Comparative study of flow rate-and material-dependent human plasma protein adsorption on oxygenator membranes and heat exchanger materials
Provisionally accepted- 1University Hospital and Faculty of Medicine, University of Tübingen, Tübingen, Germany
- 2Ension, Inc., Butler, United States
- 3University of Tübingen, Tübingen, Baden-Württemberg, Germany
Select one of your emails
You have multiple emails registered with Frontiers:
Notify me on publication
Please enter your email address:
If you already have an account, please login
You don't have a Frontiers account ? You can register here
Artificial lungs support patients with acute or chronic lung diseases. However, complications such as the activation of blood components leading to thrombosis and inflammation limit their long-term applicability. The systematic characterization of protein adhesion events on different material parts of the oxygenators at different flow rates can shed light on the initial reaction of blood to foreign materials. Miniaturized extracorporeal circuit devices with heparin-coated gas (PMP) or heat-exchange (PET) hollow-fiber membranes were exposed to high and low flow rates. Hemocompatibility and adsorption of plasma proteins were measured after one minute to six hours using mass spectroscopy analyses. Approximately 150 to 200 different proteins were present on the membranes, with almost no variation in the 10 most abundant proteins.Protein adsorption to the membrane types did not vary to a large extent, but a decreased flow rate significantly reduced the differences in protein adsorption between both membrane types and led to the adhesion of significantly higher amounts of inhibitory proteins C1INH and α1-AT.At the higher flow rate, coagulation-associated proteins adsorbed significantly more to PET membranes, whereas complement-activating-related proteins adsorbed more on PMP membranes. Our results highlight the importance of analyzing all circuit components to understand the activation of blood components during ECMO. The primary contributor to increased protein adsorption and activation of blood components was an increased flow rate.Therefore, flow rate adjustments should ideally aim to achieve optimal oxygenation levels of around 80% while minimizing protein adsorption and blood activation during ECMO. Notably, at a low flow rate, PMP HFM exhibited a significant increase in binding of complement and inflammation inhibitors, suggesting a potential benefit of lowering the flow rate apart from the general reduction in protein adsorption.
Keywords: ECMO: extracorporeal membrane oxygenation; HFM: hollow fiber membranes, PP: polypropylene, PMP: polymethyl pentene, EBS: Ension Bioactive Surface, PEI: polyethyleneimine, NAD: nitrous acid-degraded, PVC: polyvinylchloride, TAT: thrombinantithrombin III complex
Received: 18 Feb 2025; Accepted: 28 May 2025.
Copyright: © 2025 Große-Berkenbusch, Avci-Adali, Cahalan, Cahalan, Velic, Macek, Schlensak, Wendel and Stoppelkamp. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) or licensor are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
* Correspondence: Sandra Stoppelkamp, University Hospital and Faculty of Medicine, University of Tübingen, Tübingen, Germany
Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.