ORIGINAL RESEARCH article

Front. Bioeng. Biotechnol., 16 January 2026

Sec. Biomaterials

Volume 13 - 2025 | https://doi.org/10.3389/fbioe.2025.1636262

Finite element analysis of clear aligner with overhanging attachments and extended gingival coverage for interdental space closure

  • 1. Shanghai University of Medicine and Health Sciences Affiliated Zhoupu Hospital, Shanghai, China

  • 2. Intelligent inspection and diagnostics health service platform, Shanghai University of Medicine and Health Sciences, Shanghai, China

  • 3. School of Stomatology, Xuzhou Medical University, Xuzhou, China

  • 4. Department of Orthodontics, Shanghai Xuhui District Stomatological Hospital, Shanghai, China

Abstract

Objective:

Overhanging (OH) attachments were modified clear aligner (CA) attachments with an extended portion toward the root to apply force closer to the center of resistance and enable greater control over root movement, resembling power arms. This study investigated the biomechanical effects of OH attachment and partially gingival extension of CA trimline on canine movement during the closure of extraction space via finite element analysis (FEA).

Methods:

CBCT data of an adult with Angle Class I molar relationship and mild anterior crowding was applied for comparing the biomechanical effects of three attachment types (no attachment, vertical, OH) and four trimline designs (partially buccal/lingual gingival coverage). Periodontal ligament (PDL) hydrostatic stress, tooth displacement, rotational center position, and CA stress distribution were assessed via FEA.

Results:

OH attachment induced increased tooth displacement and PDL hydrostatic stress (95.5 kPa) compared to regular vertical attachment (53.1 kPa) in achieving root-controlled canine movement. OH attachment combined with a buccolingual gingival extension of CA trimline on 2–6 facilitated the most translational canine movement and lowest ratio of mesio-apical to disto-occlusal displacement (0.466, compared to 0.506 in group with no attachment and trimline extension), while simultaneously avoiding excessive aligner deformation and stress concentration.

Conclusion:

Overhanging attachment combined with partial gingival extension of CA trimline significantly enhanced the orthodontic force for premolar extraction cases involving space closure between canines and molars, as a more efficient and feasible design for canine bodily movement.

1 Introduction

Clear aligners (CA) had been increasingly being utilized in orthodontics owing to their comfort, aesthetic appeal, and ease of cleaning (Weir, 2017). Despite their popularity, challenges still remained in managing intricate tooth movements, including torque (Cheng et al., 2022a), rotation (Seo et al., 2021), and bodily movement (Zhu et al., 2022). Specifically, during the closure of extraction spaces, torque controlling was critical to the translation of tooth and prevention of undesired tipping and rotation (Cheng et al., 2022a; Seo et al., 2021; Zhu et al., 2022).

The application of attachments was a common approach to facilitate more effective control of tooth movement (Je et al., 2023; Demir, 2024). In CA orthodontics, the standard design for closing extraction spaces involved placing attachments on canines (3) and posterior teeth (5/6/7) (Cheng et al., 2022a; Jiang et al., 2020; Liu et al., 2021; Liu L. et al., 2022; Liu J. Q. et al., 2022; Cheng et al., 2022b; Wang et al., 2022) and other auxiliaries, such as traction (Pu et al., 2022) and power ridges (Hong et al., 2024). These designs adjusted torque and balanced stress to precisely optimize tooth movement (Liu J. Q. et al., 2022; Wang et al., 2022; Xia et al., 2022). Root control attachments were also designed to modulate torque, including asymmetrical attachments with opposite orientations (Gomez et al., 2015; Yokoi et al., 2019) and overhanging attachments (Hong et al., 2021). An overhanging (OH) attachment was a modified CA attachment with an extended portion toward the root to apply force closer to the center of resistance and enable greater control over root movement, resembling a power arm (Hong et al., 2021). First reported in 2021 (Hong et al., 2021), an OH attachment was designed to induce bodily movement of the incisor and close scattered diastema of anterior teeth. A latest study in 2025 (Hong et al., 2025) reported the repair of gingival recession via better controlling root movement of lower incisor using an OH attachment. So far, there had been limited researches on OH attachments and their application in other tooth positions.

The apical extension of OH attachment necessitated the gingival extension of CA for full coverage. Extension of the trimming edges of CA had been reported to enhance control over tooth movement and improve therapeutic outcomes (Elshazly et al., 2023; Elshazly et al., 2022; Elshazly et al., 2024a; Elshazly et al., 2024b), as summarized in a recently published systematic review (Nakornnoi et al., 2024). However, despite of the widely acknowledged advantages of extended trimline, the remodeling of gingival soft tissues did not synchronize with tooth movement, and a full gingiva-covering trimline design compromised the aesthetics and comfort of CA. As a result, such trimline extensions have not gained widespread popularity in clinical practice. Reducing the gingival coverage from full to partial might avoid compromising the aesthetics and comfort of CAs, but whether equivalent effectiveness could be achieved remained unknown. Existing studies had focused on integral extension of CA trimline, with a lack of comparative research on the effects of region-specific extension of CA. Furthermore, the combined effects and biomechanical behavior of OH attachments with gingival extension of CA in closing extraction spaces remained to be explored.

In this study, finite element analysis (FEA) was employed to explore the biomechanical effects of different attachment designs (vertical/OH) combined with different gingival extension designs (buccal/lingual covering different regions) during closing extraction spaces. The effects on tooth movement, PDL stress distribution, and appliance deformation were systematically analyzed and compared to improve the effectiveness of CA and minimize adverse effects during closing extraction spaces. This study could provide guidance for future design of CA to improve the orthodontic effectiveness.

2 Materials and methods

2.1 Acquisition of dentition

The use of patient imaging data in this research was in accordance with institutional ethical standards, following informed consent and approval by the Ethical Committee of the Shanghai Xuhui District Dental Center, Shanghai, China. CBCT data of a healthy adult’s dentition (with ANB angle of 4.8°, average growth pattern, Angle Class I molar relationship, and mild anterior crowding) with a slice thickness of 0.625 mm (GE Healthcare, Buckinghamshire, England) was imported into Mimics software (Version 18.0, Materialise, Leuven, Belgium) for 3 days reconstruction.

2.2 Design of CAs and attachments

As illustrated in Figure 1a, CA with a thickness of 0.5 mm was designed on the reconstructed dentition model with the distal movement of canines (3) by 0.2 mm to provide orthodontic force. Attachments were adhered in the middle of the buccal surfaces of canines (3) and second premolars (5), and the attachment designs were categorized into 3 types: vertical wedge-shaped attachment (V), overhanging attachment (OH), and no attachment (NA). As shown in Figure 1b, vertical wedge-shaped attachments were designed with the wedge surface oriented towards the occlusal direction; OH attachments were designed with 6 mm extension towards gingiva.

FIGURE 1

CA trimline designs were divided into 4 types: (A) regular CA with unextended scalloped trimline; (B) CA with scalloped trimline and 5 mm buccal extension for gingival coverage at teeth 3∼5; (C) CA with scalloped trimline and 5 mm buccal extension for gingival coverage at teeth 2∼6; (D) CA with scalloped trimline and 5 mm buccolingual extension for gingival coverage at teeth 2∼6.

In total, 12 groups of models comprising 3 attachment types and 4 CA trimline designs were created and analyzed in this study, as shown in Table 1 and Figure 1c.

TABLE 1

GroupTrimline designsAttachment types
A-NAregular CA with unextended scalloped trimlineno attachment
A-Vregular CA with unextended scalloped trimlinevertical wedge-shaped
A-OHregular CA with unextended scalloped trimlineoverhanging
B-NACA with scalloped trimline and 5 mm buccal extension at teeth 3∼5no attachment
B-VCA with scalloped trimline and 5 mm buccal extension at teeth 3∼5vertical wedge-shaped
B-OHCA with scalloped trimline and 5 mm buccal extension at teeth 3∼5overhanging
C-NACA with scalloped trimline and 5 mm buccal extension at teeth 2∼6no attachment
C-VCA with scalloped trimline and 5 mm buccal extension at teeth 2∼6vertical wedge-shaped
C-OHCA with scalloped trimline and 5 mm buccal extension at teeth 2∼6overhanging
D-NACA with scalloped trimline and 5 mm buccolingual extension at teeth 2∼6no attachment
D-VCA with scalloped trimline and 5 mm buccolingual extension at teeth 2∼6vertical wedge-shaped
D-OHCA with scalloped trimline and 5 mm buccolingual extension at teeth 2∼6overhanging

12 groups of models for FEA in this study, comprising 3 attachment types and 4 CA trimline designs.

2.3 Preprocessing of 3D reconstructed models

The reconstructed maxilla and dentition were then imported into 3-Matic Medical software (Version 9.0, Materialise, Leuven, Belgium) for further processing. Both first premolars (4) were removed to simulate subtractive tooth extraction. Alveolar bone was divided into cortical bone and cancellous bone, with 2 mm thickness of cortical bone. The alveolar bone was expanded outward by 2 mm to simulate the gingiva, and the space between the teeth and alveolar bone was expanded by 0.2 mm to simulate the PDL, according to typical parameters in existing orthodontic FEA studies (Canales et al., 2013; Cao et al., 2023; Çifter and Saraç, 2011).

3D reconstructed models of maxilla, dentition, gingiva, PDL, CA, and attachment in STL format were imported into the Geomagic software (3D-SYSTEM, United States) for surface smoothing, eliminating any potential geometric imperfections and noise. Subsequently, the output file was exported as IGES format and imported into Hypermesh software (Altair, United States) for preprocessing, including geometry repair, component organization, convergence experiment and shell meshing, material parameter setting, contact definition, and the application of boundary conditions. A symmetric dentition model with respect to the sagittal plane was adopted for analysis, with results ultimately presented as unilateral dentition.

2.4 Material properties

The material parameters used in this study were obtained from material suppliers and literature (Seo et al., 2021; Canales et al., 2013; Sarrafpour et al., 2013), as detailed in Table 2. PDL was modeled using a visco-hyperelastic-damage constitutive model (Wu et al., 2022; Natali et al., 2008). To reduce computational complexity while maintaining accuracy, the teeth were modeled as rigid bodies, as prior studies have demonstrated that the difference in predicted displacement between rigid-body and fully elastic models was less than 3%, and that tooth deformation under orthodontic forces was negligible compared to overall movement (Tamaya et al., 2021; Hamanaka et al., 2017; Kim et al., 2025).

TABLE 2

Material nameCAAttachmentGingivaCancellous boneCortical bone
Elastic Modulus/GPa125012502.8137013700
Poisson’s Ratio0.360.360.40.30.3

Mechanical parameters (elastic modulus and Poisson’s ratio) of the materials.

2.5 Loading and boundary conditions

The interaction between CA and teeth was achieved via an interference fit, applying normal hard contact and tangential frictional sliding conditions, and a friction coefficient of 0.2. Tie constraints were applied between PDL and teeth, the outer layer of PDL and alveolar bone, alveolar bone and gingiva, teeth and gingiva (Natali et al., 2008; Hahn et al., 2009; Ortún-Terrazas et al., 2020; Huang et al., 2023; Goktas et al., 2011). The symmetrical model with respect to the sagittal plane was subjected to symmetrical displacement constraints.

2.6 Construction of 3D finite element model

The general FEA software ABAQUS (DASSAULT, France) was employed as the solver and processor to perform a detailed biomechanical analysis. Element sizes of the models were determined via the convergence experiment to ensure a sufficiently refined mesh for the simulation, and the applied element types, sizes and counts, and node counts were listed in Table 3. As shown in Table 3 and Figure 2, CAs were simulated using S4R shell elements and an element size of 0.2 mm. Attachments were simulated using C3D10M elements with a size of 0.2 mm. Gingiva and PDL were modeled using C3D10H elements with a size of 0.2 mm. For alveolar bone, C3D10 elements were applied, with element sizes of 0.2 mm in regions sharing nodes with gingiva and PDL (Seo et al., 2021; AlKahlan et al., 2025), and gradual transition of element sizes to 2 mm at other regions. Dentition was simulated using R3D3 elements with a size of 0.2 mm in regions sharing nodes with PDL, and gradual transition of element sizes to 0.8 mm at other regions.

TABLE 3

ComponentElement typeElement size (mm)Element countNode count
GingivaC3D10H0.2184849299435
PDLC3D10H0.294166189790
ToothR3D30.2∼0.83795719019
Cancellous BoneC3D100.2∼2169990266400
Cortical BoneC3D100.2∼2109100185611
AttachmentC3D10M0.2578310225
CAS4R0.23777737535

Element types, sizes, counts, and node counts for the models. Element sizes were determined via the convergence experiment to ensure a sufficiently refined mesh for the simulation.

FIGURE 2

3 Results and discussions

3.1 Hydrostatic pressure in PDL

The hydrostatic pressure exerted on the PDL served as an indicator of its tensile and compressive states. Exposure to tensile or compressive stress activated the internal vascular and neural elements of the PDL under their stress conditions, thereby triggering a series of biochemical reactions and activating osteoclasts and osteoblasts, which led to bone remodeling at the interface between the PDL and the alveolar bone. Typically, bone resorption occurred on the compression side, while bone formation took place on the tension side (Sarrafpour et al., 2013; Li et al., 2021).

As illustrated in the Figure 3a, it was evident that during the root control movement of canine (3) following the extraction of first premolar (4), the disto-cervical and mesio-apical areas of PDL near were subjected to compression (+), whereas the disto-apical and mesio-cervical areas were tension (−). The maximum and mean stress values for these regions were presented in Table 4. The comparison of the maximum hydrostatic pressure in the PDL was shown in Figure 3b. Under the same CA trimline design condition, the peak hydrostatic pressure in the PDL progressively increased from the no-attachment (NA) group, to the regular vertical attachment (V) group, and further to the overhanging attachment (OH) group, with the OH group exhibiting the highest pressure. Similarly, under the same attachment design condition, the peak hydrostatic pressure also increased sequentially from the conventional CA trimline (A), to the 3∼5 buccal extension (B), to the 2∼6 buccal extension (C), and to the 2∼6 buccolingual extension (D), with the D group presenting the maximum pressure. The application of regular attachments or OH attachments, as well as the buccal or lingual extended coverage of the CA, all enhanced the orthodontic force, thereby increasing the stress values in PDL. In all, the CA with 2∼6 buccolingual gingival extensions and OH attachments exhibited the highest hydrostatic pressure in PDL.

FIGURE 3

TABLE 4

GroupLocationMaximumMean
MesialDistalMesialDistal
A-NACervical−52.053.1−39.541.9
Apical53.0−53.163.3−41.1
A-VCervical−88.281.0−71.166.6
Apical73.3−79.863.3−70.1
A-OHCervical−99.795.5−83.378.0
Apical81.0−88.065.5−74.5
B-NACervical−57.557.5−46.843.8
Apical56.2−57.246.6−47.6
B-VCervical−104.792.6−84.179.8
Apical79.8−89.765.7−73.6
B-OHCervical−107.698.6−87.384.5
Apical78.5−90.667.3−72.3
C-NACervical−62.164.8−54.451.6
Apical60.8−63.850.5−51.7
C-VCervical−106.893.4−84.980.2
Apical79.2−91.663.9−73.9
C-OHCervical−114.6107.1−96.289.5
Apical83.1−96.071.0−78.5
D-NACervical−68.666.3−54.453.0
Apical61.0−64.950.7−51.2
D-VCervical−110.7100.4−90.692.5
Apical83.6−95.571.1−79.2
D-OHCervical−115.2113.1−96.995.3
Apical86.0−91.272.8−76.4

Maximum and mean compressive (+) and tensile (−) stress in PDL of canine (unit: kPa).

The ratio of mesio-apical to disto-cervical stress of PDL reflected the tipping extent of the tooth (Bi and Shi, 2023). In an ideal situation when the canine moved distally and translationally, both mesio-cervical and mesio-apical hydrostatic stress were tensile (with negative value or zero), and both disto-cervical and disto -apical hydrostatic stress were compressive (with positive value or zero), resulting in a zero or negative value of the ratio of mesio-apical to disto-cervical stress. Therefore, the closer the ratio of mesio-apical to disto-cervical stress was to zero or negative value, the closer the tooth movement was to distal translation. In this study, the ratios of mesio-apical to disto-cervical stress in PDL of canine in different groups were shown in Figure 4, calculated with maximum stress value (Figure 4a) or mean value (Figure 4b). With the same CA trimline design, the ratios progressively decreased from no-attachment (NA) group, to regular vertical attachment (V) group, and further to OH attachment group, with the OH group exhibiting the lowest ratio, indicating OH attachment was the most effective for achieving bodily tooth movement among the three attachment designs. Similarly, with the same attachment design, the ratios decreased sequentially from the conventional CA trimline (A), to the 3∼5 buccal extension (B), to the 2∼6 buccal extension (C), and to the 2∼6 buccolingual extension (D), with the D group presenting the minimum ratio. It could be concluded that the CA with 2∼6 buccolingual gingival extensions and OH attachments represented the optimal configuration for facilitating translational canine movement.

FIGURE 4

3.2 Tooth displacement

Figure 5 presented the heatmap of tooth movement in vector diagram. In all groups, distal inclination of canine was observed, accompanied with labial tipping of incisors and mesial inclination of molars, as an outcome of normal anchorage effect. As evidenced by the upper limits of the heatmap scale, the trend in tooth displacement aligned with the prior findings for PDL hydrostatic pressure, following the pattern OH > V > NA and D > C > B > A.

FIGURE 5

Figure 6 presented the heatmap of tooth movement with contour, where the point of minimum displacement (the center of blue region) represented the center of rotation of canine (Seo et al., 2021). Distances from root apex and tooth cusp to the rotational center were respectively designated as d1 and d2. The smaller the ratio of d1/d2, the closer the rotational center was to the root apex. And a ratio of zero indicated translational movement of the tooth. The ratios of d1/d2 in different groups were shown in Figure 7a. The ratio of mesio-apical to disto-occlusal displacement of canine (Figure 7b; Table 5) also reflected the degree of root control. The smaller the ratio, the closer the tooth movement was to distal translation (Liu L. et al., 2022).

FIGURE 6

FIGURE 7

TABLE 5

GroupMesio-apical displacementDisto-occlusal displacement
A-NA0.05320.0269
A-V0.07090.0355
A-OH0.07420.0369
B-NA0.05610.0281
B-V0.07840.0389
B-OH0.08060.0382
C-NA0.06200.0307
C-V0.07960.0391
C-OH0.08060.0378
D-NA0.06390.0313
D-V0.08350.0401
D-OH0.08070.0376

Mesio-apical and disto-occlusal displacement of canine (unit: mm).

With the same CA trimline design, the ratios of d1/d2 progressively decreased from no-attachment (NA) group, to regular vertical attachment (V) group, and further to OH attachment group, with the OH group reaching the minimum value. Similarly, with the same attachment design, the ratios of d1/d2 decreased sequentially from the conventional CA trimline (A), to the 3∼5 buccal extension (B), to the 2∼6 buccal extension (C), and to the 2∼6 buccolingual extension (D), with the D group presenting the minimum ratio. In all, D-OH group exhibited the smallest ratio of d1/d2. A similar trend was observed regarding the ratio of mesio-apical to disto-occlusal displacement of canine (Figure 7b; Table 5).

The CA with extended buccal and lingual gingival coverage and OH attachments effectively reduced the ratio of d1/d2 to 0.461, which was beneficial for controlling tooth rotation and reducing the inclination of tooth during translational movement. Findings in this study were consistent with previous reports (Elshazly et al., 2023; Elshazly et al., 2022) suggesting that extending the gingival coverage of CA benefited root control movement.

3.3 Deformation of CA

The deformation of the CA was shown in Figure 8. The maximum deformation occurred at the labial side of lateral incisor in groups without attachments (NA), which was attributed to the compression caused by anchorage effect that endowed the mesial movement of canine, resulting in a labial protrusion. With attachments on 3 and 5, maximum deformation occurred around the attachment on canine, deformation of CA at the lateral incisor was reduced, and the deformation at the buccal side of molars became more pronounced. As evidenced by the upper limits of the heatmap scale, introduction of attachments reduced the overall maximum deformation of CA, indicating more evenly distributed orthodontic force. With buccal extended coverage on 2∼6, the overall maximum deformation of CA increased, while with buccolingual extended coverage on 2∼6, the overall maximum deformation of CA decreased except for the OH attachment group (D-OH). The buccolingual extension exerted a slight torsion along the dental arch curve towards the mesial direction. Moreover, upon increasing the lingual and gingival coverage, and the maximum deformation of the CA shifted to the lingual gingival extension under the first molar, resulting in a decrease in deformation of molar crowns and enhancing the retention of CA, thereby avoiding the occurrence of attachment debonding.

FIGURE 8

3.4 Stress distribution on CA

Figure 9 indicated that an increase in maximum Von-Mises stress value was observed in CAs with no attachment (NA) as the gingival coverage enlarged (D > C > B > A). Comparing A-V with A-NA, the maximum Von-Mises stress values increased after the addition of a regular vertical attachment (V), while buccal extension of CA (B-V and C-V) resulted in decreased maximum Von-Mises stress to the level of A-N. Using regular vertical attachment (V), buccolingual extension (D-V) exhibited slightly increased stress compared to buccal extension (B-V and C-V). Notably, using OH attachment, there was no significant difference in the maximum Von-Mises stress values with or without extended gingival coverage, indicating the OH attachment could minimize stress concentration in CA. The membrane material employed in this simulation was PETG, with a yield strength of 45 MPa. All designs of CA met the material strength requirements, ensuring that no plastic deformation occurred during the wearing process.

FIGURE 9

4 Conclusion

This study first reported the biomechanical effects of overhanging attachment combined with partial gingival extension of CA trimline during closing extraction spaces. The results indicate that the overhanging attachment was more effective than the regular vertical attachment in achieving root-controlled canine movement. The highest orthodontic efficiency was observed when the overhanging attachment was combined with a buccolingual gingival extension of CA trimline on 2–6, while simultaneously avoiding excessive aligner deformation and stress concentration. This study provided theoretical bases for the clinical application of overhanging attachment. The partial gingival extension of the CA trimline proposed in this study demonstrated greater feasibility in clinical application compared to the full gingiva-covering trimline design. The methodology and findings presented here provided reference and guidance for the design of future orthodontic treatment.

Statements

Data availability statement

The original contributions presented in the study are included in the article/supplementary material, further inquiries can be directed to the corresponding authors.

Author contributions

GZ: Methodology, Visualization, Formal Analysis, Writing – original draft, Writing – review and editing. XM: Visualization, Conceptualization, Supervision, Writing – original draft, Data curation. DL: Writing – review and editing, Project administration, Visualization, Validation, Supervision, Investigation, Conceptualization, Funding acquisition.

Funding

The author(s) declare that financial support was received for the research and/or publication of this article. The authors gratefully acknowledge the support of the National Natural Science Foundation of China (32201104), and the Science and Technology Commission of Shanghai Municipality (24QA2708100).

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Generative AI statement

The author(s) declare that no Generative AI was used in the creation of this manuscript.

Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

  • 1

    AlKahlanL. A.BindayelN. A.MallekA. M.BendjaballahM. Z. (2025). Extensive iterative finite element analysis of molar uprighting with the introduction of a novel method for estimating clinical treatment time. Appl. Sci.15 (12), 6463. 10.3390/app15126463

  • 2

    BiS.ShiG. (2023). The crucial role of periodontal ligament's Poisson's ratio and tension-compression asymmetric moduli on the evaluation of tooth displacement and stress state of periodontal ligament. J. Mech. Behav. Biomed. Mater148, 106217. 10.1016/j.jmbbm.2023.106217

  • 3

    CanalesC.LarsonM.GrauerD.SheatsR.StevensC.KoC. C. (2013). A novel biomechanical model assessing continuous orthodontic archwire activation. Am. J. Orthod. Dentofac. Orthop.143 (2), 281–290. 10.1016/j.ajodo.2012.06.019

  • 4

    CaoY.WangZ. W.ChenD.LiuL.LiD. X.LiN.et al (2023). The effect of space arrangement between anterior teeth on their retraction with clear aligners in first premolar extraction treatment: a finite element study. Prog. Orthod.24 (1), 39. 10.1186/s40510-023-00484-1

  • 5

    ChengY.LiuX.ChenX.LiX.FangS.WangW.et al (2022a). The three-dimensional displacement tendency of teeth depending on incisor torque compensation with clear aligners of different thicknesses in cases of extraction: a finite element study. BMC Oral Health22 (1), 499. 10.1186/s12903-022-02521-7

  • 6

    ChengY.GaoJ.FangS.WangW.MaY.JinZ. (2022b). Torque movement of the upper anterior teeth using a clear aligner in cases of extraction: a finite element study. Prog. Orthod.23 (1), 26. 10.1186/s40510-022-00421-8

  • 7

    ÇifterM.SaraçM. (2011). Maxillary posterior intrusion mechanics with mini-implant anchorage evaluated with the finite element method. Am. J. Orthod. Dentofac. Orthop.140 (5), e233–e241. 10.1016/j.ajodo.2011.06.019

  • 8

    DemirG. B. (2024). Understanding the effectiveness of attachments in clear aligner therapy: navigating design, placement, material selection and biomechanics. Australas. Orthod. J.40 (2), 63–74. 10.2478/aoj-2024-0021

  • 9

    ElshazlyT. M.KeiligL.SalvatoriD.ChavanneP.AldesokiM.BourauelC. (2022). Effect of trimming line design and edge extension of orthodontic aligners on force transmission: an in vitro study. J. Dent.125, 104276. 10.1016/j.jdent.2022.104276

  • 10

    ElshazlyT. M.SalvatoriD.ElattarH.BourauelC.KeiligL. (2023). Effect of trimming line design and edge extension of orthodontic aligners on force transmission: a 3D finite element study. J. Mech. Behav. Biomed. Mater140, 105741. 10.1016/j.jmbbm.2023.105741

  • 11

    ElshazlyT. M.BourauelC.ChavanneP.ElattarH.KeiligL. (2024a). Numerical biomechanical finite element analysis of different trimming line designs of orthodontic aligners: an in silico study. J. World Fed. Orthod.13 (2), 65–71. 10.1016/j.ejwf.2024.01.001

  • 12

    ElshazlyT. M.BourauelC.AldesokiM.SalvatoriD.AlhotanA.KeiligL.et al (2024b). Effect of attachment configuration and trim line design on the force system of orthodontic aligners: a finite element study on the upper central incisor. Orthod. Craniofac Res.27 (Suppl. 2), 131–140. 10.1111/ocr.12779

  • 13

    GoktasS.DmytrykJ. J.McFetridgeP. S. (2011). Biomechanical behavior of oral soft tissues. J. Periodontol.82 (8), 1178–1186. 10.1902/jop.2011.100573

  • 14

    GomezJ. P.PeñaF. M.MartínezV.GiraldoD. C.CardonaC. I. (2015). Initial force systems during bodily tooth movement with plastic aligners and composite attachments: a three-dimensional finite element analysis. Angle Orthod.85 (3), 454–460. 10.2319/050714-330.1

  • 15

    HahnW.DatheH.Fialka-FrickeJ.Fricke-ZechS.ZapfA.Kubein-MeesenburgD.et al (2009). Influence of thermoplastic appliance thickness on the magnitude of force delivered to a maxillary central incisor during tipping. Am. J. Orthod. Dentofac. Orthop.136 (1), 12.e1–12.e7. 10.1016/j.ajodo.2008.12.015

  • 16

    HamanakaR.YamaokaS.AnhT. N.TominagaJ. y.KogaY.YoshidaN. (2017). Numeric simulation model for long-term orthodontic tooth movement with contact boundary conditions using the finite element method. Am. J. Orthod. Dentofac. Orthop.152 (5), 601–612. 10.1016/j.ajodo.2017.03.021

  • 17

    HongK.KimW. H.Eghan-AcquahE.LeeJ. H.LeeB. K.KimB. (2021). Efficient design of a clear aligner attachment to induce bodily tooth movement in orthodontic treatment using finite element analysis. Mater. (Basel)14 (17), 4926. 10.3390/ma14174926

  • 18

    HongY. Y.KangT.ZhouM. Q.ZhongJ. Y.ChenX. P. (2024). Effect of varying auxiliaries on maxillary incisor torque control with clear aligners: a finite element analysis. Am. J. Orthod. Dentofac. Orthop.166 (1), 50–60. 10.1016/j.ajodo.2024.02.012

  • 19

    HongK. J.BaekJ. P.ParkJ. J.ParkJ. H. (2025). Repair of receded gingival tissue through lingual root movement with an overhanging attachment and clear aligners. AJO-DO Clinical Companion.

  • 20

    HuangW.DongL.YuS.ZhengY.WuH.LiC.et al (2023). Stress distribution of the modified clear twin-block aligner on the temporomandibular joint, alveolar bone and teeth: a finite element analysis. Int. Orthod.21 (4), 100815. 10.1016/j.ortho.2023.100815

  • 21

    JedlińskiM.MazurM.GrecoM.BelfusJ.GrocholewiczK.Janiszewska-OlszowskaJ. (2023). Attachments for the orthodontic aligner treatment-state of the Art-A comprehensive systematic review. Int. J. Environ. Res. Public Health20 (5), 4481. 10.3390/ijerph20054481

  • 22

    JiangT.WuR. Y.WangJ. K.WangH. H.TangG. H. (2020). Clear aligners for maxillary anterior en masse retraction: a 3D finite element study. Sci. Rep.10 (1), 10156. 10.1038/s41598-020-67273-2

  • 23

    KimK.ChoiY. K.KimS. H.KimS. S.KimY. I. (2025). Multi-step finite element simulation for clear aligner space closure: a proof-of-concept compensation protocol. Sci. Rep.15 (1), 22220. 10.1038/s41598-025-07110-6

  • 24

    LiY.ZhanQ.BaoM.YiJ.LiY. (2021). Biomechanical and biological responses of periodontium in orthodontic tooth movement: up-date in a new decade. Int. J. Oral Sci.13 (1), 20. 10.1038/s41368-021-00125-5

  • 25

    LiuL.ZhanQ.ZhouJ.KuangQ.YanX.ZhangX.et al (2021). Effectiveness of an anterior mini-screw in achieving incisor intrusion and palatal root torque for anterior retraction with clear aligners. Angle Orthod.91 (6), 794–803. 10.2319/120420-982.1

  • 26

    LiuL.SongQ.ZhouJ.KuangQ.YanX.ZhangX.et al (2022). The effects of aligner overtreatment on torque control and intrusion of incisors for anterior retraction with clear aligners: a finite-element study. Am. J. Orthod. Dentofac. Orthop.162 (1), 33–41. 10.1016/j.ajodo.2021.02.020

  • 27

    LiuJ. Q.ZhuG. y.WangY. g.ZhangB.WangS. c.YaoK.et al (2022). Different biomechanical effects of clear aligners in bimaxillary space closure under two strong anchorages: finite element analysis. Prog. Orthod.23 (1), 41. 10.1186/s40510-022-00435-2

  • 28

    NakornnoiT.SrirodjanakulW.ChintavalakornR.SantiwongP.SipiyarukK. (2024). The biomechanical effects of clear aligner trimline designs and extensions on orthodontic tooth movement: a systematic review. BMC Oral Health24 (1), 1523. 10.1186/s12903-024-05274-7

  • 29

    NataliA. N.CarnielE. L.PavanP. G.SanderF. G.DorowC.GeigerM. (2008). A visco-hyperelastic-damage constitutive model for the analysis of the biomechanical response of the periodontal ligament. J. Biomech. Eng.130 (3), 031004. 10.1115/1.2900415

  • 30

    Ortún-TerrazasJ.CegoñinoJ.Pérez Del PalomarA. (2020). In silico study of cuspid' periodontal ligament damage under parafunctional and traumatic conditions of whole-mouth occlusions. A patient-specific evaluation. Comput. Methods Programs Biomed.184, 105107. 10.1016/j.cmpb.2019.105107

  • 31

    PuL.ZhouJ.YanX.ZhouH.LiuX.YangZ.et al (2022). Orthodontic traction of an impacted maxillary third molar through a miniscrew-anchored cantilever spring to substitute the adjacent second molar with severe root resorption. J. Am. Dent. Assoc.153 (9), 884–892. 10.1016/j.adaj.2022.05.010

  • 32

    SarrafpourB.SwainM.LiQ.ZoellnerH. (2013). Tooth eruption results from bone remodelling driven by bite forces sensed by soft tissue dental follicles: a finite element analysis. PLoS One8 (3), e58803. 10.1371/journal.pone.0058803

  • 33

    SeoJ. H.Eghan-AcquahE.KimM. S.LeeJ. H.JeongY. H.JungT. G.et al (2021). Comparative analysis of stress in the periodontal ligament and center of rotation in the tooth after orthodontic treatment depending on clear aligner thickness-finite element analysis study. Mater. (Basel)14 (2), 324. 10.3390/ma14020324

  • 34

    TamayaN.KawamuraJ.YanagiY. (2021). Tooth movement efficacy of retraction spring made of a new low elastic modulus material, gum metal, evaluated by the finite element method. Mater. (Basel)14 (11), 2934. 10.3390/ma14112934

  • 35

    WangQ.DaiD.WangJ.ChenY.ZhangC. (2022). Biomechanical analysis of effective mandibular en-masse retraction using class II elastics with a clear aligner: a finite element study. Prog. Orthod.23 (1), 23. 10.1186/s40510-022-00417-4

  • 36

    WeirT. (2017). Clear aligners in orthodontic treatment. Aust. Dent. J.62 (Suppl. 1), 58–62. 10.1111/adj.12480

  • 37

    WuB.ChengK.LiuM.LiuJ.JiangD.MaS.et al (2022). Construction of hyperelastic model of human periodontal ligament based on collagen fibers distribution. J. Mech. Behav. Biomed. Mater135, 105484. 10.1016/j.jmbbm.2022.105484

  • 38

    XiaQ.HeY.JiaL.WangC.WangW.WangC.et al (2022). Assessment of labially impacted canines traction mode with clear aligners vs. fixed appliance: a comparative study based on 3D finite element analysis. Front. Bioeng. Biotechnol.10, 1004223. 10.3389/fbioe.2022.1004223

  • 39

    YokoiY.AraiA.KawamuraJ.UozumiT.UsuiY.OkafujiN. (2019). Effects of attachment of plastic aligner in closing of diastema of maxillary dentition by finite element method. J. Healthc. Eng.2019, 1–6. 10.1155/2019/1075097

  • 40

    ZhuC.LiR.YuanL.ZhengY.JinY.LiH.et al (2022). Effects of the advanced mandibular spring on mandibular retrognathia treatment: a three-dimensional finite element study. BMC Oral Health22 (1), 271. 10.1186/s12903-022-02308-w

Summary

Keywords

clear aligners, finite element analysis, overhanging attachment, bodily movement, interdental space closure

Citation

Zeng G, Ma X and Lin D (2026) Finite element analysis of clear aligner with overhanging attachments and extended gingival coverage for interdental space closure . Front. Bioeng. Biotechnol. 13:1636262. doi: 10.3389/fbioe.2025.1636262

Received

27 May 2025

Revised

30 September 2025

Accepted

21 October 2025

Published

16 January 2026

Volume

13 - 2025

Edited by

Jun Liao, University of Texas at Arlington, United States

Reviewed by

Jiahui Li, University of Pennsylvania, United States

Gizem Boztaş Demir, University of Health Sciences Diyarbakır Gazi Yaşargil Education and Research Hospital, Türkiye

Updates

Copyright

*Correspondence: Dan Lin, ; Xiaoqing Ma,

Disclaimer

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Outline

Figures

Cite article

Copy to clipboard


Export citation file


Share article

Article metrics