Introduction: This study aimed to comprehensively analyze the dosimetric parameters, plan complexity, gamma passing rates (GPRs), and most importantly, the beam-on time (BOT) of stereotactic body radiotherapy (SBRT) for small-volume inoperable early-stage centrally-located non-small-cell lung cancer (NSCLC) at a radiotherapy center. The analysis was based on both single-arc (SA) and dual-arc (DA) VMAT techniques under the deep inspiration breath hold (DIBH) scenario.
Methods: We retrospectively selected 24 cases of small-volume inoperable early-stage centrally-located NSCLC treated with SBRT under the DIBH scenario at our institution between March 2021 and June 2024. The redesigned SA-VMAT plans (SA plans) adopted the same prescription dose of 50 Gy/5 fractions and flattening-filter free (FFF) beam as the original DA-VMAT plans (DA plans). The 2-group plans (i.e., the SA and DA plans) were normalized to cover 95% of the planning target volume (PTV) and 99% of the gross tumor volume (GTV) by the prescription dose. The evaluation factors included PTV parameters (D98%, D2%, HI, CI, and R50%), organs at risk (OARs), plan complexity (segments and MUs), GPRs, and BOT.
Results: The SA technique consistently yielded superior plans. Among the PTV parameters, the SA plans were superior to the DA plans in D98%, D2%, and HI (all p < 0.05), whereas the CI and R50% of the 2-group plans were comparable (all p > 0.05), and the SA plans had an increase in the ipsilateral PBT Dmax (p < 0.05). Otherwise, the differences between other OARs were insignificant (all p > 0.05). The SA plans had reduced complexity, with mean segments and mean MUs decreasing by 18.82% and 8.15%, respectively (all p < 0.001); the GPRs did not differ significantly under the three acquisition parameters (all p > 0.05). The mean BOT was reduced by 19.70% in SA plans (p < 0.001).
Discussion: The SA plans significantly shortened the BOT while maintaining comparable plan quality, thereby improving comfort for patients with small-volume inoperable early-stage centrally located NSCLC under the DIBH scenario. Future studies should accumulate more patient data to evaluate the long-term clinical outcomes of SA plans.
1 Introduction
In recent years, with the aging of the population and advancements in screening techniques, an increasing number of patients with inoperable early-stage centrally-located non-small-cell lung cancer (NSCLC) have been identified in clinical practice (1, 2). Centrally-located NSCLC is defined as tumors where the closest point is within 2 cm of (but not abutting) the proximal bronchial tree (PBT) or within 2 cm of (whether abutting or not) the mediastinal structures (3). Stereotactic body radiotherapy (SBRT) is increasingly recognized as a key treatment modality for inoperable early-stage centrally-located NSCLC (4). SBRT is characterized by a small number of fractions, high fractional dose, high conformality, and accurate delivery (5). Studies have shown that SBRT can safely treat patients with early-stage centrally-located NSCLC and achieve favorable outcomes (3, 6–8).
A major limitation of early-stage centrally-located NSCLC SBRT treatment is the dose to mediastinal organs at risk (OARs) and healthy lungs, with the radiotoxicity being most prominent in the ipsilateral PBT and healthy lungs (9, 10). This is because the ipsilateral PBT and healthy lungs are near the target volume, and the equivalent biological dose from SBRT is large. Radiotoxicity in the ipsilateral PBT includes fatal hemoptysis (11), and radiotoxicity in the healthy lungs includes fatal radiation pneumonitis in patients with poor respiratory function, particularly those with underlying interstitial lung disease (4). Therefore, it is essential to strictly control the dose of ipsilateral PBT and healthy lungs according to the relevant standard (3), which requires using the most potent modulation technique. Volumetric-modulated arc therapy (VMAT) is the most powerful modulation technique of all linac-based SBRT treatments, providing the best possible OAR sparing (12–14), as well as a significant decrease in treatment time (15, 16). However, the VMAT technique is susceptible to respiratory motion and produces interplay effects (17). Interplay effects can lead to hot and cold spots within the target volume, and dose variations of up to 10% in a single fraction have been observed (18). Compared to conventional treatments, SBRT has fewer fractions and a smaller target volume, making it more susceptible to interplay effects (19). To mitigate these effects, respiratory management strategies are crucial, and the deep inspiration breath hold (DIBH) technique is one of the most widely used approaches (20, 21).
The DIBH technique relies on specific tools such as the Active Breathing Coordinator (ABC) and Optical Surface Monitoring System (OSMS) to achieve respiratory control, thereby reducing interplay effects and the radiation dose to OARs (22). Under the DIBH scenario, patients attempt a maximal inspiratory volume during the simulations and the treatments, which allow the mediastinal OARs, such as the ipsilateral PBT, the healthy lungs, the esophagus, and the heart, to be removed from the high-dose volume (23) and potentially increase the target dose without increasing the OAR radiotoxicity (24). Ottosson et al. (22) observed that in the lung cancer radiotherapy, DIBH scenario compared to the free-breathing state, patients had an 86.8% increase in bilateral lung volume, a 14.8% decrease in gross tumor volume (GTV), an 11.6% and 19.9% decrease in bilateral V5Gy and V20Gy, respectively, and a 12.6% decrease in mean cardiac dose. Panakis et al. (25) demonstrated that, compared to the free-breathing state, the use of the ABC reduced Dmean, V20Gy, and V13Gy by 25%, 21%, and 18%, respectively, in healthy lungs.
However, the DIBH scenario requires patients to attempt a maximal inspiratory volume during simulations and treatments and to hold their breath for a sufficient period; therefore, minimizing the beam-on time (BOT) during treatment may improve patients’ comfort and the feasibility of the DIBH technique (26). In addition to using the VMAT technique to minimize BOT, it is necessary to use a high-dose-rate flattening filter-free (FFF) beam (27). Apart from shortening treatment time, the FFF beam offers other advantages, such as a smaller penumbra and reduced head scatter (28–30),, which result in a lower out-of-field dose (31, 32). Traditionally, the dual-arc VMAT plans (DA plans) have been used to treat lung cancer (33–35). However, DA plans still require a relatively long BOT and do not represent the minimum achievable BOT.
Based on years of clinical experience, one approach to shorten the BOT of VMAT-SBRT without compromising the plan quality is to reduce the number of arcs. This is achieved through the rational adjustment of segment sequencing parameters, resulting in a remarkable reduction in BOT. Multiple pieces of literature have reported the successful application of single-arc VMAT (SA) plans in the head and neck, pelvis, and many other tumor sites (36–39). Ning et al. (39) found that when comparing the plans of 20 patients with nasopharyngeal carcinoma of DA plans, SA plans reduced the BOT by 29.8% without sacrificing target coverage, conformity index (CI) and homogeneity index (HI). Panizza et al. (38) analyzed 20 prostate cancer cases and showed that, compared with DA plans, SA plans significantly reduced the average doses to the rectum and bladder; target coverage and HI were comparable, while BOT was decreased by 22%.
However, the application of SA plans in DIBH radiotherapy for small-volume early-stage centrally-located NSCLC has rarely been reported. This study aims to explore the feasibility of SA plans for small-volume inoperable early-stage centrally-located NSCLC under the DIBH scenario by comparing the dosimetric parameters, plan complexity, gamma passing rates (GPRs), and most importantly, the BOT of SBRT treatment between SA and DA plans.
2 Methods
2.1 Patient cohort
Twenty-four consecutive patients with small-volume inoperable early-stage centrally-located NSCLC treated under ABC-DIBH scenario with SBRT were retrospectively selected from March 2021 and June 2024. This study was approved by the Ethics Committee of the Second Affiliated Hospital of Zhengzhou University (ethics number: 2023202), and written informed consent was obtained from the patients to use their anonymized data. All methods were carried out in accordance with relevant guidelines and regulations. The original DA plans were all designed using a 50Gy/5 fractions prescription and a 6 MV-FFF beam. The general characteristics of the patients are shown in Table 1.
In accordance with the RTOG 0236 and 0813 protocols (3, 40), patients were screened with the following criteria: tumors (planning target volume, PTV) no larger than 5 cm and located outside the 0.5 cm area around the PBT and within the 2 cm area or immediately adjacent to the mediastinal or pericardial pleura. Patients with tumor extent that could not be defined on CT (e.g., tumors surrounded by solid lesions or pulmonary atelectasis) and patients with ultracentral tumors were excluded (41, 42). Moreover, in light of the clinical practice at our center, patients with inoperable early-stage NSCLC that is centrally located and whose tumors have a volume exceeding 50 cc have not been subjected to the five-fraction SBRT regimen. The primary reason for this was the concern over the occurrence of severe radiation toxicities like esophagotracheal fistula.
2.2 Positioning and contouring
Each patient underwent a spiral CT scan using a big-bore CT scanner (Philips Medical Systems, Cleveland, OH) with a slice thickness of 3 mm. All patients were placed in the supine position with arms raised and hands grasping the T-handle. The chest was immobilized using a vacuum bag (Guangzhou Klarity Medical Equipment Co., Ltd., Guangzhou, China) in conjunction with a surface-guided radiotherapy (SGRT) system, and radiotherapy was performed under the guidance of the SGRT system to ensure the positional accuracy of patients during treatment. The scanning area encompassed all OARs that were subject to evaluation, along with an extra 15 cm in both the cranial and caudal directions respectively (43).
According to the RTOG recommendations (3, 44), heterogeneity corrections were required for planning, and the GTV was delineated on lung windows. The PTV was created by adding an isotropic 5 mm margin to the GTV. The OARs that needed to be delineated included the healthy lungs (ipsilateral lung and lung all that exclude the GTV), the ipsilateral PBT, the spinal cord, the esophagus, the great vessels, the heart, the ipsilateral brachial plexus, and the skin.
2.3 Planning
Based on the Monaco 5.4 (Elekta AB, Stockholm, Sweden) treatment planning system, a new SA plan was designed to replicate the corresponding DA plan to ensure that the cost function and weights of the 2-group plans (i.e., the SA and DA plans) per patient remained constant. All plans were designed with a prescription of 50Gy/5 fractions and a 6MV-FFF beam. All plans were designed so that the dose constraints for the ipsilateral PBT and the healthy lungs needed to be met first, even prior to the target coverage. All plans were rescaled to meet the coverage requirement, specifically ensuring that 95% of the PTV and more than 99% of the GTV covered by the prescription dose. Moreover, all hot spots (with the dose range between 120% and 150% of the prescription dose) were positioned within the GTV. If OARs overlapped with the PTV or GTV, the minimum dose for the PTV or GTV in the overlapping volume should be at least 70% and 90% of the prescription dose, respectively. All plans were delivered using an Infinity accelerator equipped with an Agility collimator featuring 80 pairs of leaves and an isocenter projection thickness of 5 mm.
To maximize compliance with the OAR dose constraints, each of the 2-group plans was designed with one coplanar and one noncoplanar beam: one consisted of a 210-240° arc (table angle 0°) on the ipsilateral side, and the other of a 40-60° anterior arc (table angle 90°). The specific geometric parameters for the beam at a table angle of 0° are as follows: for left lung tumors, the counterclockwise arc has a start angle of 180° and an end angle ranging from 330° to 300°; for right lung tumors, the clockwise arc has a start angle of 180° and an end angle ranging from 30° to 60°. By contrast, for the beam at a table angle of 90°, the clockwise arc is configured with a start angle of 330-340° and an end angle of 20-30°, irrespective of whether the tumor is located in the left or right lung. All beam angles should be ascertained by means of conducting trial runs of the collision test in light of the patients’ body shapes. The arc increment was set at 10°, and the collimator angles ranged from -20° to +20°. These collimator angles were adjusted to be different from each other so as to minimize the tongue-and-groove effect (45). The isocenters for the 2-group plans were placed at the center of the PTV. Dose calculations for the 2-group plans were carried out by employing the Monte Carlo algorithm, with a dose grid resolution of 2.0 mm and a statistical uncertainty of 1% per plan. To minimize variables, the 2-group plans adopted the same PTV and OAR optimization constraints. In the segment optimization parameters, the SA plans were configured with one beam and one arc, having a maximum of 200 segments per arc, a minimum segment width of 0.50 cm, and applying low fluence smoothing. Meanwhile, the original DA plans were set to feature one beam and two arcs, with a maximum of 120 segments per arc, a minimum segment width of 1.0 cm, and adopting medium fluence smoothing.
2.4 Plan evaluation factors
Dosimetric parameters for PTV and OARs were evaluated for the newly optimized SA plans and compared with the DA plans.
The PTV evaluation parameters included D98%, D2%, HI, CI, and R50%, where D98% and D2% represent the approximate minimum and approximate maximum dose of the PTV, respectively. According to the RTOG 0813 protocol, D2% decreases with decreasing PTV. At PTV = 50 cc, D2% < 77.0 Gy; at PTV < 10 cc, D2% < 57.0 Gy. D98% > 45.0 Gy. HI describes the dose homogeneity within the PTV (46), with HI = D5%/D95%, where D5% and D95% represent the dose received by greater than 5% and greater than 95% of the PTV, respectively. Because the pursuit of PTV dose homogeneity will increase the OAR dose, the maximum HI must be set carefully when the PTV does not contain tissues that need to be retained functionally (47). HI < 1.6 is generally considered appropriate based on the radiotherapy center’s situation. CI and R50% were used to compare the high and intermediate dose spillage between the 2-group plans (48). CI = PIV/TV, where PIV is the prescription isodose volume, and TV is the target volume. CI takes a value ≥ 1; the closer the CI is to 1, the better. According to the RTOG 0813 protocol, CI is < 1.5, preferably < 1.2. R50% represents the ratio of the 50% prescription isodose volume to the PTV, with R50% increasing as the PTV decreases. According to the RTOG 0813 protocol, R50% is < 5.0 for PTV = 50 cc and <7.5 for PTV <10 cc.
OARs were compared in compliance with the RTOG 0813 protocol and other literature (41, 49), including V5Gy < 60% and V20Gy < 20% in the ipsilateral lung and V5Gy < 40% and V20Gy < 10% in the lung all; V18Gy < 4 cc and Dmax < 52.5 Gy in the ipsilateral PBT; spinal cord V22.5Gy < 0.25 cc, V13.5Gy< 0.5 cc and Dmax < 30 Gy; esophagus V27.5Gy < 5 cc and Dmax < 52.5 Gy; heart V32Gy < 15 cc and Dmax < 52.5 Gy; great vessels V47Gy < 10 cc and Dmax < 52.5 Gy; ipsilateral brachial plexus V30Gy < 3 cc and Dmax < 32 Gy; skin V30Gy < 10 cc and Dmax < 32 Gy.
Plan complexity assessment was performed based on segments and MUs, as these two parameters positively correlate with the plan complexity (35). Dose delivery accuracy, i.e., GPRs, was also recorded. Based on the AAPM TG 218 report and other literature (50, 51), and in conjunction with the radiotherapy center, the qualified GPRs were set as 2%/2mm > 95%, 2%/1mm > 85%, and 1%/2mm > 90%, with data below 5% of the maximum dose excluded. BOTs for the 2-group plans were recorded while measuring GPRs using SRS MapCHECK (equipped with StereoPHAN phantom) (Sun Nuclear, Melbourne, FL).
2.5 Statistical analysis
The paired Student’s t-test (SPSS 25.0, IBM SPSS Statistics for Windows, NY, USA) was used to compare the parameters of SA and DA plans, and a p value < 0.05 was considered statistically significant.
3 Results
3.1 PTV parameters
The PTV median [range] was 23.56 [8.52-40.66] cc. The differences in PTV parameters between the SA and DA plans are shown in Table 2. The PTV homogeneity of the SA plans was significantly better than that of the DA plans, which can be seen from the statistically significant differences in D98%, D2%, and HI between the 2-group plans (all p < 0.05). Compared with the DA plans, the SA plans exceeded the PTV D98% by 1.03 Gy, and the D2% and HI were lower by 1.96 Gy and 0.04, respectively. The 2-group plans all met the PTV D2% limit (D2% < 77 Gy) and the HI limit (HI < 1.6). One of the SA plans and two of the DA plans did not meet the PTV D98% limit (D98% > 45Gy). There were no significant differences in CI and R50% between the two groups (all p > 0.05), and all CIs met the standard. All but two SA plans and one DA plan met the R50% requirement (R50% < 5.0 for PTV < 50cc).
3.2 OAR sparing
The mean (SD) values of all OAR parameters for the SA and DA plans and the corresponding mean absolute dose difference are presented in Table 3. The OAR sparing parameters for the 2-group plans all followed the RTOG 0813 protocol, except for one DA plan that had a slight overdose of V15Gy for ipsilateral PBT (V15Gy = 4.2 cc > 4 cc). In the SA plans, a significant increase in the value of the ipsilateral PBT Dmax was observed (p < 0.05). The mean absolute dose difference was 1.19 Gy (2.47%) for ipsilateral PBT Dmax. For other OARs, in particular, the difference between the ipsilateral PBT V18Gy and the healthy lungs (ipsilateral lung and lung all that exclude the GTV) were not statistically significant (all p > 0.05). Some OAR data are not presented in Table 3 because they were all 0 (spinal cord V22.5Gy, ipsilateral brachial plexus V30Gy, and skin V30Gy) or overwhelmingly 0 (spinal cord V13.5Gy, esophagus V27.5Gy).
Referring to the PTV and OAR sparing parameters, the SA technique produced plans that followed the RTOG 0813 protocol in all cases. Figure 1 compares the dose distributions, and DVH of SA and DA plans for the same case.
Figure 1. Comparison between SA (A) and DA (B) dose distribution in the axial (left) and coronal (right) planes, with isodoses ranging from 500cGy to 5500 cGy. (C) DVH comparison between the 2-group plans of the Brachial Plexus L, Great Vessels, Heart, Ipsilateral PBT, Lung L, Lung R, PTV, and Spinal Cord for SA (dashed line) and DA (solid line) plans of the same case. The low-dose line (500cGy) of the SA plan showed a certain degree of retraction, while the other dose lines were comparable to those of the DA plan. The red and yellow arrows indicate the changes in the axial and coronal planes, respectively.
3.3 Plan complexity, the GPRs and BOT
Segments and MUs describe the plan complexity. As shown in Table 4, the segments of SA plans were significantly reduced compared to DA plans [132.07 (2.10) vs. 156.93 (13.12), p < 0.001], with an average reduction of 18.82%; similarly, the MUs of SA plans were significantly reduced compared to DA plans [3090.87 (419.90) vs. 3342.79 (498.02), p < 0.001], with an average reduction of 8.15%. Thus, the complexity of SA plans decreased.
Despite the reduced complexity, there were no significant differences in GPRs between the SA plans and DA plans under different acquisition parameters (all p > 0.05). All plans had qualified GPRs under different acquisition parameters.
The mean BOT of the SA plans were significantly shorter [146.78 (22.47) vs. 182.80 (24.97), p < 0.001], with a mean reduction of 19.70%. The maximum BOT of the SA plans were 189.20 s, whereas the maximum BOT of the DA plans were 226.33 s. Figure 2 shows the difference in the per-patient BOT of the 2-group plans, and it can be seen that all SA plans had a smaller BOT than the corresponding DA plans.
4 Discussion
This study compared dosimetric parameters, plan complexity, the GPRs, and most importantly, the BOT in 24 cases of small-volume inoperable early-stage centrally-located NSCLC treated with SBRT based on SA and DA techniques under the DIBH scenario. Both techniques provided treatment plans that complied with the RTOG 0813 protocol for patients. Dosimetric analysis confirmed the feasibility of the SA plans, characterized by more homogeneous PTV doses and comparable OAR sparing to the DA plans. This is in keeping with our aim of providing superior OAR sparing, such as for the ipsilateral PBT and healthy lungs. Compared to the DA plans, the decreased complexity of the SA plans did not significantly improve GPRs. However, the SA plans provided significant BOT reduction, and BOT reduction is one of the most desirable advantages under the DIBH scenario.
In general, SA plans have shorter treatment times than DA plans (38, 39, 52). However, different literatures have different opinions on the dosimetric quality of SA plans. Zhao et al. (52) suggested that in postoperative VMAT radiotherapy for breast cancer, SA plans had some disadvantages in target dosimetric parameters (CI and HI) compared with DA plans (p < 0.05). However, there was no difference between the two in terms of OAR sparing (p > 0.05). Ning et al. (39) proposed that in VMAT radiotherapy for nasopharyngeal cancer, the SA plans showed no significant change in PTV coverage and normal tissue dose compared to the DA plans. Panizza et al. (38) compared the advantages and disadvantages of the SA and DA plans for prostate cancer radiotherapy and found that the SA plans were more advantageous overall. The SA plan had significantly lower mean rectal and bladder doses and achieved comparable target coverage and homogeneity.
Guckenberger et al. (53) designed SA and DA plans for prostate, pharyngeal, and paranasal sinus cancer, respectively, and discovered that the SA plan quality was related to the target complexity. SA plans for simple targets can guarantee the plan quality and shorten the treatment time, while DA plans must be used for complex targets to guarantee the plan quality at the cost of increasing the treatment time. Characteristics of early-stage centrally-located NSCLC targets include small-volume tumor, single tumor, absence of positive lymph nodes and distant metastases, and a more regular shape (3, 44), and thus can be considered as simple targets. This was the initial reason for exploring the application of SA plans in this study. In our study, the PTV D98% exceeded 1.03 Gy, D2% and HI were reduced by 1.96 Gy and 0.04, respectively, in the SA plans, and the OAR parameters were barely statistically different from those of the DA plans (except for a significant increase of 1.19 Gy in the ipsilateral PBT Dmax, p < 0.05), which supported and validated our view.
In the radiotherapy of NSCLC, damage to the ipsilateral PBT is an important factor limiting the practice of SBRT (3), and central airway injury is 11 times more likely to occur in central lung cancer than in peripheral lung cancer (54). Another common adverse effect is radiation-induced lung injury. The incidence of grade 3–4 lung injury with self-limiting pneumonitis, dyspnea, fever, and chest pain after SBRT treatment for NSCLC had been reported to be 2.7% -27.0%, and grade 5 and above lung injury was uncommon (55). According to the RTOG 0813 protocol (3), none of the early-stage centrally-located NSCLC patients treated with SBRT in the 10 Gy/fx group had grade 3 adverse events with minimal impact on lung function within the first year, and there was no grade 5 or higher events after one year. This study observed a significant increase in the ipsilateral PBT Dmax in the SA plans (p < 0.05). However, the absolute value exceeded by only 2.47%. None of the differences between the other OAR parameters were statistically significant (all p > 0.05). Except for one case of a slight overdose of the ipsilateral PBT V18Gy in the SA plan (V18Gy = 4.2 cc > 4cc), the other OARs were well below the requirements of the RTOG 0813 protocol and, therefore, were not expected to result in severe radiotoxicity.
Regarding plan complexity, the segments and MUs of SA plans were all significantly reduced (all p < 0.001) by 24.86 and 251.92, respectively, which may be related to the reduction in the number of arcs. Previous researchers (56, 57) observed that segments and MUs increased significantly with the number of arcs used in VMAT plans. However, more segments and MUs contribute to higher plan quality (52) and do not necessarily lead to lower GPRs (58). In our study, the differences in GPRs between SA plans and DA plans were not significant (all p > 0.05), which is consistent with the results obtained in previous studies at other sites (38, 59). Compared with the DA plans, the segments and MUs of the SA plans were reduced by 18.82% and 8.15%, respectively, but the BOT was reduced by 19.70%. This may be because the reduction in BOT is mainly caused by the decrease in the number of arcs (from two to one).
Compared with DA plans, the segment optimization parameters for SA plans were moderately relaxed. The core rationale for this adjustment is that if identical segment optimization parameters were adopted for the 2-group plans, the dosimetric parameters of SA plans would be significantly inferior to those of DA plans, thereby rendering SA plans clinically unfeasible. To maintain plan quality while achieving a moderate reduction in BOT, we appropriately relaxed the constraints on segment optimization parameters for SA plans–a strategy that is highly consistent with the findings reported in existing literature. Previous studies have confirmed that minimum segment width and fluence smoothing degree directly affect plan MUs, which in turn influences the overall plan quality (60, 61). Additionally, other study (62) has indicated that as minimum segment width increases, the dosimetric parameters of PTV gradually deviate from the prescribed dose, with a significant increase in target dose inhomogeneity (p < 0.05).
Under the DIBH scenario, a shorter BOT is more conducive to improving patient comfort and the feasibility of implementing the DIBH technique (63–65). Insufficient patient compliance constitutes a major limitation of the DIBH technique. As noted by Mah et al. (66), only half of the lung cancer patients included in their study could successfully perform this method; therefore, minimizing the BOT is a prerequisite for the application of DIBH. Compared with 3D conformal radiotherapy and fixed-field intensity-modulated radiotherapy techniques, VMAT, as a rotational therapy technique, has significantly higher efficiency in lung cancer treatment (13, 67). Specifically, in this study, the BOT of SA plans were, on average, 19.70% (36.02 s) shorter than that of DA plans, with a mean BOT of 146.78 s and a maximum BOT of 189.2 s. Although conventional views (52, 53, 59, 68) have suggested that SA plans have a slightly poorer dose distribution while shortening the BOT, our study demonstrated that SA plans can achieve dosimetric parameters comparable to or even better than DA plans. Therefore, we will adopt SA plans for more small-volume inoperable early-stage centrally-located NSCLC patients treated with SBRT under the DIBH scenario in the future.
The strategy of implementing SA plans under the DIBH scenario has shown promising progress, but there is still much work to be done in future research. Firstly, constrained by the practical experience of our center, the PTV in this study was less than 41 cc, which imposes limitations on the generalizability of SBRT for early-stage centrally-located NSCLC. Secondly, it is essential to amass a greater number of patient cases with the aim of minimizing statistical errors. Lastly, the applicability of SA plans to other cancer types, such as breast cancer and peripheral lung cancer, warrants further investigation, and related studies will be pursued.
5 Conclusions
Compared with the DA plans, the SA plans can maintain high-quality PTV parameters while achieving comparable OAR sparing, notably for the ipsilateral PBT and healthy lungs. Furthermore, the SA plans significantly reduce the BOT. These attributes, with the reduction in BOT being particularly prominent, represent the most compelling advantages of SA plans for SBRT in treating small-volume, inoperable early-stage centrally-located NSCLC under the DIBH scenario.
Data availability statement
The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.
Ethics statement
The studies involving humans were approved by the Ethics Committee of the Second Affiliated Hospital of Zhengzhou University. The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study.
Author contributions
YH: Formal Analysis, Funding acquisition, Methodology, Software, Visualization, Writing – original draft. JJY: Data curation, Investigation, Methodology, Writing – original draft. AC: Investigation, Methodology, Software, Writing – original draft. JY: Project administration, Validation, Visualization, Writing – review & editing. YL: Conceptualization, Funding acquisition, Project administration, Resources, Supervision, Writing – review & editing.
Funding
The author(s) declared that financial support was received for this work and/or its publication. This work was supported by the National Natural Science Foundation of China, (no. 11965001), Henan Province Science and Technology Development Project (no. 232102310091), Henan Province Medical Science and Technology Project (no. LHGJ20210407), Key Scientific Research Project of Colleges and Universities in Henan Province (no. 25B320023), Guidance Program for Technological Innovation in the Medical and Health Field of Zhengzhou City (no. 2025YLZDJH355).
Conflict of interest
The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declared that generative AI was not used in the creation of this manuscript.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
1. Chang JY, Senan S, Paul MA, Mehran RJ, Louie AV, Balter P, et al. Stereotactic ablative radiotherapy versus lobectomy for operable stage I non-small-cell lung cancer: a pooled analysis of two randomised trials. Lancet Oncol. (2015) 16:630–7. doi: 10.1016/s1470-2045(15)70168-3
2. Haasbeek CJA, Palma D, Visser O, Lagerwaard FJ, Slotman B, and Senan S. Early-stage lung cancer in elderly patients: a population-based study of changes in treatment patterns and survival in the Netherlands. Ann Oncol. (2012) 23:2743–47. doi: 10.1093/annonc/mds081
3. Bezjak A, Paulus R, Gaspar LE, Timmerman RD, Straube WL, Ryan WF, et al. Safety and efficacy of a five-fraction stereotactic body radiotherapy schedule for centrally located non-small-cell lung cancer: NRG oncology/RTOG 0813 trial. J Clin Oncol. (2019) 37:1316–25. doi: 10.1200/jco.18.00622
4. Nagata Y and Kimura T. Stereotactic body radiotherapy (SBRT) for Stage I lung cancer. Jpn J Clin Oncol. (2018) 48:405–09. doi: 10.1093/jjco/hyy034
5. Videtic GMM, Donington J, Giuliani M, Heinzerling J, Karas TZ, Kelsey CR, et al. Stereotactic body radiation therapy for early-stage non-small cell lung cancer: Executive Summary of an ASTRO Evidence-Based Guideline. Pract Radiat Oncol. (2017) 7:295–301. doi: 10.1016/j.prro.2017.04.014
6. Bang A and Bezjak A. Stereotactic body radiotherapy for centrally located stage I non-small cell lung cancer. Transl Lung Cancer Res. (2019) 8:58–69. doi: 10.21037/tlcr.2018.10.07
7. Schanne DH, Nestle U, Allgäuer M, Andratschke N, Appold S, Dieckmann U, et al. Stereotactic body radiotherapy for centrally located stage I NSCLC: a multicenter analysis. Strahlenther Onkol. (2015) 191:125–32. doi: 10.1007/s00066-014-0739-5
8. Sun X, Li Y, Zhu Y, Li Q, Yuan X, Lin Q, et al. Four-year follow-up outcomes after stereotactic body radiation therapy for central early-stage non-small cell lung cancer. Transl Lung Cancer Res. (2020) 9:1472–82. doi: 10.21037/tlcr-20-851
9. Temming S, Kocher M, Stoelben E, Hagmeyer L, Chang DH, Frank K, et al. Risk-adapted robotic stereotactic body radiation therapy for inoperable early-stage non-small-cell lung cancer. Strahlenther Onkol. (2018) 194:91–7. doi: 10.1007/s00066-017-1194-x
10. Schneider BJ, Daly ME, Kennedy EB, Antonoff MB, Broderick S, Feldman J, et al. Stereotactic body radiotherapy for early-stage non-small-cell lung cancer: american society of clinical oncology endorsement of the american society for radiation oncology evidence-based guideline. J Clin Oncol. (2018) 36:710–19. doi: 10.1200/jco.2017.74.9671
11. Corradetti MN, Haas AR, and Rengan R. Central-airway necrosis after stereotactic body-radiation therapy. N Engl J Med. (2012) 366:2327–9. doi: 10.1056/NEJMc1203770
12. Dwivedi S, Kansal S, Shukla J, Bharati A, and Dangwal VK. Dosimetric evaluation of different planning techniques based on flattening filter-free beams for central and peripheral lung stereotactic body radiotherapy. Biomed Phys Eng Expr. (2021) 7(6). doi: 10.1088/2057-1976/ac2f0d
13. Xhaferllari I, El-Sherif O, and Gaede S. Comprehensive dosimetric planning comparison for early-stage, non-small cell lung cancer with SABR: fixed-beam IMRT versus VMAT versus TomoTherapy. J Appl Clin Med Phys. (2016) 17:329–40. doi: 10.1120/jacmp.v17i5.6291
14. Rauschenbach BM, Mackowiak L, and Malhotra HK. A dosimetric comparison of three-dimensional conformal radiotherapy, volumetric-modulated arc therapy, and dynamic conformal arc therapy in the treatment of non-small cell lung cancer using stereotactic body radiotherapy. J Appl Clin Med Phys. (2014) 15:4898. doi: 10.1120/jacmp.v15i5.4898
15. Pokhrel D, Halfman M, and Sanford L. FFF-VMAT for SBRT of lung lesions: Improves dose coverage at tumor-lung interface compared to flattened beams. J Appl Clin Med Phys. (2020) 21:26–35. doi: 10.1002/acm2.12764
16. Cai J, Malhotra HK, and Orton CG. Point/Counterpoint. A 3D-conformal technique is better than IMRT or VMAT for lung SBRT. Med Phys. (2014) 41:040601. doi: 10.1118/1.4856175
17. Edvardsson A, Nordström F, Ceberg C, and Ceberg S. Motion induced interplay effects for VMAT radiotherapy. Phys Med Biol. (2018) 63:085012. doi: 10.1088/1361-6560/aab957
18. Court L, Wagar M, Berbeco R, Reisner A, Winey B, Schofield D, et al. Evaluation of the interplay effect when using RapidArc to treat targets moving in the craniocaudal or right-left direction. Med Phys. (2010) 37:4–11. doi: 10.1118/1.3263614
19. Dufreneix S, Kirié C, and Autret D. Evaluation of stereotactic VMAT lung treatment plans for small moving targets. Phys Med. (2023) 107:102547. doi: 10.1016/j.ejmp.2023.102547
20. Zhao F, Shen J, Lu Z, Luo Y, Yao G, Bu L, et al. Abdominal DIBH reduces the cardiac dose even further: a prospective analysis. Radiat Oncol. (2018) 13:116. doi: 10.1186/s13014-018-1062-6
21. Botticella A, Levy A, Auzac G, Chabert I, Berthold C, and Le Pechoux C. Tumour motion management in lung cancer: a narrative review. Transl Lung Cancer Res. (2021) 10:2011–17. doi: 10.21037/tlcr-20-856
22. Ottosson W, Sibolt P, Larsen C, Lykkegaard Andersen JA, Borissova S, Mellemgaard A, et al. Monte Carlo calculations support organ sparing in Deep-Inspiration Breath-Hold intensity-modulated radiotherapy for locally advanced lung cancer. Radiother Oncol. (2015) 117:55–63. doi: 10.1016/j.radonc.2015.08.032
23. Persson GF, Scherman Rydhög J, Josipovic M, Maraldo MV, Nygård L, Costa J, et al. Deep inspiration breath-hold volumetric modulated arc radiotherapy decreases dose to mediastinal structures in locally advanced lung cancer. Acta Oncol. (2016) 55:1053–6. doi: 10.3109/0284186x.2016.1142115
24. Rosenzweig KE, Hanley J, Mah D, Mageras G, Hunt M, Toner S, et al. The deep inspiration breath-hold technique in the treatment of inoperable non-small-cell lung cancer. Int J Radiat Oncol Biol Phys. (2000) 48:81–7. doi: 10.1016/s0360-3016(00)00583-6
25. Panakis N, McNair HA, Christian JA, Mendes R, Symonds-Tayler JR, Knowles C, et al. Defining the margins in the radical radiotherapy of non-small cell lung cancer (NSCLC) with active breathing control (ABC) and the effect on physical lung parameters. Radiother Oncol. (2008) 87:65–73. doi: 10.1016/j.radonc.2007.12.012
26. Josipovic M, Aznar MC, Thomsen JB, Scherman J, Damkjaer SM, Nygård L, et al. Deep inspiration breath hold in locally advanced lung cancer radiotherapy: validation of intrafractional geometric uncertainties in the INHALE trial. Br J Radiol. (2019) 92:20190569. doi: 10.1259/bjr.20190569
27. Ong CL, Dahele M, Slotman BJ, and Verbakel WF. Dosimetric impact of the interplay effect during stereotactic lung radiation therapy delivery using flattening filter-free beams and volumetric modulated arc therapy. Int J Radiat Oncol Biol Phys. (2013) 86:743–8. doi: 10.1016/j.ijrobp.2013.03.038
28. Georg D, Knöös T, and McClean B. Current status and future perspective of flattening filter free photon beams. Med Phys. (2011) 38:1280–93. doi: 10.1118/1.3554643
29. Dwivedi S, Kansal S, Dangwal VK, Bharati A, and Shukla J. Dosimetry of a 6 MV flattening filter-free small photon beam using various detectors. BioMed Phys Eng Expr. (2021) 7(4). doi: 10.1088/2057-1976/abfd80
30. Sahani G, Sharma SD, Sharma PK, Deshpande DD, Negi PS, Sathianarayanan VK, et al. Acceptance criteria for flattening filter-free photon beam from standard medical electron linear accelerator: AERB task group recommendations. J Med Phys. (2014) 39:206–11. doi: 10.4103/0971-6203.144482
31. Xiao Y, Kry SF, Popple R, Yorke E, Papanikolaou N, Stathakis S, et al. Flattening filter-free accelerators: a report from the AAPM Therapy Emerging Technology Assessment Work Group. J Appl Clin Med Phys. (2015) 16:5219. doi: 10.1120/jacmp.v16i3.5219
32. Kry SF, Bednarz B, Howell RM, Dauer L, Followill D, Klein E, et al. AAPM TG 158: Measurement and calculation of doses outside the treated volume from external-beam radiation therapy. Med Phys. (2017) 44:e391–429. doi: 10.1002/mp.12462
33. Cheung EYW, Kwong VHY, Chan FYC, Cheng DYT, Cheng JKY, Yung SHY, et al. Modified VMAT plans for locally advanced centrally located non-small cell lung cancer (NSCLC). Life (Basel). (2021) 11(10):1085. doi: 10.3390/life11101085
34. El-Sherif O, Xhaferllari I, Johnson C, Chen J, and Gaede S. Poster - Thur Eve - 63: Dosimetric impact of breathing motion in lung SBRT: Dual vs single volumetric modulated arc therapy. Med Phys. (2012) 39:4636–37. doi: 10.1118/1.4740171
35. Fernandez DJ, Sick JT, and Fontenot JD. Interplay effects in highly modulated stereotactic body radiation therapy lung cases treated with volumetric modulated arc therapy. J Appl Clin Med Phys. (2020) 21:58–69. doi: 10.1002/acm2.13028
36. Kodama T, Kudo S, Hatanaka S, Hariu M, Shimbo M, and Takahashi T. Algorithm for an automatic treatment planning system using a single-arc VMAT for prostate cancer. J Appl Clin Med Phys. (2021) 22:27–36. doi: 10.1002/acm2.13442
37. Bertelsen A, Hansen CR, Johansen J, and Brink C. Single Arc Volumetric Modulated Arc Therapy of head and neck cancer. Radiother Oncol. (2010) 95:142–8. doi: 10.1016/j.radonc.2010.01.011
38. Panizza D, Faccenda V, Arcangeli S, and De Ponti E. Treatment optimization in linac-based SBRT for localized prostate cancer: A single-arc versus dual-arc plan comparison. Cancers (Basel). (2023) 16(1):13. doi: 10.3390/cancers16010013
39. Ning ZH, Mu JM, Jin JX, Li XD, Li QL, Gu WD, et al. Single arc volumetric-modulated arc therapy is sufficient for nasopharyngeal carcinoma: a dosimetric comparison with dual arc VMAT and dynamic MLC and step-and-shoot intensity-modulated radiotherapy. Radiat Oncol. (2013) 8:237. doi: 10.1186/1748-717x-8-237
40. Timmerman R, McGarry R, Yiannoutsos C, Papiez L, Tudor K, DeLuca J, et al. Excessive toxicity when treating central tumors in a phase II study of stereotactic body radiation therapy for medically inoperable early-stage lung cancer. J Clin Oncol. (2006) 24:4833–9. doi: 10.1200/jco.2006.07.5937
41. Raman S, Yau V, Pineda S, Le LW, Lau A, Bezjak A, et al. Ultracentral tumors treated with stereotactic body radiotherapy: single-institution experience. Clin Lung Cancer. (2018) 19:e803–e10. doi: 10.1016/j.cllc.2018.06.001
42. Lindberg K, Grozman V, Karlsson K, Lindberg S, Lax I, Wersäll P, et al. The HILUS-trial-a prospective nordic multicenter phase 2 study of ultracentral lung tumors treated with stereotactic body radiotherapy. J Thorac Oncol. (2021) 16:1200–10. doi: 10.1016/j.jtho.2021.03.019
43. Benedict SH, Yenice KM, Followill D, Galvin JM, Hinson W, Kavanagh B, et al. Stereotactic body radiation therapy: the report of AAPM Task Group 101. Med Phys. (2010) 37:4078–101. doi: 10.1118/1.3438081
44. Videtic GM, Hu C, Singh AK, Chang JY, Parker W, Olivier KR, et al. A randomized phase 2 study comparing 2 stereotactic body radiation therapy schedules for medically inoperable patients with stage I peripheral non-small cell lung cancer: NRG oncology RTOG 0915 (NCCTG N0927). Int J Radiat Oncol Biol Phys. (2015) 93:757–64. doi: 10.1016/j.ijrobp.2015.07.2260
45. Potter NJ, Yan G, Liu H, Alahmad H, Kahler DL, Liu C, et al. Beam flatness modulation for a flattening filter free photon beam utilizing a novel direct leaf trajectory optimization model. J Appl Clin Med Phys. (2020) 21:142–52. doi: 10.1002/acm2.12837
46. Wilke L, Andratschke N, Blanck O, Brunner TB, Combs SE, Grosu A-L, et al. ICRU report 91 on prescribing, recording, and reporting of stereotactic treatments with small photon beams: Statement from the DEGRO/DGMP working group stereotactic radiotherapy and radiosurgery. Strahlenther und Onkol: Organ Der Deutsch Rontgengesells. (2019) 195:193–98. doi: 10.1007/s00066-018-1416-x
47. Craft D, Khan F, Young M, and Bortfeld T. The price of target dose uniformity. Int J Radiat Oncol Biol Phys. (2016) 96:913–14. doi: 10.1016/j.ijrobp.2016.07.033
48. Shaw E, Kline R, Gillin M, Souhami L, Hirschfeld A, Dinapoli R, et al. Radiation Therapy Oncology Group: radiosurgery quality assurance guidelines. Int J Radiat Oncol Biol Phys. (1993) 27:1231–9. doi: 10.1016/0360-3016(93)90548-a
49. Xhaferllari I, Chen JZ, MacFarlane M, Yu E, and Gaede S. Dosimetric planning study of respiratory-gated volumetric modulated arc therapy for early-stage lung cancer with stereotactic body radiation therapy. Pract Radiat Oncol. (2015) 5:156–61. doi: 10.1016/j.prro.2014.08.009
50. Miften M, Olch A, Mihailidis D, Moran J, Pawlicki T, Molineu A, et al. Tolerance limits and methodologies for IMRT measurement-based verification QA: Recommendations of AAPM Task Group No. 218 Med Phys. (2018) 45:e53–83. doi: 10.1002/mp.12810
51. Pulliam KB, Huang JY, Howell RM, Followill D, Bosca R, O’Daniel J, et al. Comparison of 2D and 3D gamma analyses. Med Phys. (2014) 41:021710. doi: 10.1118/1.4860195
52. Zhao LR, Zhou YB, and Sun JG. Comparison of plan optimization for single and dual volumetric-modulated arc therapy versus intensity-modulated radiation therapy during post-mastectomy regional irradiation. Oncol Lett. (2016) 11:3389–94. doi: 10.3892/ol.2016.4376
53. Guckenberger M, Richter A, Krieger T, Wilbert J, Baier K, and Flentje M. Is a single arc sufficient in volumetric-modulated arc therapy (VMAT) for complex-shaped target volumes? Radiother Oncol. (2009) 93:259–65. doi: 10.1016/j.radonc.2009.08.015
54. Timmerman R, Hu C, Michalski J, Straube W, Galvin J, Johnstone D, et al. Long-term results of RTOG 0236: a phase II trial of stereotactic body radiation therapy (SBRT) in the treatment of patients with medically inoperable stage I non-small cell lung cancer. Int J Radiat Oncol Biol Phys. (2014) 90:S30. doi: 10.1016/j.ijrobp.2014.05.135
55. Murray P, Franks K, and Hanna GG. A systematic review of outcomes following stereotactic ablative radiotherapy in the treatment of early-stage primary lung cancer. Br J Radiol. (2017) 90:20160732. doi: 10.1259/bjr.20160732
56. Jiang X, Li T, Liu Y, Zhou L, Xu Y, Zhou X, et al. Planning analysis for locally advanced lung cancer: dosimetric and efficiency comparisons between intensity-modulated radiotherapy (IMRT), single-arc/partial-arc volumetric modulated arc therapy (SA/PA-VMAT). Radiat Oncol. (2011) 6:140. doi: 10.1186/1748-717x-6-140
57. Ramar N and Meher SR. A novel method for improving VMAT plan quality and deliverability. Asian Pac J Cancer Prev. (2022) 23:2159–67. doi: 10.31557/apjcp.2022.23.6.2159
58. Lee TF, Ting HM, Chao PJ, and Fang FM. Dual arc volumetric-modulated arc radiotherapy (VMAT) of nasopharyngeal carcinomas: a simultaneous integrated boost treatment plan comparison with intensity-modulated radiotherapies and single arc VMAT. Clin Oncol (R Coll Radiol). (2012) 24:196–207. doi: 10.1016/j.clon.2011.06.006
59. Radhakrishnan S, Chandrasekaran A, Sarma Y, Balakrishnan S, and Nandigam J. Dosimetric comparison between single and dual arc-volumetric modulated arc radiotherapy and intensity modulated radiotherapy for nasopharyngeal carcinoma using a simultaneous integrated boost technique. Asian Pac J Cancer Prev. (2017) 18:1395–402. doi: 10.22034/apjcp.2017.18.5.1395
60. Jiménez-Puertas S, Rodríguez AG, Cordero SL, González TG, Chamarro JD, Hernández MH, et al. Evaluation of the minimum segment width and fluence smoothing tools for intensity-modulated techniques in Monaco treatment planning system. J Med Phys. (2024) 49:250–60. doi: 10.4103/jmp.jmp_156_23
61. Wang Y, Chen L, Zhu F, Guo W, Zhang D, and Sun W. A study of minimum segment width parameter on VMAT plan quality, delivery accuracy, and efficiency for cervical cancer using Monaco TPS. J Appl Clin Med Phys. (2018) 19:609–15. doi: 10.1002/acm2.12422
62. Wang N, Chen L, Huang G, and Sun H. Influence of minimum segment width on intensity-modulated radiotherapy plan for left-sided breast cancer after breast conserving surgery. Med (Balt). (2023) 102:e36064. doi: 10.1097/md.0000000000036064
63. Josipovic M, Persson GF, Håkansson K, Damkjær SM, Bangsgaard JP, Westman G, et al. Deep inspiration breath hold radiotherapy for locally advanced lung cancer: comparison of different treatment techniques on target coverage, lung dose and treatment delivery time. Acta Oncol. (2013) 52:1582–6. doi: 10.3109/0284186x.2013.813644
64. Mørkeset ST, Lervåg C, Lund J, and Jensen C. Clinical experience of volumetric-modulated flattening filter free stereotactic body radiation therapy of lesions in the lung with deep inspiration breath-hold. J Appl Clin Med Phys. (2022) 23:e13733. doi: 10.1002/acm2.13733
65. Boda-Heggemann J, Mai S, Fleckenstein J, Siebenlist K, Simeonova A, Ehmann M, et al. Flattening-filter-free intensity modulated breath-hold image-guided SABR (Stereotactic ABlative Radiotherapy) can be applied in a 15-min treatment slot. Radiother Oncol. (2013) 109:505–9. doi: 10.1016/j.radonc.2013.09.014
66. Mah D, Hanley J, Rosenzweig KE, Yorke E, Braban L, Ling CC, et al. Technical aspects of the deep inspiration breath-hold technique in the treatment of thoracic cancer. Int J Radiat Oncol Biol Phys. (2000) 48:1175–85. doi: 10.1016/s0360-3016(00)00747-1
67. Shrimali RK, Arunsingh M, Reddy GD, Mandal S, Arun B, Prasath S, et al. Actual gains in dosimetry and treatment delivery efficiency from volumetric modulated arc radiotherapy for inoperable, locally advanced lung cancer over five-field forward-planned intensity-modulated radiotherapy. Indian J Cancer. (2017) 54:155–60. doi: 10.4103/ijc.IJC_79_17
Keywords: BOT, DIBH, inoperable early-stage centrally-located NSCLC, SA and DA plans, SBRT, small-volume, VMAT
Citation: Huang Y, Yuan J, Chu A, Yang J and Liu Y (2026) Small-volume plan optimization of inoperable early-stage centrally-located non-small-cell lung cancer using VMAT-based SBRT under the DIBH scenario: a single-arc model or a dual-arc plan? Front. Oncol. 16:1718548. doi: 10.3389/fonc.2026.1718548
Received: 04 October 2025; Accepted: 06 January 2026; Revised: 06 January 2026;
Published: 23 January 2026.
Edited by:
James C. L. Chow, University of Toronto, CanadaReviewed by:
Nourhane Moussawi, Beirut Arab University, LebanonAwusi Kavuma, Uganda Cancer Institute, Uganda
Christopher Thomas, Nova Scotia Health Authority, Canada
Copyright © 2026 Huang, Yuan, Chu, Yang and Liu. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Jun Yang, NTI2NTEyNjNAcXEuY29t; Yibao Liu, bHliMDFAdHNpbmdodWEub3JnLmNu
†These authors have contributed equally to this work and share first authorship
Yangyang Huang1,2†