REVIEW article

Front. Neurosci., 12 November 2021

Sec. Neural Technology

Volume 15 - 2021 | https://doi.org/10.3389/fnins.2021.728092

Defining a Path Toward the Use of Fast-Scan Cyclic Voltammetry in Human Studies

  • 1. Applied Computational Neurophysiology and Neuromodulation Laboratory, Department of Neurologic Surgery, Mayo Clinic, Rochester, MN, United States

  • 2. Department of Biomedical Engineering, University of Wisconsin-Madison, Madison, WI, United States

  • 3. Division of Engineering, Mayo Clinic, Rochester, MN, United States

  • 4. Mayo Clinic Graduate School of Biomedical Sciences, Mayo Clinic, Rochester, MN, United States

  • 5. Department of Physiology and Biomedical Engineering, Mayo Clinic, Rochester, MN, United States

Abstract

Fast Scan Cyclic Voltammetry (FSCV) has been used for decades as a neurochemical tool for in vivo detection of phasic changes in electroactive neurotransmitters in animal models. Recently, multiple research groups have initiated human neurochemical studies using FSCV or demonstrated interest in bringing FSCV into clinical use. However, there remain technical challenges that limit clinical implementation of FSCV by creating barriers to appropriate scientific rigor and patient safety. In order to progress with clinical FSCV, these limitations must be first addressed through (1) appropriate pre-clinical studies to ensure accurate measurement of neurotransmitters and (2) the application of a risk management framework to assess patient safety. The intent of this work is to bring awareness of the current issues associated with FSCV to the scientific, engineering, and clinical communities and encourage them to seek solutions or alternatives that ensure data accuracy, rigor and reproducibility, and patient safety.

Introduction

Fast scan cyclic voltammetry (FSCV) is an electrochemistry technique used for over 30 years to study rapid neurotransmission in the brain of anesthetized and awake and behaving animals (; ). FSCV detects electroactive neurotransmitters by varying the electric potential between a small working electrode (on the order of a few micrometers in diameter) and a larger reference electrode (on the order of millimeters in diameter). This process oxidizes and/or reduces neurotransmitters at specific potentials and results in electrical currents with amplitude proportional to the concentration of the neurotransmitter in the extracellular space (; ; Figure 1). Under the right conditions, this enables the estimation of changes in neurotransmitter levels with high spatial and temporal resolution, sensitivity, and chemical selectivity (; ).

FIGURE 1

However, in vivo neurotransmitter measurement via FSCV can be confounded by several factors including interferant molecules with similar oxidation/reduction potentials to the neurotransmitter of interest, electrode biofouling, shifts in pH and ionic concentrations, or increased oxygenated blood flow in the electrode microenvironment. Additionally, electrical and motion artifacts can disrupt the electrochemical interface creating artificial signals that appear similar to neurotransmitter release (; ). In small animal studies, a set of guidelines known as the “Five Golden Rules” is typically used to validate and increase reliability of in vivo neurotransmitter measurements (). These Golden Rules include (i) identification of neurotransmitter-specific electrochemical signatures, (ii) additional confirmation of the chemical identity of each recorded neurotransmitter (e.g., through microdialysis at the FSCV site), (iii) anatomical validation of the recording location, (iv) kinetic validation of spontaneous or evoked changes in neurotransmitter concentration, and (v) pharmacological validation of recorded neurotransmitters (; ). These guidelines are instrumental toward ensuring the validity of pre-clinical FSCV measurements.

In the last 10 years, there has been a trend toward the use of FSCV in patients undergoing neurosurgical procedures to study human neurophysiology (). Undoubtedly, advancing the knowledge of underlying mechanisms and biological processes associated with the complex chemistry of the human brain is critical to the development of new and improved therapeutic interventions. For example, the clinical use of FSCV may allow characterization of changes in neurochemical signaling evoked by deep brain stimulation (DBS) and could potentially advance treatment of neurologic diseases (). However, clinical questions that can be currently answered with FSCV are limited by technical barriers that must be addressed to ensure that clinical studies have appropriate scientific rigor and mitigate risks to patient safety. To address potential confounds during data acquisition in humans, extensive pre-clinical testing of new technologies designed to enable human use must be performed. Additionally, there is a critical need for appropriate pre-clinical studies mimicking the clinical environment. Risks should be weighed against actual benefits of the clinical work to ensure patient safety and high-quality data. The objective of this manuscript is to initiate a discussion between the scientific, engineering, and clinical communities regarding these issues and encourage collaborative solutions that are safe and suitable for neurochemical measurements in clinical studies.

Investigational Use of Fast Scan Cyclic Voltammetry

While a clinical study may not directly benefit the patients participating in the study, studies are likely to improve future patient care. Notably, treatment improvements and benefits to patients can only be accomplished by leveraging appropriate pre-clinical studies and scientific rigor.

Large Animal Studies

Anatomical and physiological similarities, as well as behavioral correlates to human conditions often make non-human primate and swine models a better alternative for translational research (; ). Unfortunately, only a small number of FSCV studies have investigated neurotransmitter release in vivo using large animal models (Table 1).

TABLE 1

StudySpecies (common name), nGoalFSCV parametersFSCV electrodesMajor findings
Macaca nemestrina (Pigtail macaque) 2 females, 2 males1 – Determine if D-amphetamine increased current oxidation in the caudate of macaques, as previously reported in rodents.
2 – Determine if altered behavioral states induced by D-amphetamine affected oxidation peaks.
Triangular waveform from –0.2 to +0.6 V at 10 mV/sWorking electrode: Teflon coated 250 μm diameter SS wire ∼ 1 mm exposed tip Reference electrode: Ag/AgCl wiresD-Amphetamine treatment enhances dopaminergic signaling during presentation of both pleasant and unpleasant stimuli.
Macaca nemestrina (Pigtail macaque) 2 females, 1 maleDemonstrate the feasibility of using acetaminophen as in vivo internal standard for electrode calibration.Triangular waveform from –0.2 to +0.6 V at 10 mV/sWorking electrode: Teflon coated 250 μm diameter SS wire Reference electrode: Ag/AgCl wireAcetaminophen peak signal was successfully detected by FSCV in the non-human primate brain.
Callithrix jacchus (marmoset) 8 (unspecified number of males and females)Evaluate electrically-evoked dopaminergic efflux in the striatum of normal and MPTP-treated marmosets.Triphasic triangular waveform from –1.0 to +1.4 V and resting potential at 0 V at 480 V/s at 2 Hz (15 ms/scan)Working electrode: Glass insulated 7 μm diameter CFM ∼ 20–50 μm exposed tip Reference electrode: Ag/AgCl wireDopamine signaling in the striatum remains responsive to MFB electrical stimulation after dopaminergic lesion induced by MPTP.
Shon et al., 2010Sus scrofa domesticus (Swine) 4 males1 – Investigate if electrical stimulation of the STN evokes striatal dopamine release in a large animal model.
2 – Demonstrate feasibility of performing FSCV recordings in an environment similar to human OR.
Triangular waveform from –0.4 to +1.5 V at 400 V/s at 10 HzWorking electrode: glass insulated CFM (characteristics unspecified) Reference electrode: Ag/AgCl wireSTN electrical stimulation evoked intensity and frequency dependent striatal dopamine release.
Sus scrofa domesticus (Swine) Unknown number of animals – malesUse a wireless instantaneous neurotransmitter concentration measurement system to monitor electrochemical signaling in the brain.Triangular waveform from –0.4 to +1.5 V at 400 V/s at 10 HzWorking electrode: CFM (characteristics unspecified) Reference electrode: Ag/AgCl wireDopamine signaling responded in sigmoidal-like fashion to pulse intensity and pulse-width STN electrical stimulation.
Macaca mulatta (Rhesus macaque) 3 (unspecified number of males and females)Determine if changes in pH and oxygen are associated with reward and/or reward prediction.Dopamine – Triangular waveform from –0.4 or –0.6 V to +1.0 or +1.4 V at 400 V/s at 10 Hz.
Oxygen – from 0.0 to +0.8 V, a reversal to –1.4 V, and then returned to 0.0 V, at 10 Hz.
Working electrode: glass insulated 12 μm diameter CFM with 250 μm exposed tip and coated with Nafion and 4-sulfobenzene; 33 μm coated with Nafion.
Reference electrode: Ag/AgCl wire
Oxygen and pH changes were associated with the reward and cues that predicted reward. Dopamine responses evoked by reward and cues were overshadowed by pH changes detected by FSCV.
Schluter et al., 2014Macaca mulatta (Rhesus macaque) 2 malesDemonstrate the feasibility of using FSCV to measure real-time changes of dopamine levels in the striatum of macaques.Triangular waveform from –0.4 to +1.3 V at 400 V/s at 10 HzWorking electrode: fused silica capillary CFM – 125–150 μm exposed tip and 7 μm diameter. Reference electrode: Ag/AgCl wireDemonstrated striatal dopamine responses evoked by VTA/SNc and striatum electrical stimulation, and by unexpected rewards in awake monkeys.
Van Gompel et al., 2014Sus scrofa domesticus (Swine) 3 malesDetermine if extracellular adenosine concentration increases during seizure termination (in swine and humans).*Triangular waveform from –0.4 to +1.5 V at 900 V/s at 10 HzWorking electrode: CFM – 100 μm exposed tip and 7 μm diameter Reference electrode: Ag/AgCl wireIncreased adenosine levels were observed just prior to seizure termination. FSCV recordings were also performed in human patients.*
Yoshimi et al., 2015Macaca fuscata (Japanese macaques) 3 femalesMeasure changes in dopamine levels associated with reward response.Triangular waveform from –0.4 to +1.3 V at 400 V/s at 10 HzWorking electrode: fused silica CFM and glass capillary CFM – 7 μm diameter and 300 μm exposed tip. Reference electrode: Ag/AgCl wireDopamine release induced by electrical stimulation and reward signals was detected in the macaque striatum by FSCV on carbon fibers.
Macaca mulatta (Rhesus macaque) 3 malesCharacterize striatal dopamine release evoked by STN DBS as a function of stimulating and recording electrode location.Triangular waveform from –0.4 to +1.5 V at 400 V/s at 10 HzWorking electrode: CFM – 7 μm diameter and 100 μm exposed tip. Reference electrode: Ag/AgCl wireEvoked dopamine responses were higher at the stimulation of the dorsolateral posterior border of the STN.
Sus scrofa domesticus (Swine) 17 (unspecified number of males and females)Determine the functional connectivity between the medial limbic and corticolimbic circuits following fornix DBS via evoked-dopamine release in the NAc.Triangular waveform from –0.4 to +1.5 V at 400 V/s at 10 HzWorking electrode: CFM – 7 μm diameter and 100 μm exposed tip. Reference electrode: Ag/AgCl wireElectrical stimulation of the fornix induced dopamine release in the NAc and increased BOLD activity in structures along the medial-corticolimbic circuitry.
Rattus norvegicus (Sprague-Dawley rats) 40 males Sus scrofa domesticus (Swine) 12 males Macaca mulatta (Rhesus macaque) 3 malesDemonstrate proof-of-principle for wireless measurement, characterization, and control of neurotransmitter release.For dopamine and adenosine – triangular waveform from –0.4 to +1.5 V; for serotonin – N-shaped waveform: –0.4, +1.0, –0.4, +1.4 VWorking electrode: CFM – 7 μm diameter and 100 μm exposed tip. Reference electrode: Ag/AgCl wireDemonstrated successful in vivo, wireless, single or multi-channel detection of dopamine, adenosine, and serotonin, with integrated sensing and stimulation feedback capabilities.
Macaca fuscata (Japanese macaques) 3 femalesEvaluate the effects of clinically relevant STN and GPi DBS in the modulation of the activity of tonically active striatal cholinergic interneurons.Triangular waveform from –0.4 to +1.5 V at 408.6 V/s at 10 kHz for 60 s, and resting potential at 0 V.Working electrode: SS insulated CFM – 250–300 μm exposed tip. Reference electrode: Ag/AgCl wireSTN DBS, but not GPi DBS, induced striatal dopamine release that was correlated to increased activity of tonically active cholinergic striatum interneurons.
Schwerdt et al., 2017bMacaca mulatta (Rhesus macaque) 3 femalesDemonstrate the feasibility of using an integrated neurochemical modular platform for monitoring dopamine release from sensors chronically implanted in the brain of non-human primates during behavior and stimulation-evoked dopamine release.Triangular waveform from –0.4 to +1.3 V at 400 V/s at 10 HzWorking electrode: Array of chronic and acute fused silica insulated CFMs – 7 μm diameter and 150–300 μm exposed tip Reference electrode: Ag/AgCl wire or SS electrodeModular platform allowed measurements of dopamine release from multiple sites in the striatum while electrically stimulating the SNc/VTA for up to 170 days.
Settell et al., 2017Sus scrofa domesticus (Swine) 4 (unspecified number of males and females)Determine the neuromodulatory effects of VTA DBS on dopamine release in the NAc.Triangular waveform from –0.4 to +1.5 V; or N-shaped waveform: –0.4, +1.0, –0.4, +1.4 VWorking electrode: CFM – 7 μm diameter and 100 μm exposed tip. Reference electrode: not specifiedVTA DBS resulted in increased dopamine release in the NAc and increased BOLD activity in the striatum, cortical and limbic structures.
Trevathan et al., 2017Rattus norvegicus (Sprague-Dawley rats) 12 females Sus scrofa domesticus (Swine) 4 males Macaca mulatta (Rhesus macaque) 1 maleCharacterize subject-specific kinetics of stimulation-evoked dopamine release using computational modeling.Triangular waveform from –0.4 to +1.5 V at 400 V/s at 10 HzWorking electrode: CFM – 7 μm diameter and 100 μm exposed tip. Reference electrode: not specifiedDopamine dynamics in response to electrical stimulation was modeled and characterized based on non-linear increased responses of dopamine to increasing SNc/VTA stimulus intensity.

Fast scan cyclic voltammetry studies in large animals.

SS, stainless steel; FSCV, fast scan cyclic voltammetry; MPTP, 1-methyl-4-phenyl-1,2,3,6-tetrahydropyridine CFM, carbon fiber microelectrode; MFB, medial forebrain bundle; STN, subthalamic nucleus; OR, operating room; VTA, ventral tegmental area; SNc, substatia nigra pars compacta; DBS, deep brain stimulation; NAc, nucleus accumbens; BOLD, blood-oxygen-level-dependent imaging; and GPi, globus pallidus internum. *See Table 3.

Dopamine release in large animals was first measured via FSCV in 1984 in a non-human primate model (). In the study, the investigators detected increased current peaks in the oxidation range for dopamine in the caudate in response to pleasant stimuli and amphetamine administration. However, these measurements were not properly validated accordingly to the Five Golden Rules, i.e., use of rudimentary electrochemical parameters and working electrode material, which reduces reliability on the data collected.

It was not until 10 years later that FSCV was used again in non-human primates to demonstrate reduced dopaminergic release in response to electrical stimulation of the medial forebrain bundle in an anesthetized 1-methyl-4-phenyl-1,2,3,6-tetrahydropyridine marmoset model of Parkinson’s Disease (). In this study, investigators validated dopamine measurements by confirming dopamine-specific electrochemical signature, and by demonstrating anatomically, electrically, and pharmacologically evoked changes in the kinetics of dopamine concentration.

Since then, other studies using FSCV in swine and non-human primate models have utilized FSCV recording electrodes (including both working and reference electrodes), recording systems, and analysis techniques that had been well validated in small animal models (; Stamford et al., 1984; ; Suaud-Chagny et al., 1992), ensuring rigor and data reproducibility. Consistently, they have reported increased dopamine release in the caudate-putamen as a function of reward (Yoshimi et al., 2015), and electrical stimulation of the subthalamic nucleus (), fornix, and ventral tegmental area (). In addition, FSCV has been used to detect cortical adenosine during seizure termination in swine (Van Gompel et al., 2014), and to serve as the basis of a neurochemical-based closed-loop DBS system in rodent, swine, and non-human primate (). More recently, studies in non-human primates proposed new strategies to overcome challenges regarding chronic FSCV recordings in large animals, such as target optimization and spatial resolution (Schluter et al., 2014; Schwerdt et al., 2017b). However, they could not resolve other limitations such as tissue damage, glial encapsulation, electrode biofouling, and material deterioration that caused continuous signal loss ().

Overall, these pre-clinical studies show the feasibility of recording of neurotransmitters via FSCV in acute and long-term animal models. However, they do not directly address the significant barriers to clinical use of FSCV, especially in the operating room setting.

Scientific Rigor

It is not possible to apply the Five Golden Rules for neurotransmitter measurement in the human operating room. However, many of the potential confounds for FSCV measurement in clinical studies can be anticipated. Data analysis, including calibration and signal extraction algorithms designed to distinguish neurochemical signals from interferents and noise, depend mainly on the electrochemical properties of the FSCV recording system. For example, the material composition, configuration, and surface properties of FSCV electrodes define the electrochemical properties of the recording system and can affect the recorded signals (; ; ; ; ). Additionally, electrode materials and recording system commonly used in small animal models must be modified to ensure biocompatibility and the ability to reach deep brain targets. Furthermore, confirmation of measured signals accuracy via pharmacologic validation or alternate neurochemical technique in humans is largely not possible in the operating room due to increased hemorrhage risk with additional alternate neurochemical recordings (; ), and risk of pharmacological manipulations interfering with patient’s wellbeing during the procedure. Moreover, the time required for recording system stabilization and signal detection is large, but the total allowed operating room time for safe experimental procedures is limited. For example, in acute animal experiments using a single Carbon fiber microelectrode (CFM), identifying an optimal area for recording evoked neurochemical release within a given target region can take several minutes to a few hours. This process requires slowly moving the electrode to different locations within the target and allowing enough time for the signal drift to settle after each movement. In contrast, Institutional Review Boards (IRBs) typically require that the research use of FSCV adds no longer than 15–60 min to the overall procedure, which varies across institutions with an average duration of 4 h. The compressed timeframe of the operating room environment combined with patient safety concerns require the use of one single track, with very little or no time available for electrode repositioning for recording optimization. The operating room environment is also full of sources of noise and interference from electronic equipment that can affect FSCV recordings in unexpected ways. Although not exhaustive, the next sections highlight the need for appropriate pre-clinical studies addressing changes to the recording methods, experimental paradigm, and data analysis techniques beyond what is described in the FSCV literature before human neurophysiology studies can be performed.

Novel Working Electrodes Designs

The working electrode is typically composed of carbon-based materials (Table 2) to improve biocompatibility and chemical inertness, as well as to reduce background currents that interfere with the detection of neurochemical signatures associated with specific neurotransmitters (; ; ; ). CFMs have become the standard working electrode for pre-clinical FSCV recordings. The popularity of CFMs can be attributed mainly to their small size, which is critical for minimizing damage to neural tissue and reducing the impact on neurochemical transmission near the electrode (see Figure 2 for CFM and DBS electrode leads size comparison). Similarly, CFM have shown good performance in maintaining normal neurotransmitter kinetics at the measurement site and high adsorption of neurotransmitters of interest (; Wang and Michael, 2012; ). However, despite their desirable neurochemical detection properties, CFMs are not optimal for clinical applications due to their high manufacturing variability () and fragility, which increase the risk of breakage during implantation into brain tissue (Schluter et al., 2014). Similarly, surface modified CFM made by coating with polymer films such as Nafion and PEDOT provides enhanced sensitivity, selectivity, and kinetic properties for dopamine detection. However, electrode reproducibility is still poor due to the lack of coating uniformity (; ; Vreeland et al., 2015; ; ). In addition, PEDOT is not currently approved for clinical use due to the large range of PEDOT variations determined by multiple factors such as the fabrication technique, the dopant, and the functionalization of the polymer ().

TABLE 2

Electrode materialType of electrode modificationInsulationAdvantages*Disadvantages*References
Carbon fiberN/ABorosilicate glassEase of fabrication, low cost, biocompatible, recommended for acute recordings.Fragile, high risk of breakage, poor fabrication uniformity, susceptible to fouling.; ; ;
Carbon fiberN/AFused-silica capillaries sealed with epoxyGood insulating properties, increased flexibility, low cost, lower risk of breakage, small diameter, biocompatible, recommended for chronic recordings.Susceptible to fouling, poor fabrication uniformity, torsion applied to the carbon fiber not well translated to the fused-silica insulation.; ;
Carbon nanotubeSurface-modified CFMBorosilicate glass or Fused-silica capillaries sealed with epoxyHigh mechanical strength, improved electron-transfer kinetics, high sensitivity to adsorbed dopaminePoor fabrication uniformity and reproducibility.; ; Yang et al., 2017
Gold-nanoparticlesSurface-modified CFMBorosilicate glass or Fused-silica capillaries sealed with epoxyHigh electroactive surface area, fast electron-transfer kinetics, highly sensitiveIncreased foulingZhao et al., 2012;
Carbon fiberNafion-coated CFMBorosilicate glass or Fused-silica capillaries sealed with epoxyImproved selectivity and sensitivity for dopamine, fast response timePoor uniformity and reproducibility in creating Nafion film around the electrode, fouling susceptibilibility is similar to bare CFM.; ;
Carbon fiberPEDOT:Nafion-coated CFMBorosilicate glass or Fused-silica capillaries sealed with epoxyImproved selectivity and sensitivity for dopamine, fast response time, low foulingPEDOT is not approved for human useVreeland et al., 2015; ;
DiamondBoron doped conductive diamondBorosilicate glass sealed with resin or Stainless tube sealed with resin or ParyleneWide potential window, low and stable background current, low fouling, improved longevity and resistanceLow sensitivity, slow electron-transfer kinetics, costly fabrication, poor doping may lead to impurities on the surfaceYoshimi et al., 2011; ; ; ;
PlatinumMetalUnknownLess susceptible to breakage, versatile for surface modification and assembly into arrays for multi-neurotransmitter monitoringSusceptible to corrosion, passivation, increased fouling, low sensitivity.;
GoldMetalUnknownLess susceptible to breakage, versatile for surface modification and assembly into arrays or miniature gold electrodes, good sensitivity to catecholamines,Susceptible to corrosion, passivation, increased fouling, low sensitivityZachek et al., 2008; ;

Characteristics of commonly used fast scan cyclic voltammetry working electrodes in in vivo studies.

CFM, Carbon Fiber Microelectrode; PEDOT, poly(3,4-ethylenedioxythiophene).

*Relative to borosilicate glass CFM.

FIGURE 2

As an alternative to CFM and surface modified CFM, several groups have attempted to develop Boron-doped diamond (BDD) electrodes, which offer increased strength and longevity compared to CFM (; Suzuki et al., 2007; Yoshimi et al., 2015; ). However, BDD has distinct electrochemical properties and significant characterization studies must be performed to better understand and optimize BDD for neurochemical sensing in the clinical setting. To date, studies have shown that BDD electrodes are not suitable for chronic in vivo neurochemical detections due to poor sensitivity and neurochemical adsorption rate, increased fouling, and distorted signals (Yoshimi et al., 2011; ; ).

Metal electrodes such as Gold and Platinum electrodes provide increased mechanical strength being more resistant to breakage during implantation. Additionally, these materials allow for novel electrode arrays and the capability of multiple neurotransmitter simultaneous detection. However, metal electrodes are hardly used for clinical applications because of the risk of corrosion that can lead to tissue damage, passivation and increased biofouling that interfere with neurotransmitter detection (Zachek et al., 2008; ; ).

Although novel electrode materials or configurations, such as Platinum, Gold, and Nafion-coated CFM described above, provide numerous benefits, the representation of signal and artifacts in vitro and in vivo can vary significantly to those measured with CFMs (). For this reason, it is important to fully validate novel working electrode designs via large animal experiments with similar signal levels and experimental paradigms to anticipated human studies.

Reference Electrode Biocompatibility and Stability

Reference electrodes are pivotal to ensure stable, consistent, and neurotransmitter-specific measurements. Reference electrodes should not be polarized (i.e., pulled away from their open circuit potential) during FSCV to avoid sensor drift, low signal-to-noise ratio, and distorted neurochemical signals (). Thus, investigators need to ensure that polarization is avoided by increasing the surface area of the reference electrodes relative to the active CFM, as well as by using materials that have a stable in vivo reference potential and low polarizability properties. Reference electrodes for in vivo neurotransmitter measurements in animal models are typically made of Ag/AgCl, which are highly cytotoxic making them unsafe for long-term (in the order of hours/days) human use (; ; Seaton et al., 2020). To avoid the safety concerns associated with Ag/AgCl electrodes, stainless steel (SS) reference electrodes have been used in humans (; ). Unfortunately, SS is highly polarizable and has an inhomogeneous grain structure that decreases the stability of its open circuit potential by changing the protein adsorption to the electrode surface, which can shift its open circuit potential (). The use of SS reference electrodes for clinical FSCV (see Table 3) is an unresolved confound that can significantly affect measurement sensitivity and specificity. Thus, there is a critical need for the pre-clinical investigation and validation of novel reference electrodes that are not only biocompatible, but also that provide stable reference potentials prior to human use.

TABLE 3

StudyGoalFSCV parametersFSCV electrodesMajor findings
Demonstrate feasibility of sub-second dopamine detection in humans during brain surgery.Triangular waveform from –0.425 V to +1.380 at 400 V/s at 10 Hz.Working electrode: Glass insulated 7 μm CFM
Reference electrode: Stainless steel
First demonstration of real-time human dopamine release measured during a behavioral investment task.
Quantify adenosine concentrations in human subjects with essential tremor during VIM DBS.Triangular waveform from –0.4 V to +1.5 at 400 V/s at 10 Hz.Working electrode: Glass and polyamide insulated 7 μm CFM with 50 μm exposed tip or silicone insulated 30 μm CFM
Reference electrode: Stainless steel
Demonstrated changes in extracellular concentrations of adenosine in the VIM of patients undergoing VIM DBS.
Van Gompel et al., 2014Determine changes in extracellular adenosine concentration during epileptic activity (in swine and humans).*Triangular waveform from –0.4 V to +1.5 at 400 V/s at 10 Hz.Working electrode: Glass insulated 7 μm CFM
Reference electrode: Stainless steel, 24 gage
Extracellular adenosine concentration increases prior to seizure termination.
Demonstrate sensitivity of diamond-based electrode to detect and quantify neurotransmitter release in human patients undergoing DBS.Triangular waveform from –0.4 V to +1.5 at 400 V/s at 10 Hz.Working electrode: Parylene-C coated boron-doped diamond, conical shape, 50 μm at the base and 100 μm long.
Reference electrode: Stainless steel
Dopamine and adenosine release were detected using boron-doped diamond electrode in human patients undergoing DBS for tremor.
Determine whether subsecond dopamine fluctuations in the human striatum encode reward prediction errors during a sequential choice task.Triangular waveform from –0.6 to +1.4 V at 400 V/s at 10 Hz.Working electrode: 7 μm CFM insulated with a polyimide-coated fused-silica capillary
Reference electrode: Stainless steel
Demonstrated striatal dopamine changes encoding experience-dependent reward prediction error and counterfactual prediction error.
Compare BOLD activity and dopamine responses during a sequential choice task in humans.Triangular waveform from –0.6 to +1.4 V at 400 V/s at 10 Hz.Working electrode: 7 μm CFM insulated with a polyimide-coated fused-silica capillary
Reference electrode: Stainless steel
Demonstrated non-linear relationship between BOLD activity and dopamine release of subjects performing an investment task.
Demonstrate changes in the serotonergic signaling during a decision making task.Triangular waveform from –0.6 to +1.4 V at 400 V/s at 10 Hz.Working electrode: 7 μm CFM insulated with a polyimide-coated fused-silica capillary
Reference electrode: Stainless steel
Striatal serotonin correlates with decisions in a sequential investment game and may encode a strategy that modulates choice selection to mitigate risk.
Demonstrate a new approach for simultaneous
multi-neurotransmitter detection and quantification.
Triangular waveform from –0.6 to +1.4 V at 400 V/s at 10 Hz.Working electrode: 7 μm CFM insulated with a polyimide-coated fused-silica capillary
Reference electrode: Stainless steel
The elastic net approach extracted a concentration-prediction model for multiple analytes that included dopamine, serotonin, and norepinephrine.

FSCV studies in humans.

FSCV, fast scan cyclic voltammetry; CFM, carbon fiber microelectrode; VIM, ventral intermediate nucleus of thalamus; DBS, deep brain stimulation; and BOLD, blood-oxygen-level-dependent imaging. *See Table 1.

Tissue Damage and Signal Integrity

Implantation of both working and reference electrodes disrupts the blood-brain barrier, triggering a cascade of complex molecular and cellular responses such as activation of glial cells, loss of perfusion, secondary metabolic injury, neuronal degeneration, and introduction of mechanical strain, which can affect neurochemical sensing (Woolley et al., 2013; ). For example, studies have shown that brain hemorrhage associated with electrode implantation in the human brain can result in 1–2% of symptomatic hemorrhage (i.e., speech arrest, hemiplegia, agitation, and confusion) and 0.5–0.9% of permanent neurological deficit (; ). Therefore, any clinical FSCV work should minimize the risks associated with mechanical tissue damage through rigorous pre-clinical studies aimed at understanding and reducing mechanical damage to neural tissue ().

Reduction/oxidation (redox) reactions can also disrupt neurochemical diffusion to the region of the sensing electrode, changing neurochemical concentrations near the electrode and affecting detection of the neurochemical species of interest (Wang and Michael, 2012; ).

Carbon fiber microelectrodes used in pre-clinical studies have traditionally been fabricated from silica glass capillaries or tubing with a small implantation profile (; ). However, clinical FSCV studies have mimicked setups used clinically to implant electrophysiological sensing electrodes and relied on stainless steel cannulas to implant either CFM or diamond electrodes, e.g., FHC#66-IT(AO6) 165 mm SS insertion tube for placing depth electrodes along a trajectory. The effect of the relatively large tissue insult resulting from cannula insertion on measured FSCV signals is not well understood and should be explored in large animal models.

Spatial Resolution and Heterogeneity of Neurotransmitter Release

The high spatial resolution of FSCV, determined by the size of the working electrode relative to the volume of neurochemical release, is one of the driving motivators for use of FSCV to detect neurochemical signaling. Working electrodes are typically made from 7 to 30 μm diameter carbon fibers cut to 50–500 μm lengths (see Figure 2). It is the small size of the CFM that allows for rapid scan rates (>100 V/s) which enables neurotransmitter measurements on a subsecond timescale without producing large charging currents potentially harmful to the brain tissue (Zachek et al., 2009). The spatial resolution of the CFM is on the order of tens to hundreds of microns (), much higher than microdialysis () and functional imaging (). FSCV high spatial resolution allows for neurotransmitter measurement in discrete brain regions, which could potentially serve as guidance for improved targeting of DBS electrode implantation.

It must be considered that studies have shown inconsistent dopamine kinetics across different microdomains in small and large animal models (Wightman et al., 2007; ; Taylor et al., 2015), which can lead to data misinterpretation and demonstrate that FSCV recordings from a single CFM are not adequate for testing specific hypotheses of systemic neurotransmission. Thus, measurements obtained at these high resolutions may not accurately represent activity within the target brain nuclei due to sparse synaptic organization, neuronal heterogeneity, and differences in neurochemical release kinetics throughout the structures of interest. These inconstancies in neurotransmitter release across a target of interest, coupled with limited recording time in the operating room, can make finding a suitable recording site difficult for clinical studies. Sparsity and heterogeneity can be addressed by developing FSCV electrode arrays capable of simultaneously sampling multiple brain regions (Zachek et al., 2010; ; Schwerdt et al., 2017a). This, however, comes at the expense of increased risk of hemorrhage and inflammation, which must be weighed carefully against the expected benefit of the study.

Effects of Sterilization of Fast Scan Cyclic Voltammetry Electrodes on Recorded Signals

Sterilization of implantable devices is paramount to reduce the risk of infection. However, this creates a challenge for FSCV measurements as sterilization procedures can alter the geometrical and chemical surface of the CFM, compromise electrode insulation, affect electrical connections, and ultimately, hinder biocompatibility (; ; Zheng et al., 2011; ). For example, CFM working electrodes and Ag/AgCl reference electrodes are composed of heat-sensitive materials, which makes steam and dry heat sterilization inappropriate because these techniques can cause polymer degradation and changes in physical or mechanical properties (; ). Thus, sterilization procedures should not only be evaluated for sterility prior to clinical use, but also for the effect of surface modification on signal specificity and accuracy.

Effects of the Operating Room Environment on Recorded Signals

Continuous electrochemical recording is associated with baseline instability and sensor drift as a function of time, ultimately reducing measurement accuracy and reproducibility (; ). This drift can be caused by a myriad of external factors such as temperature fluctuations, changes in the chemical environment, or chemical reactions at the sensing surface (). The chemical environment can be affected by biofouling caused by extracellular proteins, redox reaction debris, cellular encapsulation around the surface of the electrode, and many other factors (; ; ; Seaton et al., 2020). To minimize these effects, several strategies such as sensor cleaning, recalibration, and even replacement have been used (; ; ). However, these are not feasible in an intraoperative setting, where available time is limited. Additionally, the relatively long period of stabilization following implantation of an FSCV electrode (; ; ) limits the time available for data collection in the operating room. Furthermore, the operating room environment has numerous sources of noise. These are known to affect electrophysiological recordings, but their effects have not been characterized for electrochemical measurements using FSCV.

Challenges for Data Analysis

One of the most challenging aspects of FSCV is the complex and variable nature of the data (; ). Signals obtained in vivo often have contributions from multiple analytes that require resolution prior to positive identification and quantification. Consistency and accuracy can be improved with automated multivariate statistical data analyses, such as principal component regression, partial least squares regression, and statistical models. Typically, information across the scan-potential window can be used to separate overlapping signals by using training sets (i.e., signals obtained from electrodes, recording sessions, and/or subjects other than those used for experimental data collection) as calibration models (, ; ; ; ; ). The effectiveness of these analysis techniques depends on the existence of well-characterized relationships between the potential at the working electrode and the measured redox currents for the neurochemical species of interest. For clinical FSCV, analysis algorithms should be validated in vivo using the exact recording setup that will be used clinically, at signal levels comparable to those normally observed in pre-clinical settings.

Recent studies adopted a data analysis algorithm built from a novel statistical model trained on in vitro data recorded against Ag/AgCl reference electrodes. However, the algorithm was applied to data recorded using SS reference electrodes in patients during a task performed in the operating room without previous pre-clinical validation (; ). Pre-clinical validation using the Five Golden Rules afore mentioned should be performed prior to clinical use of any analysis algorithms. Thus, algorithms such as the one described above remains largely unvalidated until it is tested both in vitro and in vivo in a pre-clinical model using the same type of working and reference electrodes as those that will be used clinically.

Other Considerations and Technical Challenges to Fast Scan Cyclic Voltammetry in Humans

Patient Safety

Patient safety is paramount whenever clinical studies are performed. As with all clinical studies, the onus of determining safety risk lies with the investigator and local IRB, regardless of whether the study involves the use of a device subjected to an Investigational Device Exemption or not (). While not all risk can be eliminated, it is the responsibility of the investigator and local IRB to create a risk management process in order to identify, isolate, and mitigate as much risk as possible (). In this way, not only is the well-being of the patient protected, but also is the integrity, rigor, and reproducibility of the collected data and resulting conclusions.

Material Biocompatibility

Any material or device that comes into contact with tissue, especially neural tissue, in a clinical study should be evaluated for biocompatibility. Although extensive discussion of biocompatibility testing is outside of the scope of this work, guidance for this biocompatibility testing is provided by the ISO 10993 standard (). It is important to note that even if the material itself is biocompatible, there is a risk of harmful substances to be present on the material as byproducts of manufacturing or other processing steps. In commercially marketed medical devices, these risks are mitigated by manufacturing under Good Manufacturing Practice protocols (), which help control any deviation between manufacturing batches. Currently, no medical device suppliers provide electrodes for FSCV, so it falls on the investigator to ensure consistency between materials tested for biocompatibility and those that are used for clinical FSCV work.

Electrochemical Reactions Due to Waveform Selection

Different electrode materials exhibit unique electrochemical properties that may lead to changes in the waveforms that are used for neurochemical detection. These waveforms, in turn, affect the various safety considerations listed above. Thus, testing of the electrical FSCV potential waveforms is required to determine the range of voltages that can be safely applied to the tissue without inducing excitotoxicity, electrolysis, and reduced electrode longevity. For example, harmful reactive oxygen species are generated when oxygen is reduced (negative potentials) or when water is oxidized (positive potentials; ). However, the potential at which these redox reactions occur varies depending on the water window for the specific electrode material. For voltage-controlled applications such as FSCV, the water window informs whether a particular waveform will result in electrolysis, which is damaging to tissue (; ). Therefore, all FSCV waveforms used should be evaluated for tissue damage and safety for each electrode material and configuration should be stablished.

It is also vital to ensure that the waveform does not damage the electrode itself, as material dissolution is detrimental to the longevity of the electrode (Figure 3) and the released particulate could be harmful to tissue. To prevent electrode damage, the largest potentials of a waveform should be applied to the electrode in solution in a closed-system benchtop setup over an extended period of time and optionally, at a higher frequency than planned. The beaker solution should be sampled at regular intervals and can be tested with mass spectrometry to determine whether the electrode material is dissolved into solution (). Subsequent biological tests should be conducted to determine whether the intended waveforms would damage cells due to excitotoxicity by conducting safety tests of electrical stimulation in neural tissue (, , ; Shannon, 1992). Finally, pre-clinical studies with thorough histological evaluation of tissue adjacent to and in the vicinity of the implanted electrode should be conducted prior to human studies (, ).

FIGURE 3

.

Sterilization Validation

As would be expected for any invasive medical device, sterilization validation is essential to patient safety (). Balancing this need with that of ensuring electrode integrity may require exploration of novel sterilization methods. As previously discussed, the heat-sensitive nature of some of the components commonly used for FSCV electrodes, steam or dry heat sterilization methods are not viable (; ). However, sterilization methods that do not rely on heat pose other challenges to patient safety. For example, γ radiation generates free radicals that can cause tissue damage (; ; ). Similarly, ethylene oxide and its derivatives have been reported to accumulate in some materials, potentially increasing teratogenicity risk (; ). As previously mentioned, the investigator should characterize and validate each sterilization technique for each FSCV sensor type, composition, and geometry to ensure patient safety, in addition to signal accuracy.

Discussion

Many of the safety and efficacy concerns associated with FSCV recordings in humans are due to interrelated technical challenges, which makes these challenges difficult to address. However, once the scientific and clinical benefits of the study are clearly defined and the risks to patient safety minimized, residual risks can be weighed against anticipated benefits. This risk-benefit analysis is crucial to the success of the clinical work.

Investigators should perform a compelling risk/benefit analysis to justify the clinical use of FSCV, ensure scientific rigor, and articulate the scientific and clinical benefit of FSCV studies. Addressing and mitigating risk to patient safety will help ensure patients are not put at undue risk. Of course, not all risk can be eliminated, and it is this remaining risk that must be weighed against the proposed clinical benefit. Furthermore, it is essential that both risks and benefits are appropriately and timely communicated to patients through consent forms so patients can make an informed decision about their participation.

Risk-benefit analyses are complex, so the authors propose that the FSCV community develop the necessary guidelines through partnership with the FDA to engender high-quality scientific discoveries that FSCV can offer.

Clinical studies need not demonstrate that FSCV provides direct patient benefit. However, evidence from pre-clinical studies should be used to guide approval for use in human subjects. Application of FSCV in clinical studies should be carefully evaluated to ensure that only hypothesis-driven studies that cannot be appropriately and methodically performed in animal models are conducted in human subjects. Thus, translational FSCV research should demonstrate scientific rationale, feasibility, efficacy, and safety prior to attempting clinical application.

While progress has been made in the clinical use of FSCV (Table 3), it must be stressed that electrode safety and data robustness have yet to be characterized. To date, clinical FSCV studies have employed working and reference electrodes and signal processing techniques that have not been validated by translational research. The Five Golden Rules are not directly applicable in the clinical setting. Thus, the operating room environment necessitates the reconsideration of human FSCV methodologies. Specifically, the following processes need to be performed: acute and chronic target assessment and confirmation, standardization of electrode manufacturing processes, signal verification in human recordings, and characterization of sources of electrical noise in the intra-operative environment. Addressing these issues will facilitate application of FSCV in the clinical setting, particularly if FDA-approved working and reference electrodes become commercially available.

The use of FSCV in humans must be understood as a clinical intervention, and as such, should follow the same protocols and guidelines required by the FDA to establish device safety (Figure 4). Additionally, there are many gaps in knowledge that complicate the risk/benefit analysis for the clinical use of FSCV during existing surgical procedures. Until these gaps are addressed, the likelihood of obtaining useful scientifically rigorous data is low due to all the unknowns and the quality of the recorded signal in the operating room environment. Thus, clinical FSCV should only be performed when there is a clear hypothesis that can further neuroscientific understanding in the pursuit of improved therapeutic interventions, and only after all risk factors discussed in this manuscript have been mitigated.

FIGURE 4

Publisher’s Note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Statements

Author contributions

SLB, JT, SH, and JLL contributed to the conceptualization and design of this work. SLB, JT, and SH wrote the first draft. AA wrote sections of this work. SLB was responsible for the project administration and supervision. JLL supervised the project. All authors contributed to manuscript revision, read, and approved the submitted version review and editing.

Funding

This work was supported by the National Institutes of Health, NINDS (R01 NS084975 award to JLL) and The Grainger Foundation.

Acknowledgments

We thank Kai Miller for providing human size deep brain stimulation electrode lead (Boston Scientific, Inc., Vercise DBS electrode lead) used in Figure 2.

Conflict of interest

We declare that the following investigators Kenneth T. Kishida, P. Read Montague, Charles D. Blaha, Michael Heien, Kendall H. Lee, Terry Lohrenz may present conflict of interest with the ideas presented in this review because of their past or current use of fast scan cyclic voltammetry in humans. The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

References

  • 1

    AdamsR. N. (1976). Probing brain chemistry with electroanalytical techniques.Anal. Chem.481126A1138A. 10.1021/ac50008a001

  • 2

    AriansenJ.HeienM. L. A. V.HermansA.PhillipsP. E. M.HernadiI.BermudezM.et al (2012). Monitoring extracellular pH, oxygen, and dopamine during reward delivery in the striatum of primates.Front. Behav. Neurosci.6:36. 10.3389/fnbeh.2012.00036

  • 3

    BennetK. E.TomshineJ. R.MinH.-K.ManciuF. S.MarshM. P.PaekS. B.et al (2016). A diamond-based electrode for detection of neurochemicals in the human brain.Front. Hum. Neurosci.10:102. 10.3389/fnhum.2016.00102

  • 4

    BoehlerC.CarliS.FadigaL.StieglitzT.AsplundM. (2020). Tutorial: guidelines for standardized performance tests for electrodes intended for neural interfaces and bioelectronics.Nat. Protocols1535573578. 10.1038/s41596-020-0389-382

  • 5

    BorlandL. M.MichaelA. C. (2007). “An introduction to electrochemical methods,” in Neuroscience. Electrochemical Methods for Neuroscience, edsMichaelA. C.BorlandL. M. (Boca Raton, FL: Taylor & Francis Group, LLC.).

  • 6

    BorlandL. M.ShiG.YangH.MichaelA. C. (2005). Voltammetric study of extracellular dopamine near microdialysis probes acutely implanted in the striatum of the anesthetized rat.J. Neurosci. Methods146149158. 10.1016/j.jneumeth.2005.02.002

  • 7

    BrazellM. P.KasserR. J.RennerK. J.FengJ.MoghaddamB.AdamsR. N. (1987). Electrocoating carbon fiber microelectrodes with Nafion improves selectivity for electroactive neurotransmitters.J. Neurosci. Methods22167172. 10.1016/0165-0270(87)90011-90012

  • 8

    BucherE. S.WightmanR. M. (2015). Electrochemical analysis of neurotransmitters.Ann. Rev. Anal. Chem.8239261. 10.1146/annurev-anchem-071114-40426

  • 9

    CastagnolaE.RobbinsE. M.WoeppelK. M.McGuierM.GolabchiA.TaylorI. M.et al (2020). Real-Time fast scan cyclic voltammetry detection and quantification of exogenously administered melatonin in mice brain.Front. Bioeng. Biotechnol.8:602216. 10.3389/fbioe.2020.602216

  • 10

    ChangS.-Y.JayT.MuñozJ.KimI.LeeK. H. (2012). Wireless fast-scan cyclic voltammetry measurement of histamine using WINCS – a proof-of-principle study.Analyst137, 21582165. 10.1039/c2an16038b

  • 11

    ChangA.-Y.DuttaG.SiddiquiS.ArumugamP. U. (2019). Surface fouling of ultrananocrystalline diamond microelectrodes during dopamine detection: improving lifetime via electrochemical cycling.ACS Chem. Neurosci.10313322. 10.1021/acschemneuro.8b00257

  • 12

    Cheney-ThammJ.ReiteM.AlianielloE. A.YamamotoB. K.CapitanioJ. P.FreedC. R. (1984). Caudate electrochemical response following amphetamine administration in pigtail monkeys.Life Sci.3514531460. 10.1016/0024-3205(84)90161-90169

  • 13

    Cheney-ThammJ.AlianelloE. A.FreedC. R.ReiteM. (1987). In vivo electrochemical recording of acetaminophen in non human primate brain.Life Sci.40, 375379.

  • 14

    ChongA.FischerG.DartB. R.WooleyP. H. (2015). Long-term elastic durability of polymer matrix composite materials after repeated steam sterilization.Am. J. Orthoped.44E427E433.

  • 15

    ClarkJ. J.SandbergS. G.WanatM. J.GanJ. O.HorneE. A.HartA. S.et al (2010). Chronic microsensors for longitudinal, subsecond dopamine detection in behaving animals.Nat. Methods7126129. 10.1038/nmeth.1412

  • 16

    Code of Federal Regulations (2021a). Part 812 – Investigational Device Exemptions. Title 21 Chapter I – Subchapter H. e-CFR.

  • 17

    Code of Federal Regulations (2021b). Part 820 – Quality System Regulation. Title 21 – Chapter I – Subchapter H. e-CFR.

  • 18

    CoganS. F. (2008). Neural stimulation and recording electrodes.Annu. Rev. Biomed. Eng.10275309. 10.1146/annurev.bioeng.10.061807.160518

  • 19

    CoganS. F.GuzelianA. A.AgnewW. F.YuenT. G. H.McCreeryD. B. (2004). Over-pulsing degrades activated iridium oxide films used for intracortical neural stimulation.J. Neurosci. Methods137141150. 10.1016/j.jneumeth.2004.02.019

  • 20

    CoganS. F.LudwigK. A.WelleC. G.TakmakovP. (2016). Tissue damage thresholds during therapeutic electrical stimulation.J. Neural Eng.13:021001. 10.1088/1741-2560/13/2/021001

  • 21

    DeWaeleM.OhY.ParkC.KangY. M.ShinH.BlahaC. D.et al (2017). A baseline drift detrending technique for fast scan cyclic voltammetry.Analyst14243174321. 10.1039/c7an01465a

  • 22

    EarlC. D.SautterJ.XieJ.KrukZ. L.KupschA.OertelW. H. (1998). Pharmacological characterisation of dopamine overflow in the striatum of the normal and MPTP-treated common marmoset, studied in vivo using fast cyclic voltammetry, nomifensine and sulpiride.J. Neurosci. Methods85201209. 10.1016/S0165-0270(98)00134-134

  • 23

    GanesanaM.LeeS. T.WangY.VentonB. J. (2017). Analytical techniques in neuroscience: recent advances in imaging, separation, and electrochemical methods.Anal. Chem.89314341. 10.1021/acs.analchem.6b04278

  • 24

    GloverG. H. (2011). Overview of functional magnetic resonance imaging.Neurosurg. Clin. N. Am.22133139. 10.1016/j.nec.2010.11.001

  • 25

    GodaraA.RaabeD.GreenS. (2007). The influence of sterilization processes on the micromechanical properties of carbon fiber-reinforced PEEK composites for bone implant applications.Acta Biomaterialia3209220.

  • 26

    GornaK.GogolewskiS. (2003). The effect of gamma radiation on molecular stability and mechanical properties of biodegradable polyurethanes for medical applications.Polymer Degradation Stabil.79465474.

  • 27

    HashemiP.WalshP. L.GuillotT. S.Gras-NajjarJ.TakmakovP.CrewsF. T.et al (2011). Chronically implanted, nafion-coated Ag/AgCl reference electrodes for neurochemical applications.ACS Chem. Neurosci.2658666. 10.1021/cn2000684

  • 28

    HermansA.KeithleyR. B.KitaJ. M.SombersL. A.WightmanR. M. (2008). Dopamine detection with fast-scan cyclic voltammetry used with analog background subtraction.Anal. Chem.8040404048. 10.1021/ac800108j

  • 29

    ISO10993-1:2018 (2018). Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process. Geneva: International Organization for Standardization.

  • 30

    ISO10993-6:2016 (2016). Biological Evaluation of Medical Devices. Part 6: Tests for Local Effects After Implantation.Geneva: International Organization for Standardization.

  • 31

    ISO11737-1:2018 (2018). Sterilization of Health Care Products.Geneva: International Organization for Standardization.

  • 32

    ISO14971:2019 (2019). Medical Devices - Application of Risk Management to Medical Devices.Geneva: International Organization for Standardization.

  • 33

    JackowskaK.KrysinskiP. (2013). New trends in the electrochemical sensing of dopamine.Anal. Bioanal. Chem.40537533771. 10.1007/s00216-012-6578-6572

  • 34

    Jaquins-GerstlA.MichaelA. C. (2015). A review of the effects of FSCV and microdialysis measurements on dopamine release in the surrounding tissue.Analyst14036963708. 10.1039/c4an02065k

  • 35

    JohnsonJ. A.GrayJ. H.RodebergN. T.WightmanR. M. (2017). Multivariate curve resolution for signal isolation from fast-scan cyclic voltammetric data.Anal. Chem.891054710555. 10.1021/acs.analchem.7b02771

  • 36

    JohnsonJ. A.RodebergN. T.WightmanR. M. (2016). Failure of standard training sets in the analysis of fast-scan cyclic voltammetry data.ACS Chem. Neurosci.7349359. 10.1021/acschemneuro.5b00302

  • 37

    KachoosangiR. T.ComptonR. G. (2007). A simple electroanalytical methodology for the simultaneous determination of dopamine, serotonin and ascorbic acid using an unmodified edge plane pyrolytic graphite electrode.Anal. Bioanal. Chem.38727932800. 10.1007/s00216-007-1129-y

  • 38

    KishidaK. T.SaezI.LohrenzT.WitcherM. R.LaxtonA. W.TatterS. B.et al (2016). Subsecond dopamine fluctuations in human striatum encode superposed error signals about actual and counterfactual reward.Proc. Natl. Acad. Sci. U S A.113200205.

  • 39

    KishidaK. T.SandbergS. G.LohrenzT.ComairY. G.SáezI.PhillipsP. E.et al (2011). Sub-second dopamine detection in human striatum.PLoS One6:e23291. 10.1371/journal.pone.0023291

  • 40

    KolliR.KaivosojaE.LevonK. (2015). Choice of reference electrode is critical for potentiometric whole cell-based sensor.Electroanalysis2716361641. 10.1002/elan.201400710

  • 41

    KozaiT. D. Y.Jaquins-GerstlA. S.VazquezA. L.MichaelA. C.CuiX. T. (2015). Brain tissue responses to neural implants impact signal sensitivity and intervention strategies.ACS Chem. Neurosci.64867. 10.1021/cn500256e

  • 42

    LeeK. H.ChangS. Y.JangD. P.KimI.GoerssS.Van GompelJ.et al (2011). Emerging techniques for elucidating mechanism of action of deep brain stimulation.Conf. Proc. IEEE Eng. Med. Biol. Soc.2011, 677680. 10.1109/IEMBS.2011.6090152

  • 43

    LeeK. H.LujanJ. L.TrevathanJ. K.RossE. K.BartolettaJ. J.ParkH. O.et al (2017). WINCS harmoni: closed-loop dynamic neurochemical control of therapeutic interventions.Sci. Rep.7:46675. 10.1038/srep46675

  • 44

    LindN. M.MoustgaardA.JelsingJ.VajtaG.CummingP.HansenA. K. (2007). The use of pigs in neuroscience: modeling brain disorders.Neurosci. Biobehav. Rev.31728751. 10.1016/j.neubiorev.2007.02.003

  • 45

    LohrenzT.KishidaK. T.MontagueP. R. (2016). BOLD and its connection to dopamine release in human striatum: a cross-cohort comparison.Philos. Trans. R. Soc. B: Biol. Sci.371:20150352.

  • 46

    LucasA. D.MerrittK.HitchinsV. M.WoodsT. O.McNameeS. G.LyleD. B.et al (2003). Residual ethylene oxide in medical devices and device material.J. Biomed. Mater. Res. Part B: Appl. Biomater.66548552.

  • 47

    MartinH. B.SmithB. A.AngusJ. C.LandauU.AndersonA. B. (1998). Boron-Doped diamond films for electrochemical applications.MRS Proc.555:217. 10.1557/PROC-555-217

  • 48

    McCreeryD. B.AgnewW. F.YuenT. G.BullaraL. (1990). Charge density and charge per phase as cofactors in neural injury induced by electrical stimulation.IEEE Trans. Biomed. Eng.379961001. 10.1109/10.102812

  • 49

    McCreeryD. B.AgnewW. F.YuenT. G.BullaraL. A. (1988). Comparison of neural damage induced by electrical stimulation with faradaic and capacitor electrodes.Ann. Biomed. Eng.16463481. 10.1007/bf02368010

  • 50

    McDermottM. T.McCreeryR. L. (1994). Scanning tunneling microscopy of ordered graphite and glassy carbon surfaces: electronic control of quinone adsorption.Langmuir1043074314.

  • 51

    MendesG. C.BrandaoT. R.SilvaC. L. (2007). Ethylene oxide sterilization of medical devices: a review.Am. J. Infect. Control35574581.

  • 52

    MeunierC. J.MitchellE. C.RobertsJ. G.ToupsJ. V.McCartyG. S.SombersL. A. (2018). Electrochemical selectivity achieved using a double voltammetric waveform and partial least squares regression: differentiating endogenous hydrogen peroxide fluctuations from shifts in pH.Anal. Chem.9017671776. 10.1021/acs.analchem.7b03717

  • 53

    MillarJ.StamfordJ. A.KrukZ. L.WightmanR. M. (1985). Electrochemical, pharmacological and electrophysiological evidence of rapid dopamine release and removal in the rat caudate nucleus following electrical stimulation of the median forebrain bundle.Eur. J. Pharmacol.109341348. 10.1016/0014-2999(85)90394-90392

  • 54

    MinH. K.RossE. K.JoH. J.ChoS.SettellM. L.JeongJ. H.et al (2016). Dopamine release in the nonhuman primate caudate and putamen depends upon site of stimulation in the subthalamic nucleus.J. Neurosci.36, 60226029. 10.1523/JNEUROSCI.0403-16.2016

  • 55

    MitchellE. C.DunawayL. E.McCartyG. S.SombersL. A. (2017). Spectroelectrochemical characterization of the dynamic carbon-fiber surface in response to electrochemical conditioning.Langmuir3378387846.

  • 56

    MohanarajS.WonnenbergP.CohenB.ZhaoH.HartingsM. R.ZouS.et al (2019). Gold nanoparticle modified carbon fiber microelectrodes for enhanced neurochemical detection.J. Vis. Exp.

  • 57

    MontagueP. R.KishidaK. T. (2018). Computational underpinnings of neuromodulation in humans.Cold. Spring Harb. Symp. Quant. Biol.837182. 10.1101/sqb.2018.83.038166

  • 58

    MoquinK. F.MichaelA. C. (2009). Tonic autoinhibition contributes to the heterogeneity of evoked dopamine release in the rat striatum.J. Neurochem.11014911501. 10.1111/j.1471-4159.2009.06254.x

  • 59

    MoranR. J.KishidaK. T.LohrenzT.SaezI.LaxtonA. W.WitcherM. R.et al (2018). The protective action encoding of serotonin transients in the human brain.Neuropsychopharmacology43, 14251435. 10.1038/npp.2017.304

  • 60

    NairP. D. (1995). Currently practised sterilization methods-some inadvertent consequences.J. Biomater. Appl.10121135.

  • 61

    NakajimaA.ShimoY.UkaT.HattoriN. (2017). Subthalamic nucleus and globus pallidus interna influence firing of tonically active neurons in the primate striatum through different mechanisms.Eur. J. Neurosci.4626622673.

  • 62

    NicolaiE. N.TrevathanJ. K.RossE. K.LujanJ. L.BlahaC. D.BennetK. E.et al (2017). Detection of norepinephrine in whole blood via fast scan cyclic voltammetry.IEEE Int. Symp. Med. Meas. Appl.2017111116. 10.1109/MeMeA.2017.7985859

  • 63

    OlivieriA. C. (2008). Analytical advantages of multivariate data processing. one, two, three, infinity?Anal. Chem.8057135720. 10.1021/ac800692c

  • 64

    ParkJ. H.ChungS. J.LeeC. S.JeonS. R. (2011). Analysis of hemorrhagic risk factors during deep brain stimulation surgery for movement disorders: comparison of the circumferential paired and multiple electrode insertion methods.Acta Neurochir (Wien)15315731578. 10.1007/s00701-011-0997-992

  • 65

    PatelA. N.TanS.-Y.MillerT. S.MacphersonJ. V.UnwinP. R. (2013). Comparison and reappraisal of carbon electrodes for the voltammetric detection of dopamine.Anal. Chem.851175511764.

  • 66

    PatelP. R.ZhangH.RobbinsM. T.NofarJ. B.MarshallS. P.KobylarekM. J.et al (2016). Chronic in vivo stability assessment of carbon fiber microelectrode arrays.J. Neural Eng.13:066002. 10.1088/1741-2560/13/6/066002

  • 67

    PhillipsK. A.BalesK. L.CapitanioJ. P.ConleyA.CzotyP. W.‘t HartB. A.et al (2014). Why primate models matter.Am. J. Primatol.76801827. 10.1002/ajp.22281

  • 68

    PihelK.WalkerQ. D.WightmanR. M. (1996). Overoxidized polypyrrole-coated carbon fiber microelectrodes for dopamine measurements with fast-scan cyclic voltammetry.Anal. Chem.6820842089. 10.1021/ac960153y

  • 69

    PuthongkhamP.VentonB. J. (2019). Nanodiamond coating improves the sensitivity and antifouling properties of carbon fiber microelectrodes.ACS Sensors424032411.

  • 70

    RajendrachariS. (2018). Investigation of Electrochemical Pitting Corrosion by Linear Sweep Voltammetry: A Fast and Robust Approach. Voltammetry.London: IntechOpen.

  • 71

    RamssonE. S.CholgerD.DioniseA.PoirierN.AndrusA.CurtissR. (2015). Characterization of fast-scan cyclic voltammetric electrodes using paraffin as an effective sealant with in vitro and in vivo applications.PLoS One10:e0141340. 10.1371/journal.pone.0141340

  • 72

    RiesM. D.WeaverK.RoseR. M.GuntherJ.SauerW.SealsN. (1996). Fatigue strength of polyethylene after sterilization by gamma irradiation or ethylene oxide.Clin. Orthopaed. Related Res.3338795.

  • 73

    RobbleeL. S.RoseT. L. (1990). “The electrochemistry of electrical stimulation,” in Proceedings of the 20th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, (Piscataway, NJ: IEEE), 14791480.

  • 74

    RobertsJ. G.SombersL. A. (2018). Fast-Scan cyclic voltammetry: chemical sensing in the brain and beyond.Anal. Chem.90490504. 10.1021/acs.analchem.7b04732

  • 75

    RodebergN. T.SandbergS. G.JohnsonJ. A.PhillipsP. E. M.WightmanR. M. (2017). Hitchhiker’s guide to voltammetry: acute and chronic electrodes for in vivo fast-scan cyclic voltammetry.ACS Chem. Neurosci.8221234. 10.1021/acschemneuro.6b00393

  • 76

    RossE. K.KimJ. P.SettellM. L.HanS. R.BlahaC. D.MinH.-K.et al (2016). Fornix deep brain stimulation circuit effect is dependent on major excitatory transmission via the nucleus accumbens.Neuroimage128, 138148. 10.1016/j.neuroimage.2015.12.056

  • 77

    SansurC. A.FrysingerR. C.PouratianN.FuK. M.BittlM.OskouianR. J.et al (2007). Incidence of symptomatic hemorrhage after stereotactic electrode placement.J. Neurosurg.1079981003. 10.3171/jns-07/11/0998

  • 78

    SchluterE. W.MitzA. R.CheerJ. F.AverbeckB. B. (2014). Real-time dopamine measurement in awake monkeys.PLoS One9:e98692. 10.1371/journal.pone.0098692

  • 79

    SchwerdtH. N.ShimazuH.AmemoriK.-I.AmemoriS.TierneyP. L.GibsonD. J.et al (2017b). Long-term dopamine neurochemical monitoring in primates.Proc. Natl. Acad. Sci. U S A.1141326013265. 10.1073/pnas.1713756114

  • 80

    SchwerdtH. N.KimM. J.AmemoriS.HommaD.YoshidaT.ShimazuH.et al (2017a). Subcellular probes for neurochemical recording from multiple brain sites.Lab. Chip1711041115. 10.1039/c6lc01398h

  • 81

    SeatonB. T.HillD. F.CowenS. L.HeienM. L. (2020). Mitigating the effects of electrode biofouling-induced impedance for improved long-term electrochemical measurements in vivo.Anal. Chem.9263346340. 10.1021/acs.analchem.9b05194

  • 82

    SettellM. L.TestiniP.ChoS.LeeJ. H.BlahaC. D.JoH. J.et al (2017). Functional circuitry effect of ventral tegmental area deep brain stimulation: imaging and neurochemical evidence of mesocortical and mesolimbic pathway modulation.Front. Neurosci.11:104. 10.3389/fnins.2017.00104

  • 83

    ShannonR. V. (1992). A model of safe levels for electrical stimulation.IEEE Trans. Biomed. Eng.39424426. 10.1109/10.126616

  • 84

    ShonY. M.LeeK. H.GoerssS. J.KimI. Y.KimbleC.Van GompelJ. J.et al (2010). High frequency stimulation of the subthalamic nucleus evokes striatal dopamine release in a large animal model of human DBS neurosurgery.Neurosci. Lett.475, 136140. 10.1016/j.neulet.2010.03.060

  • 85

    StamfordJ. A.KrukZ. L.MillarJ.WightmanR. M. (1984). Striatal dopamine uptake in the rat: in vivo analysis by fast cyclic voltammetry.Neurosci. Lett.51133138. 10.1016/0304-3940(84)90274-x

  • 86

    Suaud-ChagnyM. F.CherguiK.ChouvetG.GononF. (1992). Relationship between dopamine release in the rat nucleus accumbens and the discharge activity of dopaminergic neurons during local in vivo application of amino acids in the ventral tegmental area.Neuroscience496372. 10.1016/0306-4522(92)90076-E

  • 87

    SuzukiA.IvandiniT. A.YoshimiK.FujishimaA.OyamaG.NakazatoT.et al (2007). Fabrication, characterization, and application of boron-doped diamond microelectrodes for in vivo dopamine detection.Anal. Chem.7986088615. 10.1021/ac071519h

  • 88

    TaylorI. M.NesbittK. M.WaltersS. H.VarnerE. L.ShuZ.BartlowK. M.et al (2015). Kinetic diversity of dopamine transmission in the dorsal striatum.J. Neurochem.133522531. 10.1111/jnc.13059

  • 89

    TrevathanJ. K.YousefiA.ParkH. O.BartolettaJ. J.LudwigK. A.LeeK. H.et al (2017). Computational modeling of neurotransmitter release evoked by electrical stimulation: nonlinear approaches to predicting stimulation-evoked dopamine release.ACS Chem. Neurosci.8, 394410.

  • 90

    Van GompelJ. J.BowerM. R.WorrellG. A.SteadM.ChangS.-Y.GoerssS. J.et al (2014). Increased cortical extracellular adenosine correlates with seizure termination.Epilepsia55233244. 10.1111/epi.12511

  • 91

    VreelandR. F.AtcherleyC. W.RussellW. S.XieJ. Y.LuD.LaudeN. D.et al (2015). Biocompatible PEDOT:Nafion composite electrode coatings for selective detection of neurotransmitters in vivo.Anal. Chem.8726002607. 10.1021/ac502165f

  • 92

    WangY.MichaelA. C. (2012). Microdialysis probes alter presynaptic regulation of dopamine terminals in rat striatum.J. Neurosci. Methods2083439. 10.1016/j.jneumeth.2012.04.009

  • 93

    WightmanR. M.HeienM. L.WassumK. M.SombersL. A.AragonaB. J.KhanA. S.et al (2007). Dopamine release is heterogeneous within microenvironments of the rat nucleus accumbens.Eur. J. Neurosci.2620462054. 10.1111/j.1460-9568.2007.05772.x

  • 94

    WoolleyA. J.DesaiH. A.OttoK. J. (2013). Chronic intracortical microelectrode arrays induce non-uniform, depth-related tissue responses.J. Neural Eng.10:026007. 10.1088/1741-2560/10/2/026007

  • 95

    YangC.TrikantzopoulosE.JacobsC. B.VentonB. J. (2017). Evaluation of carbon nanotube fiber microelectrodes for neurotransmitter detection: correlation of electrochemical performance and surface properties.Anal. Chim. Acta965, 18. 10.1016/j.aca.2017.01.039

  • 96

    YoshimiK.KumadaS.WeitemierA.JoT.InoueM. (2015). Reward-Induced phasic dopamine release in the monkey ventral striatum and putamen.PLoS One10:e0130443. 10.1371/journal.pone.0130443

  • 97

    YoshimiK.NayaY.MitaniN.KatoT.InoueM.NatoriS.et al (2011). Phasic reward responses in the monkey striatum as detected by voltammetry with diamond microelectrodes.Neurosci. Res.714962.

  • 98

    ZachekM. K.HermansA.WightmanR. M.McCartyG. S. (2008). Electrochemical dopamine detection: comparing gold and carbon fiber microelectrodes using background subtracted fast scan cyclic voltammetry.J. Electroanal Chem. (Lausanne)614113120. 10.1016/j.jelechem.2007.11.007

  • 99

    ZachekM. K.ParkJ.TakmakovP.WightmanR. M.McCartyG. S. (2010). Microfabricated FSCV-compatible microelectrode array for real-time monitoring of heterogeneous dopamine release.Analyst13515561563. 10.1039/c0an00114g

  • 100

    ZachekM. K.TakmakovP.MoodyB.WightmanR. M.McCartyG. S. (2009). Simultaneous decoupled detection of dopamine and oxygen using pyrolyzed carbon microarrays and fast-scan cyclic voltammetry.Anal. Chem.8162586265. 10.1021/ac900790m

  • 101

    ZhaoY.LiS.-H.ChuJ.ChenY.-P.LiW.-W.YuH.-Q.et al (2012). A nano-sized Au electrode fabricated using lithographic technology for electrochemical detection of dopamine. Biosens. Bioelectron.35, 115122. 10.1016/j.bios.2012.02.030

  • 102

    ZhengJ.ClogstonJ. D.PatriA. K.DobrovolskaiaM. A.McNeilS. E. (2011). Sterilization of silver nanoparticles using standard gamma irradiation procedure affects particle integrity and biocompatibility.J. Nanomed. Nanotechnol.2011(Suppl. 5):001.

Summary

Keywords

clinical neurochemistry, deep brain stimulation, fast scan cyclic voltammetry, intraoperative, neurochemical signaling, neurophysiology

Citation

Lucio Boschen S, Trevathan J, Hara SA, Asp A and Lujan JL (2021) Defining a Path Toward the Use of Fast-Scan Cyclic Voltammetry in Human Studies. Front. Neurosci. 15:728092. doi: 10.3389/fnins.2021.728092

Received

20 June 2021

Accepted

25 October 2021

Published

12 November 2021

Volume

15 - 2021

Edited by

Grzegorz Marcin Wójcik, Marie Curie-Sklodowska University, Poland

Reviewed by

Jit Muthuswamy, Arizona State University, United States; Vassiliy Tsytsarev, University of Maryland, College Park, United States; Parastoo Hashemi, University of South Carolina, United States

Updates

Copyright

*Correspondence: Suelen Lucio Boschen,

This article was submitted to Neural Technology, a section of the journal Frontiers in Neuroscience

Disclaimer

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Outline

Figures

Cite article

Copy to clipboard


Export citation file


Share article

Article metrics