High-resolution 3D whole-heart bright- and black-blood imaging with co-registered T2 mapping at 0.55 T


Ivan Kokhanovskyi,,,Ivan Kokhanovskyi1,2,3,4Carlos Castillo-Passi,,,Carlos Castillo-Passi3,4,5,6Michael G. CrabbMichael G. Crabb3Carl GanterCarl Ganter1Simon J. LittlewoodSimon J. Littlewood3Karl P. Kunze,Karl P. Kunze3,7Dimitrios C. KarampinosDimitrios C. Karampinos1Marcus R. Makowski,Marcus R. Makowski1,2Daniel Rueckert,Daniel Rueckert8,9Claudia Prieto,,Claudia Prieto3,4,5Ren M. Botnar,,,,

René M. Botnar2,3,4,5,6*
  • 1Institute of Diagnostic and Interventional Radiology, School of Medicine and Health, Technische Universität München, Munich, Germany
  • 2Institute for Advanced Study, Technische Universität München, Munich, Germany
  • 3School of Biomedical Engineering and Imaging Science, King’s College London, London, United Kingdom
  • 4Millenium Institute for Intelligent Healthcare Engineering, Santiago, Chile
  • 5School of Engineering, Pontificia Universidad Catolica de Chile, Santiago, Chile
  • 6Institute for Biological and Medical Engineering, Pontificia Universidad Católica de Chile, Santiago, Chile
  • 7MR Research Collaborations, Siemens Healthcare Limited, Camberley, United Kingdom
  • 8Chair for AI in Healthcare and Medicine, Technical University of Munich (TUM) and TUM University Hospital, Munich, Germany
  • 9Department of Computing, Imperial College London, London, United Kingdom

Introduction: Conventional CMR exams for assessment of cardiac anatomy and tissue characterization require multiple sequential 2D acquisitions under breath-hold in different orientations, in addition to being limited to 1.5 T and 3 T.

Methods: In this study, we sought to develop a novel 3D motion-compensated free-breathing sequence for comprehensive high-resolution whole-heart assessment of cardiovascular anatomy via simultaneous bright- and black-blood imaging and co-registered T2 myocardial tissue quantification in a one-click scan at 0.55 T.

Results: Good agreement with a spin-echo reference sequence was found in the phantom for T2 mapping. In-vivo, the proposed research sequence was evaluated in 10 healthy subjects, providing great delineation of cardiac and vascular structures, good visibility of coronary arteries and accurate T2 parametric mapping in a clinically feasible time of less than 9 min.

1 Introduction

Cardiovascular diseases (CVDs) are the leading cause of mortality, responsible for a third of all deaths globally (1). Efficient screening is a key factor in the early detection of CVDs. Cardiovascular magnetic resonance (CMR) imaging is a well-established non-invasive imaging modality for a comprehensive evaluation of the structure and function of the heart, without the inherent risks associated with the exposure to ionizing radiation. In addition, quantitative myocardial tissue characterization with T1 and T2 mapping provides valuable information on the presence of myocardial fibrosis, edema, or inflammation, especially in CVDs with diffuse tissue alterations (2, 3). However, CMR examinations remain costly and complex procedures with restricted access for a significant part of the global population, especially in low- and middle-income countries (4).

With the recent introduction of high-end low-field MRI scanners, there is great potential to make CMR more accessible and affordable. Compared to mid- and high-field MRI, the installation and maintenance costs of low-field units are significantly reduced, in addition to lower weight and smaller size of low-field MRI scanners (5, 6). Overall patient comfort is improved due to the wider bore opening and reduced vibration of gradients (7). To address claustrophobia concerns, low-field magnets can be configured in open geometry systems, expanding the ability to integrate stress testing into CMR examinations. The specific absorption rate (SAR) scales with B02, resulting in reduced tissue or device heating in low-field MR scanners (8). Off-resonance and susceptibility artefacts is also reduced, enabling near-device imaging (e.g., pacemakers, sternal wires) and spiral trajectories for efficient diagnostic imaging (9).

Despite the aforementioned advantages, low-field applications still face some limitations, preventing their daily clinical use (10). Selecting the optimal scan parameters requires a trade-off between clinically acceptable scan times, image resolution and quality. Especially for low-field CMR systems with limited gradient performance, new methods for both image acquisition and reconstruction are needed (11). One of the main challenges of low-field MRI is the lower signal-to-noise ratio (SNR), which scales with B03/2, resulting in images with reduced spatial resolution or requiring longer scan time (12). In addition, motion correction algorithms successfully used at higher field strengths may be challenging to adopt at low-field due to the lower SNR of the source images. Favorable changes in relaxation parameters, with a significantly shorter T1 and a slightly longer T2 (13), at lower field strength partially compensate for SNR loss due to slower signal decay and faster magnetization recovery, which may render parametric mapping more time efficient. Balanced steady-state free precession [bSSFP (14)] sequences offer excellent SNR efficiency and high blood-to-myocardium contrast and can therefore help to compensate for the SNR disadvantage. Moreover, typical bSSFP banding artefacts caused by B0 inhomogeneity are less pronounced in low-field scanners, making it a promising choice for anatomical imaging (15).

There is still limited experience with high-resolution anatomical imaging and parametric mapping of the heart at low-field strengths, especially when performed in a single scan. Typical clinical CMR protocols require complex planning by expert personnel, involving several sequential 2D acquisitions in different orientations with multiple breath-holds, leading to long and unpredictable examination times (16). To overcome the above limitations, a combined bright- and black-bood 3D imaging sequence was developed for simultaneous lumen and wall visualization of the great thoracic vessels and cardiac structures (1719). This was further adapted to enable a combined anatomic imaging and T2 tissues quantification at 1.5 T (20). The feasibility of combined bright- and black-blood imaging has been recently demonstrated at low-field (21). In this study, we sought to develop a novel motion-compensated free-breathing CMR research sequence at 0.55 T that allows simultaneous 3D whole-heart assessment of cardiovascular anatomy via bright- and black-blood imaging and T2 myocardial tissue quantification in a single one-click scan within a clinically feasible scan time of less than 9 min.

2 Materials and methods

2.1 Pulse sequence framework

The proposed ECG-triggered research sequence consists of a repeating set of three heartbeats (HBs) arranged in an interleaved fashion, incorporating an adiabatic T2 preparation (T2prep) module accompanied by an inversion recovery (IR) pulse in the 1st, no preparation in the 2nd and another T2prep in the 3rd HB (see Figure 1A). Frequency-selective fat saturation (FatSat) pulses are employed to suppress epicardial fat in the 2nd and 3rd HBs. In the 1st HB, fat suppression is achieved by IR pulse with an appropriate inversion time (TI). 3D bSSFP acquisition is performed in coronal orientation during the mid-diastolic rest period to minimize cardiac motion by adjusting a subject-dependent trigger delay. 2D low-resolution image-based navigators (iNAVs) are acquired prior to each HB by spatially encoding ramp-up pulses to perform beat-to-beat intra-bin translational food-head and left-right motion estimation and correction, as well as respiratory binning of data (see Figure 1B) (22). A variable density 3D Cartesian spiral-like k-space trajectory with a golden angle step (VD-CASPR) is employed to collect a highly undersampled data with 100% respiratory efficiency (23). The 1st volume serves as a bright-blood image, while a black-blood volume is obtained by direct magnitude subtraction of the 1st from the 2nd volume (see Figure 1C). Co-registered T2 maps are generated voxel-wise by dictionary matching of all three volumes.

Figure 1
Diagram illustrating a cardiac MRI process. Panel A shows a heart rate graph with labels for T2prep-IR and 3D bSSFP. Panel B depicts respiratory binning with wave patterns leading to VD-CASPR binned k-space and 3D motion fields. Panel C displays acquired images: bright-blood images for three volumes with T2prep times, a signal contrast graph for dictionary matching, and resulting black-blood image and T2 map with color scale from 0 to 200.

Figure 1. (A) Sequence diagram over three interleaved heartbeats (HBs): T2prep (50 ms) IR (TI = 90 ms)—none—T2prep (30 ms). (B) Respiratory binning of data acquired with VD-CASPR to derive 3D non-rigid motion fields. Images are reconstructed in-line using motion-corrected itSENSE followed by off-line HD-PROST for denoising. (C) Contrasts are shown for the mid-coronal slice. 3D datasets for each volume can be found in the Supplementary Material. The bright-blood image is acquired in the 1nd HB, while the black-blood image is obtained through direct magnitude subtraction of the 1st from the 2nd volume. T2 maps are generated voxel-wise by dictionary matching.

2.2 Imaging parameters

T2prep modules with hyperbolic-secant adiabatic-refocusing RF pulses of 50 ms/30 ms in duration, Gaussian shaped FatSat pulses with an optimized flip angle of 180 and 26.6 ms in duration (24), TI=90ms, flip angle (FA) of bSSFP readout = 110, TR/TE=4.90ms/2.45ms, 30 segments (lines in k-space), acquisition window 150ms, 6 ramp-up pulses (also used for low-resolution iNAV spatial encoding), field of view (FOV) = 312 × 312 × 108–120 mm3, isotropic spatial resolution = 1.5 mm, VD-CASPR trajectory with an acceleration factor of 4.

2.3 Experimental protocol

The proposed framework was evaluated on the standardized cardiac T1MES phantom (25) and 10 healthy subjects (27±3 years, 3 females, heart rate (HR) of 55–78 bpm, body mass index (BMI) of 23–30 kg/m2). All data was acquired on a commercially available 0.55 T MRI scanner (MAGNETOM Free.Max, Siemens Healthineers, Forchheim, Germany) using a 6-channel small flexible coil and 9-channel spine coil, with an external ECG monitoring system (Maglife Serenity, Schiller) connected to the external trigger input. In the phantom experiment, the sequence framework was validated against a 2D IR spin-echo (SE) research sequence with SE times between 6.7 ms and 267 ms. The mean R-wave peak to R-wave peak (RR) interval for each healthy subject was extracted from an axial CINE acquisition acquired at the start of the study to obtain HR=60/RR[s]. Before undergoing the examination, written informed consent was collected from all volunteers.

2.4 Image reconstruction & denoising

Images were reconstructed using 3D non-rigid motion-corrected iterative sensitivity encoding [SENSE (26)] performed directly on the scanner in-line, providing magnitude datasets for each of three volumes separately (27). Subsequently, images were denoised using patch-based low-rank regularization [HD-PROST (28)], which groups similar patches in the multi-contrast images. The following parameters were used for HD-PROST: singular value decomposition (SVD) threshold = 0.1, voxel search window 20×20×20, voxel patch-size 6×6×6, number of selected similar patches = 20, and a voxel patch-offset = 5 when searching for similar patches.

2.5 Dictionary generation & matching

The numerical simulation of expected signals was implemented in MATLAB (The MathWorks Inc, Natick, MA, USA) using the extended phase graph [EPG (29)] method. To achieve a steady-state magnetization, the simulation involved two sets of dummy sequence repetitions without any signal sampling. The signal evolution from the third repetition was used to generate a unique, HR-specific three-point dictionary. The dictionaries were generated for a fixed T1=700ms, corresponding to a typical value of normal myocardium, with various T2s in the range [20:2:40,41:1:50,50.5:0.5:60,61:1:70,72:2:90] ms ([lower value: step size: upper value])) to cover relevant values for healthy (T2=58ms) and abnormal (T2=70ms) myocardium at 0.55 T. A short dictionary for the arterial blood with T1=1100ms and T2s in the range [95:5:250] ms was added to better separate a pool of blood from the myocardial tissue. In the phantom experiment, dictionaries with fixed mean T1 values were generated for each vial with T2s corresponding to myocardium- (A–F) and blood-like (G–I) range (see Table 1). T2 parametric values were matched voxel-wise by minimizing the least squares error (LSE) cost function between the normalized measured signal and the dictionary generated entry over three points.

Table 1
www.frontiersin.org

Table 1. T1MES phantom containing 9 agarose-based tubes with relevant cardiac T1 and T2 combinations to mimic myocardium- and blood-like material. The T1s were measured with the IR SE and T2s with SE multi-echo without IR preparation research 2D sequences at 0.55 T (see Section 2.3) and evaluated in MATLAB.

2.6 Data analysis

Anatomical image quality was measured by visibility, which is defined as the normalized signal difference between myocardium and blood signal from the 1st volume for the bright-blood image and from the 2nd–1st subtracted volume for the black-blood image. Different orientations in short-(SAx), vertical (VAx) and horizontal (HAx) long-axis views were obtained in the DICOM viewer Horos (Horos Project, Geneva, Switzerland). Multiplanar reformation and quantitative analysis of individual coronary vessels was performed using Soap-Bubble (30). Image quality was graded using the following scale (31): 0, non-diagnostic; 1, poor (limited coronary vessel visibility or noisy image); 2, average (coronary vessel visible but diagnostic confidence low); 3, good (coronary artery adequately visualised and image of diagnostic quality); and 4, excellent (coronary artery clearly depicted). The scores were assigned by a cardiology clinical research fellow with three years of experience in CMR. For quantitative analysis, vessel sharpness (VS) was calculated as the magnitude of the local change in signal intensity using a first-order derivative (32) (VS of 100% refers to a maximum change in signal intensity at the vessel border, a lower value is consistent with inferior vessel sharpness) over the first 4 cm along the right coronary artery on a bright-blood contrast. To assess the performance of the T2 mapping, the correlation plot and Bland-Altman analysis were carried out for axial images of the tubes in the phantom. The bias was adjusted by varying the amount of k-space line segments used for dictionary matching in the range [2:2:12] ([lower value: step size: upper value]) out of 30. To visualize in-vivo data, bull’s-eye plots were generated from 16-segment 3D T2 maps according to the American Heart Association model (33) (without apical segment) using the 30 slices acquired along SAx from apex to basal regions (corresponding to a length of 4.5 cm). The mean and standard deviation (SD) of the T2 values were calculated for each segment separately. The accuracy of the proposed method was determined using the coefficient of variation (CoV), defined as the SD divided by the mean value. To estimate spatial variability, T2 values for the apex, mid-cavity and basal areas were compared across all healthy subjects. The mid-cavity SAx slice was used for visual comparison.

3 Results

3.1 Sequence design optimization

A high imaging FA of 110 was chosen to achieve high SNR and good anatomical image quality at 0.55 T. For the bright-blood image, obtained directly from the 1st volume, the selected FA is a compromise between the near maximum value for arterial blood and the relatively low signals from the myocardium [Figure 2A(i)]. The FA beyond this value results in limited contrast gain compared to the quadratic increase in specific absorption rate (SAR) (12). Moreover, a slightly lower value of imaging FA is preferable for the black-blood contrast [Figure 2B(i)], which is generated from the direct magnitude subtraction of the 1st from the 2nd volume. A TI of 90 ms was employed to reduce both myocardial and epicardial fat signals while maintaining good signal from arterial blood in the 1st volume [Figure 2A(ii)]. A low dependence of TI on bight- and black-blood contrast values was observed [Figure 2A–B(ii)]. A pulse duration of 50 ms was used in the T2prep-IR module to suppress the myocardial signal in the 1st volume [Figure 2A(iii)] and simultaneously to obtain a good signal on the subtracted image [Figure 2B(iii)]. Longer T2prep generally yields lower SNR and is not beneficial for dictionary matching, which requires T2prep durations below or close to the parametric values of the mapped myocardium tissue. Sufficient T2 encoding is then provided by two T2prep modules with different durations applied prior to the bSSFP acquisition, which also has intrinsic T2-weighting. Although changing the order of T2prep modules has almost no influence on the mapping sensitivity, it affects the contrast values of the anatomical imaging based on EPG simulations. For this reason, a shorter T2prep of 30 ms was placed in the 3rd volume.

Figure 2
Graphs show signal variations by different parameters. (A) i: Signal vs. bSSFP flip angle for arterial blood, myocardium, and bright-blood. ii: Signal vs. inversion time for epicardial fat. iii: Signal vs. pulse duration. (B) i: Signal vs. bSSFP flip angle for black-blood. ii: Signal vs. inversion time. iii: Signal vs. pulse duration. Each graph includes specific signal trends for various tissues. Vertical dashed lines indicate reference points.

Figure 2. Optimization on sequence parameters using EPG simulations at 0.55 T for a HR of 60 bpm. Simulated signals of the arterial blood (T1=1100ms,T2=260ms), healthy myocardium (T1=700ms,T2=58ms) and epicardial fat (T1=180ms,T2=93ms) are shown for the 1st volume (A) and 2nd–1st subtracted volume (B). Contrast values for the bright- and black-blood are defined as a difference between the corresponding signals. In each graph one of the following parameters was varied, while the other two were kept constant: imaging flip angle (FA), inversion time (TI) and T2prep pulse duration. Signal dependency on (i) FA with fixed inversion TI of 90 ms and T2prep of 50 ms, on (ii) TI with fixed FA of 110 and T2prep of 50 ms, and on (iii) T2prep pulse duration with fixed FA of 110 and TI of 90 ms.

3.2 Phantom study

Figure 3A shows good agreement in T2 mapping with high linear correlation of y=0.99x+0.28(R2>0.99) between the proposed 3D approach and 2D SE references for a simulated HR of 60 bpm, whereby the H vial was excluded from the analysis due to unrealistically high T1 at 0.55 T (see Table 1). From the Bland-Altman analysis (see Figure 3B), a minimal bias of 0.2 ms was obtained by averaging the signal from the first 8 imaging segments (25% of the central part of k-space).

Figure 3
Panel (A) shows a scatter plot comparing T2 values from SE reference and the proposed method with a dashed line indicating equivalence. Panel (B) presents a Bland-Altman plot assessing agreement, noting a bias of 0.21 with ±1.96 SD lines. Panel (C) includes T2 maps from SE reference and proposed methods, each displaying a grid of circular regions, with a color scale from 0 to 200 ms indicating T2 values.

Figure 3. Validation of the proposed 3D T2 multi-contrast mapping sequence in the standardized T1MES phantom with simulated HR of 60 bpm at 0.55 T against a 2D SE reference scan with correlation plot (A) and Bland-Altman analysis (B). The relevant T2 myocardial region is zoomed in. (C) Phantom image of the proposed and reference sequences with the parametric values given in Table 1. The T2 value for H vial was out of range and is therefore not shown.

3.3 In-vivo

All experiments were successfully completed in healthy subjects with an average acquisition time of 8.5±0.9 min. One participant was excluded from statistical analysis due to highly variable HR and significant movement during scanning. Figure 4 illustrates reformatted bright-blood datasets to demonstrate the visibility of the left (LCA) and right (RCA) coronary arteries. Isotropic resolution of 1.5 mm allowed good depiction of the coronary arteries, including distal segments and major branches. The mean vessel sharpness of 58.6%±8.8% was obtained and the median image quality score was 3 (good: coronary artery adequately visualized and image of diagnostic quality). In Figure 5, 3D cardiovascular anatomy is displayed via bright- and black-blood imaging for two volunteers in coronal, VAx and HAx orientations along with T2 maps. Excellent delineation of cardiac and vascular structures was observed together with spatially uniform mapping in the myocardium. Figure 6 displays SAx views from apex to base for a representative subject alongside with corresponding bull’s-eye-plots (30 slices), showing mean and SD T2 values as well as CoV of the LV myocardium. In Figure 7A, bright- and black-blood images and corresponding T2 maps are shown for four representative subjects in mid-ventricular SAx slice. Comparable T2 values were measured in the septal segments across all volunteers (55.5±2.6)ms. Averaged T2 values from all 3D datasets are displayed with violin plots: (55.9±3.2)ms (apex), (55.9±2.7)ms (mid-cavity), and (55.5±2.9) ms (basal) regions (see Figure 7B). The averaged T2 values from all segments were (55.8±2.9) ms, which is consistent with literature value (13) (see Figure 7C).

Figure 4
Four medical images labeled Subject 1, Subject 2, Subject 7, and Subject 10, showing bright-blood reformats of the heart. Arrows point to highlighted areas, with labels “RCA” (Right Coronary Artery) and “LCA” (Left Coronary Artery) on Subject 1.

Figure 4. High-resolution bright-blood images are illustrated for four subjects in multiplanar reformation. The course of the left (LCA) and right (RCA) coronary arteries is indicated by white arrows.

Figure 5
Series of medical images showing bright-blood, black-blood, and T2 map images of the heart in coronal, vertical long-axis, and horizontal long-axis views. Labels indicate heart structures like the left ventricle (LV), right ventricle (RV), left atrium (LA), right atrium (RA), pulmonary artery (PA), and aorta (Ao). The T2 maps use a color scale ranging from blue to yellow to indicate varying intensities.

Figure 5. 3D reconstructed data for two healthy subjects, showing bright- and black-blood images with co-registered T2 maps in coronal, vertical long-axis (for evaluating the anterior/inferior walls and apex of the LV) and horizontal long-axis (for assessment of LV septal/lateral walls and apex, as well as RV free wall and chamber sizes) orientations. On the black-blood images different intrapericardial structures are indicated as: LV/RV—left/right ventricle, LA/RA—left/right atrium, Ao—ascending aorta, PA—pulmonary artery. To highlight relevant cardiovascular structures, a binary contrast-generated mask was applied to the T2 map dataset.

Figure 6
Cardiac MRI images displaying bright-blood and black-blood imaging sequences, alongside T2 maps, labeled as apex, mid-cavity, and basal. Below, three bullseye diagrams detail myocardial T2 relaxation parameters: mean, standard deviation, and coefficient of variation, with color-coded segments for anterior, lateral, septal, and inferior regions.

Figure 6. Bright- and black-blood images with co-registered T2 maps of SAx cross-section slices from apex to base are displayed for a healthy subject with HR of 67 bpm. The mean T2 values, SD and CoV in each of 16 segments are shown in the bull’s-eye-plot generated by projecting LV myocardial tissue onto a plane.

Figure 7
(A) Cardiovascular MRI images for four subjects, showing bright-blood, black-blood, and T2 maps, with heart rates and T2 times annotated. (B) Violin plots displaying T2 times in milliseconds across left ventricular segments: apex, mid-cavity, and basal. (C) Bar graph illustrating averaged T2 times for ten subjects, with error bars indicating variability.

Figure 7. (A) Anatomical images and mapping shown for four representative subjects in mid-cavity SAx view. (B) Violin plot comparing the mean T2 values from apical, mid-cavity and basal regions. (C) Comparison of T2 mean and SD values across all subjects.

4 Discussion

This study introduced a novel 3D whole-heart CMR sequence for simultaneous high-resolution anatomical imaging and myocardial tissue mapping at 0.55 T in an efficient and easy to plan 9 min scan. A unique sequence design allows multiple techniques to be combined in a single scan. The bright-blood volume is directly obtained from the first HB, while the black-blood volume is created by magnitude subtraction of the first from the second volume. Compared to the T2prep-IR BOOST sequence at 1.5 T (20), the preparation pulses were reordered to improve blood nulling on the directly subtracted images. As the native myocardial T1 relaxation time is shorter at 0.55 T, a shorter inversion time was used compared to 1.5 T. The number of iNAV readouts was set to 6 instead of 14 (typical value at 1.5 T) due to faster recovery of epicardial fat at low-field, as shown in (24). Further adjustments were made to determine the most appropriate combination of T2prep durations based on the prolonged T2 at low-field. By matching the signal evolution with a previously generated EPG dictionary across all three volumes, co-registered T2 maps were generated voxel-wise. The feasibility of the proposed sequence was demonstrated in a phantom study and in a cohort of healthy subjects. Precise and accurate T2 values were observed in good agreement with the SE reference scan in the phantom study. In-vivo anatomical imaging of various intrapericardial structures received high-quality scores and T2 maps provided comparable values across all subjects.

The acquisition of 3D volumes with isotropic resolution facilitated scan planning as cardiac orientations such as 2, 3 and 4 chamber and short-axis views can be obtained in a postprocessing step thereby minimizing scan complexity. The proposed sequence was implemented with in-line reconstruction, allowing the images to be viewed within approximately 3 min after the examination. To enable free-breathing acquisition, a 2D low-resolution iNAV was employed prior to each spiral interleave acquired per HB and used for respiratory binning and intrabin translational motion correction. The total scanning time is linearly related to the length of the acquisition window, the pixel size and the subject-dependent HR. Due to the high acceleration of 4 with the VD-CASPR trajectory, the 3D scan can be completed within a clinically feasible scan time of approximately 8 to 9 min. The acquisition window of the proposed sequence was kept short enough for the mid-diastolic application, a quiescent period of the cardiac cycle, resulting in less movement during the readout. Currently, the trigger delay is adjusted manually based on the CINE acquisition in the axial plane prior to the acquisition, which requires an additional planning step. To further improve image quality, the data is transferred from the scanner and denoised in a post-processing step using HD-PROST (28) in MATLAB, which takes about 10 min with current parameters. Generating one dictionary took less than a minute per subject and 3D T2 maps were available in about two minutes afterwards. Due to the 3D nature and isotropic spatial resolution, the acquired images can be reformatted into any desired view to visualize whole-heart morphology and to analyse different intrapericardial structures. Bright-blood is the reference standard for imaging cardiac structures, great vessels and coronary arteries, whereas black-blood excels in visualizing the myocardium, atrial and vessel walls. The combination of both techniques improves the assessment of cardiac morphological evaluation compared to either technique in isolation. Complex coronary artery geometry prevents visualisation of the entire anatomy in a single slice, therefore multiplanar reformation was performed in coronal and axial views for the RCA and LCA, respectively. The chosen spatial resolution was deemed sufficient to visualise small structures such as the course and anatomy of the coronary arteries. Further reduction in voxel size would reduce the SNR, which may affect dictionary matching and thus precision of the estimated T2 values.

The proposed sequence framework has several limitations. In the anterior segments of LV, the CoVs values are slightly increased possibly due to susceptibility artifacts at the heart-liver-lung interface. Because of the reduced LV-RV wall thickness in the apical septal segments, partial volume artifacts between the myocardium and blood pool are more pronounced, potentially resulting in elevated T2 values. In subjects with HR variations, mapping accuracy and precision are expected to decrease due to deviations from the steady-state signal evolution assumed in the dictionary. Future work will include validation in patients with suspected cardiovascular disease. Whilst the preliminary experiments have shown good agreement between the T2 values of the proposed sequence and the literature, further in-vivo validation against conventional 2D sequences is required for comparison. At the time the sequence was developed and implemented, there were no reference sequences available on the scanner for this purpose. To compensate for the loss of SNR and to obtain good image quality, a high FA of 110 was used for bSSFP acquisition at 0.55 T, which is particularly challenging in the scanner used. The duration of the imaging RF pulses had to be extended to achieve higher FAs, which requires lower RF voltages (24). For higher FA the expected signal behaves more robustly to off-resonance phase accumulation due to B0-inhomogeneity as shown in (34). One of the limitations is also that the T2 estimate depends on the chosen fixed T1, which may affect accuracy and precision in patients with T1s that differ from normal myocardium [e.g., T1=760ms “diseased” tissue vs. 700 ms for “healthy” tissue (13)]. To investigate its impact, additional dictionaries were generated with a coarse T1 range of [600:50:800] ms. No significant statistical difference in T2 was observed when mapping within the T1 range of the additional dictionaries (see Supplementary Material). The high FA resulted in a reduced T1 dependence of the sequence during bSSFP readout, as shown in (34), which is beneficial for T2 mapping. Since T2 elevation in the diseased myocardium is expected to increased by 10 ms to 20 ms, this effect is considered acceptable. Although the proposed sequence was designed to be less sensitive to longitudinal magnetization recovery, simultaneously acquired T1 maps could not only improve the accuracy and precision of T2 mapping but also provide complementary tissue characterization. In future studies, the sequence design will be expanded to include additional interleaves to enable simultaneous T1/T2 mapping.

5 Conclusions

In this work, we successfully demonstrated the feasibility of simultaneous whole-heart bright and black-blood anatomical imaging and T2 mapping in phantoms and healthy subjects at 0.55 T in a single 9 min scan.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

The studies involving humans were approved by Comité Ético Científico de Ciencias de la Salud UC. The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study.

Author contributions

IK: Data curation, Formal analysis, Investigation, Methodology, Writing – original draft, Writing – review & editing. CC-P: Data curation, Writing – review & editing, Investigation. MGC: Investigation, Writing – review & editing. CG: Methodology, Writing – review & editing. SJL: Formal analysis, Writing – review & editing. KPK: Methodology, Resources, Supervision, Writing – review & editing. DCK: Resources, Supervision, Writing – review & editing. MRM: Funding acquisition, Resources, Supervision, Writing – review & editing. DR: Supervision, Writing – review & editing. CP: Conceptualization, Project administration, Resources, Supervision, Writing – review & editing. RMB: Conceptualization, Funding acquisition, Project administration, Resources, Supervision, Writing – review & editing.

Funding

The author(s) declare that financial support was received for the research and/or publication of this article. The authors acknowledge financial support from: (1) Institute for Advanced Study – Technical Univeristy of Munich, (2) BHF programme grant; RG/20/1/34802 and King’s BHF Centre for Award Excellence; RE/24/130035, (3) Wellcome/EPSRC Centre for Medical Engineering; NS/A000049/1, (4) Millennium Institute for Intelligent Healthcare Engineering; ICN2021004, (5) FONDECYT; 1250261 and 1250252, (6) IMPACT, Center of Interventional Medicine for Precision and Advanced Cellular Therapy; FB210024, (7) the Department of Health through the National Institute for Health Research (NIHR) comprehensive Biomedical Research Centre award, (8) NIHR Cardiovascular MedTech Co-operative. The views expressed are those of the authors and not necessarily those of the BHF, NHS, the NIHR or the Department of Health.

Conflict of interest

KPK was employed by Siemens Healthcare Limited.

The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

The author(s) declared that they were an editorial board member of Frontiers, at the time of submission. This had no impact on the peer review process and the final decision.

Generative AI statement

The author(s) declare that no Generative AI was used in the creation of this manuscript.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fcvm.2025.1572318/full#supplementary-material.

References

1. Dattani S, Samborska V, Ritchie H, Roser M. Cardiovascular diseases. Our World in Data (2023). Available at: https://ourworldindata.org/cardiovascular-diseases. (Accessed February 05, 2025).

Google Scholar

2. Kim PK, Hong YJ, Im DJ, Suh YJ, Park CH, Kim JY, et al. Myocardial T1 and T2 mapping: techniques and clinical applications. Korean J Radiol. (2017) 18:113–31. doi: 10.3348/kjr.2017.18.1.113

PubMed Abstract | Crossref Full Text | Google Scholar

3. Messroghli DR, Moon JC, Ferreira VM, Grosse-Wortmann L, He T, Kellman P, et al. Clinical recommendations for cardiovascular magnetic resonance mapping of T1, T2, T2* and extracellular volume: a consensus statement by the society for cardiovascular magnetic resonance (SCMR) endorsed by the European association for cardiovascular imaging (EACVI). J Cardiovasc Magn Reson. (2017) 19:75. doi: 10.1186/s12968-017-0389-8

PubMed Abstract | Crossref Full Text | Google Scholar

4. World Health Organisation. Data from: Magnetic resonance imaging (MRI) units per million people [Global atlas of medical devices]. World Health Organisation – minor processing by Our World in Data (2022). (Accessed October 30, 2024).

Google Scholar

5. Sarracanie M, LaPierre CD, Salameh N, Waddington DEJ, Witzel T, Rosen MS. Low-cost high-performance MRI. Sci Rep. (2015) 5:15177. doi: 10.1038/srep15177

PubMed Abstract | Crossref Full Text | Google Scholar

6. Vosshenrich J, Breit H-C, Bach M, Merkle E. Economic aspects of low-field magnetic resonance imaging: acquisition, installation, and maintenance costs of 0.55 T systems. Radiologe. (2022) 62:400–4. doi: 10.1007/s00117-022-00986-9

PubMed Abstract | Crossref Full Text | Google Scholar

7. Rusche T, Vosshenrich J, Winkel DJ, Donners R, Segeroth M, Bach M, et al. More space, less noise—new-generation low-field magnetic resonance imaging systems can improve patient comfort: a prospective 0.55T–1.5T-scanner comparison. J Clin Med. (2022) 11:6705. doi: 10.3390/jcm11226705

PubMed Abstract | Crossref Full Text | Google Scholar

8. Simonetti OP, Ahmad R. Low-field cardiac magnetic resonance imaging: a compelling case for cardiac magnetic resonance’s future. Circ Cardiovasc Imaging. (2017) 10:e005446. doi: 10.1161/CIRCIMAGING.117.005446

PubMed Abstract | Crossref Full Text | Google Scholar

9. Campbell-Washburn AE, Ramasawmy R, Restivo MC, Bhattacharya I, Basar B, Herzka DA, et al. Opportunities in interventional and diagnostic imaging by using high-performance low-field-strength MRI. Radiology. (2019) 293:384–93. doi: 10.1148/radiol.2019190452

PubMed Abstract | Crossref Full Text | Google Scholar

10. Arnold TC, Freeman CW, Litt B, Stein JM. Low-field MRI: Clinical promise and challenges. J Magn Reson Imaging. (2022) 57:25–44. doi: 10.1002/jmri.28408

PubMed Abstract | Crossref Full Text | Google Scholar

11. Varghese J, Jin N, Giese D, Chen C, Liu Y, Pan Y, et al. Building a comprehensive cardiovascular magnetic resonance exam on a commercial 0.55 T system: a pictorial essay on potential applications. Front Cardiovasc Med. (2023) 10:1120982. doi: 10.3389/fcvm.2023.1120982

PubMed Abstract | Crossref Full Text | Google Scholar

12. Marques JP, Simonis F, Webb AG. Low-field MRI: AN MR physics perspective. J Magn Reson Imaging. (2019) 49:1528–42. doi: 10.1002/jmri.26637

PubMed Abstract | Crossref Full Text | Google Scholar

13. Campbell-Washburn AE, Varghese J, Nayak KS, Ramasawmy R, Simonetti OP. Cardiac MRI at low field strengths. J Magn Reson Imaging. (2024) 59:412–30. doi: 10.1002/jmri.28890

PubMed Abstract | Crossref Full Text | Google Scholar

14. Carr HY. Steady-state free precession in nuclear magnetic resonance. Phys Rev. (1958) 112:1693–701. doi: 10.1103/PhysRev.112.1693

Crossref Full Text | Google Scholar

15. Bieri O, Scheffler K. Fundamentals of balanced steady state free precession MRI. J Magn Reson Imaging. (2013) 38:2–11. doi: 10.1002/jmri.24163

PubMed Abstract | Crossref Full Text | Google Scholar

16. Grizzard J, Judd R, Kim R. Cardiovascular MRI in Practice: A Teaching File Approach. London: Springer (2008).

Google Scholar

17. Correia T, Ginami G, Rashid I, Nordio G, Hajhosseiny R, Ismail TF, et al. Accelerated high-resolution free-breathing 3D whole-heart T2-prepared black-blood and bright-blood cardiovascular magnetic resonance. J Cardiovasc Magn Reson. (2020) 22:88. doi: 10.1186/s12968-020-00691-3

PubMed Abstract | Crossref Full Text | Google Scholar

18. Ginami G, Neji R, Phinikaridou A, Whitaker J, Botnar RM, Prieto C. Simultaneous bright- and black-blood whole- heart MRI for noncontrast enhanced coronary lumen and thrombus visualization. Magn Reson Med. (2017) 79:1460–72. doi: 10.1002/mrm.26815

PubMed Abstract | Crossref Full Text | Google Scholar

19. Milotta G, Ginami G, Cruz G, Neji R, Prieto C, Botnar RM. Simultaneous 3D whole-heart bright-blood and black blood imaging for cardiovascular anatomy and wall assessment with interleaved T2prep-IR. Magn Reson Med. (2019) 82:312–25. doi: 10.1002/mrm.27734

PubMed Abstract | Crossref Full Text | Google Scholar

20. Milotta G, Ginami G, Bustin A, Neji R, Prieto C, Botnar RM. 3D whole-heart free-breathing qboost-T2 mapping. Magn Reson Med. (2020) 83:1673–87. doi: 10.1002/mrm.28039

PubMed Abstract | Crossref Full Text | Google Scholar

21. Castillo-Passi C, Crabb MG, Muñoz C, Tripp D, Kunze KP, Neji R, et al. Simultaneous 3D whole-heart bright-blood and black-blood imaging with inav-based non-rigid motion-corrected reconstruction at 0.55T. In: Proceedings of the International Society for Magnetic Resonance in Medicine. (2023). Vol. 31. p. 1687.

Google Scholar

22. Henningsson M, Koken P, Stehning C, Razavi R, Prieto C, Botnar RM. Whole-heart coronary mr angiography with 2D self-navigated image reconstruction. Magn Reson Med. (2012) 67:437–45. doi: 10.1002/mrm.23027

PubMed Abstract | Crossref Full Text | Google Scholar

23. Prieto C, Doneva M, Usman M, Henningsson M, Greil GF, Schaeffter T, et al. Highly efficient respiratory motion compensated free-breathing coronary mra using golden-step cartesian acquisition. J Magn Reson Imaging. (2015) 41:738–46. doi: 10.1002/jmri.24602

PubMed Abstract | Crossref Full Text | Google Scholar

24. Castillo-Passi C, Kunze KP, Crabb MG, Munoz C, Fotaki A, Neji R, et al. Highly efficient image navigator based 3D whole-heart cardiac MRA at 0.55T. Magn Reson Med. (2024) 93:689–98. doi: 10.1002/mrm.30316

PubMed Abstract | Crossref Full Text | Google Scholar

25. Captur G, Gatehouse PD, Keenan KE, Heslinga FG, Bruehl R, Prothmann M, et al. A medical device-grade T1 and ECV phantom for global T1 mapping quality assurance—the T1 mapping and ECV standardization in cardiovascular magnetic resonance (T1MES) program. J Cardiovasc Magn Reson. (2016) 18:58. doi: 10.1186/s12968-016-0280-z

PubMed Abstract | Crossref Full Text | Google Scholar

26. Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. Sense: sensitivity encoding for fast MRI. Magn Reson Med. (1999) 42:952–62. doi: 10.1002/(SICI)1522-2594(199911)42:5%3C952::AID-MRM16%3E3.0.CO;2-S

PubMed Abstract | Crossref Full Text | Google Scholar

27. Cruz G, Atkinson D, Henningsson M, Botnar RM, Prieto C. Highly efficient nonrigid motion-corrected 3D whole-heart coronary vessel wall imaging. Magn Reson Med. (2017) 77:1894–908. doi: 10.1002/mrm.26274

PubMed Abstract | Crossref Full Text | Google Scholar

28. Bustin A, da Cruz GJL, Jaubert O, López K, Botnar RM, Prieto C. High-dimensionality undersampled patch-based reconstruction (hd-prost) for accelerated multi-contrast MRI. Magn Reson Med. (2019) 81:3705–19. doi: 10.1002/mrm.27694

PubMed Abstract | Crossref Full Text | Google Scholar

29. Weigel M. Extended phase graphs: dephasing, RF pulses, and echoes – pure and simple. J Magn Reson Imaging. (2015) 41:266–95. doi: 10.1002/jmri.24619

PubMed Abstract | Crossref Full Text | Google Scholar

30. Etienne A, Botnar RM, Muiswinkel AMCV, Boesiger P, Manning WJ, Stuber M. “Soap-bubble” visualization and quantitative analysis of 3D coronary magnetic resonance angiograms. Magn Reson Med. (2002) 48:658–66. doi: 10.1002/mrm.10253

PubMed Abstract | Crossref Full Text | Google Scholar

31. Hajhosseiny R, Rashid I, Bustin A, Munoz C, Cruz G, Nazir MS, et al. Clinical comparison of sub-mm high-resolution non-contrast coronary CMR angiography against coronary CT angiography in patients with low-intermediate risk of coronary artery disease: a single center trial. J Cardiovasc Magn Reson. (2021) 23:57. doi: 10.1186/s12968-021-00758-9

PubMed Abstract | Crossref Full Text | Google Scholar

32. Botnar RM, Stuber M, Danias PG, Kissinger KV, Manning WJ. Improved coronary artery definition with t2-weighted, free-breathing, three-dimensional coronary MRA. Circulation. (1999) 99:3139–48. doi: 10.1161/01.CIR.99.24.3139

PubMed Abstract | Crossref Full Text | Google Scholar

33. Cerqueira MD, Weissman N, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. A statement for healthcare professionals from the cardiac imaging committee of the council on clinical cardiology of the American heart association. J Nucl Cardiol. (2002) 92:240–5. doi: 10.1067/mnc.2002.123122

Crossref Full Text | Google Scholar

34. Ganter C. Off-resonance effects in the transient response of ssfp sequences. Magn Reson Med. (2004) 52:368–75. doi: 10.1002/mrm.20173

PubMed Abstract | Crossref Full Text | Google Scholar

Keywords: low-field MRI, 3D whole-heart, simultaneous bright-and black-blood, high-resolution coronary angiography, T2 mapping

Citation: Kokhanovskyi I, Castillo-Passi C, Crabb MG, Ganter C, Littlewood SJ, Kunze KP, Karampinos DC, Makowski MR, Rueckert D, Prieto C and Botnar RM (2025) High-resolution 3D whole-heart bright- and black-blood imaging with co-registered T2 mapping at 0.55 T. Front. Cardiovasc. Med. 12:1572318. doi: 10.3389/fcvm.2025.1572318

Received: 6 February 2025; Accepted: 18 June 2025;
Published: 3 July 2025.

Edited by:

Casper Mihl, Maastricht University Medical Centre, Netherlands

Reviewed by:

Keigo Kawaji, Illinois Institute of Technology, United States
Ye Tian, University of Southern California, Los Angeles, United States

Copyright: © 2025 Kokhanovskyi, Castillo-Passi, Crabb, Ganter, Littlewood, Kunze, Karampinos, Makowski, Rueckert, Prieto and Botnar. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: René M. Botnar, cmVib3RuYXJAdWMuY2w=

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.