Multimodality 3D image fusion with live fluoroscopy reduces radiation dose during catheterization of congenital heart defects

  • 1. Department of Human Structure and Repair, Ghent University, Ghent, Belgium

  • 2. Department of Paediatric Cardiology, Ghent University Hospital, Ghent, Belgium

  • 3. Department of Cardiology, Ghent University Hospital, Ghent, Belgium

  • 4. Department of Paediatric Cardiology, Brussels University Hospital, Jette, Belgium

Abstract

Introduction:

Imaging fusion technology is promising as it is radiation and contrast sparing. Herein, we compare conventional biplane angiography to multimodality image fusion with live fluoroscopy using two-dimensional (2D)–three-dimensional (3D) registration (MMIF2D−3D) and assess MMIF2D−3D impact on radiation exposure and contrast volume during cardiac catheterization of patients with congenital heart disease (CHD).

Methods:

We matched institutional MMIF2D−3D procedures and controls according to patient characteristics (body mass index, age, and gender) and the seven procedure-type subgroups. Then, we matched the number of tests and controls per subgroup using chronological ordering or propensity score matching. Subsequently, we combined the matched subgroups into larger subgroups of similar procedure type, keeping subgroups with at least 10 test and 10 control cases. Air kerma (AK) and dose area product (DAP) were normalized by body weight (BW), product of body weight and fluoroscopy time (BW × FT), or product of body weight and number of frames (BW × FR), and stratified by acquisition plane and irradiation event type (fluoroscopy or acquisition). Three senior interventionists evaluated the relevance of MMIF2D−3D (5-point Likert scale).

Results:

The Overall group consisted of 54 MMIF2D−3D cases. The combined and matched subgroups were pulmonary artery stenting (StentPUL), aorta angioplasty (PlastyAO), pulmonary artery angioplasty (PlastyPUL), or a combination of the latter two (Plasty). The FT of the lateral plane reduced significantly by 69.6% for the Overall MMIF2D−3D population. AKBW and DAPBW decreased, respectively, by 43.9% and 39.3% (Overall group), 49.3% and 54.9% (PlastyAO), and 36.7% and 44.4% for the Plasty subgroup. All the aforementioned reductions were statistically significant except for DAPBW in the Overall and Plasty (sub)groups. The decrease of AKBW and DAPBW in the StentPUL and PlastyPUL subgroups was not statistically significant. The decrease in the median values of the weight-normalized contrast volume (CMCBW) in all five subgroups was not significant. Cardiologists considered MMIF2D−3D very useful with a median score of 4.

Conclusion:

In our institution, MMIF2D−3D overall enabled significant AKBW reduction during the catheterization of CHD patients and was mainly driven by reduced FT in the lateral plane. We observed significant AKBW reduction in the Plasty and PlastyAO subgroups and DAPBW reduction in the PlastyAO subgroup. However, the decrease in CMCBW was not significant.

Introduction

Congenital heart disease (CHD) is the most common birth defect (13). Cardiac catheterization has improved management of CHD patients, and owing to the increased reliability and advancements in technology, the number, types, and complexity of cardiac catheterizations have increased dramatically in recent years (47). In addition, multiple staged or repeated procedures are required during the lifetime of patients with complex CHD lesions, which significantly increases their cumulative radiation exposure (1, 8, 9). For a given radiation dose, children are three to four times more susceptible to radiation-induced malignancies. Moreover, their smaller body size results in less attenuation, and therefore relatively higher doses for deeper organs, and a relatively larger fraction of their body is exposed to the direct x-ray beam (1014). Furthermore, operators are often near the patient, thereby increased patient exposure results in increased occupational exposure. Consequently, optimizing radiation exposure of cardiac catheterization in CHD is needed. Image fusion of a three-dimensional (3D) reconstruction of volumetric imaging datasets with live fluoroscopy has the potential to reduce iodinated contrast volume and radiation dose (1517). In the last decade, 3D rotational angiography image fusion (3DRA-IF) has been more frequently used in CHD, where a 3D roadmap is obtained from a tomographic acquisition, which is subsequently overlaid on the live fluoroscopy images acquired by the same x-ray modality (16, 1822).

More recently, multimodality image fusion (MMIF) has become available. With the latter technology, a 3D roadmap can be obtained from preprocedural imaging like computed tomography (CT) or magnetic resonance imaging (MRI) (15, 2326). After segmentation, the 3D roadmap is registered with live fluoroscopy using either two fluoroscopy or cinegraphy acquisitions in different projections, i.e., 2D–3D registration (MMIF2D−3D), or using a 3DRA acquisition, i.e., 3D–3D registration (MMIF3D−3D) (26). With MMIF2D−3D only two fluoroscopy exposures are sufficient to perform registration with the volumetric dataset, resulting in nearly radiation-free registration (24, 26, 27).

The goal of the current study is to assess the impact of MMIF2D−3D on radiation exposure and contrast volume levels during cardiac catheterization in CHD patients.

Methods

The procedural and patient data included in this study were acquired from the institutional database on congenital cardiac catheterizations performed between January 2015 and October 2019. Structural interventions such as closure of atrial septal defects (ASD), ventricular septal defects (VSD), patent foramen ovale (PFO) and patent ductus arteriosus (PDA) were excluded since they were ultrasound guided, without the use of ultrasound-fluoroscopy image fusion. All procedures were performed with a biplane AlluraClarity FD20/FD10 system (Philips Healthcare, Best, The Netherlands). The diagonal of the frontal imaging detector was double the size of the lateral imaging detector. The imaging protocols available on the biplane x-ray system are split by weight in two groups, i.e., ≤40 and >40 kg. In both weight groups, the corresponding lowest dose imaging protocols are selected by default. The operators are used to working with the lowest dose settings and switch to higher dose settings only if they deem it necessary.

Multimodality image fusion

During the study period, we applied MMIF2D−3D when CT-scan or MRI imaging of sufficient quality was available for 3D reconstruction and used the conventional biplane 2D angiography (2DA) in the absence of preprocedural imaging. The MMIF2D−3D approach consisted of a four-step process and was performed using the VesselNavigator software (Philips Healthcare, Best, The Netherlands) that has been described previously in the literature (28). Briefly, the anatomy of interest was segmented. Then, the anatomical landmarks visible on both live fluoroscopy and volumetric datasets (CT or MRI) were indicated on the volumetric dataset. Afterward, two short 2D fluoroscopy projections were performed to match the position of the preprocedural 3D volume with the periprocedural anatomical position. The final step is live guidance of the procedure with the registered 3D volume transparently overlaid on top of the live fluoroscopy, during which the registration can be optimized further. The process is visualized in Figure 1, showcasing one aortic coarctation MMIF2D−3D case.

Figure 1

Matching

To reduce bias, we applied a multiple step matching process based on procedure type and patient characteristics [body mass index (BMI), age, and gender]. Initially, we subdivided the MMIF2D−3D procedures into seven subgroups: diagnostic, balloon dilatations in the aorta (BalloonAO), balloon dilatations in the pulmonary artery (BalloonPUL), stenting in the aorta (StentAO), stenting in the pulmonary artery (StentPUL), percutaneous pulmonary valve replacements (PPVR), and embolization. Subsequently, for each test group we selected a corresponding control subgroup from conventional CHD procedures of the same type performed during that time period. If the difference in the number of cases between the test and control subgroups (Δn) was less than 5, then we considered these subgroups as matched. When Δn ≥ 5 and the number of cases in the test subgroup (nTest) was larger than 5, we conducted a propensity score matching of 1:1 on BMI, age, gender, and number of treated lesions (considered only for angioplasty procedures). In addition, if nTest < 5, we selected the latest nTest cases from the chronologically ordered control subgroup as the matched control subgroup. After subgroup matching, we combined the subgroups of balloon dilatation and stenting into angioplasty subgroups for each respective region (PlastyAO and PlastyPUL) before finally combining all subgroups into one Overall group. Eventually, only the subgroups with a minimum of 10 procedures each in both the test and control cohorts were included for statistical analysis. The matching process is illustrated in Figure 2.

Figure 2

Radiation exposure data

We collected radiation exposure parameters from the Digital Imaging and Communications in Medicine (DICOM) Radiation Dose Structured Reports (RDSR) generated by the system at the end of each procedure. The parameters included cumulative dose area product (DAP), cumulative air kerma (AK), fluoroscopy time (FT), number of cinegraphy acquisitions, and number of cinegraphy frames (FR). Subsequently, DAP and AK were grouped by irradiation event type [fluoroscopy (F) and cinegraphy acquisition (A)], by acquisition plane [frontal (A) and lateral (B)], or by both. FT and FR were also grouped by acquisition plane.

As proposed by the European Guidelines on Diagnostic Reference Levels (DRLs) for Paediatric Imaging, we looked at DAP and AK normalized by body weight (DAPBW and AKBW) and by the product of body weight and fluoroscopy time (DAPBW × FT and AKBW × FT) (6). For radiation exposure associated only with cinegraphy acquisition, the product of body weight and the number of frames (BW × FR) was used for normalization (DAPABW × FR and AKABW × FR). All the collected and calculated radiation exposure parameters are summarized in Figure 3.

Figure 3

Questionnaire

The three senior experienced pediatric interventional cardiologists performing all interventions included in this study were questioned for subjective evaluations of the MMIF2D−3D cases. All catheterizations were reviewed on the Picture Archiving and Communication System (PACS) by the interventionists before filling out the three-question assessment. The first question was: “Assuming you would have performed this case using conventional biplane 2DA, are there any images missing? If yes, please state the number of missing acquisitions, their required projection angles, and if these would have been single- or bi-plane acquisitions”. We reported the number of missing angiographies as the number of spared acquisitions by the MMIF2D−3D guidance. Interventionists follow an institutional standard protocol concerning contrast media consumption (CMC) per BW administered for each acquisition. Therefore, they were asked to estimate the spared CMCBW associated with the spared acquisitions. Subsequently, screenshots of the 3D overlay registered with the fluoroscopy were shown next to the angiography images on the PACS and the second question was: “How many angiographies would you consider redundant due to the availability of the MMIF2D−3D guidance?” They were also asked to estimate the CMC associated with each redundant acquisition. The third question was: “Grade the relevance of MMIF2D−3D guidance using the following Likert scale: 1 = Misleading, 2 = Not Useful, 3 = Useful, 4 = Very Useful, 5 = Essential”.

Statistical analysis

For normally distributed variables, significance testing of differences between MMIF2D−3D and conventional 2DA was performed using an independent samples t-test with a 95% two-sided CI. A Mann–Whitney U test was performed in case of non-normality. The variables were summarized using median, 25th percentile (Q1), and 75th percentile (Q3). Comparison of categorical parameters was done using a χ2 test for association, also applied with a 95% CI. Interobserver agreement is assessed using Cohen's kappa. All statistical analyses were performed using IBM SPSS Statistics 28 (IBM Corp., Armonk, NY, USA).

Results

Between January 2015 and October 2019, a total of 420 catheterizations were conducted. Among these procedures, structural interventions for conditions such as ASD, VSD, PFO, and PDA accounted for 153 cases. Of the 267 remaining procedures, 54 and 213 were linked to the test (MMIF2D−3D-guided catheterizations) and control groups, respectively. The following results only concern significant differences, unless explicitly mentioned otherwise.

Matching

Figure 2C presents the outcome of the matching process. After initial matching, a subsequent matching phase was conducted for the Diagnostic, BalloonAO and BalloonPUL subgroups. The Diagnostic and BalloonPUL subgroups underwent propensity score matching, while chronological ordering was used for BalloonAO. Five procedure type (sub)groups were selected for analysis after completing the multiphase matching process: StentPUL, PlastyAO, PlastyPUL, Plasty, and the Overall group.

Patient and procedural characteristics

Patient characteristics are summarized using median and interquartile range (IQR) in Table 1 for the test (MMIF2D−3D) and control (Conventional 2DA) cohorts. Table 2 presents cumulative radiation exposure and CMC parameters, together with their corresponding parameters normalized by BW and BW × FT, for all the five (sub)groups. A boxplot representation excluding outliers of Tables 1, 2 is shown in Figure 4. Finally, detailed results of parameter grouping based on irradiation event type and acquisition plane for the Overall group are listed in Table 3. Supplementary Tables S1–S4 provide a similar overview of the parameters grouped by irradiation event and acquisition plane for the matched subgroups.

Table 1

Variable (unit)Conventional 2DAMMIF2D−3Dp-value
GroupOverall (n2DA = 58; nMMIF = 54)
 Weight (kg)44.0 (13.0–73.0)49.0 (21.25–66.5)0.637
 BMI (kg m−2)18.262 (15.609–23.504)19.181 (16.525–23.616)0.438
 Age (years)12.737 (2.817–28.353)15.259 (7.938–26.555)0.217
GroupStentPUL (n2DA = 10; nMMIF = 10)
 Weight (kg)20.5 (11.4–54.875)35.5 (20.5–65.75)0.212
 BMI (kg m−2)17.257 (14.836–22.393)18.115 (15.572–22.315)0.597
 Age (years)6.61 (2.817–14.205)12.027 (9.634–22.125)0.174
GroupPlastyPUL (n2DA = 19; nMMIF = 19)
 Weight (kg)16.4 (9.2–61.5)38.0 (18.6–57.5)0.199
 BMI (kg m−2)16.225 (15.05–22.923)18.553 (16.481–20.051)0.466
 Age (years)4.189 (2.032–25.189)13.093 (7.419–20.745)0.148
GroupPlastyAO (n2DA = 16; nMMIF = 12)
 Weight (kg)55.5 (6.5–74.0)51.0 (39.125–63.25)0.889
 BMI (kg m−2)19.927 (15.591–23.587)18.501 (17.238–21.491)0.889
 Age (years)14.636 (0.434–20.99)14.485 (10.771–19.851)0.610
GroupPlasty (n2DA = 35; nMMIF = 31)
 Weight (kg)39.0 (7.85–73.0)48.0 (26.0–62.0)0.362
 BMI (kg m−2)18.321 (15.05–23.388)18.553 (16.787–21.006)0.684
 Age (years)11.345 (1.108–21.06)13.293 (9.081–20.215)0.131

Patient characteristics for the Overall group and for four matched subgroups: stenting procedures of the pulmonary artery (StentPUL), angioplasty procedures (balloon dilatation and/or stenting) of the aorta (PlastyAO), angioplasty procedures of the pulmonary artery (PlastyPUL), and angioplasty procedures of the aorta and/or pulmonary artery (Plasty).

n2DA, the number of procedures in the matched control group for conventional 2DA; nMMIF, the number of procedures in the test group MMIF2D−3D.

Bold text indicates a lower median, Q1, or Q3 value in the MMIF2D−3D group compared with the corresponding value encountered in the conventional 2DA group.

Table 2

GroupVariable (Unit)Conventional 2DAMMIF2D−3Dp-value
Overall
(n2DA = 58; nMMIF = 54)
Acquisitions (#)15.5 (7.25–21.5)12.0 (7.25–18.0)0.622
FR (#)792.0 (423.5–1,221.75)671.0 (454.5–942.0)0.507
FT (min)19.033 (11.054–27.779)14.285 (8.876–30.09)0.169
DAP (mGy cm2)13,050.336 (2,838.729–39,100.954)10,307.832 (4,163.399–22,888.763)0.731
AK (mGy)169.68 (52.329–405.202)102.297 (44.624192.634)0.188
CMC (ml)60.0 (30.0–100.0)61.0 (40.0–97.5)0.916
DAPBW (mGy cm2 kg−1)442.398 (233.96–733.217)268.577 (143.544545.399)0.052
AKBW (mGy kg−1)4.839 (3.012–8.969)2.713 (1.6915.516)0.003
CMCBW (ml kg−1)1.777 (1.214–2.985)1.633 (0.7692.843)0.193
DAPBW × FT (mGy cm2 kg−1 min−1)21.459 (15.14–30.037)17.818 (14.11326.849)0.264
AKBW × FT (mGy kg−1 min−1)15.709 (11.121–21.183)11.653 (8.96518.474)0.018
StentPUL
(n2DA = 10; nMMIF = 10)
Acquisitions (#)17.0 (14.25–20.0)19.0 (14.5–25.5)0.677
FR (#)828.5 (658.25–933.5)945.0 (742.5–1,169.5)0.496
FT (min)24.283 (19.775–30.292)30.746 (15.233–38.54)0.762
DAP (mGy cm2)14,167.111 (6,157.926–26,176.615)13,017.231 (9,135.732–21,216.559)0.650
AK (mGy)187.84 (110.477–237.212)139.687 (107.26190.142)0.705
CMC (ml)72.5 (30.0–96.25)54.0 (40.0–82.5)0.791
DAPBW (mGy cm2 kg−1)487.921 (381.192–985.04)459.986 (300.059880.857)0.705
AKBW (mGy kg−1)5.908 (4.331–12.288)5.09 (2.8538.802)0.364
CMCBW (ml kg−1)3.137 (2.213–3.629)1.669 (0.7243.403)0.307
DAPBW × FT (mGy cm2 kg−1 min−1)17.795 (14.794–30.741)18.881 (15.12–21.172)0.762
AKBW × FT (mGy kg−1 min−1)15.303 (13.793–30.407)10.759 (9.16916.726)0.131
PlastyPUL
(n2DA = 19; nMMIF = 19)
Acquisitions (#)14.0 (6.0–19.0)14.0 (7.5–22.0)0.438
FR (#)638.0 (445.0–890.0)900.0 (424.5–994.0)0.293
FT (min)21.5 (14.133–29.417)30.482 (15.322–42.506)0.493
DAP (mGy cm2)8,053.284 (2,905.003–25,738.685)11,485.319 (6,257.413–29,964.438)0.389
AK (mGy)157.483 (51.774–231.524)122.701 (76.205–225.079)0.872
CMC (ml)40.0 (30.0–92.5)50.0 (40.0–77.5)0.693
DAPBW (mGy cm2 kg−1)502.994 (253.158–785.399)390.228 (268.577–721.024)0.737
AKBW (mGy kg−1)4.959 (3.101–12.973)4.693 (2.6527.744)0.249
CMCBW (ml kg−1)2.439 (1.562–3.6)1.667 (0.7352.875)0.204
DAPBW × FT (mGy cm2 kg−1 min−1)19.605 (14.93–29.486)15.232 (12.05822.598)0.249
AKBW × FT (mGy kg−1 min−1)16.111 (14.152–28.6)10.481 (8.01913.761)0.006
PlastyAO
(n2DA = 16; nMMIF = 12)
Acquisitions (#)14.0 (9.0–22.5)13.5 (10.0–16.5)0.727
FR (#)823.0 (395.75–1,149.75)652.5 (471.5–758.0)0.318
FT (min)14.133 (6.729–17.608)9.077 (7.343–10.355)0.150
DAP (mGy cm2)16,657.327 (1,961.285–32,247.914)6,045.466 (4,136.604–12,019.472)0.458
AK (mGy)184.257 (23.737–349.611)67.496 (52.739–136.279)0.458
CMC (ml)90.0 (26.25–130.0)75.0 (60.5–100.0)0.816
DAPBW (mGy cm2 kg−1)296.874 (138.95–503.661)133.762 (109.28241.567)0.046
AKBW (mGy kg−1)3.867 (2.72–5.809)1.962 (1.2112.531)0.014
CMCBW (ml kg−1)1.777 (1.463–2.698)1.675 (0.9892.345)0.330
DAPBW × FT (mGy cm2 kg−1 min−1)22.811 (20.741–29.015)16.272 (14.59525.369)0.070
AKBW × FT (mGy kg−1 min−1)18.586 (14.865–21.707)12.896 (9.85916.649)0.051
Plasty
(n2DA = 35; nMMIF = 31)
Acquisitions (#)14.0 (6.0–20.0)14.0 (9.0–18.0)0.718
FR (#)719.0 (411.5–1,022.0)688.0 (461.0–933.0)0.949
FT (min)18.333 (8.708–25.85)14.178 (8.884–31.784)0.974
DAP (mGy cm2)8,829.225 (2,527.159–30,306.158)10,007.708 (4,758.84–16,917.954)0.802
AK (mGy)157.483 (46.572–313.083)102.113 (60.661–177.917)0.500
CMC (ml)70.0 (30.0–105.0)60.0 (44.0–95.0)0.898
DAPBW (mGy cm2 kg−1)457.349 (218.506–624.543)274.384 (133.762535.505)0.225
AKBW (mGy kg−1)4.72 (2.968–7.42)2.988 (1.775.421)0.037
CMCBW (ml kg−1)2.062 (1.494–3.438)1.667 (0.7832.636)0.114
DAPBW × FT (mGy cm2 kg−1 min−1)21.455 (16.82–29.486)15.739 (13.05324.42)0.037
AKBW × FT (mGy kg−1 min−1)17.385 (14.862–23.844)11.036 (8.54915.915)0.001

Procedural characteristics, i.e., cumulative radiation exposure parameters and contrast media usage, for the Overall group and for four matched subgroups: stenting procedures of the pulmonary artery (StentPUL), angioplasty procedures (balloon dilatation and/or stenting) of the aorta (PlastyAO), angioplasty procedures of the pulmonary artery (PlastyPUL), and angioplasty procedures of the aorta and/or pulmonary artery (Plasty).

n2DA, the number of procedures in the matched control group for conventional 2DA; nMMIF, the number of procedures in the test group MMIF2D−3D; FR, number of acquisition frames; FT, fluoroscopy time; DAP, dose area product; AK, air kerma; CMC, contrast media consumption; DAPBW, dose area product normalized by body weight; AKBW, air kerma normalized by body weight; CMCBW, contrast media consumption normalized by body weight; DAPBW × FT, dose area product normalized by the product of body weight and fluoroscopy time; AKBW × FT, air kerma normalized by the product of body weight and fluoroscopy time.

Bold text indicates a lower median, Q1, or Q3 value in the MMIF2D−3D group compared with the corresponding value encountered in the conventional 2DA group. Rows with a gray background indicate p < 0.05 and therefore statistical significance.

Figure 4

Table 3

Variable (Unit)Conventional 2DA (n = 58)MMIF2D−3D (n = 54)p-value
DAP(A) (mGy cm2)7,308.795 (1,896.989–25,287.69)6,309.346 (2,684.514–16,094.01)0.771
DAP(B) (mGy cm2)3,077.668 (510.902–13,862.318)1,625.287 (593.289–5,090.852)0.183
AK(A) (mGy)73.266 (21.738–201.212)54.738 (29.928–141.209)0.736
AK(B) (mGy)52.559 (12.534–176.807)27.553 (9.28574.665)0.069
DAPF (mGy cm2)9,324.848 (2,304.053–26,872.037)7,744.786 (2,566.527–18,033.609)0.736
DAPF(A) (mGy cm2)5,477.852 (1,513.517–21,054.668)5,464.426 (2,268.587–14,404.911)0.625
DAPF(B) (mGy cm2)1,513.162 (231.323–6,200.844)678.634 (97.7732,192.279)0.061
AKF (mGy)117.267 (45.291–249.087)76.248 (29.58164.302)0.130
AKF(A) (mGy)58.701 (18.129–153.391)49.306 (27.514–127.251)0.991
AKF(B) (mGy)28.591 (3.666–83.415)9.945 (1.70929.81)0.022
DAPA (mGy cm2)2,356.695 (358.637–13,457.281)2,056.087 (932.948–5,333.384)0.958
DAPA(A) (mGy cm2)949.163 (188.26–4,314.737)936.341 (408.824–2,260.488)0.893
DAPA(B) (mGy cm2)641.767 (131.893–6,498.645)933.456 (294.006–3,158.306)0.935
AKA (mGy)23.802 (6.591–129.816)23.839 (11.112–56.454)0.816
AKA(A) (mGy)8.655 (2.606–26.268)7.281 (3.815–15.565)0.650
AKA(B) (mGy)11.05 (3.177–89.176)13.559 (4.548–47.648)0.981
FT(A)903.5 (446.25–1,351.5)758.266 (426.171–1,586.164)0.958
FT(B)159.5 (42.5–484.75)48.461 (14.531182.944)0.006
FR(A)419.5 (220.0–637.0)367.5 (242.5–503.0)0.614
FR(B)388.0 (187.25–596.5)323.0 (206.0–464.25)0.505
DAP(A)BW (mGy cm2 kg−1)228.953 (111.164–478.59)194.924 (89.853463.274)0.522
DAP(B)BW (mGy cm2 kg−1)124.627 (40.956–237.84)55.681 (23.987103.098)0.005
AK(A)BW (mGy kg−1)2.422 (1.314–4.025)1.754 (0.884–4.415)0.164
AK(B)BW (mGy kg−1)1.988 (0.878–3.454)0.841 (0.3751.7)0.001
DAPFBW (mGy cm2 kg−1)287.961 (161.619–546.347)205.032 (96.407476.367)0.077
DAPF(A)BW (mGy cm2 kg−1)209.729 (79.088–407.343)168.418 (74.944–436.627)0.709
DAPF(B)BW (mGy cm2 kg−1)54.302 (13.682–185.212)16.363 (3.5546.592)0.004
AKFBW (mGy kg−1)3.462 (1.915–5.88)1.868 (1.0824.843)0.007
AKF(A)BW (mGy kg−1)1.712 (0.918–3.555)1.555 (0.683–4.225)0.319
AKF(B)BW (mGy kg−1)0.924 (0.218–2.907)0.262 (0.0730.718)0.002
DAPABW (mGy cm2 kg−1)62.965 (30.324–192.92)46.051 (27.631109.453)0.249
DAPA(A)BW (mGy cm2 kg−1)30.142 (13.969–64.53)20.396 (14.343–40.217)0.256
DAPA(B)BW (mGy cm2 kg−1)26.207 (12.994–114.462)24.265 (13.009–62.499)0.740
AKABW (mGy kg−1)1.157 (0.487–2.072)0.619 (0.3551.313)0.048
AKA(A)BW (mGy kg−1)0.305 (0.172–0.533)0.206 (0.1410.331)0.036
AKA(B)BW (mGy kg−1)0.542 (0.263–1.33)0.408 (0.2130.985)0.289
DAP(A)BW × FT (mGy cm2 kg−1 min−1)17.569 (12.843–23.272)14.661 (11.10119.876)0.145
DAP(B)BW × FT (mGy cm2 kg−1 min−1)36.096 (23.8–70.802)59.847 (21.43–156.891)0.203
AK(A)BW × FT (mGy kg−1 min−1)0.163 (0.125–0.234)0.138 (0.1030.191)0.046
AK(B)BW × FT (mGy kg−1 min−1)0.717 (0.391–1.193)0.947 (0.367–3.043)0.501
DAPFBW × FT (mGy cm2 kg−1 min−1)15.622 (10.479–21.753)13.085 (10.499–20.162)0.244
DAPF(A)BW × FT (mGy cm2 kg−1 min−1)13.523 (9.806–19.966)11.863 (9.30317.781)0.481
DAPF(B)BW × FT (mGy cm2 kg−1 min−1)21.141 (14.476–28.528)17.937 (11.736–31.348)0.688
AKFBW × FT (mGy kg−1 min−1)0.198 (0.125–0.246)0.137 (0.1030.19)0.008
AKF(A)BW × FT (mGy kg−1 min−1)0.124 (0.095–0.198)0.125 (0.080.17)0.259
AKF(B)BW × FT (mGy kg−1 min−1)0.328 (0.252–0.477)0.278 (0.189–0.478)0.176
DAPABW × FR (mGy cm2 kg−1 fr−1)0.098 (0.064–0.164)0.075 (0.0550.126)0.129
DAPA(A)BW × FR (mGy cm2 kg−1 fr−1)0.076 (0.053–0.128)0.06 (0.0440.088)0.053
DAPA(B)BW × FR (mGy cm2 kg−1 fr−1)0.11 (0.064–0.167)0.099 (0.057–0.187)0.629
AKABW × FR (×10−3 mGy kg−1 fr−1)1.313 (1.072–2.151)1.076 (0.6931.557)0.006
AKA(A)BW × FR (×10−3 mGy kg−1 fr−1)0.843 (0.488–1.162)0.569 (0.3490.828)0.008
AKA(B)BW × FR (×10−3 mGy kg−1 fr−1)1.811 (1.404–2.821)1.521 (0.9882.789)0.144

Radiation exposure parameters grouped by irradiation event type and acquisition plane, listed as median (Q1–Q3) values for the Overall group.

DAP, dose area product; AK, air kerma; CMC, contrast media consumption; DAPBW, dose area product normalized by body weight; AKBW, air kerma normalized by body weight; CMCBW, contrast media consumption normalized by body weight; DAPBW × FT, dose area product normalized by the product of body weight and fluoroscopy time; AKBW × FT, air kerma normalized by the product of body weight and fluoroscopy time.

Bold text indicates a lower median, Q1, or Q3 value in the MMIF2D−3D group compared with the corresponding value encountered in the conventional 2DA group. Rows with a gray background indicate p < 0.05 and therefore statistical significance. The first superscript, i.e., either F or A, indicates grouping by fluoroscopy or acquisition, respectively. A missing first superscript refers to the overall (F + A) DAP and AK. The second superscript between brackets, i.e., either (A) or (B), indicates grouping by the frontal or lateral plane, respectively. A missing second superscript refers to the overall [(A) + (B)] DAP and AK of both acquisition planes combined. A missing subscript *** refers to the overall DAP and AK without normalization.

Overall group

In terms of patient characteristics, there were no significant differences observed among the Overall group. FT(B) was reduced by 69.6%, resulting in a 71.7% and 69.9% decrease of AKF(B)BW and DAPF(B)BW and consequently a reduction of AK(B)BW (57.7%), DAP(B)BW (55.3%), AKF(B) (65.2%), and AKFBW (46.1%). Furthermore, the MMIF2D−3D group demonstrated lower values for AKA(A)BW (32.4%) and AKABW (46.6%), contributing to an overall decrease in AKBW (43.9%). DAPBW decreased by 39.3%. However, this difference was not statistically significant (p = 0.052). AKFBW × FT and AKA(A)BW × FR exhibited significantly lower values, resulting in a decreased cumulative AKBW × FT.

Procedure-type subgroups

Significant differences in patient characteristics were not observed among any of the subgroups. In terms of radiation exposure quantities, similar findings to that of the Overall group were noted for the Plasty subgroup, with a 36.7% reduction in AKBW. Similarly, the PlastyAO subgroup showed reductions of 49.3% and 54.9% in AKBW and DAPBW, respectively. Both the PlastyAO and Plasty subgroups exhibited decreases in AKFBW × FT and AKA(A)BW × FR, while AKA(B)BW × FR and eventually AKABW × FR were only lower for the Plasty subgroup. These results led to a notable decrease in both AKBW × FT and DAPBW × FT within both the PlastyAO and Plasty subgroups.

In the remaining two subgroups, StentPUL and PlastyPUL, only some BW× FT or BW × FR normalized quantities showed significantly lower values. For PlastyPUL, there were eight exposure quantities showing significant reductions. These included AKF(A)BW × FT, AKF(B)BW × FT, and AKFBW × FT as well as their associated acquisition counterparts AKA(A)BW × FR, AKA(B)BW × FR, and AKABW × FR. In addition, AK(A)BW × FT and AKBW × FT also exhibited significantly lower values in this subgroup. For StentPUL, only AKF(B)BW × FT showed significantly lower values in the test cohort. Although patient characteristics were not significantly different, it is worth noting that patients were, respectively, 1.73 and 1.82 times heavier and 2.32 and 3.13 times older in the StentPUL and PlastyPUL subgroups.

Questionnaire

Question 1—spared acquisitions

Of the 54 cases, at least one spared acquisition compared with the conventional workflow was reported in 17 cases (31.5% of cases). In these cases, a median of 2.0 (IQR, 2.0–3.0) spared acquisitions were reported, which were considered to be mostly biplane. The cardiologists estimated a corresponding median saved CMC of 18.3 (IQR, 10.0–21.7) ml and a median CMC normalized to body weight of 0.82 (IQR, 0.54–1.40) ml kg−1.

Question 2—redundant acquisitions

Cardiologists reported a median of 2.0 (IQR, 2.0–3.0) redundant acquisitions in 27 procedures (50.0% of cases). The associated median redundant CMC was estimated at 30.0 (IQR, 23.5–35.8) ml, while the median redundant normalized CMC was 0.63 (IQR, 0.40–0.98) ml kg−1.

Question 3—relevance of 3D overlay

The cardiologists reported the image fusion to be very useful, with a median score of 4.0 (IQR, 3.0–4.0). None of the cases received a score of 1, indicating that the image fusion was never considered to be misleading.

For 38 cases (70.4%), either spared or redundant acquisitions were reported. There was good agreement among observers regarding the number of spared acquisitions (κ = 0.847), and substantial agreement for redundant acquisitions (κ = 0.631).

Discussion

Interventional cardiac catheterization has improved the care of patients with CHD (5). Complex congenital cardiac lesions often require long and even multiple procedures during the lifetime follow-up of patients (9, 29, 30). Therefore, these procedures should be optimized to achieve satisfactory diagnostic and therapeutic yields at as low as reasonably achievable radiation and contrast dose levels (1). The current study assessed the potentially beneficial impact of multimodality 3D image fusion, on radiation and contrast dose.

To the best of our knowledge, this study presents the largest population ever examined to compare radiation and contrast dose in MMIF2D−3D-guided procedures vs. conventional biplane 2DA among CHD patients. Furthermore, this is the first study evaluating the effect of MMIF2D−3D on radiation exposure levels grouped by irradiation event type and acquisition plane, exploiting the full potential of the data available inside the DICOM RDSR.

To reach accurate conclusions, we looked at weight-normalized exposure parameters because of the wide range of ages and body weights in the population. The reductions discussed in the following were statistically significant, unless otherwise stated.

Our findings showed a reduced FT(B) in the MMIF2D−3D cohort of the Overall group. This resulted in decreased lateral fluoroscopy-related exposure quantities [AKF(B)BW and DAPF(B)BW], which consequently reduced overall lateral plane exposure [AK(B)BW and DAP(B)BW], and eventually led to a cumulative reduction of AKBW. Similarly, AKBW was reduced for the Plasty subgroup, whereas both AKBW and DAPBW decreased in the PlastyAO subgroup. FT(B) can hence be considered the main factor driving the reduction in weight-normalized exposure parameters. In conventional biplane 2DA, a reference angiography is typically performed and used as a roadmap in both planes. This often requires biplane fluoroscopy or switching between frontal and lateral fluoroscopy to navigate guidewires, catheters, and devices to the desired location. However, with the MMIF2D−3D software, a transparent 3D roadmap registered with the frontal plane (unfortunately it cannot register with the lateral plane) is now available. This allows for movement of the frontal C-arm to different projection angles without losing registration of the 3D roadmap. In contrast to conventional biplane 2DA, most of the procedures can be done using the frontal plane without the need of lateral fluoroscopy.

The use of MMIF2D−3D resulted in a reduced FT(B), leading to a decrease in AKBW for the fusion population in the Overall, PlastyAO, and Plasty (sub)groups. However, only in the PlastyAO subgroup a DAPBW reduction was shown. This could be due to differences in imaging detector sizes between the frontal and lateral planes. As a result, AK(B) contributes more significantly to cumulative AK compared with DAP(B) contributing to cumulative DAP. Therefore, even though there may be an overall reduction in cumulative AKBW by decreased AK(B)BW in the lateral plane [which also leads to decreased DAP(B)BW], this does not necessarily reduce cumulative DAPBW.

Furthermore, the Overall and Plasty (sub)groups show lower AKFBW × FT and AKABW × FR. The reductions in exposure rate could be the result of several actions taken during the procedure. First, using lower dose imaging protocols in the MMIF2D−3D population may result in lower AK and DAP rates. The availability of MMIF2D−3D allows for lower yet adequate image quality with reduced radiation dosage. Second, utilizing larger field of views (FOVs) instead of the zooming function can lead to lower AK rates, but this also increases x-ray field areas. Because DAP is the product of these parameters and their effects cancel each other out to some extent, the overall impact on average DAP rate will be limited in this case. If operators wish to visualize the full scope of the 3D roadmap, they need to utilize the largest FOV available. Zooming in to a smaller FOV might no longer be necessary. Lastly, the use of less steep angulations in the fusion cohort, such as reducing the use of the lateral plane or employing less steep angulations in the frontal plane, can effectively decrease both AK rate and DAP rate. It is worth noting that the 3D roadmap is always accessible without x-ray exposure. This allows operators to identify the optimal viewing angle even before starting fluoroscopy. Once this optimal angle is determined, operators have the option to choose a less radiation-intensive projection while still ensuring good visibility during subsequent x-ray exposure. In addition, as mentioned earlier, operators mainly shift their focus from the steeper lateral plane to the frontal plane. In the current study, only two DAP quantities normalized by BW × FT or BW × FR were reduced in the Plasty subgroup [DAPA(A)BW × FR and DAPBW × FT], whereas many BW × FT and BW × FR normalized AK quantities exhibited a decrease, both cumulative and grouped, in multiple (sub)groups. This suggests that the most likely reason for the reduced AKFBW × FT and AKABW × FR mentioned previously is the use of larger FOVs.

This study found a reduction in the number of acquisitions and acquisition frames and their associated CMCBW, but these differences were not statistically significant. The questionnaire responses also supported these findings. Of the 54 cases, 17 reported at least one spared acquisition, while 27 had redundant acquisitions. This indicates the potential to further reduce radiation exposure and associated CMC. Furthermore, the cardiologists also confirmed the clinical benefit of MMIF2D−3D by scoring its relevance as very useful, with a median value of 4 on a 5-point Likert scale.

The literature presents similar findings regarding cumulative radiation exposure. Table 4 summarizes the key results from relevant studies for comparison purposes. In general, studies comparing CT and MRI image fusion with conventional biplane approaches tend to report lower average or median radiation exposure and contrast volume. Although more than half of the studies in Table 4 reported significantly lower values of one or several radiation exposure and CMC quantities, it is important to note that they primarily focus on a single procedure type and include small patient cohorts (15, 24, 27, 28).

Table 4

AuthorPopulationsProcedure typeRadiation exposure and CMC
Goreczny et al. (27)MMIF2D−3D (n = 7) vs. biplane 2DA (n = 8)PPVRSignificantly lower CMCBW (−60.5%), DAPBW (−66.6%) and AKBW (−68.3%)
Zablah et al. (30)MMIF2D−3D + optimized radiation settings + DrySeal sheath + ICEa (n = 38) vs. biplane 2DA (n = 38)PPVRSignificantly lower FT (−17.1%), CMCBW (−50.0%), DAPBW (−25.0%), and AK (−48.8%)
Goreczny et al. (28)MMIF2D−3D from MRI (n = 14) vs. MMIF2D−3D from CT (n = 8)RVOT Balloon Sizing, PPVRSignificantly lower DAP (−35.0%) and DAPBW (−64.5%) for MRI-guided procedures vs. CT-guided proceduresa
Goreczny et al. (31)MMIF2D−3D (n = 6) vs. biplane 2DA (n = 6)Pulmonary Vein InterventionsLower FT (−27%), CMC (−31.5%), DAP (−88.5%), and AK (−82%). The differences were not significant.
Stangenberg et al. (25)MMIF2D−3D (n = 16) vs. biplane 2DA (n = 16).
BMI matched pairs.
EVARSignificantly lower FT (−31.3%), CMC (−51.6%), and AK (−39.6%)
Tacher et al. (32)MMIF2D−3D (n = 8) vs. biplane 2DA (n = 14)EVARSignificantly lower CMC (−72.3%)
Glöckler et al. (15)3D-guided (MMIF2D−3D or 3DRA) (n = 12) vs. biplane 2DA (n = 20)CoASignificantly lower FT (−18.1%) and lower DAP (−26.4%). DAP not significantly lower.
Arar (33)MMIF2D−3D (n = 7) vs. biplane 2DA (n = 17)CoALower DAP (43.1%). DAP not significantly lower.
Ehret et al. (24)MMIF2D−3D (n = 13) vs. biplane 2DA (n = 20)Aortic Arch AngioplastySignificantly lower CMCBW (−48.1%), DAP (−51.7%), and FT (−55.9%) (nMMIF = 13, n2DA = 20)
Abu Hazeem et al. (34)MMIF2D−3D from MRI (n = 44) vs. biplane 2DA (n = 44). Weight and diagnosis matched pairsDiagnosticSignificantly lower FT (−14.6%), CMCBW (−39.4%), DAPF (−12.9%), DAP (−37.1%), and AK (−38.6%)

Summary of the main findings in the literature relevant to the current study.

PPVR, percutaneous pulmonary valve replacement; RVOT, right ventricular outflow tract; EVAR, endovascular aortic aneurysm repair; CoA, coarctation aorta.

a

Most of the MRI-guided procedures were acknowledged to be performed on a low-dose angiography unit known to significantly reduce radiation exposure during cardiac catheterization in an adult population.

Several publications in the literature compare radiation exposure associated with 3DRA imaging vs. conventional biplane 2DA (35). Some studies show higher radiation exposure with 3DRA, yet more recent research suggests lower values when using optimized exposure settings for the 3DRA imaging protocol. Some authors argue that diagnostic 3DRA is outdated for primary diagnosis and 3D guidance due to the availability of MMIF2D−3D (24, 36). However, it remains valuable when preprocedural imaging is not possible (36).

Limitations

This study has several limitations. First, there are a limited number of available MMIF2D−3D cases. The procedures involved diverse procedure types and complexities, including patients with and without preprocedural surgery and catheterizations, as well as a wide range of patient characteristics owing to the long-term follow-up. This poses challenges in achieving meaningful matching between the test and control groups. For example, StentPUL and PlastyPUL subpopulations show higher patient weight and age, yet these differences were not statistically significant. Moreover, StentPUL was the smallest procedure-type subgroup with different number of lesions, whether or not pretreated. Even though procedures were regrouped by similar procedure types, variations in patient-specific factors and technical challenges can significantly influence the imaging protocol and the projection views chosen by the operators. These choices will determine how the dose rate varies by patient thickness, fluoroscopy time, number of frames used, collimation, and FOV, and hence the eventual cumulative radiation exposure of the procedure. Nevertheless, referenced literature deals with similar limitations, while often reporting smaller sample sizes than the current study.

Furthermore, the reasons for the lower weight-normalized AK and DAP rates could not be confirmed as information regarding the use of lower dose imaging protocols and larger FOVs was unavailable in the DICOM RDSR.

Regarding the questionnaires, the numbers of the spared and redundant acquisitions were in line with the lower median number of frames and acquisitions in the MMIF2D−3D group. However, this evaluation is subjective and there was no statistically significant difference in the number of frames and acquisitions between the MMIF2D−3D and conventional 2DA cohorts.

To enable multimodality image fusion, a preprocedural cardiac CT or MRI is required. In this study, 30 cases were CT-guided and 24 were MRI-guided. The radiation exposure from the preprocedural CT should be considered for cumulative radiation exposure. However, organ or effective doses (OD or ED) are needed for both modalities to calculate the total radiation dose delivered to the patient. In this study, organ and effective doses were not assessed as the study did not impact clinical routine: MMIF2D−3D was only used when preprocedural CT or MRI was available without requesting additional imaging when unavailable. For the cardiac catheterizations included in this study, organ and effective doses could potentially be calculated using previously established DAPBW-to-ED conversion factors, which were determined for the same x-ray imaging system as the current study (13). Finally, the current study was performed in a single center and was not randomized.

Conclusion

MMIF2D−3D enables significant radiation exposure reduction in a single-center matched CHD population. It moves the focus of the operator toward the frontal plane, obviating the need for most of the lateral fluoroscopy exposure, hence reducing FT(B) and ultimately AKBW. Subgroup analysis similarly showed significant AKBW reduction in the Plasty and PlastyAO subgroups. DAPBW was significantly lower in the PlastyAO subgroup. No significantly lower CMCBW was observed in the MMIF2D−3D population for any of the analyzed (sub)groups.

Statements

Data availability statement

An anonymous dataset containing the cumulative exposure information from the RDSRs of all included procedures and supporting the conclusions of this article, can be made available by the authors upon reasonable request.

Ethics statement

This study was approved by the Ethics Committee of Ghent University Hospital (approval number 2016–1584). No written or oral consent was obtained from the patients or their legal guardians since this was a retrospective study for which data were anonymized.

Author contributions

DB: Conceptualization, Data curation, Formal Analysis, Investigation, Methodology, Resources, Software, Visualization, Writing – original draft, Writing – review & editing. KV: Data curation, Investigation, Resources, Validation, Writing – review & editing. JP: Data curation, Investigation, Resources, Validation, Writing – review & editing. LC: Data curation, Investigation, Resources, Validation, Writing – review & editing. KB: Conceptualization, Funding acquisition, Investigation, Methodology, Project administration, Resources, Supervision, Validation, Writing – review & editing. DW: Conceptualization, Funding acquisition, Investigation, Methodology, Project administration, Resources, Supervision, Validation, Writing – review & editing.

Funding

The authors declare financial support was received for the research, authorship, and/or publication of this article.

This work was supported by the Agency for Innovation by Science and Technology (IWT, Flanders, Belgium), which is now part of Flanders Innovation and Entrepreneurship (VLAIO, Flanders, Belgium) (Grant number 141670).

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

The authors declared that they were an editorial board member of Frontiers, at the time of submission. This had no impact on the peer-review process and the final decision.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fcvm.2023.1292039/full#supplementary-material

References

  • 1.

    HillKDFrushDPHanBKAbbottBGArmstrongAKDeKempRAet alRadiation safety in children with congenital and acquired heart disease: a scientific position statement on multimodality dose optimization from the image gently alliance. JACC Cardiovasc Imaging. (2017) 10:797818. 10.1016/j.jcmg.2017.04.003

  • 2.

    RoschlFPurbojoARufferACesnjevarRDittrichSGlocklerM. Initial experience with cinematic rendering for the visualization of extracardiac anatomy in complex congenital heart defects. Interact Cardiovasc Thorac Surg. (2019) 28:91621. 10.1093/icvts/ivy348

  • 3.

    TriedmanJKNewburgerJW. Trends in congenital heart disease: the next decade. Circulation. (2016) 133:271633. 10.1161/CIRCULATIONAHA.116.023544

  • 4.

    GhelaniSJGlatzACDavidSLeahyRHirschRArmsbyLBet alRadiation dose benchmarks during cardiac catheterization for congenital heart disease in the United States. JACC Cardiovasc Interv. (2014) 7:10609. 10.1016/j.jcin.2014.04.013

  • 5.

    HascoetSWarin-FresseKBaruteauAEHadeedKKarsentyCPetitJet alCardiac imaging of congenital heart diseases during interventional procedures continues to evolve: pros and cons of the main techniques. Arch Cardiovasc Dis. (2016) 109:12842. 10.1016/j.acvd.2015.11.011

  • 6.

    European guidelines on diagnostic reference levels for paediatric imaging. In: Radiation Protection. Vol. 185. (2018). Available at: https://op.europa.eu/en/publication-detail/-/publication/6e473ff5-bd4b-11e8-99ee-01aa75ed71a1/language-en (Accessed August 05, 2023).

  • 7.

    HaddadRNRizkCSalibaZFarahJ. Percutaneous closure of ventricular septal defects in children: key parameters affecting patient radiation exposure. Am J Cardiovasc Dis. (2021) 11:6572. ISSN:2160-200X/AJCD0122707.

  • 8.

    JohnsonJNHornikCPLiJSBenjaminDKJr.YoshizumiTTReimanREet alCumulative radiation exposure and cancer risk estimation in children with heart disease. Circulation. (2014) 130:1617. 10.1161/CIRCULATIONAHA.113.005425

  • 9.

    KangSLBensonL. Recent advances in cardiac catheterization for congenital heart disease. F1000Res. (2018) 7:370. 10.12688/f1000research.13021.1

  • 10.

    AlmohiyH. Paediatric computed tomography radiation dose: a review of the global dilemma. World J Radiol. (2014) 6:16. 10.4329/wjr.v6.i1.1

  • 11.

    BacherKBogaertELapereRDe WolfDThierensH. Patient-specific dose and radiation risk estimation in pediatric cardiac catheterization. Circulation. (2005) 111:839. 10.1161/01.CIR.0000151098.52656.3A

  • 12.

    BaldazziGCorazzaIRossiPLTestoniGBernardiTZannoliR. In vivo effectiveness of gadolinium filter for paediatric cardiac angiography in terms of image quality and radiation exposure. Phys Med. (2002) 18:10913.

  • 13.

    BuytaertDVandekerckhoveKPanzerJRubbensLDe WolfDBacherK. Local DRLs and automated risk estimation in paediatric interventional cardiology. PLoS One. (2019) 14:e0220359. 10.1371/journal.pone.0220359

  • 14.

    ChodickGKimKPShwarzMHorevGShalevVRonE. Radiation risks from pediatric computed tomography scanning. Pediatr Endocrinol Rev. (2009) 7:2936.

  • 15.

    GlocklerMHalbfabetaJKochAAchenbachSDittrichS. Multimodality 3D-roadmap for cardiovascular interventions in congenital heart disease—a single-center, retrospective analysis of 78 cases. Catheter Cardiovasc Interv. (2013) 82:43642. 10.1002/ccd.24646

  • 16.

    GorecznySKringsGHijaziZMFaganTBermanDKennyDet alPros, cons and future perspectives—three questions on three dimensional guidance for cardiac catheterization in congenital heart disease. Postepy Kardiol Interwencyjnej. (2019) 15:26373. 10.5114/aic.2019.87688

  • 17.

    KrishnaswamyATuzcuEMKapadiaSR. Three-dimensional computed tomography in the cardiac catheterization laboratory. Catheter Cardiovasc Interv. (2011) 77:8605. 10.1002/ccd.22740

  • 18.

    KangSLArmstrongAKringsGBensonL. Three-dimensional rotational angiography in congenital heart disease: present status and evolving future. Congenit Heart Dis. (2019) 14:104657. 10.1111/chd.12838

  • 19.

    ManicaJLBorgesMSMedeirosRFFischer LdosSBroettoGRossi FilhoRI. A comparison of radiation dose between standard and 3D angiography in congenital heart disease. Arq Bras Cardiol. (2014) 103:1317. 10.5935/abc.20140118

  • 20.

    GorecznySMorganGJDryzekPMollJAMoszuraT. Initial experience with live three-dimensional image overlay for ductal stenting in hypoplastic left heart syndrome. EuroIntervention. (2016) 12:152733. 10.4244/EIJ-D-15-00101

  • 21.

    MinderhoudSCSvan der SteltFMolenschotMMCKosterMSKringsGJBreurJ. Dramatic dose reduction in three-dimensional rotational angiography after implementation of a simple dose reduction protocol. Pediatr Cardiol. (2018) 39:163541. 10.1007/s00246-018-1943-3

  • 22.

    PetersMKringsGKosterMMolenschotMFreundMWBreurJM. Effective radiation dosage of three-dimensional rotational angiography in children. Europace. (2015) 17:6116. 10.1093/europace/euu207

  • 23.

    FaganTETruongUTJonePNBrackenJQuaifeRHazeemAAet alMultimodality 3-dimensional image integration for congenital cardiac catheterization. Methodist Debakey Cardiovasc J. (2014) 10:6876. 10.14797/mdcj-10-2-68

  • 24.

    EhretNAlkassarMDittrichSCesnjevarRRufferAUderMet alA new approach of three-dimensional guidance in paediatric cath lab: segmented and tessellated heart models for cardiovascular interventions in CHD. Cardiol Young. (2018) 28:6617. 10.1017/S1047951117002840

  • 25.

    StangenbergLShujaFCarelsenBElenbaasTWyersMCSchermerhornML. A novel tool for three-dimensional roadmapping reduces radiation exposure and contrast agent dose in complex endovascular interventions. J Vasc Surg. (2015) 62:44855. 10.1016/j.jvs.2015.03.041

  • 26.

    GorecznySDryzekPMorganGJLukaszewskiMMollJAMoszuraT. Novel three-dimensional image fusion software to facilitate guidance of complex cardiac catheterization: 3D image fusion for interventions in CHD. Pediatr Cardiol. (2017) 38:113342. 10.1007/s00246-017-1627-4

  • 27.

    GorecznySMoszuraTDryzekPLukaszewskiMKrawczukAMollJet alThree-dimensional image fusion guidance of percutaneous pulmonary valve implantation to reduce radiation exposure and contrast dose: a comparison with traditional two-dimensional and three-dimensional rotational angiographic guidance. Neth Heart J. (2017) 25:919. 10.1007/s12471-016-0941-4

  • 28.

    GorecznySDryzekPMoszuraTLukaszewskiMPodgorskiMNordmeyerSet alMagnetic resonance and computed tomography imaging fusion for live guidance of percutaneous pulmonary valve implantation. Postepy Kardiol Interwencyjnej. (2018) 14:41321. 10.5114/aic.2018.79871

  • 29.

    BardoDM. Letter by Bardo regarding article, “Cumulative radiation exposure and cancer risk estimation in children with heart disease”. Circulation. (2015) 131:e416. 10.1161/CIRCULATIONAHA.114.012646

  • 30.

    ZablahJERodriguezSALeahyRMorganGJ. Novel minimal radiation approach for percutaneous pulmonary valve implantation. Pediatr Cardiol. (2021) 42:92633. 10.1007/s00246-021-02564-7

  • 31.

    GorecznySMorganGJMcLennanDRehmanRZablahJE. Comparison of fusion imaging and two-dimensional angiography to guide percutaneous pulmonary vein interventions. Kardiol Pol. (2022) 80:4768. 10.33963/KP.a2021.0197

  • 32.

    TacherVLinMDesgrangesPDeuxJFGrunhagenTBecqueminJPet alImage guidance for endovascular repair of complex aortic aneurysms: comparison of two-dimensional and three-dimensional angiography and image fusion. J Vasc Interv Radiol. (2013) 24:1698706. 10.1016/j.jvir.2013.07.016

  • 33.

    ArarYDivekarAClarkSHussainTSebastianRHodaMet alRole of cross-sectional imaging in pediatric interventional cardiac catheterization. Children (Basel). (2022) 9:300. 10.3390/children9030300

  • 34.

    Abu HazeemAADoriYWhiteheadKKHarrisMAFogelMAGillespieMJet alX-ray magnetic resonance fusion modality may reduce radiation exposure and contrast dose in diagnostic cardiac catheterization of congenital heart disease. Catheter Cardiovasc Interv. (2014) 84:795800. 10.1002/ccd.25473

  • 35.

    van der SteltFSiegerinkSNKringsGJMolenschotMMCBreurJ. Three-dimensional rotational angiography in pediatric patients with congenital heart disease: a literature review. Pediatr Cardiol. (2019) 40:25764. 10.1007/s00246-019-02052-z

  • 36.

    SoderSWallischWDittrichSCesnjevarRPfammatterJPGlocklerM. Three-dimensional rotational angiography during catheterization of congenital heart disease—a ten years’ experience at a single center. Sci Rep. (2020) 10:6973. 10.1038/s41598-020-63903-x

Summary

Keywords

congenital heart disease, cardiac catheterization, multimodality image fusion, radiation exposure, contrast media, 3D guidance, radiation dose structured report

Citation

Buytaert D, Vandekerckhove K, Panzer J, Campens L, Bacher K and De Wolf D (2024) Multimodality 3D image fusion with live fluoroscopy reduces radiation dose during catheterization of congenital heart defects. Front. Cardiovasc. Med. 10:1292039. doi: 10.3389/fcvm.2023.1292039

Received

10 September 2023

Accepted

19 December 2023

Published

11 January 2024

Volume

10 - 2023

Edited by

Damien Kenny, Rush University, United States

Reviewed by

Federico Gutierrez-Larraya, University Hospital La Paz, Spain

Raymond N. Haddad, Assistance Publique-Hôpitaux de Paris, France

Updates

Copyright

*Correspondence: Dimitri Buytaert

Disclaimer

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Outline

Figures

Cite article

Copy to clipboard


Export citation file


Share article

Article metrics