Modified stepwise method with the guidance of QDOT MICRO catheter for mitral isthmus ablation in patients with persistent atrial fibrillation

    JL

    Jingchao Li 1

    CZ

    Chaoyu Zhao 1

    LC

    Luqian Cui 1

    XW

    Xin Wang 2

    QF

    Qianqian Feng 3

    HS

    Huihui Song 1

    CD

    Cong Ding 1

    HW

    Hailan Wang 1

    HY

    Haijia Yu 1

    YC

    Yingjie Chu 1

    SD

    Shujuan Dong 1,2

  • 1. Henan Provincial People's Hospital, Zhengzhou, China

  • 2. Henan Provincial Chest Hospital, Zhengzhou, China

  • 3. Zhoukou Central Hospital, Zhoukou, China

Article metrics

View details

79

Views

The final, formatted version of the article will be published soon.

Abstract

Background: Ethanol infusion of the vein of Marshall (EI-VOM) has elevated the success rate of mitral isthmus (MI) block in patients with persistent atrial fibrillation (PeAF). However, the procedure involve the extensive endocardial ablation and epicardial ablation, which brought the operational difficulty and risk. Material and Methods: Patients with PeAF were randomly assigned in a 1:1 ratio to either STSF catheter group using the stepwise method (STSF group) or QDM catheter group using a modified stepwise method (QDM group) for MI ablation. The modified stepwise method was as follows: step 1, The potential of VOM was mapped from endocardium using the QDM catheter. Step 2, EI-VOM. Step 3, precise endocardial ablation guiding by VOM potential. Step 4, QDM catheter was cannulated into the CS for epicardial ablation, especially the ostium of Marshall. The immediate procedural results were compared between the two groups. Results: After excluding 5 patients with unsuccessful EI-VOM, 68 patients were divided into STSF group (36 cases) and QDM group (32 cases). The potential of VOM could be clearly mapped from endocardium using a QDM catheter. Both the accumulated operation time (p=0.032) and ablation time (p<0.001) were significantly shorter in the QDM group compared to the STSF group. QDM group achieved more conduction blocks of MI after a single endocardial line ablation (71.9% vs. 36.1%, p=0.017) with fewer ablation points (p<0.001) compared to the STSF group. The block rate of the MI after endocardial ablation alone was also higher in the QDM group than in the STSF group (90.6% vs. 69.4%; P=0.019), which avoided epicardial ablation. Even if epicardial ablation is necessary, the number of ablation points on the epicardial surface in the QDM group would be fewer than in the STSF group (p<0.001). Conclusions: The QDM catheter can be used to map the potential of VOM from endocardium, thereby facilitating precise endocardial mitral isthmus linear ablation. The modified stepwise approach effectively reduces the number of endocardial ablation points and the likelihood of requiring epicardial ablation. (NCT06145906, ClinicalTrial.gov)

Summary

Keywords

Catheter Ablation, Mitral isthmus, Persistent atrial fibrillation, QDOT MICRO catheter, Vein of Marshall

Received

03 January 2026

Accepted

02 February 2026

Copyright

© 2026 Li, Zhao, Cui, Wang, Feng, Song, Ding, Wang, Yu, Chu and Dong. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) or licensor are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Shujuan Dong

Disclaimer

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Outline

Share article

Article metrics