ORIGINAL RESEARCH article

Front. Bioeng. Biotechnol., 07 November 2023

Sec. Biomechanics

Volume 11 - 2023 | https://doi.org/10.3389/fbioe.2023.1279067

Biomechanical evaluation of three implants for treating unstable femoral intertrochanteric fractures: finite element analysis in axial, bending and torsion loads

  • Department of Orthopedics, Hong Hui Hospital, Xi’an Jiaotong University, Xi’an, Shaanxi, China

Abstract

Purpose: How to effectively enhance the mechanical stability of intramedullary implants for unstable femoral intertrochanteric fractures (UFIFs) is challenging. The authors developed a new implant for managing such patients. Our aim was to enhance the whole mechanical stability of internal devices through increasing antirotation and medial support. We expected to reduce stress concentration in implants. Each implant was compared to proximal femoral nail antirotation (PFNA) via finite element method.

Methods: Adult AO/OTA 31-A2.3 fracture models were constructed, and then the new intramedullary system (NIS), PFNA, InterTan nail models were assembled. We simulated three different kinds of load cases, including axial, bending, and torsion loads. For further comparison of PFNA and the NIS, finite element analysis (FEA) was repeated for five times under axial loads of 2100 N. Two types of displacement and stress distribution were assessed.

Results: Findings showed that the NIS had the best mechanical stability under axial, bending, and torsion load conditions compared to PFNA and InterTan. It could be seen that the NIS displayed the best properties with respect to maximal displacement while PFNA showed the worst properties for the same parameter in axial loads of 2100 N. In terms of maximal stress, also the NIS exhibited the best properties while PFNA showed the worst properties in axial loads of 2100 N. For bending and torsion load cases, it displayed a similar trend with that of axial loads. Moreover, under axial loads of 2100 N, the difference between the PFNA group and the NIS group was statistically significant (p < 0.05).

Conclusion: The new intramedullary system exhibited more uniform stress distribution and better biomechanical properties compared to the PFNA and InterTan. This might provide a new and efficacious device for managing unstable femoral intertrochanteric fractures.

Introduction

Accompanied by the progress of population aging, the morbidity for hip fractures is rising every year. It is predicted to increase about 12% from 2010 to 2030 (), and half of these patients will occur in Asia by 2050 (). 41%–50% of these fractures belong to femoral intertrochanteric fractures in elderly cases (). The mortality rate of such fractures could reach 36% within 1 year (). The current recommended treatment plan is to undergo firm internal fixation within 24–48 h after injury (). This is beneficial for patients to get early rehabilitation and avoid long-term bed rest complications, and current research recommends intramedullary fixation for most patients (; ). However, the comminuted intertrochanteric fractures (especially AO/OTA 31-A2.3) account for more than 80% of UFIFs (), which are the main component of implant failure. The mechanical properties of a fixation device are the most important factors to guarantee good therapeutic effects and reduce implant failure.

Currently, the widely used implants in treating UFIFs include PFNA, Gamma3 nails and InterTan nails. The fixation techniques of these intramedullary nails are minimally invasive and easy to operate. Yet, the proximal parts of UFIFs are prone to implant lossening and even failure because of swing effects while they are fixed via a cephalomedullary nail due to the osteoporotic medullary cavity resulting from osteopenia and the advanced age (; ; ). The cephalomedullary nail of PFNA and Gamma3 is designed with only one nail, and its antirotation effect is poor. The InterTan nail has two parallel cephalomedullary nails at the neck, which can enhance its antirotation ability. However, the improvement of antirotation is limited. In addition, PFNA, Gamma3 and InterTan nails cannot provide enough medial support for patients with comminuted or defective medial wall of the femoral trochanter. The failure rate of these implants ranges from 8% to 56%, including withdrawal, cut-out, varus collapse, etc (; ; ; ). Therefore, increasing antirotation and medial support of UFIFs are vital factors for lowering the incidence of implant failure (). For these reasons, our team developed a new intramedullary implant to manage UFIFs. The proximal section of the new implant contains three nails, with two cephalomedullary nails passing into the femoral head at a specific angle and one subtrochanteric nail supporting the medial wall (Figure 1). We supposed that this design might provide good antirotation and medial support for patients with UFIFs.

FIGURE 1

Finite element analysis is a computer simulation technique to simulate the real object via applying mathematic approximate values. It has gained widespread recognition in new implant designing field of traumatic orthopedics, such as allowing for precise quantitative calculation of displacement, and for load distributions in simulated new implants and relevant bones (; ). For this study, a new intramedullary implant was designed to manage unstable femoral intertrochanteric fractures. The new intramedullary system (NIS) was compared to InterTan nail and PFNA via finite element method. The stress distribution and corresponding displacement were tested and recorded. Axial, bending, and torsion load conditions were simulated in the experiment. For further comparison of PFNA and the NIS, finite element analysis was repeated for five times under axial loads of 2100 N.

Materials and methods

Model construction and virtual surgery

This study was approved by the biomedical research ethics committee of the Xi’an Hong Hui hospital (No.202301006). Our team recruited a healthy male volunteer with the age of 65 years. The volunteer has presented written informed consent before participating in the research. A three-dimensional (3D) femoral model was built according to computed tomography (CT) scan data of the volunteer via Mimics software (Materialise, Leuven, Belgium). Voltage and current operating ranges were set to 70–140 kV and 30–800 mA for the CT equipment, respectively. Hounsfield Unit (HU) value was used to identify cortical bones. For cortical and cancellous bones, HU value was bounded by 700 (). Then, a standardized posteromedial unsupported UFIF model (AO/OTA 31-A2.3: the most unstable and common type among comminuted intertrochanteric fractures) was established according to previous studies (; ). The intertrochanteric crest and the lesser trochanter between the two osteotomy lines were removed, and part of the greater trochanter, especially the posterior part, was removed. Computer-aided design software was applied to construct three structures of fixation for UFIFs. After format conversion, the implant models were installed on the femurs. Three configurations were obtained: PFNA, InterTan nail and the NIS models. Figure 1A displays the schematic diagrams of three implants. Figure 1B shows the metal model of the new intramedullary system. For the NIS, the proximal part of the main nail is 17 mm (diameter). The specification of the main nail is 10 mm (diameter) × 170 mm (length). The specification of the sleeve, two cephalomedullary nails, and the subtrochanteric nail is 12 mm, 9 mm, 6.4 mm, and 5.0 mm in diameter for the NIS, respectively. In the NIS, the designed angle between the lower cephalomedullary nail and the main intramedullary nail is 130°. For the two cephalomedullary nails, it is 7.5°. Moreover, the designed angle between the subtrochanteric nail and the sleeve is 70°.

Finite element settings

These three different configurations all were given homogeneous and linearly isotropic material characteristics. Tetrahedral elements were applied for meshing. To evaluate the reliability of three configurations, a convergence research was performed (). In the case of maximum Degree of Freedom, the field parameters of the two types of elements, including strain energy and displacement, were in the range of 5%. In addition, the maximal stress point did not exist. Table 1 displayed the values of elements and nodes for three configuration models. In terms of material features, the elastic modulus was set at 16,800 MPa, 840 MPa, and 110,000 MPa for cortical, cancellous bones, and implants, respectively (; ). In terms of Poisson’s ratio, it was set at 0.3 with regard to cortical bones and implants while 0.2 as to cancellous bones (). These three implants were set to titanium material features as titanium and its alloys possess good biocompatibility, excellent corrosion resistance, and superior mechanical characteristics (). Frictional contacts were defined for all contact conditions. The friction coefficient was set at 0.4 based on former studies (). Figure 2 showed the boundary conditions for axial, bending, and torsion loads. The femoral condyle was tightly fixed to prohibit the overall motion of configurations. The axial loads were 2,100 N, acting vertically onto the surface of the femoral head (; ). Under bending boundary conditions, the femoral shaft and condyle were properly fixed. The loads were set as 175 N, acting laterally onto the femoral head (). Under torsion boundary conditions, with the direction of the femoral neck as the axis, the torsion loads were set at 15 Nm acting onto the femoral head ().

TABLE 1

ModelNodesElements
PFNA405406261379
InterTan506609323206
NIS494536314645

Number of nodes and elements for the three different implants.

PFNA, proximal femoral nail antirotation; NIS, new intramedullary system.

FIGURE 2

Indexes for evaluation

Displacement and stress for implants and bones were recorded and assessed. PFNA was set as the control group as it has been widely used in clinical work recently and was recognized to possess superior mechanical properties (). The variation rate was estimated using this arithmetic formula: VR =(V1 − Vn)/V1×100%. VR stands for the variation rate. Vn stands for values of InterTan, or NIS. V1 means the value for PFNA. Besides, in order to compare PFNA and the NIS deeply, finite element analysis was repeated for five times under axial loads of 2100 N.

Statistical analysis

This research used SPSS 23.0 software (IBM Co., United States) to perform statistical analysis. Finite element analysis data, including two types of displacement and stress under axial loads of 2100 N, between the PFNA and NIS groups were statistically compared using the Student’s t-test. p < 0.05 was defined as statistically significant.

Results

Maximal displacement for three fixation models

Figure 3 showed maximal displacement of three configurations under different load conditions. Specially, when axial loads of 2100 N were applied the maximal displacement for PFNA, InterTan and NIS was 17.02 mm, 15.50 mm, and 13.38 mm (0.33 mm, 0.35 mm, and 0.27 mm under bending loads of 175 N; 2.22 mm, 2.08 mm, and 1.86 mm under torsion loads of 15 Nm). Compared with PFNA, the maximal displacement reduction for NIS was 21.4% under axial loads (16.5% under bending loads and 16.5% under torsion loads). The maximal displacement of the NIS was significantly lower compared to the value of PFNA.

FIGURE 3

Maximal displacement of fracture surface (MDFS)

The values of MDFS for three different configurations were shown in Figure 4. While axial loads of 2100 N were applied, from the highest to the lowest value of this index, the three configurations were sorted as follows: PFNA, InterTan, and NIS. For bending load conditions, they were sorted as follows: InterTan, PFNA, and NIS. For torsion load conditions, they were sorted as follows: PFNA, InterTan, and NIS. The value of MDFS for the NIS was less than that of PFNA under three load cases. The MDFS reduction of the NIS relative to PFNA was 23.0% under axial loads of 2100 N while 32.3% for bending conditions. In addition, compared to PFNA, the MDFS reduction of the NIS was 12.8% when torsion loads were applied. The trend direction of MDFS was similar to that of maximal displacement for three fixation configurations. The above data indicated that the new intramedullary system possessed better biomechanical stability than PFNA for managing UFIFs.

FIGURE 4

Stress at implants for three fixation models

Figure 5 showed stress on implants for three fixation models. The maximal stress of the three implants occurred near the junction between the main nail and the cephalomedullary nail. This demonstrated that the area where the cephalomedullary nail passed through the main nail was a stress concentration area. When axial loads of 2100 N were applied, the maximal stress on implants for PFNA, InterTan and NIS was 1257.3 MPa, 1069.2 MPa, and 928.99 MPa. Compared to PFNA, the maximal stress reduction at implants for NIS was 26.1% under axial loads. Besides, the maximal stress at implants was 59.333 MPa, 96.302 MPa, and 104.64 MPa for PFNA, InterTan and NIS under bending loads while under torsion loads it was 512.3 MPa, 477.11 MPa, and 438.16 MPa, respectively. The above data demonstrated that the NIS possessed more uniform stress distribution than PFNA and InterTan for managing UFIFs.

FIGURE 5

Stress at bones

Stress at bones for three different models was displayed in Figure 6. When axial loads of 2100 N were applied, from the highest to the lowest stress at bone, the three configurations were sorted as follows: PFNA, InterTan, and NIS. Under bending load conditions, it was also sorted as follows: PFNA, InterTan and NIS. Under torsion load conditions, it was sorted as follows: InterTan, NIS, and PFNA. The maximal stress reduction of the NIS relative to PFNA reached 7.6% under axial loads of 2100 N, and 26.4% under bending loads, respectively. The trend direction of stress on bones was similar to that of stress on implants for UFIFs.

FIGURE 6

Comparison of biomechanical properties and statistical analysis

Table 2 displayed statistical analysis data between the PFNA and NIS groups. Five repeated tests were conducted under axial loads of 2100N. The maximal displacement was 13.46 ± 0.57 mm for the NIS group and 17.06 ± 0.67 mm for the PFNA group, with significant difference between the two groups (p < 0.05). The values of MDFS were 11.42 ± 0.43 mm and 14.88 ± 0.51 mm for the NIS and and PFNA groups, and the difference was significant (p < 0.05). The maximal stress at implants was 927.2 ± 25.7 MPa and 1250.1 ± 47.7 MPa, and there was significant difference between the two groups (p < 0.05). The maximal stress at bones was 357.1 ± 16.7 MPa and 381.7 ± 13.3 MPa for the NIS and PFNA groups, and the difference was statistically significant (p < 0.05). Therefore, under axial loads of 2100 N, the NIS showed better biomechanical properties than PFNA for managing UFIFs.

TABLE 2

Biomechanical propertiesPFNANIStp
Maximal displacement (mm)17.06 ± 0.6713.46 ± 0.579.1510.000
Maximal displacement of fracture surface (mm)14.88 ± 0.5111.42 ± 0.4311.5980.000
Maximal stress at implants (MPa)1250.1 ± 47.7927.2 ± 25.713.3220.000
Maximal stress at bones (MPa)381.7 ± 13.3357.1 ± 16.72.5770.037

Statistical analysis of comparison of biomechanical properties between PFNA and NIS under axial loads of 2100 N.

Notes: PFNA, stands for proximal femoral nail antirotation; NIS, stands for the new intramedullary system.

Discussion

Intramedullary fixation methods have been the mainstream management for unstable femoral intertrochanteric fractures because of minimal invasion, mild soft tissue injuries, and central fixation (; ). Among the limited intramedullary implants, PFNA is regarded as the golden standard by lots of scholars for such fractures (). However, PFNA fixation is accompanied by a high rate of surgical complications, including withdrawal, cut-out, varus collapse (). The occurrence of these complications puts patients at a relatively high risk of re-operation and mortality. Therefore, designing implants with better biomechanical properties for managing UFIFs has become a research hotspot.

Wang et al. compared the biomechanical properties of PFNA, InterTan, and proximal femur bionic nail (PFBN) for the management of elderly intertrochanteric fractures via finite element method (). Their results indicated that compared to PFNA and InterTan nails, PFBN possesses better biomechanical characteristics. There are two nails, including one pressure nail, and one tension nail in the femoral neck, which are locked at a certain angle, enhancing the biomechanical stability of PFBN. Yet, the medial support effect of PFBN for comminuted intertrochanteric fractures of the medial wall is not yet clear. Regaining medial support is vital for comminuted and unstable intertrochanteric fractures. Yet, it is sometimes unachievable in a specific operation. These comminuted intertrochanteric fractures occupy over eighty percent of unstable intertrochanteric fractures (). As a result, they have been the major cause for implant failures. Therefore, strengthening the support of the medial wall via implants may be a good scheme for such fractures. Nie et al. developed a new intramedullary implant through enhancing medial cortical support to decrease complications for UFIFs via finite element method (). It is called medial sustain nail (MSN). Their research indicated that MSN possesses better mechanical properties than PFNA. The proximal part of this implant has two nails, including one long nail and one short nail. The short nail is located on the inner side, and could provide good medial support. Yet, this design has limited antirotation effect and is prone to cut-out and loosening. Other scholars have used the cerclage cable technique to enhance the medial support of the cephalomedullary nail in patients with intertrochanteric fractures (), but whether this technique could improve fixation is still controversial (; ). Ceynowa et al.‘s study showed that the cerclage cable did not significantly enhance medial support of intertrochanteric fractures via mechanical tests (). Based on these factors, the authors have attempted to develop a new implant to better fix unstable intertrochanteric fractures.

According to traditional biomechanical theory, finite element method could transform the object being studied into a configuration formed by numbers of finite unit combinations, and the mathematical simulation analysis is gotten. The data of the object’s displacement, stress distribution, etc. could be visually reflected the whole or partial mechanical features of the configuration, and different parameters could be timely rectified. Compared to biomechanical experiments, finite element method possesses several advantages, such as short cycles, low costs, and high efficiency (). The authors performed finite element analysis of PFNA, InterTan, and NIS for fixation of AO/OTA 31-A2.3 unstable femoral intertrochanteric fractures. Based on our data, the NIS possessed the best mechanical characteristics, followed by InterTan, and then PFNA under axial, bending, and torsion load conditions. Moreover, after repeating for five times under axial loads of 2100 N, the statistical difference between the PFNA group and the NIS group was significant (p < 0.05). The unique design of the NIS might lead to these results. There are three nails in the proximal section of the implant. Two of them are distributed with the included angle of 7.5° in the femoral head. The design of 7.5° ensures that it could provide good antirotation effect while it will not penetrate the femoral neck due to excessive angle. Actually, as shown in the results under torsion loads, it did play a better antirotation effect than the single cephalomedullary nail of PFNA and the parallel and tight design of InterTan. In the PFNA and InterTan models, the junction between the cephalomedullary nail and the main nail is a stress concentration area. In the NIS model, due to the introduction of the subtrochanteric nail, the stress of the implant is dispersed to a certain extent. This reduces the risk of implant failure. In addition, the subtrochanteric nail of NIS is supported under the lesser trochanter, which can play a supporting role for the medial wall of the trochanter. This nail could effectively disperse bending and axial loads. This might enable the NIS to effectively reduce the occurrence of varus deformity in patients with UFIFs.

There are several limitations of this study. The finite element model was applied to compare the mechanical performance of PFNA, InterTan, and NIS for treating UFIFs. The experiment model and analysis process were simplified to a certain extent, and the function of soft tissues and ligaments at bones and implants has not been assessed. Currently, the mechanical evaluation and mathematical simulation of the limb just apply several single direction loads, such as axial, bending or torsion loads. In real life, under physiological conditions, the load acting on the limbs, including the hip joint, is often not a single direction load, but a combination of loads from different directions. The effect of loads on bones and implants may vary depending on the posture and motion states of the limbs. These factors might lead to certain deviations in the experimental results. In addition, the existing conclusions still require further validation through clinical research.

Conclusion

Compared with PFNA and InterTan nails, the new intramedullary system displayed better mechanical properties under axial, bending, and torsion load conditions. Thus, it may offer a better option for orthopedics to manage unstable intertrochanteric fractures.

Statements

Data availability statement

The original contributions presented in the study are included in the article/Supplementary material, further inquiries can be directed to the corresponding authors.

Ethics statement

This study was approved by the biomedical research ethics committee of the Xi’an Hong Hui hospital (No. 202301006).The volunteer has presented written informed consent before participating in the research.

Author contributions

HB: Conceptualization, Methodology, Software, Writing–review and editing. LL: Formal Analysis, Investigation, Writing–review and editing. ND: Data curation, Writing–review and editing. HX: Writing–review and editing, Methodology, Supervision. LS: Formal Analysis, Investigation, Writing–review and editing. ML: Investigation, Conceptualization, Methodology, Writing–review and editing. ZL: Conceptualization, Resources, Validation, Writing–review and editing. KZ: Conceptualization, Methodology, Supervision, Writing–review and editing. QW: Conceptualization, Supervision, Writing–review and editing. QH: Conceptualization, Methodology, Writing–original draft.

Funding

The author(s) declare financial support was received for the research, authorship, and/or publication of this article. This study was supported by the National Natural Science Foundation of Shaanxi Province (2022JQ-757), the Clinical Application-oriented Medical Innovation Foundation from National Clinical Research Center for Orthopedics, Sports Medicine and Rehabilitation and Jiangsu China-Israel Industrial Technical Research Institute Foundation (2021-NCRC-CXJJ-PY-11). The funding source has no role in study design, conduction, data collection or statistical analysis.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

  • 1

    Abdul WahabA. H.WuiN. B.Abdul KadirM. R.RamleeM. H. (2020). Biomechanical evaluation of three different configurations of external fixators for treating distal third tibia fracture: finite element analysis in axial, bending and torsion load. Comput. Biol. Med.127, 104062. 10.1016/j.compbiomed.2020.104062

  • 2

    BhandariM.SwiontkowskiM. (2017). Management of acute hip fracture. N. Engl. J. Med.377 (21), 2053–2062. 10.1056/NEJMcp1611090

  • 3

    BroxW. T.RobertsK. C.TaksaliS.WrightD. G.WixtedJ. J.TubbC. C.et al (2015). The American Academy of Orthopaedic surgeons evidence-based guideline on management of hip fractures in the elderly. J. Bone. Jt. Surg. Am.97, 1196–1199. 10.2106/JBJS.O.00229

  • 4

    CeynowaM.ZerdzickiK.KlosowskiP.PankowskiR.RocławskiM.MazurekT. (2020). The early failure of the gamma nail and the dynamic hip screw in femurs with a wide medullary canal. A biomechanical study of intertrochanteric fractures. Clin. Biomech.71, 201–207. 10.1016/j.clinbiomech.2019.11.006

  • 5

    CeynowaM.ZerdzickiK.KlosowskiP.PankowskiR.RocławskiM.MazurekT. (2021). Cerclage cable augmentation does not increase stability of the fixation of intertrochanteric fractures. A biomechanical study. Orthop. Traumatol. Surg. Res.107 (6), 103003. 10.1016/j.otsr.2021.103003

  • 6

    ChapmanT.ZmistowskiB.KriegJ.StakeS.JonesC. M.LevicoffE. (2018). Helical blade versus screw fixation in the treatment of hip fractures with cephalomedullary devices: incidence of failure and atypical "medial cutout. J. Orthop. Trauma.32 (8), 397–402. 10.1097/BOT.0000000000001193

  • 7

    ChenX.HuY.GengZ.SuJ. (2022). The "three in one" bone repair strategy for osteoporotic fractures. Front. Endocrinol. (Lausanne).13, 910602. 10.3389/fendo.2022.910602

  • 8

    GlassnerP. J.TejwaniN. C. (2011). Failure of proximal femoral locking compression plate: a case series. J. Orthop. Trauma.25, 76–83. 10.1097/BOT.0b013e3181e31ccc

  • 9

    GrønhaugK. M. L.DybvikE.MatreK.ÖstmanB.GjertsenJ. E. (2022). Intramedullary nail versus sliding hip screw for stable and unstable trochanteric and subtrochanteric fractures: 17,341 patients from the Norwegian hip fracture register. Bone. Jt. J.104-B (2), 274–282. 10.1302/0301-620X.104B2.BJJ-2021-1078.R1

  • 10

    HaidukewychG. J. (2009). Intertrochanteric fractures: ten tips to improve results. J. Bone. Jt. Surg. Am.91, 712–719.

  • 11

    HuangH.XiangC.ZengC.OuyangH.WongK. K. L.HuangW. (2015). Patient-specific geometrical modeling of orthopedic structures with high efficiency and accuracy for finite element modeling and 3D printing. Australas. Phys. Eng. Sci. Med.38, 743–753. 10.1007/s13246-015-0402-1

  • 12

    HuangQ.ZhangC.BaiH.WangQ.LiZ.LuY.et al (2023). Biomechanical evaluation of two modified intramedullary fixation system for treating unstable femoral neck fractures: a finite element analysis. Front. Bioeng. Biotechnol.11, 1116976. 10.3389/fbioe.2023.1116976

  • 13

    HungW. W.EgolK. A.ZuckermanJ. D.SiuA. L. (2012). Hip fracture management: tailoring care for the older patient. JAMA307 (20), 2185–2194. 10.1001/jama.2012.4842

  • 14

    JinW. J.DaiL. Y.CuiY. M.ZhouQ.JiangL. S.LuH. (2005). Reliability of classification systems for intertrochanteric fractures of the proximal femur in experienced orthopaedic surgeons. Injury36, 858–861. 10.1016/j.injury.2005.02.005

  • 15

    JitprapaikulsarnS.ChantarapanichN.GromprasitA.MahaisavariyaC.PatamamongkonchaiC. (2021). Single lag screw and reverse distal femur locking compression plate for concurrent cervicotrochanteric and shaft fractures of the femur: biomechanical study validated with a clinical series. Eur. J. Orthop. Surg. Traumatol.31, 1179–1192. 10.1007/s00590-020-02868-z

  • 16

    KangS. J.BaoF. L.HuangD. S.JiangT.HuY. M.LiJ. M.et al (2021). Percutaneous cerclage wiring combined with Cephalomedullary nailing for irreducible subtrochanteric fractures. Orthop. Surg.13 (6), 1899–1911. 10.1111/os.13144

  • 17

    KokoroghiannisC.AktselisI.DeligeorgisA.FragkomichalosE.PapadimasD.PappadasI. (2012). Evolving concepts of stability and intramedullary fixation of intertrochanteric fractures–a review. Injury43, 686–693. 10.1016/j.injury.2011.05.031

  • 18

    KulkarniS. G.BabhulkarS. S.KulkarniS. M.KulkarniG. S.KulkarniM. S.PatilR. (2017). Augmentation of intramedullary nailing in unstable intertrochanteric fractures using cerclage wire and lag screws: a comparative study. Injury48 (2), S18–S22. 10.1016/S0020-1383(17)30489-8

  • 19

    LewisS. R.MaceyR.LewisJ.StokesJ.GillJ. R.CookJ. A.et al (2022). Surgical interventions for treating extracapsular hip fractures in older adults: a network meta-analysis. Cochrane. Database. Syst. Rev.2 (2), CD013405. 10.1002/14651858.CD013405.pub2

  • 20

    LiJ.HanL.ZhangH.ZhaoZ.SuX.ZhouJ.et al (2019). Medial sustainable nail versus proximal femoral nail antirotation in treating AO/OTA 31-A2.3 fractures: finite element analysis and biomechanical evaluation. Injury50 (3), 648–656. 10.1016/j.injury.2019.02.008

  • 21

    LiJ.YinP.ZhangL.ChenH.TangP. (2019). Medial anatomical buttress plate in treating displaced femoral neck fracture a finite element analysis. Injury50 (11), 1895–1900. 10.1016/j.injury.2019.08.024

  • 22

    LiJ.ZhangL.ZhangH.YinP.LeiM.WangG.et al (2019). Effect of reduction quality on post-operative outcomes in 31-A2 intertrochanteric fractures following intramedullary fixation: a retrospective study based on computerised tomography findings. Int. Orthop.43 (8), 1951–1959. 10.1007/s00264-018-4098-1

  • 23

    LiJ.ZhaoX.HuX.TaoC.JiR. (2018). A theoretical analysis and finite element simulation of fixator-bone system stiffness on healing progression. J. Appl. Biomater. Funct. Mater.16, 115–125. 10.1177/2280800017750357

  • 24

    LiuY.TaoR.LiuF.WangY.ZhouZ.CaoY.et al (2010). Mid-term outcomes after intramedullary fixation of peritrochanteric femoral fractures using the new proximal femoral nail antirotation (PFNA). Injury41, 810–817. 10.1016/j.injury.2010.03.020

  • 25

    LongH.CaoR.ZhangH.QiuY.YinH.YuH.et al (2022). Incidence of hip fracture among middle-aged and older Chinese from 2013 to 2015: results from a nationally representative study. Arch. Osteoporos.17 (1), 48. 10.1007/s11657-022-01082-0

  • 26

    MahaisavariyaB.ChantarapanichN.RiansuwanK.SitthiseripratipK. (2014). Prevention of excessive medialisation of trochanteric fracture by a buttress screw: a novel method and finite element analysis. J. Med. Assoc. Thai.97(9), S127–S132.

  • 27

    MengF.YinZ.RenX.GengZ.SuJ. (2022). Construction of local drug delivery system on titanium-based implants to improve osseointegration. Pharmaceutics17 (5), 1069. 10.3390/pharmaceutics14051069

  • 28

    NieS.LiJ.LiM.HaoM.WangK.XiongY.et al (2022). Finite-element analysis of a novel cephalomedullary nail for restricted sliding to reduce risk of implant failure in unstable intertrochanteric fractures. Orthop. Surg.14 (11), 3009–3018. 10.1111/os.13497

  • 29

    ParkerM. J.HandollH. H. (2010). Gamma and other cephalocondylic intramedullary nails versus extramedullary implants for extracapsular hip fractures in adults. Cochrane. Database. Syst. Rev.26, Cd000093. 10.1002/14651858.CD000093.pub5

  • 30

    QueallyJ. M.HarrisE.HandollH. H. G.ParkerM. J. (2014). Intramedullary nails for extracapsular hip fractures in adults. Cochrane. Database. Syst. Rev.9, CD004961. 10.1002/14651858.CD004961.pub4

  • 31

    RehmeJ.WoltmannA.BrandA.von RüdenC. (2021). Does auxiliary cerclage wiring provide intrinsic stability in cephalomedullary nailing of trochanteric and subtrochanteric fractures?Int. Orthop.45 (5), 1329–1336. 10.1007/s00264-020-04795-4

  • 32

    SongH.ChangS. M.HuS. J.DuS. C. (2022). Low filling ratio of the distal nail segment to the medullary canal is a risk factor for loss of anteromedial cortical support: a case control study. J. Orthop. Surg. Res.17 (1), 27. 10.1186/s13018-022-02921-z

  • 33

    TannerD. A.KloseckM.CrillyR. G.ChesworthB.GillilandJ. (2010). Hip fracture types in men and women change differently with age. Bmc. Geriatr.10, 12. 10.1186/1471-2318-10-12

  • 34

    VicecontiM.MucciniR.BernakiewiczM.BaleaniM.CristofoliniL. (2000). Large-sliding contact elements accurately predict levels of bone implant micromotion relevant to osseointegration. J. Biomech.33 (12), 1611–1618. 10.1016/s0021-9290(00)00140-8

  • 35

    WangY.ChenW.ZhangL.XiongC.ZhangX.YuK.et al (2022). Finite element analysis of proximal femur bionic nail (PFBN) compared with proximal femoral nail antirotation and InterTan in treatment of intertrochanteric fractures. Orthop. Surg.14 (9), 2245–2255. 10.1111/os.13247

  • 36

    ZhangY.TianL.YanY.SangH.MaZ.JieQ.et al (2011). Biomechanical evaluation of the expansive cannulated screw for fixation of femoral neck fractures. Injury42 (11), 1372–1376. 10.1016/j.injury.2011.07.004

Summary

Keywords

biomechanical, unstable, intertrochanteric fracture, finite element model, implant, femur

Citation

Bai H, Liu L, Duan N, Xue H, Sun L, Li M, Li Z, Zhang K, Wang Q and Huang Q (2023) Biomechanical evaluation of three implants for treating unstable femoral intertrochanteric fractures: finite element analysis in axial, bending and torsion loads. Front. Bioeng. Biotechnol. 11:1279067. doi: 10.3389/fbioe.2023.1279067

Received

17 August 2023

Accepted

30 October 2023

Published

07 November 2023

Volume

11 - 2023

Edited by

Eiji Tanaka, Tokushima University, Japan

Reviewed by

Zhen Geng, Shanghai University, China

Minghui Yang, Beijing Jishuitan Hospital, China

Updates

Copyright

*Correspondence: Qiang Huang, ; Qian Wang,

† These authors have contributed equally to this work and share first authorship

Disclaimer

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Outline

Figures

Cite article

Copy to clipboard


Export citation file


Share article

Article metrics