Background: Recent data suggest that the true incidence of atrial fibrillation (AF) after patent foramen ovale (PFO) closure has probably been underestimated, and may differ according to the type of closing device used.
Objectives: On the basis of continuous rhythm monitoring with an implantable device, this study aims to assess the incidence of supraventricular arrhythmia following PFO closure with the Occlutech PFO device.
Methods: This is a multicentric analysis of consecutive PFO closure patients treated with an Occlutech device between 01/01/2019 and 20/03/2024, with an implantable loop recorder (IRL) (or a pacemaker or implantable cardioverter defibrillator) implanted for at least 3 months preceding the procedure, and with available follow-up for at least 1month post procedure. Primary endpoint was the incidence of patients with new onset supraventricular arrhythmia (AF, atrial flutter or any supraventricular tachycardia) lasting >30 s, post PFO closure.
Results: A total of 59 patients met the inclusion criteria. Patients were monitored (95% with ILR) during 284 days (IQR 241.5–374) before, and for 422 days (IQR 237–776) post PFO closure. Supraventricular arrhythmia post PFO closure was reported in 18 patients (31%), with median time-interval until arrhythmia occurrence of 16.5 days (IQR 13–21). A total of 88 supraventricular arrhythmia events (96.6% AF) were documented during follow-up. In 94.4% of patients with supraventricular arrhythmia, new-onset arrhythmia occurred in the first 45 days after PFO closure. Six patients (33.3%) with supraventricular arrhythmia post PFO closure, presented AF episodes beyond 60 days after PFO closure.
Conclusions: In this multicenter retrospective analysis of patients undergoing percutaneous PFO closure with the Occlutech PFO device, implantable continuous rhythm monitoring devices were able to diagnose new-onset supraventricular arrhythmia (97% AF) after PFO closure in 31% of patients. While 94% of new-onset supraventricular arrhythmia events occurred in the first 45 days post-procedure, one-third of patients with arrhythmia post PFO closure presented AF episodes beyond 60 days post procedure.
Introduction
While being a highly prevalent condition in the general population, there is currently solid evidence linking patent foramen ovale (PFO) to cryptogenic stroke. Several randomized controlled trials (RCT's) (1–3) showed a significant reduction in recurrent cryptogenic stroke by PFO closure compared to medical therapy. Consequently, PFO closure has become an established therapy for stroke prevention.
However, patient selection for PFO closure poses several challenges due to the complex interplay of factors that influence the procedure's benefits and risks. PFOs are found in ∼25% of the general population, but not all are causally linked to stroke. Identifying whether the PFO is related to the paradoxical embolism, or if the stroke resulted from another cause (such as occult atrial fibrillation) can be difficult. The use of implantable loop recorders (ILR's) significantly increases the detection rate of AF, and their systematic use has even be advocated in the work-up of stroke patients being considered for PFO closure (4, 5).
However, AF may also complicate PFO closure, with PFO-related AF most commonly occurring in the first 1–2 months after the procedure, with some authors taking 45 days as the cut-off for PFO-device related AF (6). The incidence of AF after PFO closure was initially reported to be in the range of 2.5%–5%, but these data were derived from AF diagnosis based on clinical symptoms (6). Recent trial data (7–9) from PFO patients monitored with ILR and/or external loop recorders post-procedure, have indicated a higher incidence, with AF occurring in approximately 20% of patients, often with a subclinical course.
Moreover, the incidence of AF after PFO closure seems to differ according to the type of closing device used (10). Herein, we report the results of a retrospective, multicenter registry of patients treated with the Occlutech PFO device, undergoing systematic rhythm monitoring using implantable continuous monitoring devices both before and after the procedure.
Methods and materials
Study design and population
All consecutive patients, treated with percutaneous PFO closure in 4 non-academic hospitals and 1 academic hospital between 01/01/2019 and 20/03/2024, were retrospectively reviewed for inclusion. Diagnosis of PFO was established with transthoracic and/or transesophageal echocardiography, by confirmation of significant right-to-left shunting using agitated saline injection, with and without Valsalva maneuver. A severe shunt was characterised as passage of more than 20 microbubbles. Indication for PFO closure was established by the local heart team involving a neurologist, with secondary prevention of cardio-embolic ischemic stroke, peripheral embolic events and platypnea-orthodeoxia syndrome as potential indications for closure. Patients were eligible if they were carrying an IRL (or a pacemaker or implantable cardioverter defibrillator) for at least 3 months before the PFO closure procedure. ILR use was generally part of an AF rule-out strategy. The PFO closure procedure was performed according to local standard procedures; both general or local anesthesia (+sedation) could be used, as well as imaging guiding with (mini) transesophageal or intracardiac echocardiography. Implantation of an Occlutech PFO device was mandatory for inclusion. Device size was decided by the implanting operator, with or without balloon sizing.
After the procedure, patients needed to have heart rhythm follow-up with ILR (or pacemaker/ICD) for at least 1 month, but results of ILR (or pacemaker/ICD) monitoring beyond 1 month were also collected. All ILR (or pacemaker/ICD) EGM tracings from both before and after PFO closure were reviewed by three independent cardiologists (including an electrophysiologist) for arrhythmia analysis (core laboratory analysis). Atrial fibrillation was defined as significant if an episode of irregularly irregular rhythm without distinct P waves lasting >30 s was documented on the EGM. For all patients, relevant demographic, clinical and procedure-related data were also collected from the patient file. This study was carried out in accordance with the Declaration of Helsinki. The ethical committee of the University Hospital of Brussels approved the study, which was subsequently also approved by the local ethics committees of the other enrolling institutions. Patient's consent was waived because of the retrospective character of the trial.
Study endpoints
The primary endpoint of the study was the proportion of patients with new onset, significant (lasting >30 s) supraventricular arrhythmia [AF, atrial flutter or any supraventricular tachycardia (SVT)] post PFO closure.
The secondary endpoints were the proportion of patients with new onset, significant (lasting >30 s) supraventricular arrhythmia (AF, atrial flutter, or any SVT), in the first 30 days, between 30 and 60 days, and beyond 60 days post PFO closure.
Device
The Occlutech PFO device is a self-centering device made of nitinol wires forming two flexible retention discs. Its efficacy and safety have been shown in several trials (11, 12), and it is currently one of the most frequently used devices for PFO closure.
Four device sizes are available, with the right atrial disc ranging from 18 to 35 mm. Compared to the Abbott Amplatzer PFO Occluder, it has a ball-forceps connection for attaching the device to the pusher, and it has a larger left atrial disc for each size.
Statistics
Statistical analyses was performed using R software (R Foundation for Statistical Computing, Vienna, Austria) and IBM SPSS Statistic for Windows (Armonk, NY: IBM Corp.). Normal distribution of continuous variables was assessed by Shapiro-Wilk test. Continuous variables were presented as medians [interquartiles (IQR)] for non-normally distributed variables, or means [+- standard deviation (SD)] for normally distributed variables. Categorical variables were presented as percentages.
Continuous variables were compared using the unpaired t test, Mann–Whitney or the Wilcoxon signed-rank test as appropriate. Categorical variables were compared using the chi-square test or Fisher's exact test, as appropriate. A P value less than 0.05 was considered significant. Kaplan–Meier survival curves were constructed to evaluate the time to the onset of new supraventricular arrhythmia.
Results
A total of 63 patients met the inclusion criteria, but after a secondary read of all EGM's by the core lab, 4 patients were excluded because of the presence of atrial fibrillation before PFO closure, leaving a final study population of 59 patients.
The baseline population characteristics are described in Table 1. There were no significant differences in the clinical profile of patients who developed, vs. those who did not develop supraventricular arrhythmia. The median age of the population was 55 years (IQR 45–59), with a slight male predominance (54%). The median CHA2DS2-Vasc and ROPE score were calculated based on available baseline characteristics, and were 3 (IQR 2–4) and 6 (IQR 5–7), respectively. Thirty-three patients (56%) were treated with the 27–30 mm Occlutech PFO device, 19 patients (32%) with the 23–25 mm device and 7 patients (12%) with the 31–35 mm device.
Continuous rhythm monitoring was obtained mainly using ILR (95%), with the Medtronic Reveal LINQ™ being the most frequently used device (68% of ILR use). Monitoring with a pacemaker or ICD was available in 5% of the patients (Table 2). Patients were monitored during a median period of 284 days (IQR 241.5–374), before PFO closure was performed.
After PFO closure, rhythm monitoring was available for 422 days (IQR 237–776). Analysis of implanted device monitoring, showed supraventricular arrhythmia after PFO closure (AF, atrial flutter or SVT) in 18 patients (31%) (Table 3). A total of 88 supraventricular arrhythmia events were documented during follow-up, with a median duration of 4.5 min (IQR 2–106.5): 96.6% (N = 85), 2.3% (N = 2), and 1.1% (N = 1) were identified as AF, atrial flutter, and SVT, respectively. The 3 non-AF supraventricular arrhythmia episodes all occurred in the first 30 days after closure. All supraventricular arrhythmia events beyond 30 days were AF. Patients experienced a median number of 2 (IQR 1–5.5) arrhythmia episodes, with a median cumulative supraventricular arrhythmia burden per patient of 144 min (IQR 13.75–357). The median time-interval between PFO closure and supraventricular arrhythmia occurrence was 16.5 days (IQR 13–21). Figure 1 shows the arrhythmia-free survival post PFO closure. Supraventricular arrhythmia occurred in the first 30 days after PFO closure in 88.9% (16/18) of patients, between 30 and 60 days after PFO closure in 5.6% (1/18), and beyond 60 days after PFO closure in 5.6% (1/18). Notably, 94.4% of new supraventricular arrhythmia episodes occurred in the first 45 days post procedure.
A total of 59 arrhythmia episodes were documented in the first 30 days after PFO closure, 15 episodes occurred between 30 and 60 days, and 14 beyond 60 days. The number of supraventricular arrhythmia episodes per patient declined from 3 (IQR 1–3) in the first 30 days, to 2 (IQR 1–2.5) between 30 and 60 days and 1.5 (IQR 1–2.75) beyond 60 days. While new-onset AF beyond 60 days post-procedure was diagnosed in only 1 (5.55%) of patients, AF recurrence beyond 60 days was detected in 5 (27.8%) patients with previously detected post-procedural supraventricular arrhythmia. Therefore, a total of 6 (33.3%) patients with supraventricular arrhythmia post PFO closure, presented AF episodes beyond 60 days after PFO closure. The median supraventricular arrhythmia episode duration was both 4 min in the first 30 days (IQR 2–107) and between 30 and 60 days (IQR 2–59) post closure; beyond 60 days median arrhythmia duration was 63 min (IQR 5–344.75). Figure 2 depicts the supraventricular arrhythmia burden and timing per patient.
Immediately post-PFO closure, 98.3% of patients were treated with dual anti-platelet therapy (aspirin + clopidogrel) (Supplementary Table 1). After diagnosis of supraventricular arrhythmia post PFO closure, only 55.6% of these patients were switched from antiplatelet therapy to a direct anticoagulant (DOAC).
After PFO closure, baseline anti-arrhythmic treatment with betablockers or amiodarone was prescribed in respectively 13.57% and 1.69% of patients (Supplementary Table 1). In 6 of 18 patients with supraventricular arrhythmia post PFO closure (33.3%), a new anti-arrhythmic drug was initiated during follow-up, with sotalol being the most frequently prescribed agent (57.14%).
In only 1 patient, PFO closure was associated with a peri-procedural complication (pseudo-aneurysm of the A. femoralis). For more details on hemorrhagic, thromboembolic or cardiovascular complications during follow-up, we refer to Supplementary Table 2.
Discussion
Increasing data suggest that the true incidence of AF after percutaneous PFO closure is underestimated. Earlier studies reported an incidence of AF of 2.5%–5%, but these were mainly based on the occurrence of symptomatic AF, or AF was an incidental finding during follow-up. The primary finding of this first multicenter study, utilizing continuous rhythm monitoring with implantable devices both before and after PFO closure, is the detection of supraventricular arrhythmia in 31% of patients following closure with a single device type, the Occlutech PFO device. AF was by far the most frequent arrhythmia reported, accounting for nearly 97% of arrhythmia events.
These results are in line with other recent reports, using long-term continuous rhythm monitoring. Guedeney et al. (7) reported an incidence of 21% of new onset supraventricular arrhythmia post PFO closure using continuous monitoring devices, but only 37% were monitored with an ILR. In the ILR subpopulation, an incidence of 29% was also observed. Only 8.4% of these patients had an ILR implanted prior to the procedure as part of an initial AF rule-out strategy. More recently, a study including 35 patients monitored with ILR reported an incidence of new onset AF post PFO closure of 31% (9). Notably, the median duration of rhythm monitoring prior to PFO closure was 184 days (IQR: 98; 280 days).
With a median of 284 days (IQR 241.5–374 days) of monitoring before PFO closure, and with a per-protocol core-lab second read of all EGM's pre- and post-PFO closure, our study allowed AF to be more thoroughly excluded, allowing to characterize a very selected population of true “PFO-related stroke patients”.
Our observed ILR-based incidence rate of AF is in line with previous studies using ILR for AF detection in patients with cryptogenic stroke. The CRYSTAL AF and the STROKE AF trial (4, 13), which both used ILR, observed a similar 6-fold increase rate of AF diagnosis following cryptogenic stroke, in comparison to clinical follow-up.
Whereas a CHA2DS2-Vasc score of at least 2 (because of history of stroke) theoretically warrants anticoagulation, the recent NOAH-AFNET 6 and ARTESIA trials (14, 15) suggested that anticoagulation may be less beneficial in patients with subclinical, device detected AF. Interestingly, in our population only 55.6% of patients with supraventricular arrhythmia post PFO closure were switched to a DOAC. It remains speculative if the treating physician considered these arrhythmia episodes to be of low thrombo-embolic risk, or if this was related to arrhythmia underdiagnosis. Hence, more clinical outcome data are needed in this specific subpopulation of patients with AF post PFO closure, which would help to optimize anti-thrombotic strategy. The long-term follow-up results of the PFO-AF trial (16) might shed more light on this issue.
In this study, 94% of new-onset supraventricular arrhythmia occurred in the first 45 days post procedure, with a median time-interval of 16.5 days. This timing is strongly suggestive of a device-related pathogenesis. PFO device-related AF is generally considered benign and self-limiting, resulting from local irritation, metallic component interference, tissue stretch, and perhaps nickel hypersensitivity (6).
Data from randomized trials have shown that 72% of these new-onset AF episodes spontaneously resolve within 45 days after PFO closure (17). Our data are in line with this observation. However, while only 1 patient (5.55%) experienced new-onset AF beyond 60 days post-procedure, 5 other patients with supraventricular arrhythmia arising in the first 45 days were detected with recurrent AF beyond 60 days. Thus, among the patients experiencing supraventricular arrhythmia post procedure, up to 33% suffered from AF beyond 60 days post procedure. In addition, the burden of AF (number of AF episodes, and median duration of AF episodes) remains considerable beyond 60 days (Figure 2).
While anti-arrhythmic treatment at baseline (or initiation during follow-up) was not infrequent in our study population, its effect on occurrence of supraventricular arrhythmia post PFO closure remains unclear. Interestingly, in the recently published AFLOAT trial, flecainide did not prevent atrial arrhythmia after PFO closure (18).
Nevertheless, the observations on early and late occurrence of AF post PFO closure could potentially be relevant for the management of these patients; prolonged (beyond 45–60 days) monitoring, and possibly anticoagulation, might be considered when a diagnosis of supraventricular arrhythmia is made. Future research is warranted focusing on identifying risk factors for delayed AF occurrence (and recurrence) post PFO closure. In this regard, despite lacking statistical significance in this population with relatively low sample size, early supraventricular arrhythmia burden and number of episodes could be particularly interesting for further exploration.
Despite being widely used in clinical practice, there is only sparse prior data on Occlutech PFO device-related arrhythmia. The device was also not used in the landmark randomized PFO trials. In their long-term observational follow-up data of the Occlutech PFO device, Snijder et al. reported a 6.6% incidence of new onset of clinical (thus without ILR) supraventricular arrhythmia post PFO closure, among which 2.6% occurred more than 2 years after closure (19). The Occlutech PFO device has also been compared with the Amplatzer PFO device, which shares several design similarities. This comparison included also a Holter monitoring at 1-month follow-up (20). In this observational registry, supraventricular arrhythmias occurred more frequently with the Amplatzer PFO device (17% vs. 9%). In the most recent trial from Guedeney et al. (7) using continuous rhythm monitoring, Occlutech PFO devices were used alongside mostly Abbott and Gore PFO devices, but the use of different device types (including also multi-fenestrated and ASD devices) hampered interpretation. Future trials should aim to compare the incidence rate of AF of different PFO devices.
Study limitations
This study may have a selection bias increasing the inclusion of patients with predisposition to AF, inherent to the retrospective design with inclusion limited to patients with ILR implantation before PFO closure.
The sample size was relatively small, despite the multicentric design of the study, which reduced the statistical power to perform regression analyses. As a result, we were unable to adjust for potential confounders or to explore independent predictors of early or late supraventricular arrhythmia onset, which may affect the generalizability and robustness of our findings.
Conclusion
In this multicenter retrospective analysis of patients undergoing percutaneous PFO closure with the Occlutech PFO device, implantable continuous rhythm monitoring devices were able to diagnose new-onset supraventricular arrhythmia (of which 97% AF) after PFO closure in 31% of patients. While 94% of these new-onset arrhythmia events occurred in the first 45 days post-procedure (median 16.5 days), one-third of patients with supraventricular arrhythmia post PFO closure presented AF episodes beyond 60 days after the procedure. Future research should focus on the prevention and management of late-occurring AF post PFO closure.
Data availability statement
The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.
Ethics statement
The studies involving humans were approved by the Medical Ethics Committee, UZ Brussel, Brussels, Belgium. The studies were conducted in accordance with the local legislation and institutional requirements. The ethics committee/institutional review board waived the requirement of written informed consent for participation from the participants or the participants' legal guardians/next of kin, because of the retrospective character of the trial.
Author contributions
SL: Conceptualization, Data curation, Formal analysis, Investigation, Methodology, Resources, Supervision, Writing – original draft, Writing – review & editing. ADM: Data curation, Formal analysis, Writing – original draft, Writing – review & editing. XG: Data curation, Formal analysis, Writing – original draft, Writing – review & editing. AM: Formal analysis, Methodology, Writing – original draft, Writing – review & editing. DP: Writing – original draft, Writing – review & editing. KH: Data curation, Writing – original draft, Writing – review & editing. FG: Writing – original draft, Writing – review & editing. JC-F: Data curation, Writing – original draft, Writing – review & editing. LR: Data curation, Writing – original draft, Writing – review & editing. PM: Writing – original draft, Writing – review & editing. MP: Data curation, Writing – original draft, Writing – review & editing. P-JP: Data curation, Writing – original draft, Writing – review & editing. FD: Data curation, Writing – original draft, Writing – review & editing. EW: Data curation, Project administration, Software, Writing – original draft, Writing – review & editing. JP: Data curation, Writing – original draft, Writing – review & editing. BR: Methodology, Software, Writing – original draft, Writing – review & editing. PU: Formal analysis, Methodology, Writing – review & editing. SD: Methodology, Writing – original draft, Writing – review & editing. J-FA: Methodology, Writing – review & editing. SD: Methodology, Writing – original draft, Writing – review & editing. GC: Writing – original draft, Writing – review & editing. BC: Writing – original draft, Writing – review & editing.
Funding
The author(s) declare that no financial support was received for the research and/or publication of this article.
Conflict of interest
SL received speaker/proctoring/consulting honoraria from Abbott, Occlutech and Boston Scientific. GC received speaker/proctor/teaching honoraria from Biosense Webster, Biotronik, Abbott, Medtronic, Boston Scientific.
The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declare that no Generative AI was used in the creation of this manuscript.
Publisher's note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
Supplementary material
The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fcvm.2025.1541923/full#supplementary-material
Abbreviations
PFO, patent foramen ovale; RCT, randomized controlled trial; AF, atrial fibrillation; ILR, implantable loop recorder; ICD, implantable cardioverter defibrillator; EGM, electrogram; SVT, supraventricular tachycardia; OAC, oral anticoagulation; DOAC, direct anticoagulants.
References
1. Mas JL, Derumeaux G, Guillon B, Massardier E, Hosseini H, Mechtouff L, et al. Patent foramen ovale closure or anticoagulation vs. antiplatelets after stroke. N Engl J Med. (2017) 377:1011–21. doi: 10.1056/NEJMoa1705915
2. Sondergaard L, Kasner SE, Rhodes JF, Andersen G, Iversen HK, Nielsen-Kudsk JE, et al. Patent foramen ovale closure or antiplatelet therapy for cryptogenic stroke. N Engl J Med. (2017) 377:1033–42. doi: 10.1056/NEJMoa1707404
3. Saver JL, Carroll JD, Thaler DE, Smalling RW, MacDonald LA, Marks DS, et al. Long-term outcomes of patent foramen ovale closure or medical therapy after stroke. N Engl J Med. (2017) 377:1033–42. doi: 10.1056/NEJMoa1610057
4. Sanna T, Diener HC, Passman RS, Di Lazzaro V, Bernstein RA, Morillo CA, et al. Cryptogenic stroke and underlying atrial fibrillation. N Engl J Med. (2014) 370:2478–86. doi: 10.1056/NEJMoa1313600
5. Diener HC, Wachter R, Wong A, Thijs V, Schnabel RB, Ntaios G, et al. Monitoring of atrial fibrillation prior to patent foramen ovale closure after cryptogenic stroke. Int J Stroke. (2023) 18(4):400–7. doi: 10.1177/17474930221124412
6. Apostolos A, Tsiachris D, Drakopoulou M, Trantalis G, Oikonomou G, Ktenopoulos N, et al. Atrial fibrillation after patent foramen ovale closure: incidence, pathophysiology, and management. J Am Heart Assoc. (2024) 13(9):e034249. doi: 10.1161/JAHA.124.034249
7. Guedeney P, Laredo M, Zeitouni M, Hauguel-Moreau M, Wallet T, Elegamandji B, et al. Supraventricular arrhythmia following patent foramen ovale percutaneous closure. J AM Coll Cardiol Intv. (2022) 15:2315–22. doi: 10.1016/j.jcin.2022.07.044
8. Imtiaz K, Ebrahim M, Niu J, Roberts J. The incidence of atrial fibrillation after percutaneous patent foramen ovale closure detected by implantable loop recorders. J Soc Cardiovasc Angiogr Interv. (2024) 3(7):101930. doi: 10.1016/j.jscai.2024.101930
9. von Falkenhausen AS, Keidel LM, Wischmann J, Höglinger G, Massberg S, Kääb S, et al. Atrial fibrillation after patent foramen ovale closure in patients with embolic stroke of undetermined source. Europace. (2024) 26(6):euae141. doi: 10.1093/europace/euae141
10. Vukadinović D, Scheller B, Ukena C, Ewen S, Mahfoud F, Böhm M. Device-related risk of atrial fibrillation after closure of patent foramen ovale: a systematic review and meta-analysis. Clin Res Cardiol. (2022) 111(5):583–7. doi: 10.1007/s00392-021-01964-2
11. Aytemir K, Oto A, Ozkutlu S, Kaya EB, Canpolat U, Yorgun H, et al. Early-mid term follow-up results of percutaneous closure of the interatrial septal defects with occlutech figulla devices: a single center experience. J Interv Cardiol. (2012) 25:375–81. doi: 10.1111/j.1540-8183.2012.00734.x
12. Van den Branden BJL, Post MC, Plokker HWM, Ten Berg JM, Suttorp MJ. Percutaneous atrial shunt closure using the novel occlutech figulla device: 6-month efficacy and safety. J Interv Cardiol. (2011) 24:264–70. doi: 10.1111/j.1540-8183.2010.00619.x
13. Bernstein RA, Kamel H, Granger CB, Piccini JP, Sethi PP, Katz JM, et al. Effect of long-term continuous cardiac monitoring vs. usual care on detection of atrial fibrillation in patients with stroke attributed to large- or small-vessel disease: the STROKE-AF randomized clinical trial. JAMA. (2021) 325(21):2169–77. doi: 10.1001/jama.2021.6470
14. Kirchhof P, Toennis T, Goette A, Camm AJ, Diener HC, Becher N, et al. Anticoagulation with edoxaban in patients with atrial high-rate episodes. N Engl J Med. (2023) 389(13):1167–79. doi: 10.1056/NEJMoa2303062
15. Healey JS, Lopes RD, Granger CB, Alings M, Rivard L, McIntyre WF, et al. Apixaban for stroke prevention in subclinical atrial fibrillation. N Engl J Med. (2024) 390(2):107–17. doi: 10.1056/NEJMoa2310234
16. Badoz M, Derimay F, Serzian G, Besutti M, Rioufol G, Frey P, et al. Incidence of atrial fibrillation in cryptogenic stroke with patent foramen ovale closure: protocol for the prospective, observational PFO-AF study. BMJ Open. (2023) 13(9):e074584. doi: 10.1136/bmjopen-2023-074584
17. Ntaios G, Papavasileiou V, Sagris D, Makaritsis K, Vemmos K, Steiner T, et al. Closure of patent foramen ovale versus medical therapy in patients with cryptogenic stroke or transient ischemic attack. Stroke. (2018) 49:412–8. doi: 10.1161/STROKEAHA.117.020030
18. Hauguel-Moreau M, Guedeney P, Dauphin C, Auffret V, Clerc JM, Marijon E, et al. Flecainide to prevent atrial arrhythmia after patent foramen ovale closure: AFLOAT study, a randomized clinical trial. Circulation. (2024) 150(21):1659–68. doi: 10.1161/CIRCULATIONAHA.124.071186
19. Snijder RJR, Renes LE, Suttorp MJ, Ten Berg JM, Post MC. Percutaneous patent foramen ovale closure using the occlutech figulla device: more than 1,300 patient-years of follow up. Catheter Cardiovasc Interv. (2019) 93(6):1080–4. doi: 10.1002/ccd.27984
Keywords: PFO (patent foramen ovale), atrial fibrillation, structural cardiology, occlutech PFO device, implantable loop recorder (ILR)
Citation: Lochy S, Del Monte A, Galloo X, Motoc A, Plein D, Hermans K, Gorré F, Colas-Florial J, Rosseel L, Muyldermans P, Pauwelyn M, Palmers P-J, Delvoye F, Wirix E, Podevyn J, Roosens B, Unger P, Droogmans S, Argacha J-F, De Raedt S, Chierchia GB and Cosyns B (2025) A multicenter analysis of implantable monitoring device-based diagnosis of supraventricular arrhythmia post patent foramen ovale closure: the OCCL-ILR study. Front. Cardiovasc. Med. 12:1541923. doi: 10.3389/fcvm.2025.1541923
Received: 9 December 2024; Accepted: 17 March 2025;
Published: 4 April 2025.
Edited by:
Fabien Praz, University Hospital of Bern, SwitzerlandReviewed by:
Mario Matta, AOU Città della Salute e della Scienza, ItalyVassil Traykov, Acibadem City Clinic Tokuda Hospital, Bulgaria
Copyright: © 2025 Lochy, Del Monte, Galloo, Motoc, Plein, Hermans, Gorré, Colas-Florial, Rosseel, Muyldermans, Pauwelyn, Palmers, Delvoye, Wirix, Podevyn, Roosens, Unger, Droogmans, Argacha, De Raedt, Chierchia and Cosyns. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Stijn Lochy, c3Rpam4ubG9jaHlAdXpicnVzc2VsLmJl