Continuous renal replacement therapy (CRRT) has become a cornerstone in the management of critically ill children with severe acute kidney injury (AKI) and fluid overload. However, its use in neonates and infants weighing less than 10 kg remains particularly challenging due to limited circulating volumes, higher risks of hemodynamic instability, and the scarcity of devices specifically designed for this population. In recent years, new platforms have been introduced, enabling safer extracorporeal support for the smallest patients. Although early experiences suggest that these technologies are feasible and safe, survival outcomes in this fragile group remain suboptimal, and standardized guidelines are still lacking. To address this gap, we propose the AEIOU strategy, an educational and practical framework aimed at organizing and simplifying the essential components required for safe and effective CRRT in infants under 10 kg. AEIOU summarizes: A (Alerts: indications, risks, and contraindications), E (Execution Team), I (Inputs: catheters, machines, filters, and solutions), O (Orders: prescription components such as priming, anticoagulation, blood and effluent flow rates), and U (Unified record: standardized documentation of therapy variables and outcomes). By integrating existing literature with a reproducible mnemonic, AEIOU provides a structured tool for training clinical teams, ensuring quality of care, and laying the foundation for collaborative research. Beyond CRRT, this framework may be adaptable to other extracorporeal therapies in pediatrics. The AEIOU strategy offers a simple yet comprehensive guide to improve implementation, education, and outcome monitoring of CRRT in neonates and small infants.
1 Introduction
Continuous renal replacement therapy (CRRT) has become a life-saving intervention for critically ill children with acute kidney injury (AKI) and fluid overload (1, 2). While the use of CRRT is expanding worldwide, delivering these therapies in neonates and infants weighing less than 10 kg remains particularly complex due to small extracorporeal volumes, hemodynamic instability, and the limited availability of dedicated devices (3). The safe and effective use of CRRT therefore requires highly trained multidisciplinary teams and clear educational strategies to standardize practice.
To address this need, we propose the AEIOU strategy, an educational and practical framework that organizes the essential aspects of CRRT into a simple, reproducible acronym. AEIOU highlights: A (Alerts: indications, risks, and contraindications), E (Execution Team), I (Inputs: catheters, machines, filters, and solutions), O (Orders: prescription and technical parameters), and U (Unified record: standardized documentation of performance and outcomes) (Figure 1). This structured approach is designed to simplify teaching, reduce errors, and improve adherence to best practices, particularly in pediatric intensive care units with limited experience.
In addition, contemporary data from the WE-ROCK cohort reinforce the importance of adopting a structured framework for infants <10 kg receiving CRRT (4). In this study, centers demonstrated substantial practice variability, including more than a tenfold difference in prescribed dosing. This combination of physiologic vulnerability, technical complexity, and wide inter-center variation underscores the need for structured, reproducible frameworks such as AEIOU—tools that can reduce unwarranted variability and promote safer, more consistent care in this high-risk population. The objective of this review is to summarize the current evidence on CRRT in children under 10 kg and to introduce the AEIOU strategy as a practical and educational framework to enhance safety, standardization, and implementation.
2 Special considerations for CRRT in infants <10 kg
Acute kidney injury (AKI) affects up to 30% of critically ill neonates admitted to pediatric intensive care units (PICUs) and neonatal intensive care units (NICUs) and is strongly associated with adverse outcomes, particularly when classified as severe AKI (KDIGO stage 2–3) (1, 2). One of the greatest challenges in managing these patients is the provision of renal replacement therapy. Standard hemodialysis or CRRT in this population is complicated by the small circulating blood volume relative to extracorporeal circuit volume, the risks of anticoagulation and bleeding, susceptibility to hypothermia, and the potential for errors in solute and fluid clearance that may be tolerated in adults but carry disproportionate consequences in small infants (3).
Historically, available CRRT platforms were designed for adults, with limited modifications to allow their use in children >10 kg. Over the last two decades, however, devices specifically engineered for infants <10 kg have emerged, including the Aquadex™ FlexFlow System (Nuwellis, USA), CARPEDIEM™ (Cardio-Renal Pediatric Dialysis Emergency Machine), and NIDUS™ (Newcastle Infant Dialysis and Ultrafiltration System). Of these, Aquadex™ and CARPEDIEM™ obtained FDA approval in 2020. These machines incorporate smaller filters and highly precise control of blood (Qb) and ultrafiltration (QUF) flows, thereby reducing hemodynamic instability and technical complications (4).
Early clinical experiences and case series in neonates and infants <10 kg consistently report feasibility and safety, though most studies are limited by small sample sizes (5–14). The central unanswered question remains whether these innovations will translate into improved survival in this highly vulnerable group.
Table 1 summarizes the patient characteristics, technical aspects, and prescription parameters of CRRT therapies reported in studies on the use of new technologies in infants weighing less than 10 kg.
Table 1. Summary of patient characteristics, technical aspects, and prescription parameters of CRRT in infants <10 kg.
Recent multicenter efforts, such as the ICONIC consortium, represent a critical step forward. By systematically collecting clinical and outcome data in infants treated with CARPEDIEM™, ICONIC seeks to define best practices and inform standardized treatment protocols (12). Nevertheless, there are currently no universally accepted guidelines or consensus statements for CRRT in infants <10 kg.
Collaborative registries, such as WE-ROCK (Worldwide Exploration of Renal Replacement Outcomes Collaborative in Kidney Disease), have reported modest improvements in survival among critically ill children weighing less than 10 kg who require CRRT, compared with historical cohorts. However, mortality rates and major adverse kidney events remain unacceptably high (14, 15). In this context, structured frameworks such as the AEIOU strategy may provide clinicians with a practical tool to organize key considerations and promote safe, consistent CRRT delivery, while the pediatric nephrology and critical care communities continue to build the evidence base needed to inform future guidelines.
2.1 A: alerts
2.1.1 Objectives
The primary objective of this phase is to ensure the timely recognition of clinical scenarios warranting CRRT in infants weighing <10 kg, including severe AKI, diuretic-resistant fluid overload, life-threatening electrolyte or acid–base disturbances, metabolic diseases, and selected intoxications, while simultaneously identifying the therapy's inherent risks and implementing appropriate safety precautions.
2.1.2 Indications
Technologies designed to deliver CRRT in infants weighing less than 10 kg have been employed in critically ill neonates and infants admitted to the PICU for both renal and non-renal indications. Renal indications include severe AKI with refractory uremia, hyperkalemia, or metabolic acidosis unresponsive to conventional therapy. Non-renal indications encompass fluid overload (FO), hyperammonemia, and intoxications with dialyzable substances, all of which represent the most frequent reasons for initiating CRRT in this population.
Among these, FO is the leading indication for CRRT initiation in infants <10 kg (Table 1). A recent secondary analysis from the WE-ROCK cohort included 210 patients under 10 kg who received CRRT for AKI or FO across 32 centers in seven countries (14). In this subgroup, the median FO at CRRT initiation was 16.4% (IQR 6.3–32.8). By comparison, when analyzing the entire cohort of patients <25 years, the median FO was 7.4% (IQR 2.4–18.1) (15), a value lower than previously reported in the literature (16). This observation suggests a shift in clinical reasoning toward initiating CRRT at earlier stages of FO, reflecting the well-recognized negative impact of FO on survival and renal recovery in critically ill children (17, 18).
2.1.3 Risks
Complications associated with CRRT in infants <10 kg vary across studies but commonly include hypotension (7%–23%), thrombocytopenia (22%), and circuit clotting (13%–19%) (Table 1), and despite technological advances, survival remains modest: the ppCRRT registry reported 56% mortality at PICU discharge among 84 patients <10 kg (16), while the more recent WE-ROCK cohort showed a survival of 46.1% (14). Beyond these outcomes, infants <10 kg exhibit a distinct physiologic vulnerability that heightens procedural risk. Vascular access complications are frequent due to catheter–vessel caliber mismatch; in the London cohort, insertion-site bleeding occurred in 6.7% of sessions and hypotension requiring intervention in 11.9% (3).
Metabolic instability—including rapid electrolyte shifts, acid–base disturbances, and hypothermia—arises from blood priming requirements and the disproportionate extracorporeal volume relative to circulating blood volume. In the Aquadex™ infant series, transient hypothermia, puncture-site bleeding, systemic bleeding, and occasional atrial thrombosis were noted, underscoring the hemostatic fragility of infants <5 kg (10). Collectively, these data highlight that infants <10 kg face converging hemodynamic, metabolic, and access-related risks that are intrinsic to critical illness and magnified by the technical challenges of extracorporeal support, reinforcing the need for structured frameworks that can more effectively “alert” clinicians to early risk signals and optimize timing of CRRT initiation.
2.1.4 Contraindications
The success of CRRT in infants critically depends on secure vascular access, and the inability to obtain adequate access due to severe bleeding, vascular thrombosis, infection, or extensive skin injury remains the only absolute contraindication to initiating therapy (20–22). In all other situations, contraindications are relative and context-dependent, including active bleeding with high transfusion requirements, profound hemodynamic instability not responsive to vasoactive support, or institutional limitations such as lack of neonatal-specific technology or insufficiently trained personnel, all of which may compromise safety. Importantly, most conditions traditionally viewed as barriers should not be interpreted as absolute exclusions (23); instead, they demand a careful multidisciplinary risk–benefit assessment, particularly in infants <10 kg where therapeutic margins are exceptionally narrow. This nuanced approach underscores the need for specialized training, structured frameworks such as the AEIOU strategy, and regional collaboration to expand access to CRRT while minimizing complications.
2.2 E: execution team
2.2.1 Objectives
The safe and effective delivery of CRRT in infants <10 kg requires a multidisciplinary team supported by appropriate technological and institutional resources. Beyond physicians and specialized nursing staff, essential contributors include pharmacy services, clinical laboratories, and biomedical engineering, all of whom ensure therapy safety and continuity (4). Updated institutional documents such as policies, standard operating procedures, and evidence-based clinical practice guidelines are equally critical to standardize care, minimize variability, and promote adherence to best practices. Together, these human, technological, and documentary resources form the backbone of a high-quality CRRT program in the PICU.
2.2.2 Program leadership and standardization
The successful implementation of a CRRT program for infants under 10 kg requires the designation of a dedicated program leader responsible for coordinating the multidisciplinary team, developing and updating standardized protocols, and ensuring uniform adherence to evidence-based practices. This leader—typically a pediatric intensivist or nephrologist with expertise in extracorporeal therapies—should oversee team training, protocol audits, and outcome monitoring, while fostering communication and collaboration among all participating members, including nursing, pharmacy, and biomedical engineering. The presence of a defined leadership structure promotes consistency across shifts and providers, minimizes variability in practice, and strengthens the overall quality and safety of CRRT delivery.
2.2.3 Medical and nursing teams
The delivery of CRRT in infants <10 kg requires close collaboration between pediatric intensivists, neonatologists, and pediatric nephrologists. While intensivists often lead the management in the PICU, the participation of nephrologists adds expertise in prescription and troubleshooting. In patients with difficult vascular anatomy or in very small infants, interventional radiologists or pediatric surgeons may be required to secure central venous access. Specialized nursing staff are equally essential, responsible for priming, assembling, and monitoring the circuit, as well as ensuring consistent delivery of the therapy. Ideally, nurse-to-patient ratios should be at least 1:1, although surveys indicate that this standard is not always met.
The multinational ESPNIC survey highlighted the variability of CRRT practice in Europe: intensivists were the primary prescribers in 70% of centers, nephrologists in 12%, and only 4% reported joint responsibility (20). Nursing teams in most centers were primarily in charge of machine operation and circuit management. Importantly, the survey also revealed that in the majority of cases (56%) the nurse-to-patient ratio was 1:1, while 24% reported 2:1 and only 18% of providers cared for more than one patient simultaneously. These findings underscore the heterogeneity in workforce allocation and the challenges in maintaining optimal staffing in high-demand settings.
Given that CRRT in infants <10 kg is still infrequent in many PICUs, ongoing training is crucial to maintain competence. Structured educational programs should combine theoretical modules, bedside supervision, simulation-based training, and access to digital resources. Although no universal certification standards currently exist, periodic education and skills reinforcement are essential. Notably, in the ESPNIC survey, only 61% of nurses reported having received formal training, and nearly one-third of PICUs lacked requirements for certification or recertification. These findings highlight a major gap and reinforce the need for standardized curricula and continuous professional development.
2.2.4 Competencies and safety processes
To operate CRRT safely in infants <10 kg, teams must demonstrate core competencies that include mastery of circuit priming with minimal extracorporeal volume exposure, troubleshooting of pressure alarms and flow instability, early recognition of access dysfunction, precise titration of anticoagulation strategies, and rapid interpretation of laboratory trends that signal metabolic or hemodynamic compromise. These skills must be supported by structured processes such as pre-CRRT safety checklists (confirming catheter position, machine calibration, anticoagulation plan, and fluid targets), standardized bedside monitoring bundles, and post-connection stabilization protocols. Embedding these competencies within routine workflows strengthens reliability and reduces preventable errors—an essential goal in a population with extremely narrow safety margins.
2.2.5 Pharmacy team
Pharmacists play a vital role in ensuring patient safety during CRRT. They provide guidance on dose adjustments for medications impacted by extracorporeal clearance based on estimated CRRT dose provided and drug carachteristics. In some centers, pharmacists are also involved in preparing dialysis fluids or anticoagulation solutions, which is particularly relevant in small infants where fluid composition must be precisely tailored. This collaboration minimizes medication errors and ensures individualized therapy, which is particularly relevant given that most centers are currently using commercially based CRRT solutions.
2.2.6 Clinical laboratory team
Laboratory support is essential to guide real-time adjustments during CRRT. Monitoring acid–base balance, electrolyte homeostasis, and coagulation parameters informs critical prescription changes and allows rapid response to complications. Continuous availability of laboratory services during therapy is therefore indispensable to maintain patient safety and to optimize treatment efficacy.
2.3 I: inputs
2.3.1 Objective
To ensure the safe and effective delivery of CRRT in infants <10 kg by selecting and optimizing technological resources (vascular catheters, machines, filters, extracorporeal circuits, and solutions) that minimize extracorporeal volume, reduce complications, and adapt therapy to the physiological needs of this vulnerable population.
2.3.2 Catheters
Except for NIDUS™, which operates with a single-lumen central venous line, all other neonatal CRRT systems require a double-lumen catheter to maintain adequate blood flow. Conventional devices often require catheters ≥7 Fr, while CARPEDIEM™ can be used with 4 Fr catheters, an important advantage in neonates. Reports of using two single-lumen catheters exist, but very small devices carry risks of thrombosis, poor flow, and malfunction (8, 9). Dedicated dialysis catheters are designed with sufficient rigidity to resist pump pressures and with lengths that optimize tip positioning.
The right internal jugular vein has become the preferred access due to favorable anatomy and reduced recirculation (21). In WE-ROCK, this site was used in 66% of cases, compared with 16% in ppCRRT; femoral access decreased from 73% in ppCRRT to 30% in WE-ROCK, while subclavian use remained rare (<10%) (15, 22).
2.3.3 Machines
Neonatal CRRT platforms have been specifically engineered to address the unique physiological vulnerabilities of infants <10 kg by minimizing extracorporeal circuit volume, enhancing ultrafiltration precision, and permitting exceptionally low blood-flow rates. Among these devices, CARPEDIEM™—FDA-approved for infants 2.5–10 kg (5–9)—and the HF-20 hemofilter for Prismaflex™, FDA-approved for children 8–20 kg (4), represent the only platforms with regulatory authorization for use in this population; both also hold CE Mark approval in Europe, supporting their broader international adoption. Evidence from retrospective comparisons shows that infants <5 kg treated with CARPEDIEM™ had higher survival during therapy than historical ppCRRT controls treated with standard machines, which required substantially higher vasoactive support at initiation—highlighting the hemodynamic burden of non-neonatal platforms (6–12, 19).
In contrast, Aquadex™, although FDA-approved only for patients >20 kg, has been used off-label in selected infants because of its low extracorporeal volume (33 mL) and flexible blood-flow range (5–40 mL/min), enabling treatment of children as small as 3 kg (10, 11). The NIDUS™ system, employing a syringe-driven circuit with extracorporeal volumes as low as 17 mL, is CE-approved for neonatal use in Europe and has demonstrated feasibility in neonates as small as 800 g (12). While Aquadex™ and NIDUS™ allow hemofiltration and ultrafiltration, CARPEDIEM™ is the only miniaturized platform that offers hemodialysis capability (Table 2).
2.3.4 Dialysis or replacement fluid
Most solutions approximate physiologic electrolyte concentrations, with variable potassium, calcium, and bicarbonate content. Bicarbonate-buffered solutions are preferred over lactate-based fluids for greater hemodynamic stability and improved metabolic control. When regional citrate anticoagulation is used, calcium-free solutions with lower bicarbonate concentrations help prevent metabolic alkalosis resulting from citrate metabolism. The CARPEDIEM™ system, however, offers only a single solution containing calcium and a high bicarbonate concentration.
2.4 O: orders
2.4.1 Objective
To define the essential components of CRRT prescription in infants <10 kg, ensuring individualized, safe, and effective therapy through precise programming of extracorporeal circuits, anticoagulation strategies, and machine parameters.
2.4.2 Circuit priming
Circuit priming can be performed with red blood cells, crystalloids, or colloids such as albumin. Red blood cell priming may be indicated when the extracorporeal volume (ECV) exceeds 10% of the patient's blood volume, or in cases of anemia or hemodynamic instability. However, this approach carries risks including electrolyte disturbances (hypocalcemia, hyperkalemia), acid–base disorders, thrombocytopenia, coagulopathy, and vasoplegia. A major advantage of neonatal-specific CRRT machines is their ability to maintain ECV below 10% in most cases, thereby minimizing the need for blood priming.
The following formula is used to determine the ECV:
*In neonates, the value should be multiplied by 100 (mL/kg).
2.4.3 Anticoagulation
Effective anticoagulation is essential to prevent clotting of the extracorporeal circuit, as inadequate anticoagulation reduces therapy efficacy, increases blood loss, and increases nursing workload (24). Systemic heparin remains widely used, typically administered pre-filter at 10–20 IU/kg/h with or without an initial bolus, targeting a partial thromboplastin time 1.5–2× baseline while monitoring platelets given the 1%–5% risk of heparin-induced thrombocytopenia (25). However, systemic anticoagulation carries a substantial bleeding risk (10%–50%), particularly after surgery, trauma, mucosal injury, or in patients with underlying coagulopathy (26–28).
Regional citrate anticoagulation (RCA) has become the preferred modality in many neonatal and infant programs because its anticoagulant effect is confined to the circuit. Citrate chelates calcium pre-filter to inhibit thrombin formation, while the calcium–citrate complex is removed by dialysis or metabolized to bicarbonate; systemic calcium is replaced post-filter. RCA reduces bleeding complications compared with systemic heparin (26–28). The most common solutions include ACD-A—typically infused at 1.5× blood flow to maintain in-filter ionized calcium <0.4 mmol/L—and Prismocitrate, titrated based on ionized calcium levels.
Data from the WE-ROCK cohort show that RCA was the most frequently used anticoagulation strategy in infants <10 kg (52%), followed by heparin (31%), no anticoagulation (10%), and other agents (7%) (13). By contrast, early CARPEDIEM™ studies reported more frequent heparin use, likely reflecting limited initial experience with citrate in these newer platforms (Table 1). In contemporary practice, centers such as Cincinnati Children's Hospital now use RCA in approximately 50% of CARPEDIEM™ circuits (this observation is based on personal communication with the program leadership).
Beyond heparin and citrate, prostacyclin (epoprostenol) and nafamostat serve as alternatives in selected high-risk infants. Prostacyclin provides circuit-directed anticoagulation and has demonstrated mean circuit lives of ∼50 h in acute liver failure without major bleeding, supporting its use when heparin is contraindicated (29, 30). Nafamostat, a short-acting serine-protease inhibitor, offers potent regional anticoagulation with minimal systemic effects; in a two-center cohort (158 patients, 455 filters), it achieved circuit life comparable to RCA (median 38 vs. 36 h) with similar bleeding risk and without citrate-related metabolic complications, including in patients with liver dysfunction (31). Experience with these agents in infants <10 kg remains limited to specialized centers, and their use should be individualized according to bleeding risk, local expertise, and drug availability.
2.4.4 Blood flow rates (Qb)
Unlike older children, in whom Qb during CRRT are typically set around 5 mL/kg/min, infants weighing <10 kg often require higher flows (10 mL/kg/min). This adjustment is dictated by vascular access size and the characteristics of the device employed. When using machines originally designed for adults, minimum operational flows are usually ≥60 mL/min (with HF20 minimal blood flow rate is 20 mL/min), which translates into 10–15 mL/kg/min in a small infant. In contrast, neonatal-specific technologies can operate safely at Qb as low as 20 mL/min, providing greater flexibility and hemodynamic tolerance in this population. Data from the WE-ROCK cohort confirm this practice, reporting a median Qb of 8 mL/kg/min (IQR 5.9–11.3) among infants <10 kg (14). Table 1 summarizes Qb parameters reported in studies of children treated with new-generation devices.
2.4.5 Dose
KDIGO guidelines recommend CRRT delivered doses of 20–25 mL/kg/h in adults, although higher prescriptions (30–35 mL/kg/h) are often used to offset therapy interruptions (21). In pediatrics, no large-scale trials define the optimal dose, and practice remains heterogeneous.
In the WE-ROCK cohort, the mean prescribed dose across the entire population was 41.9 mL/kg/h (IQR 30.9–59.9), with 55% receiving >40 mL/kg/h. In infants <10 kg, the mean dose was substantially higher: 63.8 mL/kg/h (IQR 49.2–88.1), with 85% receiving >40 mL/kg/h. Marked inter-center variability was reported, with prescriptions ranging from 27.8 to 187.8 mL/kg/h (14).
2.4.6 Modality
The choice of modality depends on the device and clinical context. New pediatric-specific platforms (CARPEDIEM™, Aquadex™, NIDUS™) allow for continuous venovenous hemofiltration (CVVH), hemodialysis (CVVHD), or hemodiafiltration (CVVHDF), with flexibility to individualize therapy. International surveys of European PICUs demonstrated CVVHDF as the most frequent modality (51%) (20). Similarly, in WE-ROCK, CVVHDF predominated in both the overall cohort (76%) and in infants <10 kg (69%) (14, 15). This preference reflects the ability of CVVHDF to optimize solute clearance while maintaining fluid control, particularly relevant in critically ill neonates with multiorgan dysfunction.
2.4.7 Ultrafiltration flow (QUF)
QUF represents the total ultrafiltrate removed per unit of time, comprising net ultrafiltration (UFNET) and, depending on modality, replacement or dialysate flows (29). Fluid removal is guided by clinical status, fluid balance, and degree of fluid overload. In general, net removal >5% of body weight is not recommended, and with CARPEDIEM™, hourly ultrafiltration should remain <20% of blood flow (32). Modern neonatal platforms incorporate highly precise weighing systems, capable of detecting errors as low as 1 g/h, a critical safeguard in this population where even small discrepancies can have deleterious hemodynamic or metabolic consequences. Additionally, for CARPEDIEM™, manufacturer specifications list infusion rates of 0–150, 0–600, and 0–900 mL/h for the dialysate, replacement, and ultrafiltration pumps, respectively. These parameters indicate pump capabilities rather than prescribed net ultrafiltration limits.
In summary, CRRT dosing and prescription in infants <10 kg remain an evolving field. Observational data suggest higher prescribed doses and a predominance of CVVHDF, with RCA as the preferred anticoagulation strategy. While newer devices mitigate technical limitations by reducing extracorporeal volume and enhancing ultrafiltration precision, practice variation persists. Standardization through multicenter collaboration and prospective studies will be essential to define optimal dosing strategies and improve outcomes in this fragile population (33, 34).
2.5 U: unified record
2.5.1 Objective
To establish a standardized, accessible, and integrated documentation system that improves safety, optimizes clinical decision-making, and facilitates the generation of high-quality data for research and quality improvement in pediatric CRRT.
The development of a CRRT program in infants <10 kg must extend beyond protocols and guidelines to include the systematic implementation of a unified record. This record should consolidate all aspects of prescription, delivery, monitoring, and outcomes into a practical tool, ideally embedded within the electronic medical record. Its structured completion functions as a safety checkpoint before, during, and after therapy, ensuring verification of medical orders, early identification of complications, and timely adjustment of prescriptions. Moreover, the unified record generates reliable longitudinal data that form the foundation for clinical audits, benchmarking across institutions, and the design of multicenter research initiatives. By standardizing documentation, programs not only enhance patient safety and quality of care but also strengthen the evidence base for CRRT in the smallest and most vulnerable patients (Table 3).
3 Implementation and application of the AEIOU framework
Successful adoption of the AEIOU strategy in centers caring for infants <10 kg requires deliberate integration of the framework into institutional workflows, guided by principles of implementation science. First, programs should engage key stakeholders—including intensivists, nephrologists, nursing leadership, pharmacy, and biomedical engineering—to ensure shared understanding of the framework's purpose and applicability. Readiness assessments can help identify local barriers, such as limited experience with neonatal CRRT, gaps in training, or absence of standardized protocols.
Second, centers may embed AEIOU into clinical practice by developing unit-specific checklists, flow diagrams, and order templates that reflect each domain of the framework. Simulation-based training and structured onboarding modules allow teams to apply AEIOU in controlled environments, reinforcing competencies and increasing reliability. Incorporating AEIOU elements into electronic health records—such as unified CRRT order sets or automated documentation prompts—can further enhance adherence and reduce unwarranted practice variation.
Third, programs should establish audit and feedback cycles, using metrics such as time-to-initiation, catheter-related complications, circuit lifespan, anticoagulation errors, and adherence to AEIOU components as part of continuous quality improvement. These data also support benchmarking and multicenter collaboration, enabling centers to track performance over time and identify opportunities for refinement.
4 Limitations
Although the AEIOU strategy provides a structured and pragmatic framework to organize the key elements of CRRT delivery in infants under 10 kg, several limitations must be acknowledged. First, AEIOU is not derived from prospective clinical validation but rather from expert synthesis of available evidence and practice-based experience; therefore, its applicability may vary across institutions with different resources and case volumes. Second, while the mnemonic simplifies complex processes into five domains, it may risk oversimplification, potentially overlooking patient-specific nuances such as comorbidities, pharmacologic interactions, or unique technical challenges. Third, the lack of universally standardized guidelines for CRRT in this population means that AEIOU should be regarded as a complementary educational and organizational tool, rather than a substitute for formal training or individualized clinical judgment. Finally, the effectiveness of AEIOU in improving patient outcomes remains to be demonstrated through multicenter implementation studies and integration into quality improvement initiatives.
5 Opportunities and future directions in CRRT for infants <10 kg
The AEIOU framework offers multiple opportunities to advance the quality, safety, and standardization of CRRT delivery in infants <10 kg. A key future direction is the prospective validation of the strategy across diverse centers, examining its impact on time-to-initiation, circuit longevity, anticoagulation safety, and clinically meaningful outcomes. Incorporating AEIOU into simulation curricula, structured onboarding programs, and electronic health record pathways could enhance adherence, reduce practice variability, and support consistent decision-making in high-risk infants. The framework also provides a foundation for multicenter quality improvement collaboratives, enabling benchmarking of process metrics and facilitating shared learning. As neonatal CRRT technologies continue to evolve, AEIOU can serve as a dynamic, adaptable scaffold—integrating new devices, anticoagulation approaches, and monitoring capabilities while maintaining a unified logic for bedside application. Ultimately, the future of CRRT in infants <10 kg will depend not only on technological innovation, but on the systematic implementation, refinement, and evaluation of structured strategies such as AEIOU that translate evidence into reliable, reproducible practice.
6 Conclusion
Continuous renal replacement therapy in infants weighing less than 10 kg remains one of the most demanding areas of pediatric critical care, where technological advances alone are insufficient to guarantee safe and effective therapy. The AEIOU strategy—Alerts, Execution Team, Inputs, Orders, and Unified Record—provides a structured, evidence-informed framework that translates existing knowledge into consistent bedside practice. By guiding early recognition of CRRT indications, clarifying team competencies, standardizing the selection of machines and solutions, organizing prescription principles, and promoting unified documentation, AEIOU serves as a practical tool to reduce variability and strengthen clinical decision-making in this vulnerable population.
Future progress in the field will depend on the prospective evaluation and iterative refinement of the AEIOU framework across diverse institutions, its integration into training curricula and quality-improvement initiatives, and its use as a scaffold for multicenter collaboration. Ultimately, the advancement of CRRT in infants <10 kg will require not only continued innovation in miniaturized technologies, but also the widespread adoption of structured strategies such as AEIOU that ensure safer, more reliable, and more equitable care for the smallest and most fragile patients.
Author contributions
MP: Conceptualization, Data curation, Formal analysis, Funding acquisition, Investigation, Methodology, Project administration, Resources, Software, Supervision, Validation, Visualization, Writing – original draft, Writing – review & editing. JF-S: Conceptualization, Data curation, Formal analysis, Funding acquisition, Investigation, Methodology, Project administration, Resources, Software, Supervision, Validation, Visualization, Writing – original draft, Writing – review & editing. MS: Conceptualization, Data curation, Formal analysis, Funding acquisition, Investigation, Methodology, Project administration, Resources, Software, Supervision, Validation, Visualization, Writing – original draft, Writing – review & editing. FF: Conceptualization, Data curation, Formal analysis, Funding acquisition, Investigation, Methodology, Project administration, Resources, Software, Supervision, Validation, Visualization, Writing – original draft, Writing – review & editing.
Funding
The author(s) declared that financial support was received for this work and/or its publication. JFS was supported by the Medical School at Universidad de La Sabana, (Project MED 256-2019) and the Research Department at Fundacion Cardioinfantil-Instituto de Cardiologia, Bogotá, Colombia to conduct this review and pay the journal publication costs.
Conflict of interest
The author(s) declared that this work was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Generative AI statement
The author(s) declared that generative AI was not used in the creation of this manuscript.
Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.
Publisher's note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
1. Guzzo I, de Galasso L, Karabay Bayazit A, Yildizdas D, Schmitt CP, Hayes W, et al. Acute paediatric kidney replacement therapies in Europe: demographic results from the EurAKId registry. Nephrol Dial Transplant. (2022) 37(4):770–80. doi: 10.1093/ndt/gfab280
2. Starr MC, Charlton JR, Guillet R, Reidy K, Tipple TE, Jetton JG, et al. Advances in neonatal acute kidney injury. Pediatrics. (2021) 148(5):e2021051220. doi: 10.1542/peds.2021-051220
3. Kedarnath M, Alexander EC, Deep A. Safety and efficacy of continuous renal replacement therapy for children less than 10 kg using standard adult machines. Eur J Pediatr. (2023) 182(8):3619–29. doi: 10.1007/s00431-023-05007-y
4. Menon S, Starr MC, Zang H, Collins M, Damian MA, Fuhrman D, et al. Characteristics and outcomes of children ≤10 kg receiving continuous kidney replacement therapy: a WE-ROCK study. Pediatr Nephrol. (2025) 40:253–64. doi: 10.1007/s00467-024-06438-x
5. Ronco C, Garzotto F, Brendolan A, Zanella M, Bellettato M, Vedovato S, et al. Continuous renal replacement therapy in neonates and small infants: development and first-in-human use of a miniaturised machine (CARPEDIEM). Lancet. (2014) 383(9931):1807–13. doi: 10.1016/S0140-6736(14)60799-6
6. Peruzzi L, Bonaudo R, Amore A, Chiale F, Donadio ME, Vergano L, et al. Neonatal sepsis with multi-organ failure and treated with a new dialysis device specifically designed for newborns. Case Rep Nephrol Urol. (2014) 4(2):113–9. doi: 10.1159/000363691
7. Vidal E, Cocchi E, Paglialonga F, Ricci Z, Garzotto F, Peruzzi L, et al. Continuous veno-venous hemodialysis using the cardio-renal pediatric dialysis emergency machine™: first clinical experiences. Blood Purif. (2019) 47(1–3):149–55. doi: 10.1159/000494437
8. Garzotto F, Vidal E, Ricci Z, Paglialonga F, Giordano M, Laforgia N, et al. Continuous kidney replacement therapy in critically ill neonates and infants: a retrospective analysis of clinical results with a dedicated device. Pediatr Nephrol. (2020) 35(9):1699–705. doi: 10.1007/s00467-020-04562-y
9. Battista J, De Luca D, Dit Trolli SE, Allard L, Bacchetta J, Bouhamri N, et al. CARPEDIEM® for continuous kidney replacement therapy in neonates and small infants: a French multicenter retrospective study. Pediatr Nephrol. (2023) 38:2827–37. doi: 10.1007/s00467-022-05871-0
10. Askenazi D, Ingram D, White S, Cramer M, Borasino S, Coghill C, et al. Smaller circuits for smaller patients: improving renal support therapy with Aquadex™. Pediatr Nephrol. (2016) 31(5):791–853. doi: 10.1007/s00467-015-3259-3
11. Menon S, Broderick J, Munshi R, Dill L, DePaoli B, Fathallah-Shaykh S, et al. Kidney support in children using an ultrafiltration device: a multicenter, retrospective study. Clin J Am Soc Nephrol. (2019) 14(10):1432–40. doi: 10.2215/CJN.03240319
12. Lambert H, Hiu S, Coulthard MG, Matthews JNS, Holstein EM, Crosier J, et al. The infant kidney dialysis and ultrafiltration (I-KID) study: a stepped-wedge cluster-randomized study in infants, comparing peritoneal dialysis, continuous venovenous hemofiltration, and Newcastle infant dialysis ultrafiltration system, a novel infant hemodialysis device. Pediatr Crit Care Med. (2023) 24(7):604–13. doi: 10.1097/PCC.0000000000003220
13. Slagle C, Gist K, Krallman K, Morgan J, Menon S; ICONIC Collaborative. ICONIC—improving Carpediem™ Outcomes in Neonates and Infants Through Collaboration [poster]. Cincinnati (OH): Cincinnati Children’s Hospital Medical Center (2022).
14. Mohamed TH, Morgan J, Mottes TA, Askenazi D, Jetton JG, Menon S. Kidney support for babies: building a comprehensive and integrated neonatal kidney support therapy program. Pediatr Nephrol. (2022) 38(2):345–59. doi: 10.1007/s00467-022-05768-y
15. Starr MC, Gist KM, Zang H, Ollberding NJ, Balani S, Cappoli A, et al. Continuous kidney replacement therapy and survival in children and young adults: findings from the multinational WE-ROCK collaborative. Am J Kidney Dis. (2024) 84(4):406–15. doi: 10.1053/j.ajkd.2023.12.017
16. Askenazi DJ, Goldstein SL, Koralkar R, Fortenberry JD, Baum M, Hackbarth R, et al. Continuous renal replacement therapy for children ≤10 kg: a report from the prospective pediatric continuous renal replacement therapy registry. J Pediatr. (2013) 162(3):587–592.e3. doi: 10.1016/j.jpeds.2012.08.044
17. Alobaidi R, Morgan C, Basu RK, Stenson E, Featherstone R, Majumdar SR, et al. Association between fluid balance and outcomes in critically ill children: a systematic review and meta-analysis. JAMA Pediatr. (2018) 172(3):257–68. doi: 10.1001/jamapediatrics.2017.4540
18. Perner A, Prowle J, Joannidis M, Young P, Hjortrup PB, Pettilä V. Fluid management in acute kidney injury. Intensive Care Med. (2017) 43:807–15. doi: 10.1007/s00134-017-4817-x
19. Goldstein SL, Vidal E, Ricci Z, Paglialonga F, Peruzzi L, Giordano M, et al. Survival of infants treated with CKRT: comparing adapted adult platforms with the Carpediem™. Pediatr Nephrol. (2022) 37(4):667–75. doi: 10.1007/s00467-021-05180-y
20. Daverio M, Cortina G, Jones A, Ricci Z, Demirkol D, Raymakers-Janssen P, et al. Continuous kidney replacement therapy practices in pediatric intensive care units across Europe. JAMA Netw Open. (2022) 5(12):e2246901. doi: 10.1001/jamanetworkopen.2022.46901
21. Kidney Disease: Improving Global Outcomes (KDIGO) Acute Kidney Injury Work Group. KDIGO Clinical practice guideline for acute kidney injury. Kidney Int Suppl. (2012) 2(1):1–138. doi: 10.1038/kisup.2012.1
22. Deep A. Large-sized vascular access in continuous renal replacement therapy: balancing filter life with risk of vascular thrombosis. Pediatr Crit Care Med. (2021) 22(8):766–7. doi: 10.1097/PCC.0000000000002772
23. Symons JM, Chua AN, Somers MJG, Baum MA, Bunchman TE, Benfield MR, et al. Demographic characteristics of pediatric continuous renal replacement therapy: a report of the prospective pediatric continuous renal replacement therapy registry. Clin J Am Soc Nephrol. (2007) 2(4):732–8. doi: 10.2215/CJN.03200906
24. Prada Rico M, Fernández Sarmiento J, Rojas Velasquez AM, González Chaparro LS, Gastelbondo Amaya R, Mulett Hoyos H, et al. Regional citrate anticoagulation for continuous renal replacement therapy in children. Pediatr Nephrol. (2016) 31(12):2335–43. doi: 10.1007/s00467-016-3544-9
25. Warkentin TE. Heparin-induced thrombocytopenia: diagnosis and management. Circulation. (2004) 110(18):e454–8. doi: 10.1161/01.CIR.0000147537.72829.1B
26. Mariano F, Bergamo D, Gangemi E, Hollo Z, Stella M, Triolo G. Citrate anticoagulation for continuous renal replacement therapy in critically ill patients: success and limits. Int J Nephrol. (2011) 2011:748320. doi: 10.4061/2011/748320
27. Oudemans-van Straaten HM, Kellum JA, Bellomo R. Clinical review: anticoagulation for continuous renal replacement therapy—heparin or citrate? Crit Care. (2011) 15(1):202. doi: 10.1186/cc9419
28. Oudemans-Van Straaten HM, Bosman RJ, Koopmans M, van der Voort PH, Wester JP, van der Spoel JI, et al. Citrate anticoagulation for continuous venovenous hemofiltration. Crit Care Med. (2009) 37(2):545–52. doi: 10.1097/CCM.0b013e31819598a1
29. Deep A, Zoha M, Dutta K, Kukreja P. Prostacyclin as an anticoagulant for continuous renal replacement therapy in children. Blood Purif. (2017) 43(1–3):279–89. doi: 10.1159/000452754
30. Goonasekera CD, Wang J, Bunchman TE, Deep A. Factors affecting circuit life during continuous renal replacement therapy in children with liver failure. Ther Apher Dial. (2015) 19(1):16–22. doi: 10.1111/1744-9987.12224
31. Miyaji MJ, Ide K, Takashima K, Maeno M, Krallman KA, Lazear D, et al. Comparison of nafamostat mesilate to citrate anticoagulation in pediatric continuous kidney replacement therapy. Pediatr Nephrol. (2022) 37(11):2733–42. doi: 10.1007/s00467-022-05502-8
32. Medtronic. CARPEDIEM™ (Cardio-Renal Pediatric Dialysis Emergency Machine) Technical Manual: System Description, Pump Specifications, and Operating Parameters. Minneapolis (MN): Medtronic Renal Care Solutions (2020).
33. Dubinsky S, Watt K, Saleeb S, Ahmed B, Carter C, Yeung CHT, et al. Pharmacokinetics of commonly used medications in children receiving continuous renal replacement therapy: a systematic review of current literature. Clin Pharmacokinet. (2022) 61(2):189–229. doi: 10.1007/s40262-021-01085-z
Keywords: acute kidney injury (AKI), AEIOU strategy, continuous renal replacement therapy (CRRT), extracorporeal therapies, infants <10 kg
Citation: Prada Rico M, Fernández-Sarmiento J, Santiago MJ and Flores F (2026) AEIOU strategy for continuous renal replacement therapy in infants <10 kg: a practical and educational framework. Front. Pediatr. 13:1725990. doi: 10.3389/fped.2025.1725990
Received: 15 October 2025; Revised: 8 December 2025;
Accepted: 15 December 2025;
Published: 9 January 2026.
Edited by:
Hitesh Singh Sandhu, University of Tennessee Health Science Center (UTHSC), Memphis, United StatesReviewed by:
Sasidaran Kandadamy, Dr. Mehta's Children’s Hospital, IndiaOzden Ozgur Horoz, Çukurova University, Türkiye
Copyright: © 2026 Prada Rico, Fernández-Sarmiento, Santiago and Flores. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Mayerly Prada Rio bXByYWRhckBsYWNhcmRpby5vcmc=; Jaime Fernández-Sarmiento, amFpbWVmZUB1bmlzYWJhbmEuZWR1LmNv
Mayerly Prada Rico1*