BRIEF RESEARCH REPORT article

Front. Pediatr., 22 September 2020

Sec. Children and Health

Volume 8 - 2020 | https://doi.org/10.3389/fped.2020.00588

Influence of the NICU on the Acoustic Isolation of a Neonatal Incubator

  • 1. Grupo de Investigación Entornos Acústicos, Facultad de Ingeniería y Ciencias Aplicadas, Universidad de Las Américas, Quito, Ecuador

  • 2. Laboratorio de Ingeniería Acústica, Universidad de Cádiz, Cádiz, Spain

  • 3. Unidad de Neonatología, Hospital Metropolitano, Quito, Ecuador

Abstract

The neonatal intensive care unit (NICU) is a very noisy place as compared to the intrauterine environment. To protect the neonate's health, international guidelines suggest avoiding noise levels above 45 dB in NICUs, but this recommendation is not normally met. The incubator acoustic isolation and the acoustic features of the NICU play important roles in determining the noise measured inside the incubator. In this study, the influence of two types of rooms, one with sound-absorbent covering and the other with reverberant surfaces, on the acoustic isolation of a neonatal incubator was evaluated using three acoustic isolation indexes: the level difference, the apparent sound reduction index, and the standardized level difference. Results show that the acoustic isolation of the incubator is very poor, with a level difference below 11 dBA at all frequencies. At 62.5 Hz, the level difference measured in both rooms exhibits a negative value, indicating that the incubator amplifies the noise coming from the NICU. Isolation of the incubator is poor, and the reverberation time (RT) of the containing room influences RT of the incubator, which is consequently higher when the containing room is reverberant; for example, the incubator RT in the reverberant NICU is 0.72 s higher at 500 Hz than that in a room with sound-absorbent covering.

Introduction

Numerous studies affirm that neonatal intensive care units (NICU) are noisy environments exceeding recommended sound levels (13) and that high noise levels have a significant impact on the health of preterm infants (48). The American Academy of Pediatrics uses the U.S. Environmental Protection Agency noise standards for hospitals (45 dB during the day and 35 dB at night) and recommends avoiding sound levels over 45 dB in NICU environments (9, 10). However, studies affirm that NICU noise levels exceed these recommendations between 70% (1) and 95.5% (2) of the time, exhibiting average noise levels ranging from 48 to 60.6 dBA (11, 12).

Although measurements indicating increases in NICU noise levels have been published over the years, recent studies show that this is still an unsolved problem. A study of different types of NICU facilities indicated an average equivalent continuous sound level (Leq) of 48 dBA inside the incubator, with the highest transient sounds at Leq = 81 dB (13); outside the incubator (also in different NICU types), two different studies reported measured noise levels of Leq = 55–65 dBA (13) and Leq = 54.7–58.1 dBA (14).

Several training programmes and procedures have been implemented to reduce NICU noise levels (1517), but none of them has resulted in reductions to levels below those recommended. A recent observational study (18) was undertaken with the goal of reducing NICU noise levels below the recommended 45 dBA. The maximum decrement only reduced levels to 55.3 dBA even though three different procedures were applied; these included awareness and education, environmental noise reduction measurements, and unit modification procedures.

Noisy environments affect premature infants more than full-term infants, as the peripheral auditory system is not yet fully developed in premature infants (19). The womb and the impedance mismatch between air and the embryonic fluid attenuate frequencies above 500 Hz by 40 to 50 dB, and consequently, the developing fetus is not exposed to the spectrum above this frequency (20). However, the spectral content of the noisy NICU environment is completely different from that of the intrauterine environment, and the auditory system of preterm infants is not prepared for it (21). As a consequence, preterm infants show distinctive responses to environmental stress within the NICU (22). Preterm infants cannot habituate to the acoustic stimuli, even after repeated exposures (23), and their heart rate, respiratory rate, and oxygen saturation increase in noisy environments more than those of full-term infants (4, 24).

According to Vohr (25), a stay in the NICU for more than 4 days constitutes a risk factor for hearing loss in neonates. Furthermore, the noisy NICU environment causes other adverse effects on infants' health (48). Stressful acoustic stimuli are related to future disorders in language and attention (26, 27). Heart rate, intracranial pressure, and oxygen saturation changes are some of the stressful physiologic responses (4, 2831) that may have a significant impact on infant neurological development (22).

The spectral content of NICUs shows a predominance of low frequencies (3, 32), with the addition of some mid-high frequency events resulting from equipment alarms and human voices (33). However, there is limited information on the specific effects caused by the low frequencies to which neonates are exposed. In that regard, a recent study suggested that if low frequency exposures have negative impacts on the health of exposed mice, which have a higher threshold for low frequencies than that for humans [375 (34) and 20 Hz, respectively], they could also have negative impacts on humans (35). Results show that the exposure of mice to low frequency noise with levels of 70 dB over 4 weeks causes permanent imbalance and a reduction of the number of calbindin-positive hair cells in the saccule and utricle (35).

Most studies have used Leq, which is the energy contained in the entire spectrum, to evaluate noise in NICUs and incubators; there have been some exceptions, including the studies carried out by Santos et al., Hernández Molina et al., and Fernández Zacarías et al. (3, 33, 36). However, this parameter does not indicate which frequencies are dominant. The acoustic attenuation caused by the incubator dome is crucial to the evaluation of noise within the incubator, but this aspect has largely gone unnoticed in most prior studies.

The motivation of this study is to provide an approach to improving the acoustic environment of NICUs and incubators to avoid potential neonatal health issues. Thus, the primary objective of this paper is to evaluate the influence of the NICU and incubator in the transmission of noise inside the incubator.

Methods

Physical Space and Measurement Procedures

Two different rooms containing the same incubator were studied to evaluate the incubator acoustic isolation and the influence of the containing room on the noise levels. These two rooms are located at the Universidad de Las Américas (Quito, Ecuador). The first one measured 1.79 × 3.87 × 2.52 m and was in the Sound and Acoustic Engineering Department. The second one was the CS1 room in the Simulation Medic Center of the Medicine Faculty and measured 3.87 × 6.09 × 2.5 m (Figure 1B). The first room has sound-absorbent walls with an average RT of 0.15 s, while the second one is much more reverberant (RT = 1.18 s) and is similar to a NICU room. Henceforth, the first room will be referred to as the “absorbent NICU” (Nabs) and the second one as the “reverberant NICU” (Nreverb). The designation “Nabs” is used to characterize the incubator acoustically. In Nabs, nearly all of the energy is attenuated by the absorbent walls. Conversely, the Nreverb is used to study the influence of the NICU on the incubator and represents an actual reference for the characteristics of the incubators and NICUs. Acoustic insulation measurements generate very high noise levels; therefore, they must be performed in NICUs without patients. Since there was no NICU meeting this requirement, it was necessary to use another room with similar characteristics; hence, we chose Nreverb. The dimensions of the incubator are ~0.85 × 0.41 × 0.40 m. The incubator was turned off, the access portholes were closed during measurements, and the mattress was contained inside (Figure 1A).

Figure 1

The Simulation Medic Center staff provided the facilities and gave advise on the general features and disposition of the NICU and incubators, and the Neonatal Unit of the Santa Bárbara Clinic of Quito provided the incubator studied.

Incubator acoustic isolation was measured using the standard ISO16283-1 (37) as a reference. It was used only as a reference because this standard is suitable for rooms with volumes between 10 and 250 m3, and the incubator space is significantly smaller. The incubator interior was considered as the receiver room (for both groups of measurements) because of the newborns' sensitive natures and the suggestion of the ISO standard in selecting the smallest room as a receiver (to ensure that the standardized level difference (DnT) is not overestimated) (37). To ensure straightforward data interpretation and to obtain more representative outcomes, the acoustic isolation was calculated in octave bands and not in 1/3 octave bands, as the ISO16283-1 recommends.

Incubator Acoustic Isolation in Nabs Room (Figure 1B)

The Nabs dimensions only provided for four microphone locations, requiring the sources to be placed at the corners to measure noise levels outside of the incubator. Levels were measured twice for each microphone–source combination (i.e., sixteen mic–source measurements). For measurements of the incubator energy distribution, 15 microphone positions were used inside the incubator; this matches the procedure used for small rooms in previous studies conducted by the authors (38, 39). A total of 60 noise level measurements were conducted, with four measurements in each of 15 positions. Subsequently, reverberation time (RT) and background noise were measured in the same fifteen microphone locations for each of the two impulsive sound source (balloon burst) locations.

Incubator Acoustic Isolation in Nreverb Room (Figure 1B)

Noise level measurements outside of the incubator were carried out using two sound sources and five microphone positions. Levels were measured twice for each microphone–source combination (i.e., 20 mic–source measurements). The number of measurements and mic-source combinations used to evaluate noise levels within the incubator, the background noise, and the RT were the same as those used in Nabs.

The ISO16283-1 standard recommends conducting measurements at low frequencies (from 50 to 80 Hz) when one or both rooms have a volume of <25 m3. In such small rooms, it is not easy to achieve a diffuse field, and additional measurements are required. In our case, the receiver has a volume of V = 0.147 m3 < 25 m3, so it was necessary to conduct low frequency measurements for noise level, background noise, and RT at the receiver. Noise level and background noise must be measured at least in four corners and the RT in a position other than a corner. The measurements carried out in the 15 incubator locations fulfill the previous requirements. Four additional measurements were also made in the corners of Nabsorb since its volume is below 25 m3. Although it is not required by ISO16283-1, the RTs of Nabs and Nreverb were also measured.

As a result of the measurements described above, four groups of acoustic parameters were obtained in each room for each of the octave bands studied; these included noise levels inside the incubator, Lin (Lin_63, Lin_125, Lin_250, Lin_500, Lin_1000, Lin_2000, Lin_4000), noise levels outside the incubator, Lout (Lout_63, Lout_125, Lout_250, Lout_500, Lout_1000, Lout_2000, Lout_4000), reverberation time inside the incubator, RTin (RTin_63, RTin_125, RTin_250, RTin_500, RTin_1000, RTin_2000, RTin_4000), and reverberation time outside the incubator, RTout (RTout63, RTout125, RTout250, RTout500, RTout1000, RT22000, RTout 4000).

The level difference (D), the apparent sound reduction index (R′), and DnT were calculated in this study (see ISO16283-1 for further information). The D index constitutes the simplest approach to the definition of sound insulation, and it is the level difference between the emission and receiver rooms. However, it is not suitable for comparing constructive solutions because the acoustic isolation also depends on the acoustic treatment of the receiver room. The second index is typically used to compare different materials, and it is the reference value used by manufacturers to characterize products acoustically. The third index is used for evaluating complex building solutions and for comparing in situ measurements, and it considers the RT measured in the receiver room. Considering that the scope of the present study does not include comparing different construction solutions because we were using the same incubator, indexes D and DnT were expected to yield more useful results.

The Mann–Whitney U test was applied to check the statistical significance of the differences between the measurements of the four subsets of acoustic parameters (Lin, Lout, RTin, and RTout) conducted in Nabs and Nreverb. The same test was applied to the RT measurements conducted in Nreverb, outside and inside the incubator (RTin_Nreverb and RTout_Nreverb). The acoustic indexes of both rooms were calculated with each sound source considered independently and for the overall number of measurements. The Mann–Whitney U test was also applied to determine if the differences between the acoustic indexes obtained for each room are statistically significant. Pearson's correlation coefficient was also calculated to determine if there is a relationship between the RT and the noise indicators and between RTin and RTout.

Room Impulse Response Measurements

The impulse response of a room (RIR) is a signal that contains information on the direct sound and early and late reflections arising when the room is excited by a sound source (38, 40). Additionally, the RIR describes the room sound energy and its decay; thus, by gathering the RIR of the incubator or NICU, it is possible to determine RTs and the sound pressure levels (SPLs). Subsequently, the RT and SPL may be used to estimate the acoustic isolation of the incubator.

The electroacoustic chain used for the measurements consists of a Gras 1/2" CCP free-field microphone, an external AVID sound card (i.e., AD/DA converter), and a computer-running Matlab to gather the data. In addition, the ITA toolbox (41) was used to obtain the RT and the SPL at each measurement position. For measurements of the incubator RIR, balloons were used instead of an omnidirectional source. A CESVA omnidirectional source and Gaussian white noise were used for measurements of RIR in the Nabs and Nreverb.

Results

Since the measured noise levels were 10 dB higher than the background noise, no background noise correction was applied. Measurements of RT, designated T30, were considered in this study; T30 is a measure of the time in which the SPL decays 30 dB, multiplied by 2.

The Kolmogorov–Smirnoff test was calculated to establish the goodness of fit to normality for the variables studied. Since some variables did not satisfy normality criteria, the non-parametric Mann–Whitney U test was applied as an alternative to parametric statistical studies. The data used for the analyses meet the assumptions necessary for the application of this test (42). The groups of the categorical-dependent variables are Nabs and Nreverb. The independent variables are the acoustic parameters and the acoustic isolation indexes in octave bands.

Results show that there are statistically significant differences between Nreverb and Nabs for all noise levels except for Lout_1000, Lin_1000, Lin_2000, and Lin_4000, with a significance level of 5% (Table 1). Within the frequency range, 250–4,000 Hz, the noise levels inside and outside the incubator are higher in Nreverb than they are in Nabs, but are lower in Nreverb for the other frequencies. Similarly, the RTs exhibit statistically significant differences between the rooms at all band frequencies and the values in Nreverb are higher than those in Nabs, with the exception of RTin_63. The RTs measured at 63 and 125 Hz in Nreverb are significantly higher inside the incubator than those measured outside. Still, the reverse is true (i.e., higher RTs outside the incubator) for frequencies from 500 to 4,000 Hz. No significant differences were found at 250 Hz.

Table 1

Noise levelsReverberation timesNoise indexes
Variablep-valueSubset of higher mean rankVariablep-valueSubset of higher mean rankVariablep-valueSubset of higher mean rank
Lout_630.000*NabsRTout_630.000*NreverbD630.037*Nreverb
Lout_1250.000*NabsRTout_1250.000*NreverbD1250.037*Nabs
Lout_2500.000*NreverbRTout_2500.000*NreverbD2500.037*Nabs
Lout_5000.000*NreverbRTout_5000.000*NreverbD5000.037*Nreverb
Lout_10000.080(Nreverb)RTout_10000.000*NreverbD10000.037*Nreverb
Lout_20000.000*NreverbRTout_20000.000*NreverbD20000.037*Nabs
Lout_40000.004*NreverbRTout_40000.000*NreverbD40000.037*Nreverb
Lin_630.000*NabsRTin_630.000*NabsDnT_630.050*Nreverb
Lin_1250.000*NabsRTin_1250.000*NreverbDnT_1250.827(Nabs)
Lin_2500.009*NreverbRTin_2500.000*NreverbDnT_2500.050*Nreverb
Lin_5000.000*NreverbRTin_5000.000*NreverbDnT_5000.050*Nreverb
Lin_10000.174(Nreverb)RTin_10000.000*NreverbDnT_10000.050*Nreverb
Lin_20000.692(Nreverb)RTin_20000.000*NreverbDnT_20000.050*Nreverb
Lin_40000.224(Nreverb)RTin_40000.000*NreverbDnT_40000.050*Nreverb

Mann–Whitney test applied to the acoustic parameters measured and to the noise indexes (level difference D and standardized level difference DnT) calculated for Nreverb and Nabs.

*

p-value ≤ 0.05.

Subsets of mean rank in parentheses indicate that the difference between groups is not significant.

The outcome of the Mann–Whitney U test applied to the isolation acoustic indexes shows that significant differences (p-value ≤0.05) existed between Nabs and Nreverb for all of the variables, except for Dnt_125 and R63. From the data with p-value ≤0.05, it can be concluded that the acoustic indicators calculated for Nreverb are significantly higher than those for Nabs, with the exception of D125, D250, and D2000 (which are lower).

Pearson's correlation coefficient (r) was calculated to compare the RT outside the incubator and the acoustic isolation indexes (for all the band frequencies). The correlation coefficients with the DnT index are very high for all the band frequencies (r > 0.9, p-value <0.001). High correlations also occur with D at 63, 500, 1,000, and 4,000 Hz and with R at 125 and 1,000 Hz (r > 0.9, p-value <0.001). The same trend can be observed for the correlation coefficients resulting for the RTin and Rout at high band frequencies, although the correlations at low frequencies are slightly lower (r63 = −0.494, p-value63 = 0.019; r125 = −0.515, p-value125 = 0.014).

Figure 2A shows D, R′, and DnT isolation indexes for the incubator located within Nreverb. The highest D attenuation is approximately 11 dBA at 1 kHz; the poorest attenuation is at 62.5 Hz (D = −0.25 dBA), followed by 125 Hz (D = 6.15 dBA), and the attenuation for other frequencies is approximately 8.5 dBA. A comparison of DnT and D for the Nreverb room shows that DnT is higher at all frequencies except for 4,000 Hz.

Figure 2

Figure 2B shows D, R′, and DnT indexes for each room. The D index at 62.5 Hz is also negative in Nabs (−5.13 dBA). The DnT is higher when the incubator is in Nreverb than when the incubator is in Nabs (Figure 2B).

Figure 3A shows that the average incubator RT is higher in the Nreverb than in the Nabs. However, there is an exception at 62.5 Hz; at that frequency, there was a small difference between the signal and background noise level in Nreverb, and it was not possible to measure T30. In this case, T20 was used as a substitute for T30 (T20 = time at which the SPL decays 20 dB, multiplied by 3). T30 was normally higher than T20 in the present experiments, and therefore, the T30 at 62.5 Hz could have also been higher in Nreverb than in Nabs at 62.5 Hz. Excluding that exception, the incubator RT in both rooms decreased as the frequency increased. In Nreverb, the RTs for low frequencies are higher inside the incubator than outside (Figure 3B). Furthermore, when both spaces (Nreverb and incubator) are independently excited by a 125-Hz signal, the sound persists 0.8 s longer in the incubator than it does in the NICU.

Figure 3

Discussion

The statistical analysis shows that most of the acoustic parameters measured in the two rooms are different. It is important to highlight that, although the same configuration of the sound source and measuring equipment was used in both rooms, measured noise levels for frequencies from 250 to 4,000 Hz were higher in Nreverb than they were in Nabs; this illustrates that a reverberant room amplifies noise levels. The fact that noise levels at low frequencies are lower in Nreverb results from the poor isolation of the glass walls for those frequencies, which leads to a higher noise transmission to the adjoining rooms. Therefore, similar to the case for the incubator and the room in which it is located, there is an acoustic coupling of the NICUs (with a high percentage of glass) with the adjoining rooms. This is undesirable because many electronic devices located in adjoining rooms may generate low frequency noise and the glass walls cannot protect the NICU environment from this noise.

When the incubator is in Nreverb, the highest acoustic isolation D occurs at 1 kHz and is approximately 11 dBA, which means that the incubator dome only poorly attenuates the noise coming from outside. Therefore, most of the low frequency sounds coming from the heating, ventilation, and air conditioning systems (HVAC) and other incubator engines reach the neonate easily. Similarly, and with regard to mid-high frequencies, poor attenuation results with human voices and noises from equipment with alarms. The incubator isolation is so poor that the RT of the containing room influences its RT. Pearson's correlation coefficient between the RT inside and outside the incubator is higher than 0.9 (p-value < 0.001), supporting this statement. For example, the incubator RT at 500 Hz is 0.72 s higher in Nreverb than it is when the incubator is in Nabs.

The negative D value at 62.5 Hz in Nabs and Nreverb means that the reverberant nature of the incubator causes amplification of the noise, leading to a higher noise level inside the incubator (relative to outside) at that frequency. These findings are in accordance with the suggestions made in other reports (43, 44).

DnT is an acoustic isolation index that considers the influence of the RT of the receiver room on the isolation. It is equal to D+10*log(T/T0), where T is the measured RT at the receiver and T0 is the reference RT value; the ISO16283-1 recommends a T0 = 0.5 s. Consequently, the term 10*log(T/T0) is positive when T > 0.5 s and in that case, DnT is higher than D. Comparing DnT and D for the same Nreverb room, DnT is higher at all frequencies except for 4,000 Hz, meaning that the RT inside the incubator exceeds the reference T0. For example, the RT inside the incubator for a frequency of 125 Hz is 1.59 s, three times higher than the aforementioned reference T0 (0.5 s). In Nabs, however, the DnT equation shows that the RTs at 62.5 and 125 Hz are higher than T0, but they are shorter than T0 at the other frequencies (Figure 2B).

Although the incubator DnT is higher when the incubator is in Nreverb than it is when the incubator is in Nabs (Figure 2B), this does not mean that the acoustic isolation is higher; this is also a result of the T0 used in ISO16283-1. Since the incubator acoustic isolation is so poor, sound waves go in and out through the incubator walls readily. Therefore, the RT of the room in which the incubator is located affects the incubator RT (note that this phenomenon is not usually observed in the ordinary dwelling rooms for which ISO16283-1 was designed). Since the RT of Nreverb is higher than Nabs, the incubator RT is also higher when it is in Nreverb than when in Nabs, and consequently, so is DnT.

The T0 recommended in the ISO16382-1 standard was chosen for typical dwelling rooms containing furniture, for which the RT is reasonably independent of volume and frequency and is approximately equal to 0.5 s for most frequencies (37). According to the present results, a T0 = 0.5 s is not suitable for small and reverberant rooms. In this regard, the authors of (45) conducted RT measurements in furnished and unfurnished rooms; their results showed that the average RT for unfurnished rooms is ~2.5 s at frequencies below 500 Hz and decreases to 1.3 s at frequencies of 5 kHz. In addition, the standard deviations for unfurnished rooms were higher than the furnished one, which is probably because the authors evaluated rooms with different volumes (ranging from ~10 to 200 m3) (45). Consequently, the RTs for unfurnished rooms are volume and frequency dependent. However, it is difficult to establish a reference value for the DnT equation that suits the characteristics of the incubator, since the volume is not comparable with those studied in (45). Thus, more research on the RT reference value (T0) is needed to calculate the isolation indexes for small and reverberant spaces, since neither the DnT nor the D indexes show the real effect of the RT on the acoustic isolation.

Although the traditional isolation indexes used in construction seem to be inappropriate for characterizing the acoustic isolation of neonatal incubators, the present outcomes show that incubator isolation is very poor, especially at low frequencies. This is quite problematic since many studies affirm that most of the time, there are elevated low frequency intensities in NICUs (3, 32) caused by electronic devices (46). Although the information is limited, some studies suggest that low frequency exposure may have negative effects on neonates, as is the case for animals with auditory thresholds higher than those of humans (35).

There is a strong relationship between the RT outside the incubator and the DnT index for all band frequencies (r > 0.9, p-value < 0.001), although the DnT equation does not consider the RT outside, but only inside, the incubator. Moreover, the high correlation between D (Lout-Lin) and the RT outside the incubator implies a strong influence of the RT of the room containing the incubator on incubator isolation.

Since Nreverb is a room designed for teaching activities, it has an acoustic absorbent false ceiling not normally found in traditional NICUs. Therefore, although it is quite reverberant for its volume, in a real NICU, the reverberation times obtained would have been higher.

Not only does the incubator exhibit poor isolation, but it also amplifies sound at low frequencies; therefore, the problem of high noise levels in NICUs is worsened from the perspective of the preterm infant. It would seem that incubators must be designed for improved acoustic comfort since Coston and Aune (18) have shown that reducing NICU noise levels is complicated. Conversely, another way to decrease noise levels inside incubators is to treat NICUs acoustically, since our results show that the incubator reverberation time is influenced by the NICU reverberation time; therefore, reductions in NICU RTs will lead to reductions in the RTs and SPLs of the incubator.

Conclusions

The results of the present study on the acoustic isolation and reverberant nature of neonatal incubators establish that incubators and traditional NICUs provide acoustic environments that are inappropriate for neonates. The results of the present study may provide clinical managers with evidentiary support for the need to redesign traditional NICUs.

Statements

Data availability statement

The original contributions presented in the study are included in the article, further inquiries can be directed to the corresponding author.

Author contributions

VP-R conceptualized and designed the study, carried out the acoustic measurements and initial analyses, drafted the initial manuscript, and reviewed and revised the manuscript. DN-S carried out the acoustic measurements, carried out the initial analyses, drafted the initial manuscript, and reviewed and revised the manuscript. RH-M coordinated and supervised data collection and critically reviewed the manuscript and provided important intellectual content. EJ-M assessed neonatal intensive care units and electrical devices and critically reviewed the manuscript and provided important intellectual content. All authors contributed to the article and approved the submitted version.

Funding

The research leading to these results has received funding from the V Call for Research Projects of the Universidad de Las Américas (Project reference SOA.VPR.18.04).

Acknowledgments

We would like to thank the staff of the Medical Simulation Centre of the Universidad de Las Américas for their wise advice and for allowing us to carry out the measurements at their facilities, especially considering the noise nuisance associated with the studies.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

    Abbreviations

  • D

    Level difference isolation index

  • DnT

    Standardized level difference isolation index

  • HVAC

    Heating, ventilation, and air conditioning

  • Lin

    Noise level inside the incubator

  • Lin_63, Lin_125, Lin_250, Lin_500, Lin_1000, Lin_2000, Lin_4000

    Noise level inside the incubator at the band frequencies of 63, 125, 250, 500, 1,000, 2,000, and 4,000 Hz, respectively

  • Lout

    Noise level outside the incubator

  • Lout_63, Lout_125, Lout_250, Lout_500, Lout_1000, Lout_2000, Lout_4000

    Noise level outside the incubator at the band frequencies of 63, 125, 250, 500, 1,000, 2,000, and 4,000 Hz, respectively

  • Nabs

    Room with acoustic absorbent covering

  • NICU

    Neonatal intensive care unit

  • Nreverb

    Room with reverberant surfaces

  • R′

    Apparent sound reduction index

  • RIR

    Impulsive response of a room

  • RT

    Reverberation time

  • RTin

    Reverberation time inside the incubator

  • RTin_63, RTin_125, RTin_250, RTin_500, RTin_1000, RTin_2000, RTin_4000

    Reverberation time inside the incubator at the band frequencies of 63, 125, 250, 500, 1,000, 2,000, and 4,000 Hz, respectively

  • RT in_Nreverb

    Reverberation time inside the incubator measured in the room with reverberant surfaces

  • RTout

    Reverberation time outside the incubator

  • RTout_63, RTout_125, RTout_250, RTout_500, RTout_1000, RTout_2000, RTout_4000

    Reverberation time outside the incubator at the band frequencies of 63, 125, 250, 500, 1,000, 2,000, and 4,000 Hz, respectively

  • SPL

    Sound Pressure level

  • RTout_Nreverb

    Reverberation time outside the incubator measured in the room with reverberant surfaces

  • T0

    Reference reverberation time of the standardized level difference equation defined by the ISO16283-1

  • T

    Reverberation time of the receiver room within the standardized level difference equation

  • T20

    Time in which the sound pressure level decays 20 dB multiplied by 3

  • T30

    Time in which the sound pressure level decays 30 dB multiplied by 2.

References

  • 1.

    WilliamsALDrongelenWVan LaskyRE. Noise in contemporary neonatal intensive care. J Acoust Soc Am. (2007) 2681:268190. 10.1121/1.2717500

  • 2.

    LaskyREWilliamsAL. Noise and light exposures for extremely low birth weight newborns during their stay in the neonatal intensive care unit. Pediatrics. (2009) 123:5406. 10.1542/peds.2007-3418

  • 3.

    SantosJSantosJCarvalhaisCXavierASilvaMVSantosJet al. Assessment and characterization of sound pressure levels in Portuguese neonatal intensive care units. Arch Environ Occup Heal ISSN. (2018) 73:1217. 10.1080/19338244.2017.1304883

  • 4.

    CardosoSMSKozlowskiLdeCde LacerdaABMMarquesJMRibasA. Newborn physiological responses to noise in the neonatal unit. Braz J Otorhinolaryngol. (2015) 81:5838. 10.1016/j.bjorl.2014.11.008

  • 5.

    KuhnPZoresCPebayleTHoeftALangletCEscandeBet al. Infants born very preterm react to variations of the acoustic environment in their incubator from a minimum signal-to-noise ratio threshold of 5 to 10 dBA. Pediatr Res. (2012) 71:38692. 10.1038/pr.2011.76

  • 6.

    WharradHJDavisAC. Behavioural and autonomic responses to sound in pre-term and full-term babies. Br J Audiol. (1997) 31:31529. 10.3109/03005364000000026

  • 7.

    VranekovicGHockEIsaacPCorderoL. Heart rate variability and cardiac response to an auditory stimulus. Biol Neonate. (1974) 24:6673. 10.1159/000240633

  • 8.

    StanleyNGravenM. Sound and the developing infant in the NICU: conclusions and recommendations for care. J Perinatol. (2000) 20:8893. 10.1038/sj.jp.7200444

  • 9.

    American Academy of Pediatrics. Committee on Environmental Health. Noise: a hazard for the fetus and newborn. Pediatrics. (1997) 100:7247. 10.1542/peds.100.4.724

  • 10.

    Environmental Protection Agency. Office of Noise Abatement and Control. Information on Levels of Environmental Noise Requisite to Protect Public Health and Welfare With an Adequate Margin of Safety. Washington, DC: Government Printing Office (1974).

  • 11.

    KnutsonAJ. Acceptable noise levels for neonates in the neonatal intensive care unit. In: Independent Studies and Capstone Projects Program in Audiology and Communication Sciences. Washington, DC: University School of Medicine (2012).

  • 12.

    DarcyAHancockLWareE. A descriptive study of noise in the neonatal intensive care unit ambient levels and perceptions of contributing factors. Adv Neonatal Care. (2008) 8:16575. 10.1097/01.ANC.0000324341.24841.6e

  • 13.

    AggarwaLDChawlaSRaoDBasiricoJ. Turn down the volume: a study of excessive sound levels in the neonatal intensive care unit. Pediatrics. (2019) 144:693. Available online at: https://pediatrics.aappublications.org/content/144/2_MeetingAbstract/693

  • 14.

    SmithSWOrtmannAJClarkWW. Noise in the neonatal intensive care unit: a new approach to examining acoustic events. Noise Heal. (2018) 95:12130. Available online at: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6122266/

  • 15.

    CarvalhaisCSantosJVieiraMXavierA. Is there sufficient training of health care staff on noise reduction in neonatal intensive care units? A Pilot Study From Neonoise. J Toxicol Environ Heal Part A Curr Issues. (2015) 78:897903. 10.1080/15287394.2015.1051204

  • 16.

    LiuWF. The impact of a noise reduction quality improvement project upon sound levels in the open-unit-design neonatal intensive care unit. J Perinatol. (2010) 30:48996. 10.1038/jp.2009.188

  • 17.

    WangDAubertinCBarrowmanNMoreauKDunnSHarroldJ. Reduction of noise in the neonatal intensive care unit using sound-activated noise meters. Arch Dis Child Fetal Neonatal Ed. (2014) 99:F5156. 10.1136/archdischild-2014-306490

  • 18.

    CostonADAuneC. Reducing noise in the neonatal intensive care unit. Pediatrics. (2019) 144:154. Available online at: https://pediatrics.aappublications.org/content/144/2_MeetingAbstract/154

  • 19.

    SpenceJDecasperJ. Prenatal experience with sounds influence neonatal perception of maternal voice samples. Inf Behav Dev. (1987) 10:13342. 10.1016/0163-6383(87)90028-2

  • 20.

    AbramsRMGerhardtKJ. The acoustic environemnt and physilogical responses of the fetus. J Perinatol. (2000) 20:305. 10.1038/sj.jp.7200445

  • 21.

    ZahrLKBalianS. Responses of premature infants to routine nursing interventions and noise in the NICU. Nurs Res. (1995) 44:17985. 10.1097/00006199-199505000-00009

  • 22.

    PengNBachmanJJenkinsRChenCChangY. Relationships between environmental stressors and stress biobehavioral responses of preterm infants in NICU. Adv Neonatal Care. (2013) 13:210. 10.1097/ANC.0000000000000023

  • 23.

    FieldTMDempseyJRHatchJTingGCliftonRK. Cardiac and behavioral responses to repeated tactile and auditory stimulation by preterm and term neonates. Dev Psychol. (1979) 15:40616. 10.1037/0012-1649.15.4.406

  • 24.

    HassaneinSMAEl RaggalNMShalabyAA. Neonatal nursery noise: practice-based learning and improvement. J Matern Neonatal Med. (2013) 26:3925. 10.3109/14767058.2012.733759

  • 25.

    VohrBR. Screening the Newborn for hearing loss. Wolter Kluwer; UpToDate (2020)

  • 26.

    LahavASkoeE. An acoustic gap between the NICU and womb: a potential risk for compromised neuroplasticity of the auditory system in preterm infants. Front Neurosci. (2014) 8:381. 10.3389/fnins.2014.00381

  • 27.

    LejeuneFParraJBerne-audéoudFMarcusL. Sound interferes with the early tactile manual abilities of preterm infants. Nat Publ Gr. (2016) 6:18. 10.1038/srep23329

  • 28.

    LaiTTBearerCF. Iatrogenic environmental hazards in the neonatal intensive care unit. Clin Perinatol. (2008) 35:16381. 10.1016/j.clp.2007.11.003

  • 29.

    LongJGLuceyJFPhilipAGS. Noise and hypoxemia in the intensive care nursery. Pediatrics. (1980) 65:1435. 10.1542/peds.65.2.203

  • 30.

    WachmanEMLahavA. The effects of noise on preterm infants in the NICU. Arch Dis Child Fetal Neonatal Ed. (2011) 96:F3059. 10.1136/adc.2009.182014

  • 31.

    WilliamsALSandersonMPhDLaiDPhD. Intensive care noise and mean arterial blood pressure in extremely low-birth-weight neonates. Am J Perinatol. (2008). 26:3239. 10.1055/s-0028-1104741

  • 32.

    GrayLPhilbinMK. Measuring sound in Hospital nurseries. J Perinatol. (2001) 20:99103. 10.1038/sj.jp.7200440

  • 33.

    Hernandez MolinaRFernández ZacaríasFPuyana RomeroVRodríguez MontañoVMBeira JiménezJLCueto AncelaJLet al. Análisis del ambiente sonoro e una unidad de cuidados intersivos de neonatología. In: XI Congreso Iberoamericano de Acústica; X Congreso Ibérico de Acúsica; 49oCongreso Español de Acústica -TECNIACUSTICA'18. Cádiz (2018).

  • 34.

    GarethPJLukashkinaVARussellIJLukashkin1AN. The vestibular system mediates sensation of low-frequency sounds in mice. J Assoc Res Otolaryngol. (2010) 732:72532. 10.1007/s10162-010-0230-7

  • 35.

    OhgamiNOshinoRNinomiyaHLiXKatoM. Risk assessment of neonatal exposure to low frequency noise based on balance in mice. Front Behav Neurosci. (2017) 11:17. 10.3389/fnbeh.2017.00030

  • 36.

    Fernández ZacaríasFBeira JiménezJLBustillo Velázquez-GazteluPJHernández MolinaRLubián LópezS. Noise level in neonatal incubators: A comparative study of three models. Int J Pediatr Otorhinolaryngol. (2018) 107:1504. 10.1016/j.ijporl.2018.02.013

  • 37.

    International Organization for Standardization. Insulation, ISO 16283-1:2014. Acoustics — Field Measurement of Sound Insulation in Buildings and of Building Elements — Part 1: Airborne Sound. (2014) Available online at: https://www.iso.org/standard/55997.html (accessed September 07, 2020).

  • 38.

    Núñez-SolanoDPuyana-RomeroVOrdoñez-AndradeCBravo-MonayoLGarzón-PicoC. Impulse response simulation of a small room and in situ measurements validation. In: 147th Audio Engineering Society Convention. New York, NY (2019). p. 17.

  • 39.

    Puyana-RomeroVNúñez-SolanoDHernándezRFernández-ZacaríasFBeira-JiménezJLGarzónCet al. Reverberation time measurements of a neonatal incubator. Appl Acoust. (2020) 167:107374. 10.1016/j.apacoust.2020.107374

  • 40.

    KleinerMTichyJ. Acoustics of Small Rooms (CRC Press. Taylor & Francis Group) (2014). p. 1453.

  • 41.

    BerzbornMBomhardtRKleinJRichterJ-GVorländerM. The ITA-Toolbox: an open source MATLAB toolbox for acoustic measurements and signal processing. In: 43th Annu Ger Congr Acoust (2017). p. 2225. Available online at: http://www.ita-toolbox.org/publications/ITA-Toolbox_paper2017.pdf (accessed September 07, 2020).

  • 42.

    NacharN. The Mann - Whitney U: a test for assessing whether two independent samples come from the same distribution. Tutor Quant Methods Psychol. (2008) 4:1320. 10.20982/tqmp.04.1.p013

  • 43.

    PinheiroEMGuinsburgRNabuco MA deAKakehashiTY. Noise at the neonatal intensive care unit and inside the interior of the incubator (in spanish). Rev Lat Am Enfermagem. (2011) 19:121421. 10.1590/S0104-11692011000500020

  • 44.

    RodarteMDOScochiCGSLeiteAMFujinagaCIZamberlanNECastralTC. O ruído gerado durante a manipulação das incubadoras: implicações para o cuidado de enfermagem. Rev Lat Am Enfermagem. (2005) 13:7985. 10.1590/S0104-11692005000100013

  • 45.

    MašoviDMemberSMeteÖ. Analysis of reverberation time field measurement results in building acoustics. Telfor J. (2013) 5:14550. Available online at: http://journal.telfor.rs/Published/Vol5No2/Vol5No2.aspx (accessed September 07, 2020).

  • 46.

    TamuraHOhgamiNYajimaIIidaMOhgamiKFujiiN. Chronic exposure to low frequency noise at moderate levels causes impaired balance in mice. PLoS ONE. (2012) 7:17. 10.1371/journal.pone.0039807

Summary

Keywords

acoustic environment, low frequency noise, acoustic isolation, neonatal incubator, reverberation time

Citation

Puyana-Romero V, Núñez-Solano D, Hernández-Molina R and Jara-Muñoz E (2020) Influence of the NICU on the Acoustic Isolation of a Neonatal Incubator. Front. Pediatr. 8:588. doi: 10.3389/fped.2020.00588

Received

24 April 2020

Accepted

10 August 2020

Published

22 September 2020

Volume

8 - 2020

Edited by

Kyle Jason Rehder, Duke University, United States

Reviewed by

Ameer Ahmad, Quaid-i-Azam Medical College, Bahawalpur, Pakistan; Allison Mari Dering-Anderson, University of Nebraska Medical Center, United States

Updates

Copyright

*Correspondence: Virginia Puyana-Romero

This article was submitted to Children and Health, a section of the journal Frontiers in Pediatrics

Disclaimer

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.

Outline

Figures

Cite article

Copy to clipboard


Export citation file


Share article

Article metrics