Minimally Invasive Image-Guided Gut Transport Function Measurement Probe

www.frontiersin.orgDavid O. Otuya1,2 www.frontiersin.orgEvangelia Gavgiotaki1,2 www.frontiersin.orgCamella J. Carlson1 www.frontiersin.orgSerena Q. Shi3 www.frontiersin.orgAriel J. Lee4 www.frontiersin.orgAlexander A. Krall1 www.frontiersin.orgAnita Chung1 www.frontiersin.orgCatriona G. Grant1 www.frontiersin.orgNitasha M. Bhat1 www.frontiersin.orgPeter Choy1 www.frontiersin.orgSarah L. Giddings1 www.frontiersin.orgJoseph A. Gardecki1,2 www.frontiersin.orgJay R. Thiagarajah2,5 www.frontiersin.orgSteven M. Rowe6,7 www.frontiersin.orgGuillermo J. Tearney1,2,8,9*
  • 1Wellman Center for Photomedicine, Massachusetts General Hospital, Boston, MA, United States
  • 2Harvard Medical School, Boston, MA, United States
  • 3University of Pennsylvania, Philadelphia, MA, United States
  • 4Korea Advanced Institute of Science and Technology, Daejeon, South Korea
  • 5Division of Gastroenterology, Hepatology and Nutrition, Boston Children’s Hospital, Boston, MA, United States
  • 6Department of Medicine, University of Alabama at Birmingham, Birmingham, AL, United States
  • 7Gregory Fleming James Cystic Fibrosis Research Center, Birmingham, AL, United States
  • 8Department of Pathology, Massachusetts General Hospital, Boston, MA, United States
  • 9Harvard-MIT Division of Health Sciences and Technology (HST), Boston, MA, United States

Introduction: Diseases such as celiac disease, environmental enteric dysfunction, infectious gastroenteritis, type II diabetes and inflammatory bowel disease are associated with increased gut permeability. Dual sugar absorption tests, such as the lactulose to rhamnose ratio (L:R) test, are the current standard for measuring gut permeability. Although easy to administer in adults, the L:R test has a number of drawbacks. These include an inability to assess for spatial heterogeneity in gut permeability that may distinguish different disease severity or pathology, additional sample collection for immunoassays, and challenges in carrying out the test in certain populations such as infants and small children. Here, we demonstrate a minimally invasive probe for real-time localized gut permeability evaluation through gut potential difference (GPD) measurement.

Materials and Methods: The probe has an outer diameter of 1.2 mm diameter and can be deployed in the gut of unsedated subjects via a transnasal introduction tube (TNIT) that is akin to an intestinal feeding tube. The GPD probe consists of an Ag/AgCl electrode, an optical probe and a perfusion channel all housed within a transparent sheath. Lactated Ringer’s (LR) solution is pumped through the perfusion channel to provide ionic contact between the electrodes and the gut lining. The optical probe captures non-scanning (M-mode) OCT images to confirm electrode contact with the gut lining. A separate skin patch probe is placed over an abraded skin area to provide reference for the GPD measurements. Swine studies were conducted to validate the GPD probe. GPD in the duodenum was modulated by perfusing 45 ml of 45 mM glucose.

Results: GPD values of −13.1 ± 2.8 mV were measured in the duodenum across four swine studies. The change in GPD in the duodenum with the addition of glucose was −10.5 ± 2.4 mV (p < 0.001). M-mode OCT images provided electrode-tissue contact information, which was vital in ascertaining the probe’s proximity to the gut mucosa.

Conclusion: We developed and demonstrated a minimally invasive method for investigating gastrointestinal permeability consisting of an image guided GPD probe that can be used in unsedated subjects.

Introduction

The intestinal epithelium provides a barrier against antigens and toxins from the environment and selectively allows absorption of essential nutrients, electrolytes, and water into the body. Normal intestinal permeability (IP) refers to the selective movement of ion, nutrients and water across the mucosa from the intestinal lumen into the circulating blood [1]. Several studies have shown that a breakdown of the epithelial barrier and subsequent alteration of intestinal permeability is associated with inflammation, infection and autoimmune diseases [24].

Changes in intestinal permeability have been specifically linked to a wide variety of diseases and syndromes including celiac disease [5], type II diabetes (T2D) [58]; inflammatory bowel disease (IBD) [911], irritable bowel syndrome [12,13], environmental enteric dysfunction [14,15], obesity [16,17], chronic fatigue [18], fibromyalgia [19] and colon cancer[20,21].

The standard method for measuring intestinal permeability in human subjects involves the administration of oligosaccharides of different molecular weights [22,23]. Non-metabolized pairs of sugars such as lactulose, mannitol or rhamnose are orally administered and urine elimination measured over a 3–24 h period. Although, the assessment of intestinal permeability by dual sugar absorption has been proven to be non-invasive and safe, it is time consuming since urine samples are collected between 3–24 h after sugar-probe administration and then shipped to a centralized facility for analysis, taking days to weeks [24,25]. Interpretation of the data can be confounded by a number of factors including, low urinary volumes, variable gut surface area and intestinal transit times. In addition, quantitative analysis of the urine sample requires high pressure liquid chromatography (HPLC) in combination with mass spectroscopy or direct enzyme-linked immunosorbent assay (ELISA) [25]. Lastly, the ability to collect clean urine samples without fecal contamination in certain populations such as infants, children and the elderly can be challenging, affecting the utility of the dual-sugar test for intestinal permeability evaluation in these population groups [14].

Other methods including assaying endogenous intestinal biomarkers in the serum, such as zonulin (ZO), intestinal fatty-acid binding proteins (FABPs), and tight junction (TJ) proteins, have been proposed for assessment of altered barrier integrity [26]. Bacteria or bacterial products have also been used to indicate changes of barrier integrity [27]. While simpler and less time consuming than the dual sugar method, it is currently not clear whether there is a sufficient correlation between these biomarkers and functional gut permeability in vivo [25,28].

The Ussing chamber, developed for the measurement of electrolyte and nutrient transport across epithelial tissues, offers the ability to measure permeability and ion transport in specific regions of the GI tract. However, this method measure tissues in ex vivo conditions requiring invasive collection via an intestinal biopsy sample, with subsequent issues related to tissue viability during measurements [25]. Since this method of gut permeability assessment uses small biopsies from specific areas of the gut, it is challenging to measure gut permeability over wide areas using the Ussing chamber.

There are numerous studies that have shown the applicability of potential difference (PD) measurements to record intestinal permeability alterations [2934]. These experiments used probes made of salt bridges that consisted of 3% agar in saturated KCl, either in isolation or in combination with a tube for constant Ringer’s solution perfusion, to provide ionic contact between the electrodes placed outside the body and the mucosal lining of the gut. These probes were placed in the gut invasively through surgical access or trans-orally under fluoroscopic guidance. The methods reported above used varying unrefined probes, that were not commercially available or standardized and were prone to errors due to the inability to ascertain probe-mucosal contact when measuring PD.

In this paper, we propose a gut PD measurement (GPD) probe with a capability for real-time measurement of gut potential difference, providing dynamic real-time measurements of gut permeability. We have developed an image guided GPD measurement device that obviates the need for fluoroscopic guidance and consequent radiation exposure. The probe is a small caliber device that can be deployed via a trans-nasal guidance tube or endoscope. In addition, we have eliminated the need for an agar bridge, making the GPD probe a sterilizable, transportable, reusable, and convenient to use at the bedside in humans. We present the design of the GPD probe and show results to validate its safety and efficacy in measuring gut PD.

Materials and Methods

Trans-Nasal GPD Probe

A schematic of the GPD catheter is shown in Figure 1. The GPD probe used in this study was a 1.2-mm-diameter device that can be introduced into the GI tract transnasally via an introduction tube. The probe was enclosed in 1.2 m long, 1.2 mm outer diameter (OD) and 1.0 mm internal diameter (ID) polytetrafluoroethylene (PTFE) tube (ZEUS®, SC, United States). The device consisted of a 0.5 mm OD Ag/AgCl electrode (World Precision Instruments (WPI), FL, United States ) contained in a mini cell located at the distal end of the probe. A PTFE/composite perfusion tube of OD 0.47mm and 0.26 mm ID (Microlumen®, FL, United States ) was placed inside the entire length of the probe to transport Ringer’s solution from a perfusion pump outside the body to the mini cell at the distal end of the probe. Ringer’s solution was injected via the perfusion port to provide ionic contact between the electrode and the intestinal mucosa. A single mode optical fiber (SMF) placed in the probe carried light to obtain non-scanning (M-mode) optical coherence tomography (OCT) images at the probe’s tip to determine when it is in contact with mucosa and to provide tissue morphological information. A reference probe was inserted subcutaneously or implemented via a 3 M™ Red Dot™ ECG monitoring electrode skin patch, attached over an abraded area of the skin. The GPD was defined as the voltage measured across the trans-nasal GPD probe and the reference electrode.

FIGURE 1
www.frontiersin.org

FIGURE 1. The GPD probe consists of a mini cell at the tip and the proximal connectors. (A) The tip is enclosed in a 1.2 mm outer diameter (OD) polytetrafluoroethylene (PTFE) tube with a 0.5 mm diameter Ag/AgCl electrode, an optical probe consisting of a ball lens and a single mode fiber (SMF), a perfusion tube that delivers lactated Ringer’s solution to the mini cell and perfusion channel that conducts a drip of lactated Ringer’s solution to the exteriorly located tissue. (B) The proximal end consists of a Luer lock connector to the perfusion tube from a perfusion pump, an optical connector from an OCT system to the optical probe and an electrical connector that links a wire from the Ag/AgCl to the isolation head-stage. (C, D) A Magnified view of the GPD probe.

In humans, the concept will involve inserting the probe into the gut using either OCT image guidance or an equivalent method to confirm anatomical placement [35,36]. When using OCT for placement, a separate introduction tube containing a mechanically scanning optical probe in its working channel will be deployed in the gut. The B-mode frames provided by the scanning optics will be used to localize the introduction tube in the gut. After retraction of the optics, the GPD probe will be inserted through the working channel of the introduction tube. After the probe is situated in the duodenum, fluid will be perfused via the port and the probe brought in contact with the mucosa as confirmed with OCT. GPD will then be measured.

GPD Measurement System

Figure 2A shows the GPD measurement system, which consisted of an isolation head-stage (ISO-Z, CWe Inc., PA, United States ) to which the trans-nasal GPD and reference probes were connected. The signal from the isolation head-stage was then connected to a bioamplifier (BMA-200, CWe Inc. PA, United States). The amplified signal from the bioamplifier was sent to a digitizer (Power Lab 4/26, AD instruments Inc., CO, United States) for conversion to a digital signal and logged with LabChart software (AD instruments Inc., CO, United States ) on a computer at a rate of 1 k sample/s. The GPD system was adapted from a nasal potential difference measurement system with similar equipment [37,38].

FIGURE 2
www.frontiersin.org

FIGURE 2. Experiment schematic for gut potential difference (GPD) measurement in swine. (A) PD measurement system consisting of an isolation head-stage, a bioamplifier, an analog-to-digital converter and a computer. (B) An offset zeroing setup made of a cup filled with Ringer’s solution and a skin patch. (C) The pig’s small intestine was intubated using a pediatric gastroscope and the IPD probe inserted through the working channel of the endoscope. The GPD signal was acquired via an isolation headstage and a bioamplfier in the GPD measurement system. The GPD signal was converted to a digital signal and logged using a computer. Optical coherence tomography (OCT) M-mode images were simultaneously acquired. (D) Detailed view of the duodenum and the GPD probe.

Swine Studies

The GPD probe was connected to an OCT system consisting of an Axsun OCT engine (Excelitas, MA, United States ) operating at 1,310 nm center wavelength and acquiring OCT image frames at 100 kHz [39]. The optical probe acquired a non-scanning form of OCT images known as M-mode OCT. M-mode OCT was used to ensure that the probe was in contact with the mucosal tissue while the GPD values were measured. To validate the ability of the probe to measure real-time changes in GPD, a transient gut sodium ion concentration modulator (45 mM glucose) was perfused while potential difference measurements were taken. Electrogenic absorption of glucose causes a local reduction in the sodium ion concentration on the luminal side of the mucosa resulting in a change in the apical GPD. The swine studies were approved by the Massachusetts General Hospital (MGH) Institutional Animal Care and Use Committee (IACUC) protocol number 2016N000215.

Procedure

1) Prior to the procedure, the swine were anesthetized and intubated by the staff at the Knight Surgery and Research Laboratory (KSRL) at MGH. Vitals were monitored to ensure the animal was in good physiological condition.

2) A convenient portion of the skin was shaved and abraded until interstitial fluid was exposed.

3) A skin patch was placed over the abraded area.

4) An endoscope was introduced trans-orally and advanced to the first segment of the duodenum.

5) Before any measurement was taken, the potential difference offset between the GPD probe and the skin patch was zeroed using the calibration setup shown in Figure 2B.

6) The reference cable from the isolation head-stage was then connected to the skin patch on the swine skin.

7) The GPD probe connected to the perfusion pump delivering Lactated Ringer’s solution to the probe at 1 ml/hr was inserted via the working channel of the scope.

8) Baseline GPD measurements were taken after 30 min of probe equilibration.

9) A secondary tube was inserted via the working channel in which glucose was perfused.

Results

Initially, an endoscope was safely deployed into the duodenum and through its working channel the GPD device inserted as shown in Figure 3A. Then, M-mode OCT images were acquired to ensure contact of the probe with the mucosa while PD values were recorded. The right half of Figure 3B demonstrates a representative M-mode OCT image with the red arrow indicating probe-tissue contact shown by the ring-like scattering pattern from the tissue that abuts the outside of the probe (yellow arrow). The measured average baseline GPD across four swine was −13.1 ± 2.8 mV.

FIGURE 3
www.frontiersin.org

FIGURE 3. (A) Endoscopic image of the duodenum showing the GPD probe not in contact with the mucosa and the corresponding M-mode OCT image. (B) Endoscopic image showing the IPD probe in contact with the mucosa. An M-mode OCT frame corresponding to the endoscopy frame depicting the GPD probe in contact with the duodenal mucosa, as shown by the dark ring pointed by the red arrow. (C) GPD values measured by the probe in contact and not in contact with the mucosa. (D) This figure shows a change in the gut potential difference after glucose perfusion. The baseline GPD value changed in presence of 45 mM glucose solution by as much as −10.5 mV. The single asterisk demarcates the region enclosed by the dashed rectangle before glucose infusion while the double asterisks point to the region in the rectangle depicting the potential difference across the probe and the reference skin patch after the probe was retracted from the body. (E) The graph shows baseline GPD value of the duodenum before infusion of glucose. (F) After probe retraction from the duodenum, the graph shows the GPD between the GPD probe and the skin patch was about zero, indicating that the change in GPD prior to retraction was solely due to the presence of glucose, and not because of a walk-off between the probe and the reference skin patch. The asterisks indicate the positions of the graphs in (E) and (F) in the original graph in (D).

Figure 3C shows a rapid time response of under 2 s when the GPD probe was not in contact with the mucosa was brought in contact with the tissue. We also evaluated the potential difference change that resulted from perfusion of 45 mM glucose over a duration of 60 s at a rate of 1 ml/s to stimulate gut electrolyte absorption as shown in Figure 3D. The total change in GPD was −10.5 ± 2.4 mV (p < 0.001), due to the glucose mediated active Na + transport across the mucosa. Figure 3E shows the initial baseline PD, while Figure 3F shows the PD between the probe and the reference skin patch after the device was retracted from the scope demonstrating the absence of a residual offset potential.

In addition, GPD measurements were acquired in different anatomical regions of the upper GI tract. Figure 4 demonstrates the PD values measured in three sections of the esophagus (proximal esophagus, mid esophagus and distal esophagus) measured over 15 s. The mean PD values measured in these organs was −11.6 ± 0.7, −20.4 ± 2.1, and −34.1 ± 1.3 mV for the proximal esophagus, mid esophagus and distal esophagus, respectively. In the stomach, as shown in the figure, the average PD measured in the stomach fundus, body and pyloric antral regions over a duration of 15 s was −34.2 ± 1.4, −20.2 ± 2.2, −12.7 ± 2.1 mV, respectively.

FIGURE 4
www.frontiersin.org

FIGURE 4. PD values of proximal esophagus, mid esophagus, distal esophagus, body fundus, stomach (body) and the pyloric antrum. Error bars indicate SD.

Discussion

Here, we have presented a design for a trans-nasal minimally invasive GPD measurement device. We validated the GPD probe in swine by measuring swine esophageal, stomach and duodenal potential difference. These duodenal potential difference values measured by the GPD probe approximated the PD values measured in human duodenum reported by Gustke et al [40]. Gastric PD measurements obtained with our GPD probe are different from those reported by Geall et al [31]. by about −10 to −15 mV. This discrepancy may be attributed to the lateral recumbent position of the swine during scoping that allowed for a small pool of gastric contents to remain lodged in the distal portion of the esophagus providing ionic connectivity to the stomach mucosa. The PD value measured with this probe in the proximal region of the esophagus is consistent with the potential difference values measured in the same region in humans [41]. Since there is insufficient information on GPD values in swine, further studies need to be done to establish the similarities and differences between human and swine GPD.

In addition to baseline GPD we also showed a change in duodenal PD after infusion of glucose solution and subsequent induction of electrogenic glucose uptake. This glucose absorption is coupled with sodium ion intake by the sodium/glucose cotransporter 1 (SLGT1) across the intestinal epithelium. Our results showed the expected decrease in transepithelial potential consistent with electrogenic Na+-glucose co-transport across the epithelium.

A major advantage of our device that enables blind (non-endoscopic) insertion is the integrated M-mode OCT imaging technology. M-mode imaging was able to indicate when the GPD probe was in contact with the intestinal wall. Thus, M-mode OCT allows for accurate GPD measurements by ensuring that PD values are acquired only when the device in contact with the mucosa. This feature is especially critical when the device is deployed trans-nasally, without endoscopic guidance, allowing for precise placement of the probe and accurate measurement of GPD. This was particularly important for measurement of positive potential changes that occur during intestinal electrolyte and fluid secretion, which would otherwise be erroneously attributed if the probe was not in contact with the mucosa. This will be particularly important in future studies caried out without the benefit of endoscopic guidance.

Real-time GPD measurement offers an attractive prospect for the investigation of gut transport function in a rapid and minimally invasive manner. As an example, targeted functional evaluation of ion channels such as the cystic fibrosis transmembrane conductance regulator (CFTR) can be performed with the aid of this GPD device. PD changes due to chloride transport via CFTR in response to electrochemical gradients using chloride free solution infusions with or without solutions containing cyclic adenosine monophosphate (cAMP) agonists, can be measured to ascertain the functional integrity of this critical ion channel [38]. Gut PD measurement may also play a role in further understanding the dynamics of secretory diarrheas such as cholera in humans [42] or deliver functional assessments of electrolyte and nutrient transport in a variety of disease states in infants including congenital diarrheas, allergic enteropathies and environmental enteric dysfunction [4345]. Further, gut permeability to larger molecules can be measured and localized using this device. In contrast to the dual sugar method where heterogeneity of gut permeability may be hard to address, GPD measurements can provide spatial and segmental gut permeability data that may allow dynamic mapping of permeability along the length of the intestine. Lastly, PD measurements as a means for assessing “leaky-gut” conditions, may also be appealing in low resource settings. GPD measurement offers a low-cost and more practical alternative to the dual-sugar test, which does not require advanced ELISA/chromatography analysis of collected samples.

Several challenges may still hamper gut transport evaluation using this method. One major challenge is the temperature drift between the reference and the measurement probe brought about by a change in fluid temperature as it is perfused through the perfusion channel to the gut mucosa. This drift in temperature causes a small drift in the measured PD values [46]. Also, formation of bubbles within the perfusion channel creates areas of discontinuity affecting the PD values registered, a well-known impediment to nasal PD performance [37]. To minimize the formation of gas bubbles, we degas Ringer’s solution before each GPD measurement procedure by sonication for 30 min and then placing the solution in a vacuum at a pressure of about −75 kPa for another 30 min. Another issue that may affect PD measurement in vivo is the PD fluctuation due to movement of either the probe or the subject resulting in motion artifacts on the PD signal. External potential sources of noise to the PD signal that should be considered include background radiofrequency from the environment which can be eliminated by applying a low-pass filter to PD signal with cut-off frequency about 10 Hz, as has been successful employed in NPD [37].

We also note that GPD is an indirect measurement of gut permeability. While it is true that when the gut epithelium is completely permeable, GPD will go to zero, there are other processes that could cause the PD to diminish. For example, the efflux of negative ions from the lumen or influx of positive ions to the lumen will push PD values towards zero. For this reason, in the future, it will be important to measure both transepithelial voltage and current to determine the transepithelial resistance and short-circuit current [47]. While studies have shown a correlation between dysregulation of the tight junction proteins involved in gut permeability and a change in baseline GPD [48,49], more work needs to be done in the future to understand the correlation of GPD and ionic versus macromolecular intestinal permeability.

The minimally invasive gut transport measurement probe presented here is compatible with pediatric trans-nasal introduction in a similar manner to commonly used nasogastric/nasojejunal feeding tubes (7Fr). This device can be not only be used for assessment of gut permeability through PD measurement but could also be extended for evaluation of epithelial integrity and function in the airway and other luminal organs. This GPD device is also small enough to be deployed via the working channel of ultrasmall trans-nasal endoscopes, obviating the need for sedation that is routinely done during EGD and bronchoscopy. The ability to perform unsedated minimally invasive PD assessment opens opportunities for understanding both healthy and diseased tissues across populations, as well as allowing timely diagnoses of conditions in vulnerable groups such as infants, pregnant women, and the elderly.

Data Availability Statement

The raw data supporting the conclusions of this article will be made available by the authors, without undue reservation.

Ethics Statement

The animal study was reviewed and approved by MGH IACUC.

Author Contributions

DO and EG, JT, GT wrote the article; PC, CC, and SS fabricated the PD probes; AL and SR. helped conceive and set up the initial PD measurement system, AC, CG, SR, and NB provided advice on clinical aspects of the project; JG provided help in the development of the probe and organized the swine studies; SG managed the project; JT, SR provided valuable advice on both ex vivo and in vivo studies.

Conflict of Interest

GT has a financial/fiduciary interest in SpectraWave, a company developing an OCT-NIRS intracoronary imaging system and catheter. His financial/fiduciary interest was reviewed and is managed by the Massachusetts General Hospital and Partners HealthCare in accordance with their conflict of interest policies. GT receives sponsored research from VivoLight, iLumen, Boston Scientific, Vertex, Hamamatsu, AstraZeneca, CN USA Biotech Holdings, Wayvector, and Canon Inc. GT receives catheter materials from Terumo Corporation. GT additional receives royalties from Terumo Corporation, iLumen, and MIT

The remaining authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s Note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Acknowledgments

We acknowledge the Bill and Melinda Gates Foundation for Funding this project.

References

1. Hollander D. Intestinal permeability, leaky gut, and intestinal disorders. Curr Gastroenterol Rep (1999) 1:410–6. doi:10.1007/s11894-999-0023-5

PubMed Abstract | CrossRef Full Text | Google Scholar

2. Clayburgh DR, Shen L, Turner JR. A porous defense: the leaky epithelial barrier in intestinal disease. Lab Invest (2004) 84:282–91. doi:10.1038/labinvest.3700050

PubMed Abstract | CrossRef Full Text | Google Scholar

3. Arrieta MC, Bistritz L, Meddings JB. Alterations in intestinal permeability. Gut (2006) 55:1512–20. doi:10.1136/gut.2005.085373

PubMed Abstract | CrossRef Full Text | Google Scholar

4. Thoo L, Noti M, Krebs P. Keep calm: the intestinal barrier at the interface of peace and war. Cell Death Dis (2019) 10:849. doi:10.1038/s41419-019-2086-z

PubMed Abstract | CrossRef Full Text | Google Scholar

5. Visser J, Rozing J, Sapone A, Lammers K, Fasano A. Tight Junctions, Intestinal Permeability, and Autoimmunity. Ann N Y Acad Sci (2009) 1165:195–205. doi:10.1111/j.1749-6632.2009.04037.x

PubMed Abstract | CrossRef Full Text | Google Scholar

6. Vaarala O. Leaking gut in type 1 diabetes. Curr Opin Gastroenterol (2008) 24:701–6. doi:10.1097/MOG.0b013e32830e6d98

PubMed Abstract | CrossRef Full Text | Google Scholar

7. Li X, Atkinson MA. The role for gut permeability in the pathogenesis of type 1 diabetes - a solid or leaky concept?. Pediatr Diabetes (2015) 16:485–92. doi:10.1111/pedi.12305

PubMed Abstract | CrossRef Full Text | Google Scholar

8. Cox AJ, Zhang P, Bowden DW, Devereaux B, Davoren PM, Cripps AW, et al. Increased intestinal permeability as a risk factor for type 2 diabetes. Diabetes Metab (2017) 43:163–6. doi:10.1016/j.diabet.2016.09.004

PubMed Abstract | CrossRef Full Text | Google Scholar

9. Zeissig S, Bürgel N, Günzel D, Richter J, Mankertz J, Wahnschaffe U, et al. Changes in expression and distribution of claudin 2, 5 and 8 lead to discontinuous tight junctions and barrier dysfunction in active Crohn's disease. Gut (2007) 56:61–72. doi:10.1136/gut.2006.094375

PubMed Abstract | CrossRef Full Text | Google Scholar

10. Kumar M, Leon Coria A, Cornick S, Petri B, Mayengbam S, Jijon HB, et al. Increased intestinal permeability exacerbates sepsis through reduced hepatic SCD-1 activity and dysregulated iron recycling. Nat Commun (2020) 11:483. doi:10.1038/s41467-019-14182-2

PubMed Abstract | CrossRef Full Text | Google Scholar

11. Turpin W, Lee S-H, Raygoza Garay JA, Madsen KL, Meddings JB, Bedrani L, et al. Increased Intestinal Permeability Is Associated With Later Development of Crohn's Disease. Gastroenterology (2020) 159:2092–100. doi:10.1053/j.gastro.2020.08.005

PubMed Abstract | CrossRef Full Text | Google Scholar

12. Zhou Q, Zhang B, Verne NG. Intestinal membrane permeability and hypersensitivity in the irritable bowel syndrome. Pain (2009) 146:41–6. doi:10.1016/j.pain.2009.06.017

PubMed Abstract | CrossRef Full Text | Google Scholar

13. Matricon J, Meleine M, Gelot A, Piche T, Dapoigny M, Muller E, et al. Review article: associations between immune activation, intestinal permeability and the irritable bowel syndrome. Aliment Pharmacol Ther (2012) 36:1009–31. doi:10.1111/apt.12080

PubMed Abstract | CrossRef Full Text | Google Scholar

14. Crane RJ, Jones KDJ, Berkley JA Environmental enteric dysfunction: an overview. Food Nutr Bull (2015) 36:S76–S87. doi:10.1177/15648265150361s113

PubMed Abstract | CrossRef Full Text | Google Scholar

15. Morais MBd., Silva GAPd.. Environmental enteric dysfunction and growth Jornal de Pediatria (2019) 95:85–94. doi:10.1016/j.jped.2018.11.004

PubMed Abstract | CrossRef Full Text | Google Scholar

16. Damms-Machado A, Louis S, Schnitzer A, Volynets V, Rings A, Basrai M, et al. Gut permeability is related to body weight, fatty liver disease, and insulin resistance in obese individuals undergoing weight reduction. Am J Clin Nutr (2017) 105:127–35. doi:10.3945/ajcn.116.131110

PubMed Abstract | CrossRef Full Text | Google Scholar

17. Thaiss CA, Levy M, Grosheva I, Zheng D, Soffer E, Blacher E, et al. Hyperglycemia drives intestinal barrier dysfunction and risk for enteric infection. Science (2018) 359:1376–83. doi:10.1126/science.aar3318

PubMed Abstract | CrossRef Full Text | Google Scholar

18. Lupo GFD, Rocchetti G, Lucini L, Lorusso L, Manara E, Bertelli M, et al. Potential role of microbiome in Chronic Fatigue Syndrome/Myalgic Encephalomyelits (CFS/ME). Sci Rep (2021) 11:7043. doi:10.1038/s41598-021-86425-6

PubMed Abstract | CrossRef Full Text | Google Scholar

19. Goebel A, Buhner S, Schedel R, Lochs H, Sprotte G Altered intestinal permeability in patients with primary fibromyalgia and in patients with complex regional pain syndrome. Rheumatology (2008) 47:1223–7. doi:10.1093/rheumatology/ken140

PubMed Abstract | CrossRef Full Text | Google Scholar

20. Soler AP, Miller RD, Laughlin KV, Carp NZ, Klurfeld DM, Mullin JM Increased tight junctional permeability is associated with the development of colon cancer. Carcinogenesis (1999) 20:1425–32. doi:10.1093/carcin/20.8.1425

PubMed Abstract | CrossRef Full Text | Google Scholar

21. Bekusova VV, Falchuk EL, Okorokova LS, Kruglova NM, Nozdrachev AD, Markov AG Increased paracellular permeability of tumor-adjacent areas in 1,2-dimethylhydrazine-induced colon carcinogenesis in rats. Cancer Biol Med (2018) 15:251–9. doi:10.20892/j.issn.2095-3941.2018.0016

PubMed Abstract | CrossRef Full Text | Google Scholar

22. Van Elburg RM, Uil JJ, Kokke FTM, Mulder AM, Van De Broek WGM, Mulder CJJ, et al. Repeatability of the sugar-absorption test, using lactulose and mannitol, for measuring intestinal permeability for sugars. J Pediatr Gastroenterol Nutr (1995) 20:184–8. doi:10.1097/00005176-199502000-00008

CrossRef Full Text | Google Scholar

23. Van Wijck K, Verlinden TJM, Van Eijk HMH, Dekker J, Buurman WA, Dejong CHC, et al. Novel multi-sugar assay for site-specific gastrointestinal permeability analysis: a randomized controlled crossover trial. Clin Nutr (2013) 32:245–51. doi:10.1016/j.clnu.2012.06.014

PubMed Abstract | CrossRef Full Text | Google Scholar

24. Catassi C, Fabiani E, Rätsch IM, Bonucci A, Dotti M, Coppa GV, et al. Is the sugar intestinal permeability test a reliable investigation for coeliac disease screening?. Gut (1997) 40:215–7. doi:10.1136/gut.40.2.215

PubMed Abstract | CrossRef Full Text | Google Scholar

25. Galipeau HJ, Verdu EF. The complex task of measuring intestinal permeability in basic and clinical science. Neurogastroenterol Motil (2016) 28:957–65. doi:10.1111/nmo.12871

PubMed Abstract | CrossRef Full Text | Google Scholar

26. Wells JM, Brummer RJ, Derrien M, Macdonald TT, Troost F, Cani PD, et al. Homeostasis of the gut barrier and potential biomarkers. Am J Physiology-Gastrointestinal Liver Physiol (2017) 312:G171–g193. doi:10.1152/ajpgi.00048.2015

PubMed Abstract | CrossRef Full Text | Google Scholar

27. Bischoff SC, Barbara G, Buurman W, Ockhuizen T, Schulzke J-D, Serino M, et al. Intestinal permeability - a new target for disease prevention and therapy. BMC Gastroenterol (2014) 14:189. doi:10.1186/s12876-014-0189-7

PubMed Abstract | CrossRef Full Text | Google Scholar

28. Wells JM, Brummer RJ, Derrien M, Macdonald TT, Troost F, Cani PD, et al. Homeostasis of the gut barrier and potential biomarkers. Am J Physiology-Gastrointestinal Liver Physiol (2017) 312:G171–G193. doi:10.1152/ajpgi.00048.2015

PubMed Abstract | CrossRef Full Text | Google Scholar

29. Geall MG, Summerskill WH Electric-potential difference--a neglected parameter of gut integrity and function?. Gut (1969) 10:418–21. doi:10.1136/gut.10.6.418

PubMed Abstract | CrossRef Full Text | Google Scholar

30. Barry PH, Diamond JM Junction potentials, electrode standard potentials, and other problems in interpreting electrical properties of membranes. J Membrain Biol (1970) 3:93–122. doi:10.1007/bf01868010

PubMed Abstract | CrossRef Full Text | Google Scholar

31. Geall MG, Code CF, Mcilrath DC, Summerskill WHJ Measurement of gastrointestinal transmural electric potential difference in man. Gut (1970) 11:34–7. doi:10.1136/gut.11.1.34

PubMed Abstract | CrossRef Full Text | Google Scholar

32. Isaacs PET, Horth CE, Turnberg LA The Electrical Potential Difference across Human Ileostomy Mucosa. Gastroenterology (1976) 70:52–8. doi:10.1016/s0016-5085(76)80402-7

PubMed Abstract | CrossRef Full Text | Google Scholar

33. Read NW, Smallwood RH, Levin RJ, Holdsworth CD, Brown BH Relationship between changes in intraluminal pressure and transmural potential difference in the human and canine jejunum in vivo. Gut (1977) 18:141–51. doi:10.1136/gut.18.2.141

PubMed Abstract | CrossRef Full Text | Google Scholar

34. Tarnawski A, Ivey KJ Transmucosal potential-difference profile in rat upper gastrointestinal tract. A simple model for testing gastric effects of pharmacologic agents. Can J Physiol Pharmacol (1978) 56:471–3. doi:10.1139/y78-070

PubMed Abstract | CrossRef Full Text | Google Scholar

35. Brand S, Poneros JM, Bouma BE, Tearney GJ, Compton CC, Nishioka NS Optical coherence tomography in the gastrointestinal tract. Endoscopy (2000) 32:796–803. doi:10.1055/s-2000-7714

PubMed Abstract | CrossRef Full Text | Google Scholar

36. Gora MJ, Quénéhervé L, Carruth RW, Lu W, Rosenberg M, Sauk JS, et al. Tethered capsule endomicroscopy for microscopic imaging of the esophagus, stomach, and duodenum without sedation in humans (with video). Gastrointest Endosc (2018) 88:830–40. e833. doi:10.1016/j.gie.2018.07.009

PubMed Abstract | CrossRef Full Text | Google Scholar

37. Solomon GM, Konstan MW, Wilschanski M, Billings J, Sermet-Gaudelus I, Accurso F, et al. An international randomized multicenter comparison of nasal potential difference techniques. Chest (2010) 138:919–28. doi:10.1378/chest.10-0179

PubMed Abstract | CrossRef Full Text | Google Scholar

38. Solomon GM, Bronsveld I, Hayes K, Wilschanski M, Melotti P, Rowe SM, et al. Standardized Measurement of Nasal Membrane Transepithelial Potential Difference (NPD). J Vis Exp (2018). doi:10.3791/57006

CrossRef Full Text | Google Scholar

39. Dong J, Grant C, Vuong B, Nishioka N, Gao AH, Beatty M, et al. Feasibility and Safety of Tethered Capsule Endomicroscopy in Patients With Barrett's Esophagus in a Multi-Center Study. Clin Gastroenterol Hepatol (2021). doi:10.1016/j.cgh.2021.02.008

CrossRef Full Text | Google Scholar

40. Gustke RF, Mccormick P, Ruppin H, Soergel KH, Whalen GE, Wood CM Human intestinal potential difference: recording method and biophysical implications. J Physiol (1981) 321:571–82. doi:10.1113/jphysiol.1981.sp014003

PubMed Abstract | CrossRef Full Text | Google Scholar

41. Turner KS, Powell DW, Carney CN, Orlando RC, Bozymski EM Transmural electrical potential difference in the mammalian esophagus in vivo. Gastroenterology (1978) 75:286–91. doi:10.1016/0016-5085(78)90419-5

PubMed Abstract | CrossRef Full Text | Google Scholar

42. Li C, Naren AP. CFTR chloride channel in the apical compartments: spatiotemporal coupling to its interacting partners. Integr Biol (2010) 2:161–77. doi:10.1039/b924455g

PubMed Abstract | CrossRef Full Text | Google Scholar

43. Guerrant RL, Deboer MD, Moore SR, Scharf RJ, Lima AAM. The impoverished gut-a triple burden of diarrhoea, stunting and chronic disease. Nat Rev Gastroenterol Hepatol (2013) 10:220–9. doi:10.1038/nrgastro.2012.239

PubMed Abstract | CrossRef Full Text | Google Scholar

44. Thiagarajah JR, Donowitz M, Verkman AS. Secretory diarrhoea: mechanisms and emerging therapies. Nat Rev Gastroenterol Hepatol (2015) 12:446–57. doi:10.1038/nrgastro.2015.111

PubMed Abstract | CrossRef Full Text | Google Scholar

45. Thiagarajah JR, Kamin DS, Acra S, Goldsmith JD, Roland JT, Lencer WI, et al. Advances in Evaluation of Chronic Diarrhea in Infants. Gastroenterology (2018) 154:2045–59. doi:10.1053/j.gastro.2018.03.067

PubMed Abstract | CrossRef Full Text | Google Scholar

46. Boyle MP, Diener-West M, Milgram L, Knowles M, Foy C, Zeitlin P, et al. A Multicenter Study of the Effect of Solution Temperature on Nasal Potential Difference Measurements*. Chest (2003) 124:482–9. doi:10.1378/chest.124.2.482

PubMed Abstract | CrossRef Full Text | Google Scholar

47. Srinivasan B, Kolli AR, Esch MB, Abaci HE, Shuler ML, Hickman JJ. TEER measurement techniques for in vitro barrier model systems. J Lab Autom (2015) 20:107–26. doi:10.1177/2211068214561025

PubMed Abstract | CrossRef Full Text | Google Scholar

48. Sedar AW, Forte JG. Effects of Calcium Depletion on the Junctional Complex Between Oxyntic Cells of Gastric Glands. J Cell Biol (1964) 22(1):173–88. doi:10.1083/jcb.22.1.173

CrossRef Full Text | Google Scholar

49. Forte JG, Nauss AH. Calcium Movements in Isolated Bullfrog Gastric Mucosa. Am J Physiol (1966) 211(1):239–42. doi:10.1152/ajplegacy.1966.211.1.239

CrossRef Full Text | Google Scholar

Keywords: intestinal permeability, gut potential difference, transnasal probe, m-mode OCT imaging, duodenum diagnosis

Citation: Otuya DO, Gavgiotaki E, Carlson CJ, Shi SQ, Lee AJ, Krall AA, Chung A, Grant CG, Bhat NM, Choy P, Giddings SL, Gardecki JA, Thiagarajah JR, Rowe SM and Tearney GJ (2021) Minimally Invasive Image-Guided Gut Transport Function Measurement Probe. Front. Phys. 9:735645. doi: 10.3389/fphy.2021.735645

Received: 03 July 2021; Accepted: 31 August 2021;
Published: 20 September 2021.

Edited by:

Joseph Edward Hayward, Juravinski Cancer Centre, Canada

Reviewed by:

Qiyin Fang, McMaster University, Canada
Hongki Yoo, Korea Advanced Institute of Science and Technology, South Korea

Copyright © 2021 Otuya, Gavgiotaki, Carlson, Shi, Lee, Krall, Chung, Grant, Bhat, Choy, Giddings, Gardecki, Thiagarajah, Rowe and Tearney. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Guillermo J. Tearney, gtearney@partners.org

Download