ORIGINAL RESEARCH article
Front. Sports Act. Living
Sec. Exercise Physiology
Volume 7 - 2025 | doi: 10.3389/fspor.2025.1582387
This article is part of the Research TopicImpact of Blood Flow Restriction Device Features and Methodological Considerations on Acute- and Longitudinal Responses to Blood Flow Restricted ExerciseView all 15 articles
Tourniquet Cuff Pressure During Blood Flow Restriction Exercise
Provisionally accepted- 1Northumbria University, Newcastle upon Tyne, United Kingdom
- 2Western Clinical Engineering Ltd., Vancouver British Columbia, Canada
- 3School of Biomedical Engineering and Departments of Orthopedics and Electrical and Computer Engineering, University of British Columbia, Vancouver British Columbia, Canada
Select one of your emails
You have multiple emails registered with Frontiers:
Notify me on publication
Please enter your email address:
If you already have an account, please login
You don't have a Frontiers account ? You can register here
Background: The present study examined how well different blood flow restriction (BFR) devices deliver the prescribed tourniquet cuff pressure. Methods: Fifteen participants completed four BFR exercise sessions, each with a different BFR device (Delfi Personalized Tourniquet System [PTS] for BFR, Saga, SmartCuffs, and Suji), comprising four sets of unilateral leg press (30-15-15-15 repetitions) against resistance bands with 30-second rest periods. The tourniquet cuff was secured proximally on the exercising leg, and the target pressure was set to 80% limb occlusion pressure (LOP), as measured by the device, applied continuously throughout the exercise/rest periods. Tourniquet cuff pressure was sampled at 100Hz via a pressure transducer. Results: Despite prescribing tourniquet cuff pressure at 80% LOP, the actual pressure can vary substantially and be inconsistent between individuals depending on the BFR device used. During the exercise periods, the median percentage of time pressure was within ±10% the target pressure was 95% (Delfi PTS for BFR), 25% (Saga), 26% (SmartCuffs), and 34% (Suji). During the rest periods, the median percentage of time pressure was within ±5% the target pressure was 99% (Delfi PTS for BFR) and 0% for the Saga, SmartCuffs, and Suji BFR devices. Tourniquet cuff pressure during BFR exercise behaves in a wave-like manner characterised by cyclical pressure peaks and valleys. The magnitude of pressure peaks and valleys was: Delfi PTS for BFR (89±2% and 72±3% LOP), Saga (79±9 and 58±7% LOP), SmartCuffs (79±9% and 61±7% LOP), and Suji (90±15 and 65±10% LOP). In several cases, participants experienced tourniquet pressures >100% LOP using the Saga, SmartCuffs, and Suji BFR devices, for up to ~30-55% of the exercise set duration. A progressive loss of pressure occurred throughout the BFR application period by an average of 2-4mmHg•min-1 (~1-2% LOP•min-1 ) in the Saga, SmartCuffs, and Suji BFR devices, whilst the Delfi PTS for BFR exhibited trivial pressure drifts. Conclusion: Differences between the actual and prescribed tourniquet cuff pressure can highly depend on the BFR device employed. The selection of the BFR apparatus is thus important to delivering the prescribed tourniquet cuff pressure to allow for standardisation of the relative occlusion pressure between users.
Keywords: Blood flow restriction, surgical-grade, Tourniquet, Pressure, autoregulation, Exercise
Received: 24 Feb 2025; Accepted: 24 Jul 2025.
Copyright: © 2025 Swain, McEwen, Lai and Hughes. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) or licensor are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.
* Correspondence: Luke Hughes, Northumbria University, Newcastle upon Tyne, United Kingdom
Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.