Abstract
MR quantitative T1ρ mapping has gained increasing attention due to its capability to study low-frequency motional processes and chemical exchange in biological tissues. At ultra-high fields, the chemical exchange and proton diffusion in biological tissues should be more prominent. In this study, for the first time, we aim to test the feasibility of brain T1ρ mapping at 5.0 T MR scanner and compare the T1ρ values estimated using 3.0 T and 5.0 T scanners. Preliminary experimental results show that 5.0 T achieves T1ρ-weighted images with a higher signal-to-noise ratio than those acquired at 3.0T. The SNR benefit at 5.0 T is more obvious in high-resolution imaging. The T1ρ quantifications at 5.0 T are: Corpus callosum (67.4 ± 1.9 ms), Corona radiate (71.5 ± 1.8 ms), Superior frontal gyrus (67.6 ± 2.5 ms), Putamen (58.9 ± 1.2 ms), Centrum semiovale (84.0 ± 6.3 ms). Statistical analysis results indicate that the T1ρ values at 5.0 T show no significant difference with those obtained at 3.0 T (all p > 0.05). The interfield agreements in terms of T1ρ values between 3.0 T and 5.0 T were substantial (all ICCs >0.7). The coefficients of variation for T1ρ measurements from 3.0 T to 5.0 T were all less than 6.50% (2.28%–6.32%).
Introduction
Magnetic spin-lattice relaxation in the rotation frame (referred to as T1ρ) is an emerging technique to assess neurodegenerative diseases [1]. Relative to the conventional T1 and T2 relaxations, T1ρ provides a more specific probe of motional interactions related to the exchanges that are on the time scale of the reciprocal of the spin lock field strength. It is sensitive to the local tissue microenvironment and microstructure, such as the pH and glucose levels. Previous studies have demonstrated that proton exchange between bulk water and labile protons of protein or metabolites is an important contributor to the low-frequency T1ρ dispersion in biological tissue [2–5]. Therefore, T1ρ has been considered a potential biomarker for the evaluation and early diagnoses of degenerative neurologic diseases, such as multiple sclerosis [6–8], Alzheimer’s disease [9–11], Parkinson disease [12–14], and stroke [15,16].
T1ρ quantification is performed by fitting a series of T1ρ-weighted images acquired at varying spin-lock durations (TSL), which can be obtained using most magnetic resonance imaging (MRI) sequences by including a spin-lock preparation pulse at the beginning of the sequence. The T1ρ imaging technique is sensitive to the B0 and B1 field inhomogeneities and usually has a high specific absorption rate (SAR) due to the spin-lock pulse. Therefore, most T1ρ studies were performed at typically 1.5 T or 3.0 T fields, and human studies of T1ρ quantification at ultra-high fields were rarely reported [17–19]. At the ultra-high field, chemical exchanges between sites of different chemical shifts increase rapidly with field strength and may significantly contribute to the rotating frame relaxations [20]. T1ρ relaxation occurs in response to the chemical exchange between the groups of spins, which depends on chemical shift, temperature, exchange rate of the exchanging spins, etc. [21]. It is hence well suited for probing the metabolic information at higher magnetic field strengths [18,22]. Furthermore, high-resolution T1ρ quantification can be achieved using an ultra-high field MR scanner with a high signal-to-noise ratio (SNR) and affords improved morphological detail compared to imaging at lower-field strength [7] in an effort to improve visualization of brain lesions.
Last year, the first 5.0 T ultra-high field whole-body MR system (Jupiter, United Imaging Healthcare, Shanghai, China) was delivered. Compared with the 7.0 T ultra-high field system, the magnetic field strength of 5.0 T system has fewer SAR and field inhomogeneity. These advantages may bring benefits to brain T1ρ quantification. We implemented a three-dimensional (3D) T1ρ mapping sequence in this system, using a spin-lock T1ρ preparation pulse followed by a gradient recalled echo (GRE) readout, and applied it to imaging the brain of healthy volunteers. In this study, we will report the T1ρ quantification of the human brain at 5.0 T for the first time. Also, the SNR and spatial resolution gains of brain T1ρ images obtained from 5.0 T are demonstrated, and corresponding T1ρ values obtained from 5.0 T are compared with those obtained from 3.0 T.
Materials and methods
Subjects
The experiments were approved by the local Institutional Review Board (IRB). Twelve healthy volunteers (10 males, age: 30 ± 5 years; 2 females, age: 28 ± 4 years) were recruited in this study, and informed consent was obtained from each volunteer before the scan. All subjects participated in this study were scanned at a 5.0 T scanner (Jupiter, United Imaging Healthcare, Shanghai, China) and a 3.0 T scanner (uMR 890, United Imaging Healthcare, Shanghai, China). A modified T1ρ preparation pulse sequence based on a previously reported sequence was used [17].
Imaging sequence
The imaging sequence is depicted in Figure 1. The pulse sequence starts with a 90-degree RF pulse followed by crusher gradients to reset the net magnetization, which assures the signal is constant at the beginning of T1ρ preparation pulse [23]. The following recovery time (Trec) allows the recovery of the magnetization before the T1ρ preparation pulse and the acquisition. An adiabatically prepared constant-amplitude on-resonant spin-lock preparation pulse is used for T1ρ preparation [17], which is robust to B0 and B1 inhomogeneities. It consists of a rectangular locking pulse sandwiched by an adiabatic half passage (AHP) pulse and a reverse adiabatic half passage (RAHP) pulse[24,25]. The adjusted hyperbolic secant pulses are used as the AHP and RAHP. After the T1ρ preparation pulse, the residual transverse magnetization is destroyed by a gradient crusher. Then a 3D segmented radiofrequency gradient echo (GRE) sequence is employed for signal acquisition using the centric phase encoding order [26,27].
FIGURE 1

The timing diagram of the 3D T1ρ imaging sequence. The T1ρ preparation pulse was accomplished using an adiabatically prepared constant-amplitude on-resonant spin-lock approach to compensate for the B0 and B1 inhomogeneities at 5.0 T.
The 3D T1ρ imaging was performed with spin-lock pulse amplitude B1sl = 500 Hz and TSL = 0, 25, 45, and 65 ms at the 5.0 T and 3.0 T scanners. The 5.0 T MR scanner used a local quadrature birdcage transmit and 48-channel receiver head coil [28]. At the 3.0 T MR scanner, a commercial 32-channel receiver head coil (Rx: 32 channel) was used for T1ρ imaging. Three datasets were acquired for each subject at each scanner. Two datasets were acquired with a regular resolution of 1 × 1 × 3 mm3 in the sagittal and coronal orientations, respectively. Another was acquired with a high resolution of 0.65 × 0.65 × 2.5 mm3 in the transverse orientation. The imaging parameters are depicted in Table 1.
TABLE 1
| Regular resolution | High resolution | |||
|---|---|---|---|---|
| 3.0 T | 5.0 T | 3.0 T | 5.0 T | |
| Coil | 32-ch head | 48-ch head | 32-ch head | 48-ch head |
| FOV (mm2) | 240 × 200×30 | 240 × 200×30 | 200 × 200×24 | 200 × 200×24 |
| Voxel size (mm3) | 1 × 1×3 | 1 × 1×3 | 0.65 × 0.65×2.5 | 0.65 × 0.65×2.5 |
| TR/TE (ms) | 5.55/2.00 | 7.48/3.70 | 7.89/3.90 | 7.86/3.90 |
| ETL length | 50 | 50 | 50 | 50 |
| Bandwidth (Hz/pixel) | 500 | 500 | 500 | 500 |
| Spin-lock frequency (Hz) | 500 | 500 | 500 | 500 |
| TSLs (ms) | 0,25,45,65 | 0,25,45,65 | 0,25,45,65 | 0,25,45,65 |
| Phase encoding oversampling factor | 0.15 | 0.15 | 0 | 0 |
| Slice oversampling factor | 0.1 | 0.1 | 0 | 0 |
| Block time (ms)a | 1200 | 1800 | 1200 | 2000 |
| Scan time (mins) | 13:20 | 19:82 | 13:32 | 22:32 |
Imaging parameters for T1ρ quantification of the human brain at 5.0 T and 3.0 T
Due to system hardware constraints, the block time for imaging at 5.0 T was set longer in the high-resolution scenario than that in the regular resolution scenario.
Data analysis
All T1ρ quantification and analysis were performed in Matlab R2017b (MathWorks, Natick, MA, United States). The T1ρ maps were estimated using the exponential model [29,30] by fitting the T1ρ-weighted images with different TSLs pixel-by-pixel:where is the image intensity obtained at varying TSLs, M0 is the baseline image intensity without applying the spin-lock pulse. T1ρ map was estimated using the nonlinear least-squares fitting method with the Levenberg–Marquardt algorithm [31] from the T1ρ-weighted images.
Five regions of interest (ROIs) were manually drawn on the images of each volunteer by two independent readers (with 5-year experience in neural imaging), including Corpus callosum, Corona radiate, Superior frontal gyrus, Putamen, and Centrum semiovale. The average T1ρ value in each ROI was calculated, and the correlation of the T1ρ measurements between 3.0 T and 5.0 T was analyzed.
The SNRs of T1ρ-weighted images obtained from 5.0 T were also calculated and compared with those obtained from 3.0 T. The SNR was defined as the ratio between the mean value of the imaging regions and the standard deviation (SD) of the background noise [32]. To reduce the subject bias, four ROIs were drawn on each slice to calculate the noise SD and the SNR. Final SNR (denoted as SNRT1ρ-w) was calculated as the average of these four SNRs. Mean values and SDs of SNRT1ρ-w were computed for all the volunteers in each orientation.
Statistical analysis
The difference and consistency of T1ρ values measured at 3.0 T and 5.0 T were statistically compared using the Mann-Whitney U test. The intraclass correlation coefficients (ICCs) and Bland–Altman analysis were used to evaluate the interfield agreements and to determine whether there were significant differences in terms of T1ρ values between 3.0 T and 5.0 T. The interfield agreement was considered to be poor for ICCs = 0.0–0.2, fair for ICCs = 0.2–0.4, moderate for ICCs = 0.4–0.6, substantial for ICCs = 0.6–0.8, and excellent for ICCs = 0.8–1.0. p < 0.05 was considered statistically significant. The coefficient of variation (CV = stdT1ρ/meanT1ρ) was used to assess the variability and reliability of T1ρ values measured at 3.0 T and 5.0 T. CVs <15%, between 15% and 35% and >35% were considered to be small, moderate, and large variability. Statistical analyses were performed using SPSS 25.0 (IBM, Armonk, NY) and MedCalc 20.0.22 (MedCalc Software, Mariakerke, Belgium).
Results
Imaging with regular resolution
Figure 2 shows the representative T1ρ-weighted images at TSL = 0 and 65 ms of one subject with the regular resolution at 3.0 T and 5.0 T in the sagittal and coronal orientations, respectively. The noticeable difference between the image contrasts of 3.0 T and 5.0 T might be due to the change of relaxation time with the field strength. In addition, visible noise can be observed in the T1ρ-weighted images at 3.0 T, especially at TSL = 65 ms, since the signal attenuates with the increase of TSL. The corresponding T1ρ maps of the above T1ρ-weighted images are shown in Figure 3. These maps show extensive spatial similarities present between 3.0 T and 5.0 T T1ρ values for the same volunteers, except that the T1ρ maps at 3.0 T are a little noisier than those at 5.0 T.
FIGURE 2

The T1ρ-weighted images with different TSLs collected at 3.0 T and 5.0 T.
FIGURE 3

The T1ρ maps obtained at 3.0 T and 5.0 T.
Imaging with high resolution
Figure 4 shows T1ρ-weighted images of different TSLs and the corresponding T1ρ maps in the high-resolution scenario for an axial slice from two subjects acquired at 3.0 T and 5.0 T. Similar conclusions to the regular scenario can be obtained. Furthermore, the SNRs of T1ρ-weighted images obtained at 3.0 T drop significantly due to the increased resolution. As the signal decays exponentially along the TSL direction, some image details are almost drowned out by noise, especially at longer TSLs. Correspondingly, the T1ρ maps also seem noisy. However, the T1ρ-weighted images obtained at 5.0 T still show good image detail with reasonable SNRs, and the image quality of T1ρ maps is also improved.
FIGURE 4

T1ρ-maps estimated from datasets collected at 3.0 T and 5.0 T.
T1ρ measurements and signal-to-noise ratio analysis
The T1ρ values in the tissue compartments of the selected ROIs at 5.0 T and 3.0 T are shown in Table 2. There was no significant difference between the T1ρ values obtained at 5.0 T and 3.0 T (all p > 0.05). The ICCs for evaluating the interfield agreements are also shown in Table 2. LCI and UCI are the abbreviations of the lower and upper bounds of the 95% confidence interval. The ICCs of T1ρ values in all tissue compartments between 3.0 T and 5.0 T were as follows: ICCCorpus callosum = 0.730 [95% confidence interval (CI): 0.499–0.865]; ICCCorona radiate = 0.725 (95% CI:−0.466–0.869); ICCSuperior frontal gyrus = 0.788 (95% CI: 0.576–0.901); ICCPutamen = 0.788 (95% CI: 0.461–0.882) and ICCCentrum semiovale = 0.768 (95% CI: 0.542–0.890). All ICCs between the 3.0 T and 5.0 T were >0.7, indicating substantial agreements.
TABLE 2
| 3.0 T | 5.0 T | ICC | LCI | UCI | p-value | ||
|---|---|---|---|---|---|---|---|
| Regular resolution | Corpus callosum | 66.8 ± 2.3 | 67.2 ± 1.9 | 0.730 | 0.499 | 0.865 | 0.258 |
| Corona radiate | 71.7 ± 1.9 | 71.5 ± 1.7 | 0.725 | 0.466 | 0.869 | 0.243 | |
| Superior frontal gyrus | 67.6 ± 2.1 | 68.1 ± 2.4 | 0.788 | 0.576 | 0.901 | 0.342 | |
| Putamen | 59.2 ± 1.5 | 58.7 ± 1.3 | 0.741 | 0.461 | 0.882 | 0.399 | |
| High resolution | Centrum semiovale | 84.0 ± 4.7 | 83.5 ± 5.3 | 0.768 | 0.542 | 0.890 | 0.541 |
Average T1ρ values of selected ROIs and the corresponding p-values at 3.0 T and 5.0 T
The SNR of the T1ρ-weighted images obtained at 5.0 T and 3.0 T are shown in Table 3. As expected, the image SNR at 5.0 T was higher than that at 3.0 T, and significant differences were seen between the SNRs obtained at 5.0 T and 3.0 T, with p < 0.05 for all imaging scenarios. On average, with the regular resolution, the SNR of the T1ρ-weighted images obtained at 5.0 T was 15.67% higher than that at 3.0 T, and this value increased to 46.6% in the high-resolution scenario.
TABLE 3
| Orientationa | 3.0 T | 5.0 T | p-value | |
|---|---|---|---|---|
| Regular resolution | cor | 83.7 ± 9.6 | 99.7 ± 17.1 | 0.016 |
| sag | 80.1 ± 10.4 | 90.4 ± 12.6 | 0.043 | |
| High resolution | tra | 25.9 ± 3.0 | 38.0 ± 5.5 | <0.001 |
The average SNR of the T1ρ-weighted images obtained at 3.0 T and 5.0 T
Cor, coronal plane, sag, sagittal plane, tra, transverse plane.
The coefficient of variation between the 3.0 T and 5.0 T T1ρ values for five selected ROIs are shown in Table 4. For 3.0 T, the CVs of T1ρ values for Corpus callosum, Corona radiate, Superior frontal gyrus, Putamen, Centrum semiovale were 3.42%, 2.70%, 3.14%, 2.58%, respectively, and the CVs of T1ρ values from 5.0 T were 2.79%, 2.36%, 3.48%, 2.28%. Based on the above results, the variability of T1ρ values at different field strengths for five selected ROIs was determined as small (CV < 15.0%).
TABLE 4
| 3.0 T | 5.0 T | ||
|---|---|---|---|
| Regular resolution | Corpus callosum | 3.42 | 2.79 |
| Corona radiate | 2.70 | 2.36 | |
| Superior frontal gyrus | 3.14 | 3.48 | |
| Putamen | 2.58 | 2.28 | |
| High resolution | Centrum semiovale | 5.54 | 6.32 |
The Coefficient of Variation (%) of T1ρ values for selected ROIs at 3.0 T and 5.0 T
The Bland-Altman plots of the mean interfield differences of T1ρ values (T1ρ difference = T1ρ,3.0 T- T1ρ,5.0 T) for the five tissue compartments in the aforementioned ROIs are shown in Figure 5, and the mean interfield differences of T1ρ values are as follows: T1ρ difference of Corpus callosum: mean ±1.96SD: −3.4–2.5 ms, bias: −0.4361 ms, 95% CI: −1.0214, 0.1493; T1ρ difference of Corona radiate: mean ±1.96SD: −2.6–2.7 ms, bias: 0.0616 ms, 95% CI: −0.5006, 0.6238; T1ρ difference of Superior frontal gyrus: mean ±1.96SD: −3.3–2.3 ms, bias: −0.4864 ms, 95% CI: −1.0732, 0.1004; T1ρ difference of Putamen: mean ±1.96SD: −1.4–2.4 ms, bias: 0.4784 ms, 95% CI: −0.0818, 0.8750; T1ρ difference of Centrum semiovale: mean ± SD: −4.7–6.0 ms, bias:0.6504 ms, 95% CI: −0.4798, 1.7806.
FIGURE 5

Bland–Altman plots of T1ρ difference values at 5.0 T relative to the T1ρ values at 3.0 T in the Corpus callosum, Corona radiate, Superior frontal gyrus, Putamen, and Centrum semiovale. The solid and dashed lines in the Bland-Altman plots represent the bias and ±1.96 × standard deviation, respectively.
Discussion
In this study, we demonstrated the feasibility of brain T1ρ quantification at 5.0 T and compared the T1ρ values to 3.0 T in healthy volunteers. T1ρ mapping at higher field strengths benefits from increased SNR and chemical exchange, resulting in its increased sensitivity to exchange effects, such as changes in macromolecule concentrations. In previous studies, T1ρ mapping at 7.0 T has shown great potential in studying the loss of proteoglycan and improving differentiation between knees with cartilage lesions and controls, and it is also promising for glucose metabolism studies in detecting intracerebral regions of increased glucose concentration [18,33]. However, SAR limits should be carefully taken into account since the spin lock pulse requires a large amount of RF energy. Relative to 7.0 T, T1ρ mapping at 5.0 T has met less resistance due to lower SAR of spin lock pulse. In addition, the use of the adiabatically prepared spin-lock preparation pulse can compensate for the B0 and B1 inhomogeneities at 5.0 T, which are also lower than those at 7.0 T. Thus, we can obtain reliable T1ρ maps of the brain. To our knowledge, this is the first depiction of brain T1ρ mapping at 5.0 T.
High-resolution anatomical images are desirable for better clinical diagnostic details and clearer visualization of morphological abnormalities. Additionally, it hold great potential for investigation and characterization of various pathological processes. For example, demyelination and axonal loss across various brain regions are prevalent at very early stages of multiple sclerosis [7]. These high-resolution images can improve gray matter and white matter differentiation, in an effort to classify the location of lesions in relation to the cortical/subcortical boundary [34]. While in the case of lower resolution, effects of partial of skull with this free fluid would be increased and could confound interpretation of results.
Our experimental results showed that the image SNRs of T1ρ-weighted images were significantly improved at 5.0 T, and the improvement became more obvious in the high-resolution scenario. This result was in good concordance with previous studies [34–36]. Since SNR is linearly proportional to the magnetic field strength, ultra-high field MRI affords higher spatial resolution, allowing for the visualization of small anatomic structures not previously appreciated at lower magnetic field strengths. As shown in Table 2, the T1ρ values of the brain obtained at 5.0 T are not significantly different from those obtained at 3.0 T. This indicates that the T1ρ values obtained at 5.0 T can be utilized for brain investigation, which is valuable for longitudinal investigation and comparison. Although increasing the number of averages will increase the SNR, it also enables exponential increase in the scan time. However, it may be also hard to use a different number of averages to make the scanning time of 3.0 T equal to 5.0 T. We calculated the SNR per unit of time for images at 5.0 T and 3.0 T respectively by dividing the SNR by the square root of the scanning time. Since the TE of imaging at 5.0 T is different from that at 3.0 T in the regular resolution scenario, due to different duration, gradient ramp and gradient amplitude of the readout gradient, we calculated the SNR per unit of time for images in the high resolution scenario for fair comparison. According to the SNR results in Table 3, the SNR per unit of time for images collected at 5.0 T is also higher than that at 3.0 T, with the value of 8.01 (s−1) at 5.0 T, and 7.05 (s−1) at 3.0 T, respectively.
In this study, since the imaging sequence contains the long-duration T1ρ-preparation pulse, the energy deposited into tissues needs to be addressed, especially for the ultra-high field MR systems. According to previous studies [37–39], the SAR for a single pulse used in the sequence can be estimated as:where is the duration (in ms), is the flip angle, f is a pulse shape factor determined by the type of pulse used, which equals 1 for a hard pulse, or equals 0.67 for a Gaussian pulse, or equals 2 for a sinc pulse. denotes the SAR for a 90°hard pulse with a duration of 3 ms. at 3.0 T for the head model was estimated between 0.242 W/kg (1.5 T) and 2.16 W/kg (4.1 T), and at 5.0T for the head model was slightly larger than 3.3 W/kg (4.7 T). The average SAR delivered for the pulse sequence is the sum of energy absorbed by each RF pulse divided by the total time to acquire the image:where is calculated using Eq. 2, SAR denotes the average delivered SAR over the total time period , N denotes the number of RF pulses. The Food and Drug Administration (FDA) mandated maximum SAR level that equals 3W/kg of the head during the extremities averaged over 10 min. In this study, the average SAR delivered to the brain with TSL = 65 ms approximately 2.68 W/kg at 5.0T and between 0.22 W/kg and 1.98 W/kg at 3.0 T.
In this study, TSL times were chosen empirically based on hardware limitations. The allowed maximum RF duration depends on the RF hardware of the system. In previous studies [6,39–41], the T1ρ-weighted images were produced by using an RF pulse cluster (e.g. 90°+x–SL+y –180°+y–SL-y –90°+x), it is easy to set the TSL to 100 ms due to using the composite SL pulses. In the UIH MR scanner, the deadtime between two pulses is 200us, which is much larger than other commercial MR scanners. The dephasing during the deadtime will affect the image quality of T1ρ-weighted images and may result in inaccurate T1ρ mapping. To resolve this problem and to compensate for the B0 and B1 inhomogeneity, we used a single pulse to improve the robustness of T1ρ imaging. The maximum duration allowed of a single pulse is around 80 ms, both on the 3.0 T and 5.0T UIH MR scanners. Got rid of the time for AHP and RAHP, the maximum TSL used was 65 ms in our study. In our future work, optimized T1ρ preparation pulses with longer TSL and insensitive to B0 and B1 field inhomogeneity will be studied.
Although several studies of knee cartilage showed that 7.0 T T1ρ values were slightly lower than 3.0 T T1ρ values [33,42], the dependence between the T1ρ value and the B0 field strength is not very clear for brain tissues. As is shown in Table 5, we listed the T1ρ values of white matter and gray matter at different field strengths reported in previous studies. The T1ρ values of white matter and gray matter at 3.0 T in our study are consistent with previous studies (T1ρ of the white matter: 76.07 ms, T1ρ of gray matter: 80.1 ms). And T1ρ measurements at 5.0 T showed no significant difference with that at 3.0 T, with the current imaging protocols. Theoretically, the chemical exchange and diffusion contribution to the T1ρ increases with the magnetic field, the field-dependent trend of the T1ρ relaxation times is less pronounced than that of T1 in this study, which is similar to the T2 relaxation times [43].
TABLE 5
| Magnetic field strength (T) | Authors | T1ρ values (ms) | |
|---|---|---|---|
| White matter | Gray matter | ||
| 1.5 | [12] | 80.5 ± 1.4 | 87.5 ± 1.2 |
| [10] | 82.8 ± 1.3 | 86.4 ± 4.4 | |
| [6] | 76.7 ± 1.6 | 78.2 ± 1.3 | |
| 3 | [7] | 80.4 ± 3.3 | 88.9 ± 3.4 |
| Our study | 77.52 ± 1.1 | 80.78 ± 1.2 | |
| 4 | [44] | 85.6 ± 2.4 | 77.7 ± 1.4 |
| 4.7 | [45] | 85–93 | — |
| [4] a | — | 63 ± 2 | |
| 5 | Our study | 76.07 ± 1.8 | 80.1 ± 1.7 |
| 7 | [19] | 50–100 | |
| 9.4 | [17] | 46.47 ± 1.56 | 53.42 ± 1.32 |
T1ρ values for white matter and gray matter at different magnetic field strengths.
Due to the limited literatures on T1ρ mapping in the human brain, T1ρ values of tissues for rats were added as a reference.
There were several limitations in this study. First, the scan time of T1ρ quantification is relatively long due to the requirement of acquiring multiple images with different TSLs. There might be a subject’s motion during acquisition. Fast MRI techniques such as compressed sensing and deep learning can shorten the scan time to a clinically acceptable range but have not yet been integrated into the 5.0 T scanner. Second, the slice thickness is up to 3 mm in this study, which may cause some small lesions to be missed, or inaccurate characterization of lesions due to partial volume effects in clinical applications. Future technical improvement will focus on shortening the scan time for higher-resolution imaging by using the abovementioned fast MRI techniques. Third, all of the subjects were healthy volunteers, and no patient was included in this study. Since the T1ρ values of patients may be different from that of healthy volunteers, efforts should be made to further evaluate the T1ρ quantification of brain lesions at 5.0 T for patients with neurologic diseases and investigate the ability of T1ρ mapping to determine differences between normal and patients.
Conclusion
This study confirmed the feasibility of brain T1ρ quantification at 5.0 T. There was no significant difference between the brain T1ρ values obtained at 3.0 T and 5.0 T. The SNR of T1ρ-weighted images was significantly improved at 5.0 T relative to 3.0 T, which is a benefit for high-resolution imaging and dispersion-related studies. The T1ρ quantification at 5.0 T may be reliable for clinical investigations.
Statements
Data availability statement
The original contributions presented in the study are included in the article/supplementary material, further inquiries can be directed to the corresponding authors.
Ethics statement
The studies involving human participants were reviewed and approved by the Ethics Committee at the Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences (Ethics Committee approval number: YSB-2022-Y02088). The patients/participants provided their written informed consent to participate in this study.
Author contributions
YL: conceptualization, methodology, software, writing—original draft. WW: Formal analysis, data curation, visualization, writing—original draft. YZ: investigation, formal analysis, validation. HW: supervision. HZ: supervision. DL: supervision, writing—review and editing, funding acquisition. YZ: writing—review and editing, supervision, project administration, funding acquisition.
Funding
This study is supported by the National Key R&D Program of China no. 2020YFA0712200, National Natural Science Foundation of China under grant no. 62201561, 81971611, 62125111, 81901736, and U1805261, the Guangdong Basic and Applied Basic Research Foundation under grant no. 2021A1515110540, the Key Technology and Equipment R&D Program of Major Science and Technology Infrastructure of Shenzhen under grant no. 202100102, 202100104, Shenzhen Science and Technology Program under grant no. RCYX20210609104444089.
Conflict of interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Publisher’s note
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.
References
1.
WangYXZhangQLiXChenWAhujaAYuanJ. T1ρ magnetic resonance: Basic physics principles and applications in knee and intervertebral disc imaging. Quant Imaging Med Surg (2015) 5:858–85. 10.3978/j.issn.2223-4292.2015.12.06
2.
JinTKimSG. Characterization of non-hemodynamic functional signal measured by spin-lock fMRI. Neuroimage (2013) 78:385–95. 10.1016/j.neuroimage.2013.04.045
3.
KettunenMIGrohnOHSilvennoinenMJPenttonenMKauppinenRA. Effects of intracellular pH, blood, and tissue oxygen tension on T1ρ relaxation in rat brain. Magn Reson Med (2002) 48:470–7. 10.1002/mrm.10233
4.
KettunenMISierraANarvainenMJValonenPKYla-HerttualaSKauppinenRAet alLow spin-lock field T1 relaxation in the rotating frame as a sensitive MR imaging marker for gene therapy treatment response in rat Glioma1. Radiology (2007) 243:796–803. 10.1148/radiol.2433052077
5.
MagnottaVAHeoHYDlouhyBJDahdalehNSFollmerRLThedensDRet alDetecting activity-evoked pH changes in human brain. Proc Natl Acad Sci U S A (2012) 109:8270–3. 10.1073/pnas.1205902109
6.
GonyeaJVWattsRApplebeeAAndrewsTHipkoSNickersonJPet alIn vivo quantitative whole-brain T1ρ MRI of multiple sclerosis. J Magn Reson Imaging (2015) 42:1623–30. 10.1002/jmri.24954
7.
MaSWangNFanZKaiseyMSicotteNLChristodoulouAGet alThree-dimensional whole-brain simultaneous T1, T2, and T1ρ quantification using MR Multitasking: Method and initial clinical experience in tissue characterization of multiple sclerosis. Magn Reson Med (2021) 85:1938–52. 10.1002/mrm.28553
8.
MangiaSCarpenterAFTyanAEEberlyLEGarwoodMMichaeliS. Magnetization transfer and adiabatic T1ρ MRI reveal abnormalities in normal-appearing white matter of subjects with multiple sclerosis. Mult Scler (2014) 20:1066–73. 10.1177/1352458513515084
9.
Boles PontoLLMagnottaVAMendaYMoserDJOlesonJJHarlynnELet alComparison of T1Rho MRI, glucose metabolism, and amyloid burden across the cognitive spectrum: A pilot study. J Neuropsychiatry Clin Neurosci (2020) 32:352–61. 10.1176/appi.neuropsych.19100221
10.
BorthakurASochorMDavatzikosCTrojanowskiJQClarkCM. T1ρ MRI of Alzheimer's disease. Neuroimage (2008) 41:1199–205. 10.1016/j.neuroimage.2008.03.030
11.
HarisMYadavSKRizwanASinghACaiKKauraDet alT1rho MRI and CSF biomarkers in diagnosis of Alzheimer's disease. NeuroImage: Clin (2015) 7:598–604. 10.1016/j.nicl.2015.02.016
12.
HarisMSinghACaiKJDavatzikosCTrojanowskiJQMelhemERet alT1rho (T1ρ) MR imaging in Alzheimer’ disease and Parkinson’s disease with and without dementia. J Neurol (2011) 258:380–5. 10.1007/s00415-010-5762-6
13.
MangiaSSvatkovaAMascaliDNissiMJBurtonPCBednarikPet alMulti-modal brain MRI in subjects with PD and iRBD. Front Neurosci (2017) 11:709. 10.3389/fnins.2017.00709
14.
NestrasilIMichaeliSLiimatainenTRydeenCEKotzCMNixonJPet alT1ρ and T2ρMRI in the evaluation of Parkinson's disease. J Neurol (2010) 257:964–8. 10.1007/s00415-009-5446-2
15.
JokivarsiKTHiltunenYGrohnHTuunanenPGrohnOHJKauppinenRA. Estimation of the onset time of cerebral ischemia using T1ρand T2 MRI in rats. Stroke (2010) 41:2335–40. 10.1161/strokeaha.110.587394
16.
TanYFXuJChenRYChenBXuJRenDKet alUse of T1 relaxation time in rotating frame (T1ρ) and apparent diffusion coefficient to estimate cerebral stroke evolution. J Magn Reson Imaging (2018) 48:1247–54. 10.1002/jmri.25971
17.
HerzKGandhiCSchuppertMDeshmaneASchefflerKZaissM. CEST imaging at 9.4 T using adjusted adiabatic spin-lock pulses for on- and off-resonant T1ρ-dominated Z-spectrum acquisition. Magn Reson Med (2019) 81:275–90. 10.1002/mrm.27380
18.
PaechDSchuenkePKoehlerCWindschuhJMundiyanapurathSBickelhauptSet alT1ρ-weighted dynamic glucose-enhanced MR imaging in the human brain. Radiology (2017) 285:914–22. 10.1148/radiol.2017162351
19.
SchuenkePKoehlerCKorzowskiAWindschuhJBachertPLaddMEet alAdiabatically prepared spin-lock approach for T1ρ-based dynamic glucose enhanced MRI at ultrahigh fields. Magn Reson Med (2017) 78:215–25. 10.1002/mrm.26370
20.
SpearJTGoreJC. New insights into rotating frame relaxation at high field. NMR Biomed (2016) 29:1258–73. 10.1002/nbm.3490
21.
GilaniIASepponenR. Quantitative rotating frame relaxometry methods in MRI. NMR Biomed (2016) 29:841–61. 10.1002/nbm.3518
22.
JinTIordanovaBHitchensTKModoMWangPMehrensHet alChemical exchange-sensitive spin-lock (CESL) MRI of glucose and analogs in brain tumors. Magn Reson Med (2018) 80:488–95. 10.1002/mrm.27183
23.
ChenWTChanQWangYXJ. Breath-hold black blood quantitative T1rho imaging of liver using single shot fast spin echo acquisition. Quant Imaging Med Surg (2016) 6:168–77. 10.21037/qims.2016.04.05
24.
ChenW. Artifacts correction for T1rho imaging with constant amplitude spin-lock. J Magn Reson (2017) 274:13–23. 10.1016/j.jmr.2016.11.002
25.
GarwoodMDelaBarreL. The return of the frequency sweep: Designing adiabatic pulses for contemporary NMR. J Magn Reson (2001) 153:155–77. 10.1006/jmre.2001.2340
26.
JohnsonCPThedensDRKrugerSJMagnottaVA. Three-Dimensional GRE T1ρ mapping of the brain using tailored variable flip-angle scheduling. Magn Reson Med (2020) 84:1235–49. 10.1002/mrm.28198
27.
LiXHanETBusseRFMajumdarS. In vivo T1ρ mapping in cartilage using 3D magnetization-prepared angle-modulated partitioned k-space spoiled gradient echo snapshots (3D MAPSS). Magn Reson Med (2008) 59:298–307. 10.1002/mrm.21414
28.
WeiZChenQHeQZhangXLiuXzhengHLiY. A quadrature birdcage/48-channel receiver coil assembly for human brain imaging at 5T. In: 2022 Joint Annual Meeting ISMRM-ESMRMB & ISMRT 31st Annual Meeting; London (2022).
29.
ZhuYLiuYYingLPengXWangYJYuanJet alSCOPE: Signal compensation for low-rank plus sparse matrix decomposition for fast parameter mapping. Phys Med Biol (2018) 63:185009. 10.1088/1361-6560/aadb09
30.
ZhuYLiuYYingLQiuZLiuQJiaSet alA 4-minute solution for submillimeter whole-brain T1ρ quantification. Magn Reson Med (2021) 85:3299–307. 10.1002/mrm.28656
31.
MoréJJ. The levenberg-marquardt algorithm: Implementation and theory. Numer Anal (1978) 105–16. 10.1007/BFb0067700
32.
FirbankMJCoulthardAHarrisonRMWilliamsED. A comparison of two methods for measuring the signal to noise ratio on MR images. Phys Med Biol (1999) 44:N261–4. 10.1088/0031-9155/44/12/403
33.
WyattCGuhaAVenkatachariALiXJKrugRKelleyDEet alImproved differentiation between knees with cartilage lesions and controls using 7T relaxation time mapping. J Orthopaedic Translation (2015) 3:197–204. 10.1016/j.jot.2015.05.003
34.
SpringerEDymerskaBCardosoPLRobinsonSDWeisstannerCWiestRet alComparison of routine brain imaging at 3 T and 7 T. Invest Radiol (2016) 51:469–82. 10.1097/rli.0000000000000256
35.
MetcalfMXuDOkudaDTCarvajalLSrinivasanRKelleyDAet alHigh-resolution phased-array MRI of the human brain at 7 tesla: Initial experience in multiple sclerosis patients. J Neuroimaging (2010) 20:141–7. 10.1111/j.1552-6569.2008.00338.x
36.
ZhangYYangCLiangLShiZZhuSChenCet alPreliminary experience of 5.0 T higher field abdominal diffusion-weighted MRI: Agreement of apparent diffusion coefficient with 3.0 T imaging. J Magn Reson Imaging (2022). 10.1002/jmri.28097
37.
CollinsCMLiSSmithMB. SAR and B1 field distributions in a heterogeneous human head model within a birdcage coil. Magn Reson Med (1998) 40:847–56. 10.1002/mrm.1910400610
38.
LiXHanETMaCBLinkTMNewittDCMajumdarS. In vivo 3T spiral imaging based multi-slice T1ρ mapping of knee cartilage in osteoarthritis. Magn Reson Med (2005) 54:929–36. 10.1002/mrm.20609
39.
WitscheyWR2ndBorthakurAElliottMAMellonENiyogiSWallmanDJet alArtifacts in T1ρ-weighted imaging: Compensation for B1 and B0 field imperfections. J Magn Reson (2007) 186:75–85. 10.1016/j.jmr.2007.01.015
40.
CharagundlaSRBorthakurALeighJSReddyR. Artifacts in T1ρ-weighted imaging: Correction with a self-compensating spin-locking pulse. J Magn Reson (2003) 162:113–21. 10.1016/s1090-7807(02)00197-0
41.
GramMSeethalerMGenslerDOberbergerJJakobPMNordbeckP. Balanced spin-lock preparation for B1 -insensitive and B0 -insensitive quantification of the rotating frame relaxation time T1ρ. Magn Reson Med (2021) 85:2771–80. 10.1002/mrm.28585
42.
SinghAHarisMCaiKKoganFHariharanHReddyR. High resolution T1ρ mapping of in vivo human knee cartilage at 7T. Plos One (2014) 9:e97486. 10.1371/journal.pone.0097486
43.
Oros-PeusquensAMLaurilaMShahNJ. Magnetic field dependence of the distribution of NMR relaxation times in the living human brain. Magn Reson Mater Phy (2008) 21:131–47. 10.1007/s10334-008-0107-5
44.
GrohnHIMichaeliSGarwoodMKauppinenRAGrohnOH. Quantitative T1ρ and adiabatic Carr-Purcell T2 magnetic resonance imaging of human occipital lobe at 4 T. Magn Reson Med (2005) 54:14–9. 10.1002/mrm.20536
45.
WheatonAJBorthakurACorboMCharagundlaSRReddyR. Method for reduced SAR T1ρ-weighted MRI. Magn Reson Med (2004) 51:1096–102. 10.1002/mrm.20141
Summary
Keywords
5.0 T, brain, ultra-high field strength, quantitative imaging, T1ρ
Citation
Liu Y, Wang W, Zheng Y, Wang H, Zheng H, Liang D and Zhu Y (2022) Magnetic resonance T1ρ quantification of human brain at 5.0 T: A pilot study. Front. Phys. 10:1016932. doi: 10.3389/fphy.2022.1016932
Received
11 August 2022
Accepted
23 September 2022
Published
07 October 2022
Volume
10 - 2022
Edited by
Wolfgang Bogner, Medical University of Vienna, Austria
Reviewed by
Trevor Andrews, Washington University in St. Louis, United States
Tao Jin, University of Pittsburgh, United States
Updates
Copyright
© 2022 Liu, Wang, Zheng, Wang, Zheng, Liang and Zhu.
This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
*Correspondence: Dong Liang, dong.liang@siat.ac.cn; Yanjie Zhu, yj.zhu@siat.ac.cn
‡These authors have contributed equally to this work
†ORCID: Hairong Zheng, orcid.org/org/0000-0002-8558-5102
This article was submitted to Medical Physics and Imaging, a section of the journal Frontiers in Physics
Disclaimer
All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.