Abstract
Aims:
The ovine stifle is an established model for evaluation of knee treatments, such as meniscus replacement. This study introduces a novel ovine gait simulator for pre-testing of surgical treatments prior to in vivo animal trials. Furthermore, we describe a pilot study that assessed gait kinematics and contact pressures of native ovine stifle joints and those implanted with a novel fiber-matrix reinforced polyvinyl alcohol-polyethylene glycol (PVA-PEG) hydrogel meniscus to illustrate the efficacy of the simulator.
Methods:
The gait simulator controlled femoral flexion-extension and applied a 980N axial contact force to the distal tibia, whose movement was guided by the natural ligaments. Five right ovine stifle joints were implanted with a PVA-PEG total medial meniscus replacement, fixed to the tibia via transosseous tunnels and interference screws. Six intact and five implanted right ovine stifle joints were tested for 500 k gait cycles at 1.55 Hz. Implanted stifle joint contact pressures and kinematics in the simulator were compared to the intact group. Contact pressures were measured at 55° flexion using pressure sensitive film inserted sub-meniscally. 3D kinematics were measured optically across two 30-s captures.
Results:
Peak contact pressures in intact stifles were 3.6 ± 1.0 MPa and 6.0 ± 2.1 MPa in the medial and lateral condyles (p < 0.05) and did not differ significantly from previous studies (p > 0.4). Medial peak implanted pressures were 4.3 ± 2.2 MPa (p > 0.4 versus intact), while lateral peak pressures (9.4 ± 0.8 MPa) were raised post medial compartment implantation (p < 0.01). The range of motion for intact joints was flexion/extension 37° ± 1°, varus/valgus 1° ± 1°, external/internal rotation 5° ± 3°, lateral/medial translation 2 ± 1 mm, anterior/posterior translation 3 ± 1 mm and distraction/compression 1 ± 1 mm. Ovine joint kinematics in the simulator did not differ significantly from published in vivo data for the intact group, and the intact and implanted groups were comparable (p > 0.01), except for in distraction-compression (p < 0.01).
Conclusion:
These findings show correspondence of the ovine simulator kinematics with in vivo gait parameters. The efficacy of the simulator to evaluate novel treatments was demonstrated by implanting a PVA-PEG hydrogel medial meniscal replacement, which restored the medial peak contact pressures but not lateral. This novel simulator may enable future work on the development of surgical procedures, derisking subsequent work in live animals.
1 Introduction
Pre-clinical assessment of novel orthopaedic joint procedures generally includes material characterization and biomechanical testing in the lab setting, biocompatibility evaluation and in vivo studies in large animal models prior to the commencement of clinical trials. Biomaterial characterization provides insight into the properties and performance of single or combined materials using physiologically relevant parameters but is limited by sample geometry and controlled environments. In vivo implantation in large animal models such as goats or sheep (Zur et al., 2011; ; ) provides a thorough evaluation of the design, safety, performance and biocompatibility of the novel procedure in vivo, thus derisking the procedure before clinical trials. However, such studies are costly and time-consuming, and researchers are ethically bound to reduce such studies and limit sample sizes as far as possible while satisfying regulatory requirements.
The ovine stifle joint is a well-established knee joint model for pre-clinical research in reconstruction or replacement of the cruciate and collateral ligaments, partial and total meniscus replacements, cartilage lesion repair and osteoarthritis treatment (; ; ; Zur et al., 2011; ; )
A clear gap exists in the pre-clinical assessment of novel orthopaedic joint procedures, for dynamic cadaveric testing of the ovine knee joint under simulated physiological gait, as a screening method and at a reduced cost (), before advancing to in vivo animal studies. Research studies have developed techniques for dynamic testing of meniscus replacements in cadaveric ovine stifles but were limited to low axial loads and were only run for 10 cycles (; ; ). This paper describes such a simulator, its verification in terms of loading and kinematics, then a proof of concept study in which a meniscus replacement is tested in a cadaveric ovine knee joint under simulated gait.
Meniscal surgery is one of the most common orthopaedic surgical interventions (; ; ). Although meniscectomy is the current standard of care after irreparable meniscal injury, and may alleviate symptoms in the short-term, it increases the risk of the onset of osteoarthritis (; ). In such cases, high tibial osteotomy is often performed to shift joint contact loads to the unaffected compartment in order to reduce pain and help slow OA progression in the meniscectomized compartment (). Meniscal allograft transplants (; ) and partial replacement scaffolds (; ) have been used to replace the injured meniscus, with limited success in long-term function and survivorship (; ). Novel total meniscus replacement devices aim to fill this treatment gap (Zur et al., 2011; ; ). No proven long-term joint-preserving treatment options exist for patients with irreparable meniscal damage.
The aim of this study was to develop an ovine knee joint simulator and to assess the kinematics and contact pressures of the ovine stifle joint under simulated gait. The use of the simulator will be demonstrated in a pilot study when the knee has been implanted with a novel fiber-matrix reinforced polyvinyl alcohol-polyethylene glycol (PVA-PEG) hydrogel meniscus and to compare it against intact stifles. It was hypothesized that the gait kinematics and contact pressures post meniscal replacement would not differ significantly from those of intact stifles.
2 Materials and methods
Two single-station ovine gait simulators were used (Figure 1). The joint simulator controlled femoral flexion-extension and applied a cyclic axial contact force to the distal end of the tibia, while allowing articular movement in five degrees of freedom (DOF). During knee flexion-extension, tibial movement was guided by the natural ligaments of the stifle joint, namely, the anterior (cranial) cruciate ligament, the posterior (caudal) cruciate ligament and the medial and lateral collateral ligaments, which were left intact together with surrounding soft tissue structures. This method was expected to result in stifle kinematics that were not significantly different from in vivo conditions because motion of the natural knee is restricted and stabilized by passive soft tissues (; ). The simulator design was based on a force-controlled joint simulator that has been shown to allow physiological movement of cadaveric knee joints under gait loading conditions ().
FIGURE 1
The gait simulator ran at a frequency of 1.55 Hz for 500,000 gait cycles (3.73 days), equivalent to approximately a year of normal use by adult sheep in vivo (). The simulator was mounted inside a refrigerator to maintain the stifle at 4°C–8°C to delay tissue necrosis for the duration of the test. An in-line pressure syringe pump (Graseby 3200; Smiths Medical International, MN) was used to deliver diluted sterile-filtered Bovine Calf Serum (BCS) (product 12133C; Merck Group, Germany) from a mounted 50 mL Luer Lock syringe at a rate of 2 mL per hour. The lubricant was delivered directly into the joint space throughout the test by placing an 18G × 38 mm cannula in the femoral intercondylar notch. The syringe pump was replenished with fresh lubricant every 24 h.
2.1 Ovine gait simulator design
Design parameters for the joint simulator were determined from in vivo ovine biomechanics literature (Table 1). The overall configuration of the simulator had the femoral mounting driven in flexion-extension in a vertical plane while the tibial mounting was quasi-static below it. A crank-rocker mechanism based on Grashof’s theorem imposed the flexion-extension of the femoral mounting within the physiological walking range of 45–80 degrees flexion (). All other femoral DOF were constrained. Flexion-extension of the femoral mounting about the transepicondylar axis was allowed through self-aligning bearings (Pillow Block Bearing; RS Components, England) and controlled by a connecting rod driven by a rotating crank disc (Figure 2). The crank disc was mounted onto a single-speed induction geared motor (model SD18-0083/CONT; Parvalux Electric Motors, England) with an output speed of 93 rpm, equivalent to 1.55 Hz. This provided a simplified approximation of the gait cycle, but with a shorter stance phase (34% of the gait cycle) than that observed in vivo.
TABLE 1
| Design Parameter | Design Specification | Reference Values | Source |
|---|---|---|---|
| Flexion-Extension (°) | Minimum: 45° Maximum: 80° Range: 35° | Average Min: 42°–49° Average Max: 70°–77° Range: 34° ± 5.2° | , , |
| Peak Joint Axial Load in body weight (BW) | 2BW ≡ 1 kNa | 2.27 ± 0.44 BW | , |
| Stance (° Flexion) | 45°–55° Flexion | 45°–55° Flexion | , , |
| Stance (% gait) | 34% | 62.9% ± 4.08% | , , , |
| Gait Frequency (Hz) | 1.55 Hz | 0.85–2.27 Hz | , , |
Design parameters of the ovine gait simulator.
Based on a live animal weight of 55 kg.
FIGURE 2
The distal end of the tibia was mounted on a spherical joint, in line with the tibial long mechanical axis by a snug-fitting intramedullary rod to which the sphere was attached, that allowed unconstrained internal-external rotation with minimal resistance. Proximal and distal translation of the tibia were possible through the active and passive movement of the one-way linear air cylinder (model CDQ2B63TF-50DZ; SMC, England) piston rod during the stance and swing phases respectively (Figure 1). The linear air cylinder was fixed to a slotted swing allowing varus-valgus rotation of the tibia about an axis passing through the center of the tibiofemoral joint and perpendicular to the flexion-extension axis. The medial-lateral and anterior-posterior translations in the stifle joint are small () and were accounted-for by small rotations of the spherical joint.
A cam profile, incorporated onto the crank disc, mechanically activated a roller lever pneumatic switch (model VM430-01-01, SMC, England) and controlled the gait loading profile synchronous with the flexion-extension motion of the simulator (Figure 3). The pneumatic switch actuated the linear air cylinder, which applied the axial contact force through the tibial shaft. Axial load was actuated at around 80% of the simulated gait cycle to reach a peak load of 980N in the subsequent stance phase. The peak axial load was held throughout the stance phase and then ramped down for the swing phase.
FIGURE 3
2.2 Specimen preparation
Six paired and five right fresh-frozen North of England Mule stifle joints (n = 17) of ewes aged >2 years and weighing 56–68 kg were used. Of the six paired stifles, the right stifles were tested as intact joints in the simulator and the left contralateral stifles remained untested as controls. Each right stifle was carefully dissected to preserve the surrounding muscle tissues, capsule, collateral and cruciate ligaments and patella. The femoral and tibial shafts were transected approximately 60 mm from the proximal end of the trochlear groove and 65 mm distally to the tibial tubercle, respectively. The transepicondylar axis of the femur was aligned with the flexion-extension axis of the simulator using a static alignment jig with 1.6 mm Kirschner wires that passed through the fixture along the flexion-extension axis so that their points engaged the epicondyles. The femur was then secured by three screws and potted in polymethyl methacrylate (PMMA) bone cement (Simplex Rapid, England). Correct alignment of the epicondylar axis was imperative for ensuring physiological kinematics of the stifle in the simulator (
The five unpaired right stifle joints were prepared similarly and then implanted with the TMR device. Fiber-matrix reinforced PVA-PEG hydrogel meniscal replacement implants (n = 5) were designed and manufactured for the right medial stifle of the chosen breed and weight of sheep. A modified medial parapatellar arthrotomy was performed and then the medial collateral ligament (MCL) was released and reflected distally via an epicondylar osteotomy for adequate visualization of the medial compartment while preserving the natural ligament function and joint biomechanics (
2.3 Testing lubricant
As recommended by international joint replacement wear testing standards ISO 14243-3:2014 and ASTM F732-17 (
2.4 Dynamic cadaveric testing
Six intact and five implanted right stifle joints were tested in the simulator at a peak axial contact force of 980N for 500,000 cycles at a frequency of 1.55 Hz. The six contralateral stifle joints were left untested as controls for the survivability of the menisci and any signs of anomalous cartilage damage in the intact right stifle joints post-test. Each ovine gait simulator fatigue test was started immediately after the implantation of the TMR device in each cadaveric joint, to avoid the need for an additional freeze-thaw cycle of both the cadaveric joint and the device.
At the end of the test, stifle joint kinematics were recorded in the simulator using a motion-tracking system (Optotrak Certus; Northern Digital Inc., Canada). A three-marker rigid body was rigidly attached to each of the long bones using a bone pin. The markers were attached proximal to the femoral trochlear groove and on the proximal posteromedial aspect of the tibial shaft. A probe was used to digitize anatomic landmarks marked by small bone screws to define the joint coordinate system used clinically (
Contact pressures on the medial and lateral condyles were measured statically at 55° flexion using pressure sensitive film (Prescale Low Pressure; Fujifilm, Japan) inserted under the native menisci or medial meniscus replacement implant. The surrounding soft tissues were removed, and a transverse capsulotomy was performed at the level of the tibial plateau, while preserving the cruciate and collateral ligaments, to allow the insertion of the film strip. Strips, approximately 25 mm wide and 120 mm long, were prepared as per the manufacturer’s instructions and then encased in plastic wrap to avoid fluid seepage (
Both untested left and tested right intact stifle joints were dissected to expose the menisci and the articulating cartilage surfaces by releasing the stabilizing ligaments and disarticulating the stifle. The meniscal and cartilage condition in untested joints and immediately post-test in tested intact joints was recorded photographically. In order to demonstrate cartilage surface damage such as fibrillation or erosion, the articular surfaces were then stained using black India ink (
2.5 Statistical analysis
Multiple t-tests were performed to determine statistical differences in the overall mean range of motion of the implanted group in each DOF when compared to the intact group and the in vivo kinematic data reported by
3 Results
3.1 Joint kinematics
The average joint kinematics versus percentage gait cycle in each of the six DOF were calculated for implanted joints and compared against the intact group as shown in Table 2 and Figure 4. The average range of motion of intact joints did not differ significantly from the kinematics recorded in vivo by Tapper et al. (p > 0.01 for distraction/compression, p > 0.1 for other ranges of motion) (
TABLE 2
| Group | Rotations (°) | Translations (MM) | ||||
|---|---|---|---|---|---|---|
| Flexion/Extension | Varus/Valgus | External/Internal | Lateral/Medial | Anterior/Posterior | Distraction/Compression | |
| INTACT (n = 6) | 37 ± 1 | 1 ± 1 | 5 ± 3 | 2 ± 1 | 3 ± 1 | 1 ± 1 |
| IMPLANTED (n = 5) | 37 ± 1 | 3 ± 2 | 11 ± 4 | 2 ± 1 | 2 ± 1 | 3 ± 1 |
| p-value | 0.91 | 0.28 | 0.07 | 0.68 | 0.68 | 0.008a |
The average range of motion (±standard deviation) of the intact and implanted groups in the gait simulator for the 6 DOF of the ovine stifle joint.
Shows statistical significance between groups.
FIGURE 4

Average range of motion (± standard deviation) of the intact (grey, n = 6) and implanted (blue, n = 5) groups as measured on the ovine gait simulator.
3.2 Survivability of native menisci and cartilage
No tissue damage or degeneration was observed on the medial and lateral menisci of tested joints with one exception. Slight degeneration was observed on the central inferior part of the medial meniscus of one right stifle. Comparable tissue damage was observed in the same location on the medial meniscus of the contralateral untested joint. No anomalous signs of cartilage damage were recorded on the tested joints in comparison to the untested contralateral joints following ink-staining.
3.3 Contact pressures
Peak contact pressures recorded in intact stifles were 3.6 ± 1.0 MPa and 6.0 ± 2.1 MPa in the medial and lateral condyles respectively (p < 0.05). Peak pressures were generally located on the posterior aspect of the medial condyle covered by the meniscus and distributed across the central aspect of the lateral condyle (Figure 5A).
FIGURE 5

Representative peak contact pressure maps of the medial and lateral condyles for (A) intact joints and (B) implanted joints. Measured at 55° flexion under a 980N load in intact ovine stifle joints. The solid black line shows the approximate edge of the tibial plateau in each compartment of the knee.
The measured static medial and lateral peak pressures of the intact group did not differ significantly from Fischenich et al. (2018) at 113 kg joint load at both 45° and 60° flexion and Heckelsmiller et al. (2017) (p > 0.4). Recorded medial peak pressures were significantly lower than those reported by Lee-Shee et al. (2007) (p < 0.05).
Following implantation in the medial compartment, the peak contact pressures were 4.3 ± 2.2 MPa and 9.4 ± 0.8 MPa in the medial and lateral condyles respectively, as shown in Figure 5B. Peak pressures were higher on the lateral condyle (p < 0.01). In the implanted ovine stifles, contact pressures were typically located on the central aspect of the medial condyle at the cartilage-to-cartilage contact area and distributed across the central aspect of the lateral condyle. Medial peak pressures were not significantly different between the implanted and intact groups (p > 0.4), while lateral peak pressures were significantly higher in the implanted group (p < 0.01).
4 Discussion
The most important findings of this study were that the novel ovine gait simulator led to peak contact pressures and kinematics of intact stifle joints that did not differ significantly from those reported in literature. The fiber-matrix reinforced PVA-PEG hydrogel medial meniscal replacement restored the medial peak contact pressures, but not the lateral contact pressures, thus demonstrating an example of the potential use of the simulator.
Previous research studies that performed fatigue testing of cadaveric ovine stifles for meniscus evaluation were limited to low axial contact loads, thus their results could not be compared to this study (
The peak contact pressures were of primary investigative interest for this study to confirm physiologic cartilage loading on each of the condyles. The static medial and lateral peak pressures in the intact group did not differ significantly from published literature (
Research studies have evaluated the peak contact pressures of meniscus replacement or scaffold implanted ovine stifle joints in comparison to the intact, partially or fully meniscectomized conditions (
Medial peak pressures were not significantly different between the implanted and intact groups tested in this work (p > 0.4). The medial peak pressures of the implanted group were also not significantly different from the Fischenich et al. intact group (p > 0.1) and their TPE hydrogel implanted group (p > 0.08) (Figure 6). The medial peak pressures of the implanted group in the present study were significantly lower than the Fischenich et al. meniscectomized group (p < 0.03) and not significantly different from the Fischenich et al. allograft group (p > 0.09). These findings show that the fiber-matrix reinforced PVA-PEG hydrogel meniscal replacement restored the medial peak contact pressures.
FIGURE 6

Mean ± standard deviation peak contact pressures (MPa) in the medial and lateral compartments of intact (circle), meniscectomized (downward arrow), allograft (square) and implanted (upright arrow) ovine stifle joints as measured in the current work (black) and in literature (blue). All
Similarly to the intact group, the peak contact pressures of the implanted group were significantly higher in the lateral compartment (p < 0.01) following implantation in the medial compartment. This may have resulted from a change of limb alignment or joint congruity caused by the implantation procedure. In spite of this, the peak lateral contact pressures of the PVA-PEG hydrogel implanted group were not significantly different from the Fischenich et al. allograft group (p > 0.1) and TPE hydrogel implanted group (p > 0.08).
At 55° flexion, representing the stance phase of the gait cycle, peak contact pressures of the intact group were generally located on the posterior aspect of the medial tibial plateau, whereas the lateral peak contact pressures were located along the midline of the joint. These results matched those of previous studies (
The kinematics of the ovine stifle in this study did not differ significantly from those reported in vivo (
This study has limitations: firstly, tests were performed in a short timescale, under a consistent pattern of gait and with continuous motion, and did not fully simulate in vivo conditions, including dynamic muscle function and natural lubrication of a synovial joint. Joint simulators normally apply a consistent loading/kinematic profile to simulate normal gait to test wear of joint replacements. Variations in the pattern are not usually imposed by the control system. However, the novel ovine gait simulator uses native joints rather than prostheses, and they have inherent natural variability between specimens. The simulator enables the natural kinematics of the stifle joint because the kinematics are partly controlled by the ligaments, introducing tibial rotation and AP translations, so that the articulating path is not a linear one but also includes shear. Thus, there are inter-specimen variations in the gait pattern, but not simulation of different activities. Secondly, the ovine simulator operated at refrigeration temperatures of 4°C–8°C, which are significantly lower than body temperature (39°C) and could possibly affect friction and wear during joint articulation (
The current feasibility study of the use of the ovine gait simulator to evaluate meniscus replacements was limited by a relatively small sample size (n = 5/6). Variability in the stifle joint anatomy, surgical procedure and stifle joint alignment in the ovine gait simulator (
This study found that peak contact pressures and kinematics of intact stifle joints in the ovine gait simulator were comparable to literature, thus providing support for its efficacy as a tool to evaluate novel meniscus or other surgical procedures. The pilot study with the novel fiber-matrix reinforced meniscal replacement implant found that it restored medial peak contact pressures and native joint kinematics but did not restore lateral peak contact pressures. These findings support the use of an ovine gait simulator as a screening tool prior to evaluation of novel procedures in live animals.
Statements
Data availability statement
The original contributions presented in the study are included in the article/Supplementary Material, further inquiries can be directed to the corresponding author.
Author contributions
MB: Conceptualization, Data curation, Formal Analysis, Investigation, Methodology, Software, Validation, Visualization, Writing–original draft, Writing–review and editing. SN: Investigation, Methodology, Writing–review and editing. OD: Data curation, Software, Writing–review and editing. CH: Methodology, Writing–review and editing. MA: Conceptualization, Funding acquisition, Methodology, Writing–review and editing. DD: Conceptualization, Supervision, Writing–review and editing. AA: Conceptualization, Methodology, Project administration, Resources, Supervision, Writing–review and editing.
Funding
The author(s) declare that financial support was received for the research, authorship, and/or publication of this article. This work was supported by a grant paid to Imperial College from Orthonika Ltd.
Acknowledgments
The authors thank Elena Provaggi and Hazimah Mahmud for their assistance. Manuscript content has appeared in part at the ORS Annual Meeting (
Conflict of interest
MB is an employee of and owns stock or stock options in Orthonika Ltd. MA is a director of Orthonika Ltd. AA, DD and SN own stock or stock options in Orthonika Ltd. OD is now an employee of and owns stock or stock options in Orthonika Ltd., but had no conflicts during the period in which this research was carried out.
The remaining author declares that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
The authors declare that this study received funding from Orthonika Ltd. The funder had the following additional involvement in the study: discussion of work programme and results, review of the manuscript.
The author(s) declared that they were an editorial board member of Frontiers, at the time of submission. This had no impact on the peer review process and the final decision.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
Abbreviations
BCS, Bovine Calf Serum; BW, Body Weight; PMMA, Polymethyl methacrylate; PVA-PEG, Polyvinyl alcohol-polyethylene glycol; TPE, Thermoplastic elastomer.
References
1
AbramS. G. F.JudgeA.BeardD. J.PriceA. J. (2018). Adverse outcomes after arthroscopic partial meniscectomy: a study of 700 000 procedures in the national hospital episode statistics database for England. Lancet392 (10160), 2194–2202. 10.1016/S0140-6736(18)31771-9
2
AgostinhoF. S.RahalS. C.AraújoF. A. P.ConceiçãoR. T.HussniC. A.El-WarrakA. O.et al (2012). Gait analysis in clinically healthy sheep from three different age groups using a pressure-sensitive walkway. BMC Veterinary Res.8 (1), 87. 10.1186/1746-6148-8-87
3
AllenM. J.HoultonJ. E. F.AdamsS. B.NeilR. (1998). The surgical anatomy of the stifle joint in sheep. Veterinary Surg.27 (6), 596–605. 10.1111/j.1532-950X.1998.tb00536.x
4
ASTM International (2017). Astm F732-17: standard test method for wear testing of polymeric materials used in total joint. West Conshohocken, PA: ASTM International. 10.1520/F0732-17
5
BartoloM. K.NewmanS.DandridgeOProvaggiE.AccardiM. A.DiniD.et al (2023). Ovine knee kinematics and contact pressures of A novel fibre matrix-reinforced hydrogel total meniscus replacement. Orthopaedic Proceedings105, 14. 10.1302/1358-992x.2023.9.014
6
BartoloM. K.NewmanS.DandridgeO.ProvaggiE.BraithwaiteG.AccardiM. A.et al (2021). Ovine knee kinematics and contact pressures measured in A gait simulator for meniscus replacement evaluation. ORS Annu. Meet.46, 1657–LB.
7
BartoloM. K.ProvaggiE.AthwalK. K.NewmanS.AccardiM. A.DiniD.et al (2022). Strength of interference screw fixation of meniscus prosthesis matches native meniscus attachments. Knee Surg. Sports Traumatol. Arthrosc.30 (7), 2259–2266. 10.1007/s00167-021-06772-9
8
BediA.KellyN. H.BaadM.FoxA. J. S.BrophyR. H.WarrenR. F.et al (2010). Dynamic contact mechanics of the medial meniscus as a function of radial tear, repair, and partial meniscectomy. J. Bone Jt. Surg. - Ser. A92 (6), 1398–1408. 10.2106/JBJS.I.00539
9
BortelE.CharbonnierB.HeubergerR. (2015). Development of a synthetic synovial fluid for tribological testing. Lubricants3 (4), 664–686. 10.3390/lubricants3040664
10
BrandR. A. (2005). Joint contact stress: a reasonable surrogate for biological processes?Iowa Orthop. J.25, 82–94.
11
BrophyR. H.CottrellJ.RodeoS. A.WrightT. M.WarrenR. F.MaherS. A. (2010). Implantation of a synthetic meniscal scaffold improves joint contact mechanics in a partial meniscectomy cadaver model. J. Biomed. Mater. Res. - Part A92 (3), 1154–1161. 10.1002/jbm.a.32384
12
BrzezinskiA.GhodbaneS. A.PatelJ. M.PerryB. A.GattC. J.DunnM. G. (2017). <sup/>The ovine model for meniscus tissue engineering: considerations of anatomy, function, implantation, and evaluation. Tissue Eng. Part C. Methods23 (12), 829–841. 10.1089/ten.TEC.2017.0192
13
BSI Standards Limited (2014). BS ISO 14243-3:2014 implants for surgery — wear of total knee-joint prostheses. Part 3: loading and displacement parameters for wear-testing machines with displacement control and corresponding environmental conditions for test. London, United Kingdom: BSI Standards Limited.
14
CottrellJ. M.PaulS.WanichT.WarrenR. F.WrightT. M.MaherS. A. (2008). A new technique to measure the dynamic contact pressures on the tibial plateau. J. Biomechanics41 (10), 2324–2329. 10.1016/j.jbiomech.2008.04.024
15
DesJardinsJ. D.BanksS. A.BensonL. C.PaceT.LaBergeM. (2007). A direct comparison of patient and force-controlled simulator total knee replacement kinematics. J. Biomechanics40 (15), 3458–3466. 10.1016/j.jbiomech.2007.05.022
16
EnglundM.LohmanderL. S. (2006). Meniscectomy and osteoarthritis: what is the cause and what is the effect?Future Rheumatol.1 (2), 207–215. 10.2217/17460816.1.2.207
17
FischenichK. M.PaulyH. M.LewisJ. T.BaileyT. S.Haut DonahueT. L. (2018). A hydrogel meniscal replacement: knee joint pressure and distribution in an ovine model compared to native tissue. Ann. Biomed. Eng.46 (11), 1785–1796. 10.1007/s10439-018-2069-8
18
GregoryM. H.CapitoN.KurokiK.StokerA. M.CookJ. L.ShermanS. L. (2012). A review of translational animal models for knee osteoarthritis. Arthritis2012, 1–14. 10.1155/2012/764621
19
GroodE. S.SuntayW. J. (1983). A joint coordinate system for the clinical description of three-dimensional motions: application to the knee. J. Biomechanical Eng.105 (2), 136–144. 10.1115/1.3138397
20
HeckelsmillerD. J.James RudertM.BaerT. E.PedersenD. R.FredericksD. C.GoetzJ. E. (2017). Changes in joint contact mechanics in a large quadrupedal animal model after partial meniscectomy and a focal cartilage injury. J. Biomechanical Eng.139 (5), 0545011–0545015. 10.1115/1.4036148
21
HerregodtsS.De BaetsP.VictorJ.VerstraeteM. (2015). Use of tekscan pressure sensors for measuring contact pressures in the human knee joint. Int. J. Sustain. Constr. Des.6 (2), 7. 10.21825/scad.v6i2.1123
22
HollowayJ. L. (2012). Development and characterization of UHMWPE fiber-reinforced hydrogels for meniscal replacement. Philadelphia, PA: Drexel University.
23
HoutemM. vanCloughR.KhanA.HarrisonM.BlunnG. W. (2006). Validation of the soft tissue restraints in a force-controlled knee simulator. Proc. Institution Mech. Eng. Part H J. Eng. Med.220 (3), 449–456. 10.1243/09544119JEIM57
24
HutchinsonI. D.MoranC. J.PotterH. G.WarrenR. F.RodeoS. A. (2013). Restoration of the meniscus: form and function. Am. J. Sports Med.42 (4), 987–998. 10.1177/0363546513498503
25
KellyB. T.RobertsonW.PotterH. G.XiangH. D.Simon TurnerA.LymanS.et al (2007). Hydrogel meniscal replacement in the sheep knee: preliminary evaluation of chondroprotective effects. Am. J. Sports Med.35 (1), 43–52. 10.1177/0363546506292848
26
LeeD. C.ByunS. J. (2012) “High tibial osteotomy,” in Knee surgery and related research. BioMed Central Ltd. 10.5792/ksrr.2012.24.2.61
27
Lee-SheeN. K.DickeyJ. P.HurtigM. B.HurtigB. (2007). Contact mechanics of the ovine stifle during simulated early stance in gait: an in vitro study using robotics. Veterinary Comp. Orthop. Traumatology20 (1), 70–72. 10.1055/s-0037-1616591
28
MaherS. A.RodeoS. A.PotterH. G.BonassarL. J.WrightT. M.WarrenR. F. (2011). A pre-clinical test platform for the functional evaluation of scaffolds for musculoskeletal defects: the meniscus. Musculoskelet. J. Hosp. Special Surg.7 (2), 157–163. 10.1007/s11420-010-9188-6
29
McDermottI. D.MasourosS. D.AmisA. A. (2008). Biomechanics of the menisci of the knee. Curr. Orthop.22 (3), 193–201. 10.1016/j.cuor.2008.04.005
30
MeachimG. (1972). Light microscopy of Indian ink preparations of fibrillated cartilage. Ann. Rheumatic Dis.31 (6), 457–464. 10.1136/ard.31.6.457
31
NovarettiJ. V.PatelN. K.LianJ.VaswaniR.de SaD.GetgoodA.et al (2019). Long-term survival analysis and outcomes of meniscal allograft transplantation with minimum 10-year follow-up: a systematic review. Arthroscopy35 (2), 659–667. 10.1016/j.arthro.2018.08.031
32
PatelJ. M.MerriamA. R.CulpB. M.GattC. J.DunnM. G. (2015). One-year outcomes of total meniscus reconstruction using a novel fiber-reinforced scaffold in an ovine model. Am. J. Sports Med.44 (4), 898–907. 10.1177/0363546515624913
33
RadfordW. J. P.AmisA. A.SteadA. C. (1996). The ovine stifle as a model for human cruciate ligament surgery. Veterinary Comp. Orthop. Traumatology09 (03), 134–139. 10.1055/s-0038-1632518
34
RuckstuhlK. E. (1998). Foraging behaviour and sexual segregation in bighorn sheep. Anim. Behav.56 (1), 99–106. 10.1006/anbe.1998.0745
35
SAFC Supply Solutions (2008). ProClin ® preservatives for in vitro diagnostic materials: efficacy vs. Thimerosal and sodium azide.
36
SchmitzN.LavertyS.KrausV. B.AignerT. (2010). Basic methods in histopathology of joint tissues. Osteoarthr. Cartil.18 (Suppl. 3), S113–S116. 10.1016/j.joca.2010.05.026
37
StoneK. R.Richard SteadmanJ.RodkeyW. G.Tung LiS. (1997). Regeneration of meniscal cartilage with use of a collagen scaffold. Analysis of preliminary data. J. Bone Jt. Surg. - Am.79 (12 79A), 1770–1777. 10.2106/00004623-199712000-00002
38
TapperJ. E.FukushimaS.AzumaH.ThorntonG. M.RonskyJ. L.ShriveN. G.et al (2006). Dynamic in vivo kinematics of the intact ovine stifle joint. J. Orthop. Res.24 (4), 782–792. 10.1002/jor.20051
39
TapperJ. E.PowersM. J.RonskyJ. L.CraigS.TokifumiM.FrankC. B.et al (2004). In vivo measurement of the dynamic 3-D kinematics of the ovine stifle joint. J. Biomechanical Eng.126, 301–305. 10.1115/1.1695576
40
TaylorW. R.EhrigR. M.HellerM. O.HannaS.SeebeckP.DudaG. N. (2006). Tibio-femoral joint contact forces in sheep. J. Biomechanics39 (5), 791–798. 10.1016/j.jbiomech.2005.02.006
41
TaylorW. R.PoepplauB. M.KönigC.EhrigR. M.ZachowS.DudaG. N.et al (2011). The medial-lateral force distribution in the ovine stifle joint during walking. J. Orthop. Res.29 (4), 567–571. 10.1002/jor.21254
42
VerdonkR.VerdonkP.HuysseW.ForsythR.HeinrichsE. L. (2011). Tissue ingrowth after implantation of a novel, biodegradable polyurethane scaffold for treatment of partial meniscal lesions. Am. J. Sports Med.39 (4), 774–782. 10.1177/0363546511398040
43
VerdonkR.VolpiP.VerdonkP.Van Der BrachtH.Van LaerM.AlmqvistK. F.et al (2013). Indications and limits of meniscal allografts. Injury44 (S1), S21–S27. 10.1016/S0020-1383(13)70006-8
44
VranckenA. C. T.HanninkG.MadejW.VerdonschotN.Van TienenT. G.BumaP. (2017). In vivo performance of a novel, anatomically shaped, total meniscal prosthesis made of polycarbonate urethane: a 12-month evaluation in goats. Am. J. Sports Med.45, 2824–2834. 10.1177/0363546517713687
45
WinklerP. W.RothrauffB. B.BuerbaR. A.ShahN.ZaffagniniS.AlexanderP.et al (2020). Meniscal substitution, a developing and long-awaited demand. J. Exp. Orthop.7 (1), 55. 10.1186/s40634-020-00270-6
46
ZurG.Linder-GanzE.ElsnerJ. J.ShaniJ.BrennerO.AgarG.et al (2011). Chondroprotective effects of a polycarbonate-urethane meniscal implant: histopathological results in a sheep model. Knee Surg. Sports Traumatol. Arthrosc.19 (2), 255–263. 10.1007/s00167-010-1210-5
Summary
Keywords
knee, meniscus replacement, ovine gait simulator, kinematics, contact pressures
Citation
Bartolo MK, Newman S, Dandridge O, Halewood C, Accardi MA, Dini D and Amis AA (2024) An ovine knee simulator: description and proof of concept. Front. Bioeng. Biotechnol. 12:1410053. doi: 10.3389/fbioe.2024.1410053
Received
31 March 2024
Accepted
05 June 2024
Published
27 June 2024
Volume
12 - 2024
Edited by
Andreas Martin Seitz, Ulm University Medical Center, Germany
Reviewed by
Amy Lerner, University of Rochester, United States
Christoph Kittl, University Hospital Münster, Germany
Updates

Check for updates
Copyright
© 2024 Bartolo, Newman, Dandridge, Halewood, Accardi, Dini and Amis.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Andrew A. Amis, a.amis@imperial.ac.uk
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.