Case Report: Cervical carbon fiber-reinforced PEEK pedicle screw fixation with 3D custom-made template guides for spinal oncology surgery


Fabio Cofano
&#x;,&#x;Fabio Cofano*†,‡Nicola Marengo&#x;Nicola MarengoStefano ColonnaStefano ColonnaFrancesca RizzoFrancesca RizzoFilippo LacatenaFilippo LacatenaDiego Garbossa
Diego Garbossa
  • Neurosurgery Unit, Department of Neuroscience “Rita Levi-Montalcini”, University of Turin, Turin, Italy

Cervical spine fixation in oncological and tumor-like conditions poses unique mechanical and radiological challenges. The choice of implant material is crucial to ensure optimal biomechanical stability and minimize imaging artifacts that may hinder adjuvant radiotherapy. While traditional titanium constructs provide reliable mechanical performance, they generate significant postoperative artifacts. Carbon fiber–reinforced PEEK (CFR-PEEK) implants offer comparable biomechanical strength with markedly reduced imaging interference, facilitating radiotherapy planning. However, their application in the cervical spine remains extremely limited due to narrow anatomical corridors. This technical note reports the first use of CFR-PEEK cervical pedicle screws (CPS) guided by patient-specific 3D-printed templates in the treatment of an aggressive vertebral hemangioma. The study demonstrates the technical feasibility and clinical applicability of CFR-PEEK screws with customized 3D guides for posterior cervical fixation in complex oncologic cases.

Introduction

Spinal fixation in oncological and tumor-like conditions presents specific mechanical and radiological challenges, particularly in the cervical region, where anatomical complexity and biomechanical needs are high. The choice of implant material is crucial in patients who may require frequent and extensive postoperative imaging and adjuvant radiotherapy. Traditional titanium constructs, although mechanically reliable, produce significant artifacts on CT and MRI and can interfere with radiotherapy planning or obscure local recurrence assessment (13).

Carbon fiber–reinforced polyetheretherketone (CFR-PEEK) implants have recently emerged as a valid alternative to metallic systems. These composite materials provide adequate mechanical strength, an elastic modulus closer to cortical bone, and excellent fatigue resistance, while their radiolucency allows clear visualization of bone–implant interfaces and tumor margins (4, 5). Their favorable imaging properties also allow more accurate radiotherapy planning and dose calculation and postoperative surveillance (6). Clinical studies have confirmed their safety and stability, particularly in thoracic and lumbar reconstructions, where comparable outcomes to titanium systems have been found with superior imaging and radiotherapy compatibility (710).

Despite these advantages, the application of carbon fiber–based fixation in the cervical spine remains extremely limited due to smaller anatomical corridors and the need for high mechanical precision. To date, no previous report has described the use of carbon fiber screws for fixation in the cervical region. In this technical note, we present the first known case of cervical spine fixation using carbon fiber–reinforced screws for the treatment of an aggressive vertebral hemangioma. We describe the surgical strategy, technical nuances of screw placement, and postoperative outcomes, highlighting the feasibility and potential benefits of this material in cervical spine.

Case description

A 74-year-old woman presented with a progressively worsening paraparesis that had developed insidiously over the previous four years. Given the gradual but significant decline in lower limb strength and function, she was referred for a neurosurgical evaluation. Clinical assessment raised the suspicion of a compressive spinal lesion, prompting the indication for a complete spinal MRI. The MRI revealed an area of altered signal within the D3 vertebral body, with radiological features suggestive of a vertebral hemangioma. Notably, the lesion demonstrated infiltration of the posterior epidural space and anterior bilateral paravertebral region, raising concern for extraosseous extension and aggressive behavior. To better define the lesion's characteristics and extent, a contrast-enhanced MRI of the cervico-thoracic spine was performed. This confirmed the presence of a morpho-structural alteration involving both the vertebral body and posterior elements of D3. Additionally, pathological epidural and paravertebral tissue extended from D1 to D4 on the right side, causing evident compression of the spinal cord and radiological signs of myelopathy. The imaging findings were compatible with an aggressive vertebral hemangioma with extraosseous growth. To further characterize the lesion, a cervico-thoracic CT scan was obtained, which supported the suspicion of a hypervascular, structurally disruptive lesion centered at D3.

Considering these findings, a multidisciplinary discussion was held in September 2025 involving neurosurgery and interventional neuroradiology teams. Given the extent of the lesion, spinal cord compression, and the patient's progressive neurological symptoms, a combined treatment approach was recommended. Subsequently, the patient underwent spinal angiography, which confirmed a lesional arterial supply originating from the right D3-D4 metameric artery. A preoperative embolization was performed with selective catheterization of the feeder vessel, followed by injection of polyvinyl alcohol (PVA) microparticles and placement of controlled release microcoils. Embolization achieved a satisfactory occlusion of the feeding vessels, forming a dense vascular cast and significantly reducing the lesion's vascularity (Figure 1).

Figure 1
MRI and CT scans of the spine and an angiography series. Images A to C depict spinal MRIs in various views. Images D to F show spinal CT scans with different orientations. Images G to I illustrate angiography, showing changes in blood flow before, during, and after a procedure.

Figure 1. Sagittal (A) and axial (B,C) cervical-thoracic MRI images showing morpho-structural alteration of the D3 vertebral body with posterolateral involvement, associated with epidural and paravertebral pathological tissue causing compressive myelopathy. Cervical-thoracic CT images (D–F) confirming a hypervascular lesion of the D3 vertebral body with osteostructural disruption, consistent with hemangioma. Spinal angiography confirming hypervascular D3 lesion with arterial supply originating from the right D3-D4 metameric artery (G), and subsequent selective embolization using polyvinyl alcohol microparticles and microcoils (H,I).

The following day, the patient underwent surgical intervention. A posterior cervico-thoracic fixation was performed using patient-specific, 3D-printed surgical guides (Medacta MySpine system), allowing for precise placement of carbon fiber pedicle screws at C6, C7, unilateral D4, D5, and D6.

Preoperative planning was based on a high-resolution cervical CT scan with sub-millimetric slice thickness (<1 mm). Two senior spine surgeons independently reviewed the imaging to exclude anatomical configurations incompatible with safe pedicle screw placement, including pedicle transverse angles >45°, pedicle width <3.5 mm, insufficient cancellous channel, and vertebral artery variants involving the pedicle or vertebral body. Patient-specific guides were then generated from the CT dataset. A dedicated planning platform (MySpine Surgical Planning Report, Medacta, Rancate, Switzerland) was used to define the optimal screw entry point, trajectory, and dimensions. Guides were designed to achieve secure, anatomy-matched contact with the dorsal vertebral surface through two cylindrical support interfaces, ensuring intraoperative stability (Figure 2). Templates were manufactured in Polyamide-PA12 using 3D printing and sterilized preoperatively according to the validated protocol. This well-known technology was matched with the introduction in the global market of 4.5 diameter cannulated carbon fiber-reinforced PEEK pedicle screws (Icotec, Altstätten, Switzerland), that enabled cervical pedicle screwing.

Figure 2
Surgical procedure images showing steps in a spine surgery. Image A shows exposed spine without instruments. Image B depicts placement of a spinal retraction system. Image C includes surgical tools and a gloved hand in the operating area. Image D displays placement of spinal screws with instruments in view.

Figure 2. Intraoperative images demonstrating accurate removal of soft tissue from bony landmarks, maximizing the adherence of the template to the dorsal surface of the vertebrae (A,B). Once the template is firmly secured to the vertebrae, a drill bit is used to delineate the pedicle trajectory according to the preoperative surgical plan before placing the screw (C,D).

A transpedicular approach for circumferential decompression and separation surgery was carried out at D3, consisting of bilateral facetectomy and pediculectomy, followed by vertebroplasty of the affected vertebral body using bone cement. This approach allowed for decompression of the spinal canal and partial debulking. Intraoperatively, angiomatoid tissue compressing the thecal sac was meticulously removed, including the ventral components encasing the spinal cord. The extent of decompression was verified with intraoperative ultrasound, which confirmed satisfactory relief of cord compression and reconstitution of the thecal sac. Posterior decompression was enlarged at the levels above and below according to the extent of the disease (Figure 3).

Figure 3
Three-panel image showing surgical procedures and ultrasound imagery. Panel A displays a surgical site with multiple screws implanted in the spine. Panel B shows a close-up of surgical tools and implants on red tissue. Panel C features two ultrasound images displaying different sections, marked with centimeter scales on the right.

Figure 3. Intraoperative images demonstrating posterior C6-D6 fixation using CFR-PEEK pedicle screws (A) close-up image of D3 posterior right-transpedicular circumferential decompression (B), with intraoperative ultrasound confirming satisfactory relief of cord compression (C).

The postoperative course remained uneventful. The patient began cautious mobilization with early motor improvement, and a cervical-thoracic CT scan at one month confirmed optimal positioning of all screws (Gertzbein grade A) and satisfactory vertebroplasty at D3 (Figure 4). Neurological examination showed no new deficits compared with the preoperative assessment, with initial improvement of the preexisting gait disturbance. Following multidisciplinary consultation with radiation oncologists, stereotactic radiosurgery (SRS) was deferred and considered as a potential option should radiological evidence of tumor progression emerge during subsequent follow-up. Clinical and radiological follow-up is ongoing, with progressive improvement in gait function.

Figure 4
CT scan images showing four panels labeled A through D. Panel A shows a sagittal view of the spine, highlighting dense areas suggestive of surgical intervention. Panels B, C, and D display axial views with visible metal implants or screws in the vertebrae, indicating postoperative changes.

Figure 4. Postoperative CT scan demonstrating effective decompression with D3 vertebroplasty (A,B). Axial CT images confirming Gertzbein grade A C6-C7 pedicle screws (C,D).

Discussion

The management of spinal tumors has evolved substantially over the past decades, driven by advances in oncologic therapies and continuous improvements in surgical technology. In parallel, major progress in radiotherapy has transformed the natural history and post-treatment outcomes of these diseases (1115). Today, adjuvant radiotherapy is a cornerstone of spinal tumor management (1618). However, the accuracy of radiotherapy depends not only on modern treatment platforms but also on the quality of postoperative imaging used for planning (1922).

The introduction of CFR-PEEK implants has brought meaningful benefits to spinal tumor surgery. Both in vitro and clinical studies have demonstrated that CFR-PEEK instrumentation is biomechanically comparable to titanium systems in the thoracolumbar spine, with similar bending yield load, axial compression strength, and bending stiffness (6, 2327). In a systematic review by Khan et al., the implant-related complication rate with CFR-PEEK was 7.8%, comparable to titanium constructs (28). Similarly, in a recent case series of 190 patients treated with CFR-PEEK instrumentation, mechanical implant failure occurred in 4.1% of cases at a median follow-up of 13.7 months, consistent with previously reported titanium outcomes (29).

Beyond mechanical performance, CFR-PEEK implants offer major advantages for adjuvant radiotherapy. The low radiological artifact profile allows for more accurate target definition and better optimization of treatment volumes in both photon and proton therapy (1, 2, 8). Reduced imaging distortion also shortens radiotherapy planning correction time and significantly minimizes discrepancies between planned and delivered doses (7, 30).

While CFR-PEEK constructs are well established in thoracolumbar oncologic surgery, their application in the cervical spine remains scarcely documented. Previous reports – most notably by Boriani et al. – described hybrid configurations in which thoracic CFR-PEEK rods were coupled with titanium cervical sublaminar bands positioned outside the irradiation field, achieving favorable outcomes but without addressing the feasibility of a fully carbon-based cervical screw construct (31, 32). To date, no study has reported posterior cervical fixation performed entirely with CFR-PEEK screws, nor the clinical use of CPS manufactured solely from this material.

The present case extends the clinical applicability of CFR-PEEK technology by demonstrating the feasibility of a fully CFR-PEEK posterior cervical screw construct, made possible by the recent development of small-diameter CPS specifically engineered for the cervical spine. This represents a meaningful advance compared with prior hybrid approaches, particularly in the context of spinal oncology where radiotrasparency and artifact reduction are critical.

Compared to lateral mass screws, CPS provide significant biomechanical advantages in terms of pull-out strength and primary stability. After uniplanar cyclic loading, pull-out strengths of 1,214 N vs. 332 N have been reported for CPS and LMS, respectively. Moreover, CPS reached 762 N of pull-out strength compared to 191 N for LMS under torsional loading, and 571 N vs. 289 N in flexion-extension (33, 34). These advantages are particularly valuable in spinal oncology, where bone quality often is compromised and long-term construct stability is essential.

CPS placement remains technically demanding, but recent advances have significantly improved accuracy and safety (3537). In the present case, CPS were inserted using patient-specific 3D-printed guides designed from a high-resolution volumetric preoperative CT scan. Several studies have demonstrated the safety and reliability of this technique, with accuracy rates ranging from 93% to 100% (3844). The main advantage of using 3D-printed guides lies in their collinearity with the dorsal surface of the vertebrae. Accurate soft tissue dissection and patient-specific customized templates ensure optimal stability of the guide on the vertebral surface, significantly minimizing inaccuracies related to vertebral rotation during pedicle cannulation and screw placement as described for the use of neuronavigation. Moreover, the use of unilateral guides markedly reduces paravertebral muscle retraction, lowering the risk of a muscle-pushing effect that could lead to suboptimal lateral trajectories with potential vertebral artery injury.

CPS can also be placed freehand or with intraoperative navigation assistance; however, a recent multicenter comparative study demonstrated that 3D guide–assisted placement achieved significantly higher accuracy and lower complication rates than navigation-assisted placement (97.7% vs. 85.5%, p < 0.001) (45). Moreover, these findings were confirmed by a recent meta-analysis by Bindels et al., which demonstrated no improvement in screw placement accuracy when using navigation-assisted techniques (46). The lower accuracy may be attributed both to vertebral rotation during pedicle cannulation, which is not detected by the navigation system, and to even minimal incidental shifts of the reference arrays during surgical maneuvers. In support of this hypothesis, accuracy using navigation-assisted techniques tends to decrease in the mid-cervical segments (C3–C5), where vertebral mobility and rotational potential are greater compared to the more caudal levels (C6–C7) (4749).

Limitations

The main limitation of this technique is the limited availability of screw sizes, because there is still lack of screws with less than 4.5-mm diameter produced. This means that this technique is not applicable in cases with thin cervical pedicles. Furthermore, given the time required for guide fabrication and delivery (5–10 days), the approach is also unsuitable for urgent cases due to the time required for guide fabrication. Finally, CFR-PEEK implants are substantially more expensive than titanium systems, making careful patient selection essential to maximize the cost–benefit ratio.

Conclusions

This study demonstrates the technical feasibility and clinical applicability of CFR-PEEK cervical pedicle screws for posterior fixation in oncologic surgery, supported by the precision of 3D-printed patient-specific guides. Although current limitations in screw sizing, production time, and cost restrict its widespread adoption, this approach may pave the way for broader use of radiolucent implants in complex cervical and oncologic spine surgery.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

Ethical approval was not required for the studies involving humans because it is a retrospective technical report that does not include prospective experimentation or procedures beyond standard clinical care. The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study. Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.

Author contributions

FC: Conceptualization, Supervision, Writing – original draft, Writing – review & editing. NM: Conceptualization, Visualization, Writing – original draft, Writing – review & editing. SC: Investigation, Resources, Writing – original draft, Writing – review & editing. FR: Formal analysis, Methodology, Writing – original draft, Writing – review & editing. FL: Data curation, Investigation, Writing – original draft, Writing – review & editing. DG: Project administration, Validation, Writing – original draft, Writing – review & editing.

Funding

The author(s) declared that financial support was not received for this work and/or its publication.

Conflict of interest

The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Generative AI statement

The author(s) declared that generative AI was not used in the creation of this manuscript.

Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Footnote

Abbreviations CFR-PEEK, carbon fiber reinforced polyetheretherketone; CPS, cervical pedicle screw; PVA, polyvinyl alcohol; SRS, stereotactic radiosurgery.

References

1. Müller BS, Ryang YM, Oechsner M, Düsberg M, Meyer B, Combs SE, et al. The dosimetric impact of stabilizing spinal implants in radiotherapy treatment planning with protons and photons: standard titanium alloy vs. Radiolucent carbon-fiber-reinforced PEEK systems. J Appl Clin Med Phys. (2020) 21(8):6–14. doi: 10.1002/acm2.12905

PubMed Abstract | Crossref Full Text | Google Scholar

2. Ho D, Corde S, Chen C, Saade G, Betteridge C, Mobbs R. The use of carbon fiber/polyetheretherketone (CF/PEEK) in pedicle screw fixation for spinal neoplasms—potential advantages in postoperative imaging and radiotherapy planning. J Spine Surg. (2024) 10(1):8–21. doi: 10.21037/jss-23-93

PubMed Abstract | Crossref Full Text | Google Scholar

3. Cofano F, Monticelli M, Ajello M, Zenga F, Marengo N, Di Perna G, et al. The targeted therapies era beyond the surgical point of view: what spine surgeons should know before approaching spinal metastases. Cancer Control. (2019) 26(1):1073274819870549. doi: 10.1177/1073274819870549

PubMed Abstract | Crossref Full Text | Google Scholar

4. Alvarez-Breckenridge C, de Almeida R, Haider A, Muir M, Bird J, North R, et al. Carbon fiber-reinforced polyetheretherketone spinal implants for treatment of spinal tumors: perceived advantages and limitations. Neurospine. (2023) 20(1):317–26. doi: 10.14245/ns.2244920.460

PubMed Abstract | Crossref Full Text | Google Scholar

5. Long JR, Kalani MA, Goulding KA, Ashman JB, Flug JA. Carbon-fiber-reinforced polyetheretherketone orthopedic implants in musculoskeletal and spinal tumors: imaging and clinical features. Skeletal Radiol. (2023) 52(3):393–404. doi: 10.1007/s00256-022-04069-7

PubMed Abstract | Crossref Full Text | Google Scholar

6. Cofano F, Di Perna G, Monticelli M, Marengo N, Ajello M, Mammi M, et al. Carbon fiber reinforced vs titanium implants for fixation in spinal metastases: a comparative clinical study about safety and effectiveness of the new “carbon-strategy.”. J Clin Neurosci. (2020) 75:106–11. doi: 10.1016/j.jocn.2020.03.013

PubMed Abstract | Crossref Full Text | Google Scholar

7. Poel R, Belosi F, Albertini F, Walser M, Gisep A, Lomax AJ, et al. Assessing the advantages of CFR-PEEK over titanium spinal stabilization implants in proton therapy—a phantom study. Phys Med Biol. (2020) 65(24):245031. doi: 10.1088/1361-6560/ab8ba0

PubMed Abstract | Crossref Full Text | Google Scholar

8. Ringel F, Ryang YM, Kirschke JS, Müller BS, Wilkens JJ, Brodard J, et al. Radiolucent carbon fiber–reinforced pedicle screws for treatment of spinal tumors: advantages for radiation planning and follow-up imaging. World Neurosurg. (2017) 105:294–301. doi: 10.1016/j.wneu.2017.04.091

PubMed Abstract | Crossref Full Text | Google Scholar

9. Cofano F, Di Perna G, Alberti A, Baldassarre BM, Ajello M, Marengo N, et al. Neurological outcomes after surgery for spinal metastases in symptomatic patients: does the type of decompression play a role? A comparison between different strategies in a 10-year experience. J Bone Oncol. (2021) 26:100340. doi: 10.1016/j.jbo.2020.100340

PubMed Abstract | Crossref Full Text | Google Scholar

10. Cofano F, Marengo N, Pecoraro F, Penner F, Bertero L, Zenga F. Spinal epidural capillary hemangioma: case report and review of the literature. Br J Neurosurg. (2023) 37(4):688–91. doi: 10.1080/02688697.2018.1562034

PubMed Abstract | Crossref Full Text | Google Scholar

11. Bilsky MH, Laufer I, Burch S. Shifting paradigms in the treatment of metastatic spine disease. Spine. (2009) 34(Supplement):S101–7. doi: 10.1097/BRS.0b013e3181bac4b2

PubMed Abstract | Crossref Full Text | Google Scholar

12. Patchell RA, Tibbs PA, Regine WF, Payne R, Saris S, Kryscio RJ, et al. Direct decompressive surgical resection in the treatment of spinal cord compression caused by metastatic cancer: a randomised trial. Lancet. (2005) 366(9486):643–8. doi: 10.1016/S0140-6736(05)66954-1

PubMed Abstract | Crossref Full Text | Google Scholar

13. Cofano F, Di Perna G, Zenga F, Ducati A, Baldassarre B, Ajello M, et al. The neurology-stability-epidural compression assessment: a new score to establish the need for surgery in spinal metastases. Clin Neurol Neurosurg. (2020) 195:105896. doi: 10.1016/j.clineuro.2020.105896

PubMed Abstract | Crossref Full Text | Google Scholar

14. Di Perna G, Ajello M, Colamaria A, Carbone F, Leone A, Tartara F, et al. Transpedicular approach for circumferential decompression and separation surgery as a safe and effective way to maximize surgical treatment for spinal metastases: a multicentric study. Neurosurg Focus. (2025) 58(5):E7. doi: 10.3171/2025.2.FOCUS24971

PubMed Abstract | Crossref Full Text | Google Scholar

15. Colonna S, Bianconi A, Cofano F, Prior A, Di Perna G, Palmieri G, et al. Radiofrequency ablation in vertebral body metastasis with and without percutaneous cement augmentation: a systematic review addressing the need for SPINE stability evaluation. Diagnostics. (2023) 13(6):1164. doi: 10.3390/diagnostics13061164

PubMed Abstract | Crossref Full Text | Google Scholar

16. Laufer I, Iorgulescu JB, Chapman T, Lis E, Shi W, Zhang Z, et al. Local disease control for spinal metastases following “separation surgery” and adjuvant hypofractionated or high- dose single-fraction stereotactic radiosurgery: outcome analysis in 186 patients: clinical article. SPI. (2013) 18(3):207–14. doi: 10.3171/2012.11.SPINE12111

Crossref Full Text | Google Scholar

17. Moussazadeh N, Laufer I, Yamada Y, Bilsky MH. Separation surgery for spinal metastases: effect of spinal radiosurgery on surgical treatment goals. Cancer Control. (2014) 21(2):168–74. doi: 10.1177/107327481402100210

PubMed Abstract | Crossref Full Text | Google Scholar

18. Barzilai O, Amato MK, McLaughlin L, Reiner AS, Ogilvie SQ, Lis E, et al. Hybrid surgery–radiosurgery therapy for metastatic epidural spinal cord compression: a prospective evaluation using patient-reported outcomes. Neurooncol Pract. (2018) 5(2):104–13. doi: 10.1093/nop/npx017

PubMed Abstract | Crossref Full Text | Google Scholar

19. Schneider U, Pedroni E, Lomax A. The calibration of CT hounsfield units for radiotherapy treatment planning. Phys Med Biol. (1996) 41(1):111–24. doi: 10.1088/0031-9155/41/1/009

PubMed Abstract | Crossref Full Text | Google Scholar

20. Giantsoudi D, De Man B, Verburg J, Trofimov A, Jin Y, Wang G, et al. Metal artifacts in computed tomography for radiation therapy planning: dosimetric effects and impact of metal artifact reduction. Phys Med Biol. (2017) 62(8):R49–80. doi: 10.1088/1361-6560/aa5293

PubMed Abstract | Crossref Full Text | Google Scholar

21. Di Perna G, Baldassarre B, Armocida D, De Marco R, Pesaresi A, Badellino S, et al. Application of the NSE score (neurology-stability-epidural compression assessment) to establish the need for surgery in spinal metastases of elderly patients: a multicenter investigation. Eur Spine J. (2024) 33(11):4302–15. doi: 10.1007/s00586-024-08328-0

PubMed Abstract | Crossref Full Text | Google Scholar

22. Cofano F, Giambra C, Costa P, Zeppa P, Bianconi A, Mammi M, et al. Management of extramedullary intradural spinal tumors: the impact of clinical Status, intraoperative neurophysiological monitoring and surgical approach on outcomes in a 12-year double-center experience. Front Neurol. (2020) 11:598619. doi: 10.3389/fneur.2020.598619

PubMed Abstract | Crossref Full Text | Google Scholar

23. Krätzig T, Mende KC, Mohme M, Kniep H, Dreimann M, Stangenberg M, et al. Carbon fiber–reinforced PEEK versus titanium implants: an in vitro comparison of susceptibility artifacts in CT and MR imaging. Neurosurg Rev. (2021) 44(4):2163–70. doi: 10.1007/s10143-020-01384-2

Crossref Full Text | Google Scholar

24. Huber FA, Sprengel K, Müller L, Graf LC, Osterhoff G, Guggenberger R. Comparison of different CT metal artifact reduction strategies for standard titanium and carbon-fiber reinforced polymer implants in sheep cadavers. BMC Med Imaging. (2021) 21(1):29. doi: 10.1186/s12880-021-00554-y

PubMed Abstract | Crossref Full Text | Google Scholar

25. Uri O, Folman Y, Laufer G, Behrbalk E. A novel spine fixation system made entirely of carbon-fiber-reinforced PEEK composite: an in vitro mechanical evaluation. Adv Orthop. (2020) 2020:1–7. doi: 10.1155/2020/4796136

Crossref Full Text | Google Scholar

26. Rosoklija G, Sai S, Barr GQ, Meleis AM. Evaluating long-term construct stability in metastatic spine cancer patients with carbon fiber–reinforced polyetheretherketone instrumentation. Neurosurg Focus. (2025) 58(5):E11. doi: 10.3171/2025.2.FOCUS24995

PubMed Abstract | Crossref Full Text | Google Scholar

27. Morello A, Colonna S, Lo Bue E, Chiari G, Mai G, Pesaresi A, et al. Accuracy and safety between robot- assisted and conventional freehand fluoroscope-assisted placement of pedicle screws in thoracolumbar spine: meta-analysis. Medicina (B Aires. (2025) 61(4):690. doi: 10.3390/medicina61040690

Crossref Full Text | Google Scholar

28. Khan HA, Ber R, Neifert SN, Kurland DB, Laufer I, Kondziolka D, et al. Carbon fiber–reinforced PEEK spinal implants for primary and metastatic spine tumors: a systematic review on implant complications and radiotherapy benefits: presented at the 2023 AANS/CNS joint section on disorders of the spine and peripheral nerves. Journal of Neurosurgery: Spine. (2023) 39(4):534–47. doi: 10.3171/2023.5.SPINE23106

PubMed Abstract | Crossref Full Text | Google Scholar

29. Damante M, Wilson SB, Ward J, Mensah-Arhin K, Gandhi S, Bardeesi A, et al. Durability of carbon fiber–reinforced PEEK spinal instrumentation in spine oncology: a single-institution longitudinal experience. Neurosurg Focus. (2025) 58(5):E10. doi: 10.3171/2025.2.FOCUS24976

PubMed Abstract | Crossref Full Text | Google Scholar

30. Coan JM, Gholipour A, Zhang AT, Conti S, Sorrentino D, Tobert DG. Carbon fiber versus Titanium pedicle screws: a systematic review of radiologic, biomechanical, and clinical features. Spine. (2025) 50(14):998–1009. doi: 10.1097/BRS.0000000000005356

PubMed Abstract | Crossref Full Text | Google Scholar

31. Boriani S, Pipola V, Cecchinato R, Ghermandi R, Tedesco G, Fiore MR, et al. Composite PEEK/carbon fiber rods in the treatment for bone tumors of the cervical spine: a case series. Eur Spine J. (2020) 29(12):3229–36. doi: 10.1007/s00586-020-06534-0

PubMed Abstract | Crossref Full Text | Google Scholar

32. Pipola V, Boriani S, Ghermandi R, Tedesco G, Evangelisti G, Girolami M, et al. Composite peek/carbon fiber pre-shaped rods and sublaminar bands for posterior stabilization of cervico-thoracic junction: a novel technique. J Clin Neurosci. (2020) 72:429–33. doi: 10.1016/j.jocn.2019.12.035

PubMed Abstract | Crossref Full Text | Google Scholar

33. Jones EL, Heller JG, Silcox DH, Hutton WC. Cervical pedicle screws versus lateral mass screws: anatomic feasibility and biomechanical comparison. Spine. (1997) 22(9):977–82. doi: 10.1097/00007632-199705010-00009

PubMed Abstract | Crossref Full Text | Google Scholar

34. Johnston TL, Karaikovic EE, Lautenschlager EP, Marcu D. Cervical pedicle screws vs. Lateral mass screws: uniplanar fatigue analysis and residual pullout strengths. Spine J. (2006) 6(6):667–72. doi: 10.1016/j.spinee.2006.03.019

PubMed Abstract | Crossref Full Text | Google Scholar

35. Abumi K, Itoh H, Taneichi H, Kaneda K. Transpedicular screw fixation for traumatic lesions of the middle and lower cervical spine: description of the techniques and preliminary report. J Spinal Disord. (1994) 7(1):19–28. doi: 10.1097/00002517-199407010-00003

PubMed Abstract | Crossref Full Text | Google Scholar

36. Zhou LP, Zhang RJ, Zhang WK, Kang L, Li KX, Zhang HQ, et al. Clinical application of spinal robot in cervical spine surgery: safety and accuracy of posterior pedicle screw placement in comparison with conventional freehand methods. Neurosurg Rev. (2023) 46(1):118. doi: 10.1007/s10143-023-02027-y

PubMed Abstract | Crossref Full Text | Google Scholar

37. Azimi P, Yazdanian T, Benzel EC, Azimi A, Montazeri A. 3D-printed Navigation template in cervical spine fusion: a systematic review and meta-analysis. Eur Spine J. (2021) 30(2):389–401. doi: 10.1007/s00586-020-06601-6

PubMed Abstract | Crossref Full Text | Google Scholar

38. Farshad M, Spirig JM, Winkler E, Suter D, Farshad-Amacker N, Jarvers JS, et al. Template guided cervical pedicle screw instrumentation. N Am Spine Soc J. (2022) 10:100120. doi: 10.1016/j.xnsj.2022.100120

PubMed Abstract | Crossref Full Text | Google Scholar

39. Jiang L, Dong L, Tan M, Yang F, Yi P, Tang X. Accuracy assessment of atlantoaxial pedicle screws assisted by a novel drill guide template. Arch Orthop Trauma Surg. (2016) 136(11):1483–90. doi: 10.1007/s00402-016-2530-9

PubMed Abstract | Crossref Full Text | Google Scholar

40. Kaneyama S, Sugawara T, Sumi M. Safe and accurate midcervical pedicle screw insertion procedure with the patient-specific screw guide template system. Spine (Phila Pa 1976). (2015) 40(6):E341–348. doi: 10.1097/BRS.0000000000000772

PubMed Abstract | Crossref Full Text | Google Scholar

41. Lu S, Xu YQ, Lu WW, Ni GX, Li YB, Shi JH, et al. A novel patient-specific navigational template for cervical pedicle screw placement. Spine. (2009) 34(26):E959. doi: 10.1097/BRS.0b013e3181c09985

PubMed Abstract | Crossref Full Text | Google Scholar

42. Miyamoto H, Uno K. Cervical pedicle screw insertion using a computed tomography cutout technique: technical note. Journal of Neurosurgery: Spine. (2009) 11(6):681–7. doi: 10.3171/2009.6.SPINE09352

PubMed Abstract | Crossref Full Text | Google Scholar

43. Niu G, Cheng J, Liu L, Li C, Zhou G, Chen H, et al. Individualized 3D printed navigation template- assisted atlantoaxial pedicle screws vs. Free-hand screws for the treatment of upper cervical fractures. Front Surg. (2022) 9:6–8. doi: 10.3389/fsurg.2022.932296

Crossref Full Text | Google Scholar

44. Marengo N, Colonna S, Bue EL, Pesaresi A, Saaid A, Allevi M, et al. Cervical pedicle screw placement with patient-specific 3D-printed guides: accuracy and safety in a clinical experience. Eur Spine J. (2025) 34(4):1248–55. doi: 10.1007/s00586-025-08679-2

PubMed Abstract | Crossref Full Text | Google Scholar

45. Morello A, Colonna S, Carbone F, Balagna A, Leone A, Da Rin Vidal T, et al. Cervical pedicle screws: multicenter comparison of freehand, patient-specific 3D-printed templates, loop-X navigation and OARM/CT navigation accuracy and safety. Glob Spine J. (2025) 0(0):21925682251366645. doi: 10.1177/21925682251366645 (Published online August 4).

Crossref Full Text | Google Scholar

46. Bindels BJJ, Dronkers BEG, Smits MLJ, Verlaan JJ. Accurate placement and revisions for cervical pedicle screws placed with or without navigation: a systematic review and meta-analysis. Global Spine J. (2024) 14(3):1018–37. doi: 10.1177/21925682231196456

PubMed Abstract | Crossref Full Text | Google Scholar

47. Bogduk N, Mercer S. Biomechanics of the cervical spine. I: normal kinematics. Clin Biomech. (2000) 15(9):633–48. doi: 10.1016/S0268-0033(00)00034-6

PubMed Abstract | Crossref Full Text | Google Scholar

48. Navarro-Ramirez R, Lang G, Lian X, Berlin C, Janssen I, Jada A, et al. Total navigation in spine surgery; A concise guide to eliminate fluoroscopy using a portable intraoperative computed tomography 3-dimensional navigation system. World Neurosurg. (2017) 100:325–35. doi: 10.1016/j.wneu.2017.01.025

PubMed Abstract | Crossref Full Text | Google Scholar

49. Shin HK, Jeon SR, Roh SW, Park JH. Benefits and pitfalls of O-arm navigation in cervical pedicle screw. World Neurosurg. (2022) 159:e460–5. doi: 10.1016/j.wneu.2021.12.077

PubMed Abstract | Crossref Full Text | Google Scholar

Keywords: cervical CFR-PEEK pedicle screw, CFR-PEEK, 3D template guides, spinal surgery, spinal oncology, stereotactic radiosurgery

Citation: Cofano F, Marengo N, Colonna S, Rizzo F, Lacatena F and Garbossa D (2026) Case Report: Cervical carbon fiber-reinforced PEEK pedicle screw fixation with 3D custom-made template guides for spinal oncology surgery. Front. Surg. 12:1725931. doi: 10.3389/fsurg.2025.1725931

Received: 15 October 2025; Revised: 20 November 2025;
Accepted: 30 December 2025;
Published: 28 January 2026.

Edited by:

Jon Ramm-Pettersen, Oslo University Hospital, Norway

Reviewed by:

Mirza Pojskic, University Hospital of Giessen and Marburg, Germany
Manlio Barbarisi, Università della Campania Luigi Vanvitelli, Italy

Copyright: © 2026 Cofano, Marengo, Colonna, Rizzo, Lacatena and Garbossa. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Fabio Cofano, ZmFiaW8uY29mYW5vQHVuaXRvLml0

These authors have contributed equally to this work and share first authorship

ORCID:
Fabio Cofano
orcid.org/0000-0002-5100-7478

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.