Abstract
As nucleic acid testing is playing a vital role in increasingly many research fields, the need for rapid on-site testing methods is also increasing. The test procedure often consists of three steps: Sample preparation, amplification, and detection. This review covers recent advances in on-chip methods for each of these three steps and explains the principles underlying related methods. The sample preparation process is further divided into cell lysis and nucleic acid purification, and methods for the integration of these two steps on a single chip are discussed. Under amplification, on-chip studies based on PCR and isothermal amplification are covered. Three isothermal amplification methods reported to have good resistance to PCR inhibitors are selected for discussion due to their potential for use in direct amplification. Chip designs and novel strategies employed to achieve rapid extraction/amplification with satisfactory efficiency are discussed. Four detection methods providing rapid responses (fluorescent, optical, and electrochemical detection methods, plus lateral flow assay) are evaluated for their potential in rapid on-site detection. In the final section, we discuss strategies to improve the speed of the entire procedure and to integrate all three steps onto a single chip; we also comment on recent advances, and on obstacles to reducing the cost of chip manufacture and achieving mass production. We conclude that future trends will focus on effective nucleic acid extraction via combined methods and direct amplification via isothermal methods.
1 Introduction
Nucleic acids are biomacromolecules formed of nucleotide monomers; they are among the most basic substances of life. Both deoxyribonucleic acid (DNA) and ribonucleic acid (RNA) play a vital role in the transmission of genetic information. The importance of nucleic acid detection is increasing in fields such as species authentication, point of care testing (POCT) (), diagnostics (Zhao et al., 2021), food safety (Shang et al., 2020), and forensics (), among others. Compared with conventional immunoassay (; Ma et al., 2020; ), nucleic acid testing could provide a lower detection limit after amplification, more accurate and convenient quantification, and higher specificity against pathogens that have similar shell structures and different genetic materials. In these applications, a very large number of samples need to be processed in limited time, which has prompted the development of many new detection methods. Even so, there remains a growing need for rapid on-site detection methods, since this approach provides the most reliable characterization of freshly collected samples and accelerates the entire testing procedure, with little dependence on costly apparatus. Moreover, rapid on-site methods are often easy to use and require less operator training, which means they also have advantages in civic applications. The ability to perform massive nucleic acid testing in limited time has attracted further attention due to the outbreak of COVID-19. Further development in the domain of rapid nucleic acid detection methods is crucial to improve the current health care system.
Rapid on-site nucleic acid tests generally consist of three steps: sample preparation, in which nucleic acid is extracted from the testing subject; amplification, in which the target nucleic acid is amplified via various methods; and detection, in which the amplified product is qualitatively or quantitatively analyzed. Most publications refer to ādetection timeā as the time required for the amplification and detection steps, neglecting sample preparation time. However, the three steps relate closely to one other and should not be discussed separately. For instance, the use of isothermal nucleic acid amplification methods may not dramatically decrease amplification time compared to traditional polymerase chain reaction, but it can greatly reduce sample preparation time.
Microfluidic methods have several benefits and have attracted extensive attention. They are widely used in sample preparation, amplification, and detection. On-chip methods enable automatic sample manipulation and require smaller amounts of reagent and less expensive testing devices. The high surfaceāvolume ratio and small thermal mass involved in these methods also improve the efficiency of heat transfer during amplification. For these reasons, microfluidic methods are suitable for the development of novel rapid methods for on-site detection of nucleic acid. In recent years, an increasing number of āall-in-oneā or āsample-in-answer-outā systems have been developed to perform each of these three steps on a single chip. Many review papers have been published on this topic. For instance, Wu et al. reviewed on-chip methods for DNA extraction, PCR amplification, and detection, but this was the most recent occasion on which on-chip methods were classified according to these three steps in a systematic review (Wu et al., 2014). Since then, vast improvement has occurred in isothermal amplification methods, which calls for an update. Other recent reviews have focused on different issues concerning related topics, such as on-chip cell lysis (; ; Petrou and Ladame, 2022), the role of magnetic beads in these steps (), on-chip PCR methods (), the contribution of 3D printing to on-chip amplification (Tzivelekis et al., 2021) and detection methods for all-in-one systems (Li et al., 2021), and microfluidic sensors (Sharma and Sharma, 2022). Several other reviews have also focused on these steps without limiting their scope to on-chip methods; these have covered topics such as sample preparation (Jeon et al., 2018; ), isothermal amplification (Obande and Singh, 2020; Pumford et al., 2020; Islam and Koirla, 2021; Maiti et al., 2022), and detection (; Leonardo et al., 2021; Reyes et al., 2021; ).
The aim of this paper is to offer a systematic introduction to on-chip methods of nucleic acid sample preparation, amplification, and detection that can be used in rapid on-site detection, and to discuss recent advances and challenges in the process of integrating these steps to develop an all-in-one system. The principles of various methods for cell lysis, nucleic acid purification, amplification, and detection are explained, with recent examples provided in each category. The examples selected either exhibit great potential in the domain of rapid on-site detection or employ novel designs that could help to improve performance, throughput, and integration or to reduce costs. In the final section, we discuss the integration of the three steps in all-in-one systems and present predictions on future trends for further development.
2 Rapid sample preparation methods
Samples containing target DNA or RNA fragments need to be pretreated prior to further processing. The protocol for pretreatment varies depending on nature of the sample (plant or animal cells, virus, bacteria, etc.), the environment of sample collection (tissue sample, saliva, blood, sweat, urine, etc.), and the subsequent test procedures that will be carried out. In general, however, it involves two processes: lysis (destruction of the plasma membrane) and purification (to separate the nucleic acid from inhibitors that can affect amplification). Below, we present recent advances in on-chip cell lysis and purification methods and discuss the key issues in combining these two processes on a single chip.
2.1 On-chip cell lysis methods
Plasma membranes can be deformed using a chemical reagent, mechanical force, electric field, or heat. Chemical lysis is the most common choice for on-chip lysis; under this approach, alkaline solution or surfactants are used to disrupt the plasma membrane. In alkaline lysis, the solution is first adjusted to produce an alkaline environment (often between 11.5 and 12.5 pH) so that hydroxide ions (OHā) can break down the fatty acidāglycerol ester bonds on the plasma membrane, causing it to become permeable. Subsequently, sodium dodecyl sulphate (SDS) is used to help dissolve the proteins and membrane. Although alkaline lysis is applicable to most cell types, its low reaction rate is a major disadvantage in rapid detection. Wang et al. selected E. coli and E. durans as models to study the efficiency of Gram-negative and Gram-positive bacteria lysis that could be achieved within 2Ā min (Wang et al., 2020c). They found that E. durans was barely lysed by homogeneous alkaline solution within such a limited period; E. coli was found to be lysed when pH reached 10, and was best lysed at a pH of 13. Further increase in the pH did not provide better results due to probable damage to the nucleic acid. In comparison, they reported better results with electrochemical lysis methods in 1Ā min. Surfactants or detergents, especially non-ionic surfactants that cause less damage to proteins and enzymes, are widely used for lysis of mammalian cells. However, surfactants need to be combined with lysozymes when used to lyse cells that have multiple outer layers, such as bacteria. The precise distribution and effective mixing of the sample and lysing buffer is the key issue in on-chip chemical lysis; have provided a detailed review of this question. Compared with a homogeneous liquid phase, waterāoil droplets might provide better mixing and prevent loss of lysed content. For example, Shamloo and Hassani-Gangaraj, (2020) utilized a waterāoil droplet-based chip and took advantage of secondary flows generated inside the droplet to provide better contact between the lysis buffer and sample and to keep the microchannel in good condition for reuse. Despite the convenience of chemical lysis, chips using this method often need to be equipped with a mixing unit to further decrease lysis time. Additionally, cell residues in lysis buffer might disrupt subsequent steps and have to be carefully removed.
Mechanical lysis physically destroys the plasma membrane via shear stress. In the bead milling or bead beating method, samples are ground with rigid beads made of glass, ceramic, metal, or metal oxides at a high speed; the efficiency of this method depends on the size, shape, and composition of beads (). Novel mechanical lysis methods combine nanoscale patterning with the use of acoustic or piezoelectric force to induce similar shear stress. Wang K et al. (2019) built an acoustofluidic device containing 180 pairs of sharp edges and completed cell lysis via application of high shear force created by an acoustic streaming effect. The surface of the channel was incubated with 5% Pluronic F-127 to prevent attachment of cell debris. also used acoustic streaming to induce collision between nanowires for cell lysis. Lysis efficiency of 97% was achieved after 10Ā s of stimulation with power less than 1Ā W, which suggests that this could be a gentle and effective method of lysis. Nittala et al. (2022) fabricated silicon components with various micro-patterns (pyramids, pillars, ridges, dense pointed structures, and needles) and tested their lysis efficiency as driven by piezoelectric actuation. Zhang G. et al. (2021) developed a pillar array chip to trap and lyse red blood cells via acoustic wave. Simulations of streaming velocity near the pillars and shear stress around the micropillars were provided to help regulate cell lysis. The authors also noted that the pillar array was not able to effectively trap all the cells at high cell densities. In some studies, researchers have experimented with creating microbubbles inside channels instead of nanostructures, which could simplify the design and fabrication of microchips. Frictional force generated by the gas/liquid interface could create ācavitation microstreaming,ā causing strong circulatory flow. Liu X et al. (2020) designed a microbubble array to lyse multiple cells with oscillating bubbles. The lysis efficiency of this system was measured at 97.62% after 1Ā min, similar to the results obtained by commercial chemical kits in 15Ā min. The primary drawbacks of mechanical lysis are the difficulty and high cost of fabricating delicate nanostructures and the decrease in lysis efficiency that occurs when cell debris clings to the surface of the microchannel.
Electrical cell lysis produces irreversible pores in the cell membranes through application of a high electric field. It has several advantages, in the form of short lysis time, simple channel structure, and low requirements for chemical reagents. Pandian et al. (2020) have laid out a method of predicting transmembrane potential at different coating thicknesses and voltages to help determine the optimal lysis parameters. Lysis efficiency of 98% can be achieved by following this guidance. More importantly, it is possible to selectively lyse cell membranes by controlling the strength of the electrical field, which enables the preservation of intracellular membrane structures such as the nuclear membrane (Jeon et al., 2018). Wu et al. (2020) created field gradients at the surface of polarizable active metallo-dielectric Janus particles with an external electric field, which led to selective electroporation of bacteria.
In thermal lysis, an external heater is used to continuously denature cell membranes. Qian et al. (2022) built a portable microfluidic system that lyses exosomes at 95°C for 10 min. Nguyen et al. (2022) applied a similar protocol for virus lysis. In addition, they used a vibrator to agitate the solution for 30s every minute during the heating process, which reduced the heating time from 15 min to 10 min. Bacterial cell walls are relatively more difficult to break down than those of mammalian cells, but this can still be achieved with a slight alteration to the process. Rizzo et al. (2021) applied vigorous agitation with magnetic beads that were used to capture bacteria and improved lysis efficiency. Unfortunately, the authors report only the optimal lysis parameters without indicating the extent to which agitation improved lysis efficiency. Given that the use of a heating process during nucleic acid amplification is unavoidable, thermal lysis can make direct use of the same heating device, which results in a greatly simplified system design. The greatest drawback reported for thermal lysis is the damage that it causes to certain proteins inside cells; although this may affect further analyses, it is not a major concern for nucleic acid detection. For these reasons, thermal lysis has good potential for use in rapid on-chip nucleic acid detection.
Due to the need for quicker and more effective methods of cell lysis, many studies have combined multiple lysis methods to produce further improvements. Thermal and chemical lysis methods could be combined with a simple mechanical method to facilitate heat transfer or provide better contact between the lysis buffer and sample.
2.2 On-chip nucleic acid purification methods
Purification methods can be classified into centrifugation, filtration, and magnetic methods. In addition, another direct approach is available, in which nucleic acid amplification is conducted without further purification. Centrifugation was one of the earliest purification methods to be used in biochemical research, and it lends itself perfectly to combination with a centrifugal microfluidic chip for nucleic acid purification. Homann et al. (2021) present a disk-like microfluidic chip for automatic preparation of Mycobacterium tuberculosis samples; conventional PCR tubes can be connected to the edge of the chip and purified samples collected for further examination. However, with the need for integration of additional steps on a centrifugal chip, channel and valve design become more complicated. Brassard et al. designed a test cartridge that combines pneumatic and centrifugal forces for fluidic control. This design results in a tenfold reduction in the amount of elution buffer used and a simplified channel pattern (). As a result, the costs of the reagent and chip fabrication are both reduced.
In filtration methods, a membrane is used to selectively filter the desired DNA sample from undesired lysate. A silica membrane can bind with DNA at a high salt concentration. Yoon et al. (2021) designed a syringe-based DNA extraction device that uses a silica membrane to separate out target DNA. In some cases, a commercially available filter is externally connected to the chip; this approach makes for an easy design process, but potentially presents problems with leaking (Ryzhkov et al., 2020). In other cases, the membrane serves as one layer of a sandwich structure. Lee et al. (2020) used nanoimprint technology to fabricate a silicon nitride filter membrane with a pore size of 200 nm. Under this type of approach, the membrane is tightly fixed using two sets of screws to form a sandwich structure with Teflon compartments and PDMS panels, and an electric field is used to drive negatively charged DNA and RNA across the membrane. The pore size of membrane is often in the μm range. If the microchannel is not properly sealed, the liquid tends to leak elsewhere before passing over the membrane. For this reason, on-chip filtration methods impose stringent requirements for microfabrication and assembly.
The magnetic method uses magnetic beads coated with silica or other probes that can form specific binding with target DNA; they also make use of magnetic force for further separation. Researchers have proposed various on-chip methods of this type to improve speed or efficiency. Xu et al. (2020) prepared three sub-channels for the sample, washing buffer, and eluate buffer. These sub-channels merged into a main channel, creating multi-laminar flow. By controlling the position of the magnetic field, magnetic beads carrying DNA can be made to pass through each flow sequentially. This work simplifies the procedure because the magnetic beads could pass through the main channel and provide effective separation within 3Ā s. incorporated use of an electric field into an existing magnetic separation chip; DNA extraction yield was increased by 15% thanks to the electroosmotic flow. Finally, bio-specific molecular interactions, such as antigenāantibody interaction, can be used to specifically separate target DNA. combined magnetic beads with polyclonal antibody to detect Salmonella in processed duck meat products; capture efficiency was found to surpass 95%, and the enrichment process was completed within 20Ā min.
The direct amplification method is a technique that allows nucleic acid amplification to be carried out directly after extraction. It relieves researchers of implementing the purification process and reduces the loss of original template nucleic acids to a minimum. The key to direct amplification is to reduce the impact of inhibitors in the lysate. This can be accomplished by the development of inhibitor-resistant DNA polymerase-buffer systems or by diluting the lysate solution to lower the concentration of inhibitors. used Phusion Pfu as an inhibitor-resistant DNA polymerase in the detection of Salmonella in pork samples without purification. Hedman et al. (2009) tested several DNA polymerase-buffer systems as potential alternatives to the standard AmpliTaq Gold polymerase used in PCR, and identified three systems that offered better resistance against inhibitors in saliva samples. A later study by the same authors showed that a further increase in tolerance to inhibitors can be achieved by blending two DNA polymerase-buffer systems: ExTaq Hot Start and PicoMaxx High Fidelity (Hedman et al., 2010). However, this approach increases the cost of the reagent and enzymes (by approximately 2ā3 times) and is not always applicable for different types of target DNA. Dilution of the solution leads to a decrease in the concentration of both the target nucleic acid and inhibitors. Direct amplification is possible if the concentration of target nucleic acid remains above the detection limit while inhibitors are diluted to a safe level. Satya et al. (2013) extracted DNA fragments from leaf tissue and ramie stem tissue using NaOH and Tris/TrisāHCl/TrisāEDTA and used the results for direct PCR after dilution; their protocol enables sample preparation within approximately 10ā12Ā min.
The extracted solution needs to be heavily diluted (about 1000 times) to reach the safe level of inhibitors for PCR. This requires a very large amount of target nucleic acid in the sample. Isothermal amplification methods are more tolerant of the presence of inhibitors, and the sample therefore does not require as much dilution (approximately 10 times is sufficient); or, in some cases, it may be possible to use the sample directly without dilution. Nguyen et al. (2022) demonstrated direct, on-chip RT-LAMP for the detection of SARS-CoV-2: a commercially available direct PCR kit was used to lyse the virus, and the lysate was directly transferred into a reaction chamber containing RT-LAMP reaction solution for isothermal amplification.
2.3 Integration of on-chip lysis and purification
Figure 1 provides a general illustration of various methods of cell lysis and DNA purification. Chemical and thermal lysis methods do not need any special microstructures or chip accessories and are the most compatible with on-chip lysis, but they have limitations in terms of speed and efficiency of lysis. In order to resolve this issue, researchers have simply combined mechanical lysis with chemical or thermal lysis. Since mechanical methods only provide assistance under this approach, the fabrication of complicated microstructures can be avoided. Zhao et al. (2022) added a HeLa or HUVECs cell sample to a Chelex-100 suspension; the mixture was simultaneously heated and agitated. Agitation, as discussed above, can help to improve the effectiveness of thermal lysis. Furthermore, Chelex-100 particles were in full contact with the cells to perform mechanical lysis. The chitosan modification to these particles caused them to bond with nucleic acid immediately after lysis; they could then be used for purification. Kaba et al. (2021) made use of chemical lysis facilitated by a cavitation-microstreaming effect. They used a short-pulsed laser to create cavitation bubbles in the chemical lysis buffer, which simultaneously mixed the lysis buffer and crushed the cell membranes. Sun et al. (2021) added a chemical lysis buffer containing magnetic beads to a blood sample; gas bubbles were injected to fully mix the solution and to improve both chemical lysis and DNA binding.
FIGURE 1
Centrifugal purification is very readily adapted for use as part of a microfluidic chip driven by centrifugal force. As mentioned previously, channel design for a centrifugal chip becomes very complicated with the integration of more on-chip processes. A rotating chip is also not suitable for the application of an electric field. For these reasons, chemical or thermal lysis seem to be the only choices. On the other hand, its outstanding throughput is the main reason to select a centrifugal chip design. Magnetic purification methods not only offer high-selectivity DNA extraction, but also provide assistance in reagent mixing and fluidic control. This has become the most popular purification method for on-chip applications. However, the preparation of the magnetic beads and their coating layer is challenging and expensive. present a thorough review of the preparation of magnetic beads for nucleic acid extraction.
Direct amplification methods require little pretreatment and can be combined with any lysis method. The current limitation with respect to these methods lies in the relatively high cost of commercial direct lysis or amplification kits for PCR. The high level of dilution involved in direct PCR also requires the presence of sufficient nucleic acid in the lysate. Due to their high tolerance to PCR inhibitors and outstanding speed of amplification, isothermal amplification methods have shown great potential for use in future studies. Current direct amplification methods mostly make use of commercial direct PCR kits, even in the case of isothermal amplification. We believe that more direct protocols for isothermal amplification will be developed in the coming years and that this will make direct amplification under isothermal conditions a feasible and economical solution for rapid on-site nucleic acid testing.
3 Rapid DNA amplification methods
3.1 On-chip PCR methods for rapid amplification
Traditional PCR protocols consist of three steps: a denaturation step, in which the hydrogen bonding in the double-stranded DNA template is ruptured by exposure to a high temperature (95°C); an annealing step, in which the primer interacts with the template DNA at 40°Cā60°C; and an extension step, in which the template DNA is amplified with the help of DNA polymerase at around 70°C. In some cases, the annealing and extension steps are merged to save time, but the entire process still takes approximately 1.5ā2Ā h due to the repeated thermal cycles. Many portable microfluidic devices designed for PCR have been reported to reduce the required sample volume from 20 µL in traditional PCR protocols to 5ā10 µL. The decrease in sample volume significantly decreases the thermal mass of the entire system. The high surface-to-volume ratio also enables more effective on-chip heat transfer. These features can speed up the process of applying the thermal cycles, reducing the time required from a few hours to approximately 30Ā min. Optimization of the heating/cooling process is crucial to the development of rapid amplification methods. Four common chip designs for PCR are illustrated in Figure 2; detailed accompanying descriptions are provided in this section.
FIGURE 2
3.1.1 Stationary chamber PCR chips
Stationary chamber PCR is also referred as time-domain PCR: the solution is kept stationary while the temperature of the reaction chamber is modified to apply the necessary thermal cycles. Lee et al. (2005) developed a microfluidic chip that could complete 30 thermal cycles for a 15 μL sample within a total time of 26 min and 24 s. Taking a different approach, Muddu et al. (2011) heated the reaction chamber to different temperatures at each end and induced circulation of the reaction solution. The temperature gradient provided thermal cycling and enabled completion of the PCR reaction in 10 min. A stationary chamber PCR chip requires little space for a single reaction chamber; therefore, throughput of tests could be increased through integration of multiple reaction chambers on a single chip. Chips containing multiple reaction chambers are classified as multiple-chamber PCR chips. Matsubara et al. (2004) prepared a silicon-based array chip containing 1248 microchambers, with each chamber consuming only 40 nL of sample solution. Although multiple-chamber PCR chips provide high throughput, surface treatment of the inner walls of the chamber and a tight seal are often required to avoid cross-contamination. Moreover, microfluidic chips do not offer the same precision as complex PCR devices in terms of temperature control. Precision of temperature control has a strong impact on denaturation, enzyme activity, the efficiency of extension, and (most importantly) the specificity of amplification. Temperature gradient is more important for multiple-chamber chips, since an uneven distribution could also make comparison between different chambers untrustworthy.
3.1.2 Closed-loop PCR chips
In closed-loop PCR chips, PCR solution is contained within a closed microchannel and periodically flows through three different temperature zones. The benefits of the closed-loop approach include easy control of the number of thermal cycles by simply changing the circulation time, a considerably shorter microchannel (which facilitates the construction of a miniature portable device), and good uniformity between each thermal cycle. Lok et al. (2012a) reported on the design of a magnetically actuated circular closed-loop PCR chip, in which a ferrofluid plug is driven by an external magnet that moves along the microchannel and four loops can carry out PCRs simultaneously. fabricated an identical closed-loop channel on a centrifugal disk-like chip; this design can apply 106 thermal cycles in 15Ā min. The drawbacks of closed-loop PCR chips are the inconvenience of retrieving the solution and an asymmetrical reaction procedure in concentric channels when high-throughput testing is attempted.
3.1.3 Serpentine channel PCR chips
In serpentine channel PCR chips, tight bends are applied to a microchannel to form a series of connected parallel channels, while a heating module creates three temperature zones. Trinh et al. (2018) reported on the use of a serpentine channel on a PS surface with CNC milling; the surface of chip was able to capture bacteria with styrene groups, and Escherichia coli O 157:H7 was successfully detected in raw milk. The serpentine channel design is not limited to use with a traditional square shape; it can also be employed in the form of a radial circular chip. For instance, Zhang et al. (2022) designed a radial serpentine channel and heated it using a laser. The gradient of laser intensity created three temperature zones, in which a higher temperature was induced in regions closer to the center. The longer length of the peripheral part of the channel also provided a longer extension time. Additionally, the use of soft tubing instead of a manufactured microchannel enables alteration of the number of thermal cycles. Trinh and Lee, (2018) fabricated a micro-device with a glassāpolytetrafluoroethylene (PTFE)āglass sandwich structure, in which a PTFE tube was twisted and fixed between two pieces of glass and acted as the microfluidic channel. This design enables greater flexibility in the number of thermal cycles.
Channel design has a direct impact on the time distribution of the denaturation, annealing, and extension steps. In order to increase extension time, Ragsdale et al. (2016) modified the width of the channels to impose a rapid transition from hot to cold regions and a slow transition from cold to hot regions. Li et al. (2020d) analyzed the influence of channel width-to-depth ratio and of the length ratio between the three temperature zones on temperature and flow distribution. Their results showed that the high temperature region should cover a larger area than the low temperature region to achieve optimal amplification of short DNA sequences.
3.1.4 Spiral channel PCR chips
Spiral channel PCR chips use flexible tubes as a microchannel for the reaction solution. These tubes are twisted into a spiral shape and pass through multiple temperature sectors in a planar or cylindrical design. The number of cycles applied in each experiment is controlled by altering the length of the tube and the number of twists; hence, spiral channel PCR chips are more flexible than serpentine channel PCR chips. Wu and Wu, (2019) designed a portable PCR system based on a Teflon tube wrapped around a TEC chip; not only did the novel design reduce the size of the entire system, but the heat transfer rate was also 20%ā30% faster than in conventional methods. Trinh et al. (2017) created a temperature gradient by positioning a PDMS mold swathed with PTFE tubing on a hot plate at constant temperature of 105°C; the temperature of the annealing and extension process was controlled via the height of PDMS mold.
Researchers have also used high-precision droplet manipulation techniques to create multiple droplets in the microchannel to increase throughput. Madadelahi et al. (2019) designed a high-throughput two-phase PCR device with a serpentine channel for droplet generation and a spiral channel positioned on a planar plate for thermal cycling; they concluded that such a device could reduce the time required for a PCR experiment by more than 40%.
3.2 Isothermal nucleic acid amplification chips
Isothermal nucleic acid amplification methods have emerged during the last decade; these include loop-mediated isothermal amplification (LAMP), recombinase polymerase amplification (RPA), and helicase-dependent isothermal DNA amplification (HDA), among others (Pumford et al., 2020). Some isothermal amplification methods are not affected by PCR inhibitors, making them more suitable for rapid applications. Below, we present the principles underlying three isothermal nucleic acid amplification methods that are reported to have good resistance to PCR inhibitors (Kaneko et al., 2007; Kersting et al., 2014; Nixon et al., 2014; Wang et al., 2016), along with their on-chip applications, and discuss current challenges in on-chip amplification. The principles of the isothermal amplification methods discussed are also illustrated in Figure 3.
FIGURE 3
3.2.1 Principles and comparison of isothermal amplification methods
3.2.1.1 Loop-mediated isothermal amplification
Loop-mediated isothermal amplification was first introduced by Tsugunori et al. in 2000 (Tsugunori et al., 2000). The principle of the reaction is illustrated in Figure 3. LAMP uses six regions for initial recognition of the target sequence, and then uses four primers during the subsequent amplification, elongation, and recycling processes; this provides very high specificity. In some studies, two additional primers have been used to further increase specificity (Qin Y et al., 2021). Unlike PCR, which produces exact duplicates of the template DNA, the LAMP reaction produces dumbbell-shaped or stem-loop DNA structures with different lengths. Its product displays multiple stripes in electrophoresis tests. LAMP offers excellent efficiency in DNA amplification: reported a detectable concentration of DNA after 8Ā min of amplification of a high-concentration sample (12Ā ng/assay) and after 60Ā min in the case of a low-concentration sample (1.2Ā fg/assay). Misir, (2018) compared the time required for amplification by 109 to 1010 times via PCR and LAMP; the results showed that LAMP could complete this 1Ā h sooner than traditional PCR.
3.2.1.2 Recombinase polymerase amplification
Recombinase polymerase amplification was first introduced by Piepenburg et al. (2006) in 2006. RPA begins with a binding process in which recombinase binds to a pair of primers, forming a nucleoprotein filament that can identify homologous sequences and create a D-loop structure to initiate a strand exchange reaction. Once the reaction is complete, the recombinase is released for the next pair of primers; eventually, two DNA duplexes are formed, and this process is repeated to amplify the target DNA. The recommended temperature for RPA is between 37°C and 42°C, which is the lowest temperature reported to date at which isothermal nucleic acid amplification has been achieved. The necessary temperature could be provided without the use of any heating device: in some cases, the temperature has been provided merely by body heat (), or the process has even been carried out at room temperature, with some sacrifice of accuracy (Shen et al., 2011). The amplification rate achieved by RPA is outstanding compared to other isothermal nucleic acid amplification methods. Xia et al. (2014) were able to amplify DNA to a detectable range within 10 min; in an efficiency comparison with the qPCR method, the same amplification result was achieved by qPCR after 134Ā min. Another method, known as ārecombinase-assisted amplificationā (RAA), makes use of a similar underlying principle to RPA; the main difference lies in the source of enzymes (Hou et al., 2022). In this review, we also include studies using RAA under the category of RPA.
3.2.1.3 Helicase-dependent isothermal DNA amplification
Helicase-dependent isothermal DNA amplification was first introduced by Vincent et al. in 2004 (Vincent et al., 2004); HDA uses a DNA helicase enzyme, rather than high-temperature-triggered denaturation, to generate a single-stranded DNA template for primer hybridization, which enables isothermal amplification. Earlier work used an E. coli UvrD helicase/DNA polymerase I Klenow fragment pair with two accessory proteins (MutL and SSB) to amplify DNA at 37°C, and was referred to as mesophilic HDA (mHDA) for this reason. used a thermostable helicase, Tte-UvrD, together with Bst DNA polymerase to amplify DNA at 65°C; this version of the process was therefore referred to as thermophilic HDA (tHDA). Compared with mHDA, tHDA avoids reliance on accessory proteins and offers a more sensitive amplification process, making tHDA the more efficient method.
3.2.1.4 Comparison of three isothermal amplification methods
Researchers always face the dilemma of how to choose from among the various isothermal amplification methods. Table 1 provides an overall comparison of the three isothermal amplification methods introduced above. In selecting the optimal method, the researcher must consider the length of the target sequence. Other than target length, the difficulty of primer design, amplification performance, and the cost of each test are also factors with a major impact on the use of rapid on-site detection methods. The cost is primarily dictated by the number and quantity of enzymes used during the reaction. Although on-chip methods may reduce the volume of reaction solution required, researchers have observed based on experience that there is a minimum amount required to successfully carry out the reaction. On the other hand, restrictions also arise from the detection method employed. For instance, LAMP-amplified products are fragments of different lengths with a repeating sequence of target nucleic acid; in this case, detection methods based on DNA hybridization probes cannot offer ideal detection. RPA-amplified products contain a large amount of enzyme residue, which will affect the result of electrophoresis or use of certain fluorescent dyes. For this reason, the detection of RPA-amplified products is often completed with the help of nucleic probes (Li et al., 2020b; Liu et al., 2021).
TABLE 1
| Amplification method | Reaction time | Preferable target length | Difficulty of primer design | Number of enzymes used | Reaction temperature |
|---|---|---|---|---|---|
| LAMP | 20ā60Ā min | <300 bps | Moderate | 1 | 60°Cā65°C |
| HDA | 20ā60Ā min | 70ā120 bps | Simple | 2 | 37 or 65°C |
| RPA | 10ā20Ā min | <500 bps | Difficult | 3 | 37°Cā42°C |
Comparison of different isothermal amplification methods.
Although LAMP provides excellent specificity and an excellent amplification rate, its primary drawbacks are the limitation on target DNA length and the complex structure of the amplicon. The optimal target length for LAMP is less than 300Ā bps, and it is not recommended for amplification of target DNA longer than 500Ā bps (Pumford et al., 2020). As discussed above, the shape of the amplicon brings an added challenge to the detection process. However, the byproduct magnesium pyrophosphate also allows for indirect detection; this is discussed in the third section of this review.
The primary drawback of RPA is the high level of difficulty of designing primers. Since RPA operates at a relatively low temperature, secondary structures formed by longer primers are not denatured, and this could result in unsuccessful amplification; the use of primers within the range of 30ā45 base pairs is recommended. Other key parameters affecting RPA primer design include GC content, target region, termini sequence, etc.; their influence have been discussed and explained in several articles (; Lobato and O'Sullivan, 2018; Matthew et al., 2018). No report on automatic primer design software has yet been published, but the literature contains mention of several tools that could help researchers to screen several primer candidates for further optimization (Matthew et al., 2018; Li et al., 2020b). The second drawback of RPA is its limitation on the choice of target sequence. Although RPA is capable of amplifying long sequences, up to 1.5Ā kb, it is better suited to short sequences ranging from 80 to 400Ā bps, and preferably between 100 and 200Ā bps (; Lobato and O'Sullivan, 2018). Fluorescent detection of RPA product is expensive due to the need for a fluorescence probe, which is why RPA product is often detected using lateral flow strip or other non-fluorescent detection methods.
Figure 4 illustrates the number of papers published on the theme of various amplification methods; data were obtained from the core database of the Web of Science and the MEDLINE database on 7 November 2022. As this figure demonstrates, LAMP and RPA are drawing increasing amounts of attention, while HDA is declining in relative appeal. This is mostly due to obstacles in screening for more effective helicase. Because of the absence of a heating step, HDA is limited by the low denaturation efficiency of DNA helicase, leading to low specificity and an unstable amplification rate. Several studies have introduced nanoparticles in order to suppress primer dimer formation or to prevent binding between the template and proteins as a way to improve the specificity and speed of HDA (; Sedighi et al., 2017). However, this greatly increases the cost of amplification, and researchers are still seeking more practical solutions.
FIGURE 4
3.2.2 On-chip isothermal amplification
The reaction system for isothermal amplification is quite simple, which makes the corresponding chip design simpler than in the case of on-chip PCR. The same apparatus could be shared by most on-chip isothermal amplification methods, since the only difference in the amplification protocol is the heating temperature. The major difference in chip design arises from differences in sample preparation and detection strategy. Various pumping mechanisms have been used to mix the reagent and sample, such as centrifugal mechanisms (Soares et al., 2021; Zhou et al., 2021), a syringe pump (Hardinge et al., 2020; Natsuhara et al., 2020), a peristaltic pump (Najjar et al., 2022), magnetic force (Sharma et al., 2022), capillary forces (Zhu et al., 2020), etc. A serpentine channel is also used for amplification to provide homogeneous and effective heating. Tsougeni et al. (2019) carried out an HDA reaction in a 60 cm serpentine channel with a total volume of 30 µL. The results indicated that amplification efficiency was 96% relative to data obtained by using a conventional thermal cycler. In the case of direct amplification, several teams have introduced extracted sample directly into the reaction chamber without purification (Zhu et al., 2020; Nguyen et al., 2022). Novel fabrication methods have also proposed to ease manufacture of chips and reduce costs. used 3D printing to manufacture monolithic microfluidic chips for use in RPA reaction. They present an improved post-curing protocol to avoid autofluorescence and fluorescence drift, avoiding the need for further surface treatment.
Microfluidic methods are capable of achieving detection of multiple targets in the same sample by dividing the sample into several parallel reaction chambers (Yoo et al., 2020; Wen et al., 2022). However, the high amplification rate is likely to cause cross-contamination between each chamber. Most studies have used a thin film to seal the reaction chambers after sample injection (Wen et al., 2022), but this step also requires a skilled operator to avoid cross-contamination. Wax sealing could be used to seal each reaction chamber without requiring open-lid operation. Nguyen et al. (2022) installed wax valves (melting point 58°Cā62°C) in each channel; after sample injection, a short period of heating was applied to melt the wax valves and seal the reaction chambers. Oil sealing does not require heating and is easier to control. Tian et al. (2022) preloaded mineral oil into syringes and injected this into each reaction chamber to complete the sealing process.
Many efforts have been made to increase test throughput. A disk-like centrifugal chip has a condensed fluidic pattern and can achieve very high throughput. Zhou et al. (2020) used a commercialized multi-channel disk chip containing four sets of microfluidic channels, each of which had eight reaction chambers. Their chip could test eight targets, including one negative control, in four different samples, and was used for rapid screening of emerging and re-emerging enteric coronaviruses in swine. Similarly, Nguyen et al. fabricated a centrifugal disk that could carry out 30 LAMP reactions for sex-typing (Van Nguyen and Seo, 2021).
3.3 Summary of on-chip methods of DNA amplification
The results of 40 studies of on-chip PCR systems are listed in Table 2. More studies than this were identified altogether, but we highlight only certain examples that are more recent or in which the system exhibited improved performance in rapid detection; the reader may find the remaining studies in the Supplementary Materials. Most of the PCR chips investigated were able to complete the reaction within 1Ā h, and the quickest could do so within 10Ā min. In general, continuous flow PCR chips are more time-efficient due to the removal of heating/cooling processes, and they can be easily connected to another microfluidic chip for detection of amplicon. Stationary chamber PCR chips have higher throughput. Based on the timelines reported in the listed studies, we draw the conclusion that the serpentine channel has become a mainstream design for on-chip PCR. The main reasons are the convenience of retrieving the amplification solution compared to a closed-loop structure and higher-precision temperature control compared with a spiral structure.
TABLE 2
| Driving force | Number of cycles | Single cycle time | Total time | Limit of detection | Flow rate/Temperature change rate | Sample volume | Target | Potential for high-throughput testing | References |
|---|---|---|---|---|---|---|---|---|---|
| negative pressure | 40 | 75Ā s | 62Ā min* | ā | ā | ā | Epidermal growth factor receptor gene mutation | NO | Yin et al. (2020b) |
| syringe pump | ā | ā | 8Ā min 5Ā s* | ā | 0.1ā2 μL/min | 50 μL | Porphyromonas gingivalis Tannerella forsythia, Treponema denticola | YES | Yang et al. (2022) |
| liquid pump | 40 | 90Ā s | 72Ā min* | ā | 0.4 μL/min | 5 μL | Colla corii asini | NO | Sheu et al. (2022) |
| syringe pump | 27 | 120Ā s | 56Ā min* | ā | ā | 0.5 μL | Yeast expression vector | YES | Madadelahi et al. (2019) |
| syringe pump | 30 | ā | ā | ā | 0.773ā0.889Ā mm/s | ā | ā | NO | Whulanza et al. (2017) |
| syringe pump | 25 | 90Ā s | 56Ā min* | ā 102Ā CFU | 5Ā mL/min | 10 μL | Escherichia coli O157:H7 | NO | Trinh et al. (2018) |
| syringe pump | 40 | 65 s | 48 min 20 s* | 102 CFU/mL | 0.25 mL/min | 12 μL | Salmonella | NO | Wang Y et al. (2021) |
| ā | 40 | 60Ā s | <1Ā h | 1 copies/μL | ā | 5 μL | Swine disease | YES | Jiang et al. (2021) |
| stepper motor | ā | ā | 8Ā min | ā | 0.001ā0.02Ā m/s | ā | Bacterial 16S ribosomal DNA | NO | Li et al. (2020) |
| ā | 45 | 60Ā s | 45Ā min* | 50 copies | ā | 50 μL | HPV | NO | Zhu et al. (2019) |
| ā | 45 | 75Ā s | 66Ā min 15Ā s | ā | heating rate: 15°C/s cooling rate: 10°C/s | 10 μL | HBV | YES | |
| syringe pump | 25 | 15Ā s | <13Ā min | ā | <0.5Ā mL/h | 10 μL | Genetic markers | NO | Ragsdale et al. (2016) |
| syringe pump | 50 | 26Ā s | ā | ā | ā | 10 μL | Human Genomic DNA template | ā | |
| self-activated micropump | 40 | 50Ā s | ā63Ā min* | ā | 50s/cycle | 20 μL | H7N9 | NO | Wang Z et al. (2019) |
| peristaltic pump | 30 | 75Ā s | 50Ā min | ā | heating rates: 1.5°C/s cooling rates: ā2.0°C/s | 20 μL | HPV | NO | Liu et al. (2017) |
| stepper motor | 35 | 40 s | 13 min 20 s | 125 CFU/μL | 100 V/cm | 50 μL | 16S rDNA of periodontal pathogens | NO | Li et al. (2019) |
| peristaltic pump | 30 | 10Ā s | <5Ā min | ā | 5 μL/min | 25 μL | Mouse GAPDH housekeeping gene | NO | Moschou et al. (2014) |
| self-activated micropump | 40 | ā60Ā s | 40Ā min* | ā | 8.5 μL/min | 40 µL | HBV | NO | Wu and Wu, (2019) |
| syringe pump | 40 | 60Ā s | <120Ā min* | 10 copies | 47 μL/min | 0.1ā10 μL | ā | YES | Hatch et al. (2014) |
| hydrostatic pressure | 30 | ā | ā | ā | 0.006Ā mL/min | ā | ā | NO | |
| microfluidic pump | 40 | 33Ā s | 32Ā min* | 10 copies/μL | ā | 70 μL | RNA virus | NO | |
| syringe/peristaltic pumps | 35 | 85Ā s | 85Ā min | ā | heating rate: 6°C/s cooling rate: 4°C/s | 6 μL | β-actin gene | YES | |
| syringe pump | 27 | 120Ā s | 50Ā min | ā | ā | 10 μL | ā | YES | Madadelahi et al. (2019) |
| syringe pump | 25 | 29Ā s | ā15Ā min | ā | ā | 25 μL | Synthetic pGEM-T vector inserted with the TTF-1 target gene | NO | Trinh et al. (2017) |
| syringe pump | 30 | ā | <25Ā min | ā | 2 μL/min | 20 μL | C. condimenti and E. coli O157:H7 DNA | NO | Trinh et al. (2016) |
| syringe pump | 20 | ā | ā20Ā min | 0.3 Ć 104Ā CFU | 2 μL/min | ā | E. coli O157:H7 and Salmonella spp. | NO | Trinh and Lee, (2018) |
| peristaltic pump | 36 | ā | ā30Ā min | ā | 1 μL/min | 15 μL | Human cell lines: BC-3 and IBL-1 | NO | Snodgrass et al. (2016) |
| syringe pump | 25 | ā | ā | ā | ā¤1Ā mL/h | ā | ā | ā | Thomas et al. (2017) |
| syringe pump | 40 | 40Ā s | 30ā40Ā min | ā | water phase: 0.4Ā mL/h oil phase: 1.2Ā mL/h | 50 μL | HBV | NO | Li et al. (2020) |
| pneumatic pump | 30 | 75Ā s | 47.5Ā min* | ā | ā | 10 μL | β-actin gene | YES | Trung et al. (2010) |
| magnetic force | 25 | 36Ā s | 20Ā min | 1.63 copies/μL | ā | 0.5 μL | Bacteriophage lambda | YES | Lok et al. (2012b) |
| syringe pump | 32 | ā | ā | ā | 60 uL/min: 5Ā mm/s | ā | ā | NO | |
| ā | 40 | 45Ā s | 35.5Ā min* | ā | 20°C/s | 100Ā nL | Ebola virus | NO | |
| centrifugal force | 35 | 150Ā s | 89.5Ā min* | ā | heating rate: 0.7ā0.8°C/s cooling rate: 0.9ā1.0°C/s | 7.5 μL | Human serum | YES | |
| syringe pump | 35 | 260Ā s | 165Ā min* | ā | 400 μL/min | 20 μL | C. albicans | NO | |
| piezoelectrically pumped | 35 | 20 s | <15 min | 5 fg/μL | 45 μL/min | 200 μL | E. coli DNA | YES | |
| motor | 30 | 90Ā s | <30Ā min | ā | ā | 5 μL | SSP150 DNA template | YES | Sugumar et al. (2012) |
| syringe pump | 30 | 21Ā s | ā | 100 copies/mL | 4Ā mm/s | 55 μL | Salmonella enterica Listeria monocytogenes Escherichia coli O157:H7 Staphylococcus aureus | YES | Shu et al. (2014) |
| capillary force | 50 | ā | ā30Ā min | ā | ā | 50 μL | β-actin, Escherichia coli AH1pdm influenza virus Influenza virus H1N1 | NO | Tachibana et al. (2015) |
| syringe pump | 35 | ā20Ā s | <30Ā min | ā | 0.8Ā mL/min | 10 μL | KSHV/HHV-8 | NO | Jiang et al. (2014) |
Performance of 40 on-chip PCR systems. Total times marked with * indicate a value calculated based on protocols described in the relevant article; other values are quoted directly from the article.
On-chip PCR methods are facing two challenges. The first is their weak potential in the area of direct amplification. The second is how to manage the balance between simplicity of the device and high-precision temperature control. Efforts have been made to make sure that the temperature inside each reaction chamber is the same as the value measured by the sensor and to minimize the temperature gradient among different reaction chambers to improve uniformity of the results. In continuous flow PCR chips, the temperature gradient between different temperature zones may cause non-specific amplification. Several works have reported the addition of temperature gaps to isolate temperature zones, thereby reducing the amount of heat transferred. Other researchers have simplified the system by carrying out the extension and annealing steps in the same temperature zone (72°C). On the other hand, continuous flow PCR chips do not monitor temperature dynamically, which means a less complicated heating system is required. In addition, flow velocity has an impact on temperature control in continuous flow chips, because the solution carries heat to other temperature zones as it travels through the microchannel; thus, researchers should not unthinkingly increase the flow rate to reduce reaction time (Moschou et al., 2014).
Isothermal amplification methods enable nucleic acid amplification at a constant and moderate temperature (normally below 65°), and their specificity is not strongly affected by precision of temperature control. In some cases, sunlight and body heat can provide sufficient temperature control, which has facilitated the design of portable devices (Rathore et al., 2019). Moreover, on-chip isothermal amplification has displayed better potential for use with rapid detection than on-chip PCR methods. In Table 3, we list reported details of the performance of 30 on-chip isothermal amplification systems. In general, isothermal methods can save 10ā15Ā min compared to PCR. Amplification time is generally between 15 and 40Ā min; in the quickest case, amplification can be completed in 5Ā min. Multiple reactions take place asynchronously without disturbance arising from denaturation. However, the high amplification rate is likely to cause aerosol contamination or carry-over contamination. On-chip methods can avoid open-lid operation, and they are considered to be a complementary strategy for isothermal amplification. Unfortunately, in most studies described in the listed articles, the amplicon was collected for further analysis, which would increase the possibility of false positive or aerosol contamination. Moreover, the three isothermal methods discussed here are highly tolerant to the presence of inhibitors in the lysate, which makes them suitable for direct amplification and could further reduce the total time to detection.
TABLE 3
| Amplification method | Driving force | Limit of detection | Dynamic range | Amplification time | Detection method | Sample volume | Target | References |
|---|---|---|---|---|---|---|---|---|
| LAMP | centrifugal force | 0.5 copies/μL | 0.5ā¼103 copies/μL | 60Ā min | fluorescence signals | 4 μL | SARS-CoV-2 | Tian et al. (2020) |
| LAMP | capillary force | 100Ā fg/μL | 1Ā ng/μLā¼100Ā fg/μL 108ā¼102Ā CFU/mL | 45Ā min | gel electrophoresis | 10 μL | Cryptococcus | Tian et al. (2022) |
| LAMP | syringe pump | MYSV: 11.1Ā ng/μL CCYV: 9.6Ā ng/μL | ā | 20ā60Ā min | fluorescence | 3.1 µL | Tomato yellow leaf curl virus, melon yellow spot virus, cucurbit chlorotic yellows virus | Natsuhara et al. (2020) |
| LAMP | centrifugal force | E. coli: 0.0134Ā ng/μL Salmonella spp.: 12Ā CFU/mL | ā | 30Ā min | gel electrophoresis | 100 μL | E. coli, Salmonella spp., Staphylococcus aureus, Vibrio parahaemolyticus | Zhang M et al. (2021) |
| LAMP | centrifugal force | 103 copies/μL | 103 ā¼ 106 copies/μL | 30Ā min | fluorescence | ā | HPV | Zhao et al. (2021) |
| LAMP | syringe pump | 101 copies/μL | 101 ā¼ 104 copies/μL | 45Ā min | fluorescence | ā | HBV | Zhang et al. (2018) |
| LAMP | centrifugal force | 100 copies/μL | ā | within 45Ā min | colorimetric method | 20 μL | Porcine epidemic diarrhea virus, transmissible gastroenteritis virus, porcine rotavirus, porcine circovirus type 2 | Wen et al. (2022) |
| LAMP | ā | 50Ā ng/μL | ā | 40Ā min | fluorescence | 14 μL | Carbapenemase-producing organisms | Wu et al. (2022a) |
| LAMP | ā | 102Ā PFU/200 μL | 101 ā¼ 105Ā PFU/mL | 40Ā min | fluorescence | 2 µL | Dengue virus | Yoo et al. (2020) |
| LAMP | capillary force | ā | ā | 30Ā min | colorimetric method | 3 μL | SARS-CoV-2 | |
| LAMP | capillary force | 0.34Ā fg/μL | 10ā1 Ć 10ā4Ā pg/μL | 40Ā min | colorimetric method | 1 μL | Prostate cancer 3 biomarker | Wang et al. (2020a) |
| LAMP | centrifugal force | 102ā¼103 copies | 10 ā¼ 106 copies/μL | 30Ā min | fluorescence | 10 μL | COVID-19 | Soares et al. (2020) |
| LAMP | centrifugal force | 10 copies/mL | 10 ā¼ 107 copies/μL | 40Ā min | fluorescence | ā | SARS-CoV, MERS-CoV, SARS-CoV-2, HCoV-229E, HCoV-OC43, HCoV-NL63, HCoV-HKU1 | Xiong et al. (2020) |
| LAMP | magnetic force | 500 virions/mL | 2.8 Ć 107ā28 copies/mL | 45Ā min | colorimetric method | 100 µL | HCV | Sharma et al. (2022) |
| LAMP | syringe pump | 100 to 10000 copies/μL | 5.5Ć104ā6.3Ć106 copies/mL | within 30Ā min | fluorescence | 6.4 μL | S. equi, S. zoo, EHV1 EHV4, EIV, H3N8 | Sun et al. (2020) |
| LAMP | syringe pump | 102 per chamber | 1.32 Ć 102ā.32 Ć 107 copies | 50Ā min | fluorescence | 1 μL | E. coli, S. Typhimurium, V. parahaemolyticus | Nguyen et al. (2020) |
| LAMP | push pressure | 14Ā CFU/mL | 1.4Ć101ā1.4 Ć 106Ā CFU/mL | 30Ā min | turbidity of the reaction | 250 μL | Viable Salmonella | Wang et al. (2020b) |
| RPA | capillary force | 1 copy/μL | ā | 15Ā min | lateral flow detection | 30 μL | COVID-19 | Liu et al. (2021) |
| RPA | negative pressure | 10 bacterial cells | ā | 30Ā min | fluorescence | 10 μL | E. coli, Listeria monocytogenes, Salmonella enterica | Yin et al. (2020a) |
| RPA | centrifugal force | 1 copy/μL | ā | 20Ā min | fluorescence | 75 μL | SARS-CoV-2 | |
| RPA | centrifugal force | S gene: 0.68Ā fM Orf1ab gene: 4.16Ā fM | ā | 10Ā min | fluorescence | 1 μL | SARS-CoV-2 | |
| RPA | centrifugal force | 1.02 copies/μL | 1.02ā2.04 Ć 103 copies | 20Ā min | fluorescence | 2.5 μL | Norovirus | Qin Z et al. (2021) |
| RPA | ā | 1 copy/reaction | ā | 5ā10Ā min | fluorescence | 5 μL | L. monocytogenes | Luo et al. (2022) |
| RPA | ā | 10 copies | ā | 15Ā min | fluorescence | ā | HPV | Yin et al. (2020c) |
| RPA | syringe pump | ā | ā | 10ā30Ā min | gel electrophoresis | 1 μL | E. coli | |
| HDA | push pressure | 10Ā CFU | 105 ā¼ 101Ā CFU | 30Ā min | fluorescence | 50Ā ng DNA | E. coli | Mahalanabis et al. (2010) |
| HDA | ā | 100Ā CFU | 3ā¼more than 100,000Ā CFU/swab | 60Ā min | chip image | 5 uL | Staphylococcus aureus | |
| HDA | ā | 1Ā CFU | 1ā250Ā CFU/reaction | 45Ā min | colorimetric method | 2 μL | mecA gene | Pasko et al. (2012) |
| HDA | ā | 1.25 Ć 10ā2Ā pg | 125ā2.5 Ć 10ā3Ā pg | 30Ā min | gel electrophoresis | 5Ā mL | Clostridium difficile | Huang et al. (2013) |
| HDA | syringe pump | 10 cells | below 500 cells | 30 min | gel electrophoresis | 100 μL | Foodborne pathogen | Tsougeni et al. (2019) |
Performance of 30 on-chip isothermal amplification systems.
For both amplification methods, the primary obstacle for testing of multiple targets is the addition of different primers to multiple reaction chambers. In some studies, primers have been freeze-dried and preloaded to avoid cross-contamination (; Nguyen et al., 2022). In continuous flow PCR chips, the reaction solution could be separated with an oil phase in a similar way to the method used in digital PCR to test multiple targets (Jia et al., 2018). However, the surface tension needs to be carefully calculated and high-precision flow manipulation is required to keep the droplets in shape.
4 Rapid on-chip detection methods
In this section, we discuss the principle and application of several rapid on-chip methods for the detection of different types of amplicon. The performance of these methods is evaluated according to their speed, detection limit, equipment requirements, operational difficulties, and potential for mass production and high-throughput testing.
4.1 Lateral flow assay
Lateral flow assay (LFA) or lateral flow detection (LFD) has been widely used in the analysis of DNA amplicon; it could be regarded as the simplest paper-based microfluidic detection method. The most common type of lateral flow assay comes in the form of test strips that contain a sample pad, conjugate pad, nitrocellulose filter membrane with two test lines (for the sample and a positive control), and absorbent pad. First, the amplified DNA sample is injected onto the sample pad; the strip is then put into the migration buffer. The buffer flows along the strip, driven by capillary force; the DNA sample conjugates with gold nanoparticles (AuNPs) or other labels to form a sandwich-like structure that is captured by recognition elements, such as antibodies or DNA probes, immobilized on the test lines. Finally, the redundant labels are captured on the control line to create a second red line that validates the result. Recent studies have optimized the detection limit of LFA strips to achieve a very low level. Yin et al. designed lateral flow assay methods for the detection of sheep- and pig-specific PCR products for purposes of meat authentication; the detection limit was reported to be 10Ā fg target DNA in each sample (Yin et al., 2016; Yin R. et al., 2020). The detection limit of LFA strips could be further improved by correlating the result with more sophisticated labels. For instance, used quantum dots in place of AuNPs to enhance the sensitivity and accuracy of HIV-DNA detection; the detection limit reported was 0.76 p.m. Takalkar et al. (2017) used fluorescent carbon nanoparticles as labels on the DNA probe; the detection limit was 0.4Ā fM, 4-6 orders lower than achieved in LFA based on AuNPs. Another form of lateral flow assay makes use of a cotton thread-based format. Cotton thread-based devices have advantages in terms of their small size, low price, and ease of manipulation. The principle of cotton thread-based LFA is similar to that of strip tests. The strip is replaced by a cotton thread, while the nitrocellulose filter membrane is replaced by a layer of wax. developed a cotton thread-based device for rapid detection of nucleic acid; the detection limit was reported to be 2.5Ā nM. Hydrophilic cotton thread modification could simplify the preparation of the sample pad and decrease the speed of lateral flow; such modification has been reported to produce a fourfold enhancement in the sensitivity of this method ().
LFA is one of the most successfully commercialized detection methods, with a well-developed and cost-effective mass production technique and user-friendly protocol. LFA can be adapted for use with most nucleic acid amplification methods, including PCR, LAMP, RPA, and HDA, but it is most commonly used to analyze RPA amplicons. As mentioned above, fluorescent detection of RPA amplicons requires a fluorescence probe and is more expensive than other amplification methods (). The result of LFA can be easily detected by the naked eye, and detection is not time-consuming. The detection times of four examples of LFA detection, listed in Table 4, are 2ā5Ā min, 5Ā min, 10Ā min, and 5Ā min, respectively. For all these reasons, LFA strips have been widely used in qualitative testing (species authentication, clinical diagnosis, etc.) Recent advances in lateral flow detection have overcome its difficulties in multi-target detection and quantitative detection. Ma et al. (2020) used multiple-target lateral flow dipsticks to detect RPA-amplified Staphylococcus aureus, Vibrio parahaemolyticus, and Salmonella Enteritidis; the results were read using a handheld reader, and the detection limits were 2.6 Ć 101Ā CFU/mL, 7.6 Ć 101Ā CFU/mL, and 1.29 Ć 101Ā CFU/mL, for each of these three targets, respectively. The total detection time, from amplification to reading, was 15Ā min.
TABLE 4
| Detection method | LoD | Linear range | Qualitative/Quantitative | End point/Real time | References |
|---|---|---|---|---|---|
| SERS | 102.0Ā pg/L | ā | qualitative | end point | Teixeira et al. (2020) |
| SERS | 3ā4Ā CFU/mL | 1ā108Ā CFU/mL | quantitative | end point | Zhuang et al. (2022) |
| fluorescence | E. coli, P. mirabilis, and S. typhimurium: 1Ā CFU/μL S. aureus: 10Ā CFU/μL | ā | quantitative | real time | Li et al. (2020c) |
| fluorescence | Singleplex assays: 2.5 Ć 101 DNA copies for both khe and blaNDM-1 Duplex assay: 2.5 Ć 101 DNA copies for khe and 2.5 Ć 102 DNA copies for blaNDM-1 | ā | semi-quantitative | real time | |
| fluorescence | Escherichia coli: 17.15Ā ng/μL DNA Staphylococcus aureus: 5.67Ā ng/μL DNA Pseudomonas aeruginosa: 16.47Ā ng/μL DNA | Escherichia coli: 17.15ā137.2Ā ng/μL DNA Staphylococcus aureus: 5.67ā34.02Ā ng/μL DNA Pseudomonas aeruginosa: 16.47ā197.58Ā ng/μL DNA | quantitative | end point | |
| fluorescence | 1 à 101 copies/μL | 1 à 101 to 1 à 104 copies/μL | quantitative | real time | Zhang et al. (2020) |
| fluorescence | 2.8 Ć 10 -5Ā ng/μL | ā | qualitative | end point | Zhou et al. (2022) |
| fluorescence | 5 copies/μL | ā | quantitative | end point | Yin et al. (2022) |
| fluorescence | 89Ā CFU/mL | ā | quantitative | real time | Wu et al. (2022b) |
| fluorescence | 10 DNA copies | 35Ā pg (10 haploid genome copies) to 350Ā ng (105 copies) | quantitative | real time | Khodakov et al. (2021) |
| fluorescence | 1āĆā101 copies/μL | 1āĆā101 to 1āĆā105 copies/μL | quantitative | end point | Meng et al. (2021) |
| fluorescence | 10 bacterial cells | ā | quantitative | end point | Yin et al. (2020a) |
| SPR | 10Ā pg/mL | 10ā5 to 10ā12Ā g/mL | qualitative | real time | Hsieh et al. (2022) |
| SPR | 0.1Ā nM | ā | quantitative | real time | |
| colorimetric | 5 copies/μL | ā | qualitative | end point | |
| colorimetric | ā | ā | qualitative | real time | Uddin et al. (2021) |
| colorimetric | ā | ā | quantitative | end point | Thio et al. (2022) |
| colorimetric | Sau and Sal: 102 copies/μL; Sty, Pae, and Eco: 101copies/μL | ā | qualitative | real time | Liu D et al. (2020) |
| colorimetric | 30Ā CFUĀ mLā1 | 102ā107Ā CFUā mLā1 | semi-quantitative | end point | Yu Q et al. (2021) |
| colorimetric | 100 copies/μL | ā | qualitative | end point | Wen et al. (2022) |
| colorimetric | 102ā103Ā CFUĀ mLā1 | ā | qualitative | end point | Jin et al. (2020) |
| quantum dot | 0.11Ā pmolĀ Lā1 | 0.50Ā pmolĀ Lā1ā50Ā nmolĀ Lā1 | quantitative | end point | Kokkinos et al. (2018) |
| quantum dot | 50Ā pmol | 50ā200Ā pmol | qualitative | end point | Nguyen et al. (2020b) |
| LFA | 1 copy per μL or 30 copies per sample | ā | qualitative | end point | Liu et al. (2021) |
| LFA | 0.88 TCID50/mL | 101ā103 TCID50/mL | qualitative | end point | |
| LFA | 5 Ć 103Ā CFU/mL | 105Ā CFU/mL to 103Ā CFU/mL | qualitative | end point | Lu et al. (2020) |
| LFA | 10ā100 IU/mL | ā | qualitative | real time | |
| electrochemical | 102Ā CFU/mL | 102 ā¼ 106Ā CFU/mL | quantitative | real time | Park et al. (2021) |
| electrochemical | 7.4Ā fM | ā | quantitative | real time | Liu et al. (2022) |
Performance of 30 on-chip nucleic acid detection systems.
4.2 Fluorescent and colorimetric detection
In the domain of nucleic acid detection, fluorescent dye can find applications based on its interaction with double-stranded DNA, or can be incorporated alongside other nanostructures as a fluorescent label. Fluorescent dye can directly emit light distinguishable by the naked eye, thus providing visual confirmation of DNA amplification, but the light signal increases greatly under excitation by UV light. Many sensitive devices, such as a CCD camera, fluorescent microscope, or UV-Vis spectrometer, are used for thorough analysis of fluorescent signals, but these devices often require a stable workplace to prevent interference from external light or oscillation. As a result, there are two approaches used in fluorescent rapid detection methods: qualitative end-point detection with visual confirmation; and quantitative or semi-quantitative real-time detection with portable devices.
Zhou et al. (2022) prepared a multi-chamber chip with a paper-disk and thin PDMS film for simultaneous detection of Escherichia coli O 157:H7, Salmonella spp., and Staphylococcus aureus. A portable device based on a smartphone was fabricated for end-point detection. Two UV lights powered by the smartphone battery were installed in a 3D-printed dark box; parallel installation meant that they provided uniform excitation. The detection signal was collected and analyzed by smartphone. The entire detection set cost only $120 and weighed less than 300Ā g; the detection limit was 2.8 Ć 10ā5Ā ng/μL. Li et al. (2020c) used a high-intensity light-emitting diode (LED) to excite fluorescent dye and a photomultiplier tube to record the signal in real time. Eight reaction chambers carried out a LAMP reaction for four pathogenic bacteria and a negative control; the best LoD obtained for these targets was 1Ā CFU/μL. Fluorescent dye can also be used to detect byproducts of the LAMP reaction instead of binding diretly with amplicons. During the amplification process in LAMP, Mg2+ is consumed and turned into white precipitate. However, observation of this precipitate with the naked eye is not always reliable. In many studies, fluorescent dye (such as calcein or SYBR Green) has been added to create a more conspicuous light green color for easy detection. Zhang M. et al. (2021) embedded calcein-soaked paper in the reaction chamber for direct detection of LAMP products.
Digital nucleic acid detection uses an oil phase to divide the reaction solution into thousands of droplets prior to amplification. Each droplet contains either 0 or 1 original copies of the target DNA. As a result, the fluorescent signal obtained from these droplets can be processed using Poisson statistics to estimate the amount of nucleic acid in the original sample. This method provides rapid and highly accurate quantification of target DNA. Yin et al. (2020b) designed a self-priming digital PCR chip with a four-layer structure and six detection areas; this design enables the pre-introduction of specific reaction mix into certain detection areas and lowers the instrument and reagent requirements for multiplex detection. The use of digital methods is not limited to the domain of PCR; many on-chip detection methods have been reported for digital LAMP or digital RPA. Peng et al. (2020) designed a centrifugal droplet-based chip for application of digital LAMP. Traditional centrifugal chips suffer from droplet coalescence because the droplets are tightly packed. The design by Peng et al. optimized for droplet emulsification, using oil-storage structures to control the thickness of the oil film and prevent coalescence; the reported dynamic range was 101 to 104 copies per μL. However, digital methods require a long time to be spent on droplet formation and also require high-quality fluid control, which will affect the speed and cost of on-site detection. Luo et al. (2022) prepared a hydrogel nanofluidic chip for hydrogel RPA (gRPA) as a replacement for digital RPA. Under this approach, the movement of nucleic acids is restricted inside a hydrogel, and the end-point fluorescent image of the hydrogel can be used for quantification; this method reduces total time for amplification and detection to 5Ā min. A novel ārandom overlapping theoryā was used instead of Poisson statistics to analyze the fluorescent image, and the detection limit was optimized to achieve detection of a single copy.
Despite its convenience, fluorescent dye is reported to slightly inhibit the PCR process and is more likely to bind with GC-rich sequences (Haukur et al., 2007). Moreover, fluorescent dyes are limited by their fluorescent efficiency, short lifetime, and risk of photobleaching. Quantum dots have been selected as an alternative to fluorescent dyes. Nguyen et al. (2020b) formed a quantum dotāDNA conjugate with semiconductor quantum dots and target oligonucleotides; this structure was used in a similar way to Forster resonance energy transfer (FRET) probes to quantify oligonucleic targets.
Fluorescent detection and colorimetric detection methods are capable of detecting multiple DNA targets through the careful design of labels that emit light signals at different wavelengths. Shen et al. (2022) fabricated a digital microfluidic device that detects multiple RNA targets at three different excitation wavelengths. This provides a visually clearer and more convenient method for multipurpose detection, but the high cost of multiple labels and high design complexity are disadvantages of this approach.
4.3 Non-fluorescent optical detection
Optical detection is based on various optical phenomena: emission, absorbance, and scattering of light, surface plasmon resonance, etc. The most intriguing advantage of optical detection methods is their high precision in the implementation of quantitative analysis. On the other hand, optical detection is highly demanding in terms of the need for a stable and lightproof platform to avoid disturbance from the working environment. Although the components required to control the optical behavior of the system and maintain a suitable working environment are rather expensive, some studies have succeeded in applying these methods in rapid on-site detection.
Surface-enhanced Raman spectroscopy (SERS) involves enhanced Raman scattering of molecules absorbed on or near a SERS-active surface; it is widely used in sensitive detection of biomolecules. Zhuang et al. (2022) designed a microfluidic paper-based device (μPAD) for super-sensitive SERS detection of RPA amplicons. Under this approach, SERS probes were first introduced to the μPAD; a Cas12a trans-cleavage system was added after drying. The solution then migrated within the μPAD and reached the detection zone; finally, SERS was used for sensitive detection. The entire process took less than 15 min. Teixeira et al. (2020) developed a microdroplet-LAMP-SERS platform for real-time detection of foodborne pathogens. The various lengths of LAMP products lead to variability in their SERS fingerprints, which makes direct detection very difficult. Multifunctional gold nanoparticles were used as a Raman reporter and Mg2+ chelating agent, enabling indirect detection of insoluble LAMP byproducts.
Surface plasmon resonance (SPR) is a phenomenon caused by the evanescent wave formed when monochromatic light reflects off the surface of a thin metal film (Lu et al., 2021). An SPR sensor measures the refractive index, angle of refraction, or energy of the reflected light to deduce mass change on the thin metal film. Its signal is highly proportional to the mass attached to the metal surface and provides highly accurate quantification of this. Hsieh et al. (2022) combined microfluidic PCR with the use of an SPR sensor to detect DNA sequences. DNA probes were immobilized on an SPR chip to capture PCR-amplified target sequences; the detection limit was as low as 10ā11Ā g/mL. The signal obtained within 36Ā min was equal to the signal obtained after 105Ā min using traditional methods, representing greatly improved sensitivity. designed a label-free multi-target regenerable SPR sensor for DNA detection. Flow cells were first functionalized with dextran-streptavidin; subsequently, biotinylated probes targeting different sequences were immobilized on this layer, and the SPR chip could be regenerated after each use with NaOH and glycine-HCl. Their sensor is capable of being stably regenerated more than 100 times over the course of 20Ā days while maintaining good reproducibility of test results, which means that it could function as an economical long-term detection platform.
4.4 Electrochemical detection
Electrochemical detection of nucleic acid has a variety of advantages, such as rapid response, high selectivity, good portability, a relatively low cost, and a simple system design that comes with easy manipulation. Improvements in MEMS and nanotechnology have also brought about more options and better performance in the electronic components used to build electrochemical sensors and highly integrated devices for high-throughput testing. Under this method, electrodes are used to transduce biotic information into electrical signals, including impedance, resistance, voltage, and current.
Probes that can recognize a specific target are often immobilized on the surface of electrodes to generate or enhance the detection signal. Nucleic acid strands, antibodies, enzymes, and molecular imprint polymers are the most popular choices of probe. Methods of electrode surface functionalization have been well-developed in previous studies; the binding principles include but are not limited to goldāthiol self-assembly, biotināstreptavidin interaction, electrostatic adsorption, and amide bond formation (). These probes are sometimes connected with other labels to further increase the systemās sensitivity and lower the limit of detection. The sequence-specific hybridization process of DNA is the most simple and effective probe design for detection of nucleic acid: under this approach, the complementary sequence of the target is functionalized on the electrode, and once the hybridization process begins, it will alter the electrical signal received by the sensor. fabricated a paper-based strip for electrochemical detection of single- and double-stranded DNA, in which oligonucleotides tagged with methylene blue are immobilized on the surface to enhance charge transfer; the detection limit of this method is 3Ā nM for single-stranded DNA and 7Ā nM for double-stranded DNA. Instead of hybridizing the target sequence on the probe, researchers could form DNA structures without a target sequence, then break these and rebuild a more stable structure with the presence of target DNA in order to generate a signal. Zhang et al. (2018) immobilized two hairpin probes on the electrode surface, then opened the hairpin structures through hybridization of target DNA; subsequently, cyclic amplification of the DNA was triggered for signal amplification while the target DNA was released for the next cycle. This design enabled the simultaneous detection of two target DNA sequences at fM levels of concentration. The detection process could also target the byproducts of nucleic amplification, including electrons transferred during the process, in order to simplify the mechanism. designed a sensing technique that employs a field-effect transistor for transduction to detect foodborne bacterial pathogens after a LAMP process; rather than recognizing the amplified sequences, this method measures the pH change caused by the reaction.
The easiest way to implement on-chip electrochemical detection is to install electrodes at the bottom of the reaction chamber, or to dip the electrodes directly into the reaction solution. Park et al. (2021) designed an integrated circuit for use inside a pumpless microfluidic chip; PCR-amplified product was electrochemically detected via squarewave voltammetry. Screen-printed electrodes (SPEs) enable miniaturization of the electrodes; these are also low in cost and disposable, which is important in building a highly integrated on-chip detection system with highly repeatable results (HernĆ”ndez-RodrĆguez et al., 2020). Another possibility is to fabricate the circuit directly on the chip; this approach requires high-level microfabrication technology. For instance, Horny et al. deposited a Ti/Pt metal layer onto a glass substrate to function as electrodes, then deposited another layer of CNx for surface modification (Horny et al., 2020). This layer was combined with a PDMS layer to form a detection chip for microRNA.
4.5 Summary of on-chip DNA detection
In Tables 4 and Tables 5 we present a comparison of studies of on-chip DNA detection published in recent years and evaluate the general features of the detection systems examined. Novel developments in nucleic acid testing methods have focused on parameters including LoD, sensitivity, and linearity of the signal, among others. LoD is the most important of these parameters: a lower detection limit requires fewer cycles (or less time) for amplification, which improves the speed of detection. Optical detection methods offer the best precision, but they need to compensate for the errors arising from light passing through transparent chip material. Ideally, detection methods should be able to withstand external interference, such as from vibration or light, in order to achieve better on-site performance; optical methods require many precautionary measures in this matter. LFA is the method least affected by the working environment and offers a good LoD for qualitative detection (this can be as low as the pM range), but it is not suitable for quantitative detection due to the inaccuracies introduced by the use of handheld strip readers (Yu J et al., 2021).
TABLE 5
| Method | Characterization of quantitative accuracy | Throughput | Operator skill requirements | Equipment requirements | Cost per test |
|---|---|---|---|---|---|
| LFA | Qualitative/semi-quantitative | Low | Low | Low | Low |
| Fluorescent | Semi-quantitative/quantitative | High | Medium | Medium | High |
| Optical | Excellent | High | High | High | Medium |
| Electrochemical | High | Medium | Medium | Medium | Low |
Overall evaluation of several detection methods.
The introduction of an additional separation or cleavage procedure could be an effective way to improve the outcome of the detection process. For instance, on-chip detection methods sometimes add electrophoretic separation as a form of pretreatment to improve accuracy. Li et al. (2019) combined on-chip electrophoresis with a continuous flow PCR chip to achieve rapid separation and detection of pathogen genes. Due to the short migration distance, the total time required for on-chip electrophoresis is notably short. However, on-chip preparation of the necessary gel introduces extra steps, including gel formation, flushing, bubble removal, and sterilization, that may prolong the nucleic acid testing process. CRISPR-Cas has also been used in many on-chip detection studies. CRISPR-Cas can eliminate non-specific signals from amplification and is considered to be a complementary method for isothermal amplification. used CRISPR-Cas12a-assisted RAA for SARS-CoV-2 detection. They introduced a short-strand reporter with a FAM fluorophore on 5ā² and quencher on 3ā². Every RAA amplicon could activate two CRISPR-Cas complexes; this increased the cleavage rate of the reporter sequence and released more fluorophore to enhance the signal. Similarly, used CRISPR-Cas13a to lower the LoD of RPA amplicons using LFA detection; the results were optimized to the fM level.
For on-site applications, any detection device is preferably highly integrated or simplified to minimize the overall complexity of the system. Electrochemical methods have strong potential to enable the fabrication of a highly integrated device, but research in this domain tends to focus solely on the detection step. The latest trend in electrochemical detection is the use of flexible materials to produce a wearable sensor for more convenient real-time monitoring (Mathew et al., 2021). This type of method is not compatible with on-chip sample preparation and amplification, because the direct fabrication of on-chip microelectrodes requires high-level MEMS techniques, and the implantation of flexible electrodes on a chip would mean the loss of its advantages in small size and flexibility. Handheld spectroscopic devices do offer advantages in terms of good portability, rapid response, and easy manipulation; however, minimization of the light source, probe, and spectrometer also leads to a decrease in precision. Increasingly many research groups have introduced smartphones into their methods of on-site detection, making full use of the camera for collection of data or images, the memory chip and hard drive for data processing, and the import and export interface for data transfer. Not only does this approach excel in terms of device size, it also provides a great advantage in terms of the ability to send data over WiFi or the Internet to achieve large-scale data processing.
5 All-in-one systems and future perspectives
With recent developments in on-chip methods of nucleic acid sample preparation, amplification, and detection, there has been an emergence of āall-in-oneā or āsample-in-answer-outā systems that integrate all three steps into a single microfluidic system for automatic nucleic acid detection (Li et al., 2021). Below, we identify those studies that have aimed to develop methods of increasing the speed and accuracy of detection in this type of system and analyze their advantages and disadvantages; this is followed by a discussion of future trends and challenges.
5.1 Improving speed of detection
For most application scenarios, the priority of rapid on-site detection should first be to provide sufficient sensitivity and LoD to guarantee a reliable result; the next step should be to try to improve the speed and throughput of the system as much as possible. The three steps involved in rapid nucleic acid testing are closely related to one other, suggesting that simply concatenating the fastest method for each individual step is not necessarily the optimal strategy. For instance, certain isothermal amplification methods have not been found to provide much improvement in reaction speed compared to traditional PCR methods; however, due to their high tolerance to inhibitors, these isothermal methods can save a great deal of time when they are combined with a direct amplification approach. Although many on-site applications do not require an extremely low detection limit, improved performance in the detection step could relieve the burden on the amplification step. In particular, if a lower LoD is achieved, fewer thermal cycles or less time will be required in the amplification step.
5.2 System integration
Some recent examples of system integration are presented in Figure 5 for a more vivid illustration of the possibilities. The easiest way to integrate the three steps is to connect separate chips via tubes. Tsougeni et al. (2019) connected three on-chip modules using PEEK capillary tubes; the cell lysis and DNA purification modules were fabricated on the same chip and connected to an amplification chip. This form of integration is easy to design, but a strong driving force is required to push the fluids through a long channel. Moreover, the diameter of the tube has to be very small (the aforementioned PEEK tube had an internal diameter of 150 μm and external diameter of 360 μm) due to the small volume of sample (in the range of 10 μL). A more common strategy is to design a multi-layer chip in which the three steps are completed individually in different layers (Li et al., 2020c; Tian et al., 2022). This design can help to reduce the size of chip via overlapping channels and can exploit gravity to avoid back-flow or cross-contamination.
FIGURE 5
Another issue is the pumping force through a long channel with tight bends. Although a single syringe pump may suffice, this imposes certain limitations in the case of a highly automated and multifunctional chip. Pumping mechanisms can also be combined to facilitate the entire procedure. used an electromagnetic valve-controlled peristaltic pump as well as a syringe pump; the former was used to transfer solution between different reaction chambers, and the latter was used to generate droplets for digital PCR. The drawback of combining two pumps is the large volume of the pumping system. Compared with a pneumatic peristaltic pump, which requires a vacuum pump, the choice of an electromagnetic valve-controlled peristaltic pump simplified the system design in this case. Centrifugal chips suffer from the same problem, since the centrifuge can only propel fluids from the center of the chip toward the edges. Malic et al. (2022) designed a centrifugal chip that was square-shaped rather than disk-like; a pneumatic pump was connected on one side of the chip to assist with fluidic control. Magnetic pumping is another solution to this problem, since it requires only a small piece of magnetic material and direct contact with the chip is not necessary. Furthermore, the magnetic beads involved in magnetic pumping could also help with mechanical lysis or DNA separation in the previous steps, which would make the entire system more effective.
A heating device is necessary for both PCR and isothermal amplification. Many articles pertaining to on-chip PCR provide a simulation of the heat gradient to avoid non-specific amplification (Saeed et al., 2016; Zeng et al., 2016; Li et al., 2020d). Isothermal amplification is not affected by such issues, and this is another of its advantages over on-chip PCR methods.
For DNA detection, optical and fluorescent sensing methods require an extremely lightproof working environment. In most cases, a sealed box is provided to cover the chip or the entire system (Li et al., 2020c). Electrochemical detection involves either installing wires or inserting electrodes during preparation of the chip. Some researchers have constructed a chip around a set of microelectrodes to provide a reaction chamber, but as discussed in the previous section, this approach means losing the advantages of small size and flexibility. Another strategy is to use a printed circuit board (PCB) instead of microelectrodes for electrochemical detection; the size of a PCB matches that of a microfluidic chip, and it is much cheaper compared to the use of screen-printed electrodes (Nagaraj et al., 2014). In addition, although a paper-based chip is not suitable for sample preparation and amplification, an LFA strip could also be pre-planted inside the reaction chamber of a microfluidic chip for detection purposes (Liu et al., 2021). Many works have reported the introduction of smartphones to provide power, lamination, image recording, software control, data analysis, and data transmission. This approach could reduce the use of PCBs, thus reducing system size and allowing more time to be dedicated to hardware testing and development.
5.3 Concerns relating to cost-reduction and mass production
Questions about the cost of microfluidic methods never cease to create obstacles for their development, not only in the domain of chip manufacture but also in terms of the apparatus required for pumping and detection. Studies always claim that reducing the required amount of reagent could lower the cost, but the high cost of these methods mainly arises from the use of microchannels in the µm range and the new mask or mold required for any slight change during development. Methods such as photolithography are also expensive in terms of manufacture and maintenance of the working environment and apparatus. Thanks to recent advances in fabrication technology, an increasing number of researchers are making use of high-precision 3D printing or CNC milling to design and test prototypes (Trinh et al., 2018; Sanchez et al., 2022). Once the channel design is defined, mass production via injection molding (which allows for a finest possible structure of 10ā100 μm) would reduce the price of manufacturing chips or cartridges. This could bring the cost as low as 0.1 to 0.5 dollars per chip. However, this price range may not seem likely to readers, for three reasons. First, this price is based on the authorsā experience in China, where we assume prices are cheaper than those of suppliers in other countries due to fierce competition. Second, it does not account for the materials and labor required for chip assembly, thin film sealing, surface treatment, sterilization, transportation, etc. Last but not least, most companies tend to increase their profit margins on microfluidic chips as a means of recovering the enormous costs incurred at the research stage and in preparation of the injection mold. However, in general, mass production of delicate microfluidic chips has become more feasible in recent years. It is important to point out that, even once a reliable channel design is established, a great deal of work will still be necessary in the die-sinking process (the process of preparing the injection mold), since this method requires a slope in most structures for better lift-out and any imperfection on the surface of the mold may lead to unwanted microfluidic turbulence.
Methods of chip assembly also play an important role in reducing the total cost. Traditional plasma bonding requires a highly skilled operator and assistance from a positioning device to achieve good accuracy in positioning. Furthermore, any dislocation in the assembly process may produce a defective product and is irreversible, which also increases manufacturing costs. In some studies, the chip has been assembled using screws or clips, which is reversible and user-friendly (; Satoh et al., 2018; Lee et al., 2020). Use of a multi-layer design also allows thorough and large-scale sterilization with ethylene oxide prior to assembly, which could further decrease preparation costs. In order to reduce the total cost of the chip, some researchers have installed lateral flow assay strips inside microfluidic chips for detection (Liu et al., 2021).
Based on the content of the above discussion, we believe that future trends in making speed improvements will involve the use of two lysis methods and direct amplification based on isothermal methods. Combined lysis may enable completion of lysis within 1ā2Ā min with an efficiency greater than 90%, and direct amplification with isothermal methods may enable purification to be bypassed, without the process being affected by inhibitors in the solution. Regarding the detection step, the abovementioned methods have all been found to provide rapid (within 2Ā min) or even real-time responses, but further lowering their LoD could save on amplification time and increase the speed of the full procedure.
Statements
Author contributions
CL and HJ organized the structure of the paper and were in charge of modification of the writing. JW was in charge of collecting related papers and writing the initial draft. The other authors contributied in providing insight opinion in nucleic acid testing industry and provided some related papers.
Funding
This work was supported by the National Key R&D Program of China (2021YFC2600503), the Public Welfare Technology Application Research Project of Zhejiang Province (No. LGN22C200013), the Social Public Welfare Technology Application Research Project of Ningbo (No. 2022S014), the Major Science and Technology Project of Xiaoshan District (No. 2021226), the Open Fund of the Key Laboratory of Biosafety Detection for Zhejiang Market Regulation (No.2022BS002), and the Science and Technology Innovation Activity Plan for College Students in Zhejiang Province (Nos. 2022R409A057 and 2022R409A050).
Conflict of interest
LP was employed by Zhejiang Hongzheng Testing Co., Ltd., XG was employed by Zhejiang Gongzheng Testing Center Co., Ltd., PL was employed by Key Laboratory of Biosafety detection for Zhejiang Market Regulation, Zhejiang Fangyuan Testing Group LO.T, and YT was employed by Hangzhou Tiannie Technology Co., Ltd.
The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Publisherās note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
Supplementary material
The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fbioe.2023.1020430/full#supplementary-material
References
1
AhrbergC. D.ManzA.NeuzilP. (2016). Palm-sized device for point-of-care Ebola detection. Anal. Chem.88, 4803ā4807. 10.1021/acs.analchem.6b00278
2
AnL.TangW.RanalliT. A.KimH. J.WytiazJ.KongH. (2005). Characterization of a thermostable UvrD helicase and its participation in helicase-dependent amplification. J. Biol. Chem.280, 28952ā28958. 10.1074/jbc.M503096200
3
AnN.LiK.ZhangY.WenT.LiuW.LiuG.et al (2021). A multiplex and regenerable surface plasmon resonance (MR-SPR) biosensor for DNA detection of genetically modified organisms. Talanta231, 122361. 10.1016/j.talanta.2021.122361
4
BaiX.MaX.LiM.LiX.FanG.ZhangR.et al (2020). Field applicable detection of Hepatitis B virus using internal controlled duplex recombinaseāaided amplification assay and lateral flow dipstick assay. J. Med. Virol.92, 3344ā3353. 10.1002/jmv.25778
5
BaigN.SajidM.SalehT. A. (2018). Recent trends in nanomaterial-modified electrodes for electroanalytical applications. Trac. Trends Anal. Chem.111, 47ā61. 10.1016/j.trac.2018.11.044
6
BarmanU.WiederkehrR. S.FioriniP.LagaeL.JonesB. (2018). A comprehensive methodology for design and development of an integrated microheater for on-chip DNA amplification. J. Micromech. Microeng.28, 085014. 10.1088/1361-6439/aabd2c
7
BattagliaS.PetraliaS.VicarioN.CirilloD.ConociS. (2019). An innovative silicon-chip for sensitive real time PCR improvement in pathogen detection. Anal144, 2353ā2358. 10.1039/c9an00006b
8
BehrmannO.HügleM.EckardtF.BachmannI.HellerC.SchrammM.et al (2020). 3D printed monolithic microreactors for real-time detection of Klebsiella pneumoniae and the resistance gene bla NDM-1 by recombinase polymerase amplification. Micromachines11, 595ā617. 10.3390/MI11060595
9
BhatA. I.AmanR.MahfouzM. (2022). Onsite detection of plant viruses using isothermal amplification assays. Plant Biotechnol. J.20, 1859ā1873. 10.1111/pbi.13871
10
BhiseN. S.ManoharanV.MassaS.TamayolA.GhaderiM.MiscuglioM.et al (2016). A liver-on-a-chip platform with bioprinted hepatic spheroids. Biofabrication8, 014101. 10.1088/1758-5090/8/1/014101
11
BiY.ShuM.ZhongC.LiS. Y.LiY. K.YangH. H.et al (2020). A novel SDS rinse and immunomagnetic beads separation combined with real-time loop-mediated isothermal amplification for rapid and sensitive detection of Salmonella in ready-to-eat duck meat. Food Anal. Methods13 (5), 1166ā1175. 10.1007/s12161-020-01735-1
12
BrassardD.GeisslerM.DescarreauxM.TremblayD.DaoudJ.ClimeL.et al (2019). Extraction of nucleic acids from blood: Unveiling the potential of active pneumatic pumping in centrifugal microfluidics for integration and automation of sample preparation processes. Lab. Chip19, 1941ā1952. 10.1039/c9lc00276f
13
CaoG.HuoD.ChenX.WangX.ZhouS.ZhaoS.et al (2022). Automated, portable, and high-throughput fluorescence analyzer (APHF-analyzer) and lateral flow strip based on CRISPR/Cas13a for sensitive and visual detection of SARS-CoV-2. Talanta Open123594, 123594. 10.1016/j.talanta.2022.123594
14
ChavanS. G.KimD.HwangJ.ChoiY.HongJ. W.KimJ.et al (2019). Enhanced detection of infectious pancreatic necrosis virus via lateral flow chip and fluorometric biosensors based on self-assembled protein nanoprobes. ACS Sens.4, 2937ā2944. 10.1021/acssensors.9b01362
15
ChenF.ZhangD.ZhangQ.ZuoX.FanC.ZhaoY. (2017). Zero-background helicase-dependent amplification and its application to reliable assay of telomerase activity in cancer cell by eliminating primerādimer artifacts. Chembiochem17, 1171ā1176. 10.1002/cbic.201500605
16
ChenJ. J.SheuT. S.WangY. J. (2016). Continuous-flow DNA amplification device employing microheaters. Defect Diffusion Forum366, 17ā30. 10.4028/www.scientific.net/ddf.366.17
17
ChenS.SunY.FanF.ChenS.ZhangY.ZhangY.et al (2022). Present status of microfluidic PCR chip in nucleic acid detection and future perspective. TrAC-Trends Anal. Chem.157, 116737. 10.1016/j.trac.2022.116737
18
ChenY.LiuY.ShiY.PingJ.WuJ. (2020a). Magnetic particles for integrated nucleic acid purification, amplification and detection without pipetting. TrAC-Trends Anal. Chem.127, 115912. 10.1016/j.trac.2020.115912
19
ChenY.MeiY.ZhaoX.JiangX. (2020b). Reagents-loaded, automated assay that integrates recombinase-aided amplification and Cas12a nucleic acid detection for a point-of-care test. Anal. Chem.92, 14846ā14852. 10.1021/acs.analchem.0c03883
20
ChenY.ZongN.YeF.MeiY.QuJ.JiangX. (2022). Dual-CRISPR/Cas12a-assisted RT-RAA for ultrasensitive SARS-CoV-2 detection on automated centrifugal microfluidics. Anal. Chem.94, 9603ā9609. 10.1021/acs.analchem.2c00638
21
ChenZ.ShenL.ZhaoX.ChenH.XiaoY.ZhangY.et al (2022). Acoustofluidic micromixers: From rational design to lab-on-a-chip applications. Appl. Mater Today26, 101356. 10.1016/j.apmt.2021.101356
22
ChinW. H.SunY.HĆøgbergJ.QuyenT. L.EngelsmannP.WolffA.et al (2016). Direct PCR - a rapid method for multiplexed detection of different serotypes of salmonella in enriched pork meat samples. Mol. Cell. Probes32, 24ā32. 10.1016/j.mcp.2016.11.004
23
Chiorcea-PaquimA. M.Oliveira-BrettA. M. (2021). Dna electrochemical biosensors for in situ probing of pharmaceutical drug oxidative dna damage. Sens21, 1125. 10.3390/s21041125
24
ChoiG.PrinceT.MiaoJ.CuiL.GuanW.ChoiG. (2018). Sample-to-answer palm-sized nucleic acid testing device towards low-cost malaria mass screeninget alSample-to-answer palm-sized nucleic acid testing device towards low-cost malaria mass screening. Biosens. Bioelectron.115, 83ā90. 10.1016/j.bios.2018.05.019
25
ChoiY.SongY.YongT. K.KimH.LeeK. G.LeeS. J.et al (2021). All-in-One pumpless portable genetic analysis microsystem for rapid naked-eye detection. Sensors Actuators, B344, 130307. 10.1016/j.snb.2021.130307
26
CintiS.ProiettiE.CasottoF.MosconeD.ArduiniF. (2018). Paper-based strips for the electrochemical detection of single and double stranded DNA. Anal. Chem.90, 13680ā13686. 10.1021/acs.analchem.8b04052
27
ClaudelL.ValeixN.BasmaciyanL.PereiraB.CostaD.VincentA.et al (2021). Comparative study of eleven mechanical pretreatment protocols for Cryptosporidium parvum DNA extraction from stool samples. Microorganisms9, 297. 10.3390/microorganisms9020297
28
CornelissenJ. B. W. J.De GreeffA.HeuvelinkA. E.SwartsM.SmithH. E.Van der WalF. J. (2016). Rapid detection of Streptococcus uberis in raw milk by loop-mediated isothermal amplification. J. Dairy Sci.99, 4270ā4281. 10.3168/jds.2015-10683
29
CrannellZ. A.RohrmanB.Richards-KortumR. (2014). Equipment-free incubation of recombinase polymerase amplification reactions using body heat. PloS one9, e112146. 10.1371/journal.pone.0112146
30
CuiF.ChenW.WuX.GuoZ.LiuW.ZhangW.et al (2017). Design and experiment of a PDMS-based PCR chip with reusable heater of optimized electrode. Microsyst. Technol.23, 3069ā3079. 10.1007/s00542-016-3064-3
31
CzilwikG.MessingerT.StrohmeierO.WadleS.Von StettenF.PaustN.et al (2015). Rapid and fully automated bacterial pathogen detection on a centrifugal-microfluidic LabDisk using highly sensitive nested PCR with integrated sample preparation. Lab. Chip15, 3749ā3759. 10.1039/c5lc00591d
32
DaherR. K.GaleS.MauriceB.BergeronM. G. (2016). Recombinase polymerase amplification for diagnostic applications. Clin. Chem.947, 947ā958. 10.1373/clinchem.2015.245829
33
DanaeifarM. (2022). New horizons in developing cell lysis methods: A review. Biotechnol. Bioeng.119, 3007ā3021. 10.1002/bit.28198
34
DengX.WangC.GaoY.LiJ.WenW.ZhangX.et al (2018). Applying strand displacement amplification to quantum dots-based fluorescent lateral flow assay strips for HIV-DNA detection. Biosens. Bioelectron.105, 211ā217. 10.1016/j.bios.2018.01.039
35
DeraneyR. N.SchneiderL.TripathiA. (2020). Synergistic use of electroosmotic flow and magnetic forces for nucleic acid extraction. Anal145, 2412ā2419. 10.1039/c9an02191d
36
DĆaz-GonzĆ”lezM.de la Escosura-MuƱizA. (2021). Strip modification and alternative architectures for signal amplification in nanoparticle-based lateral flow assays. Anal. Bioanal.Chem.413, 4111ā4117. 10.1007/s00216-021-03421-5
37
DongT.MaX.ShengN.QiX.ChuY.SongQ.et al (2021). Point-of-care DNA testing by automatically and sequentially performing extraction, amplification and identification in a closed-type cassette. Sensors Actuators, B327, 128919. 10.1016/j.snb.2020.128919
38
DoniaA.Furqan ShahidM.HassanS. U.ShahidR.AhmadA.JavedA.et al (2022). Integration of RT-LAMP and microfluidic technology for detection of SARS-CoV-2 in wastewater as an advanced point-of-care platform. Food Environ. Virol.1, 364ā373. 10.1007/s12560-022-09522-3
39
DuT. E.WangY.ZhangY.ZhangT.MaoX. (2015). A novel adenosine-based molecular beacon probe for room temperature nucleic acid rapid detection in cotton thread device. Anal. Chim. Acta861, 69ā73. 10.1016/j.aca.2014.12.044
40
Duarte-GuevaraC.SwaminathanV.ReddyB.HuangJ. C.LiuY. S.BashirR. (2016). On-chip electrical detection of parallel loop-mediated isothermal amplification with DG-BioFETs for the detection of foodborne bacterial pathogens. RSC Adv.6, 103872ā103887. 10.1039/c6ra19685c
41
EspulgarW. V.SaitoM.TakahashiK.UshiroS.YamamotoN.AkedaY.et al (2021). Deskilled and rapid drug-resistant gene detection by centrifugal force-assisted thermal convection PCR device. Sens21, 1225. 10.3390/s21041225
42
FarooqU.LiuX.ZhouW.HassanM.NiuL.MengL. (2021). Cell lysis induced by nanowire collision based on acoustic streaming using surface acoustic waves. Sensors Actuators, B Chem.345, 130335. 10.1016/j.snb.2021.130335
43
FernĆ”ndez-CarballoB. L.McBethC.McGuinessI.KalashnikovM.BaumC.BorrósS.et al (2018). Continuous-flow, microfluidic, qRT-PCR system for RNA virus detection. Anal. Bioanal.Chem.410, 33ā43. 10.1007/s00216-017-0689-8
44
FrechG. C.MunnsD.JenisonR. D.HickeB. J. (2012). Direct detection of nasal Staphylococcus aureus carriage via helicase-dependent isothermal amplification and chip hybridization. BMC Res. Notes.5, 430ā438. 10.1186/1756-0500-5-430
45
FuchsB. B.EatemadpourS.Martel-FoleyJ. M.StottS.TonerM.MylonakisE. (2019). Rapid isolation and concentration of pathogenic fungi using inertial focusing on a chip-based platform. Front. Cell. Infect. Microbiol.9, 27. 10.3389/fcimb.2019.00027
46
GengZ.LiS.ZhuL.ChengZ.JinM.LiuB.et al (2020). Sample-to-Answerā detection of rare ctDNA mutation from 2 mL plasma with a fully integrated DNA extraction and digital droplet PCR microdevice for liquid biopsy. Anal. Chem.92, 7240ā7248. 10.1021/acs.analchem.0c00818
47
Georgoutsou-SpyridonosM.FilippidouM.KaprouG. D.MastellosD. C.ChatzandroulisS.TserepiA. (2021). Isothermal recombinase polymerase amplification (RPA) of E. coli gDNA in commercially fabricated PCB-based microfluidic platforms. Micromachines12, 1387. 10.3390/mi12111387
48
GhalekohnehS. J.ZandM. M.BanadakiM. D.AzadB. K. D.MirAhsaniS. (2020). Novel fabrication method of a microfluidic continuous flow PCR. 10.21203/rs.3.rs-28384/v1
49
GrigorovE.KirovB.MarinovM. B.GalabovV. (2021). Review of microfluidic methods for cellular lysis. Micromachines12, 498. 10.3390/mi12050498
50
GuanJ.WangY.JinJ.ZhengG. (2022). A microfluidic chip for LAMP-based multiplex detection of pathogen. bioRxiv. 10.1007/s00216-021-03224-8
51
GuoJ.ChenS.GuoJ.MaX. (2020). Nanomaterial labels in lateral flow immunoassays for point-of-care-testing. J. mater Sci. Technol.60, 90ā104. 10.1016/j.jmst.2020.06.003
52
GuoJ.ChenS.TianS.LiuK.NiJ.ZhaoM.et al (2021). 5G-enabled ultra-sensitive fluorescence sensor for proactive prognosis of COVID-19. Biosens. Bioelectron.181, 113160. 10.1016/j.bios.2021.113160
53
HaberJ. M.GascoyneP. R. C.SokolovK. (2017). Rapid real-time recirculating PCR using localized surface plasmon resonance (LSPR) and piezo-electric pumping. Lab. Chip17, 2821ā2830. 10.1039/c7lc00211d
54
HanJ. P.SunJ.ZhuangB.LiuP.ZhaoX. C.Wan-ShuiL. I.et al (2015). The electrophoresis on microfluidic chips and its application in forensic science. Prog. Biochem. Biophys.42, 1128ā1135. 10.16476/j.pibb.2015.0184
55
HardingeP.BaxaniD. K.McCloyT.MurrayJ. A. H.CastellO. K. (2020). Bioluminescent detection of isothermal DNA amplification in microfluidic generated droplets and artificial cells. Sci. Rep.10, 21886ā21914. 10.1038/s41598-020-78996-7
56
HatchA. C.RayT.LintecumK.YoungbullC. (2014). Continuous flow real-time PCR device using multi-channel fluorescence excitation and detection. Lab. Chip14, 562ā568. 10.1039/c3lc51236c
57
HaukurG.MartinD.BangD. D.AndersW. (2007). Comparison of multiple DNA dyes for real-time PCR: Effects of dye concentration and sequence composition on DNA amplification and melting temperature. Nucleic Acids Res.35, e127. 10.1093/nar/gkm671
58
HedmanJ.NordgaardA.DufvaC.RasmussonB.AnsellR.RadstrƶmP. (2010). Synergy between DNA polymerases increases polymerase chain reaction inhibitor tolerance in forensic DNA analysis. Anal. Biochem.405, 192ā200. 10.1016/j.ab.2010.06.028
59
HedmanJ.NordgaardA.RasmussonB.AnsellR.RadstrƶmPeter. (2009). Improved forensic DNA analysis through the use of alternative DNA polymerases and statistical modeling of DNA profiles. Biotechniques47, 951ā958. 10.2144/000113246
60
HernĆ”ndez-RodrĆguezJ. F.RojasD.EscarpaA. (2020). Electrochemical sensing directions for next-generation healthcare: Trends, challenges, and frontiers. Anal. Chem.93, 167ā183. 10.1021/acs.analchem.0c04378
61
HomannA. R.NieblingL.ZehnleS.BeutlerM.DelamotteL.RothmundM. C.et al (2021). A microfluidic cartridge for fast and accurate diagnosis of Mycobacterium tuberculosis infections on standard laboratory equipment. Lab. Chip21, 1540ā1548. 10.1039/d1lc00035g
62
HornyM. C.DupuisV.SiaugueJ. M.GambyJ. (2020). Release and detection of microRNA by combining magnetic hyperthermia and electrochemistry modules on a microfluidic chip. Sens21, 185. 10.3390/s21010185
63
HouL.LiD.ZhangN.ZhaoJ.ZhaoY.SunX. (2022). Development of an isothermal recombinase-aided amplification assay for the rapid and visualized detection of Klebsiella pneumoniae. J. Sci. Food Agric.102, 3879ā3886. 10.1002/jsfa.11737
64
HsiehH. Y.ChangR.HuangY. Y.JuanP. H.TaharaH.LeeK. Y.et al (2022). Continuous polymerase chain reaction microfluidics integrated with a gold-capped nanoslit sensing chip for Epstein-Barr virus detection. Biosens. Bioelectron.195, 113672. 10.1016/j.bios.2021.113672
65
HuangS.DoJ.MahalanabisM.FanA.ZhaoL.JepealL.et al (2013). Low cost extraction and isothermal amplification of DNA for infectious diarrhea diagnosis. PLoS One8, e60059. 10.1371/journal.pone.0060059
66
IslamM. M.KoirlaD. (2021). Toward a next-generation diagnostic tool: A review on emerging isothermal nucleic acid amplification techniques for the detection of SARS-CoV-2 and other infectious viruses. Anal. Chim. Acta339338. 10.1016/j.aca.2021.339338
67
JeonH.KimS.LimG. (2018). Electrical force-based continuous cell lysis and sample separation techniques for development of integrated microfluidic cell analysis system: A review. Microelectron. Eng.198, 55ā72. 10.1016/j.mee.2018.06.010
68
JiaY.RenY.HouL.LiuW.JiangT.DengX.et al (2018). Electrically controlled rapid release of actives encapsulated in double-emulsion droplets. Lab. chip18, 1121ā1129. 10.1039/c7lc01387f
69
JiangL.MancusoM.LuZ.AkarG.CesarmanE.EricksonD. (2014). Solar thermal polymerase chain reaction for smartphone-assisted molecular diagnostics. Sci. Rep.4, 4137ā4145. 10.1038/srep04137
70
JiangY.JiangS.WuY.ZhouB.WangK.JiangL.et al (2021). Multiplex and on-site PCR detection of swine diseases based on the microfluidic chip system. BMC Vet. Res.17, 117ā118. 10.1186/s12917-021-02825-w
71
JinZ.DingG.LiG.YangG.HanY.HaoN.et al (2020). Rapid detection of foodborne bacterial pathogens using visual highāthroughput microfluidic chip. J. Chem. Technol. Biot.95, 1460ā1466. 10.1002/jctb.6331
72
KabaA. M.JeonH.ParkA.YiK.BaekS.ParkA.et al (2021). Cavitation-microstreaming-based lysis and DNA extraction using a laser-machined polycarbonate microfluidic chip. Sensors Actuators, B Chem.346, 130511. 10.1016/j.snb.2021.130511
73
KanekoH.KawanaT.FukushimaE.SuzutaniT. (2007). Tolerance of loop-mediated isothermal amplification to a culture medium and biological substances. J. Biochem. Bioph. Methods70, 499ā501. 10.1016/j.jbbm.2006.08.008
74
KerstingS.RauschV.BierF.Nickisch-RosenegkM. V. (2014). Rapid detection of plasmodium falciparum with isothermal recombinase polymerase amplification and lateral flow analysis. Malar. J.13, 99ā9. 10.1186/1475-2875-13-99
75
KhodakovD.LiJ.ZhangJ. X.ZhangD. Y. (2021). Highly multiplexed rapid DNA detection with single-nucleotide specificity via convective PCR in a portable device. Nat. Biomed. Eng.5, 702ā712. 10.1038/s41551-021-00755-4
76
KokkinosC. T.GiokasD. L.EconomouA. S.PetrouP. S.KakabakosS. E. (2018). Paper-based microfluidic device with integrated sputtered electrodes for stripping voltammetric determination of DNA via quantum dot labeling. Anal. Chem.90, 1092ā1097. 10.1021/acs.analchem.7b04274
77
LeeC. Y.LinJ. L.LiaoC. S.HuangF. C.LeeG. B. (2005). āIntegrated microfluidic systems for DNA analysis,ā in IEEE International Conference on Robotics and Biomimetics, 284ā289. 10.1109/ROBIO.2004.1521791
78
LeeK.KangJ. H.KimH. M.AhnJ.LimH.LeeJ.et al (2020). Direct electrophoretic microRNA preparation from clinical samples using nanofilter membrane. Nano Converg.7, 1ā11. 10.1186/s40580-019-0212-3
79
LeonardoS.ToldrĆ A.CampĆ sM. (2021). Biosensors based on isothermal DNA amplification for bacterial detection in food safety and environmental monitoring. Sens21, 602. 10.3390/s21020602
80
LiB.LiY.ManzA.WuW. (2020). Miniaturized continuous-flow digital PCR for clinical-level serum sample based on the 3D microfluidics and CMOS imaging device. Sensors20, 2492. 10.3390/s20092492
81
LiJ.MaB.FangJ.ZhiA.ChenE.XuY.et al (2020). Recombinase polymerase amplification (RPA) combined with lateral flow immunoassay for rapid detection of salmonella in food. Foods9, 27. 10.3390/foods9010027
82
LiN.LuY.ChengJ.XuY. (2020). A self-contained and fully integrated fluidic cassette system for multiplex nucleic acid detection of bacteriuria. Lab. Chip20, 384ā393. 10.1039/c9lc00994a
83
LiZ.BaiY.YouM.HuJ.YaoC.CaoL.et al (2021). Fully integrated microfluidic devices for qualitative, quantitative and digital nucleic acids testing at point of care. Biosens. Bioelectron.177, 112952. 10.1016/j.bios.2020.112952
84
LiZ.JuR.SekineS.ZhangD.ZhuangS.YamaguchiY. (2019). All-in-one microfluidic device for on-site diagnosis of pathogens based on an integrated continuous flow PCR and electrophoresis biochip. Lab. Chip19, 2663ā2668. 10.1039/c9lc00305c
85
LiZ.LiY.SekineS.XiH.AmanoA.ZhangD.et al (2020). Design and fabrication of portable continuous flow PCR microfluidic chip for DNA replication. Biomed. Microdevices22, 5ā7. 10.1007/s10544-019-0457-y
86
LiuB.JiangF.TaoY.ZhengT.YueY.TepperV.et al (2022). Measuring airborne antibiotic resistance genes in Swiss cities via a DNA-enabled electrochemical chip-based sensor. ACS Est. Eng.2, 1677ā1683. 10.1021/acsestengg.2c00063
87
LiuD.ShenH.ZhangY.ShenD.ZhuM.SongY.et al (2021). A microfluidic-integrated lateral flow recombinase polymerase amplification (MI-IF-RPA) assay for rapid COVID-19 detection. Lab. Chip21, 2019ā2026. 10.1039/d0lc01222j
88
LiuD.ZhuY.LiN.LuY.ChengJ.XuY. (2020). A portable microfluidic analyzer for integrated bacterial detection using visible loop-mediated amplification. Sensors Actuators, B310, 127834. 10.1016/j.snb.2020.127834
89
LiuW.ZhangM.LiuX.SharmaA.DingX. (2017). A Point-of-Need infrared mediated PCR platform with compatible lateral flow strip for HPV detection. Biosens. Bioelectron.96, 213ā219. 10.1016/j.bios.2017.04.047
90
LiuX.LiJ.ZhangL.HuangX.FarooqU.PangN.et al (2020). Cell lysis based on an oscillating microbubble array. Micromachines11, 288. 10.3390/mi11030288
91
LobatoI. M.O'SullivanC. K. (2018). Recombinase polymerase amplification: Basics, applications and recent advances. Trac. Trends Anal. Chem.98, 19ā35. 10.1016/j.trac.2017.10.015
92
LokK. S.KwokY. C.NguyenN. T. (2012a2012). Sample loading and retrieval by centrifugation in a closed-loop PCR microchip. Microchim. Acta176, 445ā453. 10.1007/s00604-011-0741-x
93
LokK. S.LeeP. P. F.KwokY. C.NguyenN. T. (2012b). Nested PCR in magnetically actuated circular closed-loop PCR microchip system. Microchim. Acta177, 111ā117. 10.1007/s00604-012-0760-2
94
LuC.Saint-PierreC.GasparuttoD.RoupiozY.RaveletC.PeyrinE.et al (2021). Melting curve analysis of aptachains: Adenosine detection with internal calibration. Biosensors11, 112. 10.3390/bios11040112
95
LuH. W.SakamuriR.KumarP.FergusonT. M.DoeblerR. W.HerringtonK. D.et al (2020). Integrated nucleic acid testing system to enable TB diagnosis in peripheral settings. Lab. Chip20, 4071ā4081. 10.1039/d0lc00445f
96
LuoZ.LuY.SunY.WangY.SuB.LinX. (2022). Digital recombinase polymerase amplification in hydrogel nanofluidic chip for ultrafast and precise quantification of pathogens in fresh food. Sens. Actuators, B132051, 132051. 10.1016/j.snb.2022.132051
97
MaB.LiJ.ChenK.YuX.SunC.ZhangM. (2020). Multiplex recombinase polymerase amplification assay for the simultaneous detection of three foodborne pathogens in seafood. Foods9, 278. 10.3390/foods9030278
98
MadadelahiM.GhazimirsaeedE.ShamlooA. (2019). Design and fabrication of a two-phase diamond nanoparticle aided fast PCR device. Anal. Chim. Acta1068, 28ā40. 10.1016/j.aca.2019.02.050
99
MahalanabisM.DoJ.AlmuayadH.ZhangJ. Y.KlapperichC. M. (2010). An integrated disposable device for DNA extraction and helicase dependent amplification. Biomed. Microdevices12, 353ā359. 10.1007/s10544-009-9391-8
100
MaitiB.AnupamaK. P.RaiP.KarunasagarI.KarunasagarI. (2022). Isothermal amplificationābased assays for rapid and sensitive detection of severe acute respiratory syndrome coronavirus 2: Opportunities and recent developments. Rev. Med. Virol.32, e2274. 10.1002/rmv.2274
101
MalicL.BrassardD.Da FonteD.NassifC.MounierM.PontonA.et al (2022). Automated sample-to-answer centrifugal microfluidic system for rapid molecular diagnostics of SARS-CoV-2. Lab. Chip22, 3157ā3171. 10.1039/d2lc00242f
102
MathewM.RadhakrishnanS.VaidyanathanA.ChakrabortyB.RoutC. S. (2021). Flexible and wearable electrochemical biosensors based on two-dimensional materials: Recent developments. Anal. Bioanal. Chem.413, 727ā762. 10.1007/s00216-020-03002-y
103
MatsubaraY.KermanK.KobayashiM.YamamuraS.MoritaY.TakamuraY.et al (2004). On-chip nanoliter-volume multiplex TaqMan polymerase chain reaction from a single copy based on counting fluorescence released microchambers. Anal. Chem.76, 6434ā6439. 10.1021/ac0497149
104
MatthewH.MattR.DanielW.JodyP.AmyI.ForrestM. S.et al (2018). PrimedRPA: Primer design for recombinase polymerase amplification assays. Bioinf35, 682ā684. 10.1093/bioinformatics/bty701
105
MengX.YuY.GongP.JinG. (2021). An integrated droplet digital PCR gene chip for absolute quantification of nucleic acid. Microfluid. Nanofluid.25, 62ā69. 10.1007/s10404-021-02465-4
106
MisirG. B. (2018). Molecular technologies and applications in seafood safety. Genet. Aquat. Org.2, 61ā67. 10.4194/2459-1831-v2_2_01
107
MoschouD.VourdasN.KokkorisG.PapadakisG.PartheniosJ.ChatzandroulisS.et al (2014). All-plastic, low-power, disposable, continuous-flow PCR chip with integrated microheaters for rapid DNA amplification. Sensors Actuators, B199, 470ā478. 10.1016/j.snb.2014.04.007
108
MudduR.HassanY. A.UgazV. M. (2011). Rapid PCR thermocycling using microscale thermal convection. J. Vis. Exp.49, 2366ā2375. 10.3791/2366
109
NagarajV. J.JacobsM.VattipalliK. M.AnnamV. P.PrasadS. (2014). Nanochannel-based electrochemical sensor for the detection of pharmaceutical contaminants in water. Environ. Sci. Process. Impacts16, 135ā140. 10.1039/c3em00406f
110
NajjarD.RainbowJ.Sharma TimilsinaS.JollyP.De PuigH.YafiaM.et al (2022). A lab-on-a-chip for the concurrent electrochemical detection of SARS-CoV-2 RNA and anti-SARS-CoV-2 antibodies in saliva and plasma. Nat. Biomed. Eng.6, 968ā978. 10.1038/s41551-022-00919-w
111
NatsuharaD.TakishitaK.TanakaK.KageA.SuzukiR.MizukamiY.et al (2020). A microfluidic diagnostic device capable of autonomous sample mixing and dispensing for the simultaneous genetic detection of multiple plant viruses. Micromachines-Basel11, 540. 10.3390/mi11060540
112
NguyenH. Q.BuiH. K.PhanV. M.SeoT. S. (2022). An internet of things-based point-of-care device for direct reverse-transcription-loop mediated isothermal amplification to identify SARS-CoV-2. Biosens. Bioelectron.195, 113655. 10.1016/j.bios.2021.113655
113
NguyenH. Q.NguyenV. D.Van NguyenH.SeoT. S. (2020). Quantification of colorimetric isothermal amplification on the smartphone and its open-source app for point-of-care pathogen detection. Sci. Rep.10, 15123ā15210. 10.1038/s41598-020-72095-3
114
NguyenT. H.SedighiA.KrullU. J.RenC. L. (2020). Multifunctional droplet microfluidic platform for rapid immobilization of oligonucleotides on semiconductor quantum dots. ACS Sens.5 (3), 746ā753. 10.1021/acssensors.9b02145
115
NittalaK. P. V.HohreiterA.LinhardE. R.DohnR.MishraS.KondaA.et al (2022). Integration of silicon chip microstructures for in-line microbial cell lysis in soft microfluidics. 10.1101/2022.10.03.510547
116
NixonG. J.SvenstrupH. F.DonaldC. E.CarderC.StephensonJ. M.Morris-JonesS.et al (2014). A novel approach for evaluating the performance of real time quantitative loop-mediated isothermal amplification-based methods. Biomol. Detect. Quantif.2, 4ā10. 10.1016/j.bdq.2014.11.001
117
ObandeG. A.SinghK. K. B. (2020). <p>Current and future perspectives on isothermal nucleic acid amplification technologies for diagnosing infections</p>. Infect. Drug Resist13, 455ā483. 10.2147/IDR.S217571
118
PandianK.Ajanth PraveenM.HoqueS. Z.SudeepthiA.SenA. K. (2020). Continuous electrical lysis of cancer cells in a microfluidic device with passivated interdigitated electrodes. Biomicrofluidics14, 064101. 10.1063/5.0026046
119
ParkY. M.ParkJ.LimS. Y.KwonY.BaeN. H.ParkJ. K.et al (2021). Integrated pumpless microfluidic chip for the detection of foodborne pathogens by polymerase chain reaction and electrochemical analysis. Sensors Actuators, B329, 129130. 10.1016/j.snb.2020.129130
120
PaskoC.HickeB.DunnJ.JaeckelH.NieuwlandtD.WeedD.et al (2012). Staph ID/R: A rapid method for determining Staphylococcus species identity and detecting the mecA gene directly from positive blood culture. J. Clin. Microbiol.50, 810ā817. 10.1128/JCM.05534-11
121
PengH.ZhuM.GaoZ.LiaoC.JiaC.WangH.et al (2020). A centrifugal microfluidic emulsifier integrated with oil storage structures for robust digital LAMP. Biomed. Microdevices.22, 18ā10. 10.1007/s10544-020-0475-9
122
PetrouL.LadameS. (2022). On-chip miRNA extraction platforms: Recent technological advances and implications for next generation point-of-care nucleic acid tests. Lab. Chip22, 463ā475. 10.1039/d1lc00868d
123
PiepenburgO.WilliamsC. H.StempleD. L.ArmesN. A. (2006). DNA detection using recombination proteins. Plos Biol.4, e204āe1121. 10.1371/journal.pbio.0040204
124
PumfordE. A.LuJ.SpaczaiI.PrasetyoM. E.ZhengE. M.ZhangH.et al (2020). Developments in integrating nucleic acid isothermal amplification and detection systems for point-of-care diagnostics. Biosens. Bioelectron.170, 112674. 10.1016/j.bios.2020.112674
125
QianJ.ZhangQ.LiuM.WangY.LuM. (2022). A portable system for isothermal amplification and detection of exosomal microRNAs. Biosens. Bioelectron.196, 113707. 10.1016/j.bios.2021.113707
126
QinY.DuanX.PengY.RuiY. (2021). Rapid detection of a novel B1-β-lactamase gene, blaAFM-1 using a loop-mediated isothermal amplification (LAMP) assay. Ann. Clin. Microb. Anti20, 80ā87. 10.1186/s12941-021-00486-z
127
QinZ.XiangX.XueL.CaiW.GaoJ.YangJ.et al (2021). Development of a novel RAA-based microfluidic chip for absolute quantitative detection of human norovirus. Microchem. J.164, 106050. 10.1016/j.microc.2021.106050
128
RagsdaleV.LiH.SantH.AmeelT.GaleB. K. (2016). A disposable, continuous-flow polymerase chain reaction device: Design, fabrication and evaluation. Biomed. Microdevices18, 62ā69. 10.1007/s10544-016-0091-x
129
RathoreH.BiyaniR.KatoH.TakamuraY.BiyaniM. (2019). Palm-size and one-inch gel electrophoretic device for reliable and field-applicable analysis of recombinase polymerase amplification. Anal. Methods11 (39), 4969ā4976. 10.1039/c9ay01476d
130
ReyesJ. B.KuimovaM. K.VilarR. (2021). Metal complexes as optical probes for DNA sensing and imaging. Curr. Opin. Chem. Biol.61, 179ā190. 10.1016/j.cbpa.2021.02.007
131
RizzoM. G.CarnazzaS.De PlanoL. M.FrancoD.NicolòM. S.ZammutoV.et al (2021). Rapid detection of bacterial pathogens in blood through engineered phages-beads and integrated Real-Time PCR into MicroChip. Sensors Actuators, B Chem.329, 129227. 10.1016/j.snb.2020.129227
132
RyzhkovV. V.ZverevA. V.EcheistovV. V.AndronicM.RyzhikovI. A.BudashovI. A.et al (2020). Cyclic on-chip bacteria separation and preconcentration. Sci. Rep.10, 21107ā21112. 10.1038/s41598-020-78298-y
133
SaeedM. A.BazazS. A.AhmedN.NaseemM.RehmanA. (2016). āDesign and simulation of microchannel for DNA amplification,ā in IEEE. In 2016 19th International Multi-Topic Conference (INMIC), 1ā5. 10.1109/INMIC.2016.7840145
134
SanchezD.HawkinsG.HinnenH. S.DayA.WoolleyA. T.NordinG. P.et al (2022). 3D printing-enabled uniform temperature distributions in microfluidic devices. Lab. Chip22, 4393ā4408. 10.1039/d2lc00612j
135
SatohT.SugiuraS.ShinK.Onuki-NagasakiR.IshidaS.KikuchiK.et al (2018). A multi-throughput multi-organ-on-a-chip system on a plate formatted pneumatic pressure-driven medium circulation platform. Lab. Chip18, 115ā125. 10.1039/c7lc00952f
136
SatyaP.MitraS.RayD. P.MahapatraB. S.KaranM.JanaS.et al (2013). Rapid and inexpensive NaOH based direct PCR for amplification of nuclear and organelle DNA from ramie (Boehmeria nivea), a bast fibre crop containing complex polysaccharides. Ind. Crops Prod.50, 532ā536. 10.1016/j.indcrop.2013.07.049
137
SedighiA.ObercC.WhitehallV.LiP. C. H. (2017). NanoHDA: A nanoparticle-assisted isothermal amplification technique for genotyping assays. Nano Res.010, 12ā21. 10.1007/s12274-016-1262-z
138
ShamlooA.Hassani-GangarajM. (2020). Investigating the effect of reagent parameters on the efficiency of cell lysis within droplets. Phys. Fluids32, 062002. 10.1063/5.0009840
139
ShangY.SunJ.YeY.ZhangJ.ZhangY.SunX. (2020). Loop-mediated isothermal amplification-based microfluidic chip for pathogen detection. Crit. Rev. Food Sci. Nutr.60, 201ā224. 10.1080/10408398.2018.1518897
140
SharmaB.SharmaA. (2022). Microfluidics: Recent advances toward lab-on-chip applications in bioanalysis. Adv. Eng. Mater24, 2100738. 10.1002/adem.202100738
141
SharmaS.ThomasE.CaputiM.AsgharW. (2022). RT-LAMP-Based molecular diagnostic set-up for rapid hepatitis C virus testing. Biosens12, 298. 10.3390/bios12050298
142
ShenC.YinH.TongZ.QiuS.LuY.WuZ.et al (2022). āDigital microfluidic chip based on direct ink writing for nucleic acid multiplex pcr detection,ā in IEEE symposium on mass storage systems and technologies (IEEE), 365ā368. 10.1109/MEMS51670.2022.9699738
143
ShenF.DavydovaE. K.DuW.KreutzJ. E.PiepenburgO.IsmagilovR. F. (2011). Digital isothermal quantification of nucleic acids via simultaneous chemical initiation of recombinase polymerase amplification reactions on SlipChip. Anal. Chem.83, 3533ā3540. 10.1021/ac200247e
144
SheuS. C.SongY. S.ChenJ. J. (2022). A portable continuous-flow polymerase chain reaction chip device integrated with arduino boards for detecting colla corii asini. Micromachines13, 1289. 10.3390/mi13081289
145
ShinD. J.TrickA. Y.HsiehY. H.ThomasD. L.WangT. H. (2018). Sample-to-answer droplet magnetofluidic platform for point-of-care hepatitis C viral load quantitation. Sci. Rep.8, 9793ā9812. 10.1038/s41598-018-28124-3
146
ShuB.ZhangC.XingD. (2014). Segmented continuous-flow multiplex polymerase chain reaction microfluidics for high-throughput and rapid foodborne pathogen detection. Anal. Chim. Acta826, 51ā60. 10.1016/j.aca.2014.04.017(
147
SnodgrassR.GardnerA.JiangL.FuC.CesarmanE.EricksonD. (2016). KS-Detectāvalidation of solar thermal PCR for the diagnosis of Kaposiās sarcoma using pseudo-biopsy samples. PLoS One11, e0147636. 10.1371/journal.pone.0147636
148
SoaresR. R.AkhtarA. S.PintoI. F.LapinsN.BarrettD.SandhG.et al (2020). Point-of-care detection of SARS-CoV-2 in nasopharyngeal swab samples using an integrated smartphone-based centrifugal microfluidic platform. medRxiv. 10.1101/2020.11.04.20225888
149
SoaresR. R.AkhtarA. S.PintoI. F.LapinsN.BarrettD.SandhG.et al (2021). Sample-to-answer COVID-19 nucleic acid testing using a low-cost centrifugal microfluidic platform with bead-based signal enhancement and smartphone read-out. Lab. Chip21, 2932ā2944. 10.1039/d1lc00266j
150
SugumarD.KongL. X.IsmailA.RavichandranM.Su YinL. (2012). Rapid multi sample DNA amplification using rotary-linear polymerase chain reaction device (PCRDisc). Biomicrofluidics6, 014119. 10.1063/1.3690469
151
SunF.GanguliA.NguyenJ.BrisbinR.ShanmugamK.HirschbergD. L.et al (2020). Smartphone-based multiplex 30-minute nucleic acid test of live virus from nasal swab extract. Lab. Chip20, 1621ā1627. 10.1039/D0LC00304B
152
SunL.SiddiqueK.WangL.LiS. (2021). Mixing characteristics of a bubble mixing microfluidic chip for genomic DNA extraction based on magnetophoresis: CFD simulation and experiment. Electrophoresis42, 2365ā2374. 10.1002/elps.202000295
153
TachibanaH.SaitoM.TsujiK.YamanakaK.TamiyaE. (2015). Self-propelled continuous-flow PCR in capillary-driven microfluidic device: Microfluidic behavior and DNA amplification. Sens. Actuators, B206, 303ā310. 10.1016/j.snb.2014.09.004
154
TakalkarS.BaryehK.LiuG. (2017). Fluorescent carbon nanoparticle-based lateral flow biosensor for ultrasensitive detection of DNA. Biosens. Bioelectron.98, 147ā154. 10.1016/j.bios.2017.06.045
155
TeixeiraA.ParisJ. L.RoumaniF.DiĆ©guezL.PradoM.EspiƱaB.et al (2020). Multifuntional gold nanoparticles for the SERS detection of pathogens combined with a LAMPāināmicrodroplets approach. Mater13, 1934. 10.3390/ma13081934
156
ThioS. K.BaeS. W.ParkS. Y. (2022). Lab on a smartphone (los): A smartphone-integrated, plasmonic-enhanced optoelectrowetting (oew) platform for on-chip water quality monitoring through LAMP assays. Sensors Actuators, B358, 131543. 10.1016/j.snb.2022.131543
157
ThomasS.OrozcoR. L.AmeelT. (2017). Microscale thermal gradient continuous-flow PCR: A guide to operation. Sens. Actuators, B247, 889ā895. 10.1016/j.snb.2017.03.005
158
TianF.LiuC.DengJ.HanZ.ZhangL.ChenQ.et al (2020). A fully automated centrifugal microfluidic system for sample-to-answer viral nucleic acid testing. Sci. China Chem.63, 1498ā1506. 10.1007/s11426-020-9800-6
159
TianY.ZhangT.GuoJ.LuH.YaoY.ChenX.et al (2022). A LAMP-based microfluidic module for rapid detection of pathogen in cryptococcal meningitis. Talanta Open236, 122827. 10.1016/j.talanta.2021.122827
160
TrinhK.ChonN. M.LeeN. Y. (2018). Fabrication of an integrated polystyrene microdevice for pre-concentration and amplification of Escherichia coli O157: H7 from raw milk. Anal. Methods10, 5071ā5077. 10.1039/C8AY01707G
161
TrinhK. T. L.LeeN. Y. (2018). Glass-polytetrafluoroethylene-glass based sandwich microdevice for continuous-flow polymerase chain reaction and its application for fast identification of foodborne pathogens. Talanta176, 544ā550. 10.1016/j.talanta.2017.07.085
162
TrinhK. T. L.ZhangH.KangD. J.KahngS. H.TallB. D.LeeN. Y. (2016). Fabrication of polymerase chain reaction plastic lab-on-a-chip device for rapid molecular diagnoses. Int. Neurourol. J.20, 38ā48. 10.5213/inj.1632602.301
163
TrinhK.WuW.LeeN. Y. (2017). Fabrication of a 3D Teflon microdevice for energy free homogeneous liquid flow inside a long microchannel and its application to continuous-flow PCR. RSC Adv.7, 10624ā10630. 10.1039/c6ra28765d
164
TrungN. B.SaitoM.TakabayashiH.VietP. H.TamiyaE.TakamuraY. (2010). Multi-chamber PCR chip with simple liquid introduction utilizing the gas permeability of polydimethylsiloxane. Sens. Actuators, B149, 284ā290. 10.1016/j.snb.2010.06.013
165
TsougeniK.KastaniaA. S.KaprouG. D.EckM.JobstG.PetrouP. S.et al (2019). A modular integrated lab-on-a-chip platform for fast and highly efficient sample preparation for foodborne pathogen screening. Sens. Actuators, B288, 171ā179. 10.1016/j.snb.2019.02.070
166
TsugunoriN.HirotoO.HarumiM.ToshihiroY.KeikoW.NobuyukiA.et al (2000). Loop-mediated isothermal amplification of DNA. Nucleic Acids Res.28, e63. 10.1093/nar/28.12.e63
167
TzivelekisC.SelbyM. P.BatetA.MadadiH.DalgarnoK. (2021). Microfluidic chip fabrication and performance analysis of 3D printed material for use in microfluidic nucleic acid amplification applications. J. Micromech. Microeng.31, 035005. 10.1088/1361-6439/abd9a9
168
UddinS. M.SayadA.ChanJ.HuynhD. H.SkafidasE.KwanP. (2021). Heater integrated lab-on-a-chip device for rapid HLA alleles amplification towards prevention of drug hypersensitivity. Sens21, 3413. 10.3390/s21103413
169
Van NguyenH.SeoT. S. (2021). High-throughput human DNA purification on a centrifugal microfluidic device for rapid forensic sex-typing. Biosens. Bioelectron.181, 113161. 10.1016/j.bios.2021.113161
170
VincentM.XuY.KongH. (2004). Helicase-dependent isothermal DNA amplification. EMBO Rep.5, 795ā800. 10.1038/sj.embor.7400200
171
WangK.WuD.WuW. (2019). A new self-activated micropumping mechanism capable of continuous-flow and real-time PCR amplification inside 3D spiral microreactor. Micromachines10, 685. 10.3390/mi10100685
172
WangL.WangR.YuY.ZhangF.WangX.YingY.et al (2016). A powerless on-the-spot detection protocol for transgenic crops within 30min, from leaf sampling up to results. Anal. Bioanal. Chem.408, 657ā662. 10.1007/s00216-015-9128-x
173
WangL. X.FuJ. J.ZhouY.ChenG.FangC.LuZ. S.et al (2020a). On-chip RT-LAMP and colorimetric detection of the prostate cancer 3 biomarker with an integrated thermal and imaging box. Talanta Open208, 120407. 10.1016/j.talanta.2019.120407
174
WangR.WuJ.HeX.ZhouP.ShenZ. (2021). A sample-in-answer-out microfluidic system for the molecular diagnostics of 24 HPV genotypes using palm-sized cartridge. Micromachines12, 263. 10.3390/mi12030263
175
WangS.LiuN.ZhengL.CaiG.LinJ. (2020a). A lab-on-chip device for the sample-in-result-out detection of viable Salmonella using loop-mediated isothermal amplification and real-time turbidity monitoring. Lab. Chip20, 2296ā2305. 10.1039/D0LC00290A
176
WangS.ZhuY.YangY.LiJ.HoffmannM. R. (2020b). Electrochemical cell lysis of gram-positive and gram-negative bacteria: DNA extraction from environmental water samples. Electrochim. Acta338, 135864. 10.1016/j.electacta.2020.135864
177
Wang YY.QiW.WangL.LinJ.LiuY. (2021). Magnetic bead chain-based continuous-flow DNA extraction for microfluidic PCR detection of Salmonella. Micromachines12, 384. 10.3390/mi12040384
178
WangZ.HuangP. H.ChenC.BachmanH.ZhaoS.YangS.et al (2019). Cell lysis via acoustically oscillating sharp edges. Lab. Chip19, 4021ā4032. 10.1039/c9lc00498j
179
WenS.ZhangJ.ZhaoR.GaoJ.WangN.LuT.et al (2022). Development of a handheld microfluidic chip for on-site multiplex detection of four porcine diarrhea-related virus. ACS Agric. Sci. Technol.2, 805ā812. 10.1021/acsagscitech.2c00105
180
WhulanzaY.AdityaR.ArvialidoR.UtomoM. S.BachtiarB. M. (2017). Ease fabrication of PCR modular chip for portable DNA detection kit. AIP Conf. Proc.1817, 040006. 10.1063/1.4976791
181
WuB.TongX.ChenB.YuanW.FuM.YangX.et al (2022a). Development of microfluidic chip-based loop-mediated isothermal amplification (LAMP) method for detection of carbapenemase producing bacteria. Microbiol. Spectr.22, e0032222. 10.1128/spectrum.00322-22
182
WuD.WuW. (2019). Battery powered portable thermal cycler for continuous-flow polymerase chain reaction diagnosis by single thermostatic thermoelectric cooler and open-loop controller. Sens19, 1609. 10.3390/s19071609
183
WuJ.KodziusR.CaoW.WenW. (2014). Extraction, amplification and detection of DNA in microfluidic chip-based assays. Microchim. Acta181, 1611ā1631. 10.1007/s00604-013-1140-2
184
WuS.DuanH.ZhangY.WangS.ZhengL.CaiG.et al (2022b). A Salmonella microfluidic chip combining non-contact eddy heater and 3D fan-shaped mixer with recombinase aided amplification. Biosens12, 726. 10.3390/bios12090726
185
WuY.FuA.YossifonG. (2020). Active particles as mobile microelectrodes for selective bacteria electroporation and transport. Sci. Adv.6, 4412. 10.1126/sciadv.aay4412
186
XiaX.YuY.WeidmannM.PanY.WangY. (2014). Rapid detection of shrimp white spot syndrome virus by real time, isothermal recombinase polymerase amplification assay. Plos One9, e104667. 10.1371/journal.pone.0104667
187
XiongH.YeX.LiY.WangL.ZhangJ.FangX.et al (2020). Rapid differential diagnosis of seven human respiratory coronaviruses based on centrifugal microfluidic nucleic acid assay. Anal. Chem.92, 14297ā14302. 10.1021/acs.analchem.0c03364
188
XuY.ZhangZ.SuZ.ZhouX.HanX.LiuQ. (2020). Continuous microfluidic purification of DNA using magnetophoresis. Micromachines11, 187. 10.3390/mi11020187
189
YangB.WangP.LiZ.TaoC.YouQ.SekineS.et al (2022). A continuous flow PCR array microfluidic chip applied for simultaneous amplification of target genes of periodontal pathogens. Lab. Chip22, 733ā737. 10.1039/d1lc00814e
190
YinJ.XiaL.ZouZ.ZhuangJ.MuY. (2022). A direct and multiplex digital PCR chip for EGFR mutation. Talanta250, 123725. 10.1016/j.talanta.2022.123725
191
YinJ.ZouZ.HuZ.ZhangS.ZhangF.WangB.et al (2020a). A āsample-in-multiplex-digital-answer-outā chip for fast detection of pathogens. Lab. Chip20, 979ā986. 10.1039/c9lc01143a
192
YinJ.ZouZ.YinF.LiangH.HuZ.FangW.et al (2020b). A self-priming digital polymerase chain reaction chip for multiplex genetic analysis. ACS Nano14, 10385ā10393. 10.1021/acsnano.0c04177
193
YinK.DingX.LiZ.ZhaoH.CooperK.LiuC. (2020c). Dynamic aqueous multiphase reaction system for one-pot CRISPR-Cas12a-based ultrasensitive and quantitative molecular diagnosis. Anal. Chem.92, 8561ā8568. 10.1021/acs.analchem.0c01459
194
YinR.SunY.WangK.FengN.XiaoM. (2020). Development of a PCR-based lateral flow strip assay for the simple, rapid, and accurate detection of pork in meat and meat products. Food Chem.318, 126541. 10.1016/j.foodchem.2020.126541
195
YinR.SunY.YuS.WangY.ZhangM.XuY.et al (2016). A validated strip-based lateral flow assay for the confirmation of sheep-specific PCR products for the authentication of meat. Food control.60, 146ā150. 10.1016/j.foodcont.2015.07.030
196
YooH. J.BaekC.LeeM. H.MinJ. (2020). Integrated microsystems for the: In situ genetic detection of dengue virus in whole blood using direct sample preparation and isothermal amplification. Analyst145, 2405ā2411. 10.1039/c9an02435b
197
YoonT.KimS.KimJ. H.ParkK. S. (2021). A syringe-based and centrifugation-free DNA extraction procedure for the rapid detection of bacteria. Chemosensors9, 167. 10.3390/chemosensors9070167
198
YuJ.WuH.HeL.TanL.JiaZ.GanN. (2021). The universal dual-mode aptasensor for simultaneous determination of different bacteria based on naked eyes and microfluidic-chip together with magnetic DNA encoded probes. Talanta225, 122062. 10.1016/j.talanta.2020.122062
199
YuQ.ZhangJ.QiuW.LiK.QianL.ZhangX.et al (2021). Gold nanorods-based lateral flow biosensors for sensitive detection of nucleic acids. Microchim. Acta.188, 133ā138. 10.1007/s00604-021-04788-z
200
ZengS.ChenB.CaiL.LiuQ. (2016). āThermal simulation for continuous-flow PCR system,ā in 2016 International Conference on Computer Science and Electronic Technology (Atlantis Press), 273ā276. 10.2991/cset-16.2016.64
201
ZhangG.LiuS.NingS.CuiW.ReedM. (2021). āParallel trapping and controllable lysis of cells Using acoustic pillar array chip,ā in IEEE international ultrasonics symposium, IUS (IEEE). 10.1109/IUS52206.2021.9593541
202
ZhangH.SongZ.PanF.HeF. (2018). A surface-confined DNA assembly enabled target recycling amplification for multiplexed electrochemical DNA detection. J. Electroanal. Chem.833, 290ā296. 10.1016/j.jelechem.2018.09.036
203
ZhangM.LiuJ.ShenZ.LiuY.SongY.LiangY.et al (2021). A newly developed paper embedded microchip based on LAMP for rapid multiple detections of foodborne pathogens. BMC Microbiol.21, 197ā213. 10.1186/s12866-021-02223-0
204
ZhangZ.ZhaoS.HuF.YangG.LiJ.TianH.et al (2020). An LED-driven AuNPs-PDMS microfluidic chip and integrated device for the detection of digital loop-mediated isothermal DNA amplification. Micromachines11, 177. 10.3390/mi11020177
205
ZhangZ.ZhaoS.JiangL.WuJ.ZhaoW.GuoX.et al (2022). A sample-to-answer DNA detection microfluidic system integrating sample pretreatment and smartphone-readable gradient plasmonic photothermal continuous-flow PCR. Analyst147, 4876ā4887. 10.1039/d2an00908k
206
ZhaoX.HuangY.LiX.YangW.LvY.SunW.et al (2022). Full integration of nucleic acid extraction and detection into a centrifugal microfluidic chip employing chitosan-modified microspheres. Talanta250, 123711. 10.1016/j.talanta.2022.123711
207
ZhaoX.LiX.YangW.PengJ.HuangJ.MiS. (2021). An integrated microfluidic detection system for the automated and rapid diagnosis of high-risk human papillomavirus. Anal146, 5102ā5114. 10.1039/D1AN00623A
208
ZhengG.GaoQ.JiangY.LuL.LiJ.ZhangX.et al (2021). Instrumentation-compact digital microfluidic reaction interface-extended loop-mediated isothermal amplification for sample-to-answer testing of Vibrio parahaemolyticus. Anal. Chem.93, 9728ā9736. 10.1021/acs.analchem.1c00917
209
ZhouL.ChenY.FangX.LiuY.DuM.LuX.et al (2020). Microfluidic-RT-LAMP chip for the point-of-care detection of emerging and re-emerging enteric coronaviruses in swine. Anal. Chim. Acta1125, 57ā65. 10.1016/j.aca.2020.05.034
210
ZhouQ. J.LuJ. F.SuX. R.JinJ. L.LiS. Y.ZhouY.et al (2021). Simultaneous detection of multiple bacterial and viral aquatic pathogens using a fluorogenic loop-mediated isothermal amplification-based dual-sample microfluidic chip. J. Fish. Dis.44, 401ā413. 10.1111/jfd.13325
211
ZhouQ.PanJ.MoL.LuoZ.QinZ.DaiZ.et al (2022). Fluorescent on-site detection of multiple pathogens using smartphone-based portable device with paper-based isothermal amplification chip. Microchim. Acta.189, 333ā410. 10.1007/s00604-022-05419-x
212
ZhuC.HuA.CuiJ.YangK.ZhuX.LiuY.et al (2019). A lab-on-a-chip device integrated DNA extraction and solid phase PCR array for the genotyping of high-risk HPV in clinical samples. Micromachines10, 537. 10.3390/mi10080537
213
ZhuY. S.ShaoN.ChenJ. W.QiW. B.LiY.LiuP.et al (2020). Multiplex and visual detection of African Swine Fever Virus (ASFV) based on Hive-Chip and direct loop-mediated isothermal amplification. Anal. Chim. Acta1140, 30ā40. 10.1016/j.aca.2020.10.011
214
ZhuangJ.ZhaoZ.LianK.YinL.WangJ.ManS.et al (2022). SERS-based CRISPR/Cas assay on microfluidic paper analytical devices for supersensitive detection of pathogenic bacteria in foods. Biosens. Bioelectron.207, 114167. 10.1016/j.bios.2022.114167
Summary
Keywords
nucleic acid testing, rapid on-site detection, nucleic acid amplification, biosensor, microfluidic chip
Citation
Wang J, Jiang H, Pan L, Gu X, Xiao C, Liu P, Tang Y, Fang J, Li X and Lu C (2023) Rapid on-site nucleic acid testing: On-chip sample preparation, amplification, and detection, and their integration into all-in-one systems. Front. Bioeng. Biotechnol. 11:1020430. doi: 10.3389/fbioe.2023.1020430
Received
16 August 2022
Accepted
12 January 2023
Published
01 February 2023
Volume
11 - 2023
Edited by
Francesca Costantini, Sapienza University of Rome, Italy
Reviewed by
Roald Tiggelaar, University of Twente, Netherlands
Jinhong Guo, University of Electronic Science and Technology of China, China
Updates
Copyright
Ā© 2023 Wang, Jiang, Pan, Gu, Xiao, Liu, Tang, Fang, Li and Lu.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Chenze Lu, chenzelu@cjlu.edu.cn
This article was submitted to Biosensors and Biomolecular Electronics, a section of the journal Frontiers in Bioengineering and Biotechnology
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.