ORIGINAL RESEARCH article

Front. Oral Health, 25 July 2025

Sec. Oral and Maxillofacial Surgery

Volume 6 - 2025 | https://doi.org/10.3389/froh.2025.1625593

This article is part of the Research TopicCurrent Trends in Oral Surgery and ImplantologyView all articles

Comparison of primary and secondary stability of a one-piece compressive conometric implant system in the posterior upper and lower jaws


Thabat R. Al-Hity,,Thabat R. Al-Hity1,2,3Bilal El-DhuwaibBilal El-Dhuwaib4Abdulsalam R. Al-ZahawiAbdulsalam R. Al-Zahawi5Sarhang Sarwat Gul,

Sarhang Sarwat Gul6,7*
  • 1Department of Maxillofacial and Oral Surgery, College of Dentistry, University of Bilad Alrafidain, Baqubah, Iraq
  • 2Department of Maxillofacial and Oral Surgery, College of Dentistry, University of Sulaimani, Sulaymaniyah, Iraq
  • 3Associate Dental Surgeon, Nationwide Health Care, NHS, Sheffield, United Kingdom
  • 4Academic Unit of Restorative Dentistry, School of Clinical Dentistry, Claremont Crescent, Sheffield, United Kingdom
  • 5Department of Conservative Dentistry, College of Dentistry, University of Sulaimani, Sulaymaniyah, Iraq
  • 6Medical Laboratory Department, College of Health and Medical Technology, Sulaimani Polytechnic University, Sulaymaniyah, Iraq
  • 7Department of Periodontics, College of Dentistry, University of Sulaimani, Sulaymaniyah, Iraq

Background: Implant stability plays a key role in establishing implant osseointegration and is a crucial factor in determining the appropriate timing for functional loading with a fixed prosthesis. This study aims to investigate the primary and secondary stability of a one-piece compressive conometric implant system in the posterior upper and lower jaws.

Materials and methods: A total of 46 compressive one-piece dental implants (DIs) from ROOTT Trade Company were placed in the upper and lower jaws. Ready-made conometric TEC (titanium) with a prosthesis was fixed to the abutment. Implant stability was measured using an implant stability tester (AnyCheck device; NeoBiotech, Republic of Korea) by resonance frequency analysis (RFA) at facial, oral, mesial, distal, and occlusal surfaces. These measures were collected at baseline immediately after implant insertion, after 3 months, and at 6 months. The Kruskal–Wallis test was used to compare between the time points.

Results: In total, 14 patients (8 men, 6 women; mean age 61.08 years; age range 35–79 years) were included. The median baseline RFA values show statistically significant differences between maxillary [55, interquartile range (IQR): 51–61] and mandibular (64, IQR: 60–67) as well as between men (59, IQR: 53–64) and women (67, IQR: 64–69). The occlusal surface shows the highest RFA compared to other surfaces. DI stability significantly increases in the maxilla and men after 6 months compared to the baseline (p = 0.0006 and 0.0001, respectively).

Conclusions: This study suggests that the one-piece compressive conometric implant system demonstrates reliable primary and secondary stability in the posterior upper and lower jaws, with a notable improvement after 6 months.

Introduction

Primary dental implant (DI) stability plays a major role in establishing implant osseointegration and is considered an important factor in determining the appropriate timing for functional loading with a fixed prosthesis. There are different methods to determine implant stability, such as clinically examining DI mobility using a manual instrument, insertion torque value (ITV), using an electronic instrument, and radiograph evaluation of the bone around the implant; however, researchers have found that these techniques are not sensitive enough. Thus, quantitative methods to measure DI stability have been introduced, such as the use of resonance frequency analysis (RFA) (1, 2). The RFA device (Osstell ISO; Integration Diagnostics AB, Gothenburg, Sweden) can measure DI stability with more valuable information, and its range of 1–99 with a value greater than 70 is considered to indicate higher DI stability (35). The damping capacity analysis (DCA) device (Periotest; Medizintectechnik Gulden, Modautal, Germany) is another non-invasive method of measuring DI stability (6). Recently, AnyCheck (Neobiotech, Seoul, Republic of Korea) was introduced as a modified DCA device to measure DI stability (again in the range of 1–99). DI stability values of 1–59 were considered low stability, values of 60–64 indicated moderate stability, and values above 65 indicated high stability (7).

Primary DI stability is achieved at the time of placement and is a mechanical phenomenon influenced by the implant design, placement technique, and the quality and quantity of the local bone. In contrast, secondary DI stability results from bone formation at the implant–tissue interface and within the surrounding bone (7, 8). In addition, the timing of DI loading—whether immediate or delayed—can impact secondary DI stability. Studies have shown that applying different approaches to immediate loading can lead to survival rates comparable with those of conventionally delayed loading (9).

Various types of prostheses are commonly retained to DIs, including removable options with precision attachments and fixed restorations retained either by cement or screws. However, both cement- and screw-retained restorations have disadvantages, including screw loosening and residual cement causing mucosal trauma (10). Conometric retention offers an alternative approach, relying on frictional contact and the elastic deformation of the coping to secure the prosthesis to the implant abutment rather than using cement or screws (11, 12).

One-piece implants with compressive threads, introduced by ROOTT Implant Company, are designed for multiple-unit restorations and can support immediate loading when adequate bone is available. Studies comparing the stability of one-piece and two-piece DIs have shown that both systems are clinically effective and offer consistent, long-term results (13). Recently, the compressive DI has been developed with conometric titanium caps (TECs) available in various sizes and lengths. However, the primary and secondary DI stability of this system has yet to be examined. Notably, the DI with TEC addresses common issues such as screw loosening and mucosal trauma caused by residual cement. Therefore, this study aimed to compare the primary and secondary stability of the compressive conometric DI (ROOTT compressive with conometric cap) in the posterior upper and lower jaws and to determine the optimal timing for functional loading.

Materials and method

This clinical trial was registered at the Clinical Trial Registry of the U.S. National Library of Medicine (registration no. NCT06800508) under the protocol title Primary and Secondary Stability of a One-Piece Compressive Implant System (University of Sulaimani Unique Protocol ID: 227/24). All patients enrolled in the study were already seeking DIs to replace missing teeth. Participation was voluntary, and all patients provided written informed consent after reading the patient information sheet. Each participant was followed up at 3 and 6 months after implantation. Recruitment took place between January 2024 and March 2025 at a private practice in Sulaimani City, Iraq, conducted by expert clinicians in implant and fixed prosthodontics. The inclusion criteria were as follows: missing teeth for more than 2 months, a minimum 10 mm of alveolar bone height, and no pathological findings at the implant site. Patients were excluded if they met any of the following criteria: systemic diseases that could impair osseointegration (e.g. uncontrolled diabetes), radiation therapy to the head or neck region within the past 12 months, smoking more than 10 cigarettes per day, pregnancy or lactation, prior bisphosphonate therapy, history of bone grafting or socket preservation at the implant site, or active periodontal infection.

Implant placement and functional loading protocol placement

Preoperative cone-beam computed tomography was performed for all patients to assess the anatomical features of the implant sites (Figure 1). Elastomeric impressions were taken of both maxillary and mandibular arches, and study casts were fabricated using dental stone. Then, an acrylic surgical guide with a metal sleeve was fabricated to localize the implant insertion site in the jawbone without a flap.

Figure 1
A composite image of dental X-rays and scans showing different views of a jaw. Panel A displays a panoramic X-ray with highlighted nerve pathways in orange. Panel B shows an axial view of the jaw with measurements. Panels C to F present cross-sectional images of individual sections of the jaw with various measurements in green, detailing distances and dimensions of anatomical structures.

Figure 1. Cone-beam computed tomography images for patients included in the study. (A) An overview of the alveolar bone level by orthopantomography. (B) Axial view for the implant sites. (C–F) Sagittal view for measuring the alveolar bone level.

After administration of infiltration local anesthesia, the surgical guide was positioned on the jaw and secured in place. A drilling sequence was performed according to the manufacturer's instructions to prepare a hole for implant placement (Figure 2). A compressive one-piece implant (fixture and abutment) from ROOTT Company, with a length of 10–12 mm, was inserted using a torque wrench set to 35 N/cm (14). Patients were excluded if the insertion torque was below 35 N/cm or if any loss of socket wall integrity, such as fenestration, was observed. A readymade conometric TEC (titanium) was then tapped onto the implant abutment using a small hammer until fully set. Each conometric TEC was selected, fitted, and marked on the abutment.

Figure 2
A close-up view of a dental procedure involving a dental implant. A metal device is being inserted into the upper gum with the aid of a dental guide.

Figure 2. Implant insertion using a customized surgical guide.

To take an impression, a transfer cup was placed on each abutment and tied to it with soft wire to prevent it moving from its position. A one-step elastomeric heavy and light body impression was then taken, and the analog was inserted into the transfer cup. The impression and marked econometric TEC were sent to the technician for construction of the ceramic fused-to-metal bridge. After receiving the bridges and marking each conometric TEC separately within 3 days with an occlusal hole to localize the abutment, the conometric TECs were returned to the corresponding abutments marked previously. The bridges were cemented with glass-ionomer luting cement onto the retained conometric TECs fixed to the DI abutment. After the cement set, the bridge was removed from the oral cavity together with the conometric TEC as one piece using a special screwdriver inserted through the hole in the bridge, which pushed the bridge in the opposite direction to the abutment without damaging the implant. The conometric TEC, removed along with the bridge as a single unit, was cleaned by removing excess cement, polished, and reinserted with light tapping (Figure 3).

Figure 3
Panel A shows a side view of a dental prosthesis in place on the lower jaw. Panel B displays dental implants positioned in the gums without the prosthesis attached. Panel C features a close-up of a two-unit dental bridge with attachments.

Figure 3. (A) After bridge cementation, (B) Removal of the bridge every time to measure the stability of the implant, and (C) The removed bridge.

Dental implant stability

DI stability was measured as RFA using an implant stability tester (AnyCheck device; Neo Biotech, Republic of Korea). Each measurement was repeated six times, following the company’s instructions that the metal rod of the AnyCheck should maintain a contact angle with the abutment within the range 0°–30° upward while measuring (Figure 4).

Figure 4
A dental procedure is being performed on a patient with a mustache. A dental tool wrapped in plastic is used inside the patient's mouth, held by a gloved hand. Another gloved hand helps to stabilize the tool. The patient is covered with a blue drape.

Figure 4. Measuring implant stability with the AnyCheck device.

The first measurement (T1) was taken immediately after implant insertion and DI stability measurements were repeated after 3 months (T2) and 6 months (T3) for all five directions [buccal (B), lingual L), mesial (M), distal (D), and occlusal (O)] after removing the bridge with the driver as mentioned earlier. The bridge was then reinserted using pressure on the corresponding abutment.

Statistical analysis

Data were assessed for normality using the Shapiro–Wilk test. The data were presented as median and interquartile range (IQR) of 25% and 75%. The Mann–Whitney U-test and Kruskal–Wallis test were used to compare the study's time points (T1, T2, and T3). The level of significance was set at p < 0.05 for the comparison of the study’s time points. All calculations were performed using SPSS software package version 26 (IBM Corp., Armonk, NY, USA).

Results

A total of 14 patients (8 men, 6 women; mean age 61.08 years; age range 35–79 years) were included in the study. Out of the 54 implants inserted, only 46 DIs, placed in the upper and lower jaws, were included in the final analysis (Figure 5). The other eight DIs were excluded from the study due to loss of integrity in the socket walls, such as fenestration and dehiscence of the supported bone and torque less than 35 N/cm. Further, participants not adhering to the study's time points were excluded.

Figure 5
Dental imaging montage featuring multiple views of dental implants. Panel A shows a panoramic X-ray with implants in the jawbone. Panel B is an axial view with angles marked from 45 to 110 degrees. Panels C to H display cross-sectional images of implants with corresponding angle labels, highlighting bone integration and density variations. Yellow markers indicate points of interest.

Figure 5. Cone-beam computed tomography images for patients after implant insertion. (A) An overview of the dental implants by orthopantomography. (B) Axial view for the implant placed. (C–H) Sagittal view of dental implants.

The median baseline RFA values shows statistically significant differences between maxillary (55, IQR: 51–61) and mandibular (64, IQR: 60–67) as well as between men (59, IQR: 53–64) and women (67, IQR: 64–69) (Figure 6).

Figure 6
Violin plot comparing RFA values for maxilla, mandibular, male, and female groups. Maxilla and male are represented in blue, mandibular and female in red. Statistical significance is marked above the pairs: 0.001 between maxilla and mandibular, and 0.0001 between male and female.

Figure 6. Comparison of baseline resonance frequency analysis between maxilla and mandibular as well as between men and women.

Of the examined surfaces, the occlusal surface showed the highest RFA compared to other surfaces for both maxillary and mandibular DIs. Meanwhile, as mentioned earlier, the RFA in the maxilla was observed to be lower than its mandibular counterparts at the baseline time point. The RFA increased at the 3- and 6-month time points for the facial, oral, and occlusal surfaces, and these increases were statistically significant at the 6-month time point (p < 0.05). However, the increase in RFA for the mandibular DIs was not statistically significant (Table 1).

Table 1
www.frontiersin.org

Table 1. Comparison of dental implant stability for the examined surfaces at the study's time points in maxillary and mandibular jaws.

Similarly, the RFA values increased at the 3- and 6-month time points compared to baseline for all examined surfaces in men. These increases at 6 months were statistically significant for all surfaces except the mesial surface (p < 0.05). In women, however, this increase in RFA was found not to be statistically significant at either the 3-month or 6-month time point when compared to baseline (Table 2).

Table 2
www.frontiersin.org

Table 2. Comparison of dental implant stability for the examined surfaces at the study's time points in men and women.

In addition, average RFA values of all DIs were compared across the study's time points. The results showed a statistically significant increase in mean RFA values for DIs in the maxilla at 6 months (median 65, IQR: 63–67) compared to baseline (median 55, IQR: 51–61, P = 0.0006) and 3 months (median 59, IQR: 56–62; p = 0.02). Although mean RFA values in the mandible increased at both 3 months (median 65, IQR: 56–69) and 6 months (median 68, IQR: 62–71), these changes were not statistically significant compared to baseline (median 64, IQR: 60–67; p > 0.05). Furthermore, no statistically significant difference was found in DI stability between the maxilla and mandible at the 6-month time point (p = 0.39) (Figure 7).

Figure 7
Violin plot showing RFA measurements over time for Maxilla and Mandibular implants. Maxilla data includes baseline, three months, and six months, with p-values of 0.0006 and 0.02. Mandibular data includes baseline, three months, and six months, with a p-value of 0.39. Each time point is displayed with distinct colors: blue for baseline, red for three months, and white for six months.

Figure 7. Comparison dental implant stability in the maxilla and mandibular at the study's time points.

On the other hand, despite increases in mean RFA values at both 3 months (median 63, IQR: 57–67) and 6 months (median 67, IQR: 63–71) in men compared to baseline (median 59, IQR: 52–64), only the 6-month RFA value showed a statistically significant increase (p = 0.0001). However, no statistically significant differences were found in women when comparing the 3-month (median 62, IQR: 53–67) and 6-month (median 64, IQR: 61–69) RFA values to baseline (median 66, IQR: 63–69; p > 0.05). Lastly, comparison of DI stability at 6 months between men and women revealed no statistically significant difference (p = 0.51) (Figure 8).

Figure 8
Violin plot comparing RFA values for males and females at baseline, three months, and six months. Males (blue and red) show a significant change from baseline to three months (p=0.0001). Females show no significant change (p=0.51).

Figure 8. Comparison of dental implant stability in men and women at the study's time points.

Discussion

This study was conducted to investigate the effect of immediate functional loading on implant stability using a compressive one-piece dental implant retained by a conometric bridge to the fixture as an alternative to cemented and screw type. The study was based on the premise that previous results on immediate functional loading of DIs have been inconclusive (3, 13, 1517). Furthermore, it is important to highlight that the compressive one-piece implant with immediate functional leading was introduced recently and has yet to be examined for its primary and secondary stability over time. The compressive one-piece DI system is specially designed to provide sufficient primary stability and is used in clinical situations involving horizontal alveolar ridge atrophy.

On the other hand, the conometric method applied in this study is an alternative to the cemented type, relying mainly on friction between the abutment and crown. The activation process occurs through applying a suitable insertion force for the conometric, which provides retentive performance equal to that of a cemented system without compromising mechanical functionality (11). Conometric techniques help overcome problems associated with cemented restoration, such as residual cement under the restoration leading to gingival inflammation and damage to the restoration during removal for any reason. This easy removal of the conometric-retained bridge without damaging the fixture or abutment enabled us to measure the stability of the DI separately and accurately at the study's time points.

Measuring DI stability before making a final decision for functional loading influences the DI's success and prognosis. Although both qualitative and quantitative methods are used to assess DI stability, a quantitative method was selected in this study to achieve greater precision (6, 18). It is well acknowledged that primary stability in a compression dental implant system depends on a combination of factors, including bone quality, implant design, insertion torque, placement technique, and bone–implant contact during insertion. Each factor contributes to the mechanical anchorage of the implant, which is essential for ensuring successful osseointegration in the early stages after placement (1923).

Generally, the results of the current study showed that DI stability increases up to 6 months, likely related to osseointegration occurring over this period (24). Significantly higher stability was detected in the mandible compared to the maxilla, which may be related to bone type (type II in the mandible and type IV in the maxilla). Indeed, bone type has been proven to affect DI stability (25). It is important to note that in the current study, DI stability in the maxilla improved significantly after 6 months, with RFA comparable to that of the mandible. It has been reported that compared with delayed loading, immediate loading is associated with a higher incidence of implant failure (13). However, the immediate implant-supported conometric retention system provides fixed retention for a complete prosthesis with four implants and a 100% survival rate over 2 years.

Although our results showed higher DI stability in women compared to men, other studies found no influence of sex on DI stability (26). This can be explained by the fact the primary stability of a compressive DI system significantly relies upon the force applied during implant insertion, which creates a compression of the surrounding bone and hence increases the DI and bone surface area (27, 28). Interestingly, DI stability in men increased at the 6-month time point, matching the increase seen in women at the same time, supporting previous reports that sex does not affect secondary DI stability (29, 30).

This study has some limitations, such as a short follow-up and a lack of comparison with delayed loading or non-compressive implant systems. Nonetheless, to the best of our knowledge, it is the first study to investigate the stability of the compressive DI system with immediate loading using a conometric retainer.

Conclusions

The results of the study indicate the reliability of both the primary and secondary stability of the one-piece compressive conometric implant system in the posterior upper and lower jaws after 6 months. In addition, secondary DI stability improved over the 6-month period. Further studies with a longer follow-up and larger sample size are needed to confirm these findings.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors upon reasonable request.

Ethics statement

The studies involving humans were approved by the Ethics Committee of College of Dentistry, University of Sulaimani. The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study. Written informed consent was obtained from the individual(s) for the publication of any potentially identifiable images or data included in this article.

Author contributions

TA-H: Methodology, Writing – original draft, Conceptualization, Investigation, Resources. BE-D: Data curation, Resources, Validation, Methodology, Writing – original draft. AA-Z: Methodology, Conceptualization, Writing – review & editing, Project administration, Formal analysis. SG: Writing – review & editing, Supervision, Formal analysis, Validation, Investigation.

Funding

The author(s) declare that no financial support was received for the research and/or publication of this article.

Acknowledgments

Thanks to Zahawi Dental Lab for performing all prosthodontic tasks.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Generative AI statement

The author(s) declare that no Generative AI was used in the creation of this manuscript.

Publisher's note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

1. Molly L. Bone density and primary stability in implant therapy. Clin Oral Implants Res. (2006) 17:124–35. doi: 10.1111/j.1600-0501.2006.01356.x

PubMed Abstract | Crossref Full Text | Google Scholar

2. Jar C, Archibald A, Gibson M, Westover L. An analytical model to measure dental implant stability with the Advanced System for Implant Stability Testing (ASIST). J Mech Behav Biomed Mater. (2024) 150:106238. doi: 10.1016/j.jmbbm.2023.106238

PubMed Abstract | Crossref Full Text | Google Scholar

3. Sarfaraz H, Johri S, Sucheta P, Rao S. Study to assess the relationship between insertion torque value and implant stability quotient and its influence on timing of functional implant loading. J Indian Prosthodont Soc. (2018) 18:139–46. doi: 10.4103/jips.jips_203_17

PubMed Abstract | Crossref Full Text | Google Scholar

4. Khan MW, Inayat N, Zafar MS, Zaigham AM. A resonance frequency analysis to investigate the impact of implant size on primary and secondary stability. Pak J Med Sci. (2024) 40:1261–6. doi: 10.12669/pjms.40.6.8213

PubMed Abstract | Crossref Full Text | Google Scholar

5. Olmedo-Gaya M-V, Romero-Olid M-N, Ocaña-Peinado FM, Vallecillo-Rivas M, Vallecillo C, Reyes-Botella C. Influence of different surgical techniques on primary implant stability in the posterior maxilla: a randomized controlled clinical trial. Clin Oral Investig. (2023) 27:3499–508. doi: 10.1007/s00784-023-04962-y

PubMed Abstract | Crossref Full Text | Google Scholar

6. Lee J, Pyo S-W, Cho H-J, An J-S, Lee J-H, Koo K-T, et al. Comparison of implant stability measurements between a resonance frequency analysis device and a modified damping capacity analysis device: an in vitro study. J Periodontal Implant Sci. (2020) 50:56–66. doi: 10.5051/jpis.2020.50.1.56

PubMed Abstract | Crossref Full Text | Google Scholar

7. Atsumi M, Park SH, Wang HL. Methods used to assess implant stability: current status. Int J Oral Maxillofac Implants. (2007) 22:743–54.17974108

PubMed Abstract | Google Scholar

8. Monje A, Ravidà A, Wang HL, Helms JA, Brunski JB. Relationship between primary/mechanical and secondary/biological implant stability. Int J Oral Maxillofac Implants. (2019) 34:s7–s23. doi: 10.11607/jomi.19suppl.g1

PubMed Abstract | Crossref Full Text | Google Scholar

9. Nkenke E, Fenner M. Indications for immediate loading of implants and implant success. Clin Oral Implants Res. (2006) 17(Suppl 2):19–34. doi: 10.1111/j.1600-0501.2006.01348.x

PubMed Abstract | Crossref Full Text | Google Scholar

10. Chee W, Jivraj S. Screw versus cemented implant supported restorations. Br Dent J. (2006) 201:501–7. doi: 10.1038/sj.bdj.4814157

PubMed Abstract | Crossref Full Text | Google Scholar

11. Bressan E, Lops D, Tomasi C, Ricci S, Stocchero M, Carniel EL. Experimental and computational investigation of Morse taper conometric system reliability for the definition of fixed connections between dental implants and prostheses. Proc Inst Mech Eng H. (2014) 228:674–81. doi: 10.1177/0954411914545556

PubMed Abstract | Crossref Full Text | Google Scholar

12. Degidi M, Nardi D, Sighinolfi G, Degidi D. The conometric concept for the definitive rehabilitation of a single posterior implant by using a conical indexed abutment: a technique. J Prosthet Dent. (2020) 123:576–9. doi: 10.1016/j.prosdent.2019.02.020

PubMed Abstract | Crossref Full Text | Google Scholar

13. Chen J, Cai M, Yang J, Aldhohrah T, Wang Y. Immediate versus early or conventional loading dental implants with fixed prostheses: a systematic review and meta-analysis of randomized controlled clinical trials. J Prosthet Dent. (2019) 122:516–36. doi: 10.1016/j.prosdent.2019.05.013

PubMed Abstract | Crossref Full Text | Google Scholar

14. Venkatakrishnan C, Bhuminathan S, Chandran CR. Dental implant insertion torque and bone density–short review. Biomed Pharmacol J. (2017) 10:1305–9. doi: 10.13005/bpj/1234

Crossref Full Text | Google Scholar

15. Daher FI, Abi-Aad HL, Dimassi HI, Baba NZ, Majzoub ZA. Factors affecting implant stability quotients at immediately and conventionally loaded implants in the posterior maxilla: a split-mouth randomized controlled trial. J Prosthodont. (2021) 30:590–603. doi: 10.1111/jopr.13296

PubMed Abstract | Crossref Full Text | Google Scholar

16. Gjelvold B, Kisch J, Chrcanovic BR. A randomized clinical trial comparing immediate loading and delayed loading of single-tooth implants: 5-year results. J Clin Med. (2021) 10:1–9. doi: 10.3390/jcm10051077

Crossref Full Text | Google Scholar

17. Pardal-Peláez B, Flores-Fraile J, Pardal-Refoyo JL, Montero J. Implant loss and crestal bone loss in immediate versus delayed load in edentulous mandibles: a systematic review and meta-analysis. J Prosthet Dent. (2021) 125:437–44. doi: 10.1016/j.prosdent.2020.01.032

PubMed Abstract | Crossref Full Text | Google Scholar

18. Chang WJ, Lee SY, Wu CC, Lin CT, Abiko Y, Yamamichi N, et al. A newly designed resonance frequency analysis device for dental implant stability detection. Dent Mater J. (2007) 26:665–71. doi: 10.4012/dmj.26.665

PubMed Abstract | Crossref Full Text | Google Scholar

19. Rues S, Schmitter M, Kappel S, Sonntag R, Kretzer JP, Nadorf J. Effect of bone quality and quantity on the primary stability of dental implants in a simulated bicortical placement. Clin Oral Investig. (2021) 25:1265–72. doi: 10.1007/s00784-020-03432-z

PubMed Abstract | Crossref Full Text | Google Scholar

20. Heimes D, Becker P, Pabst A, Smeets R, Kraus A, Hartmann A, et al. How does dental implant macrogeometry affect primary implant stability? A narrative review. Int J Implant Dent. (2023) 9:20. doi: 10.1186/s40729-023-00485-z

PubMed Abstract | Crossref Full Text | Google Scholar

21. do Vale Souza JP, de Moraes Melo Neto CL, Piacenza LT, Freitas da Silva EV, de Melo Moreno AL, Penitente PA, et al. Relation between insertion torque and implant stability quotient: a clinical study. Eur J Dent. (2021) 15:618–23. doi: 10.1055/s-0041-1725575

PubMed Abstract | Crossref Full Text | Google Scholar

22. Romero-Serrano M, Romero-Ruiz MM, Herrero-Climent M, Rios-Carrasco B, Gil-Mur J. Correlation between implant surface roughness and implant stability: a systematic review. Dent J (Basel). (2024) 12:1–18. doi: 10.3390/dj12090276

Crossref Full Text | Google Scholar

23. Karl M, Grobecker-Karl T. Effect of bone quality, implant design, and surgical technique on primary implant stability. Quintessence Int. (2018) 22:189–98. doi: 10.3290/j.qi.a39745

Crossref Full Text | Google Scholar

24. Huang Y-C, Huang Y-C, Ding S-J. Primary stability of implant placement and loading related to dental implant materials and designs: a literature review. J Dent Sci. (2023) 18:1467–76. doi: 10.1016/j.jds.2023.06.010

PubMed Abstract | Crossref Full Text | Google Scholar

25. Yoon HG, Heo SJ, Koak JY, Kim SK, Lee SY. Effect of bone quality and implant surgical technique on implant stability quotient (ISQ) value. J Adv Prosthodont. (2011) 3:10–5. doi: 10.4047/jap.2011.3.1.10

PubMed Abstract | Crossref Full Text | Google Scholar

26. Shiffler K, Lee D, Rowan M, Aghaloo T, Pi-Anfruns J, Moy PK. Effect of length, diameter, intraoral location on implant stability. Oral Surg Oral Med Oral Pathol Oral Radiol. (2016) 122:e193–8. doi: 10.1016/j.oooo.2016.06.016

PubMed Abstract | Crossref Full Text | Google Scholar

27. O’Sullivan D, Sennerby L, Meredith N. Influence of implant taper on the primary and secondary stability of osseointegrated titanium implants. Clin Oral Implants Res. (2004) 15:474–80. doi: 10.1111/j.1600-0501.2004.01041.x

Crossref Full Text | Google Scholar

28. Chong L, Khocht A, Suzuki JB, Gaughan J. Effect of implant design on initial stability of tapered implants. J Oral Implantol. (2009) 35:130–5. doi: 10.1563/1548-1336-35.3.130

PubMed Abstract | Crossref Full Text | Google Scholar

29. Vollmer A, Saravi B, Lang G, Adolphs N, Hazard D, Giers V, et al. Factors influencing primary and secondary implant stability—a retrospective cohort study with 582 implants in 272 patients. Appl Sci. (2020) 10:8084. doi: 10.3390/app10228084

Crossref Full Text | Google Scholar

30. Fu PS, Lan TH, Lai PL, Chen CH, Chen JH, Wang JC, et al. Implant stability and marginal bone level changes: a 2-year prospective pilot study. J Dent Sci. (2023) 18:1272–9. doi: 10.1016/j.jds.2023.02.018

PubMed Abstract | Crossref Full Text | Google Scholar

Keywords: compressive implant, conometric, one-piece implant, primary stability, secondary stability

Citation: Al-Hity TR, El-Dhuwaib B, Al-Zahawi AR and Gul SS (2025) Comparison of primary and secondary stability of a one-piece compressive conometric implant system in the posterior upper and lower jaws. Front. Oral Health 6:1625593. doi: 10.3389/froh.2025.1625593

Received: 9 May 2025; Accepted: 7 July 2025;
Published: 25 July 2025.

Edited by:

Francisco Correia, University of Porto, Portugal

Reviewed by:

Albraa Alolayan, Taibah University, Saudi Arabia
Nipun Dhalla, Manav Rachna International Institute of Research and Studies (MRIIRS), India

Copyright: © 2025 Al-Hity, El-Dhuwaib, Al-Zahawi and Gul. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Sarhang Sarwat Gul, c2FyaGFuZy5oYW1hQHVpbnZzdWwuZWR1Lmlx; c2FyaGFuZy5oYW1hQHNwdS5lZHUuaXE=

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.