ORIGINAL RESEARCH article

Front. Oncol., 20 January 2026

Sec. Breast Cancer

Volume 15 - 2025 | https://doi.org/10.3389/fonc.2025.1716042

Dosimetric impact of bolus thickness and immobilization mask use in postmastectomy radiotherapy: a chest wall thickness–based analysis

Qing-Jun Shang&#x;Qing-Jun ShangMeng Zhang&#x;Meng ZhangBing-Xin ZhaoBing-Xin ZhaoQi WangQi WangWen GaoWen GaoBi-Yuan ZhangBi-Yuan ZhangHai-Ji WangHai-Ji WangTian-Hui Guo*Tian-Hui Guo*
  • Department of Radiation Oncology, The Affiliated Hospital of Qingdao University, Qingdao, Shandong, China

Purpose: Postmastectomy radiotherapy (PMRT) requires a balance between optimal target coverage and organ-at-risk (OAR) sparing. Thermoplastic masks and different bolus thicknesses are frequently applied in clinical practice, yet their combined dosimetric effects and dependence on chest wall thickness remain insufficiently defined.

Methods and materials: Seventeen breast cancer patients treated with PMRT were retrospectively analyzed. For each patient, six intensity-modulated radiotherapy (IMRT) plans were generated, combining the presence or absence of a thermoplastic mask with bolus thicknesses of 0, 3, and 5 mm. Dosimetric parameters of the skin, planning target volume (PTV), OARs, and treatment efficiency (monitor units, MU) were compared across plans. Subgroup analyses were performed according to chest wall thickness (<4.1 cm, 4.1–5.0 cm, >5.0 cm). Correlations between chest wall thickness and dosimetric indices were evaluated using Spearman’s rank analysis.

Results: Application of bolus improved dose homogeneity and PTV coverage, reduced lung dose and MU, but increased skin dose. The thermoplastic mask alone raised skin dose, functioning as an unintended compensator. In patients without masks, bolus effects on skin and OARs were more pronounced, suggesting greater sensitivity to compensation thickness. Chest wall thickness demonstrated negative correlations with skin dose, lung exposure, and MU. Patients with thin chest walls (<4.1 cm) derived the greatest benefit from bolus application, whereas patients with thick chest walls (>5.0 cm) showed minimal dosimetric improvement.

Conclusions: Both thermoplastic masks and bolus significantly affect PMRT dosimetry. Bolus provides improved PTV coverage and lung sparing at the cost of higher skin dose, while thermoplastic masks act as unintended compensators. Chest wall thickness strongly influences these effects, supporting individualized selection of bolus and mask use, particularly for patients with thinner chest walls.

1 Introduction

Breast cancer is the most common malignancy in women; despite improved survival, treatment optimization remains challenging (1, 2).

The standard treatment for breast cancer often involves a combination of surgery, systemic therapy (chemotherapy, hormonal therapy, and targeted therapy), and radiotherapy (RT) (3). Post-mastectomy radiotherapy (PMRT) is essential for reducing local recurrence and improving survival, especially in high-risk patients (4, 5). Technical factors such as immobilization devices and bolus application can influence target delineation and dose distribution (6, 7), yet the specific effects of different mask applications and compensation thicknesses on radiotherapy planning remain insufficiently studied.

Dosimetric evaluation in PMRT reflects the precision and uniformity of dose distribution. This study assesses the effects of thermoplastic mask use and compensation thickness on PMRT dosimetry, focusing on skin dose, target coverage, organ-at-risk exposure, and treatment efficiency. The influence of chest wall thickness on these parameters is also analyzed to clarify its role in target coverage and normal tissue sparing.

2 Materials and methods

2.1 Clinical data

Adhering to stratified design principles—specifically methodological guidelines for stratified dosimetric studies and non-parametric test requirements that recommend ≥5 subjects per subgroup to avoid statistical inefficiency and ensure valid intragroup comparisons via the Friedman test (8) -and building on numerous high-quality radiation oncology dosimetric studies (especially those with stratified designs analogous to ours) (9, 10), we ultimately enrolled 17 breast cancer patients who underwent modified radical mastectomy and required postoperative radiotherapy at The Affiliated Hospital of Qingdao University (June–December 2024, 10 left-sided, 7 right-sided). These patients were stratified by average chest wall thickness into three subgroups: low (<4.1 cm, n=6), intermediate (4.1–5.0 cm, n=5), and high (>5.0 cm, n=6), with each subgroup comprising at least 5 patients to fulfill the purpose of our stratified research. This study was approved by the ethics committees of the Affiliated Hospital of Qingdao University (Reference No. QYFYWZLL30486). CT simulation was performed with 3-mm slices from mandible to the second lumbar vertebrae.

2.2 Target area and organ at risk sketching

For each patient, two imaging sets were created with and without a thermoplastic mask to assess immobilization and bolus effects. Image registration was performed for alignment. The clinical target volume (CTV) included the chest wall and regional lymphatic areas, and a single oncologist ensured consistency. The PTV was generated by adding a 5-mm margin to the CTV with a 2-mm subcutaneous retraction. Skin dose was assessed using 3- and 5-mm sub-PTV layers. OARs (lungs, heart, spinal cord, contralateral breast, thyroid) were delineated.

2.3 Plan design

Treatment planning was conducted using the Varian Eclipse v15.6 treatment planning system (TPS) and delivered via a Varian Trilogy linear accelerator equipped with a 60-leaf multiple leaf collimator (MLC). For each patient, a total of six fixed-jaw intensity-modulated radiotherapy (fix jaw-IMRT) plans were created, divided into two groups by bolus application status (with/without), and each group comprised 3 subgroups stratified by bolus thickness (Table 1):

Table 1
www.frontiersin.org

Table 1. Radiotherapy plan grouping according to thermoplastic mask use and bolus thickness.

All treatment plans, which were designed by the same senior medical physicist, used 6MV X-rays, a fixed dose rate of 400 MU/min, and the anisotropic analytical algorithm (AAA) for dose calculation with a default grid resolution of 2.5 mm. The prescribed dose was 50 Gy in 25 fractions (5000 cGy/25F), ensuring 95% coverage of the target volume by the isodose line for normalization.

2.4 Plan evaluation

Plans were analyzed for four dosimetric categories, and chest wall thickness was correlated with dose distribution using dose–volume histograms (DVH).

2.4.1 Dosimetric parameters of the skin structure

The skin was defined at 3 and 5 mm below the surface to reduce surface dose uncertainty, consistent with prior studies (11, 12). Parameters included Dmax, Dmin, Dmean, and skin volumes receiving ≥105% (V105%) and ≥110% (V110%) of the prescription dose.

2.4.2 Dosimetric parameters of the target volume

Parameters included:

1. Homogeneity index (HI), defined as (13):

HI = (D2 D98)/D50,

where D2, D98 , and D50 represent the minimum doses covering 2%, 98%, and 50% of the target volume, respectively. A lower HI value, approaching zero, indicates better homogeneity of the target dose distribution.

2. Conformity index (CI), defined as (14):

CI = (VT, ref/VT) × (VT, ref/Vref),

where VT, ref is the target volume covered by the prescription dose, VT is the total target volume, and Vref is the total volume covered by the prescription dose. A CI value closer to 1 indicates better conformity of the target volume.

3. PTV dose parameters, including:

a. Dmax: Maximum dose within the PTV.

b. V95%: The percentage of the PTV receiving at least 95% of the prescription dose.

2.4.3 Dosimetric parameters of OARs

1. Lungs:

a. V5, V20, V40: The percentage of lung volume receiving at least 5 Gy, 20 Gy, and 40 Gy, respectively.

b. Dmean: Mean lung dose.

2. Heart:

a. V5, V10, V20: The percentage of heart volume receiving at least 5 Gy, 10 Gy, and 20 Gy, respectively.

b. Dmean: Mean heart dose.

2.4.4 Radiation delivery efficiency

Treatment efficiency was assessed by monitor units (MU); lower MU indicates higher cost-effectiveness.

2.4.5 Chest wall thickness

Defined as the perpendicular distance from skin to rib surface. Measurements at upper (thoracic inlet), middle (midpoint between inlet and xiphoid), and lower (xiphoid) chest wall were averaged. Patients were grouped as:

a. Low chest wall thickness: < 4.1 cm.

b. Intermediate chest wall thickness: 4.1–5.0 cm.

c. High chest wall thickness: > 5.0 cm.

2.5 Statistical analysis

All statistical analyses were performed using SPSS (v19.0, IBM Corp., Armonk, NY, USA). Continuous variables are presented as medians with interquartile ranges (IQRs). Dosimetric parameters among the six radiotherapy plans were compared using the Friedman test; significant results were followed by Bonferroni-adjusted Wilcoxon signed-rank tests.

Subgroup analyses by chest wall thickness (<4.1 cm, 4.1–5.0 cm, >5.0 cm) used the same procedures. Correlations between chest wall thickness and dosimetric parameters were assessed with Spearman’s rank coefficient. All tests were two-sided, with p < 0.05 considered significant; Bonferroni correction was applied for multiple comparisons.

3 Results

Comparison of the dosimetric parameters of the six radiotherapy plans for all 17 patients are listed in Table 2.

Table 2
www.frontiersin.org

Table 2. Comparison of the dosimetric parameters of the six radiotherapy plans of patients.

3.1 Dosimetric parameters of the skin structure

Statistically significant differences were observed among the six radiotherapy plans in terms of skin dose parameters, including Dmax, Dmin, and Dmean at 3mm and 5mm depths, as well as V105% and V110% (p < 0.001 for all except Dmax at 3mm, p = 0.003).

From the Dmin and Dmean values at 3mm and 5mm, Plan 1 (P-nomask) had the lowest skin dose. Comparison between masked and non-masked plans without bolus (Plan 4 vs. Plan 1) showed a significant increase in both parameters (p < 0.01), suggesting that the thermoplastic mask itself functions as a compensating layer, raising the surface dose (Figures 1A, B, D, E).

Figure 1
Nine violin plots labeled A through I compare various measurements across six different plans represented by different colors. Each plot includes statistical significance indicators (*, **, ***) showing differences between the plans. The measurement types include maximum dose, minimum dose, and volume metrics with given criteria. A legend clarifies the color coding for each plan.

Figure 1. Dosimetric parameters of the skin structure. (A) Dmin at 3mm Depth (cGy); (B) Dmean at 3mm Depth (cGy); (C) Dmax at 3mm Depth (cGy); (D) Dmin at 5mm Depth (cGy); (E) Dmean at 5mm Depth (cGy); (F) Dmax at 5mm Depth (cGy); (G) Skin ≥V105% at 3mm (%); (H) Skin ≥V105% at 5mm (%); (I) Skin ≥V110% at 3mm (%); (J) Skin ≥V110% at 5mm (%).

Since Dmax is highly variable and influenced by outliers, its clinical relevance is limited (Figures 1C, F).

For V105% and V110%, bolus in P-nomask plans caused a nonsignificant increase, whereas all P-mask plans showed significantly higher values than Plan 1 (p < 0.01), suggesting both mask and bolus expand high-dose regions and reduce dose (Figures 1G–I).

3.2 Dosimetric parameters of the target volume

Significant differences were observed among the six radiotherapy plans in terms of HI, CI, Dmax, and V95% (p < 0.001).

3.2.1 HI

Plan 3 showed significantly lower HI than Plan 1 (p < 0.001), indicating improved dose homogeneity with bolus. No significant difference was observed between Plan 1 and Plan 4, suggesting minimal impact of the thermoplastic mask (Figure 2A).

Figure 2
Box plots labeled A, B, C, D, showing comparisons across six plans for various metrics: HI, CI, Dmax, and V95. Significant differences are indicated with asterisks. Plans are color-coded: red, blue, green, purple, gray, and orange.

Figure 2. Dosimetric parameters of the target volume. (A) HI; (B) CI; (C) Dmax (cGy); (D) V95% (%). *p < 0.05, ** p < 0.01, *** p < 0.001 (Bonferroni-adjusted Wilcoxon signed-rank test following Friedman test).

3.2.2 CI

In the P-nomask group, Plans 2 and 3 trended toward higher CI versus Plan 1, without significance. In the P-mask group, plans 5 (p < 0.05) and 6 (p < 0.001) had significantly lower CI than Plan 4, indicating that bolus reduces conformity and increases dose heterogeneity, consistent with V105% and V110% results (Figure 2B).

3.2.3 Dmax and V95%

In the P-nomask group, bolus plans (2 and 3) showed higher V95% and lower Dmax compared to Plan 1 (Dmax: Plan 1 vs. 2, p < 0.01; V95%: Plan 1 vs. 2, p < 0.01; Plan 1 vs. 3, p < 0.001), indicating improved target coverage and reduced hot spots. Similar trends were observed in the P-mask group, showing bolus enhances PTV coverage and reduces dose heterogeneity in both settings (Figures 2C, D).

3.3 Dosimetric parameters of OARs

3.3.1 Lung dosimetry

Ipsilateral lung V5 was generally similar across groups, though Plan 1 had significantly higher V5 than Plan 2 (p < 0.05), suggesting greater low-dose exposure without mask or bolus (Figure 3A). No significant differences were observed for V20 (p > 0.05), but bolus showed a trend toward dose reduction (Figure 3B). For V40, Plan 1 was significantly higher than Plans 2 (p < 0.05) and 3 (p < 0.001), indicating bolus reduces high-dose lung exposure (Figure 3C). Dmean was higher in Plan 1 than in Plans 2 (p < 0.05) and 5 (p < 0.05), showing bolus and mask use help lower lung dose (Figure 3D). Overall, non-mask, non-bolus setups increased lung exposure, while bolus effectively reduced lung dose, particularly in high-dose regions.

Figure 3
Violin plots comparing the effects of six treatment plans on ipsilateral lung metrics.   A: V5 percentage shows significant differences between plans 1-2 and others.  B: V20 percentage, illustrating variance across plans with no significant notations.  C: V40 percentage indicates significant differences, particularly among lower-numbered plans.  D: Dmean values, highlighting significant differences between plans 1, 2, and others.   Legend identifies plan conditions by color: 1:P-nomask (red), 2:P-nomask-0.3 (blue), 3:P-nomask-0.5 (green), 4:P-mask (purple), 5:P-mask-0.3 (grey), 6:P-mask-0.5 (orange).

Figure 3. Dosimetric parameters of lung. (A) Ipsilateral Lung V5 (%). (B) Ipsilateral Lung V20 (%). (C) Ipsilateral Lung V40 (%). (D) Ipsilateral Lung Dmean (cGy). * p < 0.05, ** p < 0.01, *** p < 0.001 (Bonferroni-adjustedWilcoxon signed-rank test following Friedman test).

3.3.2 Cardiac dosimetry

Cardiac dose changes were minimal and nonsignificant (P > 0.05) (Supplementary Figure S1).

3.4 Radiation delivery efficiency

A significant difference in output dose was observed among the six plans (p < 0.001). Plan 1 had a median MU of 1782.0 (IQR: 1653.0–1907.5). In the P-nomask group, bolus plans (2 and 3) showed significantly lower output doses than Plan 1 (p < 0.01), with no difference between bolus thicknesses (0.3 vs. 0.5cm; p > 0.05). Comparison of Plan 1 and Plan 4 showed no significant difference (p > 0.05), indicating that mask use alone does not affect efficiency. In the P-mask group, adding bolus (Plans 5 and 6) further reduced output dose (p < 0.01), while thickness differences remained nonsignificant (p > 0.05). Overall, bolus significantly reduces output dose, with minimal effect of thickness, and masks alone have no independent impact but enhance dose reduction when combined with bolus (Figure 4).

Figure 4
Box and violin plot comparing total MU across six conditions labeled 1 to 6, differentiated by color. Significant differences marked by asterisks indicate p-values: * < 0.05, ** < 0.01, *** < 0.001.

Figure 4. Radiation delivery efficiency. * p < 0.05, ** p < 0.01, *** p < 0.001 (Bonferroni-adjusted Wilcoxon signed-rank test following Friedman test).

3.5 Relationship between chest wall thickness and dosimetric parameters in the target area

Patients were stratified by average chest wall thickness into low (<4.1cm), intermediate (4.1–5.0cm), and high (>5.0cm) groups. Six radiotherapy plans (P-mask, P-mask-0.3, P-mask-0.5, P-nomask, P-nomask-0.3, P-nomask-0.5) were compared within each group (Tables 35; Supplementary Figures S2-S4), and correlations between chest wall thickness and target dosimetric parameters are presented in Supplementary Figure S5.

Table 3
www.frontiersin.org

Table 3. Comparison of the dosimetric parameters in the low chest wall thickness group of patients.

Table 4
www.frontiersin.org

Table 4. Comparison of the dosimetric parameters in the intermediate chest wall thickness group of patients.

Table 5
www.frontiersin.org

Table 5. Comparison of the dosimetric parameters in the high chest wall thickness group of patients.

3.5.1 Skin dosimetric parameters

In the low-thickness group, most skin parameters (Dmin, Dmean at 3mm and 5mm; V105% and V110% at 3mm) differed significantly among the six plans (p < 0.05), while Dmax and V110% at 5mm showed no significant change, suggesting limited sensitivity to compensation (Table 3, Supplementary Figure S2). In the intermediate group, only Dmax and Dmean at 3mm differed significantly (p < 0.05), indicating a modest bolus effect (Table 4, Supplementary Figure S3). In the high-thickness group, P-mask with bolus plans had higher Dmean and V105%/V110%, but inter-plan differences were smaller; some parameters (Dmin, Dmean at both depths) remained significant (p < 0.05), while Dmax and V110% at 5mm did not (Table 5, Supplementary Figure S4).

3.5.2 Target volume dosimetric parameters

CI, HI, V95%, and Dmax varied across chest wall groups. In the low-thickness group, all indices showed significant inter-plan differences (χ² = 11.60–15.05, p < 0.05), with bolus plans, especially P-mask-0.5 and P-nomask-0.5, improving HI and V95%, indicating better homogeneity and coverage (Table 3). In the intermediate group, all indices remained significant (p < 0.05), confirming bolus efficacy (Table 4). In the high-thickness group, HI and V95% differences were no longer significant (p > 0.05), though CI remained sensitive (χ² = 23.17, p < 0.001, Table 5), suggesting bolus-induced improvements diminish with thicker chest walls.

3.5.3 Ipsilateral lung dosimetric parameters

In the low-thickness group, V40 and Dmean differed significantly among plans (χ² = 13.05 and 11.32, p < 0.05), with higher doses in non-bolus plans, while V5 and V20 showed no significant differences (Table 3, Supplementary Figure S2). In the intermediate group, no lung parameters reached significance, though non-bolus plans trended toward higher Dmean (Table 4, Supplementary Figure S3). In the high-thickness group, all lung dose indices were comparable across plans (p > 0.05) and overall exposure was lower than in thinner chest walls (Table 5, Supplementary Figure S4).

3.5.4 Radiation delivery

Output dose, reflecting radiation delivery efficiency, were also compared across the three chest wall thickness groups. Output dose differed significantly among plans in all thickness groups (low: χ² = 19.93; intermediate: χ² = 18.48; high: χ² = 15.40; p < 0.01). In low- and intermediate-thickness groups, P-nomask plans had the highest output dose, while bolus plans—particularly P-mask-0.3 and P-mask-0.5—showed lower doses, indicating improved treatment efficiency (Tables 3, 4). In the high-thickness group, differences were smaller, with P-mask-0.5 yielding the lowest output dose (Table 5). Across all groups, bolus consistently reduced output dose, with minimal difference between 0.3cm and 0.5cm thicknesses.

3.5.5 Correlation analysis

As shown in Supplementary Figure S5, chest wall thickness demonstrated a negative correlation with skin and lung dose parameters. Specifically, increases in chest wall thickness were associated with decreases in skin Dmean(r=-0.25, p<0.05 in 3mm; r=-0.26, p<0.01 in 5mm), V105%(r=-0.4, p<0.001 in 3mm; r=-0.43, p<0.001 in 5mm), and V110%(r=-0.29, p<0.01 in 3mm; r=-0.29, p<0.01 in 5mm), as well as reductions in ipsilateral lung V20(r=-0.4, p<0.001), V40(r=-0.38, p<0.001), Dmean(r=-0.43, p<0,001) and MU(r=-0.24, p<0.05).

4 Discussion

This study investigated the effects of radiotherapy techniques—including thermoplastic mask use and varying bolus thickness—on target delineation and dose distribution in post-mastectomy breast cancer patients. Bolus application improved dose homogeneity and target coverage while reducing high-dose lung exposure and overall output dose, enhancing treatment efficiency, particularly in patients with thinner chest walls. In contrast, thermoplastic mask use alone increased skin dose, indicating a compensatory effect independent of bolus.

We observed a negative correlation between chest wall thickness and both skin and lung dosimetric parameters: thicker chest walls were associated with lower skin Dmean, V105%, V110%, and reduced ipsilateral lung V20, V40, Dmean, and output dose, reflecting better natural dose attenuation. Most heart dosimetric parameters showed no significant differences, likely due to the limited sample size, especially for left-sided cases.

Bolus application improved target dose distribution and reduced high-dose lung exposure but increased skin dose. For patients with skin invasion or recurrence, bolus is recommended to enhance dermal dose, but potential acute skin toxicity should be considered (15, 16). While thermoplastic masks are essential for immobilization, they also increased skin dose, consistent with Lv et al. (9). Bolus thickness had greater impact on skin and OAR doses in non-mask patients, whereas masks provided partial compensation and reduced reliance on bolus adjustments. These findings support tailoring bolus strategies to the immobilization method, enabling flexible bolus selection in masked patients while preserving dosimetric quality, treatment efficiency, and comfort.

Our study confirms the dosimetric benefits of bolus use and identifies chest wall thickness as a key factor in postmastectomy radiotherapy planning. Subgroup analysis (<4.1 cm, 4.1–5.0 cm, >5.0 cm) revealed that thinner chest wall patients showed greater variability in skin dose, target coverage, and lung exposure. Bolus improved target homogeneity, coverage, and treatment efficiency in low and intermediate chest wall groups but had minimal impact in thicker chest walls. Chest wall thickness was negatively correlated with skin and lung doses, indicating natural dose attenuation. Overall, bolus use is most advantageous in thinner chest wall patients, whereas compensation in thicker chest walls can be individualized.

This study has several limitations. The small sample size and single-center design may limit generalizability, and subjective factors in target and OAR contouring could introduce systematic errors. Additionally, the chest wall thickness stratification requires validation in larger cohorts. Future studies should involve larger, multicenter populations to improve statistical power and external validity. Advanced image segmentation and automated contouring could reduce subjectivity, and multidimensional analyses integrating positional and anatomical data may better clarify the relationship between chest wall thickness and dosimetric parameters.

This study offers important insights for optimizing radiotherapy planning in post-mastectomy breast cancer. Our results show that the judicious selection of bolus material and immobilization methods can significantly influence target dose distribution and organ-at-risk protection. The inverse correlations between chest wall thickness and lung dose, skin dose distribution, as well as delivery efficiency, provide a promising basis for individualizing radiotherapy plans. Although the number of included patients was relatively small, each patient served as their own control, thereby minimizing inter-patient variability and enhancing the reliability of intra-patient dosimetric comparisons. As an exploratory dosimetric study, these findings provide valuable insights for optimizing bolus and mask application in PMRT, while larger multicenter cohorts are warranted to confirm their generalizability. Future research should focus on validating these findings and exploring adaptive planning strategies based on chest wall anatomy, ultimately paving the way for more personalized and safer radiotherapy protocols.

Conclusion

Bolus and mask use significantly affect dosimetric outcomes in PMRT. Bolus improves homogeneity and efficiency but increases skin dose, particularly in thin-chest-wall patients. Chest wall thickness should guide bolus application to optimize PMRT planning.

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

The studies involving humans were approved by the ethics committees of the Affiliated Hospital of Qingdao University. The studies were conducted in accordance with the local legislation and institutional requirements. Written informed consent for participation was not required from the participants or the participants’ legal guardians/next of kin in accordance with the national legislation and institutional requirements.

Author contributions

QJ: Writing – original draft. M-Z: Writing – original draft, Writing – review & editing. BX-Z: Writing – review & editing. QW: Writing – review & editing. W-G: Writing – review & editing. BY-Z: Writing – review & editing. HJ-W: Writing – review & editing. TH-G: Writing – review & editing, Funding acquisition, Methodology.

Funding

The author(s) declared that financial support was not received for this work and/or its publication.

Conflict of interest

The authors declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Generative AI statement

The author(s) declared that generative AI was not used in the creation of this manuscript.

Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary Material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fonc.2025.1716042/full#supplementary-material

Supplementary Figure 1 | Dosimetric Parameters of Cardiac. (A) Heart V5 (%)-Left. (B) Heart V10 (%)-Left. (C) Heart V20 (%)-Left. (D) Heart V5 (%)-Right. (E) Heart V10 (%)-Right. (F) Heart V20 (%)-Right. (G) Heart Dmean-Left (cGy). (H) Heart Dmean-Right(cGy).

Supplementary Figure 2 | Dosimetric parameters in the low chest wall thickness group.

Supplementary Figure 3 | Dosimetric parameters in the intermediate chest wall thickness group.

Supplementary Figure 4 | Dosimetric parameters in the high chest wall thickness group.

Supplementary Figure 5 | Correlation analysis of chest wall thickness with dosimetric parameters.

References

1. Torre LA, Bray F, Siegel RL, Ferlay J, Lortet-Tieulent J, Jemal A, et al. Global cancer statistics, 2012. CA Cancer J Clin. (2015) 65:87–108. doi: 10.3322/caac.21262

PubMed Abstract | Crossref Full Text | Google Scholar

2. Ferlay J, Soerjomataram I, Dikshit R, Eser S, Mathers C, Rebelo M, et al. Cancer incidence and mortality worldwide: sources, methods and major patterns in GLOBOCAN 2012. Int J Cancer. (2015) 136:E359–86. doi: 10.1002/ijc.29210

PubMed Abstract | Crossref Full Text | Google Scholar

3. Harbeck N and Gnant M. Breast cancer. Lancet. (2017) 389:1134–50. doi: 10.1016/S0140-6736(16)31891-8

PubMed Abstract | Crossref Full Text | Google Scholar

4. Helintö M, Blomqvist C, Heikkilä P, and Joensuu H. Post-mastectomy radiotherapy in pT3N0M0 breast cancer: is it needed? Radiother Oncol. (1999) 52:213–7. doi: 10.1016/s0167-8140(99)00099-7

PubMed Abstract | Crossref Full Text | Google Scholar

5. Kirova YM, Loap P, and Fourquet A. Benefit of post mastectomy radiation therapy (PMRT) in node-positive, HER2-positive patients with breast cancer receiving anti-HER2 treatments. Int J Radiat Oncol Biol Phys. (2020) 106:511–3. doi: 10.1016/j.ijrobp.2019.12.022

PubMed Abstract | Crossref Full Text | Google Scholar

6. Qun X, Wuyun J, KuiKui Z, Qiong W, and Jing C. Which technique of positioning and immobilization is better for breast cancer patients in postmastectomy IMRT, single-pole or double-pole immobilization? J Appl Clin Med Phys. (2019) 20:168–74. doi: 10.1002/acm2.12506

PubMed Abstract | Crossref Full Text | Google Scholar

7. Dahn HM, Boersma LJ, Ruysscher DD, Meattini I, Offersen BV, Pignol JP, et al. The use of bolus in postmastectomy radiation therapy for breast cancer: A systematic review. Crit Rev Oncol Hematol. (2021) 163:103391. doi: 10.1016/j.critrevonc.2021.103391

PubMed Abstract | Crossref Full Text | Google Scholar

8. Das S, Mitra K, and Mandal M. Sample size calculation: Basic principles. Indian J Anaesth. (2016) 60:652–6. doi: 10.4103/0019-5049.190621

PubMed Abstract | Crossref Full Text | Google Scholar

9. Ran L, Guangyi Y, Yongzhi H, and Yanhong W. Dosimetric effects of supine immobilization devices on the skin in intensity-modulated radiation therapy for breast cancer: a retrospective study. BMC Cancer. (2021) 21:384. doi: 10.1186/s12885-021-08119-6

PubMed Abstract | Crossref Full Text | Google Scholar

10. Thongthanom S and Nobnop. W. Dosimetric comparison of three-dimensional (3D), intensity-modulated radiotherapy (IMRT) and hybrid IMRT for left-sided postmastectomy radiation therapy (PMRT). J Radiotherapy Pract. (2023) 22:e105. doi: 10.1017/S1460396923000250

Crossref Full Text | Google Scholar

11. Levitin R, Salari K, Squires BS, Hazy AJ, Maywood MJ, Thrasher P, et al. Near-surface dose correlates with moist desquamation and unplanned reconstructive surgery in patients with implant-based reconstruction receiving postmastectomy radiation therapy. Adv Radiat Oncol. (2023) 8:101283. doi: 10.1016/j.adro.2023.101283

PubMed Abstract | Crossref Full Text | Google Scholar

12. Donghui W, Bin L, Luxi C, Zhenghuan L, Fantu K, Huamei Y, et al. Dosimetric effects of prone immobilization devices on skin in intensity-modulated radiation therapy for gynecologic cancer: a retrospective study. BMC Cancer. (2024) 24:1464. doi: 10.1186/s12885-024-13111-x

PubMed Abstract | Crossref Full Text | Google Scholar

13. Kataria T, Sharma K, Subramani V, Karrthic KP, and Bisht SS. Homogeneity Index: An objective tool for assessment of conformal radiation treatments. J Med Phys. (2012) 37:207–13. doi: 10.4103/0971-6203.103606

PubMed Abstract | Crossref Full Text | Google Scholar

14. Feuvret L, Noël G, Mazeron JJ, and Bey P. Conformity index: a review. Int J Radiat Oncol Biol Phys. (2006) 64:333–42. doi: 10.1016/j.ijrobp.2005.09.028

PubMed Abstract | Crossref Full Text | Google Scholar

15. Abel S, Renz P, Trombetta M, Cowher M, Werts ED, Julian TB, et al. Local failure and acute radiodermatological toxicity in patients undergoing radiation therapy with and without postmastectomy chest wall bolus: Is bolus ever necessary? Pract Radiat Oncol. (2017) 7:167–72. doi: 10.1016/j.prro.2016.10.018

PubMed Abstract | Crossref Full Text | Google Scholar

16. Nichol A, Narinesingh D, Raman S, Germain F, Chan EK, Tran E, et al. The effect of bolus on local control for patients treated with mastectomy and radiation therapy. Int J Radiat Oncol Biol Phys. (2021) 110:1360–9. doi: 10.1016/j.ijrobp.2021.01.019

PubMed Abstract | Crossref Full Text | Google Scholar

Keywords: breast cancer, postmastectomy radiotherapy (PMRT), chest wall thickness, compensator thickness, thermoplastic mask, dosimetric parameters

Citation: Shang Q-J, Zhang M, Zhao B-X, Wang Q, Gao W, Zhang B-Y, Wang H-J and Guo T-H (2026) Dosimetric impact of bolus thickness and immobilization mask use in postmastectomy radiotherapy: a chest wall thickness–based analysis. Front. Oncol. 15:1716042. doi: 10.3389/fonc.2025.1716042

Received: 30 September 2025; Accepted: 23 December 2025; Revised: 11 November 2025;
Published: 20 January 2026.

Edited by:

Sharon R. Pine, University of Colorado Anschutz Medical Campus, United States

Reviewed by:

Pejman Porouhan, Sabzevar University of Medical Sciences, Iran
Lucas Gomes Sapienza, University of Florida, United States

Copyright © 2026 Shang, Zhang, Zhao, Wang, Gao, Zhang, Wang and Guo. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Tian-Hui Guo, Z3VvdGhAcWR1LmVkdS5jbg==

These authors have contributed equally to this work

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.