Purpose: The conventional implementation of proton spatially fractionated radiotherapy (SFRT) uses physical collimators with millimeter apertures to generate minibeams, creating alternating regions of high-dose peaks and low-dose valleys. Current evaluation of SFRT effectiveness predominantly relies on physical quantities, particularly the peak-to-valley dose ratio (PVDR). While high PVDR and low valley doses have been correlated with improved normal tissue sparing, this physical metric-based approach provides an incomplete picture of the treatment’s biological impact. In this work, we aim to quantify biological dose for proton minibeams created using physical collimators and critically evaluate the adequacy of physical (PVDRPHYS) as compared to biologically weighted PVDR (PVDRBIOL).
Methods: Monte Carlo simulations using TOPAS were performed to model proton minibeam arrangements with 70 and 150 MeV monoenergetic beams, as well as a Spread-Out Bragg Peak (SOBP) configuration to represent clinically relevant dose delivery. We investigated the impact of multiple collimator configurations on PVDR: collimator thickness (6.35 cm), hole diameter (1–3 mm), center-to-center (c-t-c) distances (2, 3, 4 mm for 1 mm hole, 4, 6, 8 mm for 2 mm hole) and air gaps (5, 10, and 15 cm) between the collimator and water phantom. For each configuration, 3D dose and dose-averaged LET (LETd) distributions were scored in water phantom, which were subsequently used for RBE calculation using the McNamara model with α/β ratios of 3 or 10 Gy.
Results: Our comprehensive analysis revealed significant differences between physical and biological PVDR metrics across various configurations. For both 70 and 150 MeV beams, PVDRBIOL was consistently lower than PVDRPHYS by 1-25% and 1-21%, respectively, with the most pronounced differences observed at shallow depths, smaller air gaps with larger hole diameter and c-t-c distances. Similar reductions (3.9%-26.5%) in PVDRBIOL were observed for the SOBP configuration, with the specific pattern depending on the energy composition and weighting of constituent layers.
Conclusions: The significant variations between PVDRPHYS and PVDRBIOL across different beam energies, depths, and collimator configurations demonstrate that conventional PVDR calculations based solely on physical dose may not fully represent the biological impact of proton SFRT. These findings highlight the importance of incorporating radiobiological considerations when evaluating and optimizing proton minibeam therapy, potentially leading to more biologically informed treatment planning approaches.
1 Introduction
Normal tissue toxicity remains the primary dose-limiting factor in radiation therapy, preventing the delivery of curative doses to challenging tumors such as large, radioresistant, or critically located malignancies. This fundamental constraint has driven decades of innovation in radiation delivery techniques, with spatially fractionated radiotherapy (SFRT) emerging as a promising approach to widen the therapeutic window by exploiting the superior tolerance of healthy tissues to spatially modulated versus uniform irradiation (1–5).
SFRT originated with Alban Köhler’s kilovoltage GRID therapy in 1909, which used perforated lead screens to minimize adverse radiation effects to the skin when irradiating deep-seated tumors (6, 7). More than two decades later, unaware of Köhler’s work, Dr. Liberson of U.S. Marine Hospitals used GRID therapy to treat small animals and inoperable tumors, similarly observing increased skin tolerance (8). These early investigations using perforations of different shapes in lead sheets primarily addressed the skin sparing challenges of orthovoltage treatments (9). With the advent of megavoltage radiotherapy and linear accelerators in the 1970s, skin sparing was no longer the primary concern, but new challenges emerged regarding normal tissue tolerance of organs like lung, brain, and intestines (10). This revived interest in GRID therapy, leading researchers to develop concepts compatible with megavoltage radiotherapy using specially designed Cerrobend grid matrices, successfully treating 71 patients with advanced bulky tumors using 6 MV and 25 MV photons (11). SFRT using protons represents a more recent development, beginning with micrometer-scale proton microbeam therapy (12, 13), followed by studies exploring 0.3-0.7 mm beam sizes termed proton minibeam radiation therapy (pMBRT). Prezado and Fois provided the initial proof-of-concept through Monte Carlo (MC) simulations (14), followed by experimental dosimetry studies using cyclotron-accelerated protons (15, 16) and demonstration of normal brain tissue sparing in animal models (17). Current pMBRT implementations utilize pencil beam scanning (PBS) with additional collimation to achieve millimeter-scale beamlets, as pencil beams alone produce approximately 1 cm beam sizes that require further collimation (typically using brass or tungsten plates) to maximize the dose-volume effect (18–21), employing different collimator designs [such as 3D-printed plastic plates (22)]. Beyond their material composition, collimators also differ in terms of beamlet patterning, ranging from 1D slits (19) to 2D hexagonal arrays (18). Other methods include magnetically focusing the beam to create minibeams (23–25).
Clinical effectiveness in SFRT is predominantly quantified using the peak-to-valley dose ratio (PVDR), defined as the ratio of high-dose peaks to low-dose valleys in the spatial dose distribution. Preclinical studies consistently demonstrate that higher PVDR values correlate with markedly improved normal tissue sparing (26, 27). For example, rat whole-brain pMBRT studies using 0.4 mm-wide minibeams with surface PVDR ≈ 15 achieved tolerances of single-fraction peak doses up to 58 Gy (~25 Gy mean dose) without detectable neurotoxicity, whereas conventional uniform 25 Gy proton beams caused severe brain injury (27). These remarkable tissue-sparing effects have generated significant clinical interest (28–30).
However, current SFRT evaluation relies on physical dose distributions and fails to account for biologically relevant factors that significantly influence therapeutic outcomes. For clinical practices, relative biological effectiveness (RBE) is assumed to be constant throughout the volume, with the value being equal to 1.1 relative to high-energy photons (31). However, it is well known that the RBE increases with linear energy transfer (LET); and the high LET is very characteristic of protons that have been degraded or scattered by collimators. As protons slow, their LET (and thus ionization density) rises, driving up RBE in the low−energy tail (32). Recent experimental studies (33) have documented LET increases of 3.5–7 keV/μm in collimated proton fields compared to open beams, with corresponding increases in RBE that could substantially alter the biological dose landscape. Ueno et al. (34) demonstrated that protons scattered off brass collimator edges exhibit higher dose averaged LET (LETd), while Nabha et al. measured significant LET gains near the Bragg peak when using aperture collimation (33). Smith and Hyer (35) confirmed that energy−specific collimation in PBS shifts the upper−percentile LET by a few keV/μm in trimmed beamlets, creating localized high−LET hotspots in the penumbra. These LET enhancements drive RBE increases that are not captured in conventional evaluation of pMBRT, potentially leading to systematic overestimation of normal tissue sparing when relying solely on physical dose metrics.
The clinical implications of this knowledge gap are profound. If physical PVDR (PVDRPHYS) overestimates the true biological effectiveness of spatial dose modulation, current treatment evaluation may systematically underestimate normal tissue complication probabilities, particularly for late-responding tissues with low α/β ratios that are most sensitive to RBE variations. To better interpret potential tissue-sparing effects, spatial dose modulation must be evaluated through a biological lens that accounts for variable RBE. No prior work has directly compared PVDRPHYS to a biologically weighted analogue (PVDRBIOL) that incorporates LET−driven RBE variations documented in scattered or degraded proton fields. Incorporating RBE−weighted dose calculations, by validated biophysical models into proton minibeam evaluation will reveal the accuracy of PVDRPHYS as a surrogate for therapeutic ratio and guide the development of biologically optimized treatment plans (36–38).
In this comprehensive MC simulation study, we aim to bridge the gap between physical and biological endpoints in pMBRT by (1): quantifying biological dose distributions using validated RBE models for millimeter-scale PBS minibeams shaped by physical collimators (2); directly comparing conventional PVDRPHYS with biologically weighted PVDRBIOL across a comprehensive range of beam energies and geometric configurations.
2 Materials and methods
MC simulations were performed using TOPAS (Tool for Particle Simulation) (39) version 3.7 with the default modular physics list without modification. The simulation geometry, Figure 1A, consisted of a 10×10×20 cm³ water phantom positioned below a brass collimator with variable air gaps. To ensure adequate statistical accuracy, 106 particle histories per spot were simulated. The TOPAS simulation methodology employed in this study has been extensively validated through experimental benchmarking in our previous work. Charyyev et al. demonstrated good agreement between TOPAS simulations and film dosimetry measurements for similar minibeam collimator configurations, including the same collimator thickness, hole diameter, c-t-c distances, and air gaps used in the current study (40). Cross-validation with EBT3 film measurements showed agreement within 4.4% at the entrance region and 1% in the Bragg peak region across the energy range employed in this work (18). These validation studies provide robust experimental anchoring for the MC methodology used in the present investigation.
Figure 1. (A) Schematic diagram of the MC simulation geometry. Monoenergetic (70 MeV or 150 MeV) or SOBP proton beams with 5×5 cm² field size, created using an 11×11 spot array with 5 mm interspot spacing, pass through a 6.35 cm thick brass collimator positioned at variable air gaps (5, 10, or 15 cm) above a water phantom (10×10×20 cm³). Dose and LETd scoring were performed within a central 6×6×20 cm³ region. (B) Top view of a representative hexagonal collimator pattern showing 2 mm diameter holes with 6 mm c-t-c spacing distributed across a 13.2×12.47 cm2 area. The purple dashed square delineates the 5×5 cm² proton field boundary, while the red dashed line indicates the lateral profile extraction path for PVDR analysis. (C) Representative lateral dose profile for 70 MeV proton beam at 1 cm depth demonstrating the characteristic peak-and-valley spatial dose modulation achieved through collimation. Peak doses correspond to beam paths through collimator apertures, while valley doses result from scattered radiation in inter-beam regions. This spatial fractionation pattern forms the basis for PVDR quantification across all geometric configurations studied.
Monoenergetic proton beams of 70 MeV (continuous slowing down approximation (CSDA) range ~4.1 cm in water) and 150 MeV (CSDA range ~15.8 cm in water) were modeled with a 5×5 cm² field size created using an 11×11 spot array with 5 mm interspot spacing to generate a uniform incident field. The proton beam characteristics were modeled using realistic phase space parameters derived from clinical commissioning experimental measurements (41, 42). For 70 MeV protons, the beam was characterized by: spatial beam spread σx = σy = 5.74 mm, angular divergence σx’ = σy’ = 6.91 mrad, correlation coefficients ρx = ρy = 0.798, and energy spread = 1.36%. For 150 MeV protons: spatial beam spread σx = σy = 4.11 mm, angular divergence σx’ = σy’ = 3.47 mrad, correlation coefficients ρx = ρy = 0.625, and energy spread = 0.84%. These parameters were implemented using TOPAS emittance-based source modeling with BiGaussian distribution to accurately represent the correlated spatial-angular beam characteristics at the collimator entrance. Additionally, a Spread-Out Bragg Peak (SOBP) configuration was simulated to represent clinical dose delivery. The SOBP field consisted of five energy layers ranging from 100 to 110 MeV in increments of 2–3 MeV. Each energy layer was weighted according to the values shown in Table 1 to produce a uniform dose plateau over depths of 7–9 cm in water. The primary analysis was based on a 2 Gy prescription dose, defined at the Bragg peak for pristine beams (or at the plateau for the SOBP configuration), Additional prescription dose levels of 5, 10 and 15 Gy were also investigated.
Brass collimator (density = 8.5 g/cm³, composition: 61.5% Cu, 35.5% Zn, and 3% Pb) with a thickness of 6.35 cm featured circular apertures arranged in hexagonal patterns with hole diameters of 1-, 2-, and 3-mm. Figure 1B illustrates a representative collimator design showing the hexagonal hole pattern for 2 mm diameter holes with 6 mm (0.6 cm) center-to-center (c-t-c) distance distributed across a 13.2×12.47 cm collimator area. The hexagonal arrangement was chosen as a motivation from Sammer et al. (43), who performed cell survival simulations for the various proton minibeam arrangements and showed that for a fixed tumor dose, the hexagonal beam arrangement provides the maximum benefits in normal tissue sparing. This geometric configuration spatially fractionated the initially uniform proton field into the characteristic minibeam pattern.
A comprehensive parameter study encompassed multiple configurations systematically varying the analyzed parameters for 70 MeV and 150 MeV: a) c-t-c distances scaled according to hole diameter: 1 mm holes: 2, 3, 4 mm c-t-c (corresponding to 2×, 3×, 4× hole diameter), 2 mm holes: 4, 6, 8 mm c-t-c, 3 mm holes: 6, 9, 12 mm c-t-c. b) air gap distances: 5, 10, and 15 cm between collimator exit surface and water phantom entrance, measured along the beam central axis (CAX).
The selection of proton energies, hole diameter, and c-t-c distances was based on clinical translation considerations and established optimization principles (18, 40, 43). The 70–150 MeV energy range represents clinically relevant energies for SFRT applications, with 70 and 150 MeV corresponding to approximately 4 and 16 cm range, respectively, in tissue, covering typical treatment depths for SFRT. Supra-millimeter hole diameters (1–3 mm) were selected based on practical optimization criteria that balance normal tissue sparing with clinical feasibility. While theoretically smaller apertures would enhance tissue sparing, 1–3 mm beamlets represent an optimal compromise considering several factors: beamlet width should not be smaller than setup uncertainties and beam delivery accuracy, cardiopulmonary motion effects on dose distribution, manufacturing feasibility compared to micrometric apertures, achievable dose rates and treatment times, beam economy with reduced collimator activation, and avoidance of excessively high entrance doses that can exceed Bragg peak values due to reduced scatter and loss of charged particle equilibrium. The c-t-c distances were determined to achieve uniform target dose coverage, with hexagonal arrangements requiring a σdepth/c-t-c distance ratio of approximately 0.40-0.44 for dose uniformity within 95-105% based on inverse planning studies (40, 43).
3D dose and LETd distributions were scored within a central 6×6×20 cm³ scoring region of the water phantom to focus analysis on the primary beam interaction volume while reducing computational overhead (Figure 1A). This scoring volume was positioned to capture the complete minibeam pattern. Dose and LETd distributions were scored simultaneously throughout the scoring volume using a uniform voxel grid of 0.5×0.5×1.0 mm³ and 1.0×1.0×10.0 mm³, respectively. LETd used larger scoring voxels to prevent numerical singularities in the LET calculation arising from low-dose regions. LETd calculations included contributions from primary protons and secondary protons only, including the energy deposited by associated secondary electrons but excluding heavier fragments (alpha particles, nuclear fragments), which contribute<0.1% of total dose, to maintain consistency with established clinical RBE models. LETd distributions were interpolated from the coarser scoring grid to match the finer dose scoring grid using trilinear interpolation.
The collimator design successfully generated the intended spatially fractionated dose distribution, as demonstrated in Figure 1C, which shows a representative lateral dose profile for 70 MeV minibeams at 1 cm depth displaying the characteristic peak and valley pattern. Peak doses correspond to regions aligned with collimator apertures, while valley doses represent the inter-beam regions where only scattered radiation contributes to the dose.
RBE was calculated using the McNamara (Equations 1) (44) phenomenological model:
where, is physical dose, represents intrinsic radiosensitivity, is dose averaged LET, This model was applied on a voxel-by-voxel basis using the local physical dose and corresponding LETd values. Two different α/β ratios of 3 Gy and 10 Gy were employed to represent different tissue response characteristics, with the biological dose calculated as for each voxel. The McNamara model was selected due to its comprehensive validation against experimental data and its widespread use in proton therapy research (31, 45–48). However, we have done calculations for selected subgroup of simulation scenarios for two other RBE models: Carabe-Fernandez (Equations 2–4) (49):
where, is physical dose, represents intrinsic radiosensitivity, is dose averaged LET, and (derived using V79 cell data) are defined as follows:
and Wedenberg (Equation 5) (50):
where, is physical dose, represents intrinsic radiosensitivity, is dose averaged LET.
PVDR analysis was performed using lateral dose profiles, similar to that in Figure 1C, extracted perpendicular to the minibeam arrangement through the field center. Lateral profiles were sampled at dose resolution to ensure accurate determination of peak and valley positions, with analysis performed at multiple depths from the phantom surface to near-maximum beam range.
Physical PVDR was calculated as the ratio of average peak doses to average valley doses from the physical dose profiles: . Similarly, biological PVDR was determined using RBE-weighted dose profiles: . To minimize edge effects, the analysis was restricted to the central region of the field, excluding peaks and valleys within 1 cm of the field boundary. The PVDR calculations were systematically performed across all parameter combinations, including different hole sizes, c-t-c distances, air gaps, and beam energies, enabling comprehensive evaluation of the relationship between physical and biological dose metrics in pBMRT.
Voxel-wise uncertainties in physical dose and LETd were scored in TOPAS and propagated through all calculations using standard error propagation. For PVDRPHYS, uncertainties in peak and valley doses were combined in quadrature using standard ratio error propagation. For PVDRBIOL, the uncertainty calculation required additional steps due to RBE’s dependence on both dose and LETd. Partial derivatives of the McNamara RBE model with respect to dose and LETd were calculated analytically to determine RBE uncertainty at both peak and valley locations. These RBE uncertainties were then combined with physical dose uncertainties to calculate biological dose uncertainties. Finally, PVDRBIOL uncertainty was determined by propagating the biological dose uncertainties through the PVDR calculation.
Simulations were performed on Intel® Xeon(R) Gold system with parallel processing utilizing 96 CPU cores per simulation. Total computational time was approximately 110 hours for the complete parameter study.
3 Results
Figure 2 shows a representative example of the dose and LETd analysis performed for all simulation scenarios, specifically for the configuration with 5 cm air gap, 2 mm hole diameter, and 6 mm c-t-c distance for 70 MeV protons. The top row demonstrates the physical dose (black solid line) and LETd (red dashed line) distributions, including the CAX depth profile and lateral profiles at 1, 2, and 4 cm depths. The bottom row presents the corresponding physical dose (black solid line) and RBE-weighted biological dose (red dashed line) distributions at the same positions. The lateral profiles clearly demonstrate the spatially fractionated nature of the beam delivery with distinct peaks and valleys. The LETd distribution shows elevated values in the valley regions, contributing to enhanced biological effectiveness in these areas. The biological dose distributions consistently show higher values compared to physical dose, particularly in the valley regions where the combination of moderate dose levels and elevated LETd results in significant RBE enhancement.
Figure 2. Representative dose and LETd distribution analysis for 70 MeV proton SFRT with 5 cm air gap, 2 mm hole diameter, and 6 mm c-t-c distance. Top row: Physical dose (black solid) and LETd (red dashed) distributions showing CAX depth profile and lateral profiles at 1, 2, and 4 cm depths. Bottom row: Corresponding physical dose (black solid) and RBE-weighted biological dose (red dashed) distributions. The lateral profiles clearly demonstrate the spatially fractionated beam delivery pattern with distinct peaks and valleys. This analysis was performed for all simulation scenarios in the study.
Figures 3, 4 present systematic parametric analyses of how PVDR values vary with geometric beam parameters for 70 MeV and 150 MeV protons, respectively. Both PVDRPHYS (solid lines) and PVDRBIOL (dashed lines) using the McNamara model are shown across three parameter variations. The relative uncertainties (σ/value) were less than 0.02% for PVDRPHYS at both beam energies. For PVDRBIOL, relative uncertainties ranged from 1.1% to 2.9% at 70 MeV and from 0.7% to 3.3% at 150 MeV. Due to the small magnitude of the uncertainties relative to the marker size, uncertainty bars are not displayed in figures.
Figure 3. Parametric analysis of PVDR for 70 MeV proton beams showing both PVDRPHYS (solid lines) and PVDRBIOL (dashed lines) values. Rows represent different depths: 1 cm, 2 cm, and 4 cm. Parameter variations: hole diameter (1–3 mm with c-t-c distance = 3× hole diameter, air gap = 5 cm), c-t-c distance (4–8 mm with hole diameter = 2 mm, air gap = 5 cm), and air gap (5–15 cm with hole diameter = 2 mm, c-t-c distance = 6 mm).
Figure 4. Parametric analysis of PVDR for 150 MeV proton beams with identical parameter variations as Figure 3. Rows represent depths: 5 cm, 10 cm, and 15 cm.
When varying hole diameter from 1 mm to 3 mm (with air gap fixed at 5 cm and c-t-c distance maintained at 3× hole diameter), both energies show substantial increases in PVDR. For 70 MeV at 1 cm depth, PVDR increases from approximately 5 to 80 for physical dose, while PVDRBIOL shows a similar trend but with consistently lower absolute values. The 150 MeV configuration demonstrates more modest PVDR increases but follows the same trend, with values ranging from 2 to 25 at 5 cm depth. Variation of c-t-c distance (with 5 cm air gap and 2 mm hole diameter fixed) shows the strongest influence on PVDR among all tested parameters. For 70 MeV beams, increasing c-t-c distance from 4 mm to 8 mm results in dramatic PVDR increases from approximately 5 to 65 at 1 cm depth for physical dose. The 150 MeV configuration shows similar trends with more moderate absolute values.
The air gap parameter exhibits an inverse relationship with PVDR across both energies. For 70 MeV beams (with 2 mm hole diameter and 6 mm c-t-c distance), increasing air gap from 5 cm to 15 cm reduces PVDR from approximately 25 to 16 at 1 cm depth. This inverse relationship suggests that minimal air gaps are preferred for maximizing PVDR in proton SFRT configurations. The analysis reveals distinct depth-dependent behavior for both energies. For 70 MeV, depths of 1, 2, and 4 cm show progressively decreasing PVDR values across all parameter combinations. Similarly, for 150 MeV at depths of 5, 10, and 15 cm, PVDR values decrease with increasing depth, though the 150 MeV configuration maintains more consistent PVDR values across the analyzed depth range.
Figure 5 provides a quantitative analysis of the differences between PVDRPHYS and PVDRBIOL across all parameter combinations and energies. The results demonstrate that PVDRBIOL is consistently lower than PVDRPHYS, with the magnitude of reduction depending on the specific parameter combination and energy. For maximum PVDR reduction analysis, the 70 MeV configuration shows reductions of up to 25% for optimal parameter combinations, while 150 MeV shows more modest reductions of 10-20%. The mean PVDR reduction analysis reveals similar trends with slightly lower absolute values. The c-t-c distance parameter shows the most dramatic differences between PVDRPHYS and PVDRBIOL, while air gap variations show the smallest difference. This occurs because increased c-t-c distance forces protons to traverse significantly more material to reach valley regions, reducing their energy and increasing their LETd (and thus RBE), creating larger disparities between PVDRPHYS and PVDRBIOL. Conversely, air-gap variations primarily affect beam divergence and penumbra without substantially altering proton energies, resulting in smaller PVDR differences.
Figure 5. Quantitative comparison of PVDRPHYS versus PVDRBIOL showing maximum and mean PVDR reduction percentages across all parameter combinations. Top row: Maximum PVDR reduction for 70 MeV (blue) and 150 MeV (orange) as functions of air gap, hole diameter, and c-t-c distance. Bottom row: Corresponding mean PVDR reductions. The analysis demonstrates that PVDRBIOL is consistently lower than PVDRPHYS, with the magnitude of reduction varying significantly based on beam parameters and energy. Multiple hole diameter configurations are shown for c-t-c distance analysis.
Figure 6A illustrates the depth dose profiles of the individual energy layers (100, 102, 105, 107, and 110 MeV) and their weighted superposition forming the SOBP. Figure 6B demonstrates the PVDR reduction as a function of depth for different prescription dose levels (2, 5, 10, and 15 Gy) using the McNamara model with α/β = 3 Gy. The prescription level was determined by the dose delivered to the SOBP plateau, with the 2 Gy example shown in Figure 6A. PVDR reduction ranging from 3.9% to 26.5% (for the 2 Gy prescription level) is evident with the SOBP scenario as well, with the specific pattern depending on the energy composition and weighting of constituent layers.
Figure 6. (A) CAX depth dose profiles showing individual energy layer contributions (dashed lines) and the resulting SOBP (solid black line) achieving a uniform dose plateau over 7–9 cm depth. (B) PVDR reduction as a function of depth for different dose levels using the McNamara model with α/β = 3 Gy. All calculations performed with 2 mm hole diameter, 6 mm c-t-c distance, and 5 cm air gap.
Figure 7A demonstrates the comparative analysis of dose distributions calculated using three different RBE models: McNamara (44), Carabe-Fernandez (49), and Wedenberg (50). The CAX depth dose comparison reveals that all three RBE models produce enhanced dose delivery compared to physical dose calculations, with the enhancement becoming more pronounced at greater depths where LET values increase substantially.
Figure 7. (A) CAX depth dose profiles showing physical dose (black solid line) and RBE-weighted doses calculated using McNamara (red dashed), Carabe-Fernandez (green dashed), and Wedenberg (blue dashed) models. (B) PVDR reduction comparison between different tissue radiosensitivities (α/β ratios) for all three RBE models. Solid lines represent α/β = 10 Gy (more radioresistant tissues) while dashed lines represent α/β = 3 Gy (more radiosensitive tissues). McNamara model (red circles), Carabe-Fernandez model (green squares), and Wedenberg model (blue triangles) demonstrate consistent trends with significantly higher PVDR reduction for α/β = 3 Gy compared to α/β = 10 Gy.
Figure 7B illustrates the comparative PVDR reduction analysis for both α/β = 3 Gy and α/β = 10 Gy across all three RBE models as a function of depth. The results demonstrate the substantial impact of tissue radiosensitivity on biological dose calculations. For α/β = 3 Gy, representing more radiosensitive tissues, PVDR reductions range from approximately 20-24% at 1 cm depth to 4-5% at 4 cm depth across all models. In contrast, for α/β = 10 Gy, representing more radioresistant tissues typically associated with early-responding tissues and tumors, the reductions are substantially lower, ranging from 5 to 7% at 1 cm depth to 2-3% at 4 cm depth.
4 Discussion
Among the results presented in the Results section, several key observations warrant detailed discussion. The clinical significance of the observed 20-25% reduction in PVDR must be evaluated within the context of current understanding of SFRT effectiveness, a complex undertaking in itself (51). Typical pMBRT PVDR values achieved in clinical practice range from 9-12 (52, 53). However, the field currently lacks definitive evidence establishing optimal PVDR thresholds for different anatomical sites and clinical scenarios. The NRG Oncology/AAPM Working Group consensus (30) acknowledges that PVDR parameters remain “unclear” and require further clinical validation, though general agreement exists that spatial dose modulation must be sufficient to preserve normal tissue sparing mechanisms. Despite this uncertainty in optimal PVDR values, a 20-25% reduction in PVDR represents a clinically significant finding that cannot be dismissed. Preclinical evidence demonstrates that PVDR operates within narrow therapeutic windows, with mathematical models showing therapeutic ratios varying dramatically (0.88 to 13.22) depending on achieved PVDR values (54). In the absence of established PVDR thresholds, treatment planning systems that rely solely on physical dose calculations may systematically overestimate the therapeutic benefit of proton SFRT, particularly for treatments involving late-responding critical structures where the biological dose reduction is most pronounced.
Our findings are strongly supported by the pioneering work of Meyer et al. (55), who first implemented RBE-weighted dose calculations in pMBRT using TOPAS Monte Carlo simulations and the MC damage simulation model for double strand break endpoints. Despite methodological differences—Meyer et al. used 300 μm steel slits with 1 mm center-to-center spacing while our study employed 1–3 mm circular brass apertures with 2-4× hole diameter spacing—both studies demonstrate remarkably consistent quantitative results. Meyer et al. reported that relative mean RBE-weighted entrance dose was approximately 25% lower than physical dose for proton minibeams, which closely aligns with our finding that PVDRBIOL decreases by 20-25% compared to PVDRPHYS for late-responding tissues (α/β = 3 Gy). Both studies also confirm similar depth-dependent RBE variations, with entrance RBE values around 1.02-1.04, Bragg peak RBE of 1.3-1.6, and distal edge values reaching 1.4-2.0. This convergent evidence from independent methodological approaches provides robust validation that biological dose effects significantly impact proton SFRT dosimetry. Our work extends Meyer’s foundation by systematically examining how RBE affects the fundamental PVDR parameter that defines SFRT’s therapeutic advantage, while providing multi-model validation through multiple phenomenological RBE approaches and comprehensive geometric parameter analysis across two proton energies and SOBP configurations.
An important observation from our SOBP analysis is that the resulting depth-dose distribution does not resemble the classical flat plateau achieved with conventional proton PBS. This reflects a unique characteristic of proton minibeam depth-dose curves: the peak dose decreases rapidly with depth such that the maximum dose occurs at the entrance rather than in the target region, unlike conventional PBS proton therapy. As shown in Figure 6A, the SOBP dose reaches only approximately 40% of the entrance peak dose for 2 mm minibeams, an effect that becomes more prominent as minibeam diameter decreases. While this characteristic may initially appear undesirable for normal tissue sparing, it is important to recognize that these peak dose regions exist at very small scales of ~2–3 mm and occupy only a small fraction (~10%) of the normal tissue volume. The bulk of the normal tissue (~90%) receives substantially lower valley doses.
Our choice of the McNamara model for primary RBE calculations was based on its widespread use and established popularity in the proton therapy community. However, clinicians and researchers may choose alternative RBE models depending on their goals and preferences. From our analysis, at the shallower depths McNamara model produced the highest dose enhancement, followed by the Wedenberg model, while the Carabe-Fernandez model provided the most conservative estimates. This ordering reversed at the Bragg peak depth, where the Carabe-Fernandez model produced the highest dose enhancement, followed by Wedenberg, while the McNamara model provided the most conservative estimates.
It should be noted that any chosen RBE model, including the ones we selected for this study, has inherent uncertainties and that patient-specific radiosensitivities may be different. Also, the RBE models used in this study are based on cell survival models but don’t consider other biological endpoints like vascular effects, immune responses, or repair mechanisms that might be particularly relevant for SFRT. Furthermore, current RBE models face additional limitations in hypo-fractionated SFRT conditions, as they were developed using conventional fractionation data. Even though we demonstrated PVDR reduction for different prescription levels, Figure 6B, the underlying linear-quadratic model systematically fails at high doses per fraction (>8–10 Gy) characteristic of SFRT peak regions, where vascular damage and immune activation mechanisms become dominant contributors beyond DNA damage-based calculations (56, 57). Therefore, our PVDRBIOL calculations represent first-order approximations that should be interpreted with appropriate caution in the context of these model limitations. These limitations underscore the critical need for experimental validation of RBE models under hypo-fractionated SFRT conditions (31).
The choice of α/β = 3 Gy for our primary analysis was made for several clinical reasons. First, α/β = 3 Gy represents late-responding normal tissues such as spinal cord, brain, and lung tissue, which are typically the dose-limiting structures in radiotherapy planning. For normal tissue sparing evaluation, α/β = 3 Gy provides a more conservative and clinically relevant estimate of biological effects. Second, in SFRT, the primary clinical advantage lies in sparing late-responding normal tissues, and most critical organs at risk have lower α/β ratios rather than higher. Finally, this choice aligns with literature precedent, as many proton RBE studies focusing on normal tissue effects utilize α/β ratios of 2–3 Gy. However, it is important to acknowledge that different clinical scenarios may warrant different α/β ratio selections. The comparison in Figure 7 emphasizes that the choice of α/β ratio significantly influences the magnitude of biological effects in proton SFRT, and treatment planning should consider the specific tissue types being treated and their associated radiobiological parameters.
Although extensive simulations were performed across multiple parameter combinations, we showcased the specific case of 2 mm hole diameter with 6 mm c-t-c distance and 5 cm air gap for several reasons. We believe that millimeter-sized proton beams are preferred over sub-millimeter proton beams because sub-millimeter beams cannot maintain high PVDRs at larger depths in patients due to lateral scattering effects. For this reason, peak-to-valley contrast even for 1 mm holes is not well pronounced, especially for 150 MeV protons as they travel deeper into tissue. Since PVDR values around 10 have been shown to produce minimal damage to normal tissue, a PVDR of 10 at the phantom surface was targeted as the minimum quality threshold for the selected minibeams. This target PVDR value was achieved with c-t-c distances of at least 6 mm for 2 mm hole diameter configurations. From a practical clinical standpoint, larger air gaps are generally preferred as the additional space may be needed for positioning and imaging systems. However, due to scattering effects in air, larger air gaps result in poorer PVDR values. Therefore, the smallest possible and clinically safe air gap should be considered for clinical implementation.
While this study establishes differences between PVDRPHYS and PVDRBIOL, future work is needed to determine clinical adequacy thresholds and develop specific treatment planning recommendations.
5 Conclusion
This comprehensive MC study demonstrates that biological dose calculations alter the dosimetric landscape of pMBRT compared to conventional physical dose planning. Through systematic analysis of geometric configurations across two proton energies and SOBP configuration, we have shown that incorporating RBE effects using established phenomenological models results in notable reductions in PVDRs, with PVDRBIOL decreasing by up to 26% compared to physical calculations for late-responding normal tissues (α/β = 3 Gy). While the clinical field currently lacks consensus on optimal PVDR thresholds, the magnitude of these reductions cannot be dismissed given the narrow therapeutic windows demonstrated in preclinical studies and the documented sensitivity of late-responding tissues to spatial dose modulation patterns. Our findings reveal that current treatment planning approaches relying solely on physical dose calculations may systematically overestimate the therapeutic benefit of proton SFRT, particularly for treatments involving critical structures with low α/β ratios where biological dose enhancement is most pronounced. The observed parametric dependencies, with c-t-c distance showing the strongest influence on PVDR across all models, remain consistent between physical and biological calculations, though the absolute PVDR values are consistently reduced when biological effects are considered. These results underscore the critical importance of incorporating RBE modeling into proton SFRT treatment planning systems and highlight the need for experimental validation of PVDR thresholds to optimize the clinical implementation of this promising therapeutic approach.
Data availability statement
The original contributions presented in the study are included in the article/supplementary material. Further inquiries can be directed to the corresponding author.
Author contributions
KD: Software, Formal Analysis, Writing – review & editing, Writing – original draft, Conceptualization, Investigation, Methodology. AU: Software, Writing – review & editing, Formal Analysis, Writing – original draft, Methodology. RK: Writing – original draft, Software, Formal Analysis, Writing – review & editing, Methodology. LL: Writing – review & editing, Writing – original draft, Methodology, Conceptualization, Supervision. YY: Writing – review & editing, Writing – original draft. MS: Writing – original draft, Writing – review & editing. SC: Data curation, Methodology, Formal Analysis, Investigation, Funding acquisition, Conceptualization, Writing – review & editing, Project administration, Writing – original draft, Supervision.
Funding
The author(s) declared that financial support was received for this work and/or its publication. Dr. Liyong Lin’s efforts are in part funded by the National Institutes of Health (NIH)’s 1R03CA300834.
Conflict of interest
The authors declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
The handling editor RG declared a past co-authorship with the author LL.
Generative AI statement
The author(s) declared that generative AI was not used in the creation of this manuscript.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
1. Meyer J, Eley J, Schmid TE, Combs SE, Dendale R, and Prezado Y. Spatially fractionated proton minibeams. BJR. (2019) 92:20180466. doi: 10.1259/bjr.20180466
2. Prezado Y, Grams M, Jouglar E, Martínez-Rovira I, Ortiz R, Seco J, et al. Spatially fractionated radiation therapy: a critical review on current status of clinical and preclinical studies and knowledge gaps. Phys Med Biol. (2024) 69:10TR02. doi: 10.1088/1361-6560/ad4192
3. Dilmanian FA, Zhong Z, Bacarian T, Benveniste H, Romanelli P, Wang R, et al. Interlaced x-ray microplanar beams: a radiosurgery approach with clinical potential. Proc Natl Acad Sci U.S.A. (2006) 103:9709–14. doi: 1073/pnas.0603567103
4. Lu Q, Yan W, Zhu A, Tubin S, Mourad WF, and Yang J. Combining spatially fractionated radiation therapy (SFRT) and immunotherapy opens new rays of hope for enhancing therapeutic ratio. Clin Transl Radiat Oncol. (2024) 44:100691. doi: 10.1016/j.ctro.2023.100691
5. Zhang H, Wu X, Zhang X, Chang SX, Megooni A, Donnelly ED, et al. Photon GRID radiation therapy: A physics and dosimetry white paper from the radiosurgery society (RSS) GRID/LATTICE, microbeam and FLASH radiotherapy working group. Radiat Res. (2020) 194. Available online at: https://bioone.org/journals/radiation-research/volume-194/issue-6/RADE-20-00047.1/Photon-GRID-Radiation-Therapy–A-Physics-and-Dosimetry-White/10.1667/RADE-20-00047.1.full.
6. Yan W, Khan MK, Wu X, Simone CB, Fan J, Gressen E, et al. Spatially fractionated radiation therapy: History, present and the future. Clin Transl Radiat Oncol. (2020) 20:30–8. doi: 10.1016/j.ctro.2019.10.004
7. Laissue JA, Blattmann H, and Slatkin DN. Alban köhler (1874-1947): erfinder der gittertherapie. Z für Med Phys. (2012) 22:90–9. doi: 10.1016/j.zemedi.2011.07.002
8. Liberson F. The value of a multi-perforated screen in deep X-ray therapy: A preliminary report on a new method of delivering multiple erythema doses without permanent injury to the skin. Radiology. (1933) 20:186–95. doi: 10.1148/20.3.186
9. Marks H. Clinical experience with irradiation through a grid. Radiology. (1952) 58:338–42. doi: 10.1148/58.3.338
10. Schültke E, Balosso J, Breslin T, Cavaletti G, Djonov V, Esteve F, et al. Microbeam radiation therapy - grid therapy and beyond: a clinical perspective. Br J Radiol. (2017) 90:20170073. doi: 10.1259/bjr.20170073
11. Mohiuddin M, Fujita M, Regine WF, Megooni AS, Ibbott GS, and Ahmed MM. High-dose spatially-fractionated radiation (GRID): a new paradigm in the management of advanced cancers. Int J Radiat Oncol Biol Phys. (1999) 45:721–7. doi: 10.1016/S0360-3016(99)00170-4
12. Kłodowska M, Olko P, and Waligórski MPR. Proton microbeam radiotherapy with scanned pencil-beams – Monte Carlo simulations. Physica Med. (2015) 31:621–6. doi: 10.1016/j.ejmp.2015.04.006
13. Zlobinskaya O, Girst S, Greubel C, Hable V, Siebenwirth C, Walsh DWM, et al. Reduced side effects by proton microchannel radiotherapy: study in a human skin model. Radiat Environ Biophys. (2013) 52:123–33. doi: 10.1007/s00411-012-0450-9
14. Prezado Y and Fois GR. Proton-minibeam radiation therapy: a proof of concept. Med Phys. (2013) 40:031712. doi: 10.1118/1.4791648
15. Dilmanian FA, Eley JG, and Krishnan S. Minibeam therapy with protons and light ions: physical feasibility and potential to reduce radiation side effects and to facilitate hypofractionation. Int J Radiat Oncol Biol Phys. (2015) 92:469–74. doi: 10.1016/j.ijrobp.2015.01.018
16. Peucelle C, Nauraye C, Patriarca A, Hierso E, Fournier-Bidoz N, Martínez-Rovira I, et al. Proton minibeam radiation therapy: Experimental dosimetry evaluation. Med Phys. (2015) 42:7108–13. doi: 10.1118/1.4935868
17. Prezado Y, Jouvion G, Hardy D, Patriarca A, Nauraye C, Bergs J, et al. Proton minibeam radiation therapy spares normal rat brain: Long-Term Clinical, Radiological and Histopathological Analysis. Sci Rep. (2017) 7:14403. doi: 10.1038/s41598-017-14786-y
18. Charyyev S, Artz M, Szalkowski G, Chang C, Stanforth A, Lin L, et al. Optimization of hexagonal-pattern minibeams for spatially fractionated radiotherapy using proton beam scanning. Med Phys. (2020) 47:3485–95. doi: 10.1002/mp.14192
19. De Marzi L, Patriarca A, Nauraye C, Hierso E, Dendale R, Guardiola C, et al. Implementation of planar proton minibeam radiation therapy using a pencil beam scanning system: A proof of concept study. Med Phys. (2018) 45:5305–16. doi: 10.1002/mp.13209
20. Charyyev S and Wang CKC. Assessment of ambient neutron dose equivalent in spatially fractionated radiotherapy with protons using physical collimators. Radiat Prot Dosimetry. (2020) 189:190–7. doi: 10.1093/rpd/ncaa030
21. Henry T, Ureba A, Valdman A, and Siegbahn A. Proton grid therapy: A proof-of-concept study. Technol Cancer Res Treat. (2017) 16:749–57. doi: 10.1177/1533034616681670
22. Stengl C, Arbes E, Thai LYJ, Echner G, Vedelago J, Jansen J, et al. Development and characterization of a versatile mini-beam collimator for pre-clinical photon beam irradiation. Med Phys. (2023) 50:5222–37. doi: 10.1002/mp.16432
23. Schneider T, Patriarca A, Degiovanni A, Gallas M, and Prezado Y. Conceptual design of a novel nozzle combined with a clinical proton linac for magnetically focussed minibeams. Cancers. (2021) 13:4657. doi: 10.3390/cancers13184657
24. Schneider T, De Marzi L, Patriarca A, and Prezado Y. Advancing proton minibeam radiation therapy: magnetically focussed proton minibeams at a clinical centre. Sci Rep. (2020) 10. Available online at: https://www.nature.com/articles/s41598-020-58052-0.
25. Mayerhofer M, Datzmann G, Degiovanni A, Dimov V, and Dollinger G. Magnetically focused 70 MeV proton minibeams for preclinical experiments combining a tandem accelerator and a 3 GHz linear post-accelerator. Med Phys. (2021) 48:2733–49. doi: 10.1002/mp.14854
26. Bijl HP, van Luijk P, Coppes RP, Schippers JM, Konings AWT, and van der Kogel AJ. Unexpected changes of rat cervical spinal cord tolerance caused by inhomogeneous dose distributions. Int J Radiat Oncol Biol Phys. (2003) 57:274–81. doi: 10.1016/S0360-3016(03)00529-7
27. Prezado Y, Jouvion G, Patriarca A, Nauraye C, Guardiola C, Juchaux M, et al. Proton minibeam radiation therapy widens the therapeutic index for high-grade gliomas. Sci Rep. (2018) 8. Available online at: https://www.nature.com/articles/s41598-018-34796-8.
28. Zhang H, Ma L, Lim A, Ye J, Lukas L, Li H, et al. Dosimetric validation for prospective clinical trial of GRID collimator-based spatially fractionated radiation therapy: dose metrics consistency and heterogeneous pattern reproducibility. Int J Radiat Oncol Biol Phys. (2024) 118:565–73. doi: 10.1016/j.ijrobp.2023.08.061
29. Bertho A, Ortiz R, Maurin M, Juchaux M, Gilbert C, Espenon J, et al. Thoracic proton minibeam radiation therapy: tissue preservation and survival advantage over conventional proton therapy. Int J Radiat Oncol Biol Phys. (2024) 120:579–92. doi: 10.1016/j.ijrobp.2024.04.011
30. Li H, Mayr NA, Griffin RJ, Zhang H, Pokhrel D, Grams M, et al. Overview and recommendations for prospective multi-institutional spatially fractionated radiation therapy clinical trials. Int J Radiat Oncol Biol Phys. (2024) 119:737–49. doi: 10.1016/j.ijrobp.2023.12.013
31. Paganetti H, Blakely E, Carabe-Fernandez A, Carlson DJ, Das IJ, Dong L, et al. Report of the AAPM TG-256 on the relative biological effectiveness of proton beams in radiation therapy. Med Phys. (2019) 46:e53–78. doi: 10.1002/mp.13390
32. Paganetti H. Relative biological effectiveness (RBE) values for proton beam therapy. Variations as a function of biological endpoint, dose, and linear energy transfer. Phys Med Biol. (2014) 59:R419–72. doi: 10.1088/0031-9155/59/22/R419
33. Nabha R, De Saint-Hubert M, Marichal J, Esser J, Van Hoey O, Bäumer C, et al. Biophysical characterization of collimated and uncollimated fields in pencil beam scanning proton therapy. Phys Med Biol. (2023) 68. doi: 10.1088/1361-6560/acbe8d
34. Ueno K, Matsuura T, Hirayama S, Takao S, Ueda H, Matsuo Y, et al. Physical and biological impacts of collimator-scattered protons in spot-scanning proton therapy. J Appl Clin Med Phys. (2019) 20:48–57. doi: 10.1002/acm2.12653
35. Smith BR and Hyer DE. The LET enhancement of energy-specific collimation in pencil beam scanning proton therapy. J Appl Clin Med Phys. (2025) 26:e14477. doi: 10.1002/acm2.14477
36. Zhang W, Hong X, Wu W, Wang C, Johnson D, Gan GN, et al. Multi-collimator proton minibeam radiotherapy with joint dose and PVDR optimization. Med Phys. (2025) 52:1182–92. doi: 10.1002/mp.17548
37. Ortiz R, Belshi R, De Marzi L, and Prezado Y. Proton minibeam radiation therapy for treating metastases: A treatment plan study. Med Phys. (2023) 50:2463–73. doi: 10.1002/mp.16203
38. Ortiz R, De Marzi L, and Prezado Y. Preclinical dosimetry in proton minibeam radiation therapy: Robustness analysis and guidelines. Med Phys. (2022) 49:5551–61. doi: 10.1002/mp.15780
39. Perl J, Shin J, Schümann J, Faddegon B, and Paganetti H. TOPAS: An innovative proton Monte Carlo platform for research and clinical applications: TOPAS: An innovative proton Monte Carlo platform. Med Phys. (2012) 39:6818–37. doi: 10.1118/1.4758060
40. Charyyev S. Development of 2–3 mm proton minibeams as a new form of grid radiotherapy. Atlanta: Georgia institute of Technology (2019).
41. Chang C, Huang S, Harms J, Zhou J, Zhang R, Dhabaan A, et al. A standardized commissioning framework of Monte Carlo dose calculation algorithms for proton pencil beam scanning treatment planning systems. Med Phys. (2020) 47:1545–57. doi: 10.1002/mp.14021
42. Charyyev S, Chang CW, Zhu M, Lin L, Langen K, and Dhabaan A. Characterization of 250 meV protons from the varian proBeam PBS system for FLASH radiation therapy. Int J Partic Ther. (2023) 9:279–89. doi: 10.14338/IJPT-22-00027.1
43. Sammer M, Greubel C, Girst S, and Dollinger G. Optimization of beam arrangements in proton minibeam radiotherapy by cell survival simulations. Med Phys. (2017) 44:6096–104. doi: 10.1002/mp.12566
44. McNamara AL, Schuemann J, and Paganetti H. A phenomenological relative biological effectiveness (RBE) model for proton therapy based on all publishedin vitrocell survival data. Phys Med Biol. (2015) 60:8399–416. doi: 10.1088/0031-9155/60/21/8399
45. Wieser HP, Cisternas E, Wahl N, Ulrich S, Stadler A, Mescher H, et al. Development of the open-source dose calculation and optimization toolkit matRad. Med Phys. (2017) 44:2556–68. doi: 10.1002/mp.12251
46. Kalholm F, Grzanka L, Traneus E, and Bassler N. A systematic review on the usage of averaged LET in radiation biology for particle therapy. Radiother Oncol. (2021) 161:211–21. doi: 10.1016/j.radonc.2021.04.007
47. McMahon SJ. The linear quadratic model: usage, interpretation and challenges. Phys Med Biol. (2018) 64:01TR01. doi: 10.1088/1361-6560/aaf26a
48. Wagenaar D, Tran LT, Meijers A, Marmitt GG, Souris K, Bolst D, et al. Validation of linear energy transfer computed in a Monte Carlo dose engine of a commercial treatment planning system. Phys Med Biol. (2020) 65:025006. doi: 10.1088/1361-6560/ab5e97
49. Carabe-Fernandez A, Dale RG, and Jones B. The incorporation of the concept of minimum RBE (RbEmin) into the linear-quadratic model and the potential for improved radiobiological analysis of high-LET treatments. Int J Radiat Biol. (2007) 83:27–39. doi: 10.1080/09553000601087176
50. Wedenberg M, Lind BK, and Hårdemark B. A model for the relative biological effectiveness of protons: the tissue specific parameter α/β of photons is a predictor for the sensitivity to LET changes. Acta Oncol. (2013) 52:580–8. doi: 10.3109/0284186X.2012.705892
51. Fernandez-Palomo C, Chang S, and Prezado Y. Should peak dose be used to prescribe spatially fractionated radiation therapy?-A review of preclinical studies. Cancers (Basel). (2022) 14:3625. doi: 10.3390/cancers14153625
52. Lansonneur P, Mammar H, Nauraye C, Patriarca A, Hierso E, Dendale R, et al. First proton minibeam radiation therapy treatment plan evaluation. Sci Rep. (2020) 10:7025. doi: 10.1038/s41598-020-63975-9
53. Prezado Y, Dos Santos M, Gonzalez W, Jouvion G, Guardiola C, Heinrich S, et al. Transfer of Minibeam Radiation Therapy into a cost-effective equipment for radiobiological studies: a proof of concept. Sci Rep. (2017) 7:17295. doi: 10.1038/s41598-017-17543-3
54. Zwicker RD, Meigooni A, and Mohiuddin M. Therapeutic advantage of grid irradiation for large single fractions. Int J Radiat Oncol Biol Phys. (2004) 58:1309–15. doi: 10.1016/j.ijrobp.2003.07.003
55. Meyer J, Stewart RD, Smith D, Eagle J, Lee E, Cao N, et al. Biological and dosimetric characterisation of spatially fractionated proton minibeams. Phys Med Biol. (2017) 62:9260–81. doi: 10.1088/1361-6560/aa950c
56. Kirkpatrick JP, Meyer JJ, and Marks LB. The linear-quadratic model is inappropriate to model high dose per fraction effects in radiosurgery. Semin Radiat Oncol. (2008) 18:240–3. doi: 10.1016/j.semradonc.2008.04.005
Keywords: pMBRT, SFRT, minibeams, PVDR, biological dose, GRID, protons, LET
Citation: Dere K, Usta AC, Kurnaz RS, Lin L, Yang Y, Surucu M and Charyyev S (2025) Are we overestimating proton minibeam therapy effectiveness through physical dose metrics? Front. Oncol. 15:1682511. doi: 10.3389/fonc.2025.1682511
Received: 09 August 2025; Accepted: 30 November 2025; Revised: 16 October 2025;
Published: 11 December 2025.
Edited by:
Robert Griffin, University of Arkansas for Medical Sciences, United StatesReviewed by:
Wilfredo González, University of Malaga, SpainTianyuan Dai, Shandong Cancer Hospital and Institute, China
Copyright © 2025 Dere, Usta, Kurnaz, Lin, Yang, Surucu and Charyyev. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Serdar Charyyev, Y2hhcnl5ZXZAc3RhbmZvcmQuZWR1
Kaan Dere1