ORIGINAL RESEARCH article

Front. Microbiol., 30 May 2022

Sec. Virology

Volume 13 - 2022 | https://doi.org/10.3389/fmicb.2022.893801

Harmonization of Multiple SARS-CoV-2 Reference Materials Using the WHO IS (NIBSC 20/136): Results and Implications

  • 1. Colorado School of Public Health, Center for Global Health, Aurora, CO, United States

  • 2. Division of Infectious Diseases, New York State Department of Health, Wadsworth Center, Albany, NY, United States

  • 3. Department of Pathology, Brigham and Women's Hospital, Boston, MA, United States

  • 4. Division of Rheumatology, Inflammation and Immunity, Harvard Medical School, Brigham and Women's Hospital, Boston, MA, United States

  • 5. Division of Experimental Medicine, Zuckerberg San Francisco General Hospital, University of California, San Francisco, San Francisco, CA, United States

  • 6. Biodesix, Inc., Boulder, CO, United States

  • 7. INSTAND e.V., Society for Promoting Quality Assurance in Medical Laboratories, Duesseldorf, Germany

  • 8. IQVD GmbH, Institut fuer Qualitaetssicherung in der Virusdiagnostik, Berlin, Germany

  • 9. GBD Gesellschaft fuer Biotechnologische Diagnostik mbH, Berlin, Germany

  • 10. Thermo Fisher Scientific, Waltham, MA, United States

  • 11. Department of Immunology, University of Colorado Anschutz Medical Campus, Aurora, CO, United States

  • 12. Oneworld Accuracy, Vancouver, BC, Canada

Article metrics

View details

11

Citations

4,1k

Views

1,6k

Downloads

Abstract

Background:

There is an urgent need for harmonization between severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) serology platforms and assays prior to defining appropriate correlates of protection and as well inform the development of new rapid diagnostic tests that can be used for serosurveillance as new variants of concern (VOC) emerge. We compared multiple SARS-CoV-2 serology reference materials to the WHO International Standard (WHO IS) to determine their utility as secondary standards, using an international network of laboratories with high-throughput quantitative serology assays. This enabled the comparison of quantitative results between multiple serology platforms.

Methods:

Between April and December 2020, 13 well-characterized and validated SARS-CoV-2 serology reference materials were recruited from six different providers to qualify as secondary standards to the WHO IS. All the samples were tested in parallel with the National Institute for Biological Standards and Control (NIBSC) 20/136 and parallel-line assays were used to calculate the relevant potency and binding antibody units.

Results:

All the samples saw varying levels of concordance between diagnostic methods at specific antigen–antibody combinations. Seven of the 12 candidate materials had high concordance for the spike-immunoglobulin G (IgG) analyte [percent coefficient of variation (%CV) between 5 and 44%].

Conclusion:

Despite some concordance between laboratories, qualification of secondary materials to the WHO IS using arbitrary international units or binding antibody units per milliliter (BAU/ml) does not provide any benefit to the reference materials overall, due to the lack of consistent agreeable international unit (IU) or BAU/ml conversions between laboratories. Secondary standards should be qualified to well-characterized reference materials, such as the WHO IS, using serology assays that are similar to the ones used for the original characterization of the WHO IS.

Introduction

There is an urgent need for harmonization between severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) serology platforms and assays prior to defining appropriate correlates of protection and as well inform the development of new rapid diagnostic tests that can be used for serosurveillance as new variants of concern (VOC) emerge (Berry et al., 2020; Ciotti et al., 2021; Giavarina and Carta, 2021; Infantino et al., 2021; Perkmann et al., 2021; Petrone et al., 2021; Knezevic et al., 2022).

Conversion of results from different laboratory methods to a harmonized international unit reduces the interlaboratory/method variability (Cooper et al., 2018; McDonald et al., 2018; Mattiuzzo et al., 2019, 2020; Ciotti et al., 2021; Knezevic et al., 2022). The WHO International Standards (ISs) are considered the highest quality materials to use for comparison between diagnostic methods using international units (Mattiuzzo et al., 2020). The WHO IS for SARS-CoV-2 serology standard is the National Institute for Biological Standards and Control (NIBSC) 20/136 (United Kingdom, 2020). This standard, as most biological standards, was produced in limited quantities, making it difficult to be used exclusively as a calibrant to compare results between multiple SARS-CoV-2 serology assays on a global scale. Therefore, there is a pressing need to increase the availability of appropriate reference materials that are considered equivalent to the WHO IS. Other well-characterized reference samples can be evaluated against the WHO IS to obtain a valid measurement and calibrated to the arbitrary WHO IS values of 1,000 international units per milliliter (IU/ml) for neutralization assays and 1,000 binding antibody units per milliliter (BAU/ml) (National Institute for Biological Standards Control, 2020).

We compared multiple SARS-CoV-2 serology reference materials to the WHO IS to determine their utility as secondary standards, using an international network of laboratories with high-throughput quantitative serology assays. This enabled the comparison of quantitative results between multiple serology platforms. Furthermore, each serology method can derive a BAU/ml (or IU/ml as appropriate) conversion for multiple antigen–antibody combinations within each sample that are scaled to the arbitrary 1,000 BAU/ml value assigned to the WHO IS. We also note that neutralization assays that report IU/ml may additionally be calibrated to the WHO IS.

Materials and Methods

Recruitment of Severe Acute Respiratory Syndrome Coronavirus 2 Serology Reference Materials

Between April and December 2020, 13 well-characterized and validated SARS-CoV-2 serology reference materials were recruited from six different providers (Table 1) (National Institute for Biological Standards Control, 2020; Frederick National Laboratory for Cancer Research, 2021; Oneworld Accuracy, 2021; Thermo Fisher Scientific, 2021; Windsor et al., 2021; Zeichhardt and Kammel, 2021). Reference materials were selected based on the following criteria: originally characterized by the suppliers with the relevant test's thresholds for positive and negative results, are readily available, enough panels will exist after this study to distribute for widespread use, and the providers intend to distribute their reference materials to other (primarily low-resource) laboratories. All the materials were individually evaluated against the WHO IS using previously validated diagnostic tests given in Table 2 and characterized according to the anticipated results shown in Table 1. All the reference materials and diagnostic tests were handled according to manufacturers' and the respective Clinical Laboratory Improvement Amendments (CLIA) laboratory developed test instructions.

Table 1

Institution Type of provider SARS-CoV-2 serology panel name Sample IDs Material type Anticipated results from development
University of Colorado Academic/Research COVID-19 Serology Control Panel (Windsor et al., 2021) CSCP-HR Pooled Convalescent Plasma N-IgG, Total = Reactive; RBD-IgG, Total = Highly Reactive; S-IgG, Total = Reactive
CSCP-WR Pooled Convalescent Plasma, 1:4 dilution of the CSCP_HR N-IgG, Total = Reactive; RBD-IgG, Total = Reactive; S-IgG, Total = Reactive
CSCP-NR Pre-2019 Donor Plasma Non-Reactive
NCI Frederick Lab Government Human SARS-COV-2 Serology Standard (Frederick National Laboratory for Cancer Research, n.d.) NCI Frederick Pooled Convalescent Plasma N-IgG = Reactive; N-IgM = Reactive; S-IgG = Reactive; S-IgM = Reactive
Oneworld Accuracy Commercial COVS434 | SARS-CoV-2 Serology (Oneworld Accuracy, 2021) 1WA-A Single Donor Human Plasma No Ag indication, IgG against SARS-CoV-2, Total = Reactive
1WA-B Single Donor Human Plasma No Ag indication, IgG against SARS-CoV-2, Total = Reactive
1WA-C Single Donor Human Plasma No Ag indication, IgG+IgM against SARS-CoV-2, Total = Reactive
1WA-D Pre-2019 Donor Plasma Non-Reactive
INSTAND Commercial Samples from EQA scheme (416) SARS-CoV-2 (Ak) (Zeichhardt and Kammel, n.d.) 416006 Convalescent Serum of a single donor after infection with human coronaviruses OC43 and HKU1 (single donation, blood collected 2 years after last infection) Non-Reactive
416029 Convalescent Serum of a donor after SARS-CoV-2 infection (single donation, blood collected 154 day after onset of disease) N-IgG, Total = Reactive; RBD/S-IgG, Total = Reactive
416048 Post Pfizer-BioNTech COVID-19 Vaccine donor Serum (single donation, blood collected 63 days after 2nd vaccination; no prior evidence of infection) N-IgG, Total = Non-Reactive; RBD/S-IgG, Total = Reactive
Thermo Fisher Commercial MAS™ SARS-CoV-2 IgG Positive Control Kit (Cat# 10028305) (Thermo Fisher Scientific, n.d.) ThermoFisher Pooled COVID-19 positive human plasma added to difibrinated plasma with ProClin 950 and Sodium azide N-IgG,Total = Reactive; RBD-IgG, Total = Reactive; S-IgG, Total = Reactive
National Institute for Biological Standards and Controls Government NIBSC 20/136 (National Institute for Biological Standards Control, 2020) WHO IS Pooled Convalescent Plasma 1000 BAU/mL for IgM, IgG, and IgA subtypes

Severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) serology harmonization reference material providers.

Table 2

Institution Type of lab Platform Method Antigen targets Antibodies
University of Colorado Academic/ Research Lab-Developed Test SARS-CoV Focus Reduction Neutralization Titer (FRNT) 2019 n-CoV/USA-WA1/2020 Total Ig
Biodesix, Inc. Commercial GenScript cPass Nab Neutralization(Nab) ELISA RBD Total Ig
Bio-Rad Platelia ELISA N IgG, IgM, IgA
Brigham and Women's Hospital Academic/ Clinical Laboratory Developed Test upon Quanterix Simoa HD-X platform Multiplexed Single Molecule Array (MSMA) S, RBD, N, S1 IgG, IgM, IgA
Wadsworth Center, New York State Department of Health Reference/ Public Health Lab-Developed Test upon Luminex Platform Multiplexed microsphere assay (MMA) S, RBD, N, S1, S2 IgG, IgM, IgA, Total Ig
University of Colorado Academic/ Research Lab-Developed Test Multiplex microsphere immunoarray (MIA) N, RBD, S1, S2 IgG

SARS-CoV-2 serology harmonization testing laboratories and methods.

Neutralization Assays

Severe Acute Respiratory Syndrome Coronavirus 2 Focus Reduction Neutralization Test

Vero E6 cells (ATCC, CRL-1586; Manassas, Virginia, USA) were maintained at 37°C in Dulbecco's Modified Eagle Medium (DMEM) (HyClone 11965-084; Logan, Utah, USA) supplemented with 10% fetal bovine serum and 100 U/ml penicillin-streptomycin. SARS-CoV-2 strain 2019 n-CoV/USA-WA1/2020 was obtained from ATCC. The virus was passaged once in Vero E6 cells and titrated by the focus reduction neutralization test (FRNT) on Vero E6 cells. All the work with infectious SARS-CoV-2 was performed in Biosafety Level 3 (BSL3) facilities at the University of Colorado School of Medicine.

The focus reduction neutralization test (FRNT) was performed as previously described (Annen et al., 2021; Schultz et al., 2021; Taylor et al., 2021). Vero E6 cells were seeded in 96-well plates at 104 cells/well. On the next day, serum samples were heat inactivated at 56°C for 30 min and then serially diluted (2-fold, starting at 1:10) in DMEM supplemented with 1% fetal bovine serum (FBS) and 10 mM 4-(2-Hydroxyethyl)-1-piperazine ethanesulfonic acid (HEPES) (Merck, 7365-45-9, Darmstadt, Germany). Approximately, 100 focus-forming units (FFUs) of virus were added to each well and the serum/virus mixture was incubated for 1 h at 37°C. Following co-incubation of serum and virus, medium was removed from cells and the serum/virus mixture was added to the cells for 1 h at 37°C. Serum/virus mixture was removed and cells overlaid with 1% methylcellulose (MilliporeSigma, M0512; Burlington, Massachusetts, USA) in DMEM plus 2% FBS and incubated for 24 h at 37°C. Cells were fixed with 1% paraformaldehyde (PFA) (Acros Organics, 416780030; Morris Plains, New Jersey, USA) for 1 h, washed six times with phosphate-buffered saline-0.05% Tween 20 (PBS-T), and probed with 1 μg/ml of chimeric human anti-SARS-CoV spike antibody (CR3022, Absolute Antibody, Ab01680; Oxford, UK) in Perm Wash Buffer [1X PBS/0.1% saponin/0.1% bovine serum albumin (BSA)] for 2 h at 25°C. After three washes with PBS-T, cells were incubated with goat antihuman immunoglobulin G (IgG) Fc-horseradish peroxidase (HRP) (Southern Biotech, 2014-05; Birmingham, Alabama, USA) diluted at 1:1,000 in Perm Wash Buffer for 1.5 h at 25°C. SARS-CoV-2-positive foci were visualized with TrueBlue substrate (SeraCare, 5510-0030, Milford, Massachusetts, USA) and counted using the CTL BioSpot analyzer and BioSpot software (Cellular Technology Ltd., Shaker Heights, Ohio, USA). The FRNT50 titers were calculated relative to a virus only control (no serum) set at 100%, using GraphPad Prism 9.1.2 default nonlinear curve fit constrained between 0 and 100%.

CPass α-Receptor-Binding Domain (GenScript) Neutralization Antibody Test

The cPass α-receptor-binding domain (RBD) neutralization antibody (nAb) test is a quantitative assay that specifically measures a subset of spike-binding antibodies that can block the interaction between the RBD on the SARS-CoV-2 spike protein and the human host receptor angiotensin-converting enzyme 2 (ACE2) (GenScript, 2021). The assay is performed as a blocking ELISA as described in the Food and Drug Administration (FDA) Emergency Use Authorization (EUA) instructions for use in the cPass™ SARS-CoV-2 Neutralization Antibody Detection Kit. The surrogate virus neutralization test (SVNT) cPass assay was clinically validated and shown to be 100% sensitive and specific when compared to a gold standard plaque reduction neutralization test (PRNT), with qualitative analysis results 100% in agreement (GenScript, 2021). The reference materials were diluted and preincubated 1:1 with RBD protein conjugated to HRP at 37°C for 30 min. The mixture (100 μl) was then added to a 96-well plate coated with human ACE2 receptor protein; the plate was sealed and incubated for an additional 15 min at 37°C. The plate was washed four times with 260 μl/well Wash Solution provided in the kit before addition of 100 μl per well 3,3',5,5'-Tetramethylbenzidine (TMB) substrate for 15 min at room temperature. 50 μl of 1 N sulfuric acid solution was added to each well and the optical density (OD) was measured at 450 nm using a spectrophotometer. The nAb assay readout was percent signal inhibition by neutralizing antibodies, which was calculated to be the OD value of the sample relative to the OD of the negative control subtracted from one (Tan et al., 2020; Petrone et al., 2021; Taylor et al., 2021). The positive cutoff results are ≥ 30% signal inhibition and results <30% are reported negative based on previously conducted clinical validation studies (Petrone et al., 2021).

Binding Antibody Assays

Platelia α-Nucleocapsid Total Antibody Test

The Platelia α-nucleocapsid (anti-N) total antibody test detects antibodies [IgG, immunoglobulin M (IgM), and immunoglobulin A (IgA) combined; Bio-Rad Incorporation] to the nucleocapsid protein. The assay is performed as a one-step antigen capture ELISA as described in the FDA EUA instructions for use for the Platelia SARS-CoV-2 Total Antibody Test Kit (Bio-Rad, 2021). The diluted plasma (1:5) and the WHO IS (1.5-fold serial dilution series up to 8 times, starting at 1:90 dilution) were mixed with SARS-CoV-2 nucleocapsid protein coupled with horseradish peroxidase (HRP) enzyme at a 1:1 ratio and 100 μl added to a 96-well plate coated with the nucleocapsid protein. The plate was covered with an adhesive plate sealer and incubated at 37°C for 1 h. The plate was then washed five times with the Working Washing Solution provided in the kit and 200 μl of the Enzyme Development Solution was added to each well. After a 30-min incubation in the dark at room temperature (18–30°C), the reaction was stopped by adding 100 μl per well of an acidic stopping solution and mixing thoroughly before measuring the OD at 450 nm using a spectrophotometer. The assay readout was a ratio of the specimen OD to cutoff control OD. A positive specimen-to-cutoff ratio ≥ 1.0 and <0.8 is negative and in between is reported equivocal with the recommendation of another specimen collected 3 days later. The Platelia assay has FDA EUA clearance for a qualitative interpretation of results (Bio-Rad, 2021).

Simoa Serology Assay

Simoa assays for IgG, IgA, and IgM against four SARS-CoV-2 targets (spike, S1, nucleocapsid, and RBD) were performed as previously described (Norman et al., 2020). Reference materials were diluted 1:250-, 1:1,000-, 1:4,000-, and 1:16,000-fold in Homebrew Detector/Sample Diluent (Quanterix Corporation, Product code: 101359, Billerica, Massachusetts, USA). Four antigen-conjugated capture beads were mixed and diluted in Bead Diluent (Quanterix Corporation, Product code: 101362, Billerica, Massachusetts, USA), with a total of 500,000 beads per reaction (125,000 of each bead type). Biotinylated antibodies were diluted in Homebrew Detector/Sample Diluent to final concentrations of IgG (Bethyl Laboratories A80-148B; Montgomery, Texas, USA): 7.73 ng/ml, IgA (Abcam ab214003, Waltham, Massachusetts, USA): 150 ng/ml, and IgM (Thermo Fisher Scientific, MII0401, Pittsburgh, Pennsylvania, USA): 216 ng/ml: Streptavidin-β-galactosidase (SβG) concentrate (Quanterix Corporation, Product code: 1013397, Billerica, Massachusetts, USA) was diluted to 30 pM in SβG Diluent (Quanterix Corporation, Product code: 100376, Billerica, Massachusetts, USA). The serology assay was performed on the HD-X Analyzer (Quanterix) in an automated three-step assay. Average enzymes per bead (AEB) values were calculated by the HD-X Analyzer software (Norman et al., 2020).

Multiplexed Microsphere Assay

Specimens were assessed for the presence of antibodies reactive with SARS-CoV-2 using a multiplexed microsphere assay (MMA). Recombinant SARS-CoV-2 full-length spike, nucleocapsid, S2 (The Native Antigen Company, REC31868, REC31812, and REC31807, respectively, Kidlington, Oxfordshire, UK), RBD, and S1 (Mass Biologics, https://www.umassmed.edu/massbiologics, Boston, Massachusetts, USA) subunits were covalently linked to the surface of fluorescent microspheres (Luminex Corporation, LC10047, LC10006, LC10071, LC10061, and LC10023, respectively, Austin, Texas, USA). Serum samples (25 μl at doubling dilutions from 1:50 to 1:102,400) and antigen-coupled microspheres (25 μl at 5 × 104 microspheres/ml) were mixed and incubated 30 min at 37°C. Serum-bound microspheres were washed and incubated with phycoerythrin (PE)-conjugated secondary antibody. The PE-conjugated antibodies were chosen to specifically recognize total Ig (Pan-Ig), IgM, IgA, and IgG (Southern Biotechnology Associates Incorporation, 2010–2009, 2020–2009, 2050–2009, and 2040–2009, respectively, Birmingham, Alabama, USA). After washing and final resuspension in buffer, the samples were analyzed on the FlexMap 3D analyzer using xPONENT software (Luminex Corporation, version 4.3, Austin, Texas, USA).

Multiplexed Microsphere Immunoassay (MIA)

A multiplexed microsphere immunoassay (MIA) was developed using BioLegend carboxylated LEGENDplex microbeads to simultaneously quantify IgG and IgA against the spike RBD and nucleocapsid of the Wuhan strain of SARS-CoV-2 (BEIresources.org, NR-52366, North Bethesda, Maryland, USA), three variants of concern beta gamma, delta (BEIresources.org, NR-54004/54005, North Bethesda, Maryland, USA), three season coronavirus strains (OC43, 229E, and HKU1) (BEIresources.org, NR-53713, North Bethesda, Maryland, USA), and tetanus toxoid (TT) (MilliporeSigma, #582231-25UG, St. Louis, Mosby, USA) as a positive control. Bovine serum albumin (BSA) (10%) (MilliporeSigma, #A7030, St. Louis, Mosby, USA) conjugated beads were used as a negative control. Multiplex bead protein conjugation, sample incubation, and flow cytometric analysis were performed as previously described (Schultz et al., 2021). Geometric mean fluorescence intensity (gMFI) of the IgG/IgA for each sample and dilution was captured with the CytoFLEX S Flow Cytometer (Beckman Coulter, Indianapolis, Indiana, USA) and analyzed with FlowJo (version 10.7.1; BD Biosciences, San Jose, California, USA). Prism (version 8.4.3, GraphPad) was used to plot data (Schultz et al., 2021).

Statistical Analysis

Parallel-line assay (PLA) was used to compare all the secondary standard candidate samples to the WHO IS; all the analytes were set at 1,000 IU or BAU/ml (Finney and Schild, 1966). All the samples were tested in triplicate with each diagnostic test at dilutions within each assay's given linear range for the WHO IS. Data were analyzed using PLA analysis using R 3.5.0 that we created (R Core Team, 2021). Sample results and their corresponding dilutions were log-transformed and assessed for parallelism using the relative slope calculated individually between each sample and the WHO IS. To ensure the assumption of parallelism for PLA analysis to occur, a relative slope between 0.8 and 1.2 was considered parallel and samples with relative slopes outside the range were excluded from further analysis because they violated the PLA assumption of parallel lines (Mattiuzzo et al., 2020). The relative potency was calculated for each sample whose slope was within 20% of the WHO IS slope. Relative potencies were then converted to IU or BAU/ml based on the assay used (Finney and Schild, 1966) and parametric bootstrapping was used to calculate CIs for each sample (B. Efron, 1979; Landes et al., 2019). The full reproducible code and readme file are both available at: github.com/yroell/pla. and the overview of our created PLA analysis is shown in Supplementary Figure 1 showing an overview used for each sample. IU and BAU/ml conversions were then compared for interassay variability using percent coefficient of variation (%CV) (Reed et al., 2002; Wood et al., 2012).

Results

Analysis of Samples and Binding Antibody Unit Conversions

Thirteen samples (including the WHO IS) from six different providers (Table 1) were tested using six different SARS-CoV-2 serology diagnostic platforms. Twenty-one total antigen–antibody (Ag–Ab) combinations were evaluated. Three of the platforms were multiplexed platforms targeting multiple Ag–Ab combinations. The remaining three platforms consisted of two SARS-CoV-2 neutralization tests and one nucleocapsid-specific ELISA (Table 2). Each laboratory performed serial dilutions of the WHO IS to establish the linear range of the WHO within each testing platform. All the reference samples were then serial diluted within the WHO IS linear range and tested in triplicate.

Results from each laboratory were compiled and evaluated using PLA. Reference material samples were considered “parallel” if their relative slope against the WHO IS was between 0.8 and 1.2. Samples that failed to fall within the range were excluded from further analysis. For each sample at each Ag–Ab combination, BAUs (or IUs for neutralization tests) were calculated using sample relative potency. BAU conversions for each sample are shown in Table 3 and Figure 1 summarizes the BAU or IU conversions for each sample at each analyte. The IgG-spike analyte had more consistent BAU or IU/ml conversions between methods, regardless of the sample type. For example, the multiplexed microsphere assay (MMA) and MSMA results for the oneworld accuracy a(b,c,d) (1WA-A) sample were 281 (95% CI = 280–282) and 299 (95% CI = 294–304) BAU/ml, respectively; for the 1WA-C sample, the MMA and MSMA results were 938 (95% CI = 936–940) and 1,150 (95% CI = 1,142–1,178) BAU/ml, respectively. Further, for the covid-19 serology control panel-high reactive (CSCP-HR) sample, the MMA and MSMA results were 507 (95% CI = 506–508) and 477 (95% CI = 469–485) BAU/ml, respectively, and for the covid-19 serology control panel-weak reactive (CSCP-WR) sample, the MMA and MSMA results were 141 (95% CI = 0) and 120 (95% CI = 118–122) BAU/ml, respectively. Oppositely, the IgG-nuclecapsid analytes saw wider differences in BAU or IU/ml conversions between methods within the CSCP-HR, covid-19 serology control panel-non reactive (CSCP-NR), Thermo Fisher Scientific, 416,026, and 416,048 samples.

Table 3

416006 416029 416048 1WA–A 1WA–B 1WA–C 1WA–D CSCP–HR CSCP–NR CSCP–WR NCI Frederick Thermo Fisher
Ab Ag Method BAU 95% CI BAU 95% CI BAU 95% CI BAU 95% CI BAU 95% CI BAU 95% CI BAU 95% CI BAU 95% CI BAU 95% CI BAU 95% CI BAU 95% CI BAU 95% CI
Total Ig N ELISA NA NA NA NA NA NA 1986 1958–2014 NA NA NA NA NA NA 497 594–500 NA NA NA NA 1116 1113–1119 NA NA
MMA NA NA 109 0 0 0 658 656–660 126 125–127 496 495–497 NA NA 579 577–581 NA NA 160 159–161 783 781–785 82 0
RBD MMA NA NA 49 0 1542 1,538–1,548 167 0 204 203–205 851 849–853 NA NA 615 613–617 NA NA 170 169–171 585 583–587 39 0
Nab NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA
S MMA NA NA 93 0 1860 1,854–1,866 315 314–316 260 259–261 972 969–975 NA NA 474 473–475 NA NA 129 0 980 977–983 45 0
S1 MMA NA NA 68 0 1865 1,859–1,871 294 293–295 266 265–267 936 934–938 NA NA 507 506–508 NA NA 136 0 765 763–767 36 0
S2 MMA NA NA 68 0 115 0 175 0 36 0 237 236–238 NA NA 230 229–231 NA NA 46 0 718 716–720 25 0
WV FRNT NA NA NA NA NA NA NA NA NA NA 1054 1045–1063 NA NA NA NA NA NA NA NA NA NA NA NA
IgG N MIA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA NA 745 723–767 NA NA
MMA NA NA 96 0 0 0 585 583–587 125 124–126 373 372–374 NA NA 616 614–618 NA NA 161 160–162 792 2 62 0
MSMA NA NA 35 34–36 156 149–173 251 248–254 110 106–114 419 406–432 NA NA 135 132–138 NA NA 40 39–41 856 843–869 36 35–37
RBD MIA NA NA NA NA NA NA 82 79–85 129 126–132 846 832–860 0 0 69 67–71 NA NA 7 0 489 480–498 11 12–Oct
MMA NA NA 54 0 2306 2,294–2,316 124 0 189 188–190 917 914–920 NA NA 815 813–817 NA NA 224 223–225 768 766–770 48 0
MSMA NA NA 58 56–60 2060 2,020–2,100 142 139–145 167 160–174 924 905–943 NA NA 589 576–602 NA NA 123 119–127 691 676–706 58 56–60
S MMA NA NA 97 0 2115 2,109–2,121 281 280–282 243 242–244 938 936–940 NA NA 507 506–508 NA NA 141 0 1090 1088–1092 43 0
MSMA NA NA 307 301–313 2749 2,703–2,792 299 294–304 424 415–433 1160 1,142–1,178 NA NA 477 469–485 NA NA 120 118–122 2067 2032–2102 120 118–122
S1 MIA NA NA NA NA NA NA 36 31–41 81 75–87 703 781–725 NA NA 24 23–25 NA NA NA NA 463 448–478 NA NA
MMA NA NA 74 0 2453 2,441–2,465 271 270–272 260 259–261 883 881–885 NA NA 609 608–610 NA NA 167 0 925 923–927 38 0
MSMA NA NA 75 73–77 2411 2,373–2,450 108 106–110 149 146–152 783 770–796 NA NA 393 386–400 NA NA 97 95–99 647 636–658 35 34–36
S2 MIA NA NA NA NA NA NA 29 27–31 20 19–21 92 90–94 NA NA 10 0 NA NA 1 0 443 436–450 1 0
MMA NA NA 61 0 113 0 160 159–161 18 0 210 0 NA NA 247 246–248 NA NA 43 0 1580 1576–1584 20 0
IgM N MMA NA NA 288 287–289 89 0 603 601–605 384 382–386 8360 8336–8384 NA NA 1387 1381–1393 NA NA 349 349–351 531 529–533 118 117–119
MSMA 120 112–128 417 497–437 228 211–245 NA NA NA NA 1383 1,314–1,452 18 16–20 1894 1702–2086 24 22–26 352 325–379 NA NA NA NA
RBD MMA NA NA 25 0 10 0 232 231–233 202 0 436 435–437 NA NA 375 374–376 NA NA 94 0 273 272–274 3 0
MSMA 6 0 24 0 27 0 200 198–202 174 172–176 507 503–511 4 0 372 368–376 NA NA 94 92–96 279 276–282 NA NA
S MMA NA NA 17 0 13 0 236 235–237 261 0 595 594–596 NA NA 583 582–584 NA NA 139 0 215 214–216 4 0
MSMA 13 0 21 20–22 160 158–162 333 330–336 313 308–318 819 811–827 NA NA 1280 1,268–1,292 NA NA 267 264–270 365 362–368 NA NA
S1 MMA NA NA 22 0 13 0 217 216–218 221 220–222 556 555–557 NA NA 459 458–460 NA NA 111 0 224 223–225 3 0
MSMA 4 0 14 13–15 63 61–65 310 307–313 266 262–270 299 296–302 NA NA 723 716–730 NA NA 152 151–153 213 210–216 NA NA
S2 MMA NA NA NA NA NA NA NA NA 541 533–549 198 187–209 NA NA NA NA NA NA NA NA NA NA NA NA
IgA N MMA NA NA 6 0 0 0 30 29–31 41 40–42 290 289–291 NA NA 182 181–183 NA NA 60 0 2003 1978–2028 55 0
MSMA NA NA 175 171–179 NA NA 340 335–345 347 338–356 900 891–909 NA NA 200 196–204 74 72–76 109 107–111 NA NA NA NA
RBD MMA NA NA 35 0 99 0 209 208–210 141 0 437 436–438 NA NA 211 210–212 NA NA 60 0 274 273–275 10 0
MSMA 2 0 69 68–70 224 222–226 279 277–281 211 210–212 557 552–562 1 0 216 214–218 0 0 51 49–53 NA NA NA NA
S MMA NA NA 133 132–134 326 324–328 726 723–729 332 331–333 465 463–467 NA NA 369 367–370 NA NA 97 0 432 430–434 17 0
MSMA 37 36–38 425 422–428 1659 1646–1672 800 794–806 642 638–646 941 933–949 10 0 579 574–584 NA NA 146 143–149 NA NA NA NA
S1 MMA NA NA 100 99–101 279 277–281 517 515–519 305 304–306 564 561–567 NA NA 394 391–397 NA NA 117 116–118 392 389–395 18 0
MSMA NA NA 254 252–256 844 837–851 1062 1053–1,071 473 470–476 559 554–564 NA NA 382 379–385 NA NA 90 88–92 NA NA NA NA
S2 MMA NA NA 70 69–71 12 0 134 133–135 56 0 10 0 NA NA 496 493–499 NA NA 103 102–104 194 192–196 9 0

Binding antibody unit conversions for serology harmonization samples.

Table 3 breaks down the computed BAU or IU/mL conversions and bootstrapped 95% Confidence Intervals of each method by their measured antibody and antigen analyte combination. *NA indicates the result was either not measured by the particular assay, or the relative slope fell outside of 20% and therefore not possible to convert to International Units or Binding Antibody Units/mL (BAU); Ab, antibody; Ag, antigen; N, Nucleocapsid; S, Spike; RBD, Receptor Binding Domain; WV, Whole Virus; MSMA, multiplexed single molecule array; MIA, multiplexed microsphere immunoarray; MMA,multiplexed microsphere assay; Nab, neutralization; ELISA, enzyme–linked immunosorbent assay; FRNT, focus–reduction neutralization titer. A more comprehensive breakdown of the individual samples and test methods are found in (Tables 1, 2) respectively.

Figure 1

Figure 1

Aggregated scatterplot of computed binding antibody unit conversions for each reference sample. The following samples are represented by each subfigure: (a) 1WA-A; (b) 1WA-B; (c) 1WA-C; (d) 416029; (e) 416048; (f) CSCP-HR; (g) CSCP-WR; (h) NCI Frederick; (i) ThermoFisher. MIA, multiplexed microsphere immunoarray; MMA, multiplexed microsphere assay; Nab, neutralization; ELISA, enzyme-linked immunosorbent assay; FRNT, focus-reduction neutralization titer. The following samples were removed because they were classified as “non-reactive” during testing: 1WA-D, 416006, CSCP-NR.

Interlaboratory/Method Binding Antibody Units Concordance

Once BAUs were calculated, we evaluated results for overall intermethod concordance if multiple laboratories yielded results for each Ag–Ab combination using percent coefficient of variation (%CV). Lower %CV values (<21%) indicate that results are highly agreeable between laboratories. None of the samples tested yielded universally high concordance between methods (regardless of Ag–Ab combination). For specific Ag–Ab combinations, there was no universal concordance between methods regardless of the sample tested. Samples 1WA-A, 1WA-B, and 1WA-C saw high concordance between laboratories for both the IgG and IgM bound to S, RBD, and N antigens (%CV range between 5 and 57%). CSCP-HR and CSCP-WR were highly concordant within the IgG-S combination (5 and 12%, respectively) and IgA and IgM bound to S, RBD, and N antigens (%CV range between 2 and 53%). Sample 416,029 was highly concordant between laboratories for IgG-RBD and IgG-S1 combinations (%CV 6 and 1%, respectively). Sample 416,048 saw high concordance with IgG S, S1, and RBD combinations (%CV = 19, 2, and 8%, respectively). The highest %CV value in Figure 2 was found in sample 416,006 at the IgG-N analyte, which is likely because that particular sample was acquired from a postvaccinee individual and was not highly reactive to IgG-N during its characterization (Table 1). The National Cancer Institute (NCI) Frederick sample saw good overall concordance between laboratories for all the measured analytes. The IgG-S1 of the Thermo Fisher Scientific sample was highly concordant between laboratories (%CV = 6%). Result concordance between testing methods at each Ag–Ab combination is shown in Figure 2.

Figure 2

Figure 2

Inter-method concordance of binding antibody unit conversions among reference materials for each analyte. %CV, percent coefficient of variation; Light blue, Higher concordance between methods; dark blue, lower concordance between methods; blank, not enough labs yielded PLA results to compute concordance. Thick Black outlines indicate that the particular analyte was evaluated by that sample's provider. The following samples were removed because they were classified as “non-reactive” during testing: 1WA-D, 416006, CSCP-NR; The following Ag-Ab combinations were removed due to lack of sufficient PLA data due to linearity violations: Whole-Virus-Total, S2-Total, S2-IgM, S2-Iga, S2-Total, S-Total, RBD-Total.

Discussion

We evaluated multiple candidate reference materials against the WHO IS (NIBSC 20/136) to determine whether secondary standards could be established. We then evaluated the applicability of using arbitrary BAU conversions to compare results between laboratories and serology diagnostic methods. Many seroprevalence studies use different serology assays to estimate transmission and/or herd immunity. The differences between assays make it nearly impossible to harmonize and establish a reliable limit of detection. A reference standard would theoretically allow for comparison between such studies.

A number of studies have determined that internal standards provided by the WHO for various pathogens may be useful and should be used to compare results across laboratories and diagnostic methods to help establish correlates of protection for SARS-CoV-2 and other high-threat pathogens (Cooper et al., 2018; McDonald et al., 2018; Mattiuzzo et al., 2019, 2020; Ciotti et al., 2021; Knezevic et al., 2022). For example, when assessing candidate reference materials for enterovirus serology, one study evaluating the interassay variability for both the raw neutralization titer and the calculated relative potencies found a marked decrease in interassay variability. Their calculated percent geometric coefficient of variation (%GCV) was between 30 and 94% (Cooper et al., 2018), indicating that although their candidate materials had decreased interassay variability after the results were converted to a harmonized metric, it is difficult to know what is considered an acceptable coefficient of variation across methods in this context. Two additional studies that evaluated candidate reference materials for Zika virus found similar improvements to intermethod concordance with the reference material, yet GCVs remained exceptionally high, suggesting that a threshold for acceptable intermethod concordance may be difficult, if not impossible, to establish in these contexts (Mattiuzzo et al., 2019; Berry et al., 2020).

Finally, the developers of the WHO IS conducted a robust evaluation of the candidate standard that included 125 different SARS-CoV-2 serology assays (Mattiuzzo et al., 2020; Knezevic et al., 2022). When evaluating the interassay variability of results, they stratified their comparisons into neutralization assays, ELISAs, and “other” assays relative to what is now the WHO IS. Interassay variability between neutralization assays for samples tested relative to the WHO IS did not fall below 67% (%GCV range 67–250%). The interassay variability for the WHO IS itself was 241% (Mattiuzzo et al., 2020). Similar results were found when comparing ELISA methods and there were no data that evaluated the “other” methods included in the characterization. The assignment of an arbitrary 1,000 IU for neutralization assays and 1,000 BAU/ml for other assays—despite the large interassay variability relevant to the WHO IS—does not account for the vast differences between assays. Additionally, the interassay variability between all the methods used was not presented, which, therefore, makes it difficult to fully understand how best to harmonize results between multiple laboratories in order to assess correlates of protection. This study evaluated the interassay variability relative to the WHO IS across all the methods used. We also present the variability between laboratories for multiple Ag–Ab combinations to differentiate which ones are more likely to remain consistent or be highly variable within each sample.

Other studies also suggest that SARS-CoV-2 serology tests cannot be calibrated to the same measurement “ruler” and results compared between assays (Cooper et al., 2018; Bradley et al., 2021; Castillo-Olivares et al., 2021; Giavarina and Carta, 2021; Infantino et al., 2021; Perkmann et al., 2021; Solastie et al., 2021; Knezevic et al., 2022). It is also important to note that the IU or BAU assigned to the WHO IS is arbitrary and not based on an analytical concentration measurement. Additionally, results attained using the WHO IS are highly variable between assays. Our results demonstrate that any reference material should be characterized independently for each assay and it is not advisable to compare quantitative IU or BAUs between different assays. Therefore, arbitrary BAUs that were not calculated should not be used to benchmark any characterizations made for other reference materials, especially candidate secondary standards (Bradley et al., 2021; Giavarina and Carta, 2021; Perkmann et al., 2021). International Standards are not able to account for the wide variety of reagent formulations and nuances between testing methods using a universal metric such as an IU or BAU conversion. Finally, our findings show the qualification of secondary standards using the WHO IS using the 1,000 IU or BAU as a baseline metric that does not yield consistent IU or BAU conversions between assays.

Regardless of the pathogen, many other evaluations of “candidate” reference materials from the WHO have revealed a high degree of interassay and interlaboratory variability during characterization (Bozsoky, 1963; Holder et al., 1995; Wood et al., 2012; Dimech et al., 2013; Cooper et al., 2018; McDonald et al., 2018; Mattiuzzo et al., 2019; Kempster et al., 2020; Timiryasova et al., 2020). Although these findings cannot be verified within the context of this study, our findings reinforce that SARS-CoV-2 serology reference materials face the same challenges and interpretation issues that other groups have seen (Mattiuzzo et al., 2020; Castillo-Olivares et al., 2021; Ciotti et al., 2021; Giavarina and Carta, 2021; Infantino et al., 2021; Kristiansen et al., 2021). Standardization of IU or BAU values for candidate secondary standards relative to the WHO IS could not be achieved across different laboratory assays using methods consistent with the NIBSC characterization of the WHO IS (Mattiuzzo et al., 2020). This calls into question the feasibility of standardizing different serology assays in the future and what this means when interpreting seroprevalance or distinguishing between natural infections and vaccine-induced responses.

Limitations

Some limitations are noted for this study. Among our laboratories, some were unable to yield relative potency values to use for a BAU/ml conversion for certain Ag–Ab combinations. Our criteria for PLA parallelism were more strict (relative slope = 0.8–1.2) than the standards set by the NIBSC (relative slope = 0.8–1.25) during the initial characterization of the NIBSC 20/136 because we wanted to set a more consistent range for relative slopes on either end (Mattiuzzo et al., 2020). Furthermore, the NIBSC does not clarify why they established an acceptable relative slope range of 0.8–1.25 was chosen. Manufacturing convalescent plasma/serum samples at scale is not common practice due to low volume donations and lot-to-lot differences. So, unlike molecular standards, it is difficult to generate large batches and consistent lots for harmonization or even for testing (in a postharmonization world). Two of the six methods used were neutralization assays; one did not yield relative potency for any samples tested and the other only yielded a relative potency for a single sample. Even after log, the raw candidate sample neutralization results failed to fall within the parameters to accurately perform PLA (Taylor et al., 2021).

Similar studies have used a variety of different interassay comparability methods that include, but are not limited to the Spearman's rank correlation coefficient, the Mann–Whitney U tests, and Bablok regression (McDonald et al., 2018; Castillo-Olivares et al., 2021; Giavarina and Carta, 2021; Perkmann et al., 2021). Percent coefficient of variation (%CV) is a flexible metric commonly used in clinical laboratories and the developers of International Standards to evaluate interassay, intralaboratory, and lot-to-lot variations (Reed et al., 2002; Mattiuzzo et al., 2020). Furthermore, each of the example of alternative comparison methods exclude outlier results from analysis, which biases comparisons to appear erroneously “better” in a study context where outlier laboratory results are important to consider when determining the effectiveness of candidate reference materials.

The MMA method tested the WHO standard as nonreactive (no reaction present) for IgM against the nucleocapsid and spike S2 and indeterminate (no result due to PLA violation) for IgA against the nucleocapsid. Even though the assay was sensitive enough to give values for these analytes, these numbers are below what was consider reactive. Because the standard was so low and set to 1,000 BAU/ml, any sample with detectable but similarly low quantities of an analyte will give a misleadingly high BAU/ml value and should be interpreted with caution.

Finally, each method in this study used different formulations of commercial reagents as noted in the Materials and Methods section. For coronavirus disease 2019 (COVID-19) and detection of anti-SARS-CoV-2 antibodies, the field is complicated by multiple antigen sources, multiple host experiences (one or more natural infections and/or vaccines and boosters), multiple variants, and multiple test platforms. This makes it very difficult to achieve harmony. The nuanced differences between these reagent, platforms, and host experiences might contribute to the differences between IU and BAU conversions. Serology is extremely dense with methods and tests, regardless of the pathogen, which highlights the difficulty of applying the same standards for interpretation because it does not account for the nuances that accompany a wide range of assays. This highlights the need for a more precise interpretation of reference material characterizations, so these differences can be accounted in future studies and allow for better harmonization of results between methods.

Conclusion

Harmonization of serology reference materials will increase the accessibility of reference materials—particularly in low-resource settings, provided the methods used for comparison are accurate and reliable. Our findings indicate that the arbitrary units of the WHO IS are not an accurate means to compare SARS-CoV-2 serology results between different laboratories or methods. This study also shows that even after IU or BAU conversion, candidate secondary material results are still drastically different between laboratory methods. Both the International Standards and candidate secondary standards should only be used to compare the results within the same laboratory methods, provided they are using identical testing platforms, protocols, and reagent formulations (Bradley et al., 2021; Giavarina and Carta, 2021; Perkmann et al., 2021). This must be highlighted by regulatory bodies to accurately portray the use of the WHO IS as an assay calibrator during development or external quality assurance material for intramethod comparison, not as a universal comparator (Holder et al., 1995; Infantino et al., 2021).

Finally, despite some concordance between laboratories, qualification of secondary materials to the WHO IS using arbitrary IU or BAU/ml does not provide any benefit to the reference materials overall, due to the lack of consistent agreeable IU or BAU/ml conversions between laboratories. Secondary standards should be qualified to well-characterized reference materials, such as the WHO IS, using serology assays that are similar to the ones used for the original characterization of the WHO IS. However, secondary standards are useful if qualified using similar assays as the original characterization as source traceability for they can be used for intraassay adjustments and can be used in external quality assessment to identify binding to antigen(s) presented in an assay to a reference, thereby providing intralaboratory operations (Table 4).

Table 4

Topic Recommendation(s)
Regulatory Bodies •Replace the process that qualifies candidate secondary materials to an international standard with standards or best practices set for the “characterization” process of any potential reference materials using historical development of WHO IS' as a framework. *This will elevate the quality standards for characterization of samples.*
•Regulatory bodies must also require more precise interpretation of how to use particular reference materials based on the results from their characterization. *These interpretations must take into account the nuances of reagent formulation, testing platform, and the results interpretation in a clinical setting. *
•Once these interpretations are more precise, future studies can then appropriately compare the results between seroprevalance studies for SARS-CoV-2 and potentially other incoming pathogens of interest.
Reference Material Characterization •When characterizing reference materials, the methodology, reagent formulation, and validation information must be shown and included in the interpretation of reference material testing results. Different assays with different reagent formulations might yield slightly different results.
•Establish a minimum number of laboratory methods to include when characterizing potential reference materials.
•Require that the development, manufacturing, and distribution of secondary standards align with Good Manufacturing Practices.
•Establish a minimum list of pathogens to test for when determining sample microbial bioburden.
•Establish a list of minimum requirements for “suitable assay” used to demonstrate reference material expected immunological activity.
•Establish an acceptable level of concordance (%GCV or % CV) between laboratories for the average BAU IU conversion to be considered “reliable.”
Interpretation •Clarify that reference material (international standards and secondary standards) characterization is extremely assay and context dependent, which can affect accuracy of result interpretations. Similar tests with similar reagents must be used when comparing BAU conversions, and seroprevalence study results.
•Revoke the encouraged removal of outlier method results during sample characterization. Exclusion of outlier laboratory data that fall within the PLA assumptions makes reference materials less comparable between methods which might remove the ability to adequately compare results between seroprevalance studies.
•In order to continue using any WHO IS after their supply runs out, consider the development artificial IS for serology.
•Clarify and establish that the intended use of standard reference materials is for external quality assurance schemes, comparing results between studies using similar assays or reagents, and be used as “anchors” by testing the same standards in the beginning and the end of a longitudinal research study. Which will attest to the quality of the results presented by that research study.

Recommendations for future development, use, and interpretation of International and Secondary Standards.

Funding

This study is an expansion of the original COVID-19 Serology Control Panel Study funded by the Bill & Melinda Gates Foundation.

Publisher's Note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Statements

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Author contributions

WW oversaw the entire study operation, conducted the analyses, constructed the tables, and drafted the manuscript. YR conducted the data analyses in R, built an open source R script for the parallel-line assay (github.com/yroell/pla), and built all the figures (both the manuscript and supplementary). ML provided data analysis oversight. MC assisted WW with the overall study design, execution, and result interpretation. HT and WL provided their open source parallel-line assay analysis program to use as a comparison to the R script developed by YR. WW, HZ, MK, HW, AD, and DT contributed SARS-CoV-2 serology reference materials. The remaining authors tested the samples in their respective laboratorios. All authors contributed insights, edits, revisions, and interpretations to the final manuscript.

Acknowledgments

We would like to acknowledge the Coronavirus Standards Working Group (CSWG) at the Joint Initiative for Metrology in Biology (JIMB) at Stanford University who conducted a similar study for SARS-CoV-2 molecular reference materials. Their study and expertise provided the logistical foundation that helped to ensure the success of this study's operation.

Conflict of interest

GP and LP was employed by Biodesix, Inc. HZ and MK was employed by INSTAND e.V., IQVD GmbH, and GBD Gesellschaft fuer Biotechnologische Diagnostik mbH. HW was employed by Thermo Fisher Scientific. AD and DT were employed by company Oneworld Accuracy. The authors declare that this study received funding from the Bill & Melinda Gates Foundation. The funder was not involved in the study design, collection, analysis, interpretation of data, the writing of this article or the decision to submit it for publication. The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fmicb.2022.893801/full#supplementary-material

References

  • 1

    Annen K. Morrison T. E. DomBourian M. G. McCarthy M. K. Huey L. Merkel P. A. et al . (2021). Presence and short-term persistence of SARS-CoV-2 neutralizing antibodies in COVID-19 convalescent plasma donors. Transfusion. 61, 11481159. 10.1111/trf.16261

  • 2

    B. Efron . (1979). Bootstrap methods: another look at the jackknife. Ann. Stat.7, 126. 10.1214/aos/1176344552

  • 3

    Berry N. Kempster S. Ham C. Jenkins A. Hall J. Page M. et al . (2020). Passive immunisation of convalescent human anti-Zika plasma protects against challenge with New World Zika virus in cynomolgus macaques. NPJ Vaccines. 5, 86. 10.1038/s41541-020-00234-y

  • 4

    Bio-Rad. (2021). Platelia SARS-CoV-2 Total Ab Instructions for Use. Available online at: https://www.fda.gov/media/137493/download

  • 5

    Bozsoky S (1963). The problem of standardization in rheumatoid arthritis serology. Arthritis Rheumat.6, 641649. 10.1002/art.1780060508

  • 6

    Bradley B. T. Bryan A. Fink S. L. Goecker E. A. Roychoudhury P. Huang M.-L. et al . (2021). Anti-SARS-CoV-2 antibody levels measured by the AdviseDx SARS-CoV-2 assay are concordant with previously available serologic assays but are not fully predictive of sterilizing immunity. J. Clin. Microbiol.59, e0098921. 10.1128/JCM.00989-21

  • 7

    Castillo-Olivares J. Wells D. A. Ferrari M. Chan A. C. Y. Smith P. Nadesalingam A. et al . (2021). Analysis of serological biomarkers of SARS-CoV-2 infection in convalescent samples from severe, moderate and mild COVID-19 cases. Front. Immunol.12, 748291. 10.3389/fimmu.2021.748291

  • 8

    Ciotti M. Ciccozzi M. Pieri M. Bernardini S. (2021). The COVID-19 pandemic: Viral variants and vaccine efficacy. Crit. Rev. Clin. Lab. Sci. 1–10. 10.1080/10408363.2021.1979462

  • 9

    Cooper G. Mao Q. Crawt L. Wang Y. Dougall T. Rigsby P. et al . (2018). Establishment of the 1st WHO international standard for anti-EV71 serum (Human). Biologicals.53, 3950. 10.1016/j.biologicals.2018.03.001

  • 10

    Dimech W. Arachchi N. Cai J. Sahin T. Wilson K. (2013). Investigation into low-level anti-rubella virus IgG results reported by commercial immunoassays. Clin. Vaccine Immunol. : CVI. 20, 255261. 10.1128/CVI.00603-12

  • 11

    Finney D. J. Schild H. O. (1966). Parallel line assay with successive adjustment of doses. Br. J. Pharmacol.28, 8492. 10.1111/j.1476-5381.1966.tb01876.x

  • 12

    Frederick National Laboratory for Cancer Research. (2021). The Human SARS-CoV-2 Serology Standard. Available online at: https://frederick.cancer.gov/initiatives/seronet/serology-standard

  • 13

    GenScript. (2021). CPass SARS-CoV-2 Neutralization Antibody Detection Kit Instructions for Use. Available online at: https://www.fda.gov/media/143583/download

  • 14

    Giavarina D. Carta M. (2021). Improvements and limits of anti SARS-CoV-2 antibodies assays by WHO (NIBSC 20/136) standardization. Diagnosis (Berlin, Germany). 10.1515/dx-2021-0126

  • 15

    Holder P. K. Maslanka S. E. Pais L. B. Dykes J. Plikaytis B. D. Carlone G. M. (1995). Assignment of Neisseria meningitidis serogroup A and C class-specific anticapsular antibody concentrations to the new standard reference serum CDC1992. Clin. Diagn. Lab. 2, 132137. 10.1128/cdli.2.2.132-137.1995

  • 16

    Infantino M. Pieri M. Nuccetelli M. Grossi V. Lari B. Tomassetti F. et al . (2021). The WHO international standard for COVID-19 serological tests: towards harmonization of anti-spike assays. Int. Immunopharmacol. 100, 108095. 10.1016/j.intimp.2021.108095

  • 17

    Kempster S. L. Almond N. Dimech W. Grangeot-Keros L. Huzly D. Icenogle J. et al . (2020). WHO international standard for anti-rubella: Learning from its application. Lancet. Infect. Dis.20, e17e19. 10.1016/S1473-3099(19)30274-9

  • 18

    Knezevic I. Mattiuzzo G. Page M. Minor P. Griffiths E. Nuebling M. et al . (2022). WHO international standard for evaluation of the antibody response to COVID-19 vaccines: call for urgent action by the scientific community. Lancet. Microbe. 3:e235-e240. 10.1016/S2666-5247(21)00266-4

  • 19

    Kristiansen P. A. Page M. Bernasconi V. Mattiuzzo G. Dull P. Makar K. et al . (2021). WHO International Standard for anti-SARS-CoV-2 immunoglobulin. Lancet (London, England). 397, 13471348. 10.1016/S0140-6736(21)00527-4

  • 20

    Landes R. D. Lensing S. Y. Hauer-Jensen M. (2019). Interval estimators of relative potency in toxicology and radiation countermeasure studies: Comparing methods and experimental designs. J. Biopharmaceut. Statist.29, 348358. PubMed. 10.1080/10543406.2018.1535500

  • 21

    Mattiuzzo G. Bentley E. Hassall M. Routley M. Richardson S. Bernasconi V. et al . (2020). Establishment of the WHO international standard and reference panel for anti-SARS-CoV-2 antibody. Geneva: World Health Organization

  • 22

    Mattiuzzo G. Knezevic I. Hassall M. Ashall J. Myhill S. Faulkner V. et al . (2019). Harmonization of Zika neutralization assays by using the WHO International Standard for anti-Zika virus antibody. NPJ Vaccines. 4, 42. 10.1038/s41541-019-0135-3

  • 23

    McDonald J. U. Rigsby P. Dougall T. Engelhardt O. G. (2018). Establishment of the first WHO international standard for antiserum to respiratory syncytial virus: report of an international collaborative study. Vaccine. 36, 76417649. 10.1016/j.vaccine.2018.10.087

  • 24

    National Institute for Biological Standards Control. (2020). WHO International Standard: First WHO International Standard for anti-SARS-CoV-2 immunoglobulin (human) NIBSC code: 20/136. Instructions for use. Available onlilne at: https://www.nibsc.org/documents/ifu/20-136.pdf

  • 25

    Norman M. Gilboa T. Ogata A. F. Maley A. M. Cohen L. Busch E. L. et al . (2020). Ultrasensitive high-resolution profiling of early seroconversion in patients with COVID-19. Nat. Biomed. Eng.4, 11801187. 10.1038/s41551-020-00611-x

  • 26

    Oneworld Accuracy. (2021). COVS434 |SARS-CoV-2 Serology. Available onlilne at: https://results.oneworldaccuracy.com/results/catalog/browse?cm=ACCUTESTI&y=2022&l=en&f=COVS

  • 27

    Perkmann T. Perkmann-Nagele N. Koller T. Mucher P. Radakovics A. Marculescu R. et al . (2021). Anti-spike protein assays to determine SARS-CoV-2 antibody levels: a head-to-head comparison of five quantitative assays. Microbiol. Spect.9, e0024721. 10.1128/Spectrum.00247-21

  • 28

    Petrone L. Petruccioli E. Vanini V. Cuzzi G. Najafi Fard S. Alonzi T. et al . (2021). A whole blood test to measure SARS-CoV-2-specific response in COVID-19 patients. Clin. Microbiol. Infect.27, 286.e7–286.e13. 10.1016/j.cmi.2020.09.051

  • 29

    R Core Team. (2021). R: A Language and Environment for Statistical Computing. Vienna, Austria: R Foundation for Statistical Computing. Available onlilne at: https://www.R-project.org/

  • 30

    Reed G. F. Lynn F. Meade B. D. (2002). Use of coefficient of variation in assessing variability of quantitative assays. Clin. Vaccine Immunol. 9, 12351239. 10.1128/CDLI.9.6.1235-1239.2002

  • 31

    Schultz J. S. McCarthy M. K. Rester C. Sabourin K. R. Annen K. DomBourian M. et al . (2021). Development and validation of a multiplex microsphere immunoassay using dried blood spots for SARS-CoV-2 seroprevalence: application in first responders in Colorado, USA. J. Clini. Microbiol. 59. 10.1128/JCM.00290-21

  • 32

    Solastie A. Virta C. Haveri A. Ekström N. Kantele A. Miettinen S. et al . (2021). A highly sensitive and specific SARS-CoV-2 spike- and nucleoprotein-based fluorescent multiplex immunoassay (FMIA) to measure IgG, IgA, and IgM class antibodies. Microbiol. Spect.9, e0113121. 10.1128/Spectrum.01131-21

  • 33

    Tan C. W. Chia W. N. Qin X. Liu P. Chen M. I.-C. Tiu C. et al . (2020). A SARS-CoV-2 surrogate virus neutralization test based on antibody-mediated blockage of ACE2–spike protein–protein interaction. Nat. Biotechnol.38, 10731078. 10.1038/s41587-020-0631-z

  • 34

    Taylor S. C. Hurst B. Charlton C. L. Bailey A. Kanji J. N. McCarthy M. K. et al . (2021). A new SARS-CoV-2 dual-purpose serology test: highly accurate infection tracing and neutralizing antibody response detection. J. Clini. Microbiol. 59. 10.1128/JCM.02438-20

  • 35

    Thermo Fisher Scientific. (2021). MASTM SARS-CoV-2 IgG Positive Control. Catalog No.10-028-305. Available online at: https://www.fishersci.com/shop/products/sarscov2-igg-pos-ctl-5x3ml-pk/10028305#

  • 36

    Timiryasova T. M. Hodge S. A. Zheng L. Singer A. Vincent D. Rahman M. et al . (2020). Preparation and qualification of internal rabies reference standards for use in the rabies rapid fluorescent focus inhibition test. Sci. Rep.10, 9893. 10.1038/s41598-020-66754-8

  • 37

    Windsor W. Knight V. Merkel P. Lamb M. Tucker H. Carson K. et al . (2021). COVID-19 serology control panel using the dried tube specimen method. Am. Soc. Tropical Med. Hygiene. 10.1101/2021.07.07.21260101

  • 38

    Wood J. M. Major D. Heath A. Newman R. W. Höschler K. Stephenson I. et al . (2012). Reproducibility of serology assays for pandemic influenza H1N1: Collaborative study to evaluate a candidate WHO International Standard. Vaccine. 30, 210217. 10.1016/j.vaccine.2011.11.019

  • 39

    Zeichhardt H. Kammel M. (2021). INSTAND Summary of sample properties and target values EQAS Group No. 416 Virus Immunology – SARS-CoV-2 (Ab). Availale online at: https://www.instand-ev.de/instand-ringversuche/ringversuchsprogramm/test/virusimmunologie-sars-cov-2-ak/

Summary

Keywords

SARS-CoV-2, serology, International Standards, concordance, immunology, harmonization, parallel-line assay

Citation

Windsor WJ, Roell Y, Tucker H, Cheng C-A, Suliman S, Peek LJ, Pestano GA, Lee WT, Zeichhardt H, Lamb MM, Kammel M, Wang H, Kedl R, Rester C, Morrison TE, Davenport BJ, Carson K, Yates J, Howard K, Kulas K, Walt DR, Dafni A, Taylor D and Chu M (2022) Harmonization of Multiple SARS-CoV-2 Reference Materials Using the WHO IS (NIBSC 20/136): Results and Implications. Front. Microbiol. 13:893801. doi: 10.3389/fmicb.2022.893801

Received

10 March 2022

Accepted

31 March 2022

Published

30 May 2022

Volume

13 - 2022

Edited by

Ziyong Sun, Huazhong University of Science and Technology, China

Reviewed by

Niloufar Kavian, Université Paris Descartes, France; U. G. Liebert, Leipzig University, Germany; Hassan Alkharaan, Prince Sattam Bin Abdulaziz University, Saudi Arabia; Gilberto Betancor, King's College London, United Kingdom

Updates

Copyright

*Correspondence: William Jonathan Windsor

This article was submitted to Virology, a section of the journal Frontiers in Microbiology

Disclaimer

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Outline

Figures

Cite article

Copy to clipboard


Export citation file


Share article

Article metrics