Abstract
Introduction:
Adhesive surface electrodes are worthwhile to explore in detail as alternative to subcutaneous needle electrodes to assess myogenic evoked potentials (MEP) in human and horses. Extramuscular characteristics of both electrode types and different brands are compared in simultaneous recordings by also considering electrode impedances and background noise under not mechanically secured (not taped) and taped conditions.
Methods:
In five ataxic and one non-ataxic horses, transcranial electrical MEPs, myographic activity, and noise were simultaneously recorded from subcutaneous needle (three brands) together with pre-gelled surface electrodes (five brands) on four extremities. In three horses, the impedances of four adjacent-placed surface-electrode pairs of different brands were measured and compared. The similarity between needle and surface EMGs was assessed by cross-correlation functions, pairwise comparison of motor latency times (MLT), and amplitudes. The influence of electrode noise and impedance on the signal quality was assessed by a failure rate (FR) function. Geometric means and impedance ranges under not taped and taped conditions were derived for each brand.
Results:
High coherencies between EMGs of needle-surface pairs degraded to 0.7 at moderate and disappeared at strong noise. MLTs showed sub-millisecond simultaneous differences while sequential variations were several milliseconds. Subcutaneous MEP amplitudes were somewhat lower than epidermal. The impedances of subcutaneous needle electrodes were below 900 Ω and FR = 0. For four brands, the FR for surface electrodes was between 0 and 80% and declined to below 25% after taping. A remaining brand (27G DSN2260 Medtronic) revealed impedances over 100 kΩ and FR = 100% under not taped and taped conditions.
Conclusion:
Subcutaneous needle and surface electrodes yield highly coherent EMGs and TES–MEP signals. When taped and allowing sufficient settling time, adhesive surface-electrode signals may approach the signal quality of subcutaneous needle electrodes but still depend on unpredictable conditions of the skin. The study provides a new valuable practical guidance for selection of extramuscular EMG electrodes. This study on horses shares common principles for the choice of adhesive surface or sc needle electrodes in human applications such as in intraoperative neurophysiological monitoring of motor functions of the brain and spinal cord.
Highlights
- -
Surface and sc needle recordings in horses represent extramuscular highly coherent EMG activity.
- -
Adhesive gel electrodes approach the quality of sc needle electrodes for any EMG activity under specific conditions.
- -
The signal quality depends on the adhesive electrode type.
- -
A few brands of adhesive gel electrodes can replace subcutaneous needle electrodes for TC-MEP recording when taped.
- -
Failure rates of surface-electrode MEPs are higher when compared to sc needle electrodes.
Introduction
Transcranial stimulation (TS) has become a standard technique for assessment of the motor function of the spinal cord. Using transcranial magnetic stimulation (TMS), this technique was introduced in the equine community by and evolved as a diagnostic tool in horses (, ,, ). Recently, transcranial electric stimulation (TES) was introduced as an alternative method (, , ). Both TS techniques are applied under sedation and are well tolerated and painless in horses. The elicited MEPs can be recorded either intramuscular (im) or extramuscular (em). Insulated needle electrodes with uncoated tips belong to the im class (Skinner et al., 2008). These sense the electrical activity from just a few muscle fibers and reflect characteristics of the peripheral motor neuron and are therefore appropriate for diagnostic assessment of the lower motor neuron function (Wijnberg et al., 2003; Wijnberg and Franssen, 2016). The wave shapes of im recorded transcranial MEPs (TC-MEP) show typical polyphasic patterns with relative high amplitudes, which greatly differs from recordings from extramuscular electrode types (Skinner et al., 2008). MEPs can also be recorded extramuscular either at the surface of the skin by adhesive surface electrodes or subcutaneously (sc) by uncoated needle electrodes. Extramuscular MEPs typically reflect electrical activity of many motor units representing a large population of motor neurons whereas characteristics of single motor neurons are filtered out. Therefore, em MEP recordings are appropriate for assessing the upper motor neuron function since these sense the integral activity of many axons of spinal motor tracts and interconnected neurons and are insensitive for the activity of individual lower motor neurons (Skinner et al., 2008; ).
As noticed, wave shapes and amplitudes of em and im MEPs are fundamentally different: in contrast to em recordings, im MEPs are highly polyphasic due to the dominant influence of only a few motor neurons. However, motor latency times (MLT) appear interchangeable between em and im electrodes (Skinner et al., 2008; ). Although em EMG recordings can be obtained with both surface or sc needle electrodes and thus both are applied in many human and animal studies, to our knowledge, no profound comparison between the two electrode types has been performed up until now. Only certain viewpoints on the subject have been addressed (; ).
In horses, TC-MEPs have been reported for both intramuscular and extramuscular electrodes. First transcranial elicited MEPs (TC-MEPs) in horses were recorded extramuscular by alligator clips placed on skin folds () and later on by sc needle electrodes (, , ) and im needle electrodes (, ,, ). More recently in horses, adhesive surface electrodes seem an attractive practical alternative for sc needle or alligator clip electrodes since these simply can be placed right away on the dense haired skin without any further preparation. These electrode types have successfully been applied in equine ECG recordings (Verheyen et al., 2010). This is surprising because the skin is covered with sebum, epidermal tissue, residual sweat, and a dense hair layer. These compounds have highly isolating characteristics, which are known to degrade the signal quality of surface electrodes. This is a well-known problem in human patients, which can be counteracted by skin preparation such as abrasion and prior cleansing with ethanol (; ) or more successfully by using abrasive conductive paste (). Sometimes, these actions are not sufficient to overcome the disturbing electrode noise, which can be a problem when signal amplitudes are expected to be very small like during intraoperative monitoring of patients with expected weak muscle MEPs like in paresis cases. This can occur in patients suffering from, for example, cerebral aneurysms, spinal cord tumors, and neuromuscular scoliosis. To overcome this doubt, sc electrodes are often an a priori choice in those circumstances (; ; Szelényi et al., 2010; Takeda et al., 2018). Inspired by the successful application of adhesive surface electrodes in ECGs, reported the use of adhesive surface electrodes for recording of TMS-MEPs in horses. These were well tolerated. However, only MLTs of im and em recorded MEPs from TMS could be compared, due to the study setup. This is because other MEP-characterizing parameters also strongly depend on whether the EMG recordings are from intramuscular or extramuscular origin. Therefore, a genuine conclusion on relevant characteristics of surface electrodes is precluded in that study. This problem is avoided when MEPs from surface electrodes are compared with MEPs of other extramuscular electrode types such as sc needle electrodes. Also in the aforementioned study (), MLTs were recorded sequentially in time and not pairwise and are therefore dominated by random physiological test-to-test variations and errors from magnetic coil repositioning (, ). These sequential variations can be ruled out by pairwise comparison of simultaneously recorded EMGs from sc needle and surface electrodes with a sub-millisecond precision. Finally, as mentioned by , it is known that surface electrodes can have high noise levels, which may therefore reduce the quality of the obtained signal. Especially in horses suffering from either spinal cord or upper motoneuron pathology and in which manifestation of weak MEPs is to be expected, occurrence of high noise levels might impede proper diagnosis. It is well known that background noise of sc electrodes is significantly lower when compared to the background noise of adhesive surface electrodes and that high electrode impedance most often is associated with increased electrode background noise (). Most of the noise is likely of electrochemical origin at the electrode contact interface and depends on the composition of the electrode gel (). In human for example, impedances of adhesive surface AG/AgCl electrodes can be 1–2 magnitudes higher than that reported for sc needle electrodes (; ; ). High electrode impedances also increase the sensitivity for power line noise (; ). Extra fastening of the electrodes by circumferential applied tape, further called “taping,” might reduce electrode impedance and thus reduce the background noise when compared to absent mechanical fixation (not taped condition). The low impedances of sc needle electrodes are associated with typical low background noise levels in clinical practice ().
The expected attractive features of adhesive surface electrodes in practice are still worthwhile to further explore. This was reason to set up a novel study design (1) to compare simultaneously obtained recordings of spontaneous EMG activity and TES-induced MEPs of a series of sc needle electrode and adhesive surface electrodes of different brands, (2) to assess impedance values and associated background noise of these electrodes under both not taped and taped conditions, and (3) to provide practical recommendations with respect to the tested electrode types and brands.
Materials and Methods
Materials
Six warmblood horses, consisting of two stallions, one gelding, and three mares, aged 10.07 ± 5.81 years (mean ± SD), were included in the study. The height at withers was 160.9 ± 10.2 cm (mean ± SD). Five horses showed clinical signs of muscle weakness and ataxia and were suspected of suffering from spinal or intracranial injury. These horses were subjected to a standard protocol for multipulse TES–MEP assessment in the Equine Clinic in Wolvega, the Netherlands. A 6th horse without clinical signs of weakness or ataxia was not subjected to the neurophysiological assessment and solely included for performance of impedance measurements of surface-electrode types.
Methods
Horses were subjected to a standardized neurological exam and were prepared as previously described (). The neurophysiological tests were performed under sedation by intravenous administration of detomidine (Detosedan, AST Farma B.V., Oudewater, Netherlands) and butorphanol (Butomidor, AST Farma B.V., Oudewater, Netherlands) (both 1.5–2.0 μg/kg body weight in total).
Stimulation
Two stimulation needle electrodes (L 35 mm, Ø 0.45 mm, type RMN35/0.45 Electrocap BV, Nieuwkoop, Netherlands) were inserted sc parallel to each other and caudo-rostrally on the forehead. The stimulation needle electrodes were separated 5 cm from each other, with their middle points 2.5 cm bilateral from the central location Cz on the forehead. The horses were discharged after completion of the procedure and a final clinical examination.
Transcranial electric stimulation was performed according to a standardized diagnostic protocol using biphasic multipulse trains of constant voltage generated by an intraoperative neurophysiological monitoring system (Neuro-Guard JS Center, Bedum, Netherlands). Multipulse TES was performed with three biphasic pulses per train (ppt), pulse width of (pw) 0.1 ms/phase, and interpulse interval (ipi) of 1.3 ms. The stimulation voltage was stepwise increased according to a protocol as previously defined (). At each voltage step, TS was performed 2–4 times. The diagnostic MEP parameters were retrieved at respectively 10, 20, and 30 V above the motor threshold (MT). No further TES stimulations were added to the clinical diagnostic protocol for the current study. Each TES session was extended with free-running EMG recordings and impedance measurements of the electrodes to explore differences between surface-electrode types.
Data Acquisition
Spontaneous EMG activity and TES-induced MEPs were recorded bilaterally on all four limbs from sc needle electrode pairs which were pair-by-pair placed, interspaced about 3–4 mm: (N1) Rochester 82015-PT L 12 mm 27GA Rochester Lutz, FL, United States, (N2) RLSND 110 L 13 mm, Ø 0.4 mm Rhythmlink, Columbia, SC, United States, (N3) Medtronic dual electrode 13 mm 27G DSN2260, Medtronic Xomed, Inc., Jacksonville, FL, United States (Figure 1). One pair consisted of one type, while complimentary pairs consisted of different types. To enable the pairwise EMG and MEP recordings for this study, the sc needle electrodes were overlaid by adhesive surface electrodes of type (S1) SkintactTM FS50 wet gel foam (Innsbruck, Austria) while keeping needle insertion entries a few millimeters outside the conducting pad area. The skin was left uncleansed and unabraded with the covering hair unclipped prior to placement of the surface electrodes. The sc needle and surface-electrode combinations were placed over the musculus extensor carpi radialis (ECR) (10 and 20 cm above the os carpi accessorium) and the musculus tibialis cranialis (TC) (10 and 20 cm above the medial malleolus). Sc needle electrode pairs and corresponding surface-electrode pairs were connected to the differential inputs of the physiological amplifiers of the measuring system. This allowed for simultaneous recording of four-channel needle and surface-electrode pairs, totaling eight channels. A ground needle electrode was placed sc in the neck at the right side of the horse.
FIGURE 1
The input impedance of the physiological amplifiers is 20 MΩ at 50 Hz. The common mode rejection rate is 80 dB. The analog-to-digital conversion was performed in combination with an eight-channel multiplexer using a resolution of 1 μV. The sample frequency per channel was 4.3 kHz.
EMG Recordings
For the study, each session was extended with free-running EMG recordings and impedance measurements of the electrodes to explore differences between surface-electrode types. A bandpass filter was used with a high-pass filter of 50 Hz and a low-pass filter of 2500 Hz (3-dB cutoff level). No further analog or digital filtering techniques were applied. Since in the first three horses it became obvious that the S1-type surface electrode showed relative high background noise, the idea was raised that other surface-electrode types might perform better. Therefore, it was decided to also have an orienting look at impedances of an additional series of different adhesive surface-electrode types in the 4th and 5th ataxic horses. Due to the limited available time, the number of additional surface-electrode types to be tested was restricted to a maximum of 2, leaving limited possibilities to perform repeated measurements. This was the reason to include a 6th horse in which solely surface electrodes were applied for impedance measurements without performing TES measurements. This allowed to also repeat the impedance measurements over a time frame of about 10–15 minutes, which agrees with the necessary adaptation times reported for gelled electrodes (; ) under not taped and taped conditions. This allowed in the 6th horse for execution of four impedance measurements per electrode type and impedance measurements of a total of 16 surface electrodes (4 impedances/type). In the 4th, 5th, and 6th horses, additional surface-electrode brands were attached to the skin near the Skintact electrodes. The additional tested surface-electrode brands were as follows: (S2) 3MTM Red DotTM 2259-50 solid gel soft cloth; (S3) 3MTM Red DotTM Foam Ag/AgCl Monitoring Electrode, 3M Deutschland GMBH Neuss, Germany; (S4) Medtronic Dual Electrode DSE3125 adhesive hydrogel on silver/silver chloride Medtronic-Xomed, Inc., Jacksonville, FL, United States; and (S5) AmbuTM Blue Sensor P-00-S/50 hydrogel, Ambu S/A Ballerup, Denmark. The available extra time for the sessions of the first five horses was limited to 30 min. This set a limit to the number of free-running background signal measurements of the extra surface electrodes leaving only the possibility of impedance measurements. A survey of the five surface-electrode types is given in Figure 1.
In all neurophysiological assessed horses, TES–MEPs and time series with spontaneous EMG activity and/or background noise, which still may contain low-amplitude EMG activity, were recorded in four signal pairs from all extremities using different sc needle and Skintact adhesive surface electrodes. Continuous time series of all needle and surface-electrode types were recorded not taped and sealed by tape wrapped around the limb. TES–MEP series were obtained with tape-sealed electrodes on all occasions.
Impedance Measurements
After calibration with a 2-kΩ test resistor, impedance measurements were performed of all examined electrode pairs under not taped and taped conditions. The impedance resulted from rectification and RC low-pass filtering of the amplified AC voltage as measured over electrode pairs when applying a small 2.5-kHz AC constant current. The device was part of the equipment applied for intraoperative neurophysiological monitoring and intended to check the electrode impedances in patients. The frequency of the AC current was determined by an oscilloscope at 2.5 kHz. The time interval between placement of the electrodes and impedance measurements was at least 5 minutes for adaptation of the electrode–tissue interface. Impedance measurements and background noise recordings were repeated after 10–15 min to check for reproducibility.
Data Processing
Considered TES–MEP parameters were wave shape, MLT, MEP amplitude, and cross-correlation values of MEPs. In addition, impedance, background noise, and failure rate (FR) were assessed for the different electrode types.
Motor Latency Times
The MLTs were defined as “the time lag between the onsets of stimulation and MEP response” when these were unambiguously distinguishable from baseline noise level.
TES–MEP, EMG, and Background Noise Amplitudes
The amplitudes of TES–MEPs were retrieved within the transcranial time window (TCW) between cessation of the stimulation artifact and onset of extracranially elicited late MEP responses, which are reflexes typical for horses, to preclude their interfering effects (). Amplitude parameters were measured as peak–peak differences (APP) and root-mean-square (RMS) values (ARMS).
The made assumptions of nearby identical myographic signals of sc needle and surface electrodes apply to all kinds of EMG activity as were verified for TC-MEPs, reflexes, and background EMG activity. When cross-correlations and RMS amplitudes of sweeps were compared, time windows were selected over signal parts of interest.
Background noise amplitudes were determined on silent parts of the signal recordings, where no EMG activity was observed. When necessary, time windows were adapted to exclude interference by transient EMG activity. RMS amplitudes of selected signal intervals were graphically visualized in a scatter plot visualizing the relation between background noise and electrode impedance.
Statistical Analyses
Statistical analysis was performed with SPSSTM software, version 20.0.0, IBMTM.
Comparison of EMG Wave Shapes of sc Needle and Surface Myograms and Reproducibility
The cross-correlation function Rxy(τ) quantifies the cross-correlation between sample time series x(t) and y(t) according to the equation:
where x(t) and y(t) are the time functions of myograms from the sc needle and surface electrodes or from two subsequent myograms of one channel, t is the discrete value of time, and τ is the time shift between the two time functions, and Wstart agrees with the sample number at the start and Wend with the sample number at the end of the time window W in which the cross correlation is computed. τ is computed between −14 and 14 ms in time steps of 1/fs, where fs is the sample frequency. Wstart was chosen at 19 ms to exclude influences of TES artifacts of multipulse TES and the length of W 150 ms. These choices are applied in all autocorrelation and cross-correlation calculations. The maximum of Rxy(τ) was used to quantify the correlation between sc needle and surface-electrode signal pairs or between subsequent recorded myograms. Rxy(0) is similar to Pearson’s correlation for τ = 0. When the channel numbers are different, τ is corrected for the multiplex delay bias in the 8-channel sample cycle. The algorithm is programmed in Borland C++, version 3.1. The autocorrelation and cross-correlation functions Rxx(τ), Ryy(τ), and Rxy(τ) and RMS amplitudes together with the specifications of the stimulus parameters and time window W of each sample series were transferred to SPSS databases for subsequent statistical processing. The full-width half (FWHM) of the autocorrelation and cross-correlation functions served as a measure to indicate the width of the signal waves of the MEPs, spontaneous EMG activity, and background noise and used when reviewing the relation with the reviewed time functions.
Comparison of TES–MEP Parameters of Subcutaneous Needle and Surface Electrodes
A descriptive analysis of TES–MEP parameters was performed on MLTs, amplitude ratios, and cross-correlation factors between signal pairs. For each case and muscle group (m), the following values were computed: (a) mean MLT for needle electrodes, mMLTn,m with standard deviation SDn,m (N = 6) and (b) mean differences of MLTs between surface (s) and needle (n) electrode pairs mMLTs–n,m with standard deviation SDs–n,m (N = 6). The averages were obtained from repeated stimulations (2 per voltage) at 10, 20, and 30 V above MT. The standard deviations SDn,m depend also on variations of the MLT in time where MLTs differences are computed after averaging. This sensitivity for time-varying influences is excluded in the standard deviation SDs–n,m of MLT differences of electrode pairs, which are computed before averaging. The normality of the relevant differences was graphically assessed using qq plots (MLT and MEP amplitudes in the linear and background noise amplitudes and impedances in the logarithmic domain). MLTs and their intra-individual and interindividual differences were case-wise compared in paired-samples Student t-tests under the null hypotheses of means being zero.
Failure Rate, Impedance, and Background Noise Under Not Taped and Taped Conditions
The impedances and background electrode noise and their statistical tests resulting in differences and 95% confidence intervals were processed in the logarithmic domain before back conversion in the linear domain. When converted back to the linear domain, differences become ratios. In the linear domain, Gmean becomes the geometric mean of RMS amplitude values. GMU/T denotes the improvement of the quality of the electrodes after taping defined as the ratio of not taped Gmean/taped Gmean, where the indices U refer to not taped and T to taped (sealed) conditions. Impedance range ratios are defined as the quotient of the upper and lower limits. Background amplitude-impedance pairs of electrodes under taped and not taped conditions were plotted bi-logarithmic from which a FR function, expressing the fraction of rejections of baseline noise amplitudes exceeding an ARMS = 12 μV RMS (APP ≅ 50 μV) level, was computed as running average with a window containing 8 subsequent impedances. The averages were fitted by a local regression curve using bi-weight 50% smoothing parameters in a locally weighted scatterplot smoothing (LOESS) procedure ().
In all statistical comparisons, a significance level of p = 5% was applied.
Results
Similarity Between Myographic Signals of Subcutaneous Needle and Surface Electrodes
Figures 2A–C show the expected high coherent resemblance of each of the four subsequent recordings (sweeps) of myographic signal pairs of the surface and sc needle electrode recordings for either stationary continuous EMG activity (Figure 2A) or TES–MEPs (Figures 2B,C). TES–MEPs are shown in a severely ataxic horse (Figure 2B) and in a mild ataxic horse (Figure 2C). The coherent appearance of the sc and surface signal pairs is objectified by the high correlation seen at the maxima of the cross-correlation functions Rsn(τ) in respectively Figures 2D–F. Each function belongs to one sweep. After correction for the time bias between multiplexed samples, all cross-correlation functions depict sharp delineated maxima at τ = 0. The high cross-correlation values provided in Figures 2D–F agree with the highly coherent courses of all four subsequent signal pairs depicted in Figures 2A–C.
FIGURE 2
Most correlation values are only partly affected by moderate background noise levels, unless when strong surface-electrode noise fully masks the visibility of EMG activity as seen in Figure 2G. Figure 2G shows the devastating effects of strong surface-electrode noise on continuous EMG activity, resulting in non-significant low cross-correlation values around zero as seen in graph 2 h. This belongs to one exception of the right ECR of case 2 (Table 1). Moderate noise levels partly impair the correlation as shown in case 2 (Table 1) where correlations are reduced to 0.699–0.796 and 0.847–0.899 for the right and left TC muscles. These are still significant for p < 0.01.
TABLE 1
| Muscle group | Case 1 | Case 2 | Case 3 | Case 4 | Case 5 | |||||
| 95% confidence intervals | ||||||||||
| Rns(0) | As/n | Rns(0) | As/n | Rns(0) | As/n | Rns(0) | As/n | Rns(0) | As/n | |
| ECR L R | 0.937 | 0.919 | 0.963 | 0.932 | 0.950 | 0.919 | 0.980 | 0.961 | 0.961 | 0.941 |
| 0.967 | 0.999 | 0.967 | 1.04 | 0.975 | 0.999 | 0.985 | 0.977 | 0.983 | 0.962 | |
| 0.959 | 0.952 | −0.022 | 2.37 | 0.965 | 0.944 | 0.990 | 0.999 | 0.840 | 0.953 | |
| 0.967 | 1.00 | −0.181 | 4.15 | 0.976 | 0.993 | 0.993 | 1.03 | 0.901 | 0.978 | |
| TC L R | 0.955 | 0.624 | 0.699 | 0.835 | 0.917 | 0.611 | 0.991 | 0.996 | 0.906 | 0.903 |
| 0.973 | 0.831 | 0.796 | 0.987 | 0.966 | 0.813 | 0.993 | 1.02 | 0.946 | 0.931 | |
| 0.979 | 0.927 | 0.847 | 0.773 | 0.882 | 0.927 | 0.903 | 0.881 | 0.975 | 0.979 | |
| 0.989 | 1.028 | 0.899 | 0.941 | 0.932 | 1.03 | 0.945 | 0.923 | 0.988 | 0.994 | |
95% confidence intervals of cross-correlations at τ = 0 values between surface and needle electrodes Rns(0) and amplitude ratios As/n of the RMS amplitudes of the surface and needle electrodes in five horses (cases).
The figures for each side are obtained from 2 subsequent TES-MEP recordings at 20 V above MT. The bold figures result from strong interfering noise in the surface EMG observed along with high surface-electrode impedance supporting the observation of the high amplitude ratios and not significant low correlations.
Table 1 provides a statistical overview of cross-correlation values Rns(0) and amplitude ratio’s As/n of all four muscle groups in five cases. A ratio of 1 denotes equal ARMS amplitudes in both signals of a pair. The overview of cross-correlation values of surface and sc needle MEP-signal pairs of all but one of 20 muscle groups depicted in Table 1 shows them to be highly significant since most values are close to 1.
ARMS of surface electrodes are somewhat lower than those of sc needle electrodes shown by AS/N values ranging between 0.8 and 1.0 (18/20 muscle groups) and in a left TC muscle group (case 2) between 0.7 and 0.8. The high AS/N range between 2.4 and 4.2 recorded at the right ECL muscle group of case 2 is ascribed to the high surface-electrode noise level.
The width of the bell-shaped Rns(τ) curves is related to the widths of the waves in the continuous EMG and MEP recordings. Smaller full-width-half-maximum (FWHM) values are noticed in the continuous EMG in a neighborhood of 1 ms, whereas FWHM values of MEPs are about 3–4 ms. Cross-correlation functions between subsequent TC-MEPs vary and show random variations of τ of about 2 ms. These variations appear synchronous to MLT fluctuations.
The mean of the paired differences mMLTs–n (Table 2) is in 85% of 20 muscle groups within −0.2 and 0.2 ms. SDs–ns are below 0.3 ms in 55% and below 0.6 ms in 90% of the muscle groups. This is much lower than the SDs in subsequent MLT measurements of below 3.0 ms in 70% and below 6 ms in 85% of cases. One exception with SDs–n = 2.0 ms (left TC, case 2) is found where MLTs of small strong varying prolonged polyphasic wave patterns become inaccurate due to high surface-electrode noise. The variations of SDn of sequential MLTs are about one magnitude higher than those of SDs–ns of paired differences: SDn values for thoracic limb muscles are below 3.1 and for pelvic limb muscles below 9.5 ms. The 9.5 ms is an outlying value in a case with prolonged MLTs.
TABLE 2
| Muscle group | Case 1 | Case 2 | Case 3 | Case 4 | Case 5 | |||||||||||||||
| Mean tn ts-tn ms | SD tn tn-ts ms | Mean tn tn-ts ms | SD tn tn-ts ms | Mean tn tn-ts ms | SD tn tn-ts ms | Mean tn tn-ts ms | SD tn tn-ts ms | Mean tn tn-ts ms | SD tn tn-ts ms | |||||||||||
| ECR L R | 23.1 | 0.05 | 2.79 | 0.23 | 24.6 | −0.05 | 2.16 | 0.23 | 26.6 | −0.02 | 2.85 | 0.27 | 21.8 | 0.14 | 0.44 | 0.29 | 20.9 | 0.11 | 1.31 | 0.10 |
| 23.0 | 0.16 | 3.14 | 0.37 | 24.8 | −0.27 | 1.41 | 0.34 | 27.8 | 0.00 | 2.92 | 0.35 | 22.3 | 0.09 | 1.40 | 0.23 | 20.3 | 0.08 | 0.513 | 0.57 | |
| TC L R | 46.3 | 0.11 | 9.47 | 0.93 | 49.6 | −0.90 | 2.40 | 1.97* | 60.2 | −0.05 | 4.74 | 0.32 | 44.0 | −0.17 | 6.29 | 0.28 | 34.9 | 0.03 | 1.46 | 0.22 |
| 44.3 | 0.05 | 8.72 | 0.56 | 51.4 | −0.64 | 1.37 | 0.53 | 56.2 | 0.36* | 5.95 | 0.59* | 40.2 | −0.17 | 2.33 | 0.28 | 35.2 | −0.02 | 1.19 | 0.40 | |
Survey of mean MLTs of needle electrodes (meantn) and mean of paired MLT differences (meants–tn) between surface and needle electrodes with respective standard deviations SDn and SDs–n in five horses.
* Worsened latency time readability caused by high interference noise in the surface EMG. The number of cases for each mean is 6. The ratios between the ECR and TC muscle groups 11.9x (left), 8.6 (right), 9.9 (left), and 15.7 (right) show about one magnitude differences between SDtn and SDts–tn.
Comparison of MLT
Table 2 shows means of MLTs of all muscle groups and cases together with means of paired MLT differences between surface and sc needle electrodes with the respective standard deviations SDN and SDS–N. Test-to-test variations are incorporated in SDN and excluded in SDS–N within-test MLT pairs.
Relation Between Electrode Noise and Impedances
Figure 3 depicts how the cross-correlation Rns(0) between the surface and sc needle electrode MEPs is influenced by two different electrode noise levels of respectively ARMS 18 μV (a) and ARMS = 108 μV (b).
FIGURE 3
The relation between electrode background noise amplitude and impedance is depicted in the double logarithmic scatter plot in Figure 4A for sc needle and surface electrodes in not taped and taped conditions.
FIGURE 4

(A) Scatter plot of RMS MEP amplitudes as a function of the electrode impedance (N = 114 points). (B) Running average of the failure rate (FR) over a frame width of eight subsequent impedance values. The FR values are plotted at the end of the frames. FR is computed as the percentage of MEP amplitudes within the frame above ARMS = 12 μV, which agrees with about APP = 50 μV (horizontal dashed line). The vertical dashed lines denote the impedance borders below which in panel a no background noise amplitudes exceed ARMS = 12 μV and in (B) FR = 0. (B) LOESS local regression curve of the FR as function of electrode impedance. A vertical dashed line indicates the impedance value at which the FR becomes greater than zero.
Evolution of the Surface-Electrode Impedance Over Time
The impedances of a total of 16 surface electrodes (4 impedances/type) showed over the course of 15 min highly significant (p < 0.001) decreases in time with mean rates of: not taped 95.4% and taped 97.3% decrease. This decrease was not seen in the S4-type electrodes. The impedance reduction over time in our study agrees with the 40% reduction reported by
Impedance Ranges of sc Needle and Surface Electrodes Under Not Taped and Taped Conditions
Impedance ranges of all electrode types, impedance range percentage of impedances surpassing 10 kΩ, and FRs are provided in Table 3 for all tested sc needle (N1–3) and surface electrode (S1–5) types under either not taped or taped conditions.
TABLE 3
| Electrode type | Condition | Impedance range Ω | GMU/T | p | % >10 kΩ | FR at ARMS >12 μV | ||
| Lower range limit | Gmean | Upper range limit | ||||||
| N1 | Not taped | 423 | 613 | 915 | 1.20x | 0.1 | 0% | 0% |
| Taped | 398 | 510 | 679 | 0% | 0% | |||
| N2 | Not taped | 420 | 429 | 502 | 1.04x | 0.48 | 0% | 0% |
| Taped | 375 | 409 | 509 | 0% | 0% | |||
| N3 | Not taped | 412 | 431 | 566 | 0.88x | 0.25 | 0% | 0% |
| Taped | 402 | 487 | 548 | 0% | 0% | |||
| Overall needle | 375 | 505 | 915 | 0% | 0% | |||
| S1 | Not taped | 876 | 3.6k | 14.8k | 3.65x | 0.007** | 24% | 65% |
| Taped | 312 | 986 | 4.4k | 7% | 30% | |||
| S2 | Not taped | 1.95k | 45.9k | >100k | 29.9x | 0.001** | 83% | 72% |
| Taped | 467 | 968 | 6.58k1 | 17% | 27% | |||
| S3 | Not taped | 1.35k | 2.70k | 38.5k1 | 5.93x | 0.000** | 46% | 58% |
| Taped | 338 | 456 | 854 | 23% | 0% | |||
| S4 | Not taped | >100k | >100k | >100k | 0.94x | 0.76 | 100% | 100% |
| Taped | 92k | >100k | >100k | 100% | 100% | |||
| S5 | Not taped | 1.98k | 5.72k | 12k | 5.13x | 0.000** | 50% | 69% |
| Taped | 545 | 1.11k | 2.28k | 0% | 21% | |||
Survey of electrode impedance range percentages of impedances (test frequency 2.5 kHz) surpassing 10 kΩ, and failure rates (FR) of needle (N1–N3) and surface electrode (S1–S5) types under not taped and taped conditions.
Excluded outliers: 11 case > 100 kΩ. ** Significant for P < 0.01. Gmean: geometric mean. GMU/T ratio: not taped Gmean/taped Gmean.
Background Noise
The background noise of surface electrodes usually is inversely related to the impedance. However, the relationship between both parameters shows a large variability (
Figure 4B shows the running average of the FRs at discrimination level APP > 50 μV of the total number of data pairs. The wide scatter above 900 Ω comprises a mix of acceptances and rejections of all electrodes at discrimination level APP = 50 μV. The LOESS regression curve through the points of the running average FR function visualized in Figure 4B emerges at 810 Ω (vertical dashed line) by an initial steep increase that gradually loses steepness at increasing impedances. FR levels of 20% are already reached at 1 kΩ, 50% at between 4 and 5 kΩ, and 80% at over 100 kΩ. In Table 3, the FR values are computed as ratios at a discrimination level at APP = 50 μV. The ratios represent the fraction of rejections within the impedance ranges of all electrode types and taping conditions. For all tested sc needle electrodes, the FR = 0 and corresponding impedances were well below 900 Ω. Non-significant differences at large overlapping ranges are found between needle electrode types and also between all tested hydrogel electrode types. However, there is one important exception for the Medtronic surface electrodes (type S3) which show all high impedances in a range well above 100 kΩ. The GMU/T ratios depict the improvement of the signal quality of surface electrodes by taping. Overall impedance ranges of sc needle electrodes are much smaller than those of surface electrodes. This is clearly reflected in Table 3 where range ratios are between 1.3 and 1.9x. Taping does not change the impedances of needle electrodes significantly but remains useful to prevent their dislodging.
We experienced different qualities of adhesive properties where the fixating properties of the S3 electrode was the best and worst for the S4 electrode which often detached spontaneously.
Discussion
Similarity of Myographic Recordings Between sc Needle and Surface Electrodes
The current study provides an objective comparison of different electrode types and brands and for extramuscular measurements and offers physicians a new valuable practical guidance for selection of extramuscular EMG electrodes. Besides the diagnostic application in horses, the equivalence between EMG and TES–MEPs of sc needle and surface electrodes is also still an underexplored subject in human intraoperative neurophysiological monitoring (
It was postulated that extramuscular recorded EMGs in horses from surface and sc needle electrodes share the same wave morphology and amplitudes and are about equal with ARMS,s somewhat smaller than ARMS,n. Similar findings are described for ECG recordings by surface and sc needle electrodes (
As expected, the results of our study show convincing coherent similarity between extramuscular EMGs of surface and sc needle electrodes. This pertains to continuous EMG and all TES–MEP-characterizing parameters as wave morphology, amplitudes, and MLTs. The equivalence of MEP waveforms is visible in signal pairs of continuous background EMG activity with relative low amplitudes (Figure 2A), in the dispersed polyphasic transcranial and reflex MEPs in an ataxic horse (Figure 2B), and in pathological delayed MEPs of relative larger amplitudes in a horse with mild ataxia (Figure 2C).
The high-correlation factors and amplitude ratios near AS/N = 1, depicted in Table 1 of surface-needle MEP pairs, allow to conclude that both em electrode types deliver (1) equal wave forms of MEPs and (2) about equal MEP amplitudes, with ARMS of surface electrodes somewhat lower than sc needle electrodes. The outlier with the high AS/N range of the right ECL muscle group of case 2 is ascribed to the high surface electrode noise; (3) sharp delineated maximum cross-correlation factors RNS(τ) are all at τ = 0. This implies that surface electrodes are well aligned over needle electrodes. When somewhat displaced from each other, this could theoretically cause a time shift between the two coherent signals when the motor endplate zone is not enclosed by the electrode pairs. When the velocity of traveling action potentials along muscle fibers is for example 4 m/s, the passage time difference by 4 mm displaced electrodes would be 1 ms. Accordingly, RNS(τ) is maximal at τmax = 1 ms. For relative narrow cross-correlation functions, like in Figure 2D, the maximum cross-correlation RNS(τmax) would greatly be underestimated by RNS(0). In horses, no data are available of muscle fiber conduction velocities and geometry of motor endplate zones in muscles. This was the specific reason to include the possibility to detect non-zero τmax values in the methods of this study.
For the MLT, the sequential statistics are comparable with those reported in the study of
Cross-correlation functions depict within which time-range signals remain coherent. When these time ranges are clearly contained within the sweep lengths, the sweep-to-sweep correlations of stationary noise signals, like from continuous EMG activity, are uncorrelated. This implies that the wave morphology in each sweep is different. This also applies for the continuous EMG in Figure 2A where the correlation values are very low between −0.055 and 0.085.
Basically, cross-correlations are connected to the reproducibility of the wave shapes of signals. High cross-correlations indicate a high similarity between the morphology of the signals. A cross-correlation of 1 implies a strict linear relation between signals. This means that their shapes are exact copies of each other, while their amplitudes may differ. The cross-correlation can be considered as a measure of reproducibility of the wave morphology. Low correlations indicate a poor while higher values reflect a better reproducibility of the wave shapes. The wave shapes are different when the correlation is not statistical significant.
The correlation between successive MEPs in Figure 2B is also low and varies between −0.2 and 0.5. This agrees with the poor reproducibility of highly variable complex polyphasic and delayed weak MEP waves in a marked ataxic horse. Higher correlation values between 0.2 and 0.6 in Figure 2C reflect better reproducible morphology of polyphasic MEPs in a horse with mild ataxia.
Differences in wave morphology of subsequent MEPs as expressed by the correlation between subsequent sweeps make comparison of wave parameters as MLT and MEP morphology less reliable, as has been performed in the study of
Effects of Surface-Electrode Noise
When MEP amplitudes are sufficiently high as in healthy horses, surface electrodes appear as an appealing alternative for sc needle electrodes (
The study shows that taping is very effective to improve the signal quality of surface electrodes: (1) the impedances of a few surface electrodes even underpass those of sc needle electrodes. This may be ascribed to a good adaptation of the electrolyte interface between electrode and skin in combination with the larger contact surface of the electrodes. Larger contact surfaces imply lower impedances (
This study supports the conclusion that in healthy horses where MEP amplitudes are large compared to electrode noise, adhesive surface electrodes can replace sc needle electrodes, but only when the electrodes are being taped (
The orienting look at different types of surface and sc needle electrodes permits to conclude that all types, except S4, can be used for extramuscular EMG recording after additional securing by tape. However, the zero FRs of sc needle electrodes are still not obtained for types S1, S2, S4, and S5. All impedance measurements with S4 surface electrodes remained well above 100 kΩ. With one exception of 100 kΩ, most authors accept a maximum skin impedance below 10 kΩ in practice (
Moreover, the possibility of circumferential taping applies only to muscle groups in extremities. Reduction in FRs of surface electrodes by taping elsewhere on the body is precluded. This leaves sc needle electrodes on those locations as the only reliable choice.
Conclusion
In horses, and likely also in intraoperative neurophysiological applications, sc needle and gelled surface electrodes reflect extramuscular EMGs as highly coherent EMG and MEP signals whereas adhesive surface electrodes only approach the signal quality of sc needle electrodes when taped and allowing enough settling time. However, the signal quality of surface electrodes remains vulnerable to unpredictable conditions of the skin. Differences between surface-electrode types are apparent and need to be elaborated in a larger study population and also confirmed in human patients. The current study provides a new valuable guidance for physicians for selection of extramuscular EMG electrodes when performing neuro-electrical diagnostic tests or for intraoperative use. The study shares common principles for the choice of adhesive surface or sc needle electrodes in human applications such as in intraoperative neurophysiological monitoring of motor functions of the brain and spinal cord.
Study Limitations
- -
Small number of included horses.
- -
When excluding the S4 surface electrode, the different electrode types are only studied in three horses and give not yet sufficient insight in horse-to-horse variations and statistical support for a generalized opinion on differences between four adhesive gel surface electrodes.
- -
Surface electrode characteristics may approach the subcutaneous needle electrode quality when allowing long electrode settling times like 1 h and preparation with skin cleansing, abrading, and hair clipping before starting MEP measurements. This is not examined.
Statements
Data availability statement
The datasets generated for this study are available on request to the corresponding author.
Ethics statement
The horses that were included in the study were client-owned horses that were presented to the Clinic for diagnostic work-up for complaints of spinal ataxia. With that respect horses were subjected to the standard diagnostic work-up protocol for spinal ataxia of which multipulse TES-MEP assessment is a standard part. No extra, nor additional interventions aside from the veterinary necessary work-up approach were applied in these horses. Only a set of extra (non- invasive) adhesive electrodes was placed on the legs. Impedance measurements of the electrodes were performed in conjunction with the multipulse TES-MEP assessment work-up. The 6th horse is the privately owned horse of the first author. This horse was only provided with a set of adhesive surface electrodes placed on the skin of her legs (non-invasive). No multipulse TES-MEP assessment was performed. The owner(s) informed consent for participation in this study, either written or oral was given.
Author contributions
SJ and HJ: equal contribution to this article in design, acquisition, analysis, interpretation, and writing. HB and SR: critical revision important for the intellectual content based on their professional background. CB: provision of experimental facilities, giving practical advices, and revision of the manuscript. CD: involvement in interpretation, writing, and revision.
Acknowledgments
The work was performed in the Wolvega Equine Clinic.
Conflict of interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Abbreviations
- τ
time shift parameter in autocorrelation and cross-correlation functions
- APP
peak-to-peak amplitude
- ARMS
RMS amplitude
- AS/N
RMS amplitude ratio between surface and needle electrode pairs
- dB
decibel
- ECR
musculus extensor carpi radialis
- em
extramuscular
- EMG
electromyogram
- FP–RMS
conversion factor from ARMS to APP;
- FR
failure rate
- FWHM
full-width-half-maximum
- Gmean
geometric mean of amplitude
- GMU/T
ratio of Gmean values under taped and not taped conditions
- Hz
hertz
- im
intramuscular
- ipi
interpulse interval
- m
index muscle group m
- MEP
(muscular) motor evoked potential
- MLT
motor latency time
- mMLTn
mean MLT needle electrode MEP
- mMLTs–n
mean of MLT needle-surface-electrode pair differences
- MT
motor threshold
- n
index needle electrode
- ppt
pulses per train
- pw
pulse width
- RMS
root-mean-square
- Rxx(τ)
autocorrelation function of signal x(t)
- Rxy(τ)
cross-correlation function between signals x(t) and y(t)
- Ryy(t)
autocorrelation function of signal y(t)
- s
index surface electrode
- sc
subcutaneous
- SD
standard deviation
- SNR
signal to noise ratio
- TC
musculus tibialis cranialis
- TC-MEP
transcranial MEP
- TCW
transcranial time window
- TES
transcranial electric stimulation
- TMS
transcranial magnetic stimulation
- TS
transcranial stimulation
- W
time window.
References
1
AshramY. A.YinglingC. D. (2008). Handbook of Clinical Neurophysiology.Amsterdam: Elsevier B.V.
2
AydinlarE. I.DikmenP. Y.KocakM.BaykanN.SeymenN.OzekM. M. (2018). Intraoperative neuromonitoring of motor-evoked potentials in infants undergoing surgery of the spine and spinal cord.J. Clin. Neurophysiol.3660–66. 10.1097/WNP.0000000000000523
3
Bellardine BlackC. L.StrombergK.Van BalenG. P.GhanemR. N.BreedveldR. W.TielemanR. G. (2010). Is surface ECG a useful surrogate for subcutaneous ECG?PACE Pacing Clin. Electrophysiol.33135–145. 10.1111/j.1540-8159.2009.02616.x
4
BerendsH. I.JournéeH. L. (2018). Influence of the montage of stimulation electrodes for intraoperative neuromonitoring during orthopedic spine surgery.J. Clin. Neurophysiol.35419–425. 10.1097/WNP.0000000000000498
5
BrostromS.JennumP.LoseG. (2003). Motor evoked potentials from the striated urethral sphincter: a comparison of concentric needle and surface electrodes.Neurourol. Urodyn.22123–129. 10.1002/nau.10030
6
CrumB. A.StrommenJ. A. (2008). Intraoperative Peripheral Nerve Stimulation and Recording.Amsterdam: Elsevier B.V. 10.1016/S1567-4231(07)08024-0
7
DeletisV.SalaF. (2008). Intraoperative neurophysiological monitoring of the spinal cord during spinal cord and spine surgery: a review focus on the corticospinal tracts.Clin. Neurophysiol.119248–264. 10.1016/j.clinph.2007.09.135
8
ErtekinC.UludagB.OnA.YetimalarY.ErtasM.ColakogluZ.et al (1998). Motor-evoked potentials from various levels of paravertebral muscles in normal subjects and in patients with focal lesions of the spinal cord.Spine231016–1022. 10.1097/00007632-199805010-00010
9
GeddesL. A. (1972). Interface design for bioelectrode systems.IEEE Spectr.941–48. 10.1109/MSPEC.1972.5218524
10
GirouxB.LamontagneM. (1990). Comparison between surface and intramuscular wire electrodes in isometric and dynamic conditions.J. Biomech.30397–405. 10.1016/0021-9290(89)90256-X
11
GodinD. T.ParkerP. A.ScottR. N. (1991). Noise characteristics of stainless-steel surface electrodes.Med. Biol. Eng. Comput.29585–590. 10.1007/BF02446089
12
GonzalezA. A.CheongsiatmoyJ.ShilianP.ParikhP. (2018). Comparison of transcranial motor evoked potential amplitude responses between intramuscular and subcutaneous needles in proximal thigh muscle.J. Clin. Neurophysiol.35431–435. 10.1097/WNP.0000000000000490
13
GrimnesS. (1983). Impedance measurement of individual skin surface electrodes.Med. Biol. Eng. Comput.21750–755. 10.1007/BF02464038
14
HermensH. J.FreriksB.Disselhorst-KlugC.RauG. (2000). Development of recommendations for SEMG sensors and sensor placement procedures.J. Electromyogr. Kinesiol.10361–374. 10.1016/S1050-6411(00)00027-4
15
HuigenE.PeperA.GrimbergenC. A. (2002). Investigation into the origin of the noise of surface electrodes.Med. Biol. Eng. Comput.40332–338. 10.1007/BF02344216
16
JournéeH. L.PolakH. E.de KleuverM. (2004). Influence of electrode impedance on threshold voltage for transcranial electrical stimulation in motor evoked potential monitoring.Med. Biol. Eng. Comput.42557–561. 10.1007/BF02350999
17
JournéeS. L.DelesalleC. J. G.de BruijnC. M.BergmannW.JournéeH. L. (2014). Transcranial electrical stimulation (TES) as a possible novel alternative to transcranial magnetic stimulation (TMS) to assess the motor function of the spinal cord for clinical diagnosis in horses.Equine Vet. J.46:10. 10.1111/evj.12323_20
18
JournéeS. L.JournéeH. L.de BruijnC. M.DelesalleC. J. G. (2015). Design and optimization of a novel method for assessment of the motor function of the spinal cord by multipulse transcranial electrical stimulation in horses.J. Equine Vet. Sci.35793–800. 10.1016/J.JEVS.2015.07.014
19
JournéeS. L.JournéeH. L.de BruijnC. M.DelesalleC. J. G. (2018). Multipulse transcranial electrical stimulation (TES): normative data for motor evoked potentials in healthy horses.BMC Vet. Res.14:121. 10.1186/s12917-018-1447-7
20
KanekoK.FuchigamiY.MoritaH.OfujiA.KawaiS. (1997). Effect of coil position and stimulus intensity in transcranial magnetic stimulation on human brain.J. Neurol. Sci.147155–159.
21
KanekoK.KawaiS.FuchigamiY.ShiraishiG.ItoT. (1996). Effect of stimulus intensity and voluntary contraction on corticospinal potentials following transcranial magnetic stimulation.J. Neurol. Sci.139131–136. 10.1016/0022-510X(96)00050-0
22
LaferriereP.LemaireE. D.ChanA. D. C. (2011). Surface electromyographic signals using dry electrodes.IEEE Trans. Instrum. Meas.603259–3268. 10.1109/TIM.2011.2164279
23
MacDonaldD. B.SkinnerS.ShilsJ.YinglingC. (2013). Intraoperative motor evoked potential monitoring – a position statement by the American Society of Neurophysiological Monitoring.Clin. Neurophysiol.1242291–2316. 10.1016/j.clinph.2013.07.025
24
MayhewI. G.WashbourneJ. R. (1996). Magnetic motor evoked potentials in ponies.J. Vet. Intern. Med.10326–329.
25
MerlettiR.BotterA.BaroneU. (2016). “Detection and conditioning of surface EMG signals,” in Surface Electromyography, Physiology, Engineering, and Applications, edsMerlettiR.FarinaD. (New York, NY: John Wiley & Sons Inc.), 54–90.
26
NeulohG.SchrammJ. (2009). Monitoring of motor evoked potentials compared with somatosensory evoked potentials and microvascular Doppler ultrasonography in cerebral aneurysm surgery.J. Neurosurg.100389–399. 10.3171/jns.2004.100.3.0389
27
NolletH.DeprezP.van HamL.DewulfJ.DecleirA.VanderstraetenG. (2004). Transcranial magnetic stimulation: normal values of magnetic motor evoked potentials in 84 normal horses and influence of height, weight, age and sex.Equine Vet. J.3651–57.
28
NolletH.DeprezP.Van HamL.VerschootenF.VanderstraetenG. (2002). The use of magnetic motor evoked potentials in horses with cervical spinal cord disease.Equine Vet. J.34156–163. 10.2746/042516402776767204
29
NolletH.Van HamL.DeprezP.VanderstraetenG. (2003a). Transcranial magnetic stimulation: review of the technique, basic principles and applications.Vet. J.16628–42.
30
NolletH.Van HamL.DewulfJ.VanderstraetenG.DeprezP. (2003b). Standardization of transcranial magnetic stimulation in the horse.Vet. J.166244–250.
31
PiervirgiliG.PetraccaF.MerlettiR. (2014). A new method to assess skin treatments for lowering the impedance and noise of individual gelled Ag-AgCl electrodes.Physiol. Meas.352101–2118. 10.1088/0967-3334/35/10/2101
32
RijckaertJ.PardonB.HamL.Van LoonG.vanDeprezP. (2018). Magnetic motor evoked potential recording in horses using intramuscular needle electrodes and surface electrodes.J. Equine Vet. Sci.68101–107.
33
RoyS. H.De LucaG.ChengM. S.JohanssonA.GilmoreL. D.De LucaC. J. (2007). Electro-mechanical stability of surface EMG sensors.Med. Biol. Eng. Comput.45447–457. 10.1007/s11517-007-0168-z
34
SAS Institute Inc (1999). SAS Institute Inc., SAS/STAT® User’s Guide, 8th Edn. Cary, NC: SAS Institute Inc.
35
SkinnerS. A.TransfeldtE. E.SavikK. (2008). Surface electrodes are not sufficient to detect neurotonic discharges: observations in a porcine model and clinical review of deltoid electromyographic monitoring using multiple electrodes.J. Clin. Monit. Comput.22131–139. 10.1007/s10877-008-9114-3
36
SzelényiA.HattingenE.WeidauerS.SeifertV.ZiemannU. (2010). Intraoperative motor evoked potential alteration in intracranial tumor surgery and its relation to signal alteration in postoperative magnetic resonance imaging.Neurosurgery67302–313. 10.1227/01.NEU.0000371973.46234.46
37
SzelényiA.KothbauerK. F.DeletisV. (2007). Transcranial electric stimulation for intraoperative motor evoked potential monitoring: stimulation parameters and electrode montages.Clin. Neurophysiol.1181586–1595. 10.1016/j.clinph.2007.04.008
38
TajiB.ChanA. D. C.ShirmohammadiS. (2018). Effect of pressure on skin-electrode impedance in wearable biomedical measurement devices.IEEE Trans. Instrum. Meas.671900–1912. 10.1109/TIM.2018.2806950
39
TakedaM.YamaguchiS.MitsuharaT.AbikoM.KurisuK. (2018). Intraoperative neurophysiologic monitoring for degenerative cervical myelopathy.Neurosurg. Clin. N. Am29159–167. 10.1016/j.nec.2017.09.012
40
VerheyenT.DecloedtA.ClercqD.De DeprezP.SysS. U.LoonG.et al (2010). Electrocardiography in horses – part 1: how to make a good recording.Vlaams Diergeneeskd. Tijdschr.79331–336.
41
VerinE.StrausC.DemouleA.MialonP.DerenneJ.-P.SimilowskiT. (2002). Validation of improved recording site to measure phrenic conduction from surface electrodes in humans.J. Appl. Physiol.92967–974. 10.1152/japplphysiol.00652.2001
42
WijnbergI. D.FranssenH. (2016). The potential and limitations of quantitative electromyography in equine medicine.Vet. J.20923–31. 10.1016/j.tvjl.2015.07.024
43
WijnbergI. D.van der KolkJ. H.FranssenH.BreukinkH. J. (2003). Needle electromyography in the horse compared with its principles in man: a review.Equine Vet. J.359–17.
Summary
Keywords
transcranial stimulation, equine neurology, electromyography, subcutaneous electrodes, surface electrodes
Citation
Journée SL, Journée HL, Reed SM, Berends HI, de Bruijn CM and Delesalle CJG (2020) Extramuscular Recording of Spontaneous EMG Activity and Transcranial Electrical Elicited Motor Potentials in Horses: Characteristics of Different Subcutaneous and Surface Electrode Types and Practical Guidelines. Front. Neurosci. 14:652. doi: 10.3389/fnins.2020.00652
Received
28 February 2020
Accepted
26 May 2020
Published
17 July 2020
Volume
14 - 2020
Edited by
Giancarlo Ferrigno, Politecnico di Milano, Italy
Reviewed by
Shujie Wang, Peking Union Medical College Hospital (CAMS), China; Roberto Merletti, Politecnico di Torino, Italy
Updates

Check for updates
Copyright
© 2020 Journée, Journée, Reed, Berends, de Bruijn and Delesalle.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Henricus Louis Journée, hljournee@gmail.com
†These authors have contributed equally to the study and writing of the manuscript
This article was submitted to Neural Technology, a section of the journal Frontiers in Neuroscience
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.