Validation of the Somnolyzer 24×7 automatic scoring system in children with suspected obstructive sleep apnea

Ignacio Boira
Ignacio Boira*Violeta EstebanVioleta EstebanJos Norberto Sancho-ChustJosé Norberto Sancho-ChustEsther PastorEsther PastorPaula Fernndez-MartínezPaula Fernández-MartínezAnastasiya TorbaAnastasiya TorbaEusebi ChinerEusebi Chiner
  • Department of Pneumology, Sleep Unity, San Juan de Alicante University Hospital, Alicante, Spain

Introduction: Manual scoring of polysomnography data is a laborious and complex process. Automatic scoring by current computer algorithms shows high agreement with manual scoring. The primary objective of this study was to measure the overall validity of the Somnolyzer 24×7 automatic polysomnography scoring system in children.

Materials and methods: We conducted a single-center, prospective, observational study in children undergoing diagnostic polysomnography for suspected obstructive sleep apnea (OSA) from December 2023 to December 2024. We included children aged three to 15 years with suspected obstructive sleep apnea (OSA). Each polysomnogram was scored manually by three experts and automatically by the Somnolyzer 24×7 system.

Results: Our analysis included 75 children (60% girls), of whom 9% did not have OSA, 20% had mild OSA, 31% moderate OSA, and 40% severe OSA. There was a high level of agreement between manual and automatic scoring of the respiratory disturbance index (RDI). The mean correlation (Pearson correlation coefficient) of RDI scored by the three experts was 0.93 (95% confidence interval [CI] 0.92–0.95), similar to the correlation between manual and automatic scoring (0.92, 95% CI 0.90–0.94). The correlation between the different manual scorings and between manual and automatic scoring was maintained in the different sleep stages (N1: 0.93 vs. 0.90, N2: 0.76 vs. 0.73, N3: 0.72 vs. 0.76, REM: 0.86 vs. 0.82).

Conclusion: The Somnolyzer 24×7 automatic scoring system shows strong correlation with manual scoring in respiratory events and sleep architecture. Our results suggest this system could be used for polysomnography scoring in children.

1 Introduction

Obstructive sleep apnea (OSA) is characterized by partial or total obstruction of the upper airway, which alters alveolar ventilation and sleep quality (13). In children, OSA has an estimated prevalence of 1–4%. The most common etiology is adenotonsillar hypertrophy (4), which defines the predominant phenotype. The second main phenotype involves concomitant diseases or comorbidities (with or without tonsillar hypertrophy) such as craniofacial abnormalities, neuromuscular diseases, or complex syndromes (Down syndrome, Prader-Willi syndrome, or Beckwith-Wiedemann syndrome) (5). In addition to anatomical characteristics, childhood obesity is a growing risk factor for sleep disorders in the pediatric population (6). Pediatric OSA has been associated with the development of neurocognitive and behavioral disorders, delayed growth, cardiovascular diseases, metabolic disorders, and reduced quality of life due to a proinflammatory state, increased sympathetic activity, altered coagulation, and oxidative stress (7, 8). For these reasons, early diagnosis is crucial. While polysomnography is the diagnostic gold standard (9), home respiratory polygraphy has demonstrated similar validity in detecting sleep-disordered breathing in children (10).

Manual scoring of polysomnographic data is a laborious and complex process with marked interscorer variability, especially in the identification of arousals and sleep stages (11). In response to these limitations, computerized algorithms were developed to analyze electrophysiological signals and classify sleep stages and respiratory events, achieving a high level of agreement with manual scoring by sleep medicine experts. Notably, computerized algorithms have been validated with manual scoring as a reference standard (12). An automated scoring system could be considered a valid alternative if the level of agreement between automated and manual scoring were comparable to the level of interscorer agreement observed with manual scoring. Recently, sleep medicine has benefited from Artificial Intelligence (AI), which offers innovative solutions to the limitations of traditional diagnostic methods. AI has enabled the classification of sleep stages and the detection of sleep-disordered breathing through the analysis of complex physiological data. AI can process datasets, identify patters and make predictions with high accuracy respect to manual scoring. One of the limitations of AI in pediatric sleep is the lack of pediatric dataset for training AI model. This can limit the generalizability of AI models to childen (13, 14).

The following machine learning algorithms have been employed in pediatric sleep apnea: Convolutional Neural Networks (CNNs), Support Vector Machines (SVM), Random Forest (RF), Transformer-based Model long short-term memory (LSTM)-based and Sleep Staging Model (CSleep Net). These algorithms have achieved high levels of accuracy (1315). Moeller and colleagues (16, 17) reported that U-Sleep performed an overall accuracy in sleep stages of 83.9% and a kappa value of 0.77 comparable to human experts. Somaskandhan and colleagues (16, 18) developed a combined convolutional and long short-term memory neural network architecture. Their model achieved an overall accuracy of 84.1% (Cohen’s kappa κ = 0.78), comparable to interrater reliability between manual scorers without evidence of difference between children with sleep-disordered breathing and control groups.

The most widely employed computer-assisted sleep staging system is the Somnolyzer 24×7, which has Food and Drug Administration (FDA) approval and follows the American Academy of Sleep Medicine (AASM) scoring guidelines (19). This supervised system analyzes sleep stages, respiratory events, desaturations, limb movement, and arousals, but cannot evaluate paroxysmal activity or bruxism. The steps of the analysis are: (1) artifact processing (minimization, identification, channel selection), (2) feature extraction (slow wave, sleep spindles, k complexes, delta, theta, alpha, slow and fast beta background activities, dominant alpha frequency, arousals, various artifact types, slow and rapid eye movements, eye blinks, tonic and transient muscle activity), (3) AI classifier (bi-directional long-short-term memory recurrent neural network [RNN]), and (4) rule-based sub-classification of non-rapid eye movement (NREM) sleep (configuration option). Based on RNN probability, the system uses traffic light color coding to indicate the confidence in its scoring (green: high confidence, yellow: medium confidence, red: low confidence). This automatic scoring system has been validated in several adult studies (2028), achieving very high agreement with manual analysis (Table 1). However, no studies have validated the Somnolyzer system in children.

Table 1
www.frontiersin.org

Table 1. Summary of the results of the most relevant articles comparing manual scoring with Somnolyzer system scoring.

The main objective of this study was to determine the overall validity of the Somnolyzer 24×7 automatic system for reading pediatric polysomnography data. Secondary objectives included evaluating the validity of the Somnolyzer system for classifying sleep stages and identifying respiratory events.

2 Materials and methods

2.1 Study design

We conducted a single-center, prospective, observational study in children undergoing diagnostic polysomnography for suspected OSA from December 2023 to December 2024.

2.2 Study population

Inclusion criteria were age 3–15 years, polysomnography due to suspected OSA, and high-quality recording (at least 6.5 h, with at least 3 h of sleep). Children with a primary diagnosis other than OSA (insomnia, circadian rhythm disturbances, parasomnias, narcolepsy, idiopathic or recurrent hypersomnia, restless legs syndrome, and periodic leg movement disorder) were excluded from the study. Split-night recordings were also excluded.

2.3 Variables

We collected demographic variables, clinical variables, and all sleep study parameters (respiratory disturbance index [RDI]; hypopnea index [HI]; obstructive apnea index [OAI]; central apnea index [CAI]; oxygen desaturation index [ODI]; arousal index; percentage of total sleep time with oxygen saturation below 90% [T90]; total sleep time; sleep efficiency; and time in N1, N2, N3, and REM stages). A respiratory event is scored as an apnea if there is a drop in peak signal excursion by ≥90% of the pre-event baseline and the duration of this drop lasts at least the minimum duration specified for obstructive, mixed, or central apnea. A central apnea is further defined as an event that is associated with absent inspiratory effort throughout the event, and at least one of the following: the event lasts 20 s or longer or the event lasts at least the duration of two breaths during baseline breathing and is associated with an arousal or ≥3% oxygen desaturation. An obstructive apnea is defined as a complete cessation of airflow due to upper airway obstruction, accompanied by continued respiratory effort, lasting at least 10 s or the duration of two baseline breaths, and also associated with an arousal or a decrease in oxygen saturation of 3% or more. A respiratory event is scored as a hypopnea if the peak signal excursions drop by ≥30% of the pre-event baseline, the duration of this drop lasts for at least two breaths and there is ≥3% desaturation from pre-event baseline or the event is associated with an arousal (19). We classified OSA severity according to AHI (mild OSA: AHI of 1–5/h−1, moderate OSA: AHI of 5–10 h−1, severe OSA: AHI above 10/h−1).

All polysomnograms were performed using the SomnoStar Alfa (SensorMedics, CA, USA). For automatic scoring, we used the Somnolyzer system (Philips-Respironics, Murrysville, PA, USA, version 4.0). Three pulmonologists specialized in sleep medicine performed independent manual readings following the AASM criteria for scoring sleep stages and respiratory events (19). All the scorers were blinded to each other’s assessment and to the AI output. No inter-rater variability controls were performed.

2.4 Sample size calculation

The sample size was adjusted to 75 children, given a confidence level (1-α) of 95%, a precision (d) of 3% and a power of 80%.

2.5 Statistical analysis

We analyzed the concordance between manual scoring and automatic scoring. After applying the Kolmogorov–Smirnov test for normality of distribution, we used parametric (Student’s t) or non-parametric (McWhitney) tests to compare continuous variables. We used the Chi-square test or Fisher’s test for categorical variables. To evaluate differences between manual and automatic scoring, we applied analysis of variance (ANOVA). To evaluate the level of correlation between manual and automatic scoring, we used either the Pearson or Spearman correlation coefficient, depending on the distribution of the variable. We used intraclass correlation coefficients (ICCs) and the Bland–Altman method to evaluate agreement (29). For all comparisons, p values below 0.05 indicated statistical significance. All statistical analyses were performed using IBM SPSS Statistics v25 (Arming, NY).

2.6 Ethical considerations

The study was conducted in accordance with the Declaration of Helsinki and approved by the Clinical Research Ethics Committee of San Juan de Alicante University Hospital (ref 23/016). Participants’ parents or legal guardians signed an informed consent form.

3 Results

Our study included 75 children (60% girls) with a mean age of 8 (standard deviation [SD] 4) years and a mean body mass index (BMI) of 17.5 (SD 3.8) kg/m2. Seven children (9%) were not diagnosed with OSA, 15 (20%) had mild OSA, 23 (31%) had moderate OSA, and 30 (40%) had severe OSA. Normally distributed variables were BMI; RDI; number of hypopneas; arousal index; percentage of sleep in N1, N2, and N3; duration of N1, N2, and N3; and total sleep time. Table 2 shows the polysomnographic parameters of manual and automatic scoring and the comparison of mean scores (ANOVA) between the scoring methods. Regarding to respiratory events, there were no significant differences in the means of OAI, CAI and HI between manual and automatic corrections and these differences were not relevant in the RDI (p = 0.98). Concerning to sleep stages, the N2 phase evidenced the greatest difference in mean scores compared to the other sleep stages, although this difference did not reach statistical significance. Moreover, there were also no significant differences in the comparison of means for ODI, arousals index, T90, total sleep time, and there was a trend toward statistical significance in sleep efficiency.

Table 2
www.frontiersin.org

Table 2. Analysis of variance of manual and automatic scoring.

3.1 Correlation between manual and automatic RDI scoring

We found a high correlation in RDI among the three manual scorings and between manual and automatic scoring (Figure 1a). The mean correlation (Pearson correlation coefficient) between the three manually scored RDIs was 0.93 (95% confidence interval [CI] 0.92–0.95), while the mean correlation between manual and automatic RDI scoring was 0.92 (95% CI 0.90–0.94). Therefore, interscorer correlation in the manual results was similar to the correlation between manual and automatic results.

Figure 1
www.frontiersin.org

Figure 1. Correlation matrix (left panel) and Bland–Altman (right panel) plots for the respiratory disturbance index (RDI). Pearson correlation coefficients and the average bias (95% confidence interval) are in the left and right panels, respectively.

When comparing the mean difference in manual RDI scorings versus the mean difference between manual and automatic scorings using the Bland–Altman matrix, we found no significant difference (Figure 1b). The mean difference in manually scored RDI values among the three experts was 0.04 events/h (95% CI −0.27 to 0.30), while the mean difference in manually versus automatically scored RDIs was 0.05 (95% CI −0.39 to 0.56). This mean difference in RDI had diagnostic relevance in only three patients between the three manually scored and in two patients between manual and automatic scoring and for severity classification in three patients for the mild–moderate category and none for the moderate–severe category.

We also calculated ICCs for all combinations of RDI between manual and automated scoring (Supplementary Table S1). There was a high level of agreement between RDI values recorded by the three experts (ICC: 0.93–0.98), and between manual scoring and automatic scoring (ICC: 0.91–0.99). Agreement was maintained when stratified by OSA severity and according to age (children [from 2 year to 6 years], middle childhood [from 6 to 12 years] and teens [from 12 to 16 years]).

In addition, there was substantial agreement between the three experts for OAI and CAI (ICC: 0.60–0.86 and 0.75–0.98, respectively), as well as between manual and automatic scoring (ICC: 0.67–0.86 and 0.72–0.89, respectively). However, there was lower correlation for the HI (ICC: 0.27–0.75 and 0.36–0.68, respectively). Agreement was maintained when stratified by OSA severity and according to age.

Also, there was also adequate agreement between the three experts for ODI (ICC: 0.94–1.00), arousal index (ICC: 0.68–0.94) and T90 (ICC: 0.89–0.99), as well as between manual and automatic scoring (ICC: 0.92–0.99, ICC: 0.69–0.92 and ICC: 0.90–1.00, respectively).

3.2 Correlation between manual and automatic scoring of sleep architecture

ANOVA showed no significant differences between the mean percentage and duration of the different sleep phases obtained through manual and automatic scoring (Table 2).

Figures 25 show the results of the bivariate and Bland–Altman analysis of the different sleep architecture variables, comparing the different manual scorings and the manual and automated scorings. For N1(%), the mean correlation between manual scorings was 0.93, while the mean correlation between manual and automatic scoring was 0.90. The Bland–Altman analysis (Figure 2b) showed that the mean differences between manual scorings and between manual and automatic scoring were not significantly different (−0.08% vs. 0.35%, p > 0.05).

Figure 2
www.frontiersin.org

Figure 2. Correlation matrix (left panel) and Bland–Altman (right panel) plots for the sleep stage N1 (%). Pearson correlation coefficients and the average bias (95% confidence interval) are in the left and right panels, respectively.

Figure 3
www.frontiersin.org

Figure 3. Correlation matrix (left panel) and Bland–Altman (right panel) plots for the sleep stage N2 (%). Pearson correlation coefficients and the average bias (95% confidence interval) are in the left and right panels, respectively.

Figure 4
www.frontiersin.org

Figure 4. Correlation matrix (left panel) and Bland–Altman (right panel) plots for the sleep stage N3 (%). Pearson correlation coefficients and the average bias (95% confidence interval) are in the left and right panels, respectively.

Figure 5
www.frontiersin.org

Figure 5. Correlation matrix (left panel) and Bland–Altman (right panel) plots for the sleep stage rapid eye movement (REM). Pearson correlation coefficients and the average bias (95% confidence interval) are in the left and right panels, respectively.

For N2(%), the mean correlation of manually scored results was 0.76. When we compared the mean correlation between the different manual scorings versus the mean correlation between manual and automatic scoring, we found no significant difference (0.76 vs. 0.73, p > 0.05). The Bland–Altman analysis of stage N2 (Figure 3b) showed that the mean difference between automated and manual scoring was significantly lower than the mean difference between manual scorings (0.45% vs. 4.2%, p = 0.01).

Similarly, for N3(%), there was good correlation between the different manual scorings and between manual and automatic scoring (Figure 4a). When we compared the mean differences between the different manual scorings versus between manual and automated scoring using Bland–Altman plots for N3 (Figure 4b), we found no significant difference (−0.90% vs. 0.88%, p > 0.05).

For the REM stage, there was adequate correlation between manual and automatic scoring without significant differences between the different manual scorings or between manual and automatic scoring (Spearman correlation coefficient 0.86 vs. 0.82, p > 0.05), as shown in Figure 5a. The Bland–Altman analysis (Figure 5b), which compared the mean differences between the three manual scorings versus between manual and automated scoring, showed no significant difference (−0.71% vs. 0.75%, p > 0.05). The correlation observed in sleep architecture was maintained in total sleep time (Pearson correlation coefficient between different manual scorings versus between manual and automatic scoring: 0.84 vs. 0.78, p > 0.05). The correlation was lower for sleep efficiency, without a significant difference (Spearman coefficient: 0.51 vs. 0.44, p > 0.05).

We also calculated ICCs between manual and automated scoring for all combinations of sleep stages (Supplementary Table S2). There was adequate agreement between the three experts for all sleep phases (ICC of N1[%]: 0.91–0.98; N2[%]: 0.65–0.94; N3[%]: 0.72–0.91 and ICC of REM[%]:0.85–0.97), as well as between manual and automatic scoring (ICC of N1[%]: 0.88–0.98; N2[%]: 0.78–0.92; N3[%]: 0.66–0.91 and ICC of REM[%]:0.81–0.97). The agreement was maintained when the data were stratified by OSA severity. However, when the data were stratified by age, there was a lower correlation between the manual and automatic corrections for (N1%) and N2(%) in middle childhood (from 6 to 12 years), and particularly in children (from 2 year to 6 years).

4 Discussion

Our study is the first validation study of Somnolyzer 24×7 computer-assisted sleep scoring system in pediatric population. Our results evidenced that automatic polysomnography scoring system based on AI showed strong correlation with manual scoring in respiratory events and sleep architecture.

Automatic scoring systems have been used in sleep studies for more than 20 years. The first algorithms were able to detect desaturations, snoring, heart rate, and position (30, 31), whereas current algorithms can also detect sleep stages, respiratory events, and arousals using one or more channels, evaluating the spectral power threshold of the frequency bands and detecting waveforms through pattern recognition (32, 33). Although several automatic sleep study correction systems have been validated with high correlations to manual corrections, only Somnolyzer (Philips Respironics), Ensosleep (Ensodata) and Domino (Somnomedics AG) have been certified by the AASM. Regarding Somnolyzer, the latest version includes an AI classifier with bi-directional long-short-term RNN which allows a more accurate correction. Several studies in adults evidenced a high correlation in both sleep architecture and the detection of respiratory events (2028). Despite the potential benefits of automatic scoring systems, validation of Somnolyzer 24×7 in the pediatric population has not yet been realized.

Polysomnography scoring is laborious and technically complex, with differences in the definitions of respiratory events in children versus adults. Despite the potential benefits of automatic scoring systems, Somnolyzer 24×7 has not yet been validated in the pediatric population. Automated scoring systems with proven validity and reproducibility could increase access to pediatric polysomnography in sleep units. Our validation study of the Somnolyzer 24×7 automatic scoring system showed a high level of agreement between manual and automatic scoring in sleep architecture and respiratory events. Manual scoring served as our reference standard, though it is limited by the inherent interscorer variability. To mitigate this limitation, we compared the agreement between the different manual scorings versus the agreement between manual and automatic scoring.

Regard the scoring of respiratory events, we found no significant differences when comparing the means using ANOVA for manual versus automatic correction for respiratory events and RDI. Furthermore, the analysis of the degree of agreement using the ICC revealed a high agreement in RDI, a substantial agreement in OAI and CAI, and a moderate agreement in HI. HI maintains a strong positive linear correlation between manual and automatic correction. This lower agreement in HI may be due to a detection of obstructive apneas rather than hypopneas and has no relevance to RDI. In addition, there was a strong positive linear relationship between manual and automatic scoring (Figure 1a) in RDI with a low mean difference detection which has no diagnostic relevance (Figure 1b). These results demonstrate that Somnolyzer 24×7 shows a high level of correlation with expert correction in the detection of respiratory events. Furthermore, this high level of correlation is maintained when the data is stratified by age group and OSA severity.

Concerning to sleep architecture, we found no significant differences when comparing the means using ANOVA for manual versus automatic correction for sleep stages. However, there is a tendency toward statistical significance in N2. In the analysis of the degree of agreement using the ICC there was an almost perfect agreement in all sleep stages but when the data were stratified by age, there was a lower correlation between the manual and automatic corrections for (N1%) and N2(%) in middle childhood, and particularly in children. This may be because the Somnolyzer 24×7 has only been validated in adults, there are fewer studies of pediatric patients, and sleep wave differences exist between adults and pediatric patients. During initial years of life electroencephalogram power increases in particular in the slow wave frequency range during NREM sleep. Then, maximal power values in the faster frequency bands, such as theta, alpha and beta activity, are reached at 2–5 years of age. Power in these frequency bands decreases steadily thereafter (34). This difference in the sleep waveform, especially in children, justifies the discrepancy between manual and automatic correction and also the results with a tendency toward statistical significance in the mean comparison analysis on sleep efficiency. Regarding the other results, there was a very strong positive linear relationship between manual and automatic scoring in N1(%) (Figure 2a), a strong positive linear relationship in N2(%), N3(%) and REM(%) (Figures 2a5a) with a low mean difference detection in all sleep stages which has no diagnostic relevance (Figures 2b5b). Therefore, automatic correction shows a very high degree of correlation with manual correction with respect to sleep architecture, despite limitations arising from changes in sleep waveforms in the pediatric population relative to adults.

In a validation study of the Somnolyzer system in adults (20), Punjabi and colleagues found a very strong correlation between manual and automatic scoring of AHI (Pearson correlation coefficient: 0.93), practically the same value that we obtained for RDI (Pearson correlation coefficient: 0.92). For sleep architecture, we found a stronger correlation in most sleep phases compared with Punjabi and colleagues (N1: 0.90 vs. 0.63, N2: 0.73 vs. 0.66, N3: 0.76 vs. 0.65 and REM: 0.82 vs. 0.92). This difference may be attributable to the different versions of Somnolyzer employed. While we used Somnolyzer 4.0, which classifies respiratory events and sleep architecture using artificial intelligence (RNN), Punjabi and colleagues used version 3.0. The main changes of Somnolyzer 4.0 are spindle detection from both hemispheres, improved slow-wave detection using empirical mode decomposition (improve rule-based sub-classification of NREM sleep stages), new channels and the use of LSTM and RNN classifiers with a confidence score based on RNN output probabilities.

The results obtained in our study promote the implementation of polysomnography in sleep units that do not currently perform it on pediatric patients with suspected OSA and may reduce waiting lists in units that do perform polysomnography in pediatric patients. The Somnolyzer 24×7 software is user-friendly and uses traffic light color coding to indicate the confidence of its scoring. This allows sleep reviewers to focus on the yellow and red codes, improving time efficiency. However, the implementation of automated sleep study correction systems requires caution to avoid over-reliance on AI and requires clinician training, so we propose a hybrid scoring workflow, using Somnolyzer 24×7 as a complementary to enhance and standardized sleep staging and improve patient diagnosis.

Despite our results of Somnolyzer 24×7 validation in patients with suspected OSA, further validation is required in pediatric patients with comorbidities, patients at different stages of child development and patients with sleep disorders other than OSA, such as insomnia, circadian rhythm disturbances, parasomnias, narcolepsy, idiopathic or recurrent hypersomnia, restless legs syndrome, and periodic leg movement disorder.

One main limitation of our study is that the three sleep medicine experts worked in the same clinic so we recommend the development of multicenter external validation studies. Another limitation is that we did not perform both manual and automatic joint correction. Therefore, randomized trials in pediatric population are needed to compare correction by AI alone, manual correction alone, and combined correction by AI and manual correction. A further limitation is the heterogeneity of the population, as we included children from all stages of development. However, when we analyzed the data by age group, we found that there were high correlations between respiratory events and sleep stages across the age groups, except for the correlation between manual and automatic corrections for N1% and N2%.

The main strengths of our study are its prospective and robust methodology, which has been replicated in other validation studies, and the involvement of three medical experts in manual scoring, allowing for a more reliable comparison with automated scoring. Another strength is the sample size and the favorable results obtained despite the technical difficulties of performing polysomnography in pediatric patients.

5 Conclusion

Somnolyzer 24×7 automatic polysomnography scoring system showed strong correlation with manual scoring in respiratory events and sleep architecture. By simplifying the diagnostic process, this automatic system may facilitate broader implementation of polysomnography in sleep units.

Data availability statement

The original contributions presented in the study are included in the article/Supplementary material, further inquiries can be directed to the corresponding author.

Ethics statement

The studies involving humans were approved by Clinical Research Ethics Committee of San Juan de Alicante University Hospital (ref 23/016). The studies were conducted in accordance with the local legislation and institutional requirements. Written informed consent for participation in this study was provided by the participants’ legal guardians/next of kin.

Author contributions

IB: Methodology, Data curation, Supervision, Investigation, Conceptualization, Software, Visualization, Resources, Validation, Writing – original draft, Formal analysis, Writing – review & editing, Project administration, Funding acquisition. VE: Validation, Investigation, Conceptualization, Writing – review & editing, Supervision, Methodology, Formal analysis, Funding acquisition, Visualization, Software, Resources, Data curation, Project administration, Writing – original draft. JS-C: Writing – original draft, Resources, Formal analysis, Visualization, Funding acquisition, Project administration, Data curation, Methodology, Writing – review & editing, Validation, Supervision, Conceptualization, Software, Investigation. EP: Investigation, Funding acquisition, Software, Formal analysis, Writing – original draft, Visualization, Conceptualization, Data curation, Resources, Project administration, Supervision, Methodology, Validation, Writing – review & editing. PF-M: Investigation, Supervision, Software, Funding acquisition, Conceptualization, Writing – review & editing, Resources, Writing – original draft, Formal analysis, Project administration, Data curation, Visualization, Methodology, Validation. AT: Data curation, Validation, Conceptualization, Project administration, Methodology, Supervision, Resources, Writing – original draft, Investigation, Software, Formal analysis, Writing – review & editing, Funding acquisition, Visualization. EC: Validation, Resources, Formal analysis, Conceptualization, Methodology, Visualization, Project administration, Data curation, Funding acquisition, Investigation, Supervision, Writing – review & editing, Software, Writing – original draft.

Funding

The author(s) declare that no financial support was received for the research and/or publication of this article.

Conflict of interest

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.

Generative AI statement

The authors declare that no Gen AI was used in the creation of this manuscript.

Publisher’s note

All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article, or claim that may be made by its manufacturer, is not guaranteed or endorsed by the publisher.

Supplementary material

The Supplementary material for this article can be found online at: https://www.frontiersin.org/articles/10.3389/fmed.2025.1617530/full#supplementary-material

References

1. Solano-Pérez, E, Coso, C, Castillo-García, M, Romero-Peralta, S, Lopez-Monzoni, S, Laviña, E, et al. Diagnosis and treatment of sleep apnea in children: a future perspective is needed. Biomedicines. (2023) 11:1708. doi: 10.3390/biomedicines11061708

PubMed Abstract | Crossref Full Text | Google Scholar

2. Kaditis, AG, Alonso Alvarez, ML, Boudewyns, A, Alexopoulos, EI, Ersu, R, Joosten, K, et al. Obstructive sleep disordered breathing in 2-to 18-year-old children: diagnosis and management. Eur Respir J. (2016) 47:69–94. doi: 10.1183/13993003.00385-2015

PubMed Abstract | Crossref Full Text | Google Scholar

3. Mediano, O, González Mangado, N, Montserrat, JM, Alonso-Álvarez, ML, Almendros, I, Alonso-Fernández, A, et al. Documento internacional de consenso sobre apnea obstructiva del sueño. Arch Bronconeumol. (2022) 58:52–68. doi: 10.1016/j.arbres.2021.03.017

PubMed Abstract | Crossref Full Text | Google Scholar

4. Witmans, M, and Tablizo, MA. Current concepts in pediatric obstructive sleep apnea. Children. (2023) 10:480. doi: 10.3390/children10030480

PubMed Abstract | Crossref Full Text | Google Scholar

5. Chiner, E, Sancho-Chust, JN, Pastor, E, Esteban, V, Boira, I, Castelló, C, et al. Features of obstructive sleep apnea in children with and without comorbidities. J Clin Med. (2023) 12:2418. doi: 10.3390/jcm12062418

PubMed Abstract | Crossref Full Text | Google Scholar

6. Lee, JH, and Cho, J. Sleep and obesity. Sleep Med Clin. (2022) 17:111–6. doi: 10.1016/j.jsmc.2021.10.009

PubMed Abstract | Crossref Full Text | Google Scholar

7. Brockmann, PE, and Gozal, D. Neurocognitive consequences in children with sleep disordered breathing: who is at risk? Children. (2022) 9:1278. doi: 10.3390/children9091278

PubMed Abstract | Crossref Full Text | Google Scholar

8. Galland, B, Spruyt, K, Dawes, P, McDowall, PS, Elder, D, and Schaughency, E. Sleep disordered breathing and academic performance: a meta-analysis. Pediatrics. (2015) 136:e934-46. doi: 10.1542/peds.2015-1677

PubMed Abstract | Crossref Full Text | Google Scholar

9. Vaienti, B, Di Blasio, M, Arcidiacono, L, Santagostini, A, Di Blasio, A, and Segù, M. A narrative review on obstructive sleep apnoea syndrome in pediatric population. Front Neurol. (2024) 15:1393272. doi: 10.3389/fneur.2024.1393272

Crossref Full Text | Google Scholar

10. Chiner, E, Cánovas, C, Molina, V, Sancho-Chust, JN, Vañes, S, Pastor, E, et al. Home respiratory Polygraphy is useful in the diagnosis of childhood obstructive sleep apnea syndrome. J Clin Med. (2020) 9:2067. doi: 10.3390/jcm9072067

PubMed Abstract | Crossref Full Text | Google Scholar

11. Whitney, CW, Gottlieb, DJ, Redline, S, Norman, RG, Dodge, RR, Shahar, E, et al. Reliability of scoring respiratory disturbance indices and sleep staging. Sleep. (1998) 21:749–57. doi: 10.1093/sleep/21.7.749

PubMed Abstract | Crossref Full Text | Google Scholar

12. Penzel, T, Hirshkowitz, M, Harsh, J, Chervin, RD, Butkov, N, Kryger, M, et al. Digital analysis and technical specifications. J Clin Sleep Med. (2007) 3:109–20.

Google Scholar

13. Angamuthu, AK, Nallasamy, V, Krishnan, VL, and Sughumaran, CT. Role of artificial intelligence in paediatric sleep medicine. Biosci Biotechnol Res Asia. (2025) 22:103–12. doi: 10.13005/bbra/3344

Crossref Full Text | Google Scholar

14. Vaquerizo-Villar, F, Alvarez, D, Kheirandish-Gozal, L, Gutierrez-Tobal, GC, Barroso-Garcia, V, Santamaria-Vazquez, E, et al. A convolutional neural network architecture to enhance oximetry ability to diagnose pediatric obstructive sleep apnea. IEEE J Biomed Health Inform. (2021) 25:2906–16. doi: 10.1109/JBHI.2020.3048901

PubMed Abstract | Crossref Full Text | Google Scholar

15. Pang, B, Doshi, S, Roy, B, Lai, M, Ehlert, L, Aysola, RS, et al. Machine learning approach for obstructive sleep apnea screening using brain diffusion tensor imaging. J Sleep Res. (2023) 32:e13729. doi: 10.1111/jsr.13729

PubMed Abstract | Crossref Full Text | Google Scholar

16. Bruni, O. Artificial intelligence in pediatric sleep staging: a new era or a complementary tool? Sleep. (2025) 16:zsaf067. doi: 10.1093/sleep/zsaf067

Crossref Full Text | Google Scholar

17. Moeller, AL, Perslev, M, Paulsrud, C, Thorsen, SU, Leonthin, H, Debes, NM, et al. Artificial intelligence or sleep experts: comparing polysomnographic sleep staging in children and adolescents. Sleep. (2025) 28:zsaf053. doi: 10.1093/sleep/zsaf053

Crossref Full Text | Google Scholar

18. Somaskandhan, P, Leppänen, T, Terrill, PI, Sigurdardottir, S, Arnardottir, ES, Ólafsdóttir, KA, et al. Deep learning-based algorithm accurately classifies sleep stages in preadolescent children with sleep-disordered breathing symptoms and age-matched controls. Front Neurol. (2023) 14:1162998. doi: 10.3389/fneur.2023.1162998

PubMed Abstract | Crossref Full Text | Google Scholar

19. Berry, RB, Quan, SF, Abreu, AR, Bibbs, ML, DelRosso, L, Harding, SM, et al. The AASM manual of the scoring of sleep and associated events: rules, terminology and technical specifications. Darien, IL: American Academy of Sleep Medicine (2023).

Google Scholar

20. Punjabi, NM, Shifa, N, Dorffner, G, Patil, S, Pien, G, and Aurora, RN. Computer-assisted automated scoring of Polysomnograms using the Somnolyzer system. Sleep. (2015) 38:1555–66. doi: 10.5665/sleep.5046

PubMed Abstract | Crossref Full Text | Google Scholar

21. Anderer, P, Gruber, G, Parapatics, S, Woertz, M, Miazhynskaia, T, Klosch, G, et al. An E-health solution for automatic sleep classification according to Rechtschaffen and kales: validation study of the Somnolyzer 24×7 utilizing the siesta database. Neuropsychobiology. (2005) 51:115–33. doi: 10.1159/000085205

PubMed Abstract | Crossref Full Text | Google Scholar

22. Barbanoj, MJ, Clos, S, Romero, S, Morte, A, Giménez, S, Lorenzo, JL, et al. Sleep laboratory study on single and repeated dose effects of paroxetine, alprazolam and their combination in healthy young volunteers. Neuropsychobiology. (2005) 51:134–47. doi: 10.1159/000085206

PubMed Abstract | Crossref Full Text | Google Scholar

23. Anderer, P, Moreau, A, Woertz, M, Ross, M, Gruber, G, Parapatics, S, et al. Computer-assisted sleep classification according to the standard of the American Academy of sleep medicine: validation study of the AASM version of the Somnolyzer 24×7. Neuropsychobiology. (2010) 62:250–64. doi: 10.1159/000320864

PubMed Abstract | Crossref Full Text | Google Scholar

24. Griessenberger, H, Heib, DP, Kunz, AB, Hoedlmoser, K, and Schabus, M. Assessment of a wireless headband for automatic sleep scoring. Sleep Breath. (2013) 17:747–52. doi: 10.1007/s11325-012-0757-4

PubMed Abstract | Crossref Full Text | Google Scholar

25. Magnusdottir, S, and Hilmisson, H. Ambulatory screening tool for sleep apnea: analyzing a single-lead electrocardiogram signal (ECG). Sleep Breath. (2018) 22:421–9. doi: 10.1007/s11325-017-1566-6

PubMed Abstract | Crossref Full Text | Google Scholar

26. Bakker, JP, Ross, M, Cerny, A, Vasko, R, Shaw, E, Kuna, S, et al. Scoring sleep with artificial intelligence enables quantification of sleep stage ambiguity: hypnodensity based on multiple expert scorers and auto-scoring. Sleep. (2023) 46:zsac154. doi: 10.1093/sleep/zsac154

PubMed Abstract | Crossref Full Text | Google Scholar

27. Cheng, H, Yang, Y, Shi, J, Li, Z, Feng, Y, and Wang, X. Comparison of automated deep neural network against manual sleep stage scoring in clinical data. Comput Biol Med. (2024) 179:108855. doi: 10.1016/j.compbiomed.2024.108855

PubMed Abstract | Crossref Full Text | Google Scholar

28. Gomes, DS, Seixas, C, and Cravo, J. Performance of the Alice PDx device with the Somnolyzer automated scoring algorithm for the diagnosis of obstructive sleep apnea. Cureus. (2024) 16:e52654. doi: 10.7759/cureus.52654

PubMed Abstract | Crossref Full Text | Google Scholar

29. Bland, JM, and Altman, DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. (1986) 1:307–10. doi: 10.1016/S0140-6736(86)90837-8

PubMed Abstract | Crossref Full Text | Google Scholar

30. Stoohs, R, and Guilleminault, C. MESAM 4: an ambulatory device for the detection of patients at risk for obstructive sleep apnea syndrome (OSAS). Chest. (1992) 101:1221–7. doi: 10.1378/chest.101.5.1221

PubMed Abstract | Crossref Full Text | Google Scholar

31. Rauscher, H, Popp, W, and Zwick, H. Quantification of sleep disordered breathing by computerized analysis of oximetry, heart rate and snoring. Eur Respir J. (1991) 4:655–9. doi: 10.1183/09031936.93.04060655

PubMed Abstract | Crossref Full Text | Google Scholar

32. Nakano, H, Tanigawa, T, Furukawa, T, and Nishima, S. Automatic detection of sleep-disordered breathing from a single-channel airflow record. Eur Respir J. (2007) 29:728–36. doi: 10.1183/09031936.00091206

PubMed Abstract | Crossref Full Text | Google Scholar

33. Magalang, UJ, Chen, NH, Cistulli, PA, Fedson, AC, Gíslason, T, Hillman, D, et al. Agreement in the scoring of respiratory events and sleep among international sleep centers. Sleep. (2013) 36:591–6. doi: 10.5665/sleep.2552

PubMed Abstract | Crossref Full Text | Google Scholar

34. Kurth, S, Ringli, M, Geiger, A, LeBourgeois, M, Jenni, OG, and Huber, R. Mapping of cortical activity in the first two decades of life: a high-density sleep electroencephalogram study. J Neurosci. (2010) 30:13211–9. doi: 10.1523/JNEUROSCI.2532-10.2010

PubMed Abstract | Crossref Full Text | Google Scholar

Keywords: obstructive sleep apnea, sleep disorders, children, Somnolyzer, polysomnography, artificial intelligence

Citation: Boira I, Esteban V, Sancho-Chust JN, Pastor E, Fernández-Martínez P, Torba A and Chiner E (2025) Validation of the Somnolyzer 24×7 automatic scoring system in children with suspected obstructive sleep apnea. Front. Med. 12:1617530. doi: 10.3389/fmed.2025.1617530

Received: 24 April 2025; Accepted: 03 June 2025;
Published: 18 June 2025.

Edited by:

Liang Zhao, Dalian University of Technology, China

Reviewed by:

Yunhuan Liu, Tongji University, China
Rabie Adel El Arab, Almoosa College of Health Sciences, Saudi Arabia

Copyright © 2025 Boira, Esteban, Sancho-Chust, Pastor, Fernández-Martínez, Torba and Chiner. This is an open-access article distributed under the terms of the Creative Commons Attribution License (CC BY). The use, distribution or reproduction in other forums is permitted, provided the original author(s) and the copyright owner(s) are credited and that the original publication in this journal is cited, in accordance with accepted academic practice. No use, distribution or reproduction is permitted which does not comply with these terms.

*Correspondence: Ignacio Boira, bmFjaG9ib2lyYWVucmlxdWVAaG90bWFpbC5lcw==

Disclaimer: All claims expressed in this article are solely those of the authors and do not necessarily represent those of their affiliated organizations, or those of the publisher, the editors and the reviewers. Any product that may be evaluated in this article or claim that may be made by its manufacturer is not guaranteed or endorsed by the publisher.