Introduction
Elevated resting heart rate (HR) is a recognized cardiovascular risk factor in populations with or without cardiovascular affliction (, ). High HR comes with increased myocardial oxygen demand in a decreased diastolic filling time and coronary perfusion potentially interfering with the myocardial oxygen demand/supply balance and cardiac performance. In addition, elevated HR may underlie the abnormal vascular shear and pulsatile stress associated with endothelial dysfunction, vascular stiffness and oxidative stress (, ). Similarly to blood pressure (BP), HR undergoes circadian variations primarily dependent on the sympathovagal balance. Non-dipping, i.e., blunted BP or HR decline overnight (by <10% of the daytime mean), is coupled with an adverse cardiovascular prognosis (). Although sustained relative sympathetic over-activation at night is considered to be pivotal in the pathophysiology of non-dipping HR, other factors may also participate, such as the disrupted endogenous circadian rhythmicity of the central clock in the suprachiasmatic nucleus, the cardiac clock in the sinoatrial node, or circulating neurohumoral factors, including catecholamines and glucocorticoids (–). Non-dipping HR is rather neglected during cardiovascular risk assessment and is thus omitted from risk management strategies (). Nevertheless, a restricted number of studies have shown that non-dipping HR may be of adverse prognostic value in the general population () and various cardiovascular pathologies, including hypertension (), chronic kidney disease () and type 2 diabetes (). Indeed, a 10% reduction of overnight HR decline in a general and hypertensive population was associated with an increased risk of all-cause mortality by 34 and 30%, respectively (, ). Of note, the overnight HR dip was shown to be progressively attenuated by physiological ageing (, , ). Thus, age should be taken into account in cardiovascular risk assessment using non-dipping HR to yield valid results. In fact, non-dipping HR was found to be an independent risk factor of adverse cardiovascular outcomes after adjustment for several covariates, such as age, gender, blood pressure, body mass index, co-morbidities, or treatments (, , , ). For instance, Ben-Dov et al. () calculated that the risk of non-dipping HR increased by 13% per 5 years of age and the risk of all-cause mortality categorically increased by 45% with non-dipping HR after adjustment for multiple covariates including age. Unfortunately, research into non-dipping HR is not advancing at a desirable pace; thus, only seldom has data on this topic emerged in recent decades.
Consumer-Grade Wrist-Worn Devices for Monitoring Non-Dipping HR in Hypertension
Consumer-grade wearable devices capable of monitoring health represent an ever-growing market that is expected to reach 27.5 billion USD by 2026. Wrist-worn devices, including smartwatches and fitness trackers, account for more than half of all wearable devices being used (). Various sensors and technologies are utilized by these devices to gather health parameters, but photoplethysmography (PPG) is considered quintessential. PPG is an optical measurement technique to detect blood volume changes in subcutaneous tissue based on the interaction of light (absorption or reflection) with blood in the tissue, thus allowing for a beat-to-beat pulse signal interpretation. The PPG-derived waveform is used to extrapolate several parameters, including HR, heart rhythm, and indices of autonomic balance (). The identification of motion and noise artifacts by advanced artificial intelligence, particularly machine learning, has significantly improved the accuracy of HR measurement at rest, and even during physical activity, or arrhythmia (). A set of clinical studies assessed the accuracy of HR monitoring by consumer-grade wrist-worn devices under varying physical exertion. In general, the accuracy of wrist-worn devices was best at rest (reaching 90–95% agreement with electrocardiogram, ECG), worsened with exercise, and varied with exercise type (–). Indeed, a meta-analysis of 44 validation studies comprising 738 effect sizes across 15 consumer-grade wrist-worn devices showed that HR data collected by these devices during common activities under free-living conditions (sleep, rest, treadmill, post-exercise, and daily living activities) closely resembles HR data derived from ECG or chest-strap telemetry. Importantly, the mean differences between wrist-worn devices and a criterion HR measurement (ECG or chest-strap telemetry) were minimal during night-time sleep (−0.40 bpm) and daytime rest (−0.01 bpm) or when walking/running (−0.51 bpm). A sizeable difference was found in only two vigorous training activities: cycling and resistance training (). Recently, a multiple component algorithm was developed to approximate circadian HR changes using HR and activity and sleep data measured by wrist-worn devices. A validation study comprising 47 individuals using a wrist-worn device showed this algorithm to accurately reproduce a circadian HR pattern (). Thus, the above data provide evidence that wrist-worn devices are able to measure HR across night-time sleep and common daytime activities and reproduce a circadian HR pattern under free-living conditions with high accuracy.
Interestingly, wrist-worn devices were shown to accurately and reliably determine short-term and 24-hour (24-h) heart rate variability (HRV) by extrapolation from PPG-derived waveforms (–). Time- and frequency-domain HRV indices are considered a dynamic metric of the autonomic nervous system, including cardiac autonomic balance or baroreflex sensitivity, thus making HRV a disease marker of clinical conditions with increased cardiovascular risk and a predictor of adverse cardiovascular outcomes (). Indeed, reduced circadian HRV was shown to be associated with autonomic imbalance () and reduced baroreflex sensitivity (, ). Thus, considering 24-h HRV together with other metrics simultaneously measured by wrist-worn devices (e.g., 24-h HR, non-dipping HR or activity/sleep data) may further improve cardiovascular risk assessment (). A cross-sectional study with 8,203,261 subjects from 74 countries showed that (i) diverse metrics of cardiac autonomic health can be derived from 24-h HRV measured by wrist-worn devices; and (ii) the correlation of 24-h HRV with other data obtained by these devices, such as physical activity, is feasible and can potentially be utilized to improve cardiovascular risk assessment ().
Nonetheless, PPG comes with various sources of noise that may interfere with HR measurement if left unmitigated by post-sensing processing via algorithms and data analytics. Such sources of noise can be primarily divided into three categories: (i) individual variation, e.g., skin tone, obesity, age, and gender; (ii) physiology, e.g., respiratory rate, venous pulsation, and local body temperature; and (iii) external factors, e.g., motion artifacts, ambient light, and pressure applied to the skin (). Of note, in validation studies, wrist-worn devices using PPG for HR measurement achieved high agreement with standard ECG, or chest-strap telemetry, indicative of minimal interference by noise or other artifacts ().
Importantly, wearable devices are gaining popularity among both patients and physicians, as the use of these devices by consumers nearly quadrupled in a 4-year period (from 9% in 2014 to 33% in 2018) and 2 in 3 healthcare professionals surveyed intend to use wearable devices for remote patient monitoring ().
Here, we suggest that monitoring non-dipping HR by consumer-grade wrist-worn devices may be of help in cardiovascular risk assessment in hypertension (Figure 1). Our suggestion is based on the following:
Non-dipping HR is an independent, yet somewhat neglected, cardiovascular risk factor (, , ).
Consumer-grade wrist-worn devices are able to continuously or almost continuously (HR measurements are taken every few minutes) monitor HR during a 24-h period (or even longer), encompassing daytime activities and night-time sleep. Moreover, the HR readings are reliable and accurate, with at least 90% agreement with 24-h ECG ().
Simultaneously with HR monitoring, wearable devices can track physical activity and sleep (–), thus allowing for better determination of bedtime (or sleep-time) HR decline. Indeed, in a randomized cross-over trial and validation study involving 32 healthy individuals, a wrist-worn device accurately determined sleep duration, dream sleep and slow wave sleep and had a very low bias and precision error for sleep-time HR and respiratory rate measurements when compared with ECG and polysomnography (). Proper determination of the sleep/wake cycle and sleep-time HR is essential for non-dipping HR monitoring.
Compared to conventional continuous cardiac monitors, 98% of surveyed patients deemed wrist-worn devices more convenient and 91% acknowledged they were likely to use a wrist-worn device to determine cardiac rhythm during symptoms (). Indeed, using the patient's own wearable device can eliminate the anxiety associated with using assigned conventional cardiac monitors, thus improving measurement validity. Moreover, it is of no financial burden to the healthcare provider.
Data collected by wearable devices can be automatically transmitted to smartphones for patient notification and stored in remote servers (cloud) or electronic health records to be accessed remotely by a healthcare professional. This ensures timely data collection and the avoidance of errors by subjective data reporting or secondary data entry, which may contribute to increased patient convenience, satisfaction and compliance, as well as reduced in-office visits, administrative burden and healthcare costs (, , ).
The COVID-19 pandemic brought about new challenges in delivering health care to patients with chronic illnesses, including hypertension and diabetes mellitus (43). With social distancing and shelter-in-place, outpatient care is generally provided remotely via telemedicine. The proper management of chronic illnesses is also imperative under these circumstances (44). Monitoring non-dipping HR with wearable devices can provide valuable information on cardiovascular risk in patients with chronic illnesses even during lockdowns and social distancing measures.
Importantly, HR monitoring with wearable devices could play a pivotal role in clinical research pertaining to sleep-time HR or non-dipping HR in various cardiovascular pathologies. Indeed, wearable devices can continuously and during any time interval collect data from large populations encompassing thousands of healthy subjects or subjects with hypertension or diabetes mellitus and associated cardiovascular affliction, such as ischemic heart disease, heart failure or stroke. In fact, large populations can be stratified into specific subpopulations by age, gender, demographics, and other factors, thus allowing for larger and denser datasets and advanced analyses (). Population-based data on non-dipping HR (or rising HR) at bedtime may not only improve the pathophysiological understanding of this phenomenon but also show its prognostic value in cardiovascular and even other pathologies, thus rendering non-dipping HR a risk factor and marker of prognosis. Finally, data gathered by wearable devices could help to develop a well-tailored therapeutic approach targeting non-dipping HR (, , 45, 46).
Figure 1
Conclusion
Considering the above remarks, monitoring non-dipping HR by consumer-grade wrist-worn devices can be construed as an emerging avenue for cardiovascular risk assessment in hypertension, particularly in the realm of primary care. As primary care often struggles with high healthcare costs and a shortage of physicians or cutting-edge diagnostic tools, the suggested cost-effective approach to cardiovascular risk assessment using wearable devices may be advantageous. Moreover, early identification and management of high-risk patients in primary care can curb long-term healthcare expenditures (, 47).
Publisher's Note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
Statements
Author contributions
TB conceived and drafted the manuscript. FS revised the manuscript. All authors participated in data analysis, interpretation, and approved the submitted version.
Funding
This study was supported by research grants VEGA 1/0035/19 and VEGA 2/0112/19.
Conflict of interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
References
1.
SimkoFBakaTPaulisLReiterRJ. Elevated heart rate and non-dipping heart rate as potential targets for melatonin: a review. J Pineal Res. (2016) 61:127–37. 10.1111/jpi.12348
2.
PalatiniP. Resting heart rate as a cardiovascular risk factor in hypertensive patients: an update. Am J Hypertens. (2021) 34:307–17. 10.1093/ajh/hpaa187
3.
SimkoFBakaT. Ivabradine and blood pressure reduction: underlying pleiotropic mechanisms and clinical implications. Front Cardiovasc Med. (2021) 8:607998. 10.3389/fcvm.2021.607998
4.
BöhmMReilJCDeedwaniaPKimJBBorerJS. Resting heart rate: risk indicator and emerging risk factor in cardiovascular disease. Am J Med. (2015) 128:219–28. 10.1016/j.amjmed.2014.09.016
5.
BakaTSimkoF. Non-dipping heart rate: a neglected cardiovascular risk factor based on autonomic imbalance?Auton Neurosci. (2018) 210:83–4. 10.1016/j.autneu.2018.02.001
6.
BernardiJAromolaranKAZhuHAromolaranAS. Circadian mechanisms: cardiac ion channel remodeling and arrhythmias. Front Physiol. (2021) 11:611860. 10.3389/fphys.2020.611860
7.
D'SouzaAWangYAndersonCBucchiABaruscottiMOlieslagersSet al. A circadian clock in the sinus node mediates day-night rhythms in Hcn4 and heart rate. Heart Rhythm. (2021) 18:801–10. 10.1016/j.hrthm.2020.11.026
8.
BlackND'SouzaAWangYPigginsHDobrzynskiHMorrisGet al. Circadian rhythm of cardiac electrophysiology, arrhythmogenesis, and the underlying mechanisms. Heart Rhythm. (2019) 16:298–307. 10.1016/j.hrthm.2018.08.026
9.
Ben-DovIZKarkJDBen-IshayDMeklerJBen-ArieLBursztynM. Blunted heart rate dip during sleep and all-cause mortality. Arch Intern Med. (2007) 167:2116–21. 10.1001/archinte.167.19.2116
10.
VerdecchiaPSchillaciGBorgioniCCiucciATeleraMPPedeSet al. Adverse prognostic value of a blunted circadian rhythm of heart rate in essential hypertension. J Hypertens. (1998) 16:1335–43. 10.1097/00004872-199816090-00015
11.
BiyikZYavuzYCAltintepeLCelikGGuneyIOktarSF. Non-dipping heart rate and associated factors in patients with chronic kidney disease. Clin Exp Nephrol. (2019) 23:1298–305. 10.1007/s10157-019-01782-x
12.
MagriCJXuerebRGFavaS. Non-dipping heart rate and microalbuminuria in type 2 diabetes mellitus. Cardiology. (2014) 129:28–35. 10.1159/000362714
13.
CuspidiCFacchettiRBombelliMSalaCTadicMGrassiGet al. Night-time heart rate non-dipping: clinical and prognostic significance in the general population. J Hypertens. (2018) 36:1311–7. 10.1097/HJH.0000000000001703
14.
HozawaAInoueROhkuboTKikuyaMMetokiHAsayamaKet al. Predictive value of ambulatory heart rate in the Japanese general population: the Ohasama study. J Hypertens. (2008) 26:1571–6. 10.1097/HJH.0b013e3283041172
15.
DagherLShiHZhaoYMarroucheNF. Wearables in cardiology: here to stay. Heart Rhythm. (2020) 17:889–95. 10.1016/j.hrthm.2020.02.023
16.
AllenJ. Photoplethysmography and its application in clinical physiological measurement. Physiol Meas. (2007) 28:R1–39. 10.1088/0967-3334/28/3/R01
17.
WangRBlackburnGDesaiMPhelanDGillinovLHoughtalingPet al. Accuracy of Wrist-worn heart rate monitors. JAMA Cardiol. (2017) 2:104–6. 10.1001/jamacardio.2016.3340
18.
GillinovSEtiwyMWangRBlackburnGPhelanDGillinovAMet al. Variable accuracy of wearable heart rate monitors during aerobic exercise. Med Sci Sports Exerc. (2017) 49:1697–703. 10.1249/MSS.0000000000001284
19.
EtiwyMAkhrassZGillinovLAlashiAWangRBlackburnGet al. Accuracy of wearable heart rate monitors in cardiac rehabilitation. Cardiovasc Diagn Ther. (2019) 9:262–71. 10.21037/cdt.2019.04.08
20.
PasadynSRSoudanMGillinovMHoughtalingPPhelanDGillinovNet al. Accuracy of commercially available heart rate monitors in athletes: a prospective study. Cardiovasc Diagn Ther. (2019) 9:379–85. 10.21037/cdt.2019.06.05
21.
ReddyRKPooniRZaharievaDPSenfBEl YoussefJDassauEet al. Accuracy of Wrist-worn activity monitors during common daily physical activities and types of structured exercise: evaluation study. JMIR Mhealth Uhealth. (2018) 6:e10338. 10.2196/10338
22.
ZhangYWeaverRGArmstrongBBurkartSZhangSBeetsMW. Validity of Wrist-Worn photoplethysmography devices to measure heart rate: a systematic review and meta-analysis. J Sports Sci. (2020) 38:2021–34. 10.1080/02640414.2020.1767348
23.
MorelliDBartoloniLRossiACliftonDA. A computationally efficient algorithm to obtain an accurate and interpretable model of the effect of circadian rhythm on resting heart rate. Physiol Meas. (2019) 40:095001. 10.1088/1361-6579/ab3dea
24.
ParakJTarniceriuAReneveyPBertschiMDelgado-GonzaloRKorhonenI. Evaluation of the beat-to-beat detection accuracy of PulseOn wearable optical heart rate monitor. Annu Int Conf IEEE Eng Med Biol Soc. (2015) 2015:8099–102. 10.1109/EMBC.2015.7320273
25.
DurORhoadesCNgMSElsayedRvan MourikRMajmudarMD. Design rationale and performance evaluation of the wavelet health wristband: benchtop validation of a Wrist-worn physiological signal recorder. JMIR Mhealth Uhealth. (2018) 6:e11040. 10.2196/11040
26.
HernandoDRocaSSanchoJAlesancoÁBailónR. Validation of the apple watch for heart rate variability measurements during relax and mental stress in healthy subjects. Sensors (Basel). (2018) 18:2619. 10.3390/s18082619
27.
BaekHJChoJ. Novel heart rate variability index for Wrist-worn wearable devices subject to motion artifacts that complicate measurement of the continuous pulse interval. Physiol Meas. (2019) 40:105010. 10.1088/1361-6579/ab4c28
28.
NatarajanAPantelopoulosAEmir-FarinasHNatarajanP. Heart rate variability with photoplethysmography in 8 million individuals: a cross-sectional study. Lancet Digit Health. (2020) 2:e650–7. 10.1016/S2589-7500(20)30246-6
29.
ShafferFGinsbergJP. An overview of heart rate variability metrics and norms. Front Public Health. (2017) 5:258. 10.3389/fpubh.2017.00258
30.
AccardoAMerloMSilveriGDel PopoloLDalla LiberaLRestivoLet al. Influence of ageing on circadian rhythm of heart rate variability in healthy subjects. J Cardiovasc Med. (2021) 22:405–13. 10.2459/JCM.0000000000001048
31.
UenoLMDragerLFRodriguesACRondonMUMathiasWJrKriegerEMet al. Day-night pattern of autonomic nervous system modulation in patients with heart failure with and without sleep apnea. Int J Cardiol. (2011) 148:53–8. 10.1016/j.ijcard.2009.10.023
32.
LuciniDGuzzettiSCasiraghiSPaganiM. Correlation between baroreflex gain and 24-h indices of heart rate variability. J Hypertens. (2002) 20:1625–31. 10.1097/00004872-200208000-00026
33.
OwensAP. The role of heart rate variability in the future of remote digital biomarkers. Front Neurosci. (2020) 14:582145. 10.3389/fnins.2020.582145
34.
FineJBrananKLRodriguezAJBoonya-AnantaTAjmalRamella-RomanJCet al. Sources of inaccuracy in photoplethysmography for continuous cardiovascular monitoring. Biosensors. (2021) 11:126. 10.3390/bios11040126
35.
NelsonBWAllenNB. Accuracy of consumer wearable heart rate measurement during an ecologically valid 24-hour period: intraindividual validation study. JMIR Mhealth Uhealth. (2019) 7:e10828. 10.2196/10828
36.
FergusonTRowlandsAVOldsTMaherC. The validity of consumer-level, activity monitors in healthy adults worn in free-living conditions: a cross-sectional study. Int J Behav Nutr Phys Act. (2015) 12:42. 10.1186/s12966-015-0201-9
37.
KangSGKangJMKoKPParkSCMarianiSWengJ. Validity of a commercial wearable sleep tracker in adult insomnia disorder patients and good sleepers. J Psychosom Res. (2017) 97:38–44. 10.1016/j.jpsychores.2017.03.009
38.
Asgari MehrabadiMAzimiISarhaddiFAxelinANiela-VilénHMyllyntaustaSet al. Sleep tracking of a commercially available smart ring and smartwatch against medical-grade actigraphy in everyday settings: instrument validation study. JMIR Mhealth Uhealth. (2020) 8:e20465. 10.2196/20465
39.
BerryhillSMortonCJDeanABerryhillAProvencio-DeanNPatelSIet al. Effect of wearables on sleep in healthy individuals: a randomized crossover trial and validation study. J Clin Sleep Med. (2020) 16:775–83. 10.5664/jcsm.8356
40.
RajakariarKKoshyASajeevJRobertsLTehA. Increased preference of wearable devices compared with conventional cardiac monitoring. Heart Lung Circ. (2018) 27:S170. 10.1016/j.hlc.2018.06.288
41.
IzmailovaESWagnerJAPerakslisED. Wearable devices in clinical trials: hype and hypothesis. Clin Pharmacol Ther. (2018) 104:42–52. 10.1002/cpt.966
42.
OmboniS. Connected health in hypertension management. Front Cardiovasc Med. (2019) 6:76. 10.3389/fcvm.2019.00076
43.
SimkoFBakaT. Angiotensin-converting enzyme inhibitors and angiotensin II receptor blockers: potential allies in the COVID-19 pandemic instead of a threat?Clin Sci. (2021) 135:1009–14. 10.1042/CS20210182
44.
OmboniSMcManusRJBosworthHBChappellLCGreenBBKarioKet al. Evidence and recommendations on the use of telemedicine for the management of arterial hypertension: an international expert position paper. Hypertension. (2020) 76:1368–83. 10.1161/HYPERTENSIONAHA.120.15873
45.
SimkoFBakaT. Chronotherapy as a potential approach to hypertensive patients with elevated heart rate?Br J Clin Pharmacol. (2019) 85:1861–2. 10.1111/bcp.14020
46.
BakaTSimkoF. Ivabradine reversed non-dipping heart rate in rats with l-NAME-induced hypertension. Clin Exp Pharmacol Physiol. (2019) 46:607–10. 10.1111/1440-1681.13075
47.
AgboolaSSimonsMGolasSOp den BuijsJFelstedJFischerNet al. Health care cost analyses for exploring cost savings opportunities in older patients: longitudinal retrospective study. JMIR Aging. (2018) 1:e10254. 10.2196/10254
Summary
Keywords
hypertension, heart rate, non-dipping heart rate, wearables, smartwatch, wrist-worn device
Citation
Baka T and Simko F (2021) Monitoring Non-dipping Heart Rate by Consumer-Grade Wrist-Worn Devices: An Avenue for Cardiovascular Risk Assessment in Hypertension. Front. Cardiovasc. Med. 8:711417. doi: 10.3389/fcvm.2021.711417
Received
18 May 2021
Accepted
30 June 2021
Published
23 July 2021
Volume
8 - 2021
Edited by
Stefano Omboni, Istituto Italiano di Telemedicina, Italy
Reviewed by
Giacomo Pucci, University of Perugia, Italy; Philipp Marcel Scholl, University of Freiburg, Germany
Updates
Copyright
© 2021 Baka and Simko.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Fedor Simko fedor.simko@fmed.uniba.sk
This article was submitted to Cardiovascular Epidemiology and Prevention, a section of the journal Frontiers in Cardiovascular Medicine
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.