e.g., similarity of data in each partition). DT has been applied to extract and analyse information from large datasets (i.e., data mining) and in machine learning (e.g., classification and regression) (Hastie et al., 2009; Navada et al., 2011). Classification trees are used when the target values are discrete while regression trees are used when the target values are continuous. Compared to other machine learning techniques, DT has the advantage of model interpretability. It provides an insight on the most influential data attributes related to the task at hand and helps to plan future experiments (Myles et al., 2004). Different forms of DT (e.g., classification, regression, and forests) can be considered depending on the problem and the desired output (Navada et al., 2011).
Decision trees were used as prediction models for risk factor interactions in diabetes and to identify subjects with impaired glucose metabolism (Ramezankhani et al., 2016; Hische et al., 2010). Different DT models have been used to identify vital indicators in relation to blood glucose prediction (Liu et al., 2020). One study developed a non-invasive system to detect blood glucose levels based on the conservation-of-energy method and physiological parameters (Zhang et al., 2017). The study acquired data samples from 400 participants (i.e., healthy and diabetic patients) that were then used in the model development and evaluation. Using an algorithm that combines a DT and neural network, the proposed approach was able to provide a blood glucose prediction with an accuracy of 88.53% in classifying new blood glucose samples.
A recent study considered a wearable device to estimate blood glucose based on photoplethysmography (PPG) signals (Tsai et al., 2020). The data were acquired from 9 patients with type 2 diabetes in a stable physical position (i.e., sitting). The blood glucose levels were acquired using a finger prick device (i.e., Accu-Chek). A machine learning algorithm based on decision tree (i.e., random forest) was considered to build personalized and generalized models that achieved 80% and 90%, respectively when tested with unseen data. Another study recently evaluated several ensemble machine learning models to provide a generalized blood glucose prediction (
DT based models were also utilized to predict blood glucose trends, such as hypoglycemia. One study proposed using a machine learning model based on decision trees (i.e., gradient boosting) to detect hypoglycemia using a wearable device (Maritsch et al., 2020). Using features acquired from the heart variability rate, the study developed a machine learning model based on the data acquired from one participant with type 1 diabetes. The results of the unseen samples demonstrated the possibility of detecting hypoglycemic events with 82.7% accuracy. Reddy et al. (2019) investigated the possibility of predicting hypoglycemia in 43 adults with type 1 diabetes who were performing aerobic exercise. The extracted features included physical activity, heart rate, anthropometric data, energy expenditure estimate, glucose readings, and physical activity. The results of two developed models based on decision trees showed promising results in predicting hypoglycemia with an accuracy of 86.7%.
3.3 Support Vector Machine
SVM is a supervised machine learning technique used in classification and regression. In classification problems, the SVM learns from the labeled training data how to best categorize data that belongs to one of two classes by finding the optimal hyperplane that separates them (Hearst et al., 1998; Brereton et al., 2010). The separation in SVM can be based on a linear, or non-linear combination of features depending on the complexity of the task at hand and feature dependencies. In case of a non-linear SVM, kernel functions are used to transform a problem to a linearly separable one by projecting the problem from a low-dimensional space to a high-dimensional one (Patle and Chouhan, 2013). SVM is also used in linear and non-linear regression. The principle of SVM for regression is to find a flat function that satisfies a deviation criterion from the target outputs with less restriction to minimize the errors (Smola and Schölkopf, 2004). In case of a non-linear regression, a similar technique to that used in classification is applied.
SVM has been considered in different areas and in many different applications such as in cancer genomics (Huang et al., 2018), chemistry (Ivanciuc, 2007), autism therapy (
SVM has also been applied in glucose monitoring and long-term diabetes outcome prediction (
A few studies have also considered using SVM to predict the occurrence of hypoglycemia among patients with type 1 diabetes. For example, one study used a non-invasive wearable device that measured air temperature, heart rate, and galvanic skin response to acquire data from one participant with type 1 diabetes for 2 months (Marling et al., 2016). The blood glucose data were acquired using a Dexcom CGM device and contained 34 hypoglycemic events, each lasting for 10 min or more. SVM with a linear kernel achieved the best performance, but the results were limited to one participant. Another study used a CGM device (FreeStyle Libre) in 10 participants with type 1 diabetes over 12 weeks and showed that SVM achieved a high sensitivity (78.75%) and specificity (82.15%) to predict nocturnal hypoglycemia (
3.4 Autoregressive Integrated Moving Average
ARIMA is a linear time series model used to predict or forecast future values based on past values. It is a function that includes differencing operators, and autoregressive and moving average terms (
4 Discussion
The past five years has witnessed considerable advances in the development of sensors that measure different modalities which correlate with blood glucose. Glucose levels in tears, saliva, and sweat are related to the blood glucose levels and the advances in non-invasive wearable technology have not only allowed an estimation of blood glucose levels, but also the prediction of hypoglycemia. The developed wearables have targeted the heart, skin, eyes, and mouth using various technologies such as electromagnetic and bioimpedance. Some of these solutions have been augmented with machine learning techniques which have yielded promising outcomes, especially in hypoglycemia prediction. However, there remain considerable challenges before these devices can achieve FDA approval.
4.1 Participants and Testing Conditions
A major limitation in the studies to date is the number of participants with the majority of studies being limited to a few participants. This limitation clearly affects their clinical use and hinders the generalizability of the findings. Furthermore, some studies did not even undertake trials in patients with diabetes, which limits the applicability of the proposed solutions to the target end-users. Many of the studies reported only the initial or exploratory results of the developed sensors or wearable prototypes. The majority of studies were conducted under controlled settings and were limited to subjects of a younger age. Future studies should recruit larger numbers of participants with a wider age range and especially patients with type 1 or type 2 diabetes.
4.2 Bodily Fluids Glucose Sensors
The glucose levels in these bodily fluids were reported to correlate with blood glucose concentrations, but the glucose concentrations in these bodily fluids are low compared to that found in the blood. This represents a major challenge for the development of sensors and technologies that rely on bodily fluids to estimate blood glucose levels and trends. Hence, extra considerations such as the enhancement of sensitivity and interference elimination become crucial in the development of such sensors (Yao et al., 2011). Another challenge is the way a body fluid sample is collected (e.g., stimulated or unstimulated) as it was found to influence the content of the acquired sample. For example, diagnostic biomarkers were found to be higher in unstimulated saliva compared to stimulated saliva (Miller et al., 2010). A study has also shown that unstimulated saliva is highly accurate for predicting blood glucose (
4.3 Noise Susceptibility and Physiological Effects
Some of the physiological sensors were found to be influenced by daily circadian rhythms in heart rate and sweating (Elvebakk et al., 2018). Error can also arise due to non-linear dynamics of physiological signs being measured, making sensors more susceptible to error and high noise levels (Zanon et al., 2017; Gadaleta et al., 2019). Some studies have reported the adverse impact of environmental conditions (e.g., pH, humidity, and temperature) (Park et al., 2006; Elsherif et al., 2018). The time delay of a sensor or wearable device must be minimized to capture the rapid changes in blood glucose (Marcus et al., 2020). For patch-like sensors, the multidirectional stretchability of the sensor is an issue that needs careful considerations to ensure accuracy and stability under multi stretching cycles (
4.4 Data Acquisition and Model Generalizability
A key challenge in machine learning is acquiring enough comprehensive data to train and test a model that can then be generalized to a wider population. Many studies were limited in terms of the number of participants, duration of the study, and the incidence of clinically relevant severe hypoglycemia. Additionally, most studies established their methods and techniques in relation to CGM data that may be limited in certain scenarios. There is a need to include a larger population of patients, especially with diabetes, to account for the inter-individual differences and to establish better validation of the proposed solutions. To capture meaningful data, the duration of trials for data acquisition and the incidence of hypoglycemia need to be sufficient to avoid unbalanced or skewed data (Marling et al., 2016). In the case of imbalanced data, oversampling techniques can address this and improve the accuracy (Mayo et al., 2019). Acquiring data from the same participant for longer periods allows the machine learning algorithm to combat intra-individual differences and increases overall prediction performance (
4.5 Model Interpretability
Although machine learning algorithms have great predictive potential, the majority of these algorithms are black box models and lack the means to explain their predictions. The interpretability requirement of a machine learning model in health care is crucial (
5 Conclusions and Future Directions
This review highlights the considerable progress made over the past five years in the area of non-invasive blood glucose monitoring using wearable technologies and sensors alongside machine learning algorithms. The devices have varied modalities and adopted technologies with novel approaches utilizing machine learning techniques to provide meaningful interpretations of multiple physiological parameters. However, there remain considerable limitations and challenges that hinder FDA approval and more widespread adoption of such technologies in patients. We therefore recommend future studies to focus on the following areas:
1. The recruitment of a larger number of participants especially patients with diabetes to validate the proposed techniques for use in the clinical arena.
2. Validating glucose sensors by adequate collection without contamination of bodily fluids.
3. Miniaturization of electronics and sensors for practical deployment.
4. Consistent evaluation of algorithms in personalized vs. generalized scheme, where a model is trained either on a target individual or a group of subjects.
5. The utilization of a wider set of physiological parameters for machine learning and data assimilation techniques while establishing effectiveness of each sensor’s contribution.
6. Investigating means to reduce computational cost and energy in wearable devices.
7. The development of interpretable machine learning models.
Statements
Author contributions
AYA wrote the first draft. HA, HG, AA-A, and KKS produced the figures and revised the manuscript. J-JC and RAM oversaw the work and revised the manuscript. All authors contributed to the work and approved the submitted version.
Funding
The work is supported by an NPRP grant from the Qatar National Research Fund under the grant No. NPRP 11S-0110-180247. The statements made herein are solely the responsibility of the authors.
Conflict of interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
1
AashimaBhargavS.KaushikS.DuttV.et al (2022). “A Combination of Decision Trees with Machine Learning Ensembles for Blood Glucose Level Predictions,” in Proceedings of International Conference on Data Science and Applications (Berlin, Germany: Springer), 533–548.
2
AbbasH. T.AlicL.ErraguntlaM.JiJ. X.Abdul-GhaniM.AbbasiQ. H.et al (2019). Predicting Long-Term Type 2 Diabetes with Support Vector Machine Using Oral Glucose Tolerance Test. Plos one14, e0219636. 10.1371/journal.pone.0219636
3
AbbasH.ZahedK.AlicL.ZhuY.SasangoharF.MehtaR.et al (2018). “A Wearable, Low-Cost Hand Tremor Sensor for Detecting Hypoglycemic Events in Diabetic Patients,” in 2018 IEEE International RF and Microwave Conference (RFM) (Penang, Malaysia: IEEE), 182–184. 10.1109/rfm.2018.8846546
4
AccursoF. J.Van GoorF.ZhaJ.StoneA. J.DongQ.OrdonezC. L.et al (2014). Sweat Chloride as a Biomarker of CFTR Activity: Proof of Concept and Ivacaftor Clinical Trial Data. J. Cyst. Fibros.13, 139–147. 10.1016/j.jcf.2013.09.007
5
AcharyaU. R.OhS. L.HagiwaraY.TanJ. H.AdamM.GertychA.et al (2017). A Deep Convolutional Neural Network Model to Classify Heartbeats. Comput. Biol. Med.89, 389–396. 10.1016/j.compbiomed.2017.08.022
6
AhlqvistE.PrasadR. B.GroopL. (2020). Subtypes of Type 2 Diabetes Determined from Clinical Parameters. Diabetes69, 2086–2093. 10.2337/dbi20-0001
7
AhlqvistE.StormP.KäräjämäkiA.MartinellM.DorkhanM.CarlssonA.et al (2018). Novel Subgroups of Adult-Onset Diabetes and Their Association with Outcomes: a Data-Driven Cluster Analysis of Six Variables. Lancet Diabetes & Endocrinol.6, 361–369. 10.1016/s2213-8587(18)30051-2
8
AiharaM.KubotaN.MinamiT.ShirakawaR.SakuraiY.HayashiT.et al (2020). Association between Tear and Blood Glucose Concentrations: Random Intercept Model Adjusted with Confounders in Tear Samples Negative for Occult Blood. J. Diabetes Investigation12, 266. 10.1111/jdi.13344
9
Al-JarrahO. Y.YooP. D.MuhaidatS.KaragiannidisG. K.TahaK. (2015). Efficient Machine Learning for Big Data: A Review. Big Data Res.2, 87–93. 10.1016/j.bdr.2015.04.001
10
AlbanA. Q.AyeshM.AlhaddadA. Y.Al-AliA. K.SoW. C.ConnorO.et al (2021). “Detection of Challenging Behaviours of Children with Autism Using Wearable Sensors during Interactions with Social Robots,” in 2021 30th IEEE International Conference on Robot & Human Interactive Communication (RO-MAN) (Vancouver, BC, Canada: IEEE), 852–857. 10.1109/ro-man50785.2021.9515459
11
AlbersD. J.LevineM. E.StuartA.MamykinaL.GluckmanB.HripcsakG. (2018). Mechanistic Machine Learning: How Data Assimilation Leverages Physiologic Knowledge Using Bayesian Inference to Forecast the Future, Infer the Present, and Phenotype. J. Am. Med. Inf. Assoc.25, 1392–1401. 10.1093/jamia/ocy106
12
AlbersD. J.LevineM.GluckmanB.GinsbergH.HripcsakG.MamykinaL. (2017). Personalized Glucose Forecasting for Type 2 Diabetes Using Data Assimilation. PLoS Comput. Biol.13, e1005232. 10.1371/journal.pcbi.1005232
13
AlhaddadA. Y.CabibihanJ.-J.BonariniA. (2020). Influence of Reaction Time in the Emotional Response of a Companion Robot to a Child’s Aggressive Interaction. Int. J. Soc. Robotics12, 1279. 10.1007/s12369-020-00626-z
14
AlibertiA.PupilloI.TernaS.MaciiE.Di CataldoS.PattiE.et al (2019). A Multi-Patient Data-Driven Approach to Blood Glucose Prediction. IEEE Access7, 69311–69325. 10.1109/access.2019.2919184
15
AljihmaniL.AbbasH.ZhuY.MehtaR. K.SasangoharF.ErraguntlaM.et al (2019). “Features of Physiological Tremor in Diabetic Patients,” in 2019 IEEE International Smart Cities Conference (ISC2) (Casablanca, Morocco: IEEE). 10.1109/isc246665.2019.9071646
16
AmerS.YousufM.SiddqiuiP. Q.AlamJ. (2001). Salivary Glucose Concentrations in Patients with Diabetes Mellitus-Aa Minimally Invasive Technique for Monitoring Blood Glucose Levels. Pak J. Pharm. Sci.14, 33–37.
17
American Diabetes Association (2020). 2. Classification and Diagnosis of Diabetes: Standards of Medical Care in Diabetes—2021. Diabetes Care44, S15–S33. 10.2337/dc21-S002
18
American Diabetes Association (2020a). 6. Glycemic Targets: Standards of Medical Care in Diabetes—2021. Diabetes Care44, S73–S84. 10.2337/dc21-S006
19
American Diabetes Association (2020b). 9. Pharmacologic Approaches to Glycemic Treatment: Standards of Medical Care in Diabetes—2021. Diabetes Care44, S111–S124. 10.2337/dc21-S009
20
American Diabetes Association (2013). Diagnosis and Classification of Diabetes Mellitus. Diabetes Care37, S81–S90. 10.2337/dc14-S081
21
ArakawaT.KurokiY.NittaH.ChouhanP.TomaK.SawadaS.-i.et al (2016). Mouthguard Biosensor with Telemetry System for Monitoring of Saliva Glucose: A Novel Cavitas Sensor. Biosens. Bioelectron.84, 106–111. 10.1016/j.bios.2015.12.014
22
ArakawaT.TomotoK.NittaH.TomaK.TakeuchiS.SekitaT.et al (2020). A Wearable Cellulose Acetate-Coated Mouthguard Biosensor for In Vivo Salivary Glucose Measurement. Anal. Chem.92, 12201–12207. 10.1021/acs.analchem.0c01201
23
American Diabetes Association (2021). 7. Diabetes Technology: Standards of Medical Care in Diabetes-2021. Diabetes Care44, S85–S99. 10.2337/dc21-S007
24
BaeC. W.ToiP. T.KimB. Y.LeeW. I.LeeH. B.HanifA.et al (2019). Fully Stretchable Capillary Microfluidics-Integrated Nanoporous Gold Electrochemical Sensor for Wearable Continuous Glucose Monitoring. ACS Appl. Mat. Interfaces11, 14567–14575. 10.1021/acsami.9b00848
25
BagalàF.BeckerC.CappelloA.ChiariL.AminianK.HausdorffJ. M.et al (2012). Evaluation of Accelerometer-Based Fall Detection Algorithms on Real-World Falls. PloS one7, e37062. 10.1371/journal.pone.0037062
26
BakerL. B. (2019). Physiology of Sweat Gland Function: The Roles of Sweating and Sweat Composition in Human Health. Temperature6, 211–259. 10.1080/23328940.2019.1632145
27
BariyaM.NyeinH. Y. Y.JaveyA. (2018). Wearable Sweat Sensors. Nat. Electron1, 160–171. 10.1038/s41928-018-0043-y
28
BarmanI.KongC.-R.DingariN. C.DasariR. R.FeldM. S. (2010). Development of Robust Calibration Models Using Support Vector Machines for Spectroscopic Monitoring of Blood Glucose. Anal. Chem.82, 9719–9726. 10.1021/ac101754n
29
Ben AliJ.HamdiT.FnaiechN.Di CostanzoV.FnaiechF.GinouxJ.-M. (2018). Continuous Blood Glucose Level Prediction of Type 1 Diabetes Based on Artificial Neural Network. Biocybern. Biomed. Eng.38, 828–840. 10.1016/j.bbe.2018.06.005
30
BekkinkM. O.KoenemanM.de GalanB. E.BredieS. J. (2019). Early Detection of Hypoglycemia in Type 1 Diabetes Using Heart Rate Variability Measured by a Wearable Device. Diabetes Care42, 689–692. 10.2337/dc18-1843
31
BertachiA.ViñalsC.BiagiL.ContrerasI.VehíJ.CongetI.et al (2020). Prediction of Nocturnal Hypoglycemia in Adults with Type 1 Diabetes under Multiple Daily Injections Using Continuous Glucose Monitoring and Physical Activity Monitor. Sensors20, 1705. 10.3390/s20061705
32
BiharE.WustoniS.PappaA. M.SalamaK. N.BaranD.InalS. (2018). A Fully Inkjet-Printed Disposable Glucose Sensor on Paper. npj Flex. Electron.2. 10.1038/s41528-018-0044-y
33
BoxG. E.JenkinsG. M.ReinselG. C. (2011). Time Series Analysis: Forecasting and Control, Vol. 734. Hoboken, NJ, USA: John Wiley & Sons.
34
BreretonRichardG.GavinR. L. (2010). Support Vector Machines for Classification and Regression. Analyst135 (2), 230–267.
35
BriscoeV. J.DavisS. N. (2006). Hypoglycemia in Type 1 and Type 2 Diabetes: Physiology, Pathophysiology,and Management. Physiology, Pathophysiol. Manag.24, 115–121. 10.2337/diaclin.24.3.115
36
BrockwellP. J.DavisR. A.CalderM. V. (2002). Introduction to Time Series and Forecasting, Vol. 2. Berlin, Germany: Springer.
37
BruenD.DelaneyC.FloreaL.DiamondD. (2017). Glucose Sensing for Diabetes Monitoring: Recent Developments. Sensors17, 1866. 10.3390/s17081866
38
ByvatovE.SchneiderG. (2003). Support Vector Machine Applications in Bioinformatics. Appl. Bioinforma.2, 67–77.
39
CabibihanJ.-J.ChellaliR.SoC. W. C.AldosariM.ConnorO.AlhaddadA. Y.et al (2018). “Social Robots and Wearable Sensors for Mitigating Meltdowns in Autism - A Pilot Test,” in International Conference on Social Robotics (Berlin, Germany: Springer), 103–114. 10.1007/978-3-030-05204-1_11
40
CabibihanJ.-J.JavedH.AldosariM.FrazierT.ElbashirH. (2016). Sensing Technologies for Autism Spectrum Disorder Screening and Intervention. Sensors17, 46. 10.3390/s17010046
41
Cano-GarciaH.GouzouasisI.SotiriouI.SahaS.PalikarasG.KosmasP.et al (2016). “Reflection and Transmission Measurements Using 60 Ghz Patch Antennas in the Presence of Animal Tissue for Non-invasive Glucose Sensing,” in 2016 10th European Conference on Antennas and Propagation (EuCAP) (Davos, Switzerland: IEEE), 1–3. 10.1109/eucap.2016.7481178
42
CharambaB.LiewA.CoenE.NewellJ.O’BrienT.WijnsW.et al (2021). Modelling the Relationship between Continuously Measured Glucose and Electrocardiographic Data in Adults with Type 1 Diabetes Mellitus. Endocrinol. diabetes & metabolism4, e00263. 10.1002/edm2.263
43
ChenC.DongZ.-Q.ShenJ.-H.ChenH.-W.ZhuY.-H.ZhuZ.-G. (2018). 2d Photonic Crystal Hydrogel Sensor for Tear Glucose Monitoring. ACS Omega3, 3211–3217. 10.1021/acsomega.7b02046
44
ChenJ.ZhuX.JuY.MaB.ZhaoC.LiuH. (2019). Electrocatalytic Oxidation of Glucose on Bronze for Monitoring of Saliva Glucose Using a Smart Toothbrush. Sensors Actuators B Chem.285, 56–61. 10.1016/j.snb.2019.01.017
45
ChenR.JinZ.ColónL. A. (1996). Analysis of Tear Fluid by Ce/lif: a Noninvasive Approach for Glucose Monitoring. J. Capill. Electrophor.3, 243–248.
46
ChenX.XuY.WongD. W. K.WongT. Y.LiuJ. (2015). Glaucoma Detection Based on Deep Convolutional Neural Network. Annu. Int. Conf. IEEE Eng. Med. Biol. Soc.2015, 715–718. 10.1109/EMBC.2015.7318462
47
Cheol JeongI.BychkovD.SearsonP. C. (2019). Wearable Devices for Precision Medicine and Health State Monitoring. IEEE Trans. Biomed. Eng.66, 1242–1258. 10.1109/tbme.2018.2871638
48
ChouY. H.HuangW. L.ChangC. H.YangC. C. H.KuoT. B. J.LinS. L.et al (2019). Heart Rate Variability as a Predictor of Rapid Renal Function Deterioration in Chronic Kidney Disease Patients. Nephrology24, 806–813. 10.1111/nep.13514
49
ChuJ.YangW. T.LuW. R.ChangY. T.HsiehT. H.YangF. L. (2021). 90% Accuracy for Photoplethysmography-Based Non-invasive Blood Glucose Prediction by Deep Learning with Cohort Arrangement and Quarterly Measured Hba1c. Sensors (Basel)21, 7815. 10.3390/s21237815
50
CorinoV. D. A.LaureantiR.FerrantiL.ScarpiniG.LombardiF.MainardiL. T. (2017). Detection of Atrial Fibrillation Episodes Using a Wristband Device. Physiol. Meas.38, 787–799. 10.1088/1361-6579/aa5dd7
51
CostantineJ.KanjR.EidA. (2019). Novel Non-invasive Biological, Chemical Markers and Tracers Monitoring Device in Blood Including Glucose Monitoring Using Adaptive Rf Circuits and Antenna Design. US Patent App. 16/152,990.
52
CuiY.ZhangH.ZhuJ.PengL.DuanZ.LiuT.et al (2021). Unstimulated Parotid Saliva Is a Better Method for Blood Glucose Prediction. Appl. Sci.11, 11367. 10.3390/app112311367
53
DamianoE. R.El-KhatibF. H.ZhengH.NathanD. M.RussellS. J. (2012). A Comparative Effectiveness Analysis of Three Continuous Glucose Monitors. Diabetes Care36, 251–259. 10.2337/dc12-0070
54
DarttD. A.HodgesR. R.ZoukhriD. (2005). “Tears and Their Secretion,” in Advances in Organ Biology (Amsterdam, Netherlands: Elsevier), 21–82. 10.1016/s1569-2590(05)10002-0
55
DasT.GhoshA.GuhaS.BasakP. (2017). “Early Detection of Diabetes Based on Skin Impedance Spectrogram and Heart Rate Variability Noninvasively,” in 2017 1st International Conference on Electronics, Materials Engineering and Nano-Technology (IEMENTech) (Kolkata, India: IEEE). 10.1109/iementech.2017.8076984
56
DawesC. (1974). Rhythms in Salivary Flow Rate and Composition. Int. J. Chronobiol2, 253–279.
57
de AlmeidaP. D. V.GrégioA. M.MachadoM. A.De LimaA. A.AzevedoL. R. (2008). Saliva Composition and Functions: a Comprehensive Review. J. Contemp. Dent. Pract.9, 72–80.
58
de BoerI. H. (2008). Insulin Therapy, Hyperglycemia, and Hypertension in Type 1 Diabetes Mellitus. Arch. Intern Med.168, 1867. 10.1001/archinternmed.2008.2
59
DingS.SchumacherM. (2016). Sensor Monitoring of Physical Activity to Improve Glucose Management in Diabetic Patients: A Review. Sensors16, 589. 10.3390/s16040589
60
DioufA.BouchikhiB.El BariN. (2019). A Nonenzymatic Electrochemical Glucose Sensor Based on Molecularly Imprinted Polymer and its Application in Measuring Saliva Glucose. Mater. Sci. Eng. C98, 1196–1209. 10.1016/j.msec.2019.01.001
61
DiouriO.CiglerM.VettorettiM.MaderJ. K.ChoudharyP. (2021). Hypoglycaemia Detection and Prediction Techniques: A Systematic Review on the Latest Developments. Diabetes/Metabolism Res. Rev.37, e3449. 10.1002/dmrr.3449
62
DubossonF.RanvierJ.-E.BromuriS.CalbimonteJ.-P.RuizJ.SchumacherM. (2018). The Open D1namo Dataset: A Multi-Modal Dataset for Research on Non-invasive Type 1 Diabetes Management. Inf. Med. Unlocked13, 92–100. 10.1016/j.imu.2018.09.003
63
DunnJ.RungeR.SnyderM. (2018). Wearables and the Medical Revolution. Pers. Med.15, 429–448. 10.2217/pme-2018-0044
64
DuongH. D.SohnO.-J.RheeJ. I. (2020). Development of a Ratiometric Fluorescent Glucose Sensor Using an Oxygen-Sensing Membrane Immobilized with Glucose Oxidase for the Detection of Glucose in Tears. Biosensors10, 86. 10.3390/bios10080086
65
El IdrissiT.IdriA. (2020). “Deep Learning for Blood Glucose Prediction: CNN vs LSTM,” in Computational Science and its Applications – ICCSA 2020 (Berlin, Germany: Springer International Publishing), 379–393. 10.1007/978-3-030-58802-1_28
66
EljilK. S.QadahG.PasquierM. (2016). Predicting Hypoglycemia in Diabetic Patients Using Time-Sensitive Artificial Neural Networks. Int. J. Healthc. Inf. Syst. Inf.11, 70–88. 10.4018/ijhisi.2016100104
67
ElShawiR.SherifY.Al-MallahM.SakrS. (2020). Interpretability in Healthcare: A Comparative Study of Local Machine Learning Interpretability Techniques. Comput. Intell.10.1111/coin.12410
68
ElsherifM.AlamF.SalihA. E.AlQattanB.YetisenA. K.ButtH. (2021). Wearable Bifocal Contact Lens for Continual Glucose Monitoring Integrated with Smartphone Readers. Small17, 2102876. 10.1002/smll.202102876
69
ElsherifM.HassanM. U.YetisenA. K.ButtH. (2018). Wearable Contact Lens Biosensors for Continuous Glucose Monitoring Using Smartphones. ACS Nano12, 5452–5462. 10.1021/acsnano.8b00829
70
ElvebakkO.TronstadC.BirkelandK. I.JenssenT. G.BjørgaasM. R.FrøslieK. F.et al (2018). Evaluation of Hypoglycaemia with Non-invasive Sensors in People with Type 1 Diabetes and Impaired Awareness of Hypoglycaemia. Sci. Rep.8, 14722. 10.1038/s41598-018-33189-1
71
EomT. H.LeeH. (2022). “A Wtte-Rnn Based Hypoglycemic Event Prediction,” in 2022 International Conference on Information Networking (ICOIN) (Seoul, Korea: IEEE), 378–379. 10.1109/icoin53446.2022.9687150
72
EscolarS.AbaldeaM. J.DondoJ. D.RincónF.LópezJ. C. (2016). “Early Detection of Hypoglycemia Events Based on Biometric Sensors Prototyped on FPGAs,” in Ubiquitous Computing and Ambient Intelligence (Berlin, Germany: Springer International Publishing), 133–145. 10.1007/978-3-319-48746-5_14
73
FanY.LuX.LiD.LiuY. (2016). “Video-based Emotion Recognition Using Cnn-Rnn and C3d Hybrid Networks,” in Proceedings of the 18th ACM International Conference on Multimodal Interaction, 445–450. 10.1145/2993148.2997632
74
FaruquiS. H. A.DuY.MekaR.AlaeddiniA.LiC.ShirinkamS.et al (2019). Development of a Deep Learning Model for Dynamic Forecasting of Blood Glucose Level for Type 2 Diabetes Mellitus: Secondary Analysis of a Randomized Controlled Trial. JMIR Mhealth Uhealth7, e14452. 10.2196/14452
75
GadaletaM.FacchinettiA.GrisanE.RossiM. (2019). Prediction of Adverse Glycemic Events from Continuous Glucose Monitoring Signal. IEEE J. Biomed. Health Inf.23, 650–659. 10.1109/jbhi.2018.2823763
76
GaoY.HuangY.ChenJ.LiuY.XuY.NingX. (2021). A Novel Luminescent “Nanochip” as a Tandem Catalytic System for Chemiluminescent Detection of Sweat Glucose. Anal. Chem.93, 10593–10600. 10.1021/acs.analchem.1c01798
77
GargN.SoniK.SaxenaT. K.MajiS. (2015). Applications of Autoregressive Integrated Moving Average (Arima) Approach in Time-Series Prediction of Traffic Noise Pollution. noise Cont. engng J.63, 182–194. 10.3397/1/376317
78
Geelhoed-DuijvestijnP.VegelyteD.KownackaA.AntonN.JoosseM.WilsonC. (2021). Performance of the Prototype Noviosense Noninvasive Biosensor for Tear Glucose in Type 1 Diabetes. J. Diabetes Sci. Technol.15, 1320–1325. 10.1177/1932296820964844
79
Gómez-CarmonaO.Casado-MansillaD.KraemerF. A.de IpiñaD. L.García-ZubiaJ. (2020). Exploring the Computational Cost of Machine Learning at the Edge for Human-Centric Internet of Things. Future Gener. Comput. Syst.112, 670–683.
80
GuW.ZhouY.ZhouZ.LiuX.ZouH.ZhangP.et al (2017). SugarMate. Proc. ACM Interact. Mob. Wearable Ubiquitous Technol.1, 1–27. 10.1145/3130919
81
GüemesM.RahmanS. A.HussainK. (2015). What Is a Normal Blood Glucose?Archives Dis. Child.101, 569–574.
82
GulerS. D.GannonM.SicchioK. (2016). Crafting Wearables: Blending Technology with Fashion. New York, NY, USA: Apress.
83
HabibullahM.OnindaM. A. M.BaharA. N.DinhA.WahidK. A. (2019). “Nir-spectroscopic Classification of Blood Glucose Level Using Machine Learning Approach,” in 2019 IEEE Canadian Conference of Electrical and Computer Engineering (CCECE) (Edmonton, AB, Canada: IEEE). 10.1109/ccece.2019.8861843
84
HamdiT.Ben AliJ.Di CostanzoV.FnaiechF.MoreauE.GinouxJ.-M. (2018). Accurate Prediction of Continuous Blood Glucose Based on Support Vector Regression and Differential Evolution Algorithm. Biocybern. Biomed. Eng.38, 362–372. 10.1016/j.bbe.2018.02.005
85
HannaJ.BteichM.TawkY.RamadanA. H.DiaB.AsadallahF. A.et al (2020). Noninvasive, Wearable, and Tunable Electromagnetic Multisensing System for Continuous Glucose Monitoring, Mimicking Vasculature Anatomy. Sci. Adv.6, eaba5320. 10.1126/sciadv.aba5320
86
HannaJ.CostantineJ.KanjR.TawkY.RamadanA.EidA. (2021). “A Vasculature Anatomy Inspired Flexible Slot Antenna for Continuous Non-invasive Glucose Monitoring,” in 2021 IEEE International Symposium on Antennas and Propagation and USNC-URSI Radio Science Meeting (APS/URSI) (Singapore: IEEE), 803–804. 10.1109/aps/ursi47566.2021.9704287
87
HasegawaK.WatanabeH.HisamatsuT.OhnoS.ItohH.AshiharaT.et al (2016). Early Repolarization and Risk of Arrhythmia Events in Long QT Syndrome. Int. J. Cardiol.223, 540–542. 10.1016/j.ijcard.2016.08.215
88
HastieT.TibshiraniR.FriedmanJ. (2009). The Elements of Statistical Learning: Data Mining, Inference, and Prediction. Berlin, Germany: Springer Science & Business Media.
89
HearstM. A.DumaisS. T.OsunaE.PlattJ.ScholkopfB. (1998). Support Vector Machines. IEEE Intell. Syst. Their Appl.13, 18–28. 10.1109/5254.708428
90
HeinemannL. (2008). Finger Pricking and Pain: A Never Ending Story. J. Diabetes Sci. Technol.2, 919–921. 10.1177/193229680800200526
91
HischeM.Luis-DominguezO.PfeifferA. F. H.SchwarzP. E.SelbigJ.SprangerJ. (2010). Decision Trees as a Simple-To-Use and Reliable Tool to Identify Individuals with Impaired Glucose Metabolism or Type 2 Diabetes Mellitus. Eur. J. Endocrinol.163, 565–571. 10.1530/eje-10-0649
92
HochreiterS.SchmidhuberJ. (1997). Long Short-Term Memory. Neural Comput.9, 1735–1780. 10.1162/neco.1997.9.8.1735
93
HojaijiH.ZhaoY.GongM. C.MallajosyulaM.TanJ.LinH.et al (2020). An Autonomous Wearable System for Diurnal Sweat Biomarker Data Acquisition. Lab a Chip. 10.1039/d0lc00820f
94
HongY. J.LeeH.KimJ.LeeM.ChoiH. J.HyeonT.et al (2018). Multifunctional Wearable System that Integrates Sweat-Based Sensing and Vital-Sign Monitoring to Estimate Pre-/post-exercise Glucose Levels. Adv. Funct. Mat.28, 1805754. 10.1002/adfm.201805754
95
HotmartuaR.PangestuP. W.ZakariaH.IrawanY. S. (2015). “Noninvasive Blood Glucose Detection Using Near Infrared Sensor,” in 2015 International Conference on Electrical Engineering and Informatics (ICEEI) (Denpasar, Indonesia: IEEE). 10.1109/iceei.2015.7352586
96
HuangS.CaiN.PachecoP. P.NarrandesS.WangY.XuW. (2018). Applications of Support Vector Machine (Svm) Learning in Cancer Genomics. Cancer Genomics Proteomics15, 41–51. 10.21873/cgp.20063
97
HumphreyS. P.WilliamsonR. T. (2001). A Review of Saliva: Normal Composition, Flow, and Function. J. Prosthet. Dent.85, 162–169. 10.1067/mpr.2001.113778
98
IvanciucO. (2007). Applications of Support Vector Machines in Chemistry. Rev. Comput. Chem.23, 291–400. 10.1002/9780470116449.ch6
99
JaggersJ. R.KingK. M.WatsonS. E.WintergerstK. A. (2019). Predicting Nocturnal Hypoglycemia with Measures of Physical Activity Intensity in Adolescent Athletes with Type 1 Diabetes. Diabetes Technol. Ther.21, 406–408. 10.1089/dia.2019.0048
100
JanidarmianM.Roshan FekrA.RadeckaK.ZilicZ. (2017). A Comprehensive Analysis on Wearable Acceleration Sensors in Human Activity Recognition. Sensors17, 529. 10.3390/s17030529
101
JiangF.LiS.YuY.ChengQ. S.KozielS. (2017). “Sensitivity Optimization of Antenna for Non-invasive Blood Glucose Monitoring,” in 2017 International Applied Computational Electromagnetics Society Symposium (ACES) (Suzhou, China: IEEE), 1–2.
102
JoseP. S. H.RajasekaranK.RajalakshmyP.JebastinaB. (2019). “A Non-invasive Method for Measurement of Blood Glucose Using Bio Impedance Technique,” in 2019 2nd International Conference on Signal Processing and Communication (ICSPC) (Coimbatore, India: IEEE). 10.1109/icspc46172.2019.8976732
103
JunM. H.KimS.KuB.ChoJ.KimK.YooH. R.et al (2018). Glucose-independent Segmental Phase Angles from Multi-Frequency Bioimpedance Analysis to Discriminate Diabetes Mellitus. Sci. Rep.8, 648. 10.1038/s41598-017-18913-7
104
JungD.JungD.KongS. (2017). A Lab-On-A-Chip-Based Non-invasive Optical Sensor for Measuring Glucose in Saliva. Sensors17, 2607. 10.3390/s17112607
105
JurystaC.BulurN.OguzhanB.SatmanI.YilmazT. M.MalaisseW. J.et al (2009). Salivary Glucose Concentration and Excretion in Normal and Diabetic Subjects. J. Biomed. Biotechnol.2009, 1–6. 10.1155/2009/430426
106
KališnikJ. M.AvbeljV.VratanarJ.SantarpinoG.GeršakB.FischleinT.et al (2019). Cardiac Autonomic Regulation and PR Interval Determination for Enhanced Atrial Fibrillation Risk Prediction after Cardiac Surgery. Int. J. Cardiol.289, 24–29. 10.1016/j.ijcard.2019.04.070
107
KamalrajR.NeelakandanS.Ranjith KumarM.Chandra Shekhar RaoV.AnandR.SinghH. (2021). Interpretable Filter Based Convolutional Neural Network (If-cnn) for Glucose Prediction and Classification Using Pd-Ss Algorithm. Measurement183, 109804. 10.1016/j.measurement.2021.109804
108
KamathA.MahalingamA.BraukerJ. (2009). Analysis of Time Lags and Other Sources of Error of the DexCom SEVEN Continuous Glucose Monitor. Diabetes Technol. Ther.11, 689–695. 10.1089/dia.2009.0060
109
KarpovaE. V.ShcherbachevaE. V.GalushinA. A.VokhmyaninaD. V.KaryakinaE. E.KaryakinA. A. (2019). Noninvasive Diabetes Monitoring through Continuous Analysis of Sweat Using Flow-Through Glucose Biosensor. Anal. Chem.91, 3778–3783. 10.1021/acs.analchem.8b05928
110
KatseliV.EconomouA.KokkinosC. (2021). Smartphone-addressable 3d-Printed Electrochemical Ring for Nonenzymatic Self-Monitoring of Glucose in Human Sweat. Anal. Chem.93, 3331–3336. 10.1021/acs.analchem.0c05057
111
KekadeS.HseiehC.-H.IslamM. M.AtiqueS.Mohammed KhalfanA.LiY.-C.et al (2018). The Usefulness and Actual Use of Wearable Devices Among the Elderly Population. Comput. Methods Programs Biomed.153, 137–159. 10.1016/j.cmpb.2017.10.008
112
KeshawarzA.PiropatoA. R.BrownT. L.DucaL. M.SipplR. M.WadwaR. P.et al (2018). Lower Objectively Measured Physical Activity Is Linked with Perceived Risk of Hypoglycemia in Type 1 Diabetes. J. Diabetes its Complicat.32, 975–981. 10.1016/j.jdiacomp.2018.05.020
113
KimJ.KimM.LeeM. S.KimK.JiS.KimY. T.et al (2017). Wearable Smart Sensor Systems Integrated on Soft Contact Lenses for Wireless Ocular Diagnostics. Nat. Commun.8, 14997. 10.1038/ncomms14997
114
KimJ.CampbellA. S.WangJ. (2018). Wearable Non-invasive Epidermal Glucose Sensors: A Review. Talanta177, 163–170. 10.1016/j.talanta.2017.08.077
115
KlimontovV. V.MyakinaN. E.TyanN. V. (2016). Heart Rate Variability Is Associated with Interstitial Glucose Fluctuations in Type 2 Diabetic Women Treated with Insulin. SpringerPlus5, 337. 10.1186/s40064-016-1932-z
116
KownackaA. E.VegelyteD.JoosseM.AntonN.ToebesB. J.LaukoJ.et al (2018). Clinical Evidence for Use of a Noninvasive Biosensor for Tear Glucose as an Alternative to Painful Finger-Prick for Diabetes Management Utilizing a Biopolymer Coating. Biomacromolecules19, 4504–4511. 10.1021/acs.biomac.8b01429
117
KumarS. V.VanajakshiL. (2015). Short-term Traffic Flow Prediction Using Seasonal Arima Model with Limited Input Data. Eur. Transp. Res. Rev.7, 21. 10.1007/s12544-015-0170-8
118
La BelleJ. T.AdamsA.LinC.-E.EngelschallE.PrattB.CookC. B. (2016). Self-monitoring of Tear Glucose: the Development of a Tear Based Glucose Sensor as an Alternative to Self-Monitoring of Blood Glucose. Chem. Commun.52, 9197–9204. 10.1039/c6cc03609k
119
LaiS.MuscaritoliM.AndreozziP.SgrecciaA.De LeoS.MazzaferroS.et al (2019). Sarcopenia and Cardiovascular Risk Indices in Patients with Chronic Kidney Disease on Conservative and Replacement Therapy. Nutrition62, 108–114. 10.1016/j.nut.2018.12.005
120
LeCunY.BengioY.HintonG. (2015). Deep Learning. nature521, 436–444. 10.1038/nature14539
121
LeeH.SongC.HongY. S.KimM. S.ChoH. R.KangT.et al (2017). Wearable/disposable Sweat-Based Glucose Monitoring Device with Multistage Transdermal Drug Delivery Module. Sci. Adv.3, e1601314. 10.1126/sciadv.1601314
122
LeeI.ProbstD.KlonoffD.SodeK. (2021a). Continuous Glucose Monitoring Systems - Current Status and Future Perspectives of the Flagship Technologies in Biosensor Research -. Biosens. Bioelectron.181, 113054. 10.1016/j.bios.2021.113054
123
LeeW.-C.KohE. H.KimD.-H.ParkS.-G.JungH. S. (2021b). Plasmonic Contact Lens Materials for Glucose Sensing in Human Tears. Sensors Actuators B Chem.344, 130297. 10.1016/j.snb.2021.130297
124
LeesT.Shad-KaneezF.SimpsonA. M.NassifN. T.LinY.LalS. (2018). Heart Rate Variability as a Biomarker for Predicting Stroke, Post-stroke Complications and Functionality. Biomark. Insights13, 117727191878693. 10.1177/1177271918786931
125
LekhaS.SuchethaM. (2020). Recent Advancements and Future Prospects on E-Nose Sensors Technology and Machine Learning Approaches for Non- Invasive Diabetes Diagnosis: A Review. IEEE Rev. Biomed. Eng.14, 127–138.
126
LeskoJ.SeibertS.ZhuY. (2018). “Design and Validation of a Low-Cost Non-invasive Device to Detect Overnight Hypoglycemia,” in Volume 3: Biomedical and Biotechnology Engineering (New York, NY, USA: American Society of Mechanical Engineers (ASME)). 10.1115/imece2018-86009
127
LiJ.FernandoC. (2016). Smartphone-based Personalized Blood Glucose Prediction. ICT Express2, 150–154. 10.1016/j.icte.2016.10.001
128
LiJ.IgbeT.LiuY.NieZ.QinW.WangL.et al (2018). An Approach for Noninvasive Blood Glucose Monitoring Based on Bioimpedance Difference Considering Blood Volume Pulsation. IEEE Access6, 51119–51129. 10.1109/access.2018.2866601
129
LiJ.ToboreI.LiuY.KandwalA.WangL.NieZ. (2021). Non-invasive Monitoring of Three Glucose Ranges Based on Ecg by Using Dbscan-Cnn. IEEE J. Biomed. Health Inf.25, 3340–3350. 10.1109/jbhi.2021.3072628
130
LinC.PrattB.HonikelM.JenishA.RameshB.AlkhanA.et al (2017). Toward the Development of a Glucose Dehydrogenase-Based Saliva Glucose Sensor without the Need for Sample Preparation. J. Diabetes Sci. Technol.12, 83–89. 10.1177/1932296817712526
131
LingS. H.SanP. P.NguyenH. T. (2016). Non-invasive Hypoglycemia Monitoring System Using Extreme Learning Machine for Type 1 Diabetes. ISA Trans.64, 440–446. 10.1016/j.isatra.2016.05.008
132
LiuJ.WangL.ZhangL.ZhangZ.ZhangS. (2020). Predictive Analytics for Blood Glucose Concentration: an Empirical Study Using the Tree-Based Ensemble Approach. Lht38, 835–858. 10.1108/lht-08-2019-0171
133
LiuY.XiaM.NieZ.LiJ.ZengY.WangL. (2016). “In Vivo wearable Non-invasive Glucose Monitoring Based on Dielectric Spectroscopy,” in 2016 IEEE 13th International Conference on Signal Processing (ICSP) (Chengdu, China: IEEE). 10.1109/icsp.2016.7878054
134
LoniniL.DaiA.ShawenN.SimuniT.PoonC.ShimanovichL.et al (2018). Wearable Sensors for Parkinson's Disease: Which Data Are Worth Collecting for Training Symptom Detection Models. NPJ Digit. Med.1, 64–68. 10.1038/s41746-018-0071-z
135
LuY.JiangK.ChenD.ShenG. (2019). Wearable Sweat Monitoring System with Integrated Micro-supercapacitors. Nano Energy58, 624–632. 10.1016/j.nanoen.2019.01.084
136
MalikS.KhadgawatR.AnandS.GuptaS. (2016). Non-invasive Detection of Fasting Blood Glucose Level via Electrochemical Measurement of Saliva. SpringerPlus5, 701. 10.1186/s40064-016-2339-6
137
ManoT.NagamineK.IchimuraY.ShiwakuR.FurusawaH.MatsuiH.et al (2018). Printed Organic Transistor‐Based Enzyme Sensor for Continuous Glucose Monitoring in Wearable Healthcare Applications. ChemElectroChem5, 3881–3886. 10.1002/celc.201801129
138
MarcusY.EldorR.YaronM.ShaklaiS.Ish-ShalomM.SheferG.et al (2020). Improving Blood Glucose Level Predictability Using Machine Learning. Diabetes Metab. Res. Rev.36, e3348. 10.1002/dmrr.3348
139
MaritschM.FöllS.LehmannV.BérubéC.KrausM.FeuerriegelS.et al (2020). “Towards Wearable-Based Hypoglycemia Detection and Warning in Diabetes,” in Extended Abstracts of the 2020 CHI Conference on Human Factors in Computing Systems (ACM). 10.1145/3334480.3382808
140
MarlingC.XiaL.BunescuR.SchwartzF. (2016). “Machine Learning Experiments with Noninvasive Sensors for Hypoglycemia Detection,” in Proceedings of IJCAI Workshop on Knowledge Discovery in Healthcare Data (San Francisco, CA, USA: Morgan Kaufmann Publishers Inc.), 1–6.
141
MartinssonE. (2017). WTTE-RNN: Weibull Time to Event Recurrent Neural Network A Model for Sequential Prediction of Time-To-Event in the Case of Discrete or Continuous Censored Data, Recurrent Events or Time-Varying Covariates. Master’s thesis.
142
MartinssonJ.SchliepA.EliassonB.MeijnerC.PerssonS.MogrenO. (2018). “Automaticblood Glucose Prediction with Confidence Using Recurrent Neural Networks,” in KHD@IJCAI.
143
MattssonM.-O.SimkóM. (2019). Emerging Medical Applications Based on Non-ionizing Electromagnetic Fields from 0 Hz to 10 THz. Med. Devices Evid. Res.12, 347–368. 10.2147/mder.s214152
144
MayoM.ChepulisL.PaulR. G. (2019). Glycemic-aware Metrics and Oversampling Techniques for Predicting Blood Glucose Levels Using Machine Learning. PLOS ONE14, e0225613. 10.1371/journal.pone.0225613
145
MedskerL.JainL. C. (1999). Recurrent Neural Networks: Design and Applications. Boca Raton, FL, USA: CRC Press.
146
MercanteL. A.AndreR. S.FacureM. H. M.Fugikawa-SantosL.CorreaD. S. (2021). Design of a Bioelectronic Tongue for Glucose Monitoring Using Zinc Oxide Nanofibers and Graphene Derivatives. Sensors Actuators Rep.3, 100050. 10.1016/j.snr.2021.100050
147
MhaskarH. N.PereverzyevS. V.van der WaltM. D. (2017). A Deep Learning Approach to Diabetic Blood Glucose Prediction. Front. Appl. Math. Stat.3, 14. 10.3389/fams.2017.00014
148
MillerC. D.PhillipsL. S.ZiemerD. C.GallinaD. L.CookC. B.El-KebbiI. M. (2001). Hypoglycemia in Patients with Type 2 Diabetes Mellitus. Arch. Intern Med.161, 1653. 10.1001/archinte.161.13.1653
149
MillerC. S.FoleyJ. D.BaileyA. L.CampellC. L.HumphriesR. L.ChristodoulidesN.et al (2010). Current Developments in Salivary Diagnostics. Biomarkers Med.4, 171–189. 10.2217/bmm.09.68
150
MirshekarianS.BunescuR.MarlingC.SchwartzF. (2017). Using LSTMs to Learn Physiological Models of Blood Glucose Behavior. Annu. Int. Conf. IEEE Eng. Med. Biol. Soc.2017, 2887. 10.1109/embc.2017.8037460
151
MoyerJ.WilsonD.FinkelshteinI.WongB.PottsR. (2012). Correlation between Sweat Glucose and Blood Glucose in Subjects with Diabetes. Diabetes Technol. Ther.14, 398–402. 10.1089/dia.2011.0262
152
MylesA. J.FeudaleR. N.LiuY.WoodyN. A.BrownS. D. (2004). An Introduction to Decision Tree Modeling. J. Chemom.18, 275–285. 10.1002/cem.873
153
NanayakkaraN. D.MunasinghaS. C.RuwanpathiranaG. P. (2018). “Non-invasive Blood Glucose Monitoring Using a Hybrid Technique,” in 2018 Moratuwa Engineering Research Conference (MERCon) (Moratuwa, Sri Lanka: IEEE). 10.1109/mercon.2018.8421885
154
Naranjo-HernándezD.Reina-TosinaJ.MinM. (2019). Fundamentals, Recent Advances, and Future Challenges in Bioimpedance Devices for Healthcare Applications. J. Sensors2019, 1–42.
155
NasserA. R.HasanA. M.HumaidiA. J.AlkhayyatA.AlzubaidiL.FadhelM. A.et al (2021). Iot and Cloud Computing in Health-Care: A New Wearable Device and Cloud-Based Deep Learning Algorithm for Monitoring of Diabetes. Electronics10, 2719. 10.3390/electronics10212719
156
NathansonD.SvenssonA.-M.MiftarajM.FranzénS.BolinderJ.Eeg-OlofssonK. (2021). Effect of Flash Glucose Monitoring in Adults with Type 1 Diabetes: a Nationwide, Longitudinal Observational Study of 14, 372 Flash Users Compared with 7691 Glucose Sensor Naive Controls. Diabetologia64, 1595. 10.1007/s00125-021-05437-z
157
NavadaA.AnsariA. N.PatilS.SonkambleB. A. (2011). “Overview of Use of Decision Tree Algorithms in Machine Learning,” in 2011 IEEE control and system graduate research colloquium (Shah Alam, Malaysia: IEEE), 37–42. 10.1109/icsgrc.2011.5991826
158
NawazA.ØhlckersP.SælidS.JacobsenM.Nadeem AkramM. (2016). Review: Non-invasive Continuous Blood Glucose Measurement Techniques. J. Bioinforma. Diabetes1, 1–27. 10.14302/issn.2374-9431.jbd-15-647
159
NikolaidouT.PellicoriP.ZhangJ.KazmiS.GoodeK. M.ClelandJ. G.et al (2017). Prevalence, Predictors, and Prognostic Implications of PR Interval Prolongation in Patients with Heart Failure. Clin. Res. Cardiol.107, 108–119. 10.1007/s00392-017-1162-6
160
NunesL. A. S.MussaviraS.BindhuO. S. (2015). Clinical and Diagnostic Utility of Saliva as a Non-invasive Diagnostic Fluid: a Systematic Review. Biochem. Med.25, 177–192. 10.11613/bm.2015.018
161
OhkumaT.ChalmersJ.CooperM.HametP.HarrapS.MarreM.et al (2021). The Comparative Effects of Intensive Glucose Lowering in Diabetes Patients Aged below or above 65 years: Results from the ADVANCE Trial. Diabetes, Obes. Metabolism23, 1292. 10.1111/dom.14339
162
ParkJ.KimJ.KimS. Y.CheongW. H.JangJ.ParkY. G.et al (2018). Soft, Smart Contact Lenses with Integrations of Wireless Circuits, Glucose Sensors, and Displays. Sci. Adv.4, eaap9841. 10.1126/sciadv.aap9841
163
ParkS.BooH.ChungT. D. (2006). Electrochemical Non-enzymatic Glucose Sensors. Anal. Chim. Acta556, 46–57. 10.1016/j.aca.2005.05.080
164
PatleA.ChouhanD. S. (2013). “SVM Kernel Functions for Classification,” in 2013 International Conference on Advances in Technology and Engineering (ICATE) (Mumbai, India: IEEE). 10.1109/icadte.2013.6524743
165
PienaarS. W.MalekianR. (2019). “Human Activity Recognition Using LSTM-RNN Deep Neural Network Architecture,” in 2019 IEEE 2nd Wireless Africa Conference (WAC) (Pretoria, South Africa: IEEE). 10.1109/africa.2019.8843403
166
PorumbM.StrangesS.PescapèA.PecchiaL. (2020b). Precision Medicine and Artificial Intelligence: A Pilot Study on Deep Learning for Hypoglycemic Events Detection Based on ECG. Sci. Rep.10, 170. 10.1038/s41598-019-56927-5
167
PorumbM.GriffenC.HattersleyJ.PecchiaL. (2020a). Nocturnal Low Glucose Detection in Healthy Elderly from One-Lead ECG Using Convolutional Denoising Autoencoders. Biomed. Signal Process. Control62, 102054. 10.1016/j.bspc.2020.102054
168
PrendinF.Del FaveroS.VettorettiM.SparacinoG.FacchinettiA. (2021). Forecasting of Glucose Levels and Hypoglycemic Events: Head-To-Head Comparison of Linear and Nonlinear Data-Driven Algorithms Based on Continuous Glucose Monitoring Data Only. Sensors21, 1647. 10.3390/s21051647
169
PuschmannA.WszolekZ. (2011). Diagnosis and Treatment of Common Forms of Tremor. Semin. Neurol.31, 065–077. 10.1055/s-0031-1271312
170
RajS.TripathiS.UpadhyayG.TripathiS. S.TripathiV. S. (2021). An Electromagnetic Band Gap-Based Complementary Split Ring Resonator Loaded Patch Antenna for Glucose Level Measurement. IEEE Sensors J.21, 22679–22687. 10.1109/jsen.2021.3107462
171
Rajendra AcharyaU.Paul JosephK.KannathalN.LimC. M.SuriJ. S. (2006). Heart Rate Variability: a Review. Med. Bio Eng. Comput.44, 1031–1051. 10.1007/s11517-006-0119-0
172
RamezankhaniA.HadavandiE.PournikO.ShahrabiJ.AziziF.HadaeghF. (2016). Decision Tree-Based Modelling for Identification of Potential Interactions between Type 2 Diabetes Risk Factors: a Decade Follow-Up in a Middle East Prospective Cohort Study. BMJ Open6, e013336. 10.1136/bmjopen-2016-013336
173
RanaA. Q.ChouK. L. (2015). Essential Tremor in Clinical Practice. Berlin, Germany: Springer.
174
ReddyR.ResalatN.WilsonL. M.CastleJ. R.El YoussefJ.JacobsP. G. (2019). Prediction of Hypoglycemia during Aerobic Exercise in Adults with Type 1 Diabetes. J. Diabetes Sci. Technol.13, 919–927. 10.1177/1932296818823792
175
RentkaA.KoroskenyiK.HarsfalviJ.SzekaneczZ.SzucsG.SzodorayP.et al (2017). Evaluation of Commonly Used Tear Sampling Methods and Their Relevance in Subsequent Biochemical Analysis. Ann. Clin. Biochem.54, 521–529. 10.1177/0004563217695843
176
RobsonN.BondA.WolffK. (2010). Salivary Nicotine and Cotinine Concentrations in Unstimulated and Stimulated Saliva. Afr. J. Pharm. Pharmacol.4, 061–065.
177
Rodríguez-RodríguezI.ChatzigiannakisI.RodríguezJ.-V.MaranghiM.GentiliM.Zamora-IzquierdoM.-Á. (2019). Utility of Big Data in Predicting Short-Term Blood Glucose Levels in Type 1 Diabetes Mellitus through Machine Learning Techniques. Sensors19, 4482.
178
RomeoA.MoyaA.LeungT. S.GabrielG.VillaR.SánchezS. (2018). Inkjet Printed Flexible Non-enzymatic Glucose Sensor for Tear Fluid Analysis. Appl. Mater. Today10, 133–141. 10.1016/j.apmt.2017.12.016
179
Ruiz-VargasA.ArkwrightJ.IvorraA. (2016). “A Portable Bioimpedance Measurement System Based on Red Pitaya for Monitoring and Detecting Abnormalities in the Gastrointestinal Tract,” in 2016 IEEE EMBS Conference on Biomedical Engineering and Sciences (IECBES) (Kuala Lumpur, Malaysia: IEEE), 150–154. 10.1109/iecbes.2016.7843433
180
SahaS.Cano-GarciaH.SotiriouI.LipscombeO.GouzouasisI.KoutsoupidouM.et al (2017). A Glucose Sensing System Based on Transmission Measurements at Millimetre Waves Using Micro Strip Patch Antennas. Sci. Rep.7, 6855. 10.1038/s41598-017-06926-1
181
SamirM. A.MohamedZ. A.HusseinM. A. A.AtiaA. (2021). “Applying Deep Learning to Track Food Consumption and Human Activity for Non-intrusive Blood Glucose Monitoring,” in 2021 IEEE 12th Annual Ubiquitous Computing, Electronics & Mobile Communication Conference (UEMCON) (New York, NY, USA: IEEE), 0319–0324. 10.1109/uemcon53757.2021.9666662
182
San-SegundoR.ZhangA.CebullaA.PanevS.TaborG.StebbinsK.et al (2020). Parkinson's Disease Tremor Detection in the Wild Using Wearable Accelerometers. Sensors20, 5817. 10.3390/s20205817
183
SanaiF.SahidA. S.HuvanandanaJ.SpoaS.BoyleL. H.HribarJ.et al (2021). Evaluation of a Continuous Blood Glucose Monitor: A Novel and Non-invasive Wearable Using Bioimpedance Technology. J. Diabetes Sci. Technol.2021, 19322968211054110. 10.1177/19322968211054110
184
SankhalaD.SardesaiA. U.PaliM.LinK. C.JagannathB.MuthukumarS.et al (2022). A Machine Learning-Based On-Demand Sweat Glucose Reporting Platform. Sci. Rep.12, 2442. 10.1038/s41598-022-06434-x
185
SatishS.SenK.AnandS. (2017). Impedance Spectroscopy of Aqueous Solution Samples of Different Glucose Concentrations for the Exploration of Non-invasive-continuous-blood-glucose-monitoring. MAPAN33, 185–190. 10.1007/s12647-017-0242-4
186
SchneyerL. H.YoungJ. A.SchneyerC. A. (1972). Salivary Secretion of Electrolytes. Physiol. Rev.52, 720–777. 10.1152/physrev.1972.52.3.720
187
SempionattoJ. R.MoonJ.-M.WangJ. (2021). Touch-based Fingertip Blood-free Reliable Glucose Monitoring: Personalized Data Processing for Predicting Blood Glucose Concentrations. ACS Sens.6, 1875–1883. 10.1021/acssensors.1c00139
188
SeneviratneS.HuY.NguyenT.LanG.KhalifaS.ThilakarathnaK.et al (2017). A Survey of Wearable Devices and Challenges. IEEE Commun. Surv. Tutorials19, 2573–2620. 10.1109/comst.2017.2731979
189
SeoW.ParkS.-W.KimN.JinS.-M.ParkS.-M. (2021). A Personalized Blood Glucose Level Prediction Model with a Fine-Tuning Strategy: A Proof-Of-Concept Study. Comput. Methods Programs Biomed.211, 106424. 10.1016/j.cmpb.2021.106424
190
SessaF.AnnaV.MessinaG.CibelliG.MondaV.MarsalaG.et al (2018). Heart Rate Variability as Predictive Factor for Sudden Cardiac Death. Aging10, 166–177. 10.18632/aging.101386
191
ShahidS.HussainS.KhanW. A. (2021). “Predicting Continuous Blood Glucose Level Using Deep Learning,” in Proceedings of the 14th IEEE/ACM International Conference on Utility and Cloud Computing Companion, 1–5. 10.1145/3492323.3495598
192
ShakerG.SmithK.OmerA. E.LiuS.CsechC.WadhwaU.et al (2018). Non-invasive Monitoring of Glucose Level Changes Utilizing a Mm-Wave Radar System. Int. J. Mob. Hum. Comput. Interact.10, 10–29. 10.4018/ijmhci.2018070102
193
ShendeP.KastureP. (2021). Bioconjugated Nanoflower for Estimation of Glucose from Saliva Using Nanozymes. Mater. Technol.2021, 1–7. 10.1080/10667857.2021.1898717
194
ShokrekhodaeiM.CistolaD. P.RobertsR. C.QuinonesS. (2021). Non-invasive Glucose Monitoring Using Optical Sensor and Machine Learning Techniques for Diabetes Applications. IEEE Access9, 73029–73045. 10.1109/access.2021.3079182
195
SiddiquiS. A.ZhangY.LloretJ.SongH.ObradovicZ. (2018). Pain-free Blood Glucose Monitoring Using Wearable Sensors: Recent Advancements and Future Prospects. IEEE Rev. Biomed. Eng.11, 21–35. 10.1109/rbme.2018.2822301
196
SilvaT. P.RolimL. C.Sallum FilhoC.ZimmermannL. M.MalerbiF.DibS. A. (2016). Association between Severity of Hypoglycemia and Loss of Heart Rate Variability in Patients with Type 1 Diabetes Mellitus. Diabetes Metab. Res. Rev.33, e2830. 10.1002/dmrr.2830
197
SinghP.GuptaP.JyotiK. (2019). Tasm: Technocrat Arima and Svr Model for Workload Prediction of Web Applications in Cloud. Clust. Comput.22, 619–633. 10.1007/s10586-018-2868-6
198
SmolaA. J.SchölkopfB. (2004). A Tutorial on Support Vector Regression. Statistics Comput.14, 199–222. 10.1023/b:stco.0000035301.49549.88
199
SnekhalathaU.RajalakshmiT.Vinitha SriC. H.BalachanderG.ShankarK. S. (2018). Non-invasive Blood Glucose Analysis Based on Galvanic Skin Response for Diabetic Patients. Biomed. Eng. Appl. Basis Commun.30, 1850009. 10.4015/s1016237218500096
200
StenersonM.CameronF.WilsonD. M.HarrisB.PayneS.BequetteB. W.et al (2014). The Impact of Accelerometer and Heart Rate Data on Hypoglycemia Mitigation in Type 1 Diabetes. J. Diabetes Sci. Technol.8, 64–69. 10.1177/1932296813516208
201
StrakosasX.SelbergJ.PansodteeP.YonasN.ManapongpunP.TeodorescuM.et al (2019). A Non-enzymatic Glucose Sensor Enabled by Bioelectronic pH Control. Sci. Rep.9, 10844. 10.1038/s41598-019-46302-9
202
SunQ.JankovicM. V.BallyL.MougiakakouS. G. (2018). “Predicting Blood Glucose with an LSTM and Bi-LSTM Based Deep Neural Network,” in 2018 14th Symposium on Neural Networks and Applications (NEUREL) (Belgrade, Serbia: IEEE). 10.1109/neurel.2018.8586990
203
SusanaE.RamliK.MurfiH.ApriantoroN. H. (2022). Non-invasive Classification of Blood Glucose Level for Early Detection Diabetes Based on Photoplethysmography Signal. Information13, 59. 10.3390/info13020059
204
SwapnaS.SomanK.VinayakumarR. (2018). Automated Detection of Diabetes Using CNN and CNN-LSTM Network and Heart Rate Signals. Procedia Comput. Sci.132, 1253–1262.
205
TakamatsuR.HiguchiK.MuramatsuD. (2021). “Measurement Frequency Evaluation for Bioimpedance-Based Blood-Glucose Estimation,” in 2021 IEEE 3rd Global Conference on Life Sciences and Technologies (LifeTech) (Nara, Japan: IEEE), 309–310. 10.1109/lifetech52111.2021.9391845
206
TaorminaC. R.BacaJ. T.AsherS. A.GrabowskiJ. J.FinegoldD. N. (2007). Analysis of Tear Glucose Concentration with Electrospray Ionization Mass Spectrometry. J. Am. Soc. Mass Spectrom.18, 332–336. 10.1016/j.jasms.2006.10.002
207
TehraniF.BavarianB. (2016). Facile and Scalable Disposable Sensor Based on Laser Engraved Graphene for Electrochemical Detection of Glucose. Sci. Rep.6, 27975. 10.1038/srep27975
208
ThaysenJ. H.ThornN. A.SchwartzI. L. (1954). Excretion of Sodium, Potassium, Chloride and Carbon Dioxide in Human Parotid Saliva. Am. J. Physiology-Legacy Content178, 155–159. 10.1152/ajplegacy.1954.178.1.155
209
TronstadC.ElvebakkO.StaalO. M.KalvøyH.HøgetveitJ. O.JenssenT. G.et al (2019). Non-invasive Prediction of Blood Glucose Trends during Hypoglycemia. Anal. Chim. Acta1052, 37–48. 10.1016/j.aca.2018.12.009
210
TsaiC.-W.LiC.-H.LamR. W.-K.LiC.-K.HoS. (2020). Diabetes Care in Motion: Blood Glucose Estimation Using Wearable Devices. IEEE Consum. Electron. Mag.9, 30–34. 10.1109/mce.2019.2941461
211
UpashamS.PrasadS. (2020). SLOCK (Sensor for Circadian Clock): Passive Sweat-Based Chronobiology Tracker. Lab. Chip20, 1947–1960. 10.1039/d0lc00134a
212
WangL.LinZ. Q.WongA. (2020). Covid-net: A Tailored Deep Convolutional Neural Network Design for Detection of Covid-19 Cases from Chest X-Ray Images. Sci. Rep.10, 19549. 10.1038/s41598-020-76550-z
213
WildD.von MaltzahnR.BrohanE.ChristensenT.ClausonP.Gonder-FrederickL. (2007). A Critical Review of the Literature on Fear of Hypoglycemia in Diabetes: Implications for Diabetes Management and Patient Education. Patient Educ. Couns.68, 10–15. 10.1016/j.pec.2007.05.003
214
XiaoG.HeJ.ChenX.QiaoY.WangF.XiaQ.et al (2019). A Wearable, Cotton Thread/paper-Based Microfluidic Device Coupled with Smartphone for Sweat Glucose Sensing. Cellulose26, 4553–4562. 10.1007/s10570-019-02396-y
215
XuanX.YoonH. S.ParkJ. Y. (2018). A Wearable Electrochemical Glucose Sensor Based on Simple and Low-Cost Fabrication Supported Micro-patterned Reduced Graphene Oxide Nanocomposite Electrode on Flexible Substrate. Biosens. Bioelectron.109, 75–82. 10.1016/j.bios.2018.02.054
216
YadavJ.RaniA.SinghV.MurariB. M. (2014). “Near-infrared LED Based Non-invasive Blood Glucose Sensor,” in 2014 International Conference on Signal Processing and Integrated Networks (SPIN) (Noida, India: IEEE). 10.1109/spin.2014.6777023
217
YangJ.LiL.ShiY.XieX. (2019). An ARIMA Model with Adaptive Orders for Predicting Blood Glucose Concentrations and Hypoglycemia. IEEE J. Biomed. Health Inf.23, 1251–1260. 10.1109/jbhi.2018.2840690
218
YaoH.ShumA. J.CowanM.LähdesmäkiI.ParvizB. A. (2011). A Contact Lens with Embedded Sensor for Monitoring Tear Glucose Level. Biosens. Bioelectron.26, 3290–3296. 10.1016/j.bios.2010.12.042
219
YeohE.ChoudharyP.NwokoloM.AyisS.AmielS. A. (2015). Interventions that Restore Awareness of Hypoglycemia in Adults with Type 1 Diabetes: A Systematic Review and Meta-Analysis. Diabetes Care38, 1592–1609. 10.2337/dc15-0102
220
YuanY.AliM. K.MathewsonK. J.SharmaK.FaiyazM.TanW.et al (2020). Associations between Colonic Motor Patterns and Autonomic Nervous System Activity Assessed by High-Resolution Manometry and Concurrent Heart Rate Variability. Front. Neurosci.13, 1447. 10.3389/fnins.2019.01447
221
ZahedK.SasangoharF.MehtaR.ErraguntlaM.LawleyM.QaraqeK. (2018). Investigating the Efficacy of Using Hand Tremors for Early Detection of Hypoglycemic Events: A Scoping Literature Review. Proc. Hum. Factors Ergonomics Soc. Annu. Meet.62, 1211–1215. 10.1177/1541931218621278
222
ZammittN. N.FrierB. M. (2005). Hypoglycemia in Type 2 Diabetes. Diabetes Care28, 2948–2961. 10.2337/diacare.28.12.2948
223
ZanonM.MuellerM.ZakharovP.TalaryM. S.DonathM.StahelW. A.et al (2017). First Experiences with a Wearable Multisensor Device in a Noninvasive Continuous Glucose Monitoring Study at Home, Part II: The Investigators' View. J. Diabetes Sci. Technol.12, 554–561. 10.1177/1932296817740591
224
ZapasnoyA. S.BelichenkoV. P.YakubovV. P.GorstA. V.MironchevA. S.KlokovA. V.et al (2021). Application of Broadband Microwave Near-Field Sensors for Glucose Monitoring in Biological Media. Appl. Sci.11, 1470. 10.3390/app11041470
225
ZhangJ.HodgeW.HutnickC.WangX. (2011). Noninvasive Diagnostic Devices for Diabetes through Measuring Tear Glucose. J. Diabetes Sci. Technol.5, 166–172. 10.1177/193229681100500123
226
ZhangR.LiuS.JinH.LuoY.ZhengZ.GaoF.et al (2019). Noninvasive Electromagnetic Wave Sensing of Glucose. Sensors19, 1151. 10.3390/s19051151
227
ZhangX.ZhengJ.ZhangL.LiuY.ChenG. P.WangL.et al (2018). Systemic Inflammation Mediates the Detrimental Effects of Obesity on Asthma Control. Allergy & Asthma Proc.39, 43. 10.2500/aap.2018.39.4096
228
ZhangY.ZhuJ. M.LiangY. B.ChenH. B.YinS. M.ChenZ. C. (2017). Non-invasive Blood Glucose Detection System Based on Conservation of Energy Method. Physiol. Meas.38, 325–342. 10.1088/1361-6579/aa50cf
229
ZhangZ.WangS.LiuG.HuD.YangB.DaiQ.et al (2021). Antifouling Hydrogel Film Based on a Sandwich Array for Salivary Glucose Monitoring. RSC Adv.11, 27561–27569. 10.1039/d1ra03517g
230
ZhaoJ.LinY.WuJ.NyeinH. Y. Y.BariyaM.TaiL.-C.et al (2019). A Fully Integrated and Self-Powered Smartwatch for Continuous Sweat Glucose Monitoring. ACS Sens.4, 1925–1933. 10.1021/acssensors.9b00891
231
ZhuF.AbbasS. R.BologaR.LevinN. W.KotankoP. (2020a). Effect of Decrease in Glucose Gradient on Change in Intraperitoneal Volume Using Segmental Bioimpedance during Pet. Nephrol. Dial. Transplant.35, 1183. 10.1093/ndt/gfaa142.p1183
232
ZhuT.LiK.ChenJ.HerreroP.GeorgiouP. (2020b). Dilated Recurrent Neural Networks for Glucose Forecasting in Type 1 Diabetes. J. Healthc. Inf. Res.4, 308–324. 10.1007/s41666-020-00068-2
233
ZhuT.LiK.HerreroP.ChenJ.GeorgiouP. (2018a). “A Deep Learning Algorithm for Personalized Blood Glucose Prediction,” in KHD@ IJCAI, 64–78.
234
ZhuX.JuY.ChenJ.LiuD.LiuH. (2018b). Nonenzymatic Wearable Sensor for Electrochemical Analysis of Perspiration Glucose. ACS Sens.3, 1135–1141. 10.1021/acssensors.8b00168
235
ZidaneM. A.RouaneA.HamoudaC.AmarH. (2021). Hyper-sensitive Microwave Sensor Based on Split Ring Resonator (Srr) for Glucose Measurement in Water. Sensors Actuators A Phys.321, 112601. 10.1016/j.sna.2021.112601
Summary
Keywords
diabetes mellitus, non-invasive wearables and sensors, hypoglycemia, machine learning, blood glucose management, deep learning, bodily fluids glucose
Citation
Alhaddad AY, Aly H, Gad H, Al-Ali A, Sadasivuni KK, Cabibihan J-J and Malik RA (2022) Sense and Learn: Recent Advances in Wearable Sensing and Machine Learning for Blood Glucose Monitoring and Trend-Detection. Front. Bioeng. Biotechnol. 10:876672. doi: 10.3389/fbioe.2022.876672
Received
15 February 2022
Accepted
12 April 2022
Published
12 May 2022
Volume
10 - 2022
Edited by
Bahareh Azimi, University of Pisa, Italy
Reviewed by
Roohollah Bagherzadeh, Amirkabir University of Technology, Iran
Mario Milazzo, Massachusetts Institute of Technology, United States
Updates

Check for updates
Copyright
© 2022 Alhaddad, Aly, Gad, Al-Ali, Sadasivuni, Cabibihan and Malik.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Rayaz A. Malik, ram2045@qatar-med.cornell.edu
This article was submitted to Biosensors and Biomolecular Electronics, a section of the journal Frontiers in Bioengineering and Biotechnology
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.