Skip to main content

ORIGINAL RESEARCH article

Front. Cardiovasc. Med., 08 January 2024
Sec. Coronary Artery Disease
Volume 10 - 2023 | https://doi.org/10.3389/fcvm.2023.1326091

A more-Comers populAtion trEated with an ultrathin struts polimer-free Sirolimus stent: an Italian post-maRketing study (the CAESAR registry)

Giuseppe Tarantini1* Francesco Cardaioli1,† Giuseppe De Iaco2 Bernardino Tuccillo3 Maria Carmen De Angelis3 Ciro Mauro4 Marco Boccalatte5 Antonio Trivisonno6 Flavio Ribichini7 Giuseppe Vadalà8 Giuseppe Caramanno9 Marco Caruso10 Mario Lombardi11 Dionigi Fischetti12 Alessandro Danesi13 Leonardo Abbracciavento14 Giulia Lorenzoni15 Dario Gregori15 Andrea Panza1 Luca Nai Fovino1 Giovanni Esposito16
  • 1Department of Cardiac, Thoracic, Vascular Sciences and Public Health, University of Padua, Padova, Italy
  • 2Department of Cardiology, Hospital “Cardinal G. Panico”, Tricase, Italy
  • 3U.O.C. Cardiologia, P.O. Ospedale del Mare, Napoli, Italy
  • 4Department of Cardiology, Hospital Cardarelli, Naples, Italy
  • 5Interventional Cardiology Unit, Ospedale Santa Maria Delle Grazie Pozzuoli, Napoli, Italy
  • 6Department of Cardiovascular Disease, “Antonio Cardarelli” Hospital, Campobasso, Italy
  • 7Division of Cardiology, AOUI Verona, Verona, Italy
  • 8Division of Cardiology, University Hospital Paolo Giaccone, Palermo, Italy
  • 9Interventional Cardiology, San Giovanni di Dio Hospital, Agrigento, Italy
  • 10Interventional Cardiology Unit, ARNAS Civico, G. Di Cristina Benfratelli, Palermo, Italy
  • 11Interventional Cardiology Unit, A.O. Riuniti Villa Sofia-Cervello, Palermo, Italy
  • 12Division of Cardiology, “V. Fazzi” Hospital, Lecce, Italy
  • 13Division of Cardiology, S. Spirito Hospital, Rome, Italy
  • 14Interventional Cardiology Unit, SS Annunziata Hospital, Taranto, Italy
  • 15Unit of Biostatistics, Epidemiology and Public Health, University of Padova, Padova, Italy
  • 16Division of Cardiology, Università Degli Studi di Napoli Federico II, Napoli, Italy

Introduction: The use of contemporary drug-eluting stents (DES) has significantly improved outcomes of patients with coronary artery disease (CAD) undergoing percutaneous coronary intervention (PCI). However, concerns exist regarding the long-term proinflammatory effects of durable polymer coatings used in most DES, potentially leading to long-term adverse events. First-generation polymer-free stent technologies, such as sirolimus- and probucol-eluting stents (PF-SES), have shown an excellent safety and efficacy profile. The aim of this study was to evaluate the safety and efficacy of the new ultrathin Coroflex ISAR NEO PF-SES, in a more-comers PCI population.

Methods: The CAESAR (a more-Comers populAtion trEated with an ultrathin struts polimer-free Sirolimus stent: An Italian post-maRketing study) registry is a multicenter, prospective study conducted in Italy, enrolling more-comers CAD patients undergoing PCI with the Coroflex ISAR NEO stent. Patients with left main (LM) disease, cardiogenic shock (CS), or severely reduced left-ventricular ejection fraction (LVEF) were excluded. The primary endpoint was target-lesion revascularization (TLR) at 1 year.

Results: A total of 425 patients were enrolled at 13 centers (mean age 66.9 ± 11.6 years, Diabetes mellitus 29%, acute coronary syndrome 67%, chronic total occlusion 9%). Of these, 40.9% had multivessel disease (MVD) and in 3.3% cases, the target lesion was in-stent restenosis (ISR). Clinical device success was reached in 422 (99.6%) cases. At 1 year, only two (0.5%) subjects presented ischemia-driven TLR. The 1-year rates of target vessel revascularization and MACE were 0.5% and 5.1%, respectively. Major bleeding was observed in four (1.0%) patients.

Conclusion: In this multicenter, prospective registry, the use of a new ultrathin Coroflex ISAR NEO PF-SES in a more-comers PCI population showed good safety and efficacy at 1 year.

Introduction

The development of contemporary drug-eluting stents (DES) has considerably improved clinical outcomes compared with both bare-metal stents (BMS) and earlier DES platform iterations (13). As a result, international guidelines now advocate the use of last-generation DES for patients with coronary artery disease (CAD) necessitating percutaneous coronary intervention (PCI) (4). However, concerns exist regarding the potential long-term pro-inflammatory effects of durable polymer coatings typically employed for drug release regulation of most contemporary DES, possibly leading to neo-atherosclerotic phenomena (2, 5, 6).

To overcome these drawbacks, polymer-free stent technologies have been implemented. A polymer-free stent eluting sirolimus and probucol (PF-SES) has been shown to be non-inferior to a new-generation polymer-based zotarolimus-eluting stent at long-term follow-up (2). In addition, sirolimus and probucol have been assessed on an ultrathin bare-metal platform, revealing favorable safety and efficacy profiles in a more-comers population (7). The Coroflex ISAR NEO (B. Braun, Melsungen, Germany) is the latest released ultrathin PF-SES platform, introducing a restyled design aimed at improving trackability and maximal stent expansion.

The aim of this study was to assess the safety and efficacy of the Coroflex ISAR NEO platform in a more-comers population including patients with de novo and restenotic lesions, both in native coronary arteries and coronary bypass grafts (CABG).

Methods

Study design and patient population

The CAESAR (a more-Comers populAtion trEated with an ultrathin strut polymer-free Sirolimus stent: An Italian post-maRketing study) registry is a prospective, national, multicenter, postmarket study enrolling patients with CAD undergoing PCI with the Coroflex ISAR NEO coronary stent at 13 Italian centers. According to the study protocol, we included patients ≥18 years of age with an acute coronary syndrome (ACS), chronic coronary syndrome (CCS), or objective proof of ischemia meeting the requirements for PCI (4). Either single- or multivessel stenting was allowed for treating de novo or restenotic lesions (reference diameters from 2.0 to 4.0 mm), both in native coronary arteries and CABG. Limited exclusion criteria were considered for the study [left main (LM) disease, cardiogenic shock (CS), severe calcified stenoses, and severely reduced left-ventricular ejection fraction (LVEF)], resulting in a more comprehensive cohort of PCI patients (more-comers). The complete study inclusion and exclusion criteria are provided in Table 1.

Table 1
www.frontiersin.org

Table 1. Inclusion and exclusion criteria.

Study device and procedure

The Coroflex ISAR NEO is the latest-generation PF-SES. Similarly to its predecessor (Coroflex ISAR), it is built on a premounted cobalt-chrome (Co-Cr) alloy ultrathin platform (7), while it presents a redesigned strut structure that enables greater stent expansion, maintaining at the same time good flexibility and radial force. The Coroflex ISAR NEO is available in 2.0 mm (55 µm thick) and 3.5 mm (65 µm thick) diameter sizes with a maximal expansion capability of up to 5.0 mm. Lesion pre-dilation was left to the discretion of the operators. The antiplatelet therapy after PCI had to follow the latest guidelines (4, 810). A dedicated electronic data capture system was used for data entry, and data accuracy was verified by the principal investigators at each center. Device success is defined as the successful delivery and deployment of the device and attainment of <50% diameter stenosis using only the study device. Procedural success was defined as a successful stent deployment with an antegrade thrombolysis in myocardial infarction (TIMI) grade 3 flow at the end of the procedure.

Study endpoints

The primary endpoint (EP) was target lesion revascularization (TLR) at 1 year defined as ischemia-driven revascularization of the target lesion (within the stent or the 5 mm borders adjacent to the stent). Secondary EPs were all-cause mortality, definite stent thrombosis (ST), bleeding events, major adverse cardiovascular events [MACE, a composite of all-cause and cardiac death, myocardial infarction (MI), and TLR], and target vessel failure [TVF, a composite of cardiac death, MI attributed to the target vessel, or target vessel revascularization (TVR)]. Myocardial infarctions were defined according to the fourth universal MI definition (11). The follow-up was prospectively performed at 30 days (±7 days) and 1 year (±30 days) with outpatient visits or telephone interviews. The definitions of individual endpoints can be found in the study protocol (Supplementary Appendix).

Statistical analysis

For the CAESAR study, the sample size calculation was based on the primary EP. The detailed description of the procedure regarding sample size calculation is reported in the Supplementary Material. Continuous variables are expressed as median [interquartile range (IQR)] and compared using the Mann–Whitney U test. Categorical variables are presented as counts (%) and were compared using the Chi-square test or Fisher's exact test. The Kaplan–Meier method was used to estimate survival and MACE-free survival at follow-up. Cumulative incidence functions (CIFs) were used to evaluate bleeding at follow-up accounting for competing risks. For all the analyses, a two-sided p < 0.05 was considered to be significant. The analyses were performed with the R software (R Foundation, Wien, Austria).

Results

Baseline clinical characteristics

Between August 2019 and March 2021, a total of 425 patients were enrolled in the CAESAR study at 13 centers. Detailed demographics and clinical characteristics are described in Table 2. In summary, patients were aged 66.9 ± 11.6 years and 75% were male. Diabetes mellitus was present in 29% of the patients (11% insulin-dependent). One-third of the study population had undergone previous percutaneous coronary artery revascularization. The median baseline LVEF was 55%. Clinical indications for PCI were ACS in 67% of the patients (>50% of the subjects were admitted because of acute MI), and 3% of them had a New York Heart Association (NYHA) Functional Class III–IV on admission.

Table 2
www.frontiersin.org

Table 2. Baseline clinical characteristics.

Angiographic and procedural characteristics

Overall, 492 lesions were treated with 484 PF-SES. Angiographic and procedural characteristics are reported in Tables 3, 4. A total of 492 lesions were treated in 425 patients, with 174 (40.9%) patients presenting multivessel disease (MVD). Left anterior descending (LAD) was involved in 43% of the cases. In 16 (3.4%) cases, the target lesion resulted from in-stent restenosis (ISR), while three lesions were located on a previous CABG. Chronic total occlusion (CTO) represented 9% of the total treated lesions. Coronary stenting was performed through a radial approach in 92.9% of the cases, mostly by using a 6F guiding catheter (98.8%). Pre- and post-dilation were performed in 76% and 66% of the cases, respectively. Clinical device success was reached in 422 (99.6%) patients. In 10 cases, the use of an adjunctive different stent platform was considered necessary. No intraprocedural strokes or deaths were reported. Furthermore, there were no recorded cases of acute bleeding or thrombotic complications. Four (1%) periprocedural MI were reported. A P2Y12 pretreatment strategy was applied in 235 (55.4%) patients. In 10 (2.4%) cases, intraprocedural Cangrelor infusion was administered. The time prescription for dual antiplatelet therapy (DAPT) at discharge was 1–3 months for 55 (13%) patients (all with an indication for OAC), 6 months for 75 (17.6%) subjects, and ≥12 months for the rest of the study population (Supplementary Table S1). At 1-year follow-up, 53% of the patients were taking two antiplatelet drugs.

Table 3
www.frontiersin.org

Table 3. Angiographic characteristics.

Table 4
www.frontiersin.org

Table 4. Procedural characteristics.

Outcomes

The rates of primary and secondary EP at different time points are provided in Table 5 and Supplementary Tables S2, S3. The 1-year follow-up was available for 410 (96.4%) patients. Only two (0.5%) subjects presented the primary EP, of those one was an early MI related to the target lesion, treated with PCI. Out of the 20 deaths (4.9%), one (5%) was cardiac in nature. The Kaplan–Meier curve for all-cause death is reported in Supplementary Figure S1. The 1-year rates of TVR and MACE were 0.5% (CI 0.1–0.9) and 6.1% (CI 4.6–7.2), respectively (Figure 1), while no ST was reported. Major bleeding was observed in four (1.0%) patients (Supplementary Figure S2).

Table 5
www.frontiersin.org

Table 5. One-year follow-up outcomes.

Figure 1
www.frontiersin.org

Figure 1. The Kaplan–Meier curve for MACE.

Discussion

The CAESAR study was the first real-world registry to evaluate the performance of the new polymer-free, ultrathin, sirolimus- and probucol-eluting stent (Coroflex ISAR NEO) in a more-comers population.

The key findings of this study can be summarized as follows: (1) The Coroflex ISAR NEO stent demonstrated excellent procedural performance, with high rates of procedural and clinical success, (2) At 1-year follow-up, TLR, TVR, and ST rates were very low.

The advent of DES marked a pivotal step in preventing in-stent restenosis. However, early-generation DES demonstrated an increased risk of both late and very late stent thrombosis when compared with BMS (12, 13). This was likely attributable to compromised arterial healing subsequent to stent implantation, a consequence of inflammatory reaction associated with the polymer and endothelial cell dysfunction (1416). In response, polymer-free (PF) stents were introduced to the market. The initial generations of PF stents displayed lower clinical effectiveness in comparison with the durable polymer DES, potentially owing to overly rapid drug elution without polymer control (17, 18). This challenge was effectively addressed in recent years by incorporating a secondary molecule to regulate drug release.

One such approach is the probucol- and sirolimus-coated polymer-free stents (19, 20). The first iteration exhibited favorable long-term (10-year) safety and efficacy when compared with high-performance durable polymer DES (2). The second generation of PF-SES (Coroflex ISAR), characterized by ultrathin struts (50/60 µm), was successfully tested in the international ISAR 2000 registry, demonstrating low event rates at 9-month follow-up in both CCS and ACS patients (7). Moreover, thanks to their ultrathin struts and the pro-reendothelializing effects of probucol, this type of device exhibited improved strut coverage at 3 months compared with a last-generation sirolimus-eluting stent with a bioresorbable degradable polymer (21).

The CAESAR registry was the first to evaluate the latest generation of PF-SES, the Coroflex ISAR NEO. In comparison with its predecessor, the new stent features a distinct structural backbone design that enhances visibility during the procedure and allows for broader expansion while maintaining radial stability, flexibility, strut thickness, and coating characteristics.

The device's procedural success reported in the present study was excellent, even in the presence of a high degree of target lesion complexity (>50% of patients presenting with type B2/C lesions, and almost 10% of treated lesions were CTOs), suggesting favorable stent performance in challenging anatomies as well. Regarding clinical outcomes, very low rates of adverse events were recorded during follow-up. Specifically, only two patients experienced the primary endpoint (TLR), and no case of stent thrombosis was reported. Comparing these findings with previous registries and trials involving PF-SES (as well as other platforms with ultrathin struts), the adverse stent-related event rate in this registry was notably lower (2, 7, 19, 22). This could be attributed to several factors, potentially synergistic: (1) the improved performance of the current device compared with its predecessors, (2) a relatively high 1-year mortality rate (especially due to non-cardiac causes), with death acting as a competing risk, and (3) the relatively small length of the treated lesions.

Nevertheless, the low rate of adverse events herein reported is reassuring in terms of the device's safety and efficacy profile. Furthermore, given the baseline high ischemic risk of the enrolled population (>50% of clinical presentations were ACS, and almost half of the enrolled patients had MVD), the limited number of ischemic events during follow-up is even more remarkable. The absence of mandatory angiographic follow-up might also explain the lower rates of TLR of the CAESAR study as compared with the previous trials [such as the ISAR-TEST 5 (7, 19)], potentially reducing the number of PCI indicated at angiographic follow-up.

Furthermore, it should be noted that, despite the baseline population's high ischemic risk (almost 70% of the cases were admitted because of unstable CAD), only 53% of the patients included in the study were prescribed a DAPT regimen at 12 months. Accordingly, the 1-year rate of major bleeding events was low and comparable with other previous registries (2, 7, 19, 22). Although the observational nature of the present analysis, along with its small sample size, does not allow for robust subanalysis on the impact of different DAPT regimens, the low event rates observed are encouraging.

The reported data must be viewed in light of the numerous inherent limitations of the study. Primarily, the potential sample size underestimation, combined with the less stringent data monitoring inherent to an observational study, might represent an adverse event under-reporting bias. As a consequence, the present results did not permit the execution of any subgroup analysis of those prespecified according to the study protocol. Secondly, the absence of a core laboratory for standardized assessment of baseline and procedural angiographies did not allow for a standardized angiogram evaluation. Similarly, the absence of an independent adjudication committee could have led to under-reported adverse event rates. Thirdly, the absence of angiographic follow-up and intravascular imaging data could have led to a reduced number of TLR reports and prevented a detailed struts coverage evaluation. Yet, indications of PCI for ISR in asymptomatic patients remain debatable. Furthermore, considering the relatively short follow-up period, very late stent thrombosis is not included as an endpoint in the current study and will be evaluated in subsequent analyses with a longer follow-up. Finally, it is important to note that the use of different DES platforms was necessary in 10 cases, potentially leading to biased outcomes. In conclusion, due to the specific inclusion and exclusion criteria and the utilization of a predetermined study platform, the findings from this analysis should not be generalized to more complex populations (e.g., severely calcified lesions or challenging anatomy) or to other devices, whether ultrathin or PF stents, which were not considered in this study.

Conclusions

The prospective CAESAR registry demonstrated the optimal safety and efficacy profile of the new-generation PF-SES Coroflex ISAR NEO in a more-comers population. This study contributes to the growing body of evidence supporting the use of PF-SES and showed promising results in terms of procedural success and low adverse event rates at 1-year follow-up. Further research is needed to confirm our results and explore the broader applicability of PF-SES in specific patient subsets, over an extended follow-up period.

Data availability statement

The original contributions presented in the study are included in the article and Supplementary Material. Further inquiries can be directed to the corresponding author.

Ethics statement

The studies involving humans were approved by UOSD Progetti e Ricerca Clinica, Azienda Ospedale-Università di Padova, Padua, Italy. The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study. Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.

Author contributions

GT: Project administration, Supervision, Writing – original draft, Writing – review and editing, Conceptualization, Data curation, Investigation, Methodology, Validation, Visualization. FC: Writing – original draft, Writing – review and editing, Conceptualization, Data curation, Formal Analysis, Investigation, Methodology, Supervision, Validation, Visualization. GI, BT, MA, CM, MB, AT, FR, GV, GC, MC, ML, DF, AD, and LA: Writing – review and editing, Writing – original draft. GL: Data curation, Formal Analysis, Methodology, Supervision, Writing – review & editing, Software, Validation, Writing – original draft. DG: Data curation, Formal Analysis, Methodology, Supervision, Writing – review & editing, Software, Validation, Writing – original draft. AP: Writing – review & editing, Data curation, Visualization, Writing – original draft. LN: Writing – original draft, Writing – review & editing, Conceptualization, Data curation, Formal Analysis, Investigation, Supervision, Visualization. GE: Writing – original draft, Writing – review & editing, Conceptualization, Data curation, Investigation, Methodology, Project administration, Supervision, Validation, Visualization.

Funding

The author(s) declare that no financial support was received for the research, authorship, and/or publication of this article.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

The author(s) declared that they were an editorial board member of Frontiers, at the time of submission. This had no impact on the peer review process and the final decision.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fcvm.2023.1326091/full#supplementary-material

References

1. Kaiser C, Galatius S, Erne P, Eberli F, Alber H, Rickli H, et al. Drug-eluting versus bare-metal stents in large coronary arteries. N Engl J Med. (2010) 363(24):2310–9. doi: 10.1056/NEJMoa1009406

PubMed Abstract | Crossref Full Text | Google Scholar

2. Kufner S, Ernst M, Cassese S, Joner M, Mayer K, Colleran R, et al. 10-year outcomes from a randomized trial of polymer-free versus durable polymer drug-eluting coronary stents. J Am Coll Cardiol. (2020) 76(2):146–58. doi: 10.1016/j.jacc.2020.05.026

PubMed Abstract | Crossref Full Text | Google Scholar

3. Tarantini G, Fovino LN, Varbella F, Trabattoni D, Caramanno G, Trani C, et al. A large, prospective, multicentre study of left main PCI using a latest-generation zotarolimus-eluting stent: the ROLEX study. EuroIntervention. (2023) 18(13):e1108–19. doi: 10.4244/EIJ-D-22-00454

PubMed Abstract | Crossref Full Text | Google Scholar

4. Neumann F-J, Sousa-Uva M, Ahlsson A, Alfonso F, Banning AP, Benedetto U, et al. 2018 ESC/EACTS guidelines on myocardial revascularization. Eur Heart J. (2019) 40(2):87–165. doi: 10.1093/eurheartj/ehy394

PubMed Abstract | Crossref Full Text | Google Scholar

5. Nakazawa G, Otsuka F, Nakano M, Vorpahl M, Yazdani SK, Ladich E, et al. The pathology of neoatherosclerosis in human coronary implants. J Am Coll Cardiol. (2011) 57(11):1314–22. doi: 10.1016/j.jacc.2011.01.011

PubMed Abstract | Crossref Full Text | Google Scholar

6. Otsuka F, Byrne RA, Yahagi K, Mori H, Ladich E, Fowler DR, et al. Neoatherosclerosis: overview of histopathologic findings and implications for intravascular imaging assessment. Eur Heart J. (2015) 36(32):2147–59. doi: 10.1093/eurheartj/ehv205

PubMed Abstract | Crossref Full Text | Google Scholar

7. Krackhardt F, Kočka V, Waliszewski MW, Utech A, Lustermann M, Hudec M, et al. Polymer-free sirolimus-eluting stents in a large-scale all-comers population. Open Heart. (2017) 4(2):e000592. doi: 10.1136/openhrt-2017-000592

PubMed Abstract | Crossref Full Text | Google Scholar

8. Hindricks G, Potpara T, Dagres N, Arbelo E, Bax JJ, Blomström-Lundqvist C, et al. 2020 ESC guidelines for the diagnosis and management of atrial fibrillation developed in collaboration with the European Association for Cardio-Thoracic Surgery (EACTS). Eur Heart J. (2021) 42(5):373–498. doi: 10.1093/eurheartj/ehaa612

PubMed Abstract | Crossref Full Text | Google Scholar

9. Arslan F, Damman P, Zwart B, Appelman Y, Voskuil M, Vos A de, et al. 2020 ESC guidelines on acute coronary syndrome without ST-segment elevation. Neth Heart J. (2021) 29(11):557–65. doi: 10.1007/s12471-021-01593-4

PubMed Abstract | Crossref Full Text | Google Scholar

10. Tarantini G, Fovino LN, Tellaroli P, Chieffo A, Barioli A, Menozzi A, et al. Optimal duration of dual antiplatelet therapy after second-generation drug-eluting stent implantation in patients with diabetes: the SECURITY (Second-Generation Drug-Eluting Stent Implantation Followed by Six- Versus Twelve-Month Dual Antiplatelet Therapy)-diabetes substudy. Int J Cardiol. (2016) 207:168–76. doi: 10.1016/j.ijcard.2016.01.068

PubMed Abstract | Crossref Full Text | Google Scholar

11. Thygesen K, Alpert JS, Jaffe AS, Chaitman BR, Bax JJ, Morrow DA, et al. Fourth universal definition of myocardial infarction (2018). Eur Heart J. (2019) 40(3):237–69. doi: 10.1093/eurheartj/ehy462

PubMed Abstract | Crossref Full Text | Google Scholar

12. Serruys PW, Daemen J. Late stent thrombosis. Circulation. (2007) 115(11):1433–9. doi: 10.1161/CIRCULATIONAHA.106.666826

PubMed Abstract | Crossref Full Text | Google Scholar

13. Tada T, Byrne RA, Simunovic I, King LA, Cassese S, Joner M, et al. Risk of stent thrombosis among bare-metal stents, first-generation drug-eluting stents, and second-generation drug-eluting stents. JACC Cardiovasc Interv. (2013) 6(12):1267–74. doi: 10.1016/j.jcin.2013.06.015

PubMed Abstract | Crossref Full Text | Google Scholar

14. Finn AV, Nakazawa G, Joner M, Kolodgie FD, Mont EK, Gold HK, Virmani R. Vascular responses to drug eluting stents. Arterioscler Thromb Vasc Biol. (2007) 27(7):1500–10. doi: 10.1161/ATVBAHA.107.144220

PubMed Abstract | Crossref Full Text | Google Scholar

15. Tarantini G, Cardaioli F. CHIP-PCI: ready for a prime time? Catheter Cardiovasc Interv. (2022) 100(5):721–2. doi: 10.1002/ccd.30466

PubMed Abstract | Crossref Full Text | Google Scholar

16. Tarantini G, Cardaioli F. Coronary complete revascularization strategy: a puzzle still hard to complete. Catheter Cardiovasc Interv. (2022) 99(3):639–40. doi: 10.1002/ccd.30125

PubMed Abstract | Crossref Full Text | Google Scholar

17. Mehilli J, Byrne RA, Wieczorek A, Iijima R, Schulz S, Bruskina O, et al. Randomized trial of three rapamycin-eluting stents with different coating strategies for the reduction of coronary restenosis. Eur Heart J. (2008) 29(16):1975–82. doi: 10.1093/eurheartj/ehn253

PubMed Abstract | Crossref Full Text | Google Scholar

18. Gershlick A, Scheerder ID, Chevalier B, Stephens-Lloyd A, Camenzind E, Vrints C, et al. Inhibition of restenosis with a paclitaxel-eluting, polymer-free coronary stent. Circulation. (2004) 109(4):487–93. doi: 10.1161/01.CIR.0000109694.58299.A0

PubMed Abstract | Crossref Full Text | Google Scholar

19. Massberg S, Byrne RA, Kastrati A, Schulz S, Pache J, Hausleiter J, et al. Polymer-free sirolimus- and probucol-eluting versus new generation zotarolimus-eluting stents in coronary artery disease. Circulation. (2011) 124(5):624–32. doi: 10.1161/CIRCULATIONAHA.111.026732

PubMed Abstract | Crossref Full Text | Google Scholar

20. Kufner S, Sorges J, Mehilli J, Cassese S, Repp J, Wiebe J, et al. Randomized trial of polymer-free sirolimus- and probucol-eluting stents versus durable polymer zotarolimus-eluting stents. JACC Cardiovasc Interv. (2016) 9(8):784–92. doi: 10.1016/j.jcin.2016.01.009

PubMed Abstract | Crossref Full Text | Google Scholar

21. Irurueta IO, Sucarrats SG, Molina JLB, Prado AP de, Massotti M, Ramírez MÁ C, et al. Can an ultrathin strut stent design and a polymer free, proendothelializing probucol matrix coating improve early strut healing? The FRIENDLY-OCT trial. An intra-patient randomized study with OCT, evaluating early strut coverage of a novel probucol coated polymer-free and ultra-thin strut sirolimus-eluting stent compared to a biodegradable polymer sirolimus-eluting stent. Int J Cardiol. (2022) 360:13–20. doi: 10.1016/j.ijcard.2022.04.043

PubMed Abstract | Crossref Full Text | Google Scholar

22. Leschke M, Nhan VT, Waliszewski M, Palacios V, Horváth I, Ivanov VA, et al. The “all comer” coroflex please drug-eluting stent registry in Europe and Asia—an overall and transcontinental assessment of the 10-month major adverse cardiac events. Indian Heart J. (2012) 64(5):453–61. doi: 10.1016/j.ihj.2012.08.002

PubMed Abstract | Crossref Full Text | Google Scholar

Keywords: coronary artery disease, PCI, ultrathin struts, acute coronary syndrome (ACS), drug-eluting stent (DES)

Citation: Tarantini G, Cardaioli F, De Iaco G, Tuccillo B, De Angelis MC, Mauro C, Boccalatte M, Trivisonno A, Ribichini F, Vadalà G, Caramanno G, Caruso M, Lombardi M, Fischetti D, Danesi A, Abbracciavento L, Lorenzoni G, Gregori D, Panza A, Nai Fovino L and Esposito G (2024) A more-Comers populAtion trEated with an ultrathin struts polimer-free Sirolimus stent: an Italian post-maRketing study (the CAESAR registry). Front. Cardiovasc. Med. 10:1326091. doi: 10.3389/fcvm.2023.1326091

Received: 22 October 2023; Accepted: 11 December 2023;
Published: 8 January 2024.

Edited by:

Marco Giuseppe Del Buono, Agostino Gemelli University Polyclinic (IRCCS), Italy

Reviewed by:

Rocco Vergallo, University of Genoa, Italy
Dejan Orlic, University of Belgrade, Serbia

© 2024 Tarantini, Cardaioli, De Iaco, Tuccillo, De Angelis, Mauro, Boccalatte, Trivisonno, Ribichini, Vadalà, Caramanno, CaTarantini, Cardaioli, De Iaco, Tuccillo, De Angelis, Mauro, Boccalatte, Trivisonno, Ribichini, Vadalà, Caramanno, Caruso, Lombardi, Fischetti, Danesi, Abbracciavento, Lorenzoni, Gregori, Panza, Nai Fovino and Espositoruso, Lombardi, Fischetti, Danesi, Abbracciavento, Lorenzoni, Gregori, Panza, Nai Fovino and Esposito. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Giuseppe Tarantini giuseppe.tarantini.1@unipd.it

These authors have contributed equally to this work

Abbreviations ACS, acute coronary syndrome; BMS, bare-metal stents; CABG, coronary artery bypass grafts; CAD, coronary artery disease; CCS, chronic coronary syndrome; CIF, cumulative incidence functions; CS, cardiogenic shock; CTO, chronic total occlusion; DAPT, dual antiplatelet therapy; DES, drug-eluting stents; EP, endpoint; ISR, in-stent restenosis; LAD, left anterior descending coronary artery; LM, left main coronary artery; LVEF, left-ventricular ejection fraction; MACE, major adverse cardiovascular events; MI, myocardial infarction; MVD, multivessel disease; NYHA, New York Heart Association; OAC, oral anticoagulation; PCI, percutaneous coronary intervention; PF-SES, polymer-free sirolimus-eluting stents; PF, polymer-free; ST, stent thrombosis; TLR, target lesion revascularization; TVF, target vessel failure; TVR, target vessel revascularization.

Download