Abstract
Background and Objective: Intravenous contrast-enhanced ultrasound (CEUS), using the second-generation ultrasound contrast agent SonoVue®, has been widely used in adults. In 2016, it was approved for pediatric applications by the American Food and Drug Administration (FDA). However, it has not been approved by the Chinese Food and Drug Administration (CFDA). The objective of the study was to evaluate the safety and effectiveness of CEUS in children prospectively at a single center in China.
Methods: A total of 312 cases of Chinese children were enrolled in clinical trials. Contrast agent was given intravenously with two different doses, including 2.4 ml/time and 0.03 ml/kg. All CEUS was performed for evaluating adverse effect and the diagnostic accuracy compared with the pathology and enhanced CT.
Results: All 312 subjects underwent CEUS successfully. The dose of contrast agent for CEUS was 2.4 ml between November 2015 and June 2016, which was modified to 0.03 ml/kg between July 2016 and April 2019, according to the recommendation of the FDA. With the two different doses of the contrast agent, the heart rate, respiration rate, oxygen saturation, and blood pressure of the participants had no statistically significant difference (P > 0.05) before and after administration. The blood pressure had been significantly decreased in participants who received combined anesthetic administration. Following 600 intravenous injections of the CEUS, there were three cases of transient rash and three cases of hypotension (n = 6, 1.92%). The six recovered quickly after receiving intravenous methylprednisolone and epinephrine. Most of the studies were performed for evaluating renal microcirculation and assisting renal biopsy (192/312 [61.5%]), which together had a 98.9% effectiveness in the identification of pathology in the specimens. Some studies were conducted to identify a mass in the liver, retroperitoneum, abdominal cavity, kidneys, testicles, thyroid, and so on (99/312 [31.4%]), which had a 97.6% accuracy. The other studies were conducted to identify trauma, vascular malformation, infection, hemorrhage, and so on (21/312 [6.73%]), which had a similar accuracy to enhanced CT.
Conclusion: The adverse effects of CEUS in children are similar to that in adults. The results indicate that it is safe to use SonoVue® for CEUS in pediatric patients.
Introduction
Contrast-enhanced ultrasound (CEUS) has been widely used in adults. In 2016, it was approved for intravenous use to diagnose liver lesions and intravesical use in children, basing on the big data from studies about safety and feasibility with SonoVue®/Lumason® by the FDA. However, it lacked relevant studies in China. The study was to evaluate the safety and feasibility of intravenous CEUS with SonoVue® in children under 18 years of age.
Patients And Methods
Patients
The study was performed in Shenzhen Children’s Hospital between November 2015 and April 2019. A total of 312 Chinese children were enrolled in clinical trials and 600 intravenous injections were performed. All subjects met the inclusion and exclusion criteria as follows.
Inclusion criteria
Inpatients or outpatients from birth to 18 years old.
Performance of CEUS was needed for diagnosis.
Bedside CEUS for diagnosis instead of CT and MR.
Agreed to be enrolled into the trial and signed the informed consent.
Exclusion criteria
Allergy to sulfur hexafluoride (SF6) or other components of SonoVue®.
Having an allergic history of protein or egg.
Congenital heart disease with right-to-left shunt, severe pulmonary hypertension, or uncontrolled hypertension.
Severe respiratory failure.
Without informed consent.
The prospective study with SonoVue® was approved by the Chinese Ethics Committee (No: 2016[003], ChiCTR-ONh-16009236).
The Protocol of CEUS
All US examinations, including baseline US and CEUS, were performed with 4.5–9.0-MHz C2-9 convex array probes and 5.0–9.0-MHz 9L linear probes in the US system (LogiQ E9 Ultrasound System; GE Healthcare, Milwaukee, WI). And the mechanical index (MI) during the CEUS examination was lower than 0.2.
The preventive measures for allergy were taken in trials. We weighed all subjects and calculated the doses of anti-allergic medications, including 0.01 mg/kg of 0.1% adrenaline (the maximal dose 0.3 mg) and 1∼2 mg/kg of methylprednisolone, and prepared a rescue equipment with the medications as a routine.
The contrast agent SonoVue® was administered in two different ways: 2.4 ml/time of SonoVue® before June 2016, while 0.03 ml/kg/time between July 2016 and April 2019, according to the recommendation of FDA. The contrast agents were bolus injected through peripheral veins within 3s, and then the irrigation of the catheters with 0.9% of sodium chloride solutions, keeping a velocity of about 2 ml/s, were performed. All subjects underwent different times of CEUS examinations, including 34 cases undergoing one CEUS, 270 cases undergoing two continuous CEUS, 6 cases undergoing three continuous CEUS, and 1 case undergoing four times.
The CEUS examinations and the intravenous bolus injections with SonoVue® were started simultaneously. The views and images of CEUS were acquired and stored until the microbubbles were cleared up, and the wash-in time, time to peak, and wash-out time were recorded in the whole examinations. The heart rate, respiratory rate, oxygen saturation, and blood pressure were observed and recorded respectively at three different times: before administration, immediately after administration, and 15 min after administration. As for awake participants, we inquired and made a record of any discomfort, such as pain, nausea, vomiting, and so on, besides observing the appearance and the rash. The follow-ups, which contained vital sign observation and rash management, were performed within 24 h before CEUS, 24 h after CEUS, and 72 h after CEUS. If it was necessary, for some subjects having a need for more than one intravenous CEUS, especially for the subjects with a diagnostic purpose, the next injections of SonoVue® were performed at least 20 min later.
The images were evaluated for the following: ①hepatic lesions, such as vascular malformation, abscess, hyperplasia, and so on; ②infarction and necrosis in spleen, kidney, and other solid organs; ③distinction between benign and malignant lesions; ④perfusion and blood supply of the mass and guiding the puncture; ⑤active hemorrhage of the site after puncture; ⑥complications after puncture, especially the vascular complications; and ⑦ distinguishing the hematoma from mass.
Statistical Analysis
The SPSS software (SPSS, version 21.0, IBM) was employed for statistical analysis. Means and standard deviations were used to summarize continuous data for descriptive statistics purposes. The patients’ characteristics were evaluated by using t-tests for continuous variables, while descriptions were made for categorical data. The significance was set at a p-value < 0.05.
Results
General Information and Vital Signs
Among 312 participants, 179 were male and 133 were female. The median age was 7.08 years old (from 5 days to 14.5 years) and the median weight was 23 kg (3.7–65.0 kg). 287 cases had complete records of vital sign, including heart rate (HR), respiration rate (RR), oxygen saturation (SO2), systolic blood pressure (SBP), and diastolic blood pressure (DBP), before and after administration. Ten cases of the records were shown as follows (Table 1).
Table 1
| Case number | Gender | Age | Weight | Condition | Before administration | After administration | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| HR (/min) | RR (/min) | SO2 (%) | SBP (mmHg) | DBP (mmHg) | HR (/min) | RR (/min) | SO2 (%) | SBP (mmHg) | DBP (mmHg) | |||||
| Case 1 | male | 3y3m | 15.3 | combined anesthesia | 125 | 30 | 100 | 96 | 53 | 117 | 30 | 100 | 95 | 49 |
| Case 2 | male | 3y1m | 15 | combined anesthesia | 116 | 19 | 98 | 88 | 51 | 122 | 20 | 97 | 84 | 52 |
| Case 3 | male | 9y3m | 32 | combined anesthesia | 111 | 30 | 98 | 87 | 51 | 113 | 30 | 98 | 88 | 53 |
| Case 4 | female | 13y6m | 46.3 | combined anesthesia | 111 | 25 | 98 | 102 | 71 | 114 | 24 | 99 | 107 | 72 |
| Case 5 | male | 10m | 10.6 | combined anesthesia | 108 | 24 | 100 | 132 | 60 | 105 | 24 | 100 | 105 | 61 |
| Case 6 | female | 5y8m | 17 | awake | 116 | 18 | 97 | 80 | 42 | 115 | 20 | 97 | 79 | 40 |
| Case 7 | male | 14y4m | 57 | awake | 79 | 28 | 100 | 116 | 72 | 78 | 28 | 100 | 121 | 67 |
| Case 8 | female | 5y6m | 19 | awake | 110 | 30 | 99 | 98 | 72 | 112 | 30 | 99 | 98 | 68 |
| Case 9 | male | 4m | 7.2 | sedation | 128 | 20 | 99 | 125 | 78 | 134 | 20 | 99 | 125 | 78 |
| Case 10 | male | 5y | 18 | coma | 78 | 30 | 100 | 136 | 79 | 80 | 30 | 100 | 135 | 80 |
Ten cases of the records before and after administration.
The Vital Signs Before and After Administration
All the 312 participants underwent CEUS successfully. They included: 228 cases in combined anesthesia, 63 cases in awake state, 18 cases in sedation with chloral hydrate, 3 cases in coma. Comparing the vital signs after administration with that before administration in both combined anesthetic and awake participants, there was a statistically significant difference (P < 0.05) in the combined anesthetic participants (Tables 2 and 3).
Table 2
| Subjects (n = 228) | Heart rate (/min) | Respiration rate (/min) | Oxygen saturation (%) | Systolic blood pressure (mmHg) | Diastolic blood pressure (mmHg) |
|---|---|---|---|---|---|
| Before administration | 108.79 ± 18.33 | 22.83 ± 3.84 | 98.80 ± 1.82 | 101.78 ± 15.24 | 60.03 ± 13.23 |
| After administration | 108.16 ± 17.44 | 22.80 ± 3.68 | 98.95 ± 1.36 | 99.69 ± 14.87 | 57.92 ± 13.24 |
| t | 0.834 | 0.329 | −1.388 | 3.829 | 5.086 |
| P | 0.408 | 0.743 | 0.167 | <0.001* | <0.001* |
The vital signs of the combined anesthetic participants.
*The significance was set at a p-value < 0.05 in the study.
Table 3
| Subjects (n = 63) | Heart rate (/min) | Respiration rate (/min) | Oxygen saturation (%) | Systolic blood pressure (mmHg) | Diastolic blood pressure (mmHg) |
|---|---|---|---|---|---|
| Before administration | 102.14 ± 22.00 | 23.68 ± 4.56 | 99.43 ± 1.17 | 104.94 ± 15.92 | 65.05 ± 10.81 |
| After administration | 100.27 ± 21.37 | 24.11 ± 5.08 | 99.48 ± 1.06 | 104.79 ± 14.04 | 65.68 ± 11.28 |
| t | 1.699 | −1.989 | −0.554 | 0.148 | −0.704 |
| P | 0.094 | 0.051 | 0.582 | 0.883 | 0.484 |
The vital signs of the awake participants.
The vital signs of 18 cases in sedation with chloral hydrate showed as follows: no statistically significant difference (P < 0.05) was found (Table 4).
Table 4
| Subjects (n = 18) | Heart rate (/min) | Respiration rate (/min) | Oxygen saturation (%) | Systolic blood pressure (mmHg) | Diastolic blood pressure (mmHg) |
|---|---|---|---|---|---|
| Before administration | 117.44 ± 18.64 | 23.33 ± 4.14 | 99.39 ± 0.61 | 90.78 ± 20.14 | 56.22 ± 17.20 |
| After administration | 117.89 ± 18.55 | 24.61 ± 6.04 | 99.56 ± 0.62 | 91.22 ± 20.57 | 58.94 ± 16.00 |
| t | −0.266 | −1.668 | −1.374 | −0.389 | −1.690 |
| P | 0.793 | 0.114 | 0.187 | 0.702 | 0.109 |
The vital signs of the participants in sedation with chloral hydrate.
In comatose participants, the vital signs after administration were not statistically significant different (P < 0.05) from the corresponding signs before administration: heart rate (102.67 ± 26.10 vs. 105.33 ± 26.03) beats/min,t = −4.000, P = 0.057; respiratory rate (30.00 ± 2.00 vs. 30.33 ± 1.53 breaths/min,t = −1.000, P = 0.423; oxygen saturation (95.33 ± 5.03 vs. 96.67 ± 5.77)%,t = 1.000, P = 0.423); systolic blood pressure (107.67 ± 26.69 vs. 109.00 ± 26.00 mmHg,t = −0.718, P = 0.547); and diastolic blood pressure (58.00 ± 18.52 vs. 57.67 ± 19.50 mmHg,t = 0.378, P = 0.742).
Adverse Effects
There were three cases of rash (Figure 1) and three cases of hypotension (Table 5).
Figure 1
Table 5
| Gender | Age | Clinical diagnosis | Drug batch number | Time of allergy | Allergic performance | Management | Outcome | |
|---|---|---|---|---|---|---|---|---|
| Case1 | Male | 5Y8M | Solitary hematuria | 17A011A | The first CEUS | Chest rash (Figure 1A) with stable condition | Intravenous injection of methylprednisolone 20 mg | Complete relief |
| Case 2 | Male | 7Y7M | Henoch-Schonlein purpura nephritis | 17A011A | The second CEUS | Rash in posterior neck, facial, and back | Intravenous injection of methylprednisolone 40 mg | Complete relief |
| Case 3 | Female | 7Y11M | Nephrotic syndrome | 17A023A | The second CEUS | Flushed face, and rash in posterior neck, facial, and back | Intravenous injection of methylprednisolone 40 mg | Complete relief |
| Case 4 | Male | 13Y3M | Inflammatory myofibroblastic tumor | 17A023A | The first CEUS | Rash in face, chest, and abdominal, a high heart rate (148 bpm), hypotension (66/34 mmHg) | Intravenous injection of dexamethasone 5 mg, methylprednisolone 40 mg, calcium gluconate 10 ml, adrenaline ((0.02 + 0.03+0.05) mg, 0.05 mg/kg*min) | Complete relief |
| Case 5 | Male | 6Y3M | Henoch-Schonlein purpura nephritis | 17A038A | The first CEUS | Continuous cough, hypotension (53/17 mmHg) | Intravenous injection of methylprednisolone 40 mg | Complete relief |
| Case 6 | Female | 6Y3M | Nephrotic syndrome | 17A052A | The first CEUS | Hypotension (78/47 mmHg) | Intravenous injection of methylprednisolone 40 mg | Complete relief |
The allergies and the managements.
CEUS and the Enhancement Modality
In our study, CEUS was performed for many organs, including liver, kidney, spleen, pancreas, thyroid, testicles, and so on. The arrival time (AT), time to peak (TTP), and the wash-out time were documented in Table 6.
Table 6
| Arrival time (s) | Time to peak (s) | Wash-out time (s) | |
|---|---|---|---|
| Liver (35 cases) | 8.75 ± 2.06 | 26.25 ± 1.25 | 462.50 ± 2.33.24 |
| Kidney (187 cases) | 10.17 ± 3.83 | 24.41 ± 6.53 | 450.98 ± 164.18 |
The arrival time, time to peak, and the wash-out time of liver and kidney.
With different doses, renal CEUS was performed and the wash-out time was recorded. It was showed below (Table 7).
Table 7
| Wash-out time (s) | |
|---|---|
| Dose 1 | 517.91 ± 138.78 |
| Dose 2 | 386.38 ± 102.96 |
| t | 7.001 |
| P | 0.025 |
The wash-out time of renal CEUS with different doses of contrast agents.
Dose 1: 2.4 ml/time; Dose 2: 0.03 ml/kg/time. And one case can’t acquire the wash-out time because of the severe adverse effect.
The Accuracy and Feasibility of CEUS
Among 312 subjects, 287 cases were made the terminal diagnosis by biopsy, surgery, or clinical diagnostic standards, while 25 cases were available for follow-up. (Table 8)
Table 8
| Diffuse Renal Lesions (cases) | Mass (cases) | Others (cases) | Loss of Follow-ups (cases) |
|---|---|---|---|
| Minimal change disease (MCD) (60) | Hepatoblastoma (11) | Gastrointestinal hemorrhage (2) | Hepatic mass (8) |
| Purpura nephritis (44) | Hepatocellular carcinoma (3) | Postoperative hemorrhage (3) | Renal mass (3) |
| IgA nephropathy (32) | Neuroblastoma (15) | Subcutaneous hemorrhage (1) | Cervical mass (2) |
| lupus nephritis (LN) (19) | Lymphoma (12) | Trauma (2) | Testicular mass (1) |
| Mesangial proliferative glomerulonephritis (MsPGN) (8) | Teratoma (6) | Renal segmental infarction (1) | Thyroid nodules (1) |
| Thin basement membrane nephropathy(TBMN)(7) | Yolk sac tumor (3) | Hepatic vascular malformation (2) | Pelvic mass (1) |
| Focal segmental glomerular sclerosis (FSGS) (3) | Rhabdomyosarcoma (3) | Budd-Chiari syndrome (2) | Retroperitoneal mass (1) |
| Alport syndrome (2) | Nephroblastoma (2) | Venous thrombosis (2) | Diffuse renal lesions (8) |
| Membranous nephropathy (MN) (2) | Hemangioma (4) | Infection (4) | |
| Thrombotic microangiopathy (TMA) (1) | Langerhans cell histiocytosis (LCH) (2) | Fatty liver (1) | |
| Hepatitis B virus-associated membranous nephritis (HBV-MN) (1) | Inflammatory myofibroblastic tumor (IMT) (2) | ||
| Endocapillary proliferative glomerulonephritis (EPGN) (1) | Malignant triton tumor (1) | ||
| Crescentic glomerulonephritis (1) | Clear cell sarcoma (1) | ||
| Interstitial nephritis (1) | Malignant rhabdoid tumor (MRT) (1) | ||
| Sclerosing glomerulonephritis (1) | Pheochromocytoma (1) | ||
| Cystic nephroma (1) | |||
| Failing to acquire the glomeruli (2) | Fibrosarcoma (1) | ||
| Malignant mixed germ cell tumor (1) | |||
| Neurofibroma (1) | |||
| Solid-pseudopapillary tumor (SPT) (1) | |||
| Benign lymph nodes (1) | |||
| Papillary thyroid carcinoma (PTC) (1) | |||
| Renal cyst (2) | |||
| Splenic cyst (1) | |||
| Small cell lung cancer (1) | |||
| Focal nodular hyperplasia (FNH) (1) | |||
| Glycogen storage disease (GSD) (1) |
The clinical or pathologic diagnosis of the subjects.
Figures 2–8 respectively showed the case of post-punctural active hemorrhage of kidney (Figure 2), hepatocellular carcinoma (Figure 3), inflammatory myofibroblastic tumor (IMT) (Figure 4), solid-pseudopapillary tumor of pancreas (SPT) (Figure 5), blood clot in stomach (Figure 6), thrombosis in superficial femoral artery (Figure 7), Bacillus Calmette-Guerin(BCG) vaccination disease with splenic necrosis (Figure 8), and renal segmental infarction.
Figure 2
Figure 3
Figure 4
Figure 5
Figure 6
Figure 7
Figure 8
Most of the studies were performed for evaluating renal microcirculation and assisting renal biopsy (192/312 [61.5%]), which had a 98.9% effectiveness in the identification of pathology in the specimens (Table 9).
Table 9
| Cases for renal biopsy with renal CEUS | Cases with effective pathological specimen | Cases with ineffective pathological specimen | Effective rate |
|---|---|---|---|
| 192 | 190 | 2 cases failing to acquire the glomeruli | 98.9% |
The effective rate of the pathological specimens with renal CEUS.
Some were used to identify the mass in almost the whole body, including 99 (31.4%) cases for the liver (31 cases), retroperitoneum (14 cases), abdominal cavity (14 cases), mediastinum (6 cases), kidney (10 cases), testicles (4 cases), thyroid (3 cases), pancreas (2 cases), spleen (3 cases), and so on. Except for loss to follow-up for 17 cases, it had a 97.6% accuracy as shown by the comparison of CEUS findings with pathology. The others were for identifying 21 (6.73%) cases for trauma, vascular malformation, infection, thrombosis, hemorrhage, infarction, and fatty liver, which had the same accuracy compared with enhanced CT (Table 10).
Table 10
| Cases for identifying the mass with CEUS | Loss of following up | Cases for CEUS finding fitting to the pathology | Cases unfitting to the pathology | Accurate rate |
|---|---|---|---|---|
| 99 | 17 | 80 | 2 | 97.6% |
The effective rate of the pathological specimens with renal CEUS.
Discussion
SonoVue® has been widely used in adults in the recent years, and has been proven to be safe (Piscaglia and Bolondi, 2006). However, owing to lack of relevant guideline or support from clinical trials, it has had a slow approval for clinical use in pediatric cases. The intravenous CEUS in children was off-label use, although it was reported in multiple centers in Europe (Claudon et al., 2013; Sidhu et al., 2017). In April 2016, SonoVue® was approved for non-cardiac use in pediatric patients by the FDA. In January 2017 it was approved for intravesical use, among pediatric cases. As for China, the pediatric application of SonoVue® has not been approved by the CFDA due to the lack of relevant safety data from phase IV clinical trials.
Our study institute was qualified to conduct the clinical trial, due to the qualified staff, standard equipment, and the approval of the Ethics Committee.
Our results showed that CEUS was effective for the diagnosis of most diseases, especially those that were associated with the microcirculation. (Schreiber-Dietrich and Dietrich, 2012; Darge et al., 2013; Stenzel and Mentzel, 2014; Muyi and Bei, 2018)
CEUS in pediatric interventional therapy had its own advantage. In our study, 174 cases undergoing renal CEUS provided clinically useful results pertaining to evaluating the renal microcircular perfusion and the hemorrhage after puncture. This enabled the early identification of stanch bleeding in time. Forty-six (n = 46) cases for mass puncture evaded the necrosis and facilitated the acquisition of eligible tissue for pathology. After intravenous injection, microbubble became the maker of the vascular bed, which made the blood vessels and tissue perfusion images very clear due to the enhancement, in both the great vessels as well as the capillaries. The avascular, potential necrotic and liquefied areas could be recognized due to the expansion of the vision in the interventional operation and avoiding the damage of adjacent normal tissue after enhancement. Many structures were hard to recognize in B-mode US, but were exhibited in CEUS.
In our study, CEUS was performed for evaluating the post-traumatic kidney, renal infarction, splenic cyst, splenic necrosis, splenic hemangioma, hemorrhage in stomach and duodenum, arterial thrombosis, and mass in various organs, and had a high diagnostic accuracy rate compared with the result of surgery pathology and clinical follow-up. Besides, CEUS has been already used in organ transplantation. It was demonstrated that CEUS can help to access the vessels of target organs, the perfusion of transplanted organs, and the post-transplanted complications (Pschierer et al., 2015). CEUS can be used for evaluating the mass before and after radiation ablation (Xiaoer et al., 2016), including the mass size, the boundary, the location, and the therapeutic effect. In addition to the intravenous applications, CEUS can be used for urologic diagnosis by retrograde ureteropyelography, such as vesicoureteral reflux (VUR) (Xia et al., 2012).
According to the recommendation by the FDA, the optimal dose of SonoVue® in adult was 2.4 ml/time, and the dose range was 1.0–4.8 ml depending on the target organs and the equipment (Piscaglia et al., 2012). In the previous studies, many researchers started with different contrast agent administration strategies, based on the purposes of the study, various ages and weights of subjects, and the equipment used, especially the frequency of probes. The additional dose was necessary when performing CEUS in patient with large body masses, in a deep target organ or lesion, or using a high-frequency probe. In early period, it was suggested that 0.1 ml should be added for every additional year or additional kilogram; and the dose of 2.4 kg was suitable for the subject with a weight over 24 kg in Europe (Stenzel, 2013). There were studies with a fixed dose of 0.1, 0.5, 1.2, 2.4, and 4.8 ml reported (Bonini et al., 2007; Piskunowicz et al., 2012; Jacob et al., 2013; Knieling et al., 2016). In the recent guideline in USA, the recommended dose was 0.03 ml/kg and the maximal dose was 2.4 ml in children (Lumason prescribing information website). We used different administration strategies in our study. It was found that the wash-out time was associated with the dose. The larger dose had a longer wash-out time. For reducing the detention time, the dose of 0.03 ml/kg may be the better option than the dose of 2.4 ml/time on the basis of the same diagnostic efficacy.
In our study, there were three cases of mild allergic reactions, which occurred after administering CEUS from different batches. The earliest allergy occurred within 5 min after injection of contrast agent, and relieved gradually after the immediate administration of dexamethasone injection. For the other allergic reactions, we used methylprednisolone dexamethasone, and acquired a better therapeutic effect. By now, it was known that methylprednisolone had a quicker effectiveness than dexamethasone for anti-allergy therapy. And what’s more, some researchers suggested that, using dexamethasone for prevention should be a routine before CEUS examinations (Sidhu et al., 2018). However, for the sake of safety, we didn’t take that measure in our study. Instead it was stressed that emergency preparations needed to be in place for immediate response against such adverse effects. Besides the rash, there were three cases of hypotension which happened after administration of CEUS, which was completely resolved after anti-shock therapy. Above all, intravenous CEUS with SonoVue® was generally safe.
According to our experience from the trial, we made a conclusion for the measures dealing with abrupt adverse effects. Firstly, to stop administration and to keep a suitable position—that is, lying down and raising legs when hypotension happened. Secondly, to keep the airways open and to provide oxygen should be done. And then the anti-allergy drugs should be intravenously injected, and the vital signs should be monitored in real time. For the anti-allergy therapy, we can use a dose of 1–2 mg/kg of methylprednisolone for the subjects with rash, and use a dose of 0.01 mg/kg of 0.1% adrenaline for the subjects with hypotension. And they can be used repeatedly 5–15 min after the first use if necessary. The hemodynamic change can be monitored by the arterial cannulation as well.
Conclusion
Intravenous CEUS for pediatric applications was generally safe as observed with a low frequency of adverse effects. In addition, it was associated with better diagnostic effectiveness so that it can be performed for many kinds of medical issues.
Limitation
There were limitations in our study. The study was in a single center, and lack of cooperation with other institutions. Besides, the disease entities were so complex that there was no specific CEUS modality. In the next step, we will collaborate with other institutions and make our study more standard and systematic, which are favorable for promoting the CEUS application in children and make it more feasible.
Statements
Data availability statement
The datasets generated for this study are available on request to the corresponding author.
Ethics statement
The studies involving human participants were reviewed and approved by the Chinese Ethics Committee (No: 2016[003], ChiCTR-ONh-16009236). Written informed consent to participate in this study was provided by the participants’ legal guardian/next of kin. Written informed consent was obtained from the individual(s), and minor(s)’ legal guardian/next of kin, for the publication of any potentially identifiable images or data included in the article.
Author contributions
The corresponding author BX conceived and conducted the study and gave the guidance in the whole study. The first author M-yM collected and chose the cases, performed the literature search and data extraction, and wrote the paper. The author W-IC assisted to the statistical analysis. Other authors participated in the study and provided some professional guidance. All authors read and approved the manuscript.
Conflict of interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
References
1
BoniniG.PezzottaG.MorzentiC.AgazziR.NaniR. (2007). Contrast-enhanced ultrasound with SonoVue in the evaluation of postoperative complications in pediatric liver transplant recipients. J. Ultrasound10 (2), 99–106. doi: 10.1016/j.jus.2007.02.008
2
ClaudonM.DietrichC. F.ChoiB. I.CosgroveD. O.KudoM.NolsøeC. P.et al. (2013). Guidelines and good clinical practice recommendations for contrast enhanced ultrasound (CEUS) in the liver–update 2012: a WFUMB-EFSUMB initiative in cooperation with representatives of AFSUMB, AIUM, ASUM, FLAUS and ICUS [J]. Ultraschall Med.34 (1), 11–29. doi: 10.1055/s-0032-1325499
3
DargeK.PapadopouluF.NtouliaA.BulasD. I.ColeyB. D.FordhamL. A.et al. (2013). Safety of contrast-enhanced ultrasound in children for non-cardiac applications: a review by the Society for Pediatric Radiology (SPR) and the International Contrast Ultrasound Society (ICUS). Pediatr. Radiol.43 (9), 1063–1073. doi: 10.1007/s00247-013-2746-6
4
JacobJ.DeganelloA.SellarsM. E.HadzicN.SidhuP. S. (2013). Contrast enhanced ultrasound (CEUS) characterization of grey-scale sonographic indeterminate focal liver lesions in paediatric practice. Ultraschall Med.34 (6), 529–540. doi: 10.1055/s-0033-1355785
5
KnielingF.StrobelD.RompelO.ZapkeM.Menendez-CastroC.WölfelM.et al. (2016). Spectrum, applicability and diagnostic capacity of contrast-enhanced ultrasound in pediatric patients and young adults after intravenous application – a retrospective trial. Ultraschall Med.37 (6), 619–626. doi: 10.1055/s-0042-108429
6
Lumason prescribing information websitehttp://imaging.bracco.com/sites/braccoimaging.com/files/technica_sheet_pdf/us-en-2017-01-04-spc-lumason.pdf. Accessed March 3, 2017.
7
MuyiM.BeiX. (2018). Contrast-enhanced US of Focal Liver Lesions in Children. Chin J. Med. Ultrasound (Electronic Edition)15 (2), 81–89. doi: 10.3877/cma.j.issn.1672-6448.2018.02.001
8
PiscagliaF.BolondiL. (2006). The safety of Sonovue in abdominal applications: retrospective analysis of 23188 investigations. Ultrasound Med. Biol.32 (9), 1369–1375. doi: 10.1016/j.ultrasmedbio.2006.05.031
9
PiscagliaF.NolsøeC.DietrichC. F.CosgroveD. O.GiljaO. H.Bachmannet al. (2012). The EFSUMB guidelines and recommendations on the clinical practice of contrast enhanced ultrasound (CEUS): update 2011 on non-hepatic applications. Ultraschall Med.33 (1), 33–59. doi: 10.1055/s-0031-1281676
10
PiskunowiczP.KosiakW.BatkoT. (2012). Intravenous application of secondgeneration ultrasound contrast agents in children: a review of the literature. Ultraschall Med.33 (2), 135–140. doi: 10.1055/s-0031-1281936
11
PschiererK.GrothuesD.RennertJ.da SilvaN. P.SchreyerA. G.MelterM.et al. (2015). Evaluation of the diagnostic accuracy of CEUS in children with benign and malignant liver lesions and portal vein anomalies. Clin. Hemorheol. Microcirc.61 (2), 333–345. doi: 10.3233/CH-152003
12
Schreiber-DietrichD.DietrichC. F. (2012). Contrast enhanced ultrasound (CEUS) and off-label use (in children). Ultraschall Med.33 (3), 295–296. doi: 10.1055/s-0032-1318769
13
SidhuP. S.CantisaniV.DeganelloA.DietrichC. F.DuranC.FrankeD.et al. (2017). Role of contrast-enhanced ultrasound (CEUS) in paediatric practice: An EFSUMB position statement. Ultraschall Med.38 (1), 33–43. doi: 10.1055/s-0042-110394
14
SidhuP.CantisaniV.DietrichC.GiljaO. H.SaftoiuA.BartelsE.et al. (2018). The EFSUMB guidelines and recommendations for the clinical practice of contrast-enhanced ultrasound (CEUS) in non-hepatic applications: update 2017. Ultraschall Med.39 (2), e2–e44. doi: 10.1055/a-0586-1107
15
StenzelM.MentzelH. J. (2014). Ultrasound elastography and contrast-enhanced ultrasound in infants, children and adolescents. Eur. J. Radiol.83 (9), 1560–1569. doi: 10.1016/j.ejrad.2014.06.007
16
StenzelM. (2013). Intravenous contrast-enhanced sonography in children and adolescents – a single center experience. J. Ultrasonography13 (53), 133–144. doi: 10.15557/JoU.2013.0014
17
XiaF.BeiX.ShoulinL.YanL.XiumingD.HongweiT.et al. (2012). Contrast ultrasonography in children with urinary tract diseases and its initial application. Chin. J. Ultrasound Med.3, 266–269. doi: 10.3969/j.issn.1002-0101.2012.03.029
18
XiaoerZ.MingX.XiaoyanX.LongfeiC.TengS.JieyiY.et al. (2016). Early assessment of the therapeutic response to radiofrequency ablation for hepatic local lesion with 3-dimensional contrast-enhanced ultrasound registration system: a preliminary study. Chin. J. Ultrasonography25 (6), 502–505. doi: 10.3760/cma.j.issn.1004-4477.2016.06.010
Summary
Keywords
ultrasound, contrast-enhanced ultrasound, safety, effectiveness, child
Citation
Mao M, Xia B, Chen W, Gao X, Yang J, Li S, Wang B, Mai H, Liu S, Wen F, Gan Y, Song J, Wei H, Yang W, Wu Y, Yang S, Yu W, Yu H, Fan S, Tao H, Feng X, Lin Z and Liu L (2019) The Safety and Effectiveness of Intravenous Contrast-Enhanced Sonography in Chinese Children—A Single Center and Prospective Study in China. Front. Pharmacol. 10:1447. doi: 10.3389/fphar.2019.01447
Received
22 July 2019
Accepted
12 November 2019
Published
05 December 2019
Volume
10 - 2019
Edited by
Francis Kalemeera, University of Namibia, Namibia
Reviewed by
Joao Massud, Independent Researcher, São Paulo, Brazil; Adina Turcu-Stiolica, University of Medicine and Pharmacy of Craiova, Romania
Updates
Copyright
© 2019 Mao, Xia, Chen, Gao, Yang, Li, Wang, Mai, Liu, Wen, Gan, Song, Wei, Yang, Wu, Yang, Yu, Yu, Fan, Tao, Feng, Lin and Liu.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Bei Xia, xiabeimd@Gmail.com
This article was submitted to Pharmaceutical Medicine and Outcomes Research, a section of the journal Frontiers in Pharmacology
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.