Abstract
Background:
Robotic-assisted Percutaneous Coronary Intervention (R-PCI) is an advanced technique offering potential advantages for both patients and interventional cardiology practices. In 2023, Can Tho S.I.S General Hospital implemented its first robotic system for PCI procedures. This study aimed to evaluate the safety and clinical outcomes during the early learning curve of R-PCI.
Methods:
This prospective study included all patients undergoing R-PCI with the CorPath GRX Vascular Robotic System at Can Tho S.I.S General Hospital from April to September 2023. Baseline patient characteristics, procedural details, and six-month follow-up data were collected. Primary outcomes included Clinical Success, defined as <30% residual diameter stenosis in the target vessel without major adverse cardiovascular events (MACE) during hospitalization (e.g., death, myocardial infarction, target-vessel revascularization, stroke), and robotic success, defined as achieving clinical success with partial manual support or no manual support.
Results:
Thirty-one patients (mean age 64.5 ± 10 years; 64.5% male) with 37 lesions underwent R-PCI. Robotic success was achieved in 94.6% (35/37 lesions), comprising 21 lesions (56.8%) performed as full R-PCI and 14 lesions (37.8%) requiring partial manual support (exclusively for IVUS catheter manipulation). Manual conversion occurred in 5.4% due to inadequate guiding-catheter back-up and slow-flow. No MACE occurred during hospitalization; however, one patient died from a stroke during the six-month follow-up.
Conclusion:
This study demonstrates high success rates and minimal complications for R-PCI in its early implementation phase. Nevertheless, the stroke-related death observed during follow-up highlights the need for long-term studies to comprehensively assess R-PCI's safety and efficacy.
Introduction
The integration of robotics into medical procedures presents potential advantages for both patients and healthcare professionals, driving significant expansion in their adoption across various surgical fields (1). Of particular note in cardiovascular intervention is the emerging field of Robotic-assisted Percutaneous Coronary Intervention (R-PCI) (2–6). R-PCI offers distinct benefits over traditional manual PCI by enhancing procedural outcomes through steadier and more precise movements, minimizing operator radiation exposure, and improving physician ergonomics by eliminating the necessity of wearing a heavy lead apron (2–7).
The rationale for implementing R-PCI lies in its ability to reduce procedural variability and enhance stent deployment accuracy, addressing common limitations in traditional PCI. Robotic systems in other fields, such as hip arthroplasty, have shown enhanced accuracy in implant positioning, correlating with reduced long-term complications and improved patient outcomes (8). Although comprehensive long-term studies specific to R-PCI are needed, initial findings suggest potential parallels in improving cardiovascular outcomes through precision in robotic-assisted stent placement.
Safety benefits, particularly in terms of reduced radiation exposure, are one of the primary motivations for robotic-assisted surgery adoption. As operator experience with R-PCI increases, studies show a gradual decrease in fluoroscopy time, underscoring a learning curve in optimizing procedure duration without compromising patient safety (9). Notably, while R-PCI tends to require longer fluoroscopy time than manual PCI, the total contrast volume remains similar, suggesting efficient procedural control even in the robotic setting (10).
This study assesses the safety, feasibility, and clinical success of R-PCI during the early learning phase at Can Tho S.I.S General Hospital, with a focus on clinical and robotic success rates, procedural challenges, and complications. By addressing these factors, we aim to contribute to the limited data on R-PCI's feasibility in high-volume cardiac care settings, guiding future strategies for robotic system implementation in similar clinical environments.
Methods
Inclusion criteria
From April 2023 to September 2023, all patients diagnosed with coronary artery disease (CAD) and indicated for PCI, in accordance with clinical guidelines, were selected (11). The patients provided informed consent to undergo robot-assisted PCI.
Exclusion criteria
Patients were excluded from the study if they met any of the following criteria: (1) Patients with ST segment elevation myocardial infarction (STEMI), cardiogenic shock, sepsis, coagulation disorder; (2) contraindication to standard antiplatelet therapy (heparin, aspirin, clopidogrel, or other P2Y12 inhibitors); (3) complex PCI (bifurcation lesions), CTO (Chronic Total Occlusion). The study flow chart presented in Figure 1.
Figure 1
Clinical data
Upon admission, demographic and clinical characteristics, including age, gender, body mass index (BMI), smoking status, and medical history (hypertension, atrial fibrillation, coronary artery disease, diabetes mellitus, dyslipidemia, prior cerebral stroke, and chronic kidney disease) were recorded. Additionally, laboratory measurements were conducted to assess troponin T, creatinine, and lipid profiles (total cholesterol, triglycerides, high-density lipoprotein-C, and low-density lipoprotein-C), along with echocardiographic evaluations.
Robotic system
The system CorPath GRX robotic (Corindus, Siemens Healthineers Company, Waltham, MA, USA) connected to digital subtraction angiography (DSA) is implemented for R-PCI (Figure 2). The system included the Corpath GRX control console, Corpath GRX Extrended reach arm and Corpath GRX robotic drive.
Figure 2
Procedure protocol
The R-PCI team comprised two experienced interventional cardiologists with prior training in R-PCI. One interventional cardiologist operated the robot arm via the control panel, while the other assisted in the intervention room, managing the installation of instruments (guidewires, dilator balloons, and stents) into the robotic arm system. Additionally, a nurse assisted with the equipment, and a technician was responsible for adjusting the DSA machine and the robotic system.
Before the intervention procedure, patients received a full dose of antiplatelet drugs (aspirin, clopidogrel, or ticagrelor). Unfractionated heparin was administered intravenously at a dose of 70–100 units per kilogram of body weight before the insertion of the interventional guide (11).
Cardiac catheter access was obtained via the radial artery or femoral artery. The catheter's end was connected to a pressure-measuring wire before being attached to the drive gear system. The cassette system was connected to the Y-lock and then installed into the robotic arm. Finally, the system was positioned appropriately, and the procedure commenced.
Study endpoints
The success of R-PCI implementation was evaluated based on two aspects: clinical and robotic. Clinical success was defined as the successful placement of the stent at the target lesion, with quantitative coronary angiography (QCA) showing less than 30% diameter stenosis, a TIMI flow grade of 3, and no major adverse cardiac events (MACE) during the hospital stay (12). Robotic success was defined as achieving clinical success while completing the intervention with either Full R-PCI or with partial manual support. Full R-PCI refers to the entire PCI procedure being performed exclusively from the robotic console with no manual manipulation of intracoronary devices. Partial manual support refers to necessary manual aid where the procedure is ultimately completed robotically. Manual conversion was defined as the necessity to entirely disengage the robotic system to complete the remaining PCI steps manually (2, 13).
Additionally, post R-PCI laboratory data included radiation time (minutes), patient radiation exposure (µGy·m²), procedure time (from the arterial puncture until the last interventional device was completely withdrawn from the guide catheter), the dose-area product for technical support personnel (µSv), and contrast volume (mL).
A six-month follow-up was conducted, including an assessment of target lesion failure. Furthermore, all major adverse cardiac events (MACEs) were meticulously recorded. These included cardiac death, target-vessel myocardial infarction (MI), clinically driven target lesion revascularization, any death, any MI, and any clinically driven revascularization.
Institutional ethical approval
This study was approved by the Ethical Committee of the Can Tho University of Medicine and Pharmacy with the approval numb: 23.001.NCS/PCT-HDDD. Additionally, informed consent was obtained from each participant, and the research protocol adhered to the principles outlined in the Declaration of Helsinki.
Statistical analysis
All clinical features, procedural features, and clinical outcome results are illustrated as counts and percentages for categorical variables and as mean ± standard deviation or median (interquartile range) for continuous variables.For categorical variables, comparisons between groups (e.g., gender, prevalence of risk factors, clinical success/failure rates) were performed using the Chi-square (chi2) test or Fisher's exact test when the expected frequencies were less than five. For continuous variables that exhibited a normal distribution, differences between the two independent groups were evaluated using the unpaired Student's t-test. For continuous variables that did not follow a normal distribution or for ordinal data, the Mann–Whitney U-test was used to compare the central tendency between the two independent groups. A two-sided p-value of <0.05 was considered statistically significant. Statistics were performed using Stata 15 (StataCorp. 2017. Stata Statistical Software: Release 15. College Station, TX: StataCorp LLC).
Results
In total, 608 patients underwent PCI from April 2023 to September 2023, which included 69 cases with STEMI, 159 cases with non-STEMI and 380 cases of unstable angina. Based on the lesion characteristic, 50 patients were referred for R-PCI. Finally, 31 patients agreed for R-PCI.
Demographics and clinical features
31 patients (64.5% male) meeting the predefined inclusion criteria underwent R-PCI. Of these patients, 7 exhibited a single vessle disease, 17 had two vessle disease, and 7 had three vessle disease. Overall, 37 coronary lesions were intervened with R-PCI. Baseline clinical characteristics of the cohort, including demographic and comorbidity profiles, are detailed in Table 1. Notably, due to the study's conduct within a specialized in stroke, a higher prevalence of prior cerebrovascular events was observed in comparison to similar studies.
Table 1
| Characteristics | Patient (n = 31) | Percentage (%) | |
|---|---|---|---|
| General Information | |||
| Age (years), mean ± SD | 64.5 ± 10 | ||
| <60 | 11 | 35.5 | |
| 60–69 | 9 | 29.0 | |
| 70–80 | 10 | 32.3 | |
| >80 | 1 | 3.2 | |
| Gender | Male | 20 | 64.5 |
| Female | 11 | 35.5 | |
| Smoking (n, %) | 16 | 51.6 | |
| Hypertension | 31 | 100 | |
| Hyperlipidemia | 27 | 87.1 | |
| Prior cerebral stroke | 10 | 32.3 | |
| Prior myocardial infarction | 11 | 35.5 | |
| Atrial fibrillation | 1 | 3.2 | |
| Diabetes | 11 | 35.5 | |
| Heart failure | 0 | 0 | |
| Chronic kidney disease | 3 | 9.7 | |
| Indication for percutaneous coronary intervention | |||
| Stable angina | 29 | 93.5 | |
| Acute coronary syndrome | 2 | 6.5 | |
| Laboratory parameters | |||
| Triglycerides (mmol/L) | 2.1 ± 1.1 | ||
| Total cholesterol (mmol/L) | 4.5 ± 0.9 | ||
| High-density lipoprotein-C (mmol/L) | 1.0 ± 0.3 | ||
| Low-density lipoprotein-C (mmol/L) | 2.8 ± 0.7 | ||
| Creatinine (μmol/L) | 91.6 ± 20.2 | ||
| EF (%) | 66.8 ± 7.6 | ||
Baseline demographic and clinical characteristics.
Among the 37 lesions treated, the left anterior descending (LAD) artery accounted for 23 cases, the right coronary artery (RCA) for 10 cases, while the left circumflex artery (LCX) and obtuse marginal (OM) artery accounted for three and one case, respectively. Based on the American College of Cardiology/American Heart Association (ACC/AHA) classification of coronary lesions, the majority were classified as type A (40.5%), as detailed in Table 2.
Table 2
| Characteristics | Patient (n = 31) | Percentage (%) | |
|---|---|---|---|
| Accessing sites | Right radial arterial | 25 | 80.6 |
| Left radial arterial | 0 | 0 | |
| Right femoral artery | 6 | 19.4 | |
| Stent/PCI | One | 22 | 71.0 |
| Two | 8 | 25.8 | |
| Three | 1 | 3.2 | |
| Vessel disease | 1-vessel disease | 7 | 22.6 |
| 2-vessel disease | 17 | 54.8 | |
| 3-vessel disease | 7 | 22.6 | |
| Intervened Vessel | One | 25 | 80.6 |
| Two | 6 | 19.4 | |
| Intra-vascular ultrasound (IVUS) | 14 | 37.8 | |
| Lesion characteristics (n = 37) | |||
| Lesion location | Left anterior descending artery | 23 | 62.2 |
| Left circumflex artery | 3 | 8.1 | |
| Right coronary artery | 10 | 27.0 | |
| Obtuse marginal artery | 1 | 2.7 | |
| ACC/AHA lesions classification | Type A/B1 | 23 | 62.2 |
| Type B2/C | 14 | 37.8 | |
| Lesion length | ≤20 mm | 8 | 21.6 |
| >20 mm | 29 | 78.4 | |
| Stent diameter (mm) | 3.2 ± 0.4 | ||
| Stent length (mm) | 32.0 ± 9.3 | ||
| Quantitative Coronary Angiography | 70–90% | 27 | 73.0 |
| >90% | 10 | 27.0 | |
| Calcification Degree | Severity | 4 | 10.8 |
| Mild/moderate | 14 | 37.8 | |
| No calcium | 19 | 51.4 | |
| Ballooning before stenting (yes) | 31 | 83.8 | |
| Ballooning after stenting (yes) | 28 | 75.7 | |
| Procedure time (min) | 40.8 ± 18.2 | ||
| Radiation time (min) | 17.3 ± 7.6 | ||
| Air kerma radiation dose for patient dose (µGym2) | 1,019 (629.7–1,541) | ||
| Patient's Radiation Exposure (µGym2) | 9,569 ± 6,071 | ||
| Technical supporter's dose-area product (µSV) | 0.25 | ||
| Contrast volume (mL) | 7155 ± 45.8 | ||
Procedural characteristic.
Eight patients received two stents, and one patient required the placement of three stents during a single procedure (Figure 3). Additionally, two cases necessitated conversion to manual intervention. The average procedure time was 40.8 ± 18.2 min, with a total fluoroscopy time of 17.3 ± 7.6 min. Patient radiation exposure was measured at 9,569 ± 6,071 µGy·m2, with a contrast volume of 155 ± 45.8 mL utilized.
Figure 3
Procedural outcomes
The clinical success rate was 97.3%. Only 2.7% of lesions exhibited TIMI grade 2 flow, specifically observed in one case with acute side-branch occlusion. The robotic success rate was 94.6% (35/37 lesions). Among these, full R-PCI was achieved in 21 lesions, where the entire procedure was completed without any manual contact with intracoronary devices. Partial manual support was required in 14 lesions (37.8%); notably, this assistance was strictly limited to the manual advancement and withdrawal of the IVUS catheter due to current system incompatibility. No manual support was required for the manipulation of coronary wires, balloons, or stents in any of the successful cases. Manual conversion was required for 2 lesions (5.4%) due to inadequate guiding-catheter backup and slow-flow. A comprehensive summary is presented in Table 3.
Table 3
| Characteristics | Lesion (n = 37) | Percentage (%) | |
|---|---|---|---|
| Clinical success | 36 | 97.3 | |
| Robotic success | Full R-PCI | 35 | 94.6 |
| Partial Manual Supporta | 14 | 37.8 | |
| Manual conversion | 2 | 5.4 | |
| Myocardial infarction post PCI | 1 | 2,7 | |
| Stent thrombosis post PCI | 0 | 0 | |
| Post PCI TIMI flow | TIMI I | 0 | 0 |
| TIMI II | 1 | 2.7 | |
| TIMI III | 36 | 97.3 | |
In-hospital outcomes.
IVUS only, Thrombolysis in Myocardial Infarction (TIMI), Percutaneous coronary intervention (PCI).
Hospital outcome and six months follow-up
All patients demonstrated clinical stability, with no reports of chest pain. Cardiac enzyme levels were assessed, and antiplatelet and antithrombotic therapies were prescribed according to protocol. Throughout the six-month follow-up period, there were no recorded instances of lesion failure. However, one patient experienced a stroke, resulting in mortality (68 days after discharge). Importantly, no major adverse cardiovascular events (MACE) were observed within the cohort, as outlined in Table 4.
Table 4
| Characteristics | Total (n = 31) | Percentage (%) |
|---|---|---|
| Antiplatelet therapy at discharge | ||
| Acetylic salicylic acid (aspirin) | 100% | |
| Clopidogrel | 80% | |
| Prasugrel | 10% | |
| Ticagrelor | 10% | |
| Six months clinical outcomes | ||
| Target lesion failure | 0 | 0 |
| Target vessel failure | 0 | 0 |
| Mortality | 1 | 3.2 |
| Cardiac | 0 | 0 |
| Non cardiac | 1 | 3.2 |
| Stroke | 1 | 3.2 |
| Myocardial infarction | 0 | 0 |
| Target-vessel | 0 | 0 |
| Non-target-vessel | 0 | 0 |
Discharge and six months clinical outcomes.
Discussion
During the initial implementation phase of R-PCI (April to September 2023), 31 cases involving a total of 37 lesions were treated with stenting, with no reported clinical complications attributable to robotic assistance. The clinical success rate was 97.3%, while the robotic success rate reached 94.6%. These favorable outcomes marked a notable advancement in the application of R-PCI, despite two instances requiring conversion to manual intervention.
The success rates observed in both clinical and robotic aspects align with findings reported in previous research studies, such as the Precision GRX study, where clinical success and robotic success were reported at 97.8% and 89.2%, respectively (14). Conversely, the CORA-PCI study demonstrated clinical success at 99.1% and robotic success at 91.7% (2, 14). These consistently high rates of clinical and robotic success underscore the feasibility of implementing R-PCI for routine clinical practice, emphasizing its safety and efficacy.
Furthermore, the results of the first R-PCI conducted in Japan in 2020 by Kotaro Kagiyama yielded compelling insights when comparing R-PCI utilizing the CorPath GRX robot to manual methods. The clinical success rate associated with R-PCI was found to be comparable to manual intervention, with rates of 93.3% and 94.6%, respectively, yielding a p-value of 0.97 (2, 14).
Notably, only one case (2.7%) exhibited complications (type B2/C) related to myocardial infarction due to occlusion of a small side branch, resulting in a TIMI II flow. In comparison, Pedro A. Lemos and Fabian J. Brunner reported incidence rates of 1.2% and 5.6%, respectively (6, 15). Such peri-procedural MIs stemming from SB occlusion can occur during any coronary intervention, particularly in arteries featuring severely narrowed SBs and calcification. The occlusion typically arises when dilating and intervening on the main branch, inadvertently blocking the SB. Crucially, this specific case involved a type B2/C lesion, highlighting that complications are associated with the inherent complexity of the lesion itself, rather than the robotic approach. The patient did not experience chest pain due to the small size of the occluded SB. Subsequently, the patient received medical management, and an echocardiogram revealed no adverse effects on myocardial contractility or other complications.
Two cases necessitated conversion to manual intervention due to the complexity of type C lesions. Firstly, in a LAD artery with a type C lesion characterized by calcification and diffuse stenosis, manual stenting became necessary after the robotic delivery of the stent failed. A Guidezilla catheter (Boston Scientific, Marlborough, MA, USA) was employed to support the PCI, resulting in TIMI 3 flow without any complications. Secondly, in a RCA with an abnormal origin from the anterior aspect of the right coronary sinus and a diffuse type C lesion, the catheter and wire placed robotically became dislodged during balloon delivery. Manual intervention involving the successful placement of two drug-eluting stents was subsequently performed.
The success rate of R-PCI in our study was notably high for the type A/B1 lesion group, achieving a 100% success rate. Particularly for type B2/C lesions characterized by their length, calcification, and complexity, requiring additional support from assistive devices, the intervention success rate reached 85.7%. In a study by E. Mahmud with 315 patients, 108 cases underwent R-PCI, with 78.3% of cases presenting type B2/C lesions. The study reported a conversion rate to manual intervention of 11.1% (2). This high success rate is partially attributed to the low volume of R-PCI procedures analyzed (n = 31). As this was the first clinical experience with the robotic system in Vietnam, the procedure was executed with meticulous caution and strict adherence to a learning protocol.
Furthermore, we acknowledge that a significant proportion of the lesions treated in our series were less complex (Type A/B1, 62.2%). This reflects our initial strategy: under the supervision of a manufacturer proctor, we deliberately selected less complex lesions to prioritize patient safety and facilitate a stable learning curve. Lesion complexity was then gradually increased as the interventionist gained technical proficiency with the robotic console. This cautious selection explains why Partial Manual Support was not required for the manipulation of coronary wires, balloons, or stents, unlike the findings in more complex cohorts. The only manual assistance necessitated was for the advancement of the IVUS catheter (n = 14), due to current hardware incompatibility. In comparison, Bay et al. reported a higher manual assistance rate of 19.9% (including 7.8% for Type A/B1 lesions), which often involved complex maneuvers such as bifurcation kissing or overcoming friction in long lesions scenarios that were less frequent in our early-stage experience (13).
During the 6-month follow-up period, no MACEs related to R-PCI were recorded. There was only one case of stroke and subsequent death, which was determined to be unrelated to the procedure. The recent one-year clinical outcome of R-PCI, such as those from the CORA-PCI registry reported a MACE rate of 7.8% and research conducted in Switzerland a rate of 5% (5–7). Alongside the consistently high success rates of R-PCI observed worldwide, the results from our study once again underscore its safety, efficacy, and feasibility in clinical practice.
Lessons to learn during the robotic-PCI implementation process
The Corindus robotic system, integrated with technIQ technology features such as Spin, RoR, Wiggle, dotter, and constant speed, has significantly enhanced the proficiency in navigating guidewires through intricately compromised coronary arteries. The system's exceptional capability to manipulate and stabilize the guidewire during interventions mitigates the risk of complications associated with inadvertent loss of control over the guidewire. The spin technique involves rotating the wire and is particularly effective for routine navigation through vessels with larger diameters and complex curves. In contrast, the wiggle technique employs rapid, alternating clockwise and counterclockwise rotations of the wire during advancement, making it advantageous for navigating tight or narrow segments within smaller vessels. Additionally, the “rotate on retract” method automatically applies a 270-degree rotation to the wire as it is withdrawn. This feature is especially beneficial when encountering multiple side branches, as it facilitates redirection of the wire toward the main vessel if it inadvertently enters a side branch.
Collectively, these techniques replicate the wiring practices employed by experienced interventional cardiologists, providing a valuable training platform that reduces the learning curve for new cardiologists. As Yong et al. noted in a previous study, these methods significantly enhance procedural stability and efficiency, offering substantial support to interventional cardiologists during interventions (16).
The integration of pre-programmed leads and control buttons on the robotic console effectively mitigates the issue of hand tremors, which may affect some interventional cardiologists, particularly older practitioners. Additionally, these advanced robotic systems are impervious to external influences such as psychological stress, environmental conditions, operator experience, and age (16). By providing enhanced stability during interventions, robotic systems contribute significantly to procedural precision.
Notably, the objectives of robotic-assisted percutaneous coronary interventions (R-PCI), as well as other surgical robotic systems, extend beyond stability to include minimizing complications arising from human factors, such as manual instability. Drawing from the demonstrated success of robotic systems in achieving surgical precision across various disciplines, R-PCI represents a remarkable milestone in guidewire manipulation, substantially reducing the risk of complications associated with hand tremors (17).
The utilization of the Corindus robotic system for coronary artery intervention presents numerous advantages. By allowing the interventional cardiologist to remotely control the procedure from a comfortable control room, this system effectively reduces radiation exposure for both the physician and technical support staff. This transition in workflow occurs subsequent to sheath placement and robotic arm installation. Moreover, the system seamlessly integrates with the majority of existing interventional tools and exhibits a notably high success rate for type A, B1, and B2 coronary lesions, as detailed in Table 5.
Table 5
| Characteristics | Technical aspect | ||
|---|---|---|---|
| Successful | Failure | ||
| ACC/AHA coronary lesions type | Type A/B1 | 23 (100%) | 0 (0%) |
| Type B2/C | 12 (85.7%) | 2 (14.3%) | |
| Calcification Degree | Severity | 2 (50%) | 2 (50%) |
| Mild/moderate | 14 (100%) | 0 (0%) | |
| No calcium | 19 (100%) | 0 (0%) | |
Classification of lesions.
American college of cardiology/American heart association (ACC/AHA).
Personnel and the learning curve
The inclusion of two experienced interventionalists was deemed necessary for optimal patient safety and procedural efficiency during the adoption of the system. The assistant operator primarily facilitated tasks requiring manual intervention (such as managing challenges with support tools) and received direct proctoring and training from the Senior Operator, thereby minimizing the learning curve for new physicians.
When the center's experience and proficiency with the Corindus system increase, the number of required operators has reduced to a single primary operator for routine, less complex cases, allowing the full potential of personnel reduction to be realized, especially concerning non-essential personnel within the sterile field. In these routine procedures, the remaining assistant personnel only provide minimal manual assistance strictly limited to device installation onto the cassette. This evolution aligns with the findings of other centers that have demonstrated reduced essential staff requirements after the initial implementation period.
Implementation challenges
However, there are certain limitations to consider. Despite its efficacy, several interventional support tools, including intravascular ultrasound (IVUS), optical coherence tomography (OCT), and extension catheters, can pose challenges during utilization and often necessitate manual assistance. When using IVUS in R-PCI, the IVUS catheter will be placed in the position of the coronary angioplasty balloon. During the procedure, we encountered challenges in advancing the guidewire using the robotic system. As a result, we had to manually push the guidewire with the assistance of a second interventional cardiologist present in the procedure room. Furthermore, we have not utilized OCT for coronary interventions performed with robotic systems and therefore lack experience with its application in this context. The lesions in these cases were relatively mild, and the use of OCT was not strongly indicated compared to IVUS. Current robotic systems are only partially compatible with intracoronary imaging devices, necessitating the selective and context-specific application of newer methods (7, 9, 14).
The introduction of R-PCI has exhibited encouraging indications, displaying commendable adaptability to diverse clinical scenarios (18). However, its implementation has encountered several challenges, notably in the realm of cost. Given its status as a novel and sophisticated technology, R-PCI is likely to entail substantial expenses, encompassing both initial acquisition and subsequent maintenance. Such elevated costs may render it financially prohibitive for the majority of healthcare facilities.
Moreover, the significant investment required for R-PCI can contribute to escalated procedure fees, potentially constraining patient accessibility. This financial barrier may impede many patients from availing themselves of procedures utilizing this advanced technology (19).
Limitations
While this study offers valuable insights into the feasibility and success of robotic percutaneous coronary intervention (R-PCI), several limitations must be acknowledged. The small sample size of 31 cases restricts the statistical robustness of the findings, and the selective inclusion of patients with a preexisting understanding of their coronary anatomy may limit the applicability of these results to more complex or emergent scenarios. Furthermore, the study's single-center design, conducted by a specialized team within a familiar catheterization laboratory, likely contributed to the high success rates. This controlled setting does not account for variations in expertise, equipment, or workflows that may exist across other institutions, making the learning curve and procedural adjustments specific to this center.
We acknowledge a limitation regarding data granularity: highly detailed quantification (e.g., number of pull/push actions or device setup time) was not captured prospectively and is unreliable for retrospective reconstruction. Instead, we focused on providing clear quantitative limits in the methods and confirming the 100% setup requirement and the 14 cases requiring IVUS support in the results, collectively substantiating the minimal nature of the required assistance.
Additionally, the absence of long-term follow-up data precludes evaluation of R-PCI's sustained outcomes, such as stent durability and the recurrence of coronary events. To address these limitations, future multicenter studies involving larger and more diverse patient cohorts, alongside extended follow-up periods, are essential. Such investigations would help validate the findings, address procedural variability, and assess the long-term effectiveness and cost-effectiveness of R-PCI in broader clinical contexts.
Finally, a major limitation must be noted: the CorPath GRX robotic system utilized in this study was retracted from the market by Siemens Healthineers in the end of 2023. While this affects the immediate applicability of our specific findings, the discontinuation is understood to be a strategic step to facilitate the launch of a next-generation platform, confirming the manufacturer's continued commitment to robotic PCI.
Statements
Data availability statement
The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.
Ethics statement
The studies involving humans were approved by Ethical Committee of the Can Tho University of Medicine and Pharmacy with the approval numb: 23.001.NCS/PCT-HDDD. The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study.
Author contributions
TH: Conceptualization, Investigation, Methodology, Supervision, Writing – original draft, Writing – review & editing. MN: Data curation, Investigation, Methodology, Writing – original draft. TTN: Data curation, Writing – review & editing. LT: Data curation, Writing – review & editing. HN: Data curation, Writing – review & editing. DM: Data curation, Writing – review & editing. TMN: Data curation, Writing – review & editing. DN: Data curation, Investigation, Methodology, Project administration, Writing – review & editing, Writing – original draft. CT: Conceptualization, Funding acquisition, Project administration, Resources, Supervision, Visualization, Writing – review & editing. LV: Formal analysis, Methodology, Supervision, Writing – original draft, Writing – review & editing. AT: Supervision, Writing – review & editing. KN: Writing – review & editing. TN: Methodology, Writing – review & editing, Supervision, Writing – original draft. AN: Supervision, Writing – review & editing.
Funding
The author(s) declared that financial support was not received for this work and/or its publication.
Conflict of interest
The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declared that generative AI was not used in the creation of this manuscript.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
1.
D'EttorreCMarianiAStilliARodriguez y BaenaFValdastriPDeguetAet alAccelerating surgical robotics research: a review of 10 years with the da vinci research kit. IEEE Robot Autom Mag. (2021) 28(4):56–78. 10.1109/MRA.2021.3101646
2.
MahmudENaghiJAngLHarrisonJBehnamfarOPourdjabbarAet alDemonstration of the safety and feasibility of robotically assisted percutaneous coronary intervention in Complex coronary lesions: results of the CORA-PCI study (complex robotically assisted percutaneous coronary intervention). JACC Cardiovasc Interv. (2017) 10(13):1320–7. 10.1016/j.jcin.2017.03.050
3.
SmitsonCCAngLPourdjabbarAReevesRPatelMMahmudE. Safety and feasibility of a novel, second-generation robotic-assisted system for percutaneous coronary intervention: first-in-human report. J Invasive Cardiol. (2018) 30(4):152–6.
4.
DouKFSongCXMuCWYangWXZhuCGFengLet alFeasibility and safety of robotic PCI in China: first in man experience in Asia. J Geriatr Cardiol. (2019) 16(5):401–5. 10.11909/j.issn.1671-5411.2019.05.004
5.
KagiyamaKMitsutakeYUenoTSakaiSNakamuraTYamajiKet alSuccessful introduction of robotic-assisted percutaneous coronary intervention system into Japanese clinical practice: a first-year survey at single center. Heart Vessels. (2021) 36(7):955–64. 10.1007/s00380-021-01782-6
6.
LemosPAFrankenMMarianiJJr.CaixetaAAlmeidaBOPittaFGet alSafety and effectiveness of introducing a robotic-assisted percutaneous coronary intervention program in a tertiary center: a prospective study. Cardiovasc Diagn Ther. (2022) 12(1):67–76. 10.21037/cdt-21-442
7.
HänerJDRäberLMoroCLosdatSWindeckerS. Robotic-assisted percutaneous coronary intervention: experience in Switzerland. Front Cardiovasc Med. (2023) 10:1294930. 10.3389/fcvm.2023.1294930
8.
RiceSJD’AbarnoALuuHH. Robotic-assisted total hip arthroplasty and spinopelvic parameters: a review. Hip Pelvis. (2024) 36(2):87–100. 10.5371/.hp2024.36.2.87
9.
LeungJFrenchJXuJKachwallaHKaddapuKBadieTet alRobotic assisted percutaneous coronary intervention: initial Australian experience. Heart Lung Circ. (2024) 33(4):493–9. 10.1016/j.hlc.2024.01.019
10.
DurandEEltchaninoffH. Robotic-assisted percutaneous coronary intervention: the future or the past?EuroIntervention. (2024) 20(1):19–20. 10.4244/EIJ-E-23-00064
11.
LawtonJSTamis-HollandJEBangaloreSBatesERBeckieTMBischoffJMet al2021 ACC/AHA/SCAI guideline for coronary artery revascularization: executive summary: a report of the American college of cardiology/American heart association joint committee on clinical practice guidelines. Circulation. (2022) 145(3):e4–e17. 10.1161/cir.0000000000001039
12.
LevineGNBatesERBlankenshipJCBaileySRBittlJACercekBet al2011 ACCF/AHA/SCAI guideline for percutaneous coronary intervention. Circulation. (2011) 124(23):e574–651. 10.1161/CIR.0b013e31823ba622
13.
BayBGoßlingARilingerJvonzur MühlenCHofmannFNefHet alManual support during robotic-assisted percutaneous coronary intervention. Clin Res Cardiol. (2025) 115(1):25–32. 10.1007/s00392-025-02596-6
14.
WeiszGMetzgerDCCaputoRPDelgadoJAMarshallJJVetrovecGWet alSafety and feasibility of robotic percutaneous coronary intervention: pRECISE (percutaneous robotically-enhanced coronary intervention) study. J Am Coll Cardiol. (2013) 61(15):1596–600. 10.1016/j.jacc.2012.12.045
15.
BrunnerFJWaldeyerCZengin-SahmEKondziellaCSchrageBClemmensenPet alEstablishing a robotic-assisted PCI program: experiences at a large tertiary referral center. Heart Vessels. (2022) 37(10):1669–78. 10.1007/s00380-022-02078-z
16.
YoungLKhatriJ. Robotic percutaneous coronary intervention: the good, the bad, and what is to Come. US Cardiol Rev. (2022) 16:e02. 10.15420/usc.2020.28R1
17.
KoedaYIshidaMSasakiKKikuchiSYamayaSTsujiKet alPeriprocedural and 30-day outcomes of robotic-assisted percutaneous coronary intervention used in the intravascular imaging guidance. Cardiovasc Interv Ther. (2023) 38(1):39–48. 10.1007/s12928-022-00864-0
18.
NhanVTKhoaNQThuyLTVan DuongNVan TanNNgoc TheTHet alEarly safety and mid-term clinical outcomes of technology transfer of transcatheter aortic valve implantation in patients with severe aortic valve stenosis in Vietnam: a single-center experience of 90 patients. Lancet Reg Health West Pac. (2024) 42:100956. 10.1016/j.lanwpc.2023.100956
19.
Worldbank. GDP per capita (current US$)—Viet Nam. Available online at:https://data.worldbank.org/indicator/NY.GDP.PCAP.CD?locations=VN(Accessed May 10, 2024).
Summary
Keywords
coronary artery revascularization, coronary artery disease, PCI—percutaneous coronary intervention, robotic interventions, robotic-assisted percutaneous coronary intervention
Citation
Ho TD, Nguyen MC, Nguyen TT, Tran L, Nguyen H, Mai D, Ngo TM, Nguyen DC, Tran CC, Vu L, Tran AV, Ngo K, Nguyen T and Nanjundappa A (2026) Safety and feasibility during early implementation of robotic-assisted percutaneous coronary intervention. Front. Cardiovasc. Med. 13:1731900. doi: 10.3389/fcvm.2026.1731900
Received
24 October 2025
Revised
26 December 2025
Accepted
26 January 2026
Published
25 February 2026
Volume
13 - 2026
Edited by
Tommaso Gori, Johannes Gutenberg University Mainz, Germany
Reviewed by
Jonas Häner, University Hospital of Bern, Switzerland
Fabian Johannes Brunner, University Medical Center Hamburg-Eppendorf, Germany
Updates
Copyright
© 2026 Ho, Nguyen, Nguyen, Tran, Nguyen, Mai, Ngo, Nguyen, Tran, Vu, Trang, Ngo, Nguyen and Nanjundappa.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Thuong Dung Ho dunghothuong@gmail.com
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.