SYSTEMATIC REVIEW article

Front. Cardiovasc. Med., 15 May 2023
Sec. Heart Valve Disease
Volume 10 - 2023 | https://doi.org/10.3389/fcvm.2023.1123487

Sutureless vs. rapid-deployment valve: a systemic review and meta-analysis for a direct comparison of intraoperative performance and clinical outcomes

Chenhao Wang1 Yi Xie1 Hongwei Zhang1,2 Peng Yang1,2 Yu Zhang1 Chen Lu1 Yu Liu1 Haiyue Wang1 Zhenyuan Xu1 Jia Hu1,2,3*
  • 1Department of Cardiovascular Surgery, West China Hospital, Sichuan University, Chengdu, Sichuan, China
  • 2Cardiovascular Surgery Research Laboratory, West China Hospital, Sichuan University, Chengdu, Sichuan, China
  • 3Department of Cardiovascular Surgery, Guang'an Hospital of West China Hospital, Sichuan University, Guang'an, Sichuan, China

Background: Sutureless and rapid-deployment valves are bioprostheses anchoring within the aortic annulus with few sutures, and they act as a hybrid of conventional surgical and transcatheter valves under aortic valve replacement. Considering that the 3F Enable valve is now off-market, the only two sutureless and rapid-deployment valves available on the world marketplace are the Perceval and Intuity valves. However, a direct comparison of the function of these two valves eludes researchers.

Purpose: Against this background, we performed this systematic review and meta-analysis comparing the intraoperative performance and early clinical outcomes between the Perceval valve and the Intuity valve under sutureless and rapid-deployment aortic valve replacement.

Methods: We systematically searched electronic databases through PubMed/MEDLINE, OvidWeb, Web of Science, and Cochrane Central Register of Controlled Trials (from the establishment of the database to November 17, 2022, without language restriction) for studies comparing the sutureless valve (the Perceval) and the rapid-deployment valve (the Intuity) under aortic valve replacement. Our primary outcomes were early mortality and postoperative transvalvular pressure gradients. The secondary outcomes were defined to include aortic cross-clamp and cardiopulmonary bypass time, paravalvular leak (any paravalvular leak, moderate-to-severe paravalvular leak) after aortic valve replacement, need for pacemaker implantation, postoperative neurological events (stroke), and intensive care unit stay.

Results: This meta-analysis included ten non-randomized trials with 3,526 patients enrolled (sutureless group = 1,772 and rapid-deployment group = 1,754). Quality assessments were performed, with the mean scores of the studies reading 6.90 (SD = 0.99) out of 9 according to the Newcastle–Ottawa Scale. Compared with rapid-deployment aortic valve replacement, sutureless aortic valve replacement was associated with higher mean and peak transvalvular pressure gradients postoperatively. In contrast, aortic cross-clamp and cardiopulmonary time were needed less in sutureless aortic valve replacement vs. rapid-deployment aortic valve replacement. There was no evidence of significant publication bias observed by the funnel plot and Egger's test.

Conclusions: For postoperative hemodynamics, sutureless aortic valve replacement was associated with increased mean and peak transvalvular pressure gradients compared with rapid-deployment aortic valve replacement. In sharp contrast, sutureless aortic valve replacement significantly reduced the amount of time needed for fixing the aortic cross-clamp and the cardiopulmonary bypass procedure.

Systematic Review Registration: https://www.crd.york.ac.uk/prospero/, identifier CRD42022343884.

Introduction

Aortic stenosis (AS) will become one of the most common valvular heart diseases as the population ages and life expectancy increases (1). Surgical aortic valve replacement (SAVR) is always considered the golden standard for treating AS (2). However, considering the high surgical risks involved, more than 30% of patients with severely symptomatic AS do not undergo surgery in clinical practice. Transcatheter aortic valve implantation (TAVI) has become an effective alternative established for the purpose of providing less-invasive treatment. Nevertheless, a crucial limitation of TAVI is that it is almost impossible to remove all native valve cusps or a degenerated prosthesis (36).

Recently, sutureless and rapid-deployment valves have emerged as prospective substitutes for typical valves (2). These valves are biological prostheses anchoring within the aortic annulus with at most three sutures (7, 8). With sufficient radial force to allow annular implantation without sutures in a sutureless valve and the rapid-deployment valve system providing an innovative extended balloon structure requiring only three sutures for fast deployment, these valves facilitate minimally invasive surgery and complex intervention in annulus decalcification and degenerated valve removal. Evidence from the Sutureless and Rapid-Deployment Aortic Valve Replacement International Registry (SURD-IR) enrolling more than 4,500 patients suggests that SURD-AVR is a secure and efficacious substitute for the conventional aortic valve replacement procedure (7, 9, 10).

Three sutureless and rapid-deployment prosthesis valves have received CONFORMITE EUROPEENNE (CE) market approval: the Perceval S, the Intuity, and the 3f Enable. , The 3f Enable valve was recalled in 2014 most probably because of elevated migration risks (8). Sutureless valve Perceval and rapid-deployment valve Intuity are the only two representatives of valves in SURD-AVR, both of which function well in sutureless and rapid-deployment aortic implantation by reducing aortic cross-clamp and cardiopulmonary bypass time and delivering excellent hemodynamic results (10, 11). At the same time, a previous study demonstrated that SURD-AVR was associated with an increased rate of pacemaker implantation postoperatively compared with SAVR (12). However, there are limited published data directly comparing both promising devices, and most of these data are only observational and retrospective studies rather than randomized controlled trials or only small sample studies riddled with deficiencies.

In this study, we performed a systematic review and meta-analysis to evaluate the intraoperative performance and early clinical outcomes between the sutureless and the rapid-deployment aortic valve replacement methods.

Methods

Data source and search strategy

We searched Pubmed/Medline, Ovidweb, Web of Science, and the Cochrane Central Register of Controlled Trials (CENTRAL) for relevant articles, from the date of establishment of the database to November 17, 2022, in all languages, using a combination of main terms and MeSH terms such as “aortic valve[MeSH terms]” or “heart valve prosthesis[MeSH terms]” or “aortic valve replacement” or “aortic valve implantation” and “sutureless” or “Perceval” and “rapid deployment” or “Intuity”. Next, we performed a search for additional sources of information for the literature supplement, including Google Scholar and abstracts/presentations from major international cardiovascular-relevant conferences. Finally, the reference lists of relevant works of literature were also checked for the supplement. The complete retrieval strategy is presented in Supplementary Table S1.

Study selection and data extraction

Two investigators (CW and YX) independently performed the study selection on the basis of predetermined selection criteria. Any discrepancy among the investigators was resolved by a third investigator (JH). After removing duplicates, we performed selection through two levels: the title and abstract of each searched study were screened for relevance as part of the first level, and a full-text analysis of the remaining studies was done for inclusion as the second level. Studies were considered eligible for inclusion in our systematic review and meta-analysis if they fulfilled the following criteria: (1) enrolled patients undergoing aortic valve replacement and who used both sutureless and rapid-deployment valves; (2) those who reported at least one primary outcome, defined as early mortality (30-day all-cause mortality and in-hospital mortality) and postoperative transvalvular pressure gradients (mean/peak); (3) the sample size of each group should be more than 10; (4) there should be no duplicated population figures across studies. Without any restrictions as full texts, abstract reports from important conferences that met the inclusion criteria were also considered in our study.

Using standardized data collection sheets that recorded essential items, we extracted the following data from each included study: study characteristics [publication characteristics (authors, publication year), study era, study country, study design, statistical analysis adjustment, study population], patient characteristics [age, sex, body surface area, body mass index, EuroScoreII, surgical approach (proportion of the minimally invasive approach), proportion of isolated AVR], and outcomes (primary outcomes: early mortality, transvalvular pressure gradients; secondary outcomes: aortic cross-clamp time, cardiopulmonary bypass time, paravalvular leak, pacemaker implantation, stroke, ICU stay). Data extraction was performed by two investigators (CW and YX), and discrepancies were resolved by a third investigator (JH).

Quality assessment

We assessed the overall study quality using NEWCASTLE-OTTAWA SCALE (NOS) for observational studies (13), based on the three domains: selection of participators, comparability between study groups, and outcomes. Each study in this rating system (with a maximum of 9 stars) can receive up to 1 star for each numbered entry in the Selection and Outcome categories and up to 2 stars for the majority of entries in the Comparability category. A score of 9 stars received in the study indicates a low risk of bias, and a study that receives 8 or 7 stars is assessed as having a moderate risk of bias. In contrast, an assigned score of 6 or less indicates a high risk of bias.

Outcomes

The primary outcomes of interest in the study were early mortality and transvalvular pressure gradients of the aortic valve after AVR. Early mortality was defined as 30-day all-cause mortality and in-hospital mortality. Transvalvular pressure gradients included mean transvalvular pressure gradients and peak transvalvular pressure gradients. The secondary outcomes of interest included ACC and CPB time, paravalvular leak (any paravalvular leak, moderate-to-severe paravalvular leak) after AVR, the need for pacemaker implantation, postoperative neurological events (stroke), and ICU stay.

Statistical analysis

For continuous outcomes (transvalvular pressure gradients, aortic cross-clamp time, cardiopulmonary bypass time, and ICU stay), results were presented as the mean difference (MD) with a 95% confidence interval (CI) using an Inverse Variance fixed effect model, followed by real events, significance for effect estimate (p-value), I2 statistic, and Q statistic. We estimated the mean values and standard deviations using the formula if studies reported only the median and interquartile/overall range (14). The results of dichotomous outcomes (early mortality, paravalvular leak, pacemaker implantation, and stroke) were presented as the odds ratio (OR) with a 95% confidence interval (CI) using the Mantel–Haenszel fixed effect model. Total events, significance for effect estimate (p-value), I2 statistic, and Q statistic were also presented in pooling. When a moderate-to-high heterogeneity was discovered in the trial, the random effects model with the Inverse Variance or Mantel–Haenszel method was used in continuous or dichotomous outcomes, respectively. Operative time, including the aortic cross-clamp time and cardiopulmonary time, were pooled and presented in minutes, whereas ICU stay was presented in days. The magnitude of the statistical heterogeneity between studies was assessed using the Higgins I2 test, with rates of 25%, 50%, and 75% being indicative of low, moderate, and high heterogeneity, respectively (15). Furthermore, Cochran's Q statistic was used to assess the heterogeneity between the studies. We performed the leave-one-out sensitivity analysis to explore potential sources of heterogeneity by removing individual studies each time. Subgroup analysis was also performed to further stratify outcomes. We visually assessed potential publication bias by considering the asymmetry in the funnel plots of the effect size of each estimate against the standard error. A formal calculation of the possibility of publication bias was done by using Egger's test, which defines publication bias as significant if p < 0.1 (16). All study analyses were performed using Stata 16.0 (StataCorp LLC) and Review Manager Version 5.4.1 (The Cochrane Collaboration).

Results

Study search

Our initial systematic electronic literature yielded 1,015 articles. After removing 374 duplicates, 771 articles were screened at the title/abstract level. Among these articles, 743 publications were excluded, which did not fulfill the selection criteria based on the title and abstract. With 28 articles remaining and assessed for eligibility, 10 publications were deemed eligible and included in the meta-analysis (Figure 1) (11, 1725).

FIGURE 1
www.frontiersin.org

Figure 1. Preferred reporting items for systematic reviews and meta-analyses (PRISMA) flow diagram.

Study characteristics and patient populations

The included 10 studies, nine full-text studies and one abstract with integral statistical reports, were all non-randomized studies (NRSs). Because there were three studies from the same registry, another two were used only to report supplementary data (19, 21). All studies covered 3,526 patients (sutureless group = 1,772 and rapid-deployment group = 1,754). Among these studies, propensity score matching was used in five studies (11, 17, 20, 24, 25), whereas in one study, the multivariable analysis method was used for determining early mortality in risk factor analysis (Table 1) (21). A larger proportion of male patients were enrolled in the rapid-deployment group. The mean age of patients in all studies ranged from 70 to 83 years, with most of them in their 70s (Table 2). Six studies reported about the body surface area in each group, with the rapid-deployment group having a statistically significant higher index (11, 18, 21, 22, 24, 25). One study reported data by dividing isolated AVR patients and combined AVR patients into two separate cohorts, which led us to perform a statistical analysis of these cohorts (25). All studies provided data on early mortality or transvalvular pressure gradients as primary outcomes, whereas specific secondary outcomes were unavailable in every study.

TABLE 1
www.frontiersin.org

Table 1. Study characteristics.

TABLE 2
www.frontiersin.org

Table 2. Patient characteristics.

Quality assessment

The methodological quality of each study varied, and the mean scores of the studies were 6.90 (SD = 0.99) out of 9 according to the Newcastle–Ottawa Scale (NOS), representing the included studies as moderate-to-high quality. A detailed quality assessment is presented in Supplementary Table S2.

Early mortality

All included studies reported early mortality, defined as 30-day all-cause mortality in five studies (11, 17, 18, 22, 24) and in-hospital mortality in another three studies (20, 23, 25), respectively. Effect sizes were expressed by ORs, whereas ORs were not calculated in one study because the early mortality in both groups was 0 (18). The calculated overall early mortality rate was 2.3%, being 2.5% in patients receiving Perceval valve implantation and 2.1% in those who underwent Intuity valve implantation (p = 0.31). The SU group showed no statistically significant difference in early mortality rates compared with the RD group (8 studies and 3,526 patients, OR: 1.26; 95% CI: 0.81–1.96; p= 0.31; I2 = 0%, Figure 2). No significant publication bias was observed, which was assessed by considering the asymmetry in the funnel plot visually and formally by using Egger's regression test (p = 0.5190, Supplementary Figure S5A). Finally, a sensitivity analysis was used to examine the influence of each study on the OR by excluding one individual study at one time. The exclusion of each study did not significantly change the pooled OR, and the estimates for each case were within the overall 95% confidence interval.

FIGURE 2
www.frontiersin.org

Figure 2. Odds ratio (OR) of early mortality in sutureless (SU) versus rapid-deployment (RD) aortic valve replacement. Overall pooled analysis from patients is shown. Compared with the RD group, the SU group is not associated with a significantly higher risk of early mortality (OR: 1.26; 95% CI: 0.81–1.96; p= 0.31; I2 = 0%). M–H, Mantel–Haenszel; CI, confidence interval.

Transvalvular pressure gradients

Mean transvalvular pressure gradients

Overall, the patients' mean transvalvular pressure gradients were presented in seven studies (11, 17, 20, 2225), and five studies reported the mean transvalvular pressure gradients in each size of both valve types (11, 19, 2224). The pooled analysis from seven studies covering 3,483 patients demonstrated that the SU group was associated with statistically significant higher mean transvalvular pressure gradients in patients overall, compared with the RD group (MD: 2.93; 95% CI: 2.19–3.67; p < 0.00001; I2 = 65%, Figure 3A). Next, we performed subgroup analyses by matching the sizes of the Perceval and Intuity valves to further explore the relationship between valve size and transvalvular pressure gradients and make a hierarchical contrast between the two types of valves. Subgroup 1 compared SU with RD valve sizes under radical matching by small with 21 mm, medium with 23 mm, large with 25 mm, and extralarge with 27 mm, whereas subgroup 2 compared SU with RD valve sizes under conservative matching by small with 19 mm, medium with 21 mm, large with 23 mm, and extralarge with 25 mm. Subgroup analyses demonstrated that under radical matching of valve size, the SU group was still associated with statistically significant higher mean transvalvular pressure gradients in each size-matching compared with the RD group (MD: 3.57; 95% CI: 3.20–3.94; p< 0.00001; I2 = 26%, Figure 3B). However, under conservative matching of valve size, it presented a lower mean transvalvular pressure gradient in the S SU valve than the 19 mm RD valve, but it was still significantly higher in the M, L, and XL SU valves than in the 21, 23, and 25 mm RD valves, respectively (MD: 1.68; 95% CI: 0.77–2.58; p= 0.0003; I2 = 74%, Figure 3C). No significant publication bias was observed in patients overall, which was assessed by considering the asymmetry in the funnel plot visually and formally by using Egger's regression test (p = 0.5879, Supplementary Figure S5B).

FIGURE 3
www.frontiersin.org

Figure 3. Mean difference (MD) of mean transvalvular pressure gradients (mmHg) in sutureless (SU) versus rapid-deployment (RD) aortic valve replacement. Overall pooled analyses from patients (A), subgroup 1 (B) and subgroup 2 (C) are shown. Subgroup 1 matches SU with RD valve sizes as small with 21 mm, medium with 23 mm, large with 25 mm, and extralarge with 27 mm, and subgroup 2 matches SU with RD valve sizes as small with 19 mm, medium with 21 mm, large with 23 mm, and extralarge with 25 mm. Compared with the RD group, the SU group is associated with a significantly higher mean transvalvular pressure gradient in patients overall (MD: 2.93; 95% CI: 2.19–3.67; p < 0.00001; I2 =65%), subgroup 1 (MD: 3.57; 95% CI: 3.20–3.94; p< 0.00001; I2 =26%) and subgroup 2 (MD: 1.68; 95% CI: 0.77–2.58; p= 0.0003; I2 = 74%). SD, standard deviation; IV, inverse-variance; CI, confidence interval.

Peak transvalvular pressure gradients

For peak transvalvular pressure gradients, statistical analyses demonstrated the same tendency as the mean transvalvular pressure gradients. Five studies and 3,201 patients were covered in an overall pooled analysis (11, 2225), which demonstrated that the SU group was associated with statistically significant higher peak transvalvular pressure gradients in patients overall, compared with the RD group (MD: 5.11; 95% CI: 4.45–5.78; p < 0.00001; I2 = 47%, Figure 4A). Subgroup analyses were also performed by small with 21 mm, medium with 23 mm, large with 25 mm, and extralarge with 27 mm as radical matching and small with 19 mm, medium with 21 mm, large with 23 mm, and extralarge with 25 mm as conservative matching. For radical matching, the SU group was associated with statistically significant higher peak transvalvular pressure gradients in each size-matching compared with the RD group (MD: 6.00; 95% CI: 5.34–6.65; p< 0.00001; I2 = 0%, Figure 4B). For conservative matching, the peak pressure gradients in the SU group were still significantly higher in the M, L, and XL SU valves than in the 21, 23, and 25 mm RD valves (MD: 2.86; 95% CI: 1.18–4.55; p= 0.0008; I2 = 82%, Figure 4C). No significant publication bias was observed in patients overall, which was assessed by taking into account the asymmetry in the funnel plot visually and formally by using Egger's regression test (p = 0.8425, Supplementary Figure S5C).

FIGURE 4
www.frontiersin.org

Figure 4. Mean difference (MD) of peak transvalvular pressure gradients (mmHg) in sutureless (SU) versus rapid-deployment (RD) aortic valve replacement. Overall pooled analyses from patients (A), subgroup 1 (B), and subgroup 2 (C) are shown. Subgroup 1 matches SU with RD valve sizes as small with 21 mm, medium with 23 mm, large with 25 mm, and extralarge with 27 mm. Subgroup 2 matches SU with RD valve sizes as small with 19 mm, medium with 21 mm, large with 23 mm, and extralarge with 25 mm. Compared with the RD group, the SU group is associated with significantly higher peak transvalvular pressure gradients in patients overall (MD: 5.11; 95% CI: 4.45–5.78; p < 0.00001; I2 = 47%), subgroup 1 (MD: 6.00; 95% CI: 5.34–6.65; p< 0.00001; I2 = 0%) and subgroup 2 (MD: 2.86; 95% CI: 1.18–4.55; p= 0.0008; I2 = 82%). SD, standard deviation; IV, inverse-variance; CI, confidence interval.

Secondary outcomes

For secondary outcome studies, extracted estimates were reported in the supplementary material. Overall pooled analyses from isolated AVR patients, combined AVR patients, and AVR patients demonstrated that, compared with the RD group, the SU group was associated with a significantly less aortic cross-clamp time (MD: −10.12; 95% CI: −13.90 to −6.33; p < 0.00001; I2 = 94%, Figure 5A), and similarly, with a significantly less cardiopulmonary bypass time (MD: −11.63; 95% CI: −17.14 to −6.13; p< 0.0001; I2 = 94%, Figure 5B). There were no statistically significant differences between the SU group and the RD group for any paravalvular leak (OR: 1.95; 95% CI: 1.01–3.77; p= 0.05; I2 = 75%, Supplementary Figure S1A), paravalvular leak (moderate to severe) (OR: 1.07; 95% CI: 0.61–1.87; p= 0.82; I2 = 0%, Supplementary Figure S1B), pacemaker implantation (OR: 1.16; 95% CI: 0.92–1.47; p= 0.20; I2 = 0%, Supplementary Figure S2), stroke (OR: 1.07; 95% CI: 0.70–1.64; p= 0.75; I2 = 0%, Supplementary Figure S3), and intensive care unit (ICU) stay (MD: −0.03; 95%CI: −0.37 to 0.31; p= 0.87; I2 = 75%, Supplementary Figure S4). A visual assessment of the symmetry of the funnel plots suggested that there was no significant publication bias, and a formal assessment by using Egger's test confirmed this point (Supplementary Figure S5).

FIGURE 5
www.frontiersin.org

Figure 5. Mean difference (MD) of aortic cross-clamp (ACC) (A) and cardiopulmonary bypass (CPB) (B) times in sutureless (SU) versus rapid-deployment (RD) aortic valve replacement (AVR). Overall pooled analyses from isolated AVR patients, combined AVR patients, and AVR patients are shown. Compared with the RD group, the SU group is associated with a significantly less aortic cross-clamp time (MD: −10.12; 95% CI: −13.90 to −6.33; p< 0.00001; I2 = 94%), and similarly, with a significantly less cardiopulmonary bypass time (MD: −11.63; 95% CI: −17.14 to −6.13; p< 0.0001; I2 =94%). SD, standard deviation; IV, inverse-variance; CI, confidence interval.

Discussion

In this study, we conducted a meta-analysis covering 10 non-randomized trials and 3,526 patients, highlighting two key findings. First, compared with the RD group, the SU group was associated with statistically significant higher mean and peak transvalvular pressure gradients of the aortic valve. Second, the SU group was associated with an overall decrease of ACC and CPB times for 10.12 min and 11.63 min, respectively, compared with the RD group. In terms of early mortality, paravalvular leak, moderate-to-severe paravalvular leak, pacemaker implantation, stroke, or ICU stay, data analysis revealed commonalities between the two groups.

Our honest opinion is that selecting the appropriate valve for a defined patient based on the information revealed in our study remains a challenging proposition. Although our study revealed that the two valves displayed varied hemodynamic and intraoperative performances, this did not translate into different clinical outcomes for patients. However, there is still a lack of medium- to long-term follow-up and comprehensive data to determine critical outcomes in terms of survival and major adverse cardiac and cerebral events. Therefore, it is important to have risk predictors that impact the long-term prognosis for the two valves when analyzing the advantages and disadvantages of each valve, with implications to guide clinicians in their selection.

It has been proved that SURD-AVR possesses a better hemodynamic function compared with SAVR (25). The following interpretations, according to several investigations, could account for this satisfactory observation: (1) the non-pledged sutures may contribute to a huger laminar flow; (2) as the thin stent allows the leaflets to move freely without being firmly bound to bulky stents, the Perceval valve result in the pressure gradients drops; (3) seated below the annulus, the skirt frame of the stent of the Intuity valve has a flared configuration in the left ventricular outflow, which may play a role in active constriction limitation in the left ventricular outflow tract (LVOT) (2529).

Our meta-analysis performed using both radical and conservative matching revealed that when compared with the Intuity valve, the Perceval valve had statistically significant higher mean transvalvular pressure gradients across all patients and subgroup analyses.

Theoretically, in terms of valve structure, as the Intuity valve has the valve annulus stent covered by a polyester sealing cloth (8, 30), the Perceval valve could offer a larger effective outflow orifice area, leading to its better hemodynamic performance. Nevertheless, this hypothesis is in stark contrast to our meta-analysis observation, which should be highlighted purposely. A previous study reported this theory-contradicted finding (31). If the stent in the Perceval valve undergoes compression or deformation after the prosthesis implantation procedure, it could indicate oversizing relative to the annulus or procedural misoperation by the surgeon, potentially resulting in a high gradient. This grossly oversized prosthesis mismatched with the patient tends to spring back, causing incomplete valve opening and contact loss from the annulus, which possibly results in high paravalvular leakage, besides an increase in the pressure gradients. Several published studies reported that Perceval valve rebounds were observed in clinical implantation and laboratories (32, 33). This feasible explanation for cracking the paradox of valve-pressure gradients is consistent with the trend of paravalvular leak in our meta-analysis results (SU group: 184 in 1,530; RD group: 96 in 1,542. OR: 1.95; 95% CI: 1.01–3.77; p= 0.05; I2 = 75%). Strikingly, another theoretical possibility was proposed by Campbell D. Flynn et al. to the effect that the Intuity valve that has better pressure gradients focuses on the valve skirt (34). The subannular balloon-expanded valve skirt in the Intuity valve is proposedly attributed to the recognized excellent transvalvular pressure gradients in the RD group, in which the LVOT is enlarged, promoting an increase in blood flow through the valve annulus (35, 36). Although the expandable frame skirt in the Intuity valve may enlarge the LVOT, it is certain that the stent located at the leaflet attachment margin narrows the orifice area. To sum up, our study was more inclined to conclude that the incomplete valve opening in the Perceval valve caused a higher gradient and showed a higher tendency toward paravalvular leak, for which further studies should confirm the potential mechanism.

Furthermore, it is necessary to highlight that the difference in valve gradient between these two groups (MD = 2.93 mmHg in mean aortic pressure gradients; MD = 5.11 mmHg in peak aortic pressure gradients) did not translate into differences in early clinical outcomes. In the meantime, the hemodynamic performance of the two valves needs to be further followed up and explored. Only then will it be possible to show the impact of the difference in transvalvular pressure gradients on the long-term prognosis of patients who received SU and RD-AVR. Notably, patients with smaller aortic annuli who undergo aortic valve replacement often exhibit higher transvalvular pressure gradients, and the presence of a small aortic annulus may augment the risk of patient–prosthesis mismatch (37, 38). Hence, it is plausible that the Intuity valve may offer superior postoperative benefits to patients with a small aortic annulus.

In our study, overall pooled analyses from isolated AVR patients, combined AVR patients, and AVR patients demonstrated that, compared with the RD group, the SU group was associated with significantly less aortic cross-clamp time (MD: −10.12; 95% CI: −13.90 to −6.33; p < 0.00001; I2 = 94%). We suspected that this discrepancy arose because of these two valves possessing distinct suture structures. The Perceval valve is a bovine pericardium prosthesis attached to the automated anchor used for stabilization and a fastened implantation site. When the valve is placed down to the annulus, three intercommissural sutures are used for guiding, which will be removed after valve deployment is completed (8, 39). In addition, the Perceval valve with a collapsed design may maximize visualization and simplify implantation (25). In contrast, three braided, non-pledged sutures are placed at the bottom of every valve sinus using a figure-of-eight or horizontal mattress technique without removal if the Intuity valve is selected for use in the AVR. Once annular seating is verified, the balloon will be inserted through the holder, and the stent will be deployed by inflating it to the appropriate level of pressure with saline for 10 s (40). Therefore, the Perceval valve is the only one that precisely matches the definition of “sutureless” during operation. Because of these structural and procedural differences with the Perceval valve, some opponents have argued that the Intuity valve cannot strictly be labeled as a “rapid-deployment” valve (30). However, it was noted that the magnitude assessment showed high heterogeneity, with subgroup analysis and leave-one-out sensitivity analysis being inefficient for elimination.

Postoperative mortality and morbidity are strongly associated with the duration of both ACC and CPB. A previously published study has indicated that ACC time is a critical and independent risk predictor of severe cardiovascular morbidities, with the risk increasing by 1.4% for each additional minute of ACC time (41). Kenji Lino et al. also revealed that ACC time serves as an independent risk predictor of postoperative morbidity for aortic valve replacement, with a prolonged ACC duration significantly increasing the rates of renal failure, gastrointestinal complications, pneumonia, and multiorgan failure (42). In addition, a study conducted in China has reported that CPB time is independently linked to an increased risk of acute kidney injury following surgery for acute DeBakey Type I aortic dissection (43). Therefore, for high-risk patients undergoing AVR, reducing the ACC and CPB times may confer substantial advantages in using the Perceval valve, particularly for patients with pre-existing organ damage and infections or for those undergoing redo surgery (44, 45).

Two meta-analyses (30, 34) anchored on the comparison of the sutureless and rapid-deployment aortic valves in SURD-AVR had been published before our study was done. Nevertheless, two aspects (paravalvular leak and pacemaker implantation) of our analysis presented negative results, showing slight differences with the conclusions of the two previous studies. Published studies may be responsible for causing discrepancies at different times, discrepancies in inclusion criteria, and differences in the exact definition of study outcomes. However, it is noteworthy and distinctive that compared with other studies to date, our study covers the largest period, the largest number of patients, the most significant number of included studies, all types of early clinical results, and the use of two valve size gradient matching methods, to enable a comprehensive and objective comparative analysis.

There are several limitations in our analysis that merit a scrupulous consideration. First, we included only 10 studies overall; also, we did not include any RCT. Although propensity score matching was performed in more than half of the included studies to equalize confounders in non-randomized studies similar to randomization, there is no denying the potential selection bias of our investigators. Second, SURD-IR and Germany are the majority contributors to the patient data source that we collected in the study, which means a more homogeneous region and race limit the generalizability of analysis results. Third, because follow-up was patchy across studies, there is a need for comparing the efficacy and durability of the two valves in the medium and long term. Fourth, the results of ACC and CPB times showed high heterogeneity. Even though we performed leave-one-out sensitivity analysis and subgroup analysis, we still could not well locate and reduce the source of heterogeneity. Fifth, although we performed subgroup analysis by valve size to ensure precise matching, no clear distinction could be perceived between Perceval S (Livanova PLC, London, UK) and Perceval S (Sorin Group, Saluggia, Italy) in the results of pooled estimates reported in our study. Last, potential publication bias cannot be definitively ruled out, even though both Egger's test and the funnel plots suggest no potential publication bias.

Conclusion

Although further trials and reviews are required for making a more detailed and deterministic comparison between the valves in SURD-AVR, particularly clinical outcomes in the medium and long term in practice, our findings lend support to the notion that sutureless aortic valve replacement is associated with significantly higher postoperative mean and peak transvalvular pressure gradients of the aortic valve compared with rapid-deployment aortic valve replacement in overall and subgroup analyses. Sutureless aortic valve replacement provided visible benefits to patients in terms of intraoperative performance as there was a significant reduction in ACC and CPB times compared with rapid-deployment aortic valve replacement. We also discussed the role of different risk predictors to guide valve selection. In conclusion, clinical decision-making should necessitate thoughtful valve selection for all patients prior to SURD-AVR, and in this context, it can be said that both Perceval and Intuity valves are rising stars in the bioprosthesis firmament, complementing each other very well.

Data availability statement

The original contributions presented in the study are included in the article/Supplementary Material, further inquiries can be directed to the corresponding author.

Author contributions

Conceptualization and design were done by YX, CW, HZ, CL, HW, and YZ; data collection and assembly were carried out by CW, ZX, and YL; analyses and interpretation of data were done by YX, CW, and PY; manuscript writing was done by YX and CW; reading and revising the manuscript were done by JH. All authors contributed to the article and approved the submitted version.

Funding

This study was subsidized by the National Natural Science Foundation of China (No. 81670327), 1·3·5 project for disciplines of excellence–Clinical Research Incubation Project, West China Hospital, Sichuan University (No. 2019HXFH027), and Sichuan Science and Technology Program (2019YJ0046) to perform literature searches, statistical analyses, and manuscript writing.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fcvm.2023.1123487/full#supplementary-material.

References

1. Kanwar A, Thaden JJ, Nkomo VT. Management of patients with aortic valve stenosis. Mayo Clin Proc. (2018) 93(4):488–508. doi: 10.1016/j.mayocp.2018.01.020

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Bonow RO, Leon MB, Doshi D, Moat N. Management strategies and future challenges for aortic valve disease. Lancet (London, England). (2006) 387(10025):1312–23. doi: 10.1016/S0140-6736(16)00586-9

CrossRef Full Text | Google Scholar

3. Leon MB, Smith CR, Mack M, Miller DC, Moses JW, Svensson LG, et al. Transcatheter aortic-valve implantation for aortic stenosis in patients who cannot undergo surgery. N Engl J Med. (2010) 363(17):1597–607. doi: 10.1056/NEJMoa1008232

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Gargiulo G, Sannino A, Capodanno D, Barbanti M, Buccheri S, Perrino C, et al. Transcatheter aortic valve implantation versus surgical aortic valve replacement: a systematic review and meta-analysis. Ann Intern Med. (2016) 165(5):334–44. doi: 10.7326/M16-0060

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Thourani VH, Kodali S, Makkar RR, Herrmann HC, Williams M, Babaliaros V, et al. Transcatheter aortic valve replacement versus surgical valve replacement in intermediate-risk patients: a propensity score analysis. Lancet. (2016) 387(10034):2218–25. doi: 10.1016/S0140-6736(16)30073-3

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Yerasi C, Rogers T, Forrestal BJ, Case BC, Khan JM, Ben-Dor I, et al. Transcatheter versus surgical aortic valve replacement in young, low-risk patients with severe aortic stenosis. JACC Cardiovasc Interv. (2021) 14(11):1169–80. doi: 10.1016/j.jcin.2021.03.058

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Di Eusanio M, Phan K. Sutureless aortic valve replacement. Ann Cardiothorac Surg. (2015) 4(2):123–30. doi: 10.3978/j.issn.2225-319X.2015.02.06

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Carrel T, Heinisch PP. History, development and clinical perspectives of sutureless and rapid deployment surgical aortic valve replacement. Ann Cardiothorac Surg. (2020) 9(5):375–85. doi: 10.21037/acs-2020-surd-18

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Di Eusanio M, Berretta P. The sutureless and rapid-deployment aortic valve replacement international registry: lessons learned from more than 4,500 patients. Ann Cardiothorac Surg. (2020) 9(4):289–97. doi: 10.21037/acs-2020-surd-21

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Liakopoulos OJ, Gerfer S, Weider S, Rahmanian P, Zeriouh M, Eghbalzadeh K, et al. Direct comparison of the Edwards Intuity Elite and Sorin Perceval S rapid deployment aortic valves. Ann Thorac Surg. (2018) 105(1):108–14. doi: 10.1016/j.athoracsur.2017.06.034

PubMed Abstract | CrossRef Full Text | Google Scholar

11. D'Onofrio A, Salizzoni S, Filippini C, Tessari C, Bagozzi L, Messina A, et al. Surgical aortic valve replacement with new-generation bioprostheses: sutureless versus rapid-deployment. J Thorac Cardiovasc Surg. (2020) 159(2):432–442.e1. doi: 10.1016/j.jtcvs.2019.02.135

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Jarrett CM. Permanent pacemaker insertion following transcatheter aortic valve replacement: not infrequent, not benign, and becoming predictable. J Thorac Cardiovasc Surg. (2017) 153(5):1063–4. doi: 10.1016/j.jtcvs.2016.12.014

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Muka T, Glisic M, Milic J, Verhoog S, Bohlius J, Bramer W, et al. A 24-step guide on how to design, conduct, and successfully publish a systematic review and meta-analysis in medical research. Eur J Epidemiol. (2020) 35(1):49–60. doi: 10.1007/s10654-019-00576-5

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Wan X, Wang W, Liu J, Tong T. Estimating the sample mean and standard deviation from the sample size, median, range and/or interquartile range. BMC Med Res Methodol. (2014) 14:135. doi: 10.1186/1471-2288-14-135

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Higgins JP, Thompson SG, Deeks JJ, Altman DG. Measuring inconsistency in meta-analyses. Br Med J. (2003) 327(7414):557–60. doi: 10.1136/bmj.327.7414.557

CrossRef Full Text | Google Scholar

16. Mavridis D, Salanti G. Exploring and accounting for publication bias in mental health: a brief overview of methods. Evid Based Ment Health. (2014) 17(1):11–5. doi: 10.1136/eb-2013-101700

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Nguyen A, Forcillo J, Stevens L, Bouchard D, Demers P, Carrier M. Perioperative outcomes with the intuity versus perceval valve system: a propensity-matched analysis of 2 strategies in intermediate-risk patients. Can J Cardiol. (2015) 31:S172. doi: 10.1016/j.cjca.2015.07.367

CrossRef Full Text | Google Scholar

18. Jiritano FMD, Cristodoro LMD, Malta EMD, Mastroroberto PMD. Thrombocytopenia after sutureless aortic valve implantation: comparison between intuity and perceval bioprostheses. J Thorac Cardiovasc Surg. (2016) 152(6):1631–3. doi: 10.1016/j.jtcvs.2016.07.054

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Di Eusanio M, Phan K, Berretta P, Carrel TP, Andreas M, Santarpino G, et al. Sutureless and Rapid-Deployment Aortic Valve Replacement International Registry (SURD-IR): early results from 3,343 patients. Eur J Cardiothorac Surg. (2018) 54(4):768–73. doi: 10.1093/ejcts/ezy132

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Ensminger S, Fujita B, Bauer T, Moellmann H, Beckmann A, Bekeredjian R, et al. Rapid deployment versus conventional bioprosthetic valve replacement for aortic stenosis. J Am Coll Cardiol. (2018) 71(13):1417–28. doi: 10.1016/j.jacc.2018.01.065

PubMed Abstract | CrossRef Full Text | Google Scholar

21. Berretta P, Andreas M, Carrel TP, Solinas M, Teoh K, Fischlein T, et al. Minimally invasive aortic valve replacement with sutureless and rapid deployment valves: a report from an international registry (Sutureless and Rapid Deployment International Registry). Eur J Cardiothorac Surg. (2019) 56(4):793–9. doi: 10.1093/ejcts/ezz055

PubMed Abstract | CrossRef Full Text | Google Scholar

22. Gotzmann M, Wilbring M, Charitos E, Treede H, Silaschi M. Hemodynamic comparison of sutureless and rapid-deployment valves with conventional bioprostheses. Thorac Cardiovasc Surg. (2020) 68(7):584–94. doi: 10.1055/s-0039-1683426

PubMed Abstract | CrossRef Full Text | Google Scholar

23. Hartrumpf M, Kuehnel R-U, Schroeter F, Haase R, Laux ML, Ostovar R, et al. Clinical short-term outcome and hemodynamic comparison of six contemporary bovine aortic valve prostheses. Thorac Cardiovasc Surg. (2020) 68(07):557–66. doi: 10.1055/s-0038-1676853

PubMed Abstract | CrossRef Full Text | Google Scholar

24. Liakopoulos OJ, Gerfer S, Rahmanian P, Eghbalzadeh K, Djordjevic I, Schlachtenberger G, et al. Rapid deployment aortic valve replacement with the perceval S and intuity elite. Thorac Cardiovasc Surg. (2021) 69(05):412–9. doi: 10.1055/s-0040-1716892

PubMed Abstract | CrossRef Full Text | Google Scholar

25. Berretta P, Meuris B, Kappert U, Andreas M, Fiore A, Solinas M, et al. Sutureless versus rapid deployment aortic valve replacement: results from a multicenter registry. Ann Thorac Surg. (2022) 114(3):758–65. doi: 10.1016/j.athoracsur.2021.08.037

PubMed Abstract | CrossRef Full Text | Google Scholar

26. Borger MA, Dohmen PM, Knosalla C, Hammerschmidt R, Merk DR, Richter M, et al. Haemodynamic benefits of rapid deployment aortic valve replacement via a minimally invasive approach: 1-year results of a prospective multicentre randomized controlled trial. Eur J Cardiothorac Surg. (2016) 50(4):713–20. doi: 10.1093/ejcts/ezw042

PubMed Abstract | CrossRef Full Text | Google Scholar

27. Tasca G, Vismara R, Mangini A, Romagnoni C, Contino M, Redaelli A, et al. Comparison of the performance of a sutureless bioprosthesis with two pericardial stented valves on small annuli: an in vitro study. Ann Thorac Surg. (2017) 103(1):139–44. doi: 10.1016/j.athoracsur.2016.05.089

PubMed Abstract | CrossRef Full Text | Google Scholar

28. Tabata M, Shibayama K, Watanabe H, Sato Y, Fukui T, Takanashi S. Simple interrupted suturing increases valve performance after aortic valve replacement with a small supra-annular bioprosthesis. J Thorac Cardiovasc Surg. (2014) 147(1):321–5. doi: 10.1016/j.jtcvs.2012.11.020

PubMed Abstract | CrossRef Full Text | Google Scholar

29. D'Onofrio A, Messina A, Lorusso R, Alfieri OR, Fusari M, Rubino P, et al. Sutureless aortic valve replacement as an alternative treatment for patients belonging to the “gray zone” between transcatheter aortic valve implantation and conventional surgery: a propensity-matched, multicenter analysis. J Thorac Cardiovasc Surg. (2012) 144(5):1010–6. doi: 10.1016/j.jtcvs.2012.07.040

PubMed Abstract | CrossRef Full Text | Google Scholar

30. Sohn SH, Kang Y, Kim JS, Choi JW, Jang MJ, Hwang HY. Direct comparison of rapid deployment versus sutureless aortic valve replacement: a meta-analysis. J Thorac Dis. (2021) 13(4):2203–15. doi: 10.21037/jtd-20-3548

PubMed Abstract | CrossRef Full Text | Google Scholar

31. Cerillo AG, Amoretti F, Mariani M, Cigala E, Murzi M, Gasbarri T, et al. Increased gradients after aortic valve replacement with the perceval valve: the role of oversizing. Ann Thorac Surg. (2018) 106(1):121–8. doi: 10.1016/j.athoracsur.2017.12.044

PubMed Abstract | CrossRef Full Text | Google Scholar

32. Baert J, Astarci P, Noirhomme P, de Kerchove L. The risk of oversizing with sutureless bioprosthesis in small aortic annulus. J Thorac Cardiovasc Surg. (2017) 153(2):270–2. doi: 10.1016/j.jtcvs.2016.09.035

PubMed Abstract | CrossRef Full Text | Google Scholar

33. Di Eusanio M, Saia F, Pellicciari G, Phan K, Ferlito M, Dall'Ara G, et al. In the era of the valve-in-valve: is transcatheter aortic valve implantation (TAVI) in sutureless valves feasible? Ann Cardiothorac Surg. (2015) 4(2):214–7. doi: 10.3978/j.issn.2225-319X.2014.11.12

PubMed Abstract | CrossRef Full Text | Google Scholar

34. Flynn CD, Williams ML, Chakos A, Hirst L, Muston B, Tian DH. Sutureless valve and rapid deployment valves: a systematic review and meta-analysis of comparative studies. Ann Cardiothorac Surg. (2020) 9(5):364–74. doi: 10.21037/acs-2020-surd-27

PubMed Abstract | CrossRef Full Text | Google Scholar

35. Sadri V, Bloodworth CHt, Madukauwa-David ID, Midha PA, Raghav V, Yoganathan AP. A mechanistic investigation of the EDWARDS INTUITY elite valve's hemodynamic performance. Gen Thorac Cardiovasc Surg. (2020) 68(1):9–17. doi: 10.1007/s11748-019-01154-y

PubMed Abstract | CrossRef Full Text | Google Scholar

36. Haverich A, Wahlers TC, Borger MA, Shrestha M, Kocher AA, Walther T, et al. Three-year hemodynamic performance, left ventricular mass regression, and prosthetic-patient mismatch after rapid deployment aortic valve replacement in 287 patients. J Thorac Cardiovasc Surg. (2014) 148(6):2854–60. doi: 10.1016/j.jtcvs.2014.07.049

PubMed Abstract | CrossRef Full Text | Google Scholar

37. Wagner IM, Eichinger WB, Bleiziffer S, Botzenhardt F, Gebauer I, Guenzinger R, et al. Influence of completely supra-annular placement of bioprostheses on exercise hemodynamics in patients with a small aortic annulus. J Thorac Cardiovasc Surg. (2007) 133(5):1234–41. doi: 10.1016/j.jtcvs.2006.10.074

PubMed Abstract | CrossRef Full Text | Google Scholar

38. Shalabi A, Spiegelstein D, Sternik L, Feinberg MS, Kogan A, Levin S, et al. Sutureless versus stented valve in aortic valve replacement in patients with small Annulus. Ann Thorac Surg. (2016) 102(1):118–22. doi: 10.1016/j.athoracsur.2016.01.003

PubMed Abstract | CrossRef Full Text | Google Scholar

39. Bilkhu R, Borger MA, Briffa NP, Jahangiri M. Sutureless aortic valve prostheses. Heart. (2019) 105(Suppl 2):s16–20. doi: 10.1136/heartjnl-2018-313513

PubMed Abstract | CrossRef Full Text | Google Scholar

40. Borger MA. Minimally invasive rapid deployment Edwards Intuity aortic valve implantation. Ann Cardiothorac Surg. (2015) 4(2):193–5. doi: 10.3978/j.issn.2225-319X.2014.11.12

PubMed Abstract | CrossRef Full Text | Google Scholar

41. Ranucci M, Frigiola A, Menicanti L, Castelvecchio S, de Vincentiis C, Pistuddi V. Aortic cross-clamp time, new prostheses, and outcome in aortic valve replacement. J Heart Valve Dis. (2012) 21(6):732–9.23409353

PubMed Abstract | Google Scholar

42. Iino K, Miyata H, Motomura N, Watanabe G, Tomita S, Takemura H, et al. Prolonged cross-clamping during aortic valve replacement is an independent predictor of postoperative morbidity and mortality: analysis of the Japan cardiovascular surgery database. Ann Thorac Surg. (2017) 103(2):602–9. doi: 10.1016/j.athoracsur.2016.06.060

PubMed Abstract | CrossRef Full Text | Google Scholar

43. Xu S, Liu J, Li L, Wu Z, Li J, Liu Y, et al. Cardiopulmonary bypass time is an independent risk factor for acute kidney injury in emergent thoracic aortic surgery: a retrospective cohort study. J Cardiothorac Surg. (2019) 14(1):90. doi: 10.1186/s13019-019-0907-x

PubMed Abstract | CrossRef Full Text | Google Scholar

44. Zhao C, Li Y, Pan G, Xu J, Liu S, Xiao Y. Risk factors for postoperative acute kidney injury in patients undergoing redo cardiac surgery using cardiopulmonary bypass. J Cardiovasc Dev Dis. (2022) 9(8):244. doi: 10.3390/jcdd9080244

PubMed Abstract | CrossRef Full Text | Google Scholar

45. Salsano A, Giacobbe DR, Sportelli E, Olivieri GM, Natali R, Prevosto M, et al. Aortic cross-clamp time and cardiopulmonary bypass time: prognostic implications in patients operated on for infective endocarditis. Interact Cardiovasc Thorac Surg. (2018) 27(3):328–35. doi: 10.1093/icvts/ivy085

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: Perceval, Intuity elite, aortic valve replacement, pressure gradient, CPB (cardiopulmonary bypass), aortic cross clamp

Citation: Wang C, Xie Y, Zhang H, Yang P, Zhang Y, Lu C, Liu Y, Wang H, Xu Z and Hu J (2023) Sutureless vs. rapid-deployment valve: a systemic review and meta-analysis for a direct comparison of intraoperative performance and clinical outcomes. Front. Cardiovasc. Med. 10:1123487. doi: 10.3389/fcvm.2023.1123487

Received: 14 December 2022; Accepted: 24 April 2023;
Published: 15 May 2023.

Edited by:

Matthias Bossard, Luzerner Kantonsspital, Switzerland

Reviewed by:

Aleksander Dokollari, Lankenau Medical Center, United States
Antonino S. Rubino, University of Campania Luigi Vanvitelli, Italy

© 2023 Wang, Xie, Zhang, Yang, Zhang, Lu, Liu, Wang, Xu and Hu. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Jia Hu humanjia@msn.com

Download