- Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, DC, United States
Since its early exploration in the 1950s, intracoronary imaging, including intravascular ultrasound (IVUS), optical coherence tomography (OCT), and near-infrared spectroscopy (NIRS), has revolutionized our understanding of coronary artery disease and improved percutaneous coronary intervention outcomes. These technologies have continuously evolved, enhancing our ability to diagnose and treat CAD and ultimately leading to better patient outcomes. This review focuses on the history and key developments of IVUS, OCT, and NIRS.
Introduction
Unarguably, coronary imaging has been one of the most transformative tools on the investigation of coronary artery disease (CAD). From its inception, intracoronary imaging, with the use of miniature catheters, has continuously shed light on our understanding of the atherosclerosis disease.
In this new era of DESs, not only simple lesions but also more complex lesions were treated by PCI; in this setting, intravascular imaging has played an important role as a clinical support tool for planning and assessing the final results of PCI (1). Intracoronary imaging such as intravascular ultrasound (IVUS) and optical coherence tomography (OCT) have been evaluated in clinical trials for more than two decades and they are now recommended in USA and European guidelines (Table 1). This review offers a journey through the history of the intracoronary imaging, especially developments of IVUS and OCT (Central Illustration).

Table 1. American and European guidelines recommendation for the use of intravascular ultrasound (IVUS) and optical coherence tomography (OCT).

Central Illustration. Timeline of Intracoronary Imaging and other Invasive Cardiac Procedures. A figure shows the timeline of the development and key milestones in intracoronary imaging.
Intravascular ultrasound
It is fascinating to realize that the exploration of heart structure imaging with ultrasound catheter-based devices began in 1954, just two years after Dr. Karl Theodore Dussik presented his groundbreaking paper in 1952 on the use of ultrasound for medical purposes, specifically for examining brain tumors (2). Inge Edler and Carl Hertz introduced the concept of using ultrasound to visualize the internal cardiac structures in 1954. They showed that ultrasound technology could be applied not only for static imaging, but also for monitoring the real-time movement of the heart, marking a significant advancement in medical imaging technology (3). The concept of a probe-mounted ultrasound device dates to 1959. In that year, Dean Franklin's group at the University of Washington in Seattle described an invasive ultrasound flowmeter, but it was clamped around the aorta rather than being intravascular and it was also assessed in animal experiments (4). Subsequently, a polish physician Tomasz Cieszynski, working in Wrocław, Poland, built a small ultrasonic catheter with the goal to visualize the chambers of the canine heart (5). Following these developments, an Australian scientist George Kossoff created an 8 MHz ultrasound transducer with 2 mm diameter mounter on an 8F Cournand catheter in 1966 (6). In the early 1970s, nearly a decade after the introduction of ultrasonic catheters, Nicolaas Bom and his colleagues in Rotterdam, Netherlands, made a significant contribution by developing the first intravascular phased-array catheter. This catheter used a 32-element ultrasound transducer to evaluate human heart chambers and other internal cardiac structures (7).
After the initial enthusiasm, interest in IVUS experienced a period of stagnation and was limited to internally assessing only the heart chambers. However, the scenario changed in the mid-1980s, sparked by several key developments in cardiology, including the development of the first successful percutaneous coronary interventions (8), the first implantation of coronary stent in human (8), and a growing recognition of the limitations of x-ray imaging to provide detailed information about the lumen. For that reason, IVUS was developed further. In the early 1980s, some of the first depictions of human coronary anatomy using ultrasound were produced by Sahn et al. at the University of Arizona (9). They used 9- and 12 MHz surface ultrasound probes to scan epicardial coronary arteries from outside in patients just before undergoing coronary artery bypass grafting. Shortly after, in 1986, C. Todd Sherman at Cedars-Sinai Medical Center made the first attempt to investigate coronary vessels from the inside using flexible fiberoptic angioscopes (10).
However, it wasn't until the late 1980s and early 1990s that IVUS prototypes were more extensively developed and tested. A key figure in this phase was Paul Yock, associated with the University of California, San Francisco, and Stanford University. He developed and introduced one of the first non-commercial IVUS systems (11). Often referred to as the father of IVUS, Yock's system used a 20 MHz transducer and provided some of the earliest in vivo images of generic arterial structure. His observations notably included the three-layered appearance of muscular arteries and the early thickening of the tunica intima in atheroma formation. This phenomenon, where a dark ultrasound image appears “behind” a lesion due to its high reflective properties, was a significant discovery (12). In 1989, Jonathan Tobis and John Mallery developed a similar single 20 MHz catheter system (13). This system was tested on various vascular systems, including the coronary, iliac, femoral, and tibial arteries, both before and after balloon dilation angioplasty, with evaluations conducted post-mortem. These efforts of early applications of IVUS generated some of the initial ultrasound images of arterial atherosclerotic lesions, introducing the concept of ultrasound shadowing.
Subsequently, in the early 1990s, John Hodgson and his team finalized their results on one of the first successful “phased array” IVUS trials conducted on conscious patients (14). Although their system also used 20 MHz frequency transducers (Endosonics, Rancho Cordova, CA), their catheter was slightly larger and provided a simultaneous 360° field of view Figure, in contrast to the mechanically rotated side view of earlier models. This groundbreaking phased array model was the first of its kind to be approved for intracoronary imaging in the USA, setting a standard for decades. In 1992, Carlo Di Mario was the first to apply high-frequency (40 MHz) IVUS for the study of human vascular lesions in vitro and compared it to histologic cross-sections (15). Compared with a rotating system, the phased array design offered significant advantages. It requires a smaller access sheath and is safer for use in tortuous arteries, as a mechanically rotating catheter generated friction against the arterial wall, which could hindered the generation of clear image (16, 17).
In the late 1990s and early 2000s, newer generations of IVUS systems were developed to address the limitations of earlier devices. A significant improvement was the reduction in the size of IVUS catheters, permitting easier access to smaller and more tortuous vessels (18). During this period, higher frequency transducers ranging from 40 to 60 MHz were introduced, to enhance lateral resolution (19). In 1998, (Endosonics, Rancho Cordova, CA) made a notable contribution by producing the only Food and Drug Administration (FDA)-approved system that combined balloon-transducer system that permitted stent delivery with immediate IVUS imaging post-depolyment (20). Furthermore, Yock et al. developed an innovative prototype that combined IVUS with atherectomy device, allowing imaging of regions requiring removal in real-time (21).
Beyond traditional grayscale IVUS, the late 2000s saw the utilization of backscattered IVUS radiofrequency data for tissue characterization, leading to the introduction of Virtual Histology-IVUS in 2002 (VH-IVUS) (12). This catheter-based technology uses IVUS-generated signals to create a color-coded map based on reflected signals from the artery wall (12, 22). In following years, advancements in IVUS system continued by different companies. For instance, in 2014 FDA cleared the Polaris imaging system by (Boston Scientific, Marlborough, MA) combines IVUS with fractional flow reserve (FFR) to additionally assess coronary flow. Furthermore, the integration of IVUS catheters with near-infrared spectroscopy (NIRS) has been explored, primarily for research purposes. This combination can quantify the lipid burden within coronary plaques.
Optical coherent tomography
The concept of OCT originated in the late 1980s. Adolf Fercher presented the first two-dimensional in vivo depiction of a human eye fundus along a horizontal meridian based on white light interferometric depth scans was presented at the ICO-15 SAT conference 1990 (23). This technology was then pioneered by two independent research groups: one in Japan led by Naohiro Tanno, a professor at Yamagata University, and another in the USA led by David Huang, a professor at the Massachusetts Institute of Technology (MIT). Both groups studied OCT and patented their discoveries almost simultaneously (24, 25). In 1996, a collaborative effort between Brezinski et al. at Massachusetts General Hospital and Dr. Fujimoto and Dr. Tearney at MIT led to the development of one of the first OCT catheter prototypes (26). In the same year, James Fujimoto and Mark Brezinski conducted tests on their prototypes using segments of the human abdominal aorta postmortem, and for the first time identified atherosclerotic, calcified, and thin-walled lipid-filled plaques using their catheter system (27). However, a primary limitation of their early model was that observations were made primarily in an “in air” environment, without the presence of blood in the vessel wall. This limitation would be significant in the initial models of OCT systems.
Initial experiments correlating a limited number of excised coronary and aortic specimens with histology have demonstrated OCT's capability to resolve the microstructural features of atherosclerotic plaques (24). Furthermore, in vivo animal studies have shown the effectiveness of intravascular OCT catheters in directly imaging normal rabbit aortas and swine coronary arteries (28, 29). In a significant advancement in 2002, Guillermo Tearney presented the first results of intracoronary OCT in living patients (30). He used a prototype developed at Mass General hospital, and the OCT catheter was a non-commercial 1.06 mm (3.2 Fr) modified IVUS catheter. He compared OCT images with those generated by the traditional IVUS systems of the same arteries and commented on the significantly higher resolution that allowed for identification of fine details such as intimal hyperplasia, the boundaries between the internal and elastic lamina, thin fibrous caps, and other anatomical variations that could not be detected by IVUS. Nine months later, Yabushita et al. (27) presented the first study aimed at establishing OCT criteria for characterizing atherosclerotic plaques in vitro. This was done by correlating time domain (TD) -OCT images with histological findings.
In 1996, the ZEISS company (Jena, Germany) introduced the first OCT system (OCT 1) specifically designed for eye care (28). This initial commercial OCT system was based on the time-domain technology (29). Meanwhile, the inventors of OCT—Mark Brezinski MD, PhD, James Fujimoto PhD, and Eric Swanson MS—in collaboration with Carl Zeiss America, co-founded LightLab Imaging in 1997. LightLab Imaging has played a crucial role in the advancement of OCT technology. LightLab Imaging (Westford, MA) developed the first commercially available OCT imaging system, known as the M2/M3 TD-OCT Imaging System, along with the associated ImageWire catheter. The first generation of OCT systems was based on the TD modality, where tissue depth was determined by physically altering the distance to a reference mirror. However, this initial generation faced limitations due to the necessity of a bloodless field for imaging the arterial wall and its relatively low frame rate and pullback speed, which resulted in lower quality images (30).
In response to these challenges, newer generations of OCT systems have been developed to address many of the structural and technical limitations of first-generation systems. The concept of frequency-domain (FD) OCT, often referred to as spectral-domain OCT, was first proposed by Adolf Fercher et al. in 1995 (31). This FD system utilized a fixed mirror with a variable-wavelength light source and measured data as a function of time and wavelength (32). The first in vivo application of this technology in animals was by Seok Yun in 2006, who used FD-OCT to image the microstructure of long segments of the coronary arteries in swine (33). Tearney et al. were pioneers in testing the FD system within the coronary vasculature at the Wellman Center for Photomedicine, Massachusetts General Hospital, in 2008 (34). Subsequently, (LightLab Imaging, Westford, MA) responded to the limitations of the TD modality of the M2/M3 system by introducing a FD OCT system (C7-XR) with a Dragonfly imaging catheter. This advanced system was approved by the FDA and received CE marking in 2010. See Table 2 for a technical comparison between the M3 and C7-XR systems.

Table 2. Comparison between time domain optical coherence tomography (TD-OCT) system and frequency domain-OCT (FD-Oct).
(Lightlab Imaging, Westford, MA) was sold to Goodman Co. Ltd. in 2002. On May 19, 2010, this company was acquired by (St. Jude Medical, St Paul, MN) marking a significant milestone in the development and dissemination of OCT technology. They continue to manufacture the C7XR system, along with the IllUMIEN Optis system. The IllUMIEN Optis system has advanced to simplify OCT image processing, aiding in real-time decisions regarding stent diameter and length selection. In 2011, (St. Jude Medical, St Paul, MN) received approval for its ILUMIEN system, which innovatively combines FFR and OCT technologies.
Development of a hybrid IVUS–OCT catheter
Recently, hybrid IVUS-OCT systems were developed to merge the advantages of both modalities into a single Catheter. In 2011, Yin et al. reported a modified miniaturized probe (OCT-US) that combined OCT and IVUS (35). Following this, in 2012, Li et al. introduced another hybrid system (33). Both prototypes were tested in healthy aortas of rabbits. However, these early models have limitations that preclude their use in clinical practice. The primary issues were the large size of the catheter, inability to accurately co-register OCT and IVUS images, and increased noise in the IVUS images due to electromagnetic interference from the motor.
A significant breakthrough occurred in 2018, when Sheth et al. (34) reported the first clinical use of a hybrid IVUS-OCT catheter. This iteration showcased beautifully co-registered images and met clinically acceptable specifications regarding size, speed, and resolution. The Novasight Hybrid™ system (Conavi Medical, Toronto, Canada), represents a significant advancement in this field. It was first tested in humans in 2018, and is now clinically available in the United States and Canada (34, 35). Another notable system is the dual-sensor system (TERUMO, Tokyo, Japan), which is incorporated in the AltaView and FastView systems. AltaView has received approval from the Pharmaceuticals and Medical Devices Agency (PMDA) in Japan, whereas FastView has been approved by both the PMDA and received the CE mark. Given the unique strengths and limitations of both IVUS and OCT, these hybrid catheters are theoretically poised to offer a more comprehensive evaluation of coronary artery diseases by leveraging the advantages of both modalities. As such, it raises the question of whether these novel devices will be able to outperform traditional IVUS and OCT systems in clinical practice.
Near-infrared spectroscopy
Near-infrared spectroscopy (NIRS) represents a developing imaging modality that utilizes infrared light to characterize the lipid content of vulnerable plaques (36, 37). Observations of changes in the absorption of light have been recognized as early as 1876 (38) leading to a milestone NIRS paper by Professor Frans F. Jöbsis of Duke University in 1977 (39), who is regarded as the pioneer of medical applications of NIRS. First FDA approval for a device using NIRS technology was in 1993 for cerebral oximetry device for non-invasive detection of tissues oxygen saturation (40). However, Its vascular current utility is derived from its ability to identify high-risk lesions that may be prone to rupture and cause ensuant acute coronary syndromes (41). It was first tested on arterial systems in 1993 by Cassis and Lodder (42). It was first shown that NIRS could accurately characterize low-density lipoprotein cholesterol accumulation in hypercholesterolemic rabbit aortas. In 1996, NIRS was then used in humans to image lipid content in carotid plaques exposed at the time of surgery (43).
The early reports indicated that NIRS could be useful for characterizing plaque have been presented by Werner Jaross in 1999, who compared cholesterol content determined by NIR spectroscopy vs. that determined by reversed-phase, high-pressure liquid chromatography in human aorta specimens (44). Later in 2002, other group led by Pedro Moreno at the University of Kentucky later reported the NIRS ability to identify plaque composition in human aortic atherosclerotic plaques (45). They also reported the ability of NIRS to detect an inflamed thin-cap fibroatheroma. Pedro Moreno extended the study from aortic to coronary tissues and demonstrated the ability of NIRS to differentiate normal human coronaries tissues from diseased atherosclerotic plaques (46).
In August 2005, the new ultrafast NIRS system was tested in six patients at the Lahey Clinic by Dr. Sergio Waxman and Dr. Nesto who confirmed safety of the device (41). In 2008, Garcia-Garcia et al. (47) stated that detection of coronary vulnerable plaques in vivo is essential for studying their natural history and assessing potential treatment modalities and, therefore, may have an important impact on the prevention of acute myocardial infarction and death. At the same year on 2008, FDA gave approval for the first catheter-based LipiScan(TM) Coronary Imaging System developed by (InfraReDx, Burlington, MA). In 2009, Waxman et al. (48) reported their initial results from the SPECTACL (SPECTroscopic Assessment of Coronary Lipid) trial, a first-in-human multi-center, designed to demonstrate the applicability of the lipid core-containing plaques detection algorithm in living patients. They compared catheter based NIRS signals obtained from coronaries of patients vs. to those from autopsy specimens. The initial results confirmed feasibility and safety of the intravascular NIRS system. However, intracoronary NIRS imaging system received regulatory approval in USA (FDA approval), Europe (CE marked) and Japan (PMDA) in 2010, 2011 and 2014, respectively. In 2011, (InfraReDx, Burlington, MA) announcement the launch of its TVC Imaging System™, an enhanced version of the LipiScan™ IVUS Coronary Imaging System.
NIRS has been extensively validated for lipid-rich plaque detection against the gold-standard of histology and is the only FDA-approved method to identify coronary lipids. Several, studies have demonstrated the positive association between lipid burden in culprit lesion detected by NIRS and coronary events (49–51). In 2019, the Lipid-Rich Plaque (LRP) trial conducted by Ron Waksman demonstrated that NIRS imaging of non-obstructive territories in patients undergoing cardiac catheterization can help in identifying patients and segments at higher risk for subsequent non-culprit major adverse cardiovascular events (36). Based on the results of LRP trial, FDA announced the approval of NIRS for detection of high-risk plaques and patients in 2019. Following the LRP trial, PROSPECT II trial (52) by David Erling in 2021 reported similar results.
Following the clinical success of NIRS and its integration with IVUS, efforts have expanded to combine NIRS with other high-resolution modalities such as OCT. This trend toward multimodal imaging has led to the development of new hybrid systems. While some of these technologies remain at a very early stage and are not yet approved or commercially available, recent progress has led to the development of clinically usable systems. A notable example is the HyperVue Imaging System (SpectraWAVE, Inc), which received FDA clearance and integrates next-generation optical coherence tomography (DeepOCT) with near-infrared spectroscopy (NIRS) into a single catheter. This system offers enhanced depth penetration and resolution, enabling simultaneous plaque morphological assessment and lipid detection, co-registered with the coronary angiogram. In 2024, its first-in-human use demonstrated the feasibility of high-fidelity intracoronary imaging using this combined approach with artificial intelligence–assisted interpretation (53).
The availability of co-registered OCT–NIRS imaging represents a new frontier in coronary imaging, merging compositional and structural information in a single acquisition. This innovation may offer unprecedented precision in plaque characterization, improve PCI guidance, and ultimately support new paradigms in preventive and interventional cardiology.
Conclusion
Since its inception, intracoronary imaging techniques, such as IVUS, OCT, and NIRS, have revolutionized cardiovascular medicine. We must acknowledge the groundbreaking work of the pioneers and inventors who made this possible. These high-resolution imaging tools provide detailed plaque characterization, significantly impacting percutaneous coronary intervention (PCI) optimization, and propelling our understanding of vascular biology. Although not yet fully adopted in routine clinical practice, the potential of intracoronary imaging remains vast. The integration of artificial intelligence, innovative technologies, and next-generation hardware and software promises to unlock ideal PCI outcomes, a deeper understanding of vascular biology, and, ultimately, improved prevention and treatment strategies for coronary artery disease. In addition, several hybrid technologies—such as fluorescence lifetime imaging-OCT (FLIm-OCT), intravascular photoacoustic-IVUS (IVPA-IVUS), fluorescence lifetime imaging-IVUS (FLIm-IVUS), near-infrared fluorescence-OCT (NIRF-OCT), and near-infrared fluorescence-IVUS (NIRF-IVUS) are currently under development, further underscoring the continuous innovation and evolution in the field of intravascular imaging.
Author contributions
WA: Conceptualization, Resources, Validation, Methodology, Data curation, Formal analysis, Visualization, Funding acquisition, Investigation, Writing – review & editing, Writing – original draft, Software, Supervision, Project administration. HG-G: Project administration, Funding acquisition, Conceptualization, Validation, Investigation, Writing – review & editing, Supervision, Methodology, Visualization, Formal analysis, Software, Resources, Writing – original draft, Data curation. PR: Resources, Visualization, Software, Funding acquisition, Formal analysis, Project administration, Data curation, Investigation, Methodology, Validation, Conceptualization, Writing – review & editing, Supervision. GM: Project administration, Formal analysis, Data curation, Methodology, Visualization, Validation, Investigation, Software, Funding acquisition, Supervision, Writing – review & editing, Resources, Conceptualization. RW: Resources, Data curation, Validation, Methodology, Project administration, Formal analysis, Conceptualization, Visualization, Investigation, Writing – review & editing, Funding acquisition, Supervision, Software.
Funding
The author(s) declare that no financial support was received for the research and/or publication of this article.
Conflict of interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declare that no Generative AI was used in the creation of this manuscript.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher's note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
Abbreviations
IVUS, intravascular ultrasound; OCT, optical coherence tomography; NIRS, Near-infrared spectroscopy; TD, time domain; FD, frequency domain; FFR, fractional flow reserve.
References
1. Zhang Y, Farooq V, Garcia-Garcia HM, Bourantas CV, Tian N, Dong S, et al. Comparison of intravascular ultrasound versus angiography-guided drug-eluting stent implantation: a meta-analysis of one randomised trial and ten observational studies involving 19,619 patients. EuroIntervention. (2012) 8(7):855–65. doi: 10.4244/EIJV8I7A129
2. Franklin DL, Ellis RM, Rushmer RF. Aortic blood flow in dogs during treadmill exercise. J Appl Physiol. (1959) 14(5):809–12. doi: 10.1152/jappl.1959.14.5.809
3. Cieszynski T. [Intracardiac method for the investigation of structure of the heart with the aid of ultrasonics]. Arch Immunol Ther Exp (Warsz). (1960) 8:551–7.13693633
4. Kossoff G. Diagnostic applications of ultrasound in cardiology. Australas Radiol. (1966) 10(2):101–6. doi: 10.1111/j.1440-1673.1966.tb00774.x
5. Bom N, Lancée CT, Van Egmond FC. An ultrasonic intracardiac scanner. Ultrasonics. (1972) 10(2):72–6. doi: 10.1016/0041-624X(72)90250-8
6. Grüntzig A. Transluminal dilatation of coronary-artery stenosis. Lancet. (1978) 311(8058):263. doi: 10.1016/S0140-6736(78)90500-7
7. Sigwart U, Puel J, Mirkovitch V, Joffre F, Kappenberger L. Intravascular stents to prevent occlusion and re-stenosis after transluminal angioplasty. N Engl J Med. (1987) 316(12):701–6. doi: 10.1056/NEJM198703193161201
8. Sahn DJ, Barratt-Boyes BG, Graham K, Kerr A, Roche A, Hill D, et al. Ultrasonic imaging of the coronary arteries in open-chest humans: evaluation of coronary atherosclerotic lesions during cardiac surgery. Circulation. (1982) 66(5):1034–44. doi: 10.1161/01.CIR.66.5.1034
9. Yock PG, Linker DT, Angelsen BAJ. Two-dimensional intravascular ultrasound: technical development and initial clinical experience. J Am Soc Echocardiogr. (1989) 2(4):296–304. doi: 10.1016/S0894-7317(89)80090-2
10. Tobis JM, Mallery JA, Gessert J, Griffith J, Mahon D, Bessen M, et al. Intravascular ultrasound cross-sectional arterial imaging before and after balloon angioplasty in vitro. Circulation. (1989) 80(4):873–82. doi: 10.1161/01.CIR.80.4.873
11. Hodgson JM, Graham SP, Sheehan H, Savakus AD. Percutaneous intracoronary ultrasound imaging: initial applications in patients. Echocardiography. (1990) 7(4):403–13. doi: 10.1111/j.1540-8175.1990.tb00381.x
12. Di Mario C, The SH, Madretsma S, van Suylen RJ, Wilson RA, Bom N, et al. Detection and characterization of vascular lesions by intravascular ultrasound: an in vitro study correlated with histology. J Am Soc Echocardiogr. (1992) 5(2):135–46. doi: 10.1016/S0894-7317(14)80543-9
13. Nissen SE, Yock P. Intravascular ultrasound: novel pathophysiological insights and current clinical applications. Circulation. (2001) 103(4):604–16. doi: 10.1161/01.CIR.103.4.604
14. Sung JH, Jeong JS. Development of high-frequency (>60 MHz) intravascular ultrasound (IVUS) transducer by using asymmetric electrodes for improved beam profile. Sensors. (2018) 18(12):4414. doi: 10.3390/s18124414
15. Yock PG, Fitzgerald PJ. Intravascular ultrasound: state of the art and future directions. Am J Cardiol. (1998) 81(7A):27E–32E. doi: 10.1016/S0002-9149(98)00195-7
16. Nair A, Kuban BD, Tuzcu EM, Schoenhagen P, Nissen SE, Vince DG. Coronary plaque classification with intravascular ultrasound radiofrequency data analysis. Circulation. (2002) 106(17):2200–6. doi: 10.1161/01.CIR.0000035654.18341.5E
17. Diethrich EB, Irshad K, Reid DB. Virtual histology and color flow intravascular ultrasound in peripheral interventions. Semin Vasc Surg. (2006) 19(3):155–62. doi: 10.1053/j.semvascsurg.2006.06.001
18. Fercher AF, Mengedoht K, Werner W. Eye-length measurement by interferometry with partially coherent light. Opt Lett. (1988) 13(3):186–8. doi: 10.1364/OL.13.000186
19. Huang D, Swanson EA, Lin CP, Schuman JS, Stinson WG, Chang W, et al. Optical coherence tomography. Science. (1991) 254(5035):1178–81. doi: 10.1126/science.1957169
20. Bell GI. Lilly lecture 1990. Molecular defects in diabetes mellitus. Diabetes. (1991) 40(4):413–22. doi: 10.2337/diab.40.4.413
21. Tearney GJ, Brezinski ME, Fujimoto JG, Weissman NJ, Boppart SA, Bouma BE, et al. Scanning single-mode fiber optic catheter–endoscope for optical coherence tomography. Opt Lett. (1996) 21(7):543. doi: 10.1364/OL.21.000543
22. Brezinski ME, Tearney GJ, Bouma BE, Izatt JA, Hee MR, Swanson EA, et al. Optical coherence tomography for optical biopsy: properties and demonstration of vascular pathology. Circulation. (1996) 93(6):1206–13. doi: 10.1161/01.CIR.93.6.1206
23. Brezinski ME, Tearney GJ, Bouma BE, Boppart SA, Hee MR, Swanson EA, et al. Imaging of coronary artery microstructure (in vitro) with optical coherence tomography. Am J Cardiol. (1996) 77(1):92–3. doi: 10.1016/S0002-9149(97)89143-6
24. Tearney GJ, Jang IK, Kang DH, Aretz HT, Houser SL, Brady TJ, et al. Porcine coronary imaging in vivo by optical coherence tomography. Acta Cardiol. (2000) 55(4):233–7. doi: 10.2143/AC.55.4.2005745
25. Fujimoto JG, Boppart SA, Tearney GJ, Bouma BE, Pitris C, Brezinski ME. High resolution in vivo intra-arterial imaging with optical coherence tomography. Heart. (1999) 82(2):128–33. doi: 10.1136/hrt.82.2.128
26. Jang IK, Bouma BE, Kang DH, Park SJ, Park SW, Seung KB, et al. Visualization of coronary atherosclerotic plaques in patients using optical coherence tomography: comparison with intravascular ultrasound. J Am Coll Cardiol. (2002) 39(4):604–9. doi: 10.1016/S0735-1097(01)01799-5
27. Yabushita H, Bouma BE, Houser SL, Aretz HT, Jang IK, Schlendorf KH, et al. Characterization of human atherosclerosis by optical coherence tomography. Circulation. (2002) 106(13):1640–5. doi: 10.1161/01.CIR.0000029927.92825.F6
28. Twing AH, Meyer J, Dickens H, Young MN, Shroff A. A brief history of intracoronary imaging. Curr Cardiovasc Imaging Rep. (2020) 13(6):18. doi: 10.1007/s12410-020-09538-y
29. Fercher AF, Hitzenberger CK, Kamp G, El-Zaiat SY. Measurement of intraocular distances by backscattering spectral interferometry. Opt Commun. (1995) 117(1–2):43–8. doi: 10.1016/0030-4018(95)00119-S
30. Yun SH, Tearney GJ, Vakoc BJ, Shishkov M, Oh WY, Desjardins AE, et al. Comprehensive volumetric optical microscopy in vivo. Nat Med. (2006) 12(12):1429–33. doi: 10.1038/nm1450
31. Tearney GJ, Waxman S, Shishkov M, Vakoc BJ, Suter MJ, Freilich MI, et al. Three-dimensional coronary artery microscopy by intracoronary optical frequency domain imaging. JACC: Cardiovascular Imaging. (2008) 1(6):752–61.19356512
32. Yin J, Li X, Jing J, Li J, Mukai D, Mahon S, et al. Novel combined miniature optical coherence tomography ultrasound probe for in vivo intravascular imaging. J Biomed Opt. (2011) 16(6):060505. doi: 10.1117/1.3589097
33. Li BH, Leung ASO, Soong A, Munding CE, Lee H, Thind AS, et al. Hybrid intravascular ultrasound and optical coherence tomography catheter for imaging of coronary atherosclerosis. Cathet Cardio Intervent. (2013) 81(3):494–507.
34. Sheth TN, Pinilla-Echeverri N, Mehta SR, Courtney BK. First-in-human images of coronary atherosclerosis and coronary stents using a novel hybrid intravascular ultrasound and optical coherence tomographic catheter. JACC: Cardiovascular Interventions. (2018) 11(23):2427–30.30522674
35. Akl E, Pinilla-Echeverri N, Garcia-Garcia HM, Mehta SR, Dan K, Kuku KO, et al. First in-human evaluation of a novel intravascular ultrasound and optical coherence tomography system for intracoronary imaging. Catheter Cardiovasc Interv. (2022) 99(3):686–98.34792273
36. Waksman R, Di Mario C, Torguson R, Ali ZA, Singh V, Skinner WH, et al. Identification of patients and plaques vulnerable to future coronary events with near-infrared spectroscopy intravascular ultrasound imaging: a prospective, cohort study. Lancet. (2019) 394(10209):1629–37. doi: 10.1016/S0140-6736(19)31794-5
37. Negi SI, Didier R, Ota H, Magalhaes MA, Popma CJ, Kollmer MR, et al. Role of near-infrared spectroscopy in intravascular coronary imaging. Cardiovasc Revasc Med. (2015) 16(5):299–305. doi: 10.1016/j.carrev.2015.06.001
38. Vierordt Kv. Die Quantitative Spectralanalyse in Ihrer Anwendung auf Physiologie, Physik, Chemie und Technologie. Tübingen: H. Laupp (1876).
39. Jöbsis FF. Noninvasive, infrared monitoring of cerebral and myocardial oxygen sufficiency and circulatory parameters. Science. (1977) 198(4323):1264–7. doi: 10.1126/science.929199
40. Murkin JM, Arango M. Near-infrared spectroscopy as an index of brain and tissue oxygenation. Br J Anaesth. (2009) 103(Suppl 1):i3–13. doi: 10.1093/bja/aep299
41. Caplan JD, Waxman S, Nesto RW, Muller JE. Near-infrared spectroscopy for the detection of vulnerable coronary artery plaques. J Am Coll Cardiol. (2006) 47(8):C92–6. doi: 10.1016/j.jacc.2005.12.045
42. Cassis LA, Lodder RA. Near-IR imaging of atheromas in living arterial tissue. Anal Chem. (1993) 65(9):1247–56. doi: 10.1021/ac00057a023
43. Dempsey RJ, Davis DG, Buice RG, Lodder RA. Biological and medical applications of near-infrared spectrometry. Appl Spectrosc. (1996) 50(2):18A–34A. doi: 10.1366/0003702963906537
44. Jaross W, Neumeister V, Lattke P, Schuh D. Determination of cholesterol in atherosclerotic plaques using near infrared diffuse reflection spectroscopy. Atherosclerosis. (1999) 147(2):327–37. doi: 10.1016/S0021-9150(99)00203-8
45. Moreno PR, Lodder RA, Purushothaman KR, Charash WE, O’Connor WN, Muller JE. Detection of lipid pool, thin fibrous cap, and inflammatory cells in human aortic atherosclerotic plaques by near-infrared spectroscopy. Circulation. (2002) 105(8):923–7. doi: 10.1161/hc0802.104291
46. Gardner CM, Tan H, Hull EL, Lisauskas JB, Sum ST, Meese TM, et al. Detection of lipid core coronary plaques in autopsy specimens with a novel catheter-based near-infrared spectroscopy system. JACC Cardiovasc Imaging. (2008) 1(5):638–48. doi: 10.1016/j.jcmg.2008.06.001
47. García-García HM, Gonzalo N, Granada JF, Regar E, Serruys PW. Diagnosis and treatment of coronary vulnerable plaques. Expert Rev Cardiovasc Ther. (2008) 6(2):209–22. doi: 10.1586/14779072.6.2.209
48. Waxman S, Dixon SR, L’Allier P, Moses JW, Petersen JL, Cutlip D, et al. In vivo validation of a catheter-based near-infrared spectroscopy system for detection of lipid core coronary plaques. JACC: Cardiovascular Imaging. (2009) 2(7):858–68.19608137
49. Danek BA, Karatasakis A, Karacsonyi J, Alame A, Resendes E, Kalsaria P, et al. Long-term follow-up after near-infrared spectroscopy coronary imaging: insights from the lipid cORe plaque association with CLinical events (ORACLE-NIRS) registry. Cardiovasc Revasc Med. (2017) 18(3):177–81. doi: 10.1016/j.carrev.2016.12.006
50. Madder RD, Husaini M, Davis AT, VanOosterhout S, Khan M, Wohns D, et al. Large lipid-rich coronary plaques detected by near-infrared spectroscopy at non-stented sites in the target artery identify patients likely to experience future major adverse cardiovascular events. Eur Heart J Cardiovasc Imaging. (2016) 17(4):393–9. doi: 10.1093/ehjci/jev340
51. Oemrawsingh RM, Cheng JM, García-García HM, van Geuns RJ, de Boer SPM, Simsek C, et al. Near-infrared spectroscopy predicts cardiovascular outcome in patients with coronary artery disease. J Am Coll Cardiol. (2014) 64(23):2510–8. doi: 10.1016/j.jacc.2014.07.998
52. Erlinge D, Maehara A, Ben-Yehuda O, Bøtker HE, Maeng M, Kjøller-Hansen L, et al. Identification of vulnerable plaques and patients by intracoronary near-infrared spectroscopy and ultrasound (PROSPECT II): a prospective natural history study. Lancet. (2021) 397(10278):985–95. doi: 10.1016/S0140-6736(21)00249-X
Keywords: IVUS (intravascular ultrasound), optical coherence tomography, NIRS (near infrared spectroscopy), intravascular imaging, coronary artery disease
Citation: Abusnina W, Garcia-Garcia HM, Rubio PM, Mintz GS and Waksman R (2025) Coronary intravascular imaging: a piece of history. Front. Cardiovasc. Med. 12:1632513. doi: 10.3389/fcvm.2025.1632513
Received: 4 June 2025; Accepted: 21 July 2025;
Published: 30 September 2025.
Edited by:
Christos Bourantas, Queen Mary University of London, United KingdomReviewed by:
Nicoleta-Monica Popa-Fotea, Carol Davila University of Medicine and Pharmacy, RomaniaEmrah Erdogan, Yüzüncü Yıl University, Türkiye
Copyright: © 2025 Abusnina, Garcia-Garcia, Rubio, Mintz and Waksman. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Hector M. Garcia-Garcia, aGVjdG9yLm0uZ2FyY2lhZ2FyY2lhQG1lZHN0YXIubmV0
†These authors have contributed equally to this work.