ORIGINAL RESEARCH article

Front. Chem., 07 October 2025

Sec. Analytical Chemistry

Volume 13 - 2025 | https://doi.org/10.3389/fchem.2025.1672220

This article is part of the Research TopicAdvancements in Analytical ChemistryView all 4 articles

The detection of serum-volatile organic compounds in the diagnostics of hepatocellular carcinoma using gas chromatography–ion mobility spectrometry

Xin Shen&#x;Xin Shen1Chenglin Xie&#x;Chenglin Xie2Shijia Zhang,Shijia Zhang2,3Wen Ai,Wen Ai2,3Ziran Yu,Ziran Yu2,3Xu Ye
Xu Ye1*
  • 1Hunan Cancer Hospital, The Affiliated Cancer Hospital of Xiangya School of Medicine, Central South University, Changsha, China
  • 2The Affiliated Hospital of Hunan Academy of Traditional Chinese Medicine, Changsha, China
  • 3School of Pharmacy, Hunan University of Chinese Medicine, Changsha, China

Introduction: The third most common cause of cancer-related death worldwide is hepatocellular carcinoma (HCC), a malignant liver tumor that usually arises in patients with cirrhosis and chronic liver disease. New biomarkers are required for bidirectional validation of HCC in clinical practice because of its early asymptomatic stage, high mortality rate, and rapid tumor growth, as well as the high rate of false-positive results obtained for the commonly used serum marker alpha-fetoprotein (AFP).

Methods: This study used gas chromatography–ion mobility spectrometry (GC-IMS) combined with chemometrics to analyze the differences in volatile organic compounds (VOCs) in the serum of patients with and without hepatocellular carcinoma (HCC) in order to explore new biomarkers of HCC.

Results: The GC-IMS analysis detected 31 VOCs in serum, including 8 ketones, 11 aldehydes, 6 alcohols, 2 esters, and 1 furans, many of which differed in content between the controls and the patients with HCC.

Conclusions: The VOCs obtained from the use of these analytical devices can serve as a reference for developing low-cost equipment in the future. Considering its high efficiency and low cost, the detection of volatile organic compounds can be used as an auxiliary means of diagnosis or for mass screening of the population.

1 Introduction

Hepatocellular carcinoma (HCC) is a malignant liver tumor that commonly arises in patients with chronic liver disease and cirrhosis (Foglia et al., 2023). Its risk factors mainly include viral infections (hepatitis B and C), alcohol consumption, smoking, non-alcoholic fatty liver disease, and exposure to aflatoxin (Taniguchi, 2023; Zhang et al., 2024). It is the third most common cause of cancer-related death worldwide, with high incidence rates in East and Southeast Asia, Southern Europe, and parts of Africa (Nadarevic et al., 2022). It is evident that HCC is an important public health concern, and the diagnosis and monitoring of HCC are particularly crucial.

HCC has no typical symptoms in its early stages, and the tumor progresses rapidly with a high mortality rate (Wang et al., 2024). Currently, most guidelines recommend ultrasonography (US) for early HCC detection (Sharma et al., 2024). The detection sensitivity of US for HCC lesions varies depending on physician experience, ranging from 40% to 80% (Bannaga et al., 2021). Another test that can be used to detect HCC is the serum marker alpha-fetoprotein (AFP), but this test has the drawback of low sensitivity (high false positive results) (Piñero et al., 2020). The sensitivity limit of AFP for clinical HCC diagnosis is 65%, and for preclinical prediction, it is <40% (Luo et al., 2018). HCC can also be diagnosed using non-invasive imaging, such as computed tomography (CT) and magnetic resonance imaging (MRI), which present characteristic imaging features such as arterial hyperexcitability, venous washout, and capsule enhancement (Chan et al., 2024). However, these diagnostic instruments still lack sufficient sensitivity and specificity, and they require tissue biopsy to confirm uncertain HCC lesions detected during scanning. This often leads to a delayed diagnosis, especially in resource-scarce areas (Bannaga et al., 2021).

If HCC is diagnosed at an early stage, surgical interventions such as partial liver resection or liver transplantation can effectively treat it, with a median 5-year survival rate of about 70% (Li and Wang, 2015). Therefore, how to effectively screen populations for HCC using economic, simple, and reliable diagnostic methods to achieve early diagnoses and improve its monitoring and treatment effectiveness is an urgent clinical problem that needs to be solved.

One of the methods currently gaining attention for diagnosing and monitoring cancer is to detect volatile organic compounds (VOCs) in various biological samples, such as exhaled breath, serum, urine, and bile (Daulton et al., 2021; Gui et al., 2023; Tyagi et al., 2021). VOCs are considered to be potential systemic and local biomarkers that can provide unique information on physiological and biochemical processes, indicating an individual’s health status (Liu et al., 2023; Liu et al., 2024). Therefore, various techniques for measuring volatile organic compounds (VOCs)—such as GC-MS (Chapman et al., 2023; Lubes and Goodarzi, 2018), GC-IMS (Al-Difaie et al., 2024; Liu et al., 2024), electronic nose (Wilson and Forse, 2023), and even specially trained canines (Sonoda et al., 2011; Charles et al., 2024)—can be applied to biological samples including exhaled breath, blood, urine, and other bodily fluids for disease detection and prediction (Frerichs et al., 2023).

Serum contains abundant critical biomolecular information; therefore, it plays an important role in the early screening, auxiliary diagnosis, efficacy monitoring, and prognosis evaluation of various cancers. Compared with other biological samples, serum has the following advantages: serum samples can be collected randomly, at regular intervals, or 24 h a day (Zhang et al., 2012). Compared with other biological samples for volatile compound detection, serum has relatively stable biomarkers that are less susceptible to external interference, favoring long-term preservation and repeated testing (Liu et al., 2024). Additionally, serological testing is not affected by recent medication or local gastric lesions and can reflect the patient’s status over time. Serum sampling also has a high acceptance rate among the general population.

Biomarkers in urine may be disrupted by metabolic abnormalities, including stress, alcohol, smoking, certain foods, drugs, and various environmental factors, which can affect the accuracy of test results (Tyagi et al., 2021). Bile collection is challenging and usually requires invasive methods such as surgery or endoscopy, causing patient discomfort and risk. Breath testing may be affected by various factors such as bleeding from peptic ulcers, severe atrophy of gastric mucosa, and the use of medication before testing, potentially yielding false-negative results (Wildner-Christensen et al., 2002). From this, it can be seen that serum biomarker detection, a relatively less invasive and widely accepted detection method compared with other biological sample collection approaches for VOCs detection has the advantages of easy access and reproducibility of detection. This study aimed to investigate the differences in serum VOCs between patients with HCC and patients without HCC, potentially providing evidence for developing a method for early HCC screening.

2 Materials and methods

2.1 Materials

This study recruited a total of 6 patients with HCC and 2 patients without HCC as controls. The sample numbers for HCC patients are “LC-01 to LC-06,” while the sample numbers for non-HCC patients are NG-01 and NG-02. The research plan was approved by the Ethics Committee of Hunan Cancer Hospital, and informed consent was obtained from all participating patients and their families.

Serum sample collection: All serum samples were obtained from the participants on an empty stomach on the day post-admission. The serum samples were collected in inert, separate procoagulant vessels. The collected samples were centrifuged at 6,000 rpm for 10 min. The supernatant was extracted and stored immediately at −86 °C.

2.2 Analysis by GC-IMS

VOCs were analyzed directly via GC-IMS using the FlavorSpec® Gas-Phase Ion Mobility Spectrometer from G.A.S. (Dortmund, Germany).

This study analyzed eight serum samples, comprising six from hepatocellular carcinoma (HCC) patients and two from non-HCC individuals, designated as LC1-6 and NG1-2, respectively. The samples (0.2 mL) were placed in a 20 mL headspace vial, incubated for 20 min at 60 °C, and then 500 µL was injected into the headspace using a non-shunt injection; the vials were rotated at 500 rpm for 20 min (injection needle temperature: 85 °C). Each sample was measured in three parallel groups.

An MXT-WAX capillary column (15 m × 0.53 mm × 1.0 mm, Restek Inc., USA) was used, and the VOCs were maintained at 60 °C with high-purity N2 (99.999%) as the carrier gas for 20 min. The flow rate was initially 2.00 mL min−1 for 2 min; after 8 min, it was linearly increased to 10.00 mL min−1, then 100.00 mL min−1 for 10 min, and finally held for 10 min. The runtime was 20 min, and the injection temperature was 80 °C.

IMS conditions: the IMS reagent was 3H (tritium), a 53 mm drift tube was used, the electric field intensity was 500 V cm−1, the drift tube temperature was 45 °C, the flow rate of the carrier gas (high-purity N2) was 75 mL min−1, and the positive ion mode was used.

2.3 Statistical analysis

VOCal data processing software (from G.A.S., Dortmund, Germany, version 2.0.0) was used to generate three-dimensional, two-dimensional, differential, and fingerprint spectra of VOCs. Principal component analysis (PCA) was carried out using OmicShare Tools (Mu et al., 2024). TBtools and SIMCA (Version 14.1, Umetrics, Sweden) were used for CA and PLS-DA, respectively.

3 Results

3.1 Analysis of GC-IMS results

3.1.1 Comparative analysis of VOCs in serum

Figure 1 shows the three-dimensional gas-phase ion migration spectrum of VOCs in the serum of patients with and without HCC, generated using the Reporter plugin program in the GC-IMS instrument analysis software. The three coordinate axes represent the relative drift time (X-axis), retention time (Y-axis), and signal peak intensity (Z-axis), respectively. VOCs are represented by peaks, and strength is represented by the height of the red protrusion. A higher red protrusion indicates stronger signals and a higher component content, while a lower red protrusion indicates weaker signals and a lower component content. From Figure 1, it can be seen that there are certain differences in the serum VOCs of patients with HCC relative to those of patients without HCC.

Figure 1
Three-dimensional graph showing peak intensity, retention time, and drift time. Different labels, LC-01 to LC-06 and NG-01 to NG-02, are marked on red and blue lines, indicating varying intensities and times. The x-axis represents drift time in milliseconds, the y-axis retention time in seconds, and the z-axis peak intensity.

Figure 1. Three-dimensional spectrum of VOCs in the serum samples, analyzed via GC-IMS.

Figure 2 shows a two-dimensional top view of volatile compound gas-phase ion migration in different serum samples, with the horizontal axis is a reduced ion mobility and the vertical axis representing retention time. The red vertical line represents the peak of reactant ions, with each bright spot represent in analyte ion intensity. The color and area of the bright spot represent the amount of the volatile compound content. The darker the color and the larger the area of the bright spot, the higher the content of the volatile compound. Red represents a higher content of the corresponding compound, while blue represents a lower content. Figure 2 allows the differences in the volatile compound contents between patients with HCC and patients without HCC to be visually compared.

Figure 2
A series of eight vertical heatmaps labeled LC-01 to NG-02 represents measurement data over time in seconds and drift time. Each panel shows a range of blue hues with scattered bright spots, indicating varying intensity levels. The color scale ranges from blue to red, signifying low to high values. Each subplot is annotated with measurement details, and the horizontal and vertical axes are labeled with time-related and intensity metrics, respectively.

Figure 2. Two-dimensional GC-IMS spectra of VOCs in the serum samples.

Figure 3 shows the spectrum of the NC-01 sample selected as a reference. The reason for selecting NC-01 as the reference sample is primarily that the VOCs of serum in patients without HCC represent the baseline level. By comparing these components, we can identify which components in the serum of HCC patients deviate significantly from the normal range in terms of content or proportion, thereby screening out specific biomarkers associated with HCC. This lays the foundation for efficiently identifying differences in VOCs of serum related to HCC. The spectra of the other samples were subtracted from the reference to obtain a comparison chart of the differences between different samples. As long as the volatile organic compound content in the target sample and the reference sample is the same, the subtracted background will be white. In the spectra of other target samples, red indicates that the concentration of a compound in the target sample is higher than the reference concentration, and blue indicates that it is lower. We can clearly see the differences in ionized VOCs signals between patients with HCC and patients without HCC through the differential spectra.

Figure 3
Heat map comparing measurement run times and drift time across seven sections labeled NG-01, LC-01 to LC-06, and NG-02. The color gradient ranges from blue (lower values) to red (higher values). Each section shows variations in color intensity, indicating differences in data distribution along the drift time and measurement run axes.

Figure 3. GC-IMS differential spectrum of VOCs using sample NG-01 as the reference.

3.1.2 Qualitative analysis

To compare the relative differences of VOCs in the serum samples of patients with and without HCC, the data were processed using a VOCal plugin to obtain the ion migration spectra of the samples. After searching and comparing the GC retention index (NIST 2020) database and IMS drift time database, it was found that a total of 31 VOCs were detected in the samples, including 8 ketones, 11 aldehydes, 6 alcohols, 2 esters, and 1 furans. The qualitative results are shown in Table 1.

Table 1
www.frontiersin.org

Table 1. The VOCs detected in the samples.

3.1.3 GC-IMS fingerprint analysis

To further compare the VOCs in different serum samples, fingerprint analysis was performed on all VOCs, and the results are shown in Figure 4. This Figure shows all selected signal peaks from a sample in rows, and each volatile organic compound’s signal peak in columns. A grid shows a higher content if the color is brighter. Through comparative analysis of the volatile substances detected in different serum samples, it was found that there are significant differences in the VOCs detected in the serum of patients with and without HCC. The results indicate differences in the content of n-Octanal, n-Nonanal, (E)-2-Octenal, 2-phenylacetaldehyde, (E.E)-2.4-Heptadienal, 2-Pentyl furan, 1-Octen-3-ol, 1-Octen-3-one, (E)-2-Heptenal, Heptanal, 2-Hexenal, 1-Pentanol, (E)-2-Pentenal, 1-Penten-3-ol, Pentanal, 1-Penten-3-one and (Z)-2-Penten-1-ol. Their contents are all higher in NG-01 and NG-02; that is, their contents are higher in the serum of patients without HCC and are very low in the serum of patients with HCC. The contents of tetrahydrofuran, Benzaldehyde, and 1-Butanol were higher in the LC-06 sample, while they were very low in the other patients with HCC and in patients without HCC.

Figure 4
Heatmap displaying various chemical compounds with intensity levels indicated by color ranging from deep blue to bright yellow. Rows are labeled LC-01 to NG-02, columns correspond to different compounds. Red rectangles highlight specific areas with a higher concentration of signals.

Figure 4. Fingerprint of VOCs in the serum samples.

3.2 Chemometrics

3.2.1 Principal component analysis (PCA)

PCA is a data representation method used for feature extraction and dimensionality reduction. This method converts many potentially related variables into a small number of unrelated variables, providing an overview of category separation, clustering, and outliers. PCA visualizes the model, making it easy to understand and avoiding subjective judgments. PCA of the VOCs detected in the serum samples from patients with and without HCC was performed using Origin 2024b software for LC-01, LC-02, LC-03, LC-04, LC-05, LC-06, NG-01, and NG-02. The results are shown in Figure 5. The contribution rate of PC1 is 61.7%, the contribution rate of PC2 is 12.4%, and the cumulative contribution rate reaches 74.1%. The distances between samples LC-01, LC-02, LC-03, LC-04, and LC-05 are relatively small, indicating that their differences are small. However, there is a significant gap between the six LC groups and the two NG groups, indicating that there is a large difference in serum VOCs between patients with HCC and patients without HCC. The serum VOCs of patients with HCC and patients without HCC were classified using PCA, confirming that there is a significant difference in serum VOCs between these two groups of patients. The LC-06 contains some different VOCs, which may be related to its comorbidities.

Figure 5
Principal Component Analysis (PCA) plot showing data clusters labeled LC-01 through NG-02, using PC1 (61.7%) and PC2 (12.4%) axes. Different colored ellipses represent groupings.

Figure 5. PCA score plot of VOCs in the serum samples.

3.2.2 Partial least-squares discriminant analysis (PLS-DA)

To better observe the differences between groups, a PLS-DA model was established for patients with HCC and patients without HCC. PLS-DA is a statistical method for supervised discriminant analysis that is different from PCA. It can interpret observations and make predictions for corresponding variables. Eight sets of sample data were imported into SIMCA software for analysis, and the results are shown in Figure 6. In the PLS-DA model, R2 and Q2 are used to evaluate the model reliability and predictive ability, respectively. Values of R2 and Q2 greater than 0.5 indicate acceptable model fitting, while a value close to 1 indicates strong predictive ability. The PLS-DA scoring chart shows that the PLS-DA model has good predictive ability (R2X = 0.987, R2Y = 0.976, Q2 = 0.899). The values of R2 and Q2 are close to 1, indicating good fitting accuracy. LC-01, LC-02, LC-03, LC-04, and LC-05 are distributed on the right side of the graph, while NG-01 and NG-02 are on the left side of the graph, presenting distant clusters and indicating significant differences in VOCs in the serum of patients with HCC and patients without HCC; these findings are consistent with the conclusion drawn from the PCA graph. In addition, the predicted variable importance in projection (VIP)>1 is used as a standard to measure the strength and explanatory power of the expression pattern of the compounds for the classification discrimination of each group of samples, assisting in the screening of the main distinguishing peaks between samples. As shown in Figure 7, 3-(Methylthio)-1-propene, Acetone, Acetic acid ethyl ester, 2-Pentanone, 2-Butanone, 1-Butanol, 4-Heptanone, 4-Heptanone, Cyclohexanone, Ethyl 2-methylpropanate, Cyclohexanone, 2-Butanone, n-Nonanal, 3-Hexanone, 1-Butanol, n-Octanal, Hexanal, Tetrahydrofuran, Ethanol, n-Octanal, and Benzaldehydryl are the main indicator components of the difference. To determine whether the model is overfit, we conducted 200 cross-validations to examine the R2 and Q2 values. From the graph, we observed that the slope of the line is large, indicating that the PLS-DA model is not overfit (R2 = 0.339, Q2 = −0.889; as shown in Figure 8.

Figure 6
Scatter plot depicting data points differentiated by color according to the legend on the right: LC-01 (green), LC-02 (blue), LC-03 (red), LC-04 (yellow), LC-05 (cyan), LC-06 (purple), NG-01 (orange), and NG-02 (pink). Axes labeled t[1] and t[2]. An oval shape encompasses most data points, with some outliers outside the oval.

Figure 6. PLS−DA analysis of VOCs in eight groups of serum.

Figure 7
Bar chart displaying VIP scores for various compounds, with compounds listed on the x-axis and VIP scores on the y-axis ranging from -2 to 5. Error bars represent variability.

Figure 7. VIP values of the characteristic variables.

Figure 8
Scatter plot showing R2 values in green circles and Q2 values in blue squares across 200 permutations for 9 components. Dotted lines connect the values, indicating trends with ascending slopes. Horizontal axis represents permutations, and vertical axis ranges from -1.5 to 1.5.

Figure 8. Permutation test results of VOCs in eight groups of serum.

4 Discussion

This study utilized GC-IMS combined with chemometrics to analyze the effects of serum VOCs in patients with and without HCC. A total of 31 VOCs were detected, including 8 ketones, 11 aldehydes, 6 alcohols, 2 esters, and 1 furans. By using GC-IMS technology to obtain three-dimensional, two-dimensional, and color difference fingerprint spectra, we can preliminarily showed/could potentially visualize significant differences in the volatile compounds detected in the serum of patients with HCC and patients without HCC.

According to the results of the volatile compound fingerprint spectrum, it can be concluded that one of the patients with HCC (LC-06) exhibits unique specificity in serum VOCs, mainly featuring a higher content of Tetrahydrofuran, Benzaldehyde, and 1-Butanol, while their contents are very low in the patients in the other HCC groups and patients without HCC. Among these substances, the PLS-DA model, established using SIMCA, shows that the substances with VIP values greater than 1 are Tetrahydrofuran, Benzaldehyde, and 1-Butanol.

The content of n-Octanal, n-Nonanal, (E)-2-Octenal, 2-phenylacetaldehyde, (E.E)-2.4-Heptadienal, 2-Pentyl furan, 1-Octen-3-ol, 1-Octen-3-one, (E)-2-Heptenal, Heptanal, 2-Hexenal, 1-Pentanol, (E)-2-Pentenal, 1-Penten-3-ol, Pentanal, 1-Penten-3-one, and (Z)-2-Penten-1-ol were higher in the serum of patients without HCC and very low in the serum of patients with HCC. Most of these substances are aldehydes and alcohols. The PLS-DA model shows that the substances with VIP values greater than 1 are n-nonanal.

The VIP evaluation conducted on the eight groups of serum samples in this study significantly reduced data complexity and improved the classification and diagnostic performance of the discriminant model. In this study, VIP evaluation was used to exclude over 90% of the features from the discriminative model, resulting in better classification results. This highlights the importance of variable selection for model performance. Including a large number of variables in a discriminative model may prove ineffective and counterproductive, while selecting a smaller number of variables with significant correlations can lead to better performance. The use of appropriate stoichiometric data processing, combined with analytical methods that provide a wealth of information, has been proven to be decisive in achieving satisfactory results (Rodríguez-Hernández et al., 2023).

Serum VOCs testing, with its advantages of high efficiency and low cost, has the potential to become a promising method for understanding diseases through the development of reliable, sensitive, and reproducible diagnostic tests. After many years of exploration, the research on serum biomarkers has yielded fruitful results, and the analysis of many material indicators has greatly advanced the progress of clinical diagnosis of tumors (Liu et al., 2023; Liu et al., 2024). However, in-depth exploration of the diagnostic value of VOCs in the serum of cancer patients has been lacking. The detection of serum VOC markers, has a high acceptance rate among patients and the advantages of easy access and repeatability. It has become a new approach for tumor diagnosis, especially early diagnosis. However, its effectiveness still needs to be confirmed with further validation studies.

In recent years, gas chromatography-ion mobility spectrometry (GC-IMS) has shown unique advantages as a new analytical method for the detection and analysis of volatile organic compounds (Pan, et al., 2024). The results obtained using GC-IMS show its significance as an analytical tool, as this method is much simpler than using high-end analytical methods such as GC-MS and does not require a laboratory environment. At the same time, GC-IMS combines the excellent separation performance of gas chromatography with the high sensitivity of ion mobility spectrometry, and can be created in a portable form, using nitrogen or air as the carrier gas (Dai et al., 2024; Tyagi et al., 2021; Xiang et al., 2023). It can accurately analyze VOCs without the need for tedious sample pretreatment, greatly simplifying the analysis process (Zheng et al., 2024). This innovative method is not only expected to significantly improve diagnostic efficiency, but also significantly reduce the cost of diagnosis and treatment for patients. Furthermore, its minimally invasive nature can effectively improve patient cooperation and bring more convenience to clinical diagnosis. In the future, detecting serum VOCs in combination with other screening methods can enhance diagnostic performance and deepen the analysis of VOCs and their metabolic pathways, opening new paths for tumor diagnosis and treatment. We believe that the use of serum VOCs will be an important diagnostic tool in the future.

5 Limitations

The limitation of this study is that the sample size is too small. These results will need to be further verified in the clinic.

6 Conclusion

In this study, GC-IMS combined with chemometrics was utilized to analyse and compare the VOCs detected in the serum samples of patients with and without HCC. The results showed that there were significant differences in VOCs in the serum samples of patients with HCC and patients without HCC. Some VOCs were reduced in patients with HCC, mostly aldehydes and alcohols. The VOCs obtained from the use of these analytical devices can serve as a reference for developing low-cost equipment in the future. Considering its high efficiency and low cost, the detection of volatile organic compounds can be used as an auxiliary means of diagnosis or for mass screening of the population.

Data availability statement

The original contributions presented in the study are included in the article/supplementary material, further inquiries can be directed to the corresponding author.

Ethics statement

The studies involving humans were approved by the Ethics Committee of Hunan Cancer Hospital (No. KYJJ-2023-287). The studies were conducted in accordance with the local legislation and institutional requirements. The participants provided their written informed consent to participate in this study.

Author contributions

XS: Conceptualization, Investigation, Methodology, Writing – original draft. CX: Conceptualization, Investigation, Methodology, Writing – original draft. SZ: Software, Validation, Writing – original draft. WA: Formal Analysis, Software, Visualization, Writing – original draft. ZY: Data curation, Resources, Writing – original draft. XY: Conceptualization, Funding acquisition, Project administration, Supervision, Writing – review and editing.

Funding

The author(s) declare that financial support was received for the research and/or publication of this article. This work was supported by Hunan Provincial Natural Science Foundation of China (Grant No. 2024JJ5246).

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Generative AI statement

The author(s) declare that no Generative AI was used in the creation of this manuscript.

Any alternative text (alt text) provided alongside figures in this article has been generated by Frontiers with the support of artificial intelligence and reasonable efforts have been made to ensure accuracy, including review by the authors wherever possible. If you identify any issues, please contact us.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

Al-Difaie, Z. J. J., Scheepers, M. H. M. C., Engelen, S. M. E., Lubbers, T., Havekes, B., and Bouvy, N. D. (2024). Volatile organic compounds in exhaled breath, blood, and urine detected in patients with thyroid carcinoma using gas chromatography-ion mobility spectrometry-a pilot study. J. Breath Res. 19 (1), 016009. doi:10.1088/1752-7163/ad89ef

PubMed Abstract | CrossRef Full Text | Google Scholar

Bannaga, A. S., Tyagi, H., Daulton, E., Covington, J. A., and Arasaradnam, R. P. (2021). Exploratory study using urinary volatile organic compounds for the detection of hepatocellular carcinoma. Mol. Basel, Switz. 26, 2447. doi:10.3390/molecules26092447

PubMed Abstract | CrossRef Full Text | Google Scholar

Chan, Y. T., Zhang, C., Wu, J., Lu, P., Xu, L., Yuan, H., et al. (2024). Biomarkers for diagnosis and therapeutic options in hepatocellular carcinoma. Mol. Cancer 23, 189. doi:10.1186/s12943-024-02101-z

PubMed Abstract | CrossRef Full Text | Google Scholar

Chapman, E. A., Baker, J., Aggarwal, P., Hughes, D. M., Nwosu, A. C., Boyd, M. T., et al. (2023). GC-MS techniques investigating potential biomarkers of dying in the last weeks with lung cancer. Int. J. Mol. Sci. 24 (2), 1591. doi:10.3390/ijms24021591

PubMed Abstract | CrossRef Full Text | Google Scholar

Charles, M., Ruszkiewicz, D., Eckbo, E., Bryce, E., Zurberg, T., Meister, A., et al. (2024). The science behind the nose: correlating volatile organic compound characterisation with canine biodetection of COVID-19. ERJ open Res. 10 (3), 00007–02024. doi:10.1183/23120541.00007-2024

PubMed Abstract | CrossRef Full Text | Google Scholar

Dai, Y., Liu, S., Yang, L., He, Y., Guo, X., Ma, Y., et al. (2024). Explorative study for the rapid detection of Fritillaria using gas chromatography-ion mobility spectrometry. Front. Nutr. 11, 1361668. doi:10.3389/fnut.2024.1361668

PubMed Abstract | CrossRef Full Text | Google Scholar

Daulton, E., Wicaksono, A. N., Tiele, A., Kocher, H. M., Debernardi, S., Crnogorac-Jurcevic, T., et al. (2021). Volatile organic compounds (VOCs) for the non-invasive detection of pancreatic cancer from urine. Talanta 221, 121604. doi:10.1016/j.talanta.2020.121604

PubMed Abstract | CrossRef Full Text | Google Scholar

Foglia, B., Turato, C., and Cannito, S. (2023). Hepatocellular carcinoma: latest research in pathogenesis, detection and treatment. Int. J. Mol. Sci. 24, 12224. doi:10.3390/ijms241512224

PubMed Abstract | CrossRef Full Text | Google Scholar

Frerichs, N. M., El Manouni El Hassani, S., Deianova, N., van Weissenbruch, M. M., van Kaam, A. H., Vijlbrief, D. C., et al. (2023). Fecal volatile metabolomics predict gram-negative late-onset sepsis in preterm infants: a nationwide case-control study. Microorganisms 11, 572. doi:10.3390/microorganisms11030572

PubMed Abstract | CrossRef Full Text | Google Scholar

Gui, X., Zhang, X., Xin, Y., Liu, Q., Wang, Y., Zhang, Y., et al. (2023). Identification and validation of volatile organic compounds in bile for differential diagnosis of perihilar cholangiocarcinoma. Clin. chimica acta; Int. J. Clin. Chem. 541, 117235. doi:10.1016/j.cca.2023.117235

PubMed Abstract | CrossRef Full Text | Google Scholar

Li, L., and Wang, H. (2015). Heterogeneity of liver cancer and personalized therapy. Cancer Lett. 379, 191–197. doi:10.1016/j.canlet.2015.07.018

PubMed Abstract | CrossRef Full Text | Google Scholar

Liu, Q., Li, S., Li, Y., Yu, L., Zhao, Y., Wu, Z., et al. (2023). Identification of urinary volatile organic compounds as a potential non-invasive biomarker for esophageal cancer. Sci. Rep. 13, 18587. doi:10.1038/s41598-023-45989-1

PubMed Abstract | CrossRef Full Text | Google Scholar

Liu, Q., Li, S., Mao, M., Gui, X., Zhang, Y., Zhao, Y., et al. (2024). Serum-volatile organic compounds in the diagnostics of esophageal cancer. Sci. Rep. 14, 17722. doi:10.1038/s41598-024-67818-9

PubMed Abstract | CrossRef Full Text | Google Scholar

Lubes, G., and Goodarzi, M. (2018). GC-MS based metabolomics used for the identification of cancer volatile organic compounds as biomarkers. J. Pharm. Biomed. analysis 147, 313–322. doi:10.1016/j.jpba.2017.07.013

PubMed Abstract | CrossRef Full Text | Google Scholar

Luo, P., Yin, P., Hua, R., Tan, Y., Li, Z., Qiu, G., et al. (2018). A Large-scale, multicenter serum metabolite biomarker identification study for the early detection of hepatocellular carcinoma. Hepatol. Baltim. Md 67, 662–675. doi:10.1002/hep.29561

PubMed Abstract | CrossRef Full Text | Google Scholar

Mu, H., Chen, J., Huang, W., Huang, G., Deng, M., Hong, S., et al. (2024). OmicShare tools: a zero-code interactive online platform for biological data analysis and visualization. iMeta e228, e228. doi:10.1002/imt2.228

PubMed Abstract | CrossRef Full Text | Google Scholar

Nadarevic, T., Colli, A., Giljaca, V., Fraquelli, M., Casazza, G., Manzotti, C., et al. (2022). Magnetic resonance imaging for the diagnosis of hepatocellular carcinoma in adults with chronic liver disease. Cochrane Database Syst. Rev. 5 (5), CD014798. doi:10.1002/14651858.CD014798.pub2

PubMed Abstract | CrossRef Full Text | Google Scholar

Pan, Y., Qiao, L., Liu, S., He, Y., Huang, D., Wu, W., et al. (2024). Explorative study on volatile organic compounds of cinnamon based on GC-IMS. Metabolites 14 (5), 274. doi:10.3390/metabo14050274

PubMed Abstract | CrossRef Full Text | Google Scholar

Piñero, F., Dirchwolf, M., and Pessôa, M. G. (2020). Biomarkers in hepatocellular carcinoma: diagnosis, prognosis and treatment response assessment. Cells 9, 1370. doi:10.3390/cells9061370

PubMed Abstract | CrossRef Full Text | Google Scholar

Rodríguez-Hernández, P., Cardador, M. J., Ríos-Reina, R., Sánchez-Carvajal, J. M., Galán-Relaño, Á., Jurado-Martos, F., et al. (2023). Detection of Mycobacterium tuberculosis complex field infections in cattle using fecal volatile organic compound analysis through gas chromatography-ion mobility spectrometry combined with chemometrics. Microbiol. Spectr. 11. doi:10.1128/spectrum.01743-23

PubMed Abstract | CrossRef Full Text | Google Scholar

Sharma, K. K., Mohsin, M., Mittal, P., Ali, Z., Fatma, N., Upadhyay, P., et al. (2024). Diagnosis of the initial stage of hepatocellular carcinoma: a review. Curr. Pharm. Des. 30, 1708–1724. doi:10.2174/0113816128298875240321073907

PubMed Abstract | CrossRef Full Text | Google Scholar

Sonoda, H., Kohnoe, S., Yamazato, T., Satoh, Y., Morizono, G., Shikata, K., et al. (2011). Colorectal cancer screening with odour material by canine scent detection. Gut 60 (6), 814–819. doi:10.1136/gut.2010.218305

PubMed Abstract | CrossRef Full Text | Google Scholar

Taniguchi, H. (2023). Liver cancer 2.0. Int. J. Mol. Sci. 24, 17275. doi:10.3390/ijms242417275

PubMed Abstract | CrossRef Full Text | Google Scholar

Tyagi, H., Daulton, E., Bannaga, A. S., Arasaradnam, R. P., and Covington, J. A. (2021). Urinary volatiles and chemical characterisation for the non-invasive detection of prostate and bladder cancers. Biosensors 11, 437. doi:10.3390/bios11110437

PubMed Abstract | CrossRef Full Text | Google Scholar

Wang, J. X., Lu, L. G., and Cai, X. B. (2024). Endoscopic ultrasound for the diagnosis and treatment of primary hepatocellular carcinoma. J. Dig. Dis. 25, 156–162. doi:10.1111/1751-2980.13266

PubMed Abstract | CrossRef Full Text | Google Scholar

Wildner-Christensen, M., Touborg Lassen, A., Lindebjerg, J., and Schaffalitzky de Muckadell, O. B. (2002). Diagnosis of Helicobacter pylori in bleeding peptic ulcer patients, evaluation of urea-based tests. Digestion 66 (1), 9–13. doi:10.1159/000064421

PubMed Abstract | CrossRef Full Text | Google Scholar

Wilson, A. D., and Forse, L. B. (2023). Potential for early noninvasive COVID-19 detection using electronic-nose technologies and disease-specific VOC metabolic biomarkers. Sensors Basel, Switz. 23 (6), 2887. doi:10.3390/s23062887

PubMed Abstract | CrossRef Full Text | Google Scholar

Xiang, Y., Lei, C., Hu, G., Zhou, W., Li, Y., and Huang, D. (2023). Investigation of 60Co irradiation on the volatile organic compounds from finger citron (citri sarcodactylis fructus) using GC-IMS. Foods Basel, Switz. 12 (19), 3543. doi:10.3390/foods12193543

PubMed Abstract | CrossRef Full Text | Google Scholar

Zhang, A., Sun, H., and Wang, X. (2012). Serum metabolomics as a novel diagnostic approach for disease: a systematic review. Anal. Bioanal. Chem. 404, 1239–1245. doi:10.1007/s00216-012-6117-1

PubMed Abstract | CrossRef Full Text | Google Scholar

Zhang, Y., Xu, W., Peng, C., Ren, S., and Zhang, C. (2024). Intricate effects of post-translational modifications in liver cancer: mechanisms to clinical applications. J. Transl. Med. 22, 651. doi:10.1186/s12967-024-05455-6

PubMed Abstract | CrossRef Full Text | Google Scholar

Zheng, Y., Li, F., Zhao, C., Zhu, J., Fang, Y., Hang, Y., et al. (2024). Analysis and application of volatile metabolic profiles of Escherichia coli: a preliminary GC-IMS-based study. RSC Adv. 14, 25316–25328. doi:10.1039/d4ra03601h

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: hepatocellular carcinoma, biomarkers, serum, volatile organic compounds, GC-IMS, chemometrics

Citation: Shen X, Xie C, Zhang S, Ai W, Yu Z and Ye X (2025) The detection of serum-volatile organic compounds in the diagnostics of hepatocellular carcinoma using gas chromatography–ion mobility spectrometry. Front. Chem. 13:1672220. doi: 10.3389/fchem.2025.1672220

Received: 24 July 2025; Accepted: 22 September 2025;
Published: 07 October 2025.

Edited by:

Serge Cosnier, UMR5250 Département de Chimie Moléculaire (DCM), France

Reviewed by:

Anatolii Spesyvyi, J. Heyrovsky Institute of Physical Chemistry (ASCR), Czechia
Xinru Gui, Shandong University, China

Copyright © 2025 Shen, Xie, Zhang, Ai, Yu and Ye. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Xu Ye, eWV4dUBobmNhLm9yZy5jbg==

These authors have contributed equally to this work

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.