CASE REPORT article

Front. Cardiovasc. Med., 22 July 2022

Sec. Cardiovascular Surgery

Volume 9 - 2022 | https://doi.org/10.3389/fcvm.2022.926527

A Rare Ultra-Long-Term Complication of Occluder Recanalization Due to Spontaneous Perforation of Polyvinyl Alcohol Membrane of Atrial Septal Defect Occluder: A Case Report and Review of the Literature

  • Department of Cardiovascular Surgery, The Second Xiangya Hospital, Central South University, Changsha, China

Article metrics

View details

3

Citations

3,5k

Views

767

Downloads

Abstract

Percutaneous closure of atrial septal defect (ASD) has emerged as a feasible alternative strategy to surgical repair in many cardiac centers worldwide. Occluder recanalization due to device failure is a rare and severe complication that often occurs within weeks to years after ASD closure. We reported a rare ultra-long-term complication of occluder recanalization due to delayed spontaneous perforation of polyvinyl alcohol (PVA) membrane of ASD occluder after 18 years of ASD closure. Surgical removal of the faulty device and reconstruction of the atrial septum with a bovine pericardial patch was performed. The patient was discharged and recovered uneventfully without syncope or residual shunt. The cause of this rare complication of spontaneous PVA membrane perforation of the occluder has not been fully detected. To our knowledge, this is the first report about PVA membrane perforation of an occluder that occurred soon after ASD closure.

Background

Atrial septal defect (ASD) is one of the most common congenital heart diseases in children, accounting for 7–10% of all congenital heart diseases (1). Percutaneous closure of ASD has emerged as a viable alternative to surgery when feasible in the case of favorable anatomy (2–4). The technique of catheter intervention for the treatment of ASD and patent foramen ovale was pioneered by King and Mills in 1975, which led to the invention of various devices from the late 1980s to the mid-1990s (5). The mushroom umbrella occluder (Huayi Shengjie Co., Ltd., Beijing, China) is made of two disc frames woven by a nitinol wire mesh and has a mushroom-shaped hole structure with self-expanding properties. The two discs and the waist are covered with medical polymer PVA membrane film to enhance the occlusion effect. PVA is a water-soluble polymer, which becomes insoluble for medical applications with formaldehyde or glutaraldehyde cross-links (6). The PVA membranes have a long history of permanent implantation in the medical field (2–4, 7). However, the complications of spontaneous perforation of the PVA membrane have been observed with ASD occluders made by both domestic and foreign manufacturers. Neither the previous surgeons nor the manufacturers could explain the mechanism of the PVA membrane disappearance. To date, the reports of long-term complications of occluder recanalization due to spontaneous PVA membrane perforation are extremely rare (8). Herein, we reported a 40-year-old woman with an ultra-long-term complication of massive left-to-right shunt due to spontaneous PVA membrane perforation of the occluder after 18 years of ASD closure.

Case Report

A 40-year-old female with sudden syncope was referred to our department. She has an 18-year history of percutaneous ASD (30 mm in diameter) closure with a 40-mm mushroom umbrella occluder (Huayi Shengjie Co., Ltd., Beijing, China). The surgical result was satisfactory, and no residual shunt was seen. Regular re-examination 3 years after the operation showed no abnormality. Two cesarean sections after 4 and 7 years of percutaneous ASD closure and a hysteromyoma surgery after 13 years of percutaneous ASD closure had been done. Transthoracic echocardiography and electrocardiogram during these three hospitalizations showed no abnormality. The patient in our case has not been followed up and had not undergone transthoracic echocardiography for the last 5 years. In the past 3 months, the patient had two episodes of sudden syncope, one was induced by exercise, and another had no obvious incentive, accompanied by sweating and pale complexion, which was relieved spontaneously in about 5 min. She denied chest pain, chest tightness, shortness of breathing, and palpitations. On physical examination, a soft blowing murmur was heard in the second and third costal margin of the left margin of the sternum. There were no other notable clinical findings during a physical examination and no medical, family, or psychosocial history including genetic information about cardiovascular disease. X-rays (Figures 1A,B) showed that most parts of the closure were deformed, especially the left atrial side (Figure 1B, arrow) and that the nitinol frame of the occluder was not displaced and there was no evidence of a frame fracture. The electrocardiogram demonstrated sinus rhythm and an incomplete right bundle branch block. Transthoracic echocardiography revealed a recurrent significant multi-bundle left-to-right shunt through the device and small adhesions were visible on the occluder (Figure 1C). Moreover, the enlarged right heart chambers and mild mitral valve regurgitation were detected. The patient was eventually diagnosed with a late-onset residual shunt of transcatheter device occlusion of ASD (occluder recanalization).

Figure 1

Figure 1

Chest radiography (A,B) and transthoracic echocardiography (C) preoperatively indicated that most parts of the closure were deformed, especially the left atrial side [(B), arrow]; a significant multi-bundle left-to-right shunt through the device and a small adhesion on the occluder was detected (C).

Considering the multiple perforations of the occluder with an unusual irregular surface of the discs (Figure 1B), and after a careful discussion with the patient and her family, an open heart surgery was scheduled and performed. A standard median sternotomy incision was performed. The aorta was then cannulated, followed by separate cannulas placed in the superior vena cava and inferior vena cava. After full heparinization, cardiopulmonary bypass (CPB) was routinely applied. The aorta was cross-clamped, and a cold cardioplegic solution (Del Nido) was instilled via the aortic root to arrest the heart. After the right atrium was opened, the initialized tissue of the failed occluder was sufficiently freed, the occluder was then released and completely removed, and a bovine pericardial patch was used to reconstruct the atrial septum. During the operation, most parts of the closure were deformed (Figure 2), especially the left atrial side closed to the edge of the pulmonary vein, which resulted in the possible cause of late-onset residual shunt. The PVA membranes showed multiple perforations and were partly dissolved, the diameter of the largest void was around 10 mm, and the left atrial side of the occluder was damaged more obvious than the right (Figure 2B, arrow). No thrombus and other vegetations were seen. Furthermore, according to the pathology report (Figure 3), the surface of the occluder has not been fully epithelialized, instead, there was an extensive proliferation of fibrous collagenous tissue. This may have been due to the constant flow of blood. The implanted device did not show evidence of infection, and the bacterial cultures were negative. Finally, the patient recovered uneventfully without recurrent syncope and residual shunt (Figure 4).

Figure 2

Figure 2

Intraoperative view of the procedure: After opening the right atrium, a deformed occluder was found in the atrial septum (A); the left side of the occluder was damaged more obvious than the right [(B), arrow]; The PVA membranes showed multiple perforations and were partly dissolved, and the diameter of the largest void was around 10 mm (C,D).

Figure 3

Figure 3

Postoperative histological examination with hematoxylin-eosin staining showing the proliferation of fibrous collagenous tissue (A,B). (100 ×, scale bar = 50 μm).

Figure 4

Figure 4

Postoperative transthoracic echocardiography showing that no residual shunt through the atrial septum was detected.

Discussion

Percutaneous closure of ASD has emerged as a feasible alternative to surgical repair due to its shorter recovery and less invasiveness in the late 1990s and has even become the treatment of choice in many heart centers worldwide (9, 10). It is well-known that percutaneous ASD closure also has potential short-term problems, such as occluder displacement and dislodgement (11, 12), residual shunt, arrhythmia (13), occluder device failure (14), occluder abrasion (15), bleeding or thromboembolism (16), air embolism (17), hemolysis (18), pericardial effusion or cardiac tamponade (15), mitral regurgitation (19), headache or migraine, etc. There are also some interventional complications, such as anesthesia accident, wound infection, arteriovenous fistula, and so on. With the development of medical equipment and technology, percutaneous ASD closure is known to have satisfactory short-term outcomes and complication rates. However, an important consideration for percutaneous ASD closure is the long-term and ultra-long-term complications that have emerged after its widespread adoption. Some of these, including cardiac erosion (15, 20–23), infective endocarditis (24, 25), and thromboembolism (26, 27), carry unignored morbidity and mortality (4, 28). We report a case of spontaneous perforation of the PVA membrane observed after 18 years in a patient treated with a 40-mm mushroom umbrella occlude (Huayi Shengjie Co., Ltd., Beijing, China). Similar cases have been previously reported in many versions of the PVA membrane-covered ASD occluder (29–31) (Table 1). However, to our knowledge, this is the first report about the PVA membrane perforation of occluder occurring at the latest time after ASD closure.

Table 1

First author (year) Patient age/sex Device type Device size Occluder recanalization after procedure Symptom Management References
Bozyel (2) 54/F Cardia Ultrasept septal occluder 30 mm 2 years No Surgical device removal and Gore-Tex patch repair (2)
Labombarda (3) 20/F Ultrasept II ASD occluder device 20 mm 1 months Recurrent dyspnea Surgical device removal and Gore-Tex patch repair (3)
Ramoglu (4) 4/M Cardia Ultrasept II ASD occluder 20 mm 1 week NA Surgical device removal and Gore-Tex patch repair (4)
Aubry et al. (7) 77/M ASD Ultrasept II closure device 24 mm 4 months No Surgical device removal and Gore-Tex patch repair (7)
Aubry et al. (7) 41/F ASD Ultrasept II closure device 32 mm 3 months No Surgical device removal and Gore-Tex patch repair (7)
Ten Freyhaus (8) 72/M Atrial septal defect occluder system NA 8 years Worsening dyspnea on exertion Surgical device removal and bovine pericardial patch repair (8)
Chamie et al. (9) 28/F UltraseptTM II CARDIA ASD occluder 20 mm 3 months Easy tiredness and fatigue on exertion Device-in-device technique (9)
Chamie et al. (9) 33/F UltraseptTM II CARDIA ASD occluder 16 mm 4 months NA Device-in-device technique (9)
Chamie et al. (9) 49/F UltraseptTM II CARDIA ASD occluder 16 mm 6 months NA Device-in-device technique (9)
Chamie et al. (9) 17/F UltraseptTM II CARDIA ASD occluder 14 mm 3 months NA Device-in-device technique (9)
Bartel (29) 62/F ATRIASEPT II device 24 mm 6 weeks No Surgical device removal and patch repair (29)
Bartel (29) 42/F ATRIASEPT II device 20 mm 5 weeks No Surgical device removal and patch repair (29)
Kitamura (30) 20/M Amplatzer septal occluder NA 3 years NA (infective endocarditis) Surgical device removal and bovine pericardial patch repair (30)
Weryński (31) 9/M Cardia Ultrasept occluder 20 mm 4 years NA Surgical device removal and bovine pericardial patch repair (31)
Weryński (31) 6/F Cardia Ultrasept occluder 20 mm 3 years NA Surgical device removal and bovine pericardial patch repair (31)
Aguiar Rosa (32) 39/F Ultrasept ASD Occluder 22 mm 2 years NA Device-in-device technique (32)
Bhattacharyya et al. (38) 69/M Cardia Ultrasept septal occluder 28 mm 10 months Several transient neurological events Covering the damaged
membrane with another occlusion device
(38)

Summary of reported cases of polyvinyl alcohol membrane perforation of atrial septal occluders.

NA, Note available; M, Male; F, Female.

The PVA is a synthetic polymer widely used in medical devices due to its good biocompatibility, chemical resistance, low adsorption to proteins, non-toxicity, and adhesion (6, 9, 32). Although the application of this material in ASD occluders has been generally successful, some cases of spontaneous disintegration and perforation of the PVA membrane have been described (Table 1). Neither the previous surgeons nor the manufacturers could explain the mechanism of the PVA membrane disappearance. It may be due to the incomplete endothelialization of the occluder, poor occluder placement, long-term medications, underlying immunocompromise, syndromic comorbidities, or metabolic disorders (2, 28, 32). Undoubtedly, more detailed and comprehensive follow-up data were needed to answer that question. In our patient, there was no evidence of an early and mid-term residual shunt due to device failure, but we cannot speculate on the specific cause of long-term device perforation. Similar complications of occluder recanalization due to dissolution of the PVA membrane has previously been reported as early as 1 week and as late as 8 years after implantation (4, 8). We report a case of spontaneous perforation of the PVA membrane of the occluder 18 years after surgery, so there should be no direct relationship between this complication and the time of device implantation. In our case, we did not find any evidence of infection, long-term medications, or systemic disease, so it is speculated that this may be the cause of PVA membrane degradation of the occluder, external force damage, or long-term endothelialization insufficiency. Another possible reason may be that the size of the occluder was too large, which make the endothelial tissue unsuitable for migration, and the endothelialization coverage was poor. Although complete endothelialization of ASD devices was thought to occur 3–6 months after device implantation, late incomplete endothelialization has been described based on animal and human studies (33–36). It can be seen from the disassembled occluder that the left atrial side near the edge of the pulmonary veins was more seriously damaged than the right. In addition, mitral valve regurgitation and long-term pulmonary venous blood flow scouring lead to damage of the PVA membrane of the occluder, followed by chronic dissolution. After some adverse events, the manufacturer announced the development of a new device in hopes of ameliorating this complication by adding a Gore-Tex patch between two metal disc pieces (3). The new device was proved to be safe and feasible. This conclusion was drawn from a study of 30 Mexican patients after ASD closure without occluder recanalization due to the PVA membrane perforation at follow-up for 6 (range: 1–15) months (37). However, the conclusion was only based on early follow-up data.

The patient in the presentation first experienced syncope 18 years after ASD closure, but device perforation may have occurred months or years earlier. A timeline with relevant data from the presented case was shown in Figure 5. Patients with device failure were mostly asymptomatic but may also present with progressive dyspnea, fatigue, or stroke (3, 9, 38). It has been reported that occluder failure often induces some long-term complications after ASD closure, such as thrombosis on the surface of the occluder, thromboembolism, infective endocarditis, cardiac erosion, nickel hypersensitivity, and valve damage (8, 24, 25, 28). In most reported cases of occluder recanalization, surgical removal of the faulty device and repair with a patch were preferred (2). With the development of medical equipment and technology, covering the damaged membrane with a second device with intervention is considered a viable alternative to surgery when feasible in case of favorable anatomy (9, 32, 38). This should be done at a heart center with comprehensive imaging interventional equipment and experienced interventional cardiac surgeons.

Figure 5

Figure 5

A figure showcasing a timeline with relevant data from the presented case.

Conclusion

In conclusion, for the first time, we report a rare case of severe left-to-right shunt due to spontaneous perforation of the PVA membrane of ASD occluder 18 years after ASD closure. The cause of the rare serious complication of occluder recanalization due to the PVA membrane perforation has not been clearly explained. Surgeons and manufacturers should be aware of this potential ultra-long-term complication that cannot be ignored, and they should conduct long-term systematic follow-up examinations for all patients implanted with occluder devices to reduce the potential losses of the patient while obtaining comprehensive clinical data to guide the optimization of devices and technologies.

Funding

This work was supported by the Key Project of Science and Technology of Hunan Province (No. 2020SK53420 to JY).

Publisher's Note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Statements

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

The studies involving human participants were reviewed and approved by the Ethics Committee of the Second Xiangya Hospital of Central South University. The patients/participants provided their written informed consent to participate in this study.

Author contributions

KX drafted the manuscript. CF designed the study. HZ, QW, MT, CF, and JY revised the manuscript. KX, MT, and QW were responsible for the collection of data or analysis. All authors read and approved the final manuscript.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

References

  • 1.

    Feltes TF Bacha E Beekman RH Cheatham JP Feinstein JA Gomes AS et al . et al. Indications for cardiac catheterization and intervention in pediatric cardiac disease: a scientific statement from the American Heart Association. Circulation. (2011) 123:2607–52. 10.1161/CIR.0b013e31821b1f10

  • 2.

    Bozyel S Sahin T Dervis E Aktas M Saskin H . A massive left-to-right shunt due to delayed spontaneous perforation of polyvinyl alcohol membrane of atrial septal occluder. Turk Kardiyol Dern Ars. (2017) 45:541–4. 10.5543/tkda.2017.04640

  • 3.

    Labombarda F Roule V Beygui F . Delayed spontaneous perforation of polyvinyl alcohol membrane-Covered atrial septal defect closure devices. Catheter Cardiovasc Interv. (2017) 89:E141–4. 10.1002/ccd.26704

  • 4.

    Ramoglu MG Ucar T Tutar E . Early malfunction of polyvinyl alcohol membrane of septal occluder. Catheter Cardiovasc Interv. (2016) 87:E151–153. 10.1002/ccd.26057

  • 5.

    King TD Thompson SL Steiner C Mills NL . Secundum atrial septal defect. Nonoperative closure during cardiac catheterization. JAMA. (1976) 235:2506–9. 10.1001/jama.1976.03260490024013

  • 6.

    Alexandre N Ribeiro J Gartner A Pereira T Amorim I Fragoso J et al . Biocompatibility and hemocompatibility of polyvinyl alcohol hydrogel used for vascular grafting—In vitro and in vivo studies. J Biomed Mater Res A. (2014) 102:4262–75. 10.1002/jbm.a.35098

  • 7.

    Aubry P Brochet E du Fretay XH Bouton-Brochet S Ibrahim H Arrault X et al . Early malfunction of polyvinyl alcohol membrane-covered atrial septal defect closure devices. Circ Cardiovasc Interv. (2014) 7:721–2. 10.1161/CIRCINTERVENTIONS.114.001429

  • 8.

    Ten Freyhaus H Rosenkranz S Sudkamp M Hopp HW . Dysfunction of an atrial septal defect occluder 8 years after implantation. J Interv Cardiol. (2006) 19:163–5. 10.1111/j.1540-8183.2006.00124.x

  • 9.

    Chamie F Maia J Giuliano LC . Device-in-device: A transcatheter alternative to surgical explantation of a failing atrial septal defect intracardiac prosthesis. Catheter Cardiovasc Interv. (2016) 88:239–43. 10.1002/ccd.26456

  • 10.

    Masura J Gavora P Formanek A Hijazi ZM . Transcatheter closure of secundum atrial septal defects using the new self-centering amplatzer septal occluder: initial human experience. Cathet Cardiovasc Diagn. (1997) 42:388–93.

  • 11.

    Boysan E Cicek OF Cicek MC Hamurcu Z Gurkahraman S . Surgical removal of an atrial septal occluder device embolized to the main pulmonary artery. Tex Heart Inst J. (2014) 41:91–3. 10.14503/THIJ-12-3003

  • 12.

    Truong QA Gupta V Bezerra HG Okada DR Rogers IS Holmvang G et al . Images in cardiovascular medicine. The traveling amplatzer: rare complication of percutaneous atrial septal occluder device embolism. Circulation. (2008) 118:e93–96. 10.1161/CIRCULATIONAHA.108.765586

  • 13.

    Agarwal S Bajaj NS Kumbhani DJ Tuzcu EM Kapadia SR . Meta-analysis of transcatheter closure versus medical therapy for patent foramen ovale in prevention of recurrent neurological events after presumed paradoxical embolism. JACC Cardiovasc Interv. (2012) 5:777–89. 10.1016/j.jcin.2012.02.021

  • 14.

    Qureshi AM Mumtaz MA Latson LA . Partial prolapse of a HELEX device associated with early frame fracture and mitral valve perforation. Catheter Cardiovasc Interv. (2009) 74:777–82. 10.1002/ccd.22081

  • 15.

    Divekar A Gaamangwe T Shaikh N Raabe M Ducas J . Cardiac perforation after device closure of atrial septal defects with the Amplatzer septal occluder. J Am Coll Cardiol. (2005) 45:1213–8. 10.1016/j.jacc.2004.12.072

  • 16.

    Abaci A Unlu S Alsancak Y Kaya U Sezenoz B . Short and long term complications of device closure of atrial septal defect and patent foramen ovale: meta-analysis of 28,142 patients from 203 studies. Catheter Cardiovasc Interv. (2013) 82:1123–38. 10.1002/ccd.24875

  • 17.

    Alkashkari W Albugami S Hijazi ZM . Current practice in atrial septal defect occlusion in children and adults. Expert Rev Cardiovasc Ther. (2020) 18:315–29. 10.1080/14779072.2020.1767595

  • 18.

    Yu M Shan X Bai Y Qin Y Zhao X . Percutaneous closure of a large atrial septal defect presenting with acute severe hemolysis. Clin Res Cardiol. (2019) 108:828–30. 10.1007/s00392-019-01431-z

  • 19.

    Dialetto G Covino FE Scognamiglio G Della Corte A Manduca S Iacono C Calabro R Cotrufo M . A rare complication of atrial septal occluders: diagnosis by transthoracic echocardiography. J Am Soc Echocardiogr. (2006) 19:836 e835–838. 10.1016/j.echo.2006.01.027

  • 20.

    Maimon MS Ratnapalan S Do A Kirsh JA Wilson GJ Benson LN . Cardiac perforation 6 weeks after percutaneous atrial septal defect repair using an Amplatzer septal occluder. Pediatrics. (2006) 118:e1572–75. 10.1542/peds.2005-0825

  • 21.

    Vitanova K Cleuziou J Vogt M Nobauer C Schreiber C Lange R et al . Delayed perforation of a transcatheter patent foramen ovale occluder 10 years after implantation. Ann Thorac Surg. (2015) 99:700–1. 10.1016/j.athoracsur.2014.04.117

  • 22.

    Mendirichaga R Smairat RA Sancassani R . Late tissue erosion after transcatheter closure of an atrial septal defect. Catheter Cardiovasc Interv. (2017) 89:502–4. 10.1002/ccd.26277

  • 23.

    Crawford GB Brindis RG Krucoff MW Mansalis BP Carroll JD . Percutaneous atrial septal occluder devices and cardiac erosion: a review of the literature. Catheter Cardiovasc Interv. (2012) 80:157–67. 10.1002/ccd.24347

  • 24.

    Amedro P Soulatges C Fraisse A . Infective endocarditis after device closure of atrial septal defects: case report and review of the literature. Catheter Cardiovasc Interv. (2017) 89:324–34. 10.1002/ccd.26784

  • 25.

    Nguyen AK Palafox BA Starr JP Gates RN Berdjis F . Endocarditis and incomplete endothelialization 12 years after amplatzer septal occluder deployment. Tex Heart Inst J. (2016) 43:227–31. 10.14503/THIJ-14-4949

  • 26.

    Fernando RR Koranne KP Barker CM . AMPLATZER Septal Occluder failure resulting in paradoxical cerebral embolism. Tex Heart Inst J. (2012)39:647–52.

  • 27.

    Kalayci B Kalayci S . Right atrial ball thrombus associated with atrial septal occluder device: A late complication of transcatheter atrial septal defect closure. Anatol J Cardiol. (2017) 18:E9. 10.14744/AnatolJCardiol.2017.8012

  • 28.

    Prabhu NK Turek JW Andersen ND . Commentary: Don't leave patients to their own devices: consider long-term complications after percutaneous atrial septal defect closure. JTCVS Tech. (2021) 7:242–3. 10.1016/j.xjtc.2021.02.015

  • 29.

    Bartel T Bonaros N Muller S . Device failure weeks to months after transcatheter closure of secundum type atrial septal defects. Hearts. (2010) 96:1603. 10.1136/hrt.2009.185942

  • 30.

    Kitamura H Yamamoto M Kagase A . Mobile vegetations on a poorly endothelialized atrial septal defect closure device. JTCVS Tech. (2021) 7:240–1. 10.1016/j.xjtc.2021.01.010

  • 31.

    Werynski P Sabiniewicz R . Unexpected long-term complications of atrial septal defects closure. Postepy Kardiol Interwencyjnej. (2020) 16:514–5. 10.5114/aic.2020.101782

  • 32.

    Aguiar Rosa S Ferreira F de Sousa L Fiarresga A Martins JD Galrinho A et al . Successful percutaneous closure of a residual atrial septal defect due to device failure. Rev Port Cardiol. (2017) 36:475 e471–5 e473. 10.1016/j.repc.2016.09.020

  • 33.

    Lock JE Rome JJ Davis R Van Praagh S Perry SB Van Praagh R et al . Transcatheter closure of atrial septal defects. Exp Stud Circ. (1989) 79:1091–9. 10.1161/01.CIR.79.5.1091

  • 34.

    Chen F Zhao X Zheng X Chen S Xu R Qin Y . Incomplete endothelialization and late dislocation after implantation of an Amplatzer septal occluder device. Circulation. (2011) 124:e188–9. 10.1161/CIRCULATIONAHA.110.991836

  • 35.

    Sigler M Jux C Ewert P . Histopathological workup of an Amplatzer atrial septal defect occluder after surgical removal. Pediatr Cardiol. (2006) 27:775–6. 10.1007/s00246-006-1413-1

  • 36.

    Sigler M Kriebel T Wilson N . Histological confirmation of complete endothelialisation of a surgically removed Amplatzer ASD occluder. Heart. (2006) 92:1723. 10.1136/hrt.2006.091256

  • 37.

    Mijangos-Vazquez R Garcia-Montes AJ Soto-Lopez EM Guarner-Lans V Zabal C . Atrial septal defect closure with the new Cardia Ultrasept II device with interposed Goretex patch: Mexican experience—has the perforation of Ivalon's membrane been solved?Cardiol Young. (2018) 28:709–14. 10.1017/S1047951118000100

  • 38.

    Bhattacharyya S Ilsley CD Baltabaeva A . Disintegration of polyvinyl alcohol membrane covering atrial septal defect closure device. Eur Heart J Cardiovasc Imag. (2015) 16:1153. 10.1093/ehjci/jev163

Summary

Keywords

atrial septal defect, closure, occluder recanalization, ultra-long-term complication, surgery

Citation

Xiang K, Zhuang H, Wu Q, Tang M, Yang J and Fan C (2022) A Rare Ultra-Long-Term Complication of Occluder Recanalization Due to Spontaneous Perforation of Polyvinyl Alcohol Membrane of Atrial Septal Defect Occluder: A Case Report and Review of the Literature. Front. Cardiovasc. Med. 9:926527. doi: 10.3389/fcvm.2022.926527

Received

22 April 2022

Accepted

14 June 2022

Published

22 July 2022

Volume

9 - 2022

Edited by

Hendrik Tevaearai Stahel, Bern University Hospital, Switzerland

Reviewed by

Christopher Learn, Mass General Brigham, United States; Ömer Faruk Çiçek, Selcuk University, Turkey

Updates

Copyright

*Correspondence: Chengming Fan

This article was submitted to Heart Surgery, a section of the journal Frontiers in Cardiovascular Medicine

Disclaimer

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Outline

Figures

Cite article

Copy to clipboard


Export citation file


Share article

Article metrics