Editorial on the Research Topic
Photon counting CT technology in cardiovascular imaging
Since its clinical introduction in 2021, photon-counting detector CT (PCD-CT) has demonstrated significant advantages in the evaluation of cardiovascular disease, with most supporting evidence to date derived from the only commercially available whole-body dual-source PCD-CT system (1). This technology inherently provides spectral imaging capabilities for all scans, enhances spatial resolution, and reduces electronic noise (2–6). These improvements translate into better detection of stenosis, fewer referrals for invasive coronary angiography due to superior image quality and reduced calcium blooming artifacts—especially when using ultra-high-resolution modes—compared to conventional energy-integrating detector CT (EID-CT). However, whether PCD-CT can also improve the quantitative assessment of coronary plaques remains an open question. This summary presents recent findings on the application of PCD-CT for advanced cardiac evaluation.
Mergen et al. investigated the impact of ultra-high-resolution CT angiography on the quantitative characterization of coronary plaques. Reconstructions using the Bv40 kernel with a slice thickness of 0.6 mm were used as the reference standard. Using different kernel and slice thickness settings, they found substantial changes between the standard vs. ultra-high-resolution reconstructions using a 0.2 mm slice thickness and the Bv64 kernel (p < 0.001). The findings of this study highlight the importance of determining the optimal kernel settings—potentially validated through invasive methods—and understanding the impact of using different reconstruction parameters, as these not only affect the total plaque volume but also alter the assessed plaque composition. The proportion of calcified plaque component was 85.1% in standard reconstructions but dropped to 75.2% with ultra-high-resolution imaging, while lipid-rich plaque component increased to 6.7%. These results suggest that utilizing ultra-high-resolution PCD-CT reduces blooming artifacts for improved plaque delineation and characterization (3, 7).
Vascular inflammation plays a key role in the development of atherosclerosis and the rupture of atherosclerotic plaques, ultimately leading to adverse events (8). Kahmann et al. explored the relationship between pericoronary adipose tissue (PCAT) composition and coronary artery disease (CAD) in patients with plaques in the left coronary artery. PCAT has recently emerged as a promising non-invasive marker of vascular inflammation as assessed by CT, although its characteristics can vary widely depending on patient- and imaging-related factors. This analysis revealed that certain texture features significantly distinguished CAD patients from non-CAD individuals. Similar texture changes were also found in the right coronary artery (RCA), suggesting a systemic CAD effect on PCAT. These findings imply that PCAT texture alterations can be detected before plaque formation, offering potential for early CAD risk assessment and highlighting the value of advanced imaging techniques in risk stratification and management.
A recently published study by Kahmann et al. applied radiomics texture analysis on PCD-CT to detect subtle tissue differences associated with hypercholesterolemia, offering a potential imaging biomarker for the future. Six radiomic features differentiated those with and without hypercholesterolemia independently of coronary calcification or stenosis severity. Patients with hypercholesterolemia exhibited higher average PCAT attenuation (−97.1 HU vs. −100.0 HU) and increased dense tissue markers such as High Gray-Level Emphasis and High Gray-Level Run Emphasis. Conversely, High Gray-Level Zone Emphasis was lower in the hypercholesterolemia group, suggesting a more homogeneously dense tissue distribution, potentially reflecting low-grade inflammation. These results, validated in an independent cohort, indicate that PCAT radiomics on PCD-CT can detect subclinical tissue remodeling linked to lipid disorders, opening avenues for non-invasive early cardiovascular risk biomarkers beyond plaque morphology.
Coronary calcium scoring (CACS) is a widely used risk assessment tool guiding statin therapy, yet studies comparing PCD-CT to traditional EID-CT for CACS are limited. Wolf et al. directly compared CACS between PCD-CT and EID-CT in 23 patients and a phantom model. Using standard protocols, PCD-CT yielded systematically lower Agatston scores than EID-CT (median 174.3 vs. 218.2; mean bias −41.1). Phantom experiments showed PCD-CT more accurately estimated calcium volume (66.1% overestimation vs. 77.2% for EID-CT; p = 0.0015), with excellent correlation between modalities (r = 0.99). Simulations in a larger cohort suggested that PCD-CT scoring would reclassify about 5.25% of patients to a lower cardiovascular risk category. Overall, PCD-CT offers more precise physical quantification of coronary calcifications, but caution is needed when interpreting Agatston-based risk stratification across different scanner types.
While PCD-CT enhances diagnostic accuracy for coronary artery disease assessment, especially in high-risk populations, further studies are required to determine whether these imaging improvements translate into better long-term prognosis, more accurate risk stratification based on plaque characteristics, improved correlation with invasive plaque assessment, or enhanced evaluation of coronary stents.
Author contributions
BS: Writing – original draft, Writing – review & editing. ÁV-S: Writing – review & editing, Writing – original draft. AP: Writing – original draft, Writing – review & editing. FS: Writing – review & editing, Writing – original draft.
Conflict of interest
BS and FS Speaker’s honoraria and speaker travel support from Siemens Healthineers. ÁV-S Institutional research support, speaker fees, and travel support from Siemens, institutional research support from Bayer, institutional research support, consulting fees from and shareholder at Elucid Bioimaging.
The remaining author declares that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
The author(s) declared that they were an editorial board member of Frontiers, at the time of submission. This had no impact on the peer review process and the final decision.
Generative AI statement
The author(s) declare that no Generative AI was used in the creation of this manuscript.
Publisher's note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
1. Schwartz FR, Sodickson AD, Pickhardt PJ, Sahani DV, Lev MH, Gupta R. Photon-counting CT: technology, current and potential future clinical applications, and overview of approved systems and those in Various stages of research and development. Radiology. (2025) 314(3):e240662. doi: 10.1148/radiol.240662
2. Pourmorteza A, Choux AR, Holmes TW, Schoepf UJ, van Assen M, De Cecco C, et al. Beam hardening of K-edge contrast agents: a phantom study comparing clinical energy-integrating detector and photon-counting detector CT systems. Eur Radiol Exp. (2025) 9(1):31. doi: 10.1186/s41747-024-00530-5
3. Vattay B, Boussoussou M, Vecsey-Nagy M, Kolossvary M, Juhasz D, Kerkovits N, et al. Qualitative and quantitative image quality of coronary CT angiography using photon-counting computed tomography: standard and ultra-high resolution protocols. Eur J Radiol. (2024) 175:111426. doi: 10.1016/j.ejrad.2024.111426
4. Vattay B, Szilveszter B, Boussoussou M, Vecsey-Nagy M, Lin A, Konkoly G, et al. Impact of virtual monoenergetic levels on coronary plaque volume components using photon-counting computed tomography. Eur Radiol. (2023) 33(12):8528–39. doi: 10.1007/s00330-023-09876-7
5. Vecsey-Nagy M, Tremamunno G, Schoepf UJ, Gnasso C, Zsarnoczay E, Fink N, et al. Intraindividual comparison of ultrahigh-spatial-resolution photon-counting detector CT and energy-integrating detector CT for coronary stenosis measurement. Circ Cardiovasc Imaging. (2024) 17(10):e017112. doi: 10.1161/CIRCIMAGING.124.017112
6. Vecsey-Nagy M, Varga-Szemes A, Schoepf UJ, Tremamunno G, Fink N, Zsarnoczay E, et al. Ultra-high resolution coronary CT angiography on photon-counting detector CT: bi-centre study on the impact of quantum iterative reconstruction on image quality and accuracy of stenosis measurements. Eur J Radiol. (2024) 176:111517. doi: 10.1016/j.ejrad.2024.111517
7. Vecsey-Nagy M, Tremamunno G, Schoepf UJ, Gnasso C, Zsarnoczay E, Fink N, et al. Coronary plaque quantification with ultrahigh-spatial-resolution photon-counting detector CT: intraindividual comparison with energy-integrating detector CT. Radiology. (2025) 314(3):e241479. doi: 10.1148/radiol.241479
Keywords: photon-counting CT, coronary artery disease, calcium score, phantom study, pericoronary adipose tissue (PCAT)
Citation: Szilveszter B, Varga-Szemes Á, Pourmorteza A and Schwartz FR (2025) Editorial: Photon counting CT technology in cardiovascular imaging. Front. Cardiovasc. Med. 12:1641175. doi: 10.3389/fcvm.2025.1641175
Received: 4 June 2025; Accepted: 10 June 2025;
Published: 27 June 2025.
Edited and Reviewed by: Christos Bourantas, Queen Mary University of London, United Kingdom
Copyright: © 2025 Szilveszter, Varga-Szemes, Pourmorteza and Schwartz. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Bálint Szilveszter, c3ppbHZlc3p0ZXIuYmFsaW50QGdtYWlsLmNvbQ==