Your new experience awaits. Try the new design now and help us make it even better

METHODS article

Front. Bioeng. Biotechnol.

Sec. Tissue Engineering and Regenerative Medicine

Application of Rebound Tonometer on Skin Surface Pressure Measurement: A Pilot Proof of Concept Study

Provisionally accepted
Wei  DingWei Ding1Wei  Wei BianWei Wei Bian2Lan  YangLan Yang1Ying  吴Ying 吴2Ying  HuangYing Huang1*
  • 1Department of Nursing,Shanghai Ninth People’s Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, China
  • 2Department of Plastic and Reconstructive Surgery, Shanghai Ninth People’s Hospital, Shanghai Jiao Tong University School of Medicine,, Shanghai, China

The final, formatted version of the article will be published soon.

BACKGROUND Skin tissue expansion is a widely used technique for generating additional healthy tissue to reconstruct congenital or acquired defects. Successful expansion relies on appropriate pressure management, since insufficient or excessive pressure may lead to prolonged treatment or expansion failure. Previous studies have attempted to estimate mechanical stress on expanded skin by measuring intracapsular pressure; however, this approach does not directly capture true mechanical pressure at specific skin locations. A direct method for measuring skin surface pressure remains unavailable. In this study, we evaluated the feasibility of using a rebound tonometer – a device designed for intraocular pressure measurement – for direct, site-specific assessment of skin surface pressure and for monitoring adverse events during tissue expansion. METHODS Patients with giant nevi meeting the eligibility criteria were enrolled following dilator implantation. Surface pressure measurements were obtained using the Icare® TA01i tonometer at nine predefined sites on the expanded region. The expansion protocol involved incremental saline injections of approximately 10% of the dilator's total volume per session. At each site, pressure was recorded at three timepoints: pre-inoculation, immediately post-inoculation, and 30 minutes post-inoculation. Monitoring continued throughout the expansion process until the dilator reached twice its original volume. RESULTS A total of five pediatric patients (2 boys and 3 girls), aged 3–6 years, were enrolled in this study. Each patient underwent a 34-day expansion period comprising eight or ten saline injections. No adverse events were recorded throughout the study. The rebound tonometer successfully detected dynamic changes in skin surface pressure corresponding to dilator volume increases. Furthermore, multi-point measurements across the expanded region confirmed the presence of spatially heterogeneous pressure distribution, demonstrating significant variations between different anatomical locations. CONCLUSIONS This pilot study supports the technical feasibility of using a rebound tonometer for monitoring skin surface pressure dynamics during tissue expansion. The device demonstrated favorable operational efficiency, safety profile, and measurement reproducibility throughout the expansion period. These preliminary findings suggest that rebound tonometry may represent a promising monitoring approach in tissue expansion protocols, though larger-scale validation is warranted to substantiate these observations.

Keywords: skin tissue expansion, Dilator, pressure measurement, Rebound tonometer, Children

Received: 09 Jul 2025; Accepted: 10 Nov 2025.

Copyright: © 2025 Ding, Bian, Yang, 吴 and Huang. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) or licensor are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

* Correspondence: Ying Huang, 217130@sh9hospital.org.cn

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.