PERSPECTIVE article

Front. Microbiol., 18 September 2025

Sec. Infectious Agents and Disease

Volume 16 - 2025 | https://doi.org/10.3389/fmicb.2025.1668987

Analysis of antimicrobial effects of a contactless, indirect cold atmospheric plasma-aerosol method for germ reduction on surfaces: an in vitro and in vivo study

  • 1. Faculty of Health and Healthcare Sciences, University of Applied Sciences Zwickau, Zwickau, Germany

  • 2. Department of Nursing and Health, University of Fulda, Fulda, Germany

  • 3. Wound Outpatient Clinic LKH Bregenz, Bregenz, Austria

  • 4. Research and Development, WK-MedTec GmbH, Bückeburg, Germany

  • 5. Lüsche Veterinary Clinic, Bakum, Germany

Abstract

Cold Atmospheric Plasma-Aerosol (CAP-A) technology has emerged as a novel, contactless approach for antimicrobial treatment. This study evaluates the in vivo efficacy of CAP-A in reducing microbial contamination on human skin, building on obligatory in vitro testing. In vitro results demonstrated consistent 3–4.5 log unit microbial reductions across five standard organisms. In vivo evaluation using Escherichia coli revealed a mean log reduction factor of 4.77 (SD ± 0.44), exceeding the 4-log threshold considered clinically relevant. Notably, CAP-A showed comparable efficacy to an alcohol-based reference disinfection method (p = 0.134), without associated drawbacks such as thermal effects or ozone accumulation. Results suggest that CAP-A offers equivalent in vivo efficacy compared to previously documented CAP methods while minimizing tissue damage, thermal changes, and discomfort. The results underscore the potential of CAP-A as an effective and tolerable alternative to established CAP approaches, warranting further comparative research under standardized conditions. Future studies should examine both CAP and CAP-A technologies, broadening the spectrum of tested microorganisms, incorporating additional parameters, and rigorously assessing benefits and risks. This research could elucidate the underlying mechanisms driving differences in efficacy and side effect profiles, ultimately contributing to the optimization of plasma-based treatments in clinical and industrial settings.

1 Introduction

Plasma, the fourth state of matter, is a partially ionized gas composed of charged particles. In medical applications, Cold Atmospheric Plasmas (CAP) are increasingly explored for their antimicrobial effects. Particularly, plasmas generated from oxygen and water vapor yield reactive oxygen species (ROS) capable of modifying surfaces, liquids, and even ambient air through energy transfer during ionization processes (; ; ). Until recently, most research has focused on direct plasma applications, where plasma is applied directly to the target surface. These methods have shown strong microbiocidal effects in wound sanitation, effectively reducing bacterial load regardless of wound size, location or cofactors (e.g., tobacco abuse, cardiovascular diseases) (; Apelqvist et al., 2024; ). Notably, demonstrated in vivo efficacy against both Gram-positive and Gram-negative bacteria, including resistant strains such as MRSA and MRGN, indicating that cold plasma damages microbial structures in a way that bypasses typical resistance mechanisms.

A drop in wound pH after plasma treatment further supports its antimicrobial efficacy, as pathogenic bacteria typically alkalize the wound environment. CAP treatment restores the acidic milieu necessary for optimal healing (). Moreover, studies report significantly accelerated wound healing and increased healing rates, up to 60% higher compared to supportive therapy alone (; ; ; ; ; ). According to , these improvements are independent of wound type or location. The observed efficacy is attributed primarily to the antimicrobial action, suggesting that indirect plasma methods may yield similar results.

Building on these findings, a novel indirect plasma approach - Cold Atmospheric Plasma-Aerosol (CAP-A) - has been introduced. CAP-A combines reactive species from CAP with nebulized water, enabling antimicrobial action without direct contact between the plasma source and the treated surface (; ). Energy input is limited to plasma generation, making the method safer and more flexible in clinical applications. Indirect methods like CAP-A target the transient skin flora via redox disruption and induce a “selective short circuit” in microbial membranes, collapsing the membrane potential and disabling transmembrane transport ().

Importantly, these effects are highly selective. Human cells remain unaffected due to protective antioxidant enzymes such as catalase, superoxide dismutase, and alpha-1-antitrypsin (). Toxicological evaluations of CAP-A, including the Ames test (EN ISO 10993-3) and cytotoxicity tests (EN ISO 10993-5), confirm its safety profile. No mutagenic or cytotoxic effects were detected in bacteria or human cells, respectively. Clinical use of direct CAP devices has also shown no adverse effects, reinforcing the biological selectivity and clinical safety of this emerging technology ().

The analysis builds on previous work demonstrating the microbicidal efficacy of CAP-A technology in diverse experimental and practical settings including the disinfection of thermolabile medical devices under routine hospital conditions () and the successful treatment of clinical infections in veterinary patients (). This study represents a first EN 1500-compliant in vivo investigation of CAP-A on human skin using a specified test germ for clear statistical and clinical evidence. Further, the test setting combines two standards for assessing plasma techniques. EN 1500 presents the closest standard to CAP-A for surface-active reduction of microbial load in a clinical setting and DIN SPEC 91315 defines the measurement of inactivation potency of plasma sources. The two norms applied for both chemical disinfectants (EN 1500) and plasma devices (DIN SPEC 91315) demonstrate the reliability of microbial inactivation results of CAP-A. This article critically examines how effective is the CAP-A procedure in reducing microbial contamination on human skin under in vivo conditions, following the obligatory preliminary in vitro testing as required by best practice. Furthermore, directly comparing its antimicrobial performance with the established alcohol-based reference method, this work addresses a critical gap between laboratory-based findings and clinically relevant, standardized testing.

How does the antimicrobial performance of CAP-A, evaluated in alignment with the reference alcohol-based disinfection method under EN 1500 conditions, compare with the documented efficacy of existing CAP procedures in current literature? Can these findings provide insights into the future role and integration of CAP and CAP-A methods in clinical treatment practices?

2 Subsections relevant for the subject

2.1 Working principle

The PLASMO®HEAL device (WK-MedTec GmbH, Bückeburg) operates via an indirect plasma method. A defined alternating voltage (1.45 kV, 38 kHz, sinusoidal) generates cold plasma reaction products, primarily hydroxyl radicals, from ambient air, while only negligible amounts of ozone are produced. Distilled water (stabilized according to German pharmacopoeia for hygienic reasons) is nebulized ultrasonically and activated by the plasma-enriched air, thereby increasing the electrophysical potential of the aerosol without altering its material composition. The resulting aerosol (CAP-A) is then applied to the targeted surface (e.g., skin or wounds) at a distance of 7.5 cm for three minutes, as specified in the manufacturer’s operating instructions to ensure homogeneous distribution across the treatment area and optimal antimicrobial efficacy. In accordance with EN 1500 test conditions, post-treatment sampling in this study was performed 40 min after application, allowing a preliminary assessment of the medium-term persistence of the antimicrobial effect.

2.2 In vitro testing

Microbial reduction efficacy was assessed in vitro against five standard organisms (Staphylococcus aureus ATCC 6538, Staphylococcus epidermidis ATCC 14990, Escherichia coli NCTC 10538, Pseudomonas aeruginosa ATCC 10145, and Candida albicans ATCC 10321) according to DIN 91315:2014, with each organism tested three times. Identification of the organisms was confirmed by Gram staining, metabolic profiling, and indole production. Cultures were grown on Caso agar in 60-mm Petri dishes and adjusted to McFarland standard 8 with sterile, pyrogen-free 0.9% NaCl. A microbial suspension was applied to stainless steel specimens (100 × 10 × 1 mm; surface roughness 100 μm) using a sterile swab, followed by a five-second immersion, drainage, and two-hour air drying. The specimens were then exposed to the PLASMO®HEAL aerosol for three minutes. Microorganisms were recovered by vortexing each sample in 10 mL sterile 0.9% NaCl for 30 s; serial tenfold dilutions (effective dilution from 100 to 10−3) were plated (0.1 mL per dilution) alongside untreated controls. Following incubation, colony counts (from plates with 10–100 colonies) were used to calculate microbial loads and determine the log reduction factors.

2.3 In vivo testing

In vivo antimicrobial efficacy was evaluated in accordance with EN 1500 using Escherichia coli NCTC 10538, a non-pathogenic reference strain specified for standardized hygienic handrub testing. Eighteen subjects participated as part of routine hospital hand hygiene quality assurance (not as a clinical trial). Both hands were contaminated by immersing the fingers in a test suspension for five seconds, then drained. The left hand provided a pre-treatment bacterial count, while the right hand, after contamination, was exposed to the PLASMO®HEAL aerosol for three minutes before post-treatment sampling. Serial dilutions and culture methods analogous to the in vitro procedure produced log reduction factors. E. coli, an apathogenic yet resilient organism, served as a suitable surrogate for pathogenic microbes. The number of test subjects (n = 18) was determined in accordance with the requirements of EN 1500 in order to ensure sufficient statistical significance for the evaluation of hygienic hand disinfection procedures. All procedures were conducted in accordance with relevant guidelines; participants provided written informed consent, and an ethical waiver was secured from the institutional review board (recruitment: January 2–9, 2023).

2.4 Statistical analysis

Descriptive and inferential statistics were performed using IBM Statistics Version 30 for Mac. Given the small sample size, the Shapiro–Wilk test assessed normality for the log-CFU parameter (pre: p = 0.025; post: p = 0.191) and for the log-CFU reduction factor (p < 0.001), with normal distribution confirmed only for post-treatment log-CFU values. Simulation studies support the t-test’s robustness to normality violations (). Accordingly, a paired samples t-test (α = 5%, two-tailed) compared pre- and post-treatment log-CFU values in the in vivo testing, while an independent samples t-test (α = 5%, two-tailed), preceded by Levene’s test for equality of variances (p = 0.0502), compared the log-CFU reduction factors between the CAP-A and the alcohol-based reference disinfection. The complete data set is included as an e-supplement.

3 Results

Detailed analyses of the in vitro results (Table 1) and in vivo outcomes (Table 2) provide a comprehensive assessment of the dynamic effects of this innovative wound treatment method. In the in vivo evaluation employing the CAP-A procedure, the mean log CFU prior to application was 7.23 (SD ± 0.59), which decreased to 2.52 (SD ± 0.38) post-procedure, yielding a mean log reduction factor of 4.77 (SD ± 0.44). A paired samples t-test confirmed that the reduction in log CFU values was statistically significant (p < 0.001). Furthermore, comparison of the log CFU reduction factors between the CAP-A treatment and the alcohol-based reference disinfection revealed no significant difference, as indicated by Levene’s test for homogeneity of variances (p = 0.0502) and an independent samples t-test (p = 0.134). Based on the in vivo results, the CAP-A procedure achieved a disinfection efficacy exceeding a 4-log reduction, thereby confirming its effectiveness. Moreover, the statistical analyses indicated no significant difference between the CAP-A method and the alcohol-based reference disinfection, suggesting that the two are equivalent in performance, even though current EN 1500 standards are not yet tailored to accommodate such innovative approaches.

Table 1

No.V1V2CfU/Agar plateCfU/ApproachDilutionCfU/ProbandLog CfU/ProbandLog reductions factor
Staphylococcus aureus 3 min
1.1. A-310.0 mL0.1 mL878,70010,00087,000,0007.94
1.1. B-310.0 mL0.1 mL282,80012,8003.454.49
1.2. A-310.0 mL0.1 mL565,60010,00056,000,0007.75
1.2. B-310.0 mL0.1 mL262,60012,6003.414.33
1.3. A-310.0 mL0.1 mL787,80010,00078,000,0007.89
1.3. B-310.0 mL0.1 mL222,20012,2003.344.55
Staphylococcus epidermidis 3 min
1.1. A-310.0 mL0.1 mL565,60010,00056,000,0007.75
1.1. B-310.0 mL0.1 mL313,10013,1003.494.26
1.2. A-310.0 mL0.1 mL444,40010,00044,000,0007.64
1.2. B-310.0 mL0.1 mL252,50012,5003.404.25
1.3. A-310.0 mL0.1 mL767,60010,00076,000,0007.88
1.3. B-310.0 mL0.1 mL373,70013,7003.574.31
Escherichia coli 3 min
1.1. A-310.0 mL0.1 mL929,20010,00092,000,0007.96
1.1. B-310.0 mL0.1 mL545,40015,4003.734.23
1.2. A-310.0 mL0.1 mL787,80010,00078,000,0007.89
1.2. B-310.0 mL0.1 mL343,40013,4003.534.36
1.3. A-310.0 mL0.1 mL656,50010,00065,000,0007.81
1.3. B-310.0 mL0.1 mL393,90013,9003.594.22
Pseudomonas aeruginosa 3 min
1.1. A-310.0 mL0.1 mL727,20010,00072,000,0007.86
1.1. B-310.0 mL0.1 mL262,60012,6003.414.44
1.2. A-310.0 mL0.1 mL616,10010,00061,000,0007.79
1.2. B-310.0 mL0.1 mL343,40013,4003.534.25
1.3. A-310.0 mL0.1 mL939,30010,00093,000,0007.97
1.3. B-310.0 mL0.1 mL454,50014,5003.654.32
Candida albicans 3 min
1.1. A-310.0 mL0,1 mL787,80010,0007,800,0006.89
1.1. B-310.0 mL0,1 mL313,10013,1003.493.40
1.2. A-310.0 mL0,1 mL595,90010,00059,000,0007.77
1.2. B-310.0 mL0,1 mL272,70012,7003.434.34
1.3. A-310.0 mL0,1 mL868,60010,00086,000,0007.93
1.3. B-310.0 mL0,1 mL292,90012,9003.464.47

Results of the in vitro test of the germ-reducing effect of the CAP-A method.

No., identification of the test run, where the first two digits stand for the test person number; ‘A’, test run before disinfection (initial bacterial count); ‘B’, test run after disinfection (initial bacterial count); V1, total volume approach to leaching; V2, volume of the batch from V1, which was applied to the agar plate; CfU, colony forming unit.

Table 2

No.V1V2CfU/Agar plateCfU/ApproachDilutionCfU/ProbandLog CfU/ProbandLog reductions factorMean values
Log RF
01 before10.0 mL0.1 mL464,60010,00046,000,0007.66
01 after10.0 mL0.1 mL330013002.485.195.19
02 before10.0 mL0.1 mL181,80010,00018,000,0007.26
02 after10.0 mL0.1 mL550015002.704.564.56
03 before10.0 mL0.1 mL161,600100,000160,000,0008.20
03 after10.0 mL0.1 mL141,40011,4003.155.065.06
04 before10.0 mL0.1 mL232,3001,0002,300,0006.36
04 after10.0 mL0.1 mL220012002.304.064.06
05 before10.0 mL0.1 mL848,40010,00084,000,0007.92
05 after10.0 mL0.1 mL550015002.705.235.23
06 before10.0 mL0.1 mL151,50010,00015,000,0007.18
06 after10.0 mL0.1 mL330013002.484.704.70
07 before10.0 mL0.1 mL440010,000400,0005.60
07 after10.0 mL0.1 mL110011002.003.603.60
08 before10.0 mL0.1 mL929,20010,00092,000,0007.96
08 after10.0 mL0.1 mL171,700117003.234.734.73
09 before10.0 mL0.1 mL393,90010,00039,000,0007.59
09 after10.0 mL0.1 mL550015002.704.894.89
10 before10.0 mL0.1 mL242,40010,00024,000,0007.38
10 after10.0 mL0.1 mL110011002.005.385.38
11 before10.0 mL0.1 mL171,70010,00017,000,0007.23
11 after10.0 mL0.1 mL550015002.704.534.53
12 before10.0 mL0.1 mL111,10010,00011,000,0007.04
12 after10.0 mL0.1 mL110011002.005.045.04
13 before10.0 mL0.1 mL262,60010,00026,000,0007.41
13 after10.0 mL0.1 mL220012002.305.115.11
14 before10.0 mL0.1 mL141,40010,00014,000,0007.15
14 after10.0 mL0.1 mL440014002.604.544.54
15 before10.0 mL0.1 mL141,40010,00014,000,0007.15
15 after10.0 mL0.1 mL110011002.005.155.15
16 before10.0 mL0.1 mL393,90010,00039,000,0007.59
16 after10.0 mL0.1 mL990019002.954.644.64
17 before10.0 mL0.1 mL212,10010,00021,000,0007.32
17 after10.0 mL0.1 mL330013002.484.854.85
18 before10.0 mL0.1 mL141,40010,00014,000,0007.15
18 after10.0 mL0.1 mL440014002.604.544.54
Mean value log RF4.77
Median log RF4.79

Results of the in vivo test of the germ-reducing effect of the CAP-A method.

No., identification of the test run; V1, total volume approach to leaching; V2, volume of the batch from V1, which was applied to the agar plate; CfU, colony forming unit; RF, reduction factor.

4 Discussion

This study evaluated the in vivo antimicrobial efficacy of the CAP-A procedure using the WK-MedTec GmbH PLASMO®HEAL device against the test organism E. coli. Our findings demonstrated a mean log reduction factor of 4.77, indicating a robust antimicrobial effect that is statistically significant. When considered alongside previous reports on CAP-based interventions, these results suggest that CAP-A can achieve antimicrobial reductions in a range comparable to those documented for other CAP methods, despite differences in organisms tested, exposure times, device configurations, and application sites across studies. For instance, Zimmermann () achieved a mean log reduction factor of over 3 CFU in vitro, while Boekema et al. reported an in vivo log reduction factor of 2.9 (). Similarly, demonstrated a reduction factor of 2.5 following a 2.5-min CAP treatment. In contrast, our study’s achieved value of 4.77 supports the consideration of CAP-A as a viable, EN 1500-compliant alternative to direct CAP. have shown that both CAP and CAP-A techniques are capable of producing high log reduction factors. However, where CAP procedures have been associated with heightened power input and consequential effects, such as increased production of nitrate and ozone and augmented local temperature (; ), CAP-A appears to achieve comparable bacterial inactivation while potentially reducing these secondary side effects. In addition to antimicrobial efficacy, potential skin-related adverse effects were evaluated. Skin irritation testing performed in accordance with DIN EN ISO 10993-23:2021–10 demonstrated no evidence of erythema or edema at any observation point, resulting in a Primary Irritation Index (P. I. I.) of 0.0. These findings indicate that CAP-A treatment, when applied as described in this study, is unlikely to induce skin irritation, supporting its potential suitability for repeated or prolonged clinical use. In particular, CAP-A’s lower energy requirement, combined with its high antimicrobial efficacy, positions it as a promising alternative in applications where minimizing thermal alterations and chemical by-products is desirable.

Although the current corpus of literature does not provide an exhaustive depiction of the pooled log reduction factors achievable by CAP procedures, the results presented here underscore that modern CAP-A technologies are not necessarily inferior. On the contrary, the promising log reduction factor of 4.77 achieved in our study supports the hypothesis that CAP-A may serve as an efficacious alternative to traditional CAP methods without the concomitant drawbacks of increased reactive species production and thermal side effects.

In accordance with EN 1500 and DIN SPEC 91315, E. coli NCTC 10538 was selected for the in vivo evaluation to ensure standardization and comparability with established disinfectant testing protocols. The in vitro component was designed to extend this evaluation to a broader microbial spectrum, including Gram-positive and Gram-negative bacteria, thereby enabling a representative evaluation within the framework of recognized validation standards for plasma-based and chemical disinfectants. In summary, our findings indicate that CAP-A, by achieving a significant in vivo log reduction factor while mitigating the adverse effects noted in conventional CAP treatment, holds considerable potential as an alternative antimicrobial strategy. This study provides a meaningful impetus for further investigation into the comparative efficiencies and mechanistic underpinnings of CAP versus CAP-A procedures.

5 Limitations and future research

This study has certain limitations that should be considered when interpreting the results. The in vivo evaluation was performed exclusively on healthy volunteers under standardized EN 1500 conditions, which may not fully reflect the complexity of clinical environments such as chronic wound care or high-contamination scenarios. The use of a single non-pathogenic test organism (Escherichia coli) provides a controlled model but does not account for the spectrum of microorganisms encountered in practice, including resistant strains, spore-formers, or fungal biofilms. In addition, the relatively small sample size may limit the detection of inter-individual variability, and the study did not investigate possible cumulative effects or durability of microbial reduction over time. Future research should therefore focus on assessing CAP-A performance in patient populations with different skin conditions, compromised immunity, or existing infections. Moreover, studies under realistic clinical workflow constraints, including integration within existing hygiene protocols, will be important. In this study, post-treatment microbial counts were obtained 40 min after the 3-min CAP-A application, as recommended in the device’s operating protocol. While this interval under EN 1500 conditions offers a preliminary indication of medium-term persistence, it does not substitute for a dedicated long-term evaluation of antimicrobial durability.

Given these observations, we advocate for further comparative research that examines both CAP and CAP-A technologies in parallel. Future studies should broaden the spectrum of tested microorganisms, incorporate additional parameters such as energy consumption and patient comfort, and rigorously assess both the short- and long-term benefits and risks in extended follow up studies associated with these antimicrobial methodologies. Such research could elucidate the underlying mechanisms driving the differences in efficacy and side effect profiles, ultimately contributing to the optimization of plasma-based treatments in both clinical and industrial settings.

6 Conclusion

This study demonstrates that the CAP-A method, as applied with the PLASMO®HEAL device, achieves robust antimicrobial activity both in vitro and in vivo. CAP-A achieved consistently high log reduction factors for five standard microorganisms in vitro, and in vivo tests against Escherichia coli showed a mean reduction of 4.77, exceeding the clinically relevant threshold of 4. Importantly, CAP-A’s performance is statistically equivalent to that of an alcohol-based reference method, while in addition avoiding potential drawbacks of conventional CAP techniques. These results position CAP-A as a promising alternative antimicrobial strategy with high efficacy, good safety, and broader clinical applicability. By combining contactless application and inactivation of microorganisms, CAP-A offers advantages in situations where tissue preservation and patient comfort are paramount. Future research should expand the range of target microorganisms, evaluate long-term safety, and compare CAP-A with other new disinfection technologies under standardized clinical conditions. The integration of CAP-A as standardized adjunctive therapeutic step into clinical protocols could represent an effective, patient-friendly innovation. CAP-A has the potential of improving infection control and wound care in human and veterinary medicine.

Statements

Data availability statement

The original contributions presented in the study are included in the article/supplementary material, further inquiries can be directed to the corresponding author.

Ethics statement

A confirmation has been obtained from the Ethics Committee of the West Saxon University of Applied Sciences Zwickau that no ethics approval is required for the study. The participants provided written informed consent.

Author contributions

ToS: Writing – original draft, Resources, Formal analysis, Supervision, Writing – review & editing. US: Formal analysis, Writing – original draft, Project administration, Investigation, Conceptualization. GH: Writing – review & editing, Methodology. RF: Writing – review & editing. TiS: Writing – review & editing. SK: Writing – review & editing, Data curation. MK: Writing – review & editing, Data curation. TT: Writing – original draft, Visualization, Writing – review & editing.

Funding

The author(s) declare that no financial support was received for the research and/or publication of this article.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Generative AI statement

The authors declare that no Gen AI was used in the creation of this manuscript.

Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

  • 1

    Abu RachedN.KleyS.StorckM.MeyerT.StückerM. (2023). Cold plasma therapy in chronic wounds—a multicenter, randomized controlled clinical trial (plasma on chronic wounds for epidermal regeneration study): preliminary results. J. Clin. Med.12:5121. doi: 10.3390/jcm12155121

  • 2

    ApelqvistJ.RobsonA.HelmkeA.RousseauA.BoekemaB.denBraberE.et al. (2024). Cold Plasma: An Emerging Technology for Clinical Use in Wound Healing. J Wound Management, 25:S1S84. doi: 10.35279/jowm2024.25.03.sup01

  • 3

    BoekemaB.StoopM.VligM.Van LiemptJ.SobotaA.UlrichM.et al. (2021). Antibacterial and safety tests of a flexible cold atmospheric plasma device for the stimulation of wound healing. Appl. Microbiol. Biotechnol.105, 20572070. doi: 10.1007/s00253-021-11166-5

  • 4

    BranýD.DvorskáD.HalašováE.ŠkovierováH. (2020). Cold atmospheric plasma: a powerful tool for modern medicine. Int. J. Mol. Sci.21:2932. doi: 10.3390/ijms21082932

  • 5

    DaeschleinG.NappM.LutzeS.ArnoldA.Von PodewilsS.GuembelD.et al. (2015). Skin and wound decontamination of multidrug-resistant bacteria by cold atmospheric plasma coagulation. JDDG J Dtsch Dermatol Ges.13, 143149. doi: 10.1111/ddg.12559

  • 6

    DaeschleinG.ScholzS.AhmedR.Von WoedtkeT.HaaseH.NiggemeierM.et al. (2012). Skin decontamination by low-temperature atmospheric pressure plasma jet and dielectric barrier discharge plasma. J. Hosp. Infect.81, 177183. doi: 10.1016/j.jhin.2012.02.012

  • 7

    EhlbeckJ.SchnabelU.PolakM.WinterJ.Von WoedtkeT.BrandenburgR.et al. (2011). Low temperature atmospheric pressure plasma sources for microbial decontamination. J Phys Appl Phys.44:013002. doi: 10.1088/0022-3727/44/1/013002

  • 8

    GelkerM.MrotzekJ.IchterA.Müller-GoymannC. C.ViölW. (2019). Influence of pulse characteristics and power density on stratum corneum permeabilization by dielectric barrier discharge. Biochimica et Biophysica Acta1863, 15131523. doi: 10.1016/j.bbagen.2019.05.014

  • 9

    GravesD. B.BauerG. (2018). “Key roles of reactive oxygen and nitrogen species” in Comprehensive clinical plasma medicine. eds. MetelmannH. R.Von WoedtkeT.WeltmannK. D. (Cham: Springer International Publishing), 7182. doi: 10.1007/978-3-319-67627-2_4

  • 10

    HämmerleG.AscherS.GebhardtL. (2023). Positive effects of cold atmospheric plasma on pH in wounds: a pilot study. J. Wound Care32, 530536. doi: 10.12968/jowc.2023.32.9.530

  • 11

    IsbaryG.HeinlinJ.ShimizuT.ZimmermannJ. L.MorfillG.SchmidtH.-U.et al. (2012). Successful and safe use of 2 min cold atmospheric argon plasma in chronic wounds: results of a randomized controlled trial. Br. J. Dermatol.167, 404410. doi: 10.1111/j.1365-2133.2012.10923.x

  • 12

    IsbaryG.MorfillG.SchmidtH. U.GeorgiM.RamrathK.HeinlinJ.et al. (2010). A first prospective randomized controlled trial to decrease bacterial load using cold atmospheric argon plasma on chronic wounds in patients: argon plasma significantly decreases bacteria on wounds. Br. J. Dermatol.163, 7882. doi: 10.1111/j.1365-2133.2010.09744.x

  • 13

    KurrasS.KnottenbeltD.SchmelzU.SchaalT.TischendorfT.FuchsR.et al. (2025). Cold atmospheric plasma-aerosol treatment of equine dermatophytosis: a novel therapeutic approach. Front Vet Sci.12:1651946. doi: 10.3389/fvets.2025.1651946

  • 14

    LanC.ZhuH.WangS.NieL.LiuD.ShiQ.et al. (2024). Disinfection of viruses with cold atmospheric-pressure plasmas: sources, mechanisms, and efficacy. Plasma Process. Polym.21:2300183. doi: 10.1002/ppap.202300183

  • 15

    LerougeS.WertheimerM. R.YahiaL. (2001). Plasma sterilisation: a review of parameters, mechanisms, and limitations. Plasmas Polym.6, 175188. doi: 10.1023/A:1013196629791

  • 16

    MasurK.SchmidtJ.StürmerE.Von WoedtkeT. (2018). Kaltes Plasma zur Heilung chronischer Wunden. Wund Managt.12, 253259.

  • 17

    PerniS.ShamaG.HobmanJ. L.LundP. A.KershawC. J.Hidalgo-ArroyoG. A.et al. (2007). Probing bactericidal mechanisms induced by cold atmospheric plasmas with Escherichia coli mutants. Appl. Phys. Lett90. doi: 10.1063/1.2458162

  • 18

    RaschD.GuiardV. (2004). The robustness of parametric statistical methods. Psychol. Sci.46, 175208.

  • 19

    SchaalT.SchmelzU. (2024). Author correction: plasma disinfection procedures for surfaces in emergency service vehicles: a field trial at the German red Cross. Sci. Rep.14:4315. doi: 10.1038/s41598-024-54622-8

  • 20

    SchaalT.SchmelzU.PittenF. A.TischendorfT. (2025). New approaches to disinfection of thermolabile medical devices using an indirect method with cold atmospheric plasma-aerosol. Sci. Rep15. doi: 10.1038/s41598-025-03364-2

  • 21

    SchachlJ.KönigshoferM.StoiberM.SochaM.GraslC.AbartT.et al. (2025). Cold atmospheric plasma therapy as a novel treatment for Berlin heart EXCOR pediatric cannula infections. Artif. Organs49, 256265. doi: 10.1111/aor.14869

  • 22

    StratmannB.CosteaT. C.NolteC.HillerJ.SchmidtJ.ReindelJ.et al. (2020). Effect of cold atmospheric plasma therapy vs standard therapy placebo on wound healing in patients with diabetic foot ulcers: a randomized clinical trial. JAMA Netw. Open3:e2010411. doi: 10.1001/jamanetworkopen.2020.10411

  • 23

    StrohalR.DietrichS.MittlböckM.HämmerleG. (2022). Chronic wounds treated with cold atmospheric plasmajet versus best practice wound dressings: a multicenter, randomized, non-inferiority trial. Sci. Rep.12:3645. doi: 10.1038/s41598-022-07333-x

  • 24

    TischendorfT.SchaalT.SchmelzU. (2024). Study on hand disinfection in inpatient geriatric care on the superiority of cold plasma aerosol versus alcohol-based disinfection methods in a parallel group design. Sci. Rep.14. doi: 10.1038/s41598-024-72524-7

  • 25

    ZimmermannJ. L.DumlerK.ShimizuT.MorfillG. E.WolfA.BoxhammerV.et al. (2011). Effects of cold atmospheric plasmas on adenoviruses in solution. J. Phys. Appl. Phys.44:505201. doi: 10.1088/0022-3727/44/50/505201

Summary

Keywords

cold plasma, wound treatment, chronic wounds, germ reduction, CAP, CAP-A

Citation

Schaal T, Schmelz U, Hämmerle G, Fuchs R, Schorling T, Kurras S, Koene M and Tischendorf T (2025) Analysis of antimicrobial effects of a contactless, indirect cold atmospheric plasma-aerosol method for germ reduction on surfaces: an in vitro and in vivo study. Front. Microbiol. 16:1668987. doi: 10.3389/fmicb.2025.1668987

Received

18 July 2025

Accepted

01 September 2025

Published

18 September 2025

Volume

16 - 2025

Edited by

Axel Cloeckaert, Institut National de recherche pour l’agriculture, l’alimentation et l’environnement (INRAE), France

Reviewed by

Cristin Coman, National Institute of Medical-Military Research and Development, Romania

Mirjana Mihailovic, Decision Research, United States

Akram Abbasi, Brown University, United States

Updates

Copyright

*Correspondence: Tim Tischendorf,

Disclaimer

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Outline

Cite article

Copy to clipboard


Export citation file


Share article

Article metrics