ORIGINAL RESEARCH article

Front. Oncol., 13 September 2022

Sec. Cancer Imaging and Image-directed Interventions

Volume 12 - 2022 | https://doi.org/10.3389/fonc.2022.895575

Evaluation of a new method of calculating breast tumor volume based on automated breast ultrasound

  • JM

    Jing-Jing Ma 1

  • SM

    Shan Meng 2

  • SD

    Sha-Jie Dang 3

  • JW

    Jia-Zhong Wang 4

  • QY

    Quan Yuan 5

  • QY

    Qi Yang 6

  • CS

    Can-Xu Song 5*

  • 1. Department of Internal Medicine, Xi’an Fifth Hospital, Xi’an, China

  • 2. Department of Hematology, The Second Affiliated Hospital of Xi’an Jiaotong University, Xi’an, China

  • 3. Department of Anesthesia, Shaanxi Provincial Cancer Hospital, Affiliated to Xi’an Jiaotong University, Xi’an, China

  • 4. Department of General Surgery, The Second Affiliated Hospital of Xi’an Jiaotong University, Xi’an, China

  • 5. Department of Ultrasound, Shaanxi Provincial Cancer Hospital, Affiliated to Xi’an Jiaotong University, Xi’an, China

  • 6. Department of Surgery, Shaanxi Provincial Cancer Hospital, Affiliated to Xi’an Jiaotong University, Xi’an, China

Article metrics

View details

4

Citations

2,8k

Views

817

Downloads

Abstract

Objective:

To evaluate the effectiveness and advantages of a new method for calculating breast tumor volume based on an automated breast ultrasound system (ABUS).

Methods:

A total of 42 patients (18–70 years old) with breast lesions were selected for this study. The Ivenia ABUS 2.0 (General Electric Company, USA) was used, with a probe frequency of 6–15 MHz. Adobe Photoshop CS6 software was used to calculate the pixel ratio of each ABUS image, and to draw an outline of the tumor cross-section. The resulting area (in pixels) was multiplied by the pixel ratio to yield the area of the tumor cross-section. The Wilcoxon signed rank test and Bland-Altman plot were used to compare mean differences and mean values, respectively, between the two methods.

Results:

There was no significant difference between the tumor volumes calculated by pixel method as compared to the traditional method (P>0.05). Repeated measurements of the same tumor volume were more consistent with the pixel method.

Conclusion:

The new pixel method is feasible for measuring breast tumor volume and has good validity and measurement stability.

Introduction

Breast cancer ranks first in cancer deaths among women. The latest data on global cancer burden (2020) show that breast cancer accounts for 11.7% of all new cancer cases, officially displacing lung cancer as the most prevalent cancer type worldwide (1). In China, there were 420,000 new breast cancer cases, ranking it first in the world (2). Chemotherapy is one of the most commonly prescribed treatment methods for breast cancer (3), and its effectiveness relies heavily on imaging methods to evaluate tumor volume (4).

Various methods have been used to assess tumor volume (5). Magnetic resonance imaging (MRI) has excellent soft tissue resolution, uses no radiation, and offers multi-directional and multi-sequence imaging; however, the examination is time-consuming and expensive (6). Computerized tomography (CT) examination has high spatial and density resolution (7), but uses a large radiation dose and contrast agents with negative side effects. In comparison, ultrasound examination offers the advantages of being affordable, easy to administer, and radiation-free, and is thus recognized as the preferred imaging method for breast cancer (8). It is difficult to achieve accurate measurement of dynamic and complex entities with two-dimensional ultrasound, so volumetric parameters are measured with three-dimensional ultrasound (9, 10).

Automated breast ultrasound (ABUS) is a three-dimensional ultrasound technology for breast examination (11). It uses a standardized, automated imaging system that stores image data and has good repeatability (12). Its unique advantages contribute to its important role in the diagnosis and treatment of breast tumors (13). Currently, only tumor length can be measured on the ABUS system, not cross-sectional area or volume (14). Therefore, ABUS can only estimate volume according to the ellipsoid formula using tumor length, width and height (15). It is clinically necessary to overcome this limitation to ensure accurate and stable measurement of breast tumor volumes (16). Therefore, the authors designed a new method to measure tumor volume using ABUS and evaluated its validity and measurement stability.

Materials and methods

General information

This study was approved by the ethics committee of Shaanxi Provincial Cancer Hospital (2021-137) and granted a waiver of informed consent before commencement of the study. The ABUS imaging data of 42 patients with breast tumors who underwent ABUS examination in our hospital from June 2018 to June 2021 were retrospectively analyzed. The validity and measurement stability of the pixel method were compared with those of the traditional method (length × width × height/2). Inclusion criteria: 1) female patients; 2) aged 18–70 years old; 3) with breast tumors - if there were multiple breast tumors, the one with the largest length and diameter was selected; 4) the breast tumor had a well-defined boundary; 5) the long diameter of the tumor was ≥1cm and ≤5 cm. Exclusion criteria: 1) age < 18 years old or > 70 years old; 2) breast tumor with ill-defined boundary; 3) the long diameter of breast tumor was <1cm or >5 cm.

Instrument and ABUS inspection process

The Invenia ABUS 2.0 (General Electric Company, USA) was used, with a probe frequency of 6–15 MHz, field of view of 15.3 cm, scanning length of 16.9 cm, and maximum scanning depth of 5.0 cm. The patient was instructed to lie in a supine position and breathe calmly. Preset scanning conditions were selected on the instrument according to the size of the patient’s breast. Lateral, medial, and anteroposterior scans were performed bilaterally for all patients, and upper and lower scans were added for larger breasts. After the scan the images were imported into the image viewing system that comes with ABUS for 3D reconstruction, and transverse, sagittal, and coronal cross-sections were obtained.

Tumor volume measurement

ABUS imaging data of breast tumor volume was evaluated by two physicians, each using both the pixel method and the traditional method. The traditional method is to measure the length, width, and height of the tumor using the image viewing system that comes with ABUS. Using an ABUS coronal image, the length was defined as the largest diameter of the tumor and the width was defined as the largest diameter perpendicular to the length. Using an ABUS cross-section, the height was defined as the largest diameter of the tumor perpendicular to the plane of the ABUS probe on the cross-section was selected as the height (equivalent to the anteroposterior diameter of the tumor in vivo). The volume of the tumor was calculated by length*width*height/2. For the pixel method, first the pixel ratio was calculated according to the scale of the original image, and then the tumor was outlined in order to obtain the number of pixels in each cross-section. This process is done in Adobe Photoshop CS6 software. Tumor cross-section area was then calculated by multiplying pixel ratio by the number of pixels in each cross-section. On ABUS images, each breast tumor was divided on the coronal plane at 0.1-cm intervals. For each layer, the area was measured by pixel method, and volume was calculated by multiplying area by height (the height of each layer was 0.1 cm). The total tumor volume was calculated as the sum of the volumes of each layer. Consistency and stability of the two methods were then compared.

Statistical methods

SPSS 26.0 software was used to perform all statistical analyses. All measurement data were expressed as mean ± SD, and Bland-Altman plots or Wilcoxon signed rank test were used to compare the mean values and differences of the two methods. P < 0.05 was considered statistically significant.

Results

The 42 patients studied ranged from 19 to 65 years old, with an average age of 37.3 ± 12.6 years old. Tumor lengths ranged from 1.0 to 4.9 cm with an average length of 2.17 cm.

The analysis of tumor pixels is shown in Figures 1, 2. Coronal images were imported into Adobe Photoshop CS6 software, and the area-to-pixel ratio was calculated according to the scale on the image. Figure 3 shows the mean volume of each tumor as measured by both pixel and traditional methods.

Figure 1

Figure 2

Figure 3

Patients’ ages and tumor lengths are shown in Table 1, along with the differences in tumor volumes as measured by two physicians using both pixel and traditional methods. The mean volume of 42 lesions measured by the pixel method was 1.073 times that measured by the traditional method (standard deviation: 0.266, 95% confidence interval: 0.541–1.605). There was no significant difference in the volume measured by the pixel method as compared to that measured by the traditional method, indicating that the pixel method is feasible and effective for tumor volume measurement (signed rank test, P=0.542). The stability of the two methods was assessed by comparing the mean difference between physicians. The mean difference for the pixel method was significantly less than that of the traditional method (P<0.01).

Table 1

Patient numberAgeTumor length (cm)Tumor volume (cm3, Pixel method)Difference (Pixel method)Tumor volume (cm3, Traditional method)Difference (Traditional method)
Doctor1Doctor2Doctor1Doctor2
1461.80.6760.6650.0110.5310.5600.029
2312.51.5031.5580.0551.1521.5740.422
3302.51.5261.5410.0151.2101.0590.151
4202.44.0404.0400.0003.4663.0910.375
5192.12.5352.5800.0451.7162.4430.727
6191.40.1710.1720.0010.1170.1020.015
7463.310.17410.2610.08712.07812.7080.630
8361.00.2180.2080.0100.1770.1650.012
9292.20.7580.7600.0020.6920.7500.058
10401.00.3300.3580.0280.2460.2430.003
11321.60.5820.5780.0040.7880.7330.055
12511.20.3200.3250.0050.2630.2900.027
13371.80.7300.7190.0110.5520.6190.067
14281.40.2460.2450.0010.1650.1840.019
15411.60.4440.4410.0030.3670.4360.069
16651.30.5710.5860.0150.4440.5420.098
17351.81.3841.3280.0561.1541.2110.057
18252.21.4871.4510.0361.4031.5050.102
19512.62.0172.0220.0051.9261.9050.021
20371.30.2280.2270.0010.2750.3020.027
21211.20.3470.3290.0180.4210.3880.033
22281.30.6190.6100.0090.6110.5690.042
23182.31.5081.5180.0100.7241.3550.631
24361.90.4470.4490.0020.4310.4900.059
25451.40.3620.3760.0140.2270.2430.016
26551.20.2320.2320.0000.1740.2040.030
27271.10.2250.2600.0350.1480.1470.001
28251.40.3010.3110.0100.4790.5340.055
29372.92.5382.4830.0552.2572.4180.161
30471.81.2861.3840.0980.9311.4960.565
31254.910.60610.8480.24210.02311.7491.726
32574.14.5014.0990.4028.9927.9651.027
33282.84.5504.5980.0483.4803.8250.345
34261.90.2900.3020.0120.2520.4960.244
35593.13.2993.3580.0593.5123.0090.503
36524.48.8388.7480.0909.52310.3090.786
37322.01.0771.0900.0131.2281.0430.185
38253.82.9542.9840.0303.4564.0330.577
39592.91.1641.1980.0341.7741.9830.209
40552.63.1143.0760.0382.8914.2881.397
41272.21.3511.3440.0071.6652.0130.348
42513.04.4664.4210.0453.8894.9801.091

Patient age and tumor length, and difference in tumor volumes measured by two physicians using either the pixel or traditional method.

difference values were compared using the Wilcoxon signed rank test (W-=879, W+=24, T0.01(42) =247-656, P<0.01).

In addition, the consistency of the two physicians’ measurements was assessed using Bland-Altman analysis. As shown in Figure 4, the majority of tumor volumes measured by the pixel method were within the 95% confidence range. In contrast, more tumor volumes measured by the traditional method fell outside the 95% confidence range (shown in Figure 5), further illustrating the higher consistency of the pixel method. The mean and standard deviation of the difference between the measured values of the two physician based on the pixel method was smaller, and there was no statistical difference between the measured values of the two doctors (signed rank test, P=0.300). In contrast, there was a statistically significant difference between the two doctors’ measurement values using the traditional method (signed rank test, P=0.001) (Table 2). These results show that the pixel method has smaller measurement errors, more stable results, and less subjective influence by physician.

Figure 4

Figure 5

Table 2

Pixel methodTraditional method
DifferencePDifferenceP
Mean (SD)-0.0016 (0.0816)0.300-0.194 (0.475)0.001
median (Q1, Q3)-0.002 (-0.028,0.010)-0.058 (-0.348,0.012)

Comparison between the difference of tumor volumes measured by two physicians using either the pixel or traditional method.

difference values were compared using the signed rank test.

Discussion

In this study, a new pixel method was developed to measure the volume of breast tumors based on ABUS imaging. The results demonstrate that compared with the traditional method, the pixel method offers less error and more stability in measuring the volume of breast tumors.

ABUS uses a standardized, automated system for the acquisition and storage of image data (13, 17). Tomographic images similar to MRI and CT can be obtained in transverse, sagittal, and coronal planes, providing sufficient information from which to calculate target volume (18, 19). However, the current ABUS system can only measure tumor length, and has not yet been used to measure cross-sectional area or volume. The present study used pixel ratios to calculate cross-sectional areas and whole-tumor volumes from ABUS images. Results indicated that tumor volumes calculated by the pixel method were slightly higher than, but not significantly different from, those calculated by the traditional method. This indicates that the performance of the pixel method is comparable to that of the traditional method. Bland-Altman analysis showed that more tumor volumes measured by the pixel method fell within the 95% confidence range as compared to those measured by the traditional method, further illustrating the higher consistency of the two physicians’ measurements by the pixel method. The ultimate goal of the pixel method is to enable automated measurement of tumor volumes by a computer.

Not only does the pixel method offer smaller error and greater measurement stability, but it can also measure tumors with irregular shapes. However, one disadvantage is that the measurement time is long. In this study, the time required to calculate the breast tumor volume of each patient was about 2-10 minutes. If the method were to be merged into existing clinical software, rather than having to import to Photoshop, then the calculation time would be significantly reduced (20). In addition, the present study only considered tumors with well-defined boundaries. Many malignant breast lesions are ill-defined, and whether the pixel method would be suitable for them requires further investigation.

In conclusion, the pixel method is feasible and effective to measure the volume of breast tumors, with small error and good stability of the measured value.

Funding

This work was supported by scientific research project of Shaanxi Provincial Cancer Hospital (No. SXZL-2021-001A).

Acknowledgments

We thank Medjaden Inc. for scientific editing of this manuscript.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Statements

Data availability statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics statement

This study was approved by the ethics committee of Shaanxi Provincial Cancer Hospital (2021-137). The ethics committee waived the requirement of written informed consent for participation.

Author contributions

J-JM and SM contributed to the manuscript writing. S-JD and J-ZW conducted literature research and enrolled the participants. QYa and QYu contributed to the data acquisition and analysis. C-XS contributed to the study design and revised the manuscript. All authors contributed to the article and approved the submitted version. All authors agree to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of the work are appropriately investigated and resolved.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

References

  • 1

    SungHFerlayJSiegelRLLaversanneMSoerjomataramIJemalAet al. Global cancer statistics 2020: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J Clin (2021) 71:209–49. doi: 10.3322/caac.21660

  • 2

    XiaCDongXLiHCaoMSunDHeSet al. Cancer statistics in China and united states, 2022: profiles, trends, and determinants. Chin Med J (Engl) (2022) 135:584–90. doi: 10.1097/cm9.0000000000002108

  • 3

    AlkabbanFMFergusonT. Breast cancer. In: StatPearls. Treasure Island (FL: StatPearls Publishing (2021).

  • 4

    TripathyDImSAColleoniMFrankeFBardiaAHarbeckNet al. Ribociclib plus endocrine therapy for premenopausal women with hormone-receptor-positive, advanced breast cancer (MONALEESA-7): a randomised phase 3 trial. Lancet Oncol (2018) 19:904–15. doi: 10.1016/s1470-2045(18)30292-4

  • 5

    YounIChoiSChoiYJMoonJHParkHJHamSYet al. Contrast enhanced digital mammography versus magnetic resonance imaging for accurate measurement of the size of breast cancer. Br J Radiol (2019) 92:20180929. doi: 10.1259/bjr.20180929

  • 6

    LiHYaoLJinPHuLLiXGuoTet al. MRI And PET/CT for evaluation of the pathological response to neoadjuvant chemotherapy in breast cancer: A systematic review and meta-analysis. Breast (2018) 40:106–15. doi: 10.1016/j.breast.2018.04.018

  • 7

    LiJGaoWYuBWangFWangL. Multi-slice spiral CT evaluation of breast cancer chemotherapy and correlation between CT results and breast cancerspecific gene 1. J Buon (2018) 23:378–83.

  • 8

    Dobruch-SobczakKPiotrzkowska-WróblewskaHKlimondaZRoszkowska-PurskaKLitniewskiJ. Ultrasound echogenicity reveals the response of breast cancer to chemotherapy. Clin Imaging (2019) 55:41–6. doi: 10.1016/j.clinimag.2019.01.021

  • 9

    VourtsisA. Three-dimensional automated breast ultrasound: Technical aspects and first results. Diagn Interv Imaging (2019) 100:579–92. doi: 10.1016/j.diii.2019.03.012

  • 10

    SchmachtenbergCFischerTHammBBickU. Diagnostic performance of automated breast volume scanning (ABVS) compared to handheld ultrasonography with breast MRI as the gold standard. Acad Radiol (2017) 24:954–61. doi: 10.1016/j.acra.2017.01.021

  • 11

    DeprettoCLiguoriAPrimolevoADi CosimoSCartiaFFerrantiCet al. Automated breast ultrasound compared to hand-held ultrasound in surveillance after breast-conserving surgery. Tumori (2021) 107:132–8. doi: 10.1177/0300891620930278

  • 12

    VourtsisAKachulisA. The performance of 3D ABUS versus HHUS in the visualisation and BI-RADS characterisation of breast lesions in a large cohort of 1,886 women. Eur Radiol (2018) 28:592–601. doi: 10.1007/s00330-017-5011-9

  • 13

    ChiangTCHuangYSChenRTHuangCSChangRF. Tumor detection in automated breast ultrasound using 3-d CNN and prioritized candidate aggregation. IEEE Trans Med Imaging (2019) 38:240–9. doi: 10.1109/tmi.2018.2860257

  • 14

    ZhangPMaZZhangYChenXWangG. Improved inception V3 method and its effect on radiologists’ performance of tumor classification with automated breast ultrasound system. Gland Surg (2021) 10:2232–45. doi: 10.21037/gs-21-328

  • 15

    LiYWuWChenHChengLWangS. 3D tumor detection in automated breast ultrasound using deep convolutional neural network. Med Phys (2020) 47:5669–80. doi: 10.1002/mp.14477

  • 16

    LacsonRGoodrichMEHarrisKBrawarskyPHaasJS. Assessing inaccuracies in automated information extraction of breast imaging findings. J Digit Imaging (2017) 30:228–33. doi: 10.1007/s10278-016-9927-4

  • 17

    LeeCYChangTFChouYHYangKC. Fully automated lesion segmentation and visualization in automated whole breast ultrasound (ABUS) images. Quant Imaging Med Surg (2020) 10:568–84. doi: 10.21037/qims.2020.01.12

  • 18

    LagendijkMVosELRamlakhanKPVerhoefCKoningAHJvan LankerenWet al. Breast and tumour volume measurements in breast cancer patients using 3-d automated breast volume scanner images. World J Surg (2018) 42:2087–93. doi: 10.1007/s00268-017-4432-6

  • 19

    HuangAZhuLTanYLiuJXiangJZhuQet al. Evaluation of automated breast volume scanner for breast conservation surgery in ductal carcinoma in situ. Oncol Lett (2016) 12:2481–4. doi: 10.3892/ol.2016.4924

  • 20

    MoonWKHuangYSHsuCHChang ChienTYChangJMLeeSHet al. Computer-aided tumor detection in automated breast ultrasound using a 3-d convolutional neural network. Comput Methods Programs BioMed (2020) 190:105360. doi: 10.1016/j.cmpb.2020.105360

Summary

Keywords

breast lesion, volume, automated breast ultrasound system, pixel method, surgery

Citation

Ma J-J, Meng S, Dang S-J, Wang J-Z, Yuan Q, Yang Q and Song C-X (2022) Evaluation of a new method of calculating breast tumor volume based on automated breast ultrasound. Front. Oncol. 12:895575. doi: 10.3389/fonc.2022.895575

Received

14 March 2022

Accepted

26 August 2022

Published

13 September 2022

Volume

12 - 2022

Edited by

Po-Hsiang Tsui, Chang Gung University, Taiwan

Reviewed by

Fatemeh Ghahremani, Arak University of Medical Sciences, Iran; Sumit Kumar Banchhor, National Institute of Technology Raipur, India

Updates

Copyright

*Correspondence: Can-Xu Song,

This article was submitted to Cancer Imaging and Image-directed Interventions, a section of the journal Frontiers in Oncology

Disclaimer

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Outline

Figures

Cite article

Copy to clipboard


Export citation file


Share article

Article metrics