Abstract
Background:
The accurate detection of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) is the key to control Coronavirus Disease-2019 (COVID-19). The performance of different antibody detection methods for diagnosis of COVID-19 is inconclusive.
Methods:
Between 16 February and 28 February 2020, 384 confirmed COVID-19 patients and 142 healthy controls were recruited. 24 different serological tests, including 4 enzyme-linked immunosorbent assays (EIAs), 10 chemiluminescent immunoassays (CLIAs), and 10 lateral flow immunoassays (LFIAs), were simultaneously performed.
Results:
The sensitivities of anti-SARS-CoV-2 IgG and IgM antibodies with different reagents ranged from 75 to 95.83% and 46.09 to 92.45%, respectively. The specificities of both anti-SARS-CoV-2 IgG and IgM were relatively high and comparable among different reagents, ranged from 88.03 to 100%. The area under the curves (AUCs) of different tests ranged from 0.733 to 0.984, and the AUCs of EIAs or CLIAs were significantly higher than those of LFIAs. The sensitivities of both IgG and IgM gradually increased with increase of onset time. After 3–4 weeks, the sensitivities of anti-SARS-CoV-2 IgG were maintained at a certain level but the sensitivities of IgM were gradually decreased. Six COVID-19 patients who displayed negative anti-SARS-CoV-2 results were associated with the factors such as older age, having underlying diseases, and using immunosuppressant.
Conclusion:
Besides the purpose of assessing the impact of the SARS-CoV-2 pandemic in the population, SARS-CoV-2 antibody assays may have an adjunct role in the diagnosis and exclusion of COVID-19, especially by using high-throughput technologies (EIAs or CLIAs).
Introduction
Coronavirus disease 2019 (COVID-19), the emerging infectious disease caused by a novel severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), is the greatest threat to public health worldwide in recent 2 years (Phelan et al., 2020; The, 2020; Zhu et al., 2020). Globally, as of February 11, 2022, there have been 404 million confirmed cases of COVID-19, including 5.7 million deaths, reported to the World Health Organization. To this day, COVID-19 pandemic is still the most critical problem in the global health agenda.
The rapid and accurate diagnosis of COVID-19 is the key to control the epidemic of this disease. The diagnosis of COVID-19 is mainly based on epidemiology, clinical symptoms, radiology, and laboratory pathogen detection. The clinical symptoms of COVID-19 include many typical respiratory manifestations (fever, cough, chest pain, or shortness of breath) and other manifestations (fever, muscle ache, fatigue, diarrhea, or headache) (Baj et al., 2020; Brendish et al., 2020; Pan et al., 2020), which are similar to that of influenza (Wang et al., 2014). The asymptomatic COVID-19 patients were also reported previously, increasing the difficulty of diagnosis based solely on clinical features (Rothe et al., 2020; Ralli et al., 2021; Temkin and Healthcare Worker COVID-19 Surveillance Working Group, 2021). The characteristics of radiology of COVID-19 are also unspecific and the diagnosis of which based on radiology has variation among different radiologists (Chung et al., 2020; Kuo et al., 2021). The positive detection of SARS-CoV-2 virus nucleic acid by using real-time reverse transcription polymerase chain reaction (RT-PCR) is the most important diagnostic tool for COVID-19 (Islam and Iqbal, 2020; Pascarella et al., 2020; Zowawi et al., 2021). However, nucleic acid testing has some limitations, such as requiring certified laboratories, experienced technicians and expensive equipment, long turnaround time, and the existence of false negative results (Liu et al., 2020; Sule and Oluwayelu, 2020).
Serological tests are readily available in clinical laboratories in most hospitals and are easier to carry out than molecular tests (Lisboa Bastos et al., 2020; Xiao S.Y. et al., 2020). Besides the purpose of assessing the impact of the SARS-CoV-2 pandemic in the population and evaluating antibody titers from previous exposures to SARS-CoV-2 or from vaccine treatment, serological tests were also recommended to be used for diagnosing or excluding suspected cases in the guideline of diagnosis and treatment for SARS-CoV-2 pneumonia made by Chinese National Health Commission. SARS-CoV-2-specific IgM antibody level peaks at week 3 and then declines, whereas IgG antibodies to spike protein can persist long-term, even beyond 1 year after infection (Gudbjartsson et al., 2020; Hou et al., 2020, 2021; Xiao A.T. et al., 2020). Importantly, although there were many studies focused on the role of SARS-CoV-2-specific antibodies in the diagnosis, prognosis and management of COVID-19 (Lisboa Bastos et al., 2020; Ong et al., 2021), there was rare study to evaluated the performance of different antibody detection methods in clinical practice. Given there are many commercially available anti-SARS-CoV-2 IgM and IgG detection kits developed, it is necessary to verify the accuracy of serological tests in the diagnosis of COVID-19 in clinical practice.
In the current study, we compared the performance of 24 different serological tests, which were classified as enzyme-linked immunosorbent assay (ELISA or EIA), chemiluminescent immunoassay (CLIA), and point-of-care testing (POCT) technology such as lateral flow immunoassay (LFIA), in the diagnosis of COVID-19 patients. We also analyzed the sensitivity of these methods in patients at different time after disease onset. This study is useful for developing the standards for antibody testing and further understanding the appearance and persistence of SARS-CoV-2 antibodies in COVID-19 patients.
Materials and Methods
Patients
Between 16 February and 28 February 2020, a total number of 384 COVID-19 patients and 142 healthy controls were continuously recruited from Tongji Hospital (the largest hospital in central region of China), Wuhan, China. The diagnosis of COVID-19 was according to the guideline of diagnosis and treatment for SARS-CoV-2 pneumonia made by Chinese National Health Commission. The confirmed COVID-19 patients were defined as having positive SARS-CoV-2 real-time RT-PCR results in clinical samples, together with typical clinical symptoms (fever, cough, or shortness of breath) and radiological characteristics (unilateral pneumonia, bilateral pneumonia, or ground-glass opacity). The healthy controls were defined as individuals without signs or symptoms of active disease by clinical interview and physical examination, and with negative SARS-CoV-2 real-time RT-PCR results. Five milliliter of venous blood from each participant were collected into a test tube for serum separation, which was then stored at –80°C until use. The serum was thawed and mixed before measurement. The demographic and clinical information, and laboratory results were collected from electronic medical records. This study was approved by the ethical committee of Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology (TJ-C20200128).
Severe Acute Respiratory Syndrome Coronavirus 2-Specific Antibody Detection
ELISAs
Four ELISA kits were obtained from Livzon [Anti-SARS-CoV-2 ELISA Kit (IgG), Anti-SARS-CoV-2 ELISA Kit (IgM), Livzon Pharmaceutical Group Inc., Zhuhai, China] (Livzon-EIA-IgG, Livzon-EIA-IgM) or WANTAI [Anti-SARS-CoV-2 ELISA Kit (Ab total), Anti-SARS-CoV-2 ELISA Kit (IgM), WANTAI BioPharm Group Inc., Beijing, China] (WANTAI-EIA-Ab total, WANTAI-EIA-IgM) respectively, and performed according to the manufacturers’ instructions. Briefly, –80°C stored serum samples were thawed and mixed before use. After that, the serum samples and reagents were added to 96-well microtiter plates pre-coated with SARS-CoV-2-specific antigens and detected by automated analyzers. Signal to cutoff value (S/CO) ≥ 1 is reactive (positive), and S/CO < 1 is non-reactive (negative).
Chemiluminescent Immunoassays
Similarly, –80°C stored serum samples were thawed and mixed before use. Ten anti-SARS-CoV-2 CLIA kits were obtained from InnoDx (InnoDx biotechnology Co., Ltd., Xiamen, China) (InnoDx-CLIA-Ab total, InnoDx-CLIA-IgM), Beier (Beier bioengineering Co., Ltd., Beijing, China) (Beier-CLIA-IgG, Beier-CLIA-IgM), YHLO (YHLO Biotech Co., Ltd., Shenzhen, China) (YHLO-CLIA-IgG, YHLO-CLIA-IgM), Orienter (Orienter Biotechnology Co., Ltd. Chengdu, China) (Orienter-CLIA-IgG, Orienter-CLIA-IgM), and Maccura (Maccura Biotechnology Co., Ltd., Chengdu, China) (Maccura-CLIA-IgG, Maccura-CLIA-IgM), respectively, and performed by automated chemiluminescence analyzers (Caris200, InnoDx; VI-180, Beier; iFlash 3000-C, YHLO; LA2000, Orienter; i 3000, Maccura). For InnoDx, Orienter and Maccura, S/CO ≥ 1 is reactive (positive), and S/CO < 1 is non-reactive (negative). For YHLO, the results ≥ 10 AU/ml is reactive (positive), and the results <10 AU/ml is non-reactive (negative). The cutoff value is 5 RU/ml for Beier.
Lateral Flow Immunoassays
Ten POCT anti-SARS-CoV-2 LFIA kits were obtained from Livzon (Livzon-LFIA-IgG, Livzon-LFIA-IgM), WANTAI (WANTAI-LFIA-IgG, WANTAI-LFIA-IgM), Beier (Beier-LFIA-IgG, Beier-LFIA-IgM), HEALGEN (Orient Gene Biotech Co., Ltd., Huzhou, China) (HEALGEN-LFIA-IgG, HEALGEN-LFIA-IgM), and Innovita (Innovita Biological Technology Co., Ltd., Beijing, China) (Innovita-LFIA-IgG, Innovita-LFIA-IgM), respectively, and performed according to the manufacturers’ instructions. Briefly, –80°C stored serum samples were thawed and mixed before use. After that, the serum samples were diluted and added to colloidal gold immunochromatographic strip. The results were finally read by the eyes.
For the antibody detection reagents, YHLO-CLIA-IgG, YHLO-CLIA-IgM, HEALGEN-LFIA-IgG, and HEALGEN-LFIA-IgM have received CE certification in Europe, and InnoDx-CLIA-Ab total, Innovita-LFIA-IgG, and Innovita-LFIA-IgM have been approved by the China National Medical Product Administration.
Statistical Analysis
Data analyses were performed with SPSS 21.0 (SPSS. Inc.) or GraphPad Prism 6.0.1 (GraphPad). Unless otherwise specified, the data are expressed as mean ± standard deviation (SD). Continuous variables were compared with Mann-Whitney U-test. Receiver-operator characteristic (ROC) curve analysis was used to compare the performance of different anti-SARS-CoV-2 assays for diagnosis of COVID-19. The area under the curve (AUC), sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), together with 95% confidence interval (CI), were identified. Statistical significance was determined as p < 0.05.
Results
Participant Characteristics
A total of 384 COVID-19 patients (males, 197; females, 187) were enrolled in this study. The median age was 65 years (range 5–91 years). The median time from onset of symptoms to antibody detection was 21 days (range 3–78 days). A total of 142 healthy individuals (males, 80; females, 62) who tested negative for SARS-CoV-2 were enrolled as control subjects. The median age of healthy controls was 42 years (range 2–90 years).
The Performance of 24 Severe Acute Respiratory Syndrome Coronavirus 2 Antibody Assays for Coronavirus Disease-2019 Diagnosis
The sensitivity and specificity of these 24 antibody reagents are shown in Table 1. The sensitivities of anti-SARS-CoV-2 IgG or total antibodies with different reagents ranged from 75 to 95.83%. The sensitivities of anti-SARS-CoV-2 IgM ranged from 46.09 to 92.45%. The specificities of both anti-SARS-CoV-2 IgG and IgM with different reagents were relatively high, ranged from 88.03 to 100%. The specificities of IgG and IgM in most reagents were comparable. Our data showed that most commercially available anti-SARS-CoV-2 detection kits, especially the high-throughput technologies (EIAs or CLIAs), have relatively high sensitivity and specificity for COVID-19 diagnosis.
TABLE 1
| Sensitivity (%) | Positive/Total | Specificity (%) | Negative/Total | Positive predictive value (%) | Negative predictive value (%) | |
| Livzon-EIA-IgG | 92.19 | 354/384 | 96.48 | 137/142 | 98.61 | 82.04 |
| WANTAI-EIA-Ab total | 95.83 | 368/384 | 97.18 | 138/142 | 98.92 | 89.61 |
| InnoDx-CLIA-Ab total | 93.49 | 359/384 | 99.30 | 141/142 | 99.72 | 84.94 |
| Beier-CLIA-IgG | 75.00 | 288/384 | 99.30 | 141/142 | 99.65 | 59.49 |
| YHLO-CLIA-IgG | 95.05 | 365/384 | 88.03 | 125/142 | 95.55 | 86.81 |
| Orienter-CLIA-IgG | 94.27 | 362/384 | 97.18 | 138/142 | 98.91 | 86.25 |
| Maccura-CLIA-IgG | 92.19 | 354/384 | 100 | 142/142 | 100 | 82.56 |
| Livzon-LFIA-IgG | 92.71 | 356/384 | 99.30 | 141/142 | 99.72 | 83.43 |
| WANTAI-LFIA-IgG | 83.33 | 320/384 | 95.77 | 136/142 | 98.16 | 68.00 |
| Beier-LFIA-IgG | 81.34 | 231/284* | 100 | 142/142 | 100 | 72.82 |
| HEALGEN-LFIA-IgG | 91.93 | 353/384 | 100 | 142/142 | 100 | 82.08 |
| INNOVITA-LFIA-IgG | 92.97 | 357/384 | 99.30 | 141/142 | 99.73 | 84.43 |
| Livzon-EIA-IgM | 47.14 | 181/384 | 99.30 | 141/142 | 99.45 | 40.99 |
| WANTAI-EIA-IgM | 85.68 | 329/384 | 97.89 | 139/142 | 99.10 | 71.65 |
| InnoDx-CLIA-IgM | 89.58 | 344/384 | 99.30 | 141/142 | 99.71 | 77.90 |
| Beier-CLIA-IgM | 64.32 | 247/384 | 97.18 | 138/142 | 98.41 | 50.18 |
| YHLO-CLIA-IgM | 84.11 | 323/384 | 90.85 | 129/142 | 96.13 | 67.89 |
| Orienter-CLIA-IgM | 90.36 | 347/384 | 97.89 | 139/142 | 99.14 | 78.98 |
| Maccura-CLIA-IgM | 92.45 | 355/384 | 100 | 142/142 | 100 | 83.04 |
| Livzon-LFIA-IgM | 46.09 | 177/384 | 100 | 142/142 | 100 | 40.69 |
| WANTAI-LFIA-IgM | 57.81 | 222/384 | 99.30 | 141/142 | 99.55 | 46.53 |
| Beier-LFIA-IgM | 71.48 | 203/284* | 100 | 142/142 | 100 | 63.68 |
| HEALGEN-LFIA-IgM | 69.79 | 268/384 | 100 | 142/142 | 100 | 55.04 |
| INNOVITA-LFIA-IgM | 67.71 | 260/384 | 97.18 | 138/142 | 98.48 | 52.67 |
The sensitivity and specificity of 24 SARS-CoV-2 antibody assays.
*Due to a shortage of reagents or samples, only 284 cases of COVID-19 were detected by Beier-LFIA-IgG/IgM.
ROC analysis for each assay was determined. As shown in Figure 1, the AUCs of these 24 assays ranged from 0.733 to 0.984. The AUCs of EIAs or CLIAs, no matter in IgG or IgM, were significantly higher than those of LFIAs (Figure 2). The AUCs of these 24 assays ranged in a descending order from: Orienter-CLIA-IgG > Maccura-CLIA-IgG > InnoDx-CLIA-Ab total > WANTAI-EIA-Ab total > Livzon-EIA-IgG > YHLO-CLIA-IgG > Beier-CLIA-IgG > Innovita-LFIA-IgG > HEALGEN-LFIA-IgG > Livzon-LFIA-IgG > Beier-LFIA-IgG > WANTAI-LFIA-IgG for IgG or total antibody detection; and Orienter-CLIA-IgM > Maccura-CLIA-IgM > WANTAI-EIA-IgM > InnoDx-CLIA-IgM >Beier-CLIA-IgM > YHLO-CLIA-IgM > Livzon-EIA-IgM > Beier-LFIA-IgM > HEALGEN-LFIA-IgM > Innovita-LFIA-IgM > WANTAI-LFIA-IgM > Livzon-LFIA-IgM for IgM detection.
FIGURE 1
FIGURE 2
The sensitivities of anti-SARS-CoV-2 IgM were significantly lower than those of anti-SARS-CoV-2 IgG or total antibodies (p < 0.01), but the specificities between anti-SARS-CoV-2 IgG and IgM had no difference (p = 0.568) (Figure 3). The sensitivities of both IgG and IgM were gradually increased with increase of onset time, and reached the peak after about 3–4 weeks. After that, the sensitivities of anti-SARS-CoV-2 IgG were maintained at a certain level but the sensitivities of anti-SARS-CoV-2 IgM were gradually decreased (Figure 4).
FIGURE 3
FIGURE 4
False-Negative Results of Severe Acute Respiratory Syndrome Coronavirus 2 Antibody Assays
There were 6 COVID-19 patients who had negative anti-SARS-CoV-2 results by any detection kits. We observed that days from onset to antibody detection were between 3 and 12 days in these patients. Furthermore, five of six patients had underlying diseases such as hypertension, diabetes, and coronary heart disease. One patient with systemic lupus erythematosus was undergoing immunosuppressant treatment (rituximab). The demographic and clinical characteristics of these patients are shown in Table 2. These data suggested that the false negative results of anti-SARS-CoV-2 results may be caused by the variability in the time from onset of illness to detection or the immunosuppression status in COVID-19 patients.
TABLE 2
| No. | Sex | Age (years) | Time from onset to detection (days) | Severity | Underlying disease |
| 19 | Female | 75 | 12 | Severe | Hypertension, coronary heart disease |
| 162 | Male | 34 | 5 | Mild | None |
| 191 | Female | 22 | 7 | Extremely severe | Systemic lupus erythematosus, using immunosuppressant (rituximab) |
| 208 | Female | 27 | 3 | Mild | Hyperthyreosis, pregnancy |
| 219 | Female | 66 | 7 | Severe | Hypertension, diabetes, coronary heart disease, endometrial cancer (after surgery) |
| 236 | Male | 57 | 3 | Severe | Hypertension, diabetes, lung cancer (after surgery) |
The demographic and clinical characteristics of 6 COVID-19 patients with negative anti-SARS-CoV-2 results in all 24 assays.
Discussion
The early diagnosis and isolation of COVID-19 patients are the key to control the outbreak of the disease. Given the false-negative results of SARS-CoV-2 RT-PCR is common in clinical samples, especially in patients with increased time since symptom onset or with oropharyngeal samples rather than nasopharyngeal samples, it is unsuitable for use of the method to exclude COVID-19 (Arevalo-Rodriguez et al., 2020; Wikramaratna et al., 2020). With the outbreak of COVID-19, many SARS-CoV-2 antibody detection methods based on different methodologies such as ELISA, CLIA and LFIA have been developed. The current view emphasizes that SARS-CoV-2 antibodies serve as an complement to RT-PCR in the diagnosis of acute infection (Sidiq et al., 2020). However, the performance of these antibody detection methods for the diagnosis of COVID-19 patients is inconclusive. In this study, we compared the performance of almost all current commercially available assays for anti-SARS-CoV-2 detection in China. Our data showed that the performance of high-throughput technologies including EIAs and CLIAs was superior to POCT. Moreover, most EIAs and CLIAs had high sensitivity and specificity and comparable diagnostic accuracy, which confirmed that SARS-CoV-2 antibody detection may have an adjunct role in the diagnosis of COVID-19.
Regarding using serological tests for COVID-19 diagnosis, there were two main aspects that should be considered. First, the test should have enough sensitivity and specificity to facilitate COVID-19 diagnosis. Second, technical efficiency and bio-safety also counted. Our data showed that most anti-SARS-CoV-2 IgG tests and many anti-SARS-CoV-2 IgM tests such as Orienter-CLIA-IgM and Maccura-CLIA-IgM, achieved over 90% sensitivity in the diagnosis of COVID-19, which was in accordance with previous study (Li et al., 2020). Thus, SARS-CoV-2 antibody detection was of important value in early diagnosis of COVID-19, especially in patients suspected as SARS-CoV-2 infection but with negative RT-PCR results. On the other hand, both anti-SARS-CoV-2 IgG and IgM detection had high specificity (mostly higher than 95%) for diagnosis of COVID-19. Hereafter, these data suggest that SARS-CoV-2 antibody detection plays an important role in the diagnosis and exclusion of COVID-19 patients.
Except for Beier, the sensitivities of all other anti-SARS-CoV-2 IgG detection methods were relatively high. Nevertheless, the sensitivities of anti-SARS-CoV-2 IgM detection methods varied greatly, and the sensitivities of WANTAI-EIA-IgM, InnoDx-CLIA-IgM, YHLO-CLIA-IgM, Orienter-CLIA-IgM and Maccura-CLIA-IgM were higher than those of others. Overall, the sensitivities of anti-SARS-CoV-2 IgG in different methods were higher than IgM, but the specificity of both anti-SARS-CoV-2 IgG and IgM had no difference among these methods. Generally, antigen-specific IgM can be early detected after pathogen infection and then rapidly decreases in several weeks. In contrast, IgG usually appears later but maintains at a certain level for a long time. Consistent with this notion, it was reported that IgM could be detected in peripheral blood of COVID-19 patients after 3–7 days and that IgG could be detected after 7–8 days (Zhou et al., 2020). It is worthy to note that, the median time from onset to antibody detection was 21 days in the present study, and we speculated that our enrolled patients were in the middle stage of infection or recovery period. This could be used to explain the low sensitivity of anti-SARS-CoV-2 IgM in these patients. Our findings indicated that the COVID-19 patients with decreased level of IgM but with maintained level of IgG may be in the status of recovery. These data suggest that SARS-CoV-2 antibody detection could also play an important role in the treatment monitoring and prognosis of COVID-19.
Due to the highly contagious nature of the disease, even asymptomatic carriers could spread SARS-CoV-2 virus, which made the control of COVID-19 outbreak more difficult (Rothe et al., 2020; Zou et al., 2020). Given that the sensitivities of SARS-CoV-2 antibody assays were high, these assays had great value in screening asymptomatic SARS-CoV-2 carriers. However, we still observed 6 COVID-19 patients with false-negative results of all antibody detection methods (Table 2). The reasons could be as follows. First, low concentration of antibodies could lead to false negative results. As shown in Table 2, the days from onset to antibody detection of five patients were within 7 days, while IgM and IgG levels may be below the detection limit during this period. Second, the heterogeneity of immune response to SARS-CoV-2 in different individuals may cause delayed antibody production in some individuals. Third, the patients with underlying conditions may be one of the important reasons contributing to false-negative results of antibody detection. As shown in Table 2, one patient (No. 19) was a seventy-five-year-old female who might have impaired immunity because of older age (Chandra, 2002). Another three patients (No. 191, No. 219, No. 236) had immunocompromised conditions such as diabetes, lung cancer, and undergoing immunosuppressant treatment, which could affect the producing of antibodies in these patients. In addition, previous study has shown that the possibility of false-negative results of SARS-CoV-2 antibody assays should be considered if the sample was pre-inactivated by heating, which suggests that heat inactivation prior to immunoanalysis is not recommended (Hu et al., 2020). On the other hand, the false-positive results of SARS-CoV-2 antibody assays maybe due to cross-reactivity with anti-HBV, anti-influenza, and rheumatoid factor (Tre-Hardy et al., 2020).
Some limitations of the study should be mentioned. First, since Tongji Hospital was one of designated hospitals for transfer of patients with COVID-19 from other hospitals, the enrolled patients in this study had a relatively prolonged time from onset of symptoms to admission. This is the reason why the median time from onset to antibody detection was 21 days, which could affect the results of antibody detection. Second, we did not continuously monitor the producing of antibodies in the same patients, and further study is needed.
Taken together, the present study confirms that SARS-CoV-2 antibody assays have good performance in the diagnosis and exclusion of COVID-19 patients, especially by using high-throughput technologies (EIAs or CLIAs), which suggests that antibody detection of SARS-CoV-2 may play an important role in the control of COVID-19.
Publisher’s Note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
Statements
Data availability statement
The original contributions presented in this study are included in the article/supplementary material, further inquiries can be directed to the corresponding authors.
Ethics statement
This study was approved by the ethical committee of Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology (TJ-C20200128). Written informed consent for participation was not required for this study in accordance with the national legislation and the institutional requirements.
Author contributions
SW and FW wrote the first draft of the manuscript and involved in data curation and project administration. HL managed reagents. HH and TW researched the literature and conceived the study. WW and MZ collected the specimen. SW and BY involved in experiment operation. MH, ZS, and FW involved in protocol development, gaining ethical approval, patient recruitment, and data analysis. All authors reviewed and edited the manuscript and approved the final version of the manuscript.
Acknowledgments
We thank Liming Cheng for her assistance and guidance in this research.
Conflict of interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
References
1
Arevalo-RodriguezI.Buitrago-GarciaD.Simancas-RacinesD.Zambrano-AchigP.Del CampoR.CiapponiA.et al (2020). False-negative results of initial RT-PCR assays for COVID-19: a systematic review.PLoS One15:e0242958. 10.1371/journal.pone.0242958
2
BajJ.Karakula-JuchnowiczH.TeresinskiG.BuszewiczG.CiesielkaM.SitarzE.et al (2020). COVID-19: Specific and Non-Specific Clinical Manifestations and Symptoms: The Current State of Knowledge.J. Clin. Med.9:1753. 10.3390/jcm9061753
3
BrendishN. J.PooleS.NaiduV. V.MansbridgeC. T.NortonN.BorcaF.et al (2020). Clinical characteristics, symptoms and outcomes of 1054 adults presenting to hospital with suspected COVID-19: a comparison of patients with and without SARS-CoV-2 infection.J. Infect.81937–943. 10.1016/j.jinf.2020.09.033
4
ChandraR. K. (2002). Nutrition and the immune system from birth to old age.Eur. J. Clin. Nutr.56S73–S76. 10.1038/sj.ejcn.1601492
5
ChungM.BernheimA.MeiX.ZhangN.HuangM.ZengX.et al (2020). CT Imaging Features of 2019 Novel Coronavirus (2019-nCoV).Radiology295202–207. 10.1148/radiol.2020200230
6
GudbjartssonD. F.NorddahlG. L.MelstedP.GunnarsdottirK.HolmH.EythorssonE.et al (2020). Humoral Immune Response to SARS-CoV-2 in Iceland.N. Engl. J. Med.3831724–1734. 10.1056/NEJMoa2026116
7
HouH.WangT.ZhangB.LuoY.MaoL.WangF.et al (2020). Detection of IgM and IgG antibodies in patients with coronavirus disease 2019.Clin. Transl. Immunol.9:e01136. 10.1002/cti2.1136
8
HouH.ZhangY.TangG.LuoY.LiuW.ChengC.et al (2021). Immunologic memory to SARS-CoV-2 in convalescent COVID-19 patients at 1 year postinfection.J. Allergy Clin. Immunol.1481481–1492.e2. 10.1016/j.jaci.2021.09.008
9
HuX.AnT.SituB.HuY.OuZ.LiQ.et al (2020). Heat inactivation of serum interferes with the immunoanalysis of antibodies to SARS-CoV-2.J. Clin. Lab. Anal.34:e23411. 10.1002/jcla.23411
10
IslamK. U.IqbalJ. (2020). An Update on Molecular Diagnostics for COVID-19.Front. Cell. Infect. Microbiol.10:560616. 10.3389/fcimb.2020.560616
11
KuoB. J.LaiY. K.TanM. L. M.GohX. C. (2021). Utility of Screening Chest Radiographs in Patients with Asymptomatic or Minimally Symptomatic COVID-19 in Singapore.Radiology298E131–E140. 10.1148/radiol.2020203496
12
LiZ.YiY.LuoX.XiongN.LiuY.LiS.et al (2020). Development and clinical application of a rapid IgM-IgG combined antibody test for SARS-CoV-2 infection diagnosis.J. Med. Virol.921518–1524. 10.1002/jmv.25727
13
Lisboa BastosM.TavazivaG.AbidiS. K.CampbellJ. R.HaraouiL. P.JohnstonJ. C.et al (2020). Diagnostic accuracy of serological tests for covid-19: systematic review and meta-analysis.BMJ370:m2516. 10.1136/bmj.m2516
14
LiuR.HanH.LiuF.LvZ.WuK.LiuY.et al (2020). Positive rate of RT-PCR detection of SARS-CoV-2 infection in 4880 cases from one hospital in Wuhan, China, from Jan to Feb 2020.Clin. Chim. Acta505172–175. 10.1016/j.cca.2020.03.009
15
OngD. S. Y.FragkouP. C.SchweitzerV. A.ChemalyR. F.MoschopoulosC. D.SkevakiC.et al (2021). How to interpret and use COVID-19 serology and immunology tests.Clin. Microbiol. Infect.27981–986. 10.1016/j.cmi.2021.05.001
16
PanL.MuM.YangP.SunY.WangR.YanJ.et al (2020). Clinical Characteristics of COVID-19 Patients With Digestive Symptoms in Hubei, China: A Descriptive, Cross-Sectional, Multicenter Study.Am. J. Gastroenterol.115766–773. 10.14309/ajg.0000000000000620
17
PascarellaG.StrumiaA.PiliegoC.BrunoF.Del BuonoR.CostaF.et al (2020). COVID-19 diagnosis and management: a comprehensive review.J. Intern. Med.288192–206. 10.1111/joim.13091
18
PhelanA. L.KatzR.GostinL. O. (2020). The Novel Coronavirus Originating in Wuhan, China: Challenges for Global Health Governance.JAMA323709–710. 10.1001/jama.2020.1097
19
RalliM.MorroneA.ArcangeliA.ErcoliL. (2021). Asymptomatic patients as a source of transmission of COVID-19 in homeless shelters.Int. J. Infect. Dis.103243–245. 10.1016/j.ijid.2020.12.031
20
RotheC.SchunkM.SothmannP.BretzelG.FroeschlG.WallrauchC.et al (2020). Transmission of 2019-nCoV Infection from an Asymptomatic Contact in Germany.N. Engl. J. Med.382970–971. 10.1056/NEJMc2001468
21
SidiqZ.HanifM.DwivediK. K.ChopraK. K. (2020). Benefits and limitations of serological assays in COVID-19 infection.Indian J. Tuberc.67S163–S166. 10.1016/j.ijtb.2020.07.034
22
SuleW. F.OluwayeluD. O. (2020). Real-time RT-PCR for COVID-19 diagnosis: challenges and prospects.Pan. Afr. Med. J.35:121. 10.11604/pamj.supp.2020.35.24258
23
TemkinE.Healthcare Worker COVID-19 Surveillance Working Group (2021). Extremely low prevalence of asymptomatic COVID-19 among healthcare workers caring for COVID-19 patients in Israeli hospitals: a cross-sectional study.Clin. Microbiol. Infect.27:130.e1–130.e4.
24
TheL. (2020). Emerging understandings of 2019-nCoV.Lancet395:311. 10.1016/S0140-6736(20)30186-0
25
Tre-HardyM.WilmetA.BeukingaI.DogneJ. M.DouxfilsJ.BlaironL. (2020). Validation of a chemiluminescent assay for specific SARS-CoV-2 antibody.Clin. Chem. Lab. Med.581357–1364. 10.1515/cclm-2020-0594
26
WangC.YuH.HorbyP. W.CaoB.WuP.YangS.et al (2014). Comparison of patients hospitalized with influenza A subtypes H7N9, H5N1, and 2009 pandemic H1N1.Clin. Infect. Dis.581095–1103. 10.1093/cid/ciu053
27
WikramaratnaP. S.PatonR. S.GhafariM.LourencoJ. (2020). Estimating the false-negative test probability of SARS-CoV-2 by RT-PCR.Euro Surveill.25:2000568. 10.2807/1560-7917.ES.2020.25.50.2000568
28
XiaoA. T.GaoC.ZhangS. (2020). Profile of specific antibodies to SARS-CoV-2: the first report.J. Infect.81147–178. 10.1016/j.jinf.2020.03.012
29
XiaoS. Y.WuY.LiuH. (2020). Evolving status of the 2019 novel coronavirus infection: proposal of conventional serologic assays for disease diagnosis and infection monitoring.J. Med. Virol.92464–467. 10.1002/jmv.25702
30
ZhouP.YangX. L.WangX. G.HuB.ZhangL.ZhangW.et al (2020). A pneumonia outbreak associated with a new coronavirus of probable bat origin.Nature579270–273. 10.1038/s41586-020-2012-7
31
ZhuN.ZhangD.WangW.LiX.YangB.SongJ.et al (2020). A Novel Coronavirus from Patients with Pneumonia in China, 2019.N. Engl. J. Med.382727–733. 10.1056/NEJMoa2001017
32
ZouL.RuanF.HuangM.LiangL.HuangH.HongZ.et al (2020). SARS-CoV-2 Viral Load in Upper Respiratory Specimens of Infected Patients.N. Engl. J. Med.3821177–1179. 10.1056/NEJMc2001737
33
ZowawiH. M.AlenaziT. H.AlOmaimW. S.WazzanA.AlsufayanA.HasanainR. A.et al (2021). Portable RT-PCR System: a Rapid and Scalable Diagnostic Tool for COVID-19 Testing.J. Clin. Microbiol.59:e03004–20. 10.1128/JCM.03004-20
Summary
Keywords
COVID-19, SARS-CoV-2, serological tests, lateral flow immunoassay, enzyme-linked immunosorbent assay, chemiluminescent immunoassay
Citation
Wu S, Hou H, Li H, Wang T, Wei W, Zhang M, Yin B, Huang M, Sun Z and Wang F (2022) Comparison of the Performance of 24 Severe Acute Respiratory Syndrome Coronavirus 2 Antibody Assays in the Diagnosis of Coronavirus Disease 2019 Patients. Front. Microbiol. 13:876227. doi: 10.3389/fmicb.2022.876227
Received
15 February 2022
Accepted
13 June 2022
Published
08 August 2022
Volume
13 - 2022
Edited by
Lei Zheng, Southern Medical University, China
Reviewed by
Van Thuan Hoang, Thai Binh University of Medicine and Pharmacy, Vietnam; Tatjana Vilibic-Cavlek, Croatian Institute of Public Health, Croatia
Updates
Copyright
© 2022 Wu, Hou, Li, Wang, Wei, Zhang, Yin, Huang, Sun and Wang.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Min Huang, huangminry2013@163.comZiyong Sun, zysun@tjh.tjmu.edu.cnFeng Wang, fengwang@tjh.tjmu.edu.cn
This article was submitted to Virology, a section of the journal Frontiers in Microbiology
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.