Abstract
Cell identification and enumeration are essential procedures within clinical and research laboratories. For over 150 years, quantitative investigation of body fluids such as counts of various blood cells has been an important tool for diagnostic analysis. With the current evolution of point-of-care diagnostics and precision medicine, cheap and precise cell counting technologies are in demand. This article reviews the timeline and recent notable advancements in cell counting that have occurred as a result of improvements in sensing including optical and electrical technology, enhancements in image processing capabilities, and contributions of micro and nanotechnologies. Cell enumeration methods have evolved from the use of manual counting using a hemocytometer to automated cell counters capable of providing reliable counts with high precision and throughput. These developments have been enabled by the use of precision engineering, micro and nanotechnology approaches, automation and multivariate data analysis. Commercially available automated cell counters can be broadly classified into three categories based on the principle of detection namely, electrical impedance, optical analysis and image analysis. These technologies have many common scientific uses, such as hematological analysis, urine analysis and bacterial enumeration. In addition to commercially available technologies, future technological trends using lab-on-a-chip devices have been discussed in detail. Lab-on-a-chip platforms utilize the existing three detection technologies with innovative design changes utilizing advanced nano/microfabrication to produce customized devices suited to specific applications.
Introduction
Precise determination of cells in a sample is important for a broad field of applications, such as tissue culture studies by microbiologists and disease progression in medical laboratories (OYAMA and EAGLE, ; Houwen, ). Early medical practitioners realized the importance of counting blood cells as a tool for investigation and quantitative study in healthcare (Verso, ). Prior to the advent of microscopy, bacterial microorganisms were enumerated with the naked eye by means of colony counts. This is restricted by culture growth time since most microorganisms take long to form visible colonies. Subsequently, cells were counted manually using a microscope. For over 100 years, cell biologists have been using a hemocytometer (counting chambers) to quantify cells (Bard, ). This lets the user count the number of cells manually, within a volume of liquid, under a microscope. Mass manufacturing techniques led to addition of laser etched Neubauer grids in modern slides and made them cheaper (available retail for < $ 100 from Sigma Aldrich). Despite the gradual replacement of manual counting methods with automated counters due to their ability to process multiple specimen types including blood and urine, hemocytometer is still a mainstay of all cell biology laboratories as of now (Hsiung et al., ).
Nowadays, there are three main commercial methods (detection principles) used in automated counters. The first is the automated detection and analysis using electrical properties as the basis of distinguishing cells. Coulter counter uses the impedance properties of the cell to make the analysis; cells are detected every time they cross through an aperture (orifice), typically around a few micrometers in size. Modern coulter counters are capable of measuring cells even smaller than 1 μm (Yang and Yamamoto, ). Bacterial cells, typically much smaller than mammalian cells, can be detected with coulter counters. Coulter counter was the first electronic cell counter which was widely accepted as an alternative to manual counting. Bull et al. showed clinical relevance of early coulter counters in making accurate and reproducible platelet counts (Bull et al., ).
Optical Flow cytometer, another widely used commercial method available in the market, is based on optical cell detection (Lee et al., ). It is used for management of patient diagnosis and prognosis, such as for blood cancers like leukemia and Lymphoma (Jelinek et al., ). They provide large scale, automated quantification of cells based on their optical parameters on a cell by cell basis. Laser light is illuminated on the particles in the sample stream; thereby, getting scattered and measured by optical detectors. Each event or cell in the flow produces characteristic information or parameter based on its light scattering and/or fluorescent properties, hence optical flow cytometers can simultaneously provide multi-parametric analysis with data on cell morphology and chemical characteristics (Cho et al., ). Several optical flow cytometers have an added feature which allows them to physically separate cells of interest from a heterogeneous sample, based on their optical properties and are categorized as Fluorescence assisted cell sorting (FACS) systems. They have a low coefficient of variation (CV) when counting cells below the range of 30 cells/μL (Sandhaus et al., ). Krediet et al. reported high precision and accuracy while using optical flow cytometer in comparison to manual and coulter counting methods (Krediet et al., ). They used Millipore Guava™ flow cytometer for counting algal cells and the device was able to perform quantification over a wide range of cell concentrations. The system provides automated processing (rinsing and mixing) of 96-well plate samples providing low operator time and high throughput. The Food and Drug Administration (FDA) has approved Flow cytometry as a clinical detection technique which will help manufacturers target more devices for clinical hematology and circulating tumor cell (CTC) detection (Goodin, ). Flow cytometers and Coulter counters have earned a place in research and clinical laboratories despite their bulky size and costs, which is explained due to the rise in demand for precision high-end cell counters for high throughput counts. In general, these devices can cost up to tens of thousands of dollars, which makes them unfeasible for developing countries or for portable point-of-care diagnostics.
The third most common type of device utilizes digital image analysis using a photo-microscope (O'Brien et al., ). These devices, readily available from established manufacturers like ThermoFischer scientific (i.e., Countess II automated cell counter) and BioRad (TC20 automated cell counter), contain an optical system with autofocus, capable of detecting cells using integrated software to characterize particles in terms of their fluorescent intensity and size while having a processing time of few seconds (Goldberg, ; Gupta et al., ). Image cytometry devices can either analyze static samples or perform kinetic analysis, in which case it is also referred to as imaging flow cytometry (Lanigan et al., ). Imaging flow cytometry gained commercial use after technological improvements in optics and detection systems that allowed the imaging of moving cells to stay in focus. There are some examples of small full blood counting systems utilizing image cytometry, such as HemoCUE WBC system, which can make cell counts with clinical accuracy and have been certified by the European Conformity CE marking (Lohman et al., ). Lui et al. developed a method for CTC detection in hepatocellular carcinoma using image flow cytometry (Liu et al., ). They performed fluorescent imaging of cells and enumeration of CTCs based on morphological criteria. In research laboratories, images of cells obtained using microscopy are also commonly analyzed and quantified using open source software (Hennig et al., ). Cell profiler, ImageJ, and other such open-source software, are routinely used by cell biologists for cell count from thousands of images, by performing object identification through segmentation, thresholding, recognition and division of clumped cells and other processing steps, which are user customizable.
A brief comparison of the different counting methods is summarized in Table 1. Values of sample concentration for various devices obtained by correspondence from company.
Table 1
| Detection principle | Example | Sample concentration (cells/mL) | Price range ($) | Label needed | Analysis time/throughput |
|---|---|---|---|---|---|
| Manual counting | Sigma-Aldrich, BrightLine Hemocytometer | 2 × 105-2.5 × 106 | $ | Yes—Trypan Blue, Methylene Blue, Erythrosin B, Nigrosin, Saffrarin | Low |
| SKC,Inc, C-Chip™ Disposable Hemacytometers | |||||
| Impedance system | Scepter 2.0- Sigma-Aldrich | 2 × 103-1 × 106 | $$-$$$ | NO | High |
| Multisizer 4 COULTER COUNTER | |||||
| Optical system | ACCURI c6- BD | 1 × 103-1 × 106 | $$$ | Both- Label free for scattering ; Label Needed for absorption and fluorescence | High |
| Guava- Millipore | |||||
| Image cytometer | Countess II-Fischer scientific | 1 × 104-1 × 107 | $$ | Yes- for cell identification | Low |
| Cellometer T4- Nexcelom |
Summary of commercial cell counting principles and specifications.
With the rise of microfluidic devices, several attempts have been made to develop cell counters embedded in a chip. These devices show great promise as portable and low-cost diagnostic tools (Chin et al., ). Microfluidics finds itself organically linked with cell manipulation and analysis because of the obvious size match between microfluidics- leveraging micrometer scale channels to process low volume fluid samples (Whitesides, ). However, improved stability over longer periods of time is necessary for the technology to become mainstream and find commercial success. Many studies have been done for sample enumeration in a single microfluidic platform (van Berkel et al., ). Currently, most of the microfluidics based cell counting systems available in the market, such as FX500 by Sony Biotech and Ampha z32 by Amphasys, use a microfluidic chip or cartridge integrated within conventional cytometry system. Such designs offer simplicity and flexibility since they feature a replaceable fluidics system with microchannels for sample flow, which is easy to set-up and maintain. For microfluidic devices to supplant current cytometers, it will require more development in sample preparation and detection techniques within a lab-on-a-chip microfluidic platform. The timeline in Figure 1 shows an overview of all cell counting methodologies.
Figure 1
Clinical Significance
Clinics and hospitals worldwide widely use cell counting to determine the health of a patient. Hematological analysis of body fluids can provide valuable diagnostic information and indicate medical conditions. Physician can use the results from these tests to learn a great deal about health and help in discovering the problem in its early phase. Since blood is rich with cells and proteins, among other biological substances, and provides important information about health, it has become a form of liquid biopsy performed on patients for cases such as monitoring an anemia, infection, cancer and a wide gamut of disorders. The full blood count is therefore routinely performed for any measure of diagnoses in clinics and it involves counting all the cells in the blood sample (Houwen, ). This determines the composition of various constituents within the blood such as red blood cell (RBC), white blood cell (WBC), and platelets, which give insight into the health status of a patient. For example, low white blood cell count could be a cause of a bigger underlying issue like infections with Human immunodeficiency virus (HIV), which uses CD4+ T Lymphocyte cells for replication and hence leads to critically low WBC count (Kannel et al., ). Hence, methods for CD4+ T cell enumeration are important for HIV disease staging and treatment monitoring, which makes portable label-free cell cytometers a necessity especially in resource limited setting where the disease has a higher prevalence (Cheng et al., ). Multi-parametric flow cytometry has become the mainstay of most hematologic malignancies (Braylan, ) and automated counters are standard inventory of most medical diagnostic laboratory. Automated cell counters like Sysmex XE-200 or Beckman coulter DxH 600 are extensively used by laboratory technologists for analyzing most body fluid samples due to their low turnaround times and improved precision, as opposed to manual hemocytometer. Other instruments such as the CSF cell counter GlowCyte by Advanced Instruments are marketed for specialized sample analysis in clinics and offer comprehensive results with patient reports. These devices follow validation and verification protocols set by international regulations, like the FDA in USA and diagnostic devices directive in EU, to ensure universal standards are set for the laboratory method (Verbrugge and Huisman, ). Cell counting and analysis of CTCs is also a widely popular field of research due to the prognostic information in cancer patients. For example, it was shown that the detection of CTCs using flow cytometry can be a predictor of survival in patients (Nowakowski et al., ). The use of FDA approved CTC enumeration by CellSearch™ (Menarini Silicon Biosystems) opened the door for rare cell characterization for drug development and as a prognostic marker for personalized therapies which has led to several such attempts at quantification of rare cells (Maertens et al., ).
Hemocytometer
After the modern hemocytometer was initially invented in the late nineteenth century, it has been used extensively for cell enumeration. The device evolved from being used by clinical practitioners to analyze blood samples to becoming a common tool in laboratories and clinics all over the world. It is still commonly used for counting cells due to its low cost and versatility.
The history of hemocytometer is intertwined with hematology, as the study of blood associated diseases was of interest to the early physicians (Verso, ). The first such device which resembled the modern hemocytometer was invented by Louis Charles Malassez in 1874 and this was followed by several iterations to make improvements to the design. He created a counting chamber glued to glass slide with markings that correlated the length to volume of sample. The concentration of cells could be determined by counting them and multiplying by the dilution factor. The most significant improvement to the counting efficacy was proposed by Karl Burker (1872-1957), who made the device using heavy glass slide and three cemented glass platforms on the surface along with gridding. Over time, the hemocytometer rulings and counting grid evolved to account for more diverse requirements; initial design was made for counting smaller cells, like RBCs, and was not appropriate for other kinds of cells.
Hemocytometers currently available in the market are mass manufactured with laser etched grids. They are more accurate and easier to use compared to its predecessors. The device is crafted to ensure that the area covered by the lines as well as the volume of the chamber is known. The gridded area of the hemocytometer consists of 1×1 mm squares, which are again subdivided into smaller divisions, as shown in Figure 2A. The cell suspension is loaded via capillary action into the chamber and the cells are counted manually. The number of cells can further be used to determine the concentration or density of the cells (Absher, ).
Figure 2
Traditionally, a hemocytometer was also used to supplement cell enumeration with viability analysis, shown in Figure 2B. Cell viability, which is essential to assess the kinetics of growth, is performed by the selective staining of the sample using various staining approaches. Vital exclusion stains are commonly used to mark cells that have lost membrane integrity such as Trypan Blue or inclusion stains which mark cells retaining membrane integrity such as Fluorescein Diacetate (Kwizera et al.,
Although an integral part of every laboratory in cell based research, the method is still subject to several inherent shortcomings due to its design. Instrumentation and material variation, such as the type of buffer and pipette used, can add unwanted errors to the result. Since hemocytometer still depends on manual counting, human error is inevitable in the various stages of the process (mixing, handling, and dilution). If the analyst does not have a certain level of expertise in using the hemocytometer, it may lead to less accuracy and high inter-operator errors. The standard error in a hemocytometer is described as , where n is the number of cells counted (Berkson et al.,
Automatic Cell Counting
Since the early days of cell counting, there were attempts made to automate the process to yield high throughput, while maintaining accuracy. With the development of computers, improved optical and electronic components, cameras and algorithms, it is possible to eliminate some of the errors of manual counting through automation in the cell counting process. Fully automated systems can also perform sample preparation thereby reducing operation and turnaround time.
Electrical Impedance Based Cell Counting
Cell analysis using impedance for counting particles is one of the most popular conventional methods. It is predominantly based on the “Coulter principle” which was developed by Wallace H. Coulter, in 1953, who reported a novel method to analyze individual cells in a fluidic channel (Simson,
The Coulter principle is based on the phenomenon that the electrical impedance of a particle is different from that of the buffer. An aperture is placed between two electrodes while the electrolyte forms the current path. The aperture creates a “sensing zone,” which causes a change in impedance, concurrent with electric current, as the particle passes through it (Kubitschek,
The first commercial counter, the “Coulter Counter Model A” incorporated novel components, like the invention of mercury manometer based metering system to tackle the challenge of accurately measuring the sample volume flowing through the orifice (Hogg and Cooley,
Most early implementations of coulter counter use direct current resistance between the electrodes since that generates data, which is easy to process (Graham,
Coulter principle has been widely adopted across many industry and clinical applications since its conception. (Kubitschek,
Since its early days in 1960's, technology in commercial impedance counters has evolved immensely. These modern instruments count tens of thousands of cells in just a few seconds and boast very low coefficient of variance. The dynamic range of the instrument- that is, the size of the particles that can be analyzed- depends on the aperture size and is central to the application. For example, benchtop Coulter Counter Z series™ By Beckman Coulter offers several aperture sizes to choose from, giving a range of 1 μm to 120 μm particles that can be analyzed. Hence the device can count and analyze various types of mammalian cells including blood cells, yet has the capabilities to work with bacterial or plant cells. Similarly, their another benchtop counter called coulter counter Multisizer™ is made to handle anything including industrial particles because its aperture tubes can handle particles up to 1,600 μm (Figure 3A). On the other hand, manufacturer's like EMD Millipore and ORFLO offer miniaturized impedance counters for research laboratories. EMD Millipores' pipette shaped handheld Scepter™ 2.0 aspirates sample through its disposable sensors and presents histograms on the counters screen display. Figures 3B,C show the coulter counting principle with an orifice and channel incorporated inside the scepter counter tip. More recently, relatively new companies like Izon Science™ are coming with innovative solutions like putting a nanopore hole in a plastic membrane which can mechanically open or close, instead of standard fixed aperture, thereby giving an even larger dynamic range. QNano Gold™, a device from Izon Science, were able to analyze sub-micrometer particles like viruses and micro-vesicles (van der Pol et al.,
Figure 3

Coulter counter technology used in a benchtop and handheld device. (A) Schematic of a standard benchtop coulter counter. (B) Illustration of the mechanism of detection used in benchtop coulter counter. (C) Illustration showing Scepter handheld coulter counter and its working principle.
Microfluidic Impedance Analysis
More recently, advances in impedance cytometry have been achieved using cleanroom microfabrication protocols to miniaturize sensing and detection components through the introduction of microfluidic impedance analysis systems. When microfluidics was first introduced to cell counting/particle detection, its main goal was to bring portability, low cost, and simplicity (Sun and Morgan,
One of the earliest examples of a microfluidic coulter counter employed DC voltage to count living cells (Larsen et al.,
Figure 4

Microfluidic impedance cytometers for cell detection. (A) A micro-Coulter system employing two Ag/AgCl electrodes through a salt bridge (Chun et al.,
Mikael Evander et al. used impedance spectroscopy at multiple frequencies ranging from 280 kHz to 4 MHz for determination and analysis of platelets from blood samples (Evande et al.,
Electrode Design and Particle Focusing
Various electrode design and configurations have been investigated for use as microfluidic impedance sensors. Studying electrode geometry is important to achieve better detection sensitivity because of variability in the electric field distribution created by electrode design as cells go through their trajectories and create impedance signals. For higher sensitivity, the size of the electrodes has to be comparable to the size of cells. In order to facilitate this miniaturization, microfabrication techniques have been used extensively (McDonald et al.,
Gawad et al. introduced co-planar electrodes and devised a microfluidic system to measure the spectral impedance of individual cells or particles at multiple frequencies (Gawad et al.,
Figure 5

Different electrode geometry configuration. (A) Cell flowing through co-planar electrodes. (B) Illustration of a cell flowing through a parallel electrode configuration inside a microfluidic channel. (C) Schematic diagram of needle electrodes (Mansor et al.,
The other most common electrode configuration used in impedance flow cytometry systems is the parallel overlap arrangement. Cheung et al. first fabricated parallel facing microelectrodes in a fluidic channel (Cheung et al.,
In order to mitigate the influence of positional alignment as the sample passes through electrodes, several fluid flow focusing techniques have been proposed. This ensures that the cells flow in a single stream in the interrogation region. Holmes et al. used dielectrophoresis (DEP) to focus particles within the flowing fluid and demonstrated it by counting fluorescent latex particles at rate of 250 particles/s (Holmes et al.,
Table 2
| Focusing Method | Principle | Advantage | Disadvantage |
|---|---|---|---|
| Sheath flow | 2D and 3D Hydrodynamic focusing Ligler and Kim, | Fabrication of fluidic channels without the need for active electrical circuitry | Production of large amounts of waste sheath buffer fluid Need for multichannel fluidic pumps to precisely position cells within the fluid Increased cell concentrations can result in the loss of a focused single cell stream resulting in erroneous cell counts |
| Sheath-less flow | Acoustic focusing Piyasena et al., | High volumetric throughput Precise Spatial Positioning within 3D sheath flow | Requires integration of piezoelectric devices to generate acoustic waves |
| Dielectrophoretic focusing Yu et al., | Similar efficiency to acoustic focusing | Requires electrode integration within the channel Requires sample buffer conductivity to be adjusted Depends on particle polarizability | |
| Inertial focusing Gou et al., | Passive method not requiring external driving power | Diminished performance at high cell concentrations similar to hydrodynamic focusing Results in pressure variations and consequently the shear stresses | |
| Magnetic focusing Zeng et al., | Precise spatial positioning can be achieved by extrinsic magnetic bead labeling | Few biologicals particles are diamagnetic like erythrocytes and platelets Other cell types need to be tagged/ labeled using magnetic beads Requires the integration of strong magnets to produce intense field gradients |
Summary of fluid focusing methods in microfluidic cytometry.
Microfluidic impedance flow cytometry has evolved due to developments in particle focusing and microfabrication of electrodes integrated within the fluidic channels. The high sensitivity of impedance detection has allowed detection of particles of submicron sizes (Petchakup et al.,
Optical Cytometry
Developments within the fields of laser technology, signal processing, antibody production, and fluorochrome chemistry led to the advent of optical based detection methods. The first optical cytometer was developed approximately two decades after the advent of coulter counter (Mittag and Tarnok,
Technological developments have enabled the creation of FACS systems using flow cytometry. The first fluorescently labeled antibodies using Fluorescein Isothiocyanate (FITC) was developed by Albert Coons in 1941, which led to the creation of FACS by Herzenberg and commercialized by Becton Dickinson in 1974 (Moon et al.,
Flow cytometry on cells quantifies the optical and fluorescence characteristics of individual cells in flowing liquid as they pass through an optical and fluorescence light path. The fluidic setup is comprised of a central fluid stream in which the sample is contained, surrounded by a sheath fluid which provides hydrodynamic focusing to create a single stream of particles. This is essential so that only a single cell can pass through the light path at a time to prevent errors in cell counts. Flow cytometers are now available with more customizable parts that enable optimized operation based on the cells being studied. For example, various nozzle sizes are available to control the pressure and the shear stress on cells helping to improve and preserve viability in cell sorting applications (Graves et al.,
Figure 6

Principle of optical flow cytometry. (A) As incident light beam hits the cell, different parameters such as extinction, scatter and fluorescence are measured and this interaction provides information on optical properties and composition of the cell. (B) Schematic showing the particles lined in a single stream using sheath fluid as they interact with the laser light which gets collected by detectors.
Cells can be quantified not only based on forward scatter and side scatter, but also on the basis of a fluorescence light (FL) from the fluorochromes used for the detection of target cells (Brown and Wittwer,
Flow cytometry is used in various cell enumeration applications based on the detection of the membrane, cytoplasmic and nuclear antigens (Lan et al.,
Modern optical flow cytometers have evolved immensely to be more user-friendly through the use of intuitive digital interfaces and simpler operation protocols. BD Accuri™ c6 plus is a great example of a device used widely in research for studying cell viability, absolute cell counts and monitor cell proliferation. The device uses peristaltic pump which drives the fluidics such as sheath fluid (composed of distilled water) and sample. It uses 4 fluorescence detectors for the detection of common fluorochromes such as FITC among others, and 2 scatter detectors. EMD Millipore also offers a benchtop optical flow cytometer called Guava easyCyte™. Their use of a novel microcapillary method eliminates the need for sheath fluid, hence making it more compact than Accuri™ c6. The easyCyte™ can provide enumeration of cell population using small sample volume. Attune NxT™ is another device available in the market from ThermoFisher scientific that utilizes acoustophoresis instead of hydrodynamic focusing to position cells along the flow channel. Using acoustic method, the device can achieve high throughput to perform analysis. Such commercial devices offer great precision and sensitivity for cell population enumeration. But the initial investment and large size can be challenge for implementation in remote, low resource deployment. To tackle this bottleneck, researchers have attempted to integrate optical components onto lab-on-a-chip devices for inexpensive point-of-care settings.
Microfluidic Optical Flow Counting System
Fueled by the need to reduce size and cost of flow cytometers, significant effort has been made to miniaturize and make point-of-care optical flow cytometry systems based on microfluidic chips. At the time that impedance measurement systems exploited the potential of microfluidics for low-cost cytometry analysis (as described in previous sections), researchers also focused on developing microfluidic cytometer with optical detection systems. They are similar in principle to conventional flow cytometers, in that they rely on optical system and sensors for particle detection. To achieve microfluidic optical cytometry systems, several necessary technological innovations have been proposed toward microfabrication of fluidics system for introduction of samples inside the miniaturized channel and integration of optical excitation and detection system (Cho et al.,
The mechanisms used in conventional systems for fluid flow, focusing, optical illumination and detection need to be redefined to suit microfluidic platforms. A major component of microfluidic flow cytometry is the single stream focusing of cells in the channel to avoid coincidence of multiple cells passing through the detection zone. This is accomplished through various flow focusing methods such as hydrodynamic focusing using sheath flow, dielectrophoresis and inertial microfluidics (Gong et al.,
Several materials and optical configurations have been studied for development of sensitive yet robust microdevices. Sobek et al. achieved fabrication of flow chambers by bonding together two etched silicon wafer substrates using borophosphosilicate glass (Sobek et al.,
More recently, Schafer et al. enumerated cancer cells on their microfluidic device and compared it to commercial systems (Schafer et al.,
Figure 7

Microfluidic optical flow cytometers. (A) A microscopic image indicating hydrodynamic focusing and the arrangement of the optical fibers in a flow cytometry chip (Mao et al.,
Microfluidic optical flow cytometers have brought accurate and high throughput way to enumerate and characterize cells. Advancements in microfabrication methods have allowed integration of optics and detection system to the chip. They are shown to be capable of detecting and quantifying cells of different size better than impedance based methods by labeling them with fluorescent markers. However, to make the device more practical, they need to match the low CV's and multiparametric capabilities of benchtop flow cytometers. Most of the attempts at optical microfluidic flow cytometry involve using conventional hardware such as lasers, detectors, fluid pumps, high power light source and electronics while just miniaturizing the microfluidic chip. The main disadvantages of using bulk optics for microfluidic systems are the size, susceptibility to shock, and alignment problems which prevents it from being truly portable.
Image-Based Cytometry
With the advent of computers and rise in their processing capabilities in the last few decades, along with the emergence of inexpensive high-quality camera sensors, there has been a revival in image based analysis for cell counting. This is due to the advantages it provides- easier sample preparation and handling, economical, and visual information on the morphology of the cell (Lanigan et al.,
Image based cytometry can broadly be categorized to static image cytometry and kinetic image processing performed on high resolution images from flow cytometers. Firstly, we will be looking into static image cytometry within the context of image analysis. Unlike optical flow cytometry systems, image based cytometry devices provide the advantage of identifying each cell using real images from the heterogeneous population. Most image analysis systems, such as Life TechnologiesTM CountessR, use disposable slides with appropriately stained samples, using dyes such as Trypan blue, which the system images onto a digital camera. The device is categorized as semi-automatic because the user is responsible for performing the appropriate staining protocol to optimize the reading. Such a system may need an adjustment in staining to obtain a better resolution of cells from the background image, as captured by the auto-focus camera. High Trypan blue concentration in the sample can lead to background noise and interfere with the ability of the instrument to correctly distinguish cells. However, some image cytometers, such as the Vi cell XR by Beckman Coulter allow user to load several samples simultaneously (up to 9) and perform automated analysis. This eliminates the need for the analyst to perform staining on the samples; thereby making the sample preparation steps less time consuming.
For the successful analysis of a sample it is important to have fast and accurate software which can gather data from the captured images. Figure 8 shows the workflow of steps required to gather data from a device using image-based cytometry. As described earlier, sample preparation protocols form the primary step in any image cytometry system. Appropriate selective staining with the right concentration of dye to perform the analysis is crucial. This allows for visualization of cells by the camera and get better resolution of cells from the background. This is followed by image acquisition by the system using bright field or fluorescence microscopy. This is followed by segmentation, also known as object identification, which is the most important step in the process for determining accurate cell measurements. Segmentation algorithm has to detect adjoined cell aggregates in a biological sample. Histological images of cells are difficult to analyze due to image artifacts that are introduced due to multistep preparation process as well as variation in batches due to different microscopy and staining steps. The technique must be able to extract information from images by combining local and global properties that are characterized by different pixel classes in the continuous boundary. Several methods were developed to accurately address cluster segmentation. Previous work by Sonal Kothari et al. described an edge based segmentation method that can segment complex clusters with reasonable accuracy (Kothari et al.,
Figure 8

Image cytometry workflow showing the different stages from sample preparation to data analysis.
In contrast, research in kinetic image flow cytometry offers an alternative to conventional flow cytometry through high-resolution imaging or microscopy of single cells in flowing environments. Essentially, the goal of such systems is to perform live cell imaging in a sample suspension using powerful image segmentation and feature extraction, as described previously, while combining features of flow cytometry. Hence, it allows evaluation of morphological and fluorescent data on single cell in a heterogeneous population. The limitation of static image cytometry due to challenges in image segmentation as described earlier has little effect on kinetic imaging flow cytometry as the cells are in suspension and interrogated on an individual basis. This progress was mostly due to emergence of high-speed cameras to image large number of cells per acquisition frame. For instance, McKenna et al. developed a device that utilized a parallel microchannel network along with confocal microscopy for rare cell detection (McKenna et al.,
Currently, several commercial image-based cytometers are available for research and clinical use. Multi-parametric data is obtained by applying powerful image segmentation and feature extraction algorithms to raw data. Modern image cytometers combine the power of fluorescent detection with imaging capabilities as they are highly complementary, allowing multiple parameters to be measured for individual cells. Vast array of fluorescent probes and stains makes it possible to measure spatial distribution, intensities of markers on cells, as well as general morphological quantities to generate data on each cell. Simple cell counters like TC-20TM by Bio-rad Laboratories claim to count mammalian cells in <30 s. It uses light microscopy with auto-focus that analyses multiple focus planes, along with a cell counting algorithm that is able to identify cells and exclude debris thereby calculating the total cell count. These kind of devices eliminate the need to dilute the solution (process a concentration range of 5 × 104-1 × 107 cells/mL), as usually done in hemocytometer, and reduce the error associated with sample dilutions. NC-3000™ by NucleoCounter utilizes fluorescent imaging to characterize cell properties. This is achieved by exploiting fluorescent dyes with affinity for DNA to generate images which are otherwise missed by light microscopy. Other devices such as FlowSight® Imaging Flow Cytometer combine flow cytometry capabilities with imaging and simultaneously produce dark field (side scatter), bright field and fluorescent images of each cell.
Image cytometry can measure information at single cell resolution. The method provides rich morphological data and overcomes the lack of this information from electrical impedance and optical flow cytometers. Specialty applications where specific information is required without destroying the cell, such as assays studying cell-cell interaction or morphological data make image cytometer ideal for use. The primary bottleneck of using this technology in clinical applications is the computational challenges in handling large volumes of image data that need to be processed in real time (Nitta et al.,
Conclusion
Tremendous progress has been made in each paradigm of cell enumeration. Several industry, clinical and research applications leverage the use of impedance, optical, or image analysis systems for cell counting and characterization. Each method provides advantages that need to be considered while keeping in mind the cost, throughput, accuracy, and portability of the device.
Cell analysis using a hemocytometer has been widely used for over a century, primarily due to its low cost and portability. The device is used by clinics to analyze blood samples and as a common tool in research laboratories around the world. However, due to the cumbersome process of manual counting, there was a strong push to develop automatic cell counting technologies which yielded high throughput while maintaining accuracy. Over the last two decades, automatic cell counters have undergone significant improvements due to introduction of new technologies. Coulter technology using impedance for counting particles is one of the most popular automated methods. With the help of modern micro-fabrication techniques, it is now possible to fabricate sub-micrometer apertures for counting and detecting particles as small as DNA molecules. Two decades after the advent of Coulter technology, flow cytometry developed as a result of advancements in the fields of optics, antibody production, and fluorochrome discovery. The benefits of flow cytometers over coulter counters are in terms of their ability to distinguish cells based on multiple parameters such as a wide array of cellular markers. This technology was quickly adopted commercially. Another commonly used technology is referred to as image-based cytometry. With an increase in computing capabilities and high-quality camera sensors, there has been an increase in image based analysis for cell counting. This technology is implemented through high-resolution imaging or microscopy of single cells in static and flowing environments. Using this approach, image based morphological data from a heterogeneous cell suspension is obtained using powerful segmentation and feature extraction algorithms.
Microfluidic cytometers have brought accurate and high throughput way to enumerate and characterize cells on devices with a small footprint. Advancements in microfabrication methods have allowed integration of electrodes, optics, and various detection systems on a chip. Miniaturization will eventually diminish the need for benchtop cytometry by replacing these relatively large devices with portable or even wearable platforms that provide qualitative information about the various cell types within a heterogeneous sample such as blood. In addition, such platforms would also enumerate rare cells, such as CTCs and other cell irregularities within biological fluids. Providing diagnostic information from portable home based blood counting devices without the need for a centralized diagnostic laboratory will provide user value and generate commercial interest for medical device manufacturers. These miniaturized cell counters could be accessible at low cost like modern blood sugar monitors (i.e., Blood Glucose Meter) and act as an indicator for disease status at point-of-care settings.
Statements
Author contributions
AV, AM, and MQ: designed the research. AV and AM: wrote the manuscript. MQ: supervised the work and finalized the manuscript.
Funding
We acknowledge funding from the Sheikh Hamdan Award for Medical Science Research Grant (Dubai, UAE), the New York University Abu Dhabi (NYUAD), and the NYUAD 2017 Research Enhancement Fund.
Acknowledgments
We also acknowledge Christina Johnson for designing Figure 1.
Conflict of interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
References
1
AbsherM. (1973). Tissue Culture: Methods and Applications. Massachusetts, MA: Elsevier Inc.
2
AdanA.AlizadaG.KirazY.BaranY.NalbantA. (2016). Flow cytometry: basic principles and applications. Crit Rev Biotechnol.37, 163–176. 10.3109/07388551.2015.1128876
3
AyliffeH. E.FrazierA. B.RabbittR. D. (1999). Electric impedance spectroscopy using microchannels with integrated metal electrodes. Microelectromech. Syst.8, 50–57. 10.1109/84.749402
4
BagnaninchiP. O.DrummondN. (2011). Real-time label-free monitoring of adipose-derived stem cell differentiation with electric cell-substrate impedance sensing. Proc. Nat. Acad. Sci. U.S.A.108, 6462–6467. 10.1073/pnas.1018260108
5
BardJ. (1974). Tissue culture, methods and applications. Cell2, 120–121. 10.1016/0092-8674(74)90102-0
6
BerksonJ.MagathT. B.HurnM. (1939). Error of estimate of the blood cell count as made with the hemocytometer. Am. J. Physiol.128, 300–323. 10.1152/ajplegacy.1939.128.2.309
7
BessmanJ. D.WilliamsL. J.GilmerP. R. (1982). Platelet size in health and hematologic diseas. Am. J. Clin. Pathol.78, 150–153. 10.1093/ajcp/78.2.150
8
BiggsR.MacmillanR. L. (1948). The error of the red cell count. J. Clin. Pathol.1, 288–291. 10.1136/jcp.1.5.288
9
BlasiT.HennigH.SummersH. D.TheisF. J.CerveiraJ.PattersonJ. O.et al. (2016). Label-free cell cycle analysis for high-throughput imaging flow cytometry. Nat. Commun.7:10256. 10.1038/ncomms10256
10
BraylanR. C. (2004). Impact of flow cytometry on the diagnosis and characterization of lymphomas, chronic lymphoproliferative disorders and plasma cell neoplasias. Cytometry A58A, 57–61. 10.1002/cyto.a.10101
11
BrownM.WittwerC. (2000). Flow cytometry: principles and clinical applications in hematology, in: Clinical Chemistry, eds AshwoodE. R. (Miami, FL: Clinical Chemistry), 1221–1229.
12
BullB. S.SchneidermanM. A.BrecherG. (1965). Platelet counts with coulter counter. Am. J. Clin. Pathol.44, 678–688. 10.1093/ajcp/44.6.678
13
BurandaT.WuY.SklarL. A. (2011). Quantum dots for quantitative flow cytometry, in Methods in Molecular Biology (Methods and Protocols). (Humana Press), ed WalkerJ. M. (New York, NY: Humana Press), 67–84.
14
ByerlyL.CassadaR. C.RussellR. L. (1975). Machine for rapidly counting and measuring the size of small nematodes. Rev. Sci. Inst.46:517. 10.1063/1.1134262
15
Canjun MuF. Z.ZhiyiZ.MinL.XudongC. (2011). Highly efficient dual-channel cytometric-detection of micron-sized particles in microfluidic device. Sensors Actuators B: Chem.151, 402–409. 10.1016/j.snb.2010.09.055
16
ChabinycM. L.ChiuD. T.McDonaldJ. C.StroockA. D.ChristianJ. F.KargerA. M.et al. (2001). An integrated fluorescence detection system in poly(dimethylsiloxane) for microfluidic applications. Anal. Chem.73, 4491–4498. 10.1021/ac010423z
17
ChattopadhyayP. K.RoedererM. (2012). Cytometry: Today's technology and tomorrow's horizons. Methods57, 251–258. 10.1016/j.ymeth.2012.02.009
18
ChengX.GuptaA.ChenC.TompkinsR. G.RodriguezW.TonerM. (2009). Enhancing the performance of a point-of-care CD4+ T-cell counting microchip through monocyte depletion for HIV/AIDS diagnostics. Lab Chip9, 1357–1364. 10.1039/B818813K
19
ChengX.IrimiaD.DixonM.SekineK.DemirciU.ZamirL.et al. (2007). A microfluidic device for practical label-free CD4+ T cell counting of HIV-infected subjects. Lab Chip7, 170–178. 10.1039/b612966h
20
CheungK.GawadS.RenaudP. (2005). Impedance spectroscopy flow cytometry: on-chip label-free cell differentiation. Cytometry Part A65, 124–132. 10.1002/cyto.a.20141
21
ChinC. D.LinderV.SiaS. K. (2012). Commercialization of microfluidic point-of-care diagnostic devices. Lab Chip12, 2118–2134. 10.1039/c2lc21204h
22
ChoS. H.GodinJ. M.ChenC. H.QiaoW.LeeH.LoY.-H. (2010). Review article: recent advancements in optofluidic flow cytometer. Biomicrofluidics4:43001. 10.1063/1.3511706
23
ChunH.ChungT. D.KimH. C. (2005). Cytometry and velocimetry on a microfluidic chip using polyelectrolytic salt bridges. Anal. Chem.77, 2490–2495. 10.1021/ac048535o
24
Comas-RiuJ.RiusN. (2009). Flow cytometry applications in the food industry. J. Ind. Microbiol. Biotechnol.36, 999–1011. 10.1007/s10295-009-0608-x
25
CoulterW. H. (1956). High speed automatic blood cell counter and cell size analyzer, in National Electronics Conference (Chicago, IL), 1034–1042.
26
EdmundsonI. C. (1966). Coincidence error in coulter counter particle size analysis. Nature212, 1450–1452. 10.1038/2121450b0
27
EngC.ValensteiniP. (1989). Standardization of Yeast Inocula with an electronic impedance counter. J. Clin. Microbiol.27, 2397–2399.
28
ErricoaV.DeNinnoaA.BertaniF. R.BusinaroL.BisegnaP.CaselliaF. (2017). Mitigating positional dependence in coplanar electrode Coulter-type microfluidic devices. Sensors Actuators B: Chem.247, 580–586. 10.1016/j.snb.2017.03.035
29
EtcheverryS.FaridiA.RamachandraiahH.KumarT.MargulisW.LaurellF.et al. (2017). High performance micro-flow cytometer based on optical fibres. Sci. Rep. 7:5628. 10.1038/s41598-017-05843-7
30
EvandeM.RiccoA. J.MorserJ.KovacsG. T.LeungL. L.GiovangrandiL. (2013). Microfluidic impedance cytometer for platelet analysis. Lab Chip13, 722–729. 10.1039/c2lc40896a
31
FanY. J.WuY. C.ChenY.KungY. C.WuT. H.HuangK. W.et al. (2013). Three dimensional microfluidics with embedded microball lenses for parallel and high throughput multicolor fluorescence detection. Biomicrofluidics7:44121. 10.1063/1.4818944
32
FeiP.ChenZ.MenY.LiA.ShenY.HuangY. (2012). A compact optofluidic cytometer with integrated liquid-core/PDMS-cladding waveguides. Lab Chip12, 3700–3706. 10.1039/c2lc40329c
33
FernyhoughM. E.HelterlineD. L.VierckJ. L.HillR. A.DodsonM. V. (2004). Coulter counter use in the enumeration of muscle and fat stem cells. Methods Cell Sci.25, 221–225. 10.1007/s11022-004-2382-5
34
FullerC. K.HamiltonJ.AcklerH.KrulevitchP.BoserB.EldredgeA.et al. (2000). Microfabricated multifrequenc particle impedance characterization system, in Micro Total Analysis Systems, eds van den BergA.OlthuisW.BergveldP. (Enschede: Kluwer Publisher), 265–268.
35
GarnerD. L.GledhillB. L.PinkelD.LakeS.StephensonD.DillaM. A.et al. (1983). Quantification of the X- and Y-chromosome-bearing spermatozoa of domestic animals by flow cytometry. Biol. Reproduct.28, 312–321. 10.1095/biolreprod28.2.312
36
GawadS.SchildbL.RenaudP. H. (2001). Micromachined impedance spectroscopy flow cytometer for cell analysis and particle sizing. Lab Chip1, 76–82. 10.1039/b103933b
37
GodinJ.LoY.-H. (2010). Two-parameter angular light scatter collection for microfluidic flow cytometry by unique waveguide structures. Biomed. Opt. Exp.1, 1472–1479. 10.1364/BOE.1.001472
38
GohW. J.ZouS.LeeC. K.OuY.-H.WangJ.-W.CzarnyB.et al. (2018). EXOPLEXs: chimeric drug delivery platform from the fusion of cell-derived nanovesicles and liposomes. Biomacromolecules19, 22–30. 10.1021/acs.biomac.7b01176
39
GoldbergE. M. (2008). Method, System, and Compositions for Cell Counting and Analysis. US Patent Application US20080212069A1. Washington, DC: U.S. Patent and Trademark Office.
40
GoldenJ. P.JustinG. A.NasirM.LiglerF. S. (2012). Hydrodynamic focusing – a versatile tool. Anal. Bioanal. Chem.402, 325–335. 10.1007/s00216-011-5415-3
41
GongY.FanN.YangX.PengB.JiangH. (2018). New advances in microfluidic flowcytometry. Electrophoresis0, 1–18. 10.1002/elps.201800298
42
GoodinT. (2017). FDA allows marketing of test to aid in the detection of certain leukemias and lymphomas, eds GoodinT. (US Food and Drug Administration website: US Department of Health and Services). Available online at: https://www.fda.gov/news-events/press-announcements/statement-fda-commissioner-scott-gottlieb-md-implementation-agencys-streamlined-development-and
43
GouY.JiaY.WangP.SunC. (2018). Progress of inertial microfluidics in principle and application. Sensors18, 1–26. 10.3390/s18061762
44
GrahamM. D. (2003). The coulter principle: foundation of an industry. J. Lab. Autom.8, 72–81. 10.1016/s1535-5535(03)00023-6
45
GrahamM. D. (2013). The Coulter principle: imaginary origins. Cytometry A83, 1057–1061. 10.1002/cyto.a.22398
46
GravesS. W.NolanJ. P.JettJ. H.MartinJ. C.SklarL. A. (2002). Nozzle design parameters and their effects on rapid sample delivery in flow cytometry. Cytometry part A47, 127–137. 10.1002/cyto.10056
47
GuoJ.LiH.ChenY.KangY. (2014). A microfluidic impedance cytometer on printed circuit board for low cost diagnosis. IEEE Sensors J.14, 2112–2117. 10.1109/JSEN.2013.2295399
48
GuoJ.PuiT. S.RahmanA. R. A.KangY. (2012). Numerical investigation of the performance of coulter counter with novel structure. Int. J. Info. Electron. Eng.2, 885–888. 10.7763/IJIEE.2012.V2.233
49
GuoY.BaumgartS.StärkH. J.HarmsH.MüllerS. (2017). Mass cytometry for detection of silver at the bacterial single cell level. Front. Microbiol.8:1326. 10.3389/fmicb.2017.01326
50
GuptaA.HarrisonP. J.WieslanderH.PielawskiN.KartasaloK.PartelG.et al. (2018). Deep learning in image cytometry: a review. Cytometry Part A89, 1057–1072. 10.1002/cyto.a.23701
51
HassanU.ReddyB.DamhorstG.SonoikiO.GhongeT.YangC.et al. (2015). A microfluidic biochip for complete blood cell counts at the point-of-care. Technology3, 201–213. 10.1142/S2339547815500090
52
HennigH.ReesP.BlasiT.KamentskyL.HungJ.DoaD.et al. (2017). An open-source solution for advanced imaging flow cytometry data analysis using machine learning. Methods112, 201–210. 10.1016/j.ymeth.2016.08.018
53
HeoY. J.LeeD.KangJ.LeeK.ChungW. K. (2017). Real-time image processing for microscopy-based label-free imaging flow cytometry in a microfluidic chip. Sci. Rep.7:11651. 10.1038/s41598-017-11534-0
54
HoffmanR. A.JohnsonT. S.BrittW. B. (1981). Flow cytometric electronic direct current volume and radiofrequency impedance measurements of single cells and particles. Cytometry. 1, 377–384. 10.1002/cyto.990010605
55
HoggW. R.CooleyK. D. (1964). Variable Pressure Pump for Manometric Metering Device. U. S. Patent Application US3289916A. Washington, DC: U.S. Patent and Trademark Office.
56
HolmesD.MorganH. (2010). Single cell impedance cytometry for identification and counting of CD4 T-cells in human blood using impedance labels. Anal. Chem.82, 1455–1461. 10.1021/ac902568p
57
HolmesD.MorganH.GreenN. G. (2006). High throughput particle analysis: combining dielectrophoretic particle focussing with confocal optical detection. Biosens. Bioelectron.21, 1621–1630. 10.1016/j.bios.2005.10.017
58
HolmesD.PettigrewD.RecciusC. H.GwyerJ. D.van BerkelC.HollowayJ.et al. (2009). Leukocyte analysis and differentiation using high speed microfluidic single cell impedance cytometry. Lab Chip9, 2861–3024. 10.1039/b910053a
59
HongC.LeeJ.ZhengH.HongS. S.LeeC. (2011). Porous silicon nanoparticles for cancer photothermotherapy. Nanoscale Res. Lett.6:321. 10.1186/1556-276X-6-321
60
HouwenB. (2001). The differential cell count. Lab. Hematol.7, 89–100.
61
HsiungF.MccollumT.HefnerE.RubioT. (2013). Comparision of count reproducibility, accuracy and time to results between a hemocytometer and the TC20 automated cell counter, in Technical Document BioRad, ed BioRad (California, CA: BioRad Laboratories Inc.), 1–4.
62
HurleyJ. (1970). Sizing particles with a coulter counter. Biophys. J.10, 74–79. 10.1016/S0006-3495(70)86286-5
63
JelinekT.BezdekovaR.ZatopkovaM.BurgosL.SimicekM.SevcikovaT.et al. (2017). Current applications of multiparameter flow cytometry in plasma cell disorders. Blood Cancer J.7:e617. 10.1038/bcj.2017.90
64
JungT.SchaueraU.HeusserC.NeumannC.RiegeraC. (1993). Detection of intracellular cytokines by flow cytometry. J. Immunol. Methods159, 197–207. 10.1016/0022-1759(93)90158-4
65
KannelW. B.AndersonK.WilsonP. W. (1992). White blood cell count and cardiovascular disease insights from the framingham study. J. Am. Med. Assoc.267, 1253–1256.
66
KothariS.ChaudryQ.WangM. D. (2009). Extraction of informative cell features by segmentation of densely clustered tissue images, in Annual International Conference of the IEEE Engineering and Medicine Society IEEE (Minneapolis, MN), 6706–6709.
67
KredietC. J.DeNofrioJ. C.CarusoC.BurriesciM. S.CellaK.PringleJ. R. (2015). Rapid, precise, and accurate counts of symbiodinium cells using the guava flow cytometer, and a comparison to other methods. PLoS ONE10:e0135725. 10.1371/journal.pone.0135725
68
KubitschekH. E. (1958). Electronic counting and sizing of bacteria. Nature182, 234–235.
69
KubitschekH. E. (1969). Methods in Microbiology, eds NorrisJ. R.RibbonsD. W. (Amsterdam: Elsevier).
70
KulpS. K.YangY. T.HungC. C.ChenK. F.LaiJ. P.TsengP. H.et al. (2004). 3-phosphoinositide-dependent protein kinase-1/Akt signaling represents a major cyclooxygenase-2-independent target for celecoxib in prostate cancer cells. Cancer Res.64, 1444–1451. 10.1158/0008-5472.CAN-03-2396
71
KwizeraR.AkampuriraA.KandoleT. K.NielsenK.KambuguA.MeyaD. B.et al. (2017). Evaluation of trypan blue stain in a haemocytometer for rapid detection of cerebrospinal fluid sterility in HIV patients with cryptococcal meningitis. BMC Microbiol.17:182. 10.1186/s12866-017-1093-4
72
LanH. Y.HutchinsonP.TeschG. H.MuW.AtkinsR. C. (1996). A novel method of microwave treatment for detection of cytoplasmic and nuclear antigens by flow cytometry. J. Immunol. Methods190, 1–10. 10.1016/0022-1759(95)00233-2
73
LaniganD.McLeanP. A.CurranB.LeaderM. (1993). Comparison of flow and static image cytometry in the determination of ploidy. J. Clin. Pathol.46, 135–139. 10.1136/jcp.46.2.135
74
LarsenU. D.BlankensteinG.BranebJ. (1997). Microchip coulter particle counter, in Proceedings of International Solid State Sensors and Actuators Conference (Transducers '97) (Chicago, IL), 1319–1322.
75
LastW. M.SmolJ. P. (2006). Tracking Environmental Change Using Lake Sediments.Berlin: Springer.
76
LeeG.-B.LinC.-H.ChangG.-L. (2003). Micro flow cytometers with buried SU-8/SOG optical waveguides. Sensors Actuators A: Phys.103, 165–170. 10.1016/S0924-4247(02)00305-9
77
LiglerF. S.KimJ. S. (2010). The Microflow Cytometer.New York, NY: Pan Stanford.
78
LitjensG.KooiT.BejnordiB. E.SetioA. A. A.CiompiF.GhafoorianM.et al. (2017). A survey on deep learning in medical image analysis. Med. Image Anal.42, 60–88. 10.1016/j.media.2017.07.005
79
LiuZ.GuoW.ZhangD.PangY.ShiJ.WanS.et al. (2016). Circulating tumor cell detection in hepatocellular carcinoma based on karyoplasmic ratios using imaging flow cytometry. Sci. Rep.6:39808. 10.1038/srep39808
80
LohmanA. C.Van RijnI.LindhardtC. L.VontheinR.RadesD.HollanderN. H. (2018). Preliminary results from a prospective study comparing white blood cell and neutrophil counts from a laboratory to those measured with a new device in patients with breast cancer. In Vivo32, 1283–1288. 10.21873/invivo.11378
81
LutzP.DzikW. H. (1993). Large volume hemocytometer chamber for accurate counting of white cells in WBC reduced platelets. Transfusion33, 409–412. 10.1046/j.1537-2995.1993.33593255602.x
82
MaertensY.HumbergV.ErlmeierF.SteffensS.SteinestelJ.BögemannM.et al. (2017). Comparison of isolation platforms for detection of circulating renal cell carcinoma cells. Oncotarget8, 87710–87717. 10.18632/oncotarget.21197
83
MansorM. A.TakeuchiM.NakajimaM.HasegawaY.AhmadM. R. (2017). Electrical impedance spectroscopy for detection of cells in suspensions using microfluidic device with integrated microneedles. Appl. Sci.7:170. 10.3390/app7020170
84
MaoX.NawazA. A.LinS. C.LapsleyM. I.ZhaoY.McCoyJ. P.et al. (2012). An integrated, multiparametric flow cytometry chip using “microfluidic drifting” based three-dimensional hydrodynamic focusing. Biomicrofluidics6, 1–9. 10.1063/1.3701566
85
MartiniJ.RechtM. I.HuckM.BernM. W.JohnsonN. M.KieselP. (2012). Time encoded multicolor fluorescence detection in a microfluidic flow cytometer. Lab Chip12, 5057–5062. 10.1039/c2lc40515f
86
McDonaldJ. C.DuffyD. C.AndersonJ. R.ChiuD. T.WuH.SchuellerO. J.et al. (2000). Fabrication of microfluidic systems in poly(dimethylsiloxane). Electrophoresis21, 27–40. 10.1002/(SICI)1522-2683(20000101)21:1<27::AID-ELPS27>3.0.CO;2-C
87
McKennaB. K.EvansJ. G.CheungM. C.EhrlichD. J. (2011). A parallel microfluidic flow cytometer for high-content screening. Nat. Methods8, 401–403. 10.1038/nmeth.1595
88
MernierG.DuqiE.RenaudP. (2012). Characterization of a novel impedance cytometer design and its integration with lateral focusing by dielectrophoresis. Lab Chip12, 4344–4349. 10.1039/c2lc40551b
89
MittagA.TarnokA. (2011). Advanced Optical Flow Cytometry: Methods and Disease Diagnoses, ed TuchinV. V. (New Jersey, NJ:Wiley).
90
MohanA.MarshkoleN.NairA. P.BharadwajA.PrabhakarA.SaiyedT. (2017). Development of a lab on a chip flow cytometer, in BIODEVICES, eds PeixotoN.FredA.GamboaH.VazM. (Porto: SciTePress, 179–185.
91
MoonI. S.LeeH.ParkS. D.SeogD.-H. (2010). Immunonucleochemistry: a new method for in situ detection of antigens in the nucleus of cells in culture. Cytotechnology62, 89–93. 10.1007/s10616-010-9266-0
92
NittaN.SugimuraT.IsozakiA.MikamiH.HirakiK.SakumaS.et al. (2018). Intelligent image-activated cell sorting. Cell175, 266–276. 10.1016/j.cell.2018.08.028
93
NowakowskiG. S.WitzigT. E.DingliD.TraczM. J.GertzM. A.LacyM. Q.et al. (2005). Circulating plasma cells detected by flow cytometry as a predictor of survival in 302 patients with newly diagnosed multiple myeloma. Blood106, 2276–2279. 10.1182/blood-2005-05-1858
94
OakeyJ.ApplegateR. W.ArellanoE.Di CarloD.GravesS. W.TonerM. (2010). Particle focusing in staged inertial microfluidic devices for flow cytometry. Anal. Chem.82, 3862–3867. 10.1021/ac100387b
95
O'BrienJ.HayderH.PengC. (2016). Automated quantification and analysis of cell counting procedures using imagej plugins. J. Visual. Exp.117:5471910.3791/54719
96
OngenaK.DasC.SmithJ. L.GilS.JohnstonG. (2010). Determining cell number during cell culture using the scepter cell counter. J. Visual. Exp.45:2204. 10.3791/2204
97
OYAMAV. I.EAGLEH. (1956). Measurement of cell growth in tissue culture with a phenol reagent. Exp. Biol. Med.91, 305–307. 10.3181/00379727-91-22245
98
PammeN.KoyamaR.ManzA. (2003). Counting and sizing of particles and particle agglomerates in a microfluidic device using laser light scattering: application to a particle-enhanced immunoassay. Lab Chip3, 187–192. 10.1039/b300876b
99
ParisA.NhanT.CornetE.PerolJ. P.MaletM.TroussardX. (2010). Performance evaluation of the body fluid mode on the platform Sysmex XE-5000 series automated hematology analyzer. Int. J. Lab. Hematol.32, 539–547. 10.1111/j.1751-553X.2010.01220.x
100
ParkerM. S.BarnesM. (1967). Use of the coulter counter to measure the swelling of bacterial spores during germination and outgrowth. J. Appl. Microbiol.30, 299–303. 10.1111/j.1365-2672.1967.tb00300.x
101
PedreiraC. E.CostaE. S.LecrevisseQ.van DongenJ. J.OrfaoA. (2013). Overview of clinical flow cytometry data analysis: recent advances and future challenges. Trends Biotechnol.31, 415–425. 10.1016/j.tibtech.2013.04.008
102
PetchakupC.LiK. H. H.HouH. W. (2017). Advances in single cell impedance cytometry for biomedical applications. Micromachines8, 87–107. 10.3390/mi8030087
103
PiyasenaM. E.Austin SuthanthirarajP. P.ApplegateR. W.GoumasA. M.WoodsT. A.LópezG. P.et al. (2012). Multinode acoustic focusing for parallel flow cytometry. Anal. Chem.84, 1831–1839. 10.1021/ac200963n
104
PrincenL. H.KwolekW. F. (1965). Coincidence corrections for particle size determinations with the coulter counter. Rev. Sci. Instr.36:646. 10.1063/1.1719656.
105
RamasamyS.BennetD.KimS. (2014). Drug and bioactive molecule screening based on a bioelectrical impedance cell culture platform. Int. Nanomed.9, 5789–5809. 10.2147/IJN.S71128
106
RaneA. S.RutkauskaiteJ.deMelloA.StavrakisS. (2017). High-throughput multi-parametric imaging flow cytometry. Chem3, 588–602. 10.1016/j.chempr.2017.08.005
107
Rodriguez-TrujilloR.Castillo-FernandezO.ArundellM.ValenciaA.GomilaG. (2008). High-speed particle detection in a micro-Coulter counter with two-dimensional adjustable aperture. Biosens. Bioelectron.24, 290–296. 10.1016/j.bios.2008.04.005
108
SalehO. A.SohnL. L. (2001). Quantitative sensing of nanoscale colloids using a microchip Coulter counter. Rev. Sci. Inst.72:4449. 10.1063/1.1419224
109
SandhausL. M.CiarliniP.KidricD.DillmanC.O'RiordanM. (2010). Automated cerebrospinal fluid cell counts using the sysmex XE-5000: is it time for new reference ranges?Am. J. Clin. Pathol.134, 734–738. 10.1309/AJCPABGQXSIA4SMT
110
SchaferD.GibsonE. A.SalimE. A.PalmerA. E.JimenezR.SquierJ. (2009). Microfluidic cell counter with embedded optical fibers fabricated by femtosecond laser ablation and anodic bonding. Optical Express17:6068–6073. 10.1364/OE.17.006068
111
ShriraoA. B.FritzZ.NovikE. M.YarmushG. M.SchlossR. S.ZahnJ. D.et al. (2018). Microfluidic flow cytometry: the role of microfabrication methodologies, performance and functional specification. Technology6, 1–23. 10.1142/S2339547818300019
112
SimieleM.D'AvolioA.BaiettoL.SiccardiM.SciandraM.AgatiS.et al. (2011). Evaluation of the mean corpuscular volume of peripheral blood mononuclear cells of HIV patients by a coulter counter to determine intracellular drug concentrations. Am Soc Microbiol.55, 2976–2978. 10.1128/AAC.01236-10
113
SimsonE. (2013). Wallace Coulter's life and his impact on the world. Int. J. Lab. Hematol.35, 230–236. 10.1111/ijlh.12069
114
SobekD.YoungA. M.GrayM. L.SenturiaS. D. (1993). A microfabricated flow chamber for optical measurements in fluids, in Proceedings IEEE Micro Electro Mechanical Systems (Florida, FL), 219–224.
115
SohnL. L.SalehO. A.FacerG. R.BeavisA. J.AllanR. S.NottermanD. A. (2000). Capacitance cytometry: Measuring biological cells one by one. Proc Nat Acad Sci U.S.A. 97:10687–10690. 10.1073/pnas.200361297
116
SongC.LuongT. D.KongT. F.NguyenN. T.AsundiA. K. (2011). Disposable flow cytometer with high efficiency in particle counting and sizing using an optofluidic lens. Opt.Lett.36, 657–659. 10.1364/OL.36.000657
117
SpencerD.MorganH. (2011). Positional dependence of particles in microfludic impedance cytometry. Lab Chip11, 1234–1239. 10.1039/c1lc20016j
118
SpinneyP. S.CollinsS. D.HowittD. G.SmithR. L. (2012). Fabrication and characterization of a solid-state nanopore with self-aligned carbon nanoelectrodes for molecular detection. Nanotechnology23:13550110.1088/0957-4484/23/13/135501
119
SunT.MorganH. (2010). Single-cell microfluidic impedance cytometry: a review. Microfluid. Nanofluid.8, 423–443. 10.1007/s10404-010-0580-9
120
TangW.TangD.NiZ.XiangN.YiH. (2017). Microfluidic impedance cytometer with inertial focusing and liquid electrodes for high-throughput cell counting and discrimination. Anal. Chem.89, 3154–3161. 10.1021/acs.analchem.6b04959
121
TapeinosC.MarinoA.BattagliniM.MigliorinS.BresciaR.ScarpelliniA.et al. (2019). Stimuli-responsive lipid-based magnetic nanovectors increase apoptosis in glioblastoma cells through synergic intracellular hyperthermia and chemotherapy. Nanoscale11, 72–88. 10.1039/C8NR05520C
122
TelfordW. G. (2004). Analysis of UV-excited fluorochromes by flow cytometry using near-ultraviolet laser diodes. Cytometry Part A61A, 9–17. 10.1002/cyto.a.20032
123
TestaG.PersichettiG.BerniniR. (2015). Micro flow cytometer with self-aligned 3D hydrodynamic focusing. Biomed. Opt. Express6, 54–62. 10.1364/BOE.6.000054
124
van BerkelC.GwyerJ. D.DeaneS.GreenN. G.HollowayJ.et al. (2011). Integrated systems for rapid point of care (PoC) blood cell analysis. Lab Chip11, 1249–1255. 10.1039/c0lc00587h
125
van der PolE.CoumansF.VargaZ.KrumreyM.NieuwlandR. (2013). Innovation in detection of microparticles and exosomes. J. Thromb. Haemost.11, 36–45. 10.1111/jth.12254
126
VerbruggeS. E.HuismanA. (2015). Verification and standardization of blood cell counters for routine clinical laboratory tests. Clin. Lab. Med.35, 183–196. 10.1016/j.cll.2014.10.008
127
VerellenN.VercruysseD.RochusV.BoisB. D.DusaA.KermanS.et al. (2018). Integrated metasurface photonics for miniature flow cytometry, in SPIE- Nanoscale Imaging, Sensing, and Actuation for Biomedical Applications XV, eds CartwrightA. N.NicolauD. V.FixlerD. (San Francisco, CA: SPIE).
128
VersoM. L. (1964). The evolution of blood-counting technologies. Med. History8, 149–158. 10.1017/S0025727300029392
129
WallaceH.CounterW. R. H. (1966). Signal Modulated Apparatus for Generating and Detecting Resistive and Reactive Changes in a Modulated Current Path for Particle Classification and Analysis. U.S. Patent Application US3502974A. Washington, DC: U.S. Patent and Trademark Office.
130
WangZ.El-AliJ.EngelundM.GotsaedT.Perch-NielsenI. R.MogensenK. B.et al. (2004). Measurements of scattered light on a microchip flow cytometer with integrated polymer based optical elements. Lab Chip4, 372–377. 10.1039/b400663a
131
WatkinsN. N.HassanU.DamhorstG.NiH.VaidA.RodriguezW.et al. (2013). Microfluidic CD4+ and CD8+ T lymphocyte counters for point-of-care HIV diagnostics using whole blood. Sci. Transl. Med.5:214. 10.1126/scitranslmed.3006870
132
WhitesidesG. M. (2006). The origins and the future of microfluidics. Nature442, 368–373. 10.1038/nature05058
133
XieW.NobleJ. A.ZissermanA. (2018). Microscopy cell counting with fully convolutional regression networks. Comp. Methods Biomech. Biomed. Eng. Imaging Visual.6, 283–29210.1080/21681163.2016.1149104
134
YangL.YamamotoT. (2016). Quantification of virus particles using nanopore-based resistive-pulse sensing techniques. Front. Microbiol.7:1500. 10.3389/fmicb.2016.01500
135
YuC.VykoukalJVykoukalD. M.SchwartzJ. A.ShiL, P. R. C.et al. (2005). A three-dimensional dielectrophoretic particle focusing channel for microcytometry applications. J. Microelectromech. Syst.14, 480–487. 10.1109/JMEMS.2005.844839
136
ZengJ.ChenC.VedantamP.BrownV.TzengT. R. J.XuanX. (2012). Three-dimensional magnetic focusing of particles and cells in ferrofluid flow through a straight microchannel. J. Micromech. Microeng.22, 1–8. 10.1088/0960-1317/22/10/105018
137
ZhangH.ChonC. H.PanX.LiD. (2009). Methods for counting particles in microfluidic applications. Microfluid Nanoffluid.7, 739–749. 10.1007/s10404-009-0493-7
138
ZhaoY.LiQ.HuX.LoY. (2016). Microfluidic cytometers with integrated on-chip optical systems for red blood cell and platelet counting. Biomicrofluidics10:064119. 10.1063/1.4972105
Summary
Keywords
cytometry, enumeration, biotechnology, microfluidics, microfabrication
Citation
Vembadi A, Menachery A and Qasaimeh MA (2019) Cell Cytometry: Review and Perspective on Biotechnological Advances. Front. Bioeng. Biotechnol. 7:147. doi: 10.3389/fbioe.2019.00147
Received
27 March 2019
Accepted
31 May 2019
Published
18 June 2019
Volume
7 - 2019
Edited by
Ioanna Giouroudi, National Technical University of Athens, Greece
Reviewed by
Giuseppe Maruccio, University of Salento, Italy; Attilio Marino, Istituto Italiano di Tecnologia, Italy
Updates

Check for updates
Copyright
© 2019 Vembadi, Menachery and Qasaimeh.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Mohammad A. Qasaimeh mohammad.qasaimeh@nyu.edu
This article was submitted to Nanobiotechnology, a section of the journal Frontiers in Bioengineering and Biotechnology
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.