REVIEW article

Front. Cardiovasc. Med., 24 August 2021

Sec. Cardiovascular Imaging

Volume 8 - 2021 | https://doi.org/10.3389/fcvm.2021.729786

Imaging of Cardiac Device-Related Infection

  • Department of Cardiovascular Medicine, Cleveland Clinic Foundation, Cleveland, OH, United States

Article metrics

View details

15

Citations

7k

Views

2,2k

Downloads

Abstract

Cardiac devices are frequently used in different cardiovascular conditions for the purpose of morbidity or mortality prevention. These include cardiac implantable electronic devices (CIED) like permanent pacemakers and implantable cardiac defibrillators, ventricular assistance devices (VADs), left atrial appendage occlusion (LAAO) devices like the Watchman™, atrial and ventricular septal occluders like the Amplatzer™, among others. In the past years, there has been an increase in the development of these devices as a result of a rise in the number of indications for implantation, paired with the aging and more medically complex patient population. This has led to an increase in the incidence of cardiac device-related infections, one of the most feared and serious complications which is associated with significant morbidity, mortality and financial burden. Accurate diagnosis of cardiac device-related infections is essential given the management implications which often involve removal of the infected device, removal of other prosthetic material and long-term antimicrobial therapy. Clinical and laboratory data are useful diagnostic tools but multimodality imaging is often necessary. The recently published 2020 European Heart Rhythm Association International Consensus document, which is endorsed by many expert societies, has recommended the use of multimodality imaging for the diagnosis of CIED infections. (1) This allows better disease characterization by identifying abnormal fluid collections and guiding aspiration for both diagnostic and therapeutic purposes (i.e. soft tissue ultrasound and computed tomography), evaluation for local extent of disease (i.e. transesophageal echocardiogram to evaluate for concomitant infective endocarditis), embolic manifestation of disease (i.e. computed tomography and magnetic resonance imaging) and metabolic tissue characterization (positron emission tomography and tagged white blood cell scan). (2) In addition, computed tomography (CT) allows for pre-procedural planning which has shown to be associated with better procedural outcomes.

Introduction

Cardiac devices are frequently used in different cardiovascular conditions for the purpose of morbidity or mortality prevention. These include cardiac implantable electronic devices (CIED) like permanent pacemakers and implantable cardiac defibrillators, ventricular assistance devices (VADs), left atrial appendage occlusion (LAAO) devices like the Watchman™, atrial and ventricular septal occluders like the Amplatzer™, among others. In the past years, there has been an increase in the development of these devices as a result of a rise in the number of indications for implantation, paired with the aging and more medically complex patient population. This has led to an increase in the incidence of cardiac device-related infections, one of the most feared and serious complications which is associated with significant morbidity, mortality and financial burden.

Accurate diagnosis of cardiac device-related infections is essential given the management implications which often involve removal of the infected device, removal of other prosthetic material and long-term antimicrobial therapy. Clinical and laboratory data are useful diagnostic tools but multimodality imaging is often necessary. The recently published 2020 European Heart Rhythm Association International Consensus document, which is endorsed by many expert societies, has recommended the use of multimodality imaging for the diagnosis of CIED infections (1). Some of these modalities include: (1) soft tissue ultrasound which can identify abnormal fluid collections and guide aspiration procedures, (2) transesophageal echocardiogram which can identify concomitant endocarditis with CIED infection, (3) computed tomography which can evaluate local and distant extension of disease as well as allow pre-procedural planning, (4) magnetic resonance imaging which provides information on local and embolic manifestation of disease, (5) positron emission tomography (PET) which can identify metabolically active foci and assist on the diagnosis of infection, and (6) tagged white blood cell scan which similarly to PET can evaluate areas of inflammation suggestive of infection (2).

In this article we will review the utility of multimodality imaging for the diagnosis of cardiac device-related infections using a number of clinical cases to demonstrate its practical value.

CIED Infection

CIEDs include implantable cardioverter defibrillators (ICDs), permanent pacemakers (PPM), and biventricular pacemakers which provide cardiac resynchronization therapy with or without a defibrillator (CRTs). Recently, novel devices including leadless pacemaker, subcutaneous ICD (SC-ICD) and implantable loop recorders (ILR) have been developed in an attempt to decrease the risk of infection but are not protected from infection. CIED infections are typically classified as pocket infections or systemic infections. Systemic infection may co-exist with cardiac valve infection, in which case the term CIED-related infective endocarditis (CIED-IE) is utilized (3).

Pocket Infection

Pocket infection is the most common type of CIED infection defined as infection limited to the generator pocket (4). Management typically involves entire CIED extraction (generator and transvenous leads) unless the infection is identified as early superficial site infection which can be managed with a course of antibiotics (5). Diagnosis of pocket infection is usually made clinically when there is presence of symptoms related to local inflammation. Soft tissue ultrasound can be done to assess for collections, although this is often not necessary. Pocket infection occasionally can be associated with CIED systemic infections and/or endocarditis (1). When the diagnosis pocket infection remains unclear or is equivocal, 18F-fluorodeoxyglucose positron emission tomography (18F-FDG PET) can provide important evidence. F18-FDG PET has a high diagnostic accuracy in the early diagnosis of pocket infection and is helpful in differentiating superficial from systemic infection. A recent meta-analysis showed very good sensitivity and specificity (93–96% and 97–98% respectively) for diagnosis of pocket infection which was reported to be higher than that for lead infection and endocarditis (68). The 2020 European Heart Rhythm Association International Consensus document recommend use of 18F-FDG PET-CT or WBC SPECT/CT in patients with positive blood cultures and clinically negative pocket and in patients with negative blood cultures with clinically negative pocket but high clinical suspicion (1) (Figure 1).

Figure 1

In cases where there is any concern of infection involvement beyond the superficial pocket site, as evidenced by positive blood cultures, systemic inflammatory response syndrome, purulent drainage and/or site erosion, one should suspect lead infection and further imaging work up with TTE and TEE is indicated to determine extent of infection (1, 9).

CIED Systemic Infection With and Without IE

The diagnosis of CIED systemic infection is suspected in the presence of unexplained fever, systemic symptoms, positive blood cultures and/or embolic phenomena (10). Diagnosis is typically confirmed with echocardiogram showing lead vegetations (1, 11). However, it is important to recognize that up to 13% of patients with CIED have asymptomatic lead masses, which often represent fibrin sheath and do not increase the risk of infection. Thus, the clinical presentation in such cases should be carefully evaluated and serial echocardiograms may be warranted (12). TEE has better sensitivity compared to TTE in detecting CIED infections (90 vs 22–43%). In addition, TEE allows better examination of the lead in the superior vena cava (1, 1315) (Figure 2).

Figure 2

CIED-IE most commonly affects the tricuspid valve (3, 16). This is particular important to diagnose because of its management implications that might require open heart surgery for system extraction and valvular intervention in some situations (i.e large >2cm vegetation and/or pulmonary embolism) versus percutaneoussystem extraction (5). In cases requiring system extraction only, the use of CT for pre-procedural planning has become a great asset to define adherence of leads to surrounding vasculature which would require more specialized equipment and more experienced operators for successful system extraction (17) (Figure 3).

Figure 3

In a small percentage of patients with suspected infection and non-diagnostic TEE or when TEE cannot be performed, alternative imaging methods are indicated (5, 11). A functional approach based on nuclear imaging using 18F-FDG PET-CT or radiolabeled WBC scintigraphy, has been incorporated to the 2015 ESC guidelines for management of endocarditis and the most recent 2020 European Heart Rhythm Association consensus for management of CIED infections (1, 18). The performance of 18F-FDG PET-CT in CIED infections was studied in a meta analysis by Juneau et al. and showed its high diagnostic accuracy with 87% sensitivity and 95% specificity for the diagnosis of IE (6). In addition, 18F-FDG PET-CT has the advantage of allowing imaging of multiple sites of possible infection (pocket/generator, leads) in one examination with a high diagnostic yield for septic emboli to the lungs or very rarely the bone structures which has management implications (i.e. need for longer antibiotic course with good bone penetrance in spondylodiscititis) (19, 20) (Figure 4).

Figure 4

Caution must be used when interpreting 18F-FDG PET-CT in early CIEDs (<2 months) as physiologic FDG uptake can be present without the presence of infection. Other disadvantages of using 18F-FDG PET-CT are high cost, not widely available, radiation, patient preparation and need for trained personal.

Subcutaneous ICD (S-ICD)

The S-ICD involves no hardware exposure to the intravascular system. Pocket infections are rare, found in 1.7% of patients. Systemic infections are extremely rare (21, 22). Given its recent development, S-ICD infections have limited management data with only a few case reports found in the literature (23). Management is suggested in an individual basis based on extent of infection in which case these have been managed in a similar way to traditional pocket infections.

Long-Term Mechanical Circulatory Support Devices Infection

Continuous flow ventricular support devices such as left ventricular assist devices (LVADs) and less frequently biventricular assist devices (BIVADs) are increasingly used for management of end-stage heart failure as bridge to transplantation (BTT) or destination therapy (DT) (24). As in CIEDs infection, LVADs infection is a serious complication that often involves prolong hospitalization, extended antibiotic and sometimes device extraction. For continuous-flow LVADs, infection occurs in 19–39% of patients and results in >10% of LVAD-related deaths (25).

LVAD infections are classified into two groups: LVAD-specific infections such as pump pocket infection (PPI), cannula infection and driveline infection (DLI); and VAD-related infections (endocarditis, bloodstream infection and mediastinitis). Diagnostic work up of VAD-specific and VAD-related infections remains challenging given lack of randomized data. Thus, current recommendations are based on expert consensus as published in the International Society of Heart and Lung Transplantation (ISHLT) guidelines for identification and management of LVADs infection (26).

LVAD DLI is the most common LVAD-specific infection because the driveline exit creates an entry tract for bacteria that can attach to prosthetic material with bacterial biofilm. DLI is sub-divided into deep and superficial. Both types involve soft tissue surrounding the driveline exit, whereas deep infections also involve the fascia and muscle layers (27). No imaging modality can definitely exclude deep tissue infection, thus imaging in diagnosis of LVAD infections remains a challenge and is based mostly on observational data (28). Soft tissue ultrasound can help identify fluid collections around the driveline but cannot rule out deeper infections. TTE and TEE are often not useful in this setting unless concomitant endocarditis is suspected. CT is commonly used given the ability to detect deeper DLI demonstrated by the presence of abscess and fat stranding but can be limited by metal device artifact. The sensitivity and specificity of CT has not yet been established.

18F-FDG PET-CT and tagged WBC scan have emerged as more accurate modalities for detection of LVAD infection based on mostly small non-randomized studies (2931). Correlating the anatomic findings, for example an abscess with the metabolic information is particularly useful. A major limitation of 18F-FDG PET-CT is the presence of non-specific uptake that can be present not only in early post-cardiac surgery but even years after LVAD implantation described as a post-operative inflammatory response. This tends to be characterized by a lesser degree of uptake and the reading clinician should become familiar with this presentation. Another consideration is the normal uptake pattern in patients with prosthetic materials due to foreign-body chronic low-grade inflammation. This is frequently seen when Dacron material is present in the LVAD outflow. Vaugelade et al. considered that a circumferential and homogeneous 18F-FDG uptake was a negative pattern for infection, defining positivity as a focal 18F-FDG uptake observed on both attenuation-corrected and uncorrected images in order to avoid overcorrection artifacts (31) (Figures 5, 6).

Figure 5

Figure 6

LAAO Device Infection (Watchman™ Infection)

Watchman™ device infection is very rare, with only a few cases reported in the literature (32). The incidence is estimated to be much lower than other CIED infections given that endothelization of the device is expected within 45 days of implantation. However lack of endothelization has been described and this may be associated with higher risk of infection (33). The approach to Watchman® device infection should be based on current infective endocarditis guidelines; which includes obtaining blood cultures and TEE when there is suspicion for infection (Figure 7).

Figure 7

Septal Occluder Infection

Infection of septal occluder devices is uncommon and reported as the least common complications of this procedure. Like with LAAO occlusion, only case report literature is available with as little as 3 cases of septal occluder device infection reported in the literature (34). Infection of these devices has been classified as early (within 6 months of implantation, prior to endothelization) which are thought to be secondary to introduction of microorganisms during the initial procedure; and late, when infection occurs secondary to seeding of microorganisms from distant sources like in IE. In the presence of infection involving septal occluder devices, prolonged antimicrobial therapy with constant blood culture monitoring has been reported to be adequate in the absence of device dehiscence, septal perforation or fistula formation.

Conclusion

The use of cardiac implantable electronic devices is increasing. These include permanent pacemakers and implantable cardiac defibrillators, ventricular assistance devices (VADs), left atrial appendage occlusion (LAAO) devices, atrial and ventricular septal occluders, among others. Despite its overall safety profile, CIED infection can occur and is associated with significant morbidity, mortality and financial burden. Multimodality imaging including TTE, TEE, CT, 18F-FDG PET-CT and tagged WBC can assist in establishing a diagnosis, guiding therapy and providing the patient and physician with prognostic information.

Publisher's Note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Statements

Author contributions

JA contributed in writing manuscript and collecting images. EH and WJ contributed in reviewing and making changes in manuscript and also collecting images for cases. All authors contributed to manuscript revision, read, and approved the submitted version.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

References

  • 1.

    Blomström-LundqvistCTraykovVErbaPABurriHNielsenJCBongiorniMGet al. European Heart Rhythm Association (EHRA) international consensus document on how to prevent, diagnose, and treat cardiac implantable electronic device infections-endorsed by the Heart Rhythm Society (HRS), the Asia Pacific Heart Rhythm Society (APHRS), the Latin American Heart Rhythm Society (LAHRS), International Society for Cardiovascular Infectious Diseases (ISCVID), and the European Society of Clinical Microbiology and Infectious Diseases (ESCMID) in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS). Eur Heart J. (2020) 41:201232. 10.1093/eurheartj/ehaa010

  • 2.

    De WinterFVogelaersDGemmelFDierckxRA. Promising role of 18-F-fluoro-D-deoxyglucose positron emission tomography in clinical infectious diseases. Eur J Clin Microbiol Infect Dis. (2002) 21:24757. 10.1007/s10096-002-0708-2

  • 3.

    SohailMRUslanDZKhanAHFriedmanPAHayesDLWilsonWRet al. Infective endocarditis complicating permanent pacemaker and implantable cardioverter-defibrillator infection. Mayo Clin Proc. (2008) 83:4653. 10.4065/83.1.46

  • 4.

    DaiMCaiCVaibhavVSohailMRHayesDLHodgeDOet al. Trends of cardiovascular implantable electronic device infection in 3 decades: a population-based study. JACC Clin Electrophysiol. (2019) 5:107180. 10.1016/j.jacep.2019.06.016

  • 5.

    KusumotoFMSchoenfeldMHWilkoffBLBerulCIBirgersdotter-GreenUMCarrilloRet al. 2017 HRS expert consensus statement on cardiovascular implantable electronic device lead management and extraction. Heart Rhythm. (2017) 14:e50351. 10.1016/j.hrthm.2017.09.001

  • 6.

    JuneauDGolfamMHazraSZuckierLSGarasSRedpathCet al. Positron emission tomography and single-photon emission computed tomography imaging in the diagnosis of cardiac implantable electronic device infection: a systematic review and meta-analysis. Circ Cardiovasc Imaging. (2017) 10:e005772. 10.1161/CIRCIMAGING.116.005772

  • 7.

    AhmedFZJamesJCunningtonCMotwaniMFullwoodCHooperJet al. Early diagnosis of cardiac implantable electronic device generator pocket infection using 18F-FDG-PET/CT. Eur Heart J Cardiovasc Imaging. (2015) 16:52130. 10.1093/ehjci/jeu295

  • 8.

    CantoniVSolliniMGreenRBerchiolliRLazzeriEMannarinoTet al. Comprehensive meta-analysis on [18F] FDG PET/CT and radio labeled leukocyte SPECT–SPECT/CT imaging in infectious endocarditis and cardiovascular implantable electronic device infections. Clin Transl Imaging. (2018) 6:318. 10.1007/s40336-018-0265-z

  • 9.

    Esquer GarrigosZGeorgeMPKhalilSVijayvargiyaPAbu SalehOMFriedmanPAet al. Predictors of bloodstream infection in patients presenting with cardiovascular implantable electronic device pocket infection. Open Forum Infect Dis. (2019) 6:ofz084. 10.1093/ofid/ofz084

  • 10.

    KlugDLacroixDSavoyeCGoullardLGrandmouginDHennequinJLet al. Systemic infection related to endocarditis on pacemaker leads: clinical presentation and management. Circulation. (1997) 95:2098107. 10.1161/01.CIR.95.8.2098

  • 11.

    BaddourLMBettmannMABolgerAFEpsteinAEFerrieriPGerberMAet al. Nonvalvular cardiovascular device–related infections. Circulation. (2003) 108:201531. 10.1161/01.CIR.0000093201.57771.47

  • 12.

    GolzioPGErrigoDPeyracchiaMGalloEFreaSCastagnoDet al. Prevalence and prognosis of lead masses in patients with cardiac implantable electronic devices without infection. J Cardiovasc Med. (2019) 20:3728. 10.2459/JCM.0000000000000797

  • 13.

    VilacostaISarriáCRománJASJiménezJCastilloJAIturraldeEet al. Usefulness of transesophageal echocardiography for diagnosis of infected transvenous permanent pacemakers. Circulation. (1994) 89:26847. 10.1161/01.CIR.89.6.2684

  • 14.

    MassourePLReuterSLafitteSLaborderieJBordachardPClementyJet al. Pacemaker endocarditis: clinical features and management of 60 consecutive cases. Pacing Clin Electrophysiol. (2007) 30:129. 10.1111/j.1540-8159.2007.00574.x

  • 15.

    VictorFDe PlaceCCamusCLe BretonHLeclercqCPavinDet al. Pacemaker lead infection: echocardiographic features, management, and outcome. Heart. (1999) 81:827. 10.1136/hrt.81.1.82

  • 16.

    DuvalXSelton-SutyCAllaFSalvador-MazenqMBernardYWeberMet al. Endocarditis in patients with a permanent pacemaker: a 1-year epidemiological survey on infective endocarditis due to valvular and/or pacemaker infection. Clin Infect Dis. (2004) 39:6874. 10.1086/421493

  • 17.

    PatelDSripariwuthAAbozeedMHusseinAATarakjiKGWazniOMet al. Lead location as assessed on cardiac computed tomography and difficulty of percutaneous transvenous extraction. JACC Clin Electrophysiol. (2019) 5:14328. 10.1016/j.jacep.2019.07.018

  • 18.

    HabibGLancellottiPAntunesMJBongiorniMGCasaltaJ-PDel ZottiFet al. 2015 ESC Guidelines for the management of infective endocarditis: the task force for the management of infective endocarditis of the European Society of Cardiology (ESC) endorsed by: European Association for Cardio-Thoracic Surgery (EACTS), the European Association of Nuclear Medicine (EANM). Eur Heart J. (2015) 36:3075128. 10.1093/eurheartj/ehv319

  • 19.

    AmraouiSTliliGSohalMBerteBHindiéERitterPet al. Contribution of PET imaging to the diagnosis of septic embolism in patients with pacing lead endocarditis. JACC Cardiovasc Imaging. (2016) 9:28390. 10.1016/j.jcmg.2015.09.014

  • 20.

    CautelaJAlessandriniSCammilleriSGiorgiRRichetHCasaltaJ-Pet al. Diagnostic yield of FDG positron-emission tomography/computed tomography in patients with CEID infection: a pilot study. EP Europace. (2012) 15:2527. 10.1093/europace/eus335

  • 21.

    BoersmaLBarrCKnopsRTheunsDEckardtLNeuzilPet al. Implant and midterm outcomes of the subcutaneous implantable cardioverter–defibrillator registry: the EFFORTLESS study. J Am Coll Cardiol. (2017) 70:83041. 10.1016/j.jacc.2017.06.040

  • 22.

    CalvagnaGMCeresaFPatanèS. Pocket infection as a complication of a subcutaneous implantable cardioverter–defibrillator. Int J Cardiol. (2014) 177:6168. 10.1016/j.ijcard.2014.09.119

  • 23.

    LambiasePDBarrCTheunsDAKnopsRNeuzilPJohansenJBet al. Worldwide experience with a totally subcutaneous implantable defibrillator: early results from the EFFORTLESS S-ICD registry. Eur Heart J. (2014) 35:165765. 10.1093/eurheartj/ehu112

  • 24.

    YancyCWJessupMBozkurtBButlerJCaseyDEJrDraznerMHet al. 2013 ACCF/AHA guideline for the management of heart failure: executive summary: a report of the American College of Cardiology Foundation/American Heart Association Task Force on practice guidelines. Circulation. (2013) 128:181052. 10.1161/CIR.0b013e31829e8807

  • 25.

    TopkaraVKKondareddySMalikFWangIWMannDLEwaldGAet al. Infectious complications in patients with left ventricular assist device: etiology and outcomes in the continuous-flow era. Ann Thorac Surg. (2010) 90:12707. 10.1016/j.athoracsur.2010.04.093

  • 26.

    SlaughterMSPaganiFDRogersJGMillerLWSunBRussellSDet al. Clinical management of continuous–flow left ventricular assist devices in advanced heart failure. J Heart Lung Transplant. (2010) 29:S139. 10.1016/j.healun.2010.01.011

  • 27.

    HannanMMHusainSMattnerFDanziger-IsakovLDrewRJCoreyGRet al. Working formulation for the standardization of definitions of infections in patients using ventricular assist devices. J Heart Lung Transpl. (2011) 30:37584. 10.1016/j.healun.2011.01.717

  • 28.

    KusneSMooneyMDanziger-IsakovLKaanALundLHLysterHet al. An ISHLT consensus document for prevention and management strategies for mechanical circulatory support infection. J Heart Lung Transpl. (2017) 36:113753. 10.1016/j.healun.2017.06.007

  • 29.

    KimJFellerEDChenWLiangYDilsizianV. FDG PET/CT for early detection and localization of left ventricular assist device infection: impact on patient management and outcome. JACC Cardiovas Imaging. (2019) 12:7229. 10.1016/j.jcmg.2018.01.024

  • 30.

    BernhardtAMPamirsadMABrandCReichartDTienkenMBartenMJet al. The value of fluorine-18 deoxyglucose positron emission tomography scans in patients with ventricular assist device specific infections†. Eur J Cardio-Thorac Surg. (2017) 51:10727. 10.1093/ejcts/ezx016

  • 31.

    de VaugeladeCMesguichCNubretKCamouFGreibCDournesGet al. Infections in patients using ventricular-assist devices: comparison of the diagnostic performance of (18)F-FDG PET/CT scan and leucocyte-labeled scintigraphy. J Nucl Cardiol. (2019) 26:4255. 10.1007/s12350-018-1323-7

  • 32.

    KhumriTMThibodeauJBMainML. Transesophageal echocardiographic diagnosis of left atrial appendage occluder device infection. Eur J Echocardiogr. (2007) 9:5656. 10.1016/j.euje.2007.06.005

  • 33.

    SharmaSPSinghDNakamuraDGopinathannairRLakkireddyD. Incomplete endothelialization of WatchmanTM device: predictors and implications from two cases. J Atr Fibrillation. (2019) 11:2162. 10.4022/jafib.2162

  • 34.

    AruniBSharifianAEryaziciPHerreraCJ. Late bacterial endocarditis of an Amplatzer atrial septal device. Indian Heart J. (2013) 65:4501. 10.1016/j.ihj.2013.06.002

Summary

Keywords

cardiac device, cardiac implantable electronic devices, device infection, endocarditis, transesophageal echocardiogram, positron emission tomography, computed tomography, left ventricular assist devices

Citation

Aguilera J, Hutt E and Jaber WA (2021) Imaging of Cardiac Device-Related Infection. Front. Cardiovasc. Med. 8:729786. doi: 10.3389/fcvm.2021.729786

Received

23 June 2021

Accepted

03 August 2021

Published

24 August 2021

Volume

8 - 2021

Edited by

Alessia Gimelli, Gabriele Monasterio Tuscany Foundation (CNR), Italy

Reviewed by

Maria Nazarena Pizzi, Vall d'Hebron University Hospital, Spain; Shady Abohashem, Massachusetts General Hospital and Harvard Medical School, United States

Updates

Copyright

*Correspondence: Wael A. Jaber

This article was submitted to Cardiovascular Imaging, a section of the journal Frontiers in Cardiovascular Medicine

Disclaimer

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Outline

Figures

Cite article

Copy to clipboard


Export citation file


Share article

Article metrics