Unearthed opium: development of a UHPLC-MS/MS method for the determination of Papaver somniferum alkaloids in Daunian vessels

www.frontiersin.orgFlaminia Vincenti1 www.frontiersin.orgCamilla Montesano1* www.frontiersin.orgAlessandro Ciccola1 www.frontiersin.orgIlaria Serafini1 www.frontiersin.orgGabriele Favero2 www.frontiersin.orgMatteo Pallotta1 www.frontiersin.orgFlavia Pagano1,3 www.frontiersin.orgGaia Di Francesco1 www.frontiersin.orgMartina Croce1,3 www.frontiersin.orgMaria Laura Leone4 www.frontiersin.orgItalo Maria Muntoni5 www.frontiersin.orgManuel Sergi1
  • 1Department of Chemistry, Sapienza University of Rome, Rome, Italy
  • 2Department of Environmental Biology, Sapienza University of Rome, Rome, Italy
  • 3Department of Public Health and Infectious Disease, Sapienza University of Rome, Rome, Italy
  • 4Independent Researcher in Archaeology, Foggia, Italy
  • 5Soprintendenza Archeologia, Belle Arti e Paesaggio per le Province di Barletta—Andria—Trani e Foggia, Foggia, Italy

Introduction: The analysis of organic residue in ancient vessels to investigate early-age civilization habits is an important archeological application that needs advanced analytical methods. However, these procedures should meet inherent requisites such as low sampling invasiveness and high sensitivity for trace analysis. This study deals with the development of advanced analytical methods for the detection of opium alkaloids in ceramic vessels and its first application to the study of Daunian pots dating back to the VIII–IV sec BC.

Methods: All the stages of the analytical procedure, from sampling to analysis, were carefully optimized. Concerning sampling, the traditional scraping approach was compared with a swabbing strategy which permitted minimizing sample encroachment. Extraction was based on pressurized liquid extraction or ultrasound-assisted liquid extraction, followed by dispersive liquid–liquid microextraction, which allowed concentration enrichment. On the other hand, a UHPLC-MS/MS method was specifically developed and validated to obtain reliable data. Some Daunian pots, belonging to the Ceci-Macrini private archeological collection, were selected for sample withdrawal as their iconography could suggest opium usage.

Results: Several of the analyzed samples resulted positive to thebaine and less frequently to morphine and codeine; furthermore, 70% of the analyzed items tested positive for at least one opium alkaloid. Positive findings were common to all the samples collected in the pots, suggesting that scraping and swabbing provided comparable results and validating this unusual sampling strategy. All samples were additionally analyzed by UHPLC-HRMS to further improve the confidence level of the identified compounds. The obtained results shed new light on the hypothesis of opium usage by the ancient Daunian civilization. Furthermore, this study provided suitable analytical tools for further investigations on the same topic, with a good level of confidence in the quality of the results.

1 Introduction

The peculiar properties of many natural psychoactive plants were well known in antiquity; for example, Papaver somniferum, among the most ancient and studied, has been already known since the Sumerian civilization, and the extraction of psychoactive substances, in particular opium, was known by ancient Egyptians and practiced by Cypriots during the Bronze Age, handed down to ancient Greeks, and reached up to the Romans, who reported the first cases of dependence (Nencini, 1997). The use of these substances was mainly linked to medicine and food and, most likely, was part of ritual practices and contact with the afterlife (Kritikos, 1968).

A particular case is represented by the Daunians, inhabitants of the historical geographical district of northern Puglia, which, in ancient times, with Peucezia and Messapia, constituted Apulia (Figure 1A). This civilization left us a rich ancient collection of figured stelae and vases belonging to the VIII–IV century BC (Nava, 1998; Mazzei, 2010). These ancient stelae and vases show pictorial and morphological references that could suggest the use of opioids by Daunians, probably in a religious and healing context (Leone, 2002) (Figure 1B).

FIGURE 1
www.frontiersin.org

FIGURE 1. Map showing the localization of Daunia (Apulia IT) (A); magnification of a vessel from the Ceci-Macrini collection (ID 265) whose iconography could suggest opium usage; the similarity between the decoration and the poppy capsule is highlighted (B).

Similar observations on Cypriote Bronze Age base-ring juglets led some archeologists to propose a link between these vessels and opium trade (Merrillees, 1962), given their distinctive shape, which looks like the capsule of the opium poppy. For a long period, this theory has been controversial, but recently, reliable chemical evidence demonstrated that these vessels once contained opium (Smith et al., 2018; Linares et al., 2022). In several studies, it was shown that opium alkaloids can be detected even in archeological samples through highly sensitive analytical techniques. The first findings were related to morphine (Koschel, 1996), the most abundant alkaloid in opium, but lately, the presence of morphine was questioned because of its instability (Chovanec et al., 2012). Smith et al. reported, most recently, rigorous chemical evidence of more stable opium alkaloids, i.e., papaverine and thebaine, in a sealed British Museum base-ring juglet (Smith et al., 2018), and more recently, Linares et al. published further evidence of the presence of morphinan-based alkaloids in vessels, arising from Tel Yeud burials in Israel, dating to the Late Bronze Age IIA (Linares et al., 2022). All these studies showed that advanced analytical techniques may be a high-level ally for archeologists, thus becoming crucial to answer open archeometric questions; anyway, suitable and robust analytical protocols must be used to produce reliable and incontrovertible results.

The analysis of organic residues absorbed into the ceramic walls of ancient vessels in order to determine their past contents is widespread: wine, plant oils and resins, animal fats, etc., are among the most detected compounds (Briggs et al., 2022), also for Daunian archeological contexts (Leone et al., 2015; Muntoni et al., 2021). Sampling protocols, extraction conditions, and detection techniques largely depend on the type of analytes and the objectives of the study (Kałużna-Czaplińska et al., 2016). A major distinction can be drawn between noninvasive and destructive methods, taking into account that often, it is not possible to sample an archeological remain and when possible, the sample must be considered unique and unrepeatable. Several spectroscopic techniques (mainly Raman and Fourier-transform infrared spectroscopies) fall in the first group, but the advantage of being nondestructive is balanced by the limited confidence level for analyte identification. For this reason, mass spectrometric (MS) techniques are generally preferred (Rosiak et al., 2020; Cank et al., 2021) and may be coupled with separative techniques, i.e., gas chromatography (GC) and liquid chromatography (LC). Both GC-MS (Kałużna-Czaplińska et al., 2016) and LC-MS possess numerous benefits for the analysis of chemical residues and can be used to separate complex mixtures, determine trace levels of organic residues, and/or perform quantitative experiments; however, they generally require extensive sample preparation.

The analysis of ancient pottery generally involves the withdrawal of several samples from a relevant number of vessels (Irto et al., 2022), whole potsherds, or more frequently, small amounts of material can be collected by scraping the interior surfaces of the pots (Sanjurjo-Sánchez et al., 2018). The collected residues represent very complex matrices; therefore, it is necessary to perform an adequate sample treatment to purify and isolate the target compounds. The literature data about alkaloid extraction show that solvent-based extractions are the most common. Smith et al. extracted the target analytes by ultrasonication with hydrochloric acid, while solid-phase extraction (SPE) was used for sample clean-up. A mixture of dichloromethane:methanol (2:1; v:v) (Linares et al., 2022) or chloroform:isopropanol (3:1; v:v) (Koschel, 1996) was used for extraction by other authors prior to GC-MS analysis.

The aim of this study was to corroborate the hypothesis of opium usage in ancient Daunia through the analysis of organic residues collected in several Daunian vessels belonging to the private Ceci-Macrini archeological collection (Rossi, 1979). In order to obtain rigorous evidence of the presence of opium alkaloids in these samples, a UHPLC-MS/MS method was specifically developed and validated. Furthermore, sample preparation procedures based on pressurized liquid extraction (PLE) or ultrasound-assisted liquid extraction, together with dispersive liquid–liquid microextraction (dLLME) sample enrichment, were exploited to maximize the sensitivity of the method. Two kinds of sampling approaches were carried out: scraping was compared with a less-invasive swabbing approach using commercially available wipes.

2 Materials and methods

2.1 Standards and reagents

The standards of morphine, codeine, and thebaine were supplied by LGC Standard (Milan, Italy) in methanol solution at 1 mg mL−1; internal standard morphine-d3 was purchased from the same company as a crystalline powder with a 98% degree of purity; IS working solution (IS-WS) was prepared in methanol at 100 ng mL−1.

A standard working mixture containing the three analytes at 1 μg mL−1 in methanol was obtained by the appropriate dilution of the standards and was stored at −20°C.

Water, methanol, acetonitrile, 2-propanol, and chloroform, all LC-MS grade, were supplied by Fisher Chemical (Rodano (MI), Italy). Formic acid was provided by VWR Chemicals (Milan, Italy). Cetyltrimethylammonium bromide (CTAB), sodium carbonate, sodium bicarbonate, and ammonium formate were purchased from Merck Life Science S.r.l. (Milan, Italy). A 25 mM CTAB solution in water was prepared for PLE.

2.2 Sample collection

Different samples from the 14 Daunian pots belonging to the Ceci-Macrini private collection located in Andria (Apulia, IT) were collected as reported in Table 1. The sampling protocol was performed in agreement with the Italian Ministry of Culture (MiC)—Soprintendenza Archeologia, Belle Arti e Paesaggio per le Province di Barletta-Andria-Trani e Foggia.

TABLE 1
www.frontiersin.org

TABLE 1. List of the analyzed vessels, with photographs, description, and dating information. For each vessel, several samples were collected and labeled on the basis of the sample strategy ((S)—scraping; (AP)—Alco-prep; and (PE) and (PW)—cotton pads soaked with ethanol and water, respectively.).

Different sampling techniques were implemented; in all cases, the samples were obtained in minimally invasive procedures, considering the historical, artistic, and cultural value of the vessels. To prevent exogenous contamination through hands, nitrile gloves were always worn during the handling of the vessels. First, different samples were obtained by swabbing the interior surface of the vessels with an Alco-prep® (a ready-to-use swab saturated with 70% 2-propanol) and, if allowed by the sampling surface, also with cotton pads soaked in ethanol or water. In addition, if visible debris (ground, ceramic, etc.) were found in the pot, an aliquot (200–500 mg) of the residue was collected by scraping with a disposable scalpel the interior surface of the vessel. The collected samples were labeled as follows: (S)—scraping; (AP)—Alco-prep; and (PE) and (PW)—cotton pads soaked with ethanol and water, respectively. A total of 63 samples were collected.

To avoid cross-contamination, all samples were collected in separate Eppendorf tubes, sealed, and sent to the laboratory of Sapienza University of Rome; the samples were treated separately.

2.3 Sample preparation

2.3.1 (S) sample extraction

Alkaloid extraction for (S) samples was performed by PLE. The instrument used was an ASE 200 Accelerated Solvent Extractor from Dionex (Sunnyvale, CA, United States) equipped with 24 stainless steel extraction cells of 1 mL.

An aliquot of 100 mg of each sample was homogenized with diatomaceous earth as the supporting material; the mixture, spiked with 10 μL of IS-WS at 100 ng mL-1, was used to fill the extraction cell which was secured at the top and bottom with cellulose filters. Then, the cell was loaded into the instrument. The extracting solvent consisted of a mixture of 0.25 mM CTAB:2-propanol (80:20; v:v). A single extraction cycle was performed with the following parameters: T = 150°C, p = 1,450 psi, preheat time = 2 min, heat time = 7 min, static time = 5 min, and purge time = 1 min. About 7.5 mL of the extract was collected for each sample in a separate 50 mL vial sealed with PTFE solvent-resistant septa. The extract was transferred into a 15 mL conical tube and centrifuged at 2,600 g for 5 min to promote the sedimentation of all possible solid residues; the supernatant was collected in a separate conical tube.

2.3.2 (AP), (PE), and (PW) sample extraction

All swab types were extracted by in syringe ultrasound-assisted extraction (iS-UAE), a technique specifically developed for this purpose. In brief, following the removal of the plunger, the swab was transferred in a disposable 10 mL syringe, parafilm-sealed at the luer lock; 1 mL of 2-propanol containing the IS-WS at 10 ng mL-1 was added directly into the syringe. The plunger was then loosely reassembled, and the syringe was vortex-shaken for 30 s and then immersed for 10 min in an ultrasonic bath thermostated at 25°C. By squeezing the swab using the syringe plunger, the extract was collected in a 15 mL conical tube and centrifuged as described for (S) samples. The supernatant was added to a conical tube containing 6.5 mL of a 0.25 mM CTAB solution.

2.4 Clean-up

Sample clean-up was carried out by dLLME. For each sample, the supernatant, obtained as previously described, was added with 0.7 mg of NaCl and 1 mL of carbonate buffer at pH 9.5. Once homogenized, 0.5 mL of 2-propanol (disperser solvent) and 0.2 mL of chloroform (extraction solvent) were rapidly injected into the solution. The obtained sample was vortex-mixed for 30 s, sonicated at 25°C for 10 min, and then, centrifuged at 2,600 g for 5 min at 4°C to promote chloroform sedimentation. The latter was collected, evaporated under a gentle nitrogen flow, and finally, reconstituted with 100 μL of water:methanol (50:50; v:v). 3 μL was injected into the UHPLC system.

2.5 UHPLC-MS/MS analysis

The UHPLC-MS/MS system consisted of an ExionLC AD System from SCIEX (Framingham, MA, United States) coupled with a QTRAP 6500+, hybrid quadrupole-linear ion trap mass spectrometer from SCIEX, equipped with an ESI source, operating in the positive ion mode. Data acquired were elaborated by using SCIEX OS software (version 2.0.1.).

Analyte separation was performed using an Acquity UPLC BEH C18 (100 × 2.1 mm i.d., packed with 1.7 µm particles) from Waters. The column was maintained at 40°C. Mobile phases were 5 mM ammonium formate (phase A) and 0.1% HCOOH in MeCN:MeOH (75:25; v:v) (phase B), and the flow was 400 μL min-1. The elution gradient was structured as follows: phase B was maintained at 0% for 0.5 min and then increased to 20% in 2 min, to 30% in 3.1 min, to 40% in 2 min, then to 50% in 2.5 min, and finally, to 100% in 1.1 min. This condition was maintained for 2 min. The total chromatographic run time, including the equilibration stage between two different analyses, was 16.5 min.

QTRAP source parameters were set as follows: curtain gas 30 units, ion source gas 1 and 2 at 55 and 60 units, respectively, ionspray voltage 5500 V, and source temperature 450°C.

To perform targeted identification, increasing compound identification confidence at the same time, an information-dependent acquisition (IDA) approach was developed: multiple reaction monitoring (MRM) was used as the survey scan, while enhanced product ion (EPI) was selected as the dependent scan. For morphine, thebaine, and codeine, for which standards were available, all parameters were optimized by direct infusion of the relative standards; for the other compounds of interest, MRM transitions were inferred from the literature data (Mueller et al., 2005; Nesměrák et al., 2019). The IDA criteria were set to conduct EPI scans for each detected MRM transition which exceeded 5,000 cps (selected threshold). The five most intense signals were selected during every scan. This mode has the advantage of providing a concomitant product ion spectrum when the required conditions are fulfilled. The EPI scan range was between 50 and 450 Da, while the scan rate was 10,000 Da s-1. All spectra obtained were subsequently matched with the commercial library mzCloud and an in-house library. MS parameters are listed in Table 2.

TABLE 2
www.frontiersin.org

TABLE 2. MRM method parameters..

2.6 UHPLC-HRMS/MS analysis

High-resolution mass spectrometry (HRMS) was used for untargeted analysis and to further improve the confidence level of the identified compounds. The UHPLC equipment consisted of a Dionex UltiMate® 3000 RSLC system (Thermo Fisher Scientific, San Jose, CA) comprising an auto-sampler equipped with a 100 μL loop and vacuum degasser. Analyte identification was performed on a Q-Exactive Orbitrap (Thermo Fisher Scientific, Bremen, Germany) equipped with a heated electrospray ionization source (H-ESI) operating in both negative and positive ion modes. Untargeted analyses were conducted according to a previously published UHPLC-HRMS/MS method (Vincenti et al., 2020). In brief, analyses were carried out in a full-scan/data-dependent (full MS/dd-MS2) acquisition mode, with an inclusion list to preferentially select specific ions for the MS/MS scan, as listed in Table 3. The Q-Exactive resolution was set at 35,000 FWHM for full-scan analyses, while MS/MS experiments were carried out at a resolution of 17,500 FWHM. Three different normalized collision energies were applied to obtain the MS/MS spectra, namely, 10, 30, and 50.

TABLE 3
www.frontiersin.org

TABLE 3. Inclusion list of precursor ions for HRMS ddMS2 analysis.

Compound Discoverer 3.3 was used for raw data analysis; a generic workflow for retention time alignment, component detection, and elemental composition prediction was used.

2.7 Validation

The proposed UHPLC-MS/MS quantitative method was fully validated for the target analytes, evaluating the following parameters: LOD, LOQ, linearity, selectivity, specificity, IS interference, carryover, accuracy, precision, matrix effect, and recovery.

The LOD was considered as the lowest absolute quantity that provided a signal-to-noise ratio (S/N) for the qualifier ion of 3. For this purpose, five different blank soil samples were collected and pooled together to obtain a representative blank matrix; pooled blank matrix samples were fortified with decreasing quantities of the standard in duplicate, processed as described previously, and finally, analyzed in three different chromatographic runs. For what concerns swabs, decreasing amounts of standards were spiked on a clean surface and sampled. LOQ was determined similarly, and it was considered the lowest quantity of analyte providing S/N = 10 for the qualifier ion. For cotton pads and Alco-prep, the following procedure was used: all swabs were fortified and left to dry to simulate the time elapsed between sampling and analysis.

Linearity was evaluated by preparing six calibration standards in water:methanol (50:50; v:v), and the linearity range was from LOQ to 10 ng for all analytes. The internal standard ratio was used to evaluate the calibration samples analyzed in three different chromatographic runs.

To evaluate method selectivity, three blank samples for each sampling type were processed as previously described, and the absence of any interference at analyte retention time was verified to provide an unambiguous identification of the target analytes. Even the possible interference produced by the presence of morphine-d3 was evaluated. To this aim, three samples for each sampling type were fortified with the IS-WS and treated as described previously.

Carryover was evaluated by verifying the absence of any signal above the LOD in all blank samples analyzed after the injection of the highest point of the calibration curve.

Intra-day precision and accuracy were evaluated through the analysis of several QC samples fortified with different quantities of standard, i.e., 0.5, 2.5, and 10 ng. Three different QC samples were prepared for each sample type and for each level. Precision was calculated as the relative standard deviation (RSD%), while accuracy was defined in terms of bias as the relative deviation (%) of the mean calculated quantity compared to the spiked quantity for each level. In both cases, validation parameters were considered acceptable in a range of ±20% for accuracy, while precision was required to be <20% for each analyte.

Matrix effect and recoveries were evaluated at three levels; pooled ground blank matrices, Alco-prep, and cotton pads were fortified with 20 µL of working solutions at 25, 125, and 500 ng mL-1. The matrix effect (ME%) due to ion suppression and/or enhancement was calculated for each analyte by comparing the absolute peak area obtained from the analysis of a blank sample spiked after the extraction with the absolute peak area of a reference solution at the same concentration. ME% was considered acceptable in a range of ±20%. Recoveries (R%) were calculated by comparing the absolute peak area of a sample spiked before extraction with the absolute peak area of a blank sample spiked after extraction.

3 Results and discussion

3.1 UHPLC-MS/MS

As previously described, different mass spectrometric approaches were considered in this study. A QTRAP mass spectrometer was selected to obtain high selectivity and sensitivity, taking into account that, when present, low amounts of opium alkaloids were expected to be found in organic residues from ancient vessels. The MRM-IDA-EPI acquisition mode was selected in order to selectively detect the target analytes while obtaining fragmentation spectra for the detected compounds. This last feature permits increasing the confidence level for compound identification. When analytical standards are available (this was the case for thebaine, codeine, and morphine), this additional information is not crucial; however, for putative MRM transitions, the possibility of matching the spectra with MS/MS libraries can be of utmost importance for compound identity confirmation without standards.

In summary, the MRM-IDA-EPI acquisition mode allowed obtaining several identification points (IPs), i.e., the retention time of the peak, the presence of two overlapping precursor-to-product ion transitions, the ion ratio, and possibly, a complete MS/MS spectrum that can be compared with standards or library spectra. The simultaneous collection of all these IPs led to the confirmation of the analyte’s presence in the sample. Furthermore, this approach may possibly detect additional opium-related compounds, which were not detected by MRM analysis.

3.2 UHPLC-HRMS/MS

A Q-exactive high-resolution mass spectrometer was designated for untargeted analysis and to further confirm the identity of the analytes detected in low resolution. Sample acquisition was conducted in full MS/dd-MS2; in these conditions, untargeted full-scan MS was followed by product ion scan of the five most intense peaks. This strategy allowed detecting unexpected compounds that can be putatively assigned based on the exact mass and the isotopic pattern; in addition, the presence of fragmentation spectra for the most intense ions can assist in the process of compound annotation. For analyte identification, the accurate mass of the precursor and fragment ions was taken into account, with a maximum tolerance of 5 ppm. An inclusion list with target opioids, as well as their degradation products (Nesměrák et al., 2019), was incorporated into the method in order to preferentially fragment these target compounds. The proprietary software Compound Discoverer was used to assist untargeted analysis.

The results found in low resolution were confirmed by UHPLC-HRMS/MS; no further analytes were found by untargeted analysis.

3.3 Sampling and sample preparation

Sample collection from vessels of archeological significance must be carried out with caution; low invasiveness is the watchword. Sampling of organic residues from ancient pots is typically carried out by scraping the interior surface of the containers. In the present study, scraping was only performed when visible debris were noticeable inside the vessel; this was the case for almost all the samples analyzed (Table 1).

In order to compare this established sampling approach with a less-invasive strategy based on swabbing, the same vessels were sampled with both techniques. To this aim, different types of swabs were investigated: Alco-prep are single-packed, commercially available swabs saturated with 2-propanol; two other solvents were then tested, i.e., water and ethanol, and standard cotton pads were soaked in them. The preliminary results obtained in the laboratory, by spiking the three available opioids on a ceramic mortar, showed that all the tested swabs were effective in collecting them from the surface. The results on real samples are shown in Table 4 and discussed as follows in the § real sample analyses.

TABLE 4
www.frontiersin.org

TABLE 4. UHPLC-MS/MS quantitative results for the analyzed samples. The calculated amount is reported in ng of analyte.

3.3.1 dLLME clean-up

Keeping in mind that a significant enrichment factor would be beneficial, because of the very low concentrations expected, dLLME was selected as the clean-up protocol. This liquid–liquid microextraction technique exploits ternary solvent mixtures to achieve effective extraction of medium-to-high hydrophobic compounds, with decreased amount of solvent and time; high enrichment factors are generally attained. Several parameters were monitored to maximize analyte recovery: the nature of the dispersing solvent, ionic strength, and pH were found to be particularly relevant. In the first attempt, independent of the conditions used, the recovery for morphine was extremely low. An explanation for this behavior is that morphine is charged at almost every pH value because of the co-presence of tertiary amino and phenolic hydroxy groups in its structure, so its partitioning in the organic phase is dampened. The addition of an ion pair reagent, i.e., CTAB, turned out to be a successful strategy (Figure 2); in fact, the positive charge of CTAB may interact with the negatively charged hydroxy group of morphine to promote its partitioning into the organic phase.

FIGURE 2
www.frontiersin.org

FIGURE 2. Influence of the dispersing solvent nature (A), pH (B), ionic strength (C), and CTAB concentration (D) on dLLME recoveries.

3.3.2 PLE and iS-UAE methods

Given that several types of samples were considered in this study, different strategies were developed for analyte extraction. Samples obtained by scraping were extracted by PLE, while swabs were extracted by iS-UAE. This technique was specifically designed to minimize sample handling and avoid the need of cutting the swab into several pieces to insert them in the PLE cell. Regardless of the extraction method and taking into account the complexity and variability of the analyzed samples, sample clean-up was deemed of utmost importance to ensure high quality of the data in terms of reproducibility and ion suppression.

For what concerns (S) sample extraction, PLE was selected. This technique uses solvents at high temperature and pressure to extract analytes from solid or semi-solid samples; because of the high operating temperature and pressure, the organic solvent can be replaced by aqueous solutions, significantly reducing the environmental impact of the extraction procedure. Before loading into the extraction cell, the samples were manually ground and mixed with an inert material (diatomaceous earth) to ensure their uniform dispersion, avoiding aggregation of sample particles and allowing good solvent–sample contact within the extraction cell. In order to find suitable conditions for opioid extraction from a matrix as similar as possible to the sample under investigation, blank soil samples of different types (sandy and clayish) were mixed, and several aliquots were spiked with known amounts of the available standards. Among the several parameters that may influence the PLE recovery rate, the composition of the extraction solvent is of utmost importance. Based on previous studies (Vincenti et al., 2019) and taking into account the subsequent clean-up step, water:2-propanol (90:10, v:v) with 0.25 mM CTAB was tested. This extracting solution yielded good results in terms of recovery (>50% for all the analytes) and was then selected for PLE.

As already mentioned, for (PW), (PE), and (AP) samples, a different protocol was developed. iS-UAE stands for in-syringe ultrasound-assisted extraction; in fact, in this case, the swab is packed in a disposable syringe, and isopropanol extraction is promoted by acoustic cavitation.

3.4 Method validation

LODs and LOQs were below 1 ng for all the analytes, confirming that the developed protocols have high sensitivity and are then suitable for opiate residue analysis. Values are reported in Table 5. Linearity was investigated from LOQ to 10 ng, allowing the quantification over several orders of magnitude.

TABLE 5
www.frontiersin.org

TABLE 5. LOD, LOQ, and linearity for the three analytes based on the sampling strategy.

Selectivity, together with the possible interference of IS, was evaluated by the analysis of several swab and soil blank samples. Three additional blank samples for every sampling type were fortified with morphine-d3. In addition, it must be pointed out that two blank samples, one containing only the IS, were analyzed with every batch of real samples. The absence of any signals belonging to non-deuterated morphine, as well as the absence of every signal of interest, was observed in all the analyzed blanks. These observations showed that the method is selective and the detection of opiates in real samples may not be attributable to the contamination occurred during extraction and/or analysis.

Concerning carryover, all the analyzed blanks did not show any signals at the target analyte retention time, demonstrating that the method was not affected by carryover.

Three different absolute quantities were considered to evaluate precision and accuracy, i.e., 0.5, 2.5, and 10 ng. According to the results listed in Table 6, both extraction protocols yielded optimal results and RSD% was lower than 20%, while bias ranged from −11%–28% for morphine at the lower level tested. The same level was used to verify the matrix effect and recoveries; independent of the extraction protocol, the matrix effect was shown to be not significant, while recoveries were deemed suitable. As expected, because of the use of high temperature and pressure, PLE yielded better recoveries for all the tested analytes compared to iS-UAE (Table 6).

TABLE 6
www.frontiersin.org

TABLE 6. Validation results: precision, accuracy, recoveries, and matrix effects.

3.5 Real sample analyses

In this study, several samples collected from different Daunian pots from the Ceci-Macrini collection were analyzed as described previously. A total of 14 pots were selected based on the available archeological information, and special attention was paid to vessels whose iconography could suggest opium usage. An example is shown in Figure 1.

Interestingly, several of the analyzed samples tested positive to thebaine and less frequently to morphine and codeine. Quantitative results expressed as ng of the analyte in the analyzed sample are provided in Table 4, where it can be noted that 10 vessels, or 70% of the analyzed items, tested positive for at least one opium alkaloid. Among the detected analytes, thebaine was the most recurrent, and this finding was not surprising considering that several authors showed that thebaine, together with papaverine, is among the most stable opium alkaloids (Chovanec et al., 2012; Smith et al., 2018). It should also be pointed out that, with the developed method, the recovery was higher for thebaine; this also may have contributed to the prevalent detection of this opioid in the samples. In our samples, neither with low- nor high-resolution mass spectrometry, papaverine was detected, while morphine and codeine, which were known to degrade more rapidly, were observed in some samples. An explanation for these observations can be found in the relative amounts of these alkaloids in crude opium; in fact, it has been reported that several cultivars of poppy straw have very low amounts of papaverine (Stranska et al., 2013). For what concerns morphine and codeine, the detection of these alkaloids is in accordance with other reports related to ancient vessels (Koschel, 1996; Linares et al., 2022); it can be postulated that even if morphine is relatively unstable, its detection in ancient vessels is possible due to its high concentration in the original content. This hypothesis is reinforced by the data provided by Nesmerak et al., who analyzed opium preparations aged more than 200 years and found that in some samples, morphine was still the main component (Nesměrák et al., 2019). In our samples, several factors may have contributed to the preservation of morphinan alkaloids and, especially, the fact that the vessels have been buried for a long period, with a low presence of oxygen.

In this regard, two hypotheses can be drawn for the four vessels (211, 317, 333, and 337) that showed no traces of alkaloids (<LOD): on one hand, it can be postulated that they were not used for the storage and/or consumption of opium and/or its preparations; on the other hand, it cannot be excluded that degradation may have occurred. In particular, the filter pot n. 337, which was selected because of the iconography which recalls the papaver plant, was extremely damaged both externally and internally. Conversely, it is extremely interesting that the vessel n. 265, already highlighted in Figure 1, because of the decorations that recall the poppy capsule, tested positive not only to thebaine but also to the less-stable morphine and codeine, suggesting the original presence of high amounts of opium alkaloids in this vessel. The jug 274, which showed decorations with the plant, tested positive only for thebaine.

The presence of the three alkaloids among some of the analyzed samples was confirmed by low- and high-resolution mass spectrometry; as an example, Figure 3 shows a positive finding of thebaine (a) and morphine (b).

FIGURE 3
www.frontiersin.org

FIGURE 3. Example of a positive finding (vessel 265). The retention time of the peak, the presence of two overlapping precursor-to-product ion transitions, the ion ratio, and the mirror plot of the experimental MS/MS and library spectra are shown for thebaine (A) and morphine (B).

For what concerns the different collected samples, it is very interesting to note that, generally, positive findings are common to all the samples collected in a pot, independent of the sampling strategy; similarly, no trace of alkaloids was detected in all the samples collected from the four vessels. These results suggest that scraping and swabbing provided comparable results; the latter sampling strategy is by far less diffused for organic residue analysis on archeological samples; however, considering the lower invasiveness, it could represent an interesting new approach for sample collection in an archeological context.

4 Conclusion

With the aim of verifying the presence of opioids in a selection of ancient vessels belonging to the Ceci-Macrini collection, suitable analytical methods were developed and validated.

Both targeted and untargeted analyses were performed to assess the presence of opium alkaloids, such as morphine, codeine, and thebaine. The obtained results, confirming the presence of opioids in 10 samples, shed new light on the hypothesis of opium usage in ancient Daunia. It is important to highlight that trace analyte detection was made possible by the developed analytical method, which is characterized by high sensitivity.

These results, which are included in a wide project joining analytical chemistry and archeology, are very promising, as they allowed assessing the presence of psychoactive substances in ancient samples. Furthermore, this study provided suitable analytical tools for further investigations on the same topic, with a good confidence level in the quality of the results. New analytical methods for sample preparation were presented and applied on real samples, and some interesting improvements were presented in these kinds of investigations. From an archeometrical perspective, a central result is the validation of sample collection by swabbing, and we demonstrated that it can be a valid alternative to scraping and it can be of particular relevance for archeological samples.

Future developments of the study will concern the inclusion of new target analytes such as scopolamine or other hallucinogenic substances; in fact, in the present analysis, extraction was specifically tailored for opiate detection so that other interesting molecules could be undetected. In addition, new Daunian vessels, arising from adequately documented archeological excavations, will be analyzed in future studies in order to confirm and validate the hypothesis of opium usage in ancient Daunia and, moreover, to make chronological and historical comparisons.

Data availability statement

The raw data supporting the conclusion of this article will be made available by the authors, without undue reservation.

Author contributions

FV: investigation, data acquisition, and writing—original draft; CM: writing—original draft, data analysis, and conceptualization; AC: investigation and data analysis; IS: investigation and data analysis; GF: supervision, and writing—reviewing and editing; MP: investigation and data acquisition; FP: investigation and data acquisition; GD: investigation and data acquisition; MC: investigation and data acquisition; ML: conceptualization and editing; IM: supervision and writing—reviewing and editing; and MS: conceptualization, methodology, and writing—reviewing and editing. All authors contributed to the article and approved the submitted version.

Funding

Article processing charges were funded by D-ART Srl.

Acknowledgments

The authors acknowledge the Archaeological Superintendence of Barletta-Andria-Trani e Foggia (Soprintendenza Archeologia, Belle arti e Paesaggio per le province di Barletta-Andria-Trani e Foggia) for having allowed the present investigation on a collection that is subjected to the constraints of cultural heritage in Italy. They thank Santina Ceci Macrini, who is the owner of the Ceci-Macrini collection, and Lorello Macrini and Marco Macrini, the collection curators, for having allowed the sampling of the vessels and for the support and the feedback received.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors, and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

References

Briggs, L., Demesticha, S., Katzev, S., Wylde Swiny, H., Craig, O. E., and Drieu, L. (2022). There’s more to a vessel than meets the eye: Organic residue analysis of ‘wine’ containers from shipwrecks and settlements of ancient Cyprus (4th–1st century bce). Archaeometry 64, 779–797. doi:10.1111/arcm.12747

CrossRef Full Text | Google Scholar

Cank, K. B., Henkin, J. M., Cook, A. G., and Oberlies, N. H. (2021). Droplet probe: A non-destructive residue analysis of wari ceramics from the imperial heartland. J. Archaeol. Sci. 134, 105468. doi:10.1016/j.jas.2021.105468

CrossRef Full Text | Google Scholar

Chovanec, Z., Rafferty, S., and Swiny, S. (2012). Opium for the masses. Ethnoarchaeology 4, 5–36. doi:10.1179/eth.2012.4.1.5

CrossRef Full Text | Google Scholar

Irto, A., Micalizzi, G., Bretti, C., Chiaia, V., Mondello, L., and Cardiano, P. (2022). Lipids in archaeological pottery: A review on their sampling and extraction techniques. Molecules 27, 3451. doi:10.3390/molecules27113451

PubMed Abstract | CrossRef Full Text | Google Scholar

Kałużna-Czaplińska, J., Rosiak, A., Kwapińska, M., and Kwapiński, W. (2016). Different analytical procedures for the study of organic residues in archeological ceramic samples with the use of gas chromatography-mass spectrometry. Crit. Rev. Anal. Chem. 46, 67–81. doi:10.1080/10408347.2015.1008130

PubMed Abstract | CrossRef Full Text | Google Scholar

Koschel, K. (1996). Opium alkaloids in a cypriote base ring I vessel (bulbil) of the middle Bronze age from Egypt. Ägypten Und Levante/Egypt Levant. 6, 159.

Google Scholar

Kritikos, P. G., and Papadaki, S. (1968). A history of opium in antiquity. J. Am. Pharm. Assoc. 8, 446–447. doi:10.1016/s0003-0465(16)30721-2

PubMed Abstract | CrossRef Full Text | Google Scholar

Leone, M. L. (2002). “Sacred opium botany in Daunia (southern Italy) from the VII to VI century BC,” in Journal of psychoactive plants and compounds (Trento, Italy: Museo Civico di Rovereto), 71–82. http://www.artepreistorica.com/2010/01/sacred-opium-botany-in-daunia-italy-from-the-7-to-6-centuries-bc/.

Google Scholar

Leone, M., Munzi, P., Pouzadoux, C., Garnier, N., Leguilloux, M., and Muntoni, I. M. (2015). Measure the sacred at arpi: The share of food and drink in two votive deposits, in: Proceedings of the 1st International Conference on Metrology for Archaeology, Benevento, October 2015, 425–430.

Google Scholar

Linares, V., Jakoel, E., Be’eri, R., Lipschits, O., Neumann, R., and Gadot, Y. (2022). Opium trade and use during the Late Bronze Age: Organic residue analysis of ceramic vessels from the burials of Tel Yehud, Israel. Archaeometry, 1–18. doi:10.1111/arcm.12806

CrossRef Full Text | Google Scholar

Mueller, C. A., Weinmann, W., Dresen, S., Schreiber, A., and Gergov, M. (2005). Development of a multi-target screening analysis for 301 drugs using a QTrap liquid chromatography/tandem mass spectrometry system and automated library searching. Rapid Commun. Mass Spectrom. 19, 1332–1338. doi:10.1002/rcm.1934

PubMed Abstract | CrossRef Full Text | Google Scholar

Mazzei, M. (2010). I Dauni. Archeologia dal IX al V secolo a.C. Foggia. Foggia FG, Italy, Claudio Grenzi Editore.

Google Scholar

Merrillees, R. S. (1962). Opium trade in the Bronze age levant. Antiquity 36, 287–292. doi:10.1017/S0003598X00036814

CrossRef Full Text | Google Scholar

Muntoni, I. M., and Garnier, N. (2021). “Dauniens et Samnites. La tombe du guerrier de Chiancone (Pietramontecorvino, Foggia) entre habitat et espaces funéraires,” in Manger, boire, se parfumer pour l’éternité: Rituels alimentaires et odorants en italie et en gaule du IXe siècle avant au ier siècle après J-C. Editors D. Frère, B. Del Mastro, P. Munzi, and C. Pouzadoux (Napoli: Centre Jean Bérard), 191–198.

CrossRef Full Text | Google Scholar

Nava, M. L. (1998). Le stele della Daunia: Dalle scoperte di Silvio ferri agli studi piu recenti. Milano: Electa.

Google Scholar

Nencini, P. (1997). The rules of drug taking: Wine and poppy derivatives in the ancient world. IX. Conclusions. Subst. Use Misuse 32, 2111–2119. doi:10.3109/10826089709035623

PubMed Abstract | CrossRef Full Text | Google Scholar

Nesměrák, K., Kudláček, K., Štícha, M., Kozlík, P., Červený, V., and Kunešová, J. (2019). Degradation of the opium alkaloids in pharmaceutical relics from the eighteenth century. Monatsh. Fur Chem. 150, 1593–1602. doi:10.1007/s00706-019-02461-z

CrossRef Full Text | Google Scholar

Rosiak, A., Kałużna-Czaplińska, J., and Gątarek, P. (2020). Analytical interpretation of organic residues from ceramics as a source of knowledge about our ancestors. Crit. Rev. Anal. Chem. 50, 189–195. doi:10.1080/10408347.2019.1602821

PubMed Abstract | CrossRef Full Text | Google Scholar

Rossi, F. (1979). Ceramica geometrica daunia nella collezione Ceci Macrini. Bari: Dedalo.

Google Scholar

Sanjurjo-Sánchez, J., Kaal, J., and Fenollós, J. L. M. (2018). Organic matter from bevelled rim bowls of the Middle Euphrates: Results from molecular characterization using pyrolysis-GC–MS. Microchem. J. 141, 1–6. doi:10.1016/j.microc.2018.05.001

CrossRef Full Text | Google Scholar

Smith, R. K., Stacey, R. J., Bergström, E., and Thomas-Oates, J. (2018). Detection of opium alkaloids in a Cypriot base-ring juglet. Analyst 143, 5127–5136. doi:10.1039/c8an01040d

PubMed Abstract | CrossRef Full Text | Google Scholar

Stranska, I., Skalicky, M., Novak, J., Matyasova, E., and Hejnak, V. (2013). Analysis of selected poppy (Papaver somniferum L) cultivars: Pharmaceutically important alkaloids. Ind. Crops Prod. 41, 120–126. doi:10.1016/j.indcrop.2012.04.018

CrossRef Full Text | Google Scholar

Vincenti, F., Montesano, C., Cellucci, L., Gregori, A., Fanti, F., Compagnone, D., et al. (2019). Combination of pressurized liquid extraction with dispersive liquid liquid micro extraction for the determination of sixty drugs of abuse in hair. J. Chromatogr. A 1605, 360348. doi:10.1016/j.chroma.2019.07.002

PubMed Abstract | CrossRef Full Text | Google Scholar

Vincenti, F., Montesano, C., Di Ottavio, F., Gregori, A., Compagnone, D., Sergi, M., et al. (2020). Molecular networking: A useful tool for the identification of new psychoactive substances in seizures by LC–HRMS. Front. Chem. 8, 572952. doi:10.3389/fchem.2020.572952

PubMed Abstract | CrossRef Full Text | Google Scholar

Keywords: opium alkaloids, archeological ceramics, pressurized liquid extraction, mass spectrometry, Daunian

Citation: Vincenti F, Montesano C, Ciccola A, Serafini I, Favero G, Pallotta M, Pagano F, Di Francesco G, Croce M, Leone ML, Muntoni IM and Sergi M (2023) Unearthed opium: development of a UHPLC-MS/MS method for the determination of Papaver somniferum alkaloids in Daunian vessels. Front. Chem. 11:1238793. doi: 10.3389/fchem.2023.1238793

Received: 12 June 2023; Accepted: 07 July 2023;
Published: 26 July 2023.

Edited by:

Marta Concheiro-Guisan, John Jay College of Criminal Justice, United States

Reviewed by:

Ana Miguel Fonseca Pego, John Jay College of Criminal Justice, United States
Michelle Carlin, Rutgers University Camden, United States

Copyright © 2023 Vincenti, Montesano, Ciccola, Serafini, Favero, Pallotta, Pagano, Di Francesco, Croce, Leone, Muntoni and Sergi. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Camilla Montesano, camilla.montesano@uniroma1.it

Download