Corrigendum: Analytical approaches for the evaluation of data deficient simulated leachable compounds in ENDS products: a case study

www.frontiersin.orgCameron Smith* www.frontiersin.orgMatthew Lyndon www.frontiersin.orgLena Jeong www.frontiersin.orgDanielle Lehman www.frontiersin.orgJ. Brian Jameson www.frontiersin.orgHarish Chevva www.frontiersin.orgFelix Ayala-Fierro www.frontiersin.orgDavid Cook www.frontiersin.orgKaren Carter www.frontiersin.orgMichael Oldham www.frontiersin.orgI. Gene Gillman
  • Juul Labs Inc., Washington, DC, United States

A Corrigendum on
Analytical approaches for the evaluation of data deficient simulated leachable compounds in ENDS products: a case study

by Smith C, Lyndon M, Jeong L, Lehman D, Jameson JB, Chevva H, Ayala-Fierro F, Cook D, Carter K, Oldham M and Gillman IG (2023). Front. Chem. 11:1212744. doi: 10.3389/fchem.2023.1212744

In the published article, there was an error in Eq. 1: Calculation used for Analytical Reporting of TCEQ and NNMA as published. A sample dilution factor was omitted, and therefore, values were overestimated by a factor of approximately 40. The equation previously stated:

AnalyteTCEQorNNMAPeakAreaISTDPeakArea×ISTD Conc.0.04μgmL×Fill Wt.in JUULpod780mgDevicexDensity10.8430mLmg=EstimatedConcentrationofTCEQorNNMAμgDevice(1)

The corrected equation is shown below. The e-liquid weight (approximately 50 mg or 0.05 g) in 1 mL ultrapure water containing internal standard was used in the updated equation (per 2 Experimental methods, 2.3 e-liquid sample preparation and analysis using LC-HR-MS/MS and LC-MS/MS negative mode ESI in the original publication), and density was removed as the weight to volume is addressed. Note, fill wt. in JUULpod was modified from “780 mg” to “0.780 g” to keep units consistent; however, these are the same values.

AnalyteTCEQorNNMAPeakAreaISTDPeakArea×ISTD Conc.0.04μgmL×SampleVolumemLeLiquidAliquitWt.g×Fill Wt.in JUULpod0.780gramDevice=EstimatedConcentrationofTCEQorNNMAμgDevice(1)

Regardless of the estimated concentrations in the e-liquid, the analytical approach for the calculation of the transfer efficiency discussed in the publication (3 Results, 3.3 Determination of method limits and transfer efficiency of TCEQ and NNMA) is based on a 1:40 dilution. A hypothetical example below is used to illustrate the point using Eq. 2: Calculation of Transfer Efficiency for TCEQ and NNMA.

EsitmatedConcentrationofAnalyteTCEQorNNMAinAerosolμgdeviceEstimated Concentration of AnalyteTCEQ or NNMAin JUULpod eLiquid μgdevice×100%

Example:

0.25or0.025or0.0025μgdevice10or1.0or0.1μgdevice×100%=2.5%

If estimated concentrations in the e-liquid were 10, 1.0 or 0.1 and diluted by 40 to achieve an experimental LOQ of 0.25, 0.025 or 0.0025 µg/device, respectively, the resulting transfer efficiency remains 2.5%. As stated in the publication, the analytical approach is useful when traditional analytical approaches that utilize reference standards are not available and semi-quantitative values are needed for risk assessment.

The updated equation was used to update several values reported in the original publication (see updated text corrections below). Note, no additional data was collected, nor any additional experiments performed.

A correction has been made to section 3 Results, 3.1 Confirmation analysis of TCEQ and NNMA in unflavored aged JUULpods, 3rd paragraph. The affected sentences previously stated:

“Using Eq. 1, concentrations of TCEQ and NNMA were estimated in the unflavored e-liquid removed from ambient aged JUULpods. Concentrations ranged from 4.7 to 5.7 µg/device for TCEQ and 4.9–6.0 µg/device for NNMA.”

The corrected sentence appears below:

“Using Eq. 1, concentrations of TCEQ and NNMA were estimated in the unflavored e-liquid removed from ambient aged JUULpods. Concentrations ranged from 0.118 to 0.123 µg/device for TCEQ and 0.105–0.151 µg/device for NNMA.”

A correction has been made to section 3 Results, 3.3 Determination of method limits and transfer efficiency of TCEQ and NNMA. The affected sentences previously stated:

“The calculated experimental LOQs for TCEQ and NNMA were determined to be 0.12 µg/device for both compounds. Because all values for aerosol samples collected from aged JUULpods showed no trace or detectable levels of TCEQ or NNMA, the experimentally determined LOQ of 0.12 µg/device was used for the calculation of the transfer efficiency according to Eq. 2. The transfer efficiency for TCEQ and NNMA based on estimated concentrations from the simulated leachable study (see Table 1) were calculated to be approximately 1.6% according to Eq. 2.”

The corrected sentences appears below:

“The calculated experimental LOQs for TCEQ and NNMA were determined to be 0.003 µg/device for both compounds. Because all values for aerosol samples collected from aged JUULpods show no trace or detectable levels of TCEQ or NNMA, the experimentally determined LOQ of 0.003 µg/device was used for the calculation of the transfer efficiency according to Eq. 2. The transfer efficiency for TCEQ and NNMA based on estimated concentrations was calculated to be approximately 2.0%–2.8% according to Eq. 2.”

A correction has been made to section 4 Discussion. The sentence previously stated:

“The novel analytical approach provided experimentally determined LOQs of 0.12 µg/device for each leachable compound in which estimated transfer efficiencies were calculated to be less than 2%.”

The corrected sentence appears below:

“The novel analytical approach provided experimentally determined LOQs of 0.003 µg/device for each leachable compound in which estimated transfer efficiencies were calculated to be less than 3%.”

A correction has been made to the Abstract. The sentence previously stated:

“The transfer efficiency of each leachable compound was experimentally determined to be less than 2% based on the limit of quantitation, which then could be used to define a relevant exposure limit for the toxicological risk assessment.”

The corrected sentence appears below:

“The transfer efficiency of each leachable compound was experimentally determined to be less than 3% based on the limit of quantitation, which then could be used to define a relevant exposure limit for the toxicological risk assessment.”

In the published article, there was an error in Table 1 as published. The original table read, “Estimated Concentration at 30°C/65% RH for 22 weeks (Simulated 18-month aging)” (1st column, 5th row).

The corrected Table 1 appears below.

TABLE 1
www.frontiersin.org

TABLE 1. Information on data deficient leachable compounds.

The authors apologize for these errors and state that this does not change the scientific conclusions of the article in any way. The original article has been updated.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Keywords: ENDS, e-cigarette, electronic cigarette, nicotine, leachables, risk assessment, aerosol

Citation: Smith C, Lyndon M, Jeong L, Lehman D, Jameson JB, Chevva H, Ayala-Fierro F, Cook D, Carter K, Oldham M and Gillman IG (2023) Corrigendum: Analytical approaches for the evaluation of data deficient simulated leachable compounds in ENDS products: a case study. Front. Chem. 11:1334736. doi: 10.3389/fchem.2023.1334736

Received: 07 November 2023; Accepted: 14 November 2023;
Published: 28 November 2023.

Approved by:

Frontiers Editorial Office, Frontiers Media SA, Switzerland

Copyright © 2023 Smith, Lyndon, Jeong, Lehman, Jameson, Chevva, Ayala-Fierro, Cook, Carter, Oldham and Gillman. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.