Abstract
Transcatheter tricuspid valve intervention (TTVI) has recently emerged as an alternative for the treatment of severe tricuspid regurgitation (TR). Multiple percutaneous devices have been developed in the last decade with promising early results. Among them, the coaptation devices are designed to reduce TR severity by valve leaflet plication or occupying the regurgitant orifice with a spacer. To date, the MitraClip/TriClip devices (Abbott, Santa Clara, CA, USA), the PASCAL system (Edwards Lifesciencies, Irvine, CA, USA), and the FORMA device (Edwards Lifesciencies, Irvine, CA, USA) have been used as coaptation devices for treating severe TR. The present document aimed to review the clinical evidence on coaptation devices in the field of TTVI, describing its design characteristics, main procedural steps, and early and mid-term outcomes.
Introduction
Tricuspid regurgitation (TR) has been associated with poorer clinical outcomes as heart failure and mortality in multiple cardiology settings (). Patients with chronic TR have been usually managed medically, leading to gradual annular dilatation and end-stage right ventricular (RV) heart failure. The etiology of TR is functional >80% of cases (), occurring frequently in patients with already treated left-side valvular disease. Furthermore, surgical TR replacement or repair has been associated with an increased operative risk, reaching in-hospital mortality rates of about 10% (). This higher risk is mainly related to the late referral to surgery of patients with enduring severe TR. The latter may indeed lead to an undertreated population with end-stage right heart failure, usually linked with hepatic congestion and/or late-stage kidney disease (, ). As a result, the management of isolated TR represents an unmet clinical need that has led to the development of transcatheter tricuspid valve intervention (TTVI) (). TTVI can be classified according to their mechanism of action as follows: annuloplasty devices, caval stent devices, tricuspid valve (TV) replacement, and coaptation devices. The initial experience with these first-generation transcatheter systems showed that despite the high-risk profile of TR patients undergoing TTVI, the majority of the procedures were well-tolerated with low in-hospital mortality rates (lower than surgical series) (). Among them, coaptation devices attempt to increase valve coaptation by leaflet plication or occupying the regurgitant orifice with a spacer. To date, the Mitraclip and Triclip devices (Abbott, Santa Clara, CA, USA), the PASCAL system (Edwards Lifesciencies, Irvine, CA, USA) and the FORMA device (Edwards Lifesciencies, Irvine, CA, USA) have been used as coaptation devices in the TTVI field (Figure 1) (, ). The present review document focuses on coaptation devices, describing its design characteristics, main procedural steps and the reported early and mid-term outcomes.
Figure 1
MitraClip/TriClip System
The off-label edge-to-edge repair technique with the MitraClip system has been the most common coaptation device used in the context of TTVI, mainly due to the extensive knowledge and experience acquired in mitral regurgitation. In a recent publication from the largest worldwide retrospective series of TR patients treated with TTVI (the TriValve registry; NCT03416166), the MitraClip device was used in 80% of patients (
Device Design and Procedural Planning
The MitraClip and TriClip systems consist of two parts: a steerable guide catheter (SGC) and the clip delivery system (CDS). This includes a steerable sleeve, a delivery catheter and the 4 mm (size NT) or 7 mm (size XTR) wide chrome-cobalt clip with two articulated arms to grasp and draw the valve leaflets. The 3-mm extended clip arms of the MitraClip XTR system can be useful in patients with larger coaptation gaps. The CDS are advanced through the conduit guide provided by the SGC in order to manipulate the implantable clip. In contrast to the standard MitraClip system, the SGC from the new TriClip design has two knobs for the steering maneuvers. In addition, the steerable sleeve has only one knob for tip deflection and its distal curve has a shorter radius. These iterations from the TriClip system facilitate bending and guiding maneuvers in the right atrium (RA).
Patient planning include transthoracic and transesophageal echocardiography (TEE). Also, a computed tomography may be performed to evaluate the venous system before the procedure. TEE is crucial (especially the transgastric basal short axis view) to establish the feasibility of the intervention. When opened to 120 degrees, the NTR and XTR clips have a width of 17 and 22 mm, respectively. The anatomical examination requires the evaluation of the leaflet length, the presence of tethering, and gaps (leaflet insertion should be located ≥6 and ≥9 mm for NTR and XTR, respectively). The leaflet gaps should be ≤7 mm. A standardized TEE protocol to help in the process of patient selection has been recently proposed (
Transgastric basal short-axis view (Figure 2A):
- The only 2D view that allows simultaneous visualization of all three leaflets and commissures. Commonly obtained between 25 and 50 degrees.
- Helpful to measure the tricuspid annulus size and assess leaflet motion and coaptation gaps.
- The location of vena contracta width and area along with the regurgitant jet can be visualized to assess the severity of TR.
- Useful for steering the clip intraprocedurally to the desired position within the annular plane and to rotate the clip device to achieve perpendicularity with the respective leaflet coaptation line.
Mid-esophageal TEE view (Figure 2B):
- A slow sweep along the septal leaflet allows a detailed assessment and measurement of coaptation defects and gaps as well as leaflet length, tethering, and restriction. A restricted leaflet movement, especially of the septal leaflet, frequently limits leaflet grasping.
3D-TEE imaging view (mid or lower esophageal position, Figure 2C):
- 3D imaging could be helpful for secondary assessment of both the clip perpendicularity to the coaptation line and the tissue bridge after grasping and clip closure.
- An orientation en face view identically to the transgastric view is recommended, which places the interventricular septum on the right side of the screen with the aorta located at a five o'clock position. This facilitates peri-procedural orientation and image interpretation.
Figure 2

TEE protocol to patient selection for MitraClip/TriClip procedures. (A) Transgastric basal short-axis view. ATL, anterior tricuspid leaflet; PPM, posterior papillary muscle; PTL, posterior tricuspid leaflet; RA, right atrium; RV, right ventricle; STL, septal tricuspid leaflet; TV, tricuspid valve. (B) Mid-esophageal TEE view. The left view is orientated along the anteroseptal coaptation line usually at 80–90°. The right image is obtained by a sweep from the aorta to the valve center. Ao, aorta; ATL, anterior tricuspid leaflet; RA, right atrium; RV, right ventricle; STL, septal tricuspid leaflet; TV, tricuspid valve. (C) 3D-TEE imaging view. Ao, aorta; ATL, anterior tricuspid leaflet; PTL, posterior tricuspid leaflet; STL, septal tricuspid leaflet. Reproduced with permission from Hausleiter et al. (
To date, a predominant central/anteroseptal jet and smaller TV coaptation gaps (cutoff of 7.2mm) have been identified as predictors of procedural success and greater TR tenting and effective regurgitant orifice area as predictors of procedural failure (
The edge-to-edge repair technique in the presence of a transtricuspid cardiac implantable electronic device (CIED) may represent a more challenging subset of patients due to potential interference with the MitraClip system and acoustic shadowing. While not contraindicated, careful evaluation is critical to estimate the feasibility of leaflet grasping and to exclude the CIED per se as the main etiologic factor involved in the TR mechanism.
Finally, the first-in-human report using the newer third-generation MitraClip G4 system (NT, XT, NTW, and XTW) have been published very recently (
Procedural Steps
The procedure is performed under general anesthesia with fluoroscopic and TEE guidance. Whereas the first series of cases were performed via the transjugular approach (
Two methods to reduce the TR grade with the MitraClip/TriClip system have been described. First, the triple-orifice technique (TOT), where clips are ideally placed centrally between the septal and anterior tricuspid leaflet and between the septal and posterior tricuspid leaflet. Alternatively, the bicuspidization technique (BT) can be used, where clips are placed between the septal and anterior tricuspid leaflets (
When using the standard MitraClip delivery system, the short distance between the inferior vena cava orifice to the atrial septum as well as the close proximity between the inferior vena cava and the tricuspid valve may difficult the steering of the MitraClip system vertical to the valve plane. To overcome this issue, the CDS is inserted into the SCG and then rotated (“mis-keyed”) by 90 degrees counter-clockwise (
Once the SCG is placed in the RA, the CDS is straddled in the SGC and placed inside the RA under fluoroscopy in left anterior oblique (LAO) projection (Figure 3) (
Figure 3

Intraprocedural main fluoroscopic views in transcatheter tricuspid valve intervention using the Mitraclip system. (A–D) Left anterior oblique projection. Once the CDS is straddled, turning the SGC clockwise steers away the MitraClip from the interatrial septum toward the TV. In that case, the position of the prior MitraClip placed at the TV can serve as a reference. (E,F) Right anterior oblique projection. Steering of the MitraClip toward the TV. The existing MitraClip is marked with white arrows. RAO, Right anterior oblique; LAO, Left anterior oblique. Reproduced with permission from Tang (
Both TOT and BT have also been described in the presence of a transtricuspid CIEDs (
Outcomes
The main procedural characteristics, early and mid-term outcomes in the 2 publications including the largest number of patients treated with the MitraClip/TriClip systems to date are summarized in Table 1 (
Table 1
| MitraClip/TriClip | PASCAL | FORMA | ||||
|---|---|---|---|---|---|---|
| Mehr et al. ( | Nickenig et al. ( | Fam et al. ( | Perlman et al. ( | Kodali et al. ( | ||
| Baseline characteristics | ||||||
| Age, years | 77 ± 9 | 78 ± 8 | 78 ± 6 | 76 ± 10 | 76 ± 8 | |
| Female, n (%) | 128 (51) | 56 (66) | 15 (54) | 13 (72) | 19 (66) | |
| Atrial fibrillation, n (%) | 183 (74) | 78 (92) | 26 (93) | 16 (89) | 24 (83) | |
| NYHA III-IV, n (%) | 238 (96) | 64 (75) | 28 (100) | 17 (94) | 25 (86) | |
| Transtricusid CIED, n (%) | 74 (30) | 12 (14) | 1 (3) | 3 (17) | n/a | |
| EuroSCORE II, % | 6.4 (4-14) | 8.6 ± 11 | 6.2 ± 5.2 | 9 ± 5.7 | 8.4 ± 5.3 | |
| Echocardiography | ||||||
| LVEF, % | 49 ± 14 | 59 ± 8 | 59 ± 6 | 59 ± 9 | 57 ± 12 | |
| TAPSE, mm | 16 ± 4 | 14 ± 0.3 | 16 ± 3 | 15 ± 5 | 14 ± 0.4 | |
| TR grade | TR 5+ (%) | n/a | 31/84 (37) | 12 (43) | 11 (61) | n/a |
| TR 4+ (%) | 129 (52) | 24/84 (29) | 9 (32) | 6 (33) | n/a | |
| TR 3+ (%) | 112 (45) | 24/84 (29) | 7 (25) | 1 (6) | n/a | |
| Maximum gap width, mm | 5.3 ± 3.3 | n/a | 6.9 ± 3 | n/a | n/a | |
| Vena contracta width, mm | 9.9 ± 4.1 | 17 ± 0.6 | 11.4 ± 5 | 12,1 ± 3.3 | 16 ± 0.5 | |
| EROA, cm2 | 0.70 ± 0.5 | 0.65 ± 0.3 | 1.3 ± 2.4 | 1 ± 0.6 | 2,2 ± 1.5 | |
| Mitral regurgitation ≥ moderate, n (%) | 108 (43) | 0 (0) | n/a | 4 (22) | n/a | |
| Procedural and 30-day outcomes | ||||||
| Procedural success, n (%) | 192 (77) | 76/84 (91) | 24 (86) | 16 (89) | 27 (93) | |
| Number of implants, n | 2 ± 1 | 2,2 ± 0.8 | 1,4 ± 0.6 | n/a | n/a | |
| Post-procedural tricuspid gradient, mmHg | 2.4 ± 1.5 | 2.12 ± 1.2* | 1.6 ± 1 | n/a | n/a | |
| Procedural Procedural-related Death, n (%) | 0 (0) | 0 (0) | 0 (0) | 0 (0) | 2 (7) | |
| Need for cardiac Surgery, n (%) | 1 (0.4) | 0 (0) | 0 (0) | 1 (6) | 3 (10) | |
| 30-day TR grade > moderate, n (%) | n/a | 36/83 (43) | 4/26 (15) | 7/16 (44) | n/a | |
| 30-day mortality, n (%) | n/a | 0 (0) | 2 (7) | 0 (0) | 2 (7) | |
| Follow-up | ||||||
| Follow-up time | 290 days | 6 months | 30 days | 12 months | 30 days | |
| Mortality, n (%) | 44 (19) | 4/84 (5) | 2 (7) | 0 (0) | N/A | |
| TR grade > moderate, n (%) | 46 (28) | 30/70 (43) | 4/26 (15) | 12/13 /92) | 2 (7) | |
| NYHA I-II, n (%) | 121 (69) | 63/73 (86) | 23/26 (88) | 11/14 (79) | 18/25 (72) | |
Clinical, procedural, and follow-up characteristics from the main studies using coaptation devices.
Values are mean SD or n/N (%).
At 30 days. CIED, Cardiac implantable electronic device; EROA, effective regurgitant orifice area; EuroSCORE, European System for Cardiac Operative Risk Evaluation; LVEF, left ventricular ejection fraction; NYHA, New York Heart Association; TAPSE, Tricuspid annular plane systolic excursion; TR, Tricuspid regurgitation.
Mehr et al. reported the results of 249 patients that underwent edge-to-edge repair with the MitraClip system in the TriValve registry (
The 6-month outcomes of the TRILUMINATE study have been recently published (
At 6-month follow-up, all patients but 5 survived (95%), and the clinical assessment showed a significant improvement in functional status, with most patients (86%) in NYHA class I-II (p < 0.001 vs. baseline). Also, there was a significant increase in exercise capacity, and the mean 6-min walking test (6MWT) distance increased from 278 to 340 meters (p < 0.001). In light of the results of the TRILUMINATE study, the TriClip system received the CE mark approbation for minimally invasive tricuspid valve repair.
Whereas the two aforementioned studies showed the feasibility and preliminary efficacy of the MitraClip/TriaClip systems, data regarding long-term improvement in heart failure symptoms and rehospitalization remain scarce. Orban et al recently determined the effect of transcatheter edge-to-edge repair (93% MitraClip, 7% Pascal) on the rate of heart failure hospitalization compared with the 1-year period before the intervention (
Data regarding the outcomes of the extended clip arms from the MitraClip XTR system for TR reduction are limited. Braun et al. recently reported the acute and short-term outcomes of 31 patients treated with transcatheter edge-to-edge repair using the XTR system at 2 centers (
Finally, in the only study using the MitraClip and comparing specifically patients with and without the presence of a transtricuspid CIED (a small series of 42 MitraClip recipients, 13 of them with CIEDs), no significant differences in procedural and early outcomes were found between groups (
PASCAL System
The development of the Edwards Pascal transcatheter mitral valve repair (TMVr) system brought a new alternative to edge-to-edge TR treatment (
Device Design and Procedural Planning
The PASCAL TMVr consists of a 10 mm central nitinol woven spacer that acts as a filler in the regurgitant orifice of the atrioventricular valve and is attached to the valve leaflets by two paddles and clasps. The spring-loaded paddles (25 mm width in grasping position) and clasps (10 mm length) are wider than the MitraClip system. The clasps can be operated simultaneously or independently, which can facilitate the capture of the leaflets. The system consists of a 22 French steerable guide sheath, a steerable catheter, and an implant catheter, which contains the implant at the distal end. The handle of the implant catheter controls the deployment of the device and allows the independent movement of the clasps. The device is repositionable and recapturable.
Pre-procedural planning is similar to that used for MitraClip evaluation. However, the bigger size of the device may be a disadvantage in patients with less dilated tricuspid annulus and smaller coaptation gaps. Although the device can be elongated to achieve a narrow profile that could reduce the risk of entanglement, a careful evaluation regarding the subvalvular apparatus (papillary muscles, trabeculations, or prominent moderator band) might be needed to ensure successful device navigation (especially in non-dilated ventricles).
Procedural Steps
The TTVI with the edge-to-edge technique using the PASCAL system is currently in its early stages. The framework does not vary compared to the MitraClip/TriClip device, with the procedure performed under general anesthesia with fluoroscopic and TEE guidance.
First, large-bore conventional percutaneous venous access is obtained in the right femoral vein to introduce the 22 French guide sheath. To date, all reported cases using the Pascal system in the context of TTVI have been performed through the femoral approach (
Outcomes
The main baseline characteristics, procedural results, and early outcomes from the recently published first-in-human experience are summarized in Table 1 (
While waiting for future randomized studies, Braun et al recently reported the first comparison between the two currently available edge-to-edge systems (
FORMA
The FORMA repair system aimed to improve the tricuspid valve coaptation by occupying the regurgitant orifice with a spacer, which provided a surface for native leaflets coaptation to reduce the orifice area. The device consisted of a spacer that was advanced through a rail anchored at the septal portion of the right ventricle (RV) apex (Figure 1). Of note, the FORMA system is not currently available for medical use.
Device Design and Procedural Planning
The FORMA spacer device is a foam-filled polymer balloon that passively expands via holes in the spacer shaft. A steerable delivery catheter was used to deliver the rail system to the ideal location in the RV apex. The presence of a CIED did not contraindicate the procedure.
Procedural Steps
Following left axillary or subclavian vein access, a large-bore sheath (20 French for the 12-mm device, 24 French for the 15 and 18-mm devices) was advanced to the left innominate vein/superior vena cava junction. After right ventriculography, the rail was anchored to the right ventricle and the spacer was advanced over the rail and placed within the tricuspid valve. Finally, the device was locked proximally, with the excess rail length placed in a subcutaneous pocket. The main steps of the procedure are depicted in Figure 4 (
Figure 4

Intraprocedural fluoroscopic images in transcatheter tricuspid valve intervention using the FORMA system. (A) Right ventriculography to locate the tricuspid annular plane and identify the ideal anchor location (red asterisk) on fluoroscopy. (B) Right ventricular anchoring via a steerable delivery catheter. (C) Device positioning in the tricuspid annular plane. (D) Final right ventriculography showing the device in correct position (red arrow) and reduction of tricuspid regurgitation from baseline. Reproduced with permission from Campelo-Parada et al. (
Outcomes
Table 1 summarizes the main available data regarding the FORMA device (
In the multicenter international study (
In the EFS trial (
The echocardiographic evaluation at 30 days showed a relevant reduction in TR severity in the 2 aforementioned studies. In the EFS trial, the Core Lab assessment showed significant reductions in TR grade in the 25 available patients, with a quantitative effective regurgitation orifice area reduced by 49% (from 2.1 ± 1.8 cm2 at baseline to 1.1 ± 0.9 cm2 at 30-days, p = 0.012). In the compassionate clinical use cohort, the echocardiographic evaluation showed a reduction in the grade of TR in all but one patient. However, the effective regurgitation orifice area and vena contracta assessments did not improve between 30 days and 1 year (
Finally, Asmarats et al. reported the long-term outcomes from the multicenter international cohort (
Conclusions
In conclusion, TTVI using coaptation devices is a promising technique that has demonstrated procedural feasibility, early safety, and satisfactory mid-term outcomes. However, some issues as patient and device selection, optimization of intraprocedural results, and long-term data should be addressed in future studies. If the preliminary efficacy data are confirmed in further studies, this therapy will represent a paradigm shift in the management of chronic TR.
Statements
Author contributions
All authors listed have made a substantial, direct and intellectual contribution to the work, and approved it for publication.
Acknowledgments
GM-C and AA were supported by a research grant from the Fundación Alfonso Martín Escudero (Madrid, Spain). JR-C holds the Research Chair Fondation Famille Jacques Larivière for the Development of Structural Heart Disease Interventions.
Conflict of interest
JR-C has received institutional research grants from Medtronic and Edwards Lifesciences. The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest. The reviewer FP declared a past co-authorship with one of the authors GM-C to the handling editor.
- 6MWT
6-minute walking test
- BT
Bicuspidization technique
- CIED
Cardiac implantable electronic device
- CDS
Clip delivery system
- EFS trial
early feasibility trial
- EuroSCORE
European System for Cardiac Operative Risk Evaluation
- LAO
Left anterior oblique
- IQR
Interquartile range
- KCCQ
Kansas City Cardiomyopathy Questionnaire
- NYHA
New York Heart Association
- RA
right atrium
- RAO
Right anterior oblique
- RV
Right ventricular
- SLDA
Single leaflet device attachment
- SGC
Steerable guide catheter
- TMVr
Transcatheter mitral valve repair
- TOT
Triple-orifice technique
- TR
Tricuspid regurgitation
- TTVI
Transcatheter tricuspid valve intervention
- TV
Tricuspid valve.
Abbreviations
References
1.
AsmaratsLTaramassoMRodés-CabauJ. Tricuspid valve disease: diagnosis, prognosis and management of a rapidly evolving field. Nat Rev Cardiol. (2019) 16:538–54. 10.1038/s41569-019-0186-1
2.
AlqahtaniFBerzingiCOAljohaniSHijaziMAl-HallakAAlkhouliM. Contemporary trends in the use and outcomes of surgical treatment of tricuspid regurgitation. J Am Heart Assoc. (2017) 6:e007597. 10.1161/JAHA.117.007597
3.
LauGTTanHCKritharidesL. Type of liver dysfunction in heart failure and its relation to the severity of tricuspid regurgitation. Am J Cardiol. (2002) 90:1405–9. 10.1016/S0002-9149(02)02886-2
4.
MaederMTHolstDPKayeDM. Tricuspid regurgitation contributes to renal dysfunction in patients with heart failure. J Cardiac Fail. (2008) 14:824–30. 10.1016/j.cardfail.2008.07.236
5.
AsmaratsLPuriRLatibANaviaJLRodés-CabauJ. Transcatheter tricuspid valve interventions: landscape, challenges, and future directions. J Am Coll Cardiol. (2018) 71:2935–56. 10.1016/j.jacc.2018.04.031
6.
FamNPBraunDvon BardelebenRSNabauerMRufTConnellyKAet al. Compassionate use of the PASCAL transcatheter valve repair system for severe tricus-pid regurgitation: a multicenter, observational, first-in-human experience. JACC Cardiovasc Interv. (2019) 12:2488–95. 10.1016/j.jcin.2019.09.046
7.
TaramassoMGavazzoniMPozzoliAAlessandriniHLatibAAttinger-TollerAet al. Outcomes of TTVI in patients with pacemaker or defibrillator leads: data from the trivalve registry. JACC Cardiovasc Interv. (2020) 13:554–64. 10.1016/j.jcin.2019.10.058
8.
NickenigGWeberMLurzPBardelebenRS vonSitgesMSorajjaPet al. Transcatheter edge-to-edge repair for reduction of tricuspid regurgitation: 6-month out-comes of the TRILUMINATE single-arm study. Lancet. (2019) 394:2002–11. 10.1016/S0140-6736(19)32600-5
9.
HausleiterJBraunDOrbanMLatibALurzPBoekstegersPet al. Patient selection, echocardiographic screening and treatment strategies for interven-tional tricuspid repair using the edge-to-edge repair technique. EuroIntervention. (2018) 14:645–53. 10.4244/EIJ-D-17-01136
10.
LurzPBeslerCNoackTFornerAFBevilacquaCSeeburgerJet al. Transcatheter treatment of tricuspid regurgitation using edge-to-edge repair: procedural results, clinical implications and predictors of success. EuroIntervention. (2018) 14:e290–7. 10.4244/EIJ-D-17-01091
11.
BeslerCOrbanMRommelK-PBraunDPatelMHaglCet al. Predictors of procedural and clinical outcomes in patients with symptomatic tricuspid regurgita-tion undergoing transcatheter edge-to-edge repair. JACC Cardiovasc Interv. (2018) 11:1119–28. 10.1016/j.jcin.2018.05.002
12.
MehrMTaramassoMBeslerCRufTConnellyKAWeberMYzeirajESchiaviDMangieriAVaskelyteLet al. 1-year outcomes after edge-to-edge valve repair for symptomatic tricus-pid regurgitation: results from the TriValve registry. JACC Cardiovasc Interv. (2019) 12:1451–61. 10.1016/j.jcin.2019.04.019
13.
SorajjaPCavalcanteJLGosslMBaeR. Transcatheter repair of tricuspid regurgitation with MitraClip. Prog Cardiovasc Dis. (2019) 62:488–92. 10.1016/j.pcad.2019.10.001
14.
RaghunathanDGarcia-SayanESchechterMNapierkowskiSDhobleASmallingR. First-in-human report of mitraclip G4 implantation for torrential tricuspid regurgitation and severe secondary mitral regurgitation. JACC Cardiovasc Interv. (2020) 13:1599–602. 10.1002/ccd.28777
15.
HammerstinglCSchuelerRMalasaMWernerNNickenigG. Transcatheter treatment of severe tri-cuspid regurgitation with the MitraClip system. Eur Heart J. (2016) 37:849–53. 10.1093/eurheartj/ehv710
16.
KowalskiMFranzNRitterFHofmannSStabel-MahassineCWarneckeHet al. Simultaneous transfemoral transcatheter mitral and tricuspid valve edge-to-edge repair (using MitraClip system) completed by atrial septal defect occlusion in a surgically inoperable patient. first-in-human report. Kardiochir Torako-chirurgia Pol. (2015) 12:295–7. 10.5114/kitp.2015.56776
17.
SchoferJTiburtiusCHammerstinglCDickhautP-OWittJHansenLet al. Transfemoral tricuspid valve repair using a percutaneous mitral valve repair system. J Am Coll Cardiol. (2016) 67:889–90. 10.1016/j.jacc.2015.11.047
18.
BraunDOrbanMOrbanMHaglCMassbergSNabauerMet al. Transcatheter edge-to-edge repair for severe tricuspid regurgitation using the triple-orifice technique versus the bicuspidalization technique. JACC Cardiovasc Interv. (2018) 11:1790–2. 10.1016/j.jcin.2018.05.049
19.
VismaraRGelpiGPrabhuSRomitelliPTroxlerLGManginiAet al. Transcatheter edge-to-edge treatment of functional tricuspid regurgitation in an ex vivo pulsatile heart model. J Am Coll Cardiol. (2016) 68:1024–33. 10.1016/j.jacc.2016.06.022
20.
BraunDNabauerMOrbanMOrbanMGrossLEnglmaierAet al. Transcatheter treatment of severe tricuspid regurgitation using the edge-to-edge repair technique. EuroIntervention. (2017) 12:e1837–44. 10.4244/EIJ-D-16-00949
21.
TangGHL. Tricuspid clip: step-by-step and clinical data. Interv Cardiol Clin. (2018) 7:37–45. 10.1016/j.iccl.2017.09.001
22.
PozzoliATaramassoMZuberMMaisanoF. Transcatheter tricuspid valve repair with the MitraClip system using intracardiac echocardiography: proof of concept. EuroIntervention. (2017) 13:e1452–3. 10.4244/EIJ-D-17-00360
23.
LiC-HMillanXCapelladesHDanduchLSerraAArzamendiD. Multimodality and fusion im-aging in percutaneous therapy of tricuspid regurgitation. Rev Esp Cardiol. (2019) 72:421–2. 10.1016/j.rec.2018.04.027
24.
BraunDOrbanMNabauerMOrbanMGrossLEnglmaierAet al. Transcatheter treatment of severe tricuspid regurgitation using the edge-to-edge repair tech-nique in the presence and absence of pacemaker leads. JACC Cardiovasc Interv. (2017) 10:2014–6. 10.1016/j.jcin.2017.07.010
25.
TangGHLKapleRCohenMDuttaTUndemirCAhmadHet al. First percutaneous Micra leadless pacemaker implantation and tricuspid valve repair with MitraClip NT for lead-associated severe tricuspid regurgitation. EuroIntervention. (2017) 12:e1845–8. 10.4244/EIJ-D-17-00023
26.
HahnRTZamoranoJL. The need for a new tricuspid regurgitation grading scheme. Eur Heart J Cardiovasc Imaging. (2017) 18:1342–3. 10.1093/ehjci/jex139
27.
PerlmanGPrazFPuriROfekHYeJPhilipponFet al. Transcatheter tricuspid valve repair with a new transcatheter coaptation system for the treatment of severe tricuspid regurgitation: 1-year clinical and echocardiographic results. JACC Cardiovasc Interv. (2017) 10:1994–2003. 10.1016/j.jcin.2017.06.036
28.
KodaliS. The forma early feasibility study: 30-day outcomes of transcatheter tricuspid valve therapy in patients with severe secondary tricuspid regurgitation. In: Presented at: Transcatheter Cardiovascular Therapeutics. Denver, CO (2017).
29.
OrbanMRommelK-PHoECUnterhuberMPozzoliAConnellyKAet al. Transcatheter edge-to-edge tricuspid repair for severe tricuspid regurgitation reduces hospitalizations for heart failure. JACC Heart Fail. (2020) 8:265–76. 10.1016/j.jchf.2019.12.006
30.
BraunDRommelK-POrbanMKaramNBrinkmannIBeslerCet al. Acute and short-term results of transcatheter edge-to-edge repair for severe tricuspid re-gurgitation using the MitraClip XTR system. JACC Cardiovasc Interv. (2019) 12:604–5. 10.1016/j.jcin.2018.11.028
31.
FamNPHoECZahraniMSamargandySConnellyKA. Transcatheter tricuspid valve repair with the PASCAL system. JACC Cardiovasc Interv. (2018) 11:407–8. 10.1016/j.jcin.2017.12.004
32.
BraunDLöwKOrbanMMehrMGmeinerJMassbergSet al. Mitraclip XTR VS. pascal transcatheter valve repair system for edge-to-edge repair of severe tricuspid regurgitation. J Am College Cardiol. (2020) 75:1135. 10.1016/S0735-1097(20)31762-9
33.
Campelo-ParadaFPerlmanGPhilipponFYeJThompsonCBédardEet al. First-in-man experience of a novel transcatheter repair system for treating severe tricuspid regurgitation. J Am Coll Cardiol. (2015) 66:2475–83. 10.1016/j.jacc.2015.09.068
34.
AsmaratsLPerlmanGPrazFHenseyMChrissoherisMPPhilipponFet al. Long-term outcomes of the forma transcatheter tricuspid valve repair system for the treatment of severe tricuspid regurgitation: insights from the first-in-human experience. JACC Cardio-Vasc Interv. (2019) 12:1438–47. 10.1016/j.jcin.2019.04.038
35.
FukudaSSongJ-MGillinovAMMcCarthyPMDaimonMKongsaerepongVet al. Tricuspid valve tethering predicts residual tricuspid regurgitation after tricuspid annuloplasty. Circulation. (2005) 111:975–9. 10.1161/01.CIR.0000156449.49998.51
36.
FukudaSGillinovAMMcCarthyPMMatsumuraYThomasJDShiotaT. Echocardiographic follow-up of tricuspid annuloplasty with a new three-dimensional ring in patients with functional tri-cuspid regurgitation. J Am Soc Echocardiogr. (2007) 20:1236–42. 10.1016/j.echo.2007.03.006
Summary
Keywords
tricuspid regurgitation (TR), transcatheter tricuspid intervention, FORMA, PASCAL, MitraClip device
Citation
Muntané-Carol G, Alperi A, Faroux L, Bédard E, Philippon F and Rodés-Cabau J (2020) Transcatheter Tricuspid Valve Intervention: Coaptation Devices. Front. Cardiovasc. Med. 7:139. doi: 10.3389/fcvm.2020.00139
Received
10 May 2020
Accepted
30 June 2020
Published
13 August 2020
Volume
7 - 2020
Edited by
Antonio Mangieri, Maria Cecilia Hospital, Italy
Reviewed by
Fabien Praz, Bern University Hospital, Switzerland; Damiano Regazzoli, Humanitas Research Hospital, Italy
Updates

Check for updates
Copyright
© 2020 Muntané-Carol, Alperi, Faroux, Bédard, Philippon and Rodés-Cabau.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Josep Rodés-Cabau josep.rodes@criucpq.ulaval.ca
This article was submitted to Structural Interventional Cardiology, a section of the journal Frontiers in Cardiovascular Medicine
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.