Investigation of PolyVinyl Chloride Plastisol Tissue-Mimicking Phantoms for MR- and Ultrasound-Elastography

www.frontiersin.orgSimon Chatelin1* www.frontiersin.orgElodie Breton1 www.frontiersin.orgAjeethan Arulrajah1,2 www.frontiersin.orgCéline Giraudeau2 www.frontiersin.orgBenoit Wach1 www.frontiersin.orgLaurence Meylheuc1,3 www.frontiersin.orgJonathan Vappou1
  • 1lCube, Universit de Strasbourg, CNRS UMR 7357, Strasbourg, France
  • 2IHU, Institut Hospitalo-Universitaire, Institute of Image‐Guided Surgery, Strasbourg, France
  • 3INSA, Institut National de Sciences Appliquées, Strasbourg, France

Objective: Realistic tissue-mimicking phantoms are essential for the development, the investigation and the calibration of medical imaging techniques and protocols. Because it requires taking both mechanical and imaging properties into account, the development of robust, calibrated phantoms is a major challenge in elastography. Soft polyvinyl chloride gels in a liquid plasticizer (plastisol or PVCP) have been proposed as soft tissue-mimicking phantoms (TMP) for elasticity imaging. PVCP phantoms are relatively low-cost and can be easily stored over long time periods without any specific requirements. In this work, the preparation of a PVCP gel phantom for both MR and ultrasound-elastography is proposed and its acoustic, NMR and mechanical properties are studied.

Materials and methods: The acoustic and magnetic resonance imaging properties of PVCP are measured for different mass ratios between ultrasound speckle particles and PVCP solution, and between resin and plasticizer. The linear mechanical properties of plastisol samples are then investigated over time using not only indentation tests, but also MR and ultrasound-elastography clinical protocols. These properties are compared to typical values reported for biological soft tissues and to the values found in the literature for PVCP gels.

Results and conclusions: After a period of two weeks, the mechanical properties of the plastisol samples measured with indentation testing are stable for at least the following 4 weeks (end of follow-up period 43 days after gelation-fusion). Neither the mechanical nor the NMR properties of plastisol gels were found to be affected by the addition of cellulose as acoustic speckle. Mechanical properties of the proposed gels were successfully characterized by clinical, commercially-available MR Elastography and sonoelastography protocols. PVCP with a mass ratio of ultrasound speckle particles of 0.6%–0.8% and a mass ratio between resin and plasticizer between 50 and 70% appears as a good TMP candidate that can be used with both MR and ultrasound-based elastography methods.

Introduction

Over the past 3 decades, different methods have been developed for tissue elasticity measurement using medical imaging. All elastography approaches rely on the encoding of tissue displacement as a result of a force field that can be either external or internal, static or dynamic [1]. Displacements can be encoded by using medical imaging, such as ultrasound imaging, Magnetic Resonance Imaging (MRI) or optical imaging. Tissue-mimicking phantoms (TMP) are of primary importance during the development, validation and calibration processes in elasticity imaging [2, 3] and during operator training phases. Research-based and commercially-available TMP are today proposed for elastography, and are mostly dedicated to one specific elastography modality [4].

Ideally, elastography-dedicated TMP are expected to offer the following features: 1) Their mechanical properties (such as elasticity, viscosity, anisotropy, porosity or hyperelasticity) must be well controlled and must lie within typical values of soft tissues they are supposed to mimic; 2) They should offer particular ease of use in terms of storage conditions and durability; 3) They must be compatible with the medical imaging modality for which they have been developed. Several studies have proposed elastography TMP that display particular mechanical properties beyond linear elasticity, such as viscosity [43, 65, 74], poroelasticity [5], anisotropy [610], hyperelasticity [1113], heterogeneity [1419] and TMP including dynamic flow pulsations [20]. Inclusion of anisotropy [9] or porosity could be obtained through the addition of fibrin fibers or through the use of 3D-printing, [21, 22]. Important features that need to be accounted for are the preservation process, the stability over time and the inhalation toxicity during manufacturing. Depending on the imaging method, elastography TMP are mainly composed of a matrix, solvents and other additives [3]. Studies comparing elastography measurements obtained with different imaging modalities have illustrated the need for multi-modality elastography TMP [23, 24].

Synthetic phantoms have very interesting properties and are widely used as TMP. The most commonly used synthetic TMP are styrene-ethylene/butylene-styrene (SEBS) [2, 25] and silicone-based materials (more specifically Ecoflex gels) [74, 7880]. Soft PolyVinyl Chloride (PVC) suspensions in a liquid plasticizer (plastisol or PVCP) have also been proposed as tissue-mimicking candidates for elasticity imaging [15, 26, 29, 37, 45, 69, 74, 81, 82]. These TMP are good candidates for mimicking different stages of liver fibrosis, by varying their mass ratio between resin and plasticizer MRes/Plast from 40 to 60% [26]. Soft PVCP has been proposed in robotics as TMP for needle insertion [27, 28], and also for Magnetic Resonance Elastography (MRE) [15, 24, 26, 45, 69, 81, 83, 84], sonoelastography [26, 29] and biomedical photoacoustic [3035]. PVCP has been previously compared to silicone-based Ecoflex [24] and has been shown to be a very interesting candidate for MRE in terms of MRI and dynamic viscoelastic properties [3638]. In addition to their relatively low cost, PVCP TMP are simple to prepare, stable over time at room temperature and resistant to damage caused by typical handling procedures.

However, whether the same phantoms can be used simultaneously for sonoelastography and MRE remains poorly known. This study aims at investigating several key properties of the same PVCP phantoms for application to MRE and sonoelastography. In particular, the influence of added particles for ultrasound speckle on acoustic properties, MRI relaxation times and mechanical properties is investigated. Long term stability in mechanical properties is investigated over 43 days, and phantoms with different resin and speckle concentrations are tested using commercial MR and ultrasound elastography systems.

Materials and Methods

Description of Polyvinyl Chloride Plastisol Samples

PVCP is widely used in industry, mainly textile, automobile and aeronautics. In this study, we consider its use specifically for MR and ultrasound elasticity imaging. PVCP is a combination of a PVC colloidal suspension in a liquid ester plasticizer, the role of which is to soften the final material. These gels are formed through the gelation-fusion process [39]. In the current study, after mixing the PVC resin suspension (40–80%) and the plasticizer (a Bis(2-ethylhexyl) adipate) (Plastileurre Soft and Softener, Bricoleurre, Mont-Saint-Aignan, France), the solution is heated. Complete gelation-fusion requires to reach a temperature of 160°C while the thermal degradation limit of PVCP is 190°C [40], hence the curing temperature should remain between 160°C and 190°C. Curing temperature can be achieved for instance through hot plate heating in an open beaker [28] or through bain-marie oil bath heating [31, 33, 35]. In this study, the solution is heated in an open glass beaker by means of a microwave oven (800W) with regular stirring (total heating duration of about 8 min including 5 s stirring every 2 min for the first 6 min and then every 20 s for the last 2 min) until reaching a target temperature of 180°C. Cellulose particles (Sigmacell Cellulose, Cotton linters type 50, 50 µm, SigmaAldrich, Saint Louis, MI, United states) are added as ultrasound speckle particles, once the gel has cooled down to a temperature of 80°C in order to avoid thermal degradation of cellulose. The gel is stirred with mass concentrations of 0, 0.6, 0.8, and 1% of cellulose particles. The solution is then degassed in a vacuum bell for 5 min and, poured into cylindrical molds. This process was repeated in order to make cylindrical samples of two different sizes: diameter 60 mm and height 30 mm for investigation of the imaging properties (ultrasound attenuation and MR relaxation times) and diameter 98 mm and height 75–90 mm for investigation of the mechanical properties (indentation, MRE and sonoelastography). The mass ratio between resin and plasticizer (mass fraction noted MRes/Plast) as well as mass ratio between cellulose and PVCP solution (mass fraction noted MCell/Plast) can influence the mechanical and imaging properties of the samples, as hereafter investigated. Each sample is tested at the earliest 1 day after preparation (day #1).

Influence of Acoustic Speckle on the Imaging Properties of Polyvinyl Chloride Plastisol

Influence on the Ultrasound Attenuation

The acoustic attenuation of PVCP samples is investigated using the method proposed in [41]. The reader is referred to Appendix A for more details.

In the present study, we consider the acoustic attenuation properties of PVCP at a single nominal frequency of 1 MHz. The density of water, the speed of sound in water, the density of the specimen (regardless of mass ratio between resin and plasticizer MRes/Plast, in accordance with [42], who used PVCP from the same manufacturer) and the speed of sound in the specimen are assumed to have values of 1,000 kg.m−3, 1,500 m.s−1, 1,000 kg.m−3 and 1,400 m.s−1, respectively (values from Ref. 28, who used PVCP from the same manufacturer, and in accordance with [30, 32, 40, 43]). The values for the density and speed of sound are close to those of most biological soft tissues (1,050 kg.m−3 and 1,578 m.s−1 for liver, 1,041 kg.m−3 and 1,550–1,630 m.s−1 for muscle, 1,035 kg.m−3 and 1,562 m.s−1 for brain, 928 kg.m−3 and 1,430–1,450 m.s−1 for fat [44, 46]).

The acoustic test bench is composed of a single element ceramic focused transducer (23 mm diameter, nominal frequency of 1 MHz, PA-765, Precision Acoustics Ltd, Dorchester, United Kingdom) driven by a digital wave generator (33210A, Agilent Technologies Inc., Santa Clara, CA, United States), a 0.2 mm needle hydrophone (SN2319, Precision Acoustics Ltd, Dorchester, United Kingdom) connected in series with a preamplifier and an oscilloscope (TDS 2002B, Tektronix Inc., Beaverton, OR, United States). More details on both the measurement method and the experimental set-up are proposed and illustrated in the Appendix A.

Influence on the Nuclear Magnetism Relaxation Times

The potential of PVCP as a TMP for MRE is assessed through the evaluation of its NMR relaxation times (T1 and T2). The influence of the mass ratio between resin and plasticizer MRes/Plast and of the cellulose (ultrasonic speckle) mass ratio MCell/Plast on the NMR relaxation times is evaluated.

The acquisitions are performed in a 1.5 T MRI scanner (MAGNETOM Aera, Siemens AG, Erlangen, Germany) on samples with mass ratio between resin and plasticizer MRes/Plast of 40, 50, 60, and 70%. The additional influence of ultrasonic speckle on the NMR relaxation times is also investigated by adding 0.6, 0.8, and 1%-concentrated cellulose. Overall, 16 samples are imaged.

The spin-lattice relaxation time T1 at 1.5 T is evaluated using a turbo spin echo sequence (TE/TR 6/3,000 ms, turbo factor 7, image acquisition time 97 s) with selective inversion recuperation preparation pulse and varying inversion times (TI) of 23, 50, 100, 150, 200, 250, 300, 350, 400, 500, 600, 800, 1,000, 1,500, 2,000, and 2,970 ms. The spin-spin relaxation time T2 is evaluated using a spin echo sequence (TR 2500 ms, partial Fourier 6/8, acquisition time 363 s) with varying echo times (TE) 3.5, 5, 10, 15, 20, 30, 40, 50, 60, 80, 100, 150, and 200 ms. Common relevant imaging parameters for both sequences are: matrix 192 x 192, field of view 340 mm × 340 mm, slice thickness 6 mm, bandwidth 789 Hz.px−1.

Mechanical Properties

Based on the observations reported in the previous sections, standard MRE and sonoelastography (Acoustic Radiation Force Imaging–ARFI) protocols used in vivo in clinical practice are investigated in the same TMP and compared to a reference mechanical testing approach, namely, indentation. The mechanical stability over time of the gels is reported using these three modalities at day #1 and day #43 after the beginning of the gelation-fusion process. The mass fraction of cellulose particles to PVCP solution MCell/Plast is fixed to 0.6% based on previous measurements of the acoustic attenuation.

Indentation Measurements

The linear elastic behavior is characterized against mass ratio between resin and plasticizer MRes/Plast from 50 to 80% with 10% steps.

The acquisitions are performed using a home-made dedicated indentation set-up (20 mm-diameter hemispherical shaped indenter mounted on a 1 degree-of-freedom translation motor and a force sensor) at room temperature (22°C) on cylindrical-shaped samples (98 mm and 75–90 mm in diameter and height, respectively). They are indented up to 11 mm depth at a speed of 7.6 mm.s−1 (corresponding to strain rates between 0.08 and 0.1 s−1). For this range of deformations, the mechanical behavior of PVCP can be considered linear [28]. A scheme of the experimental set-up is available in the Appendix B. The elastic modulus E is computed using the continuous stiffness measurement [47] method and the Sneddon relationship [48], as described in Ref. 49. In order to investigate the capability of PVCP to mimic biological soft tissues for shear wave elastography and to compare with the results from MRE and ARFI measurements, the shear modulus G is deduced from: G=E/2(1+ν)E/3. Each measurement is repeated five times and the mean values and standard deviations over these five measurements are reported.

In order to evaluate the stability over time of mechanical properties, indentation method is performed in the same samples over a period of 43 days (every 3–14 days) after the beginning of the gelatin-fusion process. The temperature is constant and equal to T0 ≈ 22°C for both storage and testing conditions.

Acoustic Radiation Force Imaging Measurements

Shear wave velocity cs measurements are performed using Siemens Acuson S3000 ultrasound imager together with a linear 9-L4 ultrasound transducer probe in the virtual touch quantification (VTq) ARFI mode (Siemens Medical Inc., Mountain View, CA, United States). For each sample, 10 measurements are performed and averaged, distributed along two lines at depths of 15 and 30 mm. To allow multi-modality comparison, the shear modulus is considered as G=ρcS2.

Magnetic Resonance Elastography Measurements

MRE acquisitions are performed on a 1.5 T MRI scanner (MAGNETOM Aera, Siemens AG, Erlangen, Germany) using a standard in vivo MRE protocol with motion sensitizing gradient. The mechanical waves are generated using a commercial pneumatic driver system (Resoundant®, Mayo Clinic Foundation, Rochester, MN, United States). Scanning parameters, excitation frequency f of 60.1 Hz, echo time/repetition time (TE/TR) = (14.47 ms/50 ms), flip angle 25°, slice thickness 7 mm, acquisition matrix 128 by 102, reconstruction matrix 256 by 204, resolution 1.5625 mm by 1.5625 mm, and 1 slice parallel to the vibrating plate. Assuming pure elasticity and that the stiffness is defined using the scalar shear modulus G=ρ(λf)2 (λ being the wavelength), the mechanical parameters are estimated using the local frequency algorithm [50]. The results are given in terms of mean shear modulus and their standard deviation in cylindrical regions of interest (diameter 80 mm) at the center of each phantom.

ARFI and MRE measurements were performed at day #1 (beginning of the gelatin-fusion process) and day #43. A total of 5 gels with mass ratio between resin and plasticizer varying from 40% to 80% with 10% steps was tested with both MRE and ARFI sonoelastography.

Results

Influence of Acoustic Speckle on the Ultrasound Attenuation

Depending on MRes/Plast, the attenuation values obtained at 1 MHz vary from 0.146 to 0.381, 0.458 to 0.817, 0.665 to 1.251, and 0.827 to 1.641 dB.cm−1 for MCell/Plast of 0, 0.6, 0.8, and 1%, respectively. The values are reported in Figure 1 (and further detailed in Appendix C). The results are systematically compared to typical values found in the literature not only for biological soft tissues (for instance 0.45 dB.cm−1 for liver, 0.5–1.5 dB.cm−1 for muscle, 0.58 dB.cm−1 for brain, 0.6–0.8 dB.cm−1 for fat at 1 MHz [44]) but also for similar PVCP TMP, most often over a wider frequency range [30, 32, 33, 35, 40, 51, 52]. With the exception of the values obtained for a mass ratio of resin to plasticizer MRes/Plast of 50%, the acoustic attenuation values increase with MRes/Plast. This is consistent with the observations found in the literature [32].

FIGURE 1
www.frontiersin.org

FIGURE 1. The acoustic attenuation coefficients measured at 1 MHz in the tested Plastisol phantoms (colored signs) are compared to typical values from the literature for biological soft tissues (green colored area) [44] and to similar PVCP TMP (gray and black curves) [30, 32, 33, 35, 40, 51, 52] over the acoustic frequency range 0.5–3 MHz.

Influence of Acoustic Speckle on the Nuclear Magnetism Relaxation Times

Relaxation times T1 and T2 values evaluated at 1.5 T as a function of PVC and cellulose concentrations are reported in Figure 2 (and further detailed in Appendix C). The presence of ultrasonic speckle does not appear to consistently affect T1 and T2 values of PVCP. T2 values (averaged over all cellulose mass ratios) decrease by 10% (50 ms–44 ms) between mass ratio between resin and plasticizer MRes/Plast from 40 to 50%; for MRes/Plast from 50% to 70%, no further change in T2 is observed. The T1 values of PVCP (averaged over all cellulose mass ratios) decrease from 258 to 223 ms with increasing mass ratio between resin and plasticizer MRes/Plast 40–70%. The results obtained at 1.5 T are systematically compared to typical values found in the literature for similar PVCP TMP, most often at higher magnetic field, such as at 3 T [38] and at 7 T [36]. While T2 values of PVCP are slightly shorter than those of healthy soft tissue at 1.5 T (54 ± 8 ms for liver, 35 ± 4 ms for muscle, 75–90 ms for brain, 90 ms for fat), T1 values of PVCP are significantly shorter (600 ms for liver, 1,060 ± 155 ms for muscle, 500–750 ms for brain, 200 ms for fat) [5358].

FIGURE 2
www.frontiersin.org

FIGURE 2. Longitudinal T1 and transverse T2 NMR relaxation times acquired at 1.5 T for varying mass ratio between cellulose and PVCP MRes/Plast and mass ratio between resin and plasticizer MRes/Plast. These values are compared to T1 and T2 values from the literature measured in similar PVCP TMP (not containing cellulose) at 3 [38] and 7 T [36] (gray crosses).

Mechanical Properties of the PVCP Phantom

The mechanical values obtained from indentation measurements up to 43 days after the beginning of the gelatin-fusion process are shown in Figure 3A. After a period of gelation-fusion of approximately 14 days, the linear mechanical properties are found to be stable for mass ratio between resin and plasticizer MRes/Plast 50–80%. These observations are consistent with the results obtained by MRE and dynamic mechanical analysis in [37]. After a 2-weeks stabilization period, PVCP appears to be mechanically stable over time (for at least up to 4 weeks) at room temperature, without any specific storage conditions, such as immersion or moistening.

FIGURE 3
www.frontiersin.org

FIGURE 3. The shear modulus against mass ratio resin to plasticizer is measured with indentation tests over a period of 43 days after gelation-fusion in order to study the stability of the PVCP gels in terms of mechanical properties (A). Shear moduli measured directly after gelation-fusion process (purple) and after stabilization (orange) are represented against mass ratio between resin and plasticizer, and are compared to liver [64] and brain [5963] values from the literature (B).

The mechanical values obtained from indentation measurements on day #1 of the gelatin-fusion process (purple) or averaged over all measurements obtained after day #14 once mechanical stabilization of the phantom is observed (orange) are reported in Figure 3B and in Appendix C (for the immediate and long-term elasticity). The values are compared to typical values available from the literature for both brain [5963] and liver tissues [64]. PVCP with a mass ratio MRes/Plast of 70% is an adequate candidate to mimic the elastic shear behavior of the liver tissue at small strain.

In accordance with the mechanical time-evolution observed from indentation measurements, the ARFI and MRE acquisitions are performed directly after (1 day) and 43 days after beginning of the gelatin-fusion process. The same increase over time is observed as with indentation measurements. MRE results are presented in Figure 4 and in Appendix C. The results are given in terms of mean shear modulus and their standard deviation in cylindrical regions of interest. The values are very close to those found by indentation.

FIGURE 4
www.frontiersin.org

FIGURE 4. Stiffness values measured directly (“Initial”) and 43 days (“Long term”) following the gelation-fusion process with MRE, ARFI and indentation testing. Stiffness ranges for liver [64] and brain [5963] found in the literature are represented for comparison.

The results from ARFI mechanical measurements are presented in Figure 4 and in Appendix C and compared to indentation measurements. The values at day 1 are higher than those obtained by indentation (differences of 52, 93, 21, and 18% for MRes/Plast of 50, 60, 70, and 80%, respectively) and MRE (differences of 3, 49, 14, and 9% for MRes/Plast of 50, 60, 70, and 80%, respectively). The same trend is observed at day 43 with values from ARFI measurements higher than those obtained by indentation (differences of 83, 23, 9, and 4% for MRes/Plast of 50, 60, 70, and 80%, respectively) and MRE (differences of 32, 22, 12, and 2% for MRes/Plast of 50, 60, 70, and 80%, respectively). The higher the resin mass ratio is, the closest the results are between the different methods.

General Discussion

The use of a single PVCP TMP for MRE and sonoelastography is investigated in this study. From our results, plastisol appears as a good candidate for mimicking soft tissues in terms of mechanical properties with T1 and T2 MR imaging values compatible with typical MRE pulse sequences. In addition, the ultrasound speckle can be adjusted by adding cellulose to PVCP without altering significantly its NMR properties. The current study provides indications for easily preparing PVCP phantoms for both MR and ultrasonography elastography, with readily available instruments, i.e., a microwave and a thermometer. Mechanical properties measured with indentation were found to stabilize 14 days after gelation-fusion, and remain mechanically stable until the end of our follow-up period of 43 days for PVCP gels with resin to plasticizer ratios of 50–80% and containing 0–1% mass ratio cellulose.

Many advantages for the use of PVCP in elastography phantoms can be listed, as attested by the current study: 1) short preparation time and ease-of-use; 2) fast solidification process, avoiding sedimentation of the acoustic speckle particles; 3) natural transparence to ultrasound, making it easy to control wave diffusion by adjunction of speckle particles; 4) long conservation time without any specific storage requirements, such as moistening or water bath required for other hydrogels such as PVA. Elasticity at small strains is simply controlled by varying the mass ratio between resin and plasticizer and consequently the PVC concentration. These specificities make it possible to combine both rheological and imaging properties close to those of biological soft tissues in a calibrated, robust TMP.

As previously reported, the acoustic attenuation of PVCP can mimic those of biological soft tissues through the addition of cellulose speckle particles. The results at 1 MHz are in the typical range of biological organs (0.14–1.16 dB.cm−1 at 1 MHz [51]) and a cellulose concentration of 0.6%–0.8% was found to mimic the acoustic attenuation of soft tissues. The mass ratio between resin and plasticizer MRes/Plast does not significantly influence the acoustic attenuation properties for concentrations higher than 50% in the studied PVCP. Limitations of this study in terms of measurement of the ultrasonic properties must be mentioned. First, the attenuation coefficient of PVCP TMP is characterized in this work at a single frequency (1 MHz), that lies below the center frequency of most clinical ultrasound transducers. How the studied PVCP TMP remains realistic in terms of acoustic attenuation over a wider frequency range remains unaddressed in this study. However, the values measured at 1 MHz agree with those found in the literature over a larger range of frequencies, as illustrated in Figure 1. Second, the density and speed of sound used in this study were taken from the literature [28, 30, 32, 33, 35, 40, 42, 43, 51, 52]. They are here assumed to be the same for all the mass ratios of resin to plasticizer.

NMR relaxation times of the studied PVCP phantoms were found to slightly depend on the mass ratio between resin and plasticizer. None of the tested PVCP composition reproduces both the T1 and T2 at 1.5 T of any biological soft tissues. The typical T2 value of plastisol (45 ms) lies within the lower range of T2 for biological soft tissues, including the liver and the heart. The T1 values (220–260 ms) found for plastisol at 1.5 T are lower than those of most soft tissues. These relaxation times are different than those reported in the literature in PVCP without cellulose addition [36, 38], but direct comparison is not straightforward as measurement were performed at higher magnetic fields (3 and 7 T, respectively) on different PVCP. However, the objective of this study was to evaluate whether the studied PVCP is a good TMP candidate for typical MRE acquisition sequences, rather than to reproduce NMR relaxation times of biological soft tissues. Since MRE relies on the use of phase images alone for the estimation of elasticity, and not on the T1/T2 weighted magnitude images, the studied PVCP can be used with typical MRE acquisition sequences, including the clinical hepatic MRE protocol, as demonstrated here.

PVCP gels are good surrogate for biological soft tissues from a mechanical point of view. Despite the same samples being tested across the different modalities, small differences in the stiffness values can be observed between MRE, ARFI and indentation measurements. These can be explained by the different physical protocols and assumptions between the three methods (harmonic shear wavelength, quasi-static compression and impulse shear wave velocity for MRE, indentation and ARFI, respectively). Consequently, the so-called “stiffness” does not exactly correspond to the same parameter under the same conditions. For instance, the frequency and strain rate ranges are intrinsic to the aforementioned methods and do not necessarily coincide. Previous investigations from the literature suggest that PVCP gels are likely to exhibit relatively low viscosity at the usual elastography frequency range and therefore moderate dependence of the elasticity with frequency [69, 74]. Due to different experimental strain rate (or frequency), the viscosity could be a possible reason for the differences observed between indentation, ARFI and MRE measurements. It appears that the higher the resin concentration is, the closest the results are between the different methods. This could be explained by the fact that viscous effects are less predominant compared to pure elastic effects at higher concentrations, thus resulting in decreased shear wave dispersion and therefore less sensitivity to differences in mechanical frequency content. If needed, additional investigation should be carried out in order to evaluate whether including additional compounds could increase PVCP viscosity, as oil does in gelatin phantoms [65].

The issue of cross-validation between elastography and rheological measurements or between different elastography methods is well known [18, 23, 24, 37, 62, 6669]. As previously introduced, differences in frequency range and excitation modes–harmonic vs transient-make any direct comparison particularly difficult. PVCP with added cellulose offers the possibility of cross validating indentation with MR and US elastography methods. PVCP with added cellulose is a good, stable TMP for methods that require simultaneously both MR imaging and US attenuation for internally-generated displacements, such as MR-ARFI [7072].

The stability of the mechanical properties of a phantom over time is of great importance for the development, the validation and the comparison of elastography protocols. For instance, extending the shelf-life and the mechanical stability of gelatin and agar-agar hydrogels is challenging and requires adequate storage conditions and addition of preservative, fungicides or bactericide agents [73]. The plastisol TMP of this study were found to have stable mechanical properties, as measured with indentation testing, from day 14 until day 43 (end of follow-up period) after being manufactured. Over a longer follow-up period, the mechanical properties of a PVCP gel were found stable during storage, up to six months [31].

SEBS, PDMS and silicon exhibit similarly high shelf life with no specific storage constraints. Some silicone-based materials (such as the silicone-based Ecoflex gels [74]) share other similar properties and advantages with PVCP, and a similar characterization study of their mechanical and MR/US imaging properties would be of interest. Silicone and PDMS gels have relatively similar imaging, mechanical and storage characteristics as PVCP. Indeed, they all provide a solid support to which acoustic scatterers can be easily added, they are insoluble in water, remain stable during storage and have easily controllable properties (first of all their stiffness) [32]. While the sound velocity and acoustic attenuation values of PVCP are slightly lower and higher, respectively, than those of biological soft tissues, these limitations are even greater for silicone and PDMS, with values close to 1,000 m.s−1 [46] and over 9.8 dB.cm−1 at 3 MHz [75], respectively. However, unlike silicone and PDMS gels, the preparation of PVCP gels requires curing at a given temperature range. In addition, access to the exact chemical composition of PVCP (PVC concentration, additives … ) [32] is manufacturer-dependent, which means that any direct comparison between different studies on PVCP should be performed with care. Another oil-based material, SEBS has been shown to mimic adequately acoustic (speed of sound between 1,420 and 1,464 m.s−1 and acoustic attenuation between 0.4 and 4 dB.cm−1 at 3.5 MHz, depending of the speckle particle density [25, 76]) and mechanical properties of soft tissues. Similar to PVCP gels in this study, SEBS gels were shown to be suitable TMPs for both MRE and ultrasound elastography [23], stable over time. However, the use of PVCP TMP is characterized by the simplicity and speed of its preparation process.

Conclusion

This study provides indications about how to prepare a mechanically-stable phantom using plastisol in order to mimic biological soft tissues for MRE and/or sonoelastography. On the basis of their mechanical properties, PVCP phantoms are demonstrated to be realistic TMP with NMR relaxation times compatible with MRE. With the addition of cellulose particles, it is possible to develop a calibrated TMP for both MRE and sonoelastography measurements. Mechanical properties of the tested plastisol were found stable at room temperature after 2 weeks and at least until 6 weeks after gelation-fusion. For instance, a 60–70% mass ratio between resin and plasticizer MRes/Plast sample will mimic the linear elastic behavior of liver tissue correctly. In MRE, the T2 values of the tested PVCP TMP are close to those of hepatic tissue, while with addition of 0.6% of cellulose particles, this gel becomes also a good liver-mimicking candidate in sonoelastography. Investigating the mechanical properties over a larger range of frequency using dynamic mechanical analysis [37, 69] would be a great extension for this study.

Data Availability Statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Author Contributions

SC, EB and JV wrote the manuscript. The PVCP TMP were prepared by SC. The indentation tests were developed and performed by SC, AA, LM, BW and JV. The acoustic test bench was developed, and the acoustic properties were investigated by SC, BW and JV. The NMR properties of the TMP were investigated by SC, EB and CG. SC performed final approval of the manuscript to be published and agreed to be accountable for all aspects of the work related to the accuracy and integrity. All authors agree to be accountable for the content of the work.

Funding

This work has benefitted from support of the ANR (Agence Nationale de la Recherche) by the French national program “Investissements d’Avenir”: 1) by the IDEX-Unistra (ANR-10-IDEX-0002–02) with the support of the IRIS technology platform; 2) by the LABEX-CAMI (ANR-11-LABX-004); 3) by the IHU Strasbourg (Institute of Image Guided Surgery, ANR-10-IAHU-02).

Conflict of Interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Acknowledgments

The authors would like to thank the staff of the imaging platform of the IHU-Strasbourg for their support and availability and the IRIS technological platform of the ICube laboratory for technical support.

Appendix A

Measurement of the Acoustic Attenuation Coefficient

FIGURE 5
www.frontiersin.org

FIGURE 5. llustration of the acoustic test bench for evaluation of the acoustic attenuation coefficient.

Measurement of the Acoustic Attenuation Coefficient.

The acoustic attenuation coefficient is measured at 1 MHz for PVCP samples with varying mass ratio between cellulose and PVCP MRes/Plast using the set-up illustrated in Figure 5.

The acoustic attenuation of PVCP samples is investigated using the method proposed in Ref. 41. The Fourier transform of a single-frequency plane wave propagating over two different media (water over a distance zw and a specimen over a depth z–height of the sample) can be modeled as:

U(f)=A(f)e=U0(f)e(αw+iβw)zwe(α+)zT(f)(1)

where U0(f), f, βw, αw, β, and α are the Fourier transform of the initial generated pulse wave, its frequency, the propagation and the attenuation factors of water and the specimen, respectively.

T is the overall water-specimen transmission factor defined by:

T(f)=4ρc(f)ρwcw(ρc(f)+ρwcw)2(2)

with ρw, cw, ρ, and c density and speed of sound in water, the density of the specimen and the phase velocity, respectively. According to [41], the attenuation coefficient is deduced from the amplitude spectra of the transmitted pulses with (A) or without (Aw) the specimen:

αln(T)z+1zln(AwA)(3)

The amplitude drop is mostly expressed in dB by defining: αdB=z{20log10(eαz)}=8.6886α. The attenuation coefficient of a material is generally dependent on the frequency f of the ultrasound waves. A power law αdB=afγ can be assumed for this dependency [44, 77].

Appendix B

FIGURE 6
www.frontiersin.org

FIGURE 6. Illustration of the indentation set-up.

Appendix C

Mechanical and imaging properties of PVCP.

The main results obtained in this work on PVCP TMP are summarized in Table 1 (mean and standard deviation values) for all tested properties (indentation, echographic, and MRI properties).

TABLE 1
www.frontiersin.org

TABLE 1. Mechanical, ultrasonic and MRI properties of the tested PVCP.

References

1. Vappou J. Magnetic Resonance− and ultrasound Imaging−Based elasticity imaging methods: a review. Crit Rev Biomed Eng (2012). 40(2):121–34. doi:10.1615/CritRevBiomedEng.v40.i2.30

CrossRef Full Text | Google Scholar

2. Oudry J, Lynch T, Vappou J, Sandrin L, Miette V. Comparison of four different techniques to evaluate the elastic properties of phantom in elastography: is there a gold standard? Phys Med Biol (2014). 59(19):5775–93. doi:10.1088/0031-9155/59/19/5775

CrossRef Full Text | Google Scholar

3. Cao Y, Li G-Y, Zhang X, Liu Y-L. Tissue-mimicking materials for elastography phantoms: a review, Extreme Mech Letters (2017). 17:62–70. doi:10.1016/j.eml.2017.09.009

CrossRef Full Text | Google Scholar

4. Cournane S, Fagan A, Browne JE. Review of ultrasound elastography quality control and training test phantoms. Ultrasound (2012). 20:16–23.

5. Chaudhry A, Yazdi IK, Kongari R, Tasciotti E, Righetti R. A new class of phantom materials for poroelastography imaging techniques. Ultrasound Med Biol (2016). 42(5):1230–8. doi:10.1016/j.ultrasmedbio.2015.12.013

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Namani R, Wood MD, Sakiyama-Elbert SE, Bayly PV. Anisotropic mechanical properties of magnetically aligned fibrin gels measured by magnetic resonance elastography. J Biomech (2009). 42(13):2047–53. doi:10.1016/j.jbiomech.2009.06.007

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Qin EC, Sinkus R, Geng G, Cheng S, Green M, Rae CD, et al. Combining MR elastography and diffusion tensor imaging for the assessment of anisotropic mechanical properties: a phantom study. J Magn Reson Imag (2013). 37(1):217–26. doi:10.1002/jmri.23797

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Aristizabal S, Amador C, Qiang B, Kinnick RR, Nenadic IZ, Greenleaf JF, et al. Shear wave vibrometry evaluation in transverse isotropic tissue mimicking phantoms and skeletal muscle. Phys Med Biol (2014). 59(24):7735–52. doi:10.1088/0031-9155/59/24/7735

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Chatelin S, Bernal M, Deffieux T, Papadacci C, Flaud P, Nahas A, et al. Anisotropic polyvinyl alcohol hydrogel phantom for shear wave elastography in fibrous biological soft tissue: a multimodality characterization. Phys Med Biol (2014). 59(22):6923. doi:10.1088/0031-9155/59/22/6923

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Schmidt JL, Tweten DJ, Benegal AN, Walker CH, Portnoi TE, Okamoto RJ, et al. Magnetic resonance elastography of slow and fast shear waves illuminates differences in shear and tensile moduli in anisotropic tissue. J Biomech (2016). 49(7):1042–9. doi:10.1016/j.jbiomech.2016.02.018

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Erkamp RQ, Emelianov SY, Skovoroda AR, O'Donnell M. Nonlinear elasticity imaging: theory and phantom study. IEEE Trans Ultrason Ferroelectrics Freq Contr (2004). 51(5):532–9. doi:10.1109/TUFFC.2004.1320826

CrossRef Full Text | Google Scholar

12. Gennisson J-L, Rénier M, Catheline S, Barrière C, Bercoff J, Tanter M, et al. Acoustoelasticity in soft solids: assessment of the nonlinear shear modulus with the acoustic radiation force. J Acoust Soc Am (2007). 122(6):3211–9. doi:10.1121/1.2793605

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Pavan TZ, Madsen EL, Frank GR, Adilton O Carneiro A, Hall TJ. Nonlinear elastic behavior of phantom materials for elastography. Phys Med Biol (2010). 55(9):2679–92. doi:10.1088/0031-9155/55/9/017

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Gao L, Parker KJ, Alam SK, Lerner RM. Sonoelasticity imaging: theory and experimental verification. J Acoust Soc Am (1995). 97(6):3875–86. doi:10.1121/1.412399

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Bishop J, Samani A, Sciarretta J, Plewes DB. Two-dimensional MR elastography with linear inversion reconstruction: methodology and noise analysis Phys Med Biol (2000). 45:2081–91. doi:10.1088/0031-9155/45/8/302

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Plewes DB, Bishop J, Samani A, Sciarretta J. Visualization and quantification of breast cancer biomechanical properties with magnetic resonance elastography. Phys Med Biol (2000). 45(6):1591–610. doi:10.1088/0031-9155/45/6/314

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Madsen EL, Hobson MA, Shi H, Varghese T, Frank GR. Tissue-mimicking agar/gelatin materials for use in heterogeneous elastography phantoms. Phys Med Biol (2005). 50(23):5597–618. doi:10.1088/0031-9155/50/23/013

CrossRef Full Text | Google Scholar

18. Green MA, Bilston LE, Sinkus R. In vivobrain viscoelastic properties measured by magnetic resonance elastography. NMR Biomed (2008). 21:755–64. doi:10.1002/nbm.1254

CrossRef Full Text | Google Scholar

19. Mariappan YK, Glaser KJ, Manduca A, Romano AJ, Venkatesh SK, Yin M, et al. High-frequency mode conversion technique for stiff lesion detection with Magnetic Resonance Elastography (MRE). Magn Reson Med. (2009). 62(6):1457–65. doi:10.1002/mrm.22091

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Kolipaka A, McGee KP, Araoz PA, Glaser KJ, Manduca A, Romano AJ, et al. MR elastography as a method for the assessment of myocardial stiffness: comparison with an established pressure-volume model in a left ventricular model of the heart. Magn Reson Med, (2009). 62(1):135–40. doi:10.1002/mrm.21991

PubMed Abstract | CrossRef Full Text | Google Scholar

21. Wan W, Bannerman AD, Yang L, Mak H. Poly(vinyl alcohol) cryogels for biomedical application. Adv Polym Sci (2014) 263:283–321

22. Guidetti M, Lorgna G, Hammersly M, Lewis P, Klatt D, Vena P, et al. Anisotropic composite material phantom to improve skeletal muscle characterization using magnetic resonance elastography. J Mech Behav Biomed Mater (2019). 89:199–208.

PubMed Abstract | Google Scholar

23. Oudry J, Chen J, Glaser KJ, Miette V, Sandrin L, Ehman RL. Cross-validation of magnetic resonance elastography and ultrasound-based transient elastography: a preliminary phantom study. J Magn Reson Imag (2009). 30:1145–50. doi:10.1002/jmri.21929

CrossRef Full Text | Google Scholar

24. Brinker ST, Kearney SP, Royston TJ, Klatt D. Simultaneous magnetic resonance and optical elastography acquisitions: comparison of displacement images and shear modulus estimations using a single vibration source. J Mech Behav Biomed Mater (2018). 84:135–44.

PubMed Abstract | Google Scholar

25. Oudry J, Bastard C, Miette V, Willinger R, Sandrin L. Copolymer-in-oil phantom materials for elastography. Ultrasound Med Biol (2009). 35(7):1185–97. doi:10.1016/j.ultrasmedbio.2009.01.012

PubMed Abstract | CrossRef Full Text | Google Scholar

26. Chakouch MK, Leclerc GE, Charleux F, Bensamoun SF. “Phantoms mimicking the viscoelastic behavior of healthy and fibrotic livers with Magnetic Resonance Elastography technique,” In 2015 International conference on advances in biomedical engineering (ICABME); Beirut, Lebanon (2015). p. 258–61.

Google Scholar

27. DiMaio SP, Salcudean SE. Needle insertion modeling and simulation. IEEE Trans Robot Autom (2003). 19(5):864–75. doi:10.1109/TRA.2003.817044 | Google Scholar

CrossRef Full Text | Google Scholar

28. Hungr N, Long J-A, Beix V, Troccaz J. A realistic deformable prostate phantom for multimodal imaging and needle-insertion procedures. Med Phys (2012). 39(4):2031–41. doi:10.1118/1.3692179

PubMed Abstract | CrossRef Full Text | Google Scholar

29. Montagnon E, Hadj-Henni A, Schmitt C, Cloutier G Rheological assessment of a polymeric spherical structure using a three-dimensional shear wave scattering model in dynamic spectroscopy elastography. IEEE Trans Ultrason Ferroelectrics Freq Contr (2014). 61(2):277–87. doi:10.1109/TUFFC.2014.6722613

CrossRef Full Text | Google Scholar

30. Spirou GM, Oraevsky AA, Vitkin IA, Whelan WM. Optical and acoustic properties at 1064 nm of polyvinyl chloride-plastisol for use as a tissue phantom in biomedical optoacoustics. Phys Med Biol (2005). 50(14):N141. doi:10.1088/0031-9155/50/14/N01

PubMed Abstract | CrossRef Full Text | Google Scholar

31. Bohndiek SE, Bodapati S, Van De Sompel D, Kothapalli S-R, Gambhir SS. Development and application of stable phantoms for the evaluation of photoacoustic imaging instruments. PloS One (2013). 8(9):e75533-14. doi:10.1371/journal.pone.0075533

CrossRef Full Text | Google Scholar

32 Fonseca M, Zeqiri B, Beard PC, Cox BT. Characterisation of a phantom for multiwavelength quantitative photoacoustic imaging. Phys Med Biol (2016). 61(13):4950–73. doi:10.1088/0031-9155/61/13/4950

PubMed Abstract | CrossRef Full Text | Google Scholar

33. Vogt WC, Jia C, Wear KA, Garra BS, Joshua Pfefer T. Biologically relevant photoacoustic imaging phantoms with tunable optical and acoustic properties. J Biomed Optic (2016). 21(10):101405. doi:10.1117/1.jbo.21.10.101405

PubMed Abstract | CrossRef Full Text | Google Scholar

34. Jia C, Vogt WC, Wear KA, Pfefer TJ, Garra BS. Two-layer heterogeneous breast phantom for photoacoustic imaging. J Biomed Optic (2017). 22(10):1. doi:10.1117/1.jbo.22.10.106011

PubMed Abstract | CrossRef Full Text | Google Scholar

35. Dantuma M, van Dommelen R, Manohar S. Semi-anthropomorphic photoacoustic breast phantom. Biomed Optic Express (2019). 10(11):5921. doi:10.1364/boe.10.005921

PubMed Abstract | CrossRef Full Text | Google Scholar

36. Li W, Belmont B, Greve JM, Manders AB, Downey BC, Zhang X, et al. Polyvinyl chloride as a multimodal tissue-mimicking material with tuned mechanical and medical imaging properties. Med Phys (2016). 43(10):5577–92. doi:10.1118/1.4962649

PubMed Abstract | CrossRef Full Text | Google Scholar

37. Arunachalam SP, Rossman PJ, Arani A, Lake DS, Glaser KJ, Trzasko JD, et al. Quantitative 3D magnetic resonance elastography: comparison with dynamic mechanical analysis. Magn Reson Med (2017). 77(3):1184–92. doi:10.1002/mrm.26207

CrossRef Full Text | Google Scholar

38. He Y, Qin S, Dyer BA, Zhang H, Zhao L, Chen T, et al. Characterizing mechanical and medical imaging properties of polyvinyl chloride‐based tissue‐mimicking materials. J Appl Clin Med Phys (2019). 20(7):176–83. doi:10.1002/acm2.12661

PubMed Abstract | CrossRef Full Text | Google Scholar

39. Nakajima N, Isner JD, Harrell ER, Daniels CA. Dependence of gelation and fusion behavior of poly(vinyl chloride) plastisols upon particle size and size distribution. Polym J (1981). 13(10):955–65. doi:10.1295/polymj.13.955

CrossRef Full Text | Google Scholar

40. Bakaric M, Miloro P, Zeqiri B, Cox BT, Treeby BE. The effect of curing temperature and time on the acoustic and optical properties of PVCP. IEEE Trans Ultrason Ferroelectrics Freq Contr (2020). 67(3):505–12. doi:10.1109/TUFFC.2019.2947341

PubMed Abstract | CrossRef Full Text | Google Scholar

41. He P, Zheng J. Acoustic dispersion and attenuation measurement using both transmitted and reflected pulses. Ultrasonics (2001). 39(1):27–32. doi:10.1016/S0041-624X(00)00037-8

CrossRef Full Text | Google Scholar

42. Chakouch M. Viscoelastic properties of in vivo thigh muscle and in vivo phantom using magnetic resonance elastography. [PhD thesis]. France: Université de Technologie de Compiègne (2016).

Google Scholar

43. Madsen EL, Frank GR, Krouskop TA, Varghese T, Kallel F, Ophir J. Tissue-mimicking oil-in-gelatin dispersions for use in heterogeneous phantoms. Ultrasonic Imaging (2003). 25(1):17–38. doi:10.1177/016173460302500102

CrossRef Full Text | Google Scholar

44. Duck FA. Chapter 4 - acoustic properties of tissue at ultrasonic frequencies. London, Academic Press (1990). p 73–135. | Google Scholar

CrossRef Full Text | Google Scholar

45. Leclerc GE, Debernard L, Foucart F, Robert L, Pelletier KM, Charleux F, et al. Characterization of a hyper-viscoelastic phantom mimicking biological soft tissue using an abdominal pneumatic driver with magnetic resonance elastography (MRE). J Biomech (2012). 45(6):52–7. doi:10.1016/j.jbiomech.2012.01.017

PubMed Abstract | CrossRef Full Text | Google Scholar

46. Culjat MO, Goldenberg D, Tewari P, Singh RS. A review of tissue substitutes for ultrasound imaging. Ultrasound Med Biol (2010). 36(6):861–73. doi:10.1016/j.ultrasmedbio.2010.02.012

PubMed Abstract | CrossRef Full Text | Google Scholar

47. Oliver WC, Pharr GM. An improved technique for determining hardness and elastic modulus using load and displacement sensing indentation experiments. J Mater Res (1992). 7:1564–83. doi:10.1557/JMR.1992.1564 | Google Scholar

CrossRef Full Text | Google Scholar

48. Sneddon IN. The relation between load and penetration in the axisymmetric boussinesq problem for a punch of arbitrary profile. Int J Eng Sci, (1965). 3(1):47–57. doi:10.1016/0020-7225(65)90019-4

CrossRef Full Text | Google Scholar

49. Loubet JL, Georges JM, Marchesini O, Meille G. Vickers indentation curves of magnesium oxide (MgO). J Tribol (1984). 106(1):43–8. doi:10.1115/1.3260865 | Google Scholar

CrossRef Full Text | Google Scholar

50. Manduca A, Oliphant TE, Dresner MA, Mahowald JL, Kruse SA, Amromin E, et al. Magnetic resonance elastography: non-invasive mapping of tissue elasticity. Med Image Anal (2001). 5(4):237–54. doi:10.1016/S1361-8415(00)00039-6

PubMed Abstract | CrossRef Full Text | Google Scholar

51. Maggi LE, Cortela GA, Kruger MAv, Negreira CA, Pereira WCA. Ultrasonic attenuation and speed in phantoms made of PVCP and evaluation of acoustic and thermal proper- ties of ultrasonic phantoms made of polyvinyl chloride- plastisol ( PVCP ), Iwbbio 2013, p. 18–20.

Google Scholar

52. De Carvalho IM, De Matheo LL, Costa Júnior JFS, de Melo Borba C, von Krüger MA, Catelli Infantosi AF, et al. Polyvinyl chloride plastisol breast phantoms for ultrasound imaging. Ultrasonics (2016). 70:98–106. doi:10.1016/j.ultras.2016.04.018

CrossRef Full Text | Google Scholar

53. Allmann KH, Horch R, Uhl M, Gufler H, Altehoefer C, Stark GB, et al. MR imaging of the carpal tunnel. Eur J Radiol (1997). 25(2):141–5. doi:10.1016/S0720-048X(96)01038-8

PubMed Abstract | CrossRef Full Text | Google Scholar

54. Graham SJ, Stanisz GJ, Kecojevic A, Bronskill MJ, Henkelman RM. Analysis of changes in MR properties of tissues after heat treatment. Magn Reson Med (1999). 42(6):1061–71. doi:10.1002/(SICI)1522-2594(199912)42:6<1061::AID-MRM10>3.0.CO;2-T

PubMed Abstract | CrossRef Full Text | Google Scholar

55. Cieszanowski A, Szeszkowski W, Golebiowski M, Bielecki DK, Grodzicki M, Pruszynski B. Discrimination of benign from malignant hepatic lesions based on their T2-relaxation times calculated from moderately T2-weighted turbo SE sequence. Eur Radiol (2002). 12(9):2273–9. doi:10.1007/s00330-002-1366-6

CrossRef Full Text | Google Scholar

56. Gold GE, Han E, Stainsby J, Wright G, Brittain J, Beaulieu C. Musculoskeletal MRI at 3.0 T: relaxation times and image contrast. Am J Roentgenol (2004). 183:343–51. doi:10.2214/ajr.183.2.1830343

CrossRef Full Text | Google Scholar

57. Stanisz GJ, Odrobina EE, Pun J, Escaravage M, Graham SJ, Bronskill MJ, et al. T1, T2 relaxation and magnetization transfer in tissue at 3T. Magn Reson Med (2005). 54(3):507–12. doi:10.1002/mrm605

CrossRef Full Text | Google Scholar

58. Kastler B, Vetter D. Comprendre L’IRM : manuel d’auto-apprentissage. 7th ed. Paris: Elsevier Masson (2011).

59. Bilston L, Liu Z, Phan-Thien N. Linear viscoelastic properties of bovine brain tissue in shear. Biorheology (1997). 34(6):377–85. doi:10.3233/bir-1997-34603

CrossRef Full Text | Google Scholar

60. Nicolle S, Lounis M, Willinger R, Palierne JF. Shear linear behavior of brain tissue over a large frequency range. Biorheology (2005). 42(3):209–23.

Google Scholar

61. Hrapko M, van Dommelen JA, Peters GW, Wismans JS. The mechanical behaviour of brain tissue: large strain response and constitutive modelling. Biorheology (2006). 43(5):623–36. doi:10.1109/ICElMach.2012.6350206

CrossRef Full Text | Google Scholar

62. Vappou J, Breton E, Choquet P, Goetz C, Willinger R, Constantinesco A. Magnetic resonance elastography compared with rotational rheometry for in vitro brain tissue viscoelasticity measurement. Magn Reson Mater Phy (2007). 20(5–6):273–8. doi:10.1007/s10334-007-0098-7

CrossRef Full Text | Google Scholar

63. Chatelin S, Constantinesco A, Willinger R. Fifty years of brain tissue mechanical testing: from in vitro to in vivo investigations. Biorheology (2010). 47(5–6):255–76. doi:10.3233/BIR-2010-0576

PubMed Abstract | CrossRef Full Text | Google Scholar

64. Chatelin S, Oudry J, Périchon N, Sandrin L, Allemann P, Soler L, et al. In vivo liver tissue mechanical properties by transient elastography: comparison with dynamic mechanical analysis. Biorheology (2011). 48(2):75–88. doi:10.3233/BIR-2011-0584

PubMed Abstract | CrossRef Full Text | Google Scholar

65. Nguyen MM, Zhou S, Robert J-l., Shamdasani V, Xie H. Development of oil-in-gelatin phantoms for viscoelasticity measurement in ultrasound shear wave elastography. Ultrasound Med Biol (2014). 40(1):168–76. doi:10.1016/j.ultrasmedbio.2013.08.020

PubMed Abstract | CrossRef Full Text | Google Scholar

66. Hamhaber U, Grieshaber FA, Nagel JH, Klose U. Comparison of quantitative shear wave MR-elastography with mechanical compression tests. Magn Reson Med (2003). 49(1):71–7. doi:10.1002/mrm.10343

PubMed Abstract | CrossRef Full Text | Google Scholar

67. Ringleb SI, Chen Q, Lake DS, Manduca A, Ehman RL, An K-N. Quantitative shear wave magnetic resonance elastography: comparison to a dynamic shear material test. Magn Reson Med (2005). 53(5):1197–201. doi:10.1002/mrm.20449

PubMed Abstract | CrossRef Full Text | Google Scholar

68. Okamoto RJ, Clayton EH, Bayly PV. Viscoelastic properties of soft gels: comparison of magnetic resonance elastography and dynamic shear testing in the shear wave regime. Phys Med Biol (2011). 56(19):6379–400. doi:10.1088/0031-9155/56/19/014

CrossRef Full Text | Google Scholar

69. Lefebvre PM, Koon KT, Brusseau E, Nicolle S, Palieme JF, Lambert SA, et al. Comparison of viscoelastic property characterization of plastisol phantoms with magnetic resonance elastography and high-frequency rheometry. Annu Int Conf IEEE Eng Med Biol Soc in 38th annual international conference of the IEEE engineering in medicine and biology society; (EMBC; Orlando, FL) (2016). p. 1216–9.

PubMed Abstract | Google Scholar

70. Szabo TL, Wu J. A model for longitudinal and shear wave propagation in viscoelastic media. J Acoust Soc Am (2000). 107(5):2437–46. doi:10.1121/1.428630

PubMed Abstract | CrossRef Full Text | Google Scholar

71. Souchon R, Salomir R, Beuf O, Milot L, Grenier D, Lyonnet D, et al. Transient MR elastography (t-MRE) using ultrasound radiation force: theory, safety, and initial experiments in vitro. Magn Reson Med (2008). 60(4):871–81. doi:10.1002/mrm.21718

PubMed Abstract | CrossRef Full Text | Google Scholar

72. Vappou J, Bour P, Marquet F, Ozenne V, Quesson B. MR-ARFI-based method for the quantitative measurement of tissue elasticity: application for monitoring HIFU therapy. Phys Med Biol (2018). 63(9):095018. doi:10.1088/1361-6560/aabd0d

CrossRef Full Text | Google Scholar

73. Dang J, Frisch B, Lasaygues P, Zhang D, Tavernier S, Felix N, et al. Development of an anthropomorphic breast phantom for combined PET, B-mode ultrasound and elastographic imaging. IEEE Trans Nucl Sci (2011). 58(3):660–7. doi:10.1109/TNS.2011.2105279

CrossRef Full Text | Google Scholar

74. Hadj Henni A, Schmitt C, Tremblay MÉ, Hamdine M, Heuzey MC, Carreau P, et al. Hyper-frequency viscoelastic spectroscopy of biomaterials. J Mech Behav Biomed Mater (2011). 4(7):1115–22. doi:10.1016/j.jmbbm.2011.03.020

PubMed Abstract | CrossRef Full Text | Google Scholar

75. Tsou JK, Liu J, Barakat AI, Insana MF. Role of ultrasonic shear rate estimation errors in assessing inflammatory response and vascular risk. Ultrasound Med Biol (2008). 34(6):963–72. doi:10.1016/j.ultrasmedbio.2007.11.010

PubMed Abstract | CrossRef Full Text | Google Scholar

76. Cabrelli LC, Pelissari PI, Deana AM, Carneiro AA, Pavan TZ. Stable phantom materials for ultrasound and optical imaging. Phys Med Biol (2017). 62(2):432–47. doi:10.1088/1361-6560/62/2/432

PubMed Abstract | CrossRef Full Text | Google Scholar

77. Szabo TL. Diagnostic ultrasound imaging, inside out. Elsevier Academic Press (2004).

Google Scholar

78. Yasar TK, Royston TJ, Magin RL. Wideband MR elastography for viscoelasticity model identification. Magn Reson Med (2013). 70:479–89. doi:10.1002/mrm.24495

CrossRef Full Text | Google Scholar

79. Liu Y, Yasar TK, Royston TJ. Ultra wideband (0.5 -16 kHz) MR elastography for Robust shear viscoelasticity model identification. Phys Med Biol (2015). 59(24): 7717–34. doi:10.1002/mrm.24495

CrossRef Full Text | Google Scholar

80. Piazza R, Condino S, Mattei L, Carbone M. Material characterization for elastosonographic phanthoms. Int J Comput Assisted Radiol Surg (2012). 7: 51–2. doi:10.1007/s11548-012-0729-y

CrossRef Full Text | Google Scholar

81. Leclerc GE, Charleux F, Rhein C, Latrive J-P. Viscoelastic properties of healthy and fibrotic liver with magnetic resonance elastography (MRE). J Biomech (2012). 45(1):S577. doi:10.1016/S0021-9290(12)70578-X

CrossRef Full Text | Google Scholar

82. Chartier C, Elkateb M, Mofid Y, Henni AH, Schmitt C, Bastard C, et al. Elastographic characterization of two viscoelastic phantoms using three techniques: shear wave induced resonance elastography, transient elastography and supersonic shear imaging. In: IEEE International Ultrasonics Symposium; 2014 September 3–6; Chicago, IL. IEEE. (2014). p. 1830–3

CrossRef Full Text | Google Scholar

83. Bishop J, Samani A, Sciarretta J, Luginbuhl C, Plewes DB. A signal/noise analysis of quasi-static MR elastography. IEEE Trans Med Imag. (2001). 20(11): 1183–7. doi:10.1109/42.963821

CrossRef Full Text | Google Scholar

84. Samani A, Bishop J, Plewes DB. A constrained modulus reconstruction technique for breast cancer assessment. IEEE Trans Med Imag (2001). 20(9): 877–85. doi:10.1109/42.952726

CrossRef Full Text | Google Scholar

Keywords: tissue modeling, phantom, elasticity imaging, indentation, magnetic resonance imaging, ultrasound imaging

Citation: Chatelin S, Breton E, Arulrajah A, Giraudeau C, Wach B, Meylheuc L and Vappou J (2020) Investigation of PolyVinyl Chloride Plastisol Tissue-Mimicking Phantoms for MR- and Ultrasound-Elastography. Front. Phys. 8:577358. doi: 10.3389/fphy.2020.577358

Received: 26 May 2020; Accepted: 21 August 2020;
Published: 22 December 2020.

Edited by:

Qiyin Fang, McMaster University, Canada

Reviewed by:

William Vogt, United States Food and Drug Administration, United States
Theo Zeferino Pavan, University of São Paulo, Brazil

Copyright © 2020 Chatelin, Breton, Arulrajah, Giraudeau, Wach, Meylheuc and Vappou. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.