The value of Fast Dixon combined with deep learning technology in contrast agent-free high-resolution magnetic resonance imaging of the brachial plexus

Si XieSi Xie1Xuehui DuXuehui Du2Zhitao ZhangZhitao Zhang2Guangxiang Chen
Guangxiang Chen1*
  • 1Department of Radiology, The Affiliated Hospital of Southwest Medical University, Luzhou, China
  • 2Siemens Healthineers, Shanghai, China

Introduction: This study aimed to investigate the application of Fast Dixon combined with the deep resolve gain (DRG) technique for enhancing the image quality of the brachial plexus on high-resolution MRI without the use of contrast agents.

Methods: Heavily T2–weighted Fast Dixon high-resolution coronal thin-slice magnetic resonance imaging was conducted on 19 social volunteers. Post-scan, the original data underwent reconstruction using deep learning-based denoising technology. Subjective quality scores were assigned to both the original and MIP images, and those processed with and without denoising technology were compared. The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) values for each segment of the bilateral brachial plexus were measured and analyzed to assess image quality.

Results: Subjective evaluations revealed that the quality of both original thin-slice and thin-MIP images processed with the DRG significantly outperformed those processed without the DRG (original thin-slice p = 0.005, thin-MIP p < 0.05). The bilateral SNRs and CNRs of each anatomical structure (root, trunk, cord, branch) of the brachial plexus in the experimental group with DRG were significantly greater than those in the control group without DRG (p < 0.05), as follows: the SNRs of the bilateral nerve roots increased by 35.1–36.2%, the SNRs of the bilateral nerve trunks increased by 40.6–40.8%, and the SNRs of the bilateral nerve cords and branches increased by about 40–45%. The CNR of the bilateral nerve roots increased by 43.1–44.6%, the CNR of the bilateral nerve trunks increased by 41.8–41.7%, and the CNR of the bilateral nerve cords and branches increased by 47.3–50.6% (root p < 0.001, trunk p < 0.001, cord p = 0.001, branch p = 0.011).

Conclusion: Fast Dixon T2WI can enhance the visibility of brachial plexus segments to a certain extent through DRG denoising technology, which may be an effective means to visualize the brachial plexus without contrast agent.

1 Introduction

The brachial plexus consists of the C5–C8 cervical and T1 thoracic nerve roots, traversing between the anterior and middle scalene muscles. Small branches of the plexus are located behind the pectoralis minor muscle and are distributed in the axilla, scapular region, and medial bicipital groove (1, 2). Its deep anatomical location, complex trajectory, and proximity to adjacent blood vessels present significant challenges in acquiring high-quality magnetic resonance imaging (MRI) of this structure (3).

Currently, two principal approaches are employed for brachial plexus MRI. The first is nerve imaging, which is based on diffusion-weighted imaging (DWI), leveraging the restricted movement of water molecules caused by the nerve myelin sheath and perineurium, which generates high signals on DWI (4). Additionally, diffusion tensor imaging (DTI), derived from DWI, enables visualization of the brachial plexus as fiber bundles (5, 6). The second approach relies on T2–weighted imaging, which uses differences in transverse relaxation between low-protein water molecules in the endoneurium and surrounding tissues. This method accentuates the signal from nerve fiber bundles, delineating the structural morphology of peripheral nerves (7). Among the available techniques, 3D TSE-STIR remains the clinical mainstay for brachial plexus imaging. While this sequence demonstrates substantial potential in nerve trajectory visualization through multiplanar reconstruction and maximum intensity projection (MIP) (8, 9), it faces critical limitations in radiofrequency field (RFF) homogeneity, particularly at higher field strengths. These B1 inhomogeneities frequently induce localized fat suppression failure (10), where residual fat signals obscure critical neural structures, compromising diagnostic confidence in anatomically complex regions. To address this technical limitation, the present study introduces the Fast Dixon T2WI sequence, which combines water-fat separation algorithms for enhanced background suppression, thereby improving nerve-to-background contrast, with thin-slice, zero-gap acquisition that enables high-resolution, high-contrast quasi-three-dimensional imaging of the brachial plexus without the use of exogenous contrast agents. Although the reduction in voxel volume leads to a decrease in the original image’s signal-to-noise ratio (SNR), this drawback is mitigated through the integration of an artificial intelligence (AI)-based deep learning denoising technique, thereby enhancing the overall image quality of the brachial plexus.

For MR images, the main noise sources are the electronic noise generated in the signal receiving circuit and the dielectric and inductive coupling effects caused by the conducting medium in the body. The SNR refers to the ratio of the MR signal-to-noise, which is the core parameter used to measure the quality of MR images. The traditional methods used to improve the SNR include changing the repetition time (TR), echo time (TE), FOV, basic resolution, flip angle (FA), bandwidth, etc., but are usually accompanied by a certain extent of scan time extension. The traditional approaches for improving the signal-to-noise ratio (SNR) involve parameter adjustments such as repetition time (TR), echo time (TE), field of view (FOV), matrix size, flip angle (FA), and bandwidth, but these modifications typically incur prolonged scan durations. With the rapid development of deep learning algorithms in the field of medical imaging, the reconstruction of physical–mathematical constraints on MR images has overcome the triangular challenges of the resolution, SNR, and acquisition time speed in traditional reconstruction. Its application has now appeared in various research publications (1115). The deep resolve gain (DRG) algorithm in this study leverages its inherent capability for noise reduction without requiring additional scanning time (15).

2 Materials and methods

2.1 General information

From July 2023 to December 2024, 19 healthy social volunteers, including 15 males and 4 females, participated in the study. The age range of the participants was 23–58 years, with a mean age ± standard deviation of 33.79 ± 7.72 years.

Inclusion Criteria: Eligible participants were healthy volunteers with no documented history of brachial plexus pathology, including traumatic injury, inflammatory conditions, neoplastic involvement, or congenital anomalies. Candidates were required to demonstrate normal sensorimotor function of the upper extremities and to be free from systemic comorbidities that could potentially impact neurological integrity, such as diabetes mellitus, autoimmune diseases, or metabolic disorders. All participants met MRI safety requirements, specifically the absence of pacemakers, metallic implants, or claustrophobia, and provided written informed consent.

Exclusion criteria: Exclusion criteria encompassed any history of brachial plexus injury, surgical intervention, or radiation therapy; the presence of neurological disorders affecting central or peripheral pathways, including carpal tunnel syndrome or demyelinating diseases; administration of neurotoxic medications within the preceding 6 months; chronic alcohol dependence; biochemical deficiencies compromising neural health; anatomical or physiological factors likely to compromise image quality, such as artifact-generating non-ferromagnetic implants; current pregnancy or lactation; and enrollment in other clinical trials within the last 30 days.

2.2 Equipment and methods

All the scanning protocols were conducted using a Siemens Magnetom Vida 3.0T magnetic resonance scanner, software version XA50, equipped with a 20–channel head and neck combined coil and an Ultra Flex18 small flexible coil. Prior to the examination, the participants were briefed on potential noise and heat sensations associated with the procedure to facilitate their cooperation.

During the scans, the participants were placed in a supine position with their head entering the scanner first. An Ultra Flex18 small flexible coil was placed to cover the junction of the shoulder and neck, as well as the axillary region. Magnetic resonance sandbags were symmetrically positioned on the dorsal side of the upper arms to enhance the display of distal brachial plexus signals.

A direct coronal non-interval scan was performed, covering the region from the C3 vertebral level to the humeral head in the superior–inferior direction and from the anterior edge of the vertebral body to the posterior edge of the spinal canal in the anterior–posterior direction.

The scanning parameters for the T2WI Fast Dixon sequence were as follows: repetition time (TR)/echo time (TE) = 6,220 ms/104 ms; turbo factor (ETL) = 21; parallel imaging acceleration factor (GRAPPA) = 2; average = 2; field of view (FOV) = 288 × 360 mm2; acquisition matrix = 288 × 448; slice thickness = 1.6 mm (reconstructed in-plane resolution: 0.8 × 0.8 mm2); contiguous slices (gap = 0 mm) = 46; and total scan duration = 4 min 46 s.

Upon completing the scans, the deep resolve feature was activated in the advanced reconstruction section of the reconstruction parameter card via the original dataset. The DRG denoising mode was activated, with the denoising intensity factor set to 8 and the enhancement level factor set to 5. The first group of images without DRG technology was used as the control group, while the denoised DRG images were used as the experimental group.

2.3 Data measurement and image post-processing

All the data were exported in DICOM format to RadiAnt DICOM Viewer 2023 software for measurement. Regions of interest (ROIs) of appropriate size were placed on the roots, trunks, cords, and branches of the brachial plexus, as well as on the scalene muscles, on both sides of the same thin layer along the course of the brachial plexus. The background signal intensity (SI) was determined as the average value of the four-corner areas of the image. The SIs of the roots, trunks, cords, branches, muscles, and background noise were recorded. On the basis of these measurements, the SNR and CNR for each segment of the left and right brachial plexus were calculated via the following formulas: SNR = nerve SI/background noise SI; CNR = (nerve SI – muscle SI)/background noise SI. Additionally, 19 paired pre- and post-denoising MIP image groups were reconstructed, and all thin-slice datasets were processed with 10 mm maximum intensity projection (MIP).

2.4 Image evaluations

Two musculoskeletal radiologists with 3 and 10 years of experience independently evaluated the reconstructed MIP images together with the source images in a random order and were blinded to the imaging parameters and clinical information. Like the previously published scoring system (16), the original thin-layer images were evaluated via a three-point scale: 3 points, no background noise, clear boundaries between nerves and tissues, uniform signals, and no artifact interference. Two points, mild noise, local fluctuations in the nerve signal, identifiable but local fuzzy boundaries between nerve and tissue, and mild artifacts; Score 1, the noise was significant, the structure was blurred, the nerve and tissue boundaries could not be distinguished, and artifacts covered key anatomical regions.

The thin-MIP images of the two groups were evaluated via a 4-point scale to evaluate the root, trunk, bundle, and branch (17): 4 points, the corresponding segments of the brachial plexus were continuous and complete, with clear boundaries with surrounding tissues and no artifacts or noise interference. Score 3: Each segment of the brachial plexus could be observed, and only one root/trunk/cord/branch (such as the T1 root or 1 trunk/bundle/pectoral minor branch) in the corresponding segment was slightly blurred, with mild artifacts and noise interference; Score 2: severe loss of the corresponding segments of the brachial plexus (such as T1 root loss/incomplete display of the inferior trunk/blurred medial fasciculus/visible main branches but blurred bifurcations), unclear tissue boundaries, and moderate artifact noise interference; Score 1: the corresponding segment of the brachial plexus could not distinguish the structure, and the image was severely disturbed by artifact noise. When there was disagreement in blind reviews, discrepancy in assessment was resolved through negotiation, and the final outcome was based on their consensus agreement.

2.5 Statistical methods

Data analysis was conducted via SPSSPRO statistical software. The measurement data are presented as the means ±standard deviations (X̅ ± S). The Wilcoxon signed-rank test was employed to compare the quality grading between the original thin-slice brachial plexus images and the corresponding thin-MIP reconstructed nerve segment displays. For the SNR and CNR comparisons across brachial plexus segments, the paired Student’s t test was applied when normality assumptions were satisfied (assessed via the Shapiro–Wilk test); otherwise, the Wilcoxon signed-rank test was utilized. A two-tailed p value <0.05 indicated statistical significance. Interobserver agreement was evaluated via Cohen’s kappa coefficient, which was interpreted as follows: ≤0.20 (poor), 0.21–0.40 (fair), 0.41–0.60 (moderate), 0.61–0.80 (good), and 0.81–1.00 (excellent).

3 Results

3.1 Qualitative analysis

Quantitative scores comparing image quality between two thin-slice images are systematically summarized in Table 1 and are visually represented in Figure 1. Subsequent qualitative evaluations of critical brachial plexus substructures, including roots, trunks, cords, and branch segments, are statistically documented in Table 2, with comparative analyses graphically depicted in Figure 2.

Table 1
www.frontiersin.org

Table 1. Comparison of subjective scores for the original images of the brachial plexus between the experimental and control groups.

Figure 1
www.frontiersin.org

Figure 1. Subjective rating of the original image.

Table 2
www.frontiersin.org

Table 2. Quality evaluation of root, trunk, cord, and branch segments in two groups of post thin-MIP images [n = 19, (e.g., %)].

Figure 2
www.frontiersin.org

Figure 2. MIP score for each segment of the brachial plexus.

Table 1 qualitatively shows the quality of the thin-layer images of the experimental and control groups. The score of the experimental group was greater than that of the control group, indicating that the quality of the T2WI Fast Dixon image combined with the DRG was better than that of the conventional T2WI Fast Dixon image (p = 0.005), and the difference was statistically significant.

Table 2, Quality evaluation of the root, trunk, cord and branch segments of the two groups of images after thin-MIP, p < 0.05, indicating that T2WI Fast Dixon images combined with the DRG showed statistically significant differences in the root, trunk, cord and branch of the brachial plexus after reconstruction.

Interobserver agreement was excellent, with Cohen’s kappa values of k = 1.000 for the control group and k = 0.853 for the experimental group. Thin maximum intensity projection (thin-MIP) images showed good to excellent consistency between the two observers across anatomical segments. In the control group, kappa values for the root, trunk, cord, and branch were k = 0.716, 0.835, 0.774, and 0.898, respectively. In the experimental group, the corresponding values were k = 0.776, 0.855, 0.883, and 0.897. All differences were statistically significant with p < 0.001.

Figures 3, 4 present comparative 10 mm thin-MIP images under identical window width/level settings. Panels A/C (without-DRG) and B/D (after-DRG) highlight the localized adaptive denoising capability of the DRG-integrated method within central regions (delineated by white circles), which contrasts with conventional global filtering (A/C). The DRG technique significantly improves the SNR and nerve-to-background contrast in anatomically complex areas, such as the cervicothoracic junction and supraclavicular fossa, while maintaining structural fidelity.

Figure 3
www.frontiersin.org

Figure 3. Comparison of the control and experimental groups after the addition of thin-MIP from a 35-year-old male. A,B present comparative 10 mm thin-MIP images under identical window width/level settings. B (post-DRG) highlight the localized adaptive denoising capability of the DRG-integrated method within central regions (delineated by white circles), which contrasts with A (without-DRG) a conventional global filter. The DRG technique significantly improves the SNR and nerve-to-background contrast in anatomically complex areas, such as the cervicothoracic junction and supraclavicular fossa, while maintaining structural fidelity.

Figure 4
www.frontiersin.org

Figure 4. Comparison of the control and experimental groups after thin-MIP from a 59-year-old female. C,D present comparative 10 mm thin-MIP images under identical window width/level settings. Compared with C (without-DRG), D (post-DRG) has better denoising ability and neural background contrast.

3.2 Quantitative analysis

Figure 5 illustrates the measurement scheme for the signal intensities of the right partial segments of the brachial plexus and the background noise in the two groups from another test. Compared with those in the control group (Figure 5A), the signal intensities of the roots, trunks, and cords of the brachial plexus in the experimental group (Figure 5B) were greater, whereas the background noise was significantly lower in the experimental group than in the control group.

Figure 5
www.frontiersin.org

Figure 5. Comparison of T2WI Fast Dixon measurements before and after DRG combination in displaying partial segments of the brachial plexus. A: without DRG; B: with DRG.

We statistically analyzed the average SNR and CNR values of each segment of the brachial plexus before and after the use of the DRG, and the results are shown in Table 3 and Figure 6. The bilateral SNRs and CNRs of each anatomical structure (root, trunk, cord, branch) of the brachial plexus in the experimental group with DRG were significantly greater than those in the control group without DRG (p < 0.05), as follows: the SNRs of the bilateral nerve roots increased by 35.1–36.2% (e.g., left: 57.506 vs. 42.25; right: 56.917 vs. 42.116), the SNRs of the bilateral nerve trunks increased by 40.6–40.8%, and the SNRs of the bilateral nerve cords and branches increased by about 40–45%. The CNR of the bilateral nerve roots increased by 43.1–44.6%, the CNR of the bilateral nerve trunks increased by 41.8–41.7%, and the CNR of the bilateral nerve cords and branches increased by 47.3–50.6% (root p < 0.001, trunk p < 0.001, cord p = 0.001, branch p = 0.011).

Table 3
www.frontiersin.org

Table 3. Comparison of SNR and CNR values for each segment of the brachial plexus between the control and experimental groups (x̄ ± s).

Figure 6
www.frontiersin.org

Figure 6. Comparative histogram analysis of the control and experimental groups.

4 Discussion

The current approaches to magnetic resonance neuroimaging (MRN) involve the use of high-resolution isotropic 3D imaging to capture the complex pathways of peripheral nerve structures, and sequences with strong T2 weighting or those associated with DWI are used to enhance the signal from nerve tissues while minimizing the signals from surrounding structures. Furthermore, suitable fat suppression techniques are employed to reduce the impact of fat signals (1820):

1. A brachial plexus nerve display technique based on DWI. This method primarily exploits the principle that water molecules exhibit restricted diffusion, resulting in high signals after the application of a diffusion gradient field, enabling visualization of the brachial plexus nerve. However, it has limitations in terms of spatial resolution and a low signal-to-noise ratio. The diagnostic utility of these images alone is limited, and their combination with other TSE sequences is needed to improve diagnostic efficacy (4).

2. 3D TSE STIR combined with fast spin-echo heavy T2-weighted imaging is a high-resolution technique that enables clear visualization of the brachial plexus nerves and is well-suited for both pre- and post-contrast imaging. In contrast, the T1 and T2 relaxation times of lymph nodes and blood are reduced, potentially affecting their signal characteristics. The longitudinal magnetization vectors of these background signals are suppressed after inversion, resulting in a strong contrast between the brachial plexus and surrounding tissues, which enables clear imaging of the nerves (21, 22). This method is widely used in clinical practice. However, since it involves the administration of a contrast agent, it may not be suitable for patients with poor renal function or those who are allergic to the agent. Additionally, in the neck and shoulder regions, where there is significant anatomical deformation, higher field strengths (especially above 3T) can lead to magnetic field inhomogeneities. This often results in failure of local fat suppression, causing fat signals to obscure the brachial plexus nerve, making it difficult to identify (10). The water-fat separation Dixon technique is relatively insensitive to magnetic field and radiofrequency inhomogeneities, effectively suppressing fat signals and thereby minimizing interference with brachial plexus nerve visualization. However, conventional Dixon sequences are associated with prolonged scan times, increasing susceptibility to motion artifacts and potentially compromising the clarity of anatomical detail.

The Fast Dixon technique modified by the XA version of the manufacturer, which is not sensitive to the inhomogeneity of the main magnetic field and the RFF, was used in this study. The fat signal was completely suppressed, and the display of the brachial plexus was not easily disturbed by the fat signal. Ordinary Dixon (23) completes the acquisition of opposed-phase and in-phase echoes in two anterior and posterior TRs, which are filled in two independent k-spaces, and then water and lipid images are obtained through the automatic calculation of the system to realize the separation of water and fat, and the overall scanning time is relatively long. In this study, Fast Dixon was used to acquire both opposed-phase and in-phase echoes after one TR, as shown in Figure 7. It also merged the damage gradient and readout gradient, increased the amplitude of the gradient, and reduced the ramp time. Therefore, the scanning time of T2WI Fast Dixon is almost half that of conventional T2WI Dixon. In other words, Fast Dixon can achieve higher spatial resolution and average frequency than ordinary Dixon under the same time conditions. Notably, Fast Dixon needs to be activated under high bandwidth conditions. The existence of multiple average times, higher bandwidths and smaller echo intervals also greatly reduces the probability of motion artifacts of the brachial plexus. The same application scenarios also include easy movement tissues such as the neck and pelvic soft tissue. In fact, the T2WI Fast Dixon sequence we used is essentially a 2D sequence, which is less likely to produce motion artifacts than 3D volumetric excitation is. T2WI Fast Dixon can achieve a slice thickness similar to 3D, with better intra-slice resolution than 3D TSE STIR (0.8 × 0.8 m2 vs. 1 × 1 m2) in the literature (16, 24, 25) and can provide the same interslice resolution, large FOV and post-processing possibilities as 3D STIR TSE. It allows clear visualization of the entire brachial plexus and facilitates structure recognition and image interpretation (26, 27).

Figure 7
www.frontiersin.org

Figure 7. Timing diagram of the T2WI Fast Dixon sequence. The diagram of the Fast Dixon sequence shows opposed phase and in phase echoes acquired in the same repetition time between a pair of refocusing pulses (row 2) and the corresponding readout gradient (row 3). ADC, Analog-to-digital converter.

This study demonstrates that the T2WI Fast Dixon sequence, integrated with DRG technology, significantly improves brachial plexus visualization through a dual-path architecture guided by deep learning-based noise modeling and signal enhancement, enabling simultaneous noise suppression and enhancement of SNR and CNR. As an autonomously developed intelligent denoising solution by the MR research team, the core innovations of the DRG are as follows: (1) a noise suppression pathway utilizing residual convolutional networks (ResNet) to precisely learn noise distributions (including electronic noise, motion artifacts, and dielectric effects) in low-SNR images (28), achieving joint denoising via synergistic wavelet transform (frequency domain) and pixel-level filtering (spatial domain), coupled with adversarial training (GAN) to generate noise-reduced images highly consistent with authentic high-SNR references (29); (2) a signal enhancement pathway employing attention gates to dynamically amplify local features of critical anatomical structures (e.g., brachial plexus perineurium-epineurium interfaces) while iteratively restoring noise-obscured microstructures. Some prior studies have demonstrated that deep learning via adaptive network architectures can address denoising, resolution enhancement, and scan time reduction (14, 3032), which is attributable to domain-specific knowledge embedding (e.g., neural pathway patterns) and exploitation of interslice spatial correlations (33, 34), thereby effectively constraining reconstruction challenges in generating high-quality images. Our findings align with the denoising part of the previous study. To address the persistent issue of MRI noise spatial heterogeneity (stemming from SNR depth-dependent attenuation due to receiver coil geometry and regional noise fluctuations induced by parallel imaging techniques) (35, 36), the DRG innovatively incorporates noise maps derived from raw k-space data into the reconstruction pipeline. The noninvasive extraction of noise maps eliminates additional scan time, enabling localized adaptive denoising—targeted suppression in noise-dominant regions (e.g., image centers or motion artifact zones)—while ensuring real-time processing via optimized computational frameworks. Compared with traditional global filters, the DRG markedly improves fat suppression uniformity and nerve-background contrast in anatomically complex regions. Currently, this technology has been extended to multiple domains: optimization of white matter fiber tract visualization in neuroimaging (37), microstructural enhancement of articular cartilage and tendons in musculoskeletal systems (38, 39), and the tissue edge sharpening in body imaging (40). The synergistic “denoising-enhancement” mechanism provides high-confidence imaging, marking a critical step toward the clinical translation of intelligent image reconstruction technologies.

5 Limitations

This study has several limitations. First, the sequence used is a non-contrast imaging protocol for the brachial plexus nerve. Some background signals from small blood vessels and lymph nodes are still present, which can slightly affect the display of certain segments of the brachial plexus. In addition, this study employed a single-center design with a relatively small sample size. This limitation may hinder the detection of intergroup differences due to limited sample representativeness, potentially masking true biological or clinical variations and diminishing statistical power. To overcome these constraints, future research should prioritize multicenter collaborations and the inclusion of large-scale cohorts. Multicenter frameworks can integrate geographically diverse populations, thereby enhancing demographic and phenotypic heterogeneity, whereas expanded sample sizes improve statistical power and data robustness, ultimately yielding more comprehensive, generalizable, and statistically conclusive findings. The inclusion of only healthy volunteers also limits its applicability to pathological cases, which may also have contributed to bias in this study. Deep learning performs better than radiomic features in other site case tasks, such as image segmentation, lesion detection, prognosis prediction, and multimodal image registration (4143), which also makes us full of expectations for the future brachial plexus in processing pathological changes in the brachial plexus. Overall, to improve the robustness of the study results, we should conduct future long-term follow-up studies in the future and need to include more patients with higher acceleration rates and pathological types.

6 Conclusion

T2WI Fast Dixon can enhance the visibility of brachial plexus segments to a certain extent through DRG denoising technology, which may be an effective means to display the brachial plexus without contrast agent.

Data availability statement

The original contributions presented in the study are included in the article/supplementary material, further inquiries can be directed to the corresponding author.

Ethics statement

The studies involving humans were approved by Clinical Trial Ethics Committee, Affiliated Hospital of Southwest Medical University (KY2020190). The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study.

Author contributions

SX: Writing – original draft, Writing – review & editing, Data curation, Formal analysis. XD: Formal analysis, Methodology, Writing – review & editing. ZZ: Formal analysis, Writing – review & editing. GC: Supervision, Validation, Writing – review & editing.

Funding

The author(s) declare that no financial support was received for the research and/or publication of this article.

Acknowledgments

We would like to thank Dr. Yongshu Lan of department of radiology, the Affiliated Hospital of Southwest Medical University, for his support in writing skills and logic of the article.

Conflict of interest

XD and ZZ were employed by Siemens Healthineers, Shanghai, China.

The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Generative AI statement

The authors declare that no Gen AI was used in the creation of this manuscript.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Abbreviations

2D, Two-dimensional; 3D, Three-dimensional; MIP, Maximal intensity projection; AI, Artificial intelligence; DRG, Deep resolve gain; SI, Signal intensity; CNR, Contrast-to-noise ratio; SNR, Signal-to-noise ratio; FOV, Field of view; RFF, Radio frequency field.

References

1. Uysal, II, Şeker, M, Karabulut, AK, Büyükmumcu, M, and Ziylan, T. Brachial plexus variations in human fetuses. Neurosurgery. (2003) 53:676–84. doi: 10.1227/01.neu.0000079485.24016.70

PubMed Abstract | Crossref Full Text | Google Scholar

2. Cunha, MR, Dias, AAM, Brito, JM, Cruz, CS, and Silva, SK. Anatomical study of the brachial plexus in human fetuses and its relation with neonatal upper limb paralysis. Einstein (Sao Paulo). (2020) 18:eAO5051. doi: 10.31744/einstein_journal/2020AO5051

PubMed Abstract | Crossref Full Text | Google Scholar

3. Rehman, I, Chokshi, FH, and Khosa, F. MR imaging of the brachial plexus. Clin Neuroradiol. (2014) 24:207–16. doi: 10.1007/s00062-014-0297-3

PubMed Abstract | Crossref Full Text | Google Scholar

4. Adachi, Y, Sato, N, Okamoto, T, Sasaki, M, Komaki, H, Yamashita, F, et al. Brachial and lumbar plexuses in chronic inflammatory demyelinating polyradiculoneuropathy: MRI assessment including apparent diffusion coefficient. Neuroradiology. (2011) 53:3–11. doi: 10.1007/s00234-010-0684-7

PubMed Abstract | Crossref Full Text | Google Scholar

5. Sierra-Silvestre, E, Smith, RE, Andrade, RJ, Kennedy, B, and Coppieters, MW. Microstructural changes in the median and ulnar nerve in people with and without diabetic neuropathy in their hands: a cross-sectional diffusion MRI study. Eur J Radiol. (2024) 181:111721. doi: 10.1016/j.ejrad.2024.111721

PubMed Abstract | Crossref Full Text | Google Scholar

6. Su, X, Kong, X, Liu, D, Kong, X, Alwalid, O, Wang, J, et al. Multimodal magnetic resonance imaging of peripheral nerves: establishment and validation of brachial and lumbosacral plexi measurements in 163 healthy subjects. Eur J Radiol. (2019) 117:41–8. doi: 10.1016/j.ejrad.2019.05.017

PubMed Abstract | Crossref Full Text | Google Scholar

7. Li, S, and Yuan, H. Osteoarthrography group of Chinese society of radiology of Chinese medical association. Chinese expert consensus on peripheral nerve MRI. Chin J Magn Reson Imaging. (2023) 14:1–7. doi: 10.12015/issn.1674-8034.2023.05.001

Crossref Full Text | Google Scholar

8. Szaro, P, McGrath, A, Ciszek, B, and Geijer, M. Magnetic resonance imaging of the brachial plexus. Part 1: anatomical considerations, magnetic resonance techniques, and non-traumatic lesions. Eur J Radiol Open. (2022) 9:100392. doi: 10.1016/j.ejro.2021.100392

PubMed Abstract | Crossref Full Text | Google Scholar

9. Pedrick, EG, Sneag, DB, Colucci, PG, Duong, M, and Tan, ET. Three-dimensional MR Neurography of the brachial plexus: vascular suppression with low-dose Ferumoxytol. Radiology. (2023) 307:e221087. doi: 10.1148/radiol.221087

PubMed Abstract | Crossref Full Text | Google Scholar

10. Davidson, EJ, Tan, ET, Pedrick, EG, and Sneag, DB. Brachial plexus magnetic resonance neurography: technical challenges and solutions. Investig Radiol. (2023) 58:14–27. doi: 10.1097/rli.0000000000000906

PubMed Abstract | Crossref Full Text | Google Scholar

11. Behl, N. Deep resolve—mobilizing the power of networks. Magnetom Flash. (2021) 1:29–35.

Google Scholar

12. Hammernik, K, Knoll, F, and Rueckert, D. Deep learning for parallel MRI reconstruction: overview, challenges, and opportunities. Magnetom Flash. (2019) 4:10–5.

Google Scholar

13. Herrmann, J, Gassenmaier, S, Keller, G, Koerzdoerfer, G, Almansour, H, Nickel, D, et al. Deep learning MRI reconstruction for accelerating turbo spin echo hand and wrist imaging: a comparison of image quality, visualization of anatomy, and detection of common pathologies with standard imaging. Acad Radiol. (2023) 30:2606–15. doi: 10.1016/j.acra.2022.12.042

PubMed Abstract | Crossref Full Text | Google Scholar

14. Oscanoa, JA, Middione, MJ, Syed, AB, Sandino, CM, Vasanawala, SS, and Ennis, DB. Accelerated two-dimensional phase-contrast for cardiovascular MRI using deep learning-based reconstruction with complex difference estimation. Magn Reson Med. (2023) 89:356–69. doi: 10.1002/mrm.29441

PubMed Abstract | Crossref Full Text | Google Scholar

15. Ensle, F, Kaniewska, M, Tiessen, A, Lohezic, M, Getzmann, JM, and Guggenberger, R. Diagnostic performance of deep learning-based reconstruction algorithm in 3D MR neurography. Skeletal Radiol. (2023) 52:2409–18. doi: 10.1007/s00256-023-04362-z

PubMed Abstract | Crossref Full Text | Google Scholar

16. Wang, L, Niu, Y, Kong, X, Yu, Q, Kong, X, Lv, Y, et al. The application of paramagnetic contrast-based T2 effect to 3D heavily T2W high-resolution MR imaging of the brachial plexus and its branches. Eur J Radiol. (2016) 85:578–84. doi: 10.1016/j.ejrad.2015.12.001

PubMed Abstract | Crossref Full Text | Google Scholar

17. Hu, SX, Xiao, Y, Peng, WL, Zeng, W, Zhang, Y, Zhang, XY, et al. Accelerated 3D MR neurography of the brachial plexus using deep learning-constrained compressed sensing. Eur Radiol. (2024) 34:842–51. doi: 10.1007/s00330-023-09996-0

PubMed Abstract | Crossref Full Text | Google Scholar

18. Wang, X, Harrison, C, Mariappan, YK, Gopalakrishnan, K, Chhabra, A, Lenkinski, RE, et al. MR Neurography of brachial plexus at 3.0 T with robust fat and blood suppression. Radiology. (2017) 283:538–46. doi: 10.1148/radiol.2016152842

PubMed Abstract | Crossref Full Text | Google Scholar

19. Szaro, P, Geijer, M, Ciszek, B, and McGrath, A. Magnetic resonance imaging of the brachial plexus. Part 2: traumatic injuries. Eur J Radiol Open. (2022) 9:100397. doi: 10.1016/j.ejro.2022.100397

PubMed Abstract | Crossref Full Text | Google Scholar

20. Jung, JY, Lin, Y, and Carrino, JA. An updated review of magnetic resonance neurography for plexus imaging. Korean J Radiol. (2023) 24:1114–30. doi: 10.3348/kjr.2023.0150

PubMed Abstract | Crossref Full Text | Google Scholar

21. Zhang, Y, Kong, X, Zhao, Q, Liu, X, Gu, Y, and Xu, L. Enhanced MR neurography of the lumbosacral plexus with robust vascular suppression and improved delineation of its small branches. Eur J Radiol. (2020) 129:109128. doi: 10.1016/j.ejrad.2020.109128

PubMed Abstract | Crossref Full Text | Google Scholar

22. Zhang, X, Wang, W, Liu, T, Qi, Y, and Ma, L. The effects of three different contrast agents (Gd-BOPTA, Gd-DTPA, and Gd-DOTA) on brachial plexus magnetic resonance imaging. Ann Transl Med. (2021) 9:344. doi: 10.21037/atm-21-348

PubMed Abstract | Crossref Full Text | Google Scholar

23. Chen, L, Hu, H, Chen, HH, Chen, W, Wu, Q, Wu, FY, et al. Usefulness of two-point Dixon T2-weighted imaging in thyroid-associated ophthalmopathy: comparison with conventional fat saturation imaging in fat suppression quality and staging performance. Br J Radiol. (2021) 94:20200884. doi: 10.1259/bjr.20200884

PubMed Abstract | Crossref Full Text | Google Scholar

24. Wu, W, Ding, Y, Su, Y, Wang, Y, Liu, T, Liu, D, et al. Contrast-enhanced magnetic resonance neurography for diagnosing brachial plexopathy: improved visualization and additional imaging features. Sci Rep. (2024) 14:20758. doi: 10.1038/s41598-024-71554-5

PubMed Abstract | Crossref Full Text | Google Scholar

25. Sneag, DB, Queler, SC, Campbell, G, Colucci, PG, Lin, J, Lin, Y, et al. Optimized 3D brachial plexus MR neurography using deep learning reconstruction. Skeletal Radiol. (2024) 53:779–89. doi: 10.1007/s00256-023-04484-4

PubMed Abstract | Crossref Full Text | Google Scholar

26. Reeder, SB, Yu, H, Johnson, JW, Shimakawa, A, Brittain, JH, Pelc, NJ, et al. T1-and T2-weighted fast spin-echo imaging of the brachial plexus and cervical spine with IDEAL water-fat separation. J Magn Reson Imaging. (2006) 24:825–32. doi: 10.1002/jmri.20721

PubMed Abstract | Crossref Full Text | Google Scholar

27. Grayev, A, Reeder, S, and Hanna, A. Use of chemical shift encoded magnetic resonance imaging (CSE-MRI) for high resolution fat-suppressed imaging of the brachial and lumbosacral plexuses. Eur J Radiol. (2016) 85:1199–207. doi: 10.1016/j.ejrad.2016.04.002

PubMed Abstract | Crossref Full Text | Google Scholar

28. Zhang, K, Zuo, W, Chen, Y, Meng, D, and Zhang, L. Beyond a gaussian denoiser: residual learning of deep CNN for image denoising. IEEE Trans Image Process. (2017) 26:3142–55. doi: 10.1109/tip.2017.2662206

PubMed Abstract | Crossref Full Text | Google Scholar

29. Yu, M, Guo, M, Zhang, S, Zhan, Y, Zhao, M, Lukasiewicz, T, et al. RIRGAN: an end-to-end lightweight multi-task learning method for brain MRI super-resolution and denoising. Comput Biol Med. (2023) 167:107632. doi: 10.1016/j.compbiomed.2023.107632

PubMed Abstract | Crossref Full Text | Google Scholar

30. Foreman, SC, Neumann, J, Han, J, Harrasser, N, Weiss, K, Peeters, JM, et al. Deep learning-based acceleration of compressed sense MR imaging of the ankle. Eur Radiol. (2022) 32:8376–85. doi: 10.1007/s00330-022-08919-9

PubMed Abstract | Crossref Full Text | Google Scholar

31. Ueda, T, Ohno, Y, Yamamoto, K, Iwase, A, Fukuba, T, Hanamatsu, S, et al. Compressed sensing and deep learning reconstruction for women's pelvic MRI denoising: utility for improving image quality and examination time in routine clinical practice. Eur J Radiol. (2021) 134:109430. doi: 10.1016/j.ejrad.2020.109430

PubMed Abstract | Crossref Full Text | Google Scholar

32. Bischoff, LM, Peeters, JM, Weinhold, L, Krausewitz, P, Ellinger, J, Katemann, C, et al. Deep learning super-resolution reconstruction for fast and motion-robust T2-weighted prostate MRI. Radiology. (2023) 308:e230427. doi: 10.1148/radiol.230427

PubMed Abstract | Crossref Full Text | Google Scholar

33. Pezzotti, N, Yousefi, S, Elmahdy, MS, Van Gemert, JHF, Schuelke, C, Doneva, M, et al. An adaptive intelligence algorithm for undersampled knee MRI reconstruction. IEEE Access. (2020) 8:204825–38. doi: 10.1109/ACCESS.2020.3034287

Crossref Full Text | Google Scholar

34. Zhang, J, and Ghanem, B. ISTA-net: interpretable optimization-inspired deep network for image compressive sensing In: Proceedings of the IEEE conference on computer vision and pattern recognition : IEEE (2018). 1828–37.

Google Scholar

35. Knoll, F, Hammernik, K, Zhang, C, Moeller, S, Pock, T, Sodickson, DK, et al. Deep-learning methods for parallel magnetic resonance imaging reconstruction: a survey of the current approaches, trends, and issues. IEEE Signal Process Mag. (2020) 37:128–40. doi: 10.1109/msp.2019.2950640

PubMed Abstract | Crossref Full Text | Google Scholar

36. Dubljevic, N, Moore, S, Lauzon, ML, Souza, R, and Frayne, R. Effect of MR head coil geometry on deep-learning-based MR image reconstruction. Magn Reson Med. (2024) 92:1404–20. doi: 10.1002/mrm.30130

PubMed Abstract | Crossref Full Text | Google Scholar

37. Yuan, Y, Yu, Y, Chang, J, Chu, YH, Yu, W, Hsu, YC, et al. Convolutional neural network to predict IDH mutation status in glioma from chemical exchange saturation transfer imaging at 7 tesla. Front Oncol. (2023) 13:1134626. doi: 10.3389/fonc.2023.1134626

PubMed Abstract | Crossref Full Text | Google Scholar

38. Herrmann, J, Koerzdoerfer, G, Nickel, D, Mostapha, M, Nadar, M, Gassenmaier, S, et al. Feasibility and implementation of a deep learning MR reconstruction for TSE sequences in musculoskeletal imaging. Diagnostics. (2021) 11:1484. doi: 10.3390/diagnostics11081484

PubMed Abstract | Crossref Full Text | Google Scholar

39. Kim, M, Lee, SM, Park, C, Lee, D, Kim, KS, Jeong, HS, et al. Deep learning-enhanced parallel imaging and simultaneous multislice acceleration reconstruction in knee MRI. Investig Radiol. (2022) 57:826–33. doi: 10.1097/rli.0000000000000900

PubMed Abstract | Crossref Full Text | Google Scholar

40. Pocepcova, V, Zellner, M, Callaghan, F, Wang, X, Lohezic, M, Geiger, J, et al. Deep learning-based denoising image reconstruction of body magnetic resonance imaging in children. Pediatr Radiol. (2025). doi: 10.1007/s00247-025-06230-5

PubMed Abstract | Crossref Full Text | Google Scholar

41. Yan, M, Zhang, X, Zhang, B, Geng, Z, Xie, C, Yang, W, et al. Deep learning nomogram based on Gd-EOB-DTPA MRI for predicting early recurrence in hepatocellular carcinoma after hepatectomy. Eur Radiol. (2023) 33:4949–61. doi: 10.1007/s00330-023-09419-0

PubMed Abstract | Crossref Full Text | Google Scholar

42. Kikuchi, Y, Togao, O, Kikuchi, K, Momosaka, D, Obara, M, Van Cauteren, M, et al. A deep convolutional neural network-based automatic detection of brain metastases with and without blood vessel suppression. Eur Radiol. (2022) 32:2998–3005. doi: 10.1007/s00330-021-08427-2

PubMed Abstract | Crossref Full Text | Google Scholar

43. Zhao, J, Sun, L, Zhou, X, Huang, S, Si, H, and Zhang, D. Residual-atrous attention network for lumbosacral plexus segmentation with MR image. Comput Med Imaging Graph. (2022) 100:102109. doi: 10.1016/j.compmedimag.2022.102109

PubMed Abstract | Crossref Full Text | Google Scholar

Keywords: brachial plexus, magnetic resonance imaging, Fast Dixon, deep learning, contrast agent-free

Citation: Xie S, Du X, Zhang Z and Chen G (2025) The value of Fast Dixon combined with deep learning technology in contrast agent-free high-resolution magnetic resonance imaging of the brachial plexus. Front. Neurol. 16:1558927. doi: 10.3389/fneur.2025.1558927

Received: 11 January 2025; Accepted: 09 May 2025;
Published: 04 June 2025.

Edited by:

Lian Yang, Huazhong University of Science and Technology, China

Reviewed by:

Hai-bo Qu, Sichuan University, China
Lixia Wang, Huazhong University of Science and Technology, China

Copyright © 2025 Xie, Du, Zhang and Chen. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Guangxiang Chen, Y2d4MjNseTIwMDJAMTYzLmNvbQ==

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.