ORIGINAL RESEARCH article

Front. Surg., 15 January 2026

Sec. Orthopedic Surgery

Volume 12 - 2025 | https://doi.org/10.3389/fsurg.2025.1722354

This article is part of the Research TopicSurgical Advances in Orthopedic Trauma: A Biomechanical ApproachView all 25 articles

Biomechanical evaluation of a novel hockey-stick locking plate featuring a pes anserinus-sparing design: a finite element analysis


Xiao Wang,Xiao Wang1,2Haohua LaiHaohua Lai1Songyan ZhangSongyan Zhang1Zhiheng TuZhiheng Tu1Zhaowei YinZhaowei Yin1Gadisa Musa WakoGadisa Musa Wako3Junwei Yan
Junwei Yan1*Bin Liang

Bin Liang1*
  • 1Department of Orthopaedics, Nanjing First Hospital, Nanjing Medical University, Nanjing, China
  • 2Department of Orthopaedics, Nanjing Qixia District Hospital, Nanjing, China
  • 3Department of General Surgery, Nanjing First Hospital, Nanjing Medical University, Nanjing, China

Background: Surgical fixation for Schatzker IV tibial plateau fractures presents a clinical dilemma: achieving robust stability while avoiding impingement on the pes anserinus tendons. This study evaluated the biomechanical profile of a novel hockey-stick locking plate (NHLP), which is anatomically contoured to address this challenge by being placed anteriorly.

Methods: A finite element model of a standardized Schatzker IV fracture was created. Three fixation methods were simulated: the novel hockey-stick locking plate (NHLP), the traditional T-shaped locking plate (TTLP), and the double reconstruction locking plates (DRLP). The models were subjected to four loading conditions: three physiological loads, a low axial load (500 N), a moderate combined load (1,500 N axial compression plus 150 N anterior shear force), and a high axial load (2,500 N) and a fourth “worst-case” load scenario combining a 1,700 N axial force, a 200 N anterior shear force, and a 10° varus tilt. Key biomechanical metrics, including implant stress, construct stability, fragment displacement, fracture interface mechanics and fatigue safety factor, were analyzed.

Results: Under physiological loading, the NHLP construct demonstrated the lowest peak von Mises stress on the implant. At the high axial load of 2,500 N, the peak stress on the NHLP (159.8 MPa) was 15% lower than that on the TTLP (188.1 MPa) and 35% lower than that on the DRLP (245.5 MPa). In the “worst-case” scenario, all constructs exhibited high safety factors. In terms of stability, the NHLP provided displacement comparable to that of the TTLP, and both were substantially more stable than the DRLP construct, which exhibited the largest displacement under high load. Paradoxically, the DRLP construct consistently resulted in the highest degree of implant stress and the least stability. At the fracture interface, the NHLP maintained a stable environment across all loads, with key metrics remaining within a range conducive to bone healing.

Conclusion: This finite element analysis demonstrated that the NHLP provides fracture stability while reducing peak implant stress under physiological loading. These findings support the biomechanical feasibility of its pes anserinus-sparing design, providing a strong rationale for further investigation.

1 Introduction

Tibial plateau fractures, particularly Schatzker IV fractures involving the medial plateau, present a significant treatment challenge, carrying risks of varus collapse, articular incongruity, and subsequent post-traumatic osteoarthritis (1, 2). Anatomical reduction and stable fixation are essential to permit early joint mobilization and optimize outcomes (3, 4). However, the high compressive loads across the medial tibial plateau demand robust implant support, making the choice of fixation strategy critical.

Existing implants for the medial tibia, such as T-shaped or reconstruction plates, present a well-documented compromise between mechanical strength and soft-tissue compatibility. Their profiles are often poorly adapted to the complex three-dimensional contour of the anteromedial tibial flare, leading to hardware prominence over the pes anserinus (57). This implant-tendon conflict is a significant cause of postoperative pain and may necessitate secondary surgery for hardware removal (8, 9). While attempts to mitigate this issue with smaller or manually bent plates may reduce irritation, they potentially sacrifice the mechanical stability required to resist physiological varus forces.

To address this dilemma, we designed a novel hockey-stick locking plate (NHLP) with an anatomically contoured profile intended to reduce soft-tissue impingement. The plate geometry features a crescent-like turning segment and a multiplanar proximal curve designed to fit the tibial flare while offering versatile screw options for multidirectional stability (Figure 1). This anatomical contouring is specifically intended to follow the tibial flare and avoid prominent hardware over the pes anserinus insertion site, thereby minimizing potential soft-tissue impingement (Figure 2). Before clinical consideration, its biomechanical competence must be rigorously established. Therefore, this finite element study aimed to evaluate the biomechanical performance of the NHLP construct in comparison to a traditional T-shaped locking plate (TTLP) and a double reconstruction locking plate (DRLP) on a standardized Schatzker IV fracture model. We hypothesized that under physiological loading, the NHLP construct would maintain fracture-site motion and implant stresses within clinically acceptable thresholds known to be permissive for bone healing, thereby confirming its biomechanical viability as a soft-tissue-sparing alternative.

Figure 1
Illustration of an orthopedic locking plate shown from multiple angles with measurement labels. The plate has an elongated shape with a straight distal section, a curved middle section, and a broader proximal end. Labels indicate overall length and width, distances between screw holes, and a plate thickness of 3 mm. Multiple circular screw holes are visible along the plate, with a higher density of holes at the proximal end. Different views show the plate profile and curvature, and text labels indicate material and surface treatment information.

Figure 1. Design features and specifications of the novel hockey-stick locking plate (NHLP). The figure displays key dimensions of the NHLP (including screw spacing, plate thickness, and overall length/width) alongside views from multiple angles that illustrate its anatomical profile. The plate is engineered to match the complex contour of the medial proximal tibia, with a design comprising three integrated sections: a distal straight segment with combi-holes, a crescent-shaped turning segment to follow the tibial flare, and a proximal head with multiplanar curves and universal locking holes. These holes permit the insertion of 3.5 mm locking screws at a variable angle of up to 30° from the central axis for multidirectional subchondral support. Material specifications: The implant is made of Ti-6Al-4V ELI (Grade 23) alloy with a surface anodization treatment. The undersurface of the plate is flat.

Figure 2
Illustration of a proximal tibia with an orthopedic plate fixed along the medial side. The plate is positioned anteriorly on the bone surface and secured with multiple screws. Two labeled distances are shown from the posterior edge of the plate to nearby anatomical landmarks, marked as 26 mm and 15 mm. An outline indicating the surgical incision is also visible. The relationship between the plate, bone surface, and surrounding anatomical region is emphasized through arrows and measurement markers.

Figure 2. Anatomical placement and pes anserinus-sparing principle of the novel hockey-stick locking plate (NHLP). The NHLP was conceived from an anatomy-driven design philosophy. As illustrated, its trajectory maintains a safe distance from critical soft tissue structures, positioning the plate anterior to the pes anserinus insertion zone. This placement is quantified by the distances from the plate's posterior edge to the anterior tibial crest (L1 = 26 mm) and to the inferior insertion point of the pes anserinus (L2 = 15 mm). The corresponding surgical incision, which benefits from this anterior placement, is also indicated.

2 Methods

2.1 Finite element model development

A three-dimensional (3D) solid model of the right tibia was developed from a computed tomography (CT) scan of a healthy 32-year-old male volunteer (height: 178 cm; weight: 75 kg) with no history of lower limb pathology. The scan was acquired using a 128-slice spiral CT scanner (Philips CT 6000, Netherlands) at 120 kV and 150 mA with a 0.5 mm slice thickness. The resulting DICOM data were imported into Mimics 20.0 (Materialise, Belgium) for image segmentation and 3D reconstruction of the tibial geometry. The model was subsequently exported to Creo 7.0 (PTC, USA) for geometric processing. A standardized Schatzker IV medial tibial plateau fracture was simulated by creating an oblique osteotomy, as described by Cift et al. (10). This osteotomy originated 5 cm distal to the medial joint line and extended superiorly to the medial intercondylar eminence (Figure 3), creating a fracture model consistent with previous finite element studies on this topic (11).

Figure 3
Six images showing a tibial plateau fracture model and measurement setup. The first three panels display the bone model with visible fracture lines separating different regions of the proximal tibia. The remaining panels show the joint surface with three marked points labeled for measurement. The points are positioned on the articular surface and are visually highlighted to indicate locations used for displacement assessment.

Figure 3. Finite element model of the Schatzker IV tibial plateau fracture and measurement setup. (a–c) The fracture model creation process, illustrating the standardized fracture lines based on CT data. (d–f) The three measurement points on the articular surface used for displacement analysis.

A simulation of virtual implantation was performed for the three fixation constructs (NHLP, TTLP, and DRLP) on the fractured tibia model (Figure 4). All the implants were modeled for use with 3.5 mm screws. Boolean operations were utilized in Creo 7.0 to precisely model the screw holes and ensure a conforming, interference-free interface between the implants and the bone. The finalized assembly models were then exported in STEP format for analysis.

Figure 4
Series of images showing three different internal fixation configurations applied to a tibial plateau fracture model. The first group of panels shows a single plate fixed on the medial side of the tibia. The second group shows a T-shaped plate configuration. The third group shows two plates fixed on different sides of the bone. Each configuration is presented from multiple viewing angles to show plate position and screw placement.

Figure 4. Final assembly of the three internal fixation constructs for the schatzker type IV tibial plateau fracture model. (a–d) Novel hockey-stick locking plate (NHLP) construct. (e–h) Traditional T-shaped locking plate (TTLP) construct. (i–l) Double reconstruction locking plates (DRLP) construct.

2.2 Material properties, interactions, and boundary conditions

All materials in the model were defined as homogeneous, isotropic, and linearly elastic, a standard simplification in comparative biomechanical finite element analysis (FEA) studies. The material properties of the cortical bone, cancellous bone, and titanium alloy (Ti-6Al-4V) implants were assigned on the basis of values reported in the established literature (12, 13) and are detailed in Table 1.

Table 1
www.frontiersin.org

Table 1. Parameters of the finite element models.

Interactions between components were defined to simulate physiological conditions. A surface-to-surface frictional contact was established between the fracture fragments, with a friction coefficient of 0.5 (14). A friction coefficient of 0.3 was assigned to the interface between the plate and the bone (15). To replicate the rigid fixation of a locking screw construct, a “bonded” constraint was applied to the screw-plate and screw-bone interfaces, prohibiting relative motion. The distal end of the tibia was fully constrained in all six degrees of freedom to simulate a fixed ankle joint.

2.3 Meshing and loading conditions

The models were imported into Abaqus 2020 for meshing and analysis. All the components were meshed using high-order ten-node quadratic tetrahedral elements (C3D10). To ensure results were independent of mesh density, a rigorous mesh convergence study was performed. As detailed in Table 2, this analysis confirmed that the peak implant stress stabilized and met the pre-defined <5% convergence criterion for mesh sizes of 2.0 mm and finer. Therefore, based on these findings and to balance accuracy with computational cost, a hybrid meshing strategy was adopted: a fine mesh of 1.0 mm for the implants and adjacent bone regions, and a coarser mesh of 3.0 mm for the remaining tibial shaft. This resulted in final models with total element counts ranging from approximately 250,000 to 350,000 and negligible analysis warnings for element distortion (<1%).

Table 2
www.frontiersin.org

Table 2. Mesh convergence study results for the DRLP construct under 500 N axial load.

Four loading scenarios were defined to simulate a range of physiological activities (Figure 5). The first three scenarios represented a range of physiological activities: (1) Light load (500 N axial), (2) Moderate load (1,500 N axial + 150 N anterior shear), and (3) High load (2,500 N axial). In these scenarios, the axial force was distributed 60% medially and 40% laterally, consistent with gait kinetics data (16). A fourth, “worst-case load” scenario was added based on gait analysis data to assess performance under more challenging conditions: a combined 1,700 N axial force, a 200 N anteroposterior shear force, and a 10° varus tilt (1720). The loads were applied via two reference points. Each reference point was coupled to the articular surface of the corresponding tibial condyle (medial or lateral) to simulate joint reaction forces. The anterior shear force was included to replicate the complex mechanical environment during dynamic movements (11, 2123).

Figure 5
Three schematic diagrams illustrating loading conditions applied to a tibial model. The first diagram shows vertical arrows indicating axial compression applied to the top of the bone. The second diagram shows a horizontal arrow indicating an anteriorly directed shear force. The third diagram shows the bone tilted with combined arrows indicating axial and shear loading. The lower end of the bone is shown as fixed in all diagrams.

Figure 5. Schematic of the loading conditions applied to the finite element model. (a) An axial compressive force was applied to the medial and lateral tibial plateaus. (b) An anterior shear force was applied to the tibial plateau. (c) The “worst-case” scenario, which simulated a 10° varus knee moment by applying a corresponding valgus tilt to the tibia, in combination with axial and shear forces.

2.4 Evaluation metrics

The biomechanical performance of the three constructs was compared using several key metrics. The von Mises stress (VMS) distribution on the implants was analyzed to identify areas of stress concentration. To quantitatively assess the risk of implant failure under high-stress conditions, a fatigue safety factor (FSF) was calculated for both the high Load and “worst-case load” scenarios. The FSF was defined as the ratio of the implant material's yield strength (σy) to the peak von Mises stress (σvM) observed in the implant (FSF = σy/σvM). At the fracture site, the maximum principal strain within the cancellous bone was evaluated, as interfragmentary strain is a critical stimulus for bone healing. To assess the overall structural stability, the maximum displacement of the main fracture fragment was measured. Finally, the relative displacement between the fracture surfaces was quantified to evaluate interfragmentary motion under load.

3 Results

3.1 Implant stress and overall construct stability

The biomechanical results for the three fixation constructs under all four loading conditions are summarized in Table 3.

Table 3
www.frontiersin.org

Table 3. Key biomechanical results for all constructs under all loading conditions.

Under physiological axial loads, the NHLP construct consistently registered the lowest peak VMS on the implant. Under the high-load condition of 2,500 N, the peak VMS on the NHLP was 159.8 MPa, which was 15% lower than the 188.1 MPa observed in the TTLP construct and 35% lower than the 245.5 MPa in the DRLP construct. Regarding overall stability, the maximum displacement of the main fracture fragment was similar for the NHLP and TTLP constructs under these loads, while the DRLP construct showed markedly greater displacement.

Under the “worst-case load” scenario, the peak implant VMS for the NHLP was 50.5 MPa, compared to 62.8 MPa for the TTLP and 114.4 MPa for the DRLP. The corresponding maximum fragment displacements were 5.430 mm, 5.680 mm, and 5.701 mm, respectively, which were the highest displacements recorded across all loading scenarios. The stress distribution patterns under this condition are visualized in Figure 6.

Figure 6
Six color-coded finite element images comparing three fixation constructs. The first three images show stress distributions displayed with a color scale, with different colors representing varying stress levels across the plates and bone. The last three images show displacement distributions using a similar color scale. Each image includes a legend indicating the range of values, and the constructs are shown in comparable orientations.

Figure 6. Von Mises stress (VMS) distributions and displacement magnitudes for the three internal fixation constructs under the “worst-case” loading scenario. (a–c) VMS distributions for the novel hockey-stick locking plate (NHLP), traditional T-shaped locking plate (TTLP), and double reconstruction locking plates (DRLP), respectively. (d–f) The corresponding displacement magnitudes and distributions for each construct.

The calculated FSF under the “worst-case load” was 17.23 for the NHLP, 13.85 for the TTLP, and 7.60 for the DRLP. For the high load condition, the FSF values were 5.44, 4.63, and 3.54 for the NHLP, TTLP, and DRLP, respectively.

3.2 Stress distribution within the tibia

Analysis of stress within the tibia showed that under physiological axial loads, the DRLP construct generally produced the highest peak VMS. However, under the “worst-case load”, peak tibial stresses were similar across all three groups: 73.3 MPa for the NHLP, and 73.5 MPa for both the TTLP and DRLP constructs (Table 3).

3.3 Fracture interface mechanics

The mechanical environment at the fracture interface is detailed in Table 3. The maximum principal strain under the high load condition was 0.0062 for the NHLP, 0.0056 for the TTLP, and 0.0159 for the DRLP, with distribution patterns visualized in Figure 7.

Figure 7
Six finite element images showing color-coded distributions at the fracture interface. The first three images display stress distributions using a color gradient, with higher and lower values indicated by different colors. The remaining images display strain distributions using a similar color scale. Legends with numeric values are shown, and the fracture region is visible in each image.

Figure 7. Stress and strain distributions across the fracture interface under the 2,500 N high loading condition. (a–c) Von Mises stress (VMS) distributions for the novel hockey-stick locking plate (NHLP), traditional T-shaped locking plate (TTLP), and double reconstruction locking plates (DRLP), respectively. (d–f) The corresponding maximum principal strain distributions for each construct.

Regarding interfragmentary motion, the TTLP construct demonstrated the highest stability under both high load and “worst-case” conditions. For example, at the critical lateral measurement point (Point 3), the gap displacement under high load was smallest for the TTLP, and this advantage was maintained under the “worst-case load” (0.0140 mm for TTLP vs. 0.0381 mm for NHLP). The locations of these measurement points are illustrated in Figure 3.

4 Discussion

The surgical management of Schatzker IV tibial plateau fractures often presents a clinical dilemma: achieving the robust fixation necessary for joint congruity while minimizing disruption to the vital soft-tissue structures of the medial knee, particularly the pes anserinus (3, 4, 24). Conventional plating techniques can lead to hardware prominence and tendon irritation, potentially compromising rehabilitation (57). The NHLP was conceived from an anatomy-driven design philosophy to address this specific challenge. Its design maintains a trajectory of 15–26 mm from the anterior tibial crest, which, when compared to anatomical data, places the plate anterior to the pes anserinus insertion zone (Figure 2) (25). This study, therefore, evaluates the biomechanical viability of this anatomy-conscious design.

The biomechanical performance of the NHLP appears to support the viability of its design. A primary finding under physiological loading was that the NHLP construct consistently exhibited the lowest peak VMS on the implant. For instance, under the 2,500 N high load, the peak VMS was 159.8 MPa, resulting in a superior FSF of 5.44 (Table 3). This outcome is likely a direct consequence of its geometry. The plate's multiplanar curvature, engineered to match the tibial contour, may facilitate a more efficient load transfer from the bone, thereby minimizing the stress risers that can occur at the sharp geometric transitions of plates with a more standardized shape. The elongated distal segment of the NHLP, while creating a longer lever arm was created, did not result in greater peak implant stress. It is hypothesized that the multiple points of fixation along this segment create a composite bone-implant structure that distributes the load, rather than concentrating it at a single fulcrum. In the context of repetitive physiological loading, a lower peak operational stress is a fundamental factor in mitigating the risk of fatigue-related material failure (26). The introduction of a varus moment in the “worst-case” scenario presented a different biomechanical environment. This condition led to a general decrease in implant stress for all constructs, as more load was transferred through the bone. Consequently, the calculated FSF for all three methods was markedly high, suggesting a low risk of acute implant failure under this specific challenge. In terms of stability, the NHLP demonstrated the smallest fragment displacement (5.430 mm), performing comparably to the TTLP (5.680 mm).

The performance of the DRLP construct, however, highlights that biomechanical stability is not merely a function of implant quantity, but of how effectively load is managed across the entire construct. In any fixed fracture, axial load is shared among several parallel pathways. A portion of the load travels through the implant itself, passing from the proximal screws into the plate. Another portion is transmitted directly across the fracture interface where bone-to-bone contact is present. Furthermore, a third pathway exists where load is transferred directly from the fracture fragment into the tibial metaphysis through the subchondral screws, which bridge the fracture line and engage the stable, underlying bone. The DRLP construct, despite using two plates, exhibited the highest implant stress of 245.5 MPa under high physiological load and inferior overall stability (Table 3). This paradoxical result suggests a failure in efficiently distributing the load among these pathways. With only one subchondral screw on each of its two non-synergistic plates, the construct is heavily reliant on the implant pathway. This overwhelms these two mechanically weak points, leading to high stress concentrations (Figure 6) and greater fragment displacement, as the construct fails to effectively utilize both direct bone-to-bone contact and the screw-bone bridging pathway for load sharing. In contrast, the multi-screw “raft” configurations of the TTLP and NHLP provide a more robust initial support, allowing for a more balanced load distribution across all available pathways.

The mechanical environment at the fracture interface must be interpreted within the broader context of overall stability and implant safety. Under high loading, the peak principal strains were 0.56% for the TTLP, 0.62% for the NHLP, and 1.59% for the DRLP (Table 3). Given that all these values are well below the 2% strain level typically associated with robust secondary callus formation, all three constructs created a highly stable environment favoring primary bone healing (27, 28). However, within this favorable healing context, the NHLP showed significantly lower implant stress, evidenced by a peak VMS of 159.8 MPa compared to the TTLP's 188.1 MPa (Table 3).

From a clinical standpoint, these biomechanical findings are significant as they validate the feasibility of an anatomy-conscious design. The primary impetus for the NHLP's design was not to achieve superior mechanical strength, but to respect the medial soft-tissue envelope by positioning the plate anterior to the pes anserinus. This design philosophy is an evolution of our previous clinical experience, where standard plates were manually contoured intraoperatively to achieve a similar anatomy-sparing shape (29). The central question of this study was whether a pre-contoured implant based on this principle could provide sufficient stability, or if its “cantilever-like” design would come at an unacceptable mechanical cost.

Our data answers this question favorably. The results demonstrate that the NHLP provides a level of stability comparable to established fixation methods, effectively dispelling concerns about its mechanical viability. Unexpectedly, the design also exhibited a distinct mechanical advantage: the significantly lower operational stress under physiological loading suggests a reduced long-term risk of fatigue failure. This finding is critical, as it indicates that the goal of soft-tissue preservation does not have to compromise, and may even enhance, the implant's mechanical longevity. Furthermore, the multidirectional proximal locking screws provide the versatility to create a subchondral “raft,” a well-established strategy for resisting articular subsidence (3034). Therefore, this study confirms that the NHLP is not only a mechanically sound option but one that successfully harmonizes the principles of soft-tissue respect with robust biomechanical performance, providing a strong rationale for its continued clinical investigation.

Several limitations of this study should be acknowledged. First, the findings are based on a single computational model derived from the CT data of a healthy young male. This significantly impacts the generalizability of the results, particularly regarding bone quality. The observed biomechanical advantages of the NHLP, such as its lower operational stress compared to other constructs, are contingent on the robust bone stock of this model. It is highly plausible that in an osteoporotic bone model, where bone provides less support, a greater portion of the load would be transferred to the implants, potentially diminishing or even eliminating this stress differential. Therefore, the performance of these constructs in an elderly population remains an open and critical question. Second, the loading conditions were static. This approach was intentionally chosen as the role of FEA in this stage of development is to serve as an initial, cost-effective evaluation of the design's fundamental mechanical viability. If a construct fails to demonstrate adequate performance under static peak loads, proceeding to more resource-intensive cyclic testing, either computationally or on cadaveric models, would be unwarranted. However, it must be acknowledged that this static approach omits the effects of cyclic loading, which are crucial for predicting long-term phenomena such as fatigue-related material failure and progressive screw loosening. Consequently, while our FSF calculations provide a useful proxy for fatigue resistance, a comprehensive fatigue life assessment remains a task for future mechanical testing. Finally, as previously mentioned, the model's exclusion of soft tissues means that the anticipated clinical benefit of avoiding pes anserinus impingement remains a well-founded hypothesis based on anatomical positioning. These limitations collectively define the scope of this study as a foundational, preclinical validation. Future work is essential to address these points, including FEA on varied bone geometries and densities, cadaveric mechanical testing to investigate cyclic performance, and ultimately, prospective clinical trials to confirm the NHLP's real-world efficacy and safety.

5 Conclusions

In this FEA, the NHLP maintained fracture displacement and interfragmentary strain within accepted thresholds for stability in a Schatzker IV tibial plateau fracture model. Under identical loading conditions, the NHLP resulted in a lower peak VMS on the implant than did the TTLP and DRLP. While these computational results support the mechanical feasibility of the NHLP design, they require validation through future cadaveric testing and clinical trials.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

The studies involving humans were approved by Medical Ethics Committee of Nanjing First Hospital. The studies were conducted in accordance with the local legislation and institutional requirements. Written informed consent for participation was not required from the participants or the participants' legal guardians/next of kin in accordance with the national legislation and institutional requirements.

Author contributions

XW: Writing – original draft, Writing – review & editing, Formal analysis, Data curation, Methodology, Conceptualization. HL: Visualization, Writing – review & editing, Data curation. SZ: Visualization, Writing – review & editing. ZT: Writing – review & editing, Resources. ZY: Methodology, Supervision, Writing – review & editing. GW: Writing – review & editing. JY: Writing – original draft, Writing – review & editing, Formal analysis, Data curation, Conceptualization. BL: Conceptualization, Writing – review & editing, Supervision.

Funding

The author(s) declared that financial support was not received for this work and/or its publication.

Acknowledgments

We would like to express gratitude to Gadisa Musa Wako, an English native speaker, for taking the time to revise the language.

Conflict of interest

The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Generative AI statement

The author(s) declared that generative AI was not used in the creation of this manuscript.

Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Footnote

Abbreviations NHLP, novel hockey-stick locking plate; TTLP, traditional T-shaped locking plate; DRLP, double reconstruction locking plates; 3D, three-dimensional; CT, computed tomography; FEA, finite element analysis; VMS, von Mises stress; FSF, fatigue safety factor.

References

1. Zhang Q, Zhao J, Zhang G, Tang J, Zhu W, Nie M. The comparison of clinical effect, knee function, prognosis of double plate fixation and locking plate internal fixation for tibial plateau fractures. Pak J Med Sci. (2022) 38:960–4. doi: 10.12669/pjms.38.4.5340

PubMed Abstract | Crossref Full Text | Google Scholar

2. Schatzker J, Kfuri M. Revisiting the management of tibial plateau fractures. Injury. (2022) 53:2207–18. doi: 10.1016/j.injury.2022.04.006

PubMed Abstract | Crossref Full Text | Google Scholar

3. Gonzalez LJ, Hildebrandt K, Carlock K, Konda SR, Egol KA. Patient function continues to improve over the first five years following tibial plateau fracture managed by open reduction and internal fixation. Bone Jt J. (2020) 102-B:632–7. doi: 10.1302/0301-620x.102b5.bjj-2019-1385.r1

Crossref Full Text | Google Scholar

4. Chang S-M, Hu S-J, Zhang Y-Q, Yao M-W, Ma Z, Wang X, et al. A surgical protocol for bicondylar four-quadrant tibial plateau fractures. International Orthopaedics (SICOT). (2014) 38:2559–64. doi: 10.1007/s00264-014-2487-7

PubMed Abstract | Crossref Full Text | Google Scholar

5. Johnson TR, Oquendo YA, Seltzer R, Van Rysselberghe NL, Bishop JA, Gardner MJ. Incisional negative pressure wound therapy may not protect against post-operative surgical site complications in bicondylar tibial plateau fractures. Eur J Orthop Surg Traumatol. (2024) 34:1173–81. doi: 10.1007/s00590-023-03782-w

PubMed Abstract | Crossref Full Text | Google Scholar

6. Charalambous C, Kwaees T. Anatomical considerations in hamstring tendon harvesting for anterior cruciate ligament reconstruction. Muscles Ligaments Tendons J. (2012) 4:253–7. Available online at: https://www.semanticscholar.org/paper/6fadb7550127522a0a7e86f381df61bcb598fd08 (Accessed January 06, 2026).

Google Scholar

7. Barei DP, Nork SE, Mills WJ, Coles CP, Henley MB, Benirschke SK. Functional outcomes of severe bicondylar tibial plateau fractures treated with dual incisions and medial and lateral plates. J Bone Joint Surg. (2006) 88:1713–21. doi: 10.2106/jbjs.e.00907

PubMed Abstract | Crossref Full Text | Google Scholar

8. Yabuuchi K, Kondo E, Onodera J, Onodera T, Yagi T, Iwasaki N, et al. Clinical outcomes and complications during and after medial open-wedge high tibial osteotomy using a locking plate: a 3- to 7-year follow-up study. Orthop J Sports Med. (2020) 8:2325967120922535. doi: 10.1177/2325967120922535

PubMed Abstract | Crossref Full Text | Google Scholar

9. Martin R, Birmingham TB, Willits K, Litchfield R, LeBel M-E, Giffin JR. Adverse event rates and classifications in medial opening wedge high tibial osteotomy. Am J Sports Med. (2014) 42:1118–26. doi: 10.1177/0363546514525929

PubMed Abstract | Crossref Full Text | Google Scholar

10. Cift H, Cetik O, Kalaycioglu B, Dirikoglu MH, Ozkan K, Eksioglu F. Biomechanical comparison of plate-screw and screw fixation in medial tibial plateau fractures (Schatzker 4). A model study. Orthop Traumatol Surg Res. (2010) 96:263–7. doi: 10.1016/j.otsr.2009.11.016

PubMed Abstract | Crossref Full Text | Google Scholar

11. Huang X, Zhi Z, Yu B, Chen F. Stress and stability of plate-screw fixation and screw fixation in the treatment of Schatzker type IV medial tibial plateau fracture: a comparative finite element study. J Orthop Surg Res. (2015) 10:182. doi: 10.1186/s13018-015-0325-2

PubMed Abstract | Crossref Full Text | Google Scholar

12. Anwar A, Lv D, Zhao Z, Zhang Z, Lu M, Nazir MU, et al. Finite element analysis of the three different posterior malleolus fixation strategies in relation to different fracture sizes. Injury. (2017) 48:825–32. doi: 10.1016/j.injury.2017.02.012

PubMed Abstract | Crossref Full Text | Google Scholar

13. Anwar A, Zhang Z, Lv D, Lv G, Zhao Z, Wang Y, et al. Biomechanical efficacy of AP, PA lag screws and posterior plating for fixation of posterior malleolar fractures: a three dimensional finite element study. BMC Musculoskelet Disord. (2018) 19:73. doi: 10.1186/s12891-018-1989-7

PubMed Abstract | Crossref Full Text | Google Scholar

14. Jyoti N, Mondal S, Ghosh R. Biomechanical analysis of three popular tibial designs for TAR with different implant-bone interfacial conditions and bone qualities: a finite element study. Medical Engineering Physics. (2022) 104:103812. doi: 10.1016/j.medengphy.2022.103812

PubMed Abstract | Crossref Full Text | Google Scholar

15. Araki H, Nakano T, Ono S, Yatani H. Three-dimensional finite element analysis of extra short implants focusing on implant designs and materials. Int J Implant Dent. (2020) 6:5. doi: 10.1186/s40729-019-0202-6

PubMed Abstract | Crossref Full Text | Google Scholar

16. Zheng Z, Liu Y, Zhang A, Chen H, Wan Q, Zhong L, et al. Medial-lateral translational malalignment of the prosthesis on tibial stress distribution in total knee arthroplasty: a finite element analysis. Front Bioeng Biotechnol. (2023) 11:1119204. doi: 10.3389/fbioe.2023.1119204

PubMed Abstract | Crossref Full Text | Google Scholar

17. Papaioannou G, Demetropoulos CK, King YH. Predicting the effects of knee focal articular surface injury with a patient-specific finite element model. Knee. (2010) 17:61–8. doi: 10.1016/j.knee.2009.05.001

PubMed Abstract | Crossref Full Text | Google Scholar

18. Huang N, Xu B, Wu Z, Ye X, He J, Li C, et al. Distinct effects of three knee-preserving surgeries on hip-knee-ankle alignment in patients with knee osteoarthritis. J Orthop Surg Res. (2025) 20:732. doi: 10.1186/s13018-025-06161-9

PubMed Abstract | Crossref Full Text | Google Scholar

19. Lin KM, Boyle C, Marom N, Marx RG. Graft selection in anterior cruciate ligament reconstruction. Sports Med Arthrosc Rev. (2020) 28:41–8. doi: 10.1097/jsa.0000000000000265

PubMed Abstract | Crossref Full Text | Google Scholar

20. Shao Q, MacLeod TD, Manal K, Buchanan TS. Estimation of ligament loading and anterior tibial translation in healthy and ACL-deficient knees during gait and the influence of increasing tibial slope using EMG-driven approach. Ann Biomed Eng. (2011) 39:110–21. doi: 10.1007/s10439-010-0131-2

PubMed Abstract | Crossref Full Text | Google Scholar

21. Maniar N, Schache AG, Pizzolato C, Opar DA. Muscle contributions to tibiofemoral shear forces and valgus and rotational joint moments during single leg drop landing. Scand J Med Sci Sports. (2020) 30:1664–74. doi: 10.1111/sms.13711

PubMed Abstract | Crossref Full Text | Google Scholar

22. Beaulieu ML, Ashton-Miller JA, Wojtys EM. Loading mechanisms of the anterior cruciate ligament. Sports Biomech. (2023) 22:1–29. doi: 10.1080/14763141.2021.1916578

PubMed Abstract | Crossref Full Text | Google Scholar

23. Jiang J, Xu D, Ji Z, Jia R, Wang F, Tan J, et al. Interfragmentary compression force and fixation stability of lateral tibial plateau fractures in normal and osteoporotic bones. J Orthop Res. (2024) 42:1738–47. doi: 10.1002/jor.25832

PubMed Abstract | Crossref Full Text | Google Scholar

24. Prat-Fabregat S, Camacho-Carrasco P. Treatment strategy for tibial plateau fractures: an update. EFORT Open Rev. (2016) 1:225–32. doi: 10.1302/2058-5241.1.000031

PubMed Abstract | Crossref Full Text | Google Scholar

25. Vadgaonkar R, Prameela MD, Kumar CG, Blossom V, Tonse M, Murlimanju BV, et al. Dimensions of pes anserinus of the lower extremity, an anatomical study with its surgical implications. Anat Cell Biol. (2021) 54:178–83. doi: 10.5115/acb.20.275

PubMed Abstract | Crossref Full Text | Google Scholar

26. Arcieri EV, Baragetti S, Božić Ž. Failure analysis of Ti6Al4V titanium alloy under fatigue loading: an experimental and numerical study. Eng Fail Anal. (2024) 164:108715. doi: 10.1016/j.engfailanal.2024.108715

Crossref Full Text | Google Scholar

27. Duan Z, Lu H. Effect of mechanical strain on cells involved in fracture healing. Orthop Surg. (2021) 13:369–75. doi: 10.1111/os.12885

PubMed Abstract | Crossref Full Text | Google Scholar

28. Keaveny TM, Bouxsein ML. Theoretical implications of the biomechanical fracture threshold. J Bone Miner Res. (2008) 23:1541–7. doi: 10.1359/jbmr.080406

PubMed Abstract | Crossref Full Text | Google Scholar

29. Wang X, Zhu Z, Yin Z, Xu H, Jiang D, Xiu H, et al. Early outcomes of modified hockey-stick medial plate in the treatment of Schatzker IV-VI tibial plateau fractures: a retrospective controlled study. J Orthop Surg Res. (2025) 20:345. doi: 10.1186/s13018-025-05761-9

PubMed Abstract | Crossref Full Text | Google Scholar

30. Nishiwaki M, Terasaka Y, Kiyota Y, Inaba N, Koyanagi T, Horiuchi Y. A prospective randomized comparison of variable-angle and fixed-angle volar locking plating for intra-articular distal radius fractures. J Hand Surg. (2021) 46:584–93. doi: 10.1016/j.jhsa.2021.03.014

PubMed Abstract | Crossref Full Text | Google Scholar

31. Chen MJ, Frey CS, Salazar BP, Gardner MJ, Bishop JA. Low profile fragment specific plate fixation of lateral tibial plateau fractures—a technical note. Injury. (2021) 52:1089–94. doi: 10.1016/j.injury.2020.12.037

PubMed Abstract | Crossref Full Text | Google Scholar

32. Castro-Franco AD, Mendoza-Muñoz I, González-Angeles A, Montoya-Reyes MI, Pitalúa-Díaz N. Optimization of locking plate screw angle used to treat two-part proximal humerus fractures to maintain fracture stability. Appl Sci. (2022) 12:4739. doi: 10.3390/app12094739

Crossref Full Text | Google Scholar

33. Ozkaya U, Parmaksizoglu AS. Dual locked plating of unstable bicondylar tibial plateau fractures. Injury. (2015) 46:S9–13. doi: 10.1016/j.injury.2015.05.025

PubMed Abstract | Crossref Full Text | Google Scholar

34. Hu Z, Ren W, Zhang W, Li L, Xu W. Potential problem and solution of lateral plate postposition for the posterolateral tibial plateau fracture. J Orthop Surg Res. (2023) 18:984. doi: 10.1186/s13018-023-04397-x

PubMed Abstract | Crossref Full Text | Google Scholar

Keywords: biomechanics, finite element analysis, novel hockey-stick locking plate, Schatzker classification, tibial plateau fracture

Citation: Wang X, Lai H, Zhang S, Tu Z, Yin Z, Wako GM, Yan J and Liang B (2026) Biomechanical evaluation of a novel hockey-stick locking plate featuring a pes anserinus-sparing design: a finite element analysis. Front. Surg. 12:1722354. doi: 10.3389/fsurg.2025.1722354

Received: 10 October 2025; Revised: 12 December 2025;
Accepted: 29 December 2025;
Published: 15 January 2026.

Edited by:

Wenqi Song, Shanghai Jiao Tong University, China

Reviewed by:

Fengyuan Zhao, Peking University Third Hospital, China
Yifei Yao, Shanghai Jiao Tong University, China

Copyright: © 2026 Wang, Lai, Zhang, Tu, Yin, Wako, Yan and Liang. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Junwei Yan, eWFuX2p3QHllYWgubmV0; Bin Liang, bGlhbmdiaW44MDE3M0AxMjYuY29t

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.