Skip to main content

ORIGINAL RESEARCH article

Front. Pediatr., 18 June 2019
Sec. Neonatology
Volume 7 - 2019 | https://doi.org/10.3389/fped.2019.00248

Bedside Measurement of Volatile Organic Compounds in the Atmosphere of Neonatal Incubators Using Ion Mobility Spectrometry

  • 1Department of Applied Chemistry, Reutlingen University, Reutlingen, Germany
  • 2Department of General Paediatrics and Neonatology, Saarland University Medical School, Homburg, Germany

Background: Early and non-invasive diagnosis of common diseases is of great importance in the care of preterm infants. We hypothesized that volatile organic compounds (VOC) can be successfully measured in the neonatal incubator atmosphere.

Methods: This is a feasibility study to investigate whether the discrimination of occupied and unoccupied neonatal incubators is possible by bedside measurement of volatile organic compounds (VOCs) on the neonatal intensive care unit. VOC profiles were measured in the incubator air using ion mobility spectrometry coupled to multi-capillary columns (BreathDiscovery B&S Analytik GmbH, Dortmund, Germany).

Results: Seventeen incubators occupied by preterm infants (50 measurements) and nine unoccupied neonatal incubators were sampled, using 37 room air measurements as controls. Three VOC signals that allow the discrimination between occupied and unoccupied incubators were identified. The best discrimination was reached by peak P20 exhibiting a sensitivity, specificity, positive predictive value and negative predictive value of 94.0, 88.9, 97.3, and 72.3%, respectively. Use of a decision tree improved these values to 100.0, 88.9, 98.0, and 100.0%, respectively.

Discussion: A bedside method that allows the characterization of VOC profiles in the neonatal incubator atmosphere using ion mobility spectrometry was established. Occupied and unoccupied incubators could be discriminated by characterizing VOC profiles. This technique has the potential to yield results within minutes. Thus, future studies are recommended to test the hypothesis that VOCs within neonatal incubators are useful biomarkers for non-invasive diagnostics in preterm neonates.

Introduction

More than one in ten infants is born prematurely (1, 2). The risk of complications is inversely correlated with lower gestational age (3, 4). Common complications in preterm infants are sepsis, bronchopulmonary dysplasia (BPD), retinopathy of prematurity (ROP), intraventricular hemorrhage (IVH), and necrotizing enterocolitis (NEC) (1, 48). However, diagnosis is dependent on the combination of various symptoms and invasive methods like blood analysis (610). Non-invasive diagnosis via detection of volatile organic compounds (VOCs) has been examined for sepsis, NEC, and BPD in several studies with promising results (1016). The most commonly used device is the electronic nose (eNose) that consists of an array of sensors, which allows to distinguish groups based on their specific so called “smell prints” (1115). Nevertheless, this method of pattern recognition does not allow the characterization and identification of possible biomarkers. In addition, biological samples such as tracheal aspirate (12) and feces (11, 1316) were used at defined time points. Exhaled air on the contrary is released with every breath and is therefore particularly suitable as a sample for non-invasive diagnosis and continuous monitoring. Ion mobility spectrometry (IMS), which allows the detection of VOCs in lower ppbv to pptv (ng L−1 to pg L−1) range, has the potential to be used as a bedside method (17, 18).

IMS coupled to multi-capillary columns (MCC/IMS) have been successfully used for the non-invasive diagnosis of several diseases (1922) and for the determination of characteristic VOCs and patterns from Escherichia coli (23, 24) and Pseudomonas aeruginosa (24, 25).

One major advantage of bedside IMS measurements is that results are available within minutes. However, multiple potential contaminants such as VOCs from hand sanitizers, plastic ware, and other materials must be considered to ensure the specificity of potential volatile biomarkers (2628).

Therefore, the first step toward non-invasive diagnosis from neonatal incubator air is to establish a sampling technique for neonatal incubator air and to investigate whether distinction of occupied and unoccupied neonatal incubator can be achieved by single VOCs or groups of VOCs arising from the neonatal incubator air using MCC/IMS.

Materials and Methods

This study was performed at the Department of Pediatrics, Saarland University Medical Center, Homburg (Germany) and was approved by the Ethics Committee Saarbrücken (reference 276/17). Parents' written consent was obtained prior to the measurements. Premature infants weighing <1.800 g that were treated in an incubator were enrolled in the study. The model of neonatal incubator was the THERMOCARE Vita (WY2402, Weyer GmbH, Kürten-Herweg, Germany).

For analysis of neonatal incubator air, a MCC/IMS BreathDiscovery (B&S Analytik GmbH, Dortmund, Germany), laptop computer and a synthetic air gas bottle were placed on a metal cart. The isothermal (40°C) pre-separation was accomplished with an OV-5 multi-capillary column (MCC) (Multichrom, Novosibirsk, Russia). The device and sampling parameters are given in Supplemental Table 1. The methods for VOC analysis were published earlier (17, 1921, 2931). All samples were taken with polytetrafluorethylene tubing covered by a single-use sterile glass pipette to avoid contamination between the neonatal incubators. The glass pipette was inserted with a length of proximately 10 cm through the side ports (grommets for tubes and cables of medical devices) at the head side of the neonatal incubators. According to pilot studies and the device standard, for MCC/IMS a sampling time of 20 s with a sample flow of 100 mL/min was selected. Isothermal separation was performed at 40°C with an MCC flow of 150 mL/min. Drift flow was set to 100 mL/min and the ions were detected in positive mode.

Data Assignment

A total of 17 preterm infants were measured with a sample count of 50 measurements. Each measurement was a complete MCC/IMS chromatogram consisting of 1.500 single spectra with 2.500 data points per spectra. As a reference, nine unoccupied neonatal incubators were measured with an average temperature of 36°C (32–37°C) and an average relative humidity of 60% (55–67%) as well as 41 room air measurements.

Fifteen measurements showed particularly high concentrations of ethanol that most likely originated from disinfectants. An overload was defined as the predominance of a given signal (ethanol) that overlaid neighboring peaks and that was associated with a peak constriction (Figure 1). To rule out a distortion of the analyses due to strong ethanol peaks, a two-step evaluation was chosen. First, the 15 overloaded measurements (4 room air samples and 11 preterm infant samples) were excluded for model generation (model data set). The model data set therefore included measurements of 39 preterm infants, 9 unoccupied neonatal incubators and 37 room air samples. In the second step, the primarily excluded overloaded measurements were reincluded into the dataset for model validation. The validation data set included 50 preterm infants, 9 unoccupied neonatal incubators, and 41 room air samples.

FIGURE 1
www.frontiersin.org

Figure 1. Representative chromatograms of unoccupied (A) and occupied incubators (B). The peaks that allow discrimination of the groups are indicated with arrows and rectangles. (C) shows a chromatogram with a strong ethanol peak leading to peak constriction.

In contrast to occupied incubators, which exhibit a high natural variability in VOC patterns, unoccupied incubator VOC patterns are almost constant over time. To avoid statistically overestimated significance of the unoccupied incubator peaks, no replicate measurements were included in the evaluation.

Statistical Analysis

The generated MCC/IMS data were evaluated with the software VisualNow 3.7 (B&S Analytik GmbH, Dortmund, Germany). All peaks were characterized by their specific combination of retention time per second and drift time (corresponding 1/K0-value). The peak height is correlated to the concentration (32). The databank layer 20160426_SubstanzDbNIST_122_St_layer (B&S Analytik GmbH, 2016) was used for peak referencing and determination of retention times and 1/K0-values. Box-and-Whisker plots and a rank sum test (Wilcoxon-Mann-Whitney test using Bonferroni correction) were used. Significant peaks [p < 0.05, 95% confidential interval (CI)] were used for further evaluation with decision trees (DT) (33, 34) using RapidMiner Studio Free 8.2.001 (RapidMiner GmbH, Dortmund, Germany) and principal component analysis (PCA) (35, 36) using Unscrambler X 10.4 (CAMO Software AS, Oslo, Norway).

Results

A total of 149 signals (peaks) were detected from room air, neonatal incubators and preterm infants. A reduced dataset was generated that only includes signals with significance after Bonferroni correction of p < 0.05 (95% CI) for the separation of at least two groups. These 73 peaks were included for further evaluation. Representative chromatograms are shown in Figure 1.

For the distinction of neonatal incubators and preterm infants, three peaks with a significance of at least p < 0.05 (95% CI) after Bonferroni correction were found. The average (± standard deviation) signal intensities of P20 and P113 were higher for preterm infants (50.2 ± 23.2 V, 46.8 ± 26.3 V) than for room air (15.1 ± 5.3 V, 17.5 ± 7.4 V) and neonatal incubators (21.7 ± 5.2 V, 19.0 ± 6.7 V). For P113 (p < 0.05), P20 (p < 0.01), and P26 (p < 0.01) a sensitivity/specificity of 94.9%/55.6%, 92.3%/88.9%/, and 100.0%/55.6%, as well as positive/negative predictive values of 90.2%/71.4%, 97.3%/72.7%, and 90.7%/100.0% were determined, respectively (Table 1).

TABLE 1
www.frontiersin.org

Table 1. Statistical analyses for the model set and validation set for peaks P20, P26, and P113.

The peak area for P20 differed significantly between preterm infants vs. unoccupied neonatal incubators and room air, respectively (Figure 2). For P26 unoccupied neonatal incubators showed the highest intensity with 13.2 ± 6.9 V in comparison with 5.5 ± 2.0 V and 3.0 ± 1.4 V for preterm infants and room air, respectively (Figure 3).

FIGURE 2
www.frontiersin.org

Figure 2. Intensity distribution of P20. The best threshold for the separation (26 mV) is displayed. PTI, preterm neonate infant; NI, neonatal incubator; RA, Room air.

FIGURE 3
www.frontiersin.org

Figure 3. Box-and-Whisker plot for peak P20 for original data set (nNI = 9, nPTI = 39, nRA = 37). Measurements are marked with crosses; incorrectly classified measurements are encircled. Significance levels after Bonferroni correction are indicated. PTI, preterm neonate infant; NI, neonatal incubator; RA, Room air.

All 73 significant peaks were analyzed by PCA with 5 principal components (PC) using the peak heights as a variable and each peak as a sample (Supplemental Figure 1). Using 5 principal components, the explained variance was 88.3% for the distinction of all three groups and 85% for the distinction of preterm infants and unoccupied neonatal incubators. This PCA did not allow a clear distinction between the measurement from preterm infants and unoccupied neonatal incubators (Supplemental Figure 1).

The method of calculating decision trees confirmed in the model data set and in the validation data set that the peaks P20, P26, and P113 that had been previously identified by Bonferroni correction were best suited to differentiate between preterm neonates, unoccupied neonatal incubators and room air (Supplemental Figures 2, 3).

In the model data set the decision tree reached a 100% separation of preterm infants and unoccupied neonatal incubators using peaks P20, P26, and P113 (Table 1). Moreover, the sensitivity, specificity, positive predictive value and negative predictive value all reached 100%. In the validation data set, the sensitivity, specificity, positive predictive value and negative predictive value were 100, 88.9, 100, and 98.3%. For both data sets, the peak P113 was not required to distinguish the groups via decision tree analysis.

Discussion

We demonstrate the feasibility of a novel technique that allows bedside measurements of VOC profiles using ion mobility spectrometry of the neonatal incubator atmosphere. Using this method, it was possible to distinguish VOC profiles of neonatal incubators that were occupied with preterm neonates from unoccupied neonatal incubators.

Most of the currently used biochemical diagnostic tests are based on non-volatile molecules that are solved in body fluids such as serum, plasma, saliva or urine. In particular, blood tests which are routinely required in the treatment of preterm neonates, are associated with distress during blood sampling, increased need for blood transfusion and time delays during sample processing. However, it is known that volatile organic compounds are released during many biological pathways and can serve as biomarkers in adults (37), children (38), and neonates (1116). Previous studies usually used exhaled breath (19, 39) or headspace measurements of biosamples (1116).

In this study it was confirmed that the presence of a preterm neonate is associated with changes in the VOC profile within the neonatal incubator atmosphere, using room air samples as controls. The presence of a preterm neonate was associated with changes in the VOC profile within the neonatal incubator atmosphere, using room air samples as controls. In this feasibility study 73 signals from VOCs were identified that differed significantly after Bonferroni correction between preterm infants, unoccupied neonatal incubators and room air. PCA of these 73 signals was insufficient to reliably differentiate the three groups of samples. This finding was expected since the majority of VOCs in all three groups most likely originated from environmental sources such as room air and equipment. Moreover, the number of measurements was relatively low (n = 100 measurements).

However, statistical analyses on individual peaks and on groups of peaks, including the calculation of decision trees, was successfully used to identify three peaks that allowed the distinction of occupied and unoccupied neonatal incubators. The biochemical identification of the VOCs responsible for peaks P20, P26, and P113 is part of ongoing experiments. The concentration of these analytes was too low for identification by gas chromatography/mass spectrometry (GC/MS). Therefore, sample enrichment is required in further experiments since the detection limit for MCC/IMS is lower compared to GC/MS analysis. However, a clear, reproducible identification of the peaks is provided by the specific drift and retention time, irrespective of the chemical identification of the molecule.

It was found that even the predominance of one particular environmental peak, ethanol, did not impede the analyses in our setting since the model data set (excluding measurements overloaded with ethanol) and the validation data set (including overloaded measurements) yielded similar significances for the diagnostically relevant peaks P20, P26, and P113 in this series.

However, there is a high risk that ethanol and other environmental VOCs may disturb the detection of VOCs that would be important for diagnostic use, including inflammatory diseases such as sepsis, BPD and NEC. For example, Fink et al. (40) found 3-pentanone and acetone to be significant for the distinction of sepsis vs. healthy control group in a rat model. Notably, 3-pentanone (1/K0 = 0.546 Vs/cm2; RT = 7.4 s) and acetone (1/K0 = 0.544 Vs/cm2, RT = 2.4 s) both eluate in the proximity of ethanol (1/K0 = 0.510 Vs/cm2, RT = 0.0 s) and therefore might be hidden at high ethanol concentrations. Hsieh et al. (27) demonstrated that the ethanol concentration in neonatal incubators remained in a concentration of up to 5 ppmv 30 min after the use of hand sanitizers. Ulanowska et al. (28) reported a calibration range for MCC/IMS of ethanol in the ppbv range, which is approximately 1/100–1/1,000 of the expected concentration according to Hsieh et al. (27).

Thus, the excessive sensitivity of IMS measurements and the highly variable environmental load with VOCs remain the crucial factors that need to be considered when volatile biomarkers are to be identified for diagnostic use. Potential approaches to reduce the effects of disinfectants and other potential contaminants on the measurements could be modifications of the multi-capillary flow column, or a reduced sampling time. A mathematical exclusion of the ethanol peak is also conceivable. All empty incubators were prepared and equipped identically to the occupied incubators with the only difference of the infant being present or not. Thus, we conclude that peaks P20, P26, and P113 are highly likely to be associated with the infant. Moreover, peaks P20, P26, and P113 were not detectable in potentially contaminating environmental sources such as diapers, clothing, pacifiers, medical devices, and skin care products (data not shown).

Taken together, the established method of bedside IMS measurements coupled to a multi-capillary column and the results of the presented feasibility study can be useful for the future development of non-invasive diagnostic methods using VOC analysis in the incubator atmosphere of preterm infants.

Data Availability

The datasets supporting the conclusions of this article are included within the article or are available from the authors upon request.

Ethics Statement

After written informed consent, 30 mL air were sampled from the incubator atmosphere. No additional interaction was performed. This study was approved by the Ethics Committee of the Philipps-University Marburg. AZ 91/11.

Author Contributions

MZ, SM, ET, SG-F, and JB designed the study. EK and RS supported organisational and coordinative tasks. SG-F and JS performed the experiments. JS evaluated the data. All authors contributed to writing the manuscript.

Funding

Funded by a grant from the von Behring-Roentgen Foundation, by the BMBF (SINOPE-Neo FKZ: 13GW0156E and PRIMAL Clinical Study FKZ: 01GL1746D) and by a grant from the Rhoen Klinikum AG.

Conflict of Interest Statement

Since September 2018 JB is an employee of a company producing spectrometers as used in the present paper.

The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Acknowledgments

We thank Ellen Maurer (Department of General Pediatrics and Neonatology, Saarland University Medical School, Homburg, Germany) for preparation of the necessary equipment. Moreover, JS thanks Rebecca Metzger, and Svenja Köble (Reutlingen University, Faculty of Applied Chemistry, Reutlingen, Germany) for helpful discussions.

Supplementary Material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fped.2019.00248/full#supplementary-material

Abbreviations

CI, Confidential interval; DT, Decision tree; IMS, Ion mobility spectrometry; IVH, Intraventricular hemorrhage; MCC, Multi-capillary column; MCC/IMS, Ion mobility spectrometer coupled to multi-capillary column; NI, Neonatal incubators; PCA, Principal Component Analysis; PTI, Preterm infant; R, Room air; ROP, Retinopathy of Prematurity; TDS-GC/MS, Thermo-desorption-gas chromatography coupled to mass spectrometry; VOC, Volatile organic compound.

References

1. Liu L, Oza S, Hogan D, Chu Y, Perin J, Zhu J, et al. Global, regional, and national causes of under-5 mortality in 2000–15: an updated systematic analysis with implications for the Sustainable Development Goals. Lancet. (2016) 388:3027–35. doi: 10.1016/S0140-6736(16)31593-8

PubMed Abstract | CrossRef Full Text | Google Scholar

2. World Health Organization, March of Dimes, The Partnership for Maternal, Newborn & Child Health. Born Too Soon: The Global Action Report on Preterm Birth. Geneva: World Health Organization (2012).

Google Scholar

3. Moster D, Lie RT, Markestad T. Long-term medical and social consequences of preterm birth. N Engl J Med. (2008) 359:262–73. doi: 10.1056/NEJMoa0706475

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Stensvold HJ, Klingenberg C, Stoen R, Moster D, Braekke K, Guthe HJ, et al. Neonatal morbidity and 1-year survival of extremely preterm infants. Pediatrics. (2017) 139:e20161821. doi: 10.1542/peds.2016-1821

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Institute of Medicine (US) Committee on Understanding Premature Birth and Assuring Healthy Outcomes. Mortality and acute complications in preterm infants. In: Behrman RE, Butler AS, editors. Preterm Birth: Causes, Consequences, and Prevention. Washington, DC: National Academies Press (2007). p. 313–46.

Google Scholar

6. Cohen-Wolkowiez M, Moran C, Benjamin DK, Cotten CM, Clark RH, Benjamin DK, et al. Early and late onset sepsis in late preterm infants. Pediatr Infect Dis J. (2009) 28:1052–6. doi: 10.1097/INF.0b013e3181acf6bd

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Paananen R, Husa AK, Vuolteenaho R, Herva R, Kaukola T, Hallman M. Blood cytokines during the perinatal period in very preterm infants: relationship of inflammatory response and bronchopulmonary dysplasia. J Pediatr. (2009) 154:39–43.e3. doi: 10.1016/j.jpeds.2008.07.012

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Lal CV, Ambalavanan N. Biomarkers, early diagnosis, and clinical predictors of bronchopulmonary dysplasia. Clin Perinatol. (2015) 42:739–54. doi: 10.1016/j.clp.2015.08.004

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Zea-Vera A, Ochoa TJ. Challenges in the diagnosis and management of neonatal sepsis. J Trop Pediatr. (2015) 61:1–13. doi: 10.1093/tropej/fmu079

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Niemarkt HJ, de Meij TG, van de Velde ME, van der Schee MP, van Goudoever JB, Kramer BW, et al. Necrotizing enterocolitis: a clinical review on diagnostic biomarkers and the role of the intestinal microbiota. Inflamm Bowel Dis. (2015) 21:436–44. doi: 10.1097/MIB.0000000000000184

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Berkhout DJC, Niemarkt HJ, Benninga MA, Budding AE, van Kaam AH, Kramer BW, et al. Development of severe bronchopulmonary dysplasia is associated with alterations in fecal volatile organic compounds. Pediatr Res. (2018) 83:412–9. doi: 10.1038/pr.2017.268

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Rogosch T, Herrmann N, Maier RF, Domann E, Hattesohl A, Koczulla AR, et al. Detection of bloodstream infections and prediction of bronchopulmonary dysplasia in preterm neonates with an electronic nose. J Pediatr. (2014) 165:622–4. doi: 10.1016/j.jpeds.2014.04.049

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Berkhout DJC, van Keulen BJ, Niemarkt HJ, Bessem JR, de Boode WP, Cossey V, et al. Late-onset sepsis in preterm infants can be detected preclinically by fecal volatile organic compound analysis: a prospective, multicenter cohort study. Clin Infect Dis. (2018) 68:70–7. doi: 10.1093/cid/ciy383

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Berkhout DJC, Niemarkt HJ, Buijck M, van Weissenbruch MM, Brinkman P, Benninga MA, et al. Detection of sepsis in preterm infants by fecal volatile organic compounds analysis: a proof of principle study. J Pediatr Gastroenterol Nutr. (2017) 65:e47–52. doi: 10.1097/MPG.0000000000001471

PubMed Abstract | CrossRef Full Text | Google Scholar

15. de Meij TG, van der Schee MP, Berkhout DJ, van de Velde ME, Jansen AE, Kramer BW, et al. Early detection of necrotizing enterocolitis by fecal volatile organic compounds analysis. J Pediatr. (2015) 167:562–7.e1. doi: 10.1016/j.jpeds.2015.05.044

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Garner CE, Ewer AK, Elasouad K, Power F, Greenwood R, Ratcliffe NM, et al. Analysis of faecal volatile organic compounds in preterm infants who develop necrotising enterocolitis: a pilot study. J Pediatr Gastroenterol Nutr. (2009) 49:559–65. doi: 10.1097/MPG.0b013e3181a3bfbc

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Cumeras R, Figueras E, Davis CE, Baumbach JI, Gràcia I. Review on ion mobility spectrometry. Part 1: current instrumentation. Analyst. (2015) 140:1376–90. doi: 10.1039/c4an01100g

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Armenta S, Alcala M, Blanco M. A review of recent, unconventional applications of ion mobility spectrometry (IMS). Anal Chim Acta. (2011) 703:114–23. doi: 10.1016/j.aca.2011.07.021

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Handa H, Usuba A, Maddula S, Baumbach JI, Mineshita M, Miyazawa T. Exhaled breath analysis for lung cancer detection using ion mobility spectrometry. PLoS ONE. (2014) 9:e114555. doi: 10.1371/journal.pone.0114555

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Westhoff M, Litterst P, Freitag L, Urfer W, Bader S, Baumbach JI. Ion mobility spectrometry for the detection of volatile organic compounds in exhaled breath of patients with lung cancer: results of a pilot study. Thorax. (2009) 64:744–8. doi: 10.1136/thx.2008.099465

PubMed Abstract | CrossRef Full Text | Google Scholar

21. Westhoff M, Litterst P, Maddula S, Bödeker B, Rahmann S, Davies AN, et al. Differentiation of chronic obstructive pulmonary disease (COPD) including lung cancer from healthy control group by breath analysis using ion mobility spectrometry. Int J Ion Mobil Spec. (2010) 13:131–9. doi: 10.1007/s12127-010-0049-2

CrossRef Full Text | Google Scholar

22. Westhoff M, Litterst P, Freitag L, Baumbach JI. Ion mobility spectrometry in the diagnosis of sarcoidosis: results of a feasibility study. J Physiol Pharmacol. (2007) 58 (Suppl. 5):739–51.

PubMed Abstract | Google Scholar

23. Maddula S, Blank LM, Schmid A, Baumbach JI. Detection of volatile metabolites of Escherichia coli by multi capillary column coupled ion mobility spectrometry. Anal Bioanal Chem. (2009) 394:791–800. doi: 10.1007/s00216-009-2758-0

PubMed Abstract | CrossRef Full Text | Google Scholar

24. Kunze N, Göpel J, Kuhns M, Jünger M, Quintel M, Perl T. Detection and validation of volatile metabolic patterns over different strains of two human pathogenic bacteria during their growth in a complex medium using multi-capillary column-ion mobility spectrometry (MCC-IMS). Appl Microbiol Biotechnol. (2013) 97:3665–76. doi: 10.1007/s00253-013-4762-8

PubMed Abstract | CrossRef Full Text | Google Scholar

25. Maddula S, Rabis T, Sommerwerck U, Anhenn O, Darwiche K, Freitag L, et al. Correlation analysis on data sets to detect infectious agents in the airways by ion mobility spectrometry of exhaled breath. Int J Ion Mobil Spec. (2011) 14:197–206. doi: 10.1007/s12127-011-0076-7

CrossRef Full Text | Google Scholar

26. Kuhn P, Astruc D, Messer J, Marlier L. Exploring the olfactory environment of premature newborns: a French survey of health care and cleaning products used in neonatal units. Acta Paediatr. (2011) 100:334–9. doi: 10.1111/j.1651-2227.2010.02076.x

PubMed Abstract | CrossRef Full Text | Google Scholar

27. Hsieh S, Sapkota A, Wood R, Bearer C, Kapoor S. Neonatal ethanol exposure from ethanol-based hand sanitisers in isolettes. Arch Dis Child Fetal Neonatal Ed. (2018) 103:F55–8. doi: 10.1136/archdischild-2016-311959

PubMed Abstract | CrossRef Full Text | Google Scholar

28. Ulanowska A, Ligor M, Amann A, Buszewski B. Determination of volatile organic compounds in exhaled breath by ion mobility spectrometry. Chem Anal. (2008) 53:953–65.

Google Scholar

29. Borsdorf H, Eiceman GA. Ion mobility spectrometry: principles and applications. Appl Spectrosc Rev. (2006) 41:323–75. doi: 10.1080/05704920600663469

CrossRef Full Text | Google Scholar

30. Cumeras R, Figueras E, Davis CE, Baumbach JI, Gràcia I. Review on ion mobility spectrometry. Part 2: hyphenated methods and effects of experimental parameters. Analyst. (2015) 140:1391–410. doi: 10.1039/c4an01101e

PubMed Abstract | CrossRef Full Text | Google Scholar

31. Fink T, Baumbach JI, Kreuer S. Ion mobility spectrometry in breath research. J Breath Res. (2014) 8:27104. doi: 10.1088/1752-7155/8/2/027104

PubMed Abstract | CrossRef Full Text | Google Scholar

32. Bödeker B, Baumbach JI. Analytical description of IMS-signals. Int J Ion Mobil Spec. (2009) 12:103–8. doi: 10.1007/s12127-009-0024-y

CrossRef Full Text | Google Scholar

33. Furtwängler R, Hauschild A-C, Hübel J, Rakicioglou H, Bödeker B, Maddula S, et al. Signals of neutropenia in human breath? Int J Ion Mobil Spec. (2014) 17:19–23. doi: 10.1007/s12127-014-0145-9

CrossRef Full Text | Google Scholar

34. Hauschild AC, Baumbach JI, Baumbach J. Integrated statistical learning of metabolic ion mobility spectrometry profiles for pulmonary disease identification. Genet Mol Res. (2012) 11:2733–44. doi: 10.4238/2012.July.10.17

PubMed Abstract | CrossRef Full Text | Google Scholar

35. Westhoff M, Litterst P, Maddula S, Bödeker B, Baumbach JI. Statistical and bioinformatical methods to differentiate chronic obstructive pulmonary disease (COPD) including lung cancer from healthy control by breath analysis using ion mobility spectrometry. Int J Ion Mobil Spec. (2011) 14:139–49. doi: 10.1007/s12127-011-0081-x

CrossRef Full Text | Google Scholar

36. Smolinska A, Hauschild AC, Fijten RR, Dallinga JW, Baumbach J, van Schooten FJ. Current breathomics–a review on data pre-processing techniques and machine learning in metabolomics breath analysis. J Breath Res. (2014) 8:27105. doi: 10.1088/1752-7155/8/2/027105

PubMed Abstract | CrossRef Full Text | Google Scholar

37. Christiansen A, Davidsen JR, Titlestad I, Vestbo J, Baumbach J. A systematic review of breath analysis and detection of volatile organic compounds in COPD. J Breath Res. (2016) 10:34002. doi: 10.1088/1752-7155/10/3/034002

PubMed Abstract | CrossRef Full Text | Google Scholar

38. Neerincx AH, Vijverberg SJH, Bos LDJ, Brinkman P, van der Schee MP, de Vries R, et al. Breathomics from exhaled volatile organic compounds in pediatric asthma. Pediatr Pulmonol. (2017) 52:1616–27. doi: 10.1002/ppul.23785

PubMed Abstract | CrossRef Full Text | Google Scholar

39. Darwiche K, Baumbach JI, Sommerwerck U, Teschler H, Freitag L. Bronchoscopically obtained volatile biomarkers in lung cancer. Lung. (2011) 189:445–52. doi: 10.1007/s00408-011-9324-1

PubMed Abstract | CrossRef Full Text | Google Scholar

40. Fink T, Wolf A, Maurer F, Albrecht FW, Heim N, Wolf B, et al. Volatile organic compounds during inflammation and sepsis in rats: a potential breath test using ion-mobility spectrometry. Anesthesiology. (2015) 122:117–26. doi: 10.1097/ALN.0000000000000420

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: ion mobility spectrometry, neonatal incubator, preterm infant, volatile organic compound, breath analysis

Citation: Steinbach J, Goedicke-Fritz S, Tutdibi E, Stutz R, Kaiser E, Meyer S, Baumbach JI and Zemlin M (2019) Bedside Measurement of Volatile Organic Compounds in the Atmosphere of Neonatal Incubators Using Ion Mobility Spectrometry. Front. Pediatr. 7:248. doi: 10.3389/fped.2019.00248

Received: 01 March 2019; Accepted: 30 May 2019;
Published: 18 June 2019.

Edited by:

Karel Allegaert, University Hospitals Leuven, Belgium

Reviewed by:

Tim De Meij, VU University Medical Center, Netherlands
Aakash Pandita, Sanjay Gandhi Post Graduate Institute of Medical Sciences, India

Copyright © 2019 Steinbach, Goedicke-Fritz, Tutdibi, Stutz, Kaiser, Meyer, Baumbach and Zemlin. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Sybelle Goedicke-Fritz, sybelle.goedicke-fritz@uks.eu

Co-first authors

Present Address: Jörg Ingo Baumbach, B. Braun Melsungen AG Center of Competence Breath Analysis, Dortmund, Germany

Download